Power, input and output (I/O) wiring must be in accordance with Class I, Division 2 wiring methods Article 501-4 (b) or
the National Electrical Code, NFPA 70 for installations in the U.S., or as specified in Section 18-1J2 of the Canadian
Electrical Code for installations within Canada and in accordance with the authority having juristiction.
WARNING!
1. This equipment is suitable for use in Class I, Division 2, Groups A, B, C, and D or
Non-Hazardous Locations Only.
2. WARNING - EXPLOSION HAZARD - Substituition of components may impair
suitability for Class I, Division 2.
3. WARNING - EXPLOSION HAZARD - Do not disconnect equipment unless power
has been switched off or the area is known to be Non-Hazardous.
!
This instrument contains electronic components that are susceptible to damage by static electricity. Proper handling*
procedures must be observed during the removal, installation, or handling of internal circuit boards or devices.
*Handling Procedure
1. Power to unit must be removed.
2. Personnel must be grounded, via wrist strap or other safe, suitable means, before any printed circuit board or other
internal device is installed, removed or adjusted.
3. Printed circuit boards must be transported in a conductive bag or other conductive container. Boards must not be
removed from protective enclosure until the immediate time of installation. Removed boards must be placed immediately
in protective container for transport, storage, or return to factory.
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic
designs contain components that utilize metal oxide technology (NMOS, CMOS, etc.). Experience has proven that even
small amounts of static electricity can damage or destroy these devices. Damaged components, even though they appear
!
The following instructions must be observed.
WARNING!
to function properly, may exhibit early failure.
SAFETY INSTRUCTIONS
• This instrument was designed and is checked in accordance with regulations in force EN 60950 (“Safety of
information technology equipment, including electrical business equipment”).
A hazardous situation may occur if this instrument is not used for its intended purpose or is used incorrectly.
Please note operating instructions provided in this manual.
• The instrument must be installed, operated and maintained by personnel who have been properly trained.
Personnel must read and understand this manual prior to installation and operation of the instrument.
• This instrument is internally fused. Replace the internal fuse with the following specified type and rating only:
Input PowerRecommended Fuse
115 VAC160 mA slow blow fuse
230 VAC80 mA slow blow fuse
12-24 VDC800 mA slow blow fuse
Disconnect power supply before replacing fuse!
• The manufacturer assumes no liability for damage caused by incorrect use of the instrument or for modifications
or changes made to the instrument.
Symbols Used On Unit
NumberSymbolPublicationDescription
1IEC 417, No. 5031Direct current
2IEC 417, No. 5172Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION
(equivalent to Class II of IEC 536–see annex H)
3
!
ISO 3864, No. B.3.1Caution (refer to accompanying documents)
Technical Improvements
• The manufacturer reserves the right to modify technical data without prior notice.
1. DESCRIPTION
1.1 Unit Description ...............................................................................................1
1.2 Unit Features ...................................................................................................1
13.3.2 Self Test Alarms ............................................................................68
72
ii
Unit Description1. Description
1.1 Unit Description:
The SUPERtrol-I MS-716 Flow Computer is a special version of the
SUPERtrol-1 Flow Computer which is supplied in a vehicle mount
enclosure. The SUPERtrol-I MS-716 satisfies the instrument requirements
for a variety of flowmeter types in liquid applications. Multiple flow equations
and instrument functions are available in a single unit with many advanced
features.
The alphanumeric display shows measured and calculated parameters
in easy to understand format. Single key direct access to measurements
and display scrolling is supported.
SUPERtrol-I MS-716 Flow Computer
Unit Features
The versatility of the SUPERtrol-I MS-716 permits a wide measure of
versatility within the instrument package. The various hardware inputs
and outputs can be “soft” assigned to meet a variety of common application
needs. The user “soft selects” the usage of each input/output while
configuring the instrument. Consider the following illustrative examples.
The user can assign the standard RS-232 Serial Port for data logging,
transaction printing, or for connection to a modem for remote meter
reading.
1.2 Unit Features:
The SUPERtrol-I MS-716 Flow Computer offers the following features:
• EZ-Preset or Standard Preset
• Custody Transfer with Audit Trail
• Multiple Instrument Functions
• Menu Selectable Hardware & Software Features
• Two Line LCD or VFD Display
• Isolated Outputs Standard
• Versatile RS-232 Port Standard
• DIN Enclosure with Two Piece Connector
• Optional Networking Cards
• Advanced Batching Features
1
1.3 Specifications:
SUPERtrol-I MS-716 Flow Computer
Specifications:
Environmental
Indoor Use
Altitude up to 2000m
Operating Temperature: 0°C to +50°C
(-20°C to 55°C optional)
Storage Temperature: -40°C to +85 C
Maximum Relative Humidity : 80% for temperatures
up to 31°C decreasing linearly to 50% RH at
40°C
Mains supply voltage fluctuations not to exceed
±10%
of the nominal voltage
Transient overvoltage according to INSTALLATION
CATEGORY II (see UL 3101-1 Annex J)
POLLUTION DEGREE 2 in accordance with
IEC 664 (see 3.7.3)
Materials: UL, CSA, VDE approved
UL= UL Class 1 Division 2
Approvals:CE Approved Light Industrial,
UL File #: E192404 / CUL
Display
Type: 2 lines of 20 characters
Types: Backlit LCD or VFD ordering options
Character Size: 0.3" nominal
User selectable label descriptors and units of
measure
Keypad
Keypad Type: Membrane Keypad
Keypad Rating: Sealed to Nema 4
Number of keys: 16
Enclosure
Size: See Dimensions
Seal: NEMA4X
Materials: Aluminum
Pulse Inputs:
Number of Flow Inputs: one
Configurations supported: single input with or
Pull Down Resistance: 10 KΩ to common
Trigger Level: (menu selectable)
High Level Input
Logic On:3 to 30 VDC
Logic Off:0 to 1 VDC
Low Level Input (mag pickup)
Selectable sensitivity: 10 mV & 100 mV
Minimum Count Speed: User selectable
Maximum Count Speed: Selectable: 0 to 20kHz
Overvoltage Protection: 50 VDC
Fast Transient: Protected to 500 VDC
(Capacitive Clamp)
Compensation Input
The compensation input is menu selectable for
temperature, density or not used.
Operation: Ratiometric
Accuracy: 0.01% FS
Thermal Drift: Less than 100 ppm/C
Basic Measurement Resolution: 16 bit
Update Rate: 1 update/sec minimum
Automatic Fault detection:
Signal Over-range/under-range
Current Loop Broken
RTD short
RTD open
Fault mode to user defined default settings
Transient Protection: 500 V (Capacitive Clamp)
Reverse Polarity: No ill effects
Over-Voltage Limit (Voltage Input): 50 VDC
Power Input
The factory equipped power option is internally
fused. An internal line to line filter capacitor is
provided for added transient suppression.
Order Option 1: 110VAC: 85 to 127 Vrms, 50/60
Hz
Order Option 2: 220VAC: 170 to 276 Vrms, 50/60
Hz
Order Option 3: 12VDC: 10.5 to 14 VDC
Order Option 4: 24VDC: 18 to 24 VDC
Flow Inputs:
Analog Input:
Ranges
Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA
Basic Measurement Resolution: 16 bit
Update Rate: 5 updates/sec minimum
Automatic Fault detection: Signal over/underrange,
Current Loop Broken
Calibration: Self Calibration and Auto-zero
Continuously
Extended calibration: Learns Zero and Full Scale
of each range using special test mode.
Fault Protection:
Fast Transient: 500 V Protection
(Capacitive Clamp)
Reverse Polarity: No ill effects
Over-Voltage Limit: 50 VDC Over voltage
protection
Over-Current Protection: Internally current
limited Protected to 24 VDC.
Available Input Ranges
Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA
Resistance: 100 Ohms DIN RTD
100 Ohm DIN RTD (DIN 42-760, BS 1904):
Three Wire Lead Compensation
Internal RTD linearization learns ice point
resistance
1 mA Excitation current with reverse polarity
protection
Temperature Resolution: 0.01 C
Control Inputs
Switch Inputs are menu selectable for Start, Stop,
Reset, Lock, Inhibit, Alarm Acknowledge, Print or
Not Used.
Control Input Specifications
Input Scan Rate: 10 scans per second
Logic 1: 4 - 30 VDC
Logic 0: 0 - 0.8 VDC
Transient Suppression: 500 V fast transient
(Capacitive Clamp)
Input Impedance: 100 KΩ
Control Activation: Positive Edge or Pos. Level
based on product definition
Excitation Voltage
110/220 VAC Powered Units
Menu Selectable: 5, 12 or 24 VDC @ 100mA
24 VDC Powered Units
Menu Selectable: 5 or 12 VDC @ 100mA
12 VDC Powered Units
5 VDC @ 100mA
2
SUPERtrol-I MS-716 Flow Computer
Relay Outputs
The relay outputs are menu assignable to
(Individually for each relay) Low Rate Alarm, Hi
Rate Alarm, Prewarn Alarm, Preset Alarm or
General purpose warning (security).
Number of relays: 2 (4 optional)
Contact Style: Form C contacts
Contact Ratings:2400 VAC Max., 3 Amps Max.
Fast Transient Threshold: 1000 V
Serial Communication
The serial port can be used for printing, datalogging,
modem connection and communication with a
computer.
RS-232:
The analog output is menu assignable to
correspond to the Uncompensated Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature,
Density, Volume Total, Corrected Volume Total or
Mass Total.
Type: Isolated Current Sourcing (AC power options)
Isolated I/P/C: 500 V
Available Ranges: 4-20 mA, 0-20 mA
Resolution: 12 bit
Accuracy: 0.05% FS at 20 Degrees C
Update Rate: 1 update/sec minimum
Temperature Drift: Less than 200 ppm/C
Maximum Load: 1000 ohms (at nominal line
voltage)
Compliance Effect: Less than .05% Span
60 Hz rejection: 40 dB minimum
EMI: No effect at 3 V/M
Calibration: Operator assisted Learn Mode
Averaging: User entry of DSP Averaging constant
to cause an smooth control action.
Isolated Pulse output
The isolated pulse output is menu assignable to
Uncompensated Volume Total, Compensated
Volume Total or Mass Total.
Isolation I/O/P: 500 V
Pulse Output Form: Open Collector
Maximum On Current: 125 mA
Maximum Off Voltage: 30 VDC
Saturation Voltage: 1.0 VDC
Maximum Off Current: 0.1 mA
Pulse Duration: User selectable
Pulse output buffer: 8 bit
Fault Protection
The Flow Computer can be thought of as making a
series of measurements of flow, temperature/
density sensors and then performing calculations
to arrive at a result(s) which is then updated
periodically on the display. The analog output, the
pulse output, and the alarm relays are also updated.
The cycle then repeats itself.
Step 1:Update the measurements of input signals-
Raw Input Measurements are made at each input
using equations based on input signal type selected.
The system notes the “out of range” input signal as
an alarm condition.
Step 2:Compute the Flowing Fluid Parameters-
The temperature, viscosity, and density equations
are computed as needed based on the flow
equation and input usage selected by the user.
Step 3 : Compute the Volumetric Flow-
Uncompensated flow is the term given to the flow
in volume units. The value is computed based on
the flowmeter input type selected and augmented
by any performance enhancing linearization that
has been specified by the user.
Step 4: Compute the Corrected Volume Flow at
Reference ConditionsIn the case of a corrected liquid volume flow
calculation, the corrected volume flow is computed
as required by the selected compensation equation.
Step 5 : Compute the Mass Flow-
All required information is now available to compute
the mass flow rate as volume flow times density.
Step 6: Check Flow Alarms-
The flow alarm functions have been assigned to
one of the above flow rates during the setup of the
instrument. A comparison is now made by
comparing the current flow rates against the
specified hi and low limits.
Step 7: Compute the Analog Output-
This designated flow rate value is now used to
compute the analog output.
Step 8: Compute the Flow Totals by Summation-
A flow total increment is computed for each flow
rate. This increment is computed by multiplying the
respective flow rate by a time base scaler and then
summing. The totalizer format also includes
provisions for total rollover.
Step 9: Total Preset Comparisons-
The total associated with a preset function is then
compared against the corresponding preset value
and any required control actions taken.
Step 10: Pulse Output Service-
The pulse output is next updated by scaling the
total increment which has just been determined by
the pulse output scaler and summing it to any
residual pulse output amount.
Step 11: Update Display and Printer Output-
The instrument finally runs a task to update the
various table entries associated with the front panel
display and serial outputs.
3
SUPERtrol-I MS-716 Flow Computer
Setup Mode
The setup mode is password protected by means of
a numeric lock out code established by the user. In
addition, a secret, manufacturers numeric unlock
entry sequence is available.
The system also provides a minimum implementation
of an “audit trail” which tracks significant setup
changes to the unit. This feature is increasingly being
found of benefit to users or simply required by
Weights and Measurement Officials in systems used
in commerce, trade, or “custody transfer” applications.
A software program is available which runs on a PC
using a RS-232 Serial for connection to the Flow
Computer. Illustrative examples may be down loaded
in this manner.
The setup mode has numerous subgrouping of
parameters needed for flow calculations. There is a
well conceived hierarchy to the setup parameter list.
Selections made at the beginning of the setup affect
offerings further down in the lists.
In the setup mode, the flow computer activates the
correct setup variables based on the instrument
configuration, the flow equation, and the hardware
selections made for the compensation transmitter
type, the flow transmitter type, and meter
enhancements (linearization) options selected. All
required setup parameters are enabled. All setup
parameters not required are suppressed.
A help line prompt is provided for each entry. In
addition a help message is available which may be
accessed by depressing the “HELP” key.
Maintenance Mode:
The Maintenance Mode of the SUPERtrol-I MS716 is the Test and Calibration Mode for the device.
This mode provides a number of specialized
utilities required for factory calibration, instrument
checkout on start-up, and periodic calibration
documentation.
A password is required to gain access to this
specialized mode of operation. Normally quality,
calibration, and maintenance personnel will find
this mode of operation very useful. It is also useful
for factory testing.
Many of these tests may be used during start-up
of a new system. Inputs signals may be read, and
output signals may be exercised to verify the
electrical interconnects before the entire system
is put on line.
The following action items may be performed in
the Maintenance Mode:
Print Calibration/Maintenance Report
Examine Audit Trail
Perform Keypad Checkout
Perform Display Checkout
Perform Pulse Input Checkout
Perform Pulse Output Checkout
Perform Control Input Checkout
Perform Relay Output Checkout
Perform Analog Input Checkout
Perform Analog Output Checkout
Calibrate Analog Inputs using the Learn Feature
Calibrate Analog Output using the Learn Feature
Battery Check
In the setup mode selections, several parameters
are required to be input by the operator since these
parameters are blank when the unit is received. The
user will be prompted for these necessary values for
his application.
Also note that in the setup mode are parameter
selections which have preassigned industry standard
values. The unit will assume these values unless
they are modified by the user.
Most of the process input variables have available a
“default” or emergency value which must be entered.
These are the values that the unit assumes when a
malfunction is determined to have occurred on the
corresponding input.
It is possible to enter in a nominal constant value for
temperature or density, or analog flow inputs by
placing the desired nominal value into both the lo
and hi values. This is also a convenience when
performing bench top tests without simulators.
Note that a calibration of the analog input/output
will advance the audit trail counters since it effects
the accuracy of the system.
RS-232 Serial Port
The SUPERtrol-I MS-716 has a general purpose
RS-232 Port which may be used for any one of
the following purposes:
Transaction Printing
Data Logging Internal Datalog Dumps
Remote Metering by Modem (optional)
Computer Communication Link
Configuration by Computer
Print System Setup
Print Calibration/Malfunction History
Remote Control
Instrument Setup by PC’s over Serial Port
A Diskette program is provided with the SUPERtrolI MS-716 that enables the user to rapidly configure
the SUPERtrol-I MS-716 using an Personnel
Computer. Included on the diskette are common
instrument applications which may be used as a
starting point for your application. This permits the
user to have an excellent starting point and helps
speed the user through the instrument setup.
4
SUPERtrol-I MS-716 Flow Computer
Operation of Serial Communication Port with
Printers
SUPERtrol-I MS-716’s RS-232 channel supports
a number of operating modes. One of these
modes is intended to support operation with a
printer in metering applications requiring
transaction printing, data logging and/or printing
of calibration and maintenance reports.
For transaction printing, the user defines the items
to be included in the printed document. The user
can also select what initiates the transaction print
generated as part of the setup of the instrument.
The transaction document may be initiated via a
front panel key depression, a remote contact
closure, or upon completion of a batch.
In data logging, the user defines the items to be
included in each data log as a print list. The user
can also select when or how often he wishes a
data log to be made. This is done during the
setup of the instrument as either a time of day or
as a time interval between logging.
The system setup and maintenance report lists
all the instrument setup parameters and usage
for the current instrument configuration. In
addition, the Audit trail information is presented
along with a status report listing any observed
malfunctions which have not been corrected.
The user initiates the printing of this report at a
designated point in the menu by pressing the
print key on the front panel.
Operation of Serial Port with Modems (optional)
The SUPERtrol-I MS-716 RS-232 channel supports
a number of operating modes. One of these modes
is intended to support operation with a modem in
remote metering applications.
An external modem is intentionally being used with
the SUPERtrol-I MS-716. This permits use with the
variety of modem standards worldwide while
avoiding the specialized approvals required for
equipment that is deemed to fall under the category
of telecommunication equipment.
In the modem mode, the SUPERtrol-I MS-716 is
assumed to be operating in a remote metering role.
The SUPERtrol-I MS-716 will support key items in
the Hayes Compatible “AT” Command Set. In this
role, the SUPERtrol-I MS-716 will have the following
special abilities:
0. Monitor the modem status as a task of the system
1. Instruct the modem to answer an incoming call
ATA
2. Respond to the calling modem at a compatible
baud rate and protocol
3. Perform error checking in conjunction with the
modem
4. Monitor the status of the carrier
5. Terminate the telephone connection in event the
connection is lost.
In addition, the SUPERtrol-I MS-716 will be capable
of initiating a call to a designed telephone number
in the event of a metering malfunction.
5
2. Installation
SUPERtrol-I MS-716 Flow Computer
General Mounting
Hints
2.1 General Mounting Hints:
Termination Connectors:
Minimum Wire Gauge:22 AWG
Maximum Wire Gauge:14 AWG
Voltage/current limits are limited by unit specifications.
Permanently Connected Equipment:
UL 3101-1, Section 6.12.2.1 specifies that:
•A switch or circuit breaker shall be included in the building
installation;
•It shall be in close proximity to the equipment and within easy
reach of the OPERATOR;
•It shall be marked as the disconnecting device for the equipment.
Ensure that the switch or circuit breaker chosen is suitable for the
power requirements of the unit.
2.2 Mounting Diagrams:
Vehicle Mount Dimensions
0.47
(12)
7.52 (191)
6.54 (116.1)
TOTAL
1
START
F14PRINT5GRAND
STOP
SCROLL7PRE 28F2
CLEAR
RATE2PRE 1
TIME
0 –
3
6
9
HELP
•
9.52 (241.8)
0.55
6.35 (161.3)
(14)
0.875 DIA. CONDUIT ENTRY
3 PLACES
1.82
(46.2)
3.69
(93.7)
Optional Mounting Bracket (MB)
9.13
(232)
4.61
(117)
MENU
ENTER
8.67 (220.2)
0.40
(10.2)
4.08 (103.5)
0.24 DIA. MOUNTING
THRU HOLE (4 PLACES)
0.90
(22.9)
1.82
(46.2)
9.52 (241.8)
6
START
STOP
TOTAL
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
8.34
(211.8)
3
MENU
6
ENTER
9
HELP
•
8.67 (220.2)
0.40
(10.2)
NOTE: All dimensions are
in inches (mm)
3. Applications
SUPERtrol-I MS-716 Flow Computer
Liquid Volume
3.1 Liquid Volume
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor can also be installed to correct for liquid thermal expansion (see
3.2 Corrected Volume).
Calculations:
• For Flowmeters with Pulse Outputs, Volume flow is calculated using
the flowmeter frequency output and the user entered K-Factor.
• For Flowmeters with Analog Transmitters, Volume flow is calculated
using the measured flowmeter signal and the user entered scale settings.
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total
• Analog Output
Rate or Total
• Pulse Output
Total
• Relay Outputs
Rate or Total Alarms
Liquid Volume
Illustration
Applications:
The Flow Computer can monitor actual volume flow and total of any
liquid. Flow alarms are provided via relays and datalogging is available
via analog (4-20mA) and serial outputs.
TOTAL
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
3
MENU
6
ENTER
9
HELP
•
START
STOP
Calculations
Flowmeter
Pulse Input; Average K-Factor
input frequency * time scale factor
Volume Flow =
K-Factor
Analog Input; Linear
Volume Flow = % input * Full Scale Flow
7
SUPERtrol-I MS-716 Flow Computer
Corrected
Liquid Volume
3.2 Corrected Liquid Volume
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor is installed to correct for liquid thermal expansion.
Calculations:
• Corrected Volume is calculated using the flow and temperature inputs
as well as the thermal expansion coefficient stored in the flow computer.
Use the "SET FLUID PROPERTIES" submenu to define reference
temperature and density values for standard conditions.
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total, Temperature,
Density
• Analog Output
Rate, Total, Temperature or Density
• Pulse Output
Total
• Relay Outputs
Rate , Total or Temperature Alarms
Corrected
Liquid Volume
Illustration
Applications:
Monitoring corrected volume flow and total of any liquid. Flow alarms are
provided via relays and datalogging is available via analog (4-20mA) and
serial outputs.
Actual volume is measured by the flow element (DP transmitter,
Flowmeter). Temperature is measured by the temperature transmitter. A
density transmitter can be used for direct density measurements.
Calculations:
• The density and mass flow are calculated using the reference density
and the thermal expansion coefficient of the liquid (see "SET FLUID
PROPERTIES" submenu)
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total, Temperature,
Density
• Analog Output
Rate, Total, Temperature or Density
• Pulse Output
Total
• Relay Outputs
Rate, Total or Temperature Alarms
Liquid Mass
Illustration
Applications:
Monitoring mass flow and total of any liquid. Flow alarms are provided via
relays and datalogging is available via analog (4-20mA) and serial outputs.
TOTAL
START
STOP
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
3
MENU
6
ENTER
9
HELP
•
START
STOP
TOTAL
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
3
MENU
6
ENTER
9
HELP
•
Calculations
Orifice Plate
with DP Transmitter
Volume Flow
As calculated in section 3.1
Mass Flow
Mass Flow = volume flow * density
Temp./Dens.
Tr ansmitter
9
FlowmeterTemp./Dens.
Tr ansmitter
SUPERtrol-I MS-716 Flow Computer
Batching
3.4 Batching
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor can also be installed to correct for liquid thermal expansion (see
3.2 Corrected Volume).
Calculations:
• For Flowmeters with Pulse Outputs, Volume flow is calculated using
the flowmeter frequency output and the user entered K-Factor.
• For Flowmeters with Analog Transmitters, Volume flow is calculated
using the measured flowmeter signal and the user entered scale settings.
• Corrected Volume is calculated using the flow and temperature inputs
as well as the thermal expansion coefficient stored in the flow computer.
Output Results:
•Display Results
Flow Rate, Batch Total, Non-Resettable Total, Temperature, Density
•Analog Output
Rate, Total, Temperature or Density
•Pulse Output
Total
•Relay Outputs
Batch Total, Rate, or Temperature Alarms
Batching Illustration
Applications:
Batching and monitoring flow and total of any liquid. Batching is
accomplished via relays and datalogging is available via analog (420mA) and serial outputs.
Examples using MLT787S+ Barrier (MTL4755ac for RTD)
Flow Input
Temperature Input
(4-20mA Transmitter)
Flow Input
Hazardous AreaSafe Area
4-20mA Flow
Transmitter
Q/∆P
4-20
+
–
3 4
28V
Diode
Temperature Input (4-20mA Transmitter)
Hazardous AreaSafe Area
2 1
1
2
3
4
1
24V Out
24V Out
4-20mA In
Common
Temperature Input
(RTD)
4-20mA Temp.
Transmitter
–
4-20
T
+
4 3
Temperature Input (RTD)
Hazardous AreaSafe Area
3-Wire RTD
1 2 3 4 5 6
28V
Diode
1 2
7 8 9 10 11 12 13 14 15 16 17
4
Common
8
4-20mA In
Common
4
5
RTD Excite (+)
6
RTD Sense (+)
7
RTD Sense (–)
8
12
SUPERtrol-I MS-716 Flow Computer
5. UNIT OPERATION
5.1 Front Panel Operation Concept for Run Mode
The SUPERtrol-I MS-716 is fully programmable through the front panel. Please review the
following usage summary before attempting to use the instrument.
TOTAL
RATE2PRE 1
START
STOP
1
F14PRINT5GRAND
SCROLL7PRE 28F2
CLEAR
TIME
0 –
3
MENU
6
ENTER
9
HELP
•
HELP
On-line help is provided to assist the operator in using this product. The help is available during RUN and SETUP
modes simply by pressing the HELP key. The HELP key is used to enter decimals when entering numeric values.
FUNCTION KEYS
In the RUN mode, several keys have a special, direct access feature, to display an item of interest (i.e. RATE,
TOTAL, PRE 1, etc.). Press the key to view your choice. Press the SCROLL key to return to scrolling display.
CLEARING TOTALIZER
To clear the total, you must press the TOTAL Function Key 3 times. You will be asked to verify this action. The
operator will be prompted to enter password if the unit is locked.
NOTE:In the Batcher Mode, simply press the CLEAR key to reset the total (the batcher must be stopped or finished
batching). It is not necessary to press the TOTAL Function Key first.
CLEARING GRAND TOTAL
To clear the grand total, you must press the GRAND Function Key 3 times. You will be asked to verify this action.
The supervisor will be prompted to enter the supervisor password if the unit is locked.
PRESET KEYS USING EZ PRESET
In the RUN mode, PRE 1 & PRE 2 keys are used to view and/or change the preset setpoints. To view the Presets,
simply press the desired Preset key. Press Enter then press the Clear key for direct editing of the preset setpoints.
PRESET KEYS
In the RUN mode, PRE 1 & PRE 2 keys are used to view and/or change the preset setpoints. To view the Presets,
simply press the desired Preset key. Rapidly press the Preset keys 3 times, then press the Clear key for direct editing
of the preset setpoints. Press Enter or Start to complete the Preset entry.
SCROLL
Rapidly press the Scroll key twice to setup a display list.
Press the CLEAR key to remove old scroll list.
Press the function key for the item you wish to add
Use the ∆ ∇ keys to assign the line.
PRINT
The PRINT key is used to print on demand. When the PRINT key is pressed, a user defined list of data (TOTAL,
RATE, PRE 1, etc.) is sent to the RS-232 port. A timed message of "PRINTING" will be displayed to acknowledge the
print request.
SPECIAL BATCHING KEYS
The START and STOP keys are used only when batching to start and stop batches. The CLEAR key will clear the
total without first pressing the TOTAL key (unit must be stopped). All other keys work the same in both Rate/Total
mode and Batch mode. The Start and Stop keys operation are set by the control input settings. The Start options are:
START or RESET/START. The Stop options are: STOP or STOP/RESET.
MENU KEY
The MENU key is used to enter the Setup and Test modes. Press the MENU key to enter the Setup and Test modes.
(See section 6 for Setup mode, section 8 for Test mode). The MENU key is used as "escape" in Setup and Test
Programming. Pressing the MENU key wile programming in the Sub-Menu groups will backup the display to that
Sub-Menu group heading. Pressing the MENU key while viewing the Sub-Menu groups will backup the display to the
Top Level Menu.
ACKNOWLEDGING ALARMS
Most alarm messages are self-clearing. Press the ENTER key to acknowledge and clear alarms.
NOTE: Some keys and functions are password protected. Enter the password to gain access. The passwords are
factory set as follows:Operator = 0Supervisor = 2000
13
SUPERtrol-I MS-716 Flow Computer
General
Operation
Rate/Total
Operation
Password Protection
(Rate/Total mode)
5.2 General Operation
The unit can display: Rate, Total, Grand Total, Temperature, Density, Presets
and Time of Day. The Temperature and/or Density can be displayed even if
you are using the Volumetric Flow Equation (a Temperature or Density sensor
must be installed). The unit can perform Mass or Corrected Volume equations
using a temperature or density sensor (these equations can be computed
without Temp/Dens sensors by using user defined default values). The unit can
be programmed to perform Ratemeter/Totalizer or Batching functions (see
section 6.3, SELECT INSTRUMENT Submenu).
5.3 Ratemeter/Totalizer Operation
The Ratemeter/Totalizer mode is used primarily to monitor flowrate and
accumulated total. The relays can be used to trigger flow, total, temperature or
density alarms.
5.3.1 Password Protection for Rate/Total mode
After an Operator and/or Supervisor Password is entered in the setup mode
(see section 6.3, SETUP PASSWORD submenu), the unit will be locked. The
unit will prompt the user for the password when trying to perform the following
functions:
Clear Total
Clear Grand Total
Enter Menu
Edit Preset 1 (PRE 1 Key)
Edit Preset 2 (PRE 2 Key)
Relay Operation
(Rate/Total mode)
Pulse Output
(Rate/Total mode)
The Supervisor password should be reserved for supervisors. The Supervisor
password will allow access to restricted areas of the Setup and Test menus.
5.3.2 Relay Operation in Rate/Total mode
Up to four relays are available (two standard) for alarm outputs. The relays can
be assigned to trip according to rate, total, temperature, density readings or
general system alarms. The relays can be programmed for low or high alarms.
Preset 1 (RLY1) and Preset 2 (RLY2) are easily accessible by pressing the
PRE 1 or PRE 2 key on the front panel. Preset 3 and Preset 4 are accessible
only through the setup menu.
5.3.3 Pulse Output in Rate/Total mode
The isolated pulse output (open collector) is menu assignable to Volume Total,
Corrected Volume Total or Mass Total. The pulse output duration can be set for
10mS (50 Hz max) or 100mS (5 Hz max). A pulse output scale factor (pulse
value) can be set to scale the pulse output. The pulse output is ideal for
connecting to remote totalizers or other devices such as a PLC. See section
1.3 for electrical specifications.
Analog Output
(Rate/Total mode)
5.3.4 Analog Output in Rate/Total mode
The analog output is menu assignable to correspond to the Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature, Density, Volume Total,
Corrected Volume Total or Mass Total. The analog output is ideal for "trend"
tracking using strip chart recorders or other devices.
14
SUPERtrol-I MS-716 Flow Computer
RS-232 Serial Port
(Rate/Total mode)
5.3.5 RS-232 Serial Port Operation in Rate/Total mode
The RS-232 serial port can be used for programming (using the Setup Disk)
or for communicating to printers and computers in the Operating Mode (Run
Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status
such as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Operation of RS-232 Serial Port with Printers:
Transaction Printing
For transaction printing, the user defines the items to be included in the
printed document (see section 6.3.20 SET DATA OUTPUT, Select_list). The
transaction document can be initiated by pressing the PRINT key or by a
remote contact closure.
Data Logging
In data logging, the user defines the items to be included in each data log
(see section 6.3.20 SET PRINTER OUTPUT, Select_list). The user can also
select when (time of day) or how often (print interval) the data log is to be
made (see section 6.3.19 SET PRINTER OUTPUT, Configure).
System Setup and Maintenance Report
The system setup and maintenance report lists all of the instrument setup
parameters and usage for the current instrument configuration. The audit
trail information and a status report is also printed. This report is initiated in
the Test menu (see section 8.2.3 PRINT SYSTEM SETUP).
RS-485 Serial Port
(Rate/Total mode)
5.3.6 RS-485 Serial Port (optional)
RS-485 Port Description:
The optional RS-485 card utilizes Modbus RTU protocol to access
a variety of process parameters and totalizers. The Relays can
be controlled via Modbus. In addition, action routines can be
executed. For further information, contact factory and request
RS-485 Protocol manual.
Operation of Serial Communication Port with PC
The flow computer's RS-485 channel supports a number of
Modbus RTU commands. Modbus RTU drivers are available
from third party sources for a variety of Man Machine Interface
software for IBM compatible PC's.
The user reads and writes information from/to the RS-485 using
the Modbus RTU commands. The SUPERtrol-I MS-716 then
responds to these information and command requests.
Process variables and totalizers are read in register pairs in
floating point format. Time and date are read as a series of
integer register values. Alarms are individually read as coils.
Action routines are initiated by writing to coils.
15
SUPERtrol-I MS-716 Flow Computer
5.4 Batcher Operation
The Batcher mode is used primarily to control batches. The main difference
between the Batch mode and Rate/Total mode is the relay operation. The
Batch mode allows the operator to "START" the unit via the front panel or
remote input. Once started, the relays (RLY1 & RLY2) will energize and send
power to a flow control device (i.e. solenoid valve or pump). The flow sensor
will send a signal to the unit and total accumulation will begin. When the
Prewarn value (PRE 2) is reached, Relay 2 will drop out (this is ideal for flow
slow down). When the Batch amount (PRE 1) is reached, Relay 1 will drop out
and the Batch is complete.
Several messages will be displayed during normal batch operation (i.e. Batch
Fill, Batch Stopped). The keypad is disabled for the duration of these timed
messages (approx. 2 sec).
Batcher Configuration
5.4.1 Batcher Configuration.
When the unit is programmed for batch mode, several batch operation choices
are available. These choices include: Up or Down Counting, Maximum Batch
Preset, Batch Overrun Compensation, Auto Batch Restart, Time Delay, Flow
Signal Timeout, Maximum Drain Time, Slow Start Quantity, Start or Reset/Start,
and Stop or Stop/Reset.
Standard Preset or EZ Preset
Use Standard Preset for applications in which the batch amount does not
change frequently.
Use EZ Preset in applications in which the batch amounts change frequently.
The EZ Preset mode was designed to enter presets with minimum key strokes.
Batch Count Mode
The Batch Count Mode allows the user to choose whether the unit will batch up
to a preset value or batch down from a preset value to zero.
Maximum Batch Preset
The Maximum Batch Preset allows the user to program the Maximum Batch
value allowed to be entered by the operator. If an operator should try to
program a batch higher then this value, the unit will not allow the value to be
entered and will prompt the user with an error message saying that the Maximum
Batch Preset has been exceeded.
Batch Overrun
The Batch Overrun is used for batch applications that have slow responding
valves and a consistent batching flowrate. When the Batch Overrun is set, the
unit will compensate for batch overruns by computing an averaged overrun
value from the last four batches. This average is used to internally adjust the
batch setpoint to minimize overrun.
Flow Signal Timeout
The Flow Signal Timeout allows the user to enter a timeout of 0 to 99 seconds.
If a batch is “Filling” and zero flow persists for more than the user entered time
then the batch will be aborted. This prevents over flows due to faulty flow
sensors and/or wiring.
Maximum Drain Time
The unit declares that a batch is “done” when the flow rate equals “0”. A flow
rate may be present long after the Preset Relay de-energizes due to slow
reacting valves or leaky valves. The Maximum Drain Time allows the user to
enter an amount of time (0 to 99 seconds) to wait before declaring “Batch
Done”. After the Preset Batch quantity is reached, the unit will declare “Batch
Done” when the flow rate is “0” or the Maximum Drain Time has expired. The
batch data will then be available for printing and datalogging.
16
SUPERtrol-I MS-716 Flow Computer
Slow Start Quantity
The Slow Start Quantity is a function that allows an amount to be entered for a
Slow Start up. This function requires two stage valve control. RLY 1 (slow flow)
will energize for Slow Start and RLY 2 (fast flow) will energize after the Slow
Start Quantity has been delivered. This helps reduce turbulence when filling an
empty container.
START, RESET/START and STOP, STOP/RESET
When configuring the control inputs, Control Input1 can be set for START or
RESET/START. When set for START, the unit will start batching when a signal
is applied to Control Input1 or the front panel Start key is pressed. A separate
Reset signal must be used to clear the previous batch total. When set for
RESET/START, the unit will automatically reset then start when a signal is
applied to Control Input1 or the front panel Start key is pressed (provided that
the pervious batch was completed). If a previous batch was stopped during a
batch cycle, the unit will Start from where it was stopped.
Control Input 2 can be set for STOP or STOP/RESET. When set for STOP, the
unit will stop batching when a signal is applied to Control Input 2 or the front
panel Stop key is pressed. A separate Reset signal must be used to clear the
batch total. When set for STOP/RESET, a running batch will stop when a
signal is applied to Control Input 2 or the front panel Stop key is pressed. If the
unit is Stopped or after a completed batch, the unit will reset when a signal is
applied to Control Input 2 or the front panel Stop key is pressed.
NOTE:Applying a voltage level to Control Input 2 will inhibit all Start inputs in
either mode.
Password Protection
(Batch mode)
Relay Operation
(Batch mode)
5.4.2 Password Protection for Batcher Mode
After an Operator and/or Supervisor Password is entered in the setup mode
(see section 6.3, SETUP PASSWORD submenu), the unit will be locked. The
unit will prompt the user for the password when trying to perform the following
functions:
Clear Grand Total
Enter Menu
The Supervisor password should be reserved for supervisors. The Supervisor
password will allow access to restricted areas of the Setup and Test menus.
The passwords are factory set as follows:
Operator = 0
Supervisor = 2000
5.4.3 Relay Operation in Batcher mode
Up to four relays are available (two standard) for alarm outputs. Preset 1
(RLY1) is reserved for batch amount, Preset 2 (RLY2) is reserved for prewarn.
(see section 5.4 Batcher Operation for Relay 1 & Relay 2 functions)
Preset 1 (RLY1) and Preset 2 (RLY2) are easily accessible by pressing the PRE
1 or PRE 2 key on the front panel. Preset 3 and Preset 4 are accessible only
through the setup menu.
Relays 3 and 4 can be assigned to trip according to rate, total, temperature,
overrun or alarm. When Rate is selected the relays can be programmed for low
or high alarms.
Pulse Output
(Batch mode)
Analog Output
(Batch mode)
5.4.4 Pulse Output in Batcher mode
The isolated pulse output (open collector) is menu assignable to Volume Total,
Corrected Volume Total or Mass Total. The pulse output duration can be set for
10mS (50 Hz max) or 100mS (5 Hz max). A pulse output scale factor (pulse
value) can be set to scale the pulse output. The pulse output is ideal for
connecting to remote totalizers or other devices such as a PLC. See section 1.3
for electrical specifications.
5.4.5 Analog Output in Batcher mode
The analog output is menu assignable to correspond to the Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature, Density, Volume Total,
Corrected Volume Total or Mass Total. The analog output is ideal for "trend"
tracking using strip chart recorders or other devices.
17
SUPERtrol-I MS-716 Flow Computer
RS-232 Serial Port
(Batch mode)
5.4.6 RS-232 Serial Port Operation in Batcher mode
The RS-232 serial port can be used for programming (using the Setup Disk)
or for communicating to printers and computers in the Operating Mode (Run
Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status such
as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Operation of RS-232 Serial Port with Printers:
Transaction Printing
For transaction printing, the user defines the items to be included in the
printed document (see section 6.3.20 SET DATA OUTPUT, Select_list). The
transaction document can be initiated by pressing the PRINT key, by a
remote contact closure or print at end of batch.
Data Logging
In data logging, the user defines the items to be included in each data log (see
section 6.3.20 SET PRINTER OUTPUT, Select_list). The user can also
select when (time of day) or how often (print interval) the data log is to be
made (see section 6.3.19 SET PRINTER OUTPUT, Configure).
System Setup and Maintenance Report
The system setup and maintenance report lists all of the instrument setup
parameters and usage for the current instrument configuration. The audit trail
information and a status report is also printed. This report is initiated in the
Test menu (see section 8.2.3 PRINT SYSTEM SETUP).
RS-485 Serial Port
(Batch mode)
5.4.7 RS-485 Serial Port (optional)
RS-485 Port Description:
The optional RS-485 card utilizes Modbus RTU protocol to access
a variety of process parameters and totalizers. Batches/Relays
can be controlled remotely via Modbus. In addition, action routines
can be executed. For further information, contact factory and
request RS-485 Protocol manual.
Operation of Serial Communication Port with PC
The flow computer's RS-485 channel supports a number of
Modbus RTU commands. Modbus RTU drivers are available
from third party sources for a variety of Man Machine Interface
software for IBM compatible PC's.
The user reads and writes information from/to the RS-485 using
the Modbus RTU commands. The SUPERtrol-I MS-716 then
responds to these information and command requests.
Process variables and totalizers are read in register pairs in
floating point format. Time and date are read as a series of
integer register values. Alarms are individually read as coils.
Action routines are initiated by writing to coils.
18
SUPERtrol-I MS-716 Flow Computer
6. PROGRAMMING
6.1 Front Panel Operation Concept for Program Mode
The SUPERtrol-I MS-716 is fully programmable through the front panel.
Please review the following usage summary before attempting to use
the instrument.
TOTAL
RATE2PRE 1
START
STOP
1
F14PRINT5GRAND
SCROLL7PRE 28F2
CLEAR
TIME
0 –
3
MENU
6
ENTER
9
HELP
•
How To Make Mode
Changes
How To Navigate
Through Sub-Menu
Groups
How To Select
Program Choices
Setup Mode:
MODE CHANGES
Pressing the MENU key will offer selections of RUN, SETUP, TEST.
RUN is the normal operating mode for the instrument. SETUP offers
various sub-menus used for instrument setup. TEST offers various
sub-menus for Test, Calibration and System Start-up.
Submenu GROUP NAVIGATION
Use the UP and DOWN arrow keys to navigate up and down through
the Sub-Menu groups when in the SETUP or TEST mode. Press the
ENTER key to enter a desired setup or test Sub-Menu group.
SELECTION OF ITEM
During setup, the unit will often offer multiple choices for a given topic.
The topic prompt appears on the top line of the display. The choices are
shown on the lower line of the display.
To select an item, press the key beneath the desired choice. The
selected choice will blink. Press the ENTER key to accept the
selected choice.
How To Enter
Numeric Values
How To Enter
Text Characters
NUMERIC ENTRY
The keys labeled "0 - 9", "–", ".", CLEAR and ENTER are used to enter
numerical values. A leading 0 will assume that you intend to enter a
minus "–" sign. Press the CLEAR key to clear the existing value and to
enable editing.
TEXT CHARACTER ENTRY
Some setup items (i.e. Descriptors, Units Label) require the user to
enter text characters. Press CLEAR to enable editing. The UP and
DOWN arrow keys are used to scroll through the available character
sets for each individual character. Press the ENTER key to accept the
character and advance to the next character.
19
SUPERtrol-I MS-716 Flow Computer
6.2 EZ Setup
The EZ Setup routine is a quick and easy way to configure the unit for the most
commonly used instrument functions. This setup assumes that you are
measuring Volumetric Flow using a high level, DC Pulsing flow sensor. Entering
the EZ Setup mode automatically sets many features. This may cause any
previously programmed information to be lost or reset. For a complete
customized configuration, see sections 6.3 and 6.4.
MenusDisplayNotes
6.2.1
TOP LEVEL
SETUP MENU
6.2.2
EZ Setup
Submenu
Groups
SELECT OPERATE STATE
Run Setup Test
ENTER
SELECT EZ SETUP
ENTER
ARE YOU SURE?
No Yes
ENTER
INSTRUMENT TYPE
Rate/Tot Batch
ENTER
RATE TIME BASE
Sec Min Hour Day
ENTER
RATE DEC PLACES
Select Setup to enter the instrument
setup routine.
Press ENTER to begin EZ Setup
routine.
Confirm that you want to run EZ Setup.
Caution: Any previous program
settings may be lost or reset.
Instrument Type.
Select the appropriate rate time base.
Enter the desired rate decimal location.
0
0-3 decimal places allowed.
ENTER
TOTAL VOLUME UNITS
ENTER
TOTAL DEC PLACES
ENTER
K_FACTOR TYPE
Avg LinTbl UVC
ENTER
AVERAGE KA-FACTOR
####### P/gal
LINEAR TABLE KA
Fre01:######## Hz
ENTER
LINEAR TABLE KA
Through
16 Points
KA--01:##########
ENTER
gal
0
Enter the desired totalizer units label.
Enter the desired totalizer decimal
location. 0-3 decimal places allowed.
Enter the desired K-Factor Type.
If Average selected, Enter the desired
Average K-Factor.
If LinTbl or UVC selected,
Enter the desired frequency/ K-Factor
pair for each point in the Linearization
Table.
Enter a frequency of 0 for any point
other than Fre01 to exit Linearization
Table setup.
FS ANALOG OUT 20mA
#######gal/m
ENTER
RATE 00.0 gal/m
TOTAL 0 gal
Enter the desired full scale setting for
the analog output.
Return to Run Mode
20
SUPERtrol-I MS-716 Flow Computer
6.3 Setup Menus
MenusDisplayNotes
6.3.1
Top Level Setup
Menu
6.3.2
Submenu
Groups
START
SELECT OPERATE STATE
Run Setup Test
MENU
ENTER
SELECT EZ SETUP
STOP
START
INSTRUMENT TYPE
STOP
START
SELECT FLOW EQUATION
STOP
START
SETUP INDICATORS
STOP
START
SETUP FLOW INPUT
STOP
START
Select Setup to enter the instrument setup
routine.
Refer to Page 20 for Details.
Refer to Page 22 for Details.
Refer to Pages 22 for Details.
Refer to Page 23 & 24 for Details.
Refer to Page 25, 26 & 27 for Details.
SETUP AUX INPUT
STOP
START
SET FLUID PROPERTIES
STOP
START
SETUP PULSE OUTPUT
STOP
START
SETUP ANALOG OUTPUT
STOP
START
SETUP RELAYS
STOP
START
SETUP CONTROL INPUTS
STOP
START
Refer to Pages 28 for Details.
Refer to Page 29 for Details.
Refer to Pages 29 for Details.
Refer to Page 30 for Details.
Refer to Page 30 & 31 for Details.
Refer to Page 32 for Details.
STOP
SETUP REALTIME CLOCK
STOP
START
SERIAL USAGE
STOP
START
SETUP DATALOG/PRINT
STOP
START
ADMINISTRATIVE SETUP
STOP
START
SETUP NETWORK CARD
21
Refer to Page 33 for Details.
Refer to Page 34 for Details.
Refer to Pages 35 & 36 for Details.
Refer to Page 36 for Details.
Refer to Page 37 for Details.
6.4.1
SELECT EZ SETUP
6.4 Setup Sub-Menus
SELECT EZ SETUP
STOP
START
Advance To
INSTRUMENT TYPE
SUPERtrol-I MS-716 Flow Computer
NotesDisplaySub-menus
Refer to page 20 for EZ Setup routine.
Press the DOWN (stop) key to advance to
Instrument Type. Press the UP (start) key to
advance to Administrative Setup.
6.4.2
INSTRUMENT TYPE
Rate/Tot
Batch
INSTRUMENT TYPE
ENTER
INSTRUMENT TYPE
Rate/Tot Batch
STOP
START
ENTER
Advance To
SELECT FLOW EQUATION
INSTRUMENT TYPE
ENTER
INSTRUMENT TYPE
Rate/Tot Batch
STOP
START
ENTER
SELECT PRESET TYPE
Standard EZ Preset
STOP
START
ENTER
BATCH COUNT MODE
Up Down
STOP
START
ENTER
MAXIMUM BATCH PRESET
1000.0 gal
STOP
START
ENTER
BATCH OVERRUN COMP
Off On
ENTER
Press ENTER to enter Instrument Type submenus.
Press ENTER when Rate/Total is flashing to
configure the instrument as a Ratemeter/
Totalizer.
If Rate/Tot selected, advance to Select Flow
Equation.
Press ENTER to enter Instrument Type submenus.
Press ENTER when Batch is flashing to
configure the instrument as a Batcher.
Select Standard or EZ Preset
Select UP to Reset to 0 and count up to preset.
Select DOWN to reset to Preset and count
down to 0.
Enter the maximum allowable Batch Preset.
The operator will not be able to enter a batch
preset larger than this value.
Select ON to set the unit to operate using a
Batch Overrun Compensation routine.
Select OFF to inhibit Batch Overrun
Compensation routine. (See Section 5.4)
FLOW SIGNAL TIMEOUT
10
ENTER
MAXIMUM DRAIN TIME
10
ENTER
SLOW START QUANTITY
10
ENTER
Advance To
SELECT FLOW EQUATION
Enter a timeout of 0 to 99 seconds. If a batch is
“Filling” and zero flow persists for more than this
time, the batch will be aborted.
Enter time (0-99 sec.) for Max. Drain Time. After batch
quantity is reached, “Batch Done” is declared when the
flow rate is “0” or the Maximum Drain Time has expired.
Enter a quantity for a Slow Start up. RLY 2
(slow flow) will energize for Slow Start and RLY
1 (fast flow) will energize after the Slow Start
Quantity has been delivered.
22
SUPERtrol-I MS-716 Flow Computer
Sub-menus
6.4.3
SELECT
FLOW EQUATION
6.4.4
SETUP
INDICATORS
(Total)
Display
SELECT FLOW EQUATION
ENTER
SELECT FLOW EQUATION
Volume Mass Cor/Vol
ENTER
Advance To
SETUP INDICATORS
(Total)
SETUP INDICATORS
ENTER
SETUP INDICATORS
Total Dens Rate Temp
Notes
Press ENTER to enter Select Flow Equation
submenus.
Press ENTER when desired flow equation is
flashing.
Press ENTER to begin setup of the Indicators
Press ENTER when Total is flashing to
configure the Totalizer Indicators
6.4.5
SETUP
INDICATORS
(Density)
ENTER
TOTAL DESCRIPTOR
TOTAL
ENTER
TOTAL VOLUME UNITS
gal
ENTER
TOTAL DEC PLACES
0
ENTER
Advance To
SETUP INDICATORS
(Density)
SETUP INDICATORS
Total Dens Rate Temp
ENTER
Enter the desired Total Descriptor
Enter the desired Volume Units Label for the
Totalizer.
Select the desired Total Decimal Place.
0-3 decimal places allowed.
Press ENTER when Dens is flashing to
configure the Density Indicators.
DENSITY DESCRIPTOR
DENS
ENTER
DENSITY MASS UNITS
lbs
ENTER
DENSITY DEC PLACES
0
ENTER
Advance To
SETUP INDICATORS
(Rate)
Enter the desired Density Descriptor.
Enter the desired Mass Units Label for
Density.
Select the desired Density Decimal Place.
0-3 decimal places allowed.
23
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.6
SETUP
INDICATORS
(Rate)
SETUP INDICATORS
Total Dens Rate Temp
ENTER
RATE TIME BASE
Sec Min Hour Day
ENTER
RATE DESCRIPTOR
RATE
ENTER
RATE DEC PLACES
ENTER
RATE AVG FILTER
ENTER
Press ENTER when Rate is flashing to
configure the Ratemeter Indicators
Select the desired Rate Time Base.
Enter the desired Descriptor for the
Ratemeter.
Select the desired Rate Decimal Place.
0
0-3 decimal places allowed.
Enter desired Rate Averaging Filter.
0
6.4.7
SETUP
INDICATORS
(Temperature)
QUICK UPDATE %
ENTER
Advance To
SETUP INDICATORS
(Temperature)
SETUP INDICATORS
Total Dens Rate Temp
ENTER
TEMP DESCRIPTOR
TEMP
ENTER
Enter desired Percent of Change for Quick
5
Update. If the current flowrate deviates by an
amount greater than the percentage value
entered, the Rate Averaging is inhibited.
(See Page 57 for more details.)
Press ENTER when Temp is flashing to
configure the Temperature Indicators.
Enter the desired Temperature Descriptor.
TEMPERATURE SCALE
Deg_C Deg_F
ENTER
TEMP DEC PLACES
ENTER
Advance To
SETUP FLOW INPUT
Enter the desired Temperature Scale.
Select the desired Temperature Decimal
0
Place.
24
0-3 decimal places allowed.
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.8
SETUP
FLOW INPUT
(Pulse - Ain & PS
(A=B))
NOTE:
Ain = Single Pulse
PS(A=B) = Pulse
Security
Qx1 = Quadrature
Qx2 = Quadrature x 2
SETUP FLOW INPUT
ENTER
EXCITATION VOLTAGE
5v 12v 24v
ENTER
FLOW INPUT TYPE
Pulse Analog
ENTER
PULSE INPUT TYPE
AinPS(A=B) Qx1 Qx2
ENTER
PULSE TRIGGER LEVEL
10mV 100mV 2.5V
ENTER
Press ENTER to begin setup of Flow Input.
Select the desired Excitation Voltage.
Press ENTER when Pulse is flashing to
configure the flow input for Pulse signals.
Enter the desired Pulse type. See side note.
Select the desired Input Pulse Trigger Level.
LOW PASS FILTER
40Hz 3KHz 20KHz
INPUT TERMINATION
Pullup Pulldown None
MAX WINDOW (1-99)
K_FACTOR TYPE
Avg LinTbl UVC
AVERAGE KA-FACTOR
LINEAR TABLE KA
Fre01:######## Hz
LINEAR TABLE KA
Through
16 Points
KA--01:####### P/gal
ENTER
ENTER
ENTER
ENTER
####### P/gal
ENTER
ENTER
Select the desired Low Pass Filter.
(Max. Count Speed).
Select the proper input termination.
Enter the desired Maximum Sample Window
Time (1-99 sec).
1
Enter the desired K-Factor Type.
If Avg selected, Enter the desired Average KFactor.
If LinTbl selected,
Enter the desired frequency/ K-Factor pair for
each point in the Linearization Table.
NOTE: Enter 0 for Fre value of any point
(other than Fre01) to exit the routine
and use the values entered up to that
point.
LOW FLOW RATE ALARM
####### gal/m
ENTER
HIGH FLOW RATE ALARM
####### gal/m
ENTER
Advance To
SETUP AUX INPUTS
Enter the desired volumetric Low Rate Alarm.
This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
Enter the desired volumetric High Rate Alarm.
This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
25
SUPERtrol-I MS-716 Flow Computer
SubmenusDisplayNotes
6.4.9
SETUP
FLOW INPUT
(Pulse - Quadrature,
Qx1 or Qx2)
NOTE:
Ain = Single Pulse
PS(A=B) = Pulse
Security
Qx1 = Quadrature
Qx2 = Quadrature x 2
SETUP FLOW INPUT
ENTER
EXCITATION VOLTAGE
5v 12v 24v
ENTER
FLOW INPUT TYPE
Pulse Analog
ENTER
PULSE INPUT TYPE
Ain PS(A=B) Qx1
ENTER
Qx2
PULSE TRIGGER LEVEL
10mV 100mV 2.5V
ENTER
LOW PASS FILTER
40Hz 3KHz 20KHz
ENTER
INPUT TERMINATION
Pullup Pulldown None
ENTER
Press ENTER to begin setup of Flow Input.
Select the desired Excitation Voltage.
Press ENTER when Pulse is flashing to
configure the flow input for Pulse signals.
Enter the desired Pulse type. See side note.
Select the desired Input Pulse Trigger Level.
Select the desired Low Pass Filter.
(Max. Count Speed).
Select the proper input termination.
MAX WINDOW (1-99)
ENTER
K_FACTOR TYPE
Avg LinTbl UVC
ENTER
AVERAGE KA-FACTOR
AVERAGE KB-FACTOR
LINEAR TABLE KA
Fre01:######## Hz
ENTER
LINEAR TABLE KA
Through
16 Points
KA--01:####### P/gal
ENTER
LINEAR TABLE KB
Fre01:######## Hz
ENTER
LINEAR TABLE KB
Through
16 Points
KA--01:####### P/gal
ENTER
####### P/gal
####### P/gal
Enter the desired Maximum Sample Window
Time (1-99 sec).
1
Enter the desired K-Factor Type.
If Avg selected, Enter the desired Average KFactor (KA for channel A).
Enter the desired Average K-Factor (KB for
channel B).
If LinTbl selected,
Enter the desired frequency/ K-Factor pair for
each point in the Linearization Table. (channel
A)
NOTE: Enter 0 for Fre value of any point
(other than Fre01) to exit the routine
and use the values entered up to that
point.
Enter the desired frequency/ K-Factor pair for
each point in the Linearization Table. (channel
B)
NOTE: Enter 0 for Fre value of any point
(other than Fre01) to exit the routine
and use the values entered up to that
point.
LOW FLOW RATE ALARM
####### gal/m
ENTER
HIGH FLOW RATE ALARM
####### gal/m
ENTER
Advance To
SETUP AUX INPUTS
Enter the desired volumetric Low Rate Alarm.
This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
Enter the desired volumetric High Rate Alarm.
This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
26
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.10
SETUP
FLOW INPUT
(Analog)
SETUP FLOW INPUTS
ENTER
EXCITATION VOLTAGE
5v 12v 24v
ENTER
FLOW INPUT TYPE
Pulse Analog
ENTER
ANALOG SIGNAL TYPE
Voltage Current
ENTER
ANALOG VOLTAGE RANGE
0-10V 0-5V 1-5V
ANALOG CURRENT RANGE
4-20mA 0-20mA
Press ENTER to begin setup of the Flow
Input.
Select the desired Excitation Voltage.
Press ENTER when Analog is flashing to
configure the flow input for Analog signals .
Choose Analog Signal Type.
If Voltage selected,
Choose desired Voltage Range.
If Current selected,
Choose desired Current Range.
ENTER
LINEARIZATION TYPE
Linear Sqrt LinTbl
ENTER
LINEAR TABLE KA
APR01:######## gal/m
ENTER
LINEAR TABLE KA
Through
16 Points
CFr01:########
ENTER
FLOW LOW SCALE
####### gal/m
ENTER
FLOW FULL SCALE
####### gal/m
Select the desired Linearization Type.
If LinTbl selected,
Enter the desired Apparent Input Flow (APR) /
Correction Factor (CFr) pair for each point in
the Linearization Table.
NOTE: Enter 0 for APR value of any point
(other than APR01) to exit the routine
and use the values entered up to that
point.
Enter the low flowrate corresponding to the
low analog signal.
Enter the High flowrate corresponding to the
High analog signal.
ENTER
LOW FLOW CUTOFF
####### gal/m
ENTER
LOW FLOW RATE ALARM
####### gal/m
ENTER
HIGH FLOW RATE ALARM
####### gal/m
ENTER
Advance To
SETUP AUX INPUT
Enter the desired Low Flow Cutoff.
Enter the desired volumetric Low Rate Alarm.
This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
Enter the desired volumetric High Rate Alarm.
This will trigger an alarm message if alarm
conditions occur. The relays are not affected.
27
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.11
SETUP AUX INPUT
NOTE:
When Density (Dens)
is selected, The menu
prompts will be very
similar to the Temperature prompts.
The menus will
prompt the user for
density values and
density units.
SETUP AUX INPUT
ENTER
AUX INPUT TYPE
None Dens Temp
ENTER
AUX SIGNAL TYPE
Voltage Current RTD
ENTER
INPUT SIGNAL RANGE
0-10V 0-5V 1-5V
INPUT SIGNAL RANGE
4-20mA 0-20mA
ENTER
AUX LOW SCALE
###### F
ENTER
Press ENTER to begin setup of the Auxiliary
Input.
Select Temperature to set the Auxiliary Input
for Temperature inputs.
Choose Temperature Signal Type.
Advance to "Aux Default", if RTD selected.
If Voltage selected,
Choose desired Voltage Range. Skip if RTD.
If Current selected,
Choose desired Current Range. Skip if RTD.
Enter the low temperature scale
corresponding to the low temperature signal.
Skip if RTD.
AUX FULL SCALE
###### F
ENTER
AUX DEFAULT
###### F
ENTER
AUX LOW ALARM
###### F
ENTER
AUX HIGH ALARM
###### F
ENTER
DENS EXTRACT METHOD
Therm_Coef API_2540
ENTER
Advance To
SET FLUID PROPERTIES
Enter the high temperature scale
corresponding to the high temperature signal.
Skip if RTD.
Enter the Default Temperature.
The unit will use this value if the temperature
input fails.
Enter the Low setpoint for the Temperature
Alarm.
Enter the High setpoint for the Temperature
Alarm.
Choose the Density Extraction method to be
used.
28
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.12
SET FLUID
PROPERTIES
SET FLUID PROPERTIES
ENTER
REF. DENSITY
###### lbs/g
ENTER
REF. TEMPERATURE
###### F
ENTER
EXPAN. FACTOR [xe-6]
########
ENTER
CALIBRATION DENSITY
########
ENTER
Press ENTER at this prompt to Set Fluid
Properties.
Enter the Reference Density. This is used in
the calculation of density when you have a
temp transmitter and used for corrected flow
calculation if you have a density transmitter.
Enter the Reference Temperature.
Enter the proper Expansion Factor.
(If Temp Compensated for Mass or Corrected
Volume) See Section 7.4, Calculating the
Expansion Factor.
Enter the Calibration Density. This is used in
calculation of flow for analog inputs using SQRT.
VISCOSITY COEF. A
0.000
ENTER
VISCOSITY COEF. B
0.000
ENTER
BASE DENSITY H2O@4C
###### lbs/g
ENTER
Advance To
SETUP PULSE OUTPUT
Enter the Viscosity A Coefficient. See section
7.5, Computation of Viscosity Coef. A and B.
Enter the Viscosity B Coefficient. See section
7.5, Computation of Viscosity Coef. A and B.
Enter the Base Density H2O@4C. This is
used in the centistoke calculation for UVC.
6.4.13
SETUP
PULSE OUTPUT
SETUP PULSE OUTPUT
ENTER
PULSE OUTPUT USAGE
Off Vol CVol/Mass
ENTER
PULSE WIDTH
10mS 100mS
ENTER
PULSE VALUE
####### gal/P
ENTER
Advance To
SETUP ANALOG OUTPUT
Press ENTER at this prompt to setup the
Pulse Output.
Select the desired Pulse Output Usage.
Select the desired Pulse Width for the Pulse
Output.
Enter the desired Pulse Value for the Pulse
Output (Units per Pulse).
29
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.14
SETUP
ANALOG OUTPUT
SETUP ANALOG OUTPUT
ENTER
ANALOG OUTPUT USAGE
Rate Total Temp Dens
ENTER
ANALOG OUT FLOW TYPE
Vol CVol/Mass
ENTER
ANALOG OUTPUT RANGE
4-20mA 0-20mA
ENTER
LS ANALOG OUTPUT
####### gal/m
ENTER
Press ENTER when Analog is flashing to
setup the Analog Output.
Select the desired Analog Output Usage.
Only if Rate selected & Flow EQ. = Mass, Cor/Vol
Select the desired Analog Output Flow.
Select the desired current range for the Analog
Output.
Enter desired Analog Output Low Scale Value.
NOTE: Units label will correspond with output
usage type selected.
6.4.15
SETUP RELAYS
(Relay 1 & Relay 2)
NOTE:
In Batch mode,
Relay 1 is reserved
for Preset,
Relay 2 is reserved
for Prewarn.
FS ANALOG OUT 20mA
####### gal/m
ENTER
ANALOG OUT DAMPING
0.0
ENTER
Advance To
SETUP RELAYS
SETUP RELAYS
Rly1
Rly2 Rly3 Rly4
ENTER
RELAY 1 USAGE
RATE TOTAL NA
ENTER
RELAY 1 DELAY sec
0
Enter desired Analog Output Full Scale Value.
Enter the desired Analog Output Damping
Constant.
Select the desired Relay for setup.
(Relays 3 & 4 Optional)
If Relay 1 or Relay 2 Selected,
Select Rate, Total or NA.
If Rate selected, enter desired relay activation
delay value.
ENTER
RELAY 1 MODE
LO_ALARM HI_ALARM
ENTER
RELAY 1 DURATION
#####
ENTER
RELAY 1 SETPOINT
####### gal
ENTER
RELAY 1 HYSTERESIS
##### gal/m
ENTER
Advance To
SETUP RELAYS 3, 4
Select the desired Relay Activation.
Low: Relay activates when reading is below
setpoint.
High:Relay activates when reading is above
setpoint.
If Total Selected, Enter desired Relay
Duration.
Enter the desired Setpoint. The Setpoint can
be edited in run mode using the PRE 1 key
(PRE 2 key for Relay 2).
If Rate, selected, Enter desired Relay
Hysteresis.
Settings for Relays
3 & 4 may be
entered even if
relays are not
supplied. The
settings will still
trigger display
alarms.
SETUP RELAYS
Rly1 Rly2 Rly3 Rly4
ENTER
RELAY 3 USAGE
Rate Tot Aux Ovrn NA
RELAY 4 USAGE
Rate Tot Aux Alrm NA
ENTER
RELAY 3 DELAY sec
0
ENTER
RELAY 3 MODE
LO_ALARM HI_ALARM
ENTER
RELAY 3 DURATION
#####
Select the desired Relay for setup.
(Relays 3 & 4 Optional)
If Relay 3 Selected,
Choose Rate, Total, Aux, Ovrn or NA.
If Relay 4 Selected,
Choose Rate, Total, Aux, Alrm or NA.
If Rate / Aux selected, enter desired relay
activation delay value.
Select the desired Relay Activation for Rate/Aux.
Low: Relay activates when reading is below
setpoint.
High:Relay activates when reading is above
setpoint.
If Total Selected, Enter desired Relay
Duration.
ENTER
RELAY 3 SETPOINT
Enter the desired Setpoint.
####### gal
ENTER
RELAY 3 HYSTERESIS
##### gal/m
ENTER
If Rate, selected, Enter desired Relay
Hysteresis.
Advance To
SETUP CONTROL INPUTS
RELAY NOTES & CONSIDERATIONS
1. Relay activation is based on the computed readings not the displayed value. Therefore the display
damping factor will not affect the relay response time. The RELAY DELAY feature allows the user
to enter a time delay for relay activation. This feature is very useful in applications where short over/
under range conditions are not considered alarm conditions.
2. When INSTRUMENT TYPE is set to batcher, Relay 1 is reserved for PRESET and Relay 2 is
reserved for PREWARN.
3. Setting the relays to NA (Not Assigned), will allow the relay activation to be controlled via the RS232 Serial and/or RS-485 Modbus ports.
4. Relay 3 and Relay 4 settings may be used to trigger display alarm conditions even if the relays are
not supplied.
31
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.16
SETUP CONTROL
INPUTS
(RATE/TOTAL)
SETUP CONTROL INPUTS
ENTER
SETUP CONTROL INPUTS
Input1 Input2 Input3
ENTER
CONTROL INPUT1 USAGE
INHIBIT_TOTAL NA
CONTROL INPUT2 USAGE
RESET_TOTAL NA
CONTROL INPUT3 USAGE
Prn Ack KeyLk NA
ENTER
Advance To
SETUP REALTIME CLOCK
Press Enter to begin setup of the Control
Inputs.
Select the desired Control Input for setup.
If Control Input 1 Selected,
Select Inhibit Total or NA (Not Assigned).
If Control Input 2 Selected,
Select Reset Total or NA (Not Assigned).
If Control Input 3 Selected,
Select Prn (Print), Ack (acknowledge), KeyLk
(Keylock) or NA (Not Assigned). ACK will
acknowledge and clear alarms and warning
messages.
Note: Alarms may reassert themselves if
alarm conditions are still present.
6.4.17
SETUP CONTROL
INPUTS
(BATCH)
SETUP CONTROL INPUTS
Input1 Input2 Input3
ENTER
CONTROL INPUT1 USAGE
Start Rst/Start NA
CONTROL INPUT2 USAGE
Stop Stop/Rst NA
CONTROL INPUT3 USAGE
Rst Prn KeyLk Ack NA
ENTER
Advance To
SETUP REALTIME CLOCK
Select the desired Control Input for setup.
If Control Input 1 Selected,
Select Start ,Reset/Start, NA (Not Assigned).
If Control Input 2 Selected,
Select Stop, Stop/Reset, NA (Not Assigned).
If Control Input 3 Selected,
Select Prn (Print), Ack (acknowledge), KeyLk
(Keylock) or NA (Not Assigned). ACK will
acknowledge and clear alarms and warning
messages.
Note: Alarms may reassert themselves if
alarm conditions are still present.
32
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.18
SETUP REALTIME
CLOCK
(Time)
SETUP REALTIME CLOCK
ENTER
SETUP REALTIME CLOCK
Time Date
ENTER
CLOCK TYPE
24HR 12HR
ENTER
SELECT CLOCK AM/PM
AM PM
ENTER
TIME OF DAY HH:MM:SS
##:##:##
ENTER
Press Enter to begin setup of the Realtime
Clock.
Select Time to set the time.
Select 24Hr or 12Hr clock
If 12Hr Clock,
Enter AM or PM
Enter time of day.
6.4.19
SETUP REALTIME
CLOCK
(Date)
Advance To
SETUP REALTIME CLOCK
(Date)
SETUP REALTIME CLOCK
Time Date
ENTER
DATE: MONTH,DAY,YEAR
##/##/####
ENTER
Advance To
SERIAL USAGE
Select Date to enter the date.
Enter the date. (Month, Day, Last two digits of
Year)
33
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.20
SERIAL USAGE
SERIAL USAGE
ENTER
SERIAL HARDWARE
RS232 RS485
ENTER
DEVICE ID
##
ENTER
BAUD RATE
300 600 1200 <more>
BAUD RATE
2400 4800 9600 19200
ENTER
PARITY
None Odd Even
Press Enter to begin setup of the Serial Port.
Select Serial Hardware type for standard port.
(See SETUP NETWORK CARD for RS485
Modbus option)
Select the Device ID.
Select the desired Baud Rate.
(If <more> selected)
Select the desired Parity.
ENTER
HANDSHAKING
None Softwre Hardwre
ENTER
DEVICE LINE FEED
<CR> <CR+LF>
ENTER
MODEM OPTIONS
No Yes
ENTER
MODEM INIT MASTER
No Yes
ENTER
MODEM AUTO ANSWER
No Yes
ENTER
Set the Handshake.
Choose end of line termination. Only choose
<CR> if your external device automatically
assigns a line feed for every <CR> carriage
return.
Select "Yes" if the serial port will be used to
control a modem.
Select "Yes" to have the unit engage in a
configuration conversation with the modem on
power up .
Select the desired Modem Auto Answer mode.
CALL OUT PHONE #
0
ENTER
CALL OUT TIME
##:##:##
ENTER
Continued on Next Page
Enter the Call Out Phone Number to be dialed
for "Call Out Time" or "Print On Error/Alarm".
Enter the time of day to perform Call Out
transmission.
34
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.20
SERIAL USAGE
(continued)
6.4.21
SETUP
DATALOG/PRINT
(Configure)
CALL ON ERROR/ALARM
No Yes
ENTER
NUMBER OF REDIALS
0
ENTER
HANGUP IF 2MIN INACT
No Yes
ENTER
Advance To
SETUP DATALOG/PRINT
SETUP DATALOG/PRINT
ENTER
Select "Yes" to have the unit perform a Call
Out transmission upon error/alarm condition.
Enter the number of redials to be performed
on call out time if busy or no answer. (error/
alarm tries until connected)
Select "Yes" to perform hangup if there is
inactivity for more than 2 minutes.
Press Enter to setup the Datalog/Print
information.
SETUP DATALOG/PRINT
Config Select_list
ENTER
OUTPUT FORMAT
Printer Term Dbase
ENTER
PAGE LENGTH [66 max]
66
ENTER
TOP MARGIN [60 max]
3
ENTER
DATALOG PRINT ONLY
No Yes
ENTER
PRINT TIME HH:MM:SS
00:00:00
Select Config to configure the Datalog/Print
information.
Select the type of Output Format.
Enter the desired Page Length.
If Printer selected above.
Enter the desired Top Margin.
If Printer selected above.
Select Yes to record events to the datalogger
only. Events will not be sent to the serial port.
Enter Print Time, printer will print at this time
every day. Enter 00:00:00 to inhibit print time.
ENTER
PRINT INTERVAL
00:00:00
ENTER
ENABLE PRINT KEY
NO YES
ENTER
PRINT END OF BATCH
NO YES
ENTER
Advance To
SETUP DATALOG/PRINT
(Select_list)
Enter Print Interval,
Enter 00:00:00 to inhibit print interval..
Select YES to enable Print Key.
Select NO to disable Print Key
Batch mode only.
Select Yes to print at end of batch.
35
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
6.4.22
SETUP
DATALOG/PRINT
(Select_list)
STOP
START
List Items:
TOTAL
RATE
PRE1
TEMP
GRAND
PRE2
DENS
TIME
SET DATALOG/PRINT
ENTER
SET DATALOG/PRINT
Config Select_list
PRINT LIST ITEMS
TOTAL YES
PRINT LIST ITEMS
RATE YES
PRINT LIST ITEMS
PRE 1 YES
ENTER
PRINT LIST ITEMS
DataLog size =001489
ENTER
Press enter to begin Setup Datalog/Print
routine.
Press enter when Select_list is selected to
setup print list.
Use Up and Down arrow keys to view list
status.
Press the Corresponding function key to the
items that you wish to add or remove from the
list.
Items marked with Yes will be added to the
list,
items marked with No will be removed from
the list.
The Select Print List Information display
shows the current possible Datalog size.
6.4.23
ADMINISTRATIVE
SETUP
Advance To
ADMINISTRATIVE SETUP
ADMINISTRATIVE SETUP
ENTER
OPERATOR PASSWORD
*****
ENTER
SUPERVISOR PASSWORD
*****
ENTER
SOFTWARE VERSION
vxx.xx
ENTER
Press Enter to begin Administrative Setup.
Enter Operator Password. (Factory Set to 0)
Enter Supervisor Password.
(Factory Set to 2000)
This display is used to show the software
version of the installed software.
PRODUCT ORDER CODE
MS716ST1xxxxxxx
ENTER
UNIT SERIAL NUMBER
00000
ENTER
SENSOR SERIAL NUMBER
00000
ENTER
Advance To
SETUP NETWORK CARD
This display is used to show the product order
code (model number).
This display is used to show the unit's serial
number.
This display is used to show the sensor's
serial number.
36
SUPERtrol-I MS-716 Flow Computer
6.4.24
SETUP
NETWORK CARD
(optional)
SETUP NETWORK CARD
ENTER
SELECT NTW PROTOCOL
ModbusRTU
ENTER
NETWORK DEVICE ID
1
ENTER
BAUD RATE
2400 4800 9600 19200
ENTER
PARITY
None Odd Even
ENTER
Press Enter to setup Network Card
Select desired Network Protocol.
Enter the device address on network (00-
255).
Select the desired Baud Rate.
Select the desired Parity.
Advance To
SELECT EZ SETUP?
37
7. Principle Of Operation
SUPERtrol-I MS-716 Flow Computer
General
Operation
Orifice Flowmeter
Considerations
7.1 General:
The SUPERtrol-I MS-716 Flow Computer uses several internal calculations to compute
the compensated flow based on specific data input. Several computations are performed
to arrive at the uncompensated flow, temperature, density and viscosity. This information
is then used to compute the Corrected Volume Flow or Mass Flow.
7.2 Orifice Flowmeter Considerations:
Head class flowmeters are supplied by the manufacturers with a 4-20 mA output span
which is already in flow units. The SUPERtrol-I MS-716 permits the user to enter this
flowmeter information directly. However, closely associated with this information is the
density that was assumed during flowmeter calibration. This information must also be
input if the user is to obtain maximum accuracy.
It is assumed that the user has the printout from a standardized orifice sizing program
for the particular device he will be using. Such standardized printouts list all the
necessary information which the user will then be prompted for.
Several specialized flow equations are listed that are not intended for the standard
unit but to be offered to appropriate OEMs or as special order items. These are
designated by a “†”.
Flow Equations
Note concerning Fluid Information
The user will be prompted for Fluid Information during the setup of the instrument.
The Factory will be preparing application information for several common fluid types.
† These special flow equations are not available in the standard unit. They are to be
offered to OEMs or as special order items.
39
Flow Equations
SUPERtrol-I MS-716 Flow Computer
7.3 Flow Equations: (Continued)
The above information was obtained from "Flow Measurement Engineering Handbook, 3rd
Edition" by Richard W Miller.
40
7.4 Calculating the Expansion Factor
√
SUPERtrol-I MS-716 Flow Computer
Calculating
Expansion Factor
The liquid density is a function of the flowing temperature for many fluids. This unit
solves an equation which represents this physical property of the fluid.
The information which the unit uses to describe the fluid is entered by the user in
the following variables: Reference Temperature, Reference Density, Expansion
Factor.
This information is available for many fluids in one or more of the following forms:
Fluid Specific Gravity vs. Temp. Table
Specific Gravity vs. Temp. Graph
Fluid Density vs. Temp. Table
Fluid Density vs. Temp. Graph
Begin by obtaining one of the fluid properties for the fluid you are using from
available manufacturers information or Engineering Handbooks. In some cases this
information is listed on the Material Safety Data Sheet for the fluid.
Two temperature-specific gravity pairs will be required to compute the temperature
coefficient.
The reference temperature is simply chosen by the user. Common reference
temperatures are 60° F or 15° C.
However, for cryogenic fluids, the normal boiling point may also be used. In some
cases the fluid data may list properties at 100° F, this temperature may also be used
as the reference temperature.
The reference temperature should be chosen so that it is in the application
temperature range. i.e. application temperature range -10 to 120° F, reference
temperature of 60° F chosen.
Enter the reference temperature you have chosen at this point.
The reference specific gravity corresponds to the fluid SPECIFIC GRAVITY at the
reference temperature chosen.
You may convert the fluid density information to specific gravity if it is in units other
than specific gravity. Use EQ1.
Expansion Factor
Equations
EQ1.
Spec.Grav. = Density of Fluid / Density of Water
Given the reference temperature, reference specific gravity, a second temp. and a
second Spec.Grav., the Expansion Factor (C Factor) can be computed as follows:
EQ2.Used for Liquid Mass and Corrected Volume Equations
1 - (Spec.Grav.2 / Ref.Spec.Grav.)
[]
Given the reference temperature, reference density, a second temp. and a second
density, the Expansion Factor (C Factor) can be computed as follows:
EQ3.Used for Liquid Mass and Corrected Volume Equations
[]
Temp.2 - Ref.Temp
√
1 - (Dens.2 / Ref.Dens.)
Temp.2 - Ref.Temp
x 1,000,000C =
x 1,000,000C =
41
SUPERtrol-I MS-716 Flow Computer
7.5 Computation of Viscosity Coef. A and B
Computation of
Viscosity
Coef. A & B
The flow computer solves an equation which computes the viscosity as a
function of temperature. Two parameters must be entered for this
calculation to be performed. These are the setup parameters Viscosity
Coef. A and Viscosity Coef. B. A table listing these values for common
fluids is available from KEP.
Alternately, if your intended fluid is not listed, the Viscosity Coef. A and B
can be derived from two known temperature/viscosity pairs. Begin by
obtaining this information for you intended fluid. Convert these known
points to units of Degrees F and centipoise (cP)
The information is now in a suitable form to compute the Viscosity Coef. A
and Viscosity Coef. B using the following equation based on the fluid state.
For a liquid, A and B are computed as follows:
(T1 + 459.67) • (T2 + 459.67) • ln [ cP1/cP2]
B =——————————————————————
(T2 + 459.67) - (T1 + 459.67)
cP1
A =———————————
exp [ B / ( T1 + 459.67) ]
cP • Density of Water at 4°C
NOTE: cS =——————————————
Density of Liquid
42
7.6 Linearization Table
X
SUPERtrol-I MS-716 Flow Computer
Linearization
Table
General
Information
Linearization
Table
(Pulse Inputs)
7.6.1 Linearization Table General Information
The Linearization Table is used when the flow input device gives a nonlinear
input signal. The unit uses up to 16 different points, as entered by the
operator, to form a curve for linearizing the input signal.
Notes:
1) A minimum of three points must be set up.
2) If "0" is entered for the frequency of any point other than point 1, the Flow
Computer assumes there are no more points above the points that preceded
them. The display will advance to the next setup prompt. Extrapolation is
taken from the last two nonzero points.
3) If the input frequency is above the highest or below the lowest frequency
programmed, the unit will use the last known point for the K factor in
computing the resulting actual flow.
4) Frequencies or apparent flows should be entered in ascending order.
7.6.2 Linearization Table for Pulse Inputs
The linearization table for pulse inputs programming is quite simple when
values of frequency and flow are known. The Flow Computer asks for 16
different frequencies (Freq) and 16 corresponding K factors (K). It then uses
this data to determine what the actual flow is for any given input frequency.
Usually the necessary data is provided with the flowmeter.
Linearization
Table
(Analog Inputs)
Linearization
Table
Interpolation
7.6.3 Linearization Table for Analog Inputs
The Linearization Table for Analog inputs programming is similar to the Pulse
input setup. The Flow Computer asks for 16 different flow rates (apparent
flow) and 16 corresponding Correction Factors. It then uses this data to
determine what the Actual flow is for any given apparent input signal. Again, a
minimum of three points must be set up.
Correction factor =
The same rules that applied for the Digital setup apply for the Analog setup as
well.
The Flow Computer prompts you for the Apparent input signal (APR) and a
correction factor CFr) to multiply it by to yield true actual flow.
7.6.4 Linearization Table Interpolation
The Linearization Table routine uses the entered data to determine the K
factor for any given input frequency or input flow signal. This is done by taking
the closest data points above and below the input signal, then using those
points to extrapolate the K factor (correction factor), then calculating the
uncompensated flow from the data. Below are the formulas.
Actual Flow
Apparent Input Flow
Parameters:
Determine closest point above input signal
signal = X, K factor (correction factor) = KA
Determine closest point below input signal
signal = Y, K factor (correction factor) = KB
Let input signal = H,
unknown K factor (correction factor) = KN
To find KN use this formula:
H - Y
X - Y
x (KA - KB) + KB = KN
43
KA
KN
KB
K factor
Y H
Input
8. Test, Service and Maintenance
8.1 Test Menus
SUPERtrol-I MS-716 Flow Computer
MenusDisplay
8.1.1
TOP LEVEL
TEST MENUS
START
SELECT OPERATE STATE
Run Setup Test
ENTER
Audit Trail
STOP
START
Error history
STOP
START
Print System Setup
STOP
START
Keypad Test
STOP
START
Notes
Select Test to enter the instrument
test & calibration routine.
NOTE: Supervisor (Service) password
required to gain access to this mode.
Refer to Page 40 for Details.
Refer to Page 40 for Details.
Refer to Page 40 for Details.
Refer to Page 41 Details.
Display test
STOP
START
Calibrate
STOP
START
Analog In Test
STOP
START
Pulse input test
STOP
START
Analog out test
STOP
START
Excitation out test
STOP
START
Pulse out test
Refer to Page 41 for Details.
Refer to Pages 42 - 46 for Details.
Refer to Page 46 Details.
Refer to Page 47 for Details.
Refer to Page 47 for Details.
Refer to Page 47 for Details.
Refer to Page 48 for Details.
STOP
STOP
START
Relay Test
STOP
START
Control inputs test
STOP
START
Battery Voltage Test
STOP
START
Data logger utility
Refer to Page 48 for Details.
Refer to Page 48 for Details.
Refer to Page 49 for Details.
Refer to Page 49 for Details.
44
8.2 Test Sub-Menus
SUPERtrol-I MS-716 Flow Computer
DisplayNotesSub-menus
8.2.1
Audit Trail
Submenu Group
8.2.2
Error History
Submenu Group
Audit Trail
ENTER
Audit Trail nnnnn
hh:mm:ss mm/dd/yy
MENU
Audit Trail
Error history
ENTER
Error history
Flow rate alarm low
MENU
Press Enter to view the audit trail information.
The audit trail is viewed in this format:
nnnnn= number of critical menu changes,
hh:mm:ss; mm/dd/yy = time and date of last
change.
Press Menu to get back to audit trail top-level
menu.
Press Enter to view error history.
NOTE: Press Print Key to print Error History. Printout
will include time/date of each errors first occurrence.
Press Up/Down arrow keys to scroll through
error message history. Press CLEAR to clear
entire error log.
8.2.3
Print System Setup
Submenu Group
Error history
Print System Setup
ENTER
Print System Setup
Press ENTER to print
ENTER
Print System Setup
—— Printing ——-
MENU
Print System Setup
Press Menu to get back to error history
top=level menu.
Press enter key to enter print system setup
submenu
Press enter to begin printing the system
setup.
This message will display as the data
transmission takes place.
Press Menu to get back to print system setup
top-level menu.
45
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.4
Keypad test
Submenu Group
8.2.5
Display test
Submenu Group
Keypad test
ENTER
Keypad test
Key pressed—> ENTER
MENU
Keypad test
Display test
ENTER
00000000000000000000
00000000000000000000
MENU
Press Enter to enter keypad test
Press the various keys and the display will
show the key that was pressed. Press Menu
to exit the test
Press Menu to get back to Keypad test toplevel menu.
Press Enter to enter display test.
Upon pressing enter the each digit on the
display will scroll 0-9 then A-Z. Press menu to
exit the test.
Display test
Press Menu to get back to Display test toplevel menu.
46
SUPERtrol-I MS-716 Flow Computer
ALL UNITS ARE CALIBRATED AT THE FACTORY PRIOR TO SHIPMENT
CAUTION:
This unit must be calibrated using precision and calibrated equipment.
Equipment needed is as follows: Frequency Generator, Digital Multimeter,
Precision Current/Voltage Source, Oscilloscope, Frequency Counter.
Sub-menusDisplayNotes
Calibration
Submenu Group
8.2.6
Calibrate CH1 0mA
Submenu Group
8.2.7
Calibrate CH1 20mA
Submenu Group
Calibrate
ENTER
Calibrate ch1 0mA
Iin=TB1-3 GND=TB1-4
ENTER
Calibrate ch1 0mA
CALIBRATING ——
Calibrate ch1 0mA
*** DONE ***
Calibrate ch1 0mA
Iin=TB1-3 GND=TB1-4
STOP
Calibrate ch1 20mA
Iin=TB1-3 GND=TB1-4
ENTER
Press Enter to begin the calibration routine.
(Please note the caution above)
Connect Current Source (+) TB1-3, (-) TB1-4.
Input 0mA and press Enter.
This message is displayed during calibration.
This message is displayed when the 0mA
calibration is finished.
The display will automatically return to the
Calibrate CH1 0mA submenu. Press the
Down arrow key to advance to the CH1 20mA
calibration.
Connect Current Source (+) TB1-3, (-) TB1-4.
Input 20mA and press Enter.
Calibrate ch1 20mA
0 CALIBRATING ——
Calibrate ch1 20mA
*** DONE ***
Calibrate ch1 20mA
Iin=TB1-3 GND=TB1-4
STOP
Advance to
Calibrate ch2 0mA
This message is displayed during calibration.
This message is displayed when the 20mA
calibration is finished.
The display will automatically return to the
Calibrate CH1 20mA submenu. Press the
Down arrow key to advance to the CH2 0mA
calibration.
47
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.8
Calibrate CH2 0mA
Submenu Group
8.2.9
Calibrate CH2 20mA
Submenu Group
Calibrate ch2 0mA
Iin=TB1-8 GND=TB1-4
ENTER
Calibrate ch2 0mA
0 CALIBRATING ——
Calibrate ch2 0mA
*** DONE ***
Calibrate ch2 0mA
Iin=TB1-8 GND=TB1-4
STOP
Calibrate ch2 20mA
Iin=TB1-8 GND=TB1-4
ENTER
To Calibrate: Connect Current Source (+)
TB1-8, (-) TB1-4. Input 0mA and press Enter.
This message is displayed during calibration.
This message is displayed when the 0mA
calibration is finished.
The display will automatically return to the
Calibrate CH2 0mA submenu. Press the
Down arrow key to advance to the CH2 20mA
calibration.
To Calibrate: Connect Current Source (+)
TB1-8, (-) TB1-4. Input 20mA and press
Enter.
Calibrate ch2 20mA
0 CALIBRATING ——
Calibrate ch2 20mA
*** DONE ***
Calibrate ch2 20mA
Iin=TB1-8 GND=TB1-4
STOP
Advance to
Calibrate ch1 0V
This message is displayed during calibration.
This message is displayed when the 20mA
calibration is finished.
The display will automatically return to the
Calibrate CH2 20mA submenu. Press the
Down arrow key to advance to the CH1 0V
calibration.
48
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.10
Calibrate CH1 0V
Submenu Group
8.2.11
Calibrate CH1 10V
Submenu Group
Calibrate ch1 0V
Vin=TB1-2 GND=TB1-4
ENTER
Calibrate ch1 0V
0 CALIBRATING ——
Calibrate ch1 0V
*** DONE ***
Calibrate ch1 0V
Iin=TB1-2 GND=TB1-4
STOP
Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4
ENTER
To Calibrate: Connect Voltage Source (+)
TB1-2, (-) TB1-4. Input 0V and press Enter.
This message is displayed during calibration.
This message is displayed when the 0V
calibration is finished.
The display will automatically return to the
Calibrate CH1 0V submenu. Press the Down
arrow key to advance to the CH1 10V
calibration.
To Calibrate: Connect Voltage Source (+)
TB1-2, (-) TB1-4. Input 10V and press Enter.
Calibrate ch1 10V
0 CALIBRATING ——
Calibrate ch1 10V
*** DONE ***
Calibrate ch1 10V
Iin=TB1-2 GND=TB1-4
STOP
Advance to
Calibrate ch2 0V
This message is displayed during calibration.
This message is displayed when the 10V
calibration is finished.
The display will automatically return to the
Calibrate CH1 10V submenu. Press the
Down arrow key to advance to the CH2 0V
calibration.
49
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.12
Calibrate CH2 0V
Submenu Group
8.2.13
Calibrate CH2 10V
Submenu Group
Calibrate ch2 0V
Vin=TB1-5 GND=TB1-4
ENTER
Calibrate ch2 0V
0 CALIBRATING ——
Calibrate ch2 0V
*** DONE ***
Calibrate ch2 0V
Iin=TB1-5 GND=TB1-4
STOP
Calibrate ch2 10V
Iin=TB1-5 GND=TB1-4
ENTER
To Calibrate: Connect Voltage Source (+)
TB1-5, (-) TB1-4. Input 0V and press Enter.
This message is displayed during calibration.
This message is displayed when the 0V
calibration is finished.
The display will automatically return to the
Calibrate CH2 0V top-level menu. Press the
Down arrow key to advance to the CH2 10V
calibration.
To Calibrate: Connect Voltage Source (+)
TB1-5, (-) TB1-4. Input 10V and press Enter.
8.2.14
Calibrate 100 ohm
RTD
Submenu Group
Calibrate ch2 10V
0 CALIBRATING ——
Calibrate ch2 10V
*** DONE ***
Calibrate ch2 10V
Iin=TB1-5 GND=TB1-4
STOP
Advance to
Calibrate 100ohm RTD
Calibrate 100ohm RTD
JMP TB1-6,7 100R=7,8
ENTER
Calibrate 100ohm RTD
0 CALIBRATING ——
This message is displayed during calibration.
This message is displayed when the 10V
calibration is finished.
The display will automatically return to the
Calibrate CH2 10V top-level menu. Press the
Down arrow key to advance to the 100 ohm
RTD calibration.
To Calibrate: Connect a jumper wire between
TB1-6 and TB1-7, Place a 100 ohm 0.1%
resistor between TB1-7 and TB1-8. Press
enter to calibrate.
This message is displayed during calibration.
Calibrate 100ohm RTD
*** DONE ***
Calibrate 100ohm RTD
JMP TB1-6,7 100R=7,8
STOP
Advance to
Calibrate 4mA out
This message is displayed when the RTD
calibration is finished.
The display will automatically return to the
Calibrate 100 ohm RTD top-level menu.
Press the Down arrow key to advance to the
4mA out calibration.
50
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.15
Calibrate 4mA Out
Submenu Group
8.2.16
Calibrate 20mA Out
Submenu Group
Calibrate 0mA out
+ TB1-15 - TB1-16
ENTER
Calibrate 0mA out
Enter mA: 0.00000
ENTER
Calibrate 0mA out
+ TB1-15 - TB1-16
STOP
Calibrate 20mA out
+ TB1-15 - TB1-16
ENTER
Calibrate 20mA out
Enter mA: 20.00000
ENTER
Connect ammeter to (+) TB1-15, (-) TB1-16.
Press enter.
To trim 0mA output: Press CLEAR to enable
editing and enter a small negative number (i.e. -
0.100) to force a display reading, then clear and
enter small quantity measured on your meter.
The display will return to Calibrate 0mA out.
Press the down arrow key to advance to Cal.
20mA out or repeat above if necessary.
Connect ammeter to (+) TB1-15, (-) TB1-16.
Press enter.
To trim 20mA output: Press CLEAR to enable
editing and enter the current reading that is on
the ammeter display. Press enter.
8.2.17
Analog In Test
Submenu Group
Calibrate 20mA out
+ TB1-15 - TB1-16
MENU
Calibrate
Analog In Test
ENTER
Analog In Test Volts
T2:00.000 T5:00.000
STOP
Analog In Test mA
T3:00.000 T8:00.000
STOP
The display will automatically return to the
Calibrate 20mA out submenu. Calibration is
complete.
Press the Menu key to go back to Calibrate
top-level menu.
Press enter to test the analog inputs.
To check voltage input accuracy: Use TB1-4
as Reference Ground, input 0-10 Volts to TB12 and/or TB1-5. Display should show voltage
being input. Use voltage meter to verify input.
To check current input accuracy: Use TB1-4
as Reference Ground, input 0-20mA to TB1-3
and/or TB1-8. Display should show current
being input. Use ammeter to verify input.
Analog In Test OHMS
RTD 00.000
MENU
Analog In Test
51
To check RTD input accuracy: Connect a
jumper wire between TB1-6 and TB1-7, Place
a 100 ohm 0.1% resistor between TB1-7 and
TB1-8. Display should show 100 ohms ±0.1%.
Press Menu key to return to Analog In Test
top-level menu.
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.18
Pulse input test
Submenu Group
2.5V
START
10mV
100mV
40Hz
3KHz
20kHz
STOP
START
STOP
Pulse input test
ENTER
Pulse input test
Trigger level 2.5V
ENTER
Pulse input test
count speed 3kHz
ENTER
Pulse input test
F1: 0 F2: 0
MENU
Pulse input test
Press Enter key to test the pulse input.
Use the Up/Down arrow keys to select the
appropriate trigger level.
Use the Up/Down arrow keys to select the
appropriate frequency range.
To check Pulse input accuracy: Use TB1-4 as
reference ground, input a frequency on TB1-2.
The display should show frequency being
input. Use a frequency counter to verify input.
Press Menu key to return to Pulse input test
top-level menu.
8.2.19
Analog out test
Submenu Group
8.2.20
Excitation out test
Submenu Group
Analog out test
ENTER
Analog out test
*0 4 10 15 20 mA
MENU
Analog out test
Excitation out test
ENTER
Excitation out test
*5v 12v 24v
MENU
Press Enter to test the analog output.
To simulate analog output: Connect an
ammeter to (+) TB1-15, (-) TB1-16. Press the
key under the desired setting to move the
asterisk (*). The unit should output the
selected current.
Press Menu key to return to Analog out test
top-level menu.
Press Enter to test the excitation output.
To test the excitation output: Connect a
voltmeter to (+) TB1-1, (-) TB1-4. Press the
key under the desired setting to move the
asterisk (*). The unit should output the
selected voltage.
Excitation out test
Press Menu key to return to Excitation out test
top-level menu.
52
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.21
Pulse out test
Submenu Group
8.2.22
Relay test
Submenu Group
Pulse out test
ENTER
Pulse out test
*0Hz 1Hz 10Hz 20Hz
MENU
Pulse out test
Relay Test
ENTER
Rly1 Rly2 Rly3 Rly4
Off Off Off Off
MENU
Press Enter key to test the pulse output.
To simulate a frequency on the pulse output:
Connect a frequency counter to (+)TB1-13,
(-)TB1-14. Press the key under the desired
setting to move the asterisk (*). The unit
should output the selected frequency.
Press Menu key to return to Pulse out test
top-level menu.
Press Enter to test the relays.
To manually control the relay outputs: Press
the key under the desired relay to toggle the
relays On/Off. Use an ohmmeter to check the
relay contacts.
8.2.23
Control input test
Submenu Group
Relay Test
Control inputs test
ENTER
TB1-9 TB1-10 TB1-11
Off Off Off
MENU
Control inputs test
Press Menu key to return to Relay Test toplevel menu.
Press Enter to test the control inputs.
To check the control inputs: Use TB1-12 as
reference, input a DC signal to TB1-9, TB1-10
and/or TB1-11, The Display will show ON
when input is active, OFF when inactive.
Press Menu key to return to control input test
top-level menu.
53
SUPERtrol-I MS-716 Flow Computer
Sub-menusDisplayNotes
8.2.24
Battery Voltage
test
Submenu Group
8.2.25
Data logger utility
Submenu Group
Battery Voltage Test
ENTER
Battery Voltage Test
3.312 Volts
MENU
Battery Voltage Test
Data logger utility
ENTER
Data logger utility
Log 10 958 Max
STOP
Press Enter key to view the battery voltage.
The display will show the battery voltage.
Replace battery at 2.5 VDC or below.
Press Menu key to return to battery voltage test
top-level menu.
Press Enter to use data logger utility.
The displays shows the number of Data Logs.
Press the Down arrow key to advance to PRT
(print) or CLR (clear).
Data logger utility
Log 00001 PRT CLR
MENU
Data logger utility
Press PRINT key to output data logger logs to
printer, Press CLEAR key to clear the data
logger contents.
Press Menu key to return to Data logger utility
top-level menu.
54
SUPERtrol-I MS-716 Flow Computer
8.3 Internal Fuse Replacement
Instructions:
1. Make sure you follow proper E.S.D. Precautions. All persons performing this
replacement must follow proper grounding procedures.
2. Turn the power to the unit off.
3. Remove the seal on the lower panel cover and remove screw from the lower panel
cover (see fig. 1).
4. Locate the Fuse F1 (see fig. 2) and unscrew the fuse cap.
5. Replace the fuse with the recommended size.
6. Reassemble the case and install the lower panel cover
7. Turn the unit back on.
Fuse Specifications:
110 VAC Power:160mA 250V 5x20mm TIME LAG Wickman #1950160000 or equivalent
220 VAC Power: 80mA 250V 5x20mm TIME LAG Wickman #1950080000 or equivalent
12/24 VDC Power:800mA 250V 5x20mm TIME LAG Wickman #1980800000 or equivalent
fig. 1fig. 2
TOTAL
RATE2PRE 1
START
STOP
1
F14PRINT5GRAND
SCROLL7PRE 28F2
CLEAR
TIME
0 –
3
MENU
6
ENTER
9
HELP
•
Screw
Seal
(if used)
START
STOP
TOTAL
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
Fuse
3
MENU
6
ENTER
9
HELP
•
55
9. RS-232 Serial Port
9.1 RS-232 Port Description:
The SUPERtrol-I MS-716 has a general purpose RS-232 Port which may be used for any one of the
following purposes:
Transaction Printing
Data Logging
Remote Metering by Modem (optional)
Computer Communication Link
Configuration by Computer
Print System Setup
Print Calibration/Malfunction History
9.2 Instrument Setup by PC’s over Serial Port
A Diskette program is provided with the SUPERtrol-I MS-716 that enables the user to rapidly configure
the SUPERtrol-I MS-716 using a Personal Computer. Included on the diskette are common instrument
applications which may be used as a starting point for your application. This permits the user to have
an excellent starting point and helps speed the user through the instrument setup.
SUPERtrol-I MS-716 Flow Computer
9.3 Operation of Serial Communication Port with Printers
SUPERtrol-I MS-716’s RS-232 channel supports a number of operating modes. One of these modes is
intended to support operation with a printer in metering applications requiring transaction printing, data
logging and/or printing of calibration and maintenance reports.
For transaction printing, the user defines the items to be included in the printed document. The user
can also select what initiates the transaction print generated as part of the setup of the instrument. The
transaction document may be initiated via a front panel key depression, a remote contact closure, or
upon completion of a batch.
In data logging, the user defines the items to be included in each data log as a print list. The user can
also select when or how often he wishes a data log to be made. This is done during the setup of the
instrument as either a time of day or as a time interval between logging.
The system setup and maintenance report list all the instrument setup parameters and usage for the
current instrument configuration. In addition, the Audit trail information is presented as well as a status
report listing any observed malfunctions which have not been corrected.
The user initiates the printing of this report at a designated point in the menu by pressing the print key
on the front panel.
The user may specify a “next calibration date” for periodic maintenance. The unit will automatically
remind the user that calibration is scheduled during instrument power up and in some printed documents
after that date.
9.4 SUPERtrol-I MS-716 RS-232 Port Pinout
TOTAL
1
START
F14PRINT5GRAND
STOP
SCROLL7PRE 28F2
CLEAR
RS-232
1 2 3 4 5 6
RATE2PRE 1
TIME
0 –
3
MENU
6
ENTER
9
HELP
•
1 Transmit
2 Receive
3 Ground
4 MP.
5 DCD
6 Do Not Use
56
10. RS-485 Serial Port (optional)
TOTAL
1
START
STOP
RATE2PRE 1
3
F14PRINT5GRAND
6
SCROLL7PRE 28F2
9
CLEAR
TIME
0 –
HELP
•
MENU
ENTER
1 2 3 4
10.1 RS-485 Port Description:
The SUPERtrol-I MS-716 has a an optional general purpose RS-485 Port which may
be used for any one of the following purposes:
The optional RS-485 card utilizes Modbus RTU protocol to access a variety of process parameters and totalizers. In addition, action routines can be executed. For
further information, contact factory and request RS-485 Protocol manual for SUPERtrol-I MS-716.
10.3 Operation of Serial Communication Port with PC
The ow computer's RS-485 channel supports a number of Modbus RTU commands. Refer to port pinout (below) for wiring details. Modbus RTU drivers are
available from third party sources for a variety of Man Machine Interface software for
IBM compatible PC's.
The user reads and writes information from/to the RS-485 using the Modbus RTU
commands. The SUPERtrol-I MS-716 then responds to these information and command requests.
Process variables and totalizers are read in register pairs in oating point format.
Time and date are read as a series of integer register values. Alarms are individually
read as coils. Action routines are initiated by writing to coils.
10.4 SUPERtrol-I MS-716 RS-485 Port Pinout
RS-485
1 TX/RX (+)
2 Terminating Resistor (180 Ω)
57
3 TX/RX (-)
4 Common
11. Flow Computer Setup Software
The SUPERtrol-I MS-716 setup program provides for configuring, monitoring and
controlling a SUPERtrol-I MS-716 unit.
Sample applications are stored in disk files. The setup program calls these
Templates
SUPERtrol-I MS-716 (
usage. Similarly you can load the setup in program memory from either a disk file
(
Opening
The program can monitor outputs from the unit while it is running.
The program can reset alarms and totalizers.
For assistance there are mini-helps at the bottom of each screen in the program.
There is also context sensitive help available for each screen accessible by pressing
the F1 key.
11.1 System Requirements:
. You can store the setup from the program’s memory to either the
Downloading
a file) or from the SUPERtrol-I MS-716 unit (Up
the file) or to a disk file (
SUPERtrol-I MS-716 Flow Computer
Saving
loading
the file) for later
a file).
IBM PC or compatible with 386 or higher class microprocessor
4 MB RAM
3 MB free disk space
VGA or higher color monitor at 640 x 480
Microsoft® Windows™ 3.1 or 3.11 or Windows 95™
Communication Port - RS-232
RS-232 Cable
11.2 Cable and Wiring Requirements:
The serial communication port on your PC is either a 25 pin or 9 pin connector. No
cabling is supplied with the setup software. A cable must be purchased separately or
made by the user. It is recommended to purchase a modem cable which matches
the available communication port on you PC and a 9 pin male connection for the
SUPERtrol-I MS-716 serial port.
11.3 Installation for Windows™ 3.1 or 3.11
The Setup Software includes an installation program which copies the software to
your hard drive.
Insert Setup Disk 1 in a floppy drive.
In the Program Manager, click File, and then select Run.
NOTE:For Windows 95™ Click the Start button, select Run and proceed
as follows:
Type the floppy drive letter followed by a colon (:) and a backslash (\), and the word
setup. For Example:
a:\setup
Follow the instructions on your screen.
58
11.4 Using the Flow Computer Setup Software
The setup software window consists of several menu “Tabs”. Each tab is organized
into groups containing various configuration and/or monitoring functions. To view
the tab windows, simply click on the tab. The previous tab window will be hidden as
the new tab window is brought to the foreground.
11.5 File Tab
The File Tab has three sections. Any of the options on this tab can also be
accessed from the File submenu.
SUPERtrol-I MS-716 Flow Computer
The Template Section provides for opening and saving templates. The
Save As
buttons provide the standard Windows functionality for dealing with files.
Save
and
The Load button is used to open existing templates.
only
There are two additional menu items available
from the File menu: Create new
file and Templates.
The
Create new file
, option allows for creating custom templates using the existing
template in memory as the starting point. Assign a new name for this template. The
template will be saved under this new name.
The
Template
option will bring up a list of predefined templates that can be loaded
into the program. These predefined templates are useful as a starting point when
defining custom templates.
A typical scenario using the setup program would be the following:
• Open up a predefined template from the supplied list
• Choose ‘Save As’ to save this to a new file name
• Proceed to customize the template by making any changes that are needed
• Save the template to disk (if you want to reuse this template)
• Download the template to an attached unit.
The Communications with SUPERtrol-I MS-716 Section allows the user to upload
a template file from the unit, download the program’s current template to the unit or
Compare the program’s current template with the unit.
The Print (report) Section allows the user to:
11.6 Setup Tab
The Setup tab is where majority of the SUPERtrol-I MS-716 instrument setup
modifications are done. The Setup tab is divided into five sections.
1. Configure the current Windows printer through the Select Printer option.
2. Print a Maintenance Report through the PC's printer using the Print
Maintenance option.
3. Print the current template through the PC's printer using Print Setup option.
System Section:Parameters, Display, Indicators
Input Section:Flow, Fluid, Compensations, Control Inputs
Output Section:Pulse, Currents
Relay Section:Relays
Other Settings Section:Administration, Communication, Printing
NOTE: Many setup items are enabled or disabled depending on previous setup
selections, It is important to work your way through the above list in the
order shown. Be sure to verify your selections when you are through
programming to insure that no settings were changed automatically.
59
11.7 View Tab
SUPERtrol-I MS-716 Flow Computer
The View Tab screen allows for viewing selected group items on the PC in a similar
format as shown on the unit display. Data from the following groups can be viewed
in the List of Values section:
Process Parameters (i.e. rate, temperature)
Totalizers (i.e. total, grand total)
The setup software assumes the current setup has been uploaded from the flow
computer into the PC. It is important that the setup program and the SUPERtrol-I
MS-716 unit are using the same setup information at all times or the data will be
inconsistent. It is best to upload or download the setup before using this feature.
To start the viewer, first check the boxes of items to view and then click the start
button. The data will appear in the appropriate sections and will be continuously
updated. The refresh rate is dependent on the number of items that are being
viewed and the baud rate of the connection. Data in the List of Values section can
be collapsed by clicking on the ‘minus’ sign in front of the group title. The data can
be expanded by clicking on the ‘plus’ sign in front of the group title. If a group is
collapsed and data in the group changes on refresh, the group will automatically
expand. Changing the view items requires stopping the current viewing, checking
the new selections and then restarting the viewer.
If communication errors occur while reading data from the SUPERtrol-I MS-716
device, the word ‘Error’ will appear in place of the actual value. If the connection to
the SUPERtrol-I MS-716 is lost, the viewer will time out with a message saying the
device is not responding.
The viewer will attempt to communicate with the SUPERtrol-I MS-716 device
matching the device ID set in the communications screen. If you are having trouble
establishing communication, compare settings for the PC and the flow computer.
Also verify the connections between the PC and flow computer.
11.8 Misc. Tab
This tab has three sections: Tools, Actions and Options.
The tools section contains various system administration activities such as creating/
modifying the initial sign-on screen or calibration, service test etc.
Create Sign-on, Create Print Header, Calibration, Service Test
The Actions section is used to send commands to the SUPERtrol-I MS-716 unit.
Reset Totalizers, Reset Alarms
The Options section has the following selections:
Linearization, PC Communication
Additional capabilities may be provided in the future.
NOTE: Future options appear as disabled buttons on the screen.
60
SUPERtrol-I MS-716 Flow Computer
12. Glossary Of Terms
Acknowledge & Clear Alarms
Acknowledge is used to clear alarm relays and remove any visual alarm messages from the display. In
the run mode, press the ENTER key or activate CONTROL INPUT 3 (if set for ACK) to momentarily
clear alarms and alarm messages. Alarms will reassert themselves if alarm conditions are still present.
Analog Output
The analog signal (4-20mA) that is generated by the SUPERtrol-I MS-716. It can correspond to the
Rate, Total, Temperature or Density. This output is used primarily for transmission of process
information to remote systems.
Audit Trail
The audit trail is used to track the number of changes made to the units setup program.
Batch Count Mode
Batch Count Mode specifies the user preference for count direction. The "Up" selection begins with a
value of "0" and counts up until the batch size is reached. The "Down" selection begins with a value
equal to the desired batch size and counts down to "0".
Batch Overrun
The SUPERtrol-I MS-716 offers a batch overrun compensation routine. If batch overrun occurs due to
slow valve response time, the unit will compensate for the overrun amount on the next batch. This
feature can be disabled if desired.
Batcher
An instrument which controls the dispensing of desired batch amounts. Liquid batching systems are
usually comprised of a batch controller (batcher), flowmeter and control valve. The batcher opens and
closes the valve through the use of relays and measures the amounts of liquid being dispensed via the
flowmeter.
Baud Rate
The speed of serial communication transmissions, expressed in bits per second.
C-Factor (Fluid Expansion Factor)
A parameter in a flow equation which is used to describe the relationship between density or volume and
temperature changes.
Corrected Volume Flow
The equivalently volume at a reference temperature condition which involves the measurement of liquid
volume flow using a flow sensor and temperature sensor to compensate for thermal expansion.
Custody Transfer
Weights and Measure metering codes often specify several requirements for instruments and
mechanisms to prevent and track changes in the setup of an instrument which may be used in the
commercial sale of goods. The SUPERtrol-I MS-716 tracks changes via the Audit Trail.
Data Logger
The capturing of information for later use and the mechanism for specifying the conditions where a
capture should be made.
DC Output / Excitation Voltage
An on-board DC power supply used to power peripheral sensors. The SUPERtrol-I MS-716 offers
excitation voltages of 5VDC, 12VDC or 24VDC when powered by AC voltage.
Default Value
The value to be used by the instrument if a sensor failure or out of ranch signal is detected.
Expansion Factor
See C-Factor
EZ Preset
The EZ Preset mode was designed for users who frequently change the batch amount.
Flow Alarm
A visual indication that the volumetric flowrate is above or below the flow alarm setpoint specified by
the user.
Flow Signal Timeout
The Flow Signal Timeout allows the user to enter a timeout of 0 to 99 seconds. If a batch is “Filling” and
zero flow persists for more than the user entered time then the batch will be aborted. This prevents over
flows due to faulty flow sensors and/or wiring.
61
SUPERtrol-I MS-716 Flow Computer
12. Glossary Of Terms (Continued)
Flow Equation
A flow control expression or algorithm describing a mathematical equation to be solved by a flow
computer in the desired application.
Follow, Alarm
Alarm relays which are non latching and whose output state is based solely on the comparison of the
current process value and the alarm setpoint (trip point).
Function Key
A key on a push-button panel or keyboard (whose function is described by the key label) used to
perform an instrument function or special routine.
Handshake
A means of controlling the information flow between two pieces of equipment to prevent the sending
device from transmitting information at a rate faster than what can be accepted by the receiver.
Hysteresis
The relay hysteresis is a "dead band" setting which allows the relay to remain energized for a given
amount below the setpoint. This is used to prevent relay chatter when the process value is near the
setpoint value.
Example: If the Preset is set at 100, and the hysteresis is set at 10, the relay will energize when the rate,
temp or dens. reaches 100, the relay will remain energized until the reading falls below 90.
Input Termination
Input signal lines on digital inputs often require pullup or pulldown resistor configurations to operate
properly with different sensor configurations. The SUPERtrol-I MS-716 contains such resistors and
may be enabled via the setup menu.
Inhibit Totalizer
"Inhibit Total" is a Control Input 1 setting that is used to stop the totalization. If enabled, a voltage level
on control input 1 will inhibit the total as long as the voltage is present. This feature is useful during
meter proving and in applications that provide a sensor to signal the flow computer when fluid is
present.
K-Factor
A scaling factor derived from the pulses produced by a flowmeter output, expressed in pulses per unit
(i.e. pulses/gallon)
LCD
Abbreviation for: Liquid Crystal Display
Limit Setpoint
An alarm trip point setting which specifies the value or magnitude of a process parameter necessary to
activate an alarm indicator or control relay.
Linear Flowmeter
A flow measurement device whose output is proportional to flow.
Linearization
The mathematical correction of a nonlinear device. The SUPERtrol-I MS-716 uses a linearization Table
which is made up of input/output values and makes interpolations of the table to arrive at a "linearized"
measurement.
LinTbl
Abbreviation for Linearization Table.
Low Flow Cutoff
A value set at which any flow measurements read below this value will be ignored.
Low Pass Filter
A low pass filter passes low input frequencies while blocking high frequencies. In the SUPERtrol-I MS716, this is the maximum input count speed to be encountered in an application. It is expressed in counts
per second (Hz).
Mass Flow
Mass Flow is inferred by the volumetric flow and density (or implied density) of a fluid.
62
SUPERtrol-I MS-716 Flow Computer
12. Glossary Of Terms (Continued)
Maximum Batch Preset
The Maximum Batch Preset allows the user to program the Maximum Batch value allowed to be entered
by the operator. If an operator should try to program a batch higher then this value, the unit will not allow
the value to be entered and will prompt the user with an error message saying that the Maximum Batch
Preset has been exceeded.
Maximum Drain Time
The unit declares that a batch is “done” when the flow rate equals “0”. A flow rate may be present long
after the Preset Relay de-energizes due to slow reacting valves or leaky valves. The Maximum Drain
Time allows the user to enter an amount of time (0 to 99 seconds) to wait before declaring “Batch Done”.
After the Preset Batch quantity is reached, the unit will declare “Batch Done” when the flow rate is “0” or
the Maximum Drain Time has expired. The batch data will then be available for printing and datalogging.
Max Window
The max. window time sets the maximum sample time (1 to 99 sec) for the ratemeter.
Modem Init Master
The "Modem Init Master" menu allows the user to select whether the unit will engage in a configuration
conversation with the modem on power up or impart no setup information to the modem and use it "as
is". For most users it is recommended to choose "yes" for "Modem Init Master".
Orifice Plate Flowmeter
A class of flow measurement devices where the measured signal (differential pressure) has a square law
relationship to flow.
Parity
A method for detecting errors in transmissions of serial communications data.
Preset
A set point used to trigger the relay outputs of the SUPERtrol-I MS-716.
Print Interval
The print interval allows the SUPERtrol-I MS-716 to transmit information to the serial port at selectable
time intervals.
Private Code
An operator password code which authorizes changes to the setup of the instrument but blocks access to
the Service/Calibration/Test mode. The private code also blocks the clearing of the Grand Total.
Process Parameters
Any sensor information which has been scaled to engineering units including Flow, Temperature and
Density.
Pulldown (Input Termination)
The termination of an input at which the input is pulled down to ground through a resistor. Inputs that
are terminated by this method need to be driven high with a positive voltage pulse.
Pullup (Input Termination)
The termination of an input at which the input is pulled up to a positive voltage through a resistor.
Inputs that are terminated by this method need to be pulled low with a sinking current or contact to
ground .
Pulse Output
The pulse output of the SUPERtrol-I MS-716 is available for remote accumulation of the total or sent to
peripheral devices, such as a PLC. The output can be scaled using the Pulse Output Scaling Constant.
Quad
Abbreviation for Quadrature. Quadrature signals are used for direction control. Two flowmeter signals
are output with a 90° phase shift. The counter counts UP when channel A precedes channel B, and
counts DOWN when Channel A lags Channel B.
Quick Setup
A utility that provides for rapid configuration of an instrument. The SUPERtrol-I MS-716 quick setup
provides the following:
1) Prompts the user for only critical information.
2) Automatically sets specifications to common uses.
After following the Quick Setup procedure, the unit will be operational to perform the basic
measurement. The setup can be further customized using the setup menus.
63
SUPERtrol-I MS-716 Flow Computer
12. Glossary Of Terms (Continued)
Quick Update %
This feature is used to disable the rate averaging filter when a significant change in the flow rate occurs.
The user can enter the percent of change needed to be detected to disable the averaging feature. This is
especially useful during start-up and shutdown of flow.
Rate Averaging Filter
The rate averaging filter is used to stabilize fluctuating rate displays. Higher settings provide more
averaging for a more stable display. Derived from the equation:
(OLD DATA x "Avg. Filter" + NEW DATA)
("Avg. Filter" + 1)
Ratemeter
Any device used to display the speed of a process. The ratemeter in the SUPERtrol-I MS-716 displays
flow rate.
Ref. Dens.
Abbreviation for Reference Density. This is the fluid density at reference temperature.
Ref. Temp.
Abbreviation for Reference Temperature. This represents the base or reference condition to which
corrected flow will be computed.
Reset/Start Control Input
In a batching system, a single operator activation of the START key or Control Input 1 will reset the
total then start the batch process.
Single_Pulse
The Single_Pulse setting is used for flowmeters with single pulse outputs.
Slow Start Quantity
The Slow Start Quantity is a function that allows an amount to be entered for a Slow Start up. This
function requires two stage valve control. RLY 1 (slow flow) will energize for Slow Start and RLY 2 (fast
flow) will energize after the Slow Start Quantity has been delivered. This helps reduce turbulence when
filling an empty container.
Sqrt
Abbreviation for Square Root Extraction. Used for flow elements using differential pressure
measurements.
Standard Preset
The Standard Preset mode should be used in applications where the batch amount does not change often.
Stop/Reset Control Input
In a batching system, a single operator activation of the STOP key or Control Input 2 will stop the batch
process then reset the total.
Time Constant
A damping factor for an averaging filter for the analog output. (see also Rate Averaging Filter)
Totalizer
Any device which accumulates and displays a total count.
UVC
Abbreviation for Universal Viscosity Curve. A presentation of the combined flowrate/viscosity
calibration for a turbine flowmeter.
VFD
Abbreviation for Vacuum Fluorescent Display
Visc Coef
Abbreviation for Viscosity Coefficient. One or more coefficients in an equation used to describe the
viscosity as a function of temperature for a fluid.
Volume Flow
The measurement of volumetric flow.
64
13. Diagnosis and Troubleshooting
13.1 Response of SUPERtrol-I MS-716 on Error or Alarm:
Error and warning indications which occur during operation are indicated
in the RUN mode alternately with the measured values. The SUPERtrolI MS-716 Flow Computer has three types of error:
TYPE OF ERRORDESCRIPTION
Sensor/Process AlarmsErrors detected due to sensor
Self Test ErrorsErrors detected during self test.
SUPERtrol-I MS-716 Flow Computer
failure or process alarm
conditions
System AlarmsErrors detected due to system
failure
Some alarms are self clearing. Other alarms require the user to
acknowledge and clear the alarm. Press the ENTER button to
acknowledged and clear alarms. Alarms may reassert themselves if
the alarm condition is still present.
NOTE: A historical error alarm log is available in the "Test Mode".
The following descriptions suggest possible causes and corrective
actions for each alarm message.
65
13.2 Diagnosis Flow Chart and Troubleshooting
All instruments undergo various stages of quality control during production. The last of these stages is a complete calibration carried out
on state-of-the-art calibration rigs.
A summary of possible causes is given below to help you identify
faults.
SUPERtrol-I MS-716 Flow Computer
Is there an input power
supply voltage across
Terminals 23 and 24?
Yes
Is the Display Backlight
Visible?
Yes
Are the Display
Characters Visible?
Yes
Is there a black bar
across the display?
No
No
No
No
Yes
Check the connections
according to the circuit
diagrams.
Check junction box fuses.
Check/Replace internal
fuse. If fuse is OK,
Factory Service Required.
The display may not be
visible with ambient
temperatures below -10 °C
Allow the instrument to
warm up.
Contact factory if
necessary
Check line voltage.
If voltage is OK,
Factory Service Required.
Does the display
alternate between blank
and sign on message?
No
Does the display show
an error message?
No
No system or process
errors present.
Yes
Yes
Check line voltage.
If voltage is OK,
Factory Service Required.
See section 12.3 for
cause and remedy.
66
13.3 Error & Warning Messages:
SUPERtrol-I MS-716 Flow Computer
13.3.1
Sensor/Process Alarms
Error/Warning Message
TOTALIZER ROLLOVER
AUX INPUT TOO LOW
RTD OUT OF RANGE
RATE OVERFLOW ERROR
PULSE OUT OVERFLOW
Cause
Displayed when totalizer
rolls over
4-20 mA Input current at aux
input smaller than 3.5 mA:
• Faulty Wiring
• Transmitter not set to "4-20
mA"
• Transmitter defective
Input current at RTD input
too low:
• Faulty wiring
• RTD defective
Pulse counter overflowed.
The totalizer may have lost
counts.
Calculated pulse frequency
too large:
• Pulse width setting too long
• Larger pulse scaler needed
Remedy
Acknowledge
Rollover,
Remedy not required
• Check wiring
• Check function of
sensor
• Check wiring
• Check function of
RTD sensor
• Report error to
factory
• Check application
conditions
• Check wiring
• Adjust pulse value
• Adjust pulse width
• Check process
conditions
FLOW RATE ALARM LOW
FLOW RATE ALARM HIGH
TEMP ALARM LOW
TEMP ALARM HIGH
DENSITY ALARM LOW
DENSITY ALARM HIGH
BATCH OVERRUN ALARM
MODEM NOT PRESENT
Limit value exceeded.
Batch size exceeded by
more than set limit.
The setup expects modem
usage and a modem is not
responding.
• Check application if
necessary
• Check limit value
• Adjust the limit
value if required
• Check valves in
system for proper
operation and/or
leaks
• Check limit value
• Adjust the limit
value if required
• Check setup for
proper baud rate,
parity, etc.
• Check modem
connection and
cycle power to
SUPERtrol
• Replace modem
SOFTWARE ERROR RESET
EXTENDED PFI LOCKUP
67
Watchdog Error. Transient
likely
Unit was operated with an
input power level lower than
safe operating range for an
extended period of time.
• Cycle power to
SUPERtrol
• Check data in unit.
Totalizer may have
inaccuracies
• Investigate
brownout cause.
13.3 Error & Warning Messages: (Continued)
SUPERtrol-I MS-716 Flow Computer
13.3.2
Self Test Alarms
Error/Warning Message
FLOW INPUT TOO HIGH
AUX INPUT TOO HIGH
FLOW INPUT TOO LOW
Cause
Analog input signal of the
flow input exceeded by more
than 3%:
• Sensor overranged
• Incorrect full scale setting
of flowmeter
• Function error in
transmitter or faulty wiring
Analog input signal of the
auxiliary input exceeded by
more than 3%:
• Sensor overranged
• Incorrect full scale setting
of transmitter
• Function error in
transmitter or faulty wiring
Analog input signal of the
flow input fell below the low
scale range by more than
3% of full scale value:
• Flowmeter not set to 4-20
mA
• Function error in
transmitter or faulty wiring
Remedy
• Check analog
signal range
• Check the
application
conditions
• Check wiring
• Check analog
signal range
• Check the
application
conditions
• Check wiring
• Check wiring
• Check calibration of
flowmeter
• Check function of
flowmeter
BATTERY LOW WARNING
A to D NOT CONVERTING
TIME CLOCK ERROR
CAL CHECKSUM ERROR
SETUP CHECKSUM ERROR
Battery voltage too low
Fault in analog/digital
converter
The correct time/date is no
longer shown
Calibration constants have
been corrupted
The units setup has been
corrupted
• Replace Battery
• Consult Factory for
service information
• Unit may self
correct, Press
ENTER to
acknowledge &
clear alarm
• If error reasserts,
factory service is
required
• Re-enter time and
date.
• If error occurs
again contact
factory
• Report error to
factory
• Report error to
factory
68
APPENDIX A
DECIMAL
TEMPERATURE
SCALE
TEMPERATURE
PLACES
HIGH FLOW
RATE ALARM
SUPERtrol-I MS-716 Flow Computer
QUANTITY
SLOW START
TIME
MAX. DRAIN
TIMEOUT
FLOW SIGNAL
SETUP MENUS
COMP.
BATCH
OVERRUN
TEMP
DESCRIPTOR
QUICK
UPDATE %
RATE
FILTER
AVERAGE
PLACES
RATE DECIMAL
RATE
DESCRIPTOR
BASE
RATE TIME
PLACES
DENSITY
DECIMAL
MASS UNITS
DENSITY
DESCRIPTOR
LOW FLOW
RATE ALARM
TABLE B
CHANGE
TABLE A
CHANGE
AVERAGE
KB-FACTOR
AVERAGE
KA-FACTOR
TYPE
MAX WINDOW K-FACTOR
INPUT
TERMINATION
FILTER
LOW PASS
HIGH FLOW
RATE ALARM
LOW FLOW
RATE ALARM
CUTOFF
LOW FLOW
TABLE A
CHANGE
SCALE
FLOW HIGH
SCALE
FLOW LOW
TYPE
LINEARIZATION
DENS
METHOD
EXTRACT
ALARM
AUX HIGH
ALARM
AUX LOW
AUX. DEFAULTAUX LOW
SCALE
AUX FULL
AT 4DEGC
H2O DENSITY
COEF. B
VISCOSITY
COEF. A
VISCOSITY
These functions will only appear with
appropriate settings in other functions.
RELAY
HYSTERESIS
RELAY
DAMPING
ANALOG OUT
FULL SCALE
ANALOG OUT
SETPOINT
RELAY MODE
ENTER DATESELECT
2MIN.
INACTIVE
HANGUP IF
REDIALS
NUMBER OF
ALARM
ERROR/
CALL ON
TIME
CALL OUT
CALL OUT NO
ANSWER
MODEM AUTO
MODEM
CONTROL
FEED
ITEMS
PRINT LIST
IF PRINT
CLEAR TOTAL
BATCH
PRINT END OF
KEY
ENABLE PRINT
PRINT
INTERVAL
PRINT TIME
ONLY
DATALOG
SERIAL
SENSOR
NUMBER
UNIT
SERIAL
NUMBER
PRESET
MAX. BATCH
MODE
BATCH COUNT
SELECT
PRESET TYPE
TYPE
INSTRUMENT
SELECT
EZ SETUP
TYPE
INSTRUMENT
START HERE
TOTAL
PLACES
DECIMAL
UNITS
VOLUME
TOTAL
DESCRIPTOR
SETUP
EQUATION
SELECT FLOW
INDICATORS
EQUATION
SELECT FLOW
SETUP INDICATORS
TYPE
PULSE
PULSE INPUT
RANGE
CURRENT
INPUT TYPE
SCALE
RANGE
VOLTAGE/
CURRENT
TYPE
AUX SIGNAL
TYPE
AUX INPUT
SETUP AUX INPUT
TRIGGER
VOLTAGE/
TYPE
ANALOG
TYPE
FLOW
INPUT
VOLTAGE
EXCITATION
SETUP FLOW INPUT
DENSITY
CALIBRATION
FACTOR
EXPANSION
PULSE WIDTH PULSE VALUE
TEMPERATURE
REF. DENSITYREF.
SET FLUID
USAGE
PULSE OUPUT
PROPERTIES
SETUP PULSE
LOW SCALE
ANALOG OUT
OUTPUT
ANALOG
FLOW TYPE
ANALOG OUT
ANALOG
OUPUT USAGE
OUTPUT
SETUP ANALOG
RELAY
DURATION
RELAY
DELAY
RANGE
RELAY USAGE
SETUP
RELAYS
OUTPUT
SETUP RELAYS
USAGE
INPUT 3
CONTROL
USAGE
INPUT 2
CONTROL
USAGE
INPUT 1
CONTROL
1, 2, 3
SETUP
INPUTS
1, 2, 3, 4
CONTROL
INPUTS
SETUP CONTROL
TIME OF DAY
CLOCK AM/PM
TIME CLOCK
DEVICE IDBAUD RATEPARITYHANDSHAKE DEVICE LINE
SERIAL
CLOCK TYPE
SETUP REAL
CLOCK
SETUP REAL TIME
PAGE LENGTH TOP MARGIN
OUTPUT
FORMAT
SETUP
HARDWARE
DATALOG/
SETUP
SERIAL USAGE
PRODUCT
ORDER CODE
VERSION
SOFTWARE
PASSWORD
SUPERVISOR
PRINT
OPERATOR
PASSWORD
SETUP
DATALOG/PRINT
ADMINISTRATIVE
BAUD RATEPARITY
DEVICE ID
NETWORK
SELECT
NETWORK
PROTOCOL
CARD
SETUP NETWORK
69
WARRANTY
This product is warranted against
defects in materials and workmanship for a period of two (2) years from
the date of shipment to Buyer.
The Warranty is limited to repair or
replacement of the defective unit at
the option of the manufacturer. This
warranty is void if the product has
been altered, misused, dismantled,
or otherwise abused.
SUPERtrol-I MS-716 Flow Computer
Ordering Information
Example MS716L1A0V ETMB
Series:
MS716 = ST1 Special
Display Type:
L= LCD
Input Type:
1= 110 VAC
2= 220 VAC
3= 12 VDC (10 to 14 VDC)
4= 24 VDC (14 to 28 VDC)
Relays:
A= 2 SPDT Relays
B= 4 SPDT Relays
Network Card:
0= None (STD)
2= RS485/Modbus
Mounting:
V= Vehical Mount, Skid Mount, Field Mount
ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, ARE EXCLUDED, INCLUDING BUT NOT
LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE.
Options:
ET= Extended Tempertaure
-4°F to 131°F (-20°C to 55°C)
UL= UL Class 1 Division 2
MB= Aluminum Mounting Brackets (2) (Recommended)
IM = IM-2400 Internal Modem
M = Modem Power Option
TB= RS485 Terminal Block
for Panel Mount Enclosure
Accessories:
KEPS-KEP1-32
KEP RS232 for SUPERtrol 1, SUPERtrol 1LE,
SUPERtrol 2 and LEVELtrol 2 • 32 Bit OPC/DDE
Server
KEPS-MBS32
Supports RS485 for ST1, ST1LE, ST2,
LT2, MRT, DRT & MB2 (Modbus RTU)
Modem Available, see MPP-2400 and MPP-2400N
Serial printer available, see P1000, P295
Ethernet Port Server available, see IEPS
RS-422/485 to RS-232 Communication Adaptor available, see
CA285
Remote metering and data collection software available, see
TROLlink
Kessler-Ellis Products
10 Industrial Way East
Eatontown, NJ 07724
Toll Free: 800-631-2165
Phone: (732) 935-1320
Fax: (732) 935-9344
http://www.kep.com
http://www.kepware.com
70
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.