5.4.1 Turning On the Power ............................................................................................................................. 29
9.2.2 Result display .......................................................................................................................................... 54
Illuminates during communication with external devices.
•The stability display indicates that the weighing value is
stable.
•Highlights current setting in the menu.
The indicator will mark the measuring result
Status display during drying
Heating-up period
Preset drying temperature is
reached. The drying is finished
when ∆M constant.
Heating-up period
Preset drying temperature is
reached. The drying is finished
when the preset drying time has
expired.
8 DBS-BA-e-1311
RAPID
+see chap. 9.2.1.3
SLOW
Heating-up period “pre-heating
step”
The pre-heating step is switched
on until “∆M pre-heating step” is
constant.
Temperature is lowered to the
preset drying temperature.
Preset drying temperature is
reached.
The drying is finished when the
preset completion criterion is
fulfilled.
Heating-up period
+see chap. 9.2.1.4
STEP
+see chap. 9.2.1.5
Preset drying temperature is
reached.
The drying is finished when the
preset completion criterion is
fulfilled.
Heating-up period step 1
Drying step 1
Heating-up period step 2
Drying step 2
Heating-up period step 3
Drying step 3
DBS-BA-e-1311 9
6
Previous drying time
7
8
9
Basic displays
Current temperature
Result display, see chap. 9.2.2
%
M/W [%] moisture
D/W
M/D
W/D
g
Gram
Weight display
Display % moisture
[%] dry mass:
ATRO dry mass
ATRO moisture
Gram display
0
Current menu selection
UNIT
COMSET
CODE
CAL
PRINT
Menu access via MENU button
Equipment is in Ready mode, see chap. 8.4.
Equipment is in menu „Display of results“, see chap. 9.2.2
Equipment is in menu „Interface parameters“, see chap. 11.2
Equipment is in menu „Sample denomination“, see chap. 10.1
Equipment is in menu „Adjustment“, see chap. 6
Equipment is in menu „Printer“, see chap. 11.3 – 11.5
10 DBS-BA-e-1311
a
Status display „Equipment“
Top
Centre
Bottom
Is displayed when the heated cover
is open.
Flashes when the heated cover
needs to be closed.
Goes out when heated cover is
closed
Is displayed when the heated cover
is open.
Goes out when heated cover is
closed
Flashes when drying process is
active
Indicates that a sample is in the
sample dish.
Flashes when a sample needs to be
loaded.
Goes out if there is no sample in
the sample dish.
Indicates that a sample is in the
sample dish.
Flashes if taring is required.
Goes out if no sample dish is in
place.
Indicates upon switch-on that the
dish holder is installed.
Flashes upon switch-on if the dish
holder is not installed.
DBS-BA-e-1311 11
2.1.1 Display examples
During drying:
After drying:
1
Loaded drying program + see chap 9.1
2
Previous time
3
Current temperature
4
Unit „Display of results“ + see chap. 9.2.2
5
Preset drying mode /
Status display drying
6
Current moisture content in %.
During drying with can be switched in ∆M (see chap. 9.2.1.1).
7
Drying process active + see chap. 2.1
8
Measuring result indicator
9
Measuring value (unit in the preset result display) + see chap. 9.2.2
+ see chap. 9.2.1
+ see chap. 2.1
12 DBS-BA-e-1311
2.2 Keyboard overview
Turn on/off
• Call up menu
• Selecting a menu item to the left.
• Scroll back
• Scroll to next page
• Selection of a menu item in clockwise direction
• Taring
• Zeroing
• Activate menu item
• Save setting
DBS-BA-e-1311 13
• Exit menu
• Back to moisture analysis mode
• Quit
Start drying
Complete drying
Numerical input, see chap. 7.1.1
14 DBS-BA-e-1311
3 Basic Information (General)
3.1 Proper use
The device purchased by you is designed for a fast and reliable determination of
material moisture in liquid, porous and solid materials by applying the method of
thermogravimetrics.
3.2 Improper Use
Impacts and overloading exceeding the stated maximum load (max) of the device,
minus a possibly existing tare load, must be strictly avoided.
Balance may be damage by this.
Never operate device in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-related faults and destruction of the moisture analyser.
The moisture analyser may only be used according to the described conditions.
Other areas of use must be released by KERN in writing.
3.3 Warranty
Loss of warranty due to
• Our conditions in the operation manual are ignored
• The appliance is used outside the described uses
• The appliance is modified or opened
• mechanical damage and damage caused by media, liquids
• natural wear and tear
• The appliance is improperly set up or incorrectly electrically connected
• The measuring system is overloaded
3.4 Monitoring of Test Re sources
In the framework of quality assurance the measuring-related properties of the
moisture analyser and, if applicable, the testing weight, must be checked regularly.
The responsible user must define a suitable interval as well as type and scope of this
test. Information is available on KERN’s home page (www.kern-sohn.com) with
regard to the monitoring of balance test substances and the test weights required for
this. In KERN's accredited DKD calibration laboratory test weights and balances may
be calibrated (return to the national standard) fast and at moderate cost.
DBS-BA-e-1311 15
3.5 Danger I nform ation
!
!
!
!
!
!
!
WARNING
The moisture analyser is used to analyse the moisture content of materials.
This instrument must be used exclusively for this purpose. Any other usage
may cause a risk to personnel, damage to the equipment or other material
damage.
The moisture analyser should be used mainly for the drying of aqueous
substances.
The moisture analyser may not be used in a hazardous area. The appliance may only be operated and maintained by trained personnel.Carefully read this operation manual before setup and commissioning, even if
you are already familiar with KERN appliances.
This moisture analyser is a robust precision instrument, however, it must be
handled with care to ensure many years of trouble-free operation.
Never make any modifications or design changes to the equipment
whatsoever. Always use original spare parts and accessories.
16 DBS-BA-e-1311
!
!
!
Make sure that liquids do not get in contact with the interior of the device, the
connections at the rear of the device and the peripherals (connected PC or
printer).
If you spill liquid on the device, disconnect it immediately.
Afterwards do not operate the moisture analyzer and have it checked by a
competent KERN stockist before any further use.
Never store materials on the equipment.
Never place combustible materials on, under or next to the equipment, as the
environment of the instrument heats up to a high temperature.
DBS-BA-e-1311 17
!
!
!
!
!
!
!
!
Explosive, easily flammable samples and samples that go into a chemical
reaction when subjected to heat, may not be analysed with the moisture
analyser.
Sample materials emitting toxic substances must be dried with a special
extraction system in place. Create an environment that prevents the inhalation
of vapours hazardous to health.
Sample materials developing aggressive vapours (e. G. Acids) may cause
corrosion problems on some parts of the device.
Careful when removing the sample. The sample itself, the sample dish and
the heating unit may be very hot.
Maintain sufficient space in the environment of the equipment to prevent heat
build-up (distance from the equipment 20cm, upwards 1m).
Do not operate the humidity analyser in areas with hazard of explosion Do not place anythin which may interfere the unplugging The power plug must be easily accessible
18 DBS-BA-e-1311
!
!
!
!
Hazards during and after measuri ng
Ensure correct installation of all components, see chap. 5.3 Careful when removing the sample. The sample itself, the sample dish and
the heating unit may be very hot.
Use the sample retainer at all times as it allows safe working and prevents
burns.
Individual parts of the case (e. g. the ventilation grids) may heat up
considerably during operation. Therefore, only take hold of the equipment by
the marked handles.
DBS-BA-e-1311 19
3.5.1 “Hazard information” sticker
!
!
!
!
Ventilation grilles and inspection windows may become very hot during
operation
Always open heated cover completely.Do not use any combustible sample materials.
Disconnect the equipment from the power supply prior to changing a lamp,
see chap. 13.2.2.
20 DBS-BA-e-1311
4 Transport and stora ge
4.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
4.2 Pac kaging / return transport
Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the
weighing platform, power unit etc. against shifting and
damage.
5 Unpacking, Setup and Commissioning
5.1 Instal lation Site, Location of Use
The unit is designed to achieve reliable weighing results under normal conditions of
use.
You will work accurately and fast, if you select the right location for the appliance.
DBS-BA-e-1311 21
On the installation site observe the following:
!
!
!
!
!
Remove explosive materials from the immediate vicinity.
Emerging vapours, sample dish and all parts of the sample chamber are hot!
Remove easily flammable materials from the immediate vicinity.
Protect the device against direct draughts due to open windows and doors.
Avoid extreme heat and temperature fluctuations e.g. due to installation next
to radiators.
Avoid direct sunlight
22 DBS-BA-e-1311
!
!
!
!
!
!
!
!
Never store materials on the equipment.
Protect the appliance against high humidity, vapours and dust, Major display deviations (incorrect weighing results) may be experienced
should electromagnetic fields (e.g. due to mobile phones or radio equipment),
static electricity accumulations or instable power supply occur. Change
location or remove source of interference.
Avoid static charging of the material to be weighed, weighing container and
windshield
Place the device on a firm, level surface. Avoid jarring during weighing.
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance)
may occur if a cold appliance is taken to a considerably warmer environment.
In this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
Operate the device only indoors.
DBS-BA-e-1311 23
5.2 Unpack ing and checking
Open package, take out the appliance and accessories. Verify that there has been
no damage and that all packing items are present.
5.2.1 Scope of delivery / serial accessories
1. Instrument
2. 50 disposable sample dishes
3. Dish holder
4. Wind protection ring
5. Heat shield
6. Spare fuse
7. Removal aid
8. Power cable
9. Protective cover
10. Operating instructions
11. Menu overview
12. Hexagonal socket wrench
24 DBS-BA-e-1311
5.3 Placing
The moisture analyser is delivered in pieces. Check immediately after unpacking all the
parts that the delivery is complete and assemble the individual components as described
below.
3. Install dish holder, ensuring correct positioning, on .
4. Fit sample dish.
5. Level the equipment with foot screws [1] until the air bubble of the water balance is
in the prescribed circle.
26 DBS-BA-e-1311
If both foot screws are turned simultaneously in clockwise direction, the air bubble will
Fig. 1
Fig. 3
Fig. 2
If the air bubble is at the lefthand edge, level the equipment with the righthand foot
screw, see Fig.1.
If the air bubble is at the righthand edge, level the equipment with the lefthand foot
screw, see Fig.2.
Fig. 4
move forwards (see Fig.3).
If both foot screws are turned simultaneously in anticlockwise direction, the air bubble will
move backwards (see Fig.4).
Check levelling regularly.
DBS-BA-e-1311 27
5.4 Mains connection
Power supply is provided via the supplied mains cable.
Check, whether the voltage acceptance on the scales is set correctly. Do not connect
the appliance to the power grid unless the information on the appliance (sticker)
matches the local mains voltage.
Important:
Does the labelling match the local mains voltage?
- Do not connect if mains voltages are different!
- If matching, connect the scales.
The moisture analyzer must be connected to a standard socket with earth terminal.
Do not eliminate the protective effect by using an extension lead without earth
terminal. For power supplies from power grids without earth terminals call a specialist
to establish equivalent protection according to the relevant installation regulations.
- The power plug must be always accessible.
- Before operating, check all cables for damage.
- Guide the cables so that they cannot become damaged or interfere with the
measuring process.
28 DBS-BA-e-1311
Loading...
+ 63 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.