1 Technical data
KERN
ALT
160-4B
ALT
250-4B
ALT
220-5DA
ALT
310-4B
Weighing range (max)
160 g
250 g
60 g; 220 g
310 g
Readability (d)
0.1 mg
0.1 mg
0.00001 g;
0.0001 g
0.1 mg
Reproducibility
0.1 mg
0.1 mg
0.0001 g
0.2 mg
Linearity
± 0.3 mg
± 0.3 mg
0.0003 g
± 0.3 mg
Stabilization time (typical)
4 s
4 s
4 s
4 s
Minimum unit weight at
piece counting
> 0.5 mg
> 0.5 mg
> 0.5 mg
> 0,5 mg
Warm-up time
8 hours
Adjustment weight
internal
Reference quantities at
piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, dwt, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw)
Secondary voltage powerpack
24 V AC, 500mA
Operating temperature
+ 5° C …. + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 330
Dimensions of windshield
(B x D x H) mm
160 x 140 x 205 (inside)
190 x 195 x 225 (outside)
Weighing pan (stainless steel)
Ø 80 mm
Weight kg (net)
5.7 kg
6.5 kg
Interface
RS 232C
4 ALT_B-PLT_A-PLT_F-BA-e-1333
KERN
ALT 160-4BM
ALT 250-4BM
Weighing range (max)
160 g
250 g
Readability (d)
0.1 mg
0.1 mg
Reproducibility
0.1 mg
0.1 mg
Linearity
± 0.3 mg
± 0.3 mg
Verification value (e)
1 mg
1 mg
Verification class
I
I
Minimum load (Min)
10 mg
10 mg
Stabilization time (typical)
4 s
4 s
Minimum unit weight at
piece counting
> 0.5 mg
> 0.5 mg
Warm-up time
8 hours
Adjustment weight
internal
Reference quantities at
piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, g
Secondary voltage powerpack
24 V AC, 500mA
Operating temperature
+ 5° C …. + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 330
Dimensions of windshield
(B x D x H) mm
160 x 140 x 205 (inside)
190 x 195 x 225 (outside)
Weighing pan (stainless steel)
Ø 80 mm
Weight kg (net)
5.7 kg
Interface
RS 232C
ALT_B-PLT_A-PLT_F-BA-e-1333 5
KERN
PLT 720-3A
PLT 1200-3A
PLT 6200-2A
Weighing range (max)
720 g
1200 g
6200 g
Readability (d)
1 mg
1 mg
10 mg
Reproducibility
1 mg
1 mg
10 mg
Linearity
± 2 mg
± 2 mg
± 30 mg
Stabilization time (typical)
2 s
2 s
2 s
Minimum unit weight at
piece counting
> 5 mg
> 5 mg
> 50 mg
Warm-up time
4 hours
8 hours
4 hours
Adjustment weight
internal
Reference quantities at
piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, dwt, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw)
Input Voltage
220 - 240V AC 50 Hz
Secondary voltage
powerpack
24 V AC, 500mA
Operating temperature
+ 5° C / + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 160
210 x 340 x 100
Dimensions of windshield
Ø 150, 60 high (inside)
Ø 160, 70 high (outside)
-
Weighing pan (stainless
steel)
Ø 11 cm
Ø 16 cm
Weight kg (net)
4.5 kg
Interface
RS 232C
Underfloor weighing device
Eyelet
6 ALT_B-PLT_A-PLT_F-BA-e-1333
KERN
PLT 420-3F
PLT 4200-2F
Weighing range (max)
420 g
4200 g
Readability (d)
1 mg
10 mg
Reproducibility
1 mg
10 mg
Linearity
± 4 mg
± 40 mg
Stabilization time (typical)
2 s
2 s
Minimum unit weight at
piece counting
> 5 mg
> 50 mg
Warm-up time
4 hours
Adjustment weight
internal
Reference quantities at
piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw), tol
Input Voltage
220 - 240V AC 50 Hz
Secondary voltage
powerpack
9 V 1000 mA, DC
Operating temperature
+ 5° C / + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 160
Dimensions of windshield
Ø 150, 60 high (inside)
Ø 160, 70 high (outside)
Weighing pan (stainless
steel)
Ø 11 cm
Ø 16 cm
Weight kg (net)
4 kg
Interface
RS 232C
Underfloor weighing device
-
ALT_B-PLT_A-PLT_F-BA-e-1333 7
KERN
PLT 720-3AM
PLT 6200-2AM
Weighing range (max)
720 g
6200 g
Readability (d)
1 mg
10 mg
Reproducibility
1 mg
10 mg
Linearity
2 mg
30 mg
Verification value (e)
0,01 g
0,1 g
Verification class
II
II
Minimum load (Min)
20 mg
50 mg
Stabilization time (typical)
2 s
2 s
Minimum unit weight at
piece counting
> 5 mg
> 50 mg
Warm-up time
4 hours
Adjustment weight
internal
Reference quantities at
piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, g
Input Voltage
220 - 240V AC 50 Hz
Secondary voltage
powerpack
24 V DC, 500 mA,
Operating temperature
+ 5° C / + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 160
210 x 340 x 100
Dimensions of windshield
Ø 150, 60 high (inside)
Ø 160, 70 high (outside)
-
Weighing pan (stainless
steel)
Ø 11 cm
Ø 16 cm
Weight kg (net)
4.5 kg
4.5 kg
Interface
RS 232C
Underfloor weighing device
Eyelet
8 ALT_B-PLT_A-PLT_F-BA-e-1333
2 Appliance overview
Models ALT
Models PLT, d = 1 mg
2.1 Operating side
ALT_B-PLT_A-PLT_F-BA-e-1333 9
Models PLT, d = 10 mg
Windshield
Levelling screw
Weighing pan
Bubble level
Display
Cover windshield
Keyboard
10 ALT_B-PLT_A-PLT_F-BA-e-1333
2.1.1 Back and bottom side
1. Weighing pan
2. RS232C-interface
3. Footscrews
4. Mains adapter
connection
5. Casing screws
(in models with four
feet first screw out
both rear feet)
6. Underfloor weighing
device
(remove cover)
ALT_B-PLT_A-PLT_F-BA-e-1333 11
2.2 Keyboard overview
Key
Description
Description
MENU button
Invoke main menu
How to select menu items
Press until
the acoustic
signal gets
mute
Invoke setup menu
Exit setup menu
ON/OFF
button
Turn on/off
Exit main menu
CAL-key
Adjustment
How to select menu items
PRINT button
Calculate weighing data via interface
Confirm / store settings
TARE button
Taring
Zeroing
12 ALT_B-PLT_A-PLT_F-BA-e-1333
2.2.1 Navigation buttons / numeric input
Key
Description
Function
Navigation button
Increase digit
Scroll forward in menu
Set decimal place
Navigation button
Reduce digit
Scroll backwards in menu
Change capital/small letters
Press MENU button until the acoustic signal gets
mute
Navigation button
Position digit
Delete input
Navigation button
Store
ESC
cancel
ALT_B-PLT_A-PLT_F-BA-e-1333 13
2.3 Overview of display
1
Capacity display
2
Stability display
3
Zero indicator
4
Numerical display/menu
5
Unit
6
Current time
7
Current date
Display
Description
see chap.
Stability display
+Chap. 10.3
Zero indicator
+Chap. 10.2
Balance is in percent determination mode
+Chap. 12.5
PC
Balance is in parts counting mode
+Chap. 12.1
H
Upper limit
+Chap.12.4.
L
Lower limit
DS
The balance is in density determination mode
+Chap. 12.2
g
Weighing unit Gram
Chap. 11.1
ct
Weighing unit Carat
Oz
Weighing unit Ounce
Lb
Weighing unit Pound
Dwt
Weighing unit Pennyweights
Ozt
Weighing unit Troy ounce
GN
Weighing unit Grain
tl 1
Weighing unit Hongkong Tael
tl 2
Weighing unit Singapore Tael
tl 3
Weighing unit Taiwan Tael
mo
Weighing unit Momme
14 ALT_B-PLT_A-PLT_F-BA-e-1333
Operator guidance:
8
Active application
9
Current date
10
Current time
11
Operation step to be carried out
In a selected application you are guided step-by-step through that application.
Selectable languages (D, GB, F, IT. ESP, P; see chap. 11.11).
Display example „Piece counting“
ALT_B-PLT_A-PLT_F-BA-e-1333 15
3 Basic instructions
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic balcance”, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
pan.. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“. (Example: Slowly draining fluids from a
container on the balance.)
Do not leave permanent load on the weighing pan. This may damage the measuring
system.
Impacts and overloading exceeding the stated maximum load (max) of the balance,
minus a possibly existing tare load, must be strictly avoided. Balance may be
damage by this.
Never operate balance in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
1. Our conditions in the operation manual are ignored
2. The appliance is used outside the described uses
3. The appliance is modified or opened
4. Mechanical damage or damage by media, liquids, natural wear and tear
5. The appliance is improperly set up or incorrectly electrically connected
6. The measuring system is overloaded
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this
test. Information is available on KERN’s home page (www.kern-sohn.com with
regard to the monitoring of balance test substances and the test weights required for
this. In KERN’s accredited DKD calibration laboratory test weights and balances may
be calibrated (return to the national standard) fast and at moderate cost.
16 ALT_B-PLT_A-PLT_F-BA-e-1333
4 Basic Safety Precautions
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN
balances.
All language versions contain a non-binding translation. The
original German is binding.
Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
4.1 Pay attention to the instructions in the Operation Manual
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
ALT_B-PLT_A-PLT_F-BA-e-1333 17
18 ALT_B-PLT_A-PLT_F-BA-e-1333
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
On the installation site observe the following:
Place the balance on a firm, level surface
Avoid extreme heat as well as temperature fluctuation caused by installing next
to a radiator or in the direct sunlight
Protect the balance against direct draughts due to open windows and doors
Avoid jarring during weighing
Protect the balance against high humidity, vapours and dust
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance)
may occur if a cold appliance is taken to a considerably warmer environment.
In this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
Avoid static charge of goods to be weighed and weighing container.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change location or remove
source of interference.
ALT_B-PLT_A-PLT_F-BA-e-1333 19
6.2 Unpacking, Scope of delivery
Balance, see chap. 2 Mains adapter Protective cover Operating instructions Transport Securing
Remove device and accessories carefully from packaging, remove packaging
material and place device at the planned work place. Verify that there has been no
damage and that all packing items are present.
Scope of delivery / serial accessories:
6.3 Placing
The right place is decisive for the accuracy of the weighing results of high-resolution
analytical and precision balances (see chap. 6.1).
Remove transportation lock
On the left lower side of the balance replace the golden-coloured screw with the
silver-coloured one (see label).
20 ALT_B-PLT_A-PLT_F-BA-e-1333
Install weighing pan as per illustration
Models ALT Models PLT, d = 1 mg
Models PLT, d = 10 mg
ALT_B-PLT_A-PLT_F-BA-e-1333 21
Levelling balance
Level balance with foot screws until the air bubble of the
water balance is in the prescribed circle.
Check levelling regularly.
6.4 Connect and switch-on power supply
Before connecting the mains adapter check if the printed voltage value is the same
as the local supply voltage. Only use KERN original mains adapter. Using other
makes requires consent by KERN.
Supply power to balance via mains adapter.
The balance will carry out a self-test Then the balance is set into stand-by
mode, „OFF“ is displayed.
To switch on press the ON/OFF button. During the display „CAL“ adjustment
is automatically carried out with the internal adjustment weight (see chap.
7.2). As soon as the weight display appears, the balance is ready for
weighing.
To switch off, press the ON/OFF button again. The balance returns into
stand-by mode.
6.5 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.
6.6 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1).
During this warming up time the balance must be connected to the power supply
(mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
22 ALT_B-PLT_A-PLT_F-BA-e-1333
7 Adjustment (all models excepted PLT-M)
Internal adjustm.
Adjustment with internal weight after pressing CAL
External adjustm.
Adjustment with external weight
Technical
adjustm.
Overwrite internal adjustment weight
Autom. adjustm.
Automatic adjustment with internal weight.
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Ensure that there are no objects on the weighing pan.
Issue of an adjustment log, see chap. 11.13
The balances have been set to „Automatic adjustment“ (see chap. 7.2) by factory. If
necessary, in the menu the following possibilities of adjustment can be activated.
ALT_B-PLT_A-PLT_F-BA-e-1333 23
Setting adjustment mode:
In weighing mode press the MENU button and keep it pressed until the
acoustic signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Adjustment mode“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Procedure see chap.
Internal adjustm.
Adjustment with internal
adjustment weight
Chap. 7.1
External adjustm.
Adjustment with external
adjustment weight
Chap. 7.3
Technical adjustm.
Overwrite internal
adjustment weight
Chap. 7.4
Autom. adjustm.
Automatic adjustment
with internal adjustment
weight
Chap. 7.2
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
24 ALT_B-PLT_A-PLT_F-BA-e-1333
7.1 Adjustment with internal weight after pressing the CAL button
Condition: Menu setting „Internal adjustm.“, see chap. 7
Condition: Menu setting „Autom.adjustm.“, see chap. 7
In weighing mode press the CAL button, adjustment will take place
automatically.
After successful adjustment the balance automatically returns to weighing
mode.
In case of an adjustment error (e.g. objects on the weighing pan) the
display will show an error message „CAL bUt“, repeat adjustment using
the CAL button.
7.2 Automatic adjustment with internal weight
(factory setting)
The automatic adjustment with internal weight (see chap. 7.1) is started
automatically, if the balance
after the weighing balance was disconnected from the mains after pressing ON/OFF in stand-by mode after a temperature change of 1.5 °C with non loaded weighing pan / zero
display
after a time interval of 20 min
with non loaded weighing pan / zero display
The automatic adjustment function is always enabled. However you can make a
manual adjustment at any time with the internal or an external weight,
see chap. 7.1 or chap. 7.3.
ALT_B-PLT_A-PLT_F-BA-e-1333 25
7.3 Adjustment with external adjustment weight
Condition: Menu setting „External adjustm.“, see chap. 7
At the models with internal adjustment weight the adjustment with
external weight is not recommended.
Information about test weights you will find in the internet under
http://www.kern-sohn.com
Ensure that there are no objects on the weighing pan. In weighing mode press
the CAL button.
Wait until the weighed value for the required adjustment weight appears
flashing.
During the flashing display put the required adjustment weight carefully in the
center of the weighing pan. The flashing display disappears.
After successful adjustment the balance automatically returns to weighing
mode.
Take away adjustment weight
26 ALT_B-PLT_A-PLT_F-BA-e-1333
7.4 Overwrite internal adjustment weight
Overwriting is restricted to specialist staff possessing well acquainted with
the workings of weighing scales.
Information about test weights you will find in the internet under
http://www.kern-sohn.com
Call up menu item „Techn. adjustment“, see chap. 7.
To confirm press the PRINT button and keep it pressed until the acoustic signal
gets mute.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
Ensure that there are no objects on the weighing pan. Press the CAL key
Wait until the weighing value of the adjustment weight appears flashing.
During the flashing display put the required adjustment weight carefully in the
center of the weighing pan. The flashing display disappears.
Wait for stability display, remove adjustment weight.
Keep the PRINT button pressed until the acoustic signal gets mute. The internal
adjustment weight is overwritten.
After successful overwriting, the balance automatically returns to weighing
mode.
ALT_B-PLT_A-PLT_F-BA-e-1333 27
8 Adjustment of models PLT-M
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Ensure that there are no objects on the weighing pan.
Issue of an adjustment log, see chap. 11.13
28 ALT_B-PLT_A-PLT_F-BA-e-1333
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Adjustment mode“.
Confirm by the PRINT button, „Autom. Just.“ will be displayed.
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
8.1 Automatic adjustment with internal weight
The automatic adjustment with internal weight is started automatically, if the balance
after the weighing balance was disconnected from the mains
after pressing ON/OFF in stand-by mode
after a temperature change of 1.5 °C
with non loaded weighing pan / zero display
after a time interval of 20 min with non loaded weighing pan / zero
display
The automatic adjustment function is always enabled. At any time, however, you can
carry out a manual adjustment (press CAL-button) using the internal adjustment
weight, see chap. 7.1.
Or selecting the adjustment mode in the menu:
ALT_B-PLT_A-PLT_F-BA-e-1333 29
Balances with obligation to verify must be taken out of operation if:
The weighing result of the balance is outside the error limit. Therefore, in
regular intervals load balance with known test weight (ca. 1/3 of the max. load)
and compare with displayed value.
The reverification deadline has been exceeded.
Position of the seal mark (models PLT)
9 Verification
General introduction:
According to EU directive 90/384/EEC balances must be officially verified if they are used as
follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the medical
and pharmaceutical laboratory.
c) For official purpose.
d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard.
Verification notes:
An EU type approval exists for balances described in their technical data as verifiable. If a
balance is used where obligation to verify exists as described above, it must be verified and
re-verified at regular intervals.
Re-verification of a balance is carried out according to the respective national regulations.
The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
30 ALT_B-PLT_A-PLT_F-BA-e-1333
In stand-by mode (see chap. 6.4) press ON/OFF
button.
As soon as the weight display appears, the balance is
ready for weighing.
Switching off:
Press ON/OFF. The balance changes into standby
mode, that means that the balance is now in state
ready-for-operation..
To switch-off the balance completely, separate
balance from power supply.
10 Basic Operation
10.1 Switch on off balance
Start-up:
10.2 Zeroing
Unload the balance
Press the TARE key. The zero display and the indicator [-0-] will appear.
ALT_B-PLT_A-PLT_F-BA-e-1333 31
10.3 Simple weighing
In order to obtain exact weighing results, your balance must have reached the
operating temperature (see warming up time chap. 1).
27-08-11 10:41:17
Current date/time
Weight: 50.5773 g
Measuring Value
The capacity display [] runs from below to above and proceeds
according to the loaded weight on the balance. Its full height is reached at
maximum load.
This is an analogue display of the current allocation of the weighing area.
Wait for zero display, reset to zero using TARE.
Place goods to be weighed on balance.
Wait until the stability display appears [].
Read weighing result.
When an optional printer is connected, the weighing value can be edited.
Printout examples (KERN YKB-01N):
10.4 Capacity display
32 ALT_B-PLT_A-PLT_F-BA-e-1333
10.5 Weighing with tare
When the balance is unloaded the saved taring value is displayed with
negative sign.
To delete the stored tare value, remove load from weighing pan and
press the TARE button. „Tare“ is displayed, wait until the zero display
appears.
The taring process can be repeated any number of times. The limit is
reached when the whole weighing range is exhausted.
10.5.1 Determination of the tare weight by weighing
The dead weight of any weighing container may be tared away by pressing a button,
so that the following weighing procedures show the net weight of the goods to be
weighed.
Put weighing container on the weighing pan.
Wait until stability display [] appears and then press the TARE-button;
„Tare“ is displayed.
After standstill control the zero display appears.
The weight of the container is now internally saved.
Weigh the material.
Wait until the stability display appears [ ] .
Read net weight.
Note:
ALT_B-PLT_A-PLT_F-BA-e-1333 33
1
10.6 Underfloor weighing
CAUTION
Always ensure that all suspended objects are stable enough to
hold the desired goods to be weighed safely (danger of breaking).
Never suspend loads that exceed the stated maximum load (max)
(danger of breaking)
Always ensure that there are no persons, animals or objects that
might be damaged underneath the load.
After completing the underfloor weighing the opening on the
bottom of the balance must always be closed (dust protection).
Objects unsuitable for placing on the weighing scale due to size or shape may be
weighed with the help of the flush-mounted platform.
Proceed as follows:
Switch off the balance.
Open closing cover (1) at the balance bottom.
Suspend hook for underfloor weighing carefully and completely.
Place weighing balance over an opening.
Attach load to hook and carry out weighing procedure.
34 ALT_B-PLT_A-PLT_F-BA-e-1333
11 Setup menu
Access to menu
2 sec.
In weighing mode press the MENU button and keep it pressed
until the acoustic signal gets mute.
The setup menu is displayed.
How to select
menu items
Use the navigation buttons to select the individual menu
items one by one. The cursor (►) displays the current setting.
Change settings
Acknowledge selected menu item using PRINT button, the
current setting is displayed. Each time the navigation buttons
are pressed the next setting will be displayed.
Save settings
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings
in the menu or go back to menu mode as follows.
Exit menu/
back to
weighing mode
Press the MENU button and keep it pressed until the acoustic
signal gets mute. The balance returns automatically into
weighing mode.
In the setup menu, the balance settings can be modified. This way, the balance can
be adjusted to individual weighing requirements.
Navigation in the menu:
ALT_B-PLT_A-PLT_F-BA-e-1333 35
Menu overview (all models excepted PLT-M)
Menu item
Selection
Description
Weight units
(see chap. 11.1)
g
Gram
ct
Carat
Oz
Ounce
Lb
Pound
Dwt
Pennyweights
Ozt
Troy ounce
GN
Grain
tl 1
Hongkong-Teal
tl 2
Singapur-Teal
tl 3
Taiwan-Teal
mo
Momme
Serial output
(see chap. 11.2)
Continuous
Continuous output
On demand
Output of stable weighing values after
pressing PRINT button
Generic printer
Not documented
Printer Tlp
Not documented
On demand - Glp
GLP conformant printouts after pressing the
PRINT button
Generic print. - Glp
Not documented
Printer Tlp - Glp
Not documented
Baud rate
(see chap. 11.3)
1200 Baud
Transfer speed
2400 Baud
4800 Baud
9600 Baud
Auto zero
Automatic zero point correction
(see chap. 11.4)
Auto zero OFF
Auto Zero switched off
Auto zero 1
Auto Zero range ± ½ Digit
Auto zero 2
Auto Zero range ± 3 Digits
Auto zero 3
Auto Zero range ± 7 Digits
Auto zero 3E
Auto Zero range ± 7 digits in the whole
weighing range
Filter
(see chap. 11.5)
Filter 1
Setting for dispensing
Filter 2
Sensitive and fast, very quiet set-up
location.
Filter 3
Robust but slow, busy set-up location.
Stability
(see chap. 11.6)
Stability 1
Standstill control fast / very quiet set-up
location
Stability 2
Standstill control fast + exact / quiet set-up
location
Stability 3
Standstill control exact /very set-up location.
36 ALT_B-PLT_A-PLT_F-BA-e-1333
Contrast adjustment
(see chap. 11.7)
1-15
Setting the contrast of the display
Backlight (see chap. 11.8)
on
Background illumination on
off
Background illumination off
Auto
Backlighting automatically switched off 3
seconds after achieving stable weighing
value. Changes in weight or pressing of keys
will automatically result in backlight switching
on again.
Timer OFF
Automatic switch-off function in
stand-by mode
(see chap. 11.9)
Disable
AUTO-OFF switched off
2 minutes
AUTO-OFF after 2 minutes without changing
the weight
5 minutes
AUTO-OFF after 5 minutes without changing
the weight
15 minutes
AUTO-OFF after 15 minutes without
changing the weight
Time and date
(see chap. 11.10
Setting date and time
Language
(see chap. 11.11)
English / Deutsch
Language operator guide
Français
Español
Português
English
Italiano
Calibration mode
(see chap. 11.12)
External calibration
Adjustment with external weight
Auto calibration
Automatic adjustment with internal
weight
Internal calibration
Adjustment with internal weight
after pressing CAL
Technical calib.
Overwriting the internal adjustment weight
Calibration info
(see chap. 11.13)
Record output of the last
adjustment
Return
Back to weighing mode
ALT_B-PLT_A-PLT_F-BA-e-1333 37
Menu item
Selection
Description
Units
(see chap. 11.1)
g
Gram
ct
Carat
RS 232
(see chap.11.2)
Continuous
Continuous output
PRINT button
Output of stable weighing values after
pressing PRINT key
Not documented
-
Not documented
-
PRINT button + GLP
GLP conformant printouts after pressing the
PRINT button
Not documented
-
Not documented
-
Baud rate
(see chap. 11.3)
1200 Baud
Transfer speed
2400 Baud
4800 Baud
9600 Baud
Auto zero
Automatic zero point correction
(see chap. 11.4)
Auto zero OFF
Auto Zero switched off
Auto zero 1
Auto Zero range ± ½ Digit
Filter
(see chap. 11.5)
Filter 1
Setting for dosage
Filter 2
Sensitive and fast, very quiet set-up
location.
Filter 3
Robust but slow, busy set-up location.
Stability
(see chapter 9.6).
Stability 1
Standstill control fast / very quiet set-up
location
Stability 2
Standstill control fast + exact / quiet set-up
location
Stability 3
Standstill control exact /very busy set-up
location.
Menu overview models PLT-M:
38 ALT_B-PLT_A-PLT_F-BA-e-1333
Display contrast
(see chap.11.7)
1-15
Select contrast
Background illumination of the
display (see chap. 11.8)
on
Background illumination on
off
Background illumination off
Auto
Background illumination automatically
switched off 3 seconds after achieving stable
weighing value. Changes in weight or
pressing of keys will automatically result in
background illumination switching on again.
AUTO-OFF
Automatic switch-off function in
stand-by mode
(see chap. 11.9)
Disabled
AUTO-OFF switched off
2 min
AUTO-OFF after 2 minutes without changing
the weight
5 min
AUTO-OFF after 5 minutes without changing
the weight
15 min
AUTO-OFF after 15 minutes without
changing the weight
Time and date
(see chap 11.10)
Setting date and time
Language
(see chap. 11.11)
Deutsch
Language operator guide
Français
Español
Português
English
Italiano
Adjustment mode
(see chap. 11.12)
Autom. adjustm.
Automatic adjustment with internal
weight
Adjustment log (see chap.
11.13)
Back to weighing mode
Back
ALT_B-PLT_A-PLT_F-BA-e-1333 39
11.1 Weighing Units
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Weight units“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
! In models PLT-M only gram and carat are available.
Symbol
Weighing unit
Conversion
factor 1g =
g
Gram
1.
ct
Carat
5.
Oz
Ounce
0.035273962
Lb
Pound
0.0022046226
Dwt
Pennyweights
0.643014931
Ozt
Troy ounce
0.032150747
GN
Grain
15.43235835
tl 1
Hongkong-Teal
0.02671725
tl 2
Singapur-Teal
0.02646063
tl 3
Taiwan-Teal
0.02666666
mo
Momme
0.2667
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
According to the requirements the balance can be switched over into different units.
Selected weighing unit will be retained even after disconnection from the mains.
40 ALT_B-PLT_A-PLT_F-BA-e-1333
11.2 RS232
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Serial output“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Continuous
Continuous data output
On demand
Output of stable weighing values after
pressing PRINT key
Generic printer
Not documented
Printer Tlp
Not documented
On demand - Glp
GLP conformant printouts after pressing
the PRINT button
Generic print. - Glp
Not documented
Printer Tlp - Glp
Not documented
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
ALT_B-PLT_A-PLT_F-BA-e-1333 41
11.3 Baud rate
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Baud rate“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Selectable 1200, 2400, 4800, 9600 baud.
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
42 ALT_B-PLT_A-PLT_F-BA-e-1333
11.4 Auto Zero
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Auto Zero“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
! In models PLT-Monly „Auto ZeroOff“ and „Auto Zero 1“ are available.
Display
Description
Auto Zero off
Auto Zero switched off
Auto Zero 1
Auto Zero range ± ½ Digit
Auto Zero 2
Auto Zero range ± 3 Digits
Auto Zero 3
Auto Zero range ± 7 Digits
Auto Zero 3E
Auto Zero range ± 7 digits in the whole weighing range
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
Under this menu item the automatic zero point correction can be switched on or off.
In switched-on-state the zero point is automatically corrected at drift or when dirty.
Note:
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the “stability
compensation“. (e.g. slow flow of liquids from a container placed on the balance,
evaporating processes).
When apportioning involves small variations of weight, it is advisable to switch off
this function.
ALT_B-PLT_A-PLT_F-BA-e-1333 43
11.5 Filter
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Filter“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Filter 1
Setting for dispensing
Filter 2
The balance reacts quickly and in a sensitive manner,
quiet set-up location.
Filter 3
The balance reacts slowly and in a robust manner,
busy set-up location
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
This menu item allows the balance to be set according to specific ambient conditions
and measuring purposes.
44 ALT_B-PLT_A-PLT_F-BA-e-1333
11.6 Stability
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Stability“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Stability 1
Standstill control fast - very quiet set-up location
Stability 2
Standstill control fast + exact – quiet set-up location
Stability 3
Standstill control exact / set-up location.
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
ALT_B-PLT_A-PLT_F-BA-e-1333 45
11.7 Setting the contrast of the display
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Contrast adjustment“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
0
small contrast
15
high contrast
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
To set the contrast of the display you can choose between 15 values.
46 ALT_B-PLT_A-PLT_F-BA-e-1333
11.8 Display background illumination
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Backlight“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Auto
Backlighting automatically switched off 3 seconds after
achieving stable weighing value. Changes in weight or
pressing of keys will automatically result in backlight
switching on again
On
Background illumination on
off
Background illumination on
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
ALT_B-PLT_A-PLT_F-BA-e-1333 47
11.9 Automatic switch-off function „AUTO OFF“
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Timer off“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Disabled
AUTO-OFF switched off
2 minutes
AUTO-OFF after 2 minutes without changing the
weight
5 minutes
AUTO-OFF after 5 minutes without changing the
weight
15 minutes
AUTO-OFF after 15 minutes without changing the
weight
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
48 ALT_B-PLT_A-PLT_F-BA-e-1333
11.10 Setting date and time
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Time and date“.
Acknowledge using PRINT button, the current setting is displayed.
Use navigation buttons to select the required format.
DD-MM: day / month
MM-DD: month / day
Acknowledge the format with the PRINT button and set date and time as follows.
The active digit e.g. 25 is shown underlined.
Use the navigation buttons to set the day and confirm using the PRINT button.
The active digit wanders to the month, (is shown underlined). Use the navigation
buttons to set the month and confirm using the PRINT button.
Set year, hours, minutes and seconds in the same way.
After finishing the input, press the PRINT button and keep it pressed until the
acoustic signal gets mute.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
The set date and time are displayed.
ALT_B-PLT_A-PLT_F-BA-e-1333 49
11.11 Language operator guide
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Language“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Language
Deutsch
Français
Español
Português
English
Italiano
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
50 ALT_B-PLT_A-PLT_F-BA-e-1333
11.12 Adjustment mode
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Calibration mode“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
! In models PLT-M only automatic adjustment is available.
Display
Description
Procedure see chap.
Internal
adjustm.
Adjustment with internal
weight after pressing the
CAL button
+Chap. 7.1
External
adjustm.
Adjustment with external
adjustment weight
+Chap. 7.3
Technical
adjustm.
Overwrite internal
adjustment weight
+Chap. 7.4
Autom. adjustm.
Automatic adjustment with
internal adjustment weight
+ Chap. 7.2
+ or 8.1 (PLT-M)
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
ALT_B-PLT_A-PLT_F-BA-e-1333 51
11.13 Adjustment log
Make sure that the communication parameter of balance and printer are the
same.
Communication parameters, see chap. 9.2 and 11.3
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Calibration info“.
Confirm with PRINT button; date, time, adjustment type and deviation of the last
adjustment are displayed.
If you connect an optional printer you can also print these data by pressing the
PRINT key.
Printout example (KERN YKB-01N):
27-08-11 10:41:17
Current date/time
Calibration:
27-08-11 10:11:17
Date/time of the last
adjustment
Internal calib.
Type of adjustment
Corr. : 0.21 g
Deviation from last adjustment
Press the ON/OFF key. Weighing balance returns to menu. Either make more
adjustments in the menu if required or press ON/OFF button anew; the balance
will return into weighing mode.
This function helps to print out a record of the last adjustment.
52 ALT_B-PLT_A-PLT_F-BA-e-1333
12 Main Menu
Navigation in the menu:
Access to menu
In weighing mode press MENU button. The master menu will be
displayed.
How to select
menu items
Use the navigation buttons to select the individual menu
items one by one. The cursor (►) displays the current setting.
Change settings
Acknowledge selected menu item using PRINT button, the
current setting is displayed. Each time the navigation buttons
are pressed the next setting will be displayed.
Save settings
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings
in the menu or go back to menu mode as follows.
Exit menu/
back to
weighing mode
Press the ON/OFF key.
ALT_B-PLT_A-PLT_F-BA-e-1333 53
12.1 Parts counting
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Piece counting“.
Acknowledge using PRINT button, the currently set reference quantity is
displayed.
Use the navigation buttons to select the desired setting.
Display
Description
10 off
Reference piece number 10
25 off
Reference piece number 25
50 off
Reference piece number 50
100 off
Reference piece number 100
Manual
Numerical input of reference weight, see chap. 12.1.2
Before the balance can count parts, it must know the average part weight (i.e.
reference). Proceed by putting on a certain number of the parts to be counted. The
balance determines the total weight and divides it by the number of parts, the socalled reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
As a rule:
The higher the reference quantity the higher the counting exactness.
12.1.1 Determine reference by weighing
54 ALT_B-PLT_A-PLT_F-BA-e-1333
If necessary place a weighing container, confirm the set reference piece number
by pressing the PRINT button.
Place as many pieces to add-up as required by the set reference piece number.
Acknowledge using PRINT button and wait for stability display.
As soon as the weighing result is stable, the determined average piece weight will
be taken over as reference weight.
The currently placed weight „W“, the reference weight „MPW“ and the piece number
„PCS“ are displayed.
Remove reference weight. The balance is now in parts counting mode counting
all units on the weighing pan.
ALT_B-PLT_A-PLT_F-BA-e-1333 55
If it was impossible to determine a reference due to instable goods to be
weighed or an insufficient reference weight, the following display will
appear during reference calculation.
Smallest counting weight not achieved
Increase the counted weight or select a balance with high resolution
or
If in the display “Add pieces…..“ appears, the balance asks for a
reference optimisation, as the placed number of pieces was not sufficient
for a correct reference calculation.
Place more parts, but at least the double quantity.
Press the PRINT button, the reference is calculated anew.
The placed quantity is always too
small, add more parts (double
quantity) and acknowledge using
the PRINT button. Repeat the
procedure until the piece counter
display appears.
Placed quantity sufficient for
reference determination.
Remove reference weight. The
balance is now in parts counting
mode counting all units on the
weighing pan.
56 ALT_B-PLT_A-PLT_F-BA-e-1333
Carry out parts counting
After having calculated the reference, place the pieces which will be counted.
The currently placed weight „G“, the reference weight „AUW“ and the calculated
piece number „St.“ are displayed.
The display value can be printed out by connecting an optional printer.
Printout example (KERN YKB-01N):
27-08-11 10:41:17
Current date/time
PCS
20
Calculated piece number
Weight:
199,9985 g
Weight placed on balance
MPW:
10 000 g
Reference weight
Return to weighing mode
Press the ON/OFF key
ALT_B-PLT_A-PLT_F-BA-e-1333 57
12.1.2 Numeric entry of the reference weight
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Piece counting“.
Acknowledge using PRINT button, the currently set reference quantity is
displayed.
Use the MENUbutton to select the „Manual“ setting.
Confirm by pressing the PRINT button.
Use the navigation buttons (see chap. 2.2.1) to enter the known reference piece
weight in grams. To set a decimal point press the CAL button long time. The
input is deleted by pressing the TARE button a longer time.
Confirm by pressing the PRINT button.
From here on the balance is in parts counting mode and counts all pieces which are
placed on the weighing pan (see chap.10.1.1 „Carry our parts counting“).
If the minimum parts weight is not reached, an error message will appear. Use the
ON/OFF button to return to the weighing mode and restart the procedure.
Return to weighing mode
Press the ON/OFF key
If the piece weight (reference) is known, it can be entered directly. As the balance in
this method needs not to calculate any reference, the balance after confirming the
reference piece weight will be directly in parts counting mode.
58 ALT_B-PLT_A-PLT_F-BA-e-1333
12.1.3 Automatic reference optimization
The automatic reference optimisation is not active when the reference
weight is entered numerically.
In order to improve the counting exactness, the reference can be optimised by
adding more pieces. At every reference optimisation, the reference weight is
calculated anew. As the additional pieces increase the base for the calculation, the
reference also becomes more exact.
Leave the piece number on the weighing pan after setting the reference weight
Duplicate the number of pieces on the weighing pan and wait until the acoustic
signal sounds. The reference weight is calculated anew.
Either repeat the reference optimisation by adding more pieces (max. 255
pieces) or start the counting process.
ALT_B-PLT_A-PLT_F-BA-e-1333 59
12.2 Density determination (Hydrostatic weighing)
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Density“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select setting „Solid body“.
Confirm using the PRINT button, the set density of the measuring liquid is
displayed (factory setting 1.0000 for dest. water at 20°C).
Use the navigation buttons (see chap. 2.2.1) to enter the current density of the
measuring liquid. For water see the following density table.
Confirm with PRINT button, the display for calculation of „Weight in air“ is
displayed.
Density is the relationship of weight [g] : volume [cm3]. The weight is determined by
weighing the sample in air. The volume results from the ascending force [g] of the
sample dipped in a liquid. The density [g/cm3] of that liquid is known (principle of
Archimedes).
The density is determined with help of the underfloor weighing device.
We recommend to switch off the automatic adjustment in the setup menu,
see chap. 11.12.
12.2.1 Density determination of solids
Prepare the balance as follows:
Switch off the balance
Remove the weighing pan and carefully turn over the weighing balance.
Suspend hook for underfloor weighing (optional)
Place weighing balance over an opening
Prepare suspension device
Fill measuring liquid e.g. water in a vessel e.g. beaker and temper it.
Calculate density:
60 ALT_B-PLT_A-PLT_F-BA-e-1333
Suspend sample with suspension device at the flush-mounted hook. Wait for
stability display, take over the weight value using PRINT button. The display for
determination of the "sample in water“ appears.
Immerse the sample without bubbles developing.
Make sure that the sample does not touch the beaker.
Wait for stability display, take over the weight value using PRINT button.
The density of the sample is shown.
ALT_B-PLT_A-PLT_F-BA-e-1333 61
When an optional printer is connected, the displayed value can be edited using
PRINT.
Printout example (KERN YKB-01N):
07-09-11 11:14:57
d: 8.0700 g/cm3
If at the density determination errors have appeared, „d-----„ is displayed.
1. Back to density determination mode, press MENU button.
2. Back to weighing mode, press ON/OFF button.
62 ALT_B-PLT_A-PLT_F-BA-e-1333
Density table of water:
* the asterisk means that the cipher left from the comma is reduced by 1.
Table from "Mass determination" by M. Kochsiek, M. Gläser
ALT_B-PLT_A-PLT_F-BA-e-1333 63
12.2.2 Determining density of liquids
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Density“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the setting „Liquid“.
Confirm using the PRINT button, the set density of the sinker is displayed
(factory setting 3.0000 g/cm3).
Use the navigation buttons (see chap. 2.2.1) to enter the current density of the
sinker.
Confirm by pressing the PRINT button. The display for weight determination of
the "Sinker in air“ appears.
For density determination of liquids, a sinker with known density is used (optionally
available). The sinker is weighed first in air and then in the liquid whose density is to
be determined. From the weight difference results the buoyancy from where the
software calculates the density.
Preparation:
Fill sample liquid into a vessel, e.g. a pitcher.
Heat sample liquid until temperature is constant.
Prepare a sinker with known density
Calculate density:
64 ALT_B-PLT_A-PLT_F-BA-e-1333
Suspend sinker with suspension device at the flush-mounted hook. Wait for
stability display, take over the weight value using PRINT button. The display for
weight determination of the "Sinker in sample liquid“ appears.
Immerse the sample without bubbles developing.
Make sure that the sample does not touch the beaker.
Wait for stability display, take over the weight value using PRINT button. The
density of the sample is shown.
ALT_B-PLT_A-PLT_F-BA-e-1333 65
If at the density determination errors have appeared, „d-----„ is displayed.
When an optional printer is connected, the displayed value can be edited using
PRINT.
Printout example (KERN YKB-01N):
07-09-11 11:14:57
d : 0.9984 g/cm
3
1. Back to density determination mode, press MENU button.
2. Back to weighing mode, press ON/OFF button.
66 ALT_B-PLT_A-PLT_F-BA-e-1333
12.3 Formula mode
Select formula mode „Manual“
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Formulation“.
Confirm by pressing the PRINT button.
Confirm by pressing the PRINT button. The display for weighing the first
component appears.
Weighing components
Tare when using a weighing container Weighing the first component (e.g. 15 g).
Confirm by pressing the PRINT button. The weight is added to the summation
memory and printed if an optional printer is connected.
The formula function allows weighing of components which are in a certain relation
one to the other. For check purpose, the weight of all components, as well as the
total weight (TOT) can be printed out.
The balance works with a separated memory for the weight of the weighing container
and of the formula components.
12.3.1 Formulation without using the recipe data base
ALT_B-PLT_A-PLT_F-BA-e-1333 67
Weighing the second component (e.g. 5 g).
Confirm by pressing the PRINT button. The weight is added to the summation
memory and printed if an optional printer is connected.
If required, weigh more components (max. 99) as described above.
Complete Formulation process Press the PRINT button and keep it pressed until the acoustic signal gets mute.
The total weight (TOT:) of all components will be displayed and edited on the
printer.
Printout example (KERN YKB-01N):
07-09-11 11:14:57
Date/time
Manual
Recipe mode
1. 15.00 g
Weighing 1. component
2. 5.00 g
Weighing 2. component
T = 20.00 g
Total weight
Back to formula mode, press the ON/OFF button and start the next Formulation.
Back to weighing mode, press ON/OFF button repeatedly.
68 ALT_B-PLT_A-PLT_F-BA-e-1333
12.3.2 Formulation using the recipe data base
Select formula mode
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Formula“.
Confirm by pressing the PRINT button.
Define recipe and components
1. Use the navigation buttons to select the memory space number (e.g. 01) for the
recipe. Press the PRINT button and keep it pressed until the acoustic signal gets
mute. The display for entering the recipe name appears.
2. Use the navigation buttons (see chap. 2.2.1 ) to enter the recipe name (max. 20
characters).
12.3.2.1 Store recipes
The balance contains a recipe data base, where maximally 99 recipes each with 20
components can be filed.
ALT_B-PLT_A-PLT_F-BA-e-1333 69
3. Confirm by pressing the PRINT button. The display for entering the first
component name appears.
4. Use the navigation buttons (see chap. 2.2.1) to enter the component name
(max. 11 characters).
5. Confirm by pressing the PRINT button. The display to enter the quantity appears.
6. Use the navigation buttons (see chap. 2.2.1) to enter quantity.
70 ALT_B-PLT_A-PLT_F-BA-e-1333
7. Confirm by pressing the PRINT button. The input of the minus-tolerance is
displayed.
8. Enter the negative tolerance value: ex.: -10 %
9. Confirm by pressing the PRINT button. The input of the plus-tolerance is
displayed.
10. Enter the positive tolerance value: ex.: 5 %
11. Confirm by pressing the PRINT button
12. To enter more components (max. 20) repeat step 3 – 11 respectively.
13. Once having entered all components, use the ON/OFF button to exit the formula
entry mode.
Back to weighing mode, press ON/OFF button anew.
ALT_B-PLT_A-PLT_F-BA-e-1333 71
12.3.2.2 Invoke and work off the recipes
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Formula“.
Confirm by pressing the PRINT button.
or (Example)
„Receipt 00“, or the most recently entered recipe will appear
Use navigation buttons to select the required recipe.
Acknowledge using PRINT button, the first component and its nominal value as
well as the negative and the positive tolerance value are displayed. Using the
navigation buttons , all components including their nominal value and tolerance
values can be displayed.
Select component and confirm by the PRINT button, the entry of the multiplication
factor will be displayed
After invoking a stored recipe (see chap. 12.3.2.1) the balance will be ready at once
for weighing the components. Name and nominal value, tolerance and multiplication
factor of the respective component are displayed.
72 ALT_B-PLT_A-PLT_F-BA-e-1333
1
Tolerance value
2
Total weight of all components
3
Nominal value of the component
4
Total weight of all components
5
Displaying tolerance
Weighed good below nominal value
Weighed good = nominal value
Weighed good above nominal value
Weighed good below nominal value
Weighed good = nominal value
Weighed good above nominal value
Use the navigation buttons to enter the required multiplication factor.
Use the PRINT button to confirm the selected factor:
Example for factor 1
Overview tolerance display:
Ex.: tolerance -10.0 % to +5.0 %
Ex.: without tolerance entry:
ALT_B-PLT_A-PLT_F-BA-e-1333 73
The display for weighing the first component appears.
Tare when using a weighing container. Start weighing. When the nominal value is
reached, right next to the capacity display „OK“ will appear.
If the nominal value is not reached (-) or exceeded (+), „Err 10“ will appear on the
display after pressing the PRINT button. Correct the poured quantity.
If the nominal weight is reached, press PRINT button. „Wait“ appears shortly
followed by „Tare“. Then the display passes to „W=0“ and the display for weighing
the second component appears.
Weigh-in the displayed nominal value of the second component.
74 ALT_B-PLT_A-PLT_F-BA-e-1333
Printout example (KERN YKB-01N):
07-09-11 11:14:57
Cake
Name receipt
1. 10.00 g
Weighing 1. component
Salt
Name 1. component
2. 70.00 g
Weighing 2. component
Sugar
Name 2. component
3. 0.50 g
Weighing 3. component
ABC
Name 3. component
T = 80.50 g
Total weight
Back to formula mode, press the ON/OFF button and start the next recipe.
Back to weighing mode, press ON/OFF button anew.
After weighing and confirming the last component, the total weight (TOT:) of all
components will be displayed and printed automatically.
ALT_B-PLT_A-PLT_F-BA-e-1333 75
Example for factor 2:
Call-up the required recipe as specified above
Acknowledge using PRINT button, the first component and its nominal value as
well as the negative and the positive tolerance value are displayed. Using the
navigation buttons , all components including their nominal value and tolerance
values can be displayed.
Select component and confirm by the PRINT button, the entry of the
multiplication factor will be displayed
Use the navigation buttons to enter the multiplication factor „2“.
Use the PRINT-button to confirm the selected factor:
Now the double quantity (20.000 g) appears as nominal value.
Start weighing. When the nominal value is reached, right next to the capacity
display „OK“ will appear.
76 ALT_B-PLT_A-PLT_F-BA-e-1333
If the nominal weight is reached, press PRINT button. „Wait“ appears shortly
followed by „Tare“. Then the display passes to „W=0“ and the display for weighing
the second component appears.
Now the nominal value is „40000 g“. Start weighing. When the nominal value
is reached, right next to the capacity display „OK“ will appear.
After weighing and confirming the last component, the total weight (TOT:) of all
components will be displayed and printed automatically.
ALT_B-PLT_A-PLT_F-BA-e-1333 77
Back to formula mode, press the ON/OFF button and start the next recipe.
Back to weighing mode, press ON/OFF button anew.
78 ALT_B-PLT_A-PLT_F-BA-e-1333
12.4 Check weighing
Tolerance
marker
Tone
Assessment
no
Load exceeds specified tolerance
yes
Load within specified tolerance
no
Load below specified tolerance
Tolerance
marker
Tone
Assessment
yes
Weighed goods > lower limit value
no
Weighed goods lower limit value
Tolerance
marker
Tone
Assessment
no
Weighed goods > upper limit value
yes
Weighed goods upper limit value
In checkweighing an upper and a lower limit value can be established to ensure that
the weighed goods are exactly within the established tolerance limits.
The tolerance mark () and an acoustic signal (selectable) show, if the weighed
material is within the two tolerance limits.
Display of the results
1. Lower and upper limit value 0
2. Lower limit value > 0 and upper limit value = 0
3. Lower limit value = 0 and upper limit value > 0
ALT_B-PLT_A-PLT_F-BA-e-1333 79
Settings
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Checkweighing“.
Press PRINT button to confirm, the display for entering the lower limit value will
appear.
Use the navigation buttons (see chap.2.2.1) to enter the lower limit value, e.g.
9.00 g.
Press PRINT button to confirm, the display for entering the upper limit value will
appear.
Use the navigation buttons (see chap.2.2.1) to enter the upper limit value, e.g.
11.00 g.
Confirm using the PRINT button, the display to adjust the signal tone appears.
Use the MENU-button to select the desired setting
On: Acoustic signal switched on
Off: Signal tone switched off
Confirm by pressing the PRINT button.
From now the balance is in checkweighing mode.
80 ALT_B-PLT_A-PLT_F-BA-e-1333
Start tolerance check
Tare when using a weighing container
Put on goods to be weighed, tolerance control is started
Load below specified
tolerance
Load within specified
tolerance
Load exceeds specified
tolerance
When an optional printer is connected, the displayed value can be edited using
PRINT.
Printout examples (KERN YKB-01N):
Load below specified
tolerance
Load within specified
tolerance
Load exceeds specified
tolerance
07-09-11 11:14:57
07-09-11 11:14:57
07-09-11 11:14:57
Lim.1 :
9 000 g
Lim.1 :
9 000 g
Lim.1 :
9 000 g
Lim. 2 :
11 000 g
Lim. 2 :
11 000 g
Lim. 2 :
11 000 g
Weight :
8,900 g
Weight :
10 000 g
Weight :
11,700 g
TEST : KO! ---
TEST : OK
TEST : KO! +++
Return to weighing mode
Press the ON/OFF key
ALT_B-PLT_A-PLT_F-BA-e-1333 81
12.5 Percent determination
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Percent weigh.“.
Confirm with the PRINT button, the currently set mode will be displayed.
Use the navigation buttons to select setting „Automatic“.
If using a weighing container, put this before pressing the PRINT button. The
taring process is automatically started.
Press the PRINT key.
Place the reference weight (=100 %) and press the PRINT button. As soon the
weighing value is stable, the determined weight will be taken over as reference
(100%). From now the balance is in percent determining mode.
Place weighed goods, the result display will appear.
W%
Weight of the sample in percentage
W Weight of the sample in grams
REF:
Reference weight (100%)
When an optional printer is connected, the displayed value can be edited
using PRINT. Printout example (KERN YKB-01N):
07-09-11 11:14:57
Perc.
49.95 %
Weight of the sample in
percentage
Weight:
9.990 g
Weight of the sample in grams
Refer.:
20.000 g
Reference weight (100%)
Percent determination allows weight display in percent, in relation to a reference
weight.
12.5.1 Entering the reference weight by weighing
82 ALT_B-PLT_A-PLT_F-BA-e-1333
12.5.2 Numeric entering of the reference weight
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Percent determin.“.
Confirm with the PRINT button, the currently set drying mode will be displayed.
Use the navigation buttons to select setting „Manual“.
Confirm by pressing the PRINT button.
Use the navigation buttons (see chap. 2.2.1) to enter the reference weight
(100 %) in grams.
Confirm input by pressing the PRINT button.
From now the balance is in percent determining mode.
Place weighed goods, the result display will appear.
W%
Weight of the sample in percentage
W Weight of the sample in grams
REF:
Reference weight (100%)
When an optional printer is connected, the displayed value can be edited using
PRINT. Printout example (KERN YKB-01N):
07-09-11 11:14:57
Perc.
49.95 %
Weight of the sample in
percentage
Weight:
9.990 g
Weight of the sample in grams
Refer.:
20.000 g
Reference weight (100%)
Return to weighing mode
Press the ON/OFF key
ALT_B-PLT_A-PLT_F-BA-e-1333 83
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „Animal weigh.“.
Confirm with the PRINT button, the currently set measuring time will be displayed.
Use the navigation buttons to select the wanted measuring time (5 – 90 s) and
confirm with PRINT button.
Put upon weighing goods and press the PRINT button. In the display runs a
„Countdown“ of the set measuring time.
The average value of the weighing results is displayed and remains displayed on
the screen.
When an optional printer is connected, the displayed value can be edited using
PRINT. Printout example (KERN YKB-01N):
07-09-11 11:14:57
Time
= 20 sec
Measuring time
A:
20.0052 g
Weighing result
For other weighing procedures press ON/OFF button 1x
Return to weighing mode:
Press the ON/OFF button 2x
12.6 Animal weighing function
The animal weighing function can be applied for busy weighings. During a defined
period the mean value of the weighing results is formed.
The more unquiet the weighed item, the longer the measuring time should be
selected.
84 ALT_B-PLT_A-PLT_F-BA-e-1333
12.7 Peak value function
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select menu item „Maximum load“.
Confirm using PRINT button, the taring process is automatically started. „Tare“
will be shortly displayed. From now on the balance is in peak value mode.
Load weighing pan. The maximum load value is displayed.
The peak value remains in the display until the TARE button is pressed. Then the
balance is ready for further measurements.
When an optional printer is connected, the displayed value can be edited using
PRINT. Printout example (KERN YKB-01N):
07-09-11 11:14:57
Max.:
20.0356 g
Peak value
Return to weighing mode:
Press the ON/OFF key.
This function displays the highest load value (peak value) of a weighing.
The peak value remains in the display until it will be deleted.
ALT_B-PLT_A-PLT_F-BA-e-1333 85
12.8 GLP function (Good laboratory practice)
In weighing mode press MENU button. The master menu will be displayed.
Use the navigation buttons to select the menu item „GLP.“.
Confirm by pressing the PRINT button. The display to enter the balance
identification number appears.
Use the navigation buttons (see chap. 2.2.1) to enter the balance ID (max. 18
characters).
Confirm by pressing the PRINT button. The display for entering the user ID
appears.
Use the navigation buttons (see chap. 2.2.1) to enter the user ID (max. 18
characters).
Confirm by pressing the PRINT button. The display for entering the project ID
appears.
Use the navigation buttons (see chap. 2.2.1) to enter the project ID (max. 18
characters).
To save all the information press the PRINT button and keep it pressed until the
acoustic signal gets mute. The balance returns automatically into weighing mode.
In the „GLP“ settings, information is defined, which will be edited in the measurement
records.
86 ALT_B-PLT_A-PLT_F-BA-e-1333
For GLP conformant printouts activate the menu setting „PRINT
button_GLP“, see chap. 11.2
Printout example (KERN YKB-01N):
07-09-11 11:14:57
G.L.P. Parameter
Balance ID:
TEST 1
User ID:
Miller
Project ID:
789
Weight: 199.991 g
Weighing data
Signature:
G.L.P. Parameter
ALT_B-PLT_A-PLT_F-BA-e-1333 87
RS232C
13 Data output RS 232C
Pin 1
Power +5V
Pin 2
Tx Signal
Pin 3
Rx Signal
Pin 4
busy signal
Pin 5
GND
Pin 4-6
Transmission to PC
The following conditions must be met to provide successful communication between
the weighing balance and the printer.
Use a suitable cable to connect the weighing balance to the interface of the
printer/PC.
Faultless operation requires an adequate KERN interface cable (optional).
The baud rates of balance and printer must match (see chap.11.3).
In the setup menu (see chap.11.2) activate the desired data output, you can
select
- Continuous data output
- Output of stable weighing values after pressing the PRINT button
- GLP conformant printouts after pressing the PRINT button
13.1 Technical data
8-bit ASCII Code
8 data bits, 1 stop bit, no parity bit
Baud rate optional from 1200 - 9600 Baud
13.2 Pin allocation of balance output plug
88 ALT_B-PLT_A-PLT_F-BA-e-1333
13.3 Interface
Balance
PC
Balance
PC
Input data
Busy signal
GND
2
4
5
Balance
Printer
RS232C
RS2
RS232CRS232C
RS232C
RS2
RS232C
2
3
5
6
4
3
2
7
6
20
2
3
5
6
4
2
3
5
6
4
Balance – PC 25-pole
Balance – PC 9-pole
Balance - printer
ALT_B-PLT_A-PLT_F-BA-e-1333 89
13.4 Format Data transfer
1. Character
Sign -/ blank (weighing value)
2-9. Character
Weight or other data
10-12. Character
Weighing unit
13. Character
Stability display
14. Character
Carriage return
15. Character
Line feed
1°
2°
3°
4°
5°
6°
7°
8°
9°
10°
11°
12°
13°
14°
15°r
Character
Weight
Weighing unit
Stability
CR
LF
The data set consists of the following 14 characters:
Weighing mode (continuous output or output after pressing the PRINT button)
90 ALT_B-PLT_A-PLT_F-BA-e-1333
14 Error messages
ERR01
Weight value instable or zeroing not possible.
Check the environmental conditions.
ERR02
Adjustment error, e.g. instable environmental conditions
ERR03
Adjustment error e.g. incorrect adjustment weight}
ERR04
Piece weight too small/instable
ERR05
Data transfer not possible, as weighing value is instable. Check
the environmental conditions.
ERR06
Weighing value in density determining mode instable.
Check the environmental conditions.
ERR07
Weighing value in percentage determining mode instable.
Check the environmental conditions.
ERR08
Error at internal adjustment
Weighing value in formula mode instable.
Check the environmental conditions.
“UNLOAD”:
Weighing range not achieved.
Check position of the weighing pan.
“CAL But”
Adjust balance.
Weighing range exceeded, placed load exceeds the capacity of
the balance. Unload the balance.
Weighing range not reached, e.g. weighing pan not in place
ALT_B-PLT_A-PLT_F-BA-e-1333 91
15 Service, maintenance, disposal
Before any maintenance, cleaning and repair work disconnect the
appliance from the operating voltage.
15.1 Clean
Please do not use aggressive cleaning agents (solvents or similar agents), but a
cloth dampened with mild soap suds. Ensure that no liquid penetrates into the
device. Polish with a dry soft cloth.
Loose residue sample/powder can be removed carefully with a brush or manual
vacuum cleaner.
Spilled weighing goods must be removed immediately.
15.2 Service, maintenance
The appliance may only be opened by trained service technicians who are
authorized by KERN.
Ensure that the balance is regularly calibrated, see chap. Testing instruments
control.
15.3 Disposal
Disposal of packaging and appliance must be carried out by operator according
to valid national or regional law of the location where the appliance is used.
92 ALT_B-PLT_A-PLT_F-BA-e-1333
15.4 Instant help
Fault
Possible cause
The displayed weight does
not glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
The displayed weight is
permanently changing
Draught/air movement
Glass doors not closed
Table/floor vibrations
Weighing pan has contact with other objects.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is
obviously incorrect
The display of the balance is not at zero
Adjustment is no longer correct.
The balance is on an uneven surface.
Great fluctuations in temperature.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
EC-Konformitätserklärung EC-Declaration of -Conformity
EC- Déclaration de conformité EC-Declaración de Conformidad
EC-Dichiarazione di conformità EC-Conformiteitverklaring
EC- Declaração de conformidade EC- Prohlášení o shode
EC-Deklaracja zgodnościEC-Заявлениеосоответствии
Electronic Balance: KERN ALT_B, PLT_A, PLT_F
94 ALT_B-PLT_A-PLT_F-BA-e-1333