KERN-SOHN KS DB 1-4, KS DB 20-3 Instructions

DB-BA-e-2020
Sauter GmbH
Ziegelei 1 D-72336 Balingen e-mail: info@kern-sohn.com
Phone : +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.sauter.eu
Instruction manual digital torque tester
SAUTER DB
Version 2.0
03/2020 GB
PROFESSIONAL MEASURING
2 DB-BA-e-2020
GB
SAUTER DB
V. 2.0 03/2020
Instruction manual digital torque tester
Thank you for purchasing the digital torque measuring device from SAUTER. We hope that you will be very satisfied with the high quality of the hardness tester and its extensive functionality. Although the device is a complex and technically advanced measuring instrument, it is designed for many years of trouble-free use, provided that it is used for its intended purpose and properly maintained. Before using the instrument for the first time, please read this operating manual carefully and ensure that it is always kept within easy reach during use. For any questions, requests and suggestions, please contact our customer service department at the telephone number provided.
Table of contents:
1 Before commissioning .................................................................................. 3
2 General operating information ..................................................................... 3
3 Initial commissioning .................................................................................... 3
4 Weight and dimensions ................................................................................. 3
5 Technical data ................................................................................................ 4
6 Operation of the torque measuring instrument DB .................................... 5
6.1 Before commissioning ............................................................................................................. 5
6.2 Switching on the device........................................................................................................... 5
6.3 Basic functions of the device .................................................................................................. 5
6.4 Main menu ................................................................................................................................. 9
6.5 Measurement methods .......................................................................................................... 14
7 Control buttons ............................................................................................ 15
8 Conversion factors ...................................................................................... 16
DB-BA-e-2020 3
1 Before commissioning
Before putting the device into operation, check the delivery for any transport damage to the packaging, the plastic case and the device itself. If this is the case, contact SAUTER immediately.
2 General operating information
Functions of the digital torque measuring instrument that are used most frequently (i.e. display of the torque value, peak hold function, zeroing and changing the displayed unit) can be called up by pressing a corresponding button on the front panel (see paragraph Basic functions of the measuring instrument). The setting of the measuring instrument is done with the menu key (see paragraph Main menu).
3 Initial commissioning
The SAUTER DB digital torque measuring instrument is equipped with a battery set consisting of 4 rechargeable AAA NiMH batteries. To be on the safe side, the rechargeable batteries are completely discharged for transport and separated from the power supply. Insert batteries before using the unit for the first time. To achieve maximum battery life, recharge the batteries for at least 14-16 hours using the mains adapter supplied.
4 Weight and dimensions
Torque
tester
DB 0.5-4
DB 1-4
DB 5-3
DB 10-3
DB 20-3
DB 50-2
Dimension
(LxWxH)
200x100x50mm
Weight
2.5kg
3.5kg
Packaging
Plastic case
Torque tester
DB 100-2
DB 200-2
DB 500-2
Dimension
(LxWxH)
200x100x50
Weight
3.5kg
4kg
Packaging
Plastic case
4 DB-BA-e-2020
5 Technical data
Torque tester
DB 0.5-4
DB 1-3
DB 5-3
DB 10-3
DB 20-3
DB 50-2
Maximum torque
0.5 Nm
15 Nm
5 Nm
10 Nm
20 Nm
50 Nm
Usable measuring
range
5-100% of FS (Full Scale)
Accuracy
(Composite error)
±0.5% of full scale range
Creep
±0.002 of full scale range
Non-linearity
±0.002 of full scale range
Temperature
fluctuation at zero
load
±0.002 of full scale range/°C
Operating
temperature
20±10°C
Storage and
transport
temperature
-15°C up to 65°C
Relative air humidity
Not more than 70
Torque tester
DB 50-2
DB 100-2
DB 200-2
DB 500-2
Maximum torque
50 Nm
100 Nm
200 Nm
500 Nm
Usable measuring
range
5-100% of FS (Full Scale)
Accuracy
(Composite error)
±0.5% of full scale range
Creep
±0.002 of full scale range
Non-linearity
±0.002 of full scale range
Temperature
fluctuation at zero
load
±0.002 of full scale range/°C
Operating
temperature
20±10°C
Storage and
transport
temperature
-15°C up to 65°C
Relative air
humidity
Not more than 70
DB-BA-e-2020 5
6 Operating the torque measuring instrument DB
6.1 Before commissioning
It must be ensured that the measuring plate is neither loaded nor exposed to any other force when the device is switched on.
6.2 Switching on the device
There are 8 keys arranged on the control panel.
To switch on the device, press the ON/OFF button. A brief autotest will then be performed, during which the display will show the permissible load value in Newtons.
As soon as the autotest is performed, zero values are displayed, provided that the measuring plate is unloaded. This is because the meter is automatically zeroed during the autotest procedure.
Please do not overload the load sensor! Risk of irreparable damage! If the torque value exceeds 120% of the full scale range, an acoustic signal is triggered and the symbol OL (eng. overload) appears on the display. If this is the case, remove the test load and then press the RESET key. Pressing the ON/OFF key switches the unit off.
When switching off, all current settings are saved, so that the same operating mode is selected when switching on again.
6.3 Basic functions of the device
The torque gauge can display the torque value for both clockwise (clock-wise, CW, symbol: ) and counter-clockwise (CCW, symbol: ) operation.
6 DB-BA-e-2020
6.3.1 CW/CCW torque value display Symbol
of the clockwise rotation
Load indicator beam of the clockwise rotation
Symbol
of the turning process against
clockwise
Load indicator beam of the counterclockwise rotation
The load indicator bar of the rotation is used to display the load to which the load sensor is exposed. When turning clockwise (CW), the load indicator bar moves from right to left. When turning counterclockwise (CCW), the load indicator bar moves from left to right.
6.3.2 Zeroing the display During operation it is often necessary to zero the display so that the values already
displayed are not taken into account in the next measurement. To do this, press and then release the ZERO button.
6.3.3 Changing the measuring unit Depending on the permissible load of the measuring instrument used, the following
measuring units are available: N-m, kgf-cm, kgf-m, in-lbf, ft-lbf. To change the measuring unit, press the UNITS key. Each time this key is pressed again, another measuring unit is displayed until the originally set measuring unit is displayed. During the selection process, the measurement results are automatically converted. Important! It must be remembered that not all units of measurement are always
available. This depends on the permissible load of the respective instrument.
DB-BA-e-2020 7
6.3.4 Changing the measuring mode The following measuring functions are available: Track, First Peak-Torque and Peak-
Torque. The measuring function is changed by pressing the MODE button. Pressing this key again selects the following measuring function until the originally set function is displayed.
6.3.5 Track function Press and hold the MODE button until the display shows the name of the measurement function "Track". The torque value for the rotation detected by the load sensor is displayed. The measurement results are displayed continuously:
Symbol of the track function
6.3.6 first peak torque function
Press and hold the MODE key until the display shows the designation of the measuring function "F-Peak". The first torque peak value is displayed:
Symbol of the first-peaktorque function
6.3.7 Peak torque function
Press and hold the MODE key until the display shows the name of the measurement function "Peak". The highest measured torque value is displayed:
Symbol of the Peak torque function
6.3.8 Resetting the device
Stored peak values of both types are deleted with the RESET button. This also enables the acquisition of further peak values.
8 DB-BA-e-2020
6.3.9 Display backlight
By pressing any key, the backlight of the display switches on for 60s. The same applies if the torque value detected by the load sensor will be higher than 0.5% of the full scale range.
6.3.10 Saving measured values
Measured values can be saved at any time. To do this, press the MEM/ENTER key. The memory capacity allows you to store up to 500 measurement results, including measurement units.
6.3.11 Output signal
The measurement result obtained can be transferred to the PC. It is done by pressing the PRINT key or by interrogating the instrument from the PC. The respective command can be sent via RS232 or USB.
DB-BA-e-2020 9
6.4 Main menu
Pressing the MENU/ESC key calls up the "Main Menu" screen. The arrow keys UP or DOWN are used to select desired positions in the mask. Pressing the ENTER key selects the desired sub-menu, then calls up the desired function and then enters the respective value. The arrow keys UP, DOWN, LEFT and RIGHT are also used to change numerical values. Pressing the ESC key returns to the main menu screen.
MAIN MENU
1) AUTO-OFF
2) PASS-FAIL
3) MEMORY
4) CALIBRATION
5) DIAGNOSTIC
6) ABOUT
6.4.1 AUTO-OFF (AUTOMATIC SWITCH-OFF)
Press the MENU key. The main menu mask appears on the display. Use the UP or DOWN arrow keys to position the cursor on the AUTO-OFF function. Press the ENTER key. The AUTO-OFF menu mask appears on the display. Press the ESC key to return to the main menu screen. The AUTO-OFF function helps to extend the battery life. It can be used to set the time after which the device should be switched off after the last operating activity (setting values: 5, 10 and 15 min). If this function is activated, the symbol AO appears in the main menu mask.
AUTO-OFF MENU
1) OFF
2) 5 MINUTE
3) 10 MINUTE
4) 15 MINUTE
Use the UP or DOWN arrow keys to position the cursor accordingly. Then press ENTER to select the AUTO-OFF function and return to the main menu screen.
MAIN MENU
1) AUTOMATIC SWITCH-OFF
2) PASS-FAIL
3) MEMORY
4) CALIBRATION
5) DIAGNOSIS FUNCTION
6) DEVICE INFO
10 DB-BA-e-2020
6.4.2 PASS-FAIL This function is used to set the permissible torque range. Thus, the upper and the lower
limit value of the torque range are determined. If the torque value falls below the defined range, the display shows the message PASS (OK). If the torque value detected is outside the specified range (above or below the range), the display shows FAIL (NOK). When the PASS-FAIL function is active, the display shows the symbol. To call up the PASS-FAIL menu screen, position the cursor on the PASS-FAIL function using the UP or DOWN arrow keys, then press the ENTER key. The display shows the PASS-FAIL function screen. Press the ESC key to return to the main menu screen.
PASS FAIL MENU
UPPER = 2. 5 N.m
LOWER = 1.0 N.m
Press 'Zero' key to
Clear both values.
Use the LEFT arrow key to position the cursor on the selected value. Set the desired value with the UP or DOWN arrow keys. Press and hold the key to scroll the list. Select the desired measuring unit with the RIGHT key. Press the ENTER key to save the setting made and return to the main menu screen.
The PASS-FAIL function is automatically turned off when the setting of the lower limit (LOWER) and upper limit (UPPER) is 0 N. The LOWER value must be lower than the UPPER value.
PASS FAIL MENU
UPPER LIMIT = 2. 5 N.m
LOWER LIMIT = 1.0 N.m
Press the 'Zero' key,
to delete both values.
DB-BA-e-2020 11
Example 1 LOWER LIMIT= 0 N-m, UPPER LIMIT= 20 N-m
Load
The "UPPER" LED is on. Other LEDs are off.
Upper limit
The "OK" LED is on. Other LEDs are off
Time
Example 2 LOWER LIMIT= 20 N-m, UPPER LIMIT= 0 N-m
Load
The "OK" LED is on. Other LEDs are off.
Lower limit
The "LOWER" LED wants ON. Other LEDs are off.
Time
Example 3 LOWER LIMIT= 10 N-m, UPPER LIMIT= 20 N-m
Load
The "UPPER" LED is on. Other LEDs are off.
Upper limit
The "OK" LED is on. Other LEDs are off.
Lower limit
The "LOWER" LED is on. Other LEDs are off.
Time
12 DB-BA-e-2020
6.4.3 MEMORY (MEMORY)
This function is used to view stored data records, to delete current data records, to delete all data records and to print data stored in these data records. To enter the MEMORY menu mask, first call up the main menu mask. Then use the UP or DOWN arrow keys to move the cursor to the MEMORY position and then press ENTER. The display shows the memory mask. To return to the main menu screen, press the ESC key.
Press the arrow keys UP or DOWN to change the memory mask. Press and hold the key to scroll to the position for changing the memory mask. Pressing the PRINT key controls the print output of stored data and data transmitted via serial port. To access the DELETE menu screen, press the ZERO key.
DELETE?
1) NO
2) DELETE
3) DELETE ALL
Select the desired delete option with the UP or DOWN arrow keys. If the NO option is selected and then the ENTER key is pressed, the system returns to the memory mask. If the DELETE option is selected and the ENTER key is pressed, the current stored results are deleted and the memory mask is returned to. Selecting the DELETE ALL option and pressing ENTER will delete all stored results and return to the memory mask.
DB-BA-e-2020 13
6.4.4 CALIBRATION (CALIBRATION)
Calibration of the measuring instrument is carried out by the manufacturer's service personnel. Detailed information in this regard can be obtained from SAUTER's dealer or directly from the manufacturer.
6.4.5 DIAGNOSTIC (DIAGNOSTIC FUNCTION)
This function is used to check the status of the load cell. If there are indications that the load cell transformer has been overloaded, it is possible to check the state of the load cell directly. To do this, place the measuring instrument horizontally on a level, horizontal surface and call up the main menu screen. Select the DIAGNOSTIC function with the UP or DOWN arrow keys and press ENTER. The "Diagnostic" menu mask appears on the display. To return to the main menu screen, press the ESC key.
DIAGNOSTIC
OVERLOAD COUNT: 2
Total overload measurement
ORG. OFFSET : +0.4 %
Offset value from the last calibration (%)
CUR. OFFSET : +0.4 %
Current offset value (%)
If the offset value is in the range of 5% to 10%, contact the instrument supplier to have a new calibration performed.
If the offset value exceeds 10%, contact the device supplier to have the load cell replaced. The values given are only approximate values. Calibration/repair of the instrument can be caused by various circumstances. This is due to device-specific parameters of the load cell.
14 DB-BA-e-2020
6.4.6 ABOUT (DEVICE INFO)
This function is used to display device-related information (firmware version, model, permissible load, serial number). To enter the ABOUT menu screen, first call up the main menu screen, then position the cursor on the ABOUT function using the UP or DOWN arrow keys, and then press ENTER. The display shows the About menu screen. To return to the main menu screen, press the ESC key.
ABOUT
FIRMWARE REV. : 3.0
MODEL: DB
CAPACITY: 10 N.m
S/N: 05130001
6.5 Measurement methods
In order to achieve higher accuracy of measurements, the torque ratio must correspond to the value specified for the respective device. The bending and torque loads acting on the load cell must be reduced, as they can affect the measurement result. The permissible load according to the specifications on the front of the device must not be exceeded under any circumstances. Otherwise the load cell will be permanently damaged, even in the event of a short-term overload. Such damage is excluded from the warranty.
DB-BA-e-2020 15
7 Control buttons
MENU/ESC:
Opens the menu window with the individual submenus
By pressing the MENU/ESC button in the menu, you can return to the previous
page
ZERO (zeroing):
Zeroing the display in track mode (tare function)
ENTER:
Confirm the selection in the menu
Saving a measured value in PEAK mode
PRINT (print function):
Output of the memory contents to PC or printer
RESET (delete function):
Delete the current PEAK value
UNIT (units of measurement):
Press the key briefly: Switch between N, gf, kgf, ozf, lbf, mN
Mode (mode change):
Assignment with three functions:
Track mode (continuous measurement)
Peak Mode Train
Peak Mode Print
ON / OFF:
On / Off button (press button for approx. 1 s)
16 DB-BA-e-2020
8 Conversion factoren
Unit
n.m.
p.cm.
plc
Calo-
funt
Stopo-
funt
N. m
1
10,197
0,10197
8,8507
0,73756
p.cm.
0,0980665
1
0,01
0,86796
0,07233
plc
9,80665
100
1
86,796
7,233
calofunt
0,11298
1,152
0,01152
1
0,08333
stopofunt
1,3558
13,8255
0,138255
12
1