KERN & SOHN KXE-TM User Manual

KERN & Sohn GmbH
Ziegelei 1
E-Mail: info@kern-sohn.com
Internet: www.kern-sohn.com
Operating and Installati on Instructions
Display Unit
KERN KXE-TM
Version 1.1
06/2014
GB
D-72336 Balingen
Fax: +49-[0]7433-9933-149
KXE-TM-BA_IA-e-1411
KERN KXE-TM
Display device
GB
Version 1.1 06/2014
Operating and inst a llation instr uc t ions
Contents
1 Technical data ................................................................................................... 4
2 Appliance overview ........................................................................................... 6
2.1 Keyboard overview .................................................................................................... 7
2.1.1 Numerical input via the navigation buttons ........................................................................ 8
2.2 Function keys ............................................................................................................ 8
2.3 Overview of display .................................................................................................... 9
3 Basic Information (General) ........................................................................... 10
3.1 Proper use ................................................................................................................10
3.2 Improper Use ............................................................................................................10
3.3 Warranty ...................................................................................................................10
3.4 Monitoring of Test Resources ...................................................................................11
4 Basic Safety Precautions ................................................................................ 11
4.1 Pay attention to the instructions in the Operat ion Manual .........................................11
4.2 Personnel training .....................................................................................................11
5 Transport and storage .................................................................................... 11
5.1 Testing upon acceptance ..........................................................................................11
5.2 Packaging / return t ransport .....................................................................................11
6 Unpacking and placing ................................................................................... 12
6.1 Installation Site, Location of Use ...............................................................................12
6.2 Unpacking and placing .............................................................................................12
6.3 Scope of delivery / serial accessories: ......................................................................12
6.4 Transport Securing ...................................................................................................13
6.5 Mains connection ......................................................................................................13
6.6 Battery operation ......................................................................................................13
6.7 Adjustment................................................................................................................14
6.8 Verification ................................................................................................................14
6.8.1 Position of seals and adjusting switch ............................................................................. 15
6.8.2 Software version check .................................................................................................... 15
2 KXE-TM-BA_IA-e-1411
7 Basic Operation ............................................................................................... 16
7.1 Start-up .....................................................................................................................16
7.2 Switching Off ............................................................................................................16
7.3 Zeroing .....................................................................................................................16
7.4 Simple weighing ........................................................................................................16
7.5 Switch-over weighing unit .........................................................................................17
7.6 Weighing with tar ing .................................................................................................17
7.6.1 Determination of the tare weight by weighing .................................................................. 17
7.6.2 Pre-tare deduction (PRE-TARE) ...................................................................................... 18
8 Totalization ...................................................................................................... 19
9 Function menu ................................................................................................. 22
9.1 Overview ...................................................................................................................24
10 Description of individual functions ............................................................ 27
10.1 Resetting the menu...................................................................................................27
10.2 Display higher r es olution ...........................................................................................28
10.3 Peak value function ..................................................................................................29
11 Service, maintenance, disposal .................................................................. 30
11.1 Clean ........................................................................................................................30
11.2 Ser vice, m aint enance ...............................................................................................30
11.3 Disposal ....................................................................................................................30
11.4 Er r or messages ........................................................................................................31
12 Instant help ................................................................................................... 32
13 Installation display unit / platform .............................................................. 33
13.1 Weighing system design ...........................................................................................33
13.2 How to connect t he plat form .....................................................................................
34
14 Technician menu ......................................................................................... 35
14.1 Configure / Menu item
14.2 Adjust / Menu item
14.3 Ent er password / Menu item
14.3.1 Deposit password ........................................................................................................ 41
14.3.2 Password input for access to menu ............................................................................ 42
........................................................................36
.................................................................................39
..................................................................41
14.4 Change appliance settings / menu item ................................................43
15 Enclosure ..................................................................................................... 45
15.1 Test certificate ..........................................................................................................45
KXE-TM-BA_IA-e-1411 3

1 Technical data

KERN KXE-TM
Display 6-digit
Single (Max.) 5.000 e
Solution verifiable
Dual (Max.) 10.000 e Verification class III Weighing ranges max 2 Weigh ing Units kg, t Divisions 1,2,5,…10, n Display LCD 25 mm digits with back lighting
DMS weighing cells
Input voltage 230 V, 50 / 60 Hz
Electric Supply
Mains adapter secondary voltage 9V, 1000 mA
Batteries
Admissible ambient temperature
Humidity of air < 85 % relative (not condensing) Net weight 1 kg
Min. 87 .
Max. 1000
5 x 1.5 V, AA
Service life 24 h
0°C – 40°C
4 KXE-TM-BA_IA-e-1411
Dimensions:
1. With wall support
2. Without wall support
KXE-TM-BA_IA-e-1411 5

2 Appliance overview

3 4 5 6
2
1 7 8
1. Weight display
2. Battery compartment
3. RS-232 (disabled)
4. Connection power supply (mains adapter)
5. Adjustment switch
6. Connection platform
7. Fastening screws
8. Wall bracket
6 KXE-TM-BA_IA-e-1411

2.1 Keyboard overview

Key Function
Turn on/off
Quit
Back to menu / weighing mode
Zeroing
Taring
Change between gross and net weight
Functional button (allocation see chap. 9.1 FNC-02)
Functional button (allocation see chap. 9.1 FNC-03)
Confirm/store input
KXE-TM-BA_IA-e-1411 7

2.1.1 Numerical input via the navigation buttons

Key Function
1. Scroll forward in menu
Increase flashing digit
Scroll backwards in menu
Decrease flashing digit
Digit selection to the left
Digit selection to the right
Save setting
Reject setting

2.2 Function keys

The following functions can be allocated to the buttons , or + in menu (see chap. 9):
FNC-02 FNC04
0 Printing 1 Switch-over weighing unit 2 M+ (adding up / printing) 3 MC (delete sum memory) 4 Display change 5 HR (conversion into higher resolution) 6 PRE-TARE
8 KXE-TM-BA_IA-e-1411

2.3 Overview of display

Indicator [] next to symbol displays:
PRE-TARE
TARE
M+
GROSS
that a tare value is stored Instable weighing value Data in summation memory that the displayed weighing value is a gross weight value that the displayed weight value is a PRE-TARE value Resolution 1. weighing range Resolution 2. weighing range
KXE-TM-BA_IA-e-1411 9

3 Basic Information (General)

3.1 Proper use

The display unit acquired by you is used in combination with a platform and serves to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic weighing system”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

3.2 Improper Use

Do not use display unit for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the display unit. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the weighing plate, minus a possibly existing tare load, must be strictly avoided. Both, the weighing plate and the display unit may be damaged during this process. Never operate display unit in explosive environment. The serial version is not explosion protected. Changes to the display unit's design are not permitted. This may lead to incorrect weighing results, safety-related faults and destruction of the display unit. The display unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing.

3.3 Warranty

Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
10 KXE-TM-BA_IA-e-1411

3.4 Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the display unit and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of display units’ test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and display units may be calibrated (return to the national standard) fast and at moderate cost.

4 Basic Safety Precautions

4.1 Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.

4.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

5 Transport and stora ge

5.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.

5.2 Packaging / return transport

Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning. Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices. Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
KXE-TM-BA_IA-e-1411 11

6 Unpacking and placing

For display unit, see chapter 2

6.1 Installation Site, Location of Use

The display units are designed in a way that reliable weighing results are achieved in common conditions of use. Precise and fast work is achieved by selecting the right place for your display unit and your weighing plate. On the installation site observe the following:
Place the display unit and the weighing plate on a stable, even surface.
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the display unit and the weighing plate against direct draft from open windows or doors.
Avoid jarring during weighing;
Protect the display unit and the weighing plate against high humidity, vapours
and dust.
Do not expose the display unit to extreme dampness for longer periods of time. Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.

6.2 Unpacking and placing

Take the display unit carefully out of its packaging, remove the plastic jacket and install it at the designated work space. Mount the display unit in a way that facilitates operation and where it is easy to see.

6.3 Scope of delivery / serial accessories:

Mains adapter
Plug for connecting platform
Operating instructions
12 KXE-TM-BA_IA-e-1411

6.4 Transport Securing

Transport Securing
Please note, if the display unit is used together with platform with transportation lock, this transportation lock must be released prior to use. See operating instructions attached to the respective platform.
Example of application

6.5 Mains connection

Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.

6.6 Battery operation

On the rear side of the display unit remove the battery cover and insert
5 x 1.5 V, AA batteries.
Ensure correct polarisation.
Close again the lid.
In order to ensure a long lifetime of the batteries, we recommend to enable the automatic switch-off function „FNC-08“, see chapter 9.1
KXE-TM-BA_IA-e-1411 13

6.7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each display unit with connected weighing plate must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the weighing system has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the display unit periodically in weighing operation.
+

6.8 Verification

General introduction:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
c) For official purposes d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard. Verification notes:
An EU type approval exists for the appliance described in its technical data as verifiable. If the appliance is used where obligation to verify exists as described above, it must be verified and re-verified at regular intervals. Re-verification of an appliance is carried out according to the respective national regulations. Normally the validity for verification in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
Carry out adjustment, see chapter 14.2
medical and pharmaceutical laboratory.
Verification of the device is invalid without the seals.
The seal marks attached on verified appliances point out that the appliance may only be opened and serviced by trained and authorised specialist staff. If the seal mark is destroyed, verification looses its validity. Please observe all national laws and legal regulations. In Germany a re-verification will be necessary.
14 KXE-TM-BA_IA-e-1411

6.8.1 Position of seals and adjusting switch

unlocked
locked

6.8.2 Software version check

Switch on the device by pressing and then press during the execution of the self-test procedure.
When “ver” is shown on the screen, press to display the software version specified in the type approval.
To return to the weighing mode, press several times.
KXE-TM-BA_IA-e-1411 15

7 Basic Operation

7.1 Start-up

Press and the instrument will carry out a self-test. As soon as the weight display appears, the instrument will be ready to weigh.

7.2 Switching Off

Press for 3 sec, the display disappears.

7.3 Zeroing

Resetting to zero corrects the influence of light soiling on the weighing plate.
To unload the weighing system
Press and zero display as well as indicator will appear.

7.4 Simple weighing

Place goods to be weighed on balance. Wait until the indicator over the stability display disappears. Read weighing result.
16 KXE-TM-BA_IA-e-1411

7.5 Switch-over w eighing unit

The following setting must be allocated to a functional button (see chap. 9.).
Functional
Setting
button
+
FNC-02 „1“
or
FNC-03 „1“
or
FNC-04 „1“
When pressing the functional button, the weight display changes between „kg“ and „t“.

7.6 Weighing with taring

7.6.1 Determination of the tare weight by weighing

Deposit weighing vessel. Wait for stability, then press . The zero display, the indicator above TARE and NE T will appear.
The weight of the container is now internally saved. Weigh the material, the net weight will be indicated. After removing the weighing container, the weight of the weighing container appears
as negative display. The taring process can be repeated any number of times, e.g. when adding several
components for a mixture (adding). The limit is reached when the whole weighing range is exhausted.
To change between gross weight and net weight, press .
To delete the tare value, remove load from weighing plate and press .
KXE-TM-BA_IA-e-1411 17

7.6.2 Pre-tare deduction (PRE-TARE)

or
or
The known dead weight of a weighing container can be tared off by entering its weight as pre-tare deduction in order to ensure the net weight of the goods to be weighed in subsequent weighings is always displayed. Condition: The setting for PRE-Tare must be allocated to a functional button (see chap. 9.).
Functional
Setting
button
+
FNC-02 „6“
FNC-03 „6“
FNC-04 „6
Press the functional button, „PT“ followed by „0000.00“ will be displayed.
The active digit is flashing.
Enter the known tare weight e.g. 2 kg and confirm by . Numerical input, see
chap. 2.1.1.
The entered weight will be stored as tare weight and displayed with negative sign. The indicator over PRE-TARE will appear.
Put the filled weighing container on the balance, the net weight will be displayed.
The tare value remains stored until it will be deleted by .
18 KXE-TM-BA_IA-e-1411

8 Totalization

Settings of functional menu:
In the following application example of application the following settings are allocated to the functional buttons (see chap. 9.).
Functional button
+
Setting Description
FNC-02 „4“
FNC-03 „2“
FNC-04 „3“ MC Delete total added memory
Display of all add-up data
Number weighing processes Total Weighing mode
M+: Totalization (commutation add-up mode weighing mode)
Of course you may also adapt the allocation to your individual needs. (see chapter. 9)
KXE-TM-BA_IA-e-1411 19
Add up:
1. Place weighing good A, e.g. 5 kg. Wait for stability, then press . The weight value is added-up into a sum
memory , the indicator  over M+ appears.
2. Place goods to be weighed, e.g. 7 kg. Wait for stability, then press . The weighing value is added into the
summation memory.
Add more weighed goods as described before. You can repeat this process until the capacity of the weighing system is
exhausted.
Display / switch over weighing data:
Press repeatedly to switch to a different display.
Number of weighings Indicator „PCS“ will be displayed.
Total sum
Indicator  over will be displayed
20 KXE-TM-BA_IA-e-1411
Currently placed weight / add-up mode Indicator  over M+ is displayed
Delete total added memory:
 With display „Number of weighings“ or „Total weight“ press + at the
same time. The data in the summation memory are deleted.
+
Weighing mode
Indicator  over M+ disappears
KXE-TM-BA_IA-e-1411 21

9 Function menu

Call up menu:
In weighing mode press and at the same time, the functional menu
„01 FnC“ is displayed.
Switch over between and (not documented) using and .
22 KXE-TM-BA_IA-e-1411
Navigation in the menu:
Invoke function menu , see chap. 9 and
acknowledge with .
Select the individual functions subsequently using ,
FNC-00 – FNC-08 (see chap. 9.1).
Confirm selected function by , the current setting is
displayed.
Switch into the available settings using and .
Confirm selection by , the next function is
displayed.
To exit the menu / return into weighing mode press
repeatedly.
Numerical input, see chap. 2.1.1.
KXE-TM-BA_IA-e-1411 23

9.1 Overview

0 0 0
1 1 1
Factory settings are marked by *.
No. Function
Available settings Parameters Description
FNC-00 Keyboard lock
FNC-01 Update of display
FNC-02 Allocation
Functional button
0
0000*
0 no
1
1111
0 none
1* 20 times / s
2 10 times / s 3 5 times / s 4 1 times / s 0 Print 1 Switch-over weighing unit
1 yes
2 M+ (adding up / printing) 3 MC (delete sum memory)
Display switch-over in add-up mode
Number weighing
4
processes Total Weighing mode
5*
6 PRE-TARE
HR (conversion into higher resolution)
24 KXE-TM-BA_IA-e-1411
Automatic background illumination on pressed.
FNC-03 Allocation functional
buttons
FNC-04 Allocation functional
+
buttons
0* Print
1 Switch-over weighing unit 2 M+ (adding up / printing) 3 MC (delete sum memory)
Display switch-over in add-up mode
Number weighing processes
4
Total Weighing mode
5
6 PRE-TARE
0* Print
1 Switch-over weighing unit
HR (conversion into higher resolution)
FNC-05 Display background
illumination
2 M+ (adding up / printing) 3 MC (delete sum memory)
Display switch-over in add-up mode
4
5
6 PRE-TARE
0
1*
2
Number weighing processes Total Weighing m ode
HR (conversion into higher resolution)
when weighing pate is loaded or key
Continuous background lighting
Background illumination off
KXE-TM-BA_IA-e-1411 25
FNC-06 Tone
FNC-07
Peak value function
FNC-08
Auto-power Off (Automatic power
switch-off)
FNC-09 Not documented
0 Signal tone switched off
1* Signal tone switched on
0* switched off
1 Mode 1, see chap. 11.3 2 Mode 2, see chap. 11.3
The display unit is automatically switched off within the preset time
1 – 60 min
when the display unit or the weighing bridge are not operated. Factory setting „0“
26 KXE-TM-BA_IA-e-1411

10 De s c ription of individua l functions

10.1 Resetting the menu

Reset all parameters to factory default:
to switch on the appliance. During the selftest actuate the adjustment switch and press + at the same time. „IntALL“ is displayed.
Press and keep it pressed until „End“ appears. Back to weighing mode, actuate adjustment switch.
Reset all parameters to allowed settings for verified weighing systems :
to switch on the unit. During the selftest actuate the adjustment switch and press
+ at the same time. „IntALL“ is displayed.
Press and keep it pressed until „End“ appears. Back to weighing mode, actuate adjustment switch.
KXE-TM-BA_IA-e-1411 27

10.2 Display higher resoluti on

Condition:
The following setting must be allocated to a functional button (see chap. 9.).
Functional
Setting
button
+
FNC-02 „5“
or
FNC-03 „5“
or
FNC-04 „5“
When pressing the functional button, the weight value is displayed
approx. 5 seconds in a 10 times higher resolution.
28 KXE-TM-BA_IA-e-1411

10.3 Peak value function

This function displays the highest load value (peak value) of a weighing. The peak value remains in the display until it will be deleted.
Mode 1: Menu setting FNC-07 „1“ From a weighing value of >10 d the highest load value will be displayed.
The peak value remains in the display until it will be deleted using . Then the balance is ready for further measurements.
Display
Mode 2: Menu setting FNC-07 „2“ From a weighing value of >10 d the highest load value will be displayed. If the weight
value is less than 10d, the displayed peak value will be automatically deleted. Display
KXE-TM-BA_IA-e-1411 29

11 S er vice, mai ntenance, disposal

Before appliance from the operating voltage.
any maintenance, cleaning and repair work disconnect the

11.1 Clean

Keep IP protection. Clean the stainless-steel parts with a soft cloth soaked with a cleaning agent
suitable for stainless steel.
For stainless steel parts do not use any cleaning agents which contain sodium
hydroxide solution, acetic, hydrochloric, sulphuric or citric acid.
Do not use metal brushes or cleaning sponges of steel wool, as this causes
superficial corrosion.

11.2 Service, maintenance

The appliance may only be opened by trained service technicians who are
authorized by KERN.
Ensure that the balance is regularly calibrated, see chap. Testing instruments
control.

11.3 Disposal

Disposal of packaging and appliance must be carried out by operator according
to valid national or regional law of the location where the appliance is used.
30 KXE-TM-BA_IA-e-1411

11.4 Error messages

Error
message
Err 0 Value outside the A/D changer range Err 2
Err 6 Err . Wrong password
E 1 Internal value < zero range E 2 Internal value > zero range
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
Weighing value Zero range Internal solution 0.15 µV/d range
Description
KXE-TM-BA_IA-e-1411 31

12 Instant help

The displayed weight does not glow.
(Rechargeable) batteries are inserted incorrectly or
The displayed weight is
The weighing result is obviously incorrect
In case of an error in the program process, briefly turn off the display unit and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The display unit is not switched on.
Mains power supply interrupted (mains cable
defective).
Power supply interrupted.
empty
No (rechargeable) batteries inserted.
permanently changing
Draught/air movement
Table/floor vibrations
Weighing plate has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if possible)
32 KXE-TM-BA_IA-e-1411

13 Installation display unit / platform

Installation / configuration of the weighing system must be carried out by a well acquainted specialist with the workings of weighing balances.

13.1 Weighing system design

The display unit is suitable for connection to any analogue platform in compliance with the required specifications. The following data must be established before selecting a weighing cell:
Weighing balance capacity This usually corresponds to the heaviest load to be weighed.
Preload This corresponds to the total weight of all parts that are to be placed on the weighing cell such as upper part of platform, weighing pan etc.
Total zero setting range This is composed of the start-up zero setting range (± 2%) and the zero setting range available to the user via the ZERO-key (2%). The total zero setting range equals therefore 4 % of the scale’s capacity.
The addition of weighing scales capacity, preload and the total zero setting range give the required capacity for the weighing cell. To avoid overloading of the weighing cell, include an additional safety margin.
Smallest desired display division
KXE-TM-BA_IA-e-1411 33

13.2 How to connect the platform

1
2
3
4
5
6
7
89
Load cell cable
Shield
Load cell
EXC+
SEN+
SIG+ SIG-
EXC-
SEN
-
Disconnect the display unit from the power supply. Connect load cell cable to the connecting plug. Please see diagram below for plug
allocation.
Connect the plug to the display unit.
4-(5-)wire weighing cells 6-(7-)wire weighing cells Pin 1 SIG+ Pin 1 SIG+ Pin 2, 3 EXC- Pin 2 SEN­Pin 4, 5 EXC+ Pin 3 EXC­Pin 6, 7, 8 Screen Pin 4 SEN+ Pin 9 SIG- Pin 5 EXC+ Pin 6, 7, 8 Screen Pin 9 SIG-
34 KXE-TM-BA_IA-e-1411

14 Technician menu

In verified weighing systems the access to the technician menu is
Call up menu:
Actuate adjustment switch, „SET“ followed by „01 CSP“ is displayed.
Use and to switch over between the following menu items:
locked. In order to unlock the access, the seal must be destroyed and the
adjustment switch be actuated (see chap. 6.8).
Attention: After destruction of the seal the weighing system must be re-verified by an authorised agency and a new verification wire/seal mark fitted before it can be reused for applications subject to verification.
Configure
Adjustment
Enter password
Change settings of the unit
To exit the menu / back to weighing mode, actuate adjustment switch.
KXE-TM-BA_IA-e-1411 35

14.1 Configure / Menu i te m

Navigation in the menu:
Invoke configuration menu and confirm with
. The first function is displayed.
Press to select the individual functions
one by one.
Confirm selected function such as „CSP-04“ with ,
the current setting is displayed.
Switch into the available settings using and .
Confirm selection by , the next function is displayed.
To exit the menu / back to weighing mode, actuate
adjustment switch.
Numerical input, see chap. 2.1.1.
36 KXE-TM-BA_IA-e-1411
Configuration menu overview:
Factory settings or allowed settings for verified weighing systems are marked with *.
Available settings
No. Function
CSP-00 Position decimal
point
CSP-01 Balance capacity
[Max]
CSP-02 Readability [d] /
verification value [e]
1. weighing range
Paramete rs
d 0 No decimal place d 0.0 1 decimal place d 0.00* 2 decimal places d 0 3 decimal places d 0.0000 4 decimal places
999999
000000
1 The possible resolutions and 2 5
Description
The possible capacities and factory settings depend on the connected weighing bridge.
factory settings depend on the connected weighing bridge.
CSP-03 Readability [d] /
verification value [e]
2. weighing range
10 20 50
1 The possible resolutions and
factory settings depend on the
2
5 10 20 50
connected weighing bridge.
KXE-TM-BA_IA-e-1411 37
CSP-04 Zero-Tracking
CSP-05 Movement detection
Automatic zero point correction
0.25 d 0. 25 d / 1 s*
0.5 d 0.5 d / 1 s
0.75 d 0. 75 d / 1 s 1 d 1 d / 1 s
1.25 d 1. 25 d / 2 s
1.5 d 1.5 d / 2 s
1.75 d 1. 75 d / 2 s 2 d 2 d / 2 s
no Deactivated
Digits / time
by changing the display, digits / time selectable
0.25 d 0. 25 d / 1 s*
0.5 d 0.5 d / 1 s
0.75 d 0. 75 d / 1 s 1 d 1 d / 1 s
1.25 d 1. 25 d / 2 s
1.5 d 1.5 d / 2 s
1.75 d 1. 75 d / 2 s 2 d 2 d / 2 s
no Deactivated
Adjust weighing system according to configuration.
38 KXE-TM-BA_IA-e-1411

14.2 Adjust / Menu item

The weight to be used depends on the capacity of the scale. Carry out
Actuate adjustment switch, „SET“ followed by „01 CSP“ is displayed.
Invoke menu item using or and confirm by .
The adjustment of the zero point is displayed.
adjustment as near as possible to the scale’s maximum weight. Info about test weights can be found on the Internet at: http://www.kern-sohn.com.
Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
KXE-TM-BA_IA-e-1411 39
Ensure that there are no objects on the weighing plate.
Wait for stability, then press .
Cancel the adjustment with , the appliance returns into the menu.
Having adjusted the zero point, „SPAn“ will be displayed, followed by the
currently set weight value for the adjustment weight.
After approx. 5 seconds the zero point adjustment is completed.
 Either use the displayed adjustment weight or change it with , and
(numerical input see chapter 2.1.1), the active digit is flashing.
Place the indicated adjustment weight and wait for stability.
Acknowledge with .
Cancel the adjustment with , the appliance returns into the menu.
Back to weighing mode, actuate adjustment switch.
After approx. 5 seconds the adjustment is completed.
40 KXE-TM-BA_IA-e-1411

14.3 Enter password / Menu item

The technician and function menu can be secured against non authorised access by a password query.

14.3.1 Deposit password

Actuate adjustment switch, „SET“ followed by „01 CSP“
is displayed.
Invoke menu item using or .
Acknowledge with . The currently set password is
displayed, the active digit flashes. In „0000“ (factory setting) no password is deposited.
Enter the desired password e.g. 1234 (Numeric input see
chap. 2.1.1) and confirm by .
The request to repeat the password input is displayed,
the active digit flashes.
Enter password anew and confirm with .
The appliance returns to the menu
To exit the menu / back to weighing mode, actuate
adjustment switch.
KXE-TM-BA_IA-e-1411 41

14.3.2 Password input for access to menu

Example: Invoke technician menu When invoking a menu, the password inquiry
appears‚ followed by .
Enter the password e.g. 1234 (Numeric input see chap.
2.1.1) and confirm by .
If the password is entered correctly, settings in the menu may be carried out.
If a wrong password is entered, „Err“ is displayed.
Password forgotten?
Reset the display unit to factory setting „0000“ (no password), see chap. 11.1.
42 KXE-TM-BA_IA-e-1411

14.4 Change appliance settings / menu item

Navigation in the menu:
Invoke menu item and confirm with .
The first function is displayed.
Press to select the individual functions
one by one.
Confirm selected function e.g. „CFn-05“ with , the
current setting is displayed.
Switch into the available settings using and .
Confirm selection by , the next function is displayed.
To exit the menu / back to weighing mode, actuate
adjustment switch.
Numerical input, see chap. 2.1.1.
KXE-TM-BA_IA-e-1411 43
Menu overview
Calm and stable environment The
Factory settings or allowed settings for verified weighing systems are marked with *.
Available settings
No. Function
Parameters Description
CFN-00 Taring / zeroing with
instable display
CFN-01 Set to zero during
start-up
CFN-02 Zeroing range
CFN-03 Adaptation of
weighing speed
(factory setting 6)
CFN-04 Adaptation to
environmental conditions
(factory setting 2)
0 On
1* Off
0 On
1* Off
2 %
0
9
0
always use this setting for verified weighing systems.
fast
slow
appliance works fast but is sensitive to outside influences.
5
Busy environment. The appliance works slower, but
is insensitive to outside influences.
44 KXE-TM-BA_IA-e-1411
CFN-05 Animal weighing
mode
CFN-06 Not documented CFN-07 Not documented CFN-08 Balance Type
CFN-09 Setting
multi-range balance (factory setting
5000)
CFN-10 Geo value
0
1 2
10
2
0*
1
0 - 999999
Deactivated, always use this setting.
Not documented Not documented
Always use this setting
Multi-interval balance Multi-range balance
Jump point
Not documented

15 Enclosure

15.1 Test certificate

KXE-TM-BA_IA-e-1411 45