Microsoft, Visual C++, Excel, and Windows are either registered trademarks or trademarks of
Microsoft Corporation in the United States and/or other countries.
Document number: DAQ6510-900-01Rev. A / April 2018
Safety precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by personnel who recognize shock hazards and are familiar with the safety precautions required
to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the
product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained
service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are measurement, control, and data I/O
connections, with low transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high
transient overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient
overvoltages often associated with local AC mains connections. Certain Keithley Instruments measuring instruments may be
connected to mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any
unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If
the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective
devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth)
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the
voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or
accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be
impaired.
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are defined in the
specifications and operating information and shown on the instrument panels, test fixture panels, and switching cards.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground)
connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
If a
The symbol on an instrument means caution, risk of hazard. The user must refer to the operating instructions located in the
user documentation in all cases where the symbol is marked on the instrument.
The
symbol on an instrument means warning, risk of electric shock. Use standard safety precautions to avoid personal
contact with these voltages.
The
symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this
symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be
properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains hazards that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
The CAUTION heading with the
symbol in the user documentation explains hazards that could result in moderate or minor
injury or damage the instrument. Always read the associated information very carefully before performing the indicated
procedure. Damage to the instrument may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national
safety approvals may be used if the rating and type are the same. The detachable mains power cord provided with the
instrument may only be replaced with a similarly rated power cord. Other components that are not safety-related may be
purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be
purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the
applicability of a replacement component, call a Keithley Instruments office for information.
Unless otherwise noted in product-specific literature, Keithley Instruments instruments are designed to operate indoors only, in
the following environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution
degree 1 or 2.
To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the
instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that
consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board
should be returned to the factory for proper cleaning/servicing.
Thank you for choosing a Keithley Instruments product. The DAQ6510 is a 6½ digit graphical
sampling data-acquisition system that provides high-speed digitizing and a large graphical color
touchscreen display. This DAQ offers a broad range of measurement capabilities, including 15
measurement functions. In addition to industry-leading DC accuracies, functions such as capacitance,
3 A current, and 16-bit current and voltage digitizing are included. Tying all these features together is
a large 5-inch color touchscreen display that brings users an unprecedented combination of data
visualization and interaction, enabling users to gain deeper insight into their measurements.
Contact information .................................................................. 1-2
Documentation set ................................................................... 1-2
Organization of manual sections .............................................. 1-2
The DAQ6510 provides superior measurement accuracy and the speed necessary for a broad range
of applications including system, production testing, and benchtop applications. The DAQ6510 meets
application requirements for production engineers, research and development engineers, test
engineers, and scientists.
Introduction to this manual
This manual provides detailed applications to help you achieve success with your Keithley
Instruments DAQ6510. In addition, this manual provides information about the basics of the front
panel to familiarize you with the instrument.
This manual presents an overview of each application, followed by instructions to complete the
application using the front panel, SCPI code, TSP code, or Keithley KickStart Startup Software.
More information about the commands that are used in these applications is available. Refer to the
SCPI and TSP command reference sections of the Model DAQ6510 Reference Manual. This manual
is on tek.com/keithley
.
Section
User's Manual
1: Introduction DAQ6510 Data Acquisition / Multimeter System
Extended warranty
Additional years of warranty coverage are available on many products. These valuable contracts
protect you from unbudgeted service expenses and provide additional years of protection at a fraction
of the price of a repair. Extended warranties are available on new and existing products. Contact your
local Keithley Instruments office, sales partner, or distributor for details.
Contact information
If you have any questions after you review the information in this documentation, please contact your
local Keithley Instruments office, sales partner, or distributor. You can also call the corporate
headquarters of Keithley Instruments (toll-free inside the U.S. and Canada only) at 1-800-935-5595,
or from outside the U.S. at +1-440-248-0400. For worldwide contact numbers, visit tek.com/keithley
Documentation set
The DAQ6510 documentation is supported by the following technical/product information
documentation.
•Quick Start Guide: Provides unpacking instructions, describes basic connections, reviews basic
operation information, and provides a quick test procedure to ensure the instrument is
operational.
•User’s Manual: Provides application examples that you can use as a starting point to create your
own applications.
•Reference Manual: Includes advanced operation topics, maintenance information,
troubleshooting procedures, and in-depth descriptions of programming commands.
•Accessories information: Documentation for accessories that are available for the DAQ6510.
.
Organization of manual sections
This manual is organized into the following sections:
•Using the front-panel interface: (on page 2-1
interface.
•Using a remote interface: (on page 3-1) Describes the basics of remote communications and
using the instrument web interface.
•Application examples (see below): Provides detailed examples of how to use the DAQ6510 in
some typical situations.
For the latest drivers and additional support information, see tek.com/keithley
•Troubleshooting FAQs: (on page 11-1
you troubleshoot common problems encountered with the DAQ6510.
•Next steps: (on page 12-1) Provides information about additional resources that can help you use
the DAQ6510.
The PDF version of this manual contains bookmarks for each section. The manual sections are also
listed in the Table of Contents at the beginning of this manual.
For more information about bookmarks, see Adobe
) Provides answers to frequently asked questions to help
.
) Describes the basics of using the front-panel
®
Acrobat® or Reader® help.
1-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Introduction
Data Acquisition / Multimeter System User's Manual Section 1:
Application examples
This manual provides application examples that show you how to perform tests from the front panel
and over a remote interface. The applications include:
•Making basic front-panel measurements: (on page 4-1
using a single DAQ6510 and a two-terminal device under test.
•Making basic scans: (on page 4-1) Shows the basic measure scan functionality using a single
DAQ6510.
•Scanning temperature using thermocouples: (on page 5-1) Shows how to use the DAQ6510 to
log thermocouple-based temperature measurement scans.
•Scanning low-level DCV: (on page 6-1) Shows how to use the DAQ6510 to accurately measure
DC voltage in a variety of ranges
•Scanning resistors using 4W measurement: (on page 7-1) Shows how to use the DAQ6510 to
accurately measure resistance devices
•Mixed function multi-channel scanning: (on page 8-1) Shows how to use the DAQ6510 to perform
complex multi-channel, mixed function scanning in a production-test environment.
•Speed scanning for increased production test throughput: (on page 9-1) Shows how selecting the
right multiplexer module can increase your productivity by reducing overall test time.
•Pre-scan monitoring: (on page 10-3) Shows how to configure the DAQ6510 to delay its scanning
function until the testing environment has reached a specific temperature.
) Shows the basic measure functionality
DAQ6510-900-01Rev. A / April 2018 1-3
In this section:
Menu overview ....................................................................... 2-16
Turns the instrument on or off. To turn the instrument on, press
The front panel of the DAQ6510 is shown below. Descriptions of the controls on the front panel follow
the figure.
Section 2
Front-panel overview
Figure 1: DMM6500 front panel
POWER switch
HOME key
and hold the power switch. To turn it off, press and hold the power
switch again. The LED is green when the instrument is on and the
Section
User's Manual
Opens the main menu. Press the icons on the main menu to open
Opens a menu of of preconfigured TSP scripts with a graphical
user interface.
overview information for the screen you are viewing.
Saves reading buffer data and screen snapshots to a USB flash
drive.
The DAQ6510 has a high-resolution, five-inch color touchscreen
Refer to Touchscreen display (on page 2-5) for details.
Selects the highlighted choice or allows you to edit the selected
field.
Returns to the previous screen or closes a dialog box. For
to the main menu screen.
function name on the screen.
Accesses trigger-related settings and operations. The action of
DAQ6510 Reference Manual.
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
MENU key
channel, measure, views, trigger, scripts, and system screens.
For details, refer to Menu overview (on page 2-16).
APPS key
HELP key
Opens help for the area or item that is selected on the display. If
there is no selection when you press the HELP key, it displays
USB port
drive. You can also store and retrieve scripts to and from a USB
flash drive. The flash drive must be formatted as a FAT or FAT32
Touchscreen
display. The touchscreen accesses swipe screens and menu
options. You can access additional interactive screens by
pressing the front-panel MENU, APPS, and FUNCTION keys.
ENTER key
EXIT key
example, press the EXIT key when the main menu is displayed to
return to the Home screen. When you are viewing a subscreen
(for example, the Event Log screen), press the EXIT key to return
FUNCTION key
Displays instrument functions. To select a function, touch the
TRIGGER key
the TRIGGER key depends on the instrument state. For details,
see "Switching between measurement methods" in the Model
2-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
panel overview
Use the SENSE HI and SENSE LO terminals and the INPUT
INPUT terminals
Use the INPUT HI and INPUT LO terminals for all measurements
AMPS
Use the AMPS connection with the INPUT LO terminal to
Activates the terminals on the front or rear panel. Selecting the
LED is visible.
Data Acquisition / Multimeter System User's Manual Section 2: Front-
SENSE terminals
terminals with the 4-wire resistance, 3-wire and 4-wire RTD
temperature, and DC voltage ratio functions.
except current.
measure ≤3A DC or AC
RMS
current.
TERMINALS
switch
Instrument power
Follow the steps below to connect the DAQ6510 to line power and turn on the instrument. The
DAQ6510 operates from a line voltage of 100 V to 240 V at a frequency of 50 Hz or 60 Hz. Make sure
the operating voltage in your area is compatible. The default setting is 240V.
Before plugging in the line cord, check the line setting block on the rear of the instrument. If it doesn’t
match their mains installation setting, remove the block and reinsert to select voltage.
You must turn on the DAQ6510 and allow it to warm up for at least 30 minutes to achieve rated
accuracies.
Operating the instrument on an incorrect line voltage may cause damage to the instrument,
possibly voiding the warranty.
rear panel provides the proper connections to an inserted
moduleWhen the front-panel terminals are active, the green LED
is visible. When the rear-panel terminals are active, the amber
DAQ6510-900-01Rev. A / April 2018 2-3
Section
User's Manual
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
The power cord supplied with the DAQ6510 contains a separate protective earth (safety
ground) wire for use with grounded outlets. When proper connections are made, the
instrument chassis is connected to power-line ground through the ground wire in the power
cord. In the event of a failure, not using a properly grounded protective earth and grounded
outlet may result in personal injury or death due to electric shock.
Do not replace detachable mains supply cords with inadequately rated cords. Failure to use
properly rated cords may result in personal injury or death due to electric shock.
Connect the power cord
To connect the power cord:
1. Make sure that the front-panel POWER switch is in the off (O) position.
2. Connect the female end of the supplied power cord to the AC receptacle on the rear panel.
3. Connect the male end of the power cord to a grounded AC outlet.
Turn the DAQ6510 on or off
Hazardous voltages may be present on all output and guard terminals. To prevent electrical
shock that could cause injury or death, remove power from the instrument or test system and
discharge any energy storage components (for example, capacitors or cables) before
changing any connections that might allow contact with an uninsulated conductor.
On some sensitive or easily damaged devices under test (DUTs), the instrument power-up or powerdown sequence can apply transient signals to the DUT that may affect or damage it.
When testing this type of DUT, do not make final connections to it until the instrument has completed
its power-up sequence and is in a known operating state. Disconnect it from the instrument before
turning the instrument off.
To prevent any human contact with a live conductor, connections to the DUT must be fully insulated
and the final connections to the DUT must only use safety-rated safety jack socket connectors that
do not allow bodily contact.
To turn a DAQ6510 on:
1. Disconnect any devices under test (DUTs) from the DAQ6510.
Press and hold the front-panel POWER switch to place it in the on position.
The instrument displays a status bar as the instrument powers on. The Home screen is displayed
when power on is complete.
To turn a DAQ6510 off:
1. Press and hold the front-panel POWER switch to place it in the off position.
2-4 DAQ6510-900-01Rev. A / April 2018
DAQ6510
panel overview
Data Acquisition / Multimeter System User's Manual Section 2: Front-
Touchscreen display
The touchscreen display gives you quick front-panel access to measure settings, system
configuration, instrument and test status, reading buffer information, and other instrument
functionality. The display has multiple swipe screens that you can access by swiping the front panel.
You can access additional interactive screens by pressing the front-panel MENU, APPS, and
FUNCTION keys.
Do not use sharp metal objects, such as tweezers or screwdrivers, or pointed objects, such
as pens or pencils, to touch the touchscreen. It is strongly recommended that you use only
fingers to operate the instrument. Use of clean-room gloves to operate the touchscreen is
supported.
Select items on the touchscreen
To select an item on the displayed screen:
• Press the corresponding icon on the screen.
Scroll bars
Some of the interactive screens have additional options that are only visible when you scroll down the
screen. A scroll indicator on the right side of the touchscreen identifies these screens. Swipe the
screen up or down to view the additional options.
The figure below shows a screen with a scroll bar.
Figure 2: Scroll bar
The following topics describe the DAQ6510 touchscreen in more detail.
DAQ6510-900-01Rev. A / April 2018 2-5
Section
User's Manual
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
Enter information
Some of the menu options open a keypad or keyboard that you can use to enter information. For
example, if you are creating a new reading buffer from the front panel, you see the keypad shown in
the following figure.
Figure 3: DAQ6510 front-panel keyboard for information entry
You can enter information by touching the screen to select characters and options from the keypad or
keyboard. You can move the cursor in the entry box by touching the screen. The cursor is moved to
the spot in the entry box where you touched the screen.
Adjust the backlight brightness and dimmer
You can adjust the brightness of the DAQ6510 touchscreen display and buttons from the front panel
or over a remote interface. You can also set the backlight to dim after a specified time has passed
with no front-panel activity (available from the front-panel display only). The backlight settings set
through the front-panel display are saved through a reset or power cycle.
Screen life is affected by how long the screen is on at full brightness. The higher the brightness
setting and the longer the screen is bright, the shorter the screen life.
To adjust the backlight brightness from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select the Backlight Brightness. The Backlight Brightness dialog box opens.
4. Drag the adjustment to set the backlight.
5. Select OK.
2-6 DAQ6510-900-01Rev. A / April 2018
DAQ6510
panel overview
Data Acquisition / Multimeter System User's Manual Section 2: Front-
To set the backlight dimmer from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select Backlight Dimmer. The Backlight Dimmer dialog box opens.
4. Select a dimmer setting.
Review event messages
During operation and programming, front-panel messages may be briefly displayed. Messages are
either information, warning, or error notifications. For information on event messages, refer to "Using
the event log" in the Model DAQ6510 Reference Manual.
Figure 4: Example front-panel event message
Interactive swipe screens
The DAQ6510 touchscreen display has multiple screens that you can access by swiping left or right
on the lower half of the display. The options available in the swipe screens are described in the
following topics.
Swipe screen heading bar
The heading bar of the swipe screen contains the following options.
Figure 5: DAQ6510 swipe screens, maximized and minimized
DAQ6510-900-01Rev. A / April 2018 2-7
Section
User's Manual
2
Swipe screen indicator
Each circle represents one swipe screen. As you swipe right or left, a different
3
Calculations shortcut
Select to open the CALCULATION SETTINGS menu. Only available when
4
Measure Settings
Select to open the MEASURE SETTINGS menu for the selected function. Only
5
Restore indicator
Indicates that you can swipe up to display the swipe screen.
6
Graph shortcut
Select to open the Graph screen.
Scan shortcut
Not shown. Select to open the SCAN screen.
Channel control shortcut
Not shown. Select to open the CHANNEL CONTROL screen.
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
# Screen element Description
1 Minimize indicator You can swipe down to minimize the swipe screens.
circle changes color, indicating where you are in the screen sequence. Select a
circle to move the swipe screen without swiping.
TERMINALS is set to FRONT.
shortcut
Channel Settings
available when TERMINALS is set to FRONT.
Not shown. Select to open the CHANNEL SETTINGS screen.
shortcut
2-8 DAQ6510-900-01Rev. A / April 2018
DAQ6510
panel overview
Data Acquisition / Multimeter System User's Manual Section 2: Front-
FUNCTIONS swipe screen
The FUNCTIONS swipe screen highlights the selected measure function and allows you to select a
different function.
Figure 6: FUNCTIONS swipe screen
SETTINGS swipe screen
The SETTINGS swipe screen gives you front-panel access to some instrument settings for the
selected measure function. It shows you the present settings and allows you to change them. The
available settings depend on which measure function is active.
Figure 7: SETTINGS swipe screen
To disable or enable a setting, select the box next to the setting so that it shows an X (disabled) or a
check mark (enabled).
DAQ6510-900-01Rev. A / April 2018 2-9
Section
User's Manual
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
STATISTICS swipe screen
The STATISTICS swipe screen contains information about the readings in the active reading buffer.
When the reading buffer is configured to fill continuously and overwrite old data with new data, the
buffer statistics include the data that was overwritten. To get statistics that do not include data that
has been overwritten, define a large buffer size that will accommodate the number of readings you
will make. You can use the Clear Active Buffer button on this screen to clear the data from the
active reading buffer.
Figure 8: STATISTICS swipe screen
2-10 DAQ6510-900-01Rev. A / April 2018
DAQ6510
panel overview
Data Acquisition / Multimeter System User's Manual Section 2: Front-
SECONDARY swipe screen
The SECONDARY swipe screen allows you to display the results of two measurements on the
front-panel display.
To start displaying secondary measurements, select the Second Function and select Secondary
Measure. Secondary measurements are only available in Continuous Measurement Mode and
Manual Trigger Mode. This feature is only available from the front panel of the instrument.
Refer to "Display results of two measure functions" in the Model DAQ6510 Reference Manual.
Figure 9: SECONDARY swipe screen
Depending on the selected functions, a relay may click when the instrument switches between the
measurement types. Leaving secondary measurements on for extended periods may shorten the life
of the relays.
DAQ6510-900-01Rev. A / April 2018 2-11
Section
User's Manual
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
USER swipe screen
If you program custom text, it is displayed on the USER swipe screen. For example, you can program
the DAQ6510 to show that a test is in process. Refer to "Customizing a message for the USER swipe
screen" in the DAQ6510 Reference Manual. This swipe screen is only displayed if custom text has
been displayed. Refer to "Customizing a message for the USER swipe screen" in the Model DAQ6510 Reference Manual.
Figure 10: USER swipe screen
GRAPH swipe screen
The GRAPH swipe screen shows a graphical representation of the readings in the presently selected
reading buffer.
Figure 11: GRAPH swipe screen
2-12 DAQ6510-900-01Rev. A / April 2018
DAQ6510
panel overview
Data Acquisition / Multimeter System User's Manual Section 2: Front-
To view the graph in the full screen and to access graph settings, select the graph icon on the right
side of the swipe screen header. You can also open the full-function Graph screen by pressing the
MENU key and selecting Graph under Views.
For more information about graphing measurements, see "Graphing" in the Model DAQ6510 Reference Manual.
DAQ6510-900-01Rev. A / April 2018 2-13
Section
User's Manual
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
SCAN swipe screen
The SCAN swipe screen gives you front-panel access to build a scan, edit a scan, start a scan, step
through a scan, and display scan results. You can also save the scan results to a USB flash drive.
The icon on the right side of the swipe screen heading bar is a shortcut to the Channel Scan menu.
You can also use the Channel Scan menu to build or edit a scan.
For more information about viewing a scan preview and editing or running a scan, see "Channel scan
menu" in the Model DAQ6510 Reference Manual.
This screen is only available when the rear terminals are selected.
Figure 12: SCAN swipe screen - initial view
Figure 13: SCAN swipe screen - scan results
2-14 DAQ6510-900-01Rev. A / April 2018
DAQ6510
rview
scan.
flash drive.
Runs a scan.
Incrementally steps through a scan, channel by channel.
Data Acquisition / Multimeter System User's Manual Section 2: Front-panel ove
The scan swipe screen has the following control options:
Button Description
Build Scan
Edit
Opens the SCAN screen, where you can set up a new scan.
Opens the SCAN screen, where you can change the setup of a
Save to USB
Saves the data in the scan reading buffer toa CSV file on the USB
Start Scan
Step Scan
DAQ6510-900-01Rev. A / April 2018 2-15
Section
User's Manual
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
Menu overview
To access the main menu, press the MENU key on the DAQ6510 front panel. The figure below shows
the organization of the main menu. The options change depending upon which set of terminals are
selected, front or rear.
Figure 14: DAQ6510 main menu, front terminals selected
Figure 15: DAQ6510 main menu, rear terminals selected
The main menu includes submenus that are labeled in green across the top of the display. Selecting
an option in a submenu opens an interactive screen.
2-16 DAQ6510-900-01Rev. A / April 2018
DAQ6
panel overview
510 Data Acquisition / Multimeter System User's Manual Section 2: Front-
Channel menu
The Channel menus allow you to set up and control channels and scans from the front panel.
The Channel Settings menu allows you to select and configure channels.
The Channel Control menu contains options to open and close channels.
The Channel Scan menu contains options to set up and run scans. Options include
control of groups, which are channels that are sequential and have the same
functions applied to them.
Measure Menu
The Measure menus allow you to select, configure, and perform measure operations from the front
panel. The menu content depends upon which set of terminals is selected.
With front terminals selected:
The QuickSet menu allows you to change the function and adjust performance.
The Measure Settings menu contains settings for the presently selected measure
function, which is identified by the function indicator in the upper right corner of the
menu. The available settings depend on the front-panel FUNCTION key selection.
The Calculations menu contains settings that specify the way measurement
information is processed and returned.
The Reading Buffers menu allows you to view the list of existing reading buffers
and select one to be the active buffer. You can also create, save, delete, resize,
and clear buffers from this screen.
With rear terminals selected:
The Reading Buffers menu allows you to view the list of existing reading buffers
and select one to be the active buffer. You can also create, save, delete, resize,
and clear buffers from this screen.
DAQ6510-900-01Rev. A / April 2018 2-17
Section
User's Manual
2: Front-panel overview DAQ6510 Data Acquisition / Multimeter System
Views menu
The Views menus allow you to select, configure, and view data that was gathered from measure
operations.
The Graph menu opens a screen that displays a graph of the measurements in
selected reading buffers as traces. It also contains tabs that you use to customize
the graph display.
You can also select the trigger mode and initiate the trigger model from this screen.
The Histogram menu allows you to graph the distribution of measurement data in
the selected reading buffer. It also contains tabs that you use to customize the
histogram.
The Reading Table menu allows you to view data in the selected reading buffer.
Trigger menu
The Trigger menus allow you to configure the trigger model from the front panel.
The Templates menu allows you to choose from one of several preprogrammed
trigger models. When you select a template, settings you can specify for that
template are shown in the lower part of the screen.
The Configure menu allows you to view and modify the structure and parameters
of a trigger model. You can also monitor trigger model operation.
Scripts menu
The Scripts menus allow you to configure, run, and manage scripts from the front panel. Scripts are
blocks of commands that the instrument can run as a group.
The Run menu contains a list of scripts that you can select to run immediately. You
can also copy a script to a script that runs each time the instrument power is turned
on. You can access scripts that are in the instrument or on a USB flash drive.
The Manage menu allows you to copy scripts to and from the instrument and the
USB flash drive. You can also delete scripts from the instrument or USB flash drive.
The Save Setup menu allows you to save the present settings and configuration
lists of the instrument into a configuration script. You can use this script to recall the
settings.
The options in the Record menu allow you to record your actions and store them in
a macro script. The script can be run and managed like any other script using the
options in the Scripts menu or remote commands. Note that only settings are
stored; no key presses or front-panel only options are stored.
2-18 DAQ6510-900-01Rev. A / April 2018
DAQ6510
panel overview
Data Acquisition / Multimeter System User's Manual Section 2: Front-
System menu
The menus under System in the main menu allow you to configure general instrument settings from
the DAQ6510 front panel. Among these settings are the event log, communications, backlight, time,
and password settings, calibration, and info/manage.
The Event Log menu allows you to view and clear event log entries. You can also
adjust which events are displayed or logged.
The Communication menu opens a set of tabs that contain information about the
communications settings. Most of the tabs contain settings that you can change.
The Settings menu contains general instrument settings. It includes beeper and
key click, backlight brightness and timer, time and date, system access level,
password, and reading format settings.
The Calibration menu allows you to start or manage auto calibration. Auto
calibration removes measurement errors that are caused by the effects of
temperature and time on components. You can also review factory adjustment and
verification dates.
The Info/Manage menu gives you access to version and serial number information
and settings for instrument firmware and reset functions.
DAQ6510-900-01Rev. A / April 2018 2-19
Determining the command set you will use ............................ 3-18
Using the web interface .......................................................... 3-15
Remote communications interfaces
Section 3
Using a remote interface
You can choose from one of several communication interfaces to send commands to and receive
responses from the DAQ6510.
The instrument automatically detects the type of communications interface (LAN, USB, GPIB,
RS-232, or TSP-Link) when you connect to the respective port on the rear panel of the instrument.
The GPIB, RS-232, and TSP-Link options require an optional accessory card. In most cases, you do
not need to configure anything on the instrument. In addition, you do not need to reboot if you change
the type of interface that is connected.
You can only use one communications interface to control the DAQ6510 at a time. The USB
connection takes precedence over LAN connections. For other communications interfaces, the first
interface on which the instrument receives a message takes control of the instrument. If another
interface sends a message, that interface can take control of the instrument. You may need to enter a
password to change the interface, depending on the access mode.
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
Supported remote interfaces
The DAQ6510 supports the following remote interfaces:
• Ethernet: Built-in local area network (LAN) ethernet communications.
• USB: Built-in USB communications port (Type B).
• GPIB: IEEE-488 instrumentation general purpose interface bus.
• RS-232: Serial communication data standard.
• TSP-Link: A high-speed trigger synchronization and communications bus that test system
builders can use to connect multiple instruments in a master-and-subordinate configuration.
The GPIB, RS-232, and TSP-Link interfaces require an optional communications accessory.
For details about TSP-Link, see "TSP-Link System Expansion Interface" in the Model DAQ6510 Reference Manual.
Figure 16: DAQ6510 remote interface connections
LAN communications
You can communicate with the instrument using a local area network (LAN).
When you connect using a LAN, you can use a web browser to access the internal web page of the
instrument and change some of the instrument settings. For more information, see
interface (on page 3-15).
The DAQ6510 is an LXI version 1.5 Core 2016 compliant instrument that supports TCP/IP and
complies with IEEE Std 802.3 (ethernet LAN). There is one LAN port (located on the rear panel of the
instrument) that supports full connectivity on a 10 Mbps or 100 Mbps network. The DAQ6510
automatically detects the speed.
The DAQ6510 also supports Multicast DNS (mDNS) and DNS Service Discovery (DNS-SD), which
are useful on a LAN with no central administration.
3-2 DAQ6510-900-01Rev. A / April 2018
Using the web
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
Contact your network administrator to confirm your specific network requirements before setting up a
LAN connection.
If you have problems setting up the LAN, refer to LAN troubleshooting suggestions (on page 3-16
).
Set up LAN communications on the instrument
This section describes how to set up manual or automatic LAN communications on the instrument.
Check communication settings
Before setting up the LAN configuration, you can check the communication settings on the instrument
without making any changes.
To check communication settings on the instrument:
1. Press the MENU key.
2. Under System, select Communication. The SYSTEM COMMUNICATIONS window opens.
3. Select LAN to see the settings for that interface.
4. Press the EXIT key to leave the SYSTEM COMMUNICATION window without making any
changes.
Set up automatic LAN configuration
If you are connecting to a LAN that has a DHCP server or if you have a direct connection between the
instrument and a host computer, you can use automatic IP address selection.
If you select Auto, the instrument attempts to get an IP address from a DHCP server. If this fails, it
reverts to an IP address in the range of 169.254.1.0 through 169.254.254.255.
Both the host computer and the instrument should be set to use automatic LAN configuration.
Though it is possible to have one set to manual configuration, it is more complicated to set up.
To set up automatic IP address selection using the front panel:
1. Press the MENU key.
2. Under System, select Communication.
3. Select the LAN tab.
4. For TCP/IP Mode, select Auto.
5. Select Apply Settings to save your settings.
DAQ6510-900-01Rev. A / April 2018 3-3
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
Set up manual LAN configuration
If necessary, you can set the IP address on the instrument manually.
You can also enable or disable the DNS settings and assign a host name to the DNS server.
Contact your corporate information technology (IT) department to secure a valid IP address for the
instrument when placing the instrument on a corporate network.
The instrument IP address has leading zeros, but the computer IP address cannot.
To set up manual IP address selection on the instrument:
1. Press the MENU key.
2. Under System, select Communication.
3. Select the LAN tab.
4. For TCP/IP Mode, select Manual.
5. Enter the IP Address.
6. Enter the Gateway address.
7. Enter the Subnet mask.
8. Select Apply Settings to save your settings.
Set up LAN communications on the computer
This section describes how to set up the LAN communications on your computer.
Do not change your IP address without consulting your system administrator. If you enter an
incorrect IP address, it can prevent your computer from connecting to your corporate network or it
may cause interference with another networked computer.
Record all network configurations before modifying any existing network configuration information on
the network interface card. Once the network configuration settings are updated, the previous
information is lost. This may cause a problem reconnecting the host computer to a corporate
network, particularly if DHCP is disabled.
Be sure to return all settings to their original configuration before reconnecting the host computer to a
corporate network. Contact your system administrator for more information.
Wait for the LAN status indicator to turn green
Make sure that your DAQ6510 is connected to the network by confirming that your instrument was
assigned an IP address.
To verify the LAN connection:
1. Press the MENU key.
2. Under System, select Communication.
3. Select the LAN tab.
A green LAN status indicator on the lower left of the LAN tab confirms that your instrument was
assigned an IP address.
3-4 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
Use the LXI Discovery Tool
To find the IP address of the DAQ6510, use the LXI Discovery Tool, a utility that is available from the
Resources tab of the LXI Consortium website
USB communications
To use the rear-panel USB port, you must have the Virtual Instrument Software Architecture (VISA)
layer on the host computer. See "How to install the Keithley I/O Layer" in the Model DAQ6510 Reference Manual for more information.
VISA contains a USB-class driver for the USB Test and Measurement Class (USBTMC) protocol that,
once installed, allows the Microsoft Windows operating system to recognize the instrument.
When you connect a USB device that implements the USBTMC or USBTMC-USB488 protocol to the
computer, the VISA driver automatically detects the device. Note that the VISA driver only
automatically recognizes USBTMC and USBTMC-USB488 devices. It does not recognize other USB
devices, such as printers, scanners, and storage devices.
In this section, "USB instruments" refers to devices that implement the USBTMC or
USBTMC-USB488 protocol.
.
Connect a computer to the DAQ6510 using USB
To connect the DAQ6510 to a computer using a USB connection, use Keithley Instruments
Model USB-B-1.
Each DAQ6510 needs its own USB cable to be connected to the computer.
To connect an instrument to a computer using USB:
1. Connect the Type A end of the cable to the computer.
2. Connect the Type B end of the cable to the instrument.
3. Turn on the instrument power. When the computer detects the new USB connection, the Found
New Hardware Wizard starts.
4. If the "Can Windows connect to Windows Update to search for software?" dialog box opens, click
No, and then click Next.
5. On the "USB Test and Measurement device" dialog box, click Next, and then click Finish.
DAQ6510-900-01Rev. A / April 2018 3-5
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
Communicate with the instrument
For the instrument to communicate with the USB device, you must use NI-VISATM. VISA requires a
resource string in the following format to connect to the correct USB instrument:
USB0::0x05e6::0x6510::[serial number]::INSTR
Where:
• 0x05e6: The Keithley vendor ID
• 0x6510: The instrument model number
• [serial number]: The serial number of the instrument (the serial number is also on the rear
panel)
•INSTR: Use the USBTMC protocol
To determine these parameters, you can run the Keithley Configuration Panel, which automatically
detects all instruments connected to the computer.
If you installed the Keithley I/O Layer, you can access the Keithley Configuration Panel through the
®
Microsoft
Windows® Start menu.
If you have a USB connection, you cannot switch to a LAN connection while the USB is connected.
USB takes precedence over LAN.
To use the Keithley Configuration Panel to determine the VISA resource string:
1. Click Start > Keithley Instruments > Keithley Configuration Panel. The Select Operation
dialog box is displayed.
Figure 17: Select Operation dialog box
2. Select Add.
3-6 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
3. Select Next. The Select Communication Bus dialog box is displayed.
Figure 18: Select Communication Bus dialog box
4. Select USB.
5. Click Next. The Select Instrument Driver dialog box is displayed.
Figure 19: Select Instrument Driver dialog box
6. Select Auto-detect Instrument Driver - Model.
7. Click Next. The Configure USB Instrument dialog box is displayed with the detected instrument
VISA resource string visible.
DAQ6510-900-01Rev. A / April 2018 3-7
8. Click Next. The Name Virtual Instrument dialog box is displayed.
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
Figure 20: Name Virtual Instrument dialog box
9. In the Virtual Instrument Name box, enter a name that you want to use to refer to the instrument.
10. Select Finish.
11. Select Cancel to close the Wizard.
12. Save the configuration. From the Keithley Configuration Panel, select File > Save.
Verify the instrument through the Keithley Communicator:
2. Select File > Open Instrument to open the instrument you just named.
Figure 21: Keithley Communicator Open an Instrument
3. Click OK.
4. Send a command to the instrument and see if it responds.
3-8 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
If you have a full version of NI-VISA on your system, you can run NI-MAX or the VISA Interactive
Control utility. See the National Instruments documentation for information.
GPIB communications
The DAQ6510 GPIB interface is IEEE Std 488.1 compliant and supports IEEE Std 488.2 common
commands and status model topology.
You can have up to 15 devices connected to a GPIB interface, including the controller. The maximum
cable length is the lesser of either:
• The number of devices multiplied by 2 m (6.5 ft)
• 20 m (65.6 ft)
You may see erratic bus operation if you ignore these limits.
To use the GPIB interface your computer will need a GPIB card. Check the documentation for your
GPIB controller for installation information and where to acquire drivers.
Install the KTTI-GPIB accessory card
Figure 22: KTTI-GPIB connector view
Unpack and inspect
Make sure to handle the KTTI-GPIB carefully. Always grasp the card by the side edges. Do not touch
board surfaces, components, or areas adjacent to electrical contacts. Contamination from foreign
materials such as dirt, dust, and body oils can substantially degrade card performance.
To unpack and inspect your instrument:
1. Inspect the box for damage.
2. Open the box.
3. Remove the card and inspect for any obvious signs of physical damage.
4. Report any damage to the shipping agent immediately.
Installation
Slot covers must be installed on unused slots to prevent personal contact with high-voltage
circuits. Failure to recognize and observe standard safety precautions could result in
personal injury or death due to electric shock.
DAQ6510-900-01Rev. A / April 2018 3-9
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
To install the communications card:
1. Turn the instrument off and disconnect the power line cord and any other cables connected to the
rear panel.
2. Position the instrument so that you are facing the rear panel.
3. Remove the slot cover plate from the slot on the back of the instrument. Retain the plate and
screws for future use.
4. Align the card with the connector toward the inside edge of the slot and slide the card into the
chassis. For the last ¼ inch or so, press in firmly to mate the card to the connector.
5. On each side of the card, there is a spring-loaded mounting screw. Tighten these two screws,
either by hand or with a Phillips-head screwdriver, to secure the card in the case. Do not
overtighten.
6. Reconnect the power line cable and any other cables to the rear panel.
7. Turn on the instrument.
Connect GPIB cables to your instrument
To connect a DAQ6510 to the GPIB interface, use a cable equipped with standard GPIB connectors,
as shown below.
Figure 23: GPIB connector
To allow many parallel connections to one instrument, stack the connectors. Each connector has two
screws on it to ensure that connections remain secure. The figure below shows a typical connection
diagram for a test system with multiple instruments.
To avoid possible mechanical damage, stack no more than three connectors on any one
instrument. To minimize interference caused by electromagnetic radiation, use only shielded
GPIB cables. Contact Keithley Instruments for shielded cables.
3-10 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
Figure 24: DAQ6510 Instrument GPIB connections
Additional information
Additional information is available in the KTTI-GPIB Accessory Installation Sheet.
DAQ6510-900-01Rev. A / April 2018 3-11
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
Set the GPIB address
Set the GPIB address. The default GPIB address is 16. You can set the address from one to 30 if it is
unique in the system. This address cannot conflict with an address that is assigned to another
instrument or to the GPIB controller.
GPIB controllers are usually set 0 to 21. To be safe, do not configure any instrument to have an
address of 21.
The instrument saves the address in nonvolatile memory. It does not change when you send a reset
command or when you turn the power off and on again.
To set the GPIB address from the front panel:
1. Press the MENU key.
2. Select Communication.
3. Select the GPIB tab.
4. Set the GPIB address.
5. Select OK.
RS-232
Install the KTTI-RS232 accessory card
You can also set the GPIB address using remote commands. Set the GPIB address with the SCPI
command :SYSTem:GPIB:ADDRess or the TSP command gpib.address.
Unpack and inspect
Make sure to handle the KTTI-RS232 carefully. Always grasp the card by the side edges. Do not
touch board surfaces, components, or areas adjacent to electrical contacts. Contamination from
foreign materials such as dirt, dust, and body oils can substantially degrade card performance.
To unpack and inspect your instrument:
1. Inspect the box for damage.
2. Open the box.
3. Remove the card and inspect for any obvious signs of physical damage.
4. Report any damage to the shipping agent immediately.
Installation
Slot covers must be installed on unused slots to prevent personal contact with high-voltage
circuits. Failure to recognize and observe standard safety precautions could result in
personal injury or death due to electric shock.
3-12 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
1
No connection
2
TXD, transmit data
3
RXD, receive data
4
No connection
5
GND, signal ground
6
No connection
7
No connection
8
No connection
9
No connection
Data Acquisition / Multimeter System User's Manual Section 3:
To install the communications card:
1. Turn the instrument off and disconnect the power line cord and any other cables connected to the
rear panel.
2. Position the instrument so that you are facing the rear panel.
3. Remove the slot cover plate from the slot on the back of the instrument. Retain the plate and
screws for future use.
4. Align the card with the connector toward the inside edge of the slot and slide the card into the
chassis. For the last ¼ inch or so, press in firmly to mate the card to the connector.
5. On each side of the card, there is a spring-loaded mounting screw. Tighten these two screws,
either by hand or with a Phillips-head screwdriver, to secure the card in the case. Do not
overtighten.
6. Reconnect the power line cable and any other cables to the rear panel.
7. Turn on the instrument.
Making connections
The RS-232 serial port can be connected to the serial port of a controller using a straight-through
RS232 cable terminated with DB-9 connectors. Do not use a null modem cable.
The serial port uses the transmit (TXD), receive (RXD), and signal ground (GND) lines of the RS232
standard. The figure below shows the rear panel connector for the RS232 interface. The table below
shows the pinouts for the connector.
Pin Description
Figure 25: KTTI-RS232 panel view
Figure 26: RS-232 panel connector
DAQ6510-900-01Rev. A / April 2018 3-13
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
Additional information
TSP-Link
Install the KTTI-TSP accessory card
Unpack and inspect
Make sure to handle the KTTI-TSP carefully. Always grasp the card by the side edges. Do not touch
board surfaces, components, or areas adjacent to electrical contacts. Contamination from foreign
materials such as dirt, dust, and body oils can substantially degrade card performance.
To unpack and inspect your instrument:
1. Inspect the box for damage.
2. Open the box.
3. Remove the card and inspect for any obvious signs of physical damage.
Additional information is available in the KTTI-RS232 Accessory Installation Sheet.
4. Report any damage to the shipping agent immediately.
Installation
Slot covers must be installed on unused slots to prevent personal contact with high-voltage
circuits. Failure to recognize and observe standard safety precautions could result in
personal injury or death due to electric shock.
To install the communications card:
1. Turn the instrument off and disconnect the power line cord and any other cables connected to the
rear panel.
2. Position the instrument so that you are facing the rear panel.
3. Remove the slot cover plate from the slot on the back of the instrument. Retain the plate and
screws for future use.
4. Align the card with the connector toward the inside edge of the slot and slide the card into the
chassis. For the last ¼ inch or so, press in firmly to mate the card to the connector.
5. On each side of the card, there is a spring-loaded mounting screw. Tighten these two screws,
either by hand or with a Phillips-head screwdriver, to secure the card in the case. Do not
overtighten.
6. Reconnect the power line cable and any other cables to the rear panel.
7. Turn on the instrument.
3-14 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
Making connections
Connect the TSP-Link connector to one of the TSP-Link connectors on the rear panel of the
instrument.
The TSP-Link expansion interface uses CAT5 and RJ-45 connectors to connect up to 32 devices.
Figure 27: KTTI-TSP panel view
Additional information
Additional information is available in the KTTI-TSP Accessory Installation Sheet.
Using the web interface
The DAQ6510 web interface allows you to make settings and control your instrument through a web
page. The web page includes:
• Instrument status.
• The instrument model, serial number, firmware revision, and the last LXI message.
• An ID button to help you locate the instrument.
• A virtual front panel and command interface that you can use to control the instrument.
• Download access to a CSV file that contains reading buffer data.
• Administrative options and LXI information.
The instrument web page resides in the firmware of the instrument. Changes you make through the
web interface are immediately made in the instrument.
Connect to the instrument web interface
When the LAN and instrument establish a connection, you can open the web page for the instrument.
To access the web interface:
1. Open a web browser on the host computer.
2. Enter the IP address of the instrument in the address box of the web browser. For example, if the
instrument IP address is 192.168.1.101, enter 192.168.1.101 in the browser address box.
3. Press Enter on the computer keyboard to open the instrument web page.
4. If prompted, enter a user name and password. The default is admin for both.
DAQ6510-900-01Rev. A / April 2018 3-15
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
LAN troubleshooting suggestions
If you are unable to connect to the web interface of the instrument, check the following items:
•The network cable is in the LAN port on the rear panel of the instrument, not one of the
®
TSP-Link
ports.
•The network cable is in the correct port on the computer. The LAN port of a laptop may be
disabled when the laptop is in a docking station.
• The setup procedure used the configuration information for the correct ethernet card.
• The network card of the computer is enabled.
• The IP address of the instrument is compatible with the IP address on the computer.
• The subnet mask address of the instrument is the same as the subnet mask address of the
computer.
•There is a USB cable attached between the instrument and your computer. USB communications
take precedence over LAN.
You can also try restarting the computer and the instrument.
To restart the instrument:
1. Turn the instrument's power off, and then on.
2. Wait at least 60 seconds for the network configuration to be completed.
To set up LAN communications
1. Press the MENU key.
2. Under System, select Communication.
3. Select the LAN tab.
4. Verify the settings.
If the above actions do not correct the problem, contact your system administrator.
3-16 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
Web interface Home page
Figure 28: DAQ6510 web-interface home page
The Home page of the instrument provides information about the instrument. It includes:
• The instrument model number, manufacturer, serial number, and firmware revision number.
• The TCP Raw Socket number and Telnet Port number.
• The last LXI message. The history link opens the LXI Home page.
• The ID button, which allows you to identify the instrument. Refer to Identify the instrument
page 3-18).
(on
DAQ6510-900-01Rev. A / April 2018 3-17
Section
User's Manual
3: Using a remote interface DAQ6510 Data Acquisition / Multimeter System
Identify the instrument
If you have a bank of instruments, you can select the ID button to determine which one you are
communicating with.
Before trying to identify the instrument, make sure you have a remote connection to the instrument.
To identify the instrument:
1. On each instrument in the bank, select MENU, then Communication.
2. Select the LAN tab.
3. On the web interface Home or LXI Home page, select the ID button. The button turns green and
the LXI LAN indicator on the instrument LAN tab blinks.
4. Select the ID button again to turn off the ID feature.
Determining the command set you will use
You can change the command set that you use with the DAQ6510. The remote command sets that
are available include:
• SCPI: An instrument-specific language built on the SCPI standard.
• TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
•SCPI2700: An instrument-specific language that allows you to run code developed for Keithley
Instruments Series 2700 instruments.
•SCPI2701: An instrument-specific language that allows you to run code developed for Keithley
Instruments Series 2700 instruments.
You cannot combine the command sets.
As delivered from Keithley Instruments, the DAQ6510 is set to work with the DAQ6510 SCPI
command set.
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select Command Set.
You are prompted to reboot.
3-18 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Using a remote interface
Data Acquisition / Multimeter System User's Manual Section 3:
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
This example application makes a 2-wire resistance measurement using the front panel of the
instrument.
Section 4
Making basic front-panel measurements
Set the function before making other instrument settings. Many of the settings are related to a
specific measure function. The applications in this manual use the order of operations to produce the
best results.
Equipment required for this example
Equipment required to perform this test:
• One DAQ6510
• Two insulated banana cables.
• One resistor to test, the example uses a resistor with a 9.75 kΩ rating
Section
User's Manual
4: Making basic front-panel measurements DAQ6510 Data Acquisition / Multimeter System
Device connections
Connect the DAQ6510 to the resistor in a 2-wire (local sense) configuration. In this configuration, the
device is connected between the INPUT HI and INPUT LO terminals.
To make the connections:
1. Power off the DAQ6510.
2. Connect the test leads to the front-panel INPUT HI and INPUT LO terminals, as shown in the
following figure.
Data Acquisition / Multimeter System User's Manual Section 4: Making basic front-
Basic front-panel measurements
The following procedures show you how to make a measurement, access settings for the
measurement, and view measurement data in a reading buffer.
You can make measurements continuously or manually. When you make continuous measurements,
the instrument makes measurements as quickly as possible. When you make manual measurements,
the instrument makes measurements when you press the TRIGGER key.
To make a measurement from the front panel:
1. Press the POWER switch on the front panel to turn on the instrument.
2. Make sure the terminals switch is set to FRONT.
3. From the Functions swipe screen, select 2W Ω. Measurements begin to display on the top half of
the Home screen.
4. If measurements are not displayed, press the TRIGGER key for a few seconds and select
Continuous Measurement.
To change the measure settings:
1. Press the MENU key.
2. Under Measure, select Settings.
3. Select Display Digits.
4. Select 3.5 Digits.
5. Press the HOME key. The measurement now shows 3½ digits.
To make a single measurement:
1. Press the front-panel TRIGGER key for a few seconds.
2. Select Manual Trigger Mode.
3. Press the TRIGGER key to initiate a single reading using the selected measure function.
Figure 30: Basic measurement test results
DAQ6510-900-01Rev. A / April 2018 4-3
Section
User's Manual
Channel information.
Date and time showing the month, day, hour, minute, second, and
fraction of a second.
4: Making basic front-panel measurements DAQ6510 Data Acquisition / Multimeter System
The information that appears in the reading table may vary depending on the buffer you select and
whether you are using the front or rear panel to make measurements.
Using the front panel to view the contents of a reading buffer:
1. Press the MENU key.
2. Under Views, select Reading Table. Data for the active reading buffer is displayed.
Figure 31: Reading table
When you are using the FRONT terminals to make measurements, the Channel column of the
reading table displays "Front." If you are using the REAR terminals, the Channel column displays
"Rear." If you are using a card, the Channel column displays the channel number.
3. To display data for a different reading buffer, select the buffer.
4. To view details for a specific data point, swipe the table up or down and select the data point to
view the Reading Details. If there are many data points, select an area on the reading preview
graph in the upper right corner of the screen to get closer to the data you want, and then scroll to
the data point.
5. Press the HOME key to return to the Home screen.
Column Description
Channel
Index
Reading
Time
Index number of the reading.
Measurement value.
4-4 DAQ6510-900-01Rev. A / April 2018
Thermocouple temperature scanning ....................................... 5-3
This application example demonstrates how to use the DAQ6510 to log thermocouple-based
temperature measurement scans, using internal cold-junction compensation (CJC) correction, over a
24-hour period.
Section 5
Scanning temperature using thermocouples
This type of test is typically performed when a device under test (DUT) is placed in an environmental
chamber and exposed to extreme conditions. The system captures data at different locations on the
DUT. The data is then exported from the DAQ6510 to a computer where a thermal profile is
generated. This thermal profile provides designers and consumers with a thorough understanding of
the thermal operating characteristics of their device or product.
Equipment required
• One DAQ6510
• One Model 7700 20-channel differential multiplexer module
• One computer set up for communication with the instrument
• Ten Type K thermocouples
• One USB flash drive
• One device or component to be tested
Section
User's Manual
5: Scanning temperature using thermocouples DAQ6510 Data Acquisition / Multimeter System
Device connections
Connecting thermocouples to the instrument
1. Power off the DAQ6510.
2. Remove the 7700 from the DAQ6510.
3. Remove the top cover from the 7700.
4. Connect ten Type K thermocouples to each channel as shown.
5. Route the cables out through the channels in the 7700 and secure the top cover.
6. Ensure that the DAQ6510 power is turned off.
7. Insert the 7700 into a slot on the rear of the DAQ6510.
Figure 34: DAQ6510 with 7700 multiplexer module
5-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Scanning temperature using thermocouples
Data Acquisition / Multimeter System User's Manual Section 5:
8. Press the POWER switch on the front panel to turn on the instrument.
9. Set the TERMINALS switch to rear.
10. Plug the USB flash drive into the port on the front of the instrument.
The remaining connections are the thermocouples to the DUT, which typically use soldering, thermal
compound, polyimide tape, or conductive epoxy.
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads or any device under test (DUT) that is in contact with the
conductors. It is good practice to disconnect DUTs from the instrument before powering the
instrument. Safe installation requires proper shields, barriers, and grounding to prevent
contact with test leads.
There is no internal connection between protective earth (safety ground) and the LO
terminals of the DAQ6510. Therefore, hazardous voltages (more than 30 V
) can appear on
RMS
LO terminals. This can occur when the instrument is operating in any mode. To prevent
hazardous voltage from appearing on the LO terminals, connect the LO terminal to protective
earth (safety ground) if your application allows it. You can connect the LO terminal to the
chassis ground terminal on the front panel or the chassis ground screw terminal on the rear
panel. Note that the front-panel terminals are isolated from the rear-panel terminals.
Therefore, if you are using the front-panel terminals, ground to the front-panel LO terminal. If
using the rear-panel terminals, ground to the rear panel LO terminal. Failure to follow these
guidelines can result in injury, death, or instrument damage.
Thermocouple temperature scanning
This application demonstrates how to use the DAQ6510 to measure temperature at ten different
points on a DUT every minute over a 24-hour period. During this test, all scan data is automatically
written to a USB flash drive connected to the port on the instrument.
For this application, you will:
•Configure the instrument for ten channels of Type K thermocouple measurement using internal
cold junction correction (CJC).
• Enable detection of open leads and offset compensation for more accurate readings.
• Configure the instrument to scan all channels (1 to 10) once a minute for 24 hours.
• Automatically export the data to a USB flash drive for each completed scan.
• Enable the auto-restart feature to ensure the continuation of progress in the event of power
failure.
• Check the status of the scan progress as it is running.
• Pause the scan and move the data to a computer to begin analysis, prior to scan completion.
• Analyze graphical data of the scan in progress.
• Learn how to program the instrument through the front panel user interface.
• Learn how to program the instrument using a remote communications interface with SCPI and
TSP code commands.
DAQ6510-900-01Rev. A / April 2018 5-3
Section
User's Manual
5: Scanning temperature using thermocouples DAQ6510 Data Acquisition / Multimeter System
Using the front panel
To set up the application from the front panel:
1. Press the POWER switch on the front panel to turn on the instrument.
2. Press the MENU key.
3. Under Channel, select Scan.
4. Select the + button to add a group of channels (101 to 110), and select OK.
5. Select Temperature on the Measure Functions screen.
6. Select Thermocouple, select K.
7. Scroll down and set the Reference Junction to Internal.
8. Select the Scan tab, set the Scan Count to 1440, and select OK. (One scan every minute for 24
hours = 24 hours * 60 minutes = 1440).
9. Set the Interval Between Scans to 60 s and select OK.
10. Scroll down on the Scan tab to Export to a USB flash drive. Choose After Each Scan from the
list of options.
11. Select Filename, enter scan24hr, and select OK to accept.
12. Select OK to accept the remaining File Content settings.
13. Select Power Loss Restart and select On.
14. You can now start your scan in one of two ways:
a. Select the Start icon on the SCAN screen.
b. Press the TRIGGER key and select a measurement state. Choose Initiate Scan to start the scan.
15. Select View Scan Status to go to the SCAN swipe on the HOME screen.
5-4 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Scanning temperature using thermocouples
*RST
Reset the DAQ6510
:FUNCtion 'TEMPerature',(@101:110)
Set up channel settings for Slot 1
:SENSe:TEMPerature:TRANsducer
TCouple, (@101:110)
:SENSe:TEMPerature:TCouple:TYPE K,
(@101:110)
:SENSe:TEMPerature:TCouple:RJUNction:
RSELect INTernal, (@101:110)
:SENSe:TEMPerature:ODETector ON,
(@101:110)
:ROUTe:SCAN:CREate (@101:110)
Set up Scan
:ROUTe:SCAN:COUNt:SCAN 1440
= 1440
:ROUTe:SCAN:INTerval 60.0
Set the time between scans to 60 s
:ROUTe:SCAN:EXPORT
after each scan
:ROUTe:SCAN:RESTart ON
Enable scan restart after power failure
:INIT
Trigger to start scan
:DISPlay:SCReen SWIPE_SCAN
Sends user to the SCAN swipe screen
Data Acquisition / Multimeter System User's Manual Section 5:
Using SCPI commands
This sequence of SCPI commands makes a thermocouple-based temperature scan.
You may need to make changes so that this code will run in your programming environment.
Send the following commands for this example application:
Commands Descriptions
Set the scan count to 24 hrs * 60 min/hr
"/usb1/scan24hr.csv", SCAN, ALL
Ensure data gets written to a USB drive
DAQ6510-900-01Rev. A / April 2018 5-5
Section
User's Manual
5: Scanning temperature using thermocouples DAQ6510 Data Acquisition / Multimeter System
Using TSP commands
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from tek.com/keithley
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the Model DAQ6510 Reference Manual.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the DAQ6510 uses the SCPI command set. You must select the TSP command set before
sending TSP commands to the instrument.
To enable TSP commands:
1. Press the MENU key.
2. Under System, select Settings.
3. Set the Command Set to TSP.
4. At the prompt to reboot, select Yes.
This sequence of TSP commands makes a series of temperature measurements. After the code
executes, the data is displayed in the Instrument Console of Test Script Builder.
Send the following commands for this example application:
-- Ensure that the display shows the SCAN swipe and carousel
display.changescreen(display.SCREEN_SCAN_SWIPE)
. You can install and use TSB to write
5-6 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Scanning temperature using thermocouples
Data Acquisition / Multimeter System User's Manual Section 5:
Test results
The following figures show sample results and final test measurements for this application.
1. The horizontal progress bar is completely green when the connections at the screw terminals and
at the DUT are secure. The SCAN swipe screen provides the time remaining (in seconds) and the
scan countdown.
Figure 35: DAQ6510 scan start screen
2. If you have data on the USB flash drive and want to review it prior to completion, pause the scan,
remove the drive, copy the data file to your PC, return the drive to the DAQ6510, and resume the
scan.
Figure 36: Temperature scan results
DAQ6510-900-01Rev. A / April 2018 5-7
Section
User's Manual
5: Scanning temperature using thermocouples DAQ6510 Data Acquisition / Multimeter System
3. Once the scan is complete, the SCAN swipe screen provides you with additional options. You can
select different tabs that provide graphical data analysis options. The following figure shows a
sample graph for this application.
This application example demonstrates how to use the DAQ6510 to accurately measure DC voltage
in a variety of ranges. To ensure accurate data, the NPLC (Number of Power Line Cycles) and
autozero options are used for this test.
Section 6
Scanning low-level DCV
The NPLC setting can be used to help reduce the induced noise originating from nearby AC
power-conditioning circuits. A desktop power supply or power-transmission lines would generate this
type of noise. Increasing NPLC cancels out this noise by integrating all sampled data collected in
multiples of AC signal periods (n * 1/(transmission line frequency) seconds). The more AC line cycles
used in the measurement, the more accurate the reading. The time required to conduct the scan also
increases.
The autozero function removes offset voltages that result from thermal EMFs. Thermal EMFs occur
when there is a temperature difference at junctions consisting of different materials. For example,
leads, instrument inputs, or card terminals. These EMFs adversely affect DCV measurement
accuracy by offsetting the measured voltage.
This example shows how to measure voltage in different ranges. To optimize scanning speed, you
should set a fixed range. If speed is not an issue, the measurement range can be set to auto.
Equipment required
• One DAQ6510
• One Model 7700 20-channel differential multiplexer module
• One computer set up for communication with the instrument
• One device or component to be tested
Section
User's Manual
6: Scanning low-level DCV DAQ6510 Data Acquisition / Multimeter System
Device connections
This example uses a DAQ6510 with a 7700 multiplexer module. In this example, channels 101 to 106
are connected to six DUTs (device under test) where DC voltage is measured.
Figure 38: DAQ6510 instrument and devices connections for low-level DCV
Connecting devices to the instrument
1. Power off the DAQ6510.
2. Remove the 7700 from the DAQ6510.
3. Remove the top cover from the 7700.
6-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
level DCV
Data Acquisition / Multimeter System User's Manual Section 6: Scanning low-
4. Connect six devices.
5. Route the cables out through the cable channels and secure the top cover.
6. Ensure that the DAQ6510 power is turned off.
7. Insert the 7700 into a slot on the rear of the DAQ6510.
Figure 39: DAQ6510 with 7700 multiplexer module
8. Press the POWER switch on the front panel to turn on the instrument.
9. Set the TERMINALS switch to rear.
The remaining connections are to the DUT and are at your discretion.
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads or any device under test (DUT) that is in contact with the
conductors. It is good practice to disconnect DUTs from the instrument before powering the
instrument. Safe installation requires proper shields, barriers, and grounding to prevent
contact with test leads.
There is no internal connection between protective earth (safety ground) and the LO
terminals of the DAQ6510. Therefore, hazardous voltages (more than 30 V
LO terminals. This can occur when the instrument is operating in any mode. To prevent
hazardous voltage from appearing on the LO terminals, connect the LO terminal to protective
earth (safety ground) if your application allows it. You can connect the LO terminal to the
chassis ground terminal on the front panel or the chassis ground screw terminal on the rear
panel. Note that the front-panel terminals are isolated from the rear-panel terminals.
Therefore, if you are using the front-panel terminals, ground to the front-panel LO terminal. If
using the rear-panel terminals, ground to the rear panel LO terminal. Failure to follow these
guidelines can result in injury, death, or instrument damage.
Scanning low-level DCV
This application demonstrates how to use the DAQ6510 to perform DC voltage measurement using
multiple channels on the 7700 20-channel differential multiplexer module.
For this application, you will:
• Configure channels 101 to 106 for DC voltage measurement.
• Enable autoranging and autozero on each channel.
• Set NPLC to five on each channel.
• Execute 10 scans on all selected channels.
You can program the unit through the front panel or use the remote communications interface of your
choosing (LAN, USB, GPIB, RS-232, or TSP-Link) via the SCPI and TSP code commands.
) can appear on
RMS
DAQ6510-900-01Rev. A / April 2018 6-3
Section
User's Manual
6: Scanning low-level DCV DAQ6510 Data Acquisition / Multimeter System
Using the front panel
To set up the application from the front panel:
1. Press the POWER switch on the front panel to turn on the instrument.
2. Select the REAR terminals.
3. Press the MENU key.
4. Under Channel, select Scan.
5. Select the + button to add a group of channels, select channels 101 to 106, and select OK.
6. Select DC Voltage.
7. On the Settings tab, set NPLC to 5.
8. Set Auto Zero to On.
9. On the Scan tab, set Scan Count to 10.
10. Select Start at the bottom of the left pane.
To watch active readings during scan:
1. Press the HOME key.
2. Select the arrow to the right of Watch Channel.
3. Select the channels of interest and select OK to accept.
To save the scanned measurements to a USB flash drive:
1. Select the MENU key.
2. Under the Measure column, choose Reading Buffers.
3. Insert a USB flash drive into the DAQ6510. Select Save to USB at the bottom of the pane.
4. Select OK.
Using SCPI commands
This sequence of SCPI commands executes a DCV scan on channels 101 to 106.
You may need to make changes so that this code will run in your programming environment. In the
table, the SCPI commands have a light gray background. The light green shaded code represents
pseudocode that will vary depending on the programming environments you use.
6-4 DAQ6510-900-01Rev. A / April 2018
DAQ6510
level DCV
int scanCount = 10
count
int channelCount = 6
count
int bufferSize = scanCount *
channelCount
int lastIndex
after every two seconds
string tmpBuff
scanned data every two seconds
DAQ6510
*RST
state
:TRAC:POIN bufferSize,
case)
:ROUT:SCAN:BUFF "defbuffer1"
"defbuffer1"
FUNC 'VOLT:DC', (@101:106)
Set channels functions to DCV
VOLT:DC:RANG:AUTO ON,
(@101:106)
VOLT:DC:NPLC 5, (@101:106)
Set NPLC to 5
VOLT:DC:AZER ON, (@101:106)
Set autozero function to ON
ROUT:SCAN (@101:106)
Setup the scan list
ROUT:SCAN:COUN:SCAN 10
scans
INIT
Initiate the scan
Pseudocode
for i = 1; i <= bufferSize;
delay 2000
readings to accumulate
DAQ6510
lastIndex = TRACe:ACTual?
tmpBuff = "TRACe:DATA? i,
READ
printBuffer (tmpBuffer)
to console
i = lastIndex + 1
end for
:TRAC:SAVE "/usb1/MyData.csv",
to USB
Data Acquisition / Multimeter System User's Manual Section 6: Scanning low-
Send the following commands for this example application:
Pseudocode
Commands Descriptions
Create a variable to hold the scan
Create a variable to hold channel
The number of readings
The last scan index in the buffer
A temporary buffer that store
Put the instrument in a known
"defbuffer1"
Set the buffer size (not necessary
when using the default buffer, but
added to show command use
Assign all scanned data to
Set the channels ranges to auto
Pseudocode
DAQ6510
lastIndex, "defbuffer1",
"defbuffer1"
Setup the number of repeated
Delay two seconds to allow
Query the current buffer index
Query the readings available from
index i to index lastIndex
Print readings in temporary buffer
Save buffer data from defbuffer1
DAQ6510-900-01Rev. A / April 2018 6-5
Section
User's Manual
6: Scanning low-level DCV DAQ6510 Data Acquisition / Multimeter System
Using TSP commands
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from tek.com/keithley
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the Model DAQ6510 Reference Manual.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the DAQ6510 uses the SCPI command set. You must select the TSP command set before
sending TSP commands to the instrument.
To enable TSP commands:
1. Press the MENU key.
2. Under System, select Settings.
3. Set the Command Set to TSP.
4. At the prompt to reboot, select Yes.
This sequence of TSP commands initiates a series of temperature measurements. After the code
executes, the data is displayed in the Instrument Console of Test Script Builder.
Send the following commands for this example application:
-- Reset the instrument to the default settings
reset()
channelCount = 6
scanCount = 10
bufferSize = channelCount * scanCount
defbuffer1.capacity = bufferSize
scan.buffer = defbuffer1 -- not necessary when using the default buffer, but added
This example application demonstrates how to use the DAQ6510 to accurately measure resistance
across multiple devices. To obtain the best results, the 4-wire (Kelvin) measurement method and
offset compensation are used for this test.
Section 7
Scanning resistors using 4W measurement
Typical resistance measurements made using the 2-wire method source current through the test
leads and the device under test (DUT). The voltage is measured, and the resistance is calculated.
It is difficult to obtain accurate 2-wire resistance measurements when the DUT is lower than 100 Ω.
Typical lead resistances lie in the range of 1 mΩ to 10 mΩ. When the 2-wire method is applied to lowresistance measurements, there is a small but significant voltage drop across the resistance of each
test lead. The voltage measured by the instrument is not the same as the voltage directly across the
DUT.
The 4-wire method is preferred for low-resistance measurements. With this configuration, the test
current is sourced through the DUT using one set of test leads, while a second set of SENSE leads
measures the voltage across the DUT. The voltage-sensing leads are connected as close to the
device under test as possible to avoid including the resistance of the test leads in the measurement.
Thermoelectric voltages (EMFs) can seriously affect low-resistance measurement accuracy. The
DAQ6510 can apply the offset-compensated ohms method (OCOMP), which makes one normal
resistance measurement and one using the lowest current source setting to eliminate EMFs.
For this example, you will use resistors of different low values across multiple channels of a 7700
multiplexer module and examine how the 4-wire measurement method provides a more accurate
reading than the 2-wire method. Fixed measurement ranges are applied in order to optimize scanning
speed and OCOMP is applied to correct for any EMF effects.
For comprehensive information on 4-wire resistance measurements, thermoelectric EMFs, and offset
compensation methods, see the Low Level Measurements Handbook, available on tek.com/keithley
.
Section
User's Manual
7: Scanning resistors using 4W measurement DAQ6510 Data Acquisition / Multimeter System
Equipment required
• One DAQ6510
• One Model 7700 20-channel differential multiplexer module
• One computer set up for communication with the instrument
• Six resistors of 100 Ω, 68 Ω, 10 Ω, 2.2 Ω, 0.5 Ω, and 0.2 Ω
• One device or component to be tested
7-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
ing 4W measurement
Data Acquisition / Multimeter System User's Manual Section 7: Scanning resistors us
Device connections
This example application uses the DAQ6510 to perform a 4-wire measurement using multiple
channels on a 7700 multiplexer module.
For this example you will:
•Configure channels 101, 102, 103, 104, 105, and 106 to measure resistors (in this case, 100 Ω,
68 Ω, 10 Ω, 2.2 Ω, 0.5 Ω, and 0.2 Ω).
•Pair channels 111 to 116 with channels 101 to 106, respectively, to provide the SENSE
connections needed for 4-wire measurement.
• Select fixed ranges and apply offset compensation to each channel.
• Execute 100 scans for all selected channels.
Figure 40: Schematic of 4W measurement
DAQ6510-900-01Rev. A / April 2018 7-3
Section
User's Manual
7: Scanning resistors using 4W measurement DAQ6510 Data Acquisition / Multimeter System
To highlight the impact of test lead resistance on low-resistance measurements, and how the
correction methods improve readings, the connections to each DUT consist of one meter of 22 AWG
wire.
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads or any device under test (DUT) that is in contact with the
conductors. It is good practice to disconnect DUTs from the instrument before powering the
instrument. Safe installation requires proper shields, barriers, and grounding to prevent
contact with test leads.
There is no internal connection between protective earth (safety ground) and the LO
terminals of the DAQ6510. Therefore, hazardous voltages (more than 30 V
) can appear on
rms
LO terminals. This can occur when the instrument is operating in any mode. To prevent
hazardous voltage from appearing on the LO terminals, connect the LO terminal to protective
earth (safety ground) if your application allows it. You can connect the LO terminal to the
chassis ground terminal on the front panel or the chassis ground screw terminal on the rear
panel. Note that the front-panel terminals are isolated from the rear-panel terminals.
Therefore, if you are using the front-panel terminals, ground to the front-panel LO terminal. If
using the rear-panel terminals, ground to the rear panel LO terminal. Failure to follow these
guidelines can result in injury, death, or instrument damage.
7-4 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Scanning resistors using 4W measurement
Data Acquisition / Multimeter System User's Manual Section 7:
Scanning resistors using 4-wire measurements
Using the front panel
To setup the application from the front panel:
1. Press the POWER switch on the front panel to turn on the instrument.
2. Set the TERMINALS switch to REAR.
3. Press the MENU key.
4. Under Channel, select Scan.
5. Select the + button to add groups of channels (101 to 106) and select OK.
6. Select 4W Resistance on the Measure Functions screen.
7. Select Offset Compensation, select On.
8. In the upper left corner of the screen, select the Menu icon to reveal options, select Expand Group. The left side of the pane allows you to modify the range on each channel.
9. Select channel 101 by selecting 4-Wire Res.
10. Under the Settings tab, select Range and set to 100 W.
11. Select chanel 102 by selecting 4-Wire Res and set the Range to 100 W.
12. Select chanel 103 by selecting 4-Wire Res and set the Range to 10 W.
13. Select chanel 104 by selecting 4-Wire Res and set the Range to 10 W.
14. Select chanel 105 by selecting 4-Wire Res and set the Range to 1 W.
15. Select chanel 106 by selecting 4-Wire Res and set the Range to 1 W.
16. Locate the Scan tab, and set the Scan Count to 100.
17. Select the Start button at the bottom of the left-side pane to start the scan.
You can monitor active readings during the scan or review all scanned data after the scan finishes.
To watch active readings during scan:
1. Press the HOME key.
2. Select the button to the right of Watch Channel.
3. Select the channels of interest and select OK to accept.
To review all the data after the scan finishes:
1. Press the MENU key.
2. Under the Views column, choose Reading Table.
DAQ6510-900-01Rev. A / April 2018 7-5
Section
User's Manual
*RST
Reset the DAQ6510
ROUT:SCAN:COUN:SCAN scanCnt
Set the number of times the scan is repeated
FUNC 'FRES',(@101:106)
Set function to 4-wire measurement
FRES:OCOM ON, (@101:106)
Set offset compensation to "on"
FRES:RANG 100, (@101,102)
Set the fixed range at 100 Ω
FRES:RANG 10, (@103,104)
Set the fixed range at 10 Ω
FRES:RANG 1, (@105,106)
Set the fixed range at 1 Ω
ROUT:SCAN:CRE (@101:106)
Set the scan list
TRAC:CLE
Clear the reading buffer
INIT
Initiate the scan
*WAI
Wait for scan completion
7: Scanning resistors using 4W measurement DAQ6510 Data Acquisition / Multimeter System
Using SCPI commands
This sequence of SCPI commands executes a 4-wire resistance scan.
You may need to make changes so that this code will run in your programming environment.
Send the following commands for this example application:
Commands Descriptions
7-6 DAQ6510-900-01Rev. A / April 2018
DAQ6510
Scanning resistors using 4W measurement
Data Acquisition / Multimeter System User's Manual Section 7:
Using TSP commands
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from tek.com/keithley
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the Model DAQ6510 Reference Manual.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the DAQ6510 uses the SCPI command set. You must select the TSP command set before
sending TSP commands to the instrument.
To enable TSP commands:
1. Press the MENU key.
2. Under System, select Settings.
3. Set the Command Set to TSP.
4. At the prompt to reboot, select Yes.
This sequence of TSP commands initiates a series of temperature measurements. After the code
executes, the data is displayed in the Instrument Console of Test Script Builder.
. You can install and use TSB to write
After the scan, the data in the buffer is stored on a USB flash drive.
Send the following commands for this example application:
-- Reset the instrument to the default settings
reset()
scanCount = 100
-- set up each channels function, range and offset compensation
7: Scanning resistors using 4W measurement DAQ6510 Data Acquisition / Multimeter System
Test results
The table below compares the values of scanned DUTs using 4-wire and 2-wire measurement
methods with the six resistors listed in the Instrument and Device Connection section.
The data provided in each cell of the table is the average of 100 scans.
Nominal
value (Ω)
2.2 2.20413 2.20588 2.51011
Four-wire measurement
with offset compensation
(Ω)
Four-wire measurement
with NO offset
compensation (Ω)
Two-wire measurement
(Ω)
7-8 DAQ6510-900-01Rev. A / April 2018
Mixed function multi-channel scanning .................................... 8-3
This example application demonstrates how to use the DAQ6510 to perform complex multi-channel,
mixed function scanning in a production-test environment.
The DAQ6510 can perform more than one function in a multichannel scan, providing a range of dataacquisition options in a single test.
Section 8
Mixed function multi-channel scanning
In this production environment the DAQ6510 is:
• Integrated into a test stand.
• Wired to a fixture that is connected to an active device under test (DUT).
• Quickly capturing DC volts and current, temperature, and AC volts and current.
Prior to the start of the scan, you can step through each of the configured channels on the DAQ6510,
which allows you to troubleshoot the test configuration. This allows you to view the readings of
individually closed channels to ensure that connections to the DUT are secure.
Equipment required
• One DAQ6510
• One Model 7700 20-channel differential multiplexer module
• One computer set up for communication with the instrument
• One device or component to be tested
Section
User's Manual
8: Mixed function multi-channel scanning DAQ6510 Data Acquisition / Multimeter System
Device connections
This application example uses a DAQ6510 with a 7700, 20-channel differential multiplexer configured
to monitor the following signals:
• Channel 101: AC voltage being supplied to the DUT.
• Channels 102 to 110: DC voltages at several points located on the DUT.
• Channels 111 to 112: Temperature (using Type-K thermocouples) of two voltage regulators within
the DUT.
•Channels 113 to 114: Temperature (using Type-K thermocouples) of two loads to which the DUT
is supplying power.
• Channel 121: AC current being drawn by the DUT.
• Channel 122: DC current being drawn by the load.
Figure 41: DAQ6510 Model 7700
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads or any device under test (DUT) that is in contact with the
conductors. It is good practice to disconnect DUTs from the instrument before powering the
instrument. Safe installation requires proper shields, barriers, and grounding to prevent
contact with test leads.
8-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
channel scanning
Data Acquisition / Multimeter System User's Manual Section 8: Mixed function multi-
There is no internal connection between protective earth (safety ground) and the LO
terminals of the DAQ6510. Therefore, hazardous voltages (more than 30 V
LO terminals. This can occur when the instrument is operating in any mode. To prevent
hazardous voltage from appearing on the LO terminals, connect the LO terminal to protective
earth (safety ground) if your application allows it. You can connect the LO terminal to the
chassis ground terminal on the front panel or the chassis ground screw terminal on the rear
panel. Note that the front-panel terminals are isolated from the rear-panel terminals.
Therefore, if you are using the front-panel terminals, ground to the front-panel LO terminal. If
using the rear-panel terminals, ground to the rear panel LO terminal. Failure to follow these
guidelines can result in injury, death, or instrument damage.
Mixed function multi-channel scanning
This application example uses the DAQ6510 to execute complex multi-channel, mixed function
scanning in a production-test environment.
For this application, you will:
• Configure the instrument for:
One channel each of ACV, ACI, and DCI.
Nine channels of DCV.
Four channels of temperature using Type K thermocouples using a simulated reference junction.
•Set labels on certain channels to help better identify what the measurement is monitoring at the
DUT.
• Step through the scan to check each individual channel reading prior to scanning.
• Execute 10 scans for each selected channels.
• Use the reading table to view (or print) the scanned measurements.
) can appear on
RMS
You can operate the instrument through the front panel or using SCPI or TSP code. For information
about setting up remote communications, see Remote communications interfaces (on page 3-1
).
DAQ6510-900-01Rev. A / April 2018 8-3
Section
User's Manual
8: Mixed function multi-channel scanning DAQ6510 Data Acquisition / Multimeter System
Using the front panel
To set up the application from the front panel:
1. Press the POWER switch on the front panel to turn on the instrument.
2. Set the TERMINALS switch to REAR.
3. Select Build Scan from the Scan swipe screen.
4. Select the + button to add channel 101, select OK.
5. Select AC Voltage on the Measure Functions screen.
6. Select the Detector Bandwidth option, and set the bandwidth to 30 Hz.
7. Select channel 101, scroll to Label, enter ACMains, and select OK to continue.
8. Select the + button, add channels 102 to 110, select OK.
9. Select DC Voltage.
10. Select the + button, add channels 111 to 114, select OK.
11. Select Temperature.
12. Scroll to the Temperature setting, enter a Simulated junction temperature of 23 °C, select OK.
13. Select Simulated for the Reference Junction.
14. In the top left corner of the screen, select Menu, select Expand Groups.
15. Select channel 111 and enter Reg12VTemp for the label. Select OK.
16. Select channel 112 and enter Reg5VTemp for the label. Select OK.
17. Select channel 113 and enter LoadTemp1 for the label. Select OK.
18. Select channel 114 and enter LoadTemp2 for the label. Select OK.
19. Select Menu and Collapse Groups. The AC Volt, DC Volt, and Temp groups are separated into
individual groups.
20. Select the highlighted + button to add another group of channels and add channel 121. Select
OK.
21. Select AC Current.
22. Select the + button and add channel 122, select OK.
23. Select DC Current.
24. On the Scan tab, set Scan Count to 10, select OK.
25. Press the HOME key.
26. Select Step Scan. This closes the first channel in the scan and displays the results. Use the
left/right arrows adjacent to Watch Channel to scroll through the selected channels, selecting
Step Scan on each, until all the channels have been validated.
27. Select Abort Scan.
28. Press the TRIGGER key to the right of the display to initiate the scan.
8-4 DAQ6510-900-01Rev. A / April 2018
DAQ6510
channel scanning
*RST
Put the instrument in a known state
SENS:FUNC 'VOLT:AC',(@101)
Set channel 101 for ACV
SENS:VOLT:AC:DET:BAND 30, (@101)
30Hz
SENS:FUNC 'VOLT:DC', (@102:110)
Set channels 102-110 for DCV
SENS:FUNCtion 'TEMPerature',
(@111:114)
SENS:TEMP:TRAN TC, (@111:114)
SENS:TEMP:TC:TYPE K, (@111:114)
SENS:TEMP:TC:RJUN:RSEL SIM,
(@111:114)
SENS:TEMP:TC:RJUN:SIM 23,
(@111:114)
SENS:FUNC 'CURR:AC', (@121)
Set channel 121 to measure ACI
SENS:FUNC 'CURR:DC', (@122)
ROUT:CHAN:LAB "ACSource", (@101)
ROUT:CHAN:LAB "Reg12VTemp", (@111)
ROUT:CHAN:LAB "Reg5VTemp", (@112)
Apply a label to channel 112
ROUT:CHAN:LAB "LoadTemp1", (@113)
Apply a label to channel 113
ROUT:CHAN:LAB "LoadTemp2", (@114)
ROUTe:SCAN:COUNt:SCAN 10
Set the number scans
ROUTe:SCAN:CREate
(@101:114,121,122)
INIT
*WAI
Wait for scan completion
TRAC:DATA? 1, 160, "defbuffer1",
READ, CHAN
Data Acquisition / Multimeter System User's Manual Section 8: Mixed function multi-
Using SCPI commands
This sequence of SCPI commands configures the instrument for measuring DC Voltage,
Temperature, AC Voltage, DC Current, and AC Current on different channels and then labels some of
those channels. Each channel is then scanned 10 times and the readings are returned.
You may need to make changes so that this code will run in your programming environment. In the
table, the SCPI commands have a light gray background.
Send the following commands for this example application:
Commands Descriptions
Set channel 101, low-end bandwidth to
Set channels 111-114 for Temperature
measurement using Type K
thermocouples with a simulated reference
junction set to 23°C
Set channel 122 to measure DCI
Apply a label to channel 101
Apply a label to channel 111
Apply a label to channel 114
Set the channels included in the scan
Initiate the scan
Query the readings from the instrument
DAQ6510-900-01Rev. A / April 2018 8-5
Section
User's Manual
8: Mixed function multi-channel scanning DAQ6510 Data Acquisition / Multimeter System
Using TSP commands
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from tek.com/keithley
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the Model DAQ6510 Reference Manual.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the DAQ6510 uses the SCPI command set. You must select the TSP command set before
sending TSP commands to the instrument.
To enable TSP commands:
1. Press the MENU key.
2. Under System, select Settings.
3. Set the Command Set to TSP.
4. At the prompt to reboot, select Yes.
This sequence of TSP commands makes a series of temperature measurements. After the code
executes, the data is displayed in the Instrument Console of Test Script Builder.
Send the following commands for this example application:
-- Reset the instrument to the default settings
reset()
-- Establish channel settings for the scan card configuration...
There are three different multiplex modules available for use with the DAQ6510. This application
example demonstrates how each of the multiplexer modules can impact productivity by changing test
time. The multiplexer modules all share the same basic code base for switching, scanning, and
measuring. Any limits on system speed are the result of the relays in the multiplexer that switch the
signals from the device under test (DUT) into the instrument.
Section 9
Speed scanning for increased test throughput
The Model 7700 20-Channel Differential Multiplexer Module uses electromechanical relays which
have low contact resistance and contribute only a minor offset potential (<1 Ω through the end of life
and < 500 nV, respectively). This results in the most accurate readings of the modules but with a 3
ms relay closure time, the slowest scan time in comparison to other options.
The 7703 multiplexer module uses reed relays which have low contact resistance (<1 Ω through the
end of life), but a higher contact potential (6 µV max) which contributes more signal offset and slightly
less precise readings. The benefit of this module is shorter relay close time (less than 1 ms) which
makes it approximately three times faster than the 7700.
The 7710 multiplexer module uses solid-state relays which have the highest contact resistance and
contact potential of the three options (<5 Ω and <1 µV, respectively) and are therefore the least
precise, however the 7710 has the overall speed advantage with a relay close time of less than 0.5
ms, making it twice as fast as the 7703 and at least six times faster than the 7700.
For comprehensive information on relay types, topologies, and the benefits or drawbacks of each, go
to the Keithley webpage on the tek.com/keithley
Equipment required
• One DAQ6510
• One Model 7700 20-channel differential multiplexer module
• One Model 7710 20-channel differential multiplexer module
• One computer set up for remote communication with the DAQ6510
• One device or component to be tested
and search for the Switching Handbook.
S
ection 9: Speed scanning for increased test throughput DAQ6510 Data Acquisition / Multimeter System
User's Manual
Device connections
This example assumes a setup optimized for the greatest speed where the DAQ6510 uses either the
7700 or 7710 20-channel differential multiplexer module multiple to monitor the following signals; the
example code is the same for each.
The controlling computer can use either LAN, USB, or GPIB to reach scan and data transfer speeds
comparable to those of the original factory setup. The GPIB interface requires an optional
communications accessory.
See the following image for the connection example.
Figure 43: DAQ6510 speed scanning instrument and device connection
The 7700 replaces the 7710 for the second run of the program code for comparison and assumes
the same input signals to channels 101 to 120.
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads or any device under test (DUT) that is in contact with the
conductors. It is good practice to disconnect DUTs from the instrument before powering the
instrument. Safe installation requires proper shields, barriers, and grounding to prevent
contact with test leads.
9-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
or increased test throughput
Data Acquisition / Multimeter System User's Manual Section 9: Speed scanning f
There is no internal connection between protective earth (safety ground) and the LO
terminals of the DAQ6510. Therefore, hazardous voltages (more than 30 V
LO terminals. This can occur when the instrument is operating in any mode. To prevent
hazardous voltage from appearing on the LO terminals, connect the LO terminal to protective
earth (safety ground) if your application allows it. You can connect the LO terminal to the
chassis ground terminal on the front panel or the chassis ground screw terminal on the rear
panel. Note that the front-panel terminals are isolated from the rear-panel terminals.
Therefore, if you are using the front-panel terminals, ground to the front-panel LO terminal. If
using the rear-panel terminals, ground to the rear panel LO terminal. Failure to follow these
guidelines can result in injury, death, or instrument damage.
Speed scanning for increased production test throughput
This application demonstrates how to configure the DAQ6510 to execute scans at the best scan
speed. You will eliminate certain measurement options that add to overall test time. The scan
completion times of the 7700 and 7710 20-channel differential multiplexer modules will be compared,
and you will see the speed improvement attributed to solid-state relays over electromechanical relays
used in some multiplexer modules.
For this application, you will:
• Use the sample code (either SCPI or TSP) to issue commands that:
Fix the DCV measurement range for all channels and remove the delay time introduced through
autoranging.
Eliminate the autozero feature to remove extra correction measurements from being performed.
Set the display digits to a low resolution and turn off the front panel to bypass any delays that would
come from updates/refreshes.
Turn off channel statistics calculation so that the instrument puts its processing power into data
acquisition and transmission.
Disable line synchronization.
Execute a 20-channel scan with a count of 1000 for a total sample count of 20,000 readings.
Incrementally extract the most current scan measurements until complete and either save them to a file
or print them to the computer display.
•Evaluate the time elapsed.
) can appear on
RMS
DAQ6510-900-01Rev. A / April 2018 9-3
S
ection 9: Speed scanning for increased test throughput DAQ6510 Data Acquisition / Multimeter System
User's Manual
int scanCnt = 1000
count
int sampleCnt
readings)
int chanCnt
channel count
int actualRdgs
reading count
string rcvBuffer
extracted readings
timer1.start()
elapsed time
DAQ6510
*RST
FORM:DATA ASCII
ROUT:SCAN:COUN:SCAN scanCnt
Apply the scan count
FUNC 'VOLT:DC',(@101:120)
Set function to DCV
VOLT:RANG 1, (@101:120)
Set the fixed range at 1 V
VOLT:AVER:STAT OFF,
(@101:120)
DISP:VOLT:DIG 4, (@101:120)
significant digits
VOLT:NPLC 0.0005, (@101:120)
Set fastest NPLC possible
VOLT:LINE:SYNC OFF,
(@101:120)
VOLT:AZER:STAT OFF,
(@101:120)
CALC2:VOLT:LIM1:STAT OFF,
(@101:120)
CALC2:VOLT:LIM2:STAT OFF,
(@101:120)
ROUT:SCAN:INT 0
to 0s
TRAC:CLE
Clear the reading buffer
DISP:LIGH:STAT OFF
Turn the display off
Using SCPI commands
This sequence of SCPI commands executes a 20-channel scan 1000 times and saves the data to the
controlling computer.
You may need to make changes so that this code will run in your programming environment. In the
table, the SCPI commands have a light gray background. The light green shaded code represents
pseudocode that will vary depending on the programming environments you use.
Send the following SCPI commands for this example application:
Commands Descriptions
Pseudocode
Create a variable to hold the scan
Create a variable to hold the full
sample count (total number of
Create a variable to hold the
Create a variable to hold the actual
Create a string buffer to hold
Start a timer to help capture
Put the instrument in a known state
Format data as an ASCII string
Disable background statistics
9-4 DAQ6510-900-01Rev. A / April 2018
Front panel shows only four
Turn off line sync
Turn off auto zero
Turn off limit tests
Set trigger interval between scans
DAQ6510
Speed scanning for increased test throughput
ROUT:SCAN:CRE (@101:120)
Set the scan list
chanCnt =
ROUTe:SCAN:COUNt:STEP?
sampleCnt = scanCnt * chanCnt
taken
INIT
Initiate the scan
Pseudocode
for i = 1, i < sampleCnt
incrementing of i for later
delay 500
to accumulate
DAQ6510
actualRdgs = TRACe:ACTual?
Query the actual readings captured
rcvBuffer = "TRACe:DATA? i,
READ
Pseudocode
WriteReadings("C:\myData.csv"
computer
i = actualRdgs + 1
Increment i for the next loop pass
end for
End the for loop
timer1.stop()
Stop the timer
timer1.stop – timer1.start
Calculate the elapsed time
DAQ6510
DISP:LIGH:STAT ON100
Turn the display back on
Data Acquisition / Multimeter System User's Manual Section 9:
Query the channel count
Pseudocode
DAQ6510
actualRdgs, "defbuffer1",
, rcvBuffer)
Calculate the number of readings
Setup a loop from one up to
sampleCnt, but leave the
Delay for 500 ms to allow readings
Query the readings available from i
to the value of actualRdgs
Write the extracted readings to a
file, myData.csv, on the local
DAQ6510-900-01Rev. A / April 2018 9-5
S
ection 9: Speed scanning for increased test throughput DAQ6510 Data Acquisition / Multimeter System
User's Manual
Using TSP commands
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from tek.com/keithley
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the Model DAQ6510 Reference Manual.
To use other programming environments, you may need to make changes to the example TSP code.
By default, the DAQ6510 uses the SCPI command set. You must select the TSP command set before
sending TSP commands to the instrument.
To enable TSP commands:
1. Press the MENU key.
2. Under System, select Settings.
3. Set the Command Set to TSP.
4. At the prompt to reboot, select Yes.
This sequence of TSP commands makes a series of voltage measurements. After the code executes,
the data is displayed in the Instrument Console of Test Script Builder.
Send the following commands for this example application:
-- Set up variables to be referenced during the scan
Factory run times are as follows. The test time for an equivalent setup using the Model 7703 has also
been provided so that you can see what is achievable with the reed relays.
Differential Multiplexer Module Setup Outcome
7700: 20-channels for 1000 scans at 20,000
Approximate test duration: 3 m 38.93 s at 91 rdgs/s
The 7710 multiplexer module provides readings to the DAQ6510 faster and has the speed
advantage.
Many products need to be environmentally-stressed during performance testing. This is accomplished
by placing the device in a chamber where the temperature can be set and controlled and soaking the
DUT at the required set-point conditions. Temperature changes are not instantaneous so there is
some wait time between scans. The DAQ6510 can monitor the environment until the target
temperature has been reached at which point the instrument automatically begins the scan.
Section 10
Pre-scan monitor
This application example demonstrates how to use a DAQ6510 to initiate a scan based on the
temperature of the environment around the device under test (DUT). This example will model a
situation where resistance DUTs are measured after the temperature exceeds 30 °C.
Equipment required
• One DAQ6510
• One Model 7700 20-channel differential multiplexer module
• One computer set up for communication with the instrument
• One Type K thermocouple
• Four resistors
Sect
ion 10: Pre-scan monitor DAQ6510 Data Acquisition / Multimeter System
User's Manual
Device connections
This application example uses a DAQ6510 with a Model 7700 20-channel differential multiplexer
configured to monitor a Type K thermocouple, connected to channel 101, and four DUTs (resistors)
connected to channels 102 through 105.
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads or any device under test (DUT) that is in contact with the
conductors. It is good practice to disconnect DUTs from the instrument before powering the
instrument. Safe installation requires proper shields, barriers, and grounding to prevent
contact with test leads.
There is no internal connection between protective earth (safety ground) and the LO
terminals of the DAQ6510. Therefore, hazardous voltages (more than 30 V
) can appear on
RMS
LO terminals. This can occur when the instrument is operating in any mode. To prevent
hazardous voltage from appearing on the LO terminals, connect the LO terminal to protective
earth (safety ground) if your application allows it. You can connect the LO terminal to the
chassis ground terminal on the front panel or the chassis ground screw terminal on the rear
panel. Note that the front-panel terminals are isolated from the rear-panel terminals.
Therefore, if you are using the front-panel terminals, ground to the front-panel LO terminal. If
using the rear-panel terminals, ground to the rear panel LO terminal. Failure to follow these
guidelines can result in injury, death, or instrument damage.
10-2 DAQ6510-900-01Rev. A / April 2018
DAQ6510
scan monitor
Data Acquisition / Multimeter System User's Manual Section 10: Pre-
Pre-scan monitor
The DAQ6510 uses Monitor Measurement triggering on channel 101.
For this application, you will:
•Configure the instrument to measure the temperature on channel 101 using a thermocouple. The
instrument then scans channels 102 to 105 after the specified temperature has been reached.
•Configure the scan on channels 102 to 105 using the 2-wire resistance function.
Using the front panel
To setup the application from the front panel:
1. Press the POWER switch on the front panel to turn on the instrument.
2. Select the REAR terminals.
3. Press the MENU key.
4. Under Channel, select Scan.
5. Select the + button to add a group of channels (101 to 105), and select OK.
6. Select 2W Resistance on the Measure Functions screen.
7. In the upper left corner of the screen, select Expand Group from the Menu icon.
8. Select channel 101 by selecting the adjacent 2-Wire Res button. Change the function to
Temperature.
9. Set Reference Junction to Internal.
10. In the Scan tab, change the Scan Count to 10.
11. In the Trigger tab, select Scan Start and select Monitor Measurement.
12. A second panel opens, select Above High Limit for Start Condition.
13. Set Channel to 101.
14. Set Higher Limit to 30 ˚C and select OK.
15. Select the Start button at the bottom of the left-side pane to start the scan.
16. Select View Scan Status to return to the HOME screen.
You will be able to watch the active temperature measurement on channel 101. Once the target
temperature has been reached the instrument begins scanning.
DAQ6510-900-01Rev. A / April 2018 10-3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.