15.1 Test certificate ..........................................................................................................45
KXE-TM-BA_IA-e-1411 3
1 Technical data
KERN KXE-TM
Display 6-digit
Single (Max.) 5.000 e
Solution verifiable
Dual (Max.) 10.000 e
Verification class III
Weighing ranges max 2
Weigh ing Units kg, t
Divisions 1,2,5,…10, n
Display LCD 25 mm digits with back lighting
DMS weighing cells
Input voltage 230 V, 50 / 60 Hz
Electric Supply
Mains adapter secondary voltage 9V, 1000 mA
Batteries
Admissible
ambient temperature
Humidity of air < 85 % relative (not condensing)
Net weight 1 kg
Min. 87 Ω.
Max. 1000 Ω
5 x 1.5 V, AA
Service life 24 h
0°C – 40°C
4 KXE-TM-BA_IA-e-1411
Dimensions:
1. With wall support
2. Without wall support
KXE-TM-BA_IA-e-1411 5
2 Appliance overview
3 4 5 6
2
1 7 8
1. Weight display
2. Battery compartment
3. RS-232 (disabled)
4. Connection power supply (mains adapter)
5. Adjustment switch
6. Connection platform
7. Fastening screws
8. Wall bracket
6 KXE-TM-BA_IA-e-1411
2.1 Keyboard overview
Key Function
• Turn on/off
• Quit
• Back to menu / weighing mode
• Zeroing
• Taring
• Change between gross and net weight
• Functional button (allocation see chap. 9.1 FNC-02)
• Functional button (allocation see chap. 9.1 FNC-03)
• Confirm/store input
KXE-TM-BA_IA-e-1411 7
2.1.1 Numerical input via the navigation buttons
Key Function
1. Scroll forward in menu
• Increase flashing digit
• Scroll backwards in menu
• Decrease flashing digit
Digit selection to the left
Digit selection to the right
Save setting
Reject setting
2.2 Function keys
The following functions can be allocated to the buttons , or + in
menu (see chap. 9):
FNC-02 ∼ FNC04
0 Printing
1 Switch-over weighing unit
2 M+ (adding up / printing)
3 MC (delete sum memory)
4 Display change
5 HR (conversion into higher resolution)
6 PRE-TARE
8 KXE-TM-BA_IA-e-1411
2.3 Overview of display
Indicator [] next to symbol displays:
PRE-TARE
TARE
M+
GROSS
that a tare value is stored
Instable weighing value
Data in summation memory
that the displayed weighing value is a gross weight value
that the displayed weight value is a PRE-TARE value
Resolution 1. weighing range
Resolution 2. weighing range
KXE-TM-BA_IA-e-1411 9
3 Basic Information (General)
3.1 Proper use
The display unit acquired by you is used in combination with a platform and serves to
determine the weighing value of material to be weighed. It is intended to be used as
a “non-automatic weighing system”, i.e. the material to be weighed is manually and
carefully placed in the centre of the weighing plate. As soon as a stable weighing
value is reached the weighing value can be read.
3.2 Improper Use
Do not use display unit for dynamic weighings. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the display unit. (Example: Slowly
draining fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the
measuring system.
Impacts and overloading exceeding the stated maximum load (max) of the weighing
plate, minus a possibly existing tare load, must be strictly avoided. Both, the
weighing plate and the display unit may be damaged during this process.
Never operate display unit in explosive environment. The serial version is not
explosion protected.
Changes to the display unit's design are not permitted. This may lead to incorrect
weighing results, safety-related faults and destruction of the display unit.
The display unit may only be operated in accordance with the described default
settings. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
• Our conditions in the operation manual are ignored
• The appliance is used outside the described uses
• The appliance is modified or opened
• Mechanical damage or damage by media, liquids, natural wear and tear
• The appliance is improperly set up or incorrectly electrically connected
• The measuring system is overloaded
10 KXE-TM-BA_IA-e-1411
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the display
unit and, if applicable, the testing weight, must be checked regularly. The responsible
user must define a suitable interval as well as type and scope of this test. Information
is available on KERN’s home page (www.kern-sohn.comwith regard to the
monitoring of display units’ test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and display units may be
calibrated (return to the national standard) fast and at moderate cost.
4 Basic Safety Precautions
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you
are already familiar with KERN balances.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and stora ge
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
KXE-TM-BA_IA-e-1411 11
6 Unpacking and placing
• For display unit, see chapter 2
6.1 Installation Site, Location of Use
The display units are designed in a way that reliable weighing results are achieved in
common conditions of use.
Precise and fast work is achieved by selecting the right place for your display unit
and your weighing plate.
On the installation site observe the following:
• Place the display unit and the weighing plate on a stable, even surface.
• Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
• Protect the display unit and the weighing plate against direct draft from open
windows or doors.
• Avoid jarring during weighing;
• Protect the display unit and the weighing plate against high humidity, vapours
and dust.
• Do not expose the display unit to extreme dampness for longer periods of
time. Non-permitted condensation (condensation of air humidity on the
appliance) may occur if a cold appliance is taken to a considerably warmer
environment. In this case, acclimatize the disconnected appliance for ca. 2
hours at room temperature.
• Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change location or remove
source of interference.
6.2 Unpacking and placing
Take the display unit carefully out of its packaging, remove the plastic jacket and
install it at the designated work space. Mount the display unit in a way that facilitates
operation and where it is easy to see.
6.3 Scope of delivery / serial accessories:
• Mains adapter
• Plug for connecting platform
• Operating instructions
12 KXE-TM-BA_IA-e-1411
6.4 Transport Securing
Transport Securing
Please note, if the display unit is used together with platform with transportation lock,
this transportation lock must be released prior to use.
See operating instructions attached to the respective platform.
Example of application
6.5 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage.
Only use original KERN mains adapters. Using other makes requires consent by
KERN.
6.6 Battery operation
On the rear side of the display unit remove the battery cover and insert
5 x 1.5 V, AA batteries.
Ensure correct polarisation.
Close again the lid.
In order to ensure a long lifetime of the batteries, we recommend to enable the
automatic switch-off function „FNC-08“, see chapter 9.1
KXE-TM-BA_IA-e-1411 13
6.7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each display unit with connected weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
+
6.8 Verification
General introduction:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially
verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
c) For official purposes
d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard.
Verification notes:
An EU type approval exists for the appliance described in its technical data as
verifiable. If the appliance is used where obligation to verify exists as described
above, it must be verified and re-verified at regular intervals.
Re-verification of an appliance is carried out according to the respective national
regulations. Normally the validity for verification in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
Carry out adjustment, see chapter 14.2
medical and pharmaceutical laboratory.
Verification of the device is invalid without the seals.
The seal marks attached on verified appliances point out that the appliance
may only be opened and serviced by trained and authorised specialist staff. If
the seal mark is destroyed, verification looses its validity. Please observe all
national laws and legal regulations. In Germany a re-verification will be
necessary.
14 KXE-TM-BA_IA-e-1411
6.8.1 Position of seals and adjusting switch
unlocked
locked
6.8.2 Software version check
Switch on the device by pressing and then press during the execution of
the self-test procedure.
When “ver” is shown on the screen, press to display the software version
specified in the type approval.
To return to the weighing mode, press several times.
KXE-TM-BA_IA-e-1411 15
7 Basic Operation
7.1 Start-up
Press and the instrument will carry out a self-test. As soon as the weight display
appears, the instrument will be ready to weigh.
7.2 Switching Off
Press for 3 sec, the display disappears.
7.3 Zeroing
Resetting to zero corrects the influence of light soiling on the weighing plate.
To unload the weighing system
Press and zero display as well as indicator will appear.
7.4 Simple weighing
Place goods to be weighed on balance.
Wait until the indicator over the stability display disappears.
Read weighing result.
16 KXE-TM-BA_IA-e-1411