Power, input and output (I/O) wiring must be in accordance with Class I, Division 2 wiring methods Article
501-4 (b) or the National Electrical Code, NFPA 70 for installations in the U.S., or as specied in Section 18-1J2 of the
Canadian Electrical Code for installations within Canada and in accordance with the authority having juristiction.
1. This equipment is suitable for use in Class I, Division 2, Groups A, B, C, and D or
Non-Hazardous Locations Only.
2. WARNING - EXPLOSION HAZARD - Substituition of components may impair suit-
ability for Class I, Division 2.
3. WARNING - EXPLOSION HAZARD - Do not disconnect equipment unless power
has been switched off or the area is known to be Non-Hazardous.
WARNING!
This instrument contains electronic components that are susceptible to damage by static electricity. Proper
handling* procedures must be observed during the removal, installation, or handling of internal circuit boards or devices.
*Handling Procedure
1. Power to unit must be removed.
2. Personnel must be grounded, via wrist strap or other safe, suitable means, before any printed circuit board or other
internal device is installed, removed or adjusted.
3. Printed circuit boards must be transported in a conductive bag or other conductive container. Boards must not be
removed from protective enclosure until the immediate time of installation. Removed boards must be placed immediately
in protective container for transport, storage, or return to factory.
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic designs contain components that utilize metal oxide technology (NMOS, CMOS, etc.). Experience has proven
that even small amounts of static electricity can damage or destroy these devices. Damaged components, even though
they appear to function properly, may exhibit early failure.
SAFETY INSTRUCTIONS
The following instructions must be observed.
• This instrument was designed and is checked in accordance with regulations in force EN 60950 (“Safety of
information technology equipment, including electrical business equipment”).
A hazardous situation may occur if this instrument is not used for its intended purpose or is used incorrectly.
Please note operating instructions provided in this manual.
• The instrument must be installed, operated and maintained by personnel who have been properly trained.
Personnel must read and understand this manual prior to installation and operation of the instrument.
• This instrument is internally fused. Replace the internal fuse with the following specied type and rating only:
Input Power Recommended Fuse
115 VAC 160 mA slow blow fuse
230 VAC 80 mA slow blow fuse
12-24 VDC 800 mA slow blow fuse
Disconnect power supply before replacing fuse!
• The manufacturer assumes no liability for damage caused by incorrect use of the instrument or for modications
or changes made to the instrument.
Symbols Used On Unit
Number Symbol Publication Description
1 IEC 417, No. 5031 Direct current2 IEC 417, No. 5172 Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION
(equivalent to Class II of IEC 536–see annex H)
3
ISO 3864, No. B.3.1 Caution (refer to accompanying documents)
Technical Improvements
• The manufacturer reserves the right to modify technical data without prior notice.
SUPERtrol-I Flow Computer
CONTENTS
1. DESCRIPTION
1.1 Unit Description ...............................................................................................1
1.2 Unit Features ...................................................................................................1
The SUPERtrol-I Flow Computer satises the instrument requirements for
a variety of owmeter types in liquid applications. Multiple ow equations
and instrument functions are available in a single unit with many advanced
features.
The alphanumeric display shows measured and calculated parameters
in easy to understand format. Single key direct access to measurements
and display scrolling is supported.
The versatility of the SUPERtrol-I permits a wide range of measurements
within the instrument package. The various hardware inputs and outputs
can be “soft” assigned to meet a variety of common application needs.
The user “soft selects” the usage of each input/output while conguring
the instrument. Consider the following illustrative examples.
SUPERtrol-I Flow Computer
Unit Features
The isolated analog output can be chosen to follow volume ow, corrected
volume ow, mass ow, temperature, or density by means of a menu
selection. Most hardware features are assignable by this method.
The user can assign the standard RS-232 Serial Port for data logging,
transaction printing, or for connection to a modem for remote meter
reading.
1.2 Unit Features:
The SUPERtrol-I Flow Computer offers the following features:
• EZ-Preset or Standard Preset
• Custody Transfer with Audit Trail
• Multiple Instrument Functions
• Menu Selectable Hardware & Software Features
• Two Line LCD or VFD Display
• Isolated Outputs Standard
• Versatile RS-232 Port Standard
• DIN Enclosure with Two Piece Connector
• Optional Networking Cards
• Advanced Batching Features
1
1.3 Specications:
SUPERtrol-I Flow Computer
Specications:
Environmental
Indoor Use
Altitude up to 2000m
Operating Temperature: 0°C to +50°C
(-20°C to 55°C optional)
Storage Temperature: -40°C to +85 C
Maximum Relative Humidity : 80% for
temperatures
up to 31°C decreasing linearly to 50% RH at
40°C
Mains supply voltage uctuations not to exceed
±10%
of the nominal voltage
Tr an s i e n t o v e r vo l t ag e a c c or d i n g to
INSTALLATION
CATEGORY II (see UL 3101-1 Annex J)
POLLUTION DEGREE 2 in accordance with
IEC 664 (see 3.7.3)
Materials: UL, CSA, VDE approved
UL= UL Class 1 Division 2
Approvals: CE Approved Light Industrial,
UL File #: E192404 / CUL
Display
Type: 2 lines of 20 characters
Types: Backlit LCD or VFD ordering options
Character Size: 0.3" nominal
User selectable label descriptors and units of
measure
Keypad
Keypad Type: Membrane Keypad
Keypad Rating: Sealed to Nema 4
Number of keys: 16
Enclosure
Size: See Dimensions
Seal: NEMA4X
Materials: Aluminum
Power Input
The factory equipped power option is internally
fused. An internal line to line lter capacitor is
provided for added transient suppression.
Order Option 1: 110VAC: 85 to 127 Vrms, 50/60Hz
Order Option 2: 220VAC: 170 to 276 Vrms, 50/60Hz
Order Option 3: 12VDC: 10.5 to 14 VDC
Order Option 4: 24VDC: 18 to 24 VDC
Flow Inputs:
Analog Input:
Ranges
Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA
Basic Measurement Resolution: 16 bit
Update Rate: 5 updates/sec minimum
Automatic Fault detection: Signal over/underrange,
Current Loop Broken
Calibration: Self Calibration and Auto-zero
Continuously
Extended calibration: Learns Zero and Full
Scale
of each range using special test mode.
Fault Protection:
Fast Transient: 500 V Protection
(Capacitive Clamp)
Reverse Polarity: No ill effects
Over-Voltage Limit: 50 VDC Over voltage
protection
Over-Current Protection: Internally current
limited Protected to 24 VDC.
Pulse Inputs:
Number of Flow Inputs: one
Congurations supported: single input with or
without quadrature (menu selectable)
Input Impedance: 10 KΩ nominal
Pullup Resistance: 10 KΩ to 5 VDC (menu
selectable)
Pull Down Resistance: 10 KΩ to common
Trigger Level: (menu selectable)
High Level Input
Logic On: 3 to 30 VDC
Logic Off: 0 to 1 VDC
Low Level Input (mag pickup)
Selectable sensitivity: 10 mV & 100 mV
Minimum Count Speed: User selectable
Maximum Count Speed: Selectable: 0 to 20kHz
Overvoltage Protection: 50 VDC
Fast Transient: Protected to 500 VDC
(Capacitive Clamp)
Compensation Input
The compensation input is menu selectable for
temperature, density or not used.
Operation: Ratiometric
Accuracy: 0.01% FS
Thermal Drift: Less than 100 ppm/C
Basic Measurement Resolution: 16 bit
Update Rate: 1 update/sec minimum
Automatic Fault detection:
Signal Over-range/under-range
Current Loop Broken
RTD short
RTD open
Fault mode to user dened default settings
Transient Protection: 500 V (Capacitive Clamp)
Reverse Polarity: No ill effects
Over-Voltage Limit (Voltage Input): 50 VDC
Available Input Ranges
Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA
Resistance: 100 Ohms DIN RTD
100 Ohm DIN RTD (DIN 42-760, BS 1904):
Three Wire Lead Compensation
Internal RTD linearization learns ice point
resistance
1 mA Excitation current with reverse polarity
protection
Temperature Resolution: 0.01 C
Control Inputs
Switch Inputs are menu selectable for Start, Stop,
Reset, Lock, Inhibit, Alarm Acknowledge, Print or
Not Used.
Control Input Specications
Input Scan Rate: 10 scans per second
Logic 1: 4 - 30 VDC
Logic 0: 0 - 0.8 VDC
Transient Suppression: 500 V fast transient
(Capacitive Clamp)
Input Impedance: 100 KΩ
Control Activation: Positive Edge or Pos. Level
based on product denition
Excitation Voltage
110/220 VAC Powered Units
Menu Selectable: 5, 12 or 24 VDC @ 100mA
24 VDC Powered Units
Menu Selectable: 5 or 12 VDC @ 100mA
12 VDC Powered Units
5 VDC @ 100mA
2
SUPERtrol-I Flow Computer
Relay Outputs
The relay outputs are menu assignable to
(Individually for each relay) Low Rate Alarm, Hi Rate
Alarm, Prewarn Alarm, Preset Alarm or General
purpose warning (security).
Number of relays: 2 (4 optional)
Contact Style: Form C contacts
Contact Ratings: 2400 VAC Max., 3 Amps
Max.
Fast Transient Threshold: 1000 V
Serial Communication
The serial port can be used for printing, datalogging,
modem connection and communication with a
computer.
RS-232:
Device ID: 01-99
Baud Rates: 300, 600, 1200, 2400, 4800, 9600,
19200
Parity: None, Odd, Even
Handshaking: None, Software, Hardware
Print Setup: Configurable print list and
formatting
RS-485:
Device ID: 01-247
Baud Rates: 1200, 2400, 4800, 9600, 19200
Parity: None, Odd, Even
Protocol: Modbus RTU (Half Duplex)
Analog Output
The analog output is menu assignable to correspond
to the Uncompensated Volume Rate, Corrected
Volume Rate, Mass Rate, Temperature, Density,
Volume Total, Corrected Volume Total or Mass Total.
Type: Isolated Current Sourcing (AC power
options)
Isolated I/P/C: 500 V
Available Ranges: 4-20 mA, 0-20 mA
Resolution: 12 bit
Accuracy: 0.05% FS at 20 Degrees C
Update Rate: 1 update/sec minimum
Temperature Drift: Less than 200 ppm/C
Maximum Load: 1000 ohms (at nominal line
voltage)
Compliance Effect: Less than .05% Span
60 Hz rejection: 40 dB minimum
EMI: No effect at 3 V/M
Calibration: Operator assisted Learn Mode
Averaging: User entry of DSP Averaging constant
to cause an smooth control action.
Isolated Pulse output
The isolated pulse output is menu assignable to
Uncompensated Volume Total, Compensated
Volume Total or Mass Total.
Isolation I/O/P: 500 V
Pulse Output Form: Open Collector
Maximum On Current: 125 mA
Maximum Off Voltage: 30 VDC
Saturation Voltage: 1.0 VDC
Maximum Off Current: 0.1 mA
Pulse Duration: User selectable
Pulse output buffer: 8 bit
Fault Protection
Reverse polarity: Shunt Diode
Transient Protection: 500 VDC
(Capacitive Clamp)
Operating Mode
The Flow Computer can be thought of as making a
series of measurements of ow, temperature/density
sensors and then performing calculations to arrive
at a result(s) which is then updated periodically on
the display. The analog output, the pulse output,
and the alarm relays are also updated. The cycle
then repeats itself.
Step 1: Update the measurements of input
signalsRaw Input Measurements are made at each input
using equations based on input signal type selected.
The system notes the “out of range” input signal as
an alarm condition.
Step 2: Compute the Flowing Fluid Parameters-
The temperature, viscosity, and density equations
are computed as needed based on the ow equation
and input usage selected by the user.
Step 3 : Compute the Volumetric Flow-
Uncompensated ow is the term given to the ow in
volume units. The value is computed based on the
owmeter input type selected and augmented by
any performance enhancing linearization that has
been specied by the user.
Step 4: Compute the Corrected Volume Flow at
Reference Conditions-
In the case of a corrected liquid volume ow
calculation, the corrected volume ow is computed
as required by the selected compensation
equation.
Step 5 : Compute the Mass Flow-
All required information is now available to compute
the mass ow rate as volume ow times density.
Step 6: Check Flow Alarms-
The ow alarm functions have been assigned to
one of the above ow rates during the setup of the
instrument. A comparison is now made by comparing
the current ow rates against the specied hi and
low limits.
Step 7: Compute the Analog Output-
This designated ow rate value is now used to
compute the analog output.
Step 8: Compute the Flow Totals by Summation-
A ow total increment is computed for each ow
rate. This increment is computed by multiplying
the respective ow rate by a time base scaler and
then summing. The totalizer format also includes
provisions for total rollover.
Step 9: Total Preset Comparisons-
The total associated with a preset function is then
compared against the corresponding preset value
and any required control actions taken.
Step 10: Pulse Output Service-
The pulse output is next updated by scaling the total
increment which has just been determined by the
pulse output scaler and summing it to any residual
pulse output amount.
Step 11: Update Display and Printer Output-
The instrument nally runs a task to update the
various table entries associated with the front panel
display and serial outputs.
3
SUPERtrol-I Flow Computer
Setup Mode
The setup mode is password protected by means
of a numeric lock out code established by the user.
In addition, a secret, manufacturers numeric unlock
entry sequence is available.
The system also provides a minimum implementation
of an “audit trail” which tracks signicant setup
changes to the unit. This feature is increasingly
being found of benet to users or simply required
by Weights and Measurement Ofcials in systems
used in commerce, trade, or “custody transfer”
applications.
A software program is available which runs on a
PC using a RS-232 Serial for connection to the
Flow Computer. Illustrative examples may be down
loaded in this manner.
The setup mode has numerous subgrouping of
parameters needed for ow calculations. There is
a well conceived hierarchy to the setup parameter
list. Selections made at the beginning of the setup
affect offerings further down in the lists.
In the setup mode, the ow computer activates the
correct setup variables based on the instrument
conguration, the ow equation, and the hardware
selections made for the compensation transmitter
type, the flow transmitter type, and meter
enhancements (linearization) options selected. All
required setup parameters are enabled. All setup
parameters not required are suppressed.
A help line prompt is provided for each entry. In
addition a help message is available which may be
accessed by depressing the “HELP” key.
In the setup mode selections, several parameters
are required to be input by the operator since these
parameters are blank when the unit is received. The
user will be prompted for these necessary values
for his application.
Also note that in the setup mode are parameter
selections which have preassigned industry
standard values. The unit will assume these values
unless they are modied by the user.
Most of the process input variables have available
a “default” or emergency value which must be
entered. These are the values that the unit assumes
when a malfunction is determined to have occurred
on the corresponding input.
It is possible to enter in a nominal constant value
for temperature or density, or analog ow inputs
by placing the desired nominal value into both the
lo and hi values. This is also a convenience when
performing bench top tests without simulators.
Maintenance Mode:
The Maintenance Mode of the Flow Computer is the
Test and Calibration Mode for the device. This mode
provides a number of specialized utilities required
for factory calibration, instrument checkout on startup, and periodic calibration documentation.
A password is required to gain access to this
specialized mode of operation. Normally quality,
calibration, and maintenance personnel will nd
this mode of operation very useful. It is also useful
for factory testing.
Many of these tests may be used during start-up
of a new system. Inputs signals may be read,
and output signals may be exercised to verify the
electrical interconnects before the entire system is
put on line.
The following action items may be performed in the
Maintenance Mode:
Print Calibration/Maintenance Report
Examine Audit Trail
Perform Keypad Checkout
Perform Display Checkout
Perform Pulse Input Checkout
Perform Pulse Output Checkout
Perform Control Input Checkout
Perform Relay Output Checkout
Perform Analog Input Checkout
Perform Analog Output Checkout
Calibrate Analog Inputs using the Learn Feature
Calibrate Analog Output using the Learn Feature
Battery Check
Note that a calibration of the analog input/output will
advance the audit trail counters since it effects the
accuracy of the system.
RS-232 Serial Port
The Flow Computer has a general purpose RS-232
Port which may be used for any one of the following
purposes:
Transaction Printing
Data Logging Internal Datalog Dumps
Remote Metering by Modem (optional)
Computer Communication Link
Conguration by Computer
Print System Setup
Print Calibration/Malfunction History
Remote Control
Instrument Setup by PC’s over Serial Port
A Diskette program is provided with the Flow
Computer that enables the user to rapidly congure
the Flow Computer using an Personnel Computer.
Included on the diskette are common instrument
applications which may be used as a starting point
for your application. This permits the user to have
an excellent starting point and helps speed the user
through the instrument setup.
4
SUPERtrol-I Flow Computer
Operation of Serial Communication Port with
Printers
Flow Computer’s RS-232 channel supports a
number of operating modes. One of these modes
is intended to support operation with a printer in
metering applications requiring transaction printing,
data logging and/or printing of calibration and
maintenance reports.
For transaction printing, the user denes the items
to be included in the printed document. The user
can also select what initiates the transaction print
generated as part of the setup of the instrument.
The transaction document may be initiated via
a front panel key depression, a remote contact
closure, or upon completion of a batch.
In data logging, the user denes the items to be
included in each data log as a print list. The user
can also select when or how often he wishes a data
log to be made. This is done during the setup of
the instrument as either a time of day or as a time
interval between logging.
The system setup and maintenance report lists all
the instrument setup parameters and usage for
the current instrument conguration. In addition,
the Audit trail information is presented along with
a status report listing any observed malfunctions
which have not been corrected.
Operation of Serial Port with Modems (optional)
The Flow Computer RS-232 channel supports a
number of operating modes. One of these modes
is intended to support operation with a modem in
remote metering applications.
An external modem is intentionally being used
with the Flow Computer. This permits use with
the variety of modem standards worldwide while
avoiding the specialized approvals required for
equipment that is deemed to fall under the category
of telecommunication equipment.
In the modem mode, the Flow Computer is
assumed to be operating in a remote metering role.
The Flow Computer will support key items in the
Hayes Compatible “AT” Command Set. In this role,
the Flow Computer will have the following special
abilities:
0. Monitor the modem status as a task of the
system
1. Instruct the modem to answer an incoming call
ATA
2. Respond to the calling modem at a compatible
baud rate and protocol
3. Perform error checking in conjunction with the
modem
4. Monitor the status of the carrier
5. Terminate the telephone connection in event the
connection is lost.
The user initiates the printing of this report at a
designated point in the menu by pressing the print
key on the front panel.
In addition, the Flow Computer will be capable of
initiating a call to a designed telephone number in
the event of a metering malfunction.
5
6.15
2. Installation
SUPERtrol-I Flow Computer
General Mounting
Hints
Mounting Procedure
2.1 General Mounting Hints:
The SUPERtrol-I Flow Computer should be located in an area with a clean,
dry atmosphere which is relatively free of shock and vibration. The unit is
installed in a 5.43” (138mm) wide by 2.68” (68mm) high panel cutout. (see
Mounting Dimensions) To mount the Flow Computer, proceed as follows:
a. Prepare the panel opening.
b. Slide the unit through the panel cutout until the it touches the panel.
c. Install the screws (provided) in the mounting bracket and slip the bracket
over the rear of the case until it snaps in place.
d. Tighten the screws rmly to attach the bezel to the panel. 3 in. lb. of
torque must be applied and the bezel must be parallel to the panel.
Termination Connectors:
Minimum Wire Gauge: 22 AWG
Maximum Wire Gauge: 14 AWG
Voltage/current limits are limited by unit specications.
Permanently Connected Equipment:
UL 3101-1, Section 6.12.2.1 species that:
• A switch or circuit breaker shall be included in the building
installation;
• It shall be in close proximity to the equipment and within easy
reach of the OPERATOR;
• It shall be marked as the disconnecting device for the equipment.
Ensure that the switch or circuit breaker chosen is suitable for the
power requirements of the unit.
2.2 Mounting Diagrams:
Standard MountingBezel Kit Mounting
SUPERtrol-I
Mounting Bracket
Dimensions
3.43
(87)
RATE
TOTAL
1
START
GRAND6SCROLL7PRE 28DENS
STOP
5.67 (144)
GPM
147.43
GAL
267395.749
TEMP4PRE 13RATE2TOTAL
PRINT
5
TIME
9
0
–
6.18
Dotted Line Shows Optional Bezel Kit
CLEAR•MENU
HELP
2.83
(72)
ENTER
0.28 (7.2)
0.4 (10)
(156)
0.5
(13)
SUPERtrol-I
Bezel Adaptor
Gasket
Mounting Bracket
5.43
(138)
Panel
Cutout
2.68
(68)
Dimensions are in inches (mm)
6
3. Applications
STOP
START
PRINT
5
0
–
F3
CLEAR•MENU
ENTER
HELP
F1
4
PRE 1
3
RATE
2
TOTAL
1
GRAND6SCROLL7PRE 28F2
9
Flowmeter
SUPERtrol-I Flow Computer
Liquid Volume
3.1 Liquid Volume
Measurements:
A owmeter measures the actual volume in a liquid line. A temperature
sensor can also be installed to correct for liquid thermal expansion (see
3.2 Corrected Volume).
Calculations:
• For Flowmeters with Pulse Outputs, Volume ow is calculated using the
owmeter frequency output and the user entered K-Factor.
• For Flowmeters with Analog Transmitters, Volume ow is calculated using
the measured owmeter signal and the user entered scale settings.
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total
• Analog Output
Rate or Total
• Pulse Output
Total
• Relay Outputs
Rate or Total Alarms
Liquid Volume
Illustration
Applications:
The Flow Computer can monitor actual volume ow and total of any liquid.
Flow alarms are provided via relays and datalogging is available via analog
(4-20mA) and serial outputs.
Calculations
Pulse Input; Average K-Factor
input frequency * time scale factor
Volume Flow =
K-Factor
Analog Input; Linear
Volume Flow = % input * Full Scale Flow
7
SUPERtrol-I Flow Computer
STOP
START
PRINT
5
0
–
TIME
CLEAR•MENU
ENTER
HELP
TEMP
4
PRE 1
3
RATE
2
TOTAL
1
GRAND6SCROLL7PRE 28DENS
9
FlowmeterTemperature Transmitter
Corrected
Liquid Volume
3.2 Corrected Liquid Volume
Measurements:
A owmeter measures the actual volume in a liquid line. A temperature
sensor is installed to correct for liquid thermal expansion.
Calculations:
• Corrected Volume is calculated using the ow and temperature inputs
as well as the thermal expansion coefcient stored in the ow computer.
Use the "SET FLUID PROPERTIES" submenu to dene reference
temperature and density values for standard conditions.
Monitoring corrected volume ow and total of any liquid. Flow alarms are
provided via relays and datalogging is available via analog (4-20mA) and
serial outputs.
Actual volume is measured by the flow element (DP transmitter,
Flowmeter). Temperature is measured by the temperature transmitter. A
density transmitter can be used for direct density measurements.
Calculations:
• The density and mass ow are calculated using the reference density
and the thermal expansion coefcient of the liquid (see "SET FLUID
PROPERTIES" submenu)
Batching and monitoring ow and total of any liquid. Batching is
accomplished via relays and datalogging is available via analog (4-20mA)
and serial outputs.
Examples using MLT787S+ Barrier (MTL4755ac for RTD)
Flow Input
Temperature Input
(4-20mA Transmitter)
Flow Input
Hazardous AreaSafe Area
Temperature Input (4-20mA Transmitter)
Hazardous AreaSafe Area
Temperature Input
(RTD)
Temperature Input (RTD)
Hazardous AreaSafe Area
12
SUPERtrol-I Flow Computer
5. UNIT OPERATION
5.1 Front Panel Operation Concept for Run Mode
The Flow Computer is fully programmable through the front panel. Please review the following usage summary before attempting to use the instrument.
START
STOP
TOTAL
1
GRAND6SCROLL7PRE 2
RATE
2
PRE 1
3
8
TEMP
4
DENS
9
PRINT
0
5
TIME
CLEAR
HELP
–
•
MENU
ENTER
HELP
On-line help is provided to assist the operator in using this product. The help is available during RUN and SETUP
modes simply by pressing the HELP key. The HELP key is used to enter decimals when entering numeric values.
FUNCTION KEYS
In the RUN mode, several keys have a special, direct access feature, to display an item of interest (i.e. RATE,
TOTAL, PRE 1, etc.). Press the key to view your choice. Press the SCROLL key to return to scrolling display.
CLEARING TOTALIZER
To clear the total, you must press the TOTAL Function Key 3 times. You will be asked to verify this action. The
operator will be prompted to enter password if the unit is locked.
NOTE: In the Batcher Mode, simply press the CLEAR key to reset the total (the batcher must be stopped or nished
batching). It is not necessary to press the TOTAL Function Key rst.
CLEARING GRAND TOTAL
To clear the grand total, you must press the GRAND Function Key 3 times. You will be asked to verify this action.
The supervisor will be prompted to enter the supervisor password if the unit is locked.
PRESET KEYS USING EZ PRESET
In the RUN mode, PRE 1 & PRE 2 keys are used to view and/or change the preset setpoints. To view the Presets,
simply press the desired Preset key. Press Enter then press the Clear key for direct editing of the preset setpoints.
PRESET KEYS
In the RUN mode, PRE 1 & PRE 2 keys are used to view and/or change the preset setpoints. To view the Presets,
simply press the desired Preset key. Rapidly press the Preset keys 3 times, then press the Clear key for direct editing
of the preset setpoints. Press Enter or Start to complete the Preset entry.
SCROLL
Rapidly press the Scroll key three times to setup a display list.
Press the CLEAR key to remove old scroll list.
Press the function key for the item you wish to add
Use the ∆ ∇ keys to assign the line.
PRINT
The PRINT key is used to print on demand. When the PRINT key is pressed, a user dened list of data (TOTAL,
RATE, PRE 1, etc.) is sent to the RS-232 port. A timed message of "PRINTING" will be displayed to acknowledge the
print request.
SPECIAL BATCHING KEYS
The START and STOP keys are used only when batching to start and stop batches. The CLEAR key will clear the
total without rst pressing the TOTAL key (unit must be stopped). All other keys work the same in both Rate/Total
mode and Batch mode. The Start and Stop keys operation are set by the control input settings. The Start options are:
START or RESET/START. The Stop options are: STOP or STOP/RESET.
MENU KEY
The MENU key is used to enter the Setup and Test modes. Press the MENU key to enter the Setup and Test modes.
(See section 6 for Setup mode, section 8 for Test mode). The MENU key is used as "escape" in Setup and Test
Programming. Pressing the MENU key wile programming in the Sub-Menu groups will backup the display to that
Sub-Menu group heading. Pressing the MENU key while viewing the Sub-Menu groups will backup the display to the
Top Level Menu.
ACKNOWLEDGING ALARMS
Most alarm messages are self-clearing. Press the ENTER key to acknowledge and clear alarms.
NOTE: Some keys and functions are password protected. Enter the password to gain access. The passwords are
factory set as follows: Operator = 0 Supervisor = 2000
13
SUPERtrol-I Flow Computer
General
Operation
Rate/Total
Operation
Password Protection
(Rate/Total mode)
5.2 General Operation
The unit can display: Rate, Total, Grand Total, Temperature, Density, Presets and
Time of Day. The Temperature and/or Density can be displayed even if you are
using the Volumetric Flow Equation (a Temperature or Density sensor must be
installed). The unit can perform Mass or Corrected Volume equations using a
temperature or density sensor (these equations can be computed without Temp/
Dens sensors by using user dened default values). The unit can be programmed
to perform Ratemeter/Totalizer or Batching functions (see section 6.3, SELECT
INSTRUMENT Submenu).
5.3 Ratemeter/Totalizer Operation
The Ratemeter/Totalizer mode is used primarily to monitor owrate and
accumulated total. The relays can be used to trigger ow, total, temperature or
density alarms.
5.3.1 Password Protection for Rate/Total mode
After an Operator and/or Supervisor Password is entered in the setup mode (see
section 6.3, SETUP PASSWORD submenu), the unit will be locked. The unit will
prompt the user for the password when trying to perform the following functions:
Clear Total
Clear Grand Total
Enter Menu
Edit Preset 1 (PRE 1 Key)
Edit Preset 2 (PRE 2 Key)
Relay Operation
(Rate/Total mode)
Pulse Output
(Rate/Total mode)
The Supervisor password should be reserved for supervisors. The Supervisor
password will allow access to restricted areas of the Setup and Test menus.
5.3.2 Relay Operation in Rate/Total mode
Up to four relays are available (two standard) for alarm outputs. The relays can
be assigned to trip according to rate, total, temperature, density readings or
general system alarms. The relays can be programmed for low or high alarms.
Preset 1 (RLY1) and Preset 2 (RLY2) are easily accessible by pressing the PRE
1 or PRE 2 key on the front panel. Preset 3 and Preset 4 are accessible only
through the setup menu.
5.3.3 Pulse Output in Rate/Total mode
The isolated pulse output (open collector) is menu assignable to Volume Total,
Corrected Volume Total or Mass Total. The pulse output duration can be set for
10mS (50 Hz max) or 100mS (5 Hz max). A pulse output scale factor (pulse value)
can be set to scale the pulse output. The pulse output is ideal for connecting to
remote totalizers or other devices such as a PLC. See section 1.3 for electrical
specications.
Analog Output
(Rate/Total mode)
5.3.4 Analog Output in Rate/Total mode
The analog output is menu assignable to correspond to the Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature, Density, Volume Total,
Corrected Volume Total or Mass Total. The analog output is ideal for "trend"
tracking using strip chart recorders or other devices.
14
SUPERtrol-I Flow Computer
RS-232 Serial Port
(Rate/Total mode)
5.3.5 RS-232 Serial Port Operation in Rate/Total mode
The RS-232 serial port can be used for programming (using the Setup Disk)
or for communicating to printers and computers in the Operating Mode (Run
Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status such
as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Operation of RS-232 Serial Port with Printers:
Transaction Printing
For transaction printing, the user denes the items to be included in the
printed document (see section 6.3.20 SET DATA OUTPUT, Select_list). The
transaction document can be initiated by pressing the PRINT key or by a
remote contact closure.
Data Logging
In data logging, the user denes the items to be included in each data log (see
section 6.3.20 SET PRINTER OUTPUT, Select_list). The user can also select
when (time of day) or how often (print interval) the data log is to be made (see
section 6.3.19 SET PRINTER OUTPUT, Congure).
System Setup and Maintenance Report
The system setup and maintenance report lists all of the instrument setup
parameters and usage for the current instrument conguration. The audit trail
information and a status report is also printed. This report is initiated in the
Test menu (see section 8.2.3 PRINT SYSTEM SETUP).
RS-485 Serial Port
(Rate/Total mode)
5.3.6 RS-485 Serial Port (optional)
RS-485 Port Description:
The optional RS-485 card utilizes Modbus RTU protocol to access
a variety of process parameters and totalizers. The Relays can
be controlled via Modbus. In addition, action routines can be
executed. For further information, contact factory and request
RS-485 Protocol manual.
Operation of Serial Communication Port with PC
The ow computer's RS-485 channel supports a number of
Modbus RTU commands. Modbus RTU drivers are available from
third party sources for a variety of Man Machine Interface software
for IBM compatible PC's.
The user reads and writes information from/to the RS-485 using
the Modbus RTU commands. The Flow Computer then responds
to these information and command requests.
Process variables and totalizers are read in register pairs in oating
point format. Time and date are read as a series of integer register
values. Alarms are individually read as coils. Action routines are
initiated by writing to coils.
15
SUPERtrol-I Flow Computer
5.4 Batcher Operation
The Batcher mode is used primarily to control batches. The main difference
between the Batch mode and Rate/Total mode is the relay operation. The Batch
mode allows the operator to "START" the unit via the front panel or remote input.
Once started, the relays (RLY1 & RLY2) will energize and send power to a ow
control device (i.e. solenoid valve or pump). The ow sensor will send a signal
to the unit and total accumulation will begin. When the Prewarn value (PRE 2) is
reached, Relay 2 will drop out (this is ideal for ow slow down). When the Batch
amount (PRE 1) is reached, Relay 1 will drop out and the Batch is complete.
Several messages will be displayed during normal batch operation (i.e. Batch
Fill, Batch Stopped). The keypad is disabled for the duration of these timed
messages (approx. 2 sec).
Batcher Conguration
5.4.1 Batcher Conguration.
When the unit is programmed for batch mode, several batch operation choices
are available. These choices include: Up or Down Counting, Maximum Batch
Preset, Batch Overrun Compensation, Auto Batch Restart, Time Delay, Flow
Signal Timeout, Maximum Drain Time, Slow Start Quantity, Start or Reset/Start,
and Stop or Stop/Reset.
Standard Preset or EZ Preset
Use Standard Preset for applications in which the batch amount does not change
frequently.
Use EZ Preset in applications in which the batch amounts change frequently. The
EZ Preset mode was designed to enter presets with minimum key strokes.
Batch Count Mode
The Batch Count Mode allows the user to choose whether the unit will batch up
to a preset value or batch down from a preset value to zero.
Maximum Batch Preset
The Maximum Batch Preset allows the user to program the Maximum Batch value
allowed to be entered by the operator. If an operator should try to program a
batch higher then this value, the unit will not allow the value to be entered and will
prompt the user with an error message saying that the Maximum Batch Preset
has been exceeded.
Batch Overrun
The Batch Overrun is used for batch applications that have slow responding
valves and a consistent batching owrate. When the Batch Overrun is set, the
unit will compensate for batch overruns by computing an averaged overrun value
from the last four batches. This average is used to internally adjust the batch
setpoint to minimize overrun.
Flow Signal Timeout
The Flow Signal Timeout allows the user to enter a timeout of 0 to 99 seconds.
If a batch is “Filling” and zero ow persists for more than the user entered time
then the batch will be aborted. This prevents over ows due to faulty ow sensors
and/or wiring.
Maximum Drain Time
The unit declares that a batch is “done” when the ow rate equals “0”. A ow rate
may be present long after the Preset Relay de-energizes due to slow reacting
valves or leaky valves. The Maximum Drain Time allows the user to enter an
amount of time (0 to 99 seconds) to wait before declaring “Batch Done”. After
the Preset Batch quantity is reached, the unit will declare “Batch Done” when
the ow rate is “0” or the Maximum Drain Time has expired. The batch data will
then be available for printing and datalogging.
16
SUPERtrol-I Flow Computer
Slow Start Quantity
The Slow Start Quantity is a function that allows an amount to be entered for a
Slow Start up. This function requires two stage valve control. RLY 1 (slow ow)
will energize for Slow Start and RLY 2 (fast ow) will energize after the Slow
Start Quantity has been delivered. This helps reduce turbulence when lling an
empty container.
START, RESET/START and STOP, STOP/RESET
When conguring the control inputs, Control Input1 can be set for START or
RESET/START. When set for START, the unit will start batching when a signal
is applied to Control Input1 or the front panel Start key is pressed. A separate
Reset signal must be used to clear the previous batch total. When set for RESET/
START, the unit will automatically reset then start when a signal is applied to
Control Input1 or the front panel Start key is pressed (provided that the pervious
batch was completed). If a previous batch was stopped during a batch cycle, the
unit will Start from where it was stopped.
Control Input 2 can be set for STOP or STOP/RESET. When set for STOP, the
unit will stop batching when a signal is applied to Control Input 2 or the front
panel Stop key is pressed. A separate Reset signal must be used to clear the
batch total. When set for STOP/RESET, a running batch will stop when a signal
is applied to Control Input 2 or the front panel Stop key is pressed. If the unit is
Stopped or after a completed batch, the unit will reset when a signal is applied
to Control Input 2 or the front panel Stop key is pressed.
NOTE: Applying a voltage level to Control Input 2 will inhibit all Start inputs in
either mode.
Password Protection
(Batch mode)
Relay Operation
(Batch mode)
5.4.2 Password Protection for Batcher Mode
After an Operator and/or Supervisor Password is entered in the setup mode (see
section 6.3, SETUP PASSWORD submenu), the unit will be locked. The unit will
prompt the user for the password when trying to perform the following functions:
Clear Grand Total
Enter Menu
The Supervisor password should be reserved for supervisors. The Supervisor
password will allow access to restricted areas of the Setup and Test menus.
The passwords are factory set as follows:
Operator = 0
Supervisor = 2000
5.4.3 Relay Operation in Batcher mode
Up to four relays are available (two standard) for alarm outputs. Preset 1 (RLY1) is
reserved for batch amount, Preset 2 (RLY2) is reserved for prewarn. (see section
5.4 Batcher Operation for Relay 1 & Relay 2 functions)
Preset 1 (RLY1) and Preset 2 (RLY2) are easily accessible by pressing the PRE
1 or PRE 2 key on the front panel. Preset 3 and Preset 4 are accessible only
through the setup menu.
Relays 3 and 4 can be assigned to trip according to rate, total, temperature,
overrun or alarm. When Rate is selected the relays can be programmed for low
or high alarms.
Pulse Output
(Batch mode)
Analog Output
(Batch mode)
5.4.4 Pulse Output in Batcher mode
The isolated pulse output (open collector) is menu assignable to Volume Total,
Corrected Volume Total or Mass Total. The pulse output duration can be set for
10mS (50 Hz max) or 100mS (5 Hz max). A pulse output scale factor (pulse value)
can be set to scale the pulse output. The pulse output is ideal for connecting to
remote totalizers or other devices such as a PLC. See section 1.3 for electrical
specications.
5.4.5 Analog Output in Batcher mode
The analog output is menu assignable to correspond to the Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature, Density, Volume Total,
Corrected Volume Total or Mass Total. The analog output is ideal for "trend"
tracking using strip chart recorders or other devices.
17
SUPERtrol-I Flow Computer
RS-232 Serial Port
(Batch mode)
5.4.6 RS-232 Serial Port Operation in Batcher mode
The RS-232 serial port can be used for programming (using the Setup Disk)
or for communicating to printers and computers in the Operating Mode (Run
Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status such
as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Operation of RS-232 Serial Port with Printers:
Transaction Printing
For transaction printing, the user denes the items to be included in the
printed document (see section 6.3.20 SET DATA OUTPUT, Select_list). The
transaction document can be initiated by pressing the PRINT key, by a remote
contact closure or print at end of batch.
Data Logging
In data logging, the user denes the items to be included in each data log (see
section 6.3.20 SET PRINTER OUTPUT, Select_list). The user can also select
when (time of day) or how often (print interval) the data log is to be made (see
section 6.3.19 SET PRINTER OUTPUT, Congure).
System Setup and Maintenance Report
The system setup and maintenance report lists all of the instrument setup
parameters and usage for the current instrument conguration. The audit trail
information and a status report is also printed. This report is initiated in the
Test menu (see section 8.2.3 PRINT SYSTEM SETUP).
RS-485 Serial Port
(Batch mode)
5.4.7 RS-485 Serial Port (optional)
RS-485 Port Description:
The optional RS-485 card utilizes Modbus RTU protocol to access
a variety of process parameters and totalizers. Batches/Relays can
be controlled remotely via Modbus. In addition, action routines can
be executed. For further information, contact factory and request
RS-485 Protocol manual.
Operation of Serial Communication Port with PC
The ow computer's RS-485 channel supports a number of
Modbus RTU commands. Modbus RTU drivers are available from
third party sources for a variety of Man Machine Interface software
for IBM compatible PC's.
The user reads and writes information from/to the RS-485 using
the Modbus RTU commands. The Flow Computer then responds
to these information and command requests.
Process variables and totalizers are read in register pairs in oating
point format. Time and date are read as a series of integer register
values. Alarms are individually read as coils. Action routines are
initiated by writing to coils.
18
SUPERtrol-I Flow Computer
6. PROGRAMMING
6.1 Front Panel Operation Concept for Program Mode
The Flow Computer is fully programmable through the front panel. Please
review the following usage summary before attempting to use the instrument.
How To Make Mode
Changes
How To Navigate
Through Sub-Menu
Groups
START
STOP
TOTAL
1
GRAND6SCROLL7PRE 2
RATE
2
PRE 1
3
8
TEMP
4
DENS
9
PRINT
0
5
TIME
CLEAR
HELP
•
–
MENU
ENTER
Setup Mode:
MODE CHANGES
Pressing the MENU key will offer selections of RUN, SETUP, TEST.
RUN is the normal operating mode for the instrument. SETUP offers
various sub-menus used for instrument setup. TEST offers various
sub-menus for Test, Calibration and System Start-up.
Submenu GROUP NAVIGATION
Use the UP and DOWN arrow keys to navigate up and down through
the Sub-Menu groups when in the SETUP or TEST mode. Press the
ENTER key to enter a desired setup or test Sub-Menu group.
How To Select
Program Choices
How To Enter
Numeric Values
How To Enter
Text Characters
SELECTION OF ITEM
During setup, the unit will often offer multiple choices for a given topic.
The topic prompt appears on the top line of the display. The choices are
shown on the lower line of the display.
To select an item, press the key beneath the desired choice. The
selected choice will blink. Press the ENTER key to accept the
selected choice.
NUMERIC ENTRY
The keys labeled "0 - 9", "–", ".", CLEAR and ENTER are used to enter
numerical values. A leading 0 will assume that you intend to enter a
minus "–" sign. Press the CLEAR key to clear the existing value and to
enable editing.
TEXT CHARACTER ENTRY
Some setup items (i.e. Descriptors, Units Label) require the user to
enter text characters. Press CLEAR to enable editing. The UP and
DOWN arrow keys are used to scroll through the available character
sets for each individual character. Press the ENTER key to accept the
character and advance to the next character.
19
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.