1 Technical data
Models „FOB-NS“:
KERN FOB 0.5K-4NS FOB 5K-3NS
Readability (d) 0.1 g 1 g
Weighing range (max) 500 g 5000 g
Reproducibility 0.1g 1 g
Linearity ± 0.2 g ± 2 g
Weighing Units g, lb, oz, ozt, dwt
Recommended adjustment
weight, not added (class)
Warm-up time 10 min
Stabilization time (typical) 2 sec.
Operating temperature + 10° C .... + 35° C
Humidity of air 25% - 95% (non-condensing)
Housing (B x D x H) mm 170 x 150 x 40
Weighing plate, stainless steel
(mm)
Weight kg (net) 650 g
Auto off Options: 2, 3, 4, 5, min; off
Battery 9 V block
Secondary voltage powerpack 9 V / 100 mA
IP protection IP65
500 g (M1) 5000 g (M1)
120 x 150 x 10
4 FOB-N-BA-e-1410
Models „FOB-NL“:
KERN FOB 7K-4NL FOB 10K-3NL FOB 30K-3NL
Readability (d) 0.5 g / 1 g 1 g / 2 g 2 g / 5 g
Weighing range (max) 5 kg / 7.5 kg 8 kg / 15 kg 16 kg / 30 kg
Reproducibility 0.5 g / 1 g 1 g / 2 g 2 g / 5 g
Linearity ± 1.5 g / 3 g ± 3 g / 6 g ± 6 g / 15 g
Weighing Units g, lb, oz, ozt, dwt
Recommended adjustment
weight, not added (class)
Warm-up time 30 min 30 min 30 min
Stabilization time (typical) 2 sec.
Operating temperature + 5° C …. + 35° C
Humidity of air 25% - 95% (non-condensing)
Housing (B x D x H) mm 285 x 255 x 90
Weighing plate, stainless steel
(mm)
Weight kg (net) 3611 g
Auto off 2 min
Battery 4 x 1,5 V AA
Secondary voltage powerpack 9 V / 100 mA
IP protection IP67
5 kg (M1) 10 kg (M1) 30 kg (M1)
252 x 200 x 14
FOB-N-BA-e-1410 5
2 Basic Information (General)
2.1 Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic balance”, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
plate. As soon as a stable weighing value is reached the weighing value can be read.
2.2 Improper Use
Do not use balance for dynamic weighing. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“. (Example: Slowly draining fluids from a
container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the measuring
system.
Impacts and overloadi ng exceeding the stated maximum load (max) of the balance,
minus a possibly existing tare load, must be strictly avoided. Balance may be
damage by this.
Never operate balance in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-related faul ts and des tr uct ion of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
2.3 Warranty
Warranty claims shall be voided in case
• Our conditions in the operation manual are ignored
• The appliance is used outside the described uses
• The appliance is modified or opened
• mechanical damage and damage caused by media, liquids,
natural wear and tear
• The appliance is improperly set up or incorrectly electrically connected
• The measuring system is overloaded
6 FOB-N-BA-e-1410
2.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this test.
Information is available on KERN’s home page (www.kern-sohn.com) with regard to
the monitoring of balance test substances and the test weights required for this. In
KERN's accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.
3 Basic Safety Precautions
3.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN
balances.
All language versions contain a non-binding translation.
The original German is binding.
3.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
4 T ransport and storage
4.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
4.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing plate,
power unit etc. against shifting and damage.
FOB-N-BA-e-1410 7
5 Unpacking, Setup and Commi s si oning
5.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
• Place the balance on a firm, level surface;
• Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
• Protect the balance against direct draughts due to open windows and doors;
• Avoid jarring during weighing;
• Protect the balance against high humidity, vapours and dust;
• Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (co ndensation of air humidity on the appliance)
may occur if a cold appliance is taken to a considerably warmer environment.
In this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
• Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change location or remove
source of interference.
5.2 Unpacking
Carefully remove the weighing balance from the packaging and install it at the
designated work place.
8 FOB-N-BA-e-1410
• Balance
FOB-NS
FOB-NL
5.2.1 Placing
The balance must be installed in a way that the weighing plate is exactly in horizontal
position.
5.2.2 Scope of delivery
Serial accessories:
• Weighing plate
• Mains adapter (optional)
• Protective cover
• Operating manual
5.3 Mains connection (optional)
The balance may be operated via the optional mains adapter. The stated voltage
value must be the same as the local voltage.
Only use original KERN mains adapter s . Using other makes requires consent by
KERN.
5.4 Battery operation
Unscrew battery cover on the bottom of the housing (see figure 1).
• 9 V block battery (FOB-NS) or
• 4 x 1.5V AA batteries (FOB-NL)
Put battery cover back in place and tighten it.
Fig. 1:
FOB-N-BA-e-1410 9
When you tighten the battery cover, make sure that not too much pressure
is put on the weighing plate. The load cell may be damaged.