13Help for any minor failures ................................................................................. 34
TEFC_A-BA-e-2110 3
1 Technical specification
KERNEFC 30K-3EFC 30K-3L
Product number / type TEFC 30K-3-A TEFC 30K-3L-A
Interval (d) 0.002 kg 0.002 kg
Weighing range (Max) 30 kg 30 kg
Reproducibility 0.005 kg/0.01 kg 0.005 kg/0.01 kg
Linearity ±0.004 kg (2 d) ±0.004 kg (2 d)
Minimum piece weight when
counting the number of pieces in laboratory conditions*
Minimum piece weight when
counting the number of pieces in standard conditions**
Recommended
adjustment weight
(not delivered)
1 g (0.5 d) 1 g (0.5 d)
10 g 10 g
30 kg (M1) 30 kg (M1)
Possible adjustment points 10–100% Max10–100% Max
Settling time (standard) 3 s
Heating time 30 min
Weight units kg, g, lb, oz, PCS
Air humidity max. 80%, relative (non-condensing)
Permissible
ambient temperature
Input voltage
of the device
Input voltage of
the power supply
Battery 3.7 V / 4 Ah
Rechargeable
battery operation
Display housing
dimensions [mm]
Scale plate,
stainless steel [mm]
400 × 300 × 124 mm 500 × 400 × 120 mm
operating time 80 h (illumination off)
operating time 50 h (illumination on)
252 × 162 × 57 (width × depth × height)
0°C to +40°C
5 V, 1 A
100–240 VAC; 50/60 Hz
charging time ca. 5 h
Net weight [kg] 17 kg 19 kg
Interface RS-232
4 TEFC_A-BA-e-2110
Weighing range (Max)
60 kg
60 kg
Settling time (standard)
3 s
Weight units
kg, g, lb, oz, PCS
Air humidity
max. 80%, relative (non-condensing)
Input voltage
Display housing
KERNEFC 60K-3EFC 60K-3L
Product number / type TEFC 60K-3-A TEFC 60K-3L-A
Interval (d) 0.005 kg 0.005 kg
Reproducibility 0.001 kg/0.002 kg 0.001 kg/0.002 kg
Linearity ±0.01 kg (2 d) ±0.01 kg (2 d)
Minimum piece weight when
counting the number of pieces
in laboratory conditions*
Minimum piece weight when
counting the number of pieces
in standard conditions**
Recommended
adjustment weight
(not delivered)
Possible adjustment points 10–100% Max10–100% Max
2.5 g (0.5 d) 2.5 g (0.5 d)
25 g 25 g
60 kg (F2) 60 kg (F2)
Heating time 30 min
Permissible
ambient temperature
of the device
Input voltage of
the power supply
Battery 3.7 V / 4 Ah
Rechargeable
battery operation
dimensions [mm]
Scale plate,
stainless steel [mm]
Net weight [kg] 17 kg 19 kg
Interface RS-232
400 × 300 × 124 mm 500 × 400 × 120 mm
operating time 80 h (illumination off)
operating time 50 h (illumination on)
252 × 162 × 57 (width × depth × height)
0°C to +40°C
5 V, 1 A
100–240 VAC; 50/60 Hz
charging time ca. 5 h
TEFC_A-BA-e-2110 5
Weighing range (Max)
150 kg
150 kg
300 kg
Weight units
kg, g, lb, oz, PCS
Heating time
30 min
Air humidity
max. 80%, relative (non-condensing)
Input voltage
Display housing
KERNEFC 100K-3EFC 100K-3LEFC 300K-2
Product number / type TEFC 100K-3-A TEFC 100K-3L-A TEFC 300K-2-A
Interval (d) 0.01 kg 0.01 kg 0.02 kg
Reproducibility 0.002 kg/0.005 kg 0.002 kg/0.005 kg 0.05 kg/0.1 kg
Linearity ±0.02 kg (2 d) ±0.02 kg (2 d) ±0.04 kg (2 d)
Minimum piece weight when
counting the number of pieces in laboratory conditions*
Minimum piece weight when
counting the number of pieces in standard conditions**
Recommended
adjustment weight
(not delivered)
Possible adjustment points 10–100% Max10–100% Max10–100% Max
5 g (0.5 d) 5 g (0.5 d) 10 g (0.5 d)
50 g 50 g 100 g
150 kg (F2) 150 kg (F2) 300 kg (M1)
Settling time (standard) 2 s
Permissible
ambient temperature
of the device
Input voltage of
the power supply
Battery 3.7 V / 4 Ah
Rechargeable
battery operation
dimensions [mm]
Scale plate,
stainless steel [mm]
Net weight [kg] 19 70 70
Interface RS-232
500 × 400 × 120 mm 600 × 500 × 140 mm 600 × 500 × 140 mm
operating time 80 h (illumination off)
operating time 50 h (illumination on)
252 × 162 × 57 (width × depth × height)
0°C to +40°C
5 V, 1 A
100–240 VAC; 50/60 Hz
charging time ca. 5 h
* Minimum piece weight when counting the number of pieces in laboratory conditions:
There are optimum ambient conditions to count pieces with high resolution
No diversification of the counted pieces’ weight
**Minimum piece weight when counting the number of pieces in standard conditions:
There are unsteady ambient conditions (wind gusts, vibrations)
There is diversification of the counted pieces’ weight
6 TEFC_A-BA-e-2110
Dimensions:
Models EFC 30K-3, EFC 60K-3:
Models EFC 30K-3L, EFC 60K-3L, EFC 100K-3:
Models EFC 100K-3L, EF C 300K-2:
TEFC_A-BA-e-2110 7
2 Declaration of Conformity
The valid Declaration of Conformity EC/UE is available at: