KEPCO KLR User Manual

Page 1
QUICK START GUIDE
KEPCO
An ISO 9001 Company.
KLR
KLR POWER SUPPLY
This guide gives a brief introduction to the KLR Power supply, shows simple load connections, and allows you to verify the power supply is working. The guide also shows you how to use the front panel controls to perform the most commonly used functions.
Refer to the KLR User Manual for full specifications, installation considerations and operating instruc­tions. An Installation/Operation Summary includes hyperlinked references to detailed procedure, but which can be printed as a handy reference.
Refer to the KLR Developer’s Guide for a full description of the digital interfaces, their associated driv­ers, and the SCPI command language.
The User Manual and Developer’s Guide can be downloaded from the Kepco web site at www.kepcopower.com/support/. Drivers can downloaded from www.kepcopower.com/drivers/.
CONTENTS
DESCRIPTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
OPTIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
UNPACKING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
EQUIPMENT SUPPLIED. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
ACCESSORIES. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
SAFETY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
PRELIMINARY OPERATIONAL CHECK. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
INSTALLATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Input Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Load Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Local Sensing (Factory Default). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remote Sensing Select. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Analog I/O Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
GPIB Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
RS 232 Connections (Standard Models). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
LAN Connections (E-Series Models). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
OPERATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Turning The Power Supply On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Enabling/disabling Output Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Checking Or Changing Voltage/current Setpoints. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Limit Model Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Setting Voltage Or Current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Last Setting Recall. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Viewing/changing Overvoltage or Overcurrent Protection Values. . . . . . . . . . . . . . . . . . . . . . . . 10
KLR OUTLINE DIMENSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
KEPCO, INC. 131-38 SANFORD AVENUE FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
©2013, KEPCO, INC Data subject to change without notice
http://www.kepcopower.com email: hq@kepcopower.com
228-1689 REV 1 091713
Page 2

DESCRIPTION. Five KLR Models and five KLR E-Series Models are available, as listed in Table 1.

The standard Models are GPIB and RS 232 compatible; the KLR E-Series Models (identified by the E suffix) are GPIB and LAN (ethernet, LAN) compatible. The only difference between standard and E­Series models is that LAN replaces the RS 232 interface on E-Series models. All models allow remote analog programming as well as local control. For each model a multitude of limit models can be config­ured. Limit models allow the unit to be configured to maximum settings below the rated voltage and/or rated current. When a limit model is established the unit will not accept programming values above the user-defined limits.

OPTIONS. The Rapid Output Discharge Circuit (RODC) option (suffix R) is available on all KLR mod-

els. This circuit rapidly discharges the output capacitance, thus significantly reducing response time to reductions in output voltage (see KLR User Manual for details).

TABLE 1. MODEL PARAMETERS

Model number
KLR 20-120
KLR 40-60
KLR 75-32
KLR 150-16
KLR 300-8
(1) Bandwidth: 20MHz; low frequency ripple may be higher at loads less than 30 Watts. (2) Specifications apply to both standard and E Series (Suffix E) models.
(2)
(2)
(2)
(2)
(2)
Rated
Vol tag e
Range
0-20V 1.5A 0-120A 100 mV 87%
0-40V 0.8A 0-60A 80 mV 88%
0-75V 0.4A 0-33.3A 80 mV 87%
0-150V 0.2A 0-16A 100 mV 88%
0-300V 0.1A 0-8A 150 mV 89%
Minimum
Programmable
current
Rated
Current
Range
Ripple and
noise
(1)
p-p
Efficiency
@230 Va-c

UNPACKING. This instrument has been thoroughly inspected and tested prior to packing and is

ready for operation. After careful unpacking, inspect for shipping damage before attempting to operate. Perform the PRELIMINARY OPERATIONAL CHECK. If any indication of damage is found, file an imme­diate claim with the responsible transport service.
EQUIPMENT SUPPLIED. See Table 2.

TABLE 2. EQUIPMENT SUPPLIED

ITEM
Jumper (24 AWG or larger bus wire) for local sensing connections 172-0638 2
Analog I/O port mating connector 142-0528 1
Nut with captive lockwasher (on output studs of 20V and 40V models only) 102-0175 2
Quick Start Guide* 228-1689 1
* User Manual and Developer’s Guide are available as free downloads from: www.kepcopower.com/support/opmanls.htm#klr Drivers are available as free downloads from www.kepcopower.com/drivers/drivers-dl3.htm#klr
PART NUMBER QUANTITY
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
http://www.kepcopower.com email: hq@kepcopower.com
2 228-1689 REV 1 091713
Page 3

ACCESSORIES. See Table 3.

TABLE 3. ACCESSORIES

ITEM FUNCTION
Line Cord Set (250V/20A) 2.5m long cord set, provides for source power
Line Cord Set (250V/15A) 2.5m long cord set, provides for source power
Source power connector
IEEE 488 (GPIB) Cable, 1m long Connect KLR Power Supply to GPIB bus. SNC 488-1
IEEE 488 (GPIB) Cable, 2m long Connect KLR Power Supply to GPIB bus. SNC 488-2
IEEE 488 (GPIB) Cable, 4m long Connect KLR Power Supply to GPIB bus. SNC 488-4
Chassis Slide (2 required per power supply)
Analog Connector Backshell Locks analog port mating connector to KLR via jackscrews. 108-0204
Loop Back Test Connector Used for verification of RS 232 operation. 195-0112
Null Modem Cable, 10 ft. Connect RS 232 port with controlling computer, DB9F to DB9F, 10 feet
Null Modem Cable, 1 ft. Used for Master/Slave connections. Connects RS 232 ports of stacked
LAN Patch cable Connects KLR LAN port to LAN, 10 ft long. 118-1115
Support bracket, rear Provides extra support to rear of rack-mounted unit if needed. Two
connection. Mates with NEMA 6-20R receptacle (see adjacent figure). Supports rated load power over mains voltage range of 180-265V a-c.
connection. Mates with NEMA 6-15R receptacle (see adjacent figure). Supports rated load power over mains voltage range of 180-265V a-c.
Mates with a-c input connector J1 (see Figure 1-2) for user-wired source power connection. Supports rated load power over mains volt-
age range of 180-265V a-c.
Allows rack-mounted units to slide in and out. 108-0239
long.
power supplies, DB9F to DB9F, one foot long.
brackets required per unit. Each bracket requires 2 screws and 2 washers. See Figure 4, sheet 1, Detail “C” for requirements.
NEMA 6-20R 118-1235
G
W
NEMA 6-15R 118-1137
KEPCO
PART NUMBER
142-0381
(Jonathan
375-QD Series)
118-1176
118-1220
128-2306

SAFETY. See Table 4

TABLE 4. SAFETY SYMBOLS

SYMBOL MEANING
CAUTION: RISK OF ELECTRIC SHOCK.
CAUTION: REFER TO REFERENCED PROCEDURE.
!
WARNING INDICATES THE POSSIBILITY OF BODILY INJURY OR DEATH.
CAUTION INDICATES THE POSSIBILITY OF EQUIPMENT DAMAGE.
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
091713 228-1689 REV 1 3
http://www.kepcopower.com email: hq@kepcopower.com
Page 4

PRELIMINARY OPERATIONAL CHECK. A simple operational check after unpacking and

before equipment installation is advisable to ascertain whether the power supply has suffered damage resulting from shipping.
NOTE: This test must be performed with all I/O ports disconnected, and ANALOG I/O SETTINGS
switch positions 3, 4 and 5 set to OFF (up).
1 With POWER circuit breaker set to OFF position, connect the power supply to source power (see
See “Input Connections.” on page 5.).
2 With no load connected, set POWER circuit breaker to the ON position. Each time the unit is turned
on an internal self-test is performed (see “Turning The Power Supply On.” on page 7). After the test has been successfully completed, the status display reads SET (setpoint mode), the DC VOLTS dis­play reads 0 Volts and the DC AMPERES display reads minimum Amperes (see Note 1, below).
NOTES: 1. A minimum programmed current (actual value depends on model) is required to ensure proper
operation of the power supply under all load conditions. Programmed current is automatically set to be at least the minimum current.
2. If an error indication is blinking in the Status display, refer to User Manual for an explanation of error codes.
3 Rotate VOLTAGE adjust knob clockwise. Verify that DC VOLTS display increases in large (X 100)
increments.
4 Press and rotate VOLTAGE adjust knob clockwise. Verify the DC VOLTS display increases in finer
increments than step 3, then release knob.
5 Adjust VOLTAGE adjust knob clockwise until Status display reads >MAX. Tap either VOLTAGE or
CURRENT adjust knob once to enter values. Verify Status display is blank
6 Connect a digital voltmeter (DVM) to the (M+) and (M–) terminals on the rear panel.
7 Press and release DC OUTPUT switch to enable the output. Verify DC OUTPUT indicator lights.
8 Compare the programmed output voltage value (step 5) with the voltage reading of the DVM; the dif-
ference between the two should not exceed 0.05% of the maximum voltage of the unit.
9 Compare the voltage reading of the DC VOLTS display with that of the DVM; the difference between
the two should not exceed 0.1% of the maximum voltage of the unit.
10 Enter different value for output voltage, then repeat steps 4 and 5 using different values for pro-
grammed voltage.
11 Disable the output by pressing and releasing DC OUTPUT switch; verify front panel DC VOLTS and
DC AMPERES displays read 0.0V and minimal current and the DC OUTPUT indicator is off.
12 Set POWER ON/OFF circuit breaker to OFF and disconnect from source power, load and test equip-
ment.
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
4 228-1689 REV 1 091713
http://www.kepcopower.com email: hq@kepcopower.com
Page 5

INSTALLATION. Install units in a 19 inch-wide rack. Optional slides may be used. Leave the front

and rear panels clear of obstructions to ensure adequate cooling. For parallel, series and master-slave configurations, refer to the User Manual.

INPUT CONNECTIONS. Operation at rated load power (2400W) over the full rated source voltage range (180

- 265V a-c) requires a branch service rating of 20 Amperes; Kepco recommends use of #12 AWG (2,0 mm mary wiring. For operation at output power of 2000W or less, or with source voltage not less than 200V a-c, branch service may be reduced to 15 Amperes using #14 AWG (1,6 mm
2
) wiring. Line cords and a user-wireable
2
) for pri-
source power connector are available as accessories (see Table 3). Plug the (wired) source power con­nector or line cord into the source power inlet connector at the rear panel.

LOAD CONNECTIONS. Connect the load to the (+) and (–) DC OUTPUT terminals on the rear panel

(Figure 2). (+M) and (–M) outputs are for local connection of error sense leads.
Configuration of local sensing is facilitated by pre-installed jumpers which configure the unit for local sensing as shown in Figure 2.
NOTE: Output Sense lines must be connected for proper operation, either locally, or at the load
(remote).
POWER ON/OFF
Turns unit on/off.
Circuit Breaker
gives input protection.
DC VOLTS
Display shows
output voltage or
voltage setpoint.
(Also used for
special functions).
CV
indicator lights
green when operating in
constant voltage
mode
LAN
indicator
lights green
when LAN
interface is
in use
[E-Series Models]
CC
indicator lights
amber when
operating in
constant current
mode
DC AMPERES
display shows
actual output
current or
current set point
(Also used for
special functions.)
DC OUTPUT indicator lights green to show output enabled
VOLTAGE
control/momentary
switch
Rotate to set
output voltage.
Tap to enter
SETPOINT mode.
(Also used for
other functions.)

FIGURE 1. KLR SERIES, FRONT PANEL CONTROLS AND INDICATORS

FUNCTION
momentary switch
Used to access
calibration,
GPIB address,
RS232 baud rate
(Standard models),
Limit Model and Utility functions.
Requires a thin tool
(e.g., end of paper
clip) to press
switch.
Status
Display normally
blank. Shows
active function or
blinks for error
messages.
PROTECT
momentary switch
Used to set
overvoltage and
overcurrent
protection limits.
Requires a thin
tool (e.g., end of
paper clip) to press switch.
CURRENT
control/momentary
switch
Rotate to set
output current.
Tap to enter
SETPOINT mode.
(Also used for
other functions.)
DC OUTPUT
on/off switch
enables or
disables output.
(Also used to
accept front panel
inputs.)
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
091713 228-1689 REV 1 5
http://www.kepcopower.com email: hq@kepcopower.com
Page 6

LOCAL SENSING (FACTORY DEFAULT). Unit is shipped with local sensing jumpers installed: +S

connected to +M and –M connected to –S (see Figure 2).

FIGURE 2. KLR SERIES, REAR PANEL VIEW, JUMPERS INSTALLED FOR LOCAL SENSING

REMOTE SENSING SELECT. First remove the factory-installed local sensing jumpers between +S

and +M and between –M and –S. Then connect the +S and –S lines at the load, observing proper polar­ity (see Figure 3). A high frequency bypass network consisting of two capacitors connected across the error sensing point as shown in Figure 3 is recommended to reduce noise in the sense loop.

FIGURE 3. REMOTE SENSING

KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
6 228-1689 REV 1 091713
http://www.kepcopower.com email: hq@kepcopower.com
Page 7

ANALOG I/O CONNECTIONS. The Analog I/O Port connector, located on the rear panel of the KLR

power supply (see Figure 2), provides access to analog programming inputs, status signal outputs and an external trigger input available for use with SCPI *TRG and TRIG commands.

GPIB CONNECTIONS. Your computer must have a GPIB interface card installed. Connect the power

supply to the computer’s GPIB interface card. Use a standard GPIB interface cable at the GPIB port on the rear panel (see Figure 2). The default GPIB address is 6. Refer to the User manual to change it from the front panel, or to the Developer’s Guide to change it using SCPI commands.

RS 232 CONNECTIONS (STANDARD MODELS). Connect the KLR to a modem using a Null Modem

patch cable at the RS 232 port located on the rear panel (See Figure 2). A Null Modem cable is not required for older MAC computers with D-sub serial port in which the RXD and TXD line transposition is accomplished via external hardware. The default baud rate is 38400. Refer to the User manual to change it from the front panel, or to the Developer’s Guide to change it using SCPI commands.

LAN CONNECTIONS (E-SERIES MODELS). The LAN interface allows the KLR to communicate via a

built-in web port (using web pages from a standard browser), as well as through other built-in LAN ports. Connect the KLR to the LAN via the RJ 45 LAN port located at the rear panel (See Figure 2). Refer to the User manual to configure the LAN interface from the front panel or through the web interface. Refer to the Developer’s Guide for the associated SCPI commands

OPERATION. Additional features covered in the User Manual are: Quick Boot (eliminating the power-

up displays), use of the internal relay, operation via the LAN interface or analog signals and setting coarse/fine adjustment preference of the VOLTAGE and CURRENT controls. An Installation/Operation Summary is also included in the User manual. The Developer’s Guide covers the GPIB and RS 232 and LAN interfaces, including the use of the drivers available from www.kepcopower.com/drivers/.

TURNING THE POWER SUPPLY ON. To turn the power supply on, set POWER ON/OFF circuit

breaker (Figure 1) to ON. If an error occurs, an error code is displayed in the status display (see User Manual.)
!
CAUTION: DO NOT attempt to “tease” the switch contacts closed by repetitive press-and-
release action. The circuit breaker is “trip-free” design; if overload exists, contacts cannot be held closed by actuator. After turning the unit off, allow at least 5 seconds before turning the unit on again.
1 When the power supply is turned on, it performs a self-test and displays the following information:
Status display flashes KLR for 2 seconds while DC VOLTS and DC AMPERES displays show the max- imum voltage and current capability of the unit.
Status display flashes
Status display flashes installed.
Status display flashes LMT for 2 seconds while DC VOLTS and DC AMPERES displays show the max- imum voltage and current allowed for a previously programmed Limit model.
Status display flashes ously stored OVP and OCP trigger levels.
LAN for 2 seconds if unit is an E-Series model
RODC for 2 seconds if unit has the Rapid Output Discharge Circuit option
PROT for 2 seconds while DC VOLTS and DC AMPERES displays show previ-
2 After a successful self test, the default conditions upon power up are as follows:
Status display shows SET,
Output is disabled (green DC OUTPUT indicator is off).
DC VOLTS and DC AMPERES displays show programmed output conditions: 0 Volts, minimum Amperes (if DIP switch position 3 was ON when the unit was turned off, previously saved setpoint val­ues are displayed).
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
091713 228-1689 REV 1 7
http://www.kepcopower.com email: hq@kepcopower.com
Page 8
NOTE: A minimum programmed current is required to ensure proper operation of the power supply under all
load conditions. Programmed current is automatically set to be at least the minimum current (actual value depends on model, see Table 1).
Since the output is off, Constant Voltage (CV) mode indicator (green LED) and Constant Current (CC) indicator (amber LED) are off.

ENABLING/DISABLING OUTPUT POWER. When the power supply is turned on, the output is auto-

matically disabled (DC OUTPUT LED is off), the DC VOLTS and DC AMPERES displays show the pro- grammed output voltage and current set points and “
SET” appears in the status display.
1 To enable the output, first exit
SET mode by tapping either the VOLTAGE or CURRENT control, then
press and release the DC OUTPUT switch. The associated green LED lights to indicate output power is applied to the load. Each time you exit
SET mode, the setpoint values are stored for possible
recall if the power supply is turned off or the output is disabled.
2 To disable the output, press and release the DC OUTPUT switch again. The DC OUTPUT indicator
goes off.
NOTE: While in the setpoint mode (Status display reads
SET) the output cannot be turned on (press-
ing DC OUTPUT has no effect). If output was on while setpoint mode was entered, pressing DC OUTPUT will disable the output.

CHECKING OR CHANGING VOLTAGE/CURRENT SETPOINTS.

CAUTION: When the output is disabled, the DC VOLTS and DC AMPERES displays show the
actual output voltage and current. Before enabling the output ALWAYS check the set­points to avoid possible damage to the load.
1 Tap either the VOLTAGE or CURRENT control. The status display reads SET, and the DC VOLTS
and DC AMPERES displays show the stored setpoints.
2 To accept the displayed value, tap the associated adjustment control again. To change the value,
rotate the control (press the control in while rotating for fine adjustment; or see User Manual to set coarse/fine preference), then tap the adjustment control again to accept the new setting.
NOTE: Before changing the setpoint, note the displayed setpoint. If you decide not to change the
value after rotating the control, you must rotate the control to the value noted, then tap the adjustment control to accept.

LIMIT MODEL SETTING. The limit model establishes a maximum programmable voltage and current

for the unit within the
maximum voltage and current ratings listed in Table 1. The unit will not accept pro-
grammed values beyond these values.
1 Using a thin tool (e.g., a paper clip), press the FUNCTION switch until the status display reads
LMT.
The DC VOLTS and DC AMPERES displays show the programmed maximum voltage and current of the limit model.
2 To exit, rotate either VOLTAGE or CURRENT control until Status display reads
EXIT, then tap DC
OUTPUT switch. To change the limit model settings refer to the User Manual.
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
8 228-1689 REV 1 091713
http://www.kepcopower.com email: hq@kepcopower.com
Page 9

SETTING VOLTAGE OR CURRENT. The VOLTAGE and CURRENT controls adjust output voltage

and current limit, respectively, when the unit is in constant voltage (CV) mode and adjust voltage limit and output current, respectively, when the unit is in constant current (CC) mode. The mode (CV or CC) is determined by the load together with the programmed settings. As long as the voltage across the load produces a current that is less than the programmed Current setpoint, the unit operates in CV mode (voltage programmed to voltage setpoint, current limited by current setpoint). If the load changes to the point that current through the load reaches the current setpoint, the unit automatically enters CC mode (current programmed to current setpoint, voltage limited by voltage setpoint).
Output voltage or current can be set at the front panel in two ways: Real-time adjustment or Setpoint adjustment.

Real-time Voltage/Current Adjustment. Rotating the associated control will change the output voltage or current in real-time only if the output is enabled. If the unit is in constant voltage mode (CV indicator lit) the DC VOLTS display shows the actual output voltage as the VO LTAG E control is rotated. Similarly, if the unit is in constant current mode (CC indicator lit) the DC AMPERES display shows the actual output current as the CURRENT control is rotated.

If the unit is in CV mode, rotating the CURRENT control will affect the current limit even though the DC
AMPERES display does not change since it is showing actual output current. Similarly, rotating the VOLTAGE control while in CC mode affects the voltage limit. To change the limits to a precise value,
refer to SETPOINT ADJUSTMENT.
Voltage and current settings are not allowed to exceed either a) the limit model setting or b) 80% of the protection value. Attempts to set voltage or current beyond these limits will not be accepted. The status display shows either >MAX for limit model, >OVP for voltage protection, or >OCP for current protection to indicate which limit is being exceeded.

Setpoint Voltage/Current Adjustment.

1 Tap either the VOLTAGE or CURRENT control to initiate setpoint adjustment (status display reads
SET). The previous setpoint is visible on the corresponding DC VOLTS or DC AMPERES display.
2 Rotate the corresponding control to change the setpoint as viewed on the corresponding LED dis-
play. Tap the adjustment control again to accept the new setting.
In the setpoint mode, both coarse and fine adjustment (depending on whether the control is pressed in while rotating) is available. The coarse adjustment is approximately 100X the fine adjustment resolu­tion. Refer to User Manual to set coarse/fine preference.
Setpoint adjust can be done with output either on or off, however the output can not be enabled while setpoint is active.
Voltage and current settings are not allowed to exceed either a) the limit model setting or b) 80% of the protection value. Attempts to set voltage or current beyond these limits will not be accepted. The status display shows either
>MAX for limit model, >OVP for voltage protection, or >OCP for current protection to
indicate which limit is being exceeded.
If you decide not to change the value after rotating the control, the previous setpoint value will be restored after about 20 seconds of inactivity. Then tap either adjustment control to accept.
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
091713 228-1689 REV 1 9
http://www.kepcopower.com email: hq@kepcopower.com
Page 10

LAST SETTING RECALL. KLR is capable of saving the last setpoint values for voltage and current

prior to unit shutdown, and recalling them when the unit is next turned on. Enable this function prior to power-up by setting Analog I/O Switch 3 to the ON (down) position (see Figure 2). Upon next power-up, each setpoint entry for voltage and current is stored in non-volatile memory, with latest settings displac­ing prior ones. When the power supply is turned off, the last settings are saved and automatically retrieved at next power-up as the starting setpoints.
Recall is for SETPOINT ADJUSTMENTS only; output changes entered as REALTIME ADJUSTMENTS are not saved.
To disable this function, turn the unit off and set Analog I/O Switch 3 to the OFF (up) position. At next power-up, the stored value buffer is cleared and setpoints will default to zero volts and minimum cur­rent.
If the limit model for the power supply is altered, the last settings are reset to zero volts and minimum current.

VIEWING/CHANGING OVERVOLTAGE OR OVERCURRENT PROTECTION VALUES.

Overvoltage and Overcurrent protection values can be individually programmed. The range for overvolt­age and overcurrent values are 0.2 to 1.2 x E
max and 0.72 to 1.2 x IOmax.
O
If the output voltage/current exceeds the overvoltage/overcurrent protection value, the protection circuit latches the output off, flashes an overvoltage (OVP) or overcurrent (OCP) protection error message on the status display and sets a status bit that can be retrieved through the RS 232, GPIB or LAN port. The unit must be cycled on and off to restore the output.
The maximum values are 1.2 x EOmax for overvoltage protection (OVP), and 1.2 x IOmax for overcur­rent protection (OCP).
A built-in feature prevents the unit from encroaching upon the above margins for OVP and OCP. For exam- ple, if OVP and OCP are set to 12V and 72A, respectively, programming of the output is automatically limited to 10V and 60A.
When a limit model is defined, OVP and OCP are automatically set 20% and 20%, respectively, above the maximum programmable values established by the limit model. If OVP or OCP are lowered by the user, valid settings will actually be lower than the limit model limits because values within 20% of OVP or 20% of OCP are prohibited.
If either adjustment exceeds maximum programmable volts or amps, or if the adjustment is within 20% of OVP or 20% of OCP, the status display will show
>MAX.
1 Using a thin tool (e.g., end of paper clip), press and hold the PROTECT switch. The output is
switched off, setting voltage to zero Volts, and current to a minimal value, and the status display shows
PROT. The DC VOLTS and DC AMPERES displays show the corresponding overvoltage and
overcurrent protection setpoints while the PROTECT switch is held in.
2 To change the values, operate the corresponding adjustment control while holding the PROTECT
switch in. When the PROTECT switch is released, the protection values showing in the DC VOLTS and DC AMPERES displays are entered as the new protection values.
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
10 228-1689 REV 1 091713
http://www.kepcopower.com email: hq@kepcopower.com
Page 11
FIGURE 4. KLR OUTLINE DIMENSIONS (SHEET 1 OF 2)
KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
091713 228-1689 REV 1 11
http://www.kepcopower.com email: hq@kepcopower.com
Page 12

FIGURE 4. KLR OUTLINE DIMENSIONS (SHEET 2 OF 2)

KEPCO, INC. 131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
http://www.kepcopower.com email: hq@kepcopower.com
12 228-1689 REV 1 091713
Loading...