KERN & SOHN FKTF User Manual

KERN & Sohn GmbH
Ziegelei 1
E-Mail: info@kern-sohn.com
Internet: www.kern-sohn.com
Operating instructions
Finished packaging regulations scale
KERN FKTF
Version 1.1
01/2015
GB
D-72336 Balingen
Fax: +49-[0]7433-9933-149
FKTF-BA-e-1511
KERN FKTF
Finished packaging regulations scale
GB
Version 1.1 01/2015
Operating Instructions
Contents
1 Technical Data ........................................................................................................... 3
2 Basic Information (General) ..................................................................................... 7
2.1 Proper use .............................................................................................................................................. 7
2.2 Improper Use ......................................................................................................................................... 7
2.3 Warranty ................................................................................................................................................. 7
2.4 Monitoring of Test Resources ............................................................................................................. 7
3 Basic Safety Precautions ......................................................................................... 8
3.1 Pay attention to the instructions in the Operation Manual ............................................................... 8
3.2 Personnel training ................................................................................................................................. 8
4 Transportation & Storage ......................................................................................... 8
4.1 Testing upon acceptance ..................................................................................................................... 8
4.2 Packaging / return transport ................................................................................................................ 8
5 Unpacking, Setup and Commissioning ................................................................... 9
5.1 Installation Site, Location of Use ......................................................................................................... 9
5.2 Unpacking .............................................................................................................................................. 9
5.3 Mains connection .................................................................................................................................. 9
5.4 Connection of peripheral devices ........................................................................................................ 9
5.5 Initial Commissioning ........................................................................................................................... 9
5.6 Adjustment ........................................................................................................................................... 10
5.7 Adjustment ........................................................................................................................................... 10
5.8 Verification ........................................................................................................................................... 10
6 Operation ................................................................................................................. 11
6.1 Start the scale and turn on the screen .............................................................................................. 11
6.2 Select Operator/Article ........................................................................................................................ 15
6.3 Start the measurement ....................................................................................................................... 16
6.4 Creating products with desired specificatio ns ................................................................................ 18
6.5 Activate operator ................................................................................................................................. 20
6.6 Operating mode settings .................................................................................................................... 20
7 Data output RS 232 C .............................................................................................. 22
7.1 Interface RS 232C ................................................................................................................................ 22
7.1.1 Explanation of the data transfer ............................................................................................................ 23
8 Service, maintenance, disposal ............................................................................. 23
8.1 Cleaning ............................................................................................................................................... 23
8.2 Service, maintenance .......................................................................................................................... 23
8.3 Disposal ................................................................................................................................................ 23
9 Instant help .............................................................................................................. 24
10 Declaration of conformity ....................................................................................... 25
2 FKTF-BA-e-1511
Taring range (subtractive)
Stabilization time (typical)
Housing (B x D x H) mm
Weighing plat e stain­less steel mm
Battery operation with 6 x 1.5 V, size "C"

1 Technical Data

KERN FKTF 3K0.5LM FKTF 6K1LM FKTF 12K2LM
Accuracy class III III III Readability (d) 0.5 g 1 g 2 g
Verification value (e) 0.5 g 1 g 2 g Weighing r ange (max) 3.000 g 6.000 g 12.000 g Minimum load (Min) 10 g 20 g 40 g
3.000 g 6.000 g 12.000 g Reproducibility 0,5 g 0,5 g 1 g Linearity ±0,5 g ±0,5 g ±1 g Adjustment points 1 / 2 / 3 kg 2 / 5 / 6 kg 2 / 5 / 10 / 12 kg
Recommended adjust­ing weight M1 (not
3 kg 5 kg 10 kg
supplied) Humidity of air max. 80% rel. (non-condensing)
3 sec.
Allowable ambient tem­perature
+10 °C ... + 40 °C
Warm-up time 2 h
350 x 390 x 120
Vibration filter yes
340 x 240 Units see menu
Weight kg (net) 6,5
Data interface yes (RS232)
Article regular places 400
Number of operators 40
FKTF-BA-e-1511 3
yes
Taring range (subtr ac­tive)
Stabilization time (typical)
Housing (B x D x H) mm
6 x 1.5 V, size "C"
KERN FKTF 30K5LM FKTF 60K10LM
Accuracy class III III Readability (d) 5 g 10 g
Verification value (e) 5 g 10 g Weighing r ange (max) 30.000 g 60.000 g Minimum load (Min) 100 g 200 g
30.000 g 60.000 g Reproducibility 2,5 g 5 g Linearity ±2,5 g ±5 g Adjustment points 10 / 20 / 30 kg 20 / 50 / 60 k g
Recommended adjust­ing weight M1 (not supplied)
Humidity of air max. 80% rel. (non-condensing)
Allowable ambient tem­perature
Warm-up time 2 h
Vibration filter yes Weighing plat e stain-
less steel mm Units see menu
Weight kg (net)
20 kg + 10 kg 50 kg
3 sec.
+10 °C ... + 40 °C
350 x 390 x 120
340 x 240
6,5
Data interface Article regular places Number of operators
Battery operation with
4 FKTF-BA-e-1511
yes (RS232)
400
40
yes
Nominal filling quan­tity Q in g or ml
Admissible minus deviation
5 to 50
9 - 50 to 100
-
4.5
100 to 200
4.5 - 200 to 300
- 9 300 to 500
3 - 500 to 1,000
-
15
1,000 to 10,000
1.5
-
Packaged Goods Regulation (PGR/FPVO). To weighing scales specially designed for packaged goods applies
§ 22 of the Packaged Goods Reg ulat ion PG R. This paragraph says: (1) Commercially produced, packaged g oods labelled by weight or volume of the same nom-
inal filling quantity must be produced in such as way that the filling quantity at the point of production
1. on average does not drop below the nominal filling quant it y and
2. does not exceed the values for negative deviation from t he nom inal filling quantity stipu­lated in section 3.
(2) Packaged goods labelled by weight or volume of the s am e nom inal weight mus t be brought into the territor y covered by this reg ulat ion only if t he filling quantity at the point of production
1. on average does not drop below the nominal filling quant it y and
2. does not exceed the values for negative deviation from t he nom inal filling quantity stipu­lated in section 3.
(3) The allowable minus deviations are as follows:
N
in % of QN in g or ml
When applying this table, any values for adm issible minus deviations calculated in weight or volume units that are stated in percent should always be rounded up to 0.1 gram or 0.1 milli­litre. Minus deviations should not exceed 2 per cent of the packaged goods.
(4) Packaged goods labelled by weight or volume of the same nom inal filling quantity may only be launched commercially if the minus deviation from the nominal filling quantity does not exceed double the amount of the values stipulated on the table under sec t ion 3.
These rules give the manufacturers of packaged goods a clear standard for production con­trol in manufacture and the consumer confidence that such products actually contain the filling quantity labelled on the package by the manufacturer.
FKTF-BA-e-1511 5
Nominal filling quantity QN
for packaged goods in g or ml
Maximum admissible verification value
in g
Less than 10
0.1
For 10 to less than 50
0.2
Nominal filling quantity QN
For packaged goods in g or ml
Maximum admissible verification value
in g
For 50 to less than 150
0.5
For 150 to less than 500
1.0
For 500 to less than 2500
2.0
For 2500 and more
5.0
Packaged goods manufactured and tested according to the rules stipulated by the Pack­aged Goods Regulation bear the following marking on their label:
Weighing scales to be used in packaged goods control must have type approval according to category III combined with a valid verification.
The choice of a non-automatic weighing scale should be on the one hand according to the required maximum weight and on the other hand the maximum admissible verification value (e) for the nominal filling quantity to be tested: According to the admissible maximum num­ber of errors for the verified weighed result (e):
Finally, it should be noted that shorter re-verification periods are applied in many European States for weighing scales used in package controlling as opposed to those for regular weighing scales of verification category III. Usually this is every year or every two years.
6 FKTF-BA-e-1511

2 Basic Information (General)

It is absolutely necessary that you read and understand the operating instructions prior to installation and commissioning and follow the instructions during the process!

2.1 Proper use

The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

2.2 Improper Use

Do not use balance for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance (Example: Slowly draining fluids from a container on the balance.). Do not leave permanent load on the weighing plate. This may damage the measuring sys­tem. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion pro­tected. The structure of the balance may not be modified. This may lead to incorrect weighing re­sults, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.

2.3 Warranty

Warranty claims shall be voided in case
- Our conditions in the operation manual are ignored
- The appliance is used outside the described uses
- The appliance is modified or opened
- mechanical damage and damage caused by media, liquids
- natural wear and tear
- The appliance is improperly set up or incorrectly electrically connected
- The measuring system is overloaded

2.4 Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must de­fine a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. Our accredited DKD calibration laboratory offers fast and inexpensive adjustment for test weights and weighing balances (reset to na­tional normal weight).
FKTF-BA-e-1511 7

3 Basic Safety Precautions

3.1 Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and commis­sioning, even if you are already familiar with KERN balances.
Versions in other languages are non-binding translations. The only binding version is the original document in German.

3.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

4 Transportation & Storage

4.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance it­self when unpacking for possible visible damage. In case of visible damage have the damage verified by the messenger’s signature. Do not alter goods or packaging and do not remove any parts of the delivery. Report the damage immediately (within 24 hours) in writing to the parcel service.

4.2 Packaging / return transport

Keep all parts of the original packaging for a possibly required return. Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile parts. Reattach possibly supplied transport securing devices. Secure all parts such as the glass wind screen, the weighing platform,
power unit etc. against shifting and damage.
8 FKTF-BA-e-1511

5 Unpacking, Setup and Commissioning

5.1 Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
- Place the balance on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing next to a
radiator or in the direct sunlight;
- Protect the balance against direct draughts due to open windows and doors;
- Avoid jarring during weighing;
- Protect the balance against high humidity, vapors and dust;
- Do not expose the device to extreme dampness for longer periods of time. Non-
permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, accli­matize the disconnected appliance for ca. 2 hours at room temperature.
- Avoid static charging of the material to be weighed, weighing container and wind-
shield.
If electro-magnetic fields or static charge occur, or if the power supply is unstable major de­viations on the display (incorrect weighing results) are possible. In that case, the location must be changed.

5.2 Unpacking

Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.

5.3 Mains connection

Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.

5.4 Connection of peripheral devices

Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use acces­sories and peripheral devices by KERN, as they are ideally tuned to your balance.

5.5 Initial Commissioning

A warming up time of 2 hours after switching on stabilizes the measuring values. The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter “Adjustment”.
FKTF-BA-e-1511 9

5.6 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle
- to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out during the initial start-up, after change in location and variation of surrounding temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.

5.7 Adjustment

With an adjustment weight, the weighing accuracy can be checked and re-adjusted at any time.
Attention: In the verified balances the adjustment is not possible. Procedure when a djusting:
Observe stable environmental conditions. A short warming up time of ca. 15 minutes is rec­ommended for stabilization.

5.8 Verification

General introduction: According to EU directive 90/384/EEC balances must be verified if they are used as follows
(legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the medical
and pharmaceutical laboratory. c) For official purposes d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard. Verification instructions An EU type approval exists for balances described in their technical data as verifiable. If a
balance is used where obligation to verify exists as described above, it must verified and re­verified in regular intervals. Re-verification of a balance is carried out according to the respective national regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
10 FKTF-BA-e-1511
After contact the Touches and

6 Operation

6.1 Start the scale and turn on the screen

Start scale by touching the touch panel
Start-up screen:
start up the scale the following screen appears.
To start the measurement it must necessarily be given the Article and the Operator.
Please press the reverse button.
FKTF-BA-e-1511 11
12 FKTF-BA-e-1511
FKTF-BA-e-1511 13
14 FKTF-BA-e-1511
After pressing the operator key se-
Now in the second information row
Select from the displayed article

6.2 Select Operator/Article

Select Operator:
lect fields will appear for authorized operators.
Select the current operator. If no operator is available, they must
be re-entered via MENU / Memory / Operator / Description.
Select Article:
the selected user appears (see ar­row).
Now press the ARTICLE button.
selection fields the right item by pressing.
Are no items available, they must be re-entered via MENU / Memory / Item / Item*.
FKTF-BA-e-1511 15
Now in the first line of information
To start the measurement, please
The weighing mode is displayed at

6.3 Start the measurement

the selected Item and the Item code appears (see arrows).
press STRT / MEAS key
The different types of measure­ment (average Tara, Tara previ­ous, Tara afterwards) can be se­lected in each article.
In this example, "Tara average"
the top right. Well the first target must be
placed.
16 FKTF-BA-e-1511
After the first measurement object
After the last measurement object
has been placed it will be confirmed with TAKE GROS. The balance jumps automatically to the second target.
At any time you can stop the measurement with QUIT MEAS.
OK = measured value is within the tolerance (between TU and TO) + = passed the upper limit (TO)
- = below the lower limit (TU)
the scale displays "Ready" and the series of measurements can be printed with PRNT END.
By pressing, the scale also jumps back to the start screen.
The measured values are stored for the daily and weekly statistics, if they are activated under MENU / Configuration / Setting control mode / Statistic.
FKTF-BA-e-1511 17
You can enter the specific values of
On the first page of the article the Here, the tolerance data as set-

6.4 Creating products with des i r ed specifications

each item under MENU / Memory / Item / Item*.
following data can be entered: Item: memory location number
Number: article number or code Machine: enter the machine number Tara: weighing type (Tara previous, Tara afterwards, Tara average) Description: product name Density: density value of the article
You can use the arrow key to scroll to the second page.
point, upper limit, lower limit, lower limit 1 and lower limit 2 can be en­tered.
Under "Quantity", the number of the tested measurement objects can be entered.
The arrow key can be used to scroll to the previous page or to the next page.
18 FKTF-BA-e-1511
The automatic limit value calculation is active if under MENU / Configu-
On page 3 of the item you will get
By clicking the "HOME" button you get back to
ration / Setting control mode / Toleranc e the "tolerance" is set to auto­matically and the verification switch is active. After entering a setpoint the ohter limits are automatically calculated according to the statutory FPVO regulation.
insight into the respective daily and weekly statistics. If desired, these can be printed.
Caution: The daily and weekly statistics are only visible if you have been acti­vated under MENU / Configuration / Setting control mode / Statistic.
With "Manual tare" you can enter the desired automatic tare value.
Generally:
the start-up screen.
The "back arrow" key allows one page back.
With "CLR" all product data will be deleted.
FKTF-BA-e-1511 19
Under MENU / Memory / Operator/
Please press on the start-up screen
Then select in the menu the "Con-

6.5 Activate operator

Activated you can activate the cre­ated operator, so that they are visi­ble on the start-up page under "OPERATOR" and can be selected.

6.6 Operating mode settings

the "MENU CAL" button.
figuration" button.
20 FKTF-BA-e-1511
Now, "Setting control mode" appears.
Under Setting control mode, then you
This also choose.
can adjust various settings:
Automatic:
Automatic mode can be enabled and disabled.
Tolerance: When "automatic" is ac-
tivated, the tolerances are automati­cally calculated according to the stat­utory FPVO regulation. The verification switch must then be activated in this case!
Histogram:
The histogram printout will be acti­vated and displayed in the respective product data.
Statistic:
The statistic printout will be activated and displayed in the respective prod­uct data.
Selection: Under Selection you can change the view of the article. With “list”, the Items in "ARTICLE" on the start-up screen are displayed with a field of activity. With "search" you can search the articles via the keyboard.
FKTF-BA-e-1511 21
5 1

7 Data output RS 232 C

Technical data:
8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bit, no parity bit Baud rate selectable from 2400, 4800, 9600 Baud (factory setting) and 19200 Baud. Sub-D plug 9-channel required For operation with interface faultless operation is only ensured with the correct KERN –
interface cable (max. 2m)
Pin allocation of the balance output socket (front view)
Sub-D jack 9-channel
Pin 2: Transmit data Pin 3: Receive data
9 6
Pin 5: Signal ground

7.1 Interface RS 232C Data output via interface RS 232C

General
The previous condition for the data transfer between balance and a peripheral device (e.g. printer, PC ...) is that the appliances are set to the same interface parameters (e.g. baud rate, parity ...).
22 FKTF-BA-e-1511
Bit.Nr.
1 2 3 4 5 6 7 8 9
10
11
12
13
14
15
16
17
18
19
20 N N N N N B B B B B B B B B B B 0 • 0 0
Bit.Nr.
21
22
23
24
25
26 E E E E CR
LF

7.1.1 Explanation of the data transfer

Each data transfer is structured as follows:
N = Numerator B*: = Blank or for autotare on in zero range. B, 0, ., g: = Blank or weighing value giving unit according to loading of the balance E = Unit CR: = Carriage Return LF: = Line Feed

8 Service, maintenance, disposal

8.1 Cleaning

Before cleaning, please disconnect the appliance from the operating voltage. Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth
dampened with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods m ust be removed immediately.

8.2 Service, maintenance

The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.

8.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to valid na­tional or regional law of the location where the appliance is used.
FKTF-BA-e-1511 23
The displayed weight does not glow.
The balance is not switched on.
rupted (mains cable not plugged in/faulty).
Power supply interrupted.
The displayed weight is permanently changing
Table/floor vibrations
Electromagnetic fields / static charging (choose possible)
The weighing value is obviously
The display of the balance is not at zero
wrong
Adjustment is no longer correct.
Electromagnetic fields / static charging (choose possible)

9 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The mains supply connection has been inter-
Draught/air movement
Weighing plate has contact with other objects.
different location/switch off interfering device if
Great fluctuations in temperature.
different location/switch off interfering device if
Should other error messages occur, switch balance off and then on again. If the error mes­sage remains inform manufacturer.
24 FKTF-BA-e-1511
KERN & Sohn GmbH
D-72322 Balingen-Frommern
E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0
Internet: www.kern-sohn.de
D
Konformitäts­erklärung
Wir er klären hiermit, dass das Produk t, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
GB
Declaration of conformity
We hereby declare that the product to which this declaration ref ers conforms with the following standards.
CZ
Prohlášení o shode
Tímto prohlašujem e, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela respons abilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferi­sce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja
zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy,
jest zgodny z poniższymi normami.
RUS
Заявление о
соответствии
Мы заявляем, что продукт, к которому относится данная декларация,
соответствует перечисленным ниже нормам.
EU Directive
Standards
2004/108/EC
EN 55022: 2010/AC:2011(Limit class B) EN
EN 45501: 1992-10+AC: 1993-08
2006/95/EC
EN 60950
2011/65/EU
EN 50581:2012
Signature
Datum
17.07.2014
Ort der Ausstellung
Place of issue
72336 Balingen
Albert Sauter KERN & Sohn GmbH
Geschäftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com

10 Declar a t ion of conformity

Postfach 4052
Fax: 0049-[0]7433-9933-149
Declaration of conformit y
EC-Konformitätserklärung EC-Declaration of Conformity
Déclaration de conformité EC-Declaración de Conformidad
EC­EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии

Electronic Balance: KERN FKTF

55011: 2009+A1:2010 (Limit class B) EN 61000-3-2: 2006-04+A1: 2009 + A2: 2009 EN 61000-3-3: 2008 EN 55024: 1998-09+A1: 2001-10+A2: 2003-10 OIML R 76-1:2006
Date
Signatur
FKTF-BA-e-1511 25