Power, input and output (I/O) wiring must be in accordance with Class I, Division 2 wiring methods Article 501-4 (b) or
the National Electrical Code, NFPA 70 for installations in the U.S., or as specified in Section 18-1J2 of the Canadian
Electrical Code for installations within Canada and in accordance with the authority having juristiction.
WARNING!
1. This equipment is suitable for use in Class I, Division 2, Groups A, B, C, and D or
Non-Hazardous Locations Only.
2. WARNING - EXPLOSION HAZARD - Substituition of components may impair
suitability for Class I, Division 2.
3. WARNING - EXPLOSION HAZARD - Do not disconnect equipment unless power
has been switched off or the area is known to be Non-Hazardous.
!
This instrument contains electronic components that are susceptible to damage by static electricity. Proper handling*
procedures must be observed during the removal, installation, or handling of internal circuit boards or devices.
*Handling Procedure
1. Power to unit must be removed.
2. Personnel must be grounded, via wrist strap or other safe, suitable means, before any printed circuit board or other
internal device is installed, removed or adjusted.
3. Printed circuit boards must be transported in a conductive bag or other conductive container. Boards must not be
removed from protective enclosure until the immediate time of installation. Removed boards must be placed immediately
in protective container for transport, storage, or return to factory.
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic
designs contain components that utilize metal oxide technology (NMOS, CMOS, etc.). Experience has proven that even
small amounts of static electricity can damage or destroy these devices. Damaged components, even though they appear
!
The following instructions must be observed.
WARNING!
to function properly, may exhibit early failure.
SAFETY INSTRUCTIONS
• This instrument was designed and is checked in accordance with regulations in force EN 60950 (“Safety of
information technology equipment, including electrical business equipment”).
A hazardous situation may occur if this instrument is not used for its intended purpose or is used incorrectly.
Please note operating instructions provided in this manual.
• The instrument must be installed, operated and maintained by personnel who have been properly trained.
Personnel must read and understand this manual prior to installation and operation of the instrument.
• This instrument is internally fused. Replace the internal fuse with the following specified type and rating only:
Input PowerRecommended Fuse
115 VAC160 mA slow blow fuse
230 VAC80 mA slow blow fuse
12-24 VDC800 mA slow blow fuse
Disconnect power supply before replacing fuse!
• The manufacturer assumes no liability for damage caused by incorrect use of the instrument or for modifications
or changes made to the instrument.
Symbols Used On Unit
NumberSymbolPublicationDescription
1IEC 417, No. 5031Direct current
2IEC 417, No. 5172Equipment protected throughout by DOUBLE
INSULATION or REINFORCED INSULATION
(equivalent to Class II of IEC 536–see annex H)
3
!
ISO 3864, No. B.3.1Caution (refer to accompanying documents)
Technical Improvements
• The manufacturer reserves the right to modify technical data without prior notice.
1. DESCRIPTION
1.1 Unit Description ...............................................................................................1
1.2 Unit Features ...................................................................................................1
13.3.2 Self Test Alarms ............................................................................68
72
ii
Unit Description1. Description
1.1 Unit Description:
The SUPERtrol-I MS-716 Flow Computer is a special version of the
SUPERtrol-1 Flow Computer which is supplied in a vehicle mount
enclosure. The SUPERtrol-I MS-716 satisfies the instrument requirements
for a variety of flowmeter types in liquid applications. Multiple flow equations
and instrument functions are available in a single unit with many advanced
features.
The alphanumeric display shows measured and calculated parameters
in easy to understand format. Single key direct access to measurements
and display scrolling is supported.
SUPERtrol-I MS-716 Flow Computer
Unit Features
The versatility of the SUPERtrol-I MS-716 permits a wide measure of
versatility within the instrument package. The various hardware inputs
and outputs can be “soft” assigned to meet a variety of common application
needs. The user “soft selects” the usage of each input/output while
configuring the instrument. Consider the following illustrative examples.
The user can assign the standard RS-232 Serial Port for data logging,
transaction printing, or for connection to a modem for remote meter
reading.
1.2 Unit Features:
The SUPERtrol-I MS-716 Flow Computer offers the following features:
• EZ-Preset or Standard Preset
• Custody Transfer with Audit Trail
• Multiple Instrument Functions
• Menu Selectable Hardware & Software Features
• Two Line LCD or VFD Display
• Isolated Outputs Standard
• Versatile RS-232 Port Standard
• DIN Enclosure with Two Piece Connector
• Optional Networking Cards
• Advanced Batching Features
1
1.3 Specifications:
SUPERtrol-I MS-716 Flow Computer
Specifications:
Environmental
Indoor Use
Altitude up to 2000m
Operating Temperature: 0°C to +50°C
(-20°C to 55°C optional)
Storage Temperature: -40°C to +85 C
Maximum Relative Humidity : 80% for temperatures
up to 31°C decreasing linearly to 50% RH at
40°C
Mains supply voltage fluctuations not to exceed
±10%
of the nominal voltage
Transient overvoltage according to INSTALLATION
CATEGORY II (see UL 3101-1 Annex J)
POLLUTION DEGREE 2 in accordance with
IEC 664 (see 3.7.3)
Materials: UL, CSA, VDE approved
UL= UL Class 1 Division 2
Approvals:CE Approved Light Industrial,
UL File #: E192404 / CUL
Display
Type: 2 lines of 20 characters
Types: Backlit LCD or VFD ordering options
Character Size: 0.3" nominal
User selectable label descriptors and units of
measure
Keypad
Keypad Type: Membrane Keypad
Keypad Rating: Sealed to Nema 4
Number of keys: 16
Enclosure
Size: See Dimensions
Seal: NEMA4X
Materials: Aluminum
Pulse Inputs:
Number of Flow Inputs: one
Configurations supported: single input with or
Pull Down Resistance: 10 KΩ to common
Trigger Level: (menu selectable)
High Level Input
Logic On:3 to 30 VDC
Logic Off:0 to 1 VDC
Low Level Input (mag pickup)
Selectable sensitivity: 10 mV & 100 mV
Minimum Count Speed: User selectable
Maximum Count Speed: Selectable: 0 to 20kHz
Overvoltage Protection: 50 VDC
Fast Transient: Protected to 500 VDC
(Capacitive Clamp)
Compensation Input
The compensation input is menu selectable for
temperature, density or not used.
Operation: Ratiometric
Accuracy: 0.01% FS
Thermal Drift: Less than 100 ppm/C
Basic Measurement Resolution: 16 bit
Update Rate: 1 update/sec minimum
Automatic Fault detection:
Signal Over-range/under-range
Current Loop Broken
RTD short
RTD open
Fault mode to user defined default settings
Transient Protection: 500 V (Capacitive Clamp)
Reverse Polarity: No ill effects
Over-Voltage Limit (Voltage Input): 50 VDC
Power Input
The factory equipped power option is internally
fused. An internal line to line filter capacitor is
provided for added transient suppression.
Order Option 1: 110VAC: 85 to 127 Vrms, 50/60
Hz
Order Option 2: 220VAC: 170 to 276 Vrms, 50/60
Hz
Order Option 3: 12VDC: 10.5 to 14 VDC
Order Option 4: 24VDC: 18 to 24 VDC
Flow Inputs:
Analog Input:
Ranges
Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA
Basic Measurement Resolution: 16 bit
Update Rate: 5 updates/sec minimum
Automatic Fault detection: Signal over/underrange,
Current Loop Broken
Calibration: Self Calibration and Auto-zero
Continuously
Extended calibration: Learns Zero and Full Scale
of each range using special test mode.
Fault Protection:
Fast Transient: 500 V Protection
(Capacitive Clamp)
Reverse Polarity: No ill effects
Over-Voltage Limit: 50 VDC Over voltage
protection
Over-Current Protection: Internally current
limited Protected to 24 VDC.
Available Input Ranges
Voltage: 0-10 VDC, 0-5 VDC, 1-5 VDC
Current: 4-20 mA, 0-20 mA
Resistance: 100 Ohms DIN RTD
100 Ohm DIN RTD (DIN 42-760, BS 1904):
Three Wire Lead Compensation
Internal RTD linearization learns ice point
resistance
1 mA Excitation current with reverse polarity
protection
Temperature Resolution: 0.01 C
Control Inputs
Switch Inputs are menu selectable for Start, Stop,
Reset, Lock, Inhibit, Alarm Acknowledge, Print or
Not Used.
Control Input Specifications
Input Scan Rate: 10 scans per second
Logic 1: 4 - 30 VDC
Logic 0: 0 - 0.8 VDC
Transient Suppression: 500 V fast transient
(Capacitive Clamp)
Input Impedance: 100 KΩ
Control Activation: Positive Edge or Pos. Level
based on product definition
Excitation Voltage
110/220 VAC Powered Units
Menu Selectable: 5, 12 or 24 VDC @ 100mA
24 VDC Powered Units
Menu Selectable: 5 or 12 VDC @ 100mA
12 VDC Powered Units
5 VDC @ 100mA
2
SUPERtrol-I MS-716 Flow Computer
Relay Outputs
The relay outputs are menu assignable to
(Individually for each relay) Low Rate Alarm, Hi
Rate Alarm, Prewarn Alarm, Preset Alarm or
General purpose warning (security).
Number of relays: 2 (4 optional)
Contact Style: Form C contacts
Contact Ratings:2400 VAC Max., 3 Amps Max.
Fast Transient Threshold: 1000 V
Serial Communication
The serial port can be used for printing, datalogging,
modem connection and communication with a
computer.
RS-232:
The analog output is menu assignable to
correspond to the Uncompensated Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature,
Density, Volume Total, Corrected Volume Total or
Mass Total.
Type: Isolated Current Sourcing (AC power options)
Isolated I/P/C: 500 V
Available Ranges: 4-20 mA, 0-20 mA
Resolution: 12 bit
Accuracy: 0.05% FS at 20 Degrees C
Update Rate: 1 update/sec minimum
Temperature Drift: Less than 200 ppm/C
Maximum Load: 1000 ohms (at nominal line
voltage)
Compliance Effect: Less than .05% Span
60 Hz rejection: 40 dB minimum
EMI: No effect at 3 V/M
Calibration: Operator assisted Learn Mode
Averaging: User entry of DSP Averaging constant
to cause an smooth control action.
Isolated Pulse output
The isolated pulse output is menu assignable to
Uncompensated Volume Total, Compensated
Volume Total or Mass Total.
Isolation I/O/P: 500 V
Pulse Output Form: Open Collector
Maximum On Current: 125 mA
Maximum Off Voltage: 30 VDC
Saturation Voltage: 1.0 VDC
Maximum Off Current: 0.1 mA
Pulse Duration: User selectable
Pulse output buffer: 8 bit
Fault Protection
The Flow Computer can be thought of as making a
series of measurements of flow, temperature/
density sensors and then performing calculations
to arrive at a result(s) which is then updated
periodically on the display. The analog output, the
pulse output, and the alarm relays are also updated.
The cycle then repeats itself.
Step 1:Update the measurements of input signals-
Raw Input Measurements are made at each input
using equations based on input signal type selected.
The system notes the “out of range” input signal as
an alarm condition.
Step 2:Compute the Flowing Fluid Parameters-
The temperature, viscosity, and density equations
are computed as needed based on the flow
equation and input usage selected by the user.
Step 3 : Compute the Volumetric Flow-
Uncompensated flow is the term given to the flow
in volume units. The value is computed based on
the flowmeter input type selected and augmented
by any performance enhancing linearization that
has been specified by the user.
Step 4: Compute the Corrected Volume Flow at
Reference ConditionsIn the case of a corrected liquid volume flow
calculation, the corrected volume flow is computed
as required by the selected compensation equation.
Step 5 : Compute the Mass Flow-
All required information is now available to compute
the mass flow rate as volume flow times density.
Step 6: Check Flow Alarms-
The flow alarm functions have been assigned to
one of the above flow rates during the setup of the
instrument. A comparison is now made by
comparing the current flow rates against the
specified hi and low limits.
Step 7: Compute the Analog Output-
This designated flow rate value is now used to
compute the analog output.
Step 8: Compute the Flow Totals by Summation-
A flow total increment is computed for each flow
rate. This increment is computed by multiplying the
respective flow rate by a time base scaler and then
summing. The totalizer format also includes
provisions for total rollover.
Step 9: Total Preset Comparisons-
The total associated with a preset function is then
compared against the corresponding preset value
and any required control actions taken.
Step 10: Pulse Output Service-
The pulse output is next updated by scaling the
total increment which has just been determined by
the pulse output scaler and summing it to any
residual pulse output amount.
Step 11: Update Display and Printer Output-
The instrument finally runs a task to update the
various table entries associated with the front panel
display and serial outputs.
3
SUPERtrol-I MS-716 Flow Computer
Setup Mode
The setup mode is password protected by means of
a numeric lock out code established by the user. In
addition, a secret, manufacturers numeric unlock
entry sequence is available.
The system also provides a minimum implementation
of an “audit trail” which tracks significant setup
changes to the unit. This feature is increasingly being
found of benefit to users or simply required by
Weights and Measurement Officials in systems used
in commerce, trade, or “custody transfer” applications.
A software program is available which runs on a PC
using a RS-232 Serial for connection to the Flow
Computer. Illustrative examples may be down loaded
in this manner.
The setup mode has numerous subgrouping of
parameters needed for flow calculations. There is a
well conceived hierarchy to the setup parameter list.
Selections made at the beginning of the setup affect
offerings further down in the lists.
In the setup mode, the flow computer activates the
correct setup variables based on the instrument
configuration, the flow equation, and the hardware
selections made for the compensation transmitter
type, the flow transmitter type, and meter
enhancements (linearization) options selected. All
required setup parameters are enabled. All setup
parameters not required are suppressed.
A help line prompt is provided for each entry. In
addition a help message is available which may be
accessed by depressing the “HELP” key.
Maintenance Mode:
The Maintenance Mode of the SUPERtrol-I MS716 is the Test and Calibration Mode for the device.
This mode provides a number of specialized
utilities required for factory calibration, instrument
checkout on start-up, and periodic calibration
documentation.
A password is required to gain access to this
specialized mode of operation. Normally quality,
calibration, and maintenance personnel will find
this mode of operation very useful. It is also useful
for factory testing.
Many of these tests may be used during start-up
of a new system. Inputs signals may be read, and
output signals may be exercised to verify the
electrical interconnects before the entire system
is put on line.
The following action items may be performed in
the Maintenance Mode:
Print Calibration/Maintenance Report
Examine Audit Trail
Perform Keypad Checkout
Perform Display Checkout
Perform Pulse Input Checkout
Perform Pulse Output Checkout
Perform Control Input Checkout
Perform Relay Output Checkout
Perform Analog Input Checkout
Perform Analog Output Checkout
Calibrate Analog Inputs using the Learn Feature
Calibrate Analog Output using the Learn Feature
Battery Check
In the setup mode selections, several parameters
are required to be input by the operator since these
parameters are blank when the unit is received. The
user will be prompted for these necessary values for
his application.
Also note that in the setup mode are parameter
selections which have preassigned industry standard
values. The unit will assume these values unless
they are modified by the user.
Most of the process input variables have available a
“default” or emergency value which must be entered.
These are the values that the unit assumes when a
malfunction is determined to have occurred on the
corresponding input.
It is possible to enter in a nominal constant value for
temperature or density, or analog flow inputs by
placing the desired nominal value into both the lo
and hi values. This is also a convenience when
performing bench top tests without simulators.
Note that a calibration of the analog input/output
will advance the audit trail counters since it effects
the accuracy of the system.
RS-232 Serial Port
The SUPERtrol-I MS-716 has a general purpose
RS-232 Port which may be used for any one of
the following purposes:
Transaction Printing
Data Logging Internal Datalog Dumps
Remote Metering by Modem (optional)
Computer Communication Link
Configuration by Computer
Print System Setup
Print Calibration/Malfunction History
Remote Control
Instrument Setup by PC’s over Serial Port
A Diskette program is provided with the SUPERtrolI MS-716 that enables the user to rapidly configure
the SUPERtrol-I MS-716 using an Personnel
Computer. Included on the diskette are common
instrument applications which may be used as a
starting point for your application. This permits the
user to have an excellent starting point and helps
speed the user through the instrument setup.
4
SUPERtrol-I MS-716 Flow Computer
Operation of Serial Communication Port with
Printers
SUPERtrol-I MS-716’s RS-232 channel supports
a number of operating modes. One of these
modes is intended to support operation with a
printer in metering applications requiring
transaction printing, data logging and/or printing
of calibration and maintenance reports.
For transaction printing, the user defines the items
to be included in the printed document. The user
can also select what initiates the transaction print
generated as part of the setup of the instrument.
The transaction document may be initiated via a
front panel key depression, a remote contact
closure, or upon completion of a batch.
In data logging, the user defines the items to be
included in each data log as a print list. The user
can also select when or how often he wishes a
data log to be made. This is done during the
setup of the instrument as either a time of day or
as a time interval between logging.
The system setup and maintenance report lists
all the instrument setup parameters and usage
for the current instrument configuration. In
addition, the Audit trail information is presented
along with a status report listing any observed
malfunctions which have not been corrected.
The user initiates the printing of this report at a
designated point in the menu by pressing the
print key on the front panel.
Operation of Serial Port with Modems (optional)
The SUPERtrol-I MS-716 RS-232 channel supports
a number of operating modes. One of these modes
is intended to support operation with a modem in
remote metering applications.
An external modem is intentionally being used with
the SUPERtrol-I MS-716. This permits use with the
variety of modem standards worldwide while
avoiding the specialized approvals required for
equipment that is deemed to fall under the category
of telecommunication equipment.
In the modem mode, the SUPERtrol-I MS-716 is
assumed to be operating in a remote metering role.
The SUPERtrol-I MS-716 will support key items in
the Hayes Compatible “AT” Command Set. In this
role, the SUPERtrol-I MS-716 will have the following
special abilities:
0. Monitor the modem status as a task of the system
1. Instruct the modem to answer an incoming call
ATA
2. Respond to the calling modem at a compatible
baud rate and protocol
3. Perform error checking in conjunction with the
modem
4. Monitor the status of the carrier
5. Terminate the telephone connection in event the
connection is lost.
In addition, the SUPERtrol-I MS-716 will be capable
of initiating a call to a designed telephone number
in the event of a metering malfunction.
5
2. Installation
SUPERtrol-I MS-716 Flow Computer
General Mounting
Hints
2.1 General Mounting Hints:
Termination Connectors:
Minimum Wire Gauge:22 AWG
Maximum Wire Gauge:14 AWG
Voltage/current limits are limited by unit specifications.
Permanently Connected Equipment:
UL 3101-1, Section 6.12.2.1 specifies that:
•A switch or circuit breaker shall be included in the building
installation;
•It shall be in close proximity to the equipment and within easy
reach of the OPERATOR;
•It shall be marked as the disconnecting device for the equipment.
Ensure that the switch or circuit breaker chosen is suitable for the
power requirements of the unit.
2.2 Mounting Diagrams:
Vehicle Mount Dimensions
0.47
(12)
7.52 (191)
6.54 (116.1)
TOTAL
1
START
F14PRINT5GRAND
STOP
SCROLL7PRE 28F2
CLEAR
RATE2PRE 1
TIME
0 –
3
6
9
HELP
•
9.52 (241.8)
0.55
6.35 (161.3)
(14)
0.875 DIA. CONDUIT ENTRY
3 PLACES
1.82
(46.2)
3.69
(93.7)
Optional Mounting Bracket (MB)
9.13
(232)
4.61
(117)
MENU
ENTER
8.67 (220.2)
0.40
(10.2)
4.08 (103.5)
0.24 DIA. MOUNTING
THRU HOLE (4 PLACES)
0.90
(22.9)
1.82
(46.2)
9.52 (241.8)
6
START
STOP
TOTAL
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
8.34
(211.8)
3
MENU
6
ENTER
9
HELP
•
8.67 (220.2)
0.40
(10.2)
NOTE: All dimensions are
in inches (mm)
3. Applications
SUPERtrol-I MS-716 Flow Computer
Liquid Volume
3.1 Liquid Volume
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor can also be installed to correct for liquid thermal expansion (see
3.2 Corrected Volume).
Calculations:
• For Flowmeters with Pulse Outputs, Volume flow is calculated using
the flowmeter frequency output and the user entered K-Factor.
• For Flowmeters with Analog Transmitters, Volume flow is calculated
using the measured flowmeter signal and the user entered scale settings.
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total
• Analog Output
Rate or Total
• Pulse Output
Total
• Relay Outputs
Rate or Total Alarms
Liquid Volume
Illustration
Applications:
The Flow Computer can monitor actual volume flow and total of any
liquid. Flow alarms are provided via relays and datalogging is available
via analog (4-20mA) and serial outputs.
TOTAL
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
3
MENU
6
ENTER
9
HELP
•
START
STOP
Calculations
Flowmeter
Pulse Input; Average K-Factor
input frequency * time scale factor
Volume Flow =
K-Factor
Analog Input; Linear
Volume Flow = % input * Full Scale Flow
7
SUPERtrol-I MS-716 Flow Computer
Corrected
Liquid Volume
3.2 Corrected Liquid Volume
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor is installed to correct for liquid thermal expansion.
Calculations:
• Corrected Volume is calculated using the flow and temperature inputs
as well as the thermal expansion coefficient stored in the flow computer.
Use the "SET FLUID PROPERTIES" submenu to define reference
temperature and density values for standard conditions.
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total, Temperature,
Density
• Analog Output
Rate, Total, Temperature or Density
• Pulse Output
Total
• Relay Outputs
Rate , Total or Temperature Alarms
Corrected
Liquid Volume
Illustration
Applications:
Monitoring corrected volume flow and total of any liquid. Flow alarms are
provided via relays and datalogging is available via analog (4-20mA) and
serial outputs.
Actual volume is measured by the flow element (DP transmitter,
Flowmeter). Temperature is measured by the temperature transmitter. A
density transmitter can be used for direct density measurements.
Calculations:
• The density and mass flow are calculated using the reference density
and the thermal expansion coefficient of the liquid (see "SET FLUID
PROPERTIES" submenu)
Output Results:
• Display Results
Flow Rate, Resettable Total, Non-Resettable Total, Temperature,
Density
• Analog Output
Rate, Total, Temperature or Density
• Pulse Output
Total
• Relay Outputs
Rate, Total or Temperature Alarms
Liquid Mass
Illustration
Applications:
Monitoring mass flow and total of any liquid. Flow alarms are provided via
relays and datalogging is available via analog (4-20mA) and serial outputs.
TOTAL
START
STOP
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
3
MENU
6
ENTER
9
HELP
•
START
STOP
TOTAL
RATE2PRE 1
1
F14PRINT5GRAND
SCROLL7PRE 28F2
TIME
CLEAR
0 –
3
MENU
6
ENTER
9
HELP
•
Calculations
Orifice Plate
with DP Transmitter
Volume Flow
As calculated in section 3.1
Mass Flow
Mass Flow = volume flow * density
Temp./Dens.
Tr ansmitter
9
FlowmeterTemp./Dens.
Tr ansmitter
SUPERtrol-I MS-716 Flow Computer
Batching
3.4 Batching
Measurements:
A flowmeter measures the actual volume in a liquid line. A temperature
sensor can also be installed to correct for liquid thermal expansion (see
3.2 Corrected Volume).
Calculations:
• For Flowmeters with Pulse Outputs, Volume flow is calculated using
the flowmeter frequency output and the user entered K-Factor.
• For Flowmeters with Analog Transmitters, Volume flow is calculated
using the measured flowmeter signal and the user entered scale settings.
• Corrected Volume is calculated using the flow and temperature inputs
as well as the thermal expansion coefficient stored in the flow computer.
Output Results:
•Display Results
Flow Rate, Batch Total, Non-Resettable Total, Temperature, Density
•Analog Output
Rate, Total, Temperature or Density
•Pulse Output
Total
•Relay Outputs
Batch Total, Rate, or Temperature Alarms
Batching Illustration
Applications:
Batching and monitoring flow and total of any liquid. Batching is
accomplished via relays and datalogging is available via analog (420mA) and serial outputs.
Examples using MLT787S+ Barrier (MTL4755ac for RTD)
Flow Input
Temperature Input
(4-20mA Transmitter)
Flow Input
Hazardous AreaSafe Area
4-20mA Flow
Transmitter
Q/∆P
4-20
+
–
3 4
28V
Diode
Temperature Input (4-20mA Transmitter)
Hazardous AreaSafe Area
2 1
1
2
3
4
1
24V Out
24V Out
4-20mA In
Common
Temperature Input
(RTD)
4-20mA Temp.
Transmitter
–
4-20
T
+
4 3
Temperature Input (RTD)
Hazardous AreaSafe Area
3-Wire RTD
1 2 3 4 5 6
28V
Diode
1 2
7 8 9 10 11 12 13 14 15 16 17
4
Common
8
4-20mA In
Common
4
5
RTD Excite (+)
6
RTD Sense (+)
7
RTD Sense (–)
8
12
SUPERtrol-I MS-716 Flow Computer
5. UNIT OPERATION
5.1 Front Panel Operation Concept for Run Mode
The SUPERtrol-I MS-716 is fully programmable through the front panel. Please review the
following usage summary before attempting to use the instrument.
TOTAL
RATE2PRE 1
START
STOP
1
F14PRINT5GRAND
SCROLL7PRE 28F2
CLEAR
TIME
0 –
3
MENU
6
ENTER
9
HELP
•
HELP
On-line help is provided to assist the operator in using this product. The help is available during RUN and SETUP
modes simply by pressing the HELP key. The HELP key is used to enter decimals when entering numeric values.
FUNCTION KEYS
In the RUN mode, several keys have a special, direct access feature, to display an item of interest (i.e. RATE,
TOTAL, PRE 1, etc.). Press the key to view your choice. Press the SCROLL key to return to scrolling display.
CLEARING TOTALIZER
To clear the total, you must press the TOTAL Function Key 3 times. You will be asked to verify this action. The
operator will be prompted to enter password if the unit is locked.
NOTE:In the Batcher Mode, simply press the CLEAR key to reset the total (the batcher must be stopped or finished
batching). It is not necessary to press the TOTAL Function Key first.
CLEARING GRAND TOTAL
To clear the grand total, you must press the GRAND Function Key 3 times. You will be asked to verify this action.
The supervisor will be prompted to enter the supervisor password if the unit is locked.
PRESET KEYS USING EZ PRESET
In the RUN mode, PRE 1 & PRE 2 keys are used to view and/or change the preset setpoints. To view the Presets,
simply press the desired Preset key. Press Enter then press the Clear key for direct editing of the preset setpoints.
PRESET KEYS
In the RUN mode, PRE 1 & PRE 2 keys are used to view and/or change the preset setpoints. To view the Presets,
simply press the desired Preset key. Rapidly press the Preset keys 3 times, then press the Clear key for direct editing
of the preset setpoints. Press Enter or Start to complete the Preset entry.
SCROLL
Rapidly press the Scroll key twice to setup a display list.
Press the CLEAR key to remove old scroll list.
Press the function key for the item you wish to add
Use the ∆ ∇ keys to assign the line.
PRINT
The PRINT key is used to print on demand. When the PRINT key is pressed, a user defined list of data (TOTAL,
RATE, PRE 1, etc.) is sent to the RS-232 port. A timed message of "PRINTING" will be displayed to acknowledge the
print request.
SPECIAL BATCHING KEYS
The START and STOP keys are used only when batching to start and stop batches. The CLEAR key will clear the
total without first pressing the TOTAL key (unit must be stopped). All other keys work the same in both Rate/Total
mode and Batch mode. The Start and Stop keys operation are set by the control input settings. The Start options are:
START or RESET/START. The Stop options are: STOP or STOP/RESET.
MENU KEY
The MENU key is used to enter the Setup and Test modes. Press the MENU key to enter the Setup and Test modes.
(See section 6 for Setup mode, section 8 for Test mode). The MENU key is used as "escape" in Setup and Test
Programming. Pressing the MENU key wile programming in the Sub-Menu groups will backup the display to that
Sub-Menu group heading. Pressing the MENU key while viewing the Sub-Menu groups will backup the display to the
Top Level Menu.
ACKNOWLEDGING ALARMS
Most alarm messages are self-clearing. Press the ENTER key to acknowledge and clear alarms.
NOTE: Some keys and functions are password protected. Enter the password to gain access. The passwords are
factory set as follows:Operator = 0Supervisor = 2000
13
SUPERtrol-I MS-716 Flow Computer
General
Operation
Rate/Total
Operation
Password Protection
(Rate/Total mode)
5.2 General Operation
The unit can display: Rate, Total, Grand Total, Temperature, Density, Presets
and Time of Day. The Temperature and/or Density can be displayed even if
you are using the Volumetric Flow Equation (a Temperature or Density sensor
must be installed). The unit can perform Mass or Corrected Volume equations
using a temperature or density sensor (these equations can be computed
without Temp/Dens sensors by using user defined default values). The unit can
be programmed to perform Ratemeter/Totalizer or Batching functions (see
section 6.3, SELECT INSTRUMENT Submenu).
5.3 Ratemeter/Totalizer Operation
The Ratemeter/Totalizer mode is used primarily to monitor flowrate and
accumulated total. The relays can be used to trigger flow, total, temperature or
density alarms.
5.3.1 Password Protection for Rate/Total mode
After an Operator and/or Supervisor Password is entered in the setup mode
(see section 6.3, SETUP PASSWORD submenu), the unit will be locked. The
unit will prompt the user for the password when trying to perform the following
functions:
Clear Total
Clear Grand Total
Enter Menu
Edit Preset 1 (PRE 1 Key)
Edit Preset 2 (PRE 2 Key)
Relay Operation
(Rate/Total mode)
Pulse Output
(Rate/Total mode)
The Supervisor password should be reserved for supervisors. The Supervisor
password will allow access to restricted areas of the Setup and Test menus.
5.3.2 Relay Operation in Rate/Total mode
Up to four relays are available (two standard) for alarm outputs. The relays can
be assigned to trip according to rate, total, temperature, density readings or
general system alarms. The relays can be programmed for low or high alarms.
Preset 1 (RLY1) and Preset 2 (RLY2) are easily accessible by pressing the
PRE 1 or PRE 2 key on the front panel. Preset 3 and Preset 4 are accessible
only through the setup menu.
5.3.3 Pulse Output in Rate/Total mode
The isolated pulse output (open collector) is menu assignable to Volume Total,
Corrected Volume Total or Mass Total. The pulse output duration can be set for
10mS (50 Hz max) or 100mS (5 Hz max). A pulse output scale factor (pulse
value) can be set to scale the pulse output. The pulse output is ideal for
connecting to remote totalizers or other devices such as a PLC. See section
1.3 for electrical specifications.
Analog Output
(Rate/Total mode)
5.3.4 Analog Output in Rate/Total mode
The analog output is menu assignable to correspond to the Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature, Density, Volume Total,
Corrected Volume Total or Mass Total. The analog output is ideal for "trend"
tracking using strip chart recorders or other devices.
14
SUPERtrol-I MS-716 Flow Computer
RS-232 Serial Port
(Rate/Total mode)
5.3.5 RS-232 Serial Port Operation in Rate/Total mode
The RS-232 serial port can be used for programming (using the Setup Disk)
or for communicating to printers and computers in the Operating Mode (Run
Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status
such as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Operation of RS-232 Serial Port with Printers:
Transaction Printing
For transaction printing, the user defines the items to be included in the
printed document (see section 6.3.20 SET DATA OUTPUT, Select_list). The
transaction document can be initiated by pressing the PRINT key or by a
remote contact closure.
Data Logging
In data logging, the user defines the items to be included in each data log
(see section 6.3.20 SET PRINTER OUTPUT, Select_list). The user can also
select when (time of day) or how often (print interval) the data log is to be
made (see section 6.3.19 SET PRINTER OUTPUT, Configure).
System Setup and Maintenance Report
The system setup and maintenance report lists all of the instrument setup
parameters and usage for the current instrument configuration. The audit
trail information and a status report is also printed. This report is initiated in
the Test menu (see section 8.2.3 PRINT SYSTEM SETUP).
RS-485 Serial Port
(Rate/Total mode)
5.3.6 RS-485 Serial Port (optional)
RS-485 Port Description:
The optional RS-485 card utilizes Modbus RTU protocol to access
a variety of process parameters and totalizers. The Relays can
be controlled via Modbus. In addition, action routines can be
executed. For further information, contact factory and request
RS-485 Protocol manual.
Operation of Serial Communication Port with PC
The flow computer's RS-485 channel supports a number of
Modbus RTU commands. Modbus RTU drivers are available
from third party sources for a variety of Man Machine Interface
software for IBM compatible PC's.
The user reads and writes information from/to the RS-485 using
the Modbus RTU commands. The SUPERtrol-I MS-716 then
responds to these information and command requests.
Process variables and totalizers are read in register pairs in
floating point format. Time and date are read as a series of
integer register values. Alarms are individually read as coils.
Action routines are initiated by writing to coils.
15
SUPERtrol-I MS-716 Flow Computer
5.4 Batcher Operation
The Batcher mode is used primarily to control batches. The main difference
between the Batch mode and Rate/Total mode is the relay operation. The
Batch mode allows the operator to "START" the unit via the front panel or
remote input. Once started, the relays (RLY1 & RLY2) will energize and send
power to a flow control device (i.e. solenoid valve or pump). The flow sensor
will send a signal to the unit and total accumulation will begin. When the
Prewarn value (PRE 2) is reached, Relay 2 will drop out (this is ideal for flow
slow down). When the Batch amount (PRE 1) is reached, Relay 1 will drop out
and the Batch is complete.
Several messages will be displayed during normal batch operation (i.e. Batch
Fill, Batch Stopped). The keypad is disabled for the duration of these timed
messages (approx. 2 sec).
Batcher Configuration
5.4.1 Batcher Configuration.
When the unit is programmed for batch mode, several batch operation choices
are available. These choices include: Up or Down Counting, Maximum Batch
Preset, Batch Overrun Compensation, Auto Batch Restart, Time Delay, Flow
Signal Timeout, Maximum Drain Time, Slow Start Quantity, Start or Reset/Start,
and Stop or Stop/Reset.
Standard Preset or EZ Preset
Use Standard Preset for applications in which the batch amount does not
change frequently.
Use EZ Preset in applications in which the batch amounts change frequently.
The EZ Preset mode was designed to enter presets with minimum key strokes.
Batch Count Mode
The Batch Count Mode allows the user to choose whether the unit will batch up
to a preset value or batch down from a preset value to zero.
Maximum Batch Preset
The Maximum Batch Preset allows the user to program the Maximum Batch
value allowed to be entered by the operator. If an operator should try to
program a batch higher then this value, the unit will not allow the value to be
entered and will prompt the user with an error message saying that the Maximum
Batch Preset has been exceeded.
Batch Overrun
The Batch Overrun is used for batch applications that have slow responding
valves and a consistent batching flowrate. When the Batch Overrun is set, the
unit will compensate for batch overruns by computing an averaged overrun
value from the last four batches. This average is used to internally adjust the
batch setpoint to minimize overrun.
Flow Signal Timeout
The Flow Signal Timeout allows the user to enter a timeout of 0 to 99 seconds.
If a batch is “Filling” and zero flow persists for more than the user entered time
then the batch will be aborted. This prevents over flows due to faulty flow
sensors and/or wiring.
Maximum Drain Time
The unit declares that a batch is “done” when the flow rate equals “0”. A flow
rate may be present long after the Preset Relay de-energizes due to slow
reacting valves or leaky valves. The Maximum Drain Time allows the user to
enter an amount of time (0 to 99 seconds) to wait before declaring “Batch
Done”. After the Preset Batch quantity is reached, the unit will declare “Batch
Done” when the flow rate is “0” or the Maximum Drain Time has expired. The
batch data will then be available for printing and datalogging.
16
SUPERtrol-I MS-716 Flow Computer
Slow Start Quantity
The Slow Start Quantity is a function that allows an amount to be entered for a
Slow Start up. This function requires two stage valve control. RLY 1 (slow flow)
will energize for Slow Start and RLY 2 (fast flow) will energize after the Slow
Start Quantity has been delivered. This helps reduce turbulence when filling an
empty container.
START, RESET/START and STOP, STOP/RESET
When configuring the control inputs, Control Input1 can be set for START or
RESET/START. When set for START, the unit will start batching when a signal
is applied to Control Input1 or the front panel Start key is pressed. A separate
Reset signal must be used to clear the previous batch total. When set for
RESET/START, the unit will automatically reset then start when a signal is
applied to Control Input1 or the front panel Start key is pressed (provided that
the pervious batch was completed). If a previous batch was stopped during a
batch cycle, the unit will Start from where it was stopped.
Control Input 2 can be set for STOP or STOP/RESET. When set for STOP, the
unit will stop batching when a signal is applied to Control Input 2 or the front
panel Stop key is pressed. A separate Reset signal must be used to clear the
batch total. When set for STOP/RESET, a running batch will stop when a
signal is applied to Control Input 2 or the front panel Stop key is pressed. If the
unit is Stopped or after a completed batch, the unit will reset when a signal is
applied to Control Input 2 or the front panel Stop key is pressed.
NOTE:Applying a voltage level to Control Input 2 will inhibit all Start inputs in
either mode.
Password Protection
(Batch mode)
Relay Operation
(Batch mode)
5.4.2 Password Protection for Batcher Mode
After an Operator and/or Supervisor Password is entered in the setup mode
(see section 6.3, SETUP PASSWORD submenu), the unit will be locked. The
unit will prompt the user for the password when trying to perform the following
functions:
Clear Grand Total
Enter Menu
The Supervisor password should be reserved for supervisors. The Supervisor
password will allow access to restricted areas of the Setup and Test menus.
The passwords are factory set as follows:
Operator = 0
Supervisor = 2000
5.4.3 Relay Operation in Batcher mode
Up to four relays are available (two standard) for alarm outputs. Preset 1
(RLY1) is reserved for batch amount, Preset 2 (RLY2) is reserved for prewarn.
(see section 5.4 Batcher Operation for Relay 1 & Relay 2 functions)
Preset 1 (RLY1) and Preset 2 (RLY2) are easily accessible by pressing the PRE
1 or PRE 2 key on the front panel. Preset 3 and Preset 4 are accessible only
through the setup menu.
Relays 3 and 4 can be assigned to trip according to rate, total, temperature,
overrun or alarm. When Rate is selected the relays can be programmed for low
or high alarms.
Pulse Output
(Batch mode)
Analog Output
(Batch mode)
5.4.4 Pulse Output in Batcher mode
The isolated pulse output (open collector) is menu assignable to Volume Total,
Corrected Volume Total or Mass Total. The pulse output duration can be set for
10mS (50 Hz max) or 100mS (5 Hz max). A pulse output scale factor (pulse
value) can be set to scale the pulse output. The pulse output is ideal for
connecting to remote totalizers or other devices such as a PLC. See section 1.3
for electrical specifications.
5.4.5 Analog Output in Batcher mode
The analog output is menu assignable to correspond to the Volume Rate,
Corrected Volume Rate, Mass Rate, Temperature, Density, Volume Total,
Corrected Volume Total or Mass Total. The analog output is ideal for "trend"
tracking using strip chart recorders or other devices.
17
SUPERtrol-I MS-716 Flow Computer
RS-232 Serial Port
(Batch mode)
5.4.6 RS-232 Serial Port Operation in Batcher mode
The RS-232 serial port can be used for programming (using the Setup Disk)
or for communicating to printers and computers in the Operating Mode (Run
Mode).
PC Communications:
The Setup Disk also allows the user to query the unit for operating status such
as Flow Rate, Flow Total, Temperature, Density, Presets, etc.
Operation of RS-232 Serial Port with Printers:
Transaction Printing
For transaction printing, the user defines the items to be included in the
printed document (see section 6.3.20 SET DATA OUTPUT, Select_list). The
transaction document can be initiated by pressing the PRINT key, by a
remote contact closure or print at end of batch.
Data Logging
In data logging, the user defines the items to be included in each data log (see
section 6.3.20 SET PRINTER OUTPUT, Select_list). The user can also
select when (time of day) or how often (print interval) the data log is to be
made (see section 6.3.19 SET PRINTER OUTPUT, Configure).
System Setup and Maintenance Report
The system setup and maintenance report lists all of the instrument setup
parameters and usage for the current instrument configuration. The audit trail
information and a status report is also printed. This report is initiated in the
Test menu (see section 8.2.3 PRINT SYSTEM SETUP).
RS-485 Serial Port
(Batch mode)
5.4.7 RS-485 Serial Port (optional)
RS-485 Port Description:
The optional RS-485 card utilizes Modbus RTU protocol to access
a variety of process parameters and totalizers. Batches/Relays
can be controlled remotely via Modbus. In addition, action routines
can be executed. For further information, contact factory and
request RS-485 Protocol manual.
Operation of Serial Communication Port with PC
The flow computer's RS-485 channel supports a number of
Modbus RTU commands. Modbus RTU drivers are available
from third party sources for a variety of Man Machine Interface
software for IBM compatible PC's.
The user reads and writes information from/to the RS-485 using
the Modbus RTU commands. The SUPERtrol-I MS-716 then
responds to these information and command requests.
Process variables and totalizers are read in register pairs in
floating point format. Time and date are read as a series of
integer register values. Alarms are individually read as coils.
Action routines are initiated by writing to coils.
18
SUPERtrol-I MS-716 Flow Computer
6. PROGRAMMING
6.1 Front Panel Operation Concept for Program Mode
The SUPERtrol-I MS-716 is fully programmable through the front panel.
Please review the following usage summary before attempting to use
the instrument.
TOTAL
RATE2PRE 1
START
STOP
1
F14PRINT5GRAND
SCROLL7PRE 28F2
CLEAR
TIME
0 –
3
MENU
6
ENTER
9
HELP
•
How To Make Mode
Changes
How To Navigate
Through Sub-Menu
Groups
How To Select
Program Choices
Setup Mode:
MODE CHANGES
Pressing the MENU key will offer selections of RUN, SETUP, TEST.
RUN is the normal operating mode for the instrument. SETUP offers
various sub-menus used for instrument setup. TEST offers various
sub-menus for Test, Calibration and System Start-up.
Submenu GROUP NAVIGATION
Use the UP and DOWN arrow keys to navigate up and down through
the Sub-Menu groups when in the SETUP or TEST mode. Press the
ENTER key to enter a desired setup or test Sub-Menu group.
SELECTION OF ITEM
During setup, the unit will often offer multiple choices for a given topic.
The topic prompt appears on the top line of the display. The choices are
shown on the lower line of the display.
To select an item, press the key beneath the desired choice. The
selected choice will blink. Press the ENTER key to accept the
selected choice.
How To Enter
Numeric Values
How To Enter
Text Characters
NUMERIC ENTRY
The keys labeled "0 - 9", "–", ".", CLEAR and ENTER are used to enter
numerical values. A leading 0 will assume that you intend to enter a
minus "–" sign. Press the CLEAR key to clear the existing value and to
enable editing.
TEXT CHARACTER ENTRY
Some setup items (i.e. Descriptors, Units Label) require the user to
enter text characters. Press CLEAR to enable editing. The UP and
DOWN arrow keys are used to scroll through the available character
sets for each individual character. Press the ENTER key to accept the
character and advance to the next character.
19
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.