KERN-SOHN IOC 100K-3, BID 600K-1DS, BID 600K-1D, BID 1T-4DS, BID 1T-4D, IOC 600K-2, IOC 10K-4, IOC 60K-3L, IOC 30K-4, IOC 60K-3, IOC 10K-4L, IOC 6K-4, IOC 100K-3L, BID 3T-3D, BID 3T-3DL, IOC 300K-3 Instructions
Functions
Display LCD 24 mm digits with back lighting
Load cell resistance
Range calibration We recommend ≥ 50 % max.
Data output RS232
Electric Supply
Display unit
(W x D x H) mm
Tolerance weighing, totalling, parts counting,
percentage calculation, animal weighing
Input voltage power unit 100 V - 240 V, 50 / 60 Hz
Input voltage device 12V, 1000mA
30,000 d
87 Ω
260 x 115 x 70
Admissible
ambient temperature
Humidity of air max. 80 % (not condensing)
Net weight 0.8 kg
Operating time backlight on 22 h
Rechargeable battery
(optional)
RS 232 interface Standard
Operating time backlight off 36 h
-10°C – 40°C
Loading time 3 h
4 KIB-TM-BA_IA-e-1912
2 Appliance overview
1. Weight display
2. Weighing unit
3. Keyboard
4. Connection of mains adapt er
5. RS232 interface
6. Input connection load cell cable
KIB-TM-BA_IA-e-1912 5
2.1 Keyboard overview
Button Function
• Turn on/off
• Zeroing
Navigation button
Navigation button
Navigation button
Navigation button
C
• Confirm entry
• Taring
• At numeric input increase flashing digit
• Scroll forward in menu
• Display sum total
• Digit selection to the right
• Add weighing value to summation memory
• Digit selection to the left
• Calculate weighing data via interface
• Delete
• Change between gross and net weight
ESC
+
+
+
6 KIB-TM-BA_IA-e-1912
• Back to menu/weighing mode
• Call up animal weighing function
• Retrieve tolerance weighing
• Delete total added memory
to access further digits and change them by
2.1.1 Numeric entry via navigation keys
Press , the current setting will be displayed. The first digit will be flashing
and is ready for changing.
If you do not wish to change the first digit, press and the second digit will
start flashing.
Each time you press , the display will move to the subsequent digit, after
the last digit the display will return to the first digit.
To change the selected (flashing) digit, press repeatedly until the desired
value is displayed. Then press
.
Complete your entry by .
KIB-TM-BA_IA-e-1912 7
2.2 Overview of display
Display Significance
W1 Weighing range 1
W2 Weighing range 2
Stability display
GROSS Gross weight
NET Net weight
PT Pre-Tare
hold Hold function
Pcs Parts counting
Kg Weighing unit
Rechargeable battery very low
Zero indicator
8 KIB-TM-BA_IA-e-1912
Totalization
Indicators for weighing with tolerance range
3 Basic Information (General)
3.1 Proper use
The display unit acquired by you is used in combination with a weighing plate and
serves to determine the weighing value of material to be weighed. It is intended to be
used as a “non-automatic weighing system”, i.e. the material to be weighed is
manually and carefully placed in the centre of the weighing plate. As soon as a stable
weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use display unit for dynamic weighing. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the display unit. (Example: Slowly
draining fluids from a container on the balance.)
Do not leave permanent load on the weighing pan. This may damage the measuring
system.
Impacts and overloading exceeding the stated maximum load (max) of the weighing
plate, minus a possibly existing tare load, must be strictly avoided. Both, the weighing
plate and the display unit may be damaged during this process.
Never operate display unit in explosive environment. The serial version is not
explosion protected.
Changes to the display unit's design are not permitted. This may lead to incorrect
weighing results, safety-related faul ts and des tr uct ion of the display unit.
The display unit may only be operated in accordance with the described default
settings. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
• Our conditions in the operation manual are ignored
• The appliance is used outside the described uses
• The appliance is modified or opened
• Mechanical damage or damage by media, liquids, natural wear and tear
• The appliance is improperly set up or incorrectly electrically connected
• The measuring system is overloaded
KIB-TM-BA_IA-e-1912 9
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the display
unit and, if applicable, the testing weight, must be checked regularly. The responsible
user must define a suitable interval as well as type and scope of this test. Information
is available on KERN’s home page (www.kern-sohn.com with regard to the
monitoring of display units’ test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and display units may be
calibrated (return to the national standard) fast and at moderate cost.
4 Basic Safety Precautions
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN balances.
4.2 Personnel training
The appliance may only be operated and maintained by train ed per sonnel.
5 T ransport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the
weighing platform, power unit etc. against shifting and
damage.
10 KIB-TM-BA_IA-e-1912
• Display Unit
6 Unpacking and installation
6.1 Installation Site, Location of Use
The display units are designed in a way that reliable weighing results are achieved in
common conditions of use.
Precise and fast work is achieved by selecting the right place for your display unit
and your weighing plate.
On the installation site observe the following:
• Place the display unit and the weighing plate on a stable, even surface.
• Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
• Protect the display unit and the weighing plate against direct draft from open
windows or doors.
• Avoid jarring during weighing;
• Protect the display unit and the weighing plate against high humidity, vapours
and dust.
• Do not expose the display unit to extreme dampness for longer periods of
time. Non-permitted condensation (condensation of air humidity on the
appliance) may occur if a cold appliance is taken to a considerably warmer
environment. In this case, acclimatize the disconnected appliance for ca. 2
hours at room temperature.
• Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change location or remove
source of interference.
6.2 Unpacking
Carefully remove the display unit from packaging, remove plastic cover and place it
in the designated work area.
and place it in the designated work area.
Mount the display unit in a way that facilitates operation and where it is easy to see.
6.3 Scope of delivery / serial accessories:
• Mains adapter
• Table leg
• Wall bracket
• Operating manual
KIB-TM-BA_IA-e-1912 11
Transport Securing
6.4 Transportation lock (illustration example)
Please note: if the display unit is used together with platform with transportation lock,
this transportation lock must be released prior to use:
6.5 Error message
As soon as an error message appears in the balance display,
the balance must not more be used, e.g. Err 4
6.6 Placing
Mount the display unit in a way that facilitates operation and where it is easy to see.
There are many ways of positioning the display unit, such as free-standing or wall-
mounted (optional).
Wall-mounted (optional) Free standing
12 KIB-TM-BA_IA-e-1912
Version with support base/wall bracket
Version with tripod EOC-A05 (optional)
In order to raise the display, the display unit can be mounted on an optional stand.
Usage with fitting panel EOC-A03 (optional):
1. Platform
2. Fitting panel
3. Display Unit
KIB-TM-BA_IA-e-1912 13
Check, whether the voltage acceptance on the scales is set correctly.
Do not connect the scales to the power mains unless the information on
consent by KERN.
6.7 Mains connection
6.8 Storage battery operation (optional)
Before the first use, the battery should be charged by connecting it to the mains
power supply for at least 12 hours.
If the weight display shows the flashing symbol , this is an indication that the
capacity of the rechargeable battery is almost exhausted. Charge the battery with the
help of the supplied power pack.
Select a country-specific power plug and insert it in the mains adapter.
the scales (sticker) matches the local mains voltage.
Only use KERN original mains adapter. Using other makes requires
Important:
Before starting your weighing balance, check the mains cable for
damage.
Ensure that the power unit does not come into contact with liquids.
Ensure access to mains plug at all times.
14 KIB-TM-BA_IA-e-1912
Observe stable environmental conditions. Stabilisation requires a certain
warm-up time.
6.9 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each display unit with conn ec ted weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
•Prepare the required adjustment weight.The adjustment weight to be
Adjustment switch:
The adjustment switch is located in the battery cartridge(see the photo).
used depends on the capacity of the weighing system. Carry out
adjustment as near as possible to the weighing system’s maximum
weight. Info about test weights can be found on the I nter n et at:
http://www.kern-sohn.com.
•
Adjustment switch in the “ADJ” setting:
• All menu items are available.
Adjustment switch in the “LOCK” position:
• Some of the menu items are locked (see the section “Menu overview”)
To unblock these menu items, place the adjustment switch in the ”ADJ”
position.
KIB-TM-BA_IA-e-1912 15
Call up menu:
Switch-on balance and during the selftest press .
„Pn“ will be displayed
Press , , , subsequently, the first menu block
„PO CHK“ will be displayed.
Press repeatedly until „P3 CAL“ will be displayed.
Acknowledge with . „CoUnt“ is displayed.
Press repeatedlyuntil „CAL“ will be displayed.
Acknowledge using , the current setting is displayed.
Press to select the desired setting and confirm by .
noLin = adjustment
LineAr = linearization, see chap. 6.10
16 KIB-TM-BA_IA-e-1912
• In balances with a resolution of > 15 000 dividing steps carrying out a
see chapter 3.4“Monitoring of test equipment”.
How to carry out adjustment:
Confir m me nu sett ing „noLin“ by .
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press .
The currently set adjustment weight will be displayed.
Either use adjustment weight displayed
or
change with the help of the navigation keys (see chap.2.1.1),
the active digit will be flashing.
Confirm value for adjustment weight by pressing .
Carefully place adjustment weight in the centre of the
(example)
weighing plate. Wait for stability display, then press .
”PASS“ will be shown briefly, followed by the weight of the
placed adjustment weight.
Remove adjustment weight, balance will return into weighing
mode automatically. An adjusting error or incorrect
adjustment weight will be indicated by the error message;
repeat adjustment procedure.
6.10 Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range. If linearity deviation is discovered during a monitoring of test resources, you
can improve this by means of linearization.
linearisation is recommended.
• Carrying out linearization is restricted to specialist staff possessing
well acquainted with the workings of weighing scales.
• The test weights to be used must be adapted to the weighing scale’s
specifications; see chapter “Monitoring of test equipment”.
• Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
• After successful linearisation you will have to carry out calibration;
KIB-TM-BA_IA-e-1912 17
Remove adjustment weight, balance will return into weighing
Procedure:
Call-up menu item P3 CALCalLiner, see chap. 6.9
Confirm by , the password query „Pn“ will be displayed.
Press , , subsequently.
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press .
When “Ld 1“ is displayed, put the first adjustment weight
(1/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press .
When “Ld 2“ is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press .
When “Ld 3“ is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform.
Wait for stability display, then press .
”PASS“ will be shown briefly, followed by the weight of the
placed adjustment weight.
mode automatically. An adjusting error or incorrect
adjustment weight will be indicated by the error message;
repeat adjustment procedure.
18 KIB-TM-BA_IA-e-1912
Overloading exceeding the stated maximum load (max) of the device, minus a
7 Operation
7.1 Start-up
Press , the appliance will carry out a self-test. As soon as the weight display
appears, the instrument will be ready to weigh.
7.2 Switching Off
Press , the display will disappear.
7.3 Zeroing
Resetting to zero corrects the influence of light soiling on the weighing plate. The unit
is equipped with an automatic zero setting function. Therefore the unit can be reset to
zero at any time as follows:
To unload the weighing system
Press and zero display as well as indicator will appear.
7.4 Simple weighing
Place goods to be weighed on balance.
Wait for stability display .
Read weighing result.
Overload warning
possibly existing tare load, must be strictly avoided. This could damage the
instrument.
Exceeding the maximum load is indicated by the display
and an audio sound. Unload weighing system or reduce preload.
KIB-TM-BA_IA-e-1912 19
7.5 Switch-over weighing unit
How to enable weighing units:
Invoke menu item P5 Unt, see chap. 8
Press and the first weighing unit with the current setting
will be displayed.
To enable [on] / disable [off] the displayed weighing unit,
press
Acknowledge with . The next unit with the current setting
will be displayed.
To enable [on] / disable [off] the displayed weighing unit,
press
Acknowledge with .
Repeat seque nc e for each weighing unit.
Note:
„tj“ and „Hj“ cannot be activated at the same time,
only either ... or ... .
Press back to weighing mode several times.
Switch-over weighing unit:
Keep pressed, the display changes over to the weighing
units activated before (e.g. kg lb)
20 KIB-TM-BA_IA-e-1912
7.6 Weighing with tare
Deposit weighing vessel. After successful standstill control press the button.
Zero display and indicator NET appear.
The weight of the container is now internally saved.
Weigh the material, the net weight will be indicated.
The weight of the weighing container will be displayed as a minus number after
removing the weighing container.
The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The limit
is reached when the taring range capacity (see type plate)is full.
To change between gross weight and net weight, press .
To delete the tare value, remove load from weighing plate and press .
7.6.1 Pre-Tare
It is possible to enter a pre-tare value beforehand.
Ensure that there are no objects on the weighing pan.
Press and hold until 0.0 is displayed and the left digit is flashing
Use the arrow keys to enter the pre-tare value and confirm by pressing
The pre-tare value will be shown as a negative value.
KIB-TM-BA_IA-e-1912 21
An acoustic signal sounds when load is beyond tolerance
7.7 Weighing with tolerance range
You can set an upper or lower limit when weighing with tolerance range and thus
ensure that the weighed load remains exactly within the set limits.
During tolerance tests such as dosing, portioning and sorting the unit will indicate
exceeded or undershot limits by emitting an optical or acoustic signal.
Audio signal:
The acoustic signal depends on the settings in menu block „BEEP“.
Options:
• no
• ok
Acoustic signal turned off
An acoustic signal sounds when load is within tolerance
limits
• ng
limits
Optical signal:
The indicators show whether the load is within the two set tolerance
limits.
Target quantity / target weight within tolerance range
Target quantity / target weight exceeds maximum tolerance limit
The settings for tolerance check may be called up either via menu block „P0 CHK“
(see chap. 8) or faster via the key combination
+
22 KIB-TM-BA_IA-e-1912
7.7.1 Tolerance check for target weight
Settings
Press and at the same time in weighing mode.
Press until the display for entering the lower limit value
appears.
Press , the current setting will be displayed.
To enter the lower limit, e. g. 1000 Kg, press the navigation
keys (See chap. 2.1.1); the currently enabled digit will be
flashing.
Confir m input by .
Press repeatedly until is displayed.
Press , the current setting for the upper limit will be
displayed.
Press the navigation keys (See chap. 2.1.1) to enter the
upper limit, e.g. 1100 kg; the currently enabled digit will be
flashing.
Confir m input by .
Press repeatedly until is displayed.
KIB-TM-BA_IA-e-1912 23
Press , the current setting for the acoustic signal will be
shown.
Select desired setting (no, ok, ng, s. chap. 8) by .
Confir m input by .
Press ; weighing system is in tolerance weighing mode.
From here evaluation takes place whether the goods to be
weighed are within the two tolerance limits.
Weighing with tolerance range
Tare when using a weighing container.
Put on goods to be weighed, tolerance control is started. The signal lights
indicate whether the load is within the two set limits.
Load below specified
tolerance
Indicator will be
displayed
• The tolerance control is not active when the weight is under 20d.
• To delete limits, enter “00.000 kg“.
Load within specified
tolerance
Indicator will be
displayed
Load exceeds specified
tolerance
Indicator will be
displayed
24 KIB-TM-BA_IA-e-1912
7.7.2 Tolerance check for target quantity
Settings
Press and at the same time in weighing mode.
Press until the display for entering the lower limit value
appears.
Press , the current setting will be displayed.
To enter the lower limit, e. g. 75 items, press the navigation
buttons (see chap. 2.1.1); the currently enabled digit will be
flashing.
Confir m input by .
Press repeatedly until is displayed.
Press , the current setting for the upper limit will be
displayed.
To enter the upper limit, e. g. 100 items, press the navigation
buttons (see chap. 2.1.1); the currently enabled digit will be
flashing.
Confir m input by .
Press repeatedly until is displayed.
KIB-TM-BA_IA-e-1912 25
Press , the current setting for the acoustic signal will be
shown.
Select desired setting (no, ok, ng, s. chap. 8) by .
Confir m input by .
Press ; weighing system is in tolerance weighing mode.
From here evaluation takes place whether the goods to be
weighed are within the two tolerance limits.
Weighing with tolerance range
Set item weight, see chap. 7.10.
Tare when using a weighing container.
Put on goods to be weighed, tolerance control is started. The signal lights
indicate whether the load is within the two set limits.
Load below specified
tolerance
Indicator will be
displayed
• The tolerance control is not active when the weight is under 20d.
• To delete limits, enter „00000 PCS“.
Load within specified
tolerance
Indicator will be
displayed
Load exceeds specified
tolerance
Indicator will be
displayed
26 KIB-TM-BA_IA-e-1912
• Menu setting:
• The totalizing function is not active when the weight is under 20d.
7.8 Manual totalizing
With this function the individual weighing values are added into the summation
memory by pressing and edited, when an optional printer is connected.
Add up:
Place load A.
Wait until the stability display appears, then press . The weight value will
be saved and printed if an optional printer is connected.
Remove the weighed good. More weighed goods can only be added when the
display ≤ zero.
Place good to be weighed B.
Wait until the stability display appears, then press . The weight value will be
added to the summation memory and printed, as required. The number of
weighings, the whole weight and the currently placed weight will be shown in
succession.
„P2 COM“ „MODE“ „PR2““, see chap. 8
Add more weighed goods as described before.
Please note that the weighi ng system must be unloaded between the individual
weighing procedures.
This process may be repeated 99 times or till such time as the capacity of the
weighing system has been exhausted.
Display and edit sum „Total“:
Press , the number of weighings, followed by the total weight will be displayed
for 2 sec. Press to print out this display.
KIB-TM-BA_IA-e-1912 27
Sample log (KERN YKB-01N):
Delete weighing data:
Press and at the same time The data in the summation memory are
deleted.
Menu setting “P2 COM“ “Lab 2“ / Prt 4 - 7“
First weighing
Second weighing
Third weighing
Number of
+
weighings/
Total
For additional sample logs see chap. 0
28 KIB-TM-BA_IA-e-1912
• Menu settings:
7.8
7.9 Automatic adding-up
With this function the individual weighing values are automatically added into the
summation memory when the bala nce is unloaded without pressi n g and edited,
when an optional printer is connected.
„P2 COM „MODE“ „AUTO““, s. chap. 8
Indicator will be displayed.
Add up:
Place load A.
After the standstill control sounds a signal tone. The weighing value will be added
to the summation memory and printed.
Remove the weighed good. More weighed goods can only be added when the
display ≤ zero.
Place good to be weighed B.
After the standstill control sounds a signal tone. The weighing value will be added
to the summation memory and printed, as required. The number of weighings,
followed by the currently placed weight will be shown in succession.
Add more weighed goods as described before.
Please note that the weighing system must be unloaded between the individual
weighing procedures.
This process may be repeated 99 times or till such time as the capacity of the
weighing system has been exhausted.
For how to display and delete weighing data as well as sample logs see chap.
KIB-TM-BA_IA-e-1912 29
7.10 Parts counting
Before the balance can count parts, it must know the average part weight (i.e.
reference). Proceed by putting on a certain number of the parts to be counted. The
balance determines the total weight and divides it by the number of parts, the socalled reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
As a rule:
The higher the reference quantity the higher the counting exactness.
In weighing mode , press and hold until the message
„P 10“ appears that is used to set the reference quantity.
Use to set the desired reference quantity (such as 100),
options include P 10, P 20, P 50, P100, P 200.
Place as many items to be counted (such as 100 items) as
demanded by the set reference quantity and confirm by .
The weighing scales calculate the reference weight. The
current quantity (such as 100 items) will be displayed.
Remove reference weight. The balance is from now in parts
counting mode counting all units on the weighing plate.
Back to Weighing mode by .
30 KIB-TM-BA_IA-e-1912
7.11 Percent weighing
Percentage weighing allows you to display the weight in percent related to
reference weight.
Press in weighing mode (approx. 3 sec.) until “rEF 10%“
is displayed.
Use to set the desired percentage value to be applied
Place a sample matching the set percentage value on the
weighing platform and pr ess . “-------%“ will be briefly
displayed.
The percentage value for the sample will be displayed.
Remove reference weight
The display returns to „0.0 %“
Put on specimen
In the display appears the percentage of the specimen with
reference to the reference weight.
Back to weighing mode by pressing the button.
KIB-TM-BA_IA-e-1912 31
7.12 Animal weighing
The animal weighing function is suitable for weighing restless loads.
The weighing system will display a mean value derived from several weighing results.
The animal weighing program can be enabled by either calling up menu block
„P4 OTH“ „ANM“ „ON“ (see chap. 8) or faster via key combination.
+
The indicator shows HOLD as long as the animal weighing function remains enabled.
Place the load on the weighing system and wait until the scale is steady.
Press and at the same time; you will hear an acoustic signal, indicating
that the mean value function is ena bled.
Whilst averaging is taking place you can add or remove loads as the measuring
value will be constantly updated.
To deactivate the animal weighing function press and at the same time.
32 KIB-TM-BA_IA-e-1912
7.13 Lock keyboard
In the menu item „P4 OTH“ „LOCK“ (see chap. 8) the keyboard lock can be
enabled/disabled.
Whilst the function is enabled the keyboard will self-lock after no key has been
pressed for 10 minutes. „K-LCK“ will be displayed as soon as a key is pressed.
To disable the lock, keep , and pressed at the same time (2 s) until
„U LCK“ appears.
7.14 Display background illumination
Keep pressed (3s) until „setbl“ appears.
Press again, the current setting will be displayed.
Press to select the desired setting.
bl on Continuous background lighting
bl off Background illumination off
bl Auto Automatic background illumination on when weighing pate is
loaded or key pressed.
Either save by or cancel input by pressing .
Back to Weighing mode by .
KIB-TM-BA_IA-e-1912 33
7.15 Automatic switch-off function „AUTO OFF“
The unit is automatically switched off within the preset time when the display unit or
the weighing bridge are not operated.
Keep pressed (3s) until „setbl“ appears.
Press to call up AUTO OFF-function
Press , the current setting will be displayed.
Press to select the desired setting.
of 0 Function deactivated
of 3 Weighing sy s tem w il l be turned off after 3 mi n .
of 5 Weighing sy s tem w il l be turned off after 5 mi n .
of 15 Weighing system wi ll be turned off a fter 15 mi n.
of 30 Weighing system will be turned off after 30 min.
Either save by or cancel input by pressing .
Back to Weighing mode by .
34 KIB-TM-BA_IA-e-1912
A B b C D E F G H h I J K
L
7.16 Setting time and date
To change date and time go to menu item "P8 ind" "dAtE" or "tIME"
(See chap. 8):
Setting date:
Select menu item "dAtE"
Press to confirm and the last date entered will be displayed. The first digit is
flashing
(example)
Use the navigation keys, as described in chap. 2.1.1, to enter the current date
and press to confirm.
The current date will be shown in standby mode.
Setting time:
Select menu item "tiME"
Press to confirm and the last time entered will be shown. The first digit is
flashing.
(example)
Use the navigation keys, as described in chap. 2.1.1, to enter the current time
and press to confirm.
The time will be shown in standby mode.
7.17 Alphabet
Letters are ordered in the following order:
l M N O P Q R S U V W X Y Z
KIB-TM-BA_IA-e-1912 35
Enter the password (see above.)
selected one after the other.
current setting will be displayed.
.
8 Menu
8.1 Navigation in the menu:
Call up menu
Switch-on balance and during the selftest press .
Press , , , subsequently, the first menu
block „PO CHK“ will be displayed.
From the weighing mode:
Press and hold until Pn appears.
Select menu block
With help of , the individual menu items can be
Select setting
Confirm selected menu item by pressing . The
Change settings To change to the available settings, press the
navigations keys as described in chap. 2.1.1.
Acknowledge setting /
exit the menu
Return to weighing
mode
36 KIB-TM-BA_IA-e-1912
Either save by pressing or cancel by pressing
Press repeatedly to exit menu.
Menu block
Main menu
Menu item
Submenu
Acoustic signal for weighing with tolerance range
switched off
Audio sound when weighed load is within
tolerance limits
Audio sound when weighed load is beyond
tolerance limits
Relay pilot light
P1 rEF1
P2 COM
MODE
8.2 Menu overview
Available settings / explanation
PO CHK
Weighing with
tolerance range,
see chap. 7.7
Zero point
settings
nEt H Upper limit value „Tolerance check weighing“, input see
chap. 7.7.1
nEt L
PCS H
Lower limit value „Tolerance check weighing“, input see
chap. 7.7.1
Upper limit value „Tolerance check counting“, input
see chap. 7.7.2
PCS L
Lower limit value „Tolerance check counting“, input
see chap. 7.7.2
BEEP no
ok
nG
rELAY on
oFF
A2n0 Automatic zero point correction (Autozero) by changing the
display, digits selectable (0, 0.5d, 1d, 2d, 4d)
0AUto Zero setting range
Load range where the display after switching-on the
balance is set to zero. Selectable 0, 2, 5, 10, 20, 30, 50,
100 %
0rAGE Zero setting range
0tArE Automatic taring „on / off“, taring range adjustable in menu
Interface
parameter
Load range where the display is set to zero by pressing
. Selectable 0, 2, 4, 10 , 20* , 50, 100%.
item „0Auto“.
CONT S0 offContinuous data output,
S0 on
ST1 One output for stable weighing value
STC Continuous data output of stable weighing
values
PR1
• Output after pressing
• Precondition for alibi memory
PR2 Manual totalizing, see chap. 7.8
Press and the weighing value will be added
to the summation memory and issued.
selectable „sending 0“, yes / no
KIB-TM-BA_IA-e-1912 37
AUTO*
For automatic totalizing see chap. 7.9
memory on unloading of weighing scale.
ASK
Remote control instructions
wirel
Not documented
BAUD
Available Baudrate: 600, 1200, 2400, 4800, 9600*
Pr
7E1
7 bits, even parity
7o1
7 bits, odd parity
8n1*
8 bits, no parity
PTYPE
tPUP*
Standard printer setting
LP50
Not documented
KCP
KERN Communication Protocol
LAb
LAb x
For data output format,
see table below. 1
Prt
Prt x
LAnG
eng*
Standard settings English
chn
Not documented
P3 CAL1
COUNT
Display internal resolution
DECI
Position of the decimal dot
DUAL
Setting balance type, capacity (Max) and readability (d)
off
Single-range balance
R1 inc
Readability
R1 cap
Capacity
on
Dual range balance
R1 inc
Readability 1st weighing range
R1 cap
Capacity 1st weighing range
R2 inc
Readability 2nd weighing range
R2 cap
Capacity 2nd weighing range
CAL
noLin
For adjustment, see chap. 6.9.2
Liner
For linearization, see chap. 6.10.2
GrA
Gravitational constant at place of installation
GrB
Gravitational constant at place of manufacture
on
Keyboard lock enabled, see chap. 7.13
off*
Keyboard lock disabled
on
Animal weighing enabled, see chap. 7.12
off*
Animal weighing disabled
SCr
on
watch as screensaver enabled
off*
watch as screensaver disabled
Configuration
data
see chap. 12.4
This function is used to issue and add individual
weighing values automatically to the summation
LOCK
ANM1
P4 OTH
38 KIB-TM-BA_IA-e-1912
kg
on* off g on
off*
lb
on
off*
oz
on
off*
tJ
on
off
HJ
on
off
dAtE
Setting date: Format: TTMMJJ
tIME
Setting time: Format: HHMMSS
ALibi
Alibi memory
dAtA
Number of saved records
rdAtA
Read the record value
ErASE
Delete all data
ExPT
Export data (USB stick)
PrEt
Enter pre-tare value
485
ModE
2disP,
Count
Export mode (2nd display)
bAUd
600,
9600
Baud rate
Pr
7o1
7 Bit, odd Parity, 1 Stop bit
7E1
7 Bit, equal Parity, 1 Stop bit
8n1
8 Bit, no Parity, 1 Stop bit
io
i_tSt
Test input
o_tSt
Test output
oPt
intF
USB,
WiFi, EnEt
Select connections
ModE
(output)
no, CoUnt (USB, Bt, Wi-Fi, EnEt)
no, Expt (UdiSK)
iP_1
IP addresses KIB-TM
iP_2
iP_3
iP_4
MASK_1
Subnet mask
MASK_2
MASK_3
MASK_4
GAtE_1
KIB-TM Gateway
GAtE_2
GAtE_3
GAtE_4
P5 Unt1
Change weighing
unit
see chap. 7.5
P6 xcl1
P7 rst1
Factory setting
P8 ind
P9 Prt
Not documented
Use to reset balance settings to factory default.
1200,
2400,
4800,
KIB-TM-BA_IA-e-1912 39
UdiSK, Bt,
oPt
riP_1
remote (IP-Adresse PC)
riP_2
riP_3
riP_4
rPort
Remote port (Port for communication
between PC and KIB-TM
SSid_1
SSID
SSid_2
PSW_1
WLAN Passwo rd
PSW_2
Lab
Continuation menu item P9 Prt
P9 Prt
Factory settings are marked by *.
1
Function blocked when the adjustment switch is in the position “balance is
calibratable” (adjus tment switch in the “L OC K ” pos ition).
NT Net weight
TW Tare weight
NO Number weighing processes
Total Total of all individual
weighings
40 KIB-TM-BA_IA-e-1912
9 Servicing, maintenance, disposal
9.1 Cleaning
• Before cleaning, disconnect the appliance from the operating voltage.
• Do not use aggressive detergents (solvents or similar).
9.2 Servicing, maintenance
The appliance may only be opened by trained service technicians who are authorized
by KERN.
Before opening, disconnect from power supply.
9.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.
KIB-TM-BA_IA-e-1912 41
Error message
Description
9.4 Error messages
Excess load if weight exceeds capacity of +9d
Underweight (less than 20 d)
The weight is too low (less than -20 d)
Zero setting range exceeded on start-up of balance.
Zero setting range during start-up of weighing scale or on
pressing of exceeded.
Incorrectly connect ed verification plug
Is displayed on setting weighing scale to zero with ,
without load
For parts counting and percentage calculation:
Weighed value < zero
For add-up:
Total number of weighings above 999
For add-up:
Total weight above 999999
Adjustment failed
Keyboard locked
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
42 KIB-TM-BA_IA-e-1912
Keyboard unlocked
Capacity of batteries exhausted.
(Battery voltage below 5.7 V, automatic shutdown
happens at less than 5.4 V)
Pin1
RX
Input
Pin2
TX
Output
Pin3
GND
Signal ground
Pin4
N/C
Not connected
10 RS 232 interface
You can print weighing data automatically via the RS 232C interface or manually by
pressing via the interface according to the setting in the menu.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communi cati on between
the weighing system and the printer.
• Use a suitable cable to connect the display unit to the interface of the printer.
Faultless operation requires an adequate KERN interface cable.
• Communication parameters (ba ud ra te, bit s and parity) of display unit and
printer must match. For a detailed description of interface parameters, please
refer to chapter 8, Menu block “P2 COM“
10.1 Technical data
Connection 4 pin d-subminiature bushing
Baud rate Optional 600/1200/2400/4800/9600
Parity 8 bits, no parity / 7 bits, even parity / 7 bits, odd parity
KCP (KERN communication protocol) contains the commands that are used to
control the KERN balances via the interface.
• Menu setting P2 Com Mode ASK
• Menu setting P2 Com PTYPE KCP
• Finish commands with CR/LF character.
• Consult the KCP manual for more inform ati o n , av ai labl e on
our KERN website (www.kern-sohn.com).
The following commands are supported:
@ Cancel
List all implemented KCP commands
Query KCP level and KCP versions
Query device information (type, capacity)
Query device software version
Set serial number (default value is K123456)
Query SW-Identification number
46 KIB-TM-BA_IA-e-1912
Send stable weight value
Send weight value immediately and repeat
Z
ZI
D
D˽“˽“
DW
K
SR
T
MM
TAC
TI
Tare immediately
Zero
Zero immediately
Display: Write text to display
Clear Display (after D-Command)
Display: Show weight
Keys: Set configuration
Send weight value on weight change (send and repeat)
Tare
Query/preset tare weight value
Clear tare value
Polling-Intervall
• The time between periodic inquiries or when sending requests
(queries) by the interface must be longer than 100 ms.
KIB-TM-BA_IA-e-1912 47
The displayed weight does
not glow.
• Mains power supply interrupted (mains cable
defective).
• Power supply interrupted.
• (Rechargeable) batteries are inserted incorrectly or
empty
• No (rechargeable) batteries inserted.
The displayed weight is
• Table/floor vibrations
• Weighing pan has contact with other objects.
possible)
The weighing result is
obviously incorrect
• Adjustment is no longer correct.
• Great fluctuations in temperature.
possible)
11 Instant help
In case of an error in the program process, briefly turn off the display unit and
disconnect from power supply. The weighing process must then be restarted from the
beginning.
Help:
Fault Possible cause
• The display unit is not switched on.
permanently changing
Should other error messages occur, switch display unit off and then on again. If the
error message remains inform manufacturer.
different location/switch off interfering device if
48 KIB-TM-BA_IA-e-1912
Installation / configuration of a weighing system must be carried out by a
12 Installing display unit / weighi ng bridge
•
12.1 Technical data
Supply voltage 12 V / 1000mA
Max. signal voltage 5V
Zeroing range 0-2mV
Sensitivity ≥ 0.15uV/d
Resistance parameter 87 - 1100 Ω
12.2 Weighing system design
The display unit is suitable for connection to any analogue load cell in compliance
with the required specifications.
The following data must be established before selecting a load cell:
well acquainted specialist with the workings of weighing balances.
•Weighing balance capacity
This usually corresponds to the heaviest load to be weighed.
•Preload
This corresponds to the total weight of all parts that are to be placed on the
weighing cell such as upper part of platform, weighing pan etc.
•Total zero setting range
This is composed of the start-up zero setting range (± 2%) and the zero
setting range available to the user via the ZERO-key (2%). The total zero
setting range equals therefore 4 % of the scale’s capacity.
The addition of weighing scales capacity, preload and the total zero setting
range give the required capacity for the weighing cell.
To avoid overloading of the weighing cell, include an additional safety
margin.
•Smallest desired display division
KIB-TM-BA_IA-e-1912 49
In 4-wire load cells or connection boxes (junction boxes), bridge the solder
EXC+
SEN+
SEN-
EXC-
SIG+
SIG-
SHIELD
7
6
3
2
5
4
1
LOAD
CELL
12.3 How to connect the platform
Disconnect the display unit from the power supply.
Solder the individual wires of the load cell cable to the printed circuit board (See
fig.).
joints J1 and J2.
50 KIB-TM-BA_IA-e-1912
12.4 Configur e display unit
+ For menu overview see chap. 8.
Call up menu:
Switch-on balance and during the selftest press . „Pn“
will be displayed
Press , , , subsequently, the first menu block
„PO CHK“ will be displayed.
Press repeatedly until „P3 CAL“ will be displayed.
Acknowledge with . „CoUnt“ is displayed.
Navigation in the menu
With help of , the individual menu items can be selected
one after the other.
Confirm selected menu item by pressing . The current
setting will be displayed.
To change to the available settings, press the navigations
keys as described in chap. 2.1.1.
Either save by pressing or reject by pressing .
Press repeatedly to exit menu.
Parameter selection
1. Display internal resolution
Press , the internal resolution will be shown.
Return to menu by .
Use to select another men u ite m.
KIB-TM-BA_IA-e-1912 51
2. Position decimal point
Press , the currently set position of the decimal dot is
displayed.
Press to select the desired setting. Selectable 0, 0.0,
0.00, 0.000, 0.0000.
Confirm input by .
Use to select another men u ite m.
3. Weighing scales type, capacity and readability
Press , the current setting will be displayed.
Select des ir ed s etti ng by .
„off“ Single-range balance
„on“ Dual range balance
Press to confirm, the display for entering readability (for
dual range scales for the first weighing range) appears.
Press , the current setting will be displayed.
52 KIB-TM-BA_IA-e-1912
Select des ir ed s etti ng with and acknowledge by .
Press , the display for entering capacity will appear (at
dual range balance for the first ra nge).
Press , the current setting will be displayed.
Using the navigation buttons (see chap. 2.1.1) select the
desired setting, the active digit is flashing.
Acknowledge with .
In a single-range balance the entry of capacity / readability
is finished.
either in single-range balance
Press , the unit will return to the menu Press to call
up next menu item „CAL“.
or
In a dual range balance enter readability and capacity of
the second weighing range.
Press , the display for entering the readability of the
second weighing range will appear.
Press , the current setting will be displayed.
Select des ir ed s etti ng with and acknowledge by .
KIB-TM-BA_IA-e-1912 53
6.10 for linearisation
Press , the display for entering the capacity of the
second weighing range will appear.
Press , the current setting will be displayed.
Using the navigation buttons (see chap. 2.1.1) select the
desired setting, the active digit is flashing.
Confir m input by .
Press , the unit will return to the menu
Press to call up next menu item „CAL“.
4. Adjustment or linearisation
Adjustment or linearisation is required after entering
configuration data.
For carrying out adjustment see chap. 6.9 step 4 or chap.
Acknowledge using , the current setting is displayed.
Press to confirm, pres s to select the desired setting
noLin = Adjustment
LineAr = Linearisation
13 Conformity explanation/ t est certificate
The curre nt EC/EU Conformity declaration can be found online in:
www.kern-sohn.com/ce
54 KIB-TM-BA_IA-e-1912