KERN-SOHN ORM 1RS, ORM 50BM Instructions

Catalogue
Introduction
Display & Button
Preparations before Operation
Booting and Measurement
Calibration
Switch the Scales and Temperature Units
Turn Off
Maintenance & Preservation
Appendix
Before operating your instrument, please
read this manual properly.
1.Introduction
Panel Descriptions
Stainless Steel Sample Groove
LCD Display Screen
Battery Hatch
The Packing Accessories Packaging
x1 The Instructions x1 AAA Batteries x1
Dropper
x1 Square Cover x1 Digital Refractometer x1
Cover
Prism
Removable Shell
2.Display and Buttons
Display Areas and Buttons
Temperature Display Area
Multi-function Display Area
Battery Capacity
Host Display Area
Turn on Button
Converting Different Scales& Temperature Unit
Calibrating“Zero Point”
NotePlease replace the battery when the is displayed.
Refractive Index (RI)
% or ‰ Unit
3.Preparations before Operation
3.1
Install the Battery
Put 1 piece of 1.5V battery into the cabin in the right way and recover the cabin again.
Open the battery cabin by pushing the battery hatch in the direction of the arrow.
Install the Wrist Strap
3.2
Properly Install the wrist strap into the hole at the bottom of the instrument.
4.Booting and Measurement
4.1
Booting
The multi-function display area shows the current scale number.
Press "READ" Button to Turn On
If there is no sample on the drip, it will display “dash”
Note :
1. When used outdoors, please avoid strong light so as not to affect the measurement accuracy.
2. Please keep the instrument in a stable and still statement and position.
4.2
Measurement After turn on, clear the distilled water and dry the sample plate, drip 0.2~0.3ml sample then close the cover to measure.
Showing Beyond Measuring Range
Measuring Result
Press the "READ" one time the meter will measure one time.
If press the "Read" button for 2 seconds, the instrument would make the automatic measurements upon programmed times (default 15times), the final value is the average of 15 times’ measurements. After measurements, the multi-functions display area would return back to scale showing status.
The multi-function area would show remaining times during the automatic measurement.
5.Zero Calibration
The meter only supports pure water calibration. The calibration method is as following drop
0.2-0.3ml pure water then close the cover to measure.
LCD Flashing
Press 'CAL' button for 2 seconds to enter the calibration status Press “CAL” button for 2-3 seconds till see the ‘CAL’ flashing
Display
Press "CAL" button once again during the ‘CAL’ flashing to start to calibrate. When the calibration is completed showing as following. If no any operations for 10seconds the instrument would return back to booting status.
If fail to complete the calibration, multi-function display area would show an error code.
Temperature exceed the limitations, other error codes could be checked in the appendix error code page.
6.Switch the Scales and Temperature Units
6.1
Scales Converting
The instrument could support 10 scales maximally.
Current Scale Number
Press to convert the scales and the values.
Temperature System Converting
6.2
Press button for 2 seconds, temperature unit would be converted.
Current Scale Unit
Temperature Unit
If exceed the temperature limitations, the signs “HHH” or “LLL” would show.
7.Turn Off
If without any operations for 1 minute, the instrument would be automatically turned off.
8.Maintenance and Preservation
1. Please clean and wash the sample plate with distilled water and dry it with soft cleaning cloth or paper towel after finishing the measuring one kind sample.
2. Never left the remains and residuals of samples in the sample plate for long time.
3. After finishing measurements of the corrosive liquid, please clean the sample plate as quick as possible to avoid the irreparable damage of the prism and metal surface of the plate.
4. Please use soft cleaning cloth or paper towel to clean the sample plate to avoid scribing the prism’s glass.
5. When the dropper and dust-free cloth are not used, please clean it with clean water and put it in the packing box after drying.
6. If no using the instrument for a long time, please remove the battery, and preserved in a cool and dry environment.
Appendix:
Performance:
Range
0.0%~50.0% ±0.2%
Brix
0.0%~90.0%
1.3330~1.4200
R.I.
1.3330~1.5177
Temperature
Dimension
Net weight
0.0~40.0℃
32.0~104.0℉
121*58*25mm
90g (excluding battery)
The Error Codes Table:
code Instructions
A01
A02
A03
Beyond the scope of calibration temperature.0.0~40.0)
During calibration, no solution or solution wrong.
This instrument has a hardware failure.
Accuracy
±0.2%
±0.0003
±0.0003
±0.5℃
±0.9℉
Description of Scales Numbering:
Resolution
0.1%
0.1%
0.0001
0.0001
0.1℃
0.1℉
2020/01/01 V1.0