KERN CME 300-2, CME 3000-1, CME 1000-2, CME 100-3, CME 6000-1 Operating Instruction

...
KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Operating instruction Counting scales/Counting system
KERN CDE/CME/CDEE
Version 1.4 04/2010 GB
CDE/CME/CDEE-BA-e-1014
CDE/CME/CDEE-BA-e-1014
2
GB
KERN CDE/CME/CDEE
Version 1.4 04/2010
Operating instruction Counting scales/Counting system
Table of Contents
1 Technical data................................................................................................ 4
2 Declaration of conformity ............................................................................. 8
3 Basic Information (General).......................................................................... 9
3.1 Proper use................................................................................................................................. 9
3.2 Improper Use ............................................................................................................................ 9
3.3 Warranty .................................................................................................................................... 9
3.4 Monitoring of Test Resources................................................................................................. 9
4 Basic Safety Precautions ............................................................................ 10
4.1 Pay attention to the instructions in the Operation Manual ................................................ 10
4.2 Personnel training.................................................................................................................. 10
5 Transport and storage................................................................................. 10
5.1 Testing upon acceptance ...................................................................................................... 10
5.2 Packaging................................................................................................................................ 10
6 Unpacking, Setup and Commissioning ..................................................... 10
6.1 Installation Site, Location of Use.......................................................................................... 10
6.2 Unpacking ............................................................................................................................... 11
6.2.1 Placing...................................................................................................................................... 11
6.2.2 Scope of delivery ...................................................................................................................... 11
6.2.3 Assembly instructions for the use of a tripod (KERN CDE only, optional)............................... 12
6.3 Mains connection ................................................................................................................... 13
6.4 Operation using a (rechargeable) battery (optional) .......................................................... 13
6.5 Connection of peripheral devices......................................................................................... 13
6.6 Initial Commissioning ............................................................................................................ 13
6.7 Adjustment.............................................................................................................................. 14
6.8 Adjustment.............................................................................................................................. 14
7 Operation...................................................................................................... 15
7.1 Overview of display................................................................................................................ 15
7.2 Keyboard overview................................................................................................................. 15
7.3 Weighing ................................................................................................................................. 16
7.4 Taring....................................................................................................................................... 16
7.5 PRETARE - Function..............................................................................................................16
7.6 Plus/minus weighings............................................................................................................17
7.7 Parts counting ........................................................................................................................ 17
7.7.1 Automatic reference optimization............................................................................................. 18
7.8 Percent weighing.................................................................................................................... 18
CDE/CME/CDEE-BA-e-1014
3
8 The menu...................................................................................................... 19
8.1 Invoke menu............................................................................................................................ 19
8.2 Exit menu ................................................................................................................................ 19
8.3 Menu overview........................................................................................................................ 20
8.4 Description of individual functions ...................................................................................... 22
8.4.1 Weighing units (Unit) ................................................................................................................ 22
8.4.2 Backlighting of display (CDE models only) .............................................................................. 23
8.4.3 Dosing and Zero-tracking ......................................................................................................... 24
8.4.4 Selection of the adjustment weight........................................................................................... 25
8.4.5 Reset to factory setting............................................................................................................. 25
9 Data output RS 232 C .................................................................................. 26
9.1 Technical data......................................................................................................................... 26
9.2 Pin allocation of the balance output socket (front view) ................................................... 26
9.3 Interface parameter ................................................................................................................ 26
9.3.1 Data transfer mode................................................................................................................... 26
9.3.2 Baud rate .................................................................................................................................. 27
9.3.3 Selection printed edition ........................................................................................................... 27
9.4 Explanation of the data transfer ........................................................................................... 28
9.4.1 AU PC...................................................................................................................................... 29
9.5 Output on bar code printer.................................................................................................... 31
10 Service, maintenance, disposal.................................................................. 32
10.1 Cleaning .................................................................................................................................. 32
10.2 Service, maintenance............................................................................................................. 32
10.3 Disposal................................................................................................................................... 32
11 Instant help................................................................................................... 33
12 Counting system CDEE ............................................................................... 34
12.1 Introduction............................................................................................................................. 34
12.2 Technical data......................................................................................................................... 34
12.3 Basic structure ....................................................................................................................... 34
12.4 Installation............................................................................................................................... 35
12.5 Balance settings..................................................................................................................... 35
12.6 Counting with both scales .................................................................................................... 35
CDE/CME/CDEE-BA-e-1014
4
1 Technical data
KERN CME 100-3 CME 300-2 CME 1000-2
Readability (d) 0,001 g 0,01 g 0,01 g
Weighing range (max) 100 g 300 g 1000 g
Reproducibility 0,001 g 0,01 g 0,01 g
Minimum piece weight 0,002 g 0,02 g 0,02 g
Linearity ± 0,003 g ± 0,02 g ± 0,03 g
Recommended adjustment weight, not included (class) For details on “Selecting an adjustment weight“ see chapter 8.4.4
100 g (F 2) 300 g (M 1) 1000 g (F2)
Warming up time (operating temperature)
2 h
Reference unit weights at piece count
5, 10, 20
Stabilization time (typical) 2-3 sec
Electric Supply
300 mA/9V 300 mA/9V 300 mA/9V
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Weighing platform, synthetic material
Ø 105 mm Ø 150 mm Ø 150 mm
Dimensions of the housing (B x D x H)
170 x 240 x 39
Data interface RS 232C RS 232C RS 232C
Weight kg (net) 0,7 0,7 0,7
Battery operation 9 V- compound battery (optional)
CDE/CME/CDEE-BA-e-1014
5
KERN CME 3000-1 CME 6000-1
Readability (d) 0,1 g 0,1 g
Weighing range (max) 3000 g 6000g
Reproducibility 0,1 g 0,1 g
Minimum piece weight 0,2 g 0,2 g
Linearity ± 0,2 g ± 0,3 g
Recommended adjustment weight, not included (class) For details on “Selecting an adjustment weight“ see chapter 8.4.4
3000 g (M 1) 6000 g (M 1)
Warming up time (operating temperature)
2 h
Reference unit weights at piece count
5, 10, 20
Stabilization time (typical) 2-3 sec
Electric Supply
300mA/9V 300mA/9V
Operating temperature + 5° C .... + 35° C
Humidity of air max. 80 % (not condensing)
Weighing platform, synthetic material
Ø 150 mm Ø 150 mm
Dimensions of the housing (B x D x H)
170 x 240 x 39
Data interface RS 232C RS 232C
Weight kg (net) 0,7 0,5
Battery operation 9 V- compound battery (optional)
CDE/CME/CDEE-BA-e-1014
6
KERN CDE 35K1 CDE 35K1L CDE 60K2
Readability (d) 1 g 1 g 2 g
Weighing range (max) 35 kg 35 kg 60 kg
Reproducibility 1 g 1 g 2 g
Minimum piece weight 2 g 2 g 4 g
Linearity 3 g 3 g 6 g
Recommended adjustment weight, not included (class) For details on “Selecting an adjustment weight“ see chapter 8.4.4
30 kg (F2) 20 kg (F2) 60 kg (F2)
Warming up time (operating temperature)
2 h
Reference unit weights at piece count
5, 10, 20 5, 10, 20 5, 10, 20
Stabilization time (typical)
2-3 sec
Electric Supply
DC 9V/300 mA
Operating temperature
+ 5° C .... + 35° C
Humidity of air
max. 80 % (not condensing)
Weighing plate, stainless steel
315x305x85 mm
Terminal 200 x 100 x 55
Dimensions of the housing (B x D x H)
Platform 315 x 305 x 85
RS 232C Data interface
Connection reference balance
Weight kg (net)
5 16 5
Battery operation
9 V- compound battery (optional)
Rechargeable battery (optional)
Working life 15 h with display backlighting/ charging
time 10 h
CDE/CME/CDEE-BA-e-1014
7
KERN CDE 60K2L CDE 150K5 CDE 150K5L
Readability (d) 2 g 5 g 5 g
Weighing range (max) 60 kg 150 kg 150 kg
Reproducibility 2 g 5 g 5 g
Minimum piece weight 4 g 10 g 10 g
Linearity 6 g 15 g 15 g
Recommended adjustment weight, not included (class) For details on “Selecting an adjustment weight“ see chapter 8.4.4
60 kg (F2) 150 kg (F2) 150 kg (F2)
Warming up time (operating temperature)
2 h
Reference unit weights at piece count
5, 10, 20 5, 10, 20 5, 10, 20
Stabilization time (typical)
2-3 sec
Electric Supply
DC 9V/300 mA
Operating temperature
+ 5° C .... + 35° C
Humidity of air
max. 80 % (not condensing)
Weighing plate, stainless steel
315 x 305 x 85 mm
Terminal 200 x 100 x 55
Dimensions of the housing (B x D x H)
Platform 315 x 305 x 85
RS 232C
Data interface
Connection reference balance
Weight kg (net)
16 5 16
Battery operation
9 V- compound battery (optional)
Rechargeable battery (optional)
Working life 15 h with display backlighting/ charging
time 10 h
CDE/CME/CDEE-BA-e-1014
8
2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
EC-Konformitätserklärung EC-Declaration of -Conformity EC-
Déclaration de conformité EC-Declaración de Conformidad
EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
D
Konformitäts­erklärung
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
GB
Declaration of conformity
We hereby declare that the product to which this declaration refers conforms with the following standards.
CZ
Prohlášení o shode
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami.
RUS
Заявление о соответствии
Мы заявляем, что продукт, к которому относится данная декларация, соответствует перечисленным ниже нормам.
Electronic Scale: KERN CDE, CME
Mark applied EU Directive Standards
2004/108/EC EN 55022: 1998+A1+A2
EN 61000-3-2: 2000+A2 EN 61000-3-3: 1995+A1 EN 55024: 1998+A1+A2
Date: 27.10.2008 Signature:
Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
CDE/CME/CDEE-BA-e-1014
9
3 Basic Information (General)
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage and damage caused by media, liquids
Natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
CDE/CME/CDEE-BA-e-1014
10
4 Basic Safety Precautions
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
5.2 Packaging
Keep all parts of the original packaging in case you need to return the appliance. Only use original packaging for returning. Before sending, disconnect all connected cables and loose/movable parts. Attach possibly existing transport safeguards. Secure all parts, e.g. weighing plate, mains adapter etc., to prevent slipping and damage.
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
CDE/CME/CDEE-BA-e-1014
11
Therefore, observe the following for the installation site:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing next to
a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time. Non-
permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed and weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.
6.2 Unpacking
Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.
6.2.1 Placing
The balance must be installed in a way that the weighing plate is exactly in horizontal
position.
6.2.2 Scope of delivery Serial accessories:
Models CME Models CDE
Balance • Terminal
Mains power supply Platform
Operating Manual • Mains power supply
Protective cover
Operating Manual
Loading...
+ 24 hidden pages