Readout (d) 5 g 10 g 20 g
Weighing range (max) 10 kg 20 kg 50 kg
Taring range (subtractive) 10 kg 20 kg 50 kg
Reproducibility
Linearity
Adjusting weight
(not included), adjustment/
tolerance as per M3
Stabilisation time (typical) 2 sec.
Optimal temp. of operation + 5° C .... + 35° C
Humidity max. 80 % (non-condensing)
Size (W x D x H) mm 320 x 300 x 60
Weighing plate mm 320 x 260
Weight kg (net) 1,7
5 g 10 g 20 g
± 10 g ± 20 g ± 400 g
5 kg
(M3)
10 kg
(M3)
20 kg
(M3)
ECE-BA-e-1120
3
2 Fundamental information (general)
2.1 Intended use
The balance you have acquired serves to determine the weighing value of the material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
plate. The weighing value can be read off after a stable weighing value has been obtained.
2.2 Inappropriate use
Do not use the balance for dynamic weighing. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance.)
Do not leave a permanent load on the weighing plate. This can damage the measuring equipment.
Be sure to avoid impact shock and overloading the balance in excess of the prescribed maximum load rating (max.), minus any possible tare weight that is already
present. This could cause damage to the balance.
Never operate the balance in hazardous locations. The series design is not explosion-proof.
Structural alterations may not be made to the balance. This can lead to incorrect
weighing results, faults concerning safety regulations as well as to destruction of the
balance.
The balance may only be used in compliance with the described guidelines. Varying
areas of application/planned use must be approved by KERN in writing.
2.3 Guarantee
The guarantee is not valid following
- non-observation of our guidelines in the operating instructions
- use outside the described applications
- alteration to or opening of the device
- mechanical damage and damage caused by media, liquids
- natural wear and tear
- inappropriate erection or electric installation
- overloading of the measuring equipment
4
ECE-BA-e-1120
2.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight
must be checked at regular intervals within the scope of quality assurance. For this
purpose, the answerable user must define a suitable interval as well as the nature
and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at
a reasonable price in KERN’s accredited DKD calibration laboratory (return to national normal).
3 Fundamental safety information
3.1 Observe the information in the operating instructions
Please read the operating instructions carefully before erecting and commissioning,
even if you already have experience with KERN balances.
3.2 Staff training
The device may only be operated and looked after by trained members of staff.
4 Transport and storage
4.1 Acceptance check
Please check the packaging immediately upon delivery and the device during unpacking for any visible signs of external damage.
4.2 Packaging / return transport
Ö Keep all parts of the original packaging for a possibly required re-
turn.
Ö Only use original packaging for returning.
Ö Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Ö Reattach possibly supplied transport securing devices.
Ö Secure all parts such as the glass wind screen, the weighing plat-
form, power unit etc. against shifting and damage.
ECE-BA-e-1120
5
5 Unpacking, installation and commissioning
5.1 Place of installation, place of use
The balance is constructed in such a way that reliable weighing results can be achieved under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and
precisely.
Therefore please observe the following at the place of installation:
- Place the balance on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
- Protect the balance against direct draughts due to open windows and doors;
- Avoid jarring during weighing;
- Protect the balance against high humidity, vapours and dust;
- Do not expose the device to extreme dampness for longer periods of time. In-
admissible bedewing (condensation of air moisture on the device) can occur if
a cold device is taken into a significantly warmer environment. In this case,
please acclimatise the device for approx. 2 hours at room temperature after it
has been disconnected from the mains.
- Avoid static charging of the material to be weighed, weighing container and
windshield.
Major display deviations (incorrect weighing results) are possible if electromagnetic
fields occur as well as due to static charging and instable power supply. It is then necessary to change the location.
5.2 Unpacking the Balance
Carefully remove the balance from the packaging and out of the plastic covering,
then place the balance on the assigned position.
5.2.1 Installation
Install the balance in such a fashion that the weighing plate is absolutely horizontal.
6
ECE-BA-e-1120
5.2.2 Package volume
Serial fittings:
KERN ECB
• Balance
• Operating manual
• Power adaptor
5.3 Mains supply/ Battery operation
5.3.1 Mains supply
Electric power supply is by means of the external mains supply circuit. The printed
voltage level must comply with the local voltage.
Only use original KERN mains supply circuits. The use of other makes is subject to
approval by Kern.
Mains adapter output voltage: 9V DC (polarity: inner terminal is negative)
5.3.2 Battery operation
Open the battery cover on the underside of the balance. Insert 6 x 1.5volt (AA) batteries.
Replace the battery cover. To preserve battery life, the balance switches off
automatically after 3 minutes if no weighing has taken place.
For battery operation the balance has an automatic switchoff function which can be
activated and deactivated in the menu.
When the battery power is used up the display will show “LO“. Press the ON/OFF
key and change the batteries at once.
When the balance is not in use for a longer period of time remove batteries and keep
them separately. Leakage of battery liquid might damage the balance.
The optionally supplied battery is charged with the supplied power supply.
The battery has a service life of c. 100 hours.
The charging period for total charge is c. 10 hours.
ECE-BA-e-1120
7
Rechargeable Battery Installation:
1. Remove battery cover
2. Slide spacer to left or right, lever down on corner, then remove the
spacer
8
3. Insert and connect rechargeable battery
ECE-BA-e-1120
5.4 Initial start-up
A warm-up time of 5 minutes stabilises the measured values after switching on.
The accuracy of the balance depends on the local acceleration of the fall.
Please be sure to observe the information in the chapter on adjusting.
5.5 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated – in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location (
only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out during the initial start-up, after change in location and variation of surrounding temperature. It is also recommendable to adjust
the balance periodically during weighing operation in order to obtain exact measured
values.
5.6 Adjusting
Using a precision weight, the accuracy of the balance can be checked at any time
and adjusted.
Adjustment procedure:
Check that the surrounding conditions are stable.
A short warm-up time of about 5 minutes is recommended for stabilisation.
Operation Display
Turn balance on using ON/OFF button
Hold down TARE button for 3 seconds
The weight value for adjusting the scales will appear in the
display.
Place the required adjustment weight on the weighing plat-
form (see section 1, “Technical data“)
Shortly afterwards, this will appear:
If adjustment has not worked correctly or if the wrong ad-
justment weight was used then this will be displayed:
In this case switch the power off and on and repeat the
adjustment process.
Remove the weight from the weighing platform and the
display will automatically return to normal weighing mode.
Adjustment has been successfully completed.
0,00
10,00
F
E
0,00
ECE-BA-e-1120
9
6 Operation
6.1 Weighing
• The balance is switched on by pressing the "ON/OFF" key.
• The balance performs auto-diagnostics (for 2 s) and then displays "0.00".
• If the weighing object is heavier than the weighing range the display indicates
"E" (=Overload).
• The balance is switched off by pressing the "ON/OFF" key once.
6.2 Taring (TARE key)
• Switch the balance with the “ON/OFF” key on and wait until display appears
as "0.00". Position the tare load and press the “TARE” key.
• The balance display becomes "0.00".
• The item to be weighed can now be placed in the tare container and the
weight can be read off.
10
ECE-BA-e-1120
7 Maintenance, upkeep, disposal
7.1 Cleaning
Please disconnect the device from the operating voltage before cleaning.
Only use a cloth dampened with mild suds and not aggressive cleaning agents (sol-
vents or similar). Please ensure that fluids are not able to get into the device and rub
off using a clean, soft cloth.
Loose sample residue/powder can be removed carefully using a brush or hand vacuum cleaner.
Remove any spilt material to be weighed immediately.
7.2 Maintenance, upkeep
The device may only be opened by trained service engineers authorised by KERN.
Disconnect from the mains supply before opening.
7.3 Disposal
The operating company shall dispose of the packaging and the device in compliance
with the valid national or regional law of the operating location.
ECE-BA-e-1120
11
8 Troubleshooting
The balance should be switched off for a short time following an interruption in the
programme sequence and disconnected from the mains supply. It is then necessary
to repeat the weighing process from the beginning.
Help:
Interruption Possible cause
Weight display is not illuminated.
The weight display changes continually
The weighing result is obviously incorrect
• The balance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
• Power supply interrupted. .
• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with for-
eign matter.
• Electromagnetic fields / static charging
(choose different location/switch off interfering device if possible)
• The balance display is not set to zero
Switch the balance off if other error messages should appear and then switch on again. Contact the manufacturer if the error message does not disappear.
12
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging
(choose different location/switch off interfering device if possible)
ECE-BA-e-1120
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.