KERN & SOHN EOE XL User Manual

KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Platform/floor scales
KERN EOB / EOE_L / EOE_XL / EOS
Version 3.0 07/2010 GB
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
KERN EOB / EOE_L / EOE_XL /
GB
EOS
Version 3.0 07/2010
Table of Contents
1 Technical Specifications ............................................................................... 4
2 Appliance overview ....................................................................................... 9
2.1 Overview of displays................................................................................................11
2.2 Keyboard overview...................................................................................................11
3 Basic Information (General)........................................................................ 12
3.1 Proper use.................................................................................................................12
3.2 Improper Use ............................................................................................................12
3.3 Warranty ....................................................................................................................12
3.4 Monitoring of Test Resources.................................................................................13
4 Basic Safety Precautions ............................................................................ 13
4.1 Pay attention to the instructions in the Operation Manual................................... 13
Operating instructions Platform/floor scales
4.2 Personnel training .................................................................................................... 13
5 Transportation & Storage............................................................................ 13
5.1 Testing upon acceptance ........................................................................................13
5.2 Packaging / return transport ...................................................................................13
6 Unpacking, Setup and Commissioning ..................................................... 14
6.1 Installation Site, Location of Use ............................................................................14
6.2 Unpacking and placing ............................................................................................14
6.2.1 Scope of delivery / serial accessories...................................................................................... 15
6.2.2 Assembly instructions for the use of the wall fixture ................................................................15
6.3 Mains connection .....................................................................................................16
6.4 Battery operation......................................................................................................16
6.5 Initial Commissioning ..............................................................................................16
7 Adjustment ................................................................................................... 17
2
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
8 Operation...................................................................................................... 18
Start-up..................................................................................................................................18
Switching Off ........................................................................................................................18
Weighing ...............................................................................................................................18
Switch-over weighing unit..................................................................................................18
Taring ....................................................................................................................................19
Hold function (animal weighing function)..........................................................................20
Plus/minus weighing............................................................................................................21
9 Menu ............................................................................................................. 22
10 Service, maintenance, disposal.................................................................. 23
10.1 Cleaning ....................................................................................................................23
10.2 Service, maintenance...............................................................................................23
10.3 Disposal.....................................................................................................................23
11 Error messages, troubleshooting guide .................................................... 23
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
3

1 Technical Specifications

Models KERN EOB
KERN
Readability (d) 5 g 10 g 20 g 20 g Weighing range (max) 15 kg 35 kg 60 kg 60 kg Reproducibility 5 g 10 g 20 g 20 g Linearity ±10 g ±20 g ±40 g ±40 g
EOB
15K5
EOB
35K10
EOB
60K20
EOB
60K20L
Recommended adjustment weight, not added (class)
Stabilization time (typical) 2.5 sec. Warm-up time 10 min Supply voltage 220 V- 240 V, AC 50 Hz Power pack
secondary voltage Battery 6 x 1.5 V AA (=9V) Battery operation period 220 h Auto-Off 3 min Weighing Units kg, lb Ambient temperature + 10°C …+ 35°C Humidity of air max. 80 % (not condensing)
10 kg
(M2)
20 kg
(M2)
9V, 100 mA
40 kg
(M2)
40 kg
(M2)
Display unit (B x D x H) mm
Cable length display unit 180 cm 180 cm 180 cm 270 cm Weighing surface mm 314.5 x 304.5 550 x 550 Weight kg (net) 4.2 13.5
4
210 x 110 x 45
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
KERN
EOB
150K50
EOB
150K50L
EOB
150K50XL
Readability (d) 50 g 50 g 50 g Weighing range (max) 150 kg 150 kg 150 kg Reproducibility 50 g 50 g 50 g Linearity ±100 g ±100 g ±100 g Recommended
adjustment weight, not added (class)
100 kg
(M2)
100 kg
(M2)
100 kg
(M2)
Stabilization time (typical) 2.5 sec. Warm-up time 10 min Supply voltage 220 V- 240 V, AC 50 Hz Power pack
secondary voltage
9V, 100 mA
Battery 6 x 1.5 V AA (=9V) Battery operation period 220 h Auto-Off 3 min Weighing Units kg, lb Ambient temperature
+ 10°C …+ 35°C
Humidity of air max. 80 % (not condensing) Display unit
(B x D x H) mm
210 x 110 x 45
Cable length display unit 180 cm 270 cm 270 cm Weighing surface mm 314.5 x 304.5 550 x 550 945 x 505 Weight kg (net) 4.2 13.5 19.5
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
5
KERN
EOB
300K100A
EOB
300K100L
EOB
300K100XL
Readability (d) 100 g 100 g 100 g Weighing range (max) 300 kg 300 kg 300 kg Reproducibility 100 g 100 g 100 g Linearity ±200 g ±200 g ±200 g Recommended
adjustment weight, not added (class)
200 kg
(M2)
200 kg
(M2)
200 kg
(M2)
Stabilization time (typical) 2.5 sec. Warm-up time 10 min. Supply voltage 220 V- 240 V, AC 50 Hz Power pack
secondary voltage
9V, 100 mA
Battery (optional) 6 x 1.5 V AA (=9V) Battery operation period 220 h Auto-Off 3 min Weighing Units kg, lb Ambient temperature
+ 10°C …+ 35°C
Humidity of air max. 80 % (not condensing) Display unit
(B x D x H) mm
210 x 110 x 45
Cable length display unit 180 cm 270 cm 270 cm Weighing surface mm 314.5 x 304.5 550 x 550 945 x 505 Weight kg (net) 4.2 13.5 19.5
6
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
Models KERN EOE
KERN
EOE
150K50L
EOE
150K50XL
EOE
300K100L
EOE
300K100XL
Readability (d) 50 g 50 g 100 g 100 g Weighing range (max) 150 kg 150 kg 300 kg 300 kg Reproducibility 50 g 50 g 100 g 100 g Linearity ±100 g ±100 g ±200 g ±200 g Recommended
adjustment weight, not added (class)
100 kg
(M2)
100 kg
(M2)
200 kg
(M2)
200 kg
(M2)
Stabilization time (typical) 2.5 sec. Warm-up time 10 min. Supply voltage 220 V- 240 V, AC 50 Hz Power pack
secondary voltage
9V, 100 mA
Battery (optional) 6 x 1.5 V AA (=9V) Battery operation period 220 h Auto-Off 3 min Weighing Units kg, lb Ambient temperature + 10°C …+ 35°C Humidity of air max. 80 % (not condensing) Display unit
(B x D x H) mm
210 x 110 x 45
Cable length display unit 270 cm Weighing surface mm
505 x 505 945 x 505 505 x 505 945 x 505
Weight kg (net) 12.5 19.5 12.5 19.5
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
7
Models KERN EOS
KERN
EOS
150K50XL
EOS
300K100XL
Readability (d) 50 g 100 g Weighing range (max) 150 kg 300 kg Reproducibility 50 g 100 g Linearity ±100 g ±200 g Recommended
adjustment weight, not added (class)
100 kg
(M2)
200 kg
(M2)
Stabilization time (typical) 2.5 sec. Warm-up time 10 min. Supply voltage 220 V- 240 V, AC 50 Hz Power pack
secondary voltage
9V, 100 mA
Battery (optional) 6 x 1.5 V AA (=9V) Battery operation period 220 h Auto-Off 3 min Weighing Units kg, lb Ambient temperature + 10°C …+ 35°C Humidity of air max. 80 % (not condensing) Display unit
(B x D x H) mm
210 x 110 x 45
Cable length display unit 270 cm Weighing surface mm
900 x 550 900 x 550
Weight kg (net) 18.8 18.8
8
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

2 Appliance overview

Models EOB
Weighing plate, stainless steel
Models EOE
Weighing plate varnished steel
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
9
Models EOS
Weighing plate, stainless steel
Anti-slip rubber mat
Comfortable transport due to 2 rollers and 1 grab handle
10
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

2.1 Overview of displays

No. Description
1 Is displayed, when the battery capacity is exhausted. 2 Balance zero display:
Should the balance not display exactly zero despite empty scale pan, press the TARE button. The balance will be set to zero after a short standby time.
3 Stability display:
If the display shows the stability display [ο] the balance is in a stable
status. The [ο] indication disappears if the condition is unstable. 4 Hold/animal weighing function active, see chap. 9 5
Weighing unit [kg ' lb], see chap. 8 „Switch-over weighing unit“ 6 Stored tare value, see chap. 8 „Taring“

2.2 Keyboard overview

Key Function
Turn on/off balance
Hold/ animal weighing function
Tare balance
Switch-over weighing unit
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
11

3 Basic Information (General)

3.1 Proper use

The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.

3.2 Improper Use

Do not use balance for dynamic add-on weighing procedures, if small amounts of goods to be weighed are removed or added. The “stability compensation“ installed in the balance may result in displaying an incorrect measuring value! (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.

3.3 Warranty

Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described use
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
12
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

3.4 Monitoring of Test Resources

In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.

4 Basic Safety Precautions

4.1 Pay attention to the instructions in the Operation Manual

Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.

4.2 Personnel training

The appliance may only be operated and maintained by trained personnel.

5 Transportation & Storage

5.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.

5.2 Packaging / return transport

Ö Keep all parts of the original packaging for a possibly
required return.
Ö Only use original packaging for returning. Ö Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Ö Reattach possibly supplied transport securing devices.
Ö Secure all parts such as the glass wind screen, the
weighing platform, power unit etc. against shifting and damage.
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
13

6 Unpacking, Setup and Commissioning

6.1 Installation Site, Location of Use

The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be changed.

6.2 Unpacking and placing

Open package, take out the appliance and accessories. Verify that there has been no damage and that all packing items are present. The balance must be installed in a way that the weighing plate is exactly in horizontal position. Mount the display unit in a way that facilitates operation and where it is easy to see.
14
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

6.2.1 Scope of delivery / serial accessories

Platform and display unit, (see chap. 2)
4 x adjustment feet (excepted platform size 314.5 x 304.5 cm)
Mains power supply
Wall fixture (with fixing screws)
Operating instructions

6.2.2 Assembly instructions for the use of the wall fixture

EOB/EOE_L/EOE_XL/EOS-BA-e-1030
15

6.3 Mains connection

Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.

6.4 Battery operation

On the rear side of the display unit remove the battery cover and connect 6 x 1.5V mignon cells. Reinsert the battery cover. In order to save the battery, the balance switches automatically off after 3 minutes without weighing. Further shutdown times can be set in the Menu (Function "A.OFF"), see Section 9. If the batteries are exhausted, the battery symbol will appear (see chap. 2.1). Switch­off balance and replace batteries at once. If the balance is not used for a longer time, take out the batteries and store them separately. Leaking battery liquid could damage the balance.

6.5 Initial Commissioning

In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1). During this warming up time the balance must be connected to the power supply (mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
16
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

7 Adjustment

As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation. Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization.
Ö In weighing mode press the UNIT key and keep it pressed until „CAL“ followed by
the weight value of the required adjustment weight in displayed. „LOAd“ is displayed alternatively.
Ö
Ö Carefully place the required adjustment weight (see tab. 1 „Technical data“) in
the centre of the weighing plate Press TARE key, adjustment is started.
Ö
Ö Wait until the stability display appears, after that the adjustment is finished. The
balance will automatically return to Weighing mode upon successful completion of the adjustment and will display the weight of the adjustment weight.
Ö Take away adjustment weight.
Ù
Example
An error message will be displayed in the event of an adjustment error or
incorrect adjustment weight. Remove the adjustment weight and repeat the
adjustment process.
Keep the adjustment close to the balance. Daily control of the weighing
exactness is recommended for quality-relevant applications.
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
17

8 Operation

Start-up

Switching Off

Ö Press the ON/OFF key.
The balance will carry out a self-test. The balance is ready for weighing when the weight display appears.
Should the balance not display exactly zero despite empty weighing plate, press the TARE button. The balance will be set to zero after a short standby time.
Ö Press ON/OFF button, the display disappears

Weighing

Switch-over weighing unit

Ö Place goods to be weighed on balance Ö Wait for stability display Ö Read weighing result.
Ö By pressing the UNIT button the weighing result can be
displayed in another unit [kg ' lb].
Unit conversion: 1 kg = 2.20462 lb
18
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

Taring

Ö Place an empty weighing container, the weight of the
weighing container will be displayed.
Ö Press the TARE button, the zero display disappears. The
indicator [S] under [NET] is displayed. The tare weight is saved until it is deleted.
Ø
Ö Weigh the material, the net weight will be indicated.
Delete tare
The taring process can be repeated any number of times, e.g. when adding several components for a mixture (adding). The limit is reached when the whole weighing range is exhausted.
The weight of the weighing container will be displayed as a minus number after removing the weighing container.
The tare weight is saved until it is deleted.
Ö Unload the balance and press the TARE button, the zero
display appears.
Ø
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
19

Hold function (animal weighing function)

The balance has an integrated animal weighing function (mean value calculation). Using this function it is possible to weigh domestic or small animals exactly (min. load 1% of the max. one), although they do not stand quiet on the weighing plate.
Ö Put upon weighing goods and press the HOLD key. In the
Ö This value remains in the display until the HOLD button is
display the indicator [S] starts flashing. During this time the balance takes up several measured values and displays then the calculated mean value.
Ø
pressed again. The indicator [S] under [HOLD] goes out, the balance returns into normal weighing mode.
Ö Press the HOLD button once more to repeat this function
as often as required.
There is no average value calculation in the event of too much movement
(heavy display oscillation).
20
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

Plus/minus weighing

Ö For example for item weight check, production check etc.
switch-on the balance using ON/OFF button and wait for zero display.
Ö Put the nominal weight on the weighing plate and tare to
zero using TARE key. Remove the nominal weight.
Ø
Ö Put the test objects subsequently on the weighing plate, the
respective deviation from the nominal weight is displayed with the respective sign to „+“ and „-“ (e.g. + 0.037 g).
According to the same procedure also packages with the same weight can be produced, referring to a nominal weight.
Ö Back to weighing mode by pressing the TARE button.
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
21

9 Menu

Changes are only allowed on the functions (weighing units) and (automatic switch-off function).
Ö In weighing mode keep TARE button depressed for approx. 3 sec until "SETUP"
is displayed followed by "UNIT"
Ö Press the HOLD button as often as necessary until the required function is
displayed.
Ö Press the TARE button to confirm the selected function. The current setting will
be displayed. Using the HOLD or TARE button, select the desired parameter. Use the HOLD button to return to the menu.
Ö To exit the Menu and save, press the HOLD button as often as required until
"END" is displayed and then confirm by pressing TARE. The balance returns automatically into weighing mode.
Navigation:
Not documented
Not documented
22
Factory settings are marked by *.
EOB/EOE_L/EOE_XL/EOS-BA-e-1030

10 Service, maintenance, disposal

10.1 Cleaning

Before cleaning, disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Take care that the device is not penetrated by fluids and polish it with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Remove spilled weighing goods immediately.

10.2 Service, maintenance

The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.

10.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.

11 Error messages, troubleshooting guide

Error message Function
LO
ERRE
ErrEd
Err
ErrL
Capacity of batteries exhausted EEPROM error display unit EEPROM error Overload Minimum weight under min. value
EOB/EOE_L/EOE_XL/EOS-BA-e-1030
23
In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The displayed weight does not glow.
The displayed weight is permanently changing
-
The weighing value is obviously wrong
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
Power supply interrupted.
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with
foreign matter.
Electromagnetic fields / static charging (choose different location/switch off interfering device if possible)
The display of the balance is not at zero
Adjustment is no longer correct.
The balance is on an uneven surface.
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
24