10 Instant help ................................................................................................... 24
MPB / MXS -BA-e-1113
4
1 Technical Data
KERN
MPB300K100 / P
MXS300K100
Readability (d)
100 g
Weighing range (max)
300 kg
300kg
Recommended adjustment weight, (Class)
200 kg
(M2)
200 kg
(M2)
Stabilization time (typical)
2 – 3 sec.
Warm-up time
10 min
Electric Supply
Mains adapter 9V / 100 mA
Battery operation 6 x 1.5V, size AA
Service life 170h
Auto Off
After 3 min without load change (adjustable)
Operating temperature
+ 5°C … + 35°C
Storage temperature
- 20°C … + 60°C
Humidity of air
max. 80 % (not condensing)
Platform (B x D x H )mm
315x300x60
550 x 550x80
Terminal (B x T x H) mm
210 x 110 x 48
Balance ready for operation
(B x D x H) mm
Without stand: 315x300x60
with stand: 315x440x1010
550 x 550x80
Weight kg (net)
Without stand: 4.0
with stand : 5.8
14.0
MPB / MXS -BA-e-1113
5
2 Basic Information (General)
2.1 Proper use
These balance is used to calculate the weight of people in the private sphere.
Utilisation of this balance in the medical field is not permissible.
A person must stand on the weighing plate for the weighing procedure. As soon as a
stable weighing value is reached the weighing value can be read.
When using balance with height measurement, ensure that the top flap is turned
downwards immediately after use in order to avoid risk of injury.
The balance should be checked for correct condition prior to each utilisation by a
person familiar with proper operation of the balance.
2.2 Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly draining
fluids from a container on the balance. )
Do not leave permanent load on the weighing plate. This may damage the measuring system.
Impacts and overloading exceeding the stated maximum load (max) of the balance,
minus a possibly existing tare load, must be strictly avoided. This could cause damage to the balance.
Never operate balance in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
2.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
mechanical damage and damage caused by media, liquids
natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
Dropping the balance
MPB / MXS -BA-e-1113
6
2.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com) with regard to the
monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.
Using measuring technology to check the accuracy of the measuring device is recommended for personal balances with body height measurement but is not absolutely essential as the calculation of the human body height is always subject to a great
deal of inaccuracy.
MPB / MXS -BA-e-1113
7
3 Basic Safety Precautions
3.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you
are already familiar with KERN balances.
3.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
4 Transportation & Storage
4.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
4.2 Packaging
Keep all parts of the original packaging in case you need to return the appliance.
Only use original packaging for returning.
Before sending, disconnect all connected cables and loose/movable parts.
Attach possibly existing transport safeguards. Secure all parts e.g. weighing platform,
power pack, stand, operating panel etc. against slipping and damage.
MPB / MXS -BA-e-1113
8
5 Unpacking, Setup and Commissioning
5.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
- Place the balance on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
- Protect the balance against direct draughts due to open windows and doors;
- Avoid jarring during weighing;
- Protect the balance against high humidity, vapors and dust;
- Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance)
may occur if a cold appliance is taken to a considerably warmer environment.
In this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
- Avoid static charge of the balance and of the person to be weighed.
- Avoid contact with water.
If electro-magnetic fields or static charge occur, or if the power supply is unstable
major deviations on the display (incorrect weighing results) are possible. In that case,
the location must be changed.
MPB / MXS -BA-e-1113
9
5.2 Unpacking
Remove the individual components of the balance or the complete balance from the
packaging with care and install at the intended location. When using the power pack,
ensure that the power cable does not produce a risk of stumbling.
5.2.1 Balance assembly and installation
MPB personal balance with wall fixing device:
Scope of delivery:
MPB personal balance with stand:
MPB / MXS -BA-e-1113
10
MXS adipose balance:
Scope of delivery:
Instructions for attaching an optimum body-height measuring device to the
model MPB with stand
A separately available body-height measuring device can be screwed onto this model. Follow the operating instructions for the body-height measuring device for this
purpose.
Instructions for attaching an external stand to Model MPB without stand and
Model MXS.
MPB / MXS -BA-e-1113
11
Secure wall fixing device with screws at the top on the aluminium section
Secure circular plate with screws on the aluminium section
Remove the two lateral rubber plugs from the display unit.
Secure the display unit with the two rotary knobs to the fixing device.
Position display unit with the rotary knobs.
Secure cable with cable clips
General instructions for installing the aforementioned balance.
Install the personal balance in the intended location exactly horizontally. Align the
adipose balance horizontally by means of the height-adjustable rubber feet.
When installing or transporting balances with large and heavy platforms (weighing
plate folded upwards), ensure that the balance do not fall over or get damaged.
MPB / MXS -BA-e-1113
12
5.2.2 Scope of delivery
Serial accessories:
Mains power supply Operating Manual
5.2.3 Assembly instructions for models with wall fixing device
(Personal balance, adipose balance )
MPB / MXS -BA-e-1113
13
5.3 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage. Only KERN original power packs should be used.
5.4 Battery operation (insertion and removal)
On models where the back of the display unit is not directly accessible, remove the
two black rotary knobs from both sides of the display unit in order to open the battery
compartment and remove the display unit from the holder. Remove battery cover
from under the display unit. Insert 6 x 1.5V AA batteries into the holder. Replace battery cover and screw the display unit back into the holder using the black rotary
knobs.
In order to save the battery, the balance switches automatically off after 3 minutes
without weighing. Further shutdown times can be set in the Menu
(Function "A.OFF"), see chapter 5.7.
If the batteries are run down, "LO" appears in the display. Press [ON/OFF] -key and
replace the batteries immediately.
If the balance is not used for a longer time, take out the batteries and store them
separately. Leaking battery liquid could damage the balance.
5.5 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1). During this warming up time the balance must be connected to the power supply (mains, accumulator
or battery) and be switched on.
The accuracy of the balance depends on the local acceleration of gravity.
5.6 Adjustment (only dealers authorised by KERN)
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out during the initial start-up, after change in
location and variation of surrounding temperature. To receive accurate measuring
values it is also recommended to adjust the balance periodically in weighing operation.
Use the adjustment weight (See Technical Data) to check and adjust the accuracy of
the balance at any time.
MPB / MXS -BA-e-1113
14
Operation
Display
Turn on balance by pressing the [ON/OFF] key
Keep [0] button depressed for approx. 3 sec until
"SETUP" is displayed followed by "9rAd"
Press the [ TARE ] button as often as required until
"CAL ib" is displayed
Press the [HOLD] button.
Press the [TARE] key. ( Zero count mode )
Press the [0] button repeatedly until "CAL 0" is
displayed.
Press the [HOLD] button.
Press the [TARE] key.
Enter the required size of the adjustment weight (See
Section 1, "Technical Data"): Press the [HOLD] button
to select the item to be changed and press the [TARE]
button to select the numerical value.
Confirm by pressing the [0] button.
5.6.1 Procedure when adjusting (only dealers authorised by KERN)
Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization.
MPB / MXS -BA-e-1113
15
Place the adjustment weight carefully in the centre of
the weighing plate and a numerical value will appear in
the display. Press the [HOLD] button. The adjustment
process is started.
The balance will automatically return to Weighing
mode upon successful completion of the adjustment
and will display the weight of the adjustment weight.
Take away adjustment weight.
Turn off balance by pressing the [ON/OFF] key.
MPB / MXS -BA-e-1113
16
5.7 Menu overview (only dealers authorised by KERN)
Navigation in the menu:
With the balance switched on, keep the [0] button depressed for approx.
3 sec until "SETUP" is displayed followed by "9rAd".
Press the [ TARE ] button as often as necessary until the required function is
displayed.
Press the [HOLD] button to confirm the selected function. The first parameter
will be displayed. Press the [HOLD] button to select the required parameter
and confirm by pressing the [ TARE ] button.
To exit the Menu and save, press the [ TARE ] button as often as required until
"" is displayed and then confirm by pressing the [HOLD] button. The balance
returns automatically into weighing mode.
Select using the [HOLD] and the [ TARE]-button
x factory setting
MPB / MXS -BA-e-1113
17
Partition steps, weighing range (max.) and readout (d)
Balance zero display: Should the balance not display
exactly zero despite an empty scale pan, press the
[0] button. Your balance will be set to zero after a
short standby time.
2
[ο]
Stability display:
If the display shows the stability display [ο] the balance
is in a stable status. The [ο] indication disappears if the
condition is unstable.
3
Illuminates in the event of power supply via mains adaptor
4
BMI
Calculated BMI value
5
HOLD
Hold/Save function active
6
NET
Net weight will be displayed
7
WEIGHT
Current weight value will be displayed
6.1 Controls 4-button terminal
6.1.1 Display
Overview of display
MPB / MXS -BA-e-1113
19
Key
Description
ON/OFF
Turn on/off balance
0
Balance will be reset to 0.0 kg.
Possible up to max. 2 % or 100% of maximum load
(can be selected in the menu)
In BMI Mode: Acceptance of input
HOLD
Hold function/Calculation of a stable weight value
BMI
Function call-up: Body Mass Index
In BMI Mode: Moving the input point
TARE
Tare balance
In BMI Mode: Reducing the input value
6.1.2 Keyboard overview
MPB / MXS -BA-e-1113
20
7 Using the balance
7.1 Weighing
Turn on balance by pressing the [ON/OFF] key. The balance carry out a seg-
ment test, then the program version is displayed. Your balance is ready to
weigh as soon as the “0.0 kg" display appears.
Information: The [0] key can be used to set the balance to zero at any
time.
Place person in the centre of the balance or place baby in the scale pan. Wait
until the standstill display (ο) appears, then read the weighing result.
Note:
If the person is heavier than the weighing range, "Err" (=overload) will appear in the
display.
7.2 Taring
The tare weight of any preloads can be deducted by pressing a button so that the
actual weight of the person is displayed in subsequent weighings.
The balance will display 0 if, for example, a rubber mat has been placed
on the weighing plate.
To start the taring process press the [TARE] key. The weight will now be
saved internally and 0.0 kg will be displayed.
Place the person in the middle of the weighing plate.
Now read the weight from the display.
Information:
The balance is able to only store one taring value at a time.
When the balance is unloaded the saved taring value is displayed with negative sign.
Remove all items from the weighing plate in order to delete the stored tare value and
subsequently press the [TARE] key.
MPB / MXS -BA-e-1113
21
7.3 Hold function (Standstill function)
The balance has an integrated standstill function (mean value calculation). With this
function it is possible to weigh people accurately even if they do not stand still on the
weighing plate.
Note: There is no average value calculation in the event of too much movement.
Turn on balance by pressing the [ON/OFF] key. The balance will carry out a
self-test Your balance is ready to weigh as soon as the “0.0 kg" display appears.
Place the person in the middle of the weighing plate.
Press the [HOLD] button. A triangle starts to flash in the display, during this
time the balance will record several measuring values and will then display the
calculated average value.
Press the [HOLD] button repeatedly to reset the balance to normal Weighing
mode.
Press the [HOLD] button once more to repeat this function as often as re-
quired.
7.4 Calculation of the Body Mass Index
After the balance have displayed 0.0 kg at standstill, the person should stand in the
middle of the weighing plate. Wait until the weighing value has stabilised. Then keep
the BMI button depressed for 3 seconds. Now enter the body height.
It should be noted that a reliable calculation of the BMI is only possible with a body
height of between 100cm and 250cm and a weight >10kg.
The lefthand figure of the body height last entered will flash in the display.
e.g. „1“70.0. Press the Hold button to reduce the item and the Tare button to reduce the value. Press the 0button to confirm the input and then the BMI of the
person will be displayed.
When the BMI value is displayed, it is presented in the display with an arrow pointing
to BMI. To return to normal Weighing mode, press the0button once more and
the arrow pointing to BMI will disappear.
MPB / MXS -BA-e-1113
22
Categorie
BMI ( kg/m2)
Risk of diseases associated with
overweight
Underweight
< 18.5
low
Normal weight
18.5 – 24.9
Average
Overweight
Pre-adipose
Adipose degree I
Adipose degree II
Adipose degree III
> 25.0
25.0 – 29.9
30.0 – 34.9
35.0 – 39.9
> 40
A bit high
High
High
Very high
7.4.1 Classification of BMI values
Weight classification for adults over 18 years of age using the BMI in accordance
with WHO, 2000 EK IV and WHO 2004.
MPB / MXS -BA-e-1113
23
8 Error message
Messages can be displayed when the balance are switched on or in operation.
Lo : Batteries are run down
ErrL : The balance detects underload
ooooo:
The weighing plate was loaded during switch-on, empty the weighing plate.
Err: Overload, too high load on weighing plate
9 Service, maintenance, disposal
9.1 Cleaning
Disconnect the unit from the mains power supply prior to cleaning.
Do not use aggressive cleaning agents (solvents or similar) but only a damp cloth
with mild soapsuds, or household cleaner. Ensure that no liquid gets inside the unit
and rub with a soft, dry cloth.
Loose dirt can be removed carefully with a brush or hand-held vacuum cleaner.
Balance must not be tilted or rotated for cleaning, as damage may occur.
Remove dirt immediately.
9.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized by KERN.
Before opening, disconnect from power supply.
9.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.
MPB / MXS -BA-e-1113
24
Fault
Possible cause
The displayed weight does
not glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
(Rechargeable) batteries are inserted incorrectly or
empty
No (rechargeable) batteries inserted.
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with foreign bodies
or is not correctly positioned.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is obviously incorrect
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
10 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
MPB / MXS -BA-e-1113