KERN & SOHN KXS-TM User Manual

KERN & Sohn GmbH
Ziegelei 1
E-Mail: info@kern-sohn.com
Internet: www.kern-sohn.com
Operating and Inst allation Instructions Display Unit
KERN KXS/KXG-TM
11/2014
Version 2.0
GB
D-72336 Balingen
Fax: +49-[0]7433-9933-149
KXS/KXG-TM-BA_IA-e-1420
GB
KERN KXS/KXG-TM
Display unit
Version 2.0 11/2014
Operating and installa t ion instructions
Contents
1 Technical data ................................................................................................... 5
2 Appliance overview ........................................................................................... 6
2.1 Keyboard overview .................................................................................................... 7
2.1.1 Models KXS-TM: ................................................................................................................ 7
2.1.2 Models KXG-TM: ............................................................................................................... 8
2.2 Overview of display ...................................................................................................10
3 Basic Information (Gen eral) ........................................................................... 12
3.1 Proper use ................................................................................................................12
3.2 Improper Use ............................................................................................................12
3.3 Warranty ...................................................................................................................12
3.4 Monitoring of Test Resources ...................................................................................13
4 Basic Safety Precauti o ns ................................................................................ 13
4.1 Pay attention to the instructions in the Operation Manual .........................................13
4.2 Personnel training .....................................................................................................13
5 Transport and storage .................................................................................... 13
5.1 Testing upon acceptance ..........................................................................................13
5.2 Packaging / return transport ......................................................................................13
6 Unpacking and placing ................................................................................... 14
6.1 Installation Site, Location of Use ...............................................................................14
6.2 Unpacking and placing ..............................................................................................14
6.3
6.4 Transport Securing ...................................................................................................15
6.5 Placing ......................................................................................................................15
6.6 Mains connection ......................................................................................................15
6.7 Rechargeable battery operation (Factory option) ......................................................15
6.8 Adjustment ................................................................................................................16
6.9 Verification ................................................................................................................16
Scope of delivery / serial accessories: ......................................................................14
2 KXS/KXG-TM-BA_IA-e-1420
7 Basic Operation ............................................................................................... 18
7.1 Start-up .....................................................................................................................18
7.2 Switching Off ............................................................................................................18
7.3 Zeroing .....................................................................................................................18
7.4 Simple weighing ........................................................................................................18
7.5 Switch-over weighing unit .........................................................................................19
7.6 Weighing with tare ....................................................................................................19
7.6.1 Taring ............................................................................................................................... 19
7.6.2 Delete tare ....................................................................................................................... 19
7.6.3 Numerical input of tare (PRE-TARE) ............................................................................... 20
7.7 Gross/Net call-up ......................................................................................................20
8 Applications: .................................................................................................... 21
8.1 Parts counting ...........................................................................................................21
8.2 Totalization ...............................................................................................................23
9 Menu ................................................................................................................. 27
10 Indicator settings (menu block "01 FnC") ................................................. 28
10.1 Backlighting of the display (function "FnC 01") ..........................................................31
10.2 Automatic power cut "Auto power off (function „FnC 02) ..........................................32
10.3 Weighing with tolerance range (function "FnC 03 „) ..................................................33
10.4 Resetting to factory settings (function "FnC 04 „) ......................................................
10.5 Automatic zero-adjusting/ stability and response settings (function "FnC 05) ..........38
10.6 Data Hold functions (function "FnC 06) ....................................................................40
10.6.1 Animal weighing function ............................................................................................. 41
10.7 Automatic reference optimisation with the unit count (function „FnC 07) ..................46
10.8 Automatic storage of zero point (function "FnC 10) ..................................................47
10.9 Function of the F-key (function „FnC 12) ..................................................................48
10.10 Auto Zero ..............................................................................................................49
37
11 External adjustment "Unverified weighing systems „ (menu block "02 EC”) 50
12 Settings RS 232 (menu block "03 rS1”) ..................................................... 53
12.1 Baud rate (function "rS1 01) ....................................................................................56
12.2 Parity (function "rS1 02) ...........................................................................................57
12.3 Output format (function "rS1 03) ..............................................................................58
12.4 Output format (function „rS1 04) ..............................................................................60
12.5 Continuous output rate (function „rS1 05) ................................................................61
12.6 Automatic data output with zero display (function "rS1 06) ......................................62
12.7 Condition for data output (function "rS1 08”) .............................................................63
12.8 6/ 7-digit data format (function "rS1 09”) ...................................................................64
12.9 Set the date/ time (function "rS1 10) ........................................................................65
13 Instruction mode .......................................................................................... 68
KXS/KXG-TM-BA_IA-e-1420 3
14 Data output format ....................................................................................... 70
14.1 7-digit data format (first decimal place not included) .................................................70
14.2 6-digit data format (first decimal place not included) .................................................70
15 Bluetooth (Factory option) .......................................................................... 71
16 Servicing, maintenance, disposal .............................................................. 75
16.1 Clean ........................................................................................................................75
16.2 Servicing, maintenance .............................................................................................75
16.3 Disposal ....................................................................................................................75
16.4 Error messages ........................................................................................................75
17 Instant help ................................................................................................... 76
18 Installation display appliance/ load cell ..................................................... 77
18.1 Weighing system design ...........................................................................................77
18.2 Attach the load cell ...................................................................................................78
18.3 Attach RS232 ...........................................................................................................79
19 Service menu ............................................................................................... 80
19.1 Configure weighing system / menu block ..............................................82
19.1.1 Unverified weighing systems (function „CSP 01”) ....................................................... 83
19.1.2 Verified balancing systems (function "CSP 02) .......................................................... 87
19.1.3 Set the buffeting point (function „CSP 03) ................................................................. 91
19.2 Linearization/ menu block .....................................................................93
19.3 Adjustment of "verified weighing systems / menu block " ...................97
19.4 Amend the local gravitation zone/ menu block .................................... 100
19.5 Appliance settings/ menu block ..................................................... 101
19.5.1 Parameter „adjustment to site conditions (function „CFn 01”) ................................. 102
19.5.2 Configuration „verified/ unverified (function „CFn 02) ............................................. 105
19.5.3 Resetting to zero when switched on (function "CFn 04) .......................................... 106
19.5.4 Amend the DATA Hold function of verified weighing system (function „CFn 05“) .... 107
19.5.5 Auto zero (verified weighing systems)/ function "CFn 0 6 ........................................ 109
4 KXS/KXG-TM-BA_IA-e-1420

1 Technical data

KERN KXS-TM KXG-TM
Display 6-digit
Single (Max.) 10,000 e
Solution verifiable
Dual (Max.) 5,000 e Resolution non-verifiable 30,000 d Verification class III Weighing ranges Max. 2 Weighing Units g, kg Divisions 1,2,5,…10, n Display LCD 55 mm digits with back lighting DMS weighing cells
Electric Supply
Rechargeable battery optional
Factory option
Admissible ambient temperature
Humidity of air < 85 % relative (not condensing) Net weight 2500 g 2000 g Dimensions Width x Depth
x Height, (mm)
Interfaces Factory option
Operating time (backlight on) 40 h Operating time (backlight off) 80 h
Max. 8 x 350
Input voltage 110-230 V AC
Built-in power supply unit
6 V, 4.5 Ah
Loading time 12 h
-10°C – 40°C
232 x 170 x 80
RS232: KXS-A04 RS485: KXS-A01
Bluetooth: KXS-A02
Protection type IP 68 according to DIN 60529
KXS/KXG-TM-BA_IA-e-1420 5

2 Appliance overview

1
Wireless
2
Status of rechargeable battery
3
Wall bracket
4
Locking screw
5
Display
6
Keyboard
7
Electric Supply
8
RS 232 / RS485 (Factory option)
9
Connection to the load cell
6 KXS/KXG-TM-BA_IA-e-1420

2.1 Keyboard overview

2.1.1 Models KXS-TM:

Button Description Function
ON/OFF button
F-key
BG/ NET key
UNIT-key
TARE button
ZERO key
1. Turn on/off
Functional button (allocation see chap. 10.1 FNC-03)
Gross weight switch over to net weight
Weighing unit
Taring
Confirm entry
Zeroing
Functional button (allocation see chap. 10.1
M+/ PRINT key
KXS/KXG-TM-BA_IA-e-1420 7
FNC-02)
2.1.1.1 Navigation keys and their function in the menu
Increase flashing digit

2.1.2 Models KXG-TM:

Scroll forward in menu
Decrease flashing digit
Scroll backwards in menu
Digit selection to the right
Take over selected setting
Digit selection to the left
Button Description Function
ON/OFF button
F-key
8 KXS/KXG-TM-BA_IA-e-1420
Turn on/off
Functional button (allocation see chap. 10.1
FNC-03)
BG/ NET key
Gross weight switch over to net weight
UNIT-key
TARE button
ZERO key
M+/ PRINT key
2.1.2.1 Navigation keys and their function in the menu
Increase flashing digit
Scroll forward in menu
Weighing unit
Taring
Confirm entry
Zeroing
Functional button (allocation see chap. 10.1
FNC-02)
Decrease flashing digit
Scroll backwards in menu
Digit selection to the right
Take over selected setting
Digit selection to the left
KXS/KXG-TM-BA_IA-e-1420 9

2.2 Overview of display

HI / OK / LO Indicators for weighing with tolerance range
Kg Current selected unit of weight „kilograms
Lb Current selected unit of weight „pound
Pcs Indicator for counting
Zero indicator
Net The displayed weighing value is a net weighing value
Status of rechargeable battery
10 KXS/KXG-TM-BA_IA-e-1420
Non verified weighing systems:
3
4
5
7
2
3
 
6
7
Indicator [] next to symbol displays:
1
that the weight value is stable
GROSS
2
that the displayed weighing value is a gross weight value that the balance is set to unit counting mode
PRE-
TARE
HOLD
M+
that a PRE-TARE value is stored
that the indicated weight value is held in the display, until it is
deleted.
6
that data are stored in an accumulative memory
T hat the current selected unit of weight is „oz.
Verified weighing systems:
Indicator [] next to symbol displays:
1
that the weight value is stable
GROSS
that the displayed weighing value is a gross weight value that the balance is set to unit counting mode
PRE-
TARE
4
that a PRE-TARE value is stored
5
Display of the current weighing range of the connected platform
M+
that data are stored in an accumulative memory
KXS/KXG-TM-BA_IA-e-1420 11

3 Basic Information (General)

3.1 Proper use

The display unit acquired by you is used in combination with a load cell and serves to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic weighing system”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

3.2 Improper Use

Do not use weighing system for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the display unit. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing pan. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the weighing plate, minus a possibly existing tare load, must be strictly avoided. Both, the w ei g hi ng plate and the display unit may be damaged during this process. Never operate display unit in explosive environment. The serial version is not explosion protected. Changes to the display unit's design are not permitted. This may lead to incorrect weighing results, safety-related faul ts and des tr uct ion of the display unit. The display unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing.

3.3 Warranty

Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
12 KXS/KXG-TM-BA_IA-e-1420

3.4 Monitoring of Test Resources

Keep all parts of the original packaging for a possibly required
and remove loose/mobile
Secure all parts such as the glass wind screen, the weighing
In the framework of quality assurance the measuring-related properties of the display unit and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of display units’ test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and display units may be calibrated (return to the national standard) fast and at moderate cost.

4 Basic Safety Precautions

4.1 Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN balances.
All language versions contain a non-binding translation.
The original German is binding.

4.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

5 T ransport and storage

5.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.

5.2 Packaging / return transport

return. Only use original packaging for returning. Prior to dispatch disconnect all cables
parts. Reattach possibly supplied transport securing devices.
platform, power unit etc. against shifting and damage.
KXS/KXG-TM-BA_IA-e-1420 13

6 Unpacking and placing

For display unit, see chapter 2

6.1 Installation Site, Location of Use

The display units are designed in a way that reliable weighing results are achieved in common conditions of use. Precise and fast work is achieved by selecting the right place for your display unit and your weighing plate. On the installati on site obs erve the following:
Place the weighing system on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the display unit and the weighing plate against direct dra ft fr om ope n windows or doors.
Avoid jarring during weighing;
Protect the display unit and the weighing plate against high humidity, vapours
and dust.
Do not expose the display unit to extreme dampness for longer periods of time. Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.

6.2 Unpacking and placing

Take the display unit carefully out of its packaging, remove the plastic jacket and install it at the designated work space. Mount the displ ay unit in a way that facilitates operation and where it is easy to see.

6.3 Scope of delivery / serial accessories:

Plug for connection to load cell
Operating manual
14 KXS/KXG-TM-BA_IA-e-1420

6.4 Transport Securing

Remove the transportation safety device at the four marked positions:
Transport Securing

6.5 Placing

Mount the display unit in a way that facilitates operation and where it is easy to see.
In order to raise the display, the display device can be mounted on an optional stand.

6.6 Mains connection

Power is supplied via the external power supply unit. Insert the mains plug into the wall socket.
Once connected, the instrument performs a self check. If the zero display shows the device is ready for use.

6.7 Rechargeable battery operation (Factory option)

Charge the battery for at least 12 hours before initial use. The battery symbol indicates the current charge level of the batteries. If the symbol
flashes, the capacity of the battery will soon be exhausted. The weighing scale will remain ready for operation for a few more hours before switching off in order to save battery. Fully charge the battery pack.
KXS/KXG-TM-BA_IA-e-1420 15

6.8 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each display unit with connected weighing plate must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the weighing system has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the display unit periodically in weighing operation.
How to carry out adjustment:
Unverified weighing systems, see chapter 11Verified weighing systems, see chapter 19.3

6.9 Verification

General introduction:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory. c) For official purposes d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard. Verification notes:
An EU type approval exists for the appliance described in its technical data as verifiable. If the appliance is used where obligation to verify exists as described above, it must be verified and re-verified at regular intervals. Re-verification of an appliance is carried out according to the respective national regulations. Normally the validity for verification in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
Verification of the device is invalid without the seals.
The seal marks / verification wire attached on verified appliances p oi nt out that the appliance may only be opened and serviced by trained and authorised specialist staff. If the sealing (paper seal/ verification wire) is damaged, the verification validity expires. Please observe all national laws and legal regulations. In Germany a re-verification will be necessary.
16 KXS/KXG-TM-BA_IA-e-1420
Sealing:
KXS/KXG-TM-BA_IA-e-1420 17

7 Basic Operation

7.1 Start-up

Press on/ off key, the equipment completes a self check. As soon as the weight
display appears, the instrument will be ready to weigh.

7.2 Switching Off

Press the on/ off key approx. 3 seconds, the display will turn off.

7.3 Zeroing

Resetting to zero corrects the influence of light soiling on the weighing plate.
Manual
To unload the weighing system Press the ZERO button, the zero display and the indicator will appear.
Automatic
The automatic zero tracking menu, can be switched on or off, function "FNC 13"
see chapter 10.10. When the balance is cleared the zero point is corrected automatically.

7.4 Simple weighing

Place goods to be weighed on balance. Wait until the indicator over the stability display appears. Read weighing result.
18 KXS/KXG-TM-BA_IA-e-1420

7.5 Switch-over weighing unit

Press the UNIT-key The weight gauge changes between „kgand "g".

7.6 Weighing with tare

After taring no numeric input of a tare weight can follow.

7.6.1 Taring

Deposit weighing vessel. Wait for stability display, then press the TARE button
Zero display and indicator net appear.
The weight of the container is now internally saved. Weigh the material, the net weight will be indicated. After removing the weighing container, the weight of the weighing container
appears as negative display.
The taring process can be repeated any number of times, e.g. when adding
several components for a mixture (adding). The limit is reached when the whole weighing range is exhausted.
With the BG/ NET key you can switch between gross weight and net weight, see
chapter 7.7.
After the numeric input of a tare weight, the system can only be tared if „tare weight > PRE TARE value

7.6.2 Delete tare

Unload balance and press the TARE key.
The NET indicator turns off, the zero display shows.
KXS/KXG-TM-BA_IA-e-1420 19

7.6.3 Numerical input of tare (PRE-TARE)

The known dead weight of a weighing container can be tared off by entering its weight as pre-tare deduction in order to ensure the net weight of the goods to be weighed in subsequent weighings is always displayed.
Press the TARE- key, the active position flashes.
Enter the known tare weight, e.g. 2 kg and acknowledge with TARE button.
Numerical input, see chap. 2.1.1.
The entered weight will be stored as tare weight and displayed with negative sign. The indicator  over PRE-TARE will appear.
Put the filled weighing container on the balance, the net weight will be displayed. The tare value remains stored until it is deleted with the TARE key.

7.7 Gross/Net call-up

By repeated pressing of the BG/ NET key you can change between the gross and net indicator values. At the indicator „gross weight the indi c ator a ppe ar s above LARGE. In the „Net weight“ display the indicator appears next to NET.
20 KXS/KXG-TM-BA_IA-e-1420

8 Applications:

8.1 Parts counting

Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by putting on a certain number of the parts to be counted. The balance determines the total weight and divides it by the number of parts, the so­called reference quantity. Counting is then carried out on the basis of the calculated average piece weight. As a rule: The higher the reference quantity the higher the counting exactness.
Determine the unit weight and count
1. In the weighing mode press the unit key, in order to activate the counting
function. The indicator Pcs is indicated.
2. Press the BG/NET-key repeatedly until the desired reference number of items
(e.g. 100) appears, select between C 10, C 20, C 50, C100, C 200.
3. Place as many items to be counted (such as 10 items) as demanded by the
set reference quantity and confirm by UNIT key. The weighing scales calculate
the reference weight. The current number of items (e.g. 10 pieces) is
indicated.
4. Remove reference weight. The balance is from now in parts counting mode
counting all units on the weighing plate.
KXS/KXG-TM-BA_IA-e-1420 21
Switch between number of items and weight
With the UNIT key you can switch between number of items and weight:
By switching the scale off and then on again the scale starts in the unit counting mode. The current reference weight remains stored until the reference is set again or the power supply is interrupted.
The counting function is not active, if the weight is under 20d or if below the minimum piece weight (< 0.2 D) is reached.
Automatic reference o pt i miz at ion
In order to improve the counting exactness, the referenc e can be opt imised by adding more pieces. At every reference optimisation, the reference weight is calculated anew. As the additional pieces increase the base for the calculation, the reference also becomes more exact.
If the number of placed parts are 5 pieces more than the reference, then the
automatic reference optimisation starts. The reference weight is calculated anew.
22 KXS/KXG-TM-BA_IA-e-1420

8.2 Totalization

F_M 10 (M+ accumulating mode 1):
F_M 11 (M+ accumulating mode 2):
Printout example (KERN YKB-01N)
Printout example (KERN YKB-01N)
2. Weighing
2. Weighing
Output
Output
Required settings in the menu:
1. Function „rSi 04: Setting „M_Pl u s, see chapter 12.4
2. Function „rSi 03, setting F_M 10 or F_M11 (sample print-out see following overview , see chapter 12.3.
Total = sum of net weight of each weight reading
1. Weighing
Total = Sum of gross weight of each weight reading
1. Weighing
The totalizing function is not active when the weight is under 20d.
KXS/KXG-TM-BA_IA-e-1420 23
Accumulation memory
Sum of net weight of each weight reading
Accumulation memory
Gross weight of each weight reading
Add up:
Place weighing good A, e.g. 5 kg.
Wait for stability, then press M + / PRINT key. . The weight is added to the accumulative memory and printed if an optional printer is connected. The number of weighings, followed by the total weight will be indicated. The indicator above M+ shows.
Remove the weighed good. More weighed goods can only be added when the
display ≤ zero.
24 KXS/KXG-TM-BA_IA-e-1420
Place goods to be weighed, e.g. 7 kg.
Wait for stability then press the M+/PRINT, it calculates. The weight value is added to the accumulative memory and given out when connecting an optional printer. Number of weighings, followed by the total weight will be displayed for 2 sec. Then the current weight value appears, the indicator above M + appears.
Add more weighed goods as described before. Please note that the weighing
system must be unloaded between the individual weighing procedures.
You can repeat this process until the capacity of the weighing system is
exhausted.
KXS/KXG-TM-BA_IA-e-1420 25
Display and output sum „Total“:
When the scale is unloaded (zero indicator) press the M+/PRINT key, the number
of weight readings followed by the total weight are indicated for 2 seconds. Press M+/PRINT to print out this display.
Delete accumulative memory:
1. With M+/PRINT : At the display “Number of weighings” press M+/PRINT key again. The data in the accumulative memory are deleted. The indicator  above M+ turns off.
2. With F in the menu use the function "FnC 12" in setting "MC" to allocate the F-key with the function "Delete accumulative memory”, see chapter 10.9 When the scale is unloaded (zero indicator) press the F-key, the data in the accumulative memory are deleted. The indicator  above M+ turns off.
26 KXS/KXG-TM-BA_IA-e-1420

9 Menu

The menu consists of the following blocks:
Access to menu: In the weighing mode press the BG/NET and ZERO key simultaneously, "01 FnC"
is indicated.
Settings of the display device External adjustment Settings of the interface Exit menu / back to weighing mode.
KXS/KXG-TM-BA_IA-e-1420 27

10 Indicat or s e t ti ngs (me nu block "01 FnC")

Press TARE key
Access to menu: In the weighing mode press BG/ NET and ZERO key simultaneously "01 FnC" is
indicated, see chapter 9.
Keys and their function in the menu:
/
/
/
/
Call up the navigation in the menu/ function:
Increase flashing digit
Scroll forward in menu
Decrease flashing digit
Scroll backwards in menu
Digit selection to the right
Digit selection to the left
With navigation keys Select function, (FNC 00  FNC 13)
Back to menu: Select function FNC 00 and confirm with TARE key.
28 KXS/KXG-TM-BA_IA-e-1420
Exit menu / back to weighing mode:
Leave the example function "FnC 01"
Select with the navigation keys "FNC 00” and confirm with TARE key.
KXS/KXG-TM-BA_IA-e-1420 29
Overview:
FnC 00 Exit menu FnC 01 Display background illumination FnC 02 Automatic power switch-off FnC 03 Weighing with tolerance range / set limit values FnC 04 Reset to default setting FnC 05 Automatic zero tracking / stability and response settings FnC 06 Data-Hold function FnC 07 Automatic reference optimization with counting of items FnC 08 Not documented FnC 09 Not documented FnC 10 Automatic saving of zero point (not available in the case of
weighing systems with type approval).
FnC 11 Not documented FnC 12 Function of the F-key FnC 13 Auto-Zero function
(not available with balancing systems with approved design)
30 KXS/KXG-TM-BA_IA-e-1420

10.1 Backlighting of the display (function "FnC 01")

Call up function "FnC 01", see chapter 10 „Navigation in the menu“.
Press the TARE-key , the current setting for the back lighting of the display is
indicated.
Select desired settings with TARE or UNIT key.
bl off Display background illumination off bl on Starting from a weight > 10 d the back lighting of the display is
switched on automatically. If there is no activity at the appliance for10 seconds or at the zero indicator the back lighting is switched off autom atically.
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
KXS/KXG-TM-BA_IA-e-1420 31

10.2 Automatic power cut "Auto power off (function „FnC 02)

The indicator is switched off automatically within the adjusted time,
Call up function "FnC 02", see chapter 10 "Navigation in the menu”.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
A off0 Function switched off A off1
if the weighing system is not used and/ or the weight < 10 D.
A off9
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
Selectable between 1 - 9 minut es .
32 KXS/KXG-TM-BA_IA-e-1420

10.3 Weighing with tolerance range (function "FnC 03 „)

You can set an upper or lower tolerance limit when weighing with tolerance range and thus ensure that the weighed load remains exactly within the set tolerance limits. Tolerance checks such as dosing, portioning and sorting the display above or below the tolerance levels with a visual and acoustic signal (depending on the setting in the menu).
Optical signal:
Three indicators show whether the material being weighed is located within the two tolerance limits. The indicators provide the following information:
Weight less than nominal weight and below the lower tolerance
Weight within tolerance limits
Weight more than the nominal w ei g ht and above the upper tolerance
KXS/KXG-TM-BA_IA-e-1420 33
Settings:
Call up function "FnC 03", see chapter 10 "Navigation in the menu“.
Press the TARE key, the current settings of the upper tolerance is indicated.
With the navigation keys (see chapter XX) enter the desired value e.g. 3 kg.
Press the TARE key repeatedly until the display for the input of the lower
tolerance appears.
With the navigation keys (see chapter 2.1.1) Enter the desired value e.g. 2.9 kg.
34 KXS/KXG-TM-BA_IA-e-1420
Press the TARE key until the display for the settings of the audio signal appears.
The current setting will be displayed.
With the navigation keys (see chapter 2.1.1) select the desired setting.
Settings of audio signal:
Selectable mode Description
(a) ⇒
(b) ⇒
(c) ⇒
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
1 = Audio signal switched on 0 = Audio signal switched off
1 = Audio signal sounds with stable weight values 0 = Audio signal sounds with unstable weight values
1 = If the object to be weighed is within the tolerance range,
the acoustic signal sounds will be emitted.
0 = If the material being weighed lies outside the tolerance
range and > 10 d the acoustic signal sounds.
KXS/KXG-TM-BA_IA-e-1420 35
Weighing with tolerance range:
Tare when using a weighing container. Put on goods to be weighed, tolerance control is started. The indicators show
whether the load is within the two set tolerance limits.
Load below specified tolerance
Indicator illuminated
To deactivate the function for the tolerance levels enter the value „00.000 kg.
Load within specified tolerance
Indicator illuminated
Load exceeds specified tolerance
Indicator illuminated
36 KXS/KXG-TM-BA_IA-e-1420

10.4 Resetting to factory settings (function "FnC 04 „)

The following functions are reset to factory setting with "FnC 04 „.
1. External adjustment weight
2. Settings for weighing with tolerance range
3. Filter settings
4. Settings of number of items
Call up function "FnC 04", see chapter 10 "Navi g ati on in the menu“.
Press the TARE key. Select desired settings with TARE or UNIT key.
return No resetting to factory settings format Reset to default setting
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
KXS/KXG-TM-BA_IA-e-1420 37
10.5 Automatic zero-adjusting/ stability and response settings
for verified weighing systems see chapter 18.5.1.
(function "FnC 05)
This function is only available for unverified weighing systems (CFn 02 = 0),
Call up function "FnC 05", see chapter 10 "Navigation in the menu“.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
Zero 0
Zero 9
Confirm with TARE key, the current stability settings are indicated.
Switch Automatic zero-adjusting switch on and off and select zero range.
You can select 0 (switched off), 1d, 2d, 3d, 4d, 5d, 6d, 7d, 8d, 9d
38 KXS/KXG-TM-BA_IA-e-1420
FiL 0
the stability of the display and the degree of reaction of the weighing
FiL 9
Confirm with the TARE key, the current response level is displayed.
The higher the value, the more priority is given to stability. You can select 0-9. Factory setting "0".
3 db 0
3 db 9
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
The higher the value, the more priority is given to response. You can select 0-9. Factory setting "5".
With the functions "FiL" and "3 DB", there is the possibility to select
system on the requirements of specific applications or environmental conditions. Please note that in general slowing down reaction times result in higher stability of the set data handling, while speeding up reaction times have an influence on the stability deterioration.
KXS/KXG-TM-BA_IA-e-1420 39

10.6 Data Hold functions (function "FnC 06)

(changes are possible with function "CFn 05" see chapter 18.5.4.).
Peak value function
is deleted with any key.
„Stable hold 1“ mode
up to the manipulation of any key in the display.
„Stable hold 2“ mode until the load falls under 10d.
Animal weighing
chapter 10.6.1
With verified balancing systems (Cfn 02 = 1) the setting „hold = 0 cannot be altered with the function „FnC 06
Call up function "FnC 06", see chapter 10 "Navigation in the menu“.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
hold 0 Function switched off (factory setting) hold 1
hold 2
hold 3
hold 4
This function indicates the highest load factor ( peak val ue) of a continuously rising load. The peak value remains in the display until it
The weight value is held automatically after reaching a stable value
The weight value is held in the display after reaching a stable value
This function is suitable for jerky weighing procedures, see following
40 KXS/KXG-TM-BA_IA-e-1420
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode function, see chap. 10.

10.6.1 Animal weighing function

With this function jerky balancing goods can be weighed, e.g. living animals. The scale calculates an aver ag e v al ue by the number o f adjusted weighings and displays this until the scale is cleared (display < 10d).
Settings:
Call up setting "hold 4", see. chapter 10.6.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
1%
Range of display fluctuation, you can select 1 -100%. Factory setting „10“
100%
KXS/KXG-TM-BA_IA-e-1420 41
Press the TARE key repeatedly until the current setting "number of weight
readings" is indicated.
1
Number of weight readings, which are consulted for the average value calculation, you can select 1, 2, 4, 8, 16, 32, 64. Factory setting "8
64
Confirm with the TARE key, the display returns to the menu.
Select with the navigation keys "FNC 00” and confirm with TARE key.
42 KXS/KXG-TM-BA_IA-e-1420
From now on the scale is in animal weighing mode.
Animal weighing:
When the animal weighing function is activated horizontal segments appear at the zero display.
Place goods to be weighed on the balance. The scale calculates the average
value by the number of adjusted weight readings. The indicator appears above HOLD.
For further measurements unload the scale.
KXS/KXG-TM-BA_IA-e-1420 43
Return to weighing mode:
In the weighing mode press the BG/NET and ZERO key simultaneously, "01 FnC"
is indicated. Press the TARE key, „FNC 00” is indicated.
Select „FnC 00“ with the navigation keys and confirm with the TARE key. „hold 4“
is indicated.
Select „hold 0“ with the navigation keys and confirm with the TARE key.
Confirm with the TARE key, the display returns back to the menu.
Select with the navigation keys "FNC 00” and confirm with TARE key.
44 KXS/KXG-TM-BA_IA-e-1420
KXS/KXG-TM-BA_IA-e-1420 45

10.7 Automatic reference optimisation with the unit count (function „FnC 07 )

determines the weight per unit again, if the number of parts has
Call up function „FnC 07“, see chapter 10 "Navigation in the menu“.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
AvErG 0 No reference optimisation AvErG 1 Reference optimisation is activated. The device automati cal l y
increased
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
46 KXS/KXG-TM-BA_IA-e-1420

10.8 Automatic storage of zero point (function "FnC 10)

This function is only available for unverified weighing systems (CFn 02 = 0).
Call up function "FnC 10", see chapter 10 "Navigation in the menu“.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
off Function switched off (factory setting) on If the function is switched on, the last zero point is stored. After
switching the appliance off and on or after a power failure the equipment with the stored zero point continues to work.
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
KXS/KXG-TM-BA_IA-e-1420 47

10.9 Function of the F-key (function „FnC 12)

The weight value 10 x higher resolution is displayed when the F key is
Call up function "FnC 12", see chapter 10 "Navigation in the menu“.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
3 different functions can be assigned to the F-key:
MC The accumulative memory is cleared when the F-key is pressed. hr
pressed.
t-tP
PRE-TARE
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
48 KXS/KXG-TM-BA_IA-e-1420

10.10 Auto Zero

For verified weighing systems see chapter 18.5.5
This function is used to tare small variations in weight automatically. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“. (e.g. slow flow of liquids from a container placed on the balance, evaporating processes). When apportioning involves small variations of weight, it is advisable to switch off this function.
This function is only available for unverified weighing systems (CFn 02 = 0).
Call up function "FnC 13", see chapter 10 "Navigation in the menu“.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
off Function switched off (factory setting) on W hen the balance is cleared the zero point is corrected automatically.
Confirm with the TARE key, the display returns back to the menu.
Return to weighing mode, see chap. 10.
KXS/KXG-TM-BA_IA-e-1420 49
11 External a djust m ent "Unverified weighing s ystems „
warm-up time.
(menu block "02 EC”)
Prepare the required adjustment weight. The weight to be used depends
Procedure:
In the weighing mode press the BG/NET and ZERO key simultaneously, "01 FnC"
is indicated. Select "02 EC" with the navigation keys.
on the capacity of the scal e. Carry out adjustment as near as possible to the scale’s maximum weight. Info about test weights can be found on the Internet at: http://www.kern-sohn.com.
Observe stable environmental conditions. Stabilisation requires a certain
Press the TARE button, the currently set adjustment weight is displayed. The last
digit flashes.
To change the adjustment weight select the digit to be changed with BG/ NET key, the active digit flashes. Enter the desired value with the navigation keys. Press the TARE key repeatedly, until "-----“ is indicated.
50 KXS/KXG-TM-BA_IA-e-1420
Ensure that there are no objects on the weighing pan. Press the TARE key.
Press the BG/ NET key again, the display returns to the menu "02 EC".
The weight value of the required adjustment weight will be displayed.
Carefully place the adjusting weight into the centre of the balanci ng plate and
press the TARE key. After successful adjustment, an acoustic signal sounds, the scale will automatically return to the me nu. Tak e away adjustment weight .
Cancel the adjustment :
Press the BG/ NET key repeatedly, until „PASS is indicated.
KXS/KXG-TM-BA_IA-e-1420 51
Return to weighing mode:
52 KXS/KXG-TM-BA_IA-e-1420

12 Settings RS 23 2 (menu bl oc k "0 3 rS1”)

Press TARE key
Access to menu:
In the weighing mode press the BG/NET and ZERO key simultaneously, "01 FnC"
is indicated. Select "03 rS1" with the navigation keys.
Call up the navigation in the menu/ function:
Select the function with navigation keys (rS1 00 rS1 13)
Zurück ins Menü: Select function rbl 00 and confirm with the TARE key.
KXS/KXG-TM-BA_IA-e-1420 53
Keys and their function in the menu: see chap. 10
Exit menu / back to weighing mode:
For example to leave function "rS1 01"
Select "rS1 00" with the navigation keys and confirm with the TARE key.
54 KXS/KXG-TM-BA_IA-e-1420
Overview:
rS1 00 Exit menu rS1 01 Baud rate rS1 02 Parity rS1 03 Output format rS1 04 Output mode rS1 05 Continuous output rate rS1 06 Automatic data output with zero display. rS1 07 Not documented (Factory setting “0”) rS1 08 Set output stable/instable weighing values rS1 09 6 /7 -digit data format rS1 10 How to set date/time rS1 11 Set the date format rS1 12 Activate R-S 485 rS1 13 Not documented
KXS/KXG-TM-BA_IA-e-1420 55

12.1 Baud rate (function "rS1 01)

Call up function "rS1 01", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
b600
b19200
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
You can select 600/1200/2400/4800/9600 (bit/ second)
56 KXS/KXG-TM-BA_IA-e-1420

12.2 Parity (function "rS1 02)

P n81
P E71
P o71
Call up function "rS1 01", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
No parity, 8 bits, 1 stop bit Even parity 7 bits, 1 stop bit Odd parity, 7 bits, 1 stop bit
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
KXS/KXG-TM-BA_IA-e-1420 57

12.3 Output format (function "rS1 03)

Call up function "rS1 01", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
F_M 00
Factory
settings
Display value with status Sample printout:
ST, G + 1.000 kg
F_M 01 Gross weight with status F_M 02 Net weight with status
Display value without status
F_M 03
sample printout: + 1.000
F_M 04 Gross weight without status F_M 05 Net weight without status
58 KXS/KXG-TM-BA_IA-e-1420
F_M 06 Hi /Lo/ OK status + display value without status
ST
Stable value
US
Instable value
G
Gross weight
N
Net weight
T
Tare weight
„Accumulation
TOTAL
Total of all individual weighings
sample printout:
Lo (001) 001+ 2.000 OK (010) 010+ 3.000 Hi (100) 100+ 4.000
F_M 07 Hi /Lo/ OK status + gross weight without status F_M 08 Hi/ Lo/ Ok status + net weight without status F_M 09 Tare weight
F_M 10
M+ accumulation mode 1, sample printout see chapter 8.2 Total = sum of net weight of each weight reading
M+ accumulation mode 2, sample printout see chapter 8.2
F_M 11
Total = Sum of gross weight of each weight reading
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
Symbols:
TICKET NO Number of weight readings
KXS/KXG-TM-BA_IA-e-1420 59

12.4 Output format (function „rS1 04)

Comand
Stream
Auto
Automatic issue of first stable weighing value
3. Place next weight.
rS-off
M_Plus
Call up function "rS1 04", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
Data output via remote control commands Continuous output
Sequence of operations:
1. Place weight, issue of first stable weighing value
2. Renewed output only possible after weight was removed.
Interface deactivated Add-up mode
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
60 KXS/KXG-TM-BA_IA-e-1420

12.5 Continuous output rate (function „rS1 05)

Call up function "rS1 05", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
rPS 1
rPS 16 MAII Data transmission rate >16 times/ second
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
Data transmission rate 1, 2, 4, 8, 16 times/ second Factory setting rPS 4
KXS/KXG-TM-BA_IA-e-1420 61

12.6 Automatic data output with zero display (function "rS1 06 )

Call up function "rS1 06", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
-b 00
-b 99
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
Range, below which appliance has to go before the scale prints, you can select 0 - 99 division steps.
Factory setting "- b05".
62 KXS/KXG-TM-BA_IA-e-1420

12.7 Condition for data output (function "rS1 08”)

Call up function "rS1 08", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
Stb-P
Factory setting
Output only with stable weight values (no output with unstable weight values and "OL „)
StoL-p Output only for stable weight values (including „OL“) ALL-P Output of all weight values
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
KXS/KXG-TM-BA_IA-e-1420 63

12.8 6/ 7-digit data format (function "rS1 09”)

Call up function "rS1 09", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
S iII
6 Digits
Factory setting
SEvEn 7 Digits
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
64 KXS/KXG-TM-BA_IA-e-1420

12.9 Set the date/ time (function "rS1 10)

Call up function "rS1 09", see chapter 12.
Press the TARE-key, current setting appears. Set the date with TARE or UNIT key.
Press the TARE key repeatedly until the last digit flashes, the display for the time
setting appears.
Press the TARE key repeatedly until the last light flashes, the display returns back
to the menu.
To go back into the weighing mode see chapter 12.
KXS/KXG-TM-BA_IA-e-1420 65
Set the date format (function "rS1 11) Call up function "rS1 11", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
y_m_d Year/ month/ day
d _m_j Day/Month/Year
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
66 KXS/KXG-TM-BA_IA-e-1420
RS 485 ID input (function "rS1 12)
Call up function "rS1 12", see chapter 12.
Press the TARE-key, current setting appears. Select desired settings with TARE or UNIT key.
Id 00
Id 99
Confirm with the TARE key, the display returns back to the menu.
To go back into the weighing mode see chapter 12.
ID input, you can select 0 - 99 With the input „ID 00 the remote control
instructions are the same as with RS232, see chapter 13
KXS/KXG-TM-BA_IA-e-1420 67

13 Instruc ti on m ode

RS 485:
During input „Id 00“(function RS1 12, see chapter 12.12) the instruction format is
the same as with RS232.
During input "ID 01-99“ "@ID „m ust be placed in front of the remote control
command, so that the respective scale reacts to the instruction; e. g. to set the scale to zero with ID 99 the instruction must be sent to "@99MZ“ followed by the control characters <CR><LF>.
Orders:
MZ Zeroing MT Taring MG Display gross weight MN Indicate net weight CT Delete tare SC Continuous data output SA Automatic data output SO Instruction mod e % Stop continuous data output and start instruction mode. UA* Switch over to the first weighing unit. UB* Switch over to the second weighing unit* UC* Switch over to the third weighing unit* UD* Switch over to the forth weighing unit* UE* Switch over to the fifth weighing unit* UF* Switch over to the sixth weighing unit* RW Send current weight value RG Send gross weight RN Send net weight RT Send tare weight RB Send the display value without status RH Send gross weight without status RI Send net weight without status
68 KXS/KXG-TM-BA_IA-e-1420
RJ Hi /Lo/ OK status + display value without status
Hi/ Lo/ OK status
Sample display
RS
Show tolerance limit : Class (00 ~ 99) *  : LO or HI
WS
Set tolerance limit : Class (00 ~ 99) *  : LO or HI XXXXXX: Enter the limit value
#######
Example: Value
Position decimal point
1 2 3 4 5
6
1
Example:
Lo (001) 001+ 2.000 OK (010) 010+ 3.000 Hi (100) 100+ 4.000
RK Hi/Lo/ OK status + gross weight without status (example see "RJ"). RL Rear one/ Lo/ OK status + net weight without status (example see "RJ").
RS
HI Display of the adjusted upper limit value LO Display adjusted lower limit value
Example: Command RS02LO<CR><LF>
Response RS02LOXXXXXX<CR><LF>
WS XXXXXX
HI Enter the value for the upper limit LO Enter the value for the lower limit
Example: Command WS00HI001000<CR><LF>
Response WS00HI001000<CR><LF>
Display
*dependent on model Should it be sent continuously, the command must be preceded by %. If stable weight values are to be sent, the command must be preceded by #.
CR LF
KXS/KXG-TM-BA_IA-e-1420 69

14 Data output format

Byte0
Byte1
Byte2
+/- 1 . 2 3 . 4 5 6
CR
LF
Byte0
Byte1
Byte2
+/- . 2 3 . 4 5 6 CR
LF

14.1 7-digit data format (first decimal place not included)

Weight value with status
Gross S T , G S , + 0 1 2 3 4 5 6 7 SP SP o z Net S T , N T , + 1 . 2 3 . 4 5 6 t l . g Tare S T , T R , + 0 1 2 . 3 4 5 6 SP SP k g Plus OL O L , G S , + SP SP SP SP SP SP SP SP SP SP SP SP Minus OL O L , G S , - SP SP SP SP SP SP SP SP SP SP SP SP Unstable U S , G S , + 0 1 2 3 4 . 5 6 SP SP l b
Weight value without status
G / N + 1 . 2 3 . 4 5 6 G / N + 0 1 2 3 4 5 . 6 G / N + 0 1 2 . 3 4 5 6 Plus OL + SP SP SP SP SP SP SP SP Minus OL - SP SP SP SP SP SP SP SP
Hi /Lo/ OK status + display value without status:
CR LF
CR LF
Byte0: HI 30H / 31H Byte1: OK 30H / 31H Byte2: Lo 30H / 31H

14.2 6-digit data format (first decimal place not included)

Weight value with status
Gross S T , G S , + 1 2 3 4 5 6 7 SP SP o z Net S T , N T , + . 2 3 . 4 5 6 t l . g Tare S T , T R , + 1 2 . 3 4 5 6 SP SP k g Plus OL O L , G S , + SP SP SP SP SP SP SP SP SP SP SP Minus OL O L , G S , - SP SP SP SP SP SP SP SP SP SP SP Unstable U S , G S , + 1 2 3 4 . 5 6 SP SP l b
Weight value without status
G / N + . 2 3 . 4 5 6 G / N + 1 2 3 4 5 . 6 G / N + 1 2 . 3 4 5 6 Plus OL + SP SP SP SP SP SP SP Minus OL - SP SP SP SP SP SP SP
CR LF
Hi /Lo/ OK status + display value without status:
CR LF
Byte0: HI 30H / 31H Byte1: OK 30H / 31H Byte2: Lo 30H / 31H
70 KXS/KXG-TM-BA_IA-e-1420

15 Bluetooth (Fa c t ory option)

Click in the task border with activated Bluetooth.
Click on „Add device“.
KXS/KXG-TM-BA_IA-e-1420 71
Click on "Enter pairing code of the device"
Enter code 1111
72 KXS/KXG-TM-BA_IA-e-1420
Click on „Next“
Click on „Close“
KXS/KXG-TM-BA_IA-e-1420 73
Display Bluetooth network appliance
Double c lic k
74 KXS/KXG-TM-BA_IA-e-1420

16 Servicing, maintenance, disposal

Before any maintenance, cleaning and repair work disconnect the appliance from the oper ati ng v ol tag e.

16.1 Clean

Keep IP protection. Clean the stainless-steel parts with a soft cloth soaked with a cleaning agent
suitable for stainless steel.
For stainless steel parts do not use any cleaning agents which contain sodium
hydroxide solution, acetic, hydrochloric, sulphuric or citric acid.
Do not use metal brushes or cleaning sponges of steel wool, as this causes
superficial corrosion.

16.2 Servicing, maintenance

The appliance may only be opened by trained service technicians who are
authorized by KERN.
Ensure that the balance is regularly calibrated, see chap. Monitoring of test
resources.

16.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.

16.4 Error messages

Error
message
E 0 EEPROM error value outside of A/ D transducer range E 1 Zero points above the zero adjusting range E 2 Zero points below zero adjusting range E 4 A/D converter oL Overload
-oL Underload oF Internal value < zero range
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
Description
KXS/KXG-TM-BA_IA-e-1420 75

17 Instant hel p

The displayed weight does not glow.
(Rechargeable) batteries are inserted incorrectly or
The displayed weight is
The weighing result is obviously incorrect
In case of an error in the program process, briefly turn off the display unit and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The display unit is not switched on.
Mains power supply interrupted (mains cable
defective).
Power supply interrupted.
empty
No (rechargeable) batteries inserted.
permanently changing
Draught/air movement
Table/floor vibrations
Weighing pan has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
76 KXS/KXG-TM-BA_IA-e-1420

18 Install a ti on display appliance/ load cell

Installation / configuration of the weighing system must be carried out by a well acquainted specialist with the workings of weighing balances.

18.1 Weighing system design

The display unit is suitable for connection to any analogue platform in compliance with the required specifications. The following data must be established before selecting a weighing cell:
Weighing balance capacity This usually corresponds to the heaviest load to be weighed.
Preload This corresponds to the total weight of all parts that are to be placed on the weighing cell such as upper part of platform, weighing pan etc.
Total zero setting range This is composed of the start-up zero setting range and the zero setting range available to the user via the ZERO-key. The addition of weighing scales capacity, preload and the total zero setting range give the required capacity for the weighing cell. To avoid overloading of the weighing cell, include an additional safety margin.
Smallest desired display division
Verification ability, if necessary
menu setting CFn 02 = 1, see chapter 18.5.2
KXS/KXG-TM-BA_IA-e-1420 77

18.2 Attach the load cell

Disconnect the display unit from the power supply. Attach the load cell cables according to the following illustration.
4-wire load cells:
When using 4-wire load cells SEN + and SEN - must not be connected.
6-wire load cells:
78 KXS/KXG-TM-BA_IA-e-1420

18.3 Attach RS232

KXS/KXG-TM-BA_IA-e-1420 79

19 Service menu

ADJ
LOCK
SWA1
Verified weighing systems:
In verified weighing systems the access to the service menu is locked.
In order to unlock the access, the sealing must be destroyed (see chapter 6.9)
and the housing opened. On the plate move the adjusting switch SWA1 to position „ADJ”. Switch on equipment, is indicated.
After the configuration, the access to the service menu must be locked again. Also move the adjusting switch to the position „LOCK“. At the same time it leaves the service mode, the appliance returns to the weighing mode.
Attention: After destruction of the sealing the weighing system must be re-verified by an authorised agency and a new sealing fitted before it can be reused for applications subject t o v er i ficati on.
Unverified weighing systems
With unverified weighing systems the housing does not have to be opened,
the access to the service menu is protected only by a password ("002011“, not changeable).
Use the on/off key to switch on an appliance with pressed UNIT key.
Press TARE button, „--------„ is displayed.
With the navigation keys (see chapter 2.1.1) Enter password "002011".
At incorrect input the scale completes a self check and returns back into the weighing mode.
Confirm with T ARE key. The first menu block is displayed. With the navigation keys (see chapter 2.1.1) select desired menu block.
To exit the menu / return to the weighing mode, switch device off
and on again with the ON/ OFF key.
80 KXS/KXG-TM-BA_IA-e-1420
The service menu consists of the following blocks:
Configure the weighing system, see chapter 18.1 Adjustment of verified weighing systems, see chapter 18.3
Navigation:
For linearisation see chapter 18.2 Appliance settings of verified weighing systems, see chapter 18.5
Alter the gravitation zone, see chapter 18.4
KXS/KXG-TM-BA_IA-e-1420 81

19.1 Configure weighing system / menu block

Press TARE key
Call up the menu block , see chapter 18.
Call up the navigation in the menu/ function:
Select function using navigation keys, (CSP 00 CSP 03)
Back to menu: Select function "CSP 00" and confirm with the TARE key.
Overview:
CSP 00 Exit menu CSP 01 Unverified weighing systems, see chapter 18.1.1 CSP 02 For verified weighing systems see chapter 18.1.2 CSP 03 Type of scale, see chapter 18.1.3
82 KXS/KXG-TM-BA_IA-e-1420

19.1.1 Unverified weighing systems (function „CSP 01”)

(a)
Standard weighing unit, only "kg „, "g „or "lb „can be selected (parameters 0, 1 or 2)
(b)
Second weighing unit, in which can be toggled with the UNIT key.
(c)
Third weighing unit, in which can be toggled with the UNIT key
(d)
Forth weighing unit, in which can be toggled with the UNIT key
(e)
Fifth weighing unit, in which can be toggled with the UNIT key
(f)
Number of weighing units in which can be toggled with the UNIT key
Parameters
1 2 3 4 5 6 7
8
Unit
kg g lb
oz
GN
dwt
ct
Symbol
kg g lb
7
6
6
6
Indicators , see ch. 2.3
Parameter:
(0)
kg
(5)
Oz
(1)
g
(6)
GN
(2)
lb
(7)
dwt
(4)
Not documented
(k)
ct
Menu setting „unverified” CFN 02 = 0, see chapters. 18.5.2.
Call up function "CSP 01", see chapter 18.1 "Navigation in the menu“.
Press the TARE key, the current set capacity / resolution is displayed for 2
seconds (display sample Max 15 kg / d 0.01 kg), followed by the display of "u00000" to adjust the w eig hi ng units.
Settings of weighing units:
Description
Not
documented
KXS/KXG-TM-BA_IA-e-1420 83
Use the navigation buttons to select desired settings
Max.
Input
15.000 kg
015000
1500.0 g
015000
6.000 lb
006000
Example 020002:
Standard weighing unit "kg (a), second weighing unit "lb (b), number of weighing units "2 (f)
Confirm with TARE key, the display "C00000 appears on the input capacity
(max).
Example 015000 ⇒ 15.000kg
With the navigation keys (see chapter 2.1.1) enter the desired capacity.
Examples:
Press the TARE key repeatedly until the display "d00000" appears for the input of
"resolution (d), decimal places and type of scale".
With the navigation keys (see chapter 2.1.1) select desired settings.
84 KXS/KXG-TM-BA_IA-e-1420
Settings:
Adjustment
3 15.000 kg
1 1500.0 g
3 6.000 lb
Multi-interval balance
Multi-range balance
Two sections
Multi-interval balance
Multi-range balance
Three sections
(q)
Not documented
Description
(m) ⇒
(n) ⇒
Readability (d), select from 1, 2, or 5 steps Number of decimal places, select from 0 ~ 5 Examples:
(o) ⇒
(p) ⇒
0 , 1 2 3 0 1
Single-range balance
Divide into maximum of two sections (subranges ½ max) Divide into maximum of three sections (subranges 1/6 and 2/3 max)
Multi-interval balance Multi-range balance
(0) ⇒
Factory setting
KXS/KXG-TM-BA_IA-e-1420 85
Example 132000:
1 ⇒ Readability 1 step 3 ⇒ 3 decimal places 2 ⇒ 0 ⇒ 0 ⇒
Divide into maximum of two sections Multi-interval balance Division multiplication factor
Press the TARE key, the newly set capacity / resolution is displayed for 2 seconds
(display sample maximum 15 kg / d 0.001 kg). The display automatically returns back to the menu.
With the navigation keys (see chapter 2.1.1) select "CSP 00".
Confirm with T ARE key.
To exit the menu / return to the weighing mode, switch off and on again with the
ON/ OFF key.
86 KXS/KXG-TM-BA_IA-e-1420

19.1.2 Verified balancing systems (function "CSP 02)

Parameter:
For menu settings "verified" CFN 02 = 1, see chapter. 18.5.2.
Call up function "CSP 02", see chapter 18.1 "Navigation in the menu.
Press the TARE key, the current set capacity / resolution is displayed for 2
seconds (display sample Max 15 kg / d 0.01 kg), followed by the display of "unit" of the current set weighing units.
Settings of weighing units:
Description
(a) ⇒ (b) ⇒
Number of weighing units in which can be toggled with the UNIT key Standard weighing unit, only „kg, „g can be selected (parameters 0 or 1)
(0) ⇒
kg
(1) ⇒
g
KXS/KXG-TM-BA_IA-e-1420 87
Use the navigation buttons to select desired settings
Example unit 20:
Number of balancing units "2 (a), standard weighing unit „kg(b),
Confirm with TARE key, the current standard weighing unit is indicated for 2
seconds (e.g. „kg) followed by the display „000001 for adjustment to the capacity (max.).
Settings maximum:
Description
(c) ⇒
(d-h) ⇒
(i-n) ⇒
Standard weighing unit Capacity (max) Weight value of adjusting weight
With the navigation keys (see chapter 2.1.1) enter the desired capacity.
Example 015000 ⇒ 15.000kg
Confirm with TARE key, the display "000002" appears for the input of the weight
value for adjusting weight.
Enter the desired weight value with the navigation keys.
88 KXS/KXG-TM-BA_IA-e-1420
0
0
Standard weighing unit Always use setting of "0.
Example 015000 ⇒ 15.000kg
Confirm with the TARE key, the display "000003" appears for the input of
"resolution (d), decimal places and type of scale".
Settings:
Description
(o) ⇒ (p) ⇒
(q) ⇒
(r) ⇒
(s) ⇒
Readability (d), select from 1, 2, or 5 steps Number of decimal places, select from 0 ~ 5
1 ⇒ 2 ⇒ 3 ⇒ 4 ⇒ 5 ⇒
0
0 , 1
2 3
1
0.0
0.00
0.000
0.0000
0.00000
Single-range balance
Divide into maximum of two sections (subranges ½ max) Divide into maximum of three sections (subranges 1/6 and 2/3 max)
Order of weighing units in which can be toggled using the UNIT key. Always use setting of "1.
(t) ⇒
Enter the desired parameters with the navigation keys.
KXS/KXG-TM-BA_IA-e-1420 89
0 1
Multi-in terval balance
Multi-range balance
Example 530010:
5 ⇒
Readability of 5 steps
3 ⇒ 0 ⇒ 0 ⇒ 1 ⇒ 0 ⇒
3 decimal places d = 0.005 Standard weighing unit „kg Single-range balance Sequence of the weighing units Multi-interval balance
Press the TARE key, the newly set capacity / resolution is displayed for 2 seconds
(display sample maximum 15 kg / d 0.005kg). The display automatically returns back to the menu.
Select „CSP 00“ with the navigation keys.
Confirm with T ARE key.
Exit menu / jumping back to weighing mode:
Move the adjusting switch to the position „LOCK“, see chapter 18.
After the configuration linearize and/or adjust the weighing system.
90 KXS/KXG-TM-BA_IA-e-1420

19.1.3 Set the buffeting point (function „CSP 03)

Call up function „CSP 03” to amend the buffeting point.
With the functions „CSP 01” and „CSP 02” the weighing range, by default, can be divided into:
Single-range balance Divid e into maximum of two sections (subranges ½ max) Divid e into maximum of three sections (subranges 1/6 and 2/3 max)
Call up function „CSP 03“, see chapter 18.1 "Navigation in the menu“.
Press the TARE key, the display shows "unit 0" to enter the weighing unit.
With the navigation keys (see chapter 2.1.1) enter desired unit, see chapter
18.1.1 and/or 18.1.2.
Confirm with the TARE key, the current set buffeting point is displayed.
With the navigation keys (see chapter 2.1.1) select the desired setting.
KXS/KXG-TM-BA_IA-e-1420 91
Confirm with the TARE key, the current set second buffeting point (only during
allocation into 3 sections) is indicated.
Use the navigation keys to select the desired settings
Confirm with the TARE key, the display returns back to the menu.
Select „CSP 00“ with the navigation keys.
Confirm with T ARE key.
To exit the menu / return back to the weighing mode, see chapter 18.1.1 and/or
18.1.2.
92 KXS/KXG-TM-BA_IA-e-1420

19.2 Linearization/ menu block

Carrying out linearization is restricted to specialist staff possessing
see chapter 3.4 “Monitoring of test resources”.
Linearity shows the greatest deviation of a weight display on the scale to the value of the respective test weight according to plus and minus over the entire weighing range. If linearity deviation is discovered during a monitoring of test resources, you can improve this by means of linearization.
Procedure:
1. Call up function , see chapter 18.
2. Press the TARE key.
Ensure that there are no objects on the weighing pan.
well acquainted with the workings of weighing scales.
The test weights to be used must be adapted to the weighing scale’s specifications; see chapter “monitoring of test resources”.
Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
After successful linearisation you will have to carry out calibration;
3. Press the TARE key.
4. With display "L1 1" set the „first adjusting weight carefully into the centre of the weighing plate". With the navigation keys (see chapter 2.1.1) Enter the desired factor, by which the following adjustment weight is to increase. Press the TARE key.
5. With display "L2 1" set the second adjustment weight carefully into the centre of the weighing plate. With the navigation keys (see chapter 2.1.1) Enter the desired factor, by which the following adjustment weight is to increase. Press the TARE key.
KXS/KXG-TM-BA_IA-e-1420 93
6. When „L3 1“ is displayed, place the third adjustment weight carefully in the centre of the weighing plate. Enter the desired factor, by which the following adjusting weight is to increase, with the navigation keys. Press the TARE key.
7. Repeat step 6 depending on the number of desired linearization points. The number of desired points of linearization is completed by pressing the BG/ NET key.
Confirm with the TARE key, the display returns back to the menu.
To exit the menu / return back to the weighing mode, see chapter 18.1.1 and/or
18.1.2.
With the L0, L1 display, the linearization can be cancelled with the BG/ NET key.
With the L2, L3, L4, L5, L6, L7 or L8 display, the linearization can be cancelled by pressing the BG/ NET key (e.g. L3 corresponds to a 3-point linearization).
94 KXS/KXG-TM-BA_IA-e-1420
Setting "factor”
1 =
All weight values are equal
2 =
The following weight value is twice as high as the first.
3 =
The following weight value is three times higher than the first
4 =
….5 times…
5 =
….6 times…
6 =
….7 times…
7 =
….8 times…
8 =
….9 times…
A =
….10 times…
B =
….11 times…
C =
….12 times…
D =
….13 times…
E =
….14 times…
F =
….15 times…
Display
Button
Explication
Examples:
Example 1: 4-point linearization/ 30 kg (10 kg, 10 kg, 10 kg)
Display Button Explication
03 CLn TARE Linearization function L0 TARE Zero point L11 TARE First adjustment weight (10 kg) L21 TARE Second adjustment weight (10 kg) L31 TARE Third adjustment weight (10 kg) L41 BG/NEt Terminate linearization (4-point linearization) 03 CLn
Example 2: 5-point linearization/ 30 kg (5 kg, 10 kg, 10 kg, 5kg)
03 CLn TARE Linearization function L0 TARE Zero point L11 TARE First adjustment weight (5 kg) L22 TARE Second adjustment weight (10 kg); factor 2, i.e. twice as high as the first L32 TARE Third adjustment weight (10 kg); factor 2, i.e. twice as high as the first L41 Fourth adjustment weight (5 kg), factor 1, i.e., the same size as the first L51 BG/NEt Terminate linearization (5-point linearization) 03 CLn
KXS/KXG-TM-BA_IA-e-1420 95
Example 4: 5-point linearization/ 30 kg (5kg, 5kg, 10 kg, 10kg)
Display
Button
Explication
Display
Button
Explication
03 CLn TARE Linearization function L0 TARE zero point L11 TARE First adjustment weight (5 kg) L21 TARE Second adjustment weight (5kg), factor 1, i.e., the same size as the first L32 TARE Third adjustment weight (10 kg); factor 2, i.e. twice as high as the first L42 Fourth adjustment weight (10 kg); factor 2, i.e. twice as high as the first L51 BG/NEt Terminate linearization (5-point linearization) 03 CLn
Example 4: 7-Point linearization / 30 kg (1 kg, 2 kg, 5 kg, 10 kg, 2 kg, 10 kg)
03 CLn TARE Linearization function L0 TARE zero point L11 TARE First adjustment weight (1 kg) L22 TARE Second adjustment weight (2 kg); factor 2, i.e. twice as high as the first L3 5 TARE Third adjustment weight (5 kg); Factor 5 i.e. 5 times as high as the first L4 A TARE Forth adjustment weight (10 kg); Factor A i.e. 10 times as high as the first L52 TARE Fifth adjustment weight (5kg); factor 2, i.e. twice as high as the first L6 A TARE Sixth adjustment weight (10 kg); Factor A i.e. 10 times as high as the first L71 BG/NEt Terminate linearization (7-point linearization) 03 CLn
96 KXS/KXG-TM-BA_IA-e-1420

19.3 Adjustment of "verified weighing systems / menu block "

warm-up time.
Press TARE key
Call up the navigation in the menu/ function:
Call up function , see chapter 18.
Using navigation keys select function, (CAL 00 CAL 03)
Back to menu: Select function "CAL 00" and confirm with the TARE key.
Overview:
Prepare the required adjustment weight. The weight to be used depends on the capacity of the scale. Carry out adjustment as near as possible to the scale’s maximum weight. Info about test weights can be found on the Internet at: http://www.kern-sohn.com.
Observe stable environmental conditions. Stabilisation requires a certain
CAL 00 Exit menu CAL 01 Adjustment with external adjustment weight CAL 02 Not documented (adjustment with gravitational constant)
KXS/KXG-TM-BA_IA-e-1420 97
Perform external adjustment / function "CAL 01":
Call up function "CAL 01”
Press the TARE button, the currently set adjustment weight is displayed. The last
digit flashes.
To change the adjustment weight select the digit to be changed with BG/ NET key, the active digit flashes. Enter the desired value with the navigation keys. Press the TARE key repeatedly, until "-----“ is indicated.
Ensure that there are no objects on the weighing pan. Press the TARE key.
The weight value of the required adjustment weight will be displayed.
98 KXS/KXG-TM-BA_IA-e-1420
Carefully place the adjusting weight into the centre of the balancing plate and
press the TARE key. After successful adjustment, an acoustic signal sounds.
Remove the adjustment weight. Press any key, the scale will return to the menu..
With the navigation keys (see chapter 2.1.1) select "CSP 00".
Confirm with T ARE key.
To exit the menu, move the adjustment switch into the „LOCK” position, see
chapter 18.
KXS/KXG-TM-BA_IA-e-1420 99

19.4 Amend the local gravitation zone/ menu block

Press TARE key
Call up function , see chapter 18.
The current setting will be displayed.
Using the navigation buttons, enter the local gravitational zone.
Confirm with the TARE key, the display returns back to the menu.
To exit the menu, move the adjustment switch into the „LOCK position, see
chapter 18.
100 KXS/KXG-TM-BA_IA-e-1420