KEP 531, 532 Data sheet

531 Series
Features
• Compact and Low-Cost Temperature Display
• Temperature Display in °C or °F
• EEPROM Data Backup on Power Failure
• Galvanic Isolation with Reverse Polarity Protection
Temperature Display for Pt100 and Ni100 RTD's
• Screw Terminal Connectors: pitch 5 mm
• Display Hold Input
Specifications:
Supply voltage: 10-30 V DC, galvanically isolated with
Current draw: max. 40 mA Display: 5-digit display, red LED‘s; height 8 mm Measuring rate: 5 measurements/second Display refresh: 1-2 times per second Data backup: EEPROM Housing: housing for control panel 48 x 24 mm acc. to
Ambient temp.: –20 to +65 °C EMC: according to EC EMC directive 89/36/EEC Interference emissions:
Interference resistance:
Protection: NEMA4 / IP65 (front) Weight: app. 50 g Circuit type: 2-wire, 3-wire and 4-wire connection
DIGITAL PANEL METERS
Input: Pt100 or Ni100 RTD with sensor breakage
Wiring:
reverse polarity protection
DIN 43 700; RAL 7021, dark grey
EN 50081-2/EN 55 011 Class B
EN 6100-6-2
technique, programmable
monitoring
Power Supply
1 2 3 4 5 6 7
2-Wire RTD 3-Wire RTD 4-Wire RTD
1 2 3 4 5 6 7 1 2 3 4 5 6 7 1 2 3 4 5 6 7
• Easy Programming and Operation
•5 Measurements/second
Control inputs: High: 4-30 V DC, Low: 0-2 V DC Supply current: 1 mA Supply line: 2-wire: max 20 , programmable 3-wire, 4-
wire: max 20 , no balancing required
Temp. ranges: Pt100 acc. to DIN IEC 751:
–199.9 °C to +850.0 °C –327.8 °F to +1562.0 °F
Ni100 acc. to DIN 43760:
–60.0 °C to +250.0 °C
–76.0 °F ... +482.0 °F Resolution: 0.1°C (0.1°F) or 1°C (1°F) Linearity error: Pt100 < 0.1 % for entire measuring range at
an ambient temperature of 20 °C Ni100 < 0.2 % for entire measuring range at an ambient temperature of 20 °C
Temp. drift: 0.1 K/KAmbient
Order #:
531 = Temperature Display with RTD Input
Accessories
N7 - Explosion proof housing (see accessories section) E200 - Outdoor Enclosure (see accessories section)
3
3 • Latch input 2 • 0 V DC (GND)
2
1 • 10 to 30 V DC
1
Panel Cutout: 0.876" x 1.78" (22.3 x 45.2mm) or 0.99" x 1.97" (25 x 50mm) with adaptor provided
.1.88"
0.944"
(24mm)
(48mm)
1.23456
0.866"x1.77" (22x45mm)
2.32"
(59mm)
0.759"max. (19.3mm)
0.157" (4mm)
0.255"
(6.5mm)
234567
1
Page 92 • Industrial Instruments • II-15 Kessler-Ellis Products • 800-631-2165
532 Series
Features
• Compact and Low-Cost Temperature Display
• Temperature Display in °C or °F
• MIN/MAX Value Retention
• EEPROM Data Backup on Power Failure
• Galvanic Isolation with Reverse Polarity Protection
Temperature Display for J, K and N Thermocouples
• Screw Terminal Connectors: pitch 5 mm
• Display Hold Input
•5 Measurements/second
Specifications:
Supply voltage: 10-30 V DC, galvanically isolated with
reverse polarity protection Current draw: max. 40 mA Display: 5-digit display, red 7-segment LED‘s; height
8 mm Measuring rate: 5 measurements/second Display refresh: 1-2 times per second Data backup: EEPROM Housing: housing for control panel 48 x 24 mm acc. to
DIN 43 700; RAL 7021, dark grey Ambient temp.: –20 to +65 °C EMC: according to EC EMC directive 89/36/EEC Interference emissions:
EN 50081-2/EN 55 011 Class B Interference resistance:
EN 6100-6-2 Protection: NEMA4 / IP65 (front) Weight: app. 50 g Input: Thermocouple Sensor
J (Fe-CuNi)
K (Ni-CrNi)
N (NiCrSi-NiSi)
with sensor breakage monitoring Control inputs: High: 4-30 V DC, Low: 0-2 V DC Supply current: 1 mA Supply line: 2-wire: max 20 , programmable 3-wire, 4-
wire: max 20 , no balancing required
• J, K, N Thermocouples with External or Internal Cold Junction Compensation
• Easy Programming and Operation
Temp. ranges: according to DIN IEC 584
Resolution: 0.1°C (0.1°F) or 1°C (1°F) Linearity error: < 0.4 % for entire measuring range at an
Cold junction error:
Temp. drift: 0.1 K/KAmbient
Wiring:
Order #:
532 = Temperature Display with thermocouple Input
Accessories
N7 - Explosion proof housing (see accessories section) E200 - Outdoor Enclosure (see accessories section)
J (Fe-CuNi) –210.0 °C to +1200.0 °C
K (Ni-CrNi) –200.0 °C ... +1372.0 °C
N (NiCrSi-NiSi) –200.0 °C ... +1300.0 °C
ambient temperature of 20 °C
±1.0 °C typ. ±3.0 °C
Power Supply
1 2 3 4 5 6 7
3
3 • Latch input 2 • 0 V DC (GND)
2
1 • 10 to 30 V DC
1
–346.0°F ... +2192.0 °F
–328.0 °F ... +2501.6 °F
–328.0 °F ... +2370.0 °F
2-Wire RTD
1 2 3 4 5 6 7
n.c.
n.c.
+–
DIGITAL PANEL METERS
Panel Cutout: 0.876" x 1.78" (22.3 x 45.2mm) or 0.99" x 1.97" (25 x 50mm) with adaptor provided
.1.88"
0.944"
(24mm)
(48mm)
1.23456
Kessler-Ellis Products • 800-631-2165 II-15 • Industrial Instruments • Page 93
0.866"x1.77" (22x45mm)
2.32"
(59mm)
0.759"max. (19.3mm)
0.157" (4mm)
0.255"
(6.5mm)
234567