KERN & SOHN KFN-TM User Manual

KERN & Sohn GmbH
Ziegelei 1
E-Mail: info@kern-sohn.com
Operating and Installation Instructi ons Display devices
KERN KFB/KFN-TM
Version 2.3
07/2013
GB
D-72336 Balingen
Fax: +49-[0]7433-9933-149
KFB/KFN-TM-BA_IA-e-1323
KERN KFB/KFN -TM
Display units
GB
Version 2.3 07/2013
Operating and installa t ion instructions
Contents
1 Technical data ................................................................................................ 4
2 Appliance overview ....................................................................................... 5
2.1 Keyboard overview ..................................................................................................................... 7
2.1.1 Numerical input via the navigation buttons................................................................................. 8
2.2 Overview of display .................................................................................................................... 8
3 Basic Information (General) .......................................................................... 9
3.1 Proper use .................................................................................................................................. 9
3.2 Improper Use .............................................................................................................................. 9
3.3 Warranty ..................................................................................................................................... 9
3.4 Monitoring of Test Resources .................................................................................................. 10
4 Basic Safety Precautions ............................................................................ 10
4.1 Pay attention to the instructions in the Operation Manual........................................................ 10
4.2 Personnel training ..................................................................................................................... 10
5 Transport and storage ................................................................................. 10
5.1 Testing upon acceptance ......................................................................................................... 10
5.2 Packaging / return transport ..................................................................................................... 10
6 Unpacking and placing ................................................................................ 11
6.1 Installation Site, Location of Use .............................................................................................. 11
6.2 Unpacking ................................................................................................................................. 11
6.3 Scope of delivery / serial accessories: ..................................................................................... 11
6.4 Transportation lock (illustration example) ................................................................................. 12
6.5 Error message .......................................................................................................................... 12
6.6 Placing ...................................................................................................................................... 13
6.7 Mains connection ...................................................................................................................... 14
6.8 Storage battery operation (optional) ......................................................................................... 14
6.9 Adjustment ................................................................................................................................ 15
6.9.1 Verified weighing systems ........................................................................................................ 15
6.9.2 Non verifiable weighing syst ems .............................................................................................. 18
6.10 Linearization ............................................................................................................................. 19
6.10.1 Verified weighing systems:................................................................................................... 19
6.10.2 Non-verified weighing systems ............................................................................................ 20
6.11 Verification ................................................................................................................................ 21
2 KFB/KFN-TM-BA_IA-e-1323
7 Operation ...................................................................................................... 24
7.1 Start-up ..................................................................................................................................... 24
7.2 Switching Off ............................................................................................................................ 24
7.3 Zeroing ..................................................................................................................................... 24
7.4 Simple weighing ....................................................................................................................... 24
7.5 Switch-over weighing unit (only not verifiable weighing systems) ........................................... 25
7.6 Weighing with tare .................................................................................................................... 26
7.7 Weighing with tolerance range ................................................................................................. 27
7.7.1 Tolerance check for target weight ............................................................................................ 28
7.7.2 Tolerance check for target quantity .......................................................................................... 30
7.8 Manual totalizing ....................................................................................................................... 32
7.9 Automatic adding-up ................................................................................................................. 34
7.10 Parts counting ........................................................................................................................... 35
7.11 Animal weighing ....................................................................................................................... 36
7.12 Lock keyboard .......................................................................................................................... 37
7.13 Display background illum inatio n ............................................................................................... 37
7.14 Automatic switch-off function „AUTO OFF“ .............................................................................. 38
8 Menu ............................................................................................................. 39
8.1 Overview non verifiable weighing systems (contacts of circuit board [K1] not short-circuited) 40
8.2 Overview verified weighing systems (contacts of circuit board [K1] short-circuited by means of jumper) 43
9 Service, maintenance, disposal .................................................................. 46
9.1 Clean ........................................................................................................................................ 46
9.2 Service, maintenance ............................................................................................................... 46
9.3 Disposal .................................................................................................................................... 46
9.4 Error messages ........................................................................................................................ 46
10 Data output RS 232C ................................................................................... 48
10.1 Technical data .......................................................................................................................... 48
10.2 Printer mode ............................................................................................................................. 48
10.3 Output log (continuous output) ................................................................................................. 49
10.4 Remote control instructions ...................................................................................................... 49
11 Instant help ................................................................................................... 50
12 Installing display unit / weighi ng bridge .................................................... 51
12.1 Technical data .......................................................................................................................... 51
12.2 Weighing system design........................................................................................................... 51
12.3 How to connect the platform ..................................................................................................... 52
12.4 Configure display unit ............................................................................................................... 53
12.4.1 Verified weighing systems (contacts of circuit board [K1] short-circuited by means of jumper) 53
12.4.2 Non verifiable weighing systems (contacts of circuit board [K1] not short-circ u it ed ) .......... 59
13 Declaration of Conformity / Test Certificate .............................................. 63
KFB/KFN-TM-BA_IA-e-1323 3
1 Technical data
KERN KFB-TM KFN-TM
Display 5 ½ - digit Resolution (verified) 6000
Single (Max.) 6.000 e
Dual (Max.) 3.000 e
Resolution (non-verified) 30.000 Weighing ranges 2 Divisions 1,2,5,…10n Weighing Units kg
Functions Display LCD 52 mm digits with back lighting DMS weighing cells Range calibration We recommend ≥ 50 % max.
Data output RS232
Electric Supply
Housing 250 x 160 x 58 266 x 165 x 96 Admissible
ambient temperature Net weight 1.5 kg 2 kg
Rechargeable battery (optional)
Operating / charge time
Weighing with tolerance range, Totalizing, Animal weighing
80-100 . Max. 4 item per 350 Ω;
Sensitivity 2-3 mV/V
Input voltage 220 V – 240 V, 50 Hz
Power pack secondary voltage 12V, 500mA
0°C – 40°C (non-verified)
-10°C – 40°C (verified)
35 h / 12 h 90 h / 12 h
RS 232 interface Standard Option
Tripod KERN BFS-07, option Support base
incl. wall bracket
IP protection -
4 KFB/KFN-TM-BA_IA-e-1323
Standard
IP 67 as per DIN 60529
(rechargeable battery
operation only)
2 Appliance overview
KFB-TM: Synthetic finis h
1. Status of rechargeable battery
2. Keyboard
3. Weight display
4. Tolerance margin, see chap. 7.7
5. Weighing unit
6. RS-232
7. Input connection load cell cable
8. Guide rail support base / stand
9. End stop support base / stand
10. Mains adapter connection
11. Adjustment switch
KFB/KFN-TM-BA_IA-e-1323 5
KFN-TM: Stainless steel finish
1. Status of rechargeable battery
2. Keyboard
3. Weight display
4. For tolerance mark see chap. 7.7
5. Weighing unit
6. Input connection load cell cable
7. Mains adapter connection
6 KFB/KFN-TM-BA_IA-e-1323
2.1 Keyboard overview
Key Function
Turn on/off
Zeroing
Navigation button
Navigation key
Navigation key
Navigation key
C
Confirm entry
Taring
At numeric input increase flashing digit
Scroll forward in menu
Display sum total
Digit selection to the right
Add weighing value to summation memory
Digit select ion to the left
Calculate weighing data via interface
Delete
Change between gross and net weight
ESC
KFB/KFN-TM-BA_IA-e-1323 7
Back to menu/weighing mode
Call up an imal weighing funct ion
Call up weighing with tolerance range
Delete total added memory
2.1.1 Numerical input via the navigation buttons
Press and current setting will be displayed. The first digit will be flashing
and is ready for changing.
If you do not wish to change the first digit, press and the second digit will
start flashing. Each time you press , the display will move to the subsequent digit, after the
last digit the display will return to the first digit.
To change the selected (flashing) digit, press repeatedly until the desired
value is displayed. Then press to access further digits and change them by
.
Complete your entry by .
2.2 Overview of display
Display Significance
Battery very low
STABLE Stability display
ZERO Zero indicator
GROSS Gross weight
NET Net weight
AUTO Automatic add-up enabled
Kg Weighing unit
M+ Totalisation
LED
+ / / -
8 KFB/KFN-TM-BA_IA-e-1323
Indicators for weighing with tolerance range
3 Basic Information (General)
3.1 Proper use
The display unit acquired by you is used in combination with a weighing plate and serves to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic weighing system”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use display unit for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results ca n be displayed due to the “stability compensation“ in the display unit. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing pan. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the weighing plate, minus a possibly existing tare load, must be strictly avoided. Both, the weighi ng plate and the display unit may be damaged during this process. Never operate display unit in explosive environment. The serial version is not explosion protected. Changes to the display unit's design are not permitted. This may lead to incorrect weighing results, safety-related faul ts and des tr ucti on o f the dis pl ay unit. The display unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opene d
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
KFB/KFN-TM-BA_IA-e-1323 9
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the display unit and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of display units’ test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and display units may be calibrated (return to the national standard) fast and at moderate cost.
4 Basic Safety Precautions
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 T ransport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning. Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices. Secure all parts such as the glass wind screen, the
weighing platform, power unit etc. against shifting and damage.
10 KFB/KFN-TM-BA_IA-e-1323
Display Unit
6 Unpacking and placing
6.1 Installation Site, Location of Use
The display units are designed in a way that reliable weighing results are achieved in common conditions of use. Precise and fast work is achieved by selecting the right place for your display unit and your weighing plate.
On the installation site observe the following:
Place the display unit and the weighing plate on a stable, even surface.
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the display unit and the weighing plate against direct draf t fr om ope n windows or doors.
Avoid jarring during weighing;
Protect the display unit and the weighing plate against high humidity, vapours
and dust.
Do not expose the display unit to extreme dampness for longer periods of time. Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.
6.2 Unpacking
Take the display unit carefully out of its packaging, remove the plastic jacket and install it at the designated work space.
6.3 Scope of delivery / serial accessories:
Mains adapter
Support base incl. wall bracket
Operating instructions
KFB/KFN-TM-BA_IA-e-1323 11
Transport Securing
6.4 Transportation lock (illustration example)
Please note: if the display unit is used together with platform with transportation lock, this transportation lock must be released prior to use:
6.5 Error message
As soon as an error message appears in the balance display, the balance must not more be used, e.g. Err 4
12 KFB/KFN-TM-BA_IA-e-1323
6.6 Placing
Mount the display unit in a way that facilitates operation and where it is easy to see.
Usage with support base (KFB-TM only)
Push support base holder in guide rail [8] up to end stop [9], see chap. 2.
Usage with wall mount (KFB-TM only)
Use the wall mount to affix the display unit to the wall.
Using with tripod (optional)
An optional tripod (KERN BFS-07) is available if the display unit is to be mounted in a raised position.
KFB/KFN-TM-BA_IA-e-1323 13
6.7 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.
6.8 Storage battery operation (optional)
Before the first use, the battery should be charged by connecting it to the mains power supply for at least 12 hours.
If the weight display shows , this is an indication that the capacity of the rechargeable battery is almost exhausted. The unit will be ready for operation for approx. another 10 hours before switching off automatically. Charge the battery with the help of the supplied power pack. The LED display informs you during loading about the loading status of the rechargeable battery.
red: Voltage has dropped below prescribed minimum. green: Battery is completely discharged yellow: Charging storage battery
To conserve energy, enable the automatic switch-o ff function „AUTO OFF“, see chap. 7.14.
14 KFB/KFN-TM-BA_IA-e-1323
6.9 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each display unit with connected weighing plate must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the weighing system has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the display unit periodically in weighing operation.
In weighing systems with a resolution of < 15 000 dividing steps an adjustment is recommended. In weighing systems with a resolution of > 15 000 dividing steps a linearisation is recommended (see chap. 6.10).
Prepare the required adjustment weight. The weight to be used dep ends on the capacity of the scale. Carry out adjustment as near as possible to the scale’s maximum weight. Info about test weights can be found on the Internet at: http://www.kern-sohn.com.
Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
6.9.1 Verified weighing systems
In verified weighing systems the menu item for adjustment „P2 mode“ is blocked.
KERN KFB-TM
To disable the access lock, destroy the seal and actuate the adjustment switch. Position of the adjustment switch see chap. 6.11
KERN KFN-TM To override the blocked access you will have to destroy the seal before calling
up the menu and to short-circuit the two contacts on the circuit board [K2], using a jumper (See chap. 6.11).
Attention: After destruction of the seal the weighing system must be re-veri fied by an authorised agency and a new verification wire/seal mark fitted before it can be reused for applications subject to verification.
KFB/KFN-TM-BA_IA-e-1323 15
Call up menu:
1. Switch-on balance and during the selftest press .
2. Press , , subsequently, the first menu block
„PO CHK“ will be displayed.
3. Press repeatedly until „P2 mode“ will be displayed.
For the KFB-TM model operate the adjustment switch.
4. Press and select the set weighing scales type by .
= Single-range balance
= Dual range balance
= Multi-interval balance
5. Acknowledge with .
6. Press repeatedly until „CAL“ will be displayed.
7. Confirm with and select setting „noLin“ by .
16 KFB/KFN-TM-BA_IA-e-1323
How to carry out an adjustment:
Confirm menu setting „noLin“ by .
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press . The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with . Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press .
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return into weighing mode automatically. An adjusting error or incorrect adjusting weight will be indicated by the error message; repeat adjustment procedure.
KFB/KFN-TM-BA_IA-e-1323 17
Call up menu:
6.9.2 Non verifiable weighing systems
1. Switch-on balance and during the selftest press .
2. Press subsequently , , the first menu block
„PO CHK“ will be displayed.
3. Press repeatedly until „P3 CAL“ will be displayed.
4. Confirm with ; press repeatedly until „CAL“
appears.
5. Acknowledge using , the current setting is displayed.
Press to confirm; press to select setting.
noLin = adjustment LineAr = linearization, see chap. 6.10
How to carry out adjustment:
Confirm menu setting „noLin“ by .
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press . The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with . Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press .
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return into weighing mode automatically. An adjusting error or incorrect adjusting weight will be indicated by the error message; repeat adjustment procedure.
18 KFB/KFN-TM-BA_IA-e-1323
In balances with a resolution of > 15 000 dividing steps carrying out a
the adjustment switch s ee chap. 6.11
6.10 Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value of the respective test weight according to plus and minus over the entire weighi ng range. If linearity deviation is discovered during a testing instrument control, you can improve this by means of linearization.
6.10.1 Verified weighing systems:
linearisation is recommended.
Carrying out linearization is restricted to specialist staff possessing well acquainted with the workings of weighing scales.
The test weights to be used must be adapted to the weighing scale’s specifications; see chapter “testing instruments control”.
Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
After successful linearisation you will have to carry out calibration; see chapter “testing instruments control”.
The adjustment is locked for verified balances. To disable the access lock, destroy the seal and actuate the adjustment switch. Position of
Menu item P2 modeCalCall up liner, see chap. 6.9.1
Confirm by , the password query „Pn“ will be displayed.
Press subsequently , , or , , .
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press . When “Ld 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform. Wait for stability display, then press .
When “Ld 2“ is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform. Wait for stability display, then press .
When “Ld 3“ is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform. Wait for stability display, then press .
KFB/KFN-TM-BA_IA-e-1323 19
After linearisation the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return into weighing mode automatically.
6.10.2 Non-verified weighing systems
Call-up menu item P3 CALCalLiner , see c hap. 6.9.1
Confirm by , the password query „Pn“ will be displayed.
Press , , or , , subsequently.
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press . When “Ld 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform. Wait for stability display, then press .
When “Ld 2“ is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform. Wait for stability display, then press .
When “Ld 3“ is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform. Wait for stability display, then press .
After a successful linearisation the balance will carry out a
self-test. Remove adjusting weight during selftest, bal a nce will return into weighing mode automatically.
20 KFB/KFN-TM-BA_IA-e-1323
6.11 Verification
General introduction: According to EU directive 90/384/EEC balances must be officially verified if they are used as follows (legal ly controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purpose. d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification notes:
An EU Qualification Approval is in existence for verified weighing systems. If a balance is used where obligation to verify exists as described above, it must be verified and re-verified at regular intervals. Reverification is carried out according to the relevant national statutory regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
Verification of the weighing system is invalid without the "seal".
KFB/KFN-TM-BA_IA-e-1323 21
[K1]
Notes on verified weighing systems KFB-TM:
Access to conductor plate:
Remove seal
Open display unit
The application of the display unit as a weighing system able to be veri fie d
requires that the contacts of the circuit board are short-circuited with the help of a jumper [K1]. For non verifiable weighing systems remove the jumper.
In verified weighing systems the menu item for adjustment, „P2 mode“ will be blocked.
To disable the access lock, destroy the seal and actuate the adjustment switch.
Position of seals and adjusting switch
1. Self-destroying seal mark
2. Adjustment switch
3. Cover of adjustment switch
4. Self-destroying seal mark
22 KFB/KFN-TM-BA_IA-e-1323
[K1]
[K2]
KFN-TM:
Access to conductor plate:
Remove seal
Open display unit
The application of the display unit as a weighing system able to be verified
requires that the contacts of the circuit board are short-circuited with the help of a jumper [K1]. For non verifiable weighing systems remove the jumper.
To adjust, short-circuit the contacts of the circuit board, using a jumper [K2].
KFB/KFN-TM-BA_IA-e-1323 23
Overloading exceeding the stated maximum load (max) of the device, minus a
7 Operation
7.1 Start-up
Press and the instrument will carry out a self-test. As soon as the weight
display appears, the instrument will be ready to weigh.
7.2 Switching Off
Press and the display will disappear.
7.3 Zeroing
Resetting to zero corrects the influence of light soiling on the weighing plate. The unit is equipped with an automatic zero setting function. Therefore the unit can be reset to zero at any time as follows:
To unload the weighing system Press and zero display as well as indicator ZERO will appear.
7.4 Simple weighing
Place goods to be weighed on balance. Wait until stability display STABLE appears. Read weighing result.
Overload warning
possibly existing tare load, must be strictly avoided. This could damage the instrument. Exceeding maximum load is indicated by the display of „----“ and an audio sound. Unload weighing system or reduce preload.
24 KFB/KFN-TM-BA_IA-e-1323
7.5 Switch-over we i ghing unit (only not verifiable weighing systems)
How to enable weighing units:
Call-up menu item P5 Unt, see chap. 8.1
Press and the first weighing unit with the current setting
will be displayed.
To enable [on] / disable [off] the displayed weighing unit,
press
Acknowledge with . The next unit with the current setting
will be displayed.
To enable [off] / disable [on] the displayed weighing unit,
press .
Acknowledge with . Repeat sequence for each weighing unit.
Note: „tj“ and „Hj“ cannot be activated at the same time, only either
... or ... .
Return to weighing mode using
Switch-over weighing unit:
Keep pressed, the display changes over to the
weighing units activated before (e.g. kg  lb)
KFB/KFN-TM-BA_IA-e-1323 25
7.6 Weighing with tare
Deposit weighing vessel. After successful standstill control press the button.
Zero display and indicator NET appear.
The weight of the container is now internally saved.
Weigh the material, the net weight will be indicated. The weight of the weighing container will be displayed as a minus number after
removing the weighing container.
The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The limit is reached when the taring range capacity (see type plate)is full.
To change between gross weight and net weight, press . To delete the tare value, remove load from weighing plate and press .
26 KFB/KFN-TM-BA_IA-e-1323
An acoustic signal sounds when load is beyond tolerance
7.7 Weighing with tolerance range
You can set an upper or lower limit when weighing with tolerance range and thus ensure that the weighed load remains exactly within the set limits. During tolerance tests such as dosing, portioning and sorting the unit will indicate exceeded or undershot limits by emitting an optical or acoustic signal.
Audio signal:
The acoustic signal depends on the settings in menu block „BEEP“. Options:
no
ok
ng
Optical signal:
Three colour signal lights indicate whether the load is within the two tolerance limits. The signal lights provide the following information:
Acoustic signal turned off An acoustic signal sounds when load is within tolerance
limits
limits
+
Goods to be weighed above tolerance limit
Goods to be weighed within tolerance range
Goods to be weighed below tolerance limit
-
Red signal light glowing
Green signal light glowing
Red signal light glowing
The settings for tolerance weighing may be called up either via menu block „P0 CHK“ (see chap. 8) or faster via the key combination
KFB/KFN-TM-BA_IA-e-1323 27
and the current setting for the acoustic signal will
7.7.1 Tolerance check for target weight
Settings
Press and at the same time in weighing mode.
Press until the display for entering the lower limit value
appears.
Press , the current setting will be displayed. To enter the lower limit, e. g. 1000 Kg, press the navigation
keys (See chap. 2.1.1); the currently enabled digit will be flashing.
Confirm input by .
Press repeatedly until is displayed.
Press , the current setting for the upper limit will be
displayed.
Press the navigation keys (See chap. 2.1.1) to enter the
upper limit, e.g. 1,100 kg; the currently enabled digit will be flashing.
Confirm input by .
Press repeatedly until is displayed.
Press
be shown.
Select desired setting (no, ok, ng) by .
Confirm input by .
28 KFB/KFN-TM-BA_IA-e-1323
Press ; weighing system is in tolerance weighing mode.
From here evaluation takes place whether the goods to be weighed are within the two tolerance limits.
Weighing with tolerance range
Tare when using a weighing container. Put on goods to be weighed, tolerance control is started. The signal lights
indicate whether the load is within the two set limits.
Load below specified tolerance
Red signal light
next to „-“ ON illuminated
The tolerance control is not active when the weight is under 20d.
To delete limits, enter “00.000 kg“.
Load within specified tolerance
Green signal light
next to „„ illuminated
Load exceeds specified tolerance
Red signal light
next to „+“ ON illuminated
KFB/KFN-TM-BA_IA-e-1323 29
Press and the current setting for the acoustic signal will
7.7.2 Tolerance check for target quantity
Settings
Press and at the same time in weighing mode.
Press until the display for entering the lower limit value
appears.
Press , the current setting will be displayed. To enter the lower limit, e. g. 75 items, press the navigation
buttons (see chap. 2.1.1); the currently enabled digit will be flashing.
Confirm input by .
Press repeatedly until is displayed.
Press , the current setting for the upper limit will be
displayed.
To enter the upper limit, e. g. 100 items, press the navigation
buttons (see chap. 2.1.1); the currently enabled digit will be flashing.
Confirm input by .
Press repeatedly until is displayed.
be shown.
Select desired setting (no, ok, ng) by .
Confirm input by .
30 KFB/KFN-TM-BA_IA-e-1323
Press ; weighing system is in tolerance weighing mode.
From here evaluation takes place whether the goods to be weighed are within the two tolerance limits.
Weighing with tolerance range
Set item weight, see chap. 7.10. Tare when using a weighing container.
Put on goods to be weighed, tolerance control is started. The signal lights
indicate whether the load is within the two set limits.
Load below specified tolerance
Red signal light
next to „-“ ON illuminated
The tolerance control is not active when the weight is under 20d.
To delete limits, enter „00000 PCS“.
Load within specified tolerance
Green signal light
next to „„ illuminated
Load exceeds specified tolerance
Red signal light
next to „+“ ON illuminated
KFB/KFN-TM-BA_IA-e-1323 31
Menu setting:
The totalizing function is not active when the weight is under 20d.
7.8 Manual totalizing
With this function the individual weighing values are added into the summation memory by pressing and edited, when an optional printer is connected.
P1 COM“ or „P2 COMMODE“ PR2““, see chap. 8
Add up:
Place weighing goods A.
Wait until the stability display STABLE appears, then press . The weight value will be saved and printed if an optional printer is connected.
Remove the weighed good. More weighed goods can only be added when the
display ≤ zero.
Place goods to be weighed B.
Wait until the stability display appears, then press . Weighing value will be added to summation memory and possibly printed. The number of weighing actions, followed by the total weight will be displayed for 2 sec.
Add more weighed goods as described before.
Please note that the weighing system must be unloaded between the individual weighing procedures.
This process may be repeated 99 times or till such time as the capacity of the
weighing system has been exhausted.
Display and output sum „Total“:
Press , number of weighing, followed by the total weight will be displayed for
2 sec. Press to print out this display.
32 KFB/KFN-TM-BA_IA-e-1323
Menu setting
Menu setting
Delete weighing data:
Press and at the same time The data in the summation memory are
deleted.
Printout example KERN YKB-01N, verified weighing system:
„P1 COM“ or „P2 COM“ „Lab 2“ / Prt 7“
„P1 COM“ or „P2 COM“ „Lab 0“ / Prt 0“
1 First weighing 2 Second weighing 3 Third weighing
4 Number of weighings / total
+
KFB/KFN-TM-BA_IA-e-1323 33
Menu settings:
7.8.
7.9 Automatic adding-up
With this function the individual weighing values are automatically added into the summation memory when the bala nce is unloaded without pressing and edited,
when an optional printer is connected.
P1 COM“ or „P2 COM MODE“ AUTO““, see chap. 8 Der Indikator AUTO wird angezeigt.
Add up:
Place weighing goods A.
After the standstill control sounds a signal tone. The weighing value will be added to the summation memory and printed.
Remove the weighed good. More weighed goods can only be added when the
display ≤ zero.
Place goods to be weighed B.
After the standstill control sounds a signal tone. The weighing value will be added to the summation memory and printed. Number of weighing, followed by the total weight will be displayed for 2 sec.
Add more weighed goods as described before.
Please note that the weighing system must be unloaded between the individual weighing procedures.
This process may be repeated 99 times or till such time as the capacity of the
weighing system has been exhausted.
Display and delete the weighing data, as well as printout examples see chap.
34 KFB/KFN-TM-BA_IA-e-1323
7.10 Parts c ounti ng
Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by putting on a certain number of the parts to be counted. The balance determines the total weight and divides it by the number of parts, the so­called reference quantity. Counting is then carried out on the basis of the calculated average piece weight. As a rule: The higher the reference quantity the higher the counting exactness.
In weighing mode , press and hold until the message „P
10“ appears that is used to set the reference quantity.
Use to set the desired reference quantity (such as 100),
options include P 10, P 20, P 50, P100, P 200.
Place as many items to be counted (such as 100 items) as
demanded by the set reference quantity and confirm by The weighing scales calculate the reference weight. The current quantity (such as 100 items) will be displayed.
Remove reference weight. The balance is from now in parts
counting mode counting all units on the weighing plate.
Back to Weighing mode by .
KFB/KFN-TM-BA_IA-e-1323 35
7.11 Animal weighing
The animal weighing function is suitable for weighing restless loads. The weighing system will display a mean value derived from several weighing results.
The animal weighing program can be enabled by either calling up menu block „P3 OTH“ or „P4 OTHANMON“ (See chap. 8) or faster via key combination.
The indicator shows HOLD as long as the animal weighing function remains enabled.
Place the load on the weighing system and wait until the scale is steady. Press and at the same time; you will hear an acoustic signal, indicating
that the animal weighing function is enabled. Whilst averaging is taking place you can add or remove loads as the measuring value will be constantly updated.
To deactivate the animal weighing function press and at the same time.
36 KFB/KFN-TM-BA_IA-e-1323
7.12 Lock keyboard
To enable/disable the keyboard lock go to menu item „P3 OTH“ orP4 OTH“ LOCK“, see chap.8. Whilst the function is enabled the keyboard will self-lock after no key has been pressed for 10 minutes. „K-LCK“ will be displayed as soon as a key is pressed.
To disable the lock, press , and hold plus (2 s) until „U LCK“ appears.
7.13 Display background illumination
Keep pressed (3s) until „setbl“ appears.
Press again, the current setting will be displayed.
Use to select the desired setting.
bl on Continu ous bac kg r oun d li g hti ng bl off Background illumination off bl Auto Automatic background illumination on when weighing pate is
loaded or key pressed.
Either save by or cancel by pressing .
Back to weighing mode by .
KFB/KFN-TM-BA_IA-e-1323 37
7.14 Automatic switch-off function „AUTO OFF“
The unit is automatically switched off within the preset time when the display unit or the weighing bridge are not operated.
Keep pressed (3s) until „setbl“ appears.
Press to call up AUTO OFF-function
Press , the current setting will be displayed. Use to select the desired setting.
of 0 AUTO OFF - function disabled of 3 Weighing system wi l l be turned off a fter 3 min . of 5 Weighing system wi l l be turned off a fter 5 min . of 15 Weighing system will be turned off after 15 mi n. of 30 Weighing system will be turned off after 30 min.
Either save by or cancel by pressing .
Back to weighing mode by .
38 KFB/KFN-TM-BA_IA-e-1323
current setting will be displayed.
8 Menu
The application of the display unit as a verified weighing system requires that you short-circuit the two contacts [K1] of the circuit board, using a jumper. To that effect, a menu for verified weighing systems is available. For menu layout see chap. 8.2. There is no jumper for weighing systems that cannot be verified. To that effect, a menu is available for weighing systems that cannot be verified, Menu layout see chap. 8.1
Navigation in the menu:
Call up menu
Switch-on balance and during the selftest press .
Press , , subsequently, the first menu
block „PO CHK“ will be displayed.
Select menu block
Select setting
Change settings
Acknowledge setting / exit the menu
Return to weighing mode
With help of , the individual menu items can be
selected one after the other.
Confirm selected menu item by pressing . The
To change to the available settings, press the
navigations keys as described in chap. 2.1.
Either save by pressing or cancel by pressing .
Press repeatedly to exit menu.
KFB/KFN-TM-BA_IA-e-1323 39
Menu block Main menu
Menu item Submenu
Acoustic signal for weighing with tolerance range switched off
P1 REF
P2 COM
MODE
S0 off
Continuous data output, selectable “send zero” yes / no
S0 on
ST1
One output for stable weighing value
STC
Continuous data output of stable weighing values
PR1 Output after pressing
PR2
Manual totalizing, see chap. 7.8.
added to the summation memory and issued.
8.1 Overview non verifiable weighing systems (contacts of circuit board [K1] not short-circuited)
Available settings / explanation
PO CHK
Weighing with tolerance range, see chap. 7.7
Zero point settings
nEt H Upper limit value „Tolerance check weighing“, input see
chap. 7.7.1
nEt LO Lower limit value „Tolerance check weighing“, input see
chap. 7.7.1
PCS H Upper limit value „Tolerance check counting“, input see
chap. 7.7.2
PCS L Lower limit value „Tolerance check counting“, input see
chap. 7.7.2
BEEP no
ok Audio sound when load is within tolerance limits nG Audio sound when load is beyond tolerance limits
A2n0 Automatic zero point correction (Autozero) by changing the
display, digits selectable (0.5d, 1d, 2d, 4d)
0AUto Zero setting range
Load range where the display after switching-on the balance is set to zero.
Selectable 0, 2, 5, 10, 20, 30, 50, 100 %
0rAGE Zero setting range
Load range where the display is set to zero by pressing
. Selectable 0, 2, 4, 10 , 20* , 50, 100%.
0tArE Automatic taring „on / off“, taring range adjustable in menu
item „0Auto“. SPEEd Not documented Zero Zero point setting
CONT
Interface parameter
Press and the weighing value will be
40 KFB/KFN-TM-BA_IA-e-1323
AUTO*
For automatic add-up see chap. 7.9.
memory on unloading of weighing scale.
wirel
Not documented
BAUD
Available Baudrate: 600, 1200, 2400, 4800, 9600*
Pr
7E1
7 bits, even parity
7o1
7 bits, odd parity
8n1*
8 bits, no parity
PTYPE tPUP*
Standard printer setti n g
LP50
Not documented
Lab
Lab x
Prt
Prt x
LAnG
eng*
Standard settings English
chn
P3 CAL
COUNT
Display internal resolution
DECI
Position of the decimal dot
DUAL
Setting balance type, capacity (Max) and readability (d)
off Single-range balance
R1 inc
Readability
R1 cap
Capacity
on
Dual range balance
R1 inc
Readability 1st weighing range
R1 cap
Capacity 1st weighing range
R2 inc
Readability 2nd weighing range
R2 cap
Capacity 2nd weighing range
CAL
noLin
For adjustment, see chap. 6.9.2
Liner
For linearization, see chap. 6.10.2
GrA
Not documented
on
Keyboard lock enabled, see chap. 7.11
off*
Keyboard lock disable d
on
Animal weighing enabled, see chap. 7.10
off*
Animal weighing disabled
This function is used to issue and add individual weighing values automatically to the summation
Configuration data see chap. 12.4
ASK
For remote control commands, see chap. 10.4
For data output format, see chap.8.2, tab. 1
P4 OTH
LOCK ANM
KFB/KFN-TM-BA_IA-e-1323 41
kg on* off
g on off*
lb on off*
oz on off*
tJ on off
HJ on off
P5 Unt
Switch-over weighing unit, see chap. 7.5
P6 xcl Not documented P7 rst
Use to reset balance settings to factory default.
P8 uwb Not documented
Factory settings are marked by *.
42 KFB/KFN-TM-BA_IA-e-1323
Menu block Main menu
Menu item Submenu
Acoustic signal for weighing with tolerance range switched off
Audio sound when load is within tolerance limits
Audio sound when load is beyond tolerance limits
P1 COM
MODE
CONT S0 off
Continuous data output, selectable “send zero” yes / no
S0 on
ST1
One output for stable weighing value
Continuous data output of stable weighing values
PR1
Output after pressing
PR2
Manual totalizing, see chap. 7.8
added to the summation memory and issued.
AUTO
8.2 Overview verified weighing systems (contacts of circuit board [K1] short-circuited by means of jumper)
In verified weighing systems the access to „P2 mode and „P4 tAr“ is locked.
KERN KFB-TM:
To disable the access lock, destroy the seal and actuate the adjustment switch. Position of the adjustment switch see chap. 6.11.
KERN KFN-TM: In order to unlock the access, the seal must be destroyed and both contacts of the
printed circuit board [K2] must be shor t -circuited by a jumper, see chap. 6.11. Attention:
After destruction of the seal the weighing system must be re-veri fied by an authorised agency and a new verification wire/seal mark fitted before it can be reused for applications subject to verification.
PO CHK
Weighing with tolerance range, see chap. 7.7
Interface parameter
Available settings / explanation
nEt H Upper limit value „Tolerance check weighing“, input see
chap. 7.7.1
nEt LO Lower limit value „Tolerance check weighing“, input see
chap. 7.7.1
PCS H Upper limit value „Tolerance check counting“, input see
chap. 7.7.2
PCS L Lower limit value „Tolerance check counting“, input see
chap. 7.7.2
BEEP
no ok ng
STC
KFB/KFN-TM-BA_IA-e-1323 43
Press and the weighing value will be
For automatic totalizing see chap. 7.9 This function is used to issue and add individual weighing values automatically to the summation memory on unloading of weighing scale.
ASK
For remote control commands, see chap. 10.4
wireless
Not documented
baud
Available Baudrate: 600, 1200, 2400, 4800, 9600
Pr 7E1
7 bits, even parity
7o1
7 bits, odd parity
8n1
8 bits, no parity
PtYPE tPUP
Standard printer setting
LP50
Not documented
Lab
Lab x
Prt
Prt x
Lang Eng*
Chn
P2 mode
SiGr
Single-range balance
COUNT
Display internal resolution
DECI
Position of the decimal dot
Div.
Readability [d] / verification value[s]
CAP
Balance capacity [Max]
noLin
Adjustment, see chap. 6.9
LinEr
Linearisation, see chap. 6.10
GrA
Not documented
dUAL 1
Dual range balance
scales will remain in 2nd range.
COUNT
Display internal resolution
DECI
Position of the decimal dot
Readability [d] / verification value [e]
1. weighing range
Readability [d] / verification value [e]
2. weighing range
Weighing scale capacity [max]
1. Weighing range
Weighing scale capacity [max]
2. Weighing range
noLin
Adjustment, see chap. 6.9
LinEr
For linearization, see chap. 6.10
GrA
Not documented
Konfigurations­daten
Details see following table 1
Standard setting English
CAL
Balance with two weighing ranges and different maximum load and weighing ranges and interval sizes but only one load­supporting pan, whereby each range extends from zero to the respective maximum capacity. When load is removed, weighing
div 1
div.
div 2 CAP 1
CAP
CAP 2
CAL
44 KFB/KFN-TM-BA_IA-e-1323
dUAL 2
Multi-interval balance
changed during loading and unloading.
COUNT
Display internal resolution
DECI
Position of the decimal dot
Readability [d] / verification value [e]
1. weighing range
Readability [d] / verification value [e]
2. weighing range
Weighing scale capacity [max]
1. Weighing range
Weighing scale capacity [max]
2. Weighing range
noLin
Adjustment, see chap. 0
LinEr
Linearisation, see chap. 6.10
GrA
Not documented
P3 OTH
on
Keyboard lock enabled
off
Keyboard lock disabled
on
Animal weighing enabled
off
Animal weighing disabled
P4 tAr
Confirm input by .
P5 St
Follow up tare
St on
Follow up tare switched on
St off
Follow up tare switched off
P6 SP
7.5, 15, 30
Not documented
Lab
**********************
**********************
**********************
**********************
GS / GW
Gross weight
NO
Number weighing processes
NT
Net weight
TOTAL
Total of all individual weighings
TW
Tare weight
Weighing scales with one weighing range subdivided into partial weighing ranges, each providing a different scale interval. The scale interval depends on the applied load and is automatically
div 1
div.
div 2 CAP 1
CAP
CAP 2
CAL
s. Kap. 7.10 / 7.11
ANM
LOCK
Restricted taring range
Press , the current setting will be displayed. Using the navigation buttons (see chap. 2.1.1) select the desired setting, the active digit is flashing.
Tab. 1. Printout examples Standard printer
Prt
0~3
******************* GS: 5.000kg *******************
0 1 2 3
******************* NT: 5.000kg TW: 5.000kg GW: 10.000kg *******************
********************** GS: 5.000kg TOTAL: 10.000kg **********************
NT: 5.000kg TW: 5.000kg GW: 10.000kg TOTAL: 10.000kg
4~7
******************* No.: 1 GS: 5.000kg *******************
KFB/KFN-TM-BA_IA-e-1323 45
******************* No.: 1 NT: 5.000kg TW: 5.000kg GW: 10.000kg *******************
********************** No.: 1 GS: 5.000kg TOTAL: 10.000kg **********************
No.: 1 NT: 5.000kg TW: 5.000kg GW: 10.000kg TOTAL: 10.000kg
Value outside the A/D changer
9 Service, maintenance, disposal
9.1 Clean
Before cleaning, disconnect the appliance from the operating voltage.
Do not use aggressive detergents (solvents or similar).
9.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.
9.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
9.4 Error messages
Error
message
Description Possible causes
- - - - -
Maximum load exceeded
- - ol - -
Err 1 Err 2 Err 4 Zeroing range exceeded due
Err 5 Keyboard error Err 6
Incorrect data input Incorrect time entry
to switching-on balance or pressing (normally 4%
max)
range
Unload weighing system or reduce preload.
Follow format “yy:mm:dd“
Follow format “hh:mm:ss“
Object on the weighing plate
Overload when zeroing
Weighing plate not installed
Damaged weighing cell
Err 9 Stability display does not
appear
46 KFB/KFN-TM-BA_IA-e-1323
Damaged electronics
Check the environmental
conditions.
Err 10 Communication error Err 15 Gravitation error Err 17 Taring range exceeded Fai l h /
Fai l l Err P Ba lo /
Lo ba
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
Adjustment error
Printer error
Battery very low
No data
Range 0.9 ~ 1.0
Reduce load
Repeat adjustment.
Check communication paramet er s
Recharge battery
KFB/KFN-TM-BA_IA-e-1323 47
ST, GS 1.000kg
10 Da t a out put RS 2 32C
You can print weighing data automatically via the RS 232C interface or manually by pressing via the interface according to the setting in the menu.
This data exchange is asynchronous using ASCII - Code. The following conditions must be met to provide successful communication between
the weighing system and the printer.
Use a suitable cable to connect the display unit to the interface of the printer. Faultless operation requires an adequate KERN interface cable.
Communication parameters (baud rate, bits and parity) of display unit and printer must match. For a detailed description of interface parameters see chap. 8, menu block „P1 COM“ or ‚“P2 COM“
10.1 Technic al da ta
Connection 9 pin d-subminiature bushing
Pin 2 input Pin 3 output Pin 5 signal earth
Baud rate Optional 600/1200/2400/4800/9600 Parity 8 bits, no parity / 7 bits, even parity / 7 bits, odd parity
10.2 Printer mode
Printout examples (KERN YKB-01N):
Weighing
48 KFB/KFN-TM-BA_IA-e-1323
ST
Stable value
US
Instable value
GS / GW
Gross weight
NT
Net weight
TW
Tare weight
NO
Number weighing processes
TOTAL
Total of all individual weighings
<lf>
Space line
<lf>
Space line
ST,GS 1.000KG
****************************
Symbols:
Counting
PCS 100 ****************************
10.3 Output log (continuous output)
Weighing
HEADER1 : ST=STABLE, US=UNSTABLE HEADER2: NT=NET, GS=GROSS
10.4 Remote control instructions
Command Function Printout examples
S Stable weighing value for the weight is
sent via the RS232 interface
W Weighing value for the weight (stabl e or
ST,GS 1.000KG
US,GS 1.342KG
unstable) is sent via the RS232 interface
T No data are sent, the balance carries out
the tare function.
Z No data are sent, the zero-display
appears.
P Quantity will be sent via the RS232-
interface
KFB/KFN-TM-BA_IA-e-1323 49
-
-
10PCS
The displayed weight does not glow.
Mains power supply interrupted (mains cable defective).
Power supply interrupted.
(Rechargeable) batteries are inserted incorrectly or
empty
No (rechargeable) batteries inserted.
The displayed weight is
Table/floor vibrations
Weighing pan has contact with other objects.
possible)
The weighing result is obviously incorrect
Adjustment is no longer correct.
Great fluctuations in temperature.
possible)
11 Instant help
In case of an error in the program process, briefly turn off the display unit and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The display unit is not switched on.
permanently changing
Should other error messages occur, switch display unit off and then on again. If the error message remains inform manufacturer.
Draught/air movement
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
The display of the balance is not at zero
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
50 KFB/KFN-TM-BA_IA-e-1323
12 Installing displa y unit / weighi ng bridge
Installation / configuration of a weighing system must be carried out
12.1 Technic al da ta
Supply voltage: 5 V/150mA Max. signal voltage 0-10 mV Zeroing range 0-2 mV Sensitivity 2-3 mV/V Resistance parameter 80 - 100 Ω, max 4 items per 350 Ω load cell
by a well acquainted specialist with the workings of weighing balances.
12.2 Weighing system design
The display unit is suitable for connection to any analogue platform in compliance with the required specifications. The following data must be established before selecting a weighing cell:
Weighing balance capacity This usually corresponds to the heaviest load to be weighed.
Preload This corresponds to the total weight of all parts that are to be placed on the weighing cell such as upper part of platform, weighing pan etc.
Total zero setting range This is composed of the start-up zero setting range (± 2%) and the zero setting range available to the user via the ZERO-key (2%). The total zer o setting range equals therefore 4 % of the scale’s capacity.
The addition of weighing scales capacity, preload and the tot al z er o sett ing range give the required capacity for the weighing cell. To avoid overloading of the weighing cell, include an additional safety margin.
Smallest desired display division
Verifiability, if required
The application of the display unit as a verified weighing system requires that you short-circuit the two contacts [K1] of the circuit board, using a jumper; for position see cha p. 6.11. Remove the jumper for weighing systems not able to be verified.
KFB/KFN-TM-BA_IA-e-1323 51
PIN
Loadcell
6- conductor
4- conductor
7
EXC+
6
SEN+
5
EXC-
4
SEN-
3
SHIELD
SHIELD
2
SIG-
SIG-
1
SIG+
SIG+
EXC+
SEN+
SEN-
EXC-
SIG+
SIG-
SHIELD
7 6 3 2 5 4 1
LOAD CELL
12.3 How to connect the platform
Disconnect the display unit from the power supply. Solder the individual leads of the load cell cable onto the circuit board. See
diagrams below.
EXC+
52 KFB/KFN-TM-BA_IA-e-1323
EXC-
12.4 Configure displ ay unit
12.4.1 Verified weighing systems (contacts of circuit board [K1] short-circuited by means of jumper)
For menu overview see chap. 8.2. In verified weighing systems the menu item for calibration „P2 mode“ is blocked.
KERN KFB-TM:
To disable the access lock, destroy the seal and actuate the adjustment switch. Position of the adjustment switch see chap. 6.11
KERN KFN-TM: To override the blocked access you will have to destroy the seal before calling up the
menu and to short-circuit the two contacts on the circuit board [K2], using a jumper (See chap. 6.11).
Attention: After destruction of the seal the weighing system must be re-veri fied by an authorised agency and a new verification wire/seal mark fitted before it can be reused for applications subject to verification.
Call up menu:
Switch-on balance and during the selftest press . Press , , subsequently, the first menu block
„PO CHK“ will be displayed.
Press repeatedly until „P2 mode“ will be displayed. Operate the adjustment switch (models KFB-TM).
Press and use to select the weighing scales type.
Single-range balance
Dual range balance Multi-interval balance
KFB/KFN-TM-BA_IA-e-1323 53
Example single range scales (d = 10 g, max. 30 kg)
Confirm selected weighing scales type by pressing ; the
first menu item „COUNT“ will be shown.
1. Display internal resolution
Press , the internal resolution will be shown.
Return to menu by .
Press to select the next menu item.
2. Position decimal point
Press , the currently set position of the decimal dot is
displayed.
Press to select the desired setting.
Options 0, 0.0, 0.00, 0.000, 0.0000.
Confirm input by .
Press to select the next menu item.
3. Readability
Press and current setting will be displayed.
Select desired setting by . Options 1, 2, 5, 10, 20, 50.
Confirm entry by .
Press to select the next menu item.
54 KFB/KFN-TM-BA_IA-e-1323
6.10.1 for linearisation
4. Capacity
Press , the current setting will be displayed.
Using the navigation buttons (see chap. 2.1.1) select the desired setting, the active digit is flashing.
Confirm input by .
Press to select the next menu item.
5. Adjustment / linearization Adjustment or linearization is required after entering configuration data. For carrying out adjustment see chap. 6.9.1/step 6 or chap.
KFB/KFN-TM-BA_IA-e-1323 55
Example dual range scales (d = 2 / 5 g, max. 6 / 15 kg)
Confirm selected weighing scales type by ; the first
menu item „COUNT“ will be shown.
1. Display internal resolution
Press , the internal resolution will be shown.
Return to menu by .
Press to select the next menu item.
2. Position decimal point
Press , the currently set position of the decimal dot is
displayed.
Use to select the desired setting.
Options 0, 0.0, 0.00, 0.000, 0.0000.
Confirm input by .
Press to select the next menu item.
56 KFB/KFN-TM-BA_IA-e-1323
3. Readability
Press , the display used to enter readability/verification
value for first weighing range will appear.
Press , the current setting will be displayed.
Select desired setting with and acknowledge by .
Press to enter the next menu item for
readability/verification value for second weighing range.
Press and current setting will be displayed.
Select desired setting w i th and acknowledge by .
Press , the unit will return to the menu
Press to select the next menu item.
KFB/KFN-TM-BA_IA-e-1323 57
6.10.1 for linearisation
4. Capacity
Press and the display for entering the capacity for the
first weighing range will appear.
Press and current setting will be displayed.
Select desired setting with and acknowledge by .
Press to select the next menu item used to enter the
capacity for the second weighing range.
Press and current setting will be displayed.
Select desired setting with and acknowledge by .
Press , the unit will return to the menu
Use toselect next menu item.
5. Adjustment / linearization Adjustment or linearization is required after entering configuration data. For carrying out adjustment see chap. 6.9.1/step 6 or chap.
Acknowledge using , the current setting is displayed. Acknowledge by , select desired setting with
= Adjustment
= Linearisation
58 KFB/KFN-TM-BA_IA-e-1323
Press repeatedly to exit menu.
12.4.2 Non verifiable weighing systems (contacts of circuit board [K1] not short-circuited )
+ For menu overview see chap. 8.1.
Call up menu
Switch-on balance and during the selftest press .
Press , , subsequently , the first menu block
„PO CHK“ will be displayed.
Press repeatedly until „CAL“ will be displayed.
Press , the first menu item „COUNT“ will be displayed.
Navigation in the menu
With help of , the individual menu items can be
selected one after the other.
Confirm selected menu item by pressing . The current
setting will be displayed.
To change to the available settings, press the navigations
keys as described in chap. 2.1.1.
Either save by pressing or cancel by pressing .
KFB/KFN-TM-BA_IA-e-1323 59
Parameter selection
1. Display internal resolution
Press , the internal resolution will be shown.
Return to menu by .
Use to select another menu item.
2. Position decimal point
Press , the currently set position of the decimal dot is
displayed.
To make changes using the navigation keys (See chap.
2.1.1), select the desired setting. Options 0, 0.0, 0.00,
0.000, 0.0000.
Confirm input by .
Use to select another menu item.
3. Weighing scales type, capacity and readability
Press and current setting will be displayed.
Select desired setting by .
„off“ Single-range balance „on“ Dual range balance
Press to confirm, the display for entering readability (for
dual range scales for the first weighing range) appears.
Press , the current setting will be displayed.
60 KFB/KFN-TM-BA_IA-e-1323
the current setting will be shown (such as max. =
Select desired setting with and acknowledge by .
Press , the display for entering capacity will appear (at
dual range balance for the first ra nge).
Press ,
2000kg).
Using the navigation buttons (see chap. 2.1.1) select the
desired setting, the active digit is flashing.
Acknowledge with .
In a single-range balance the entry of capacity / readability is finished.
either in single-range balance
Press , the unit will return to the menu Press to call
up next menu item „CAL“.
or
In a dual range balance enter readability/verification value and capacity of the second weighing range.
Press , the display for entering the capacity of the
second weighing range will appear.
Press , the current setting will be displayed. Using the navigation buttons (see chap. 2.1.1) select the
desired setting, the active digit is flashing.
Confirm input by .
KFB/KFN-TM-BA_IA-e-1323 61
6.10.2 for linearisation
Press , the display for entering the readability of the
second weighing range will appear.
Press , the current setting will be displayed.
Select desired setting w i th and acknowledge by .
Press , the unit will return to the menu
Press to call next menu item.
4. Adjustment or linearisation Adjustment or linearisation is required after entering configuration data. For carrying out adjustment see chap. 6.9.2/step 4 or chap.
Acknowledge using , the current setting is displayed. Press to confirm, press to select the desired
setting noLin = Adjustment LineAr = Linearisation
62 KFB/KFN-TM-BA_IA-e-1323
KERN & Sohn GmbH
D-72322 Balingen-
sohn.de
Phone: 0049-[0]7433- 9933-
D
Konformitäts­erklärung
Wir erklären hierm it, dass das Produkt, auf das sich diese Erk lärung bezieht, mit den nachstehenden Normen übereinstimmt.
EN
Declaration of conformity
We hereby declare that the product to which this dec laration refers conforms to the following standards.
CZ
Prohlášení o shode
Tímto prohlašujem e, že v ýrobek , k terého se toto proh lášení týká , je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela res ponsabilité que le produit, auquel se r apporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformità
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami.
RUS
Заявление о
соответствии
Мы заявляем, что продукт, к которому относится данная декларация,
соответствует перечисленным ниже нормам.
EU Directive
Standards
2004/108/EC
EN55022: 2006 A1:2007 EN55024: 1998+A1:2001+A2:2003
2006/95/EC
EN 60950-1:2006 EN 60065:2002+A1:2006
Signature
Datum
08.04.2013
Ort der Ausstellung
Place of issue
72336 Balingen
Albert Sauter KERN & Sohn GmbH
Geschäftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com
13 De c laration of Conformity / Tes t Ce rt ificate
Frommern Postbox 4052 E-Mail: info@kern-
0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Declaration of conformity
EC Declaration of Conformity EC-Declaration of -Conformity EC- Déclaration de conformité EC-Declaración de Conformidad EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
Electronic Balances: KERN KFB-TM, KFN-TM, BFB, BFN, IFB, NFB, SFB, UFA, UFB, UFN
EN61000-3-3:1995+A1:2001+A2:2005
Date
KFB/KFN-TM-BA_IA-e-1323 63
Signatur
TEST CERTIFICATE
No. DK0199-R76-11.04
Instrument type KFN-TM / KFB-TM
Test item device Non-automatic Weighing Indicator
Issued by DELTA Danish Electronics, Light & Acoustics
In accordance with Paragraph 8.1 of the European Standard on metrological aspects
Fractional factor (pi) 0.5 (refer to 3.5.4 of the standard).
Issued to Kern & Sohn GmbH
Manufacturer Kern & Sohn GmbH
In respect of A family of indicators tested as a module of a weighing instrument.
Characteristics Suitable as a non-automatic weighing instrument with the following
Description and The A/D device is described and documented in the annex to documentation this certificate.
Remarks Summary of tests involved: See test report no. DANAK-
This test certificate cannot be quoted in an EU type approval certificate without permis­sion from the holder of the certificate mentioned above.
The annex comprises 7 pages.
Issued on 2011-03-16 Signatory: J. Hovgård
EU - Notified Body No. 0199
of non-automatic weighing instruments EN 45501:1992.
Ziegelei 1 D 72336 Balingen-Frommern GERMANY
characteristics:
Self indicating with single-interval, multi-interval or
multi-range Accuracy class III or IIII Verification scale interval: e
= Maxi/ni
i
Maximum number of verification scale intervals: n = 6000 for single-interval
n = 2×3000 for multi-interval and
multi-range,
however limited to 1000 for Class IIII Min. input voltage per VSI: 1 µV The essential characteristics are described in the annex.
1910568, DANAK-1910388 and NMi 709226.
DELTA
Danish Electronics,
Light & Acoustics
Venlighedsvej 4
2970 Hørsholm
Denmark
Tel. (+45) 72 19 40 00
Fax (+45) 72 19 40 01
www.delta.dk
VAT No. DK 12275110
Annex page 1 of 7 Annex to Test Certificate No. DK0199-R76-11.04
1. Name and type of instrument
The indicators KFN-TM / KFB-TM are a family of weighing indicators suitable to be incorporated in non-automatic weighing instruments, class III or class IIII, with single-interval, multi-interval or multi-range.
2. Description of the construction and function
2.1 Construction
The electronic indicator consists of a single circuit board, SMD populated on both sides as the A/D­interface circuits, the microprocessor and the voltage regulation are placed on one side and the LCD display on the other side.
The LCD-display has indication for: Stable, zero, gross, net, tare, and weight unit (kg, g, t), and 5½ digits with a height of 52 mm.
The enclosure is made of stainless steel for the KFN-TM indicator or of ABS plastics for KFB-TM.
The front of the enclosure has an on/off key plus 6 keys for operating the functions of the indicator.
All instrument calibration and metrological setup data are stored in the non-volatile memory.
The indicators are power supplied with 9 - 12 VDC - normally supplied by external 230 VAC to 9 - 12 VDC adapter. An optional internal battery can be factory installed.
As part of the indicators EMC protection ferrites shall be placed as follows:
Externally around the DC supply cable near its connection to the indicator (min. 1 turn).
Internal on cable between power plug and main board (4 turns).
Internal on cable between load cell connector and main board (min. 2 turns).
Software
The software version is displayed during the start-up of the indicator. The tested software version is 1.07.
Sealing
The configuration and calibration data can only be changed if the calibration jumper is installed on the circuit board.
2.2 Function
The devices are a microprocessor based electronic weighing indicators for connection of strain gauge load cells.
List of devices:
Self test
Determination and indication of stable equilibrium
Initial zero-setting ± 10% of Max
Semi-automatic zero-setting ± 2% of Max
Automatic zero-tracking ± 2% of Max
Issued by DELTA
Annex page 2 of 7 Annex to Test Certificate No. DK0199-R76-11.04
Indication of zero
Semi-automatic subtractive tare
Acting upon significant fault
Weighing unstable samples
Real time clock (optional)
3. Technical data
3.1 Indicator
Type KFN-TM / KFB-TM
Accuracy class III or IIII
Weighing range Single-interval, multi-interval or multi-range
Maximum number of verification scale intervals (n) 6000 for single-interval
2×3000 for multi-interval and multi-range, however limited to 1000 for Class IIII
Minimum input voltage per VSI 1 µV
Maximum capacity of interval or range (Max
Verification scale interval, e
= Maxi / ni
i
Initial zero-setting range: ± 10 % of Max
Maximum tare effect: 100 % of Max
Fractional factor (pi) 0.5
Excitation voltage 5 VDC
Circuit for remote sense Active, (see below)
Minimum input impedance 87 ohm
Maximum input impedance 1600 ohm
Connecting cable to load cell(s): See Section 3.1.1
Supply voltage: 9 - 12 VDC
Operating temperature range Min / Max = -10 °C / +40 °C
Peripheral interface(s) See Section 4
): ni × ei
i
230 VAC using external Vac/2Vdc adapter
3.1.1 Connecting cable between the indicator and the junction box for load cell(s), if any
3.1.1.1 4-wire system
Line 4 wires, shielded Maximum length The certified length of the load cell cable, which shall be con-
nected directly to the indicator.
3.1.1.2 6-wire system
Line 6 wires, shielded
2
Maximum length 227 m/mm Maximum resistance per wire 3.8 ohm
Issued by DELTA
Annex page 3 of 7 Annex to Test Certificate No. DK0199-R76-11.04
4. Interfaces
4.1 Load cell interface
Refer to section 3.1.1.
Any load cell(s) can be used for instruments under this certificate provided the following conditions are met:
There is a respective test certificate (EN 45501) or an OIML Certificate of Conformity (R60)
issued for the load cell by a Notified Body responsible for type examination under the Directive 2009/23/EC.
The certificate contains the load cell types and the necessary load cell data required for the
manufacturer’s declaration of compatibility of modules (WELMEC 2, Issue 5, 2009, section
11), and any particular installation requirements. A load cell marked NH is allowed only if humidity testing to EN 45501 has been performed.
The compatibility of load cells and indicator is established by the manufacturer by means of
the compatibility of modules form, contained in the above WELMEC 2 document, or the like, at the time of EC verification or declaration of EC conformity of type.
The load transmission must conform to one of the examples shown in the WELMEC 2.4 Guide
for load cells.
4.2 Peripheral interfaces
The indicator may be equipped with one or more of the following protective interfaces that have not to be secured.
RS-232C
Analogue output (0 - 10 V / 4 - 20 mA)
Digital output
Blue Tooth
The peripheral interfaces are characterised “Protective interfaces” according to paragraph 8.4 in the Directive.
5. Conditions for use
Legal use of the indicator for automatic weighing or as counting device is not allowed with reference to this test certificate.
Issued by DELTA
Annex page 4 of 7 Annex to Test Certificate No. DK0199-R76-11.04
6. Location of seals and inscriptions
Seals shall bear the verification mark of a notified body or alternative mark of the manufacturer ac­cording to ANNEX II, section 2.3 of the Directive 2009/23/EC. The seals shall be placed so that the enclosure can not be opened.
Location of CE mark of conformity:
The CE mark of conformity is placed on the overlay on the rear side of the device.
Inscription on the overlay:
Type, accuracy class, Temp. -10 °C / +40 °C, Certificate No. DK0199-R76-11.04.
Other inscriptions on the overlay:
Manufacturer’s name and/or logo, Part No, Supply voltage.
7. Tests
The indicator has been tested according to EN 45501 and WELMEC 2.1 Guide for testing of indicators.
Examination / tests
Temperature tests: 20 / 40 / -10 / 5 / 20 (tested at minimum input-voltage sensitivity)
Temperature effect on no-load indication (tested at minimum input-voltage sensitivity)
Stability of equilibrium
Repeatability
Warm-up time
Voltage variations
Short time power reductions
Electrical bursts
Electrostatic discharges
Immunity to radiated electromagnetic fields
Damp heat, steady state
Span stability
Checklist
Maximum load cell cable length and impedance of cable to load cell
Load cell interface measurements with interruptions of the sense circuit
The test item fulfilled the maximum permissible errors at all tests.
Issued by DELTA
Annex page 5 of 7 Annex to Test Certificate No. DK0199-R76-11.04
8. Documentation
Contents of the technical documentation held by the notified body:
8.1 Product specification
Manuals and descriptions
Drawings
Etc.
8.2 Examination report
OIML R76 report no. DANAK-1910568, DANAK-1910388 and NMi 709226.
8.3 Test results
Report no. DANAK-1910568, DANAK-1910388 and NMi 709226.
Issued by DELTA
Annex page 6 of 7 Annex to Test Certificate No. DK0199-R76-11.04
9. Pictures
Figure 1 Sealing of KFN-TM.
a self destroyable sticker/seal shall be used.
After remove the label, you will find VOID on housing, or
Issued by DELTA
Annex page 7 of 7 Annex to Test Certificate No. DK0199-R76-11.04
Issued by DELTA
Figure 2 Sealing of KFB-TM.