KEPCO EL Series User Manual

Page 1
INSTRUCTION MANUAL
An ISO 9001 Company.
NON-VOLATILE MEMORY
SAVE, CLEAR AND RESTORE PROCEDURES

I — INTRODUCTION

This manual documents the procedures required to save, clear and restore data stored in EEPROM (Electrically Erasable Programmable Read-Only Memory), the non-vol­atile memory (referred to as NV RAM) of Kepco’s EL Series Electronic Loads.
The EL Series uses two (2) types of memory to store cus­tomer data within the processor chip. The processor chip contains a volatile Random Access Memory (RAM) and a Flash (EEPROM) Memory. The Flash memory is divided into regions, Program Memory, and Variable and Settings memory. Power cycling the EL will erase all data stored in RAM memory during the power up self test. However, Flash memory must be erased or reset to the original manufactur­ing state by using the procedure in section IV of this docu­ment.
The EL Series uses EEPROM to store the unit’s identity, calibration constants, front panel settings and certain other operating parameters. Though the command sequences for storing this information are not published, clearing NV RAM guards against the possibility that this data can be changed in the field by a sophisticated or properly trained individual.
The EL Calibration Save and Restore program must first be downloaded and installed (see Section II). Each time the
NV RAM
(EEPROM)
SAVE, CLEAR
AND RESTORE
NV RAM is to be cleared, the calibration data must first be saved (see Section III), then the NV RAM can be cleared (see Section IV).
Once the Clear function is executed, the unit will no longer function and will not accept any front panel inputs until the Restore function is executed (see Section V). The Restore function restores the previously saved calibration settings. Upon request from Kepco, the original factory calibration constants can be accessed and the unit can be restored to the original factory calibration condition.
Kepco can provide a Certificate of Volatility which insures that the factory calibration constants of the specific unit are saved by Kepco for a minimum of 10 years and can be for­warded to the customer after contacting Kepco application support. If the Certificate Of Volatility is not ordered, the factory calibration constants are only available for two years after the unit was manuafactured.
The NV RAM Clear feature will not function until it is enabled. It is enabled at time of order by requesting a Cer­tificate of Volatility. If not requested at time of order, contact Kepco application support to order a Certiifcate of Volatility and receive the instructions to enable the feature.

II — INSTALLING CALIBRATION DATA SAVE AND RESTORE PROGRAM

NOTE: The EL Calibration Save and Restore Program
only needs to be downloaded and installed once.
1. Download the EL Calibration Save and Restore pro­gram from www.kepcopower.com/utility/ to a Windows­compatible computer, running Windows XP or later.
2. Extract zipped files into a temporary directory.
KEPCO, INC. " 131-38 SANFORD AVENUE " FLUSHING, NY. 11355 U.S.A. " TEL (718) 461-7000 " FAX (718) 767-1102
http://www.kepcopower.com " email: hq@kepcopower.com
©2013, KEPCO, INC 1 Data subject to change without notice 243-1353
3. Click on setup to install the program by following the on­screen prompts. An icon will be saved on the desktop, and the program will stored in c:\Program files\kepco\el_restore.
4. When data is saved it will be stored where the program was installed. When restoring data, the EL must be con­nected to the same computer used to save the data.
Page 2

III — SAVING CALIBRATION DATA

Once the Calibration Data Save and Restore Program has been installed (see Section II), the calibration data can be saved.
Performing this procedure generates a copy of the current values of the non-volatilie memory. This copy, if used to restore the unit, does not meet the requirements of the Certificate Of Volatility of the US government. By provid­ing the model and serial number of the unit, the factory non-volatile ram contents file can be obtained from Kepco Application Engineering support for the first two years after manuafacture. If a Certificate Of Volatility is ordered for the unit, the file will be available for 10 years mininum.
of the Kepco EL load with the message:
saving caldata, please wait ...
If there is a connection problem, an Error window (Fig­ure 3) will appear, which will display: Kepco not found. Click OK to continue.
1. Launch the Program from the Windows Start menu, by selecting All programs - Kepco - calsave.
2. When the Kepco EL Cal Save Restore window (Fig­ure 1) appears, select the serial port to which the EL load is connected.

FIGURE 1. KEPCO EL CAL SAVE RESTORE WINDOW

3. Select SAVE SETTINGS to save the EL calibration data.
4. If the connection is made, the Standard Input/Output window opens (Figure 2), showing the Identifier string
FIGURE 2. STANDARD INPUT/OUTPUT WINDOW,
SAVING CALDATA

FIGURE 3. ERROR WINDOW

5. When the data has been saved, the Cal File Saved window opens (see Figure 4), entitled Kepco EL xxxxxxxx where xxxxxxxx is the serial number of the EL Load, and the message Cal File Saved is dis­played. Click OK to end the program.

FIGURE 4. CAL FILE SAVED WINDOW

KEPCO, INC. " 131-38 SANFORD AVENUE " FLUSHING, NY. 11355 U.S.A. " TEL (718) 461-7000 " FAX (718) 767-1102
http://www.kepcopower.com " email: hq@kepcopower.com
2 243-1353 010914
Page 3

IV — CLEARING NV RAM

CAUTION:
Clearing NV RAM will render the unit inoperative
until the restore procedure is executed. Calibration
data must be saved prior to clearing NV RAM!
CAUTION:
Step 7 offers a final opportunity
to abort the NV RAM clear.
To execute NV RAM clear proceed to step 8.
The Clear function is executed from the EL front panel. Safeguards are in place to avoid accidental or inadvertent clearing of NV RAM. Executing the Clear function initiates a series of erasures to NV RAM, including erasing unused program space, followed by rewriting the program space with multiple data patterns,
If execution of the Clear function is interrupted by a power cycle, the clear operation is completed without notifying the user.
1. Power unit off.
2. While depressing and holding the CONTROL knob, power up the unit.
3. Once unit is powered up: VOLTS display shows (NV). AMPS display shows (CLER).
4. Release the CONTROL knob.
5. Press and hold LOAD pushbutton for approximately one second, then release it.
VOLTS display shows (ARE).
AMPS display shows (YOU).
KW display shows (SUR).
7. To exit without clearing, rotate CONTROL knob. The unit reverts to the normal power-up sequence without clearing NV RAM.
CAUTION:
Executing step 8 will clear NV RAM
and render the unit inoperative until
the Restore procedure is executed.
8. To clear NV RAM, press LOAD pushbutton, hold for approximately one second, then release LOAD push­button.:
VOLTS display shows (CLRG).
AMPS display shows (NVRM).
Unit performs NV RAM clear, including writing multiple patterns to prevent recovery of data, then restarts auto­matically.
During power on tests, the erased state of NV memory causes a rewrite of original firmware defaults, including the backup copy.
The Restore function (see Section V) must be executed before the unit can be used.
6. Proceed to step 7 to cancel the clear of NV RAM. Proceed to step 8 to continue.
KEPCO, INC. " 131-38 SANFORD AVENUE " FLUSHING, NY. 11355 U.S.A. " TEL (718) 461-7000 " FAX (718) 767-1102
010914 243-1353 3
http://www.kepcopower.com " email: hq@kepcopower.com
Page 4

V — RESTORING NV RAM CALIBRATION DATA

The following procedure restores calibration data that has been previously saved, but can also be used to restore factory calibration. To restore factory calibration, a calibra­tion data file specific to that particular serial no. must first be obtained from Kepco, and saved at c:\Program files\kepco\el_restore. Factory calibration data is available for two (2) years from date of purchase (10 years if a Cer­tificate of Volatility has been ordered).
1. To restore NV RAM calibration data, launch the Pro­gram from the Windows Start menu, by selecting All programs - Kepco - calsave.
When the Kepco EL Cal Save Restore window (Figure 5)
appears, select the serial port to which the EL load is connected.
NOTE: The EL must be connected to the same computer
used to save the data.
3. The Standard Input/Output window opens (see Fig- ure 7) showing that the restore process is started, list­ing the new settings.
4. When done, the Cal File Restored window opens (see Figure 8), entitled Kepco EL xxxxxxxx where xxxxxxxx is the serial number of the EL Load, and the message Cal File Restored is displayed. Click OK to end the program.

FIGURE 5. KEPCO EL CAL SAVE RESTORE WINDOW

2. Click the RESTORE SETTINGS button (see Figure 5). The Restore Cal Data window opens (Figure 6. Enter the serial number of the unit to be restored and click
OK.

FIGURE 6. RESTORE CAL DATA WINDOW

FIGURE 7. STANDARD INPUT/OUTPUT WINDOW,
RESTORING CALDATA

FIGURE 8. CAL FILE RESTORED WINDOW

KEPCO, INC. " 131-38 SANFORD AVENUE " FLUSHING, NY. 11355 U.S.A. " TEL (718) 461-7000 " FAX (718) 767-1102
http://www.kepcopower.com " email: hq@kepcopower.com
4 243-1353 010914
Loading...