1 Technical data
KERN PCB 40-3 PCB 60-3 PCB 100-3 PCB 200-2
Readability (d) 0.001 g 0.001 g 0.001 g 0.01 g
Weighing range (max) 40 g 60 g 100 g 200 g
Taring range (subtractive) 40 g 60 g 100 g 200 g
Reproducibility 0.001 g 0.001 g 0.001 g 0.01 g
Linearity ± 0.003 g ± 0.003 g ± 0.003 g ± 0.02 g
Minimum unit weight at
piece counting
Warm-up time 2 hours 2 hours 2 hours 30 minutes
Reference quantities at
piece counting
Weighing Units
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm 163 x 245 x 79
0.002 g 0.002 g 0.002 g 0.02 g
5, 10, 25, 50
Details „Weighing units“ chapter 7.9
40g (F1) 50g (F2) 100 g (F1) 200 g (M1)
Weighing plate mm Ø 81 Ø 81 Ø 81 Ø 105
Weight kg (net) 1.3 1.3 1.3 1.3
Underfloor weighing Standard
Flush-mounted hook -
36
PCB-BA-defsi-0610
KERN PCB 400-1PCB 400-2PCB 600-2 PCB 800-2
Readability (d) 0.1 g 0.01 g 0.01 g 0.01 g
Weighing range (max) 400 g 400 g 600 g 800 g
Taring range (subtractive) 400 g 400 g 600 g 800 g
Reproducibility 0.1 g 0.01 g 0.01 g 0.01 g
Linearity ± 0.2 g ± 0.03 g ± 0.03 g ± 0.03 g
Minimum unit weight at
piece counting
Warm-up time 10 minutes2 hours 2 hours 2 hours
Reference quantities at
piece counting
Weighing Units
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm 163 x 245 x 79
0.2 g 0.02 g 0.02 g 0.02 g
5, 10, 25, 50
Details „Weighing units“ chapter 7.9
400g (M2) 400g (F2) 500 g (F2) 700 g (F1)
Weighing plate mm 130 x 130 Ø 105 Ø 105 Ø 105
Weight kg (net) 1.3 1.3 1.3 1.3
Underfloor weighing Standard
Flush-mounted hook -
PCB-BA-defsi-0610
37
KERN PCB 1000-1 PCB 1000-2 PCB 2000-1 PCB 4000-0
Readability (d) 0.1 g 0.01 g 0.1 g 1 g
Weighing range (max) 1,000 g 1,000 g 2,000 g 4,000 g
Taring range (subtractive) 1,000 g 1,000 g 2,000 g 4,000 g
Reproducibility 0.1 g 0.01 g 0.1 g 1 g
Linearity ± 0.2 g ± 0.03 g ± 0.2 g ± 2.0 g
Minimum unit weight at
piece counting
Warm-up time 30 minutes2 hours 30 minutes 10 minutes
Reference quantities at
piece counting
Weighing Units
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm 163 x 245 x 79
0.2 g 0.02 g 0.2 g 2.0 g
5, 10, 25, 50
Details „Weighing units“ chapter 7.9
1,000g
(M1)
1,000g (F1)
2,000 g
(M1)
4,000 g (M2)
Weighing plate mm 130 x 130 Ø 105 130 x 130 150 x 170
Weight kg (net) 1.3 1.3 1.3 1.3
Underfloor weighing Standard
Flush-mounted hook -
38
PCB-BA-defsi-0610
KERN PCB 4000-1 PCB 6000-0PCB 6000-1 PCB 8000-1
Readability (d) 0.1 g 1 g 0.1 g 0.1 g
Weighing range (max) 4,000 g 6,000 g 6,000 g 8,000 g
Taring range (subtractive) 4,000 g 6,000 g 6,000 g 8,000 g
Reproducibility 0.1 g 1 g 0.1 g 0.1 g
Linearity ± 0.3 g ± 2 g ± 0.3 g ± 0.3 g
Minimum unit weight at piece
count
Warm-up time 2 hours 2 hours 2 hours 2 hours
Reference unit weights at piece
count
Weighing Units
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 3 sec.
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Housing (B x D x H) mm 163 x 245 x 79
0.2g 2 g 0.2 g 0.2 g
5, 10, 25, 50
Details „Weighing units“ chapter 7.9
4,000 g
(F2)
5,000 g (M2)5,000 g (F2) 7,000 g (F1)
Weighing plate mm 150 x 170 150 x 170 150 x 170 150 x 170
Weight kg (net) 1.3 1.3 1.3 1.3
Underfloor weighing Standard
Flush-mounted hook -
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformità per apparecchi contrassegnati con la marcatura CE
English
Deutsch
Français
Español
Italiano
We hereby declare that the product to which this declaration refers conforms to the
following standards.
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Manifestamos en la presente que el producto al que se refiere esta declaración está de
acuerdo con las normas siguientes
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è
conforme alle norme di seguito citate.
EN 61000-3-2 : 2000
EN 61000-3-3 : 1995+A1 : 2001
EN 55024 : 1998+A1 : 2001
Gottl. KERN & Sohn GmbH
Management
40
PCB-BA-defsi-0610
3 Basic Information (General)
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
plate. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance. (Example: Slowly
draining fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the measuring
system.
Impacts and overloading exceeding the stated maximum load (max) of the balance,
minus a possibly existing tare load, must be strictly avoided. Balance may be
damage by this.
Never operate balance in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
• Our conditions in the operation manual are ignored
• The appliance is used outside the described uses
• The appliance is modified or opened
• Mechanical damage and damage caused by media, liquids,
natural wear and tear
• The appliance is improperly set up or incorrectly electrically connected
• The measuring system is overloaded
PCB-BA-defsi-0610
41
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this test.
Information is available on KERN’s home page (www.kern-sohn.com with regard to
the monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.
4 Basic Safety Precautions
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you
are already familiar with KERN balances.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
5.2 Packaging
Keep all parts of the original packaging in case you need to return the appliance.
Only use original packaging for returning.
Before sending, disconnect all connected cables and loose/movable parts.
Attach possibly existing transport safeguards. Secure all parts, e.g. weighing plate,
power unit etc., to prevent slipping and damage.
42
PCB-BA-defsi-0610
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
• Place the balance on a firm, level surface;
• Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
• Protect the balance against direct draughts due to open windows and doors;
• Avoid jarring during weighing;
• Protect the balance against high humidity, vapours and dust;
• Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance)
may occur if a cold appliance is taken to a considerably warmer environment.
In this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
• Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change location or remove
source of interference.
6.2 Unpacking
Carefully remove the balance from the packaging, remove plastic cover and setup
balance at the intended workstation.
6.2.1 Setup
The balance must be installed in a way that the weighing plate is exactly in horizontal
position.
PCB-BA-defsi-0610
43
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.