KERN & SOHN FH 500 EXT User Manual

Sauter GmbH
Ziegelei 1
Instruction Manual
FH 10– 500 EXT
LCD
Pilot lamp
RS 232 port
Force sensor
B
L
D-72336 Balingen E-Mail: info@sauter.eu
Thank you for buying a SAUTER force gauge. We hope you are pleased with your high quality force gauge with its big functional range. If you have any queries, wishes or helpful suggestions, do not hesitate to call our service number.
„Sensor Outside“ means the cell is outside the body.
1. Included in Delivery
- SAUTER FH, incl. rechargeable battery
- Carrying Case
- Charger
- Standard attachments, 5 pieces. M3x 8 screws
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Set
Dimensions in mm for the instruments FH 10 EXT up to FH 500 EXT:
Sensor
Outside
FH_10-500-BA-e-1213 1
Sauter GmbH
Instruction Manual
FH 10– 500 EXT
100 mm
H
Dimensions of the sensor L x B x H: 74 x 50 x 22 mm Cable length between display and sensor: 2 m
Ziegelei 1 D-72336 Balingen E-Mail: info@sauter.eu
4. Technical Data
- Accuracy: ± 0,5 % of capacity
- Data Sampling Rate: 2.000 Hz
- Weight: 640 g
5. Display
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Operation
(1) Measuring Result (2) Measuring Units (3) Activation of PRINT Function (4) Indication of power charging status
PEAK or AUTO-PEAK Mode (5) Average value of stored peak values (6) Force direction
External sensor with accessories for compression tests
Important annotation:
By pressing the RESET key (on the right side of housing, see illustration on the left), individual settings and memorised values can be re-set or erased, in example for a new start of the instrument after an operating error.
(7) Occupancy of storing spaces (8) AVERAGE- or Saving Mode
Operating keys
ON / OFF:
ON / OFF key
(For ON, press 1 sec.)
2. Working Conditions
10°C to 30°C / 15% up to 80% humidity
3. Electrical Power Supply
Either by recharchable battery or current power supply
Current power supply:
- Connection by power adapter
- Rechargeable batteries are charged simultaneously
Rechargeable battery pack for mobile applications:
- Type: Ni 8.4V / 600 mAh
- Charging time: approx. 1 hour
FH_10-500-BA-e-1213 2
UNIT:
- Press short: Select unit:
- Press for 2 sec.: Display return
ZERO:
Three functions:
- Zeros the measuring result (Tara function)
- Cleans the peak value (in Peak mode)
- Saves a setting (in SET mode
N, kg or lb
Sauter GmbH
Ziegelei 1
Instruction Manual
FH 10– 500 EXT
Hi Lt
Lo dT
Lo Pe
P.OFF
HoldT
rS232
StoP
D-72336 Balingen E-Mail: info@sauter.eu
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
SET:
- 1 x Press: Upper Limit [HidT / Hi Lt]. To
change press: ▲ or (see section High/Low Limit function)
- 1 x Press: Lower Limit [ LodT]. To change
press ▲ or ▼ (see s ection High/Low Limit function )
- 1 x Press: Minimum limit to save peak
values in the instrument’s memory. [LE.SET / Lo Pe]. Please see section „Min
Limit Peak Save“. To change press: ▲ or ▼ (Only in „Peak-Mode“ active) (see
section min. limit peak save)
- 1 x Press: Sets the Auto-STOP force value
that stops the movement of a test stand. Adjustment of the force value. To change press: ▲ or ▼ (only active in „Battery­Mode“)
- 1 x Press: Sets the AUTO-OFF force value
that stops the movement of a test stand. To change press: ▲ or ▼ (Only in „Battery­Mode“ active)
- 1 x Press: Peak-Freeze-Time [PE.2E /
A.PE / HoldT]: Time period in which a peak value is being shown in the displ ay in sec. To change press: ▲ or ▼
- 1 x Press: Data Transfer [rS232] to PC
(PC) or to the printer (Print) or (i n version U
5.1) to the test stand (stand)
- PC: Pressing SET once: Saves the Settings
- PRINT: Pressing SET twice: sending data to the printer
- STAND: Pressing SET three times: sending a signal to the Test Stand to stop its movement (in version U 5.1)
Back Light:
FH_10-500-BA-e-1213 3
PEAK:
Three functions
- Track mode (continuous measurement)
- Peak mode (capture of maximum values)
- Auto-Peak mode, same as Peak-mode, only
MEMORY:
DELETE
PRINT:
High / Low Limit Function
LEDs to display OK / NOT OK Tests
Lower than lower limt ▲ Higher than higher limit
This function allows efficient testing of OK / NOT OK measurements of similar or identical testing objects. A lower and an upper limit value can be defined. The instrument compares the individual m easuring results with these limit values and shows the OK or NOT OK result by green or red light diode and by sound. To set these limit values, please see the SET Menu in section „Operating keys“.
Measurement (Track Modus)
Display (1) shows the continuous force in a defined direction (6)
without the „Min limit peak save“ function
Saves the peak values to calculate the average value (please see section “Memory mode”
Deletes stored peak values (only active in „Memory mode“)
Sends the stored peak values to a PC or Printer (please see section „Data Output”)
To zero the display, press:
Peak-Hold Function (Peak Mode)
Please press:
Sauter GmbH
Ziegelei 1
Instruction Manual
FH 10– 500 EXT
Pin
Signal
Illustration
2
TxD
Output signal
3
RxD
Input signal
5
GND
Ground
6
+1.6 to + 2 V
Over upper limit
7
+1.6 to + 2 V
Lower than lower limit
8
+1.6 to + 2 V
OK
D-72336 Balingen E-Mail: info@sauter.eu
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Auto-Peak-Hold-Function(Auto-Peak Mode)
Please press:
Min. limit peak save
6.1 Output Protocol
RS-232 Parameter:
Baud rate: 9600 Data-Bit: 8
This function allows to eliminate unwanted „Pre-Peak values“ that are lower than the m ain peak value (Fp). The “Min limit peak save” value (Fo) takes care, that these “Pre-Peak values” are not saved.
The „Min limit peak save“ function is only possible in Peak-Mode.
To set this Min limit value, please see the SET Menu in section „Operating keys“.
Memory mode and average value (from up to 10 peak values)
Saving peak values in the instrument
Activating the „AUTO PEAK Function“ by PEAK key Deactivating the „Average Function“ by MEMORY
key
Now, all peak values are stored automatically in the
instrument
To browse through the stored values, please use the
▲ or ▼ keys. (The values will be shown in the upper display segment)
By pressing the MEMORY key, the average value of
the stored peak values can be shown (in the upper display segment)
To delete every stored value, press the▼-key in the
AVERAGE-Mode
6. Configuration of RS 232
FH_10-500-BA-e-1213 4
SUB-D 9pm
Parity: none Stop-Bit: 1
The measured value is requested by the PC by the ASCII Sign “9”.
The measured value that comes from the instrument has this format:
e.g. 0011.70 means -11,70 Newton, if Newton is the selected unit ||-----| | |____> the other 6 places describe the measured value as ASCII-Signs |_______> the first place desribes the direction of the force (0 = minus = Pressure; 1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
|_______> the first place describes the direction of the force (0 = minus = Pressure; 1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
7. Warning
Intended use
The instrument you have acquired serves to determine the measuring value of the material to be measured. It is intended to be used as a “non-automatic“ instrument, i.e. the material to be measured is manually and carefully attached at the instrument. The measuring value can be read off after a stable measuring value has been obtained.
Inappropriate use
Do not use the instrument for dynamic measuring. In the event that small quantities are removed or added to the
Sauter GmbH
Ziegelei 1
Instruction Manual
FH 10– 500 EXT
D-72336 Balingen E-Mail: info@sauter.eu
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
material to be measured, incorrect measuri ng results can be displayed due to the “stability compensation“ in the instrument. (Example: Slow draining off of liquid from a container suspended from the instrument). Do not attach a continuous load. This can damage the measuring unit as well as the parts, relevant to safety. Prevent jolts, torsion and oscillation (e.g. by appending slopingly) of all kinds. Be sure to prevent overloading the instrument in excess of the stated maximum load (max.), minus any tare weight that may possibly exist. This could damage the instrument (risk of breakage). Important:
• Always make sure that there are no people or materials below the load that could be injured or damaged!
• The instrument is not suitable for measuring people. Do not use as baby scales!
• The instrument does not comply with the medical product law (MPG). Never operate the instrument in hazardous locations. The series design is not explosion-proof. Structural alterations may not be effected to the instrument. This can lead to incorrect measuring results, faults concerning safety regulations as well as to destruction of the instrument. The instrument may only be used in compliance with the described guidelines. Varying areas of application/ planned use must be approved by SAUTER in writing.
Guarantee
The guarantee is not valid following
• Non-observation of our guidelines in the operating instructions
• Alteration to or opening of the device
• Mechanical damage and damage caused by media, liquids
• Natural wear and tear
• Inappropriate erection or electric installati on
• overloading of the measuring equipment
Monitoring the test substances
The metrology features of the instrument and any possible available adjusting weight must be checked at regular
intervals within the scope of quality assurance. For this purpose, the answerable user must define a suitable interval as well as the nature and scope of this check. Information is available on the home page (www.KERN-
sohn.com) with regard to the monitoring of instrument test
substances and the test weights required for this. Test weights and instruments can be adjusted quickly and at a reasonable price in KERN’s accredited DKD calibration laboratory (return to national normal).
Fundamental safety information
Do not use the hanging instrument to transport loads. Prevent jolts, torsion and oscillation (e.g. by appending slopingly) of all kinds. Never use the hanging instrument over the maximum permitted weight (!! Danger of breaking!!). Always make sure that there are no living beings or materials below the load that could be injured or damaged. The hanging electronic instruments from the SAUTER instrument are only suitable for hand-held use or use in a test stand. They are not suitable for hanging from a mechanical hook, e.g. a crane hook.
Observe the information in the operating instructions please read the operating instructions carefully before erecting and commissioning, even if you already have experience with S AUTER instruments.
Staff training
The device may only be operated and looked after by trained members of staff.
Major display deviations (incorrect measuring results) are possible if electromagnetic fields occur as well as due to static charging and instable power supply. It is then necessary to change the location.
Unpacking
Carefully remove the instrument from its packaging, remove the plastic wrapping and position the instrument in its intended working location.
FH_10-500-BA-e-1213 5
Sauter GmbH
Ziegelei 1
Instruction Manual
FH 10– 500 EXT
1
Start the
Press ON/OFF
The green
2
Enter the
Directly after pressing
The lower
3
Instrument
Press SET
XXX N is
4 Select the
Press ▼▲
The
5
Save
Press SET
6 Enter the
Press MEMOR Y
The upper
7
Chose the
Press UNIT and
The weight
8
Save the
Press SET and UNIT
D-72336 Balingen E-Mail: info@sauter.eu
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
8. CE Declaration of conformity
FH Adjustment Procedure
instrument
adjustment procedure
Type
a (Back to the
normal mode ???)
instrument type
ON/OFF, press PEAK and PRINT together for more times and very short intervals until the red light is on.
Directly after pressing PEAK + PRINT
(If you are in the operation mode again, the instrument has to be shut off. Start again with
1. Eventually pressing the buttons faster)
light is on
red light is on
shown
correspondi ng value to the instr. appears in the display
FH_10-500-BA-e-1213 6
instrument type
adjustment sequence
adjustment weight
adjustment weight
9 Attach the
weight to the instrument (by hanging)
with▼▲ specify the calibration weight in Newton (X kg * 9,81)
simultaneously
Press ZERO The
red light is on
is written in the display
instrument changes to operation mode. Adjustment procedure is finished
Loading...