KERN & SOHN ALT 250-4B User Manual

KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com
Operating instructions Analytical and precision balances
KERN ALT_B-PLT_A-PLT_F
Version 3.3
08/2013 GB
ALT_B-PLT_A-PLT-F-BA-e-1333
GB
KERN ALT_B-PLT_A-PLT_F
Version 3.3 08/2013
Operating instructions Electronic analytical and precision balances
Contents
1 Technical data ................................................................................................ 4
2 Appliance overview ....................................................................................... 9
2.1 Operating side .......................................................................................................................... 9
2.1.1 Back and bottom side ............................................................................................................... 11
2.2 Keyboard overview ................................................................................................................. 12
2.2.1 Navigation buttons / numeric input ........................................................................................... 13
2.3 Overview of display ................................................................................................................ 14
3 Basic instructions ........................................................................................ 16
3.1 Proper use ............................................................................................................................... 16
3.2 Improper Use........................................................................................................................... 16
3.3 Warranty .................................................................................................................................. 16
3.4 Monitoring of Test Resources ............................................................................................... 16
4 Basic Safety Precautions ............................................................................ 17
4.1 Pay attention to the instructions in the Operation Manual ................................................ 17
4.2 Personnel training .................................................................................................................. 17
5 Transport and storage ................................................................................. 17
5.1 Testing upon acceptance ...................................................................................................... 17
5.2 Packaging / return transport ................................................................................................. 17
6 Unpacking, Setup and Commissioning ..................................................... 19
6.1 Installation Site, Location of Use .......................................................................................... 19
6.2 Unpacking, Scope of delivery ............................................................................................... 20
6.3 Placing ..................................................................................................................................... 20
6.4 Connect and switch-on power supply .................................................................................. 22
6.5 Connection of peripheral devices ......................................................................................... 22
6.6 Initial Commissioning ............................................................................................................ 22
7 Adjustment (all models excepted PLT-M) .................................................. 23
7.1 Adjustment with internal weight after pressing the CAL button ....................................... 25
7.2 Automatic adjustment with internal weight (factory setting) ............................................. 25
7.3 Adjustment with external adjustment weight ...................................................................... 26
7.4 Overwrite internal adjustment weight .................................................................................. 27
8 Adjustment of models PLT-M ..................................................................... 28
8.1 Automatic adjustment with internal weight ......................................................................... 29
9 Verification ................................................................................................... 30
10 Basic Operation ........................................................................................... 31
10.1 Switch on off balance ............................................................................................................. 31
10.2 Zeroing..................................................................................................................................... 31
10.3 Simple weighing ..................................................................................................................... 32
10.4 Capacity display ..................................................................................................................... 32
10.5 Weighing with tare .................................................................................................................. 33
10.5.1 Determination of the tare weight by weighing....................................................................... 33
10.6 Underfloor weighing ............................................................................................................... 34
2 ALT_B-PLT_A-PLT_F-BA-e-1333
11 Setup menu .................................................................................................. 35
11.1 Weighing Units ....................................................................................................................... 40
11.2 RS232 ....................................................................................................................................... 41
11.3 Baud rate ................................................................................................................................. 42
11.4 Auto Zero ................................................................................................................................. 43
11.5 Filter ......................................................................................................................................... 44
11.6 Stability .................................................................................................................................... 45
11.7 Setting the contrast of the display ........................................................................................ 46
11.8 Display background illumination .......................................................................................... 47
11.9 Automatic switch-off function „AUTO OFF“ ........................................................................ 48
11.10 Setting date and time ............................................................................................................. 49
11.11 Language operator guide ...................................................................................................... 50
11.12 Adjustment mode ................................................................................................................... 51
11.13 Adjustment log ........................................................................................................................ 52
12 Main Menu .................................................................................................... 53
12.1 Parts counting ........................................................................................................................ 54
12.1.1 Determine reference by weighing ......................................................................................... 54
12.1.2 Numeric entry of the reference weight ................................................................................. 58
12.1.3 Automatic reference optimization ......................................................................................... 59
12.2 Density determination (Hydrostatic weighing) .................................................................... 60
12.2.1 Density determination of solids ............................................................................................ 60
12.2.2 Determining density of liquids .............................................................................................. 64
12.3 Formula mode ......................................................................................................................... 67
12.3.1 Formulation without using the recipe data base ................................................................... 67
12.3.2 Formulation using the recipe data base ............................................................................... 69
12.4 Check weighing ...................................................................................................................... 79
12.5 Percent determination ............................................................................................................ 82
12.5.1 Entering the reference weight by weighing ........................................................................... 82
12.5.2 Numeric entering of the reference weight ............................................................................ 83
12.6 Animal weighing function ...................................................................................................... 84
12.7 Peak value function ................................................................................................................ 85
12.8 GLP function (Good laboratory practice) ............................................................................. 86
13 Data output RS 232C ................................................................................... 88
13.1 Technical data ......................................................................................................................... 88
13.2 Pin allocation of balance output plug .................................................................................. 88
13.3 Interface ................................................................................................................................... 89
13.4 Format Data transfer .............................................................................................................. 90
14 Error messages ............................................................................................ 91
15 Service, maintenance, disposal .................................................................. 92
15.1 Clean ........................................................................................................................................ 92
15.2 Service, maintenance ............................................................................................................. 92
15.3 Disposal ................................................................................................................................... 92
15.4 Instant help ............................................................................................................................. 93
16 Declaration of conformity ........................................................................... 94
ALT_B-PLT_A-PLT_F-BA-e-1333 3
1 Technical data
KERN
ALT
160-4B
ALT
250-4B
ALT
220-5DA
ALT
310-4B
Weighing range (max)
160 g
250 g
60 g; 220 g
310 g
Readability (d)
0.1 mg
0.1 mg
0.00001 g;
0.0001 g
0.1 mg
Reproducibility
0.1 mg
0.1 mg
0.0001 g
0.2 mg
Linearity
± 0.3 mg
± 0.3 mg
0.0003 g
± 0.3 mg
Stabilization time (typical)
4 s
4 s
4 s
4 s
Minimum unit weight at piece counting
> 0.5 mg
> 0.5 mg
> 0.5 mg
> 0,5 mg
Warm-up time
8 hours
Adjustment weight
internal
Reference quantities at piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, dwt, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw)
Secondary voltage powerpack
24 V AC, 500mA
Operating temperature
+ 5° C …. + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 330
Dimensions of windshield (B x D x H) mm
160 x 140 x 205 (inside)
190 x 195 x 225 (outside)
Weighing pan (stainless steel)
Ø 80 mm
Weight kg (net)
5.7 kg
6.5 kg
Interface
RS 232C
4 ALT_B-PLT_A-PLT_F-BA-e-1333
KERN
ALT 160-4BM
ALT 250-4BM
Weighing range (max)
160 g
250 g
Readability (d)
0.1 mg
0.1 mg
Reproducibility
0.1 mg
0.1 mg
Linearity
± 0.3 mg
± 0.3 mg
Verification value (e)
1 mg
1 mg
Verification class
I
I
Minimum load (Min)
10 mg
10 mg
Stabilization time (typical)
4 s
4 s
Minimum unit weight at piece counting
> 0.5 mg
> 0.5 mg
Warm-up time
8 hours
Adjustment weight
internal
Reference quantities at piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, g
Secondary voltage powerpack
24 V AC, 500mA
Operating temperature
+ 5° C …. + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 330
Dimensions of windshield (B x D x H) mm
160 x 140 x 205 (inside)
190 x 195 x 225 (outside)
Weighing pan (stainless steel)
Ø 80 mm
Weight kg (net)
5.7 kg
Interface
RS 232C
ALT_B-PLT_A-PLT_F-BA-e-1333 5
KERN
PLT 720-3A
PLT 1200-3A
PLT 6200-2A
Weighing range (max)
720 g
1200 g
6200 g
Readability (d)
1 mg
1 mg
10 mg
Reproducibility
1 mg
1 mg
10 mg
Linearity
± 2 mg
± 2 mg
± 30 mg
Stabilization time (typical)
2 s
2 s
2 s
Minimum unit weight at piece counting
> 5 mg
> 5 mg
> 50 mg
Warm-up time
4 hours
8 hours
4 hours
Adjustment weight
internal
Reference quantities at piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, dwt, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw)
Input Voltage
220 - 240V AC 50 Hz
Secondary voltage powerpack
24 V AC, 500mA
Operating temperature
+ 5° C / + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 160
210 x 340 x 100
Dimensions of windshield
Ø 150, 60 high (inside)
Ø 160, 70 high (outside)
-
Weighing pan (stainless steel)
Ø 11 cm
Ø 16 cm
Weight kg (net)
4.5 kg
Interface
RS 232C
Underfloor weighing device
Eyelet
6 ALT_B-PLT_A-PLT_F-BA-e-1333
KERN
PLT 420-3F
PLT 4200-2F
Weighing range (max)
420 g
4200 g
Readability (d)
1 mg
10 mg
Reproducibility
1 mg
10 mg
Linearity
± 4 mg
± 40 mg
Stabilization time (typical)
2 s
2 s
Minimum unit weight at piece counting
> 5 mg
> 50 mg
Warm-up time
4 hours
Adjustment weight
internal
Reference quantities at piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, g, gn, lb, mo, oz, ozt,
tl (HK), tl (Singap. Malays), tl (Tw), tol
Input Voltage
220 - 240V AC 50 Hz
Secondary voltage powerpack
9 V 1000 mA, DC
Operating temperature
+ 5° C / + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 160
Dimensions of windshield
Ø 150, 60 high (inside)
Ø 160, 70 high (outside)
Weighing pan (stainless steel)
Ø 11 cm
Ø 16 cm
Weight kg (net)
4 kg
Interface
RS 232C
Underfloor weighing device
-
ALT_B-PLT_A-PLT_F-BA-e-1333 7
KERN
PLT 720-3AM
PLT 6200-2AM
Weighing range (max)
720 g
6200 g
Readability (d)
1 mg
10 mg
Reproducibility
1 mg
10 mg
Linearity
2 mg
30 mg
Verification value (e)
0,01 g
0,1 g
Verification class
II
II
Minimum load (Min)
20 mg
50 mg
Stabilization time (typical)
2 s
2 s
Minimum unit weight at piece counting
> 5 mg
> 50 mg
Warm-up time
4 hours
Adjustment weight
internal
Reference quantities at piece counting
10, 25, 50, 100, freely selectable
Weighing Units
ct, g
Input Voltage
220 - 240V AC 50 Hz
Secondary voltage powerpack
24 V DC, 500 mA,
Operating temperature
+ 5° C / + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
210 x 340 x 160
210 x 340 x 100
Dimensions of windshield
Ø 150, 60 high (inside)
Ø 160, 70 high (outside)
-
Weighing pan (stainless steel)
Ø 11 cm
Ø 16 cm
Weight kg (net)
4.5 kg
4.5 kg
Interface
RS 232C
Underfloor weighing device
Eyelet
8 ALT_B-PLT_A-PLT_F-BA-e-1333
2 Appliance overview
Models ALT
Models PLT, d = 1 mg
2.1 Operating side
ALT_B-PLT_A-PLT_F-BA-e-1333 9
Models PLT, d = 10 mg
Windshield
Levelling screw
Weighing pan
Bubble level
Display
Cover windshield
Keyboard
10 ALT_B-PLT_A-PLT_F-BA-e-1333
2.1.1 Back and bottom side
1. Weighing pan
2. RS232C-interface
3. Footscrews
4. Mains adapter connection
5. Casing screws (in models with four feet first screw out both rear feet)
6. Underfloor weighing device (remove cover)
ALT_B-PLT_A-PLT_F-BA-e-1333 11
2.2 Keyboard overview
Key
Description
Description
MENU button
Invoke main menu How to select menu items
Press until the acoustic signal gets mute
Invoke setup menu Exit setup menu
ON/OFF
button
Turn on/off Exit main menu
CAL-key
Adjustment How to select menu items
PRINT button
Calculate weighing data via interface Confirm / store settings
TARE button
Taring Zeroing
12 ALT_B-PLT_A-PLT_F-BA-e-1333
2.2.1 Navigation buttons / numeric input
Key
Description
Function
Navigation button
Increase digit Scroll forward in menu
Set decimal place
Navigation button
Reduce digit Scroll backwards in menu
Change capital/small letters Press MENU button until the acoustic signal gets mute
Navigation button
Position digit
Delete input
Navigation button
Store
ESC
cancel
ALT_B-PLT_A-PLT_F-BA-e-1333 13
2.3 Overview of display
1
Capacity display
2
Stability display
3
Zero indicator
4
Numerical display/menu
5
Unit
6
Current time
7
Current date
Display
Description
see chap.
Stability display
+ Chap. 10.3
Zero indicator
+ Chap. 10.2
Balance is in percent determination mode
+ Chap. 12.5
PC
Balance is in parts counting mode
+ Chap. 12.1
H
Upper limit
+ Chap.12.4.
L
Lower limit
DS
The balance is in density determination mode
+ Chap. 12.2
g
Weighing unit Gram
Chap. 11.1
ct
Weighing unit Carat
Oz
Weighing unit Ounce
Lb
Weighing unit Pound
Dwt
Weighing unit Pennyweights
Ozt
Weighing unit Troy ounce
GN
Weighing unit Grain
tl 1
Weighing unit Hongkong Tael
tl 2
Weighing unit Singapore Tael
tl 3
Weighing unit Taiwan Tael
mo
Weighing unit Momme
14 ALT_B-PLT_A-PLT_F-BA-e-1333
Operator guidance:
8
Active application
9
Current date
10
Current time
11
Operation step to be carried out
In a selected application you are guided step-by-step through that application. Selectable languages (D, GB, F, IT. ESP, P; see chap. 11.11).
Display example „Piece counting“
ALT_B-PLT_A-PLT_F-BA-e-1333 15
3 Basic instructions
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic balcance”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing pan.. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing pan. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
1. Our conditions in the operation manual are ignored
2. The appliance is used outside the described uses
3. The appliance is modified or opened
4. Mechanical damage or damage by media, liquids, natural wear and tear
5. The appliance is improperly set up or incorrectly electrically connected
6. The measuring system is overloaded
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
16 ALT_B-PLT_A-PLT_F-BA-e-1333
4 Basic Safety Precautions
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
All language versions contain a non-binding translation. The original German is binding.
Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach possibly supplied transport securing devices. Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
4.1 Pay attention to the instructions in the Operation Manual
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
ALT_B-PLT_A-PLT_F-BA-e-1333 17
18 ALT_B-PLT_A-PLT_F-BA-e-1333
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
On the installation site observe the following:
Place the balance on a firm, level surface Avoid extreme heat as well as temperature fluctuation caused by installing next
to a radiator or in the direct sunlight
Protect the balance against direct draughts due to open windows and doors
Avoid jarring during weighing Protect the balance against high humidity, vapours and dust Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed and weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.
ALT_B-PLT_A-PLT_F-BA-e-1333 19
6.2 Unpacking, Scope of delivery
Balance, see chap. 2 Mains adapter Protective cover Operating instructions Transport Securing
Remove device and accessories carefully from packaging, remove packaging material and place device at the planned work place. Verify that there has been no damage and that all packing items are present.
Scope of delivery / serial accessories:
6.3 Placing
The right place is decisive for the accuracy of the weighing results of high-resolution analytical and precision balances (see chap. 6.1).
Remove transportation lock On the left lower side of the balance replace the golden-coloured screw with the
silver-coloured one (see label).
20 ALT_B-PLT_A-PLT_F-BA-e-1333
Install weighing pan as per illustration
Models ALT Models PLT, d = 1 mg
Models PLT, d = 10 mg
ALT_B-PLT_A-PLT_F-BA-e-1333 21
Levelling balance
Level balance with foot screws until the air bubble of the
water balance is in the prescribed circle.
Check levelling regularly.
6.4 Connect and switch-on power supply
Before connecting the mains adapter check if the printed voltage value is the same as the local supply voltage. Only use KERN original mains adapter. Using other makes requires consent by KERN.
Supply power to balance via mains adapter. The balance will carry out a self-test Then the balance is set into stand-by
mode, „OFF“ is displayed.
To switch on press the ON/OFF button. During the display „CAL“ adjustment
is automatically carried out with the internal adjustment weight (see chap.
7.2). As soon as the weight display appears, the balance is ready for weighing.
To switch off, press the ON/OFF button again. The balance returns into
stand-by mode.
6.5 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply. With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
6.6 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1). During this warming up time the balance must be connected to the power supply (mains, accumulator or battery). The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
22 ALT_B-PLT_A-PLT_F-BA-e-1333
7 Adjustment (all models excepted PLT-M)
Internal adjustm.
Adjustment with internal weight after pressing CAL
External adjustm.
Adjustment with external weight
Technical adjustm.
Overwrite internal adjustment weight
Autom. adjustm.
Automatic adjustment with internal weight.
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization.
Ensure that there are no objects on the weighing pan. Issue of an adjustment log, see chap. 11.13
The balances have been set to „Automatic adjustment“ (see chap. 7.2) by factory. If necessary, in the menu the following possibilities of adjustment can be activated.
ALT_B-PLT_A-PLT_F-BA-e-1333 23
Setting adjustment mode:
In weighing mode press the MENU button and keep it pressed until the
acoustic signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Adjustment mode“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
Display
Description
Procedure see chap.
Internal adjustm.
Adjustment with internal adjustment weight
Chap. 7.1
External adjustm.
Adjustment with external adjustment weight
Chap. 7.3
Technical adjustm.
Overwrite internal adjustment weight
Chap. 7.4
Autom. adjustm.
Automatic adjustment with internal adjustment weight
Chap. 7.2
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
24 ALT_B-PLT_A-PLT_F-BA-e-1333
7.1 Adjustment with internal weight after pressing the CAL button
Condition: Menu setting „Internal adjustm.“, see chap. 7
Condition: Menu setting „Autom. adjustm.“, see chap. 7
In weighing mode press the CAL button, adjustment will take place
automatically.
After successful adjustment the balance automatically returns to weighing
mode.
In case of an adjustment error (e.g. objects on the weighing pan) the
display will show an error message „CAL bUt“, repeat adjustment using the CAL button.
7.2 Automatic adjustment with internal weight (factory setting)
The automatic adjustment with internal weight (see chap. 7.1) is started automatically, if the balance
after the weighing balance was disconnected from the mains after pressing ON/OFF in stand-by mode after a temperature change of 1.5 °C with non loaded weighing pan / zero
display
after a time interval of 20 min
with non loaded weighing pan / zero display
The automatic adjustment function is always enabled. However you can make a manual adjustment at any time with the internal or an external weight, see chap. 7.1 or chap. 7.3.
ALT_B-PLT_A-PLT_F-BA-e-1333 25
7.3 Adjustment with external adjustment weight
Condition: Menu setting „External adjustm.“, see chap. 7 At the models with internal adjustment weight the adjustment with
external weight is not recommended.
Information about test weights you will find in the internet under
http://www.kern-sohn.com
Ensure that there are no objects on the weighing pan. In weighing mode press
the CAL button.
Wait until the weighed value for the required adjustment weight appears
flashing.
During the flashing display put the required adjustment weight carefully in the
center of the weighing pan. The flashing display disappears.
After successful adjustment the balance automatically returns to weighing
mode.
Take away adjustment weight
26 ALT_B-PLT_A-PLT_F-BA-e-1333
7.4 Overwrite internal adjustment weight
Overwriting is restricted to specialist staff possessing well acquainted with
the workings of weighing scales.
Information about test weights you will find in the internet under
http://www.kern-sohn.com
Call up menu item „Techn. adjustment“, see chap. 7.  To confirm press the PRINT button and keep it pressed until the acoustic signal
gets mute.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
Ensure that there are no objects on the weighing pan. Press the CAL key
Wait until the weighing value of the adjustment weight appears flashing.
During the flashing display put the required adjustment weight carefully in the
center of the weighing pan. The flashing display disappears.
Wait for stability display, remove adjustment weight.
Keep the PRINT button pressed until the acoustic signal gets mute. The internal
adjustment weight is overwritten.
After successful overwriting, the balance automatically returns to weighing
mode.
ALT_B-PLT_A-PLT_F-BA-e-1333 27
8 Adjustment of models PLT-M
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Ensure that there are no objects on the weighing pan. Issue of an adjustment log, see chap. 11.13
28 ALT_B-PLT_A-PLT_F-BA-e-1333
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons to select the menu item „Adjustment mode“.
Confirm by the PRINT button, „Autom. Just.“ will be displayed.
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.
8.1 Automatic adjustment with internal weight
The automatic adjustment with internal weight is started automatically, if the balance
after the weighing balance was disconnected from the mains after pressing ON/OFF in stand-by mode after a temperature change of 1.5 °C
with non loaded weighing pan / zero display
after a time interval of 20 min with non loaded weighing pan / zero
display
The automatic adjustment function is always enabled. At any time, however, you can carry out a manual adjustment (press CAL-button) using the internal adjustment weight, see chap. 7.1.
Or selecting the adjustment mode in the menu: