Any unauthorized reproduction, photocopy, or use the information herein, in whole or in part,
without the prior written approval of Keithley Instruments, Inc. is strictly prohibited.
All Keithley Instruments product names are trademarks or registered trademarks of Keithley
Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective
holders.
Document number: 2651A-900-01 Rev. A / March 2011
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using
the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained
service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I and
Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most
measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or
to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient
over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are
for connection to Category I sources unless otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any
unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If
the circuit is capable of operating at or above 1000V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective
devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth)
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the
voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions, or the safety of
11/07
the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating
information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
If a
The ! symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions located in the
user documentation in all cases where the symbol is marked on the instrument.
The symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal
contact with these voltages.
The
The
If this
symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
symbol indicates a connection terminal to the equipment frame.
Hg
symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be
properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national
safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be
purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be
purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the
applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled
according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the
factory for proper cleaning/servicing.
Organization of manual sections .............................................. 1-2
Model 2651A applications ........................................................ 1-2
Thank you for choosing a Keithley Instruments product. The Model 2651A High Power System
SourceMeter
®
Instrument provides manufacturers of electronic components and semiconductor
devices a high-power source-measure unit (SMU). A SMU is an instrument that combines source and
measurement capabilities in a single instrument. This combination simplifies test processes by
eliminating synchronization and connection issues associated with multiple instrument solutions. A
Model 2651A provides a scalable, high throughput, highly cost-effective solution for precision DC,
pulse, and low frequency AC source-measure testing that also maintains code compatibility with the
Series 2600A.
Extended warranty
Additional years of warranty coverage are available on many products. These valuable contracts
protect you from unbudgeted service expenses and provide additional years of protection at a fraction
of the price of a repair. Extended warranties are available on new and existing products. Contact your
local Keithley Instruments representative for details.
Introduction to this manual
This manual provides detailed tutorials to help you achieve success with your Keithley Instruments
Model 2651A. In addition, the basics of the two simplest interfaces, the front panel and the web
interface, are provided to familiarize you with the instrument. Select and run all examples in Sections
5, 6, and 7 that are relevant to your intended use and to the equipment you are using.
Some of the examples in this manual may use unfamiliar commands and concepts. For detailed
information about these, refer to the Reference Manual (part number 2651A-901-01) on the Product
Information CD-ROM that came with your instrument.
Section 1: Introduction Model 2651A High Power System SourceMeter® Instrument User's Manual
CD-ROM contents
Two CD-ROMs are shipped with each Model 2651A order. The Model 2651A Quick Start Guide,
User's Manual, and Reference Manual are provided in PDF format on the Model 2651A Product
Information CD-ROM.
basic operation information. If you are new to Keithley Instruments equipment, refer to the Quick
Start Guide to take the steps needed to unpack, set up, and verify operation.
•User's Manual: Provides application examples. If you need a starting point to begin creation of
applications, refer to the User's Manual for a variety of specific examples.
•Reference Manual: Includes advanced operation topics and maintenance information.
Programmers looking for a command reference, and users looking for an in-depth description of
the way the instrument works (including troubleshooting and optimization), should refer to the
Reference Manual.
A second CD-ROM contains the Test Script Builder script development software (Keithley
Instruments part number KTS-850). Use this CD-ROM to install the Test Script Builder Integrated
Development Environment. This software not only provides an environment to develop a test
program, but also the ability to load the test program onto the instrument. Running a program loaded
on the instrument eliminates the need to send individual commands from the host computer to the
instrument when running a test.
Organization of manual sections
Bookmarks for each manual section have been provided in the PDF (see Adobe® Acrobat® or
Reader
Contents located at the beginning of this manual.
®
help for information about bookmarks). The manual sections are also listed in the Table of
Model 2651A applications
In addition to being a stand-alone instrument, the Keithley Instruments Model 2651A High Power
System SourceMeter
This manual provides application examples that guide you through several common
instrument-to-instrument scenarios. These applications are presented after the summary information
about the High Power System SourceMeter
•Making basic front-panel measurements (on page 4-1): Demonstrates the basic measurement
function using a single Model 2651A and a two-terminal device under test (DUT).
•Capturing high power pulse waveforms (on page 5-1)Capturing high power pulse waveforms:
Demonstrates how to use the Keithley Instruments Model 2651A connected to a control
computer. This example introduces configuration of the instrument's trigger model to make
high-speed measurements of an LED.
®
Instrument can intelligently connect to other instruments and multiple devices.
®
Instrument, and include:
1-2 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 1: Introduction
•Characterization of power discretes (on page 6-1): Demonstrates how to use the Keithley
Instruments Model 2651A with a second Series 2600A System SourceMeter
®
instrument to
characterize power MOSFETs and IGBTs. This example introduces the interaction between two
instruments' trigger models. The application performs pulsed measurements, collecting IV curve
data and placing it in a Microsoft
®
Excel® compatible file format for graphic and analysis.
•Increasing SMU current sourcing ability (on page 7-1): Demonstrates how to combine two
Keithley Instruments Model 2651A instruments with an additional Series 2600A System
SourceMeter
current up to 100 A.
®
instrument, and characterizes the resistance of a power MOSFET by sweeping
Front panel ............................................................................... 2-1
Introduction
Before starting this section, complete the tasks outlined in the Model 2651A Quick Start Guide. Once
you have completed those tasks, read this section, which provides enough basic information about
the Model 2651A front-panel interface to work through the examples provided in this manual.
Section 2
Using the front-panel interface
The front panel of the Keithley Instruments Model 2651A contains the following items:
(1) The POWER key
(2) The display
(3) The navigation wheel
(4) The OUTPUT ON/OFF control
(5) The setup and control keys
You can use the keys, display, and the navigation wheel
and set values.
For additional information about the front panel, see the “General operation” section in the Model
2651A Reference Manual.
Front panel
The front panel of the Model 2651A is shown below.
to access, view, and edit the menu items
Section 2: Using the front-panel interface Model 2651A High Power System SourceMeter® Instrument User's Manual
CONFIG
Figure 1: Front panel
2
+3.21000V
SrcA:+20.0000 A LimAL10.0000V
78 90+/-
SRC MEAS LIMIT MODE
DISPLAY
4
POWER
CONFIG
56
DIGITS
SPEED
12 3
LOAD
RUN
REL
FILTER
0000
STORE
RECALL
1
ARM
AUTO
LOCAL
TRIG
EXIT
MENU
ENTER
5
3
D
E
O
I
T
T
/
E
H
2651A HIGH POWER SYSTEM SourceMeter
N
S
T
U
E
P
R
R
P
E
U
T
S
N
H
E
/
T
T
I
O
E
D
P
T
U
U
T
CURSOR
O
O
F
N
F
/
O
®
4
(1) The POWER key. Press this key to turn the instrument on (|). Press it again to turn the instrument
off (0).
(2) The display. During operation, the display provides readings and information about the selected
measurement and configuration. It also shows the control status (local or remote). If REM is
displayed, the instrument is being controlled remotely (through GPIB, LAN, or USB). If REM is not
displayed, control is through the front panel.
During setup, the display shows menu choices that you can use to configure the instrument.
The items listed below represent the possible display indicators and what they mean.
Indicator Meaning
EDIT: Unit is in the source editing mode
ERR: Questionable reading or invalid calibration step
REM: Unit is in remote mode
TALK: Unit is addressed to talk
LSTN: Unit is addressed to listen
SRQ: Service request is asserted
REL: Relative mode is enabled
FILT: Digital filter is enabled
AUTO: Source or measure autorange is selected
* (asterisk): Readings are being stored in the buffer
(3) The navigation wheel . Turn the navigation wheel to scroll to a menu option or to change the
selected value.
Push the navigation wheel
cases, pressing the navigation wheel
to open menus or to select a menu option or a value. In most
performs the same action as pressing the ENTER key.
(4) The OUTPUT ON/OFF control. Press this control to turn the Model 2651A source output on or off.
The output indicator will light when the source is on.
(5) The setup and control keys provide front panel control and configuration. The following figure
2-2 2651A-900-01 Rev. A / March 2011
illustrates each key's location. The table following the figure contains a definition of each key.
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 2: Using the front-panel interface
Figure 2: Setup and control keys
78 90+/-
DISPLAY
CONFIG
SRC MEASLIMIT MODE
4
56
DIGITS
SPEED
12 3
LOAD
RUN
REL
STORE
FILTER
0000
RECALL
TRIG
MENU
LOCAL
EXIT
ENTER
AUTO
CURSOR
Key descriptions
Key Description
DISPLAY Toggles between the various source-measure displays and the user message mode.
CONFIG Configures a function or operation.
SRC Selects the source function (V or A) and places the cursor in the source field for
MEAS
LIMIT Places the cursor in the compliance limit field for editing. Also selects the limit value
MODE Directly controls the mode.
DIGITS Cycles through display resolution (4-1/2, 5-1/2, or 6-1/2 digits).
SPEED Selects either the fast or integrating A/D converter. When the integrating A/D
REL Controls relative measurements. This allows a baseline value to be subtracted from
FILTER Enables or disables the digital filter. You can use this filter to reduce reading noise.
LOAD Loads test for execution.
RUN Runs the last selected factory or user-defined test code.
STORE Accesses reading buffers and takes readings.
RECALL Recalls data (or statistics) from CHANA-BUFF1 or CHANA-BUFF2.
TRIG Triggers readings.
MENU Accesses the main menu. The main menu can be used to configure many facets of
EXIT
(LOCAL)
ENTER Accepts the current selection or brings up the next menu option. In most cases,
Number
keys
editing.
Cycles through measure functions (V, A, Ω, or W).
to edit (V, A, or W).
converter is selected, this key also sets the measurement speed and accuracy by
controlling the measurement aperture.
a reading.
operation.
Cancels selection and backs out of the menu structure. Also used as a LOCAL key
to take the unit out of remote operation.
pressing the ENTER key is the same as pressing the navigation wheel
.
When enabled, the number keys (0-9, +/-, 0000) allow direct numeric entry in the
EDIT mode. Press the navigation wheel
to enter EDIT mode (see To change a
value using the numeric keypad (on page 2-4)).
2651A-900-01 Rev. A / March 2011 2-3
Section 2: Using the front-panel interface Model 2651A High Power System SourceMeter® Instrument User's Manual
To change a value using the navigation wheel
1. Turn the navigation wheel
to go to the character you want to change (the character blinks
:
when selected).
2. Press the navigation wheel
3. Turn the navigation wheel
to edit that character.
to change the value.
4. Press the navigation wheel to enter the change.
5. Repeat these steps as needed to change the value.
6. Press the ENTER key or the navigation wheel
when finished changing all the characters.
To change a value using the numeric keypad:
NOTE
The numeric entry method may only be used if the numeric keypad is enabled.
1. If the keypad is disabled, press the MENU key, then select DISPLAY > NUMPAD > ENABLE.
2. Use the CURSOR arrow keys to move the cursor to the value that you want to edit.
3. Press any of the desired number keys (0-9, +/-, 0000). The cursor moves to the next value on
the right.
4. Repeat the above steps as required to set the desired values.
5. Press the ENTER key to select the value or press the EXIT (LOCAL) key to cancel the change.
6. (Optional) Press the EXIT (LOCAL) key to return to the main menu.
The Model 2651A web interface allows you to review instrument status, control the instrument, and
upgrade the instrument over a LAN connection.
The instrument web page resides in the firmware of the instrument. Changes you make through the
web interface are immediately made in the instrument.
Section 3
Using the web interface
Many examples in this manual and in the Model 2651A Reference Manual can be run through the
TSB Embedded page of the instrument web interface.
Connect to the instrument web interface
NOTE
The instrument web interface requires the web browser plug-in JavaTM SE Runtime Environment
Version
ROM that came with your instrument.
The instrument web interface uses Java applets and, depending on your browser security settings,
may require your permission to download and install them.
To connect to the instrument web interface, you must have a LAN connection from the computer to
the instrument. See "LAN concepts and settings" in the Model 2651A Reference Manual for more
information about configuring the Model 2651A for a LAN connection, connecting the Model 2651A to
the LAN, and establishing a LAN connection to the instrument.
6 or later. Installation files are also available on the Model 2651A Product Information CD-
Section 3: Using the web interface Model 2651A High Power System SourceMeter® Instrument User's Manual
Once the Model 2651A is configured correctly and connected to the LAN, you can use the LXI®
Discover
automatic IP address selection. Once identified, you can double-click the IP address in the LXI
Discovery Browser to open the web interface for the instrument.
To locate the LXI Discovery Browser on the Keithley website:
1. Select the Support tab.
2. In the model number box, type 2651A.
3. From the list, select Software and click the search icon. A list of software applications for the
4. See the readme file included with the application for more information.
y Browser to identify the IP addresses of LXI certified instruments that are set up for
Model 2651A is displayed.
1. Confirm the LAN light on the instrument is
as part of the LAN RJ-45 connector.
2. Open an Internet browser, such as Microsoft
only).
3. If you do not know the IP address, press the MENU key on the instrument front panel and then
select LAN > STATUS >IP-ADDRESS.
4. In the Address box of the Internet browser, enter the IP address of the instrument and press
Enter.
The home page of the instrument web interface is displayed.
Web interface home page
The Welcome page of the web interface gives you basic information about the instrument, including:
•The instrument model, serial number, firmware revision, calibration date, and LXI information
•An ID button to help you locate the instrument
•Links to the instrument web applications, including TSB Embedded, Reading Buffers, Flash
Upgrade, and TSP
®
Express.
illuminated. The LAN light is located on the rear panel
®
Windows® Internet Explorer® (version 6.0 or later
IP configuration tab
The IP Configuration tab provides access to the Model 2651A LAN settings. Refer to the Model
2651A Reference Manual for additional information about IP configuration.
3-2 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 3: Using the web interface
NOTE
You must reload the page if you change the gateway or subnet mask from the Modify IP configuration
page. If you change the IP address, you must type the new IP address in the address bar before you
can use the web interface again.
Figure 3: Select IP configuration
TSB Embedded
TSB Embedded is a web application that includes a command-line interface that you can use to issue
commands and interact with the instrument. TSB Embedded also provides a convenient way to create
and manage user scripts. TSB Embedded resides in the instrument. You can use TSB Embedded to
run some of the application examples in other sections of this manual. If you can access the web
interface home page, but cannot use TSB Embedded (or TSP Express), make sure you have the
TM
Java
Create a script using TSB Embedded
If you are using TSB Embedded to create scripts, you do not need to use the commands
loadscript or loadandrunscript and endscript.
SE Runtime Environment (JRE) Version 6 or later installed on your computer.
NOTE
2651A-900-01 Rev. A / March 2011 3-3
Section 3: Using the web interface Model 2651A High Power System SourceMeter® Instrument User's Manual
Exercise: Create and run a script with TSB Embedded
The following programming example illustrates the setup and command sequence of a basic
source-measure procedure with the following parameters:
• Source function and range: Volts, autorange
• Source output level: 5 V
• Current compliance limit: 10 mA
• Measure function and range: Current, 10 mA
To create and run a sample script with TSB Embedded:
1. From the navigation on the left side of the web interface, select TSB Embedded. The TSB
Embedded page is displayed.
Figure 4: Select TSB Embedded
2. To create the example script, enter the name of the TSP script: basic_source_measure at the
location shown by (1) in the following illustration.
3-4 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 3: Using the web interface
Section 3: Using the web interface Model 2651A High Power System SourceMeter® Instrument User's Manual
NOTE
Commands and parameters for the Model 2651A are case-sensitive. It is important to type in the
commands exactly as shown to avoid syntax and execution errors.
4. Click Save Script (3). The script is added to the User Scripts list on the left.
Quick Tip
Standard edit functions, such as copy, cut, and paste, work in TSB Embedded.
5. Clear the buffer (4):
•In the console, type the following and then press Enter.
smua.nvbuffer1.clear()
•If successful, the command will appear in the Instrument Output box.
6. Set the buffer to append readings (4):
•In the console, type the following and then press Enter.
smua.nvbuffer1.appendmode = 1
•If successful, the command will appear in the Instrument Output box.
7. Set the buffer to collect timestamps (4):
•In the console, type the following and then press Enter.
smua.nvbuffer1.collecttimestamps = 1
•If successful, the command will appear in the Instrument Output box.
8. Run the script:
• Select the script in the User Scripts list.
• Click Run.
9. The Instrument Output box displays any error messages and output from the script.
10. If no errors appear, but readings do, click Run a few times to populate the buffer. Each time the
script is run, readings appear in the Instrument Output box and are also placed in the buffer.
Script management options
Existing scripts are listed in the User Scripts box on the left side of the web interface. To delete a
script, click the name of the script and click Delete. The script is deleted from the User Scripts list and
from the nonvolatile memory of the instrument. To stop operation of a script, click Abort Script.
To export the selected script to the computer, click Export to PC. Choose the directory in which to
save the script and click Save. Scripts are saved to a file with the extension tsp. TSP files are native
to Test Script Builder or TSB Embedded, but they can be opened and edited in any text editor.
To import scripts from the computer, click Import to PC. Select the directory that contains the file.
You can only import files with the extension tsp.
To clear the name box and the box that contains the script, click Clear. You can type the name of a
script in the TSP Script box and click View Script to view the contents of the script.
3-6 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 3: Using the web interface
Reading buffers
The Reading Buffers tab provides access to the Model 2651A reading buffers. The data used in this
example was created and placed in the buffer by the Exercise: Create and run a script with TSB
Embedded (on page 3-4).
Exercise: Retrieve readings from a buffer
To retrieve readings from a populated buffer:
1. From the navigation area on the left side of the web interface, select Reading Buffers. The
Reading Buffers embedded page is displayed.
Figure 6: Select reading buffer
2. Select smua.nvbuffer1 (1).
3. In the Recall Attributes area (near the bottom of the page), select Values and Time Stamp (2).
2651A-900-01 Rev. A / March 2011 3-7
4. Click Download (3). The table displays the present buffer data.
Section 3: Using the web interface Model 2651A High Power System SourceMeter® Instrument User's Manual
Figure 7: Downloading the reading buffer
1
2
3
5. Notice that the Source Value column is not populated. To collect source values, before taking
readings, use TSB Embedded to set the following attribute:
smua.nvbuffer1.collectsourcevalues = 1.
TSP Express
Figure 8: Sample downloaded reading buffer
The TSP® Express tab provides access to the TSP Express Launch page. From this page, click the
Launch button to start TSP Express.
3-8 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 3: Using the web interface
NOTE
Only one TSB Embedded or TSP Express session can be running and connected to the same
instrument.
Figure 9: Select TSP Express
To run TSP Express:
1. From the navigation area on the left side of the web interface, select TSP Express. The TSP
Express page is displayed.
2. Click the Launch button. The TSP Express window opens.
3. There is a help pane on the right side of the main window. You can adjust the width of the help
pane by sliding the vertical bar. Click and follow the step-by-step example to become familiar with
TSP Express (there is a link to the example in the TSP Express Introduction's first paragraph).
2651A-900-01 Rev. A / March 2011 3-9
Section 3: Using the web interface Model 2651A High Power System SourceMeter® Instrument User's Manual
Making front-panel measurements ........................................... 4-2
Introduction
You can use the Model 2651A High Power System SourceMeter® Instrument to make basic
measurements from the front panel. In this example, measurements are made on a 10 kΩ resistor.
Similar measurements can be made on any two-terminal device under test (DUT) if appropriate
source values are used.
Section 4
Making basic front-panel measurements
Equipment needed for this example
To run this example, you will need the following equipment:
• Model 2651A High Power System SourceMeter
• Additional cable and connector assemblies as required to make connections to the DUT (for
example, Model 2651A-KIT). See Device connections
connections.
Device connections
Connections from the Model 2651A High Power System SourceMeter® Instrument to the DUT are
shown in the diagram below. Proper care should be taken to ensure good contact through all
connections.
Figure 11: Two-wire resistance connections
SENSE/GUARD
S
GGGG
LO
S
HI
®
Instrument
(on page 4-1) for a schematic of required
OUTPUT
LO HI
40V, 50A MAX.
HI
DUT
LO
Section 4: Making basic front-panel measurements Model 2651A High Power System SourceMeter® Instrument User's Manual
Making front-panel measurements
Use the following procedure to configure the instrument and make measurements from the front
panel.
Step 1: Select and set source level
Perform the following steps to select the voltage source and set its value to 10 V:
1. Press the SRC key as needed to select the V-Source, as indicated by the units in the source field
on the display. The flashing digit (cursor) indicates which value is selected for editing.
2. Move the cursor to the digit to change, then press the navigation wheel
mode, as indicated by the EDIT indicator.
to enter the EDIT
3. Use the RANGE keys to select the 10 V range. If using a different source value, use the lowest
possible range for the best accuracy.
4. Set the source value to 10.0000 V, and then press the ENTER key or the navigation wheel
complete editing.
to
Step 2: Set compliance limit
Perform the following steps to edit the compliance limit value to 10 mA:
1. Press the LIMIT key.
2. Move the cursor to the digit to change, then press the navigation wheel
to enter the EDIT
mode, as indicated by the EDIT indicator.
3. Enter the limit value of 10.000 mA.
4. Press the ENTER key or the navigation wheel
to complete editing.
Step 3: Select measurement function and range
To select measurement function and range:
1. Select the current measurement function by pressing the MEAS key as needed.
2. Enable autorange by pressing the AUTO key (the AUTO indicator lights). Alternatively, you can
set manual ranging using the up or down RANGE keys.
Step 4: Turn output on
4-2 2651A-900-01 Rev. A / March 2011
Turn the output on by pressing the OUTPUT ON/OFF control. The OUTPUT indicator light turns on.
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 4: Making basic front-panel measurements
Step 5: Observe readings on the display
1. Observe the readings on the display. Press the TRIG key if necessary to trigger the instrument to
begin taking readings. The readings are on the top line, and source and limit values are on the
bottom line. For the 10 kΩ resistor, typical display values are:
1.00000mA SrcA: +10.0000 V LimA:010.0000mA
2. Press the MEAS key several times to display measured voltage, resistance, power, and current.
Typical values for the 10 kΩ resistor are:
Step 6: Turn output off
When finished, turn the output off by pressing the OUTPUT ON/OFF control. The OUTPUT indicator
light will turn off.
10.0000 V, 10.0000 kΩ, 10.0000 mW, and 1.00000 mA
2651A-900-01 Rev. A / March 2011 4-3
Capturing high power pulse waveforms
In this section:
Introduction to capturing waveforms ........................................ 5-1
Equipment needed for this example......................................... 5-1
Set up communication.............................................................. 5-1
Configuring the trigger model ................................................... 5-2
Example program code ............................................................ 5-4
Example program usage .......................................................... 5-9
Introduction to capturing waveforms
Section 5
The Model 2651A High Power System SourceMeter® Instrument offers a unique feature. In addition to
a traditional integrating ADC, the Model 2651A also features a fast-sampling ADC. The fast ADC of
the Model 2651A allows the instrument to take measurements at high speed, with capabilities similar
to an oscilloscope. This allows you to capture data at up to one million samples per second with more
than 5000 readings in a single burst. This feature allows simultaneous current and voltage
measurements without any additional hardware, and offers 18-bit resolution providing for high
accuracy, even at high speed. The fast ADC of the Model 2651A allows this instrument to be used in
a broad range of applications, such as negative bias temperature instability (NBTI) testing and
transient thermal analysis.
In this example, we will demonstrate how to use the fast ADC of the Model 2651A to capture both the
current and voltage waveforms of high power pulses.
Equipment needed for this example
To run this test, you will need the following equipment:
• Model 2651A High Power System SourceMeter
• GPIB cable or Ethernet cable to connect the Model 2651A to a computer
• Additional cable and connector assemblies as required to make connections to the device under
test (DUT) (for example, Model 2651A-KIT); see Device connections
schematic of required connections
®
Instrument
(on page 5-2) for a
Set up communication
The communication setup is illustrated in the following diagram. GPIB is used as an example, but this
application can be run using any of the supported communication interfaces for the instrument.
Section 5: Capturing high power pulse waveforms Model 2651A High Power System SourceMeter® Instrument User's Manual
CONFIG
Figure 12: GPIB communication example for fast ADC
D
E
O
GPIB
POWER
+3.21000V
SrcA:+20.0000 A LimAL10.0000V
78 90+/-
SRC MEAS LIMIT MODE
DISPLAY
4
56
DIGITS
REL
SPEED
12 3
LOAD
RUN
FILTER
0000
STORE
RECALL
CONFIG
I
T
T
/
E
H
2651A HIGH POWER SYSTEM SourceMeter
ARM
LOCAL
TRIG
EXIT
MENU
ENTER
N
S
T
U
E
P
R
R
P
E
U
T
S
N
H
E
/
T
T
I
O
E
D
P
T
U
CURSOR
AUTO
O
O
N
/
O
®
U
T
F
F
Controller
Device connections
Make the connections as illustrated in the following figure. To ensure good contact, use care when
making connections.
SENSE/GUARD
LO
Figure 13: Connections for fast ADC example
Model 2651A
OUTPUT
S
GGGG
S
HI
Equivalent
LO HI
40V, 50A MAX.
HI
S HI
HI
S HI
LO
LED
(DUT)
S LO
SMU
LED
(DUT)
LO
S LO
Configuring the trigger model
To output precisely timed high power pulses and accurately capture the pulse with the fast ADC, the
trigger model of the Model 2651A High Power System SourceMeter
the trigger model, we can output pulses as large as 50 A with pulse widths as small as 100 μs and
position our measurements with microsecond resolution. The image below is a diagram of the trigger
model used in this example.
5-2 2651A-900-01 Rev. A / March 2011
®
Instrument must be used. Using
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 5: Capturing high power pulse waveforms
Figure 14: Trigger model configuration for fast ADC example
Model 2651A
Local node*
trigger.timer[1].
stimulus
passthrough = true
trigger.timer[2].
stimulus
passthrough = false
0.03 s
3 ms
EVENT_ID
count = 4
EVENT_ID
count = 1
smua.trigger.
Idle
Arm layer
arm.stimulus
Trigger layer
source.stimulus
measure.stimulus
endpulse.stimulus
arm.count = 1
SWEEP_COMPLETE_EVENT_ID
count = 5
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
IDLE_EVENT_ID
SWEEPING_EVENT_ID
ARMED_EVENT_ID
* Local node: Since only one instrument is being used in this example, the local node is used.
In this example, timer 1 is used to control the time between pulses while timer 2 is used to control the
pulse width. Timer 1 also controls the start of the measurements, starting them at the same time it
triggers the pulse and thus overlapping the measurements with the pulse.
When the trigger model of the Model 2651A is initialized, the following occurs:
1. The source-measure unit (SMU) trigger model:
• Leaves the idle state
• Flows through the Arm Layer
• Enters the Trigger Layer
• Outputs the ARMED event trigger
• Pauses at the Source Event, where it waits for an event trigger
2. Timer 1 receives the ARMED event trigger and begins its countdown while passing the event
trigger through to both the SMU's Source Event and Measure Event simultaneously.
3. The SMU Measure Event receives the event trigger from timer 1 and begins to take
measurements. Because measurements have been configured to be asynchronous, the Measure
Event begins its execution before the Source Event has completed and will continue to take
measurements even after the End Pulse Event has executed. For more information, see "Using
the remote trigger model" in the Model 2651A Reference Manual.
4. The SMU Source Event receives the event trigger from timer 1, begins to output the pulse, and
then waits the programmed source delay (if any) before it outputs the SOURCE_COMPLETE
event to timer 2 and allows the SMU trigger model to continue.
2651A-900-01 Rev. A / March 2011 5-3
Section 5: Capturing high power pulse waveforms Model 2651A High Power System SourceMeter® Instrument User's Manual
5. Timer 2 receives the SOURCE_COMPLETE_EVENT_ID event trigger from timer 1 and begins its
countdown.
6. The SMU trigger model continues to the End Pulse Event, where it pauses and waits for an event
trigger from timer 2.
7. The timer 2 countdown expires and timer 2 outputs an event trigger to the SMU End Pulse Event.
8. The SMU End Pulse Event receives the event trigger from timer 2 and outputs the falling edge of
the pulse, after which it allows the SMU trigger model to continue.
9. The SMU trigger model then compares the current Trigger Layer loop iteration with the trigger
count:
•If the current iteration is less than the trigger count, the Trigger Layer repeats and the SMU trigger
model reaches Source Event, where it waits for another trigger from timer 1. The trigger model then
repeats from step 3.
•If the current iteration is equal to the trigger count, the SMU trigger model exits the Trigger Layer,
passes through the Arm Layer, and returns to idle.
Example program code
NOTE
The example code is designed to be run from Test Script Builder or TSB Embedded. It can also be
run from other programming environments, such as Microsoft® Visual Studio® or National Instruments
LabVIEWTM. However, modification of the example code may be required to do so.
The following script contains all the code necessary to output and capture a current or voltage pulse
using the Model 2651A fast ADC. The script sets up the source-measure unit (SMU), configures the
trigger model, prepares the reading buffers, executes the test, and outputs the collected data in a
format that can be copied and pasted from the instrument console to a Microsoft
spreadsheet.
The script is written using Test Script Processor (TSP®) functions rather than as a single block of
inline code. TSP functions are similar to functions in other programming languages, such as
Microsoft
®
Visual C® or Visual Basic®, and must be called before the code contained in them is
executed. Because of this, running the script alone will not execute the test. To execute the test, run
the script to load the functions into test script memory and then call the functions. Refer to the
documentation for Test Script Builder or TSB Embedded
(on page 3-3) for directions on how to run
scripts and enter commands using the instrument console.
®
Excel®
The script contains comments that describe what is being performed by the lines of code and
documentation for the functions in the script.
5-4 2651A-900-01 Rev. A / March 2011
--[[
Title: Fast ADC Usage
Description: This script is designed to output pulses and capture
both the current and the voltage of the pulse using the fast ADC of
the Model 2651A High Power System SourceMeter instrument.
Equipment Needed:
1x 2651A
]]
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 5: Capturing high power pulse waveforms
--[[
Name: CapturePulseV(pulseLevel, pulseWidth, pulseLimit, numPulses)
Description:
This function outputs voltage pulses with a 1% duty cycle and
performs measurements using the fast ADC to capture each pulse in
its entirety. At the conclusion of the pulse train, the data is
returned to the instrument console in a Microsoft Excel compatible format.
Parameters:
pulseLevel: The voltage level of the pulse in volts
pulseWidth: The width of the pulse in seconds
100e-6 <= pulseWidth <= 4e-3
pulseLimit: The current limit of the pulse in amps
numPulses: The number of pulses to output
Example Usage:
CapturePulseV(5, 300e-6, 50, 5)
]]
function CapturePulseV(pulseLevel, pulseWidth, pulseLimit, numPulses)
if (numPulses == nil) then numPulses = 1 end
-- Configure the SMU
reset()
smua.reset()
smua.source.func = smua.OUTPUT_DCVOLTS
smua.sense = smua.SENSE_REMOTE
smua.source.rangev = pulseLevel
smua.source.levelv = 0 -- The bias level
smua.source.limiti = 5 -- The DC limit
-- Use a measure range that is as large as the biggest
-- possible pulse
smua.measure.rangei = pulseLimit
smua.measure.rangev = pulseLevel
-- Select the fast ADC for measurements
smua.measure.adc = smua.ADC_FAST
-- Set the measure count to be 1.25 times the width of the pulse
-- to ensure we capture the entire pulse plus falling edge.
smua.measure.count = (pulseWidth / smua.measure.interval) * 1.25
-- Pulses will all be the same level so set start and stop to
-- the same value and the number of points in the sweep to 2
smua.trigger.source.linearv(pulseLevel, pulseLevel, 2)
smua.trigger.source.limiti = pulseLimit
smua.trigger.measure.action = smua.ASYNC
-- We want to start the measurements before the source action takes
-- place, so we must configure the ADC to operate asynchronously of
-- the rest of the SMU trigger model actions
-- Measure I and V during the pulse
smua.trigger.measure.iv(smua.nvbuffer1, smua.nvbuffer2)
-- Return the output to the bias level at the end of the pulse/sweep
smua.trigger.endpulse.action = smua.SOURCE_IDLE
smua.trigger.endsweep.action = smua.SOURCE_IDLE
smua.trigger.count = numPulses
smua.trigger.arm.stimulus = 0
smua.trigger.source.stimulus = trigger.timer[1].EVENT_ID
smua.trigger.measure.stimulus = trigger.timer[1].EVENT_ID
smua.trigger.endpulse.stimulus = trigger.timer[2].EVENT_ID
smua.trigger.source.action = smua.ENABLE
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 5: Capturing high power pulse waveforms
--[[
Name: CapturePulseI(pulseLevel, pulseWidth, pulseLimit,
numPulses)
Description:
This function outputs current pulses with a 1% duty cycle and performs
measurements using the fast ADC to capture each pulse in its entirety.
At the conclusion of the pulse train, the data is returned to the
instrument console in a Microsoft Excel compatible format.
Parameters:
pulseLevel: The current level of the pulse in amps
pulseWidth: The width of the pulse in seconds
100e-6 <= pulseWidth <= 4e-3
pulseLimit: The voltage limit of the pulse in volts
numPulses: The number of pulses to output
Example Usage:
CapturePulseI(50, 300e-6, 10, 5)
]]
function CapturePulseI(pulseLevel, pulseWidth, pulseLimit, numPulses)
if (numPulses == nil) then
numPulses = 1
end
-- Configure the SMU
reset()
smua.reset()
smua.source.func = smua.OUTPUT_DCAMPS
smua.sense = smua.SENSE_REMOTE
smua.source.rangei = pulseLevel
smua.source.leveli = 0 -- The bias level
smua.source.limitv = 10 -- The DC limit
-- Use a measure range that is as large as the biggest possible pulse.
smua.measure.rangev = pulseLimit
smua.measure.rangei = pulseLevel
-- Select the fast ADC for measurements
smua.measure.adc = smua.ADC_FAST
-- Set the measure count to be 1.25 times the width of the pulse
-- to ensure we capture the entire pulse plus falling edge.
smua.measure.count = pulseWidth / smua.measure.interval) * 1.25
-- Pulses will all be the same level so set start and stop to
-- the same value and the number of points in the sweep to 2
smua.trigger.source.lineari(pulseLevel, pulseLevel, 2)
smua.trigger.source.limitv = pulseLimit
smua.trigger.measure.action = smua.ASYNC
-- We want to start the measurements before the source action takes
-- place, so we must configure the ADC to operate asynchronously of
-- the rest of the SMU trigger model actions
-- Measure I and V during the pulse
smua.trigger.measure.iv(smua.nvbuffer1, smua.nvbuffer2)
-- Return the output to the bias level at the end of the pulse
smua.trigger.endpulse.action = smua.SOURCE_IDLE
smua.trigger.endsweep.action = smua.SOURCE_IDLE
smua.trigger.count = numPulses
smua.trigger.arm.stimulus = 0
smua.trigger.source.stimulus = trigger.timer[1].EVENT_ID
smua.trigger.measure.stimulus = trigger.timer[1].EVENT_ID
smua.trigger.endpulse.stimulus = trigger.timer[2].EVENT_ID
smua.trigger.source.action = smua.ENABLE
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 5: Capturing high power pulse waveforms
--[[
Name: PrintPulseData()
Description;
This function prints the data contained in smua.nvbuffer1 and
smua.nvbuffer2 in a format that is copy and paste compatible with
Microsoft Excel.
]]
function PrintPulseData()
print("Timestamp\tVoltage\tCurrent")
for i=1, smua.nvbuffer1.n do
print(string.format("%g\t%g\t%g", smua.nvbuffer1.timestamps[i],
smua.nvbuffer2[i], smua.nvbuffer1[i]))
end
end
Example program usage
The functions in this script allow the parameters of the test to be adjusted without rewriting and
rerunning the script. To execute the test, either call CapturePulseV() (to capture voltage) or
CapturePulseI() (to capture current), passing in the appropriate values as parameters.
The parameters of the CapturePulseV() function are in the following table:
CapturePulseV() parameters
Parameter Units Description
pulseLevel
pulseWidth
pulseLimit
numPulses
Volts The voltage level of the pulse
Seconds The width of the pulse
Amperes The current limit of the pulses to output
Not applicable The number of pulses to output
An example call to this function is as follows:
CapturePulseV(10, 300e-6, 50, 5)
This call will output five 10 V pulses with a 300 μs pulse width. The pulses will be limited to 50 A and
have a 1 percent duty cycle. At the completion of the pulsed outputs, the source-measure unit (SMU)
output is turned off. The resulting data from this test will be returned in a Microsoft
®
Excel® compatible
format (you can cut and paste the output from the console) that can be used for graphing and
analysis.
The parameters of the CapturePulseI() function are contained in the following table:
CapturePulseI() parameters
Parameter Units Description
pulseLevel Amperes The current level of the pulse
pulseWidth Seconds The width of the pulse
pulseLimit Volts The voltage limit of the pulses to output
numPulses Not applicable The number of pulses to output
2651A-900-01 Rev. A / March 2011 5-9
Section 5: Capturing high power pulse waveforms Model 2651A High Power System SourceMeter® Instrument User's Manual
An example call to this function is as follows:
CapturePulseI(50, 300e-6, 50, 5)
This call will output five 50 A pulses with a 300 μs pulse width. The pulses will be limited to 10 V and
have a 1% duty cycle. At the completion of the pulsed outputs, the SMU output is turned off. The
resulting data from this test will be returned in a Microsoft
®
Excel® compatible format (you can cut and
paste the output from the console) that can be used for graphing and analysis.
5-10 2651A-900-01 Rev. A / March 2011
Characterization of power discretes
In this section:
Introduction to power discrete I-V curve testing ....................... 6-1
Equipment needed for this example......................................... 6-2
Set up communication.............................................................. 6-2
Configuring the trigger model ................................................... 6-5
Example program code .......................................................... 6-10
Example program usage ........................................................ 6-16
Introduction to power discrete I-V curve testing
Section 6
A common test performed on power MOSFETs and IGBTs characterizes the I-V performance of the
device under test (DUT) at various gate voltages. In this test, the gate of the device is held at a
constant voltage while the voltage on the drain (or collector) of the device is swept and current is
measured. This process is repeated for several different gate voltages, and the resulting data is
plotted to show a series of I-V curves for the device. The following figure illustrates an example of this
plot for a power MOSFET device.
Figure 15: Example set of IV curves for a power MOSFET device
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
To avoid device self-heating, this test is usually performed using pulsed measurements. This method
pulses voltage across the device for a short duration. During this pulse, after the pulse has settled,
the current is measured. The duration of the pulse and the duty cycle in which the pulses are applied
to the device are controlled, minimizing self-heating. A single Model 2651A High Power System
SourceMeter
®
Instrument can pulse-measurement test with precision timing at currents up to 50 A
and pulse widths as low as 100 µs.
This example demonstrates how to perform pulsed measurements and collect I-V curves using the
Model 2651A High Power System SourceMeter
®
could be used to test IGBTs.
Equipment needed for this example
To run this test, you will need the following equipment:
• Model 2651A High Power System SourceMeter
• Series 2600A System SourceMeter
• TSP-Link
®
cable (Keithley Instruments part number CA-180-3A)
®
Instrument (see NOTE)
• GPIB cable or Ethernet cable to connect the Model 2651A to a computer
• Additional cable and connector assemblies as required to make connections to the device under
test (DUT) (for example, Model 2651A-KIT and Model 2600-KIT); see Device connections
page 6-4) for a schematic of required connections
NOTE
For this application, Series 2600A System SourceMeter® Instrument refers to any of the following
instrument models: Models 2601A/2602A, 2611A/2612A, or 2635A/2636A.
Instrument on a MOSFET. The same procedure
®
Instrument
(on
Set up communication
The communication setup is illustrated in the following diagram. GPIB is used as an example, but this
application can be run using any of the supported communication interfaces for the instruments. A
TSP-Link
Series 2600A (on TSP-Link node 2) are sent over the TSP-Link interface.
6-2 2651A-900-01 Rev. A / March 2011
®
connection enables communication between the two instruments. Commands for the
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 6: Characterization of power discretes
Figure 16: GPIB communication overview
GPIB
Model 2651A
(TSP-Link Node #1)
TSP-Link
Controller
Series 26xxA*
(TSP-Link Node #2)
* Model 260xA, Model 261xA,
or Model 263xA
When using the TSP-Link communication interface, each instrument must have a unique TSP-Link
node number. Configure the node number for the:
• Model 2651A to 1
• Series 2600A to 2
To set the TSP-Link node number using the front-panel interface:
1. Press the MENU key.
2. Select TSPLink.
3. Select NODE.
4. Use the navigation wheel
5. Press the ENTER key to save the TSP-Link node number.
to adjust the node number.
On the Model 2651A SMU #1 (TSP-Link node #1), perform a TSP-Link reset to update it with the
linked instruments:
1. Press the MENU key.
2. Select TSPLink.
3. Select RESET.
NOTE
If error 1205 is generated during the TSP-link reset, ensure that the Series 2600A has a unique
TSP-Link node number.
Quick Tip
You can also perform a TSP-Link reset from the remote command interface by sending
tsplink.reset() to the Model 2651A. This method is used in the example program.
2651A-900-01 Rev. A / March 2011 6-3
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
Device connections
To run the test, make the connections as illustrated in the following figure. To ensure good contact,
use care when making connections.
NOTE
For best results, all connections should be left floating (no connections should be tied to ground).
Figure 17: Connections for IV curve characterization
Model 2651A
(master node #1)
Equivalent
D
SENSE/GUARD
S
GGGG
LO
OUTPUT
S
HI
LO HI
40V, 50A MAX.
LO
HI
S HI
SMU
(subordinate
node #2)
LO
S LO
G
Gate resistor
(if required)
HI
S
S HI
SMU
(master
node #1)
LO
S LO
HI
S HI
Series 26xxA*
(subordinate node #2)
D
N-Channel
G
power
MOSFET
LO
S
S LO
LO
LO G HI G G G
S
LO
S
HI
Gate resistor
(if required)
S LO
* Model 260xA, Model 261xA,
or Model 263xA
NOTE
During high-current pulsing through the channel of the device, the gate may begin to oscillate. This
oscillation creates an unstable voltage on the gate, which may cause unstable measurements through
the device channel. You can install a resistor to dampen this oscillation and stabilize the gate. Install
the resistor between the gate of the device and the force and sense HI of the Model 2651A. If the
gate still remains unstable, enable High-C mode on the Model 2651A source-measure unit (SMU).
6-4 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 6: Characterization of power discretes
Configuring the trigger model
Correctly configuring the trigger model allows you to achieve pulsed measurements with precise
timing and current levels up to 50 A on the Model 2651A. The following figure illustrates this test's
overall trigger model configuration, including the interactions between the Model 2651A and the
Series 2600A instrument. Refer to the Example program code
of this trigger model.
In this example, the trigger model of each instrument is configured to perform a linear sweep.
The Series 2600A is configured to sweep the voltage applied to the gate of the device.
The Model 2651A is configured to sweep the voltage across the drain (or collector) of the device. The
Model 2651A is configured to repeat its sweep and will do so as many times as there are steps in the
sweep of the Series 2600A.
For example, if the Model 2651A instrument’s sweep contains four points and the Series 2600A
instrument’s sweep contains three points, the Model 2651A instrument’s sweep will run three times,
for a total of 12 points. To synchronize the sweeps between instruments, the trigger lines of the TSP-
®
interface are used. The following figure illustrates the 12-point sweep example.
Link
(on page 6-10) for an implementation
Figure 18: SMU output from configured trigger model
4V
Gate
(Series 26xxA)
0V
4V
Drain
or Collector
(Model 2651A)
0V
2651A-900-01 Rev. A / March 2011 6-5
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
Figure 19: Trigger model configuration for IV curve characterization
Model 2651A (master)
node[1].
trigger.timer[1].
stimulus
passthrough = true
trigger.timer[2].
stimulus
passthrough = false
0.03 s
0.001 s
EVENT_ID
count = 20
EVENT_ID
count = 1
smua.trigger.
Idle
Arm layer
arm.stimulus
Trigger layer
source.stimulus
measure.stimulus
endpulse.stimulus
IDLE_EVENT_ID
SWEEPING_EVENT_ID
arm.count = 5
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
count = 21
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
tsplink.trigger[1].
mode = tsplink.TRIG_SYNCHRONOUSM
tsplink.trigger[2].
mode = tsplink.TRIG_FALLING
Series 26xxA (subordinate)*
node[2].
tsplink.trigger[1].
mode = tsplink.TRIG_SYNCHRONOUSA
tsplink.trigger[2].
mode = tsplink.TRIG_FALLING
EVENT_ID
stimulus
EVENT_ID
stimulus
EVENT_ID
stimulus
EVENT_ID
stimulus
smua.trigger.
Idle
Arm layer
arm.stimulus
Trigger layer
source.stimulus
measure.stimulus
endpulse.stimulus
IDLE_EVENT_ID
SWEEPING_EVENT_ID
arm.count = 1
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
count = 5
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
* Model 260xA, Model 261xA, or Model 263xA
6-6 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 6: Characterization of power discretes
Model 2651A trigger model operation
The following figure illustrates this test's Model 2651A trigger model configuration.
Figure 20: Model 2651A trigger model for IV curve characterization
Model 2651A (master)
node[1].
trigger.timer[1].
stimulus
passthrough = true
trigger.timer[2].
stimulus
passthrough = false
0.03 s
0.001 s
EVENT_ID
count = 20
EVENT_ID
count = 1
smua.trigger.
Idle
Arm layer
arm.stimulus
Trigger layer
source.stimulus
measure.stimulus
endpulse.stimulus
IDLE_EVENT_ID
SWEEPING_EVENT_ID
arm.count = 5
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
count = 21
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
tsplink.trigger[1].
mode = tsplink.TRIG_SYNCHRONOUSM
tsplink.trigger[2].
mode = tsplink.TRIG_FALLING
EVENT_ID
stimulus
EVENT_ID
stimulus
TSP-Link bus
2651A-900-01 Rev. A / March 2011 6-7
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
When the configured trigger model of the Model 2651A is initiated, the following occurs:
1. The source-measure unit (SMU) trigger model passes through the Arm Layer, enters the Trigger
Layer, then outputs the ARMED event trigger.
2. The ARMED event trigger triggers TSP-Link
®
trigger 1, which outputs a trigger on TSP-Link line 1.
This trigger is meant to tell the Series 2600A that it should step the gate SMU.
3. The SMU trigger model continues until it reaches the Source Event and then pauses and waits for
an event trigger.
4. A trigger comes back on TSP-Link trigger 1, indicating that the gate step is complete. An event
trigger is output from TSP-Link trigger 1 and input to timer 1.
5. Timer 1 receives the event trigger and begins its countdown while simultaneously passing the
event trigger.
6. The SMU Source Event Detector receives the event trigger from timer 1, sets the SMU output to
the pulse level, waits for a programmed source delay (if any), and outputs the
SOURCE_COMPLETE event to timer 2.
7. Timer 2 receives the SOURCE_COMPLETE event trigger and begins its countdown.
8. The SMU trigger model moves to the Measure Event, where it waits for a programmed measure
delay (if any), then takes a measurement.
9. The SMU trigger model moves to the End Pulse Event, where it stops and waits for an event
trigger from timer 2.
10. The timer 2 countdown expires and outputs an event trigger to the SMU End Pulse Event.
11. The End Pulse Event Detector receives the event trigger from timer 2, sets the SMU output to the
bias level, and then continues.
12. The trigger model checks the current Trigger Layer iteration number against the trigger count.
•If the iteration number is less than the trigger count, the trigger model stays in the trigger layer and
wraps around to the Source Event, where it waits again for an event trigger from timer 1 (as it did in step
3). This time, however, timer 2 will not receive an event trigger from TSP-Link trigger 1. Despite this, the
trigger model will continue running because the timer 1 count was set to one less than the trigger count.
Therefore, timer 1 will continue outputting event triggers while the trigger model is in the Trigger Layer,
outputting one trigger event every time the countdown expires. The trigger model continues as
described in step 6.
•If the iteration number is greater than the trigger count, the SMU trigger model exits the Trigger Layer
and goes into the Arm Layer.
13. Upon entering the Arm Layer, the SMU trigger model outputs the SWEEP_COMPLETE event
trigger.
14. TSP-Link trigger 2 receives the SWEEP_COMPLETE event trigger and outputs a trigger on TSPLink line 2. This trigger is meant to signal the Series 2600A that it is acceptable to continue, and it
no longer needs to hold the current gate level.
15. The trigger model checks the current Arm Layer iteration number against the arm count.
•If the iteration number is less than the arm count, the trigger model stays in the Arm Layer and begins
the entire process again from step 1.
•If the iteration number is greater than the arm count, the trigger model continues, exits the Arm Layer,
and returns to an iIdle state.
6-8 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 6: Characterization of power discretes
Series 26xxA instrument's trigger model operation
The following figure illustrates the Series 26xxA trigger model configuration for this test.
Figure 21: Series 26xxA instrument's trigger model for IV curve characterization
TSP-Link bus
Series 26xxA (subordinate)*
node[2].
tsplink.trigger[1].
mode = tsplink.TRIG_SYNCHRONOUSA
tsplink.trigger[2].
mode = tsplink.TRIG_FALLING
EVENT_ID
stimulus
EVENT_ID
stimulus
smua.trigger.
Idle
Arm layer
arm.stimulus
Trigger layer
source.stimulus
measure.stimulus
endpulse.stimulus
IDLE_EVENT_ID
SWEEPING_EVENT_ID
arm.count = 1
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
count = 5
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
* Model 260xA, Model 261xA, or Model 263xA
When the configured trigger model of the Series 2600A instrument is initiated, the following occurs:
1. The source-measure unit (SMU) trigger model passes through the Arm Layer, enters the Trigger
Layer, reaches the Source Event, pauses, and then waits for an event trigger from TSP-Link
®
trigger 1.
2. A TSP-Link trigger is received by TSP-Link trigger 1. Because TSP-Link trigger 1 is configured as
a synchronous acceptor, it latches the line low and outputs an event trigger to the SMU Source
Event.
3. The SMU Source Event Detector receives the event trigger from TSP-Link trigger 1, sets the
SMU output to the step level, waits a programmed source delay (if any) and continues.
4. The SMU trigger model moves to the Measure Event, where it waits a programmed measure
delay (if any), takes a measurement, and then outputs the MEASURE_COMPLETE event trigger.
5. TSP-Link trigger 1 receives the MEASURE_COMPLETE event trigger and releases the TSP-link
trigger line, telling the Model 2651A that it has completed its step and that the Model 2651A may
begin its sweep.
6. The SMU trigger model moves to the End Pulse Event, where it stops and waits for an event
trigger from TSP-Link trigger 2.
7. A TSP-Link trigger is received by TSP-Link trigger 2, indicating that the Model 2651A instrument
sweep has completed and it is safe to go to the next step. TSP-Link trigger 2 outputs an event
trigger to the SMU End Pulse Event Detector.
8. The SMU End Pulse Event Detector receives an event trigger from TSP-Link trigger 2, completes
2651A-900-01 Rev. A / March 2011 6-9
the End Pulse Action, and then continues.
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
9. The trigger model checks the current Trigger Layer iteration number against the trigger count.
•If the iteration number is less than the trigger count, the trigger model stays in the trigger layer and
wraps around to the Source Event, where it waits again for an event trigger from TSP-Link Trigger 1.
The trigger model repeats from step 2.
•If the iteration number is greater than the trigger count, the SMU trigger model exits the Trigger Layer
and goes into the Arm Layer.
10. Because the arm count is set to 1, the trigger model exits the Arm Layer and returns to an idle
state.
Example program code
NOTE
The example code is designed to be run from Test Script Builder or TSB Embedded. It can also be
run from other programming environments, such as Microsoft® Visual Studio® or National Instruments
LabVIEWTM. However, modification of the example code may be required to do so.
The following is a Test Script Processor (TSP®) script that performs a series of sweeps to collect I-V
curve data for power MOSFETs and IGBT devices. The script contains all the code necessary to:
•Initialize the TSP-Link
®
connection
•Configure the output and trigger models of the source-measure unit (SMU), as described in the
previous section
• Prepare the reading buffers
• Execute the test
• Return the collected data in a format that can be copied from the instrument console and pasted
into Microsoft
The script is written using Test Script Processor (TSP®) functions rather than as a single block of
inline code. TSP functions are similar to functions in other programming languages, such as
Microsoft
®
Visual C® or Visual Basic®, and must be called before the code contained in them is
executed. Because of this, running the script alone will not execute the test. To execute the test, run
the script to load the functions into test script memory and then call the functions. Refer to the
documentation for Test Script Builder or TSB Embedded for directions on how to run scripts and enter
commands using the instrument console.
The script contains comments that describe what is being performed by the lines of code and
documents the functions in the script.
Lines starting with node[2]. are commands that are being sent to the Series 26xxA through the
TSP-Link interface. All other commands are executed on the Model 2651A.
®
Excel®
6-10 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 6: Characterization of power discretes
--[[
Title: IV Curves Example Script
Description: This script will perform a series of IV curves on a
MOSFET or IGBT device and will return the data in a Microsoft Excel
compatible format for graphing and analysis.
Description: This function will perform a series of pulsed sweeps
on a MOSFET or IGBT device to generate a series of IV Curves that
characterize the device.
Note: To avoid device oscillations, a series resistor on the gate
terminal of the device may be required.
Parameters:
gstart: The starting voltage of the gate sweep
gstop: The ending voltage of the gate sweep
gsteps: The number of steps in the gate sweep
dstart: The starting voltage of the drain sweep
dstop: The ending voltage of the drain sweep
dsteps: The number of steps in the drain sweep
pulseWidth: The width of the drain pulse in seconds
pulsePeriod: The time in seconds between the start of
consecutive drain pulses in the sweep
pulseLimit: The current limit in amperes of the drain pulse
function IV_Curves(gstart, gstop, gsteps, dstart, dstop, dsteps, pulseWidth,
pulsePeriod, pulseLimit)
reset()
tsplink.reset(2) -- Verify that at least two nodes are present
-- Select the source range that is large enough to fit all values of the sweep
smua.source.levelv = 0 -- Sets the drain bias level
smua.source.limiti = 5
smua.measure.rangev = smua.source.rangev
smua.measure.rangei = (pulseLimit == "off") and 50 or pulseLimit
-- Select a measure range large enough to fit pulses up to the current limit
2651A-900-01 Rev. A / March 2011 6-11
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
-- NPLC can be increased to improve measurement accuracy.
-- However, it should remain small enough to fit the measurement
-- within the width of the settled part of the pulse.
smua.measure.delay = (pulseWidth - ((1/localnode.linefreq) *
smua.measure.nplc)) - 50e-6
-- Set the measure delay so that the measurement is
-- taken at the end of the pulse before the falling edge
-- Timer 1 controls the pulse period
trigger.timer[1].count = (dsteps <= 1) and 1 or (dsteps - 1)
-- If dsteps <= 1 then use 1 for the count else use dsteps - 1
trigger.timer[1].delay = pulsePeriod
trigger.timer[1].passthrough = true
trigger.timer[1].stimulus = tsplink.trigger[1].EVENT_ID
trigger.timer[1].clear()
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
-- Prepare the Gate SMU (26xxA) reading buffers
node[2].smua.nvbuffer1.clear()
node[2].smua.nvbuffer1.collectsourcevalues = 1
if node[2].smua.nvbuffer1.fillmode ~= nil then
node[2].smua.nvbuffer1.fillmode = node[2].smua.FILL_ONCE
end
node[2].smua.nvbuffer2.clear()
node[2].smua.nvbuffer2.collectsourcevalues = 1
if node[2].smua.nvbuffer1.fillmode ~= nil then
node[2].smua.nvbuffer1.fillmode = node[2].smua.FILL_ONCE
end
-- The SMUs are configured and ready to run the test
-- Outputs on
node[2].smua.source.output = node[2].smua.OUTPUT_ON
smua.source.output = smua.OUTPUT_ON
if errorqueue.count > 0 then
print("Errors were encountered")
reset()
return
end
-- Start the 26xxA trigger model
node[2].smua.trigger.initiate()
-- Start the 2651A trigger model
smua.trigger.initiate()
waitcomplete() -- Wait until the sweeps are complete
-- Outputs off
smua.source.output = smua.OUTPUT_OFF
node[2].smua.source.output = node[2].smua.OUTPUT_OFF
-- Return the data
PrintIVcurveData(gsteps, dsteps)
end
6-14 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 6: Characterization of power discretes
--[[
PrintIVcurveData(gsteps, dsteps)
Description: This function will output the data collected by the
IV_Curves() function in a Microsoft Excel compatible format. For each
step of the gate, this function will output three columns that contain
the drain sweep data and the gate data in the first row.
Parameters:
gsteps: The number of steps in the gate sweep
dsteps: The number of steps in the drain sweep
Example Usage:
PrintIVcurveData(5, 21)
--]]
function PrintIVcurveData(gsteps, dsteps)
line1 = ""
line2 = ""
for i = 1,gsteps do
line1 = line1 .. string.format("Vgs = %0.2f\t%g\t%g\t",
node[2].smua.nvbuffer1.sourcevalues[i], node[2].smua.nvbuffer2[i],
node[2].smua.nvbuffer1[i])
line2 = line2 .. "Source Value\tVoltage\tCurrent\t"
end
print(line1)
print(line2)
for i = 1, dsteps do
line = ""
for j = 1, gsteps do
line = line .. string.format("%g\t%g\t%g\t",
smua.nvbuffer1.sourcevalues[(j - 1) * dsteps + i],
smua.nvbuffer2[(j - 1) * dsteps + i],
smua.nvbuffer1[(j - 1) * dsteps + i])
end
print(line)
end
end
2651A-900-01 Rev. A / March 2011 6-15
Section 6: Characterization of power discretes Model 2651A High Power System SourceMeter® Instrument User's Manual
Example program usage
The functions of this script allow the sweep parameters of the test to be adjusted without having to
rewrite and rerun the script. A test can be executed by calling the function IV_Curves(gstart, gstop, gsteps, dstart, dstop, dsteps, pulseWidth, pulsePeriod, pulseLimit)
with the appropriate values passed to it in its parameters.
IV_Curves() parameters
Parameter Units Description
gstart
gstop
gsteps
dstart
dstop
dsteps
pulseWidth
pulsePeriod
pulseLimit
An example call to this function is as follows:
IV_Curves(5, 9, 5, 0, 10, 21, 300e-6, 30e-3, 50)
Volts The level of the first step in the gate sweep
Volts The level of the last step in the gate sweep
Volts The number of steps in the gate sweep
Volts The level of the first step in the drain/collector sweep
Volts The level of the last step in the drain/collector sweep
Not
applicable
Seconds The width of the pulse in the drain/collector sweep
Seconds The time between the start of consecutive pulses in the
Amperes The current limit of the pulses in the drain/collector sweep
The number of steps in the drain/collector sweep
drain/collector sweep
This call sweeps the drain of the device from 0 V to 10 V in 21 points. The points of this sweep are
gathered using pulsed measurements with a pulse width of 300 µs and a pulse period of 30 ms for a
1 percent duty cycle. These pulses are limited to a maximum of 50 A. This sweep is performed at five
different gate voltages, starting with 5 V and ending at 9 V. The resulting data from this test will be
returned in a Microsoft
®
Excel® compatible format (you can cut and paste the output from the console)
that can be used for graphing and analysis.
6-16 2651A-900-01 Rev. A / March 2011
Increasing SMU current sourcing ability
In this section:
Introduction to combining SMUs .............................................. 7-1
Combining SMUs to increase current sourcing capability ........ 7-2
Equipment needed for this example......................................... 7-3
Set up communication.............................................................. 7-3
Configuring the trigger model ................................................... 7-6
Example program code ............................................................ 7-9
Example program usage ........................................................ 7-16
Introduction to combining SMUs
Section 7
A common test performed on power MOSFETs characterizes the resistance of the device under test
(DUT). In this test, the gate of the DUT is held at a constant voltage (representing the on-state of the
device) while a current is swept through the drain and the voltage is measured. The resulting data can
be plotted to show drain current versus Rds(on) for the DUT. The following figure is an example of
this plot.
Figure 22: Example Rds(on) curves for a power MOSFET device
0.02
0.019
0.018
0.017
0.016
0.015
Rds(on)
0.014
0.013
0.012
0.011
0.01
0 10 20 30 40 50 60 70 80 90 100
Ids (amperes)
Vgs = 10 V
Vgs = 6 V
Section 7: Increasing SMU current sourcing ability Model 2651A High Power System SourceMeter® Instrument User's Manual
To avoid device self-heating, this test is usually performed using pulsed measurements. This method
pulses current across the device for a short duration. During this pulse (after the pulse has settled),
the voltage is measured. The duration of the pulse and the duty cycle in which the pulses are applied
to the device are controlled, minimizing self-heating. A single Model 2651A High Power System
SourceMeter
®
Instrument can pulse-measurement test with precision timing at currents up to 50 A
and pulse widths as low as 100 µs.
To increase the pulse output, you can combine two Model 2651A High Power System SourceMeter
Instruments to output a pulse that is greater than that of a single Model 2651A. This example
demonstrates how to perform an Rds(on) sweep up to 100 A using two Model 2651A High Power
System SourceMeter
®
Instruments.
Combining SMUs to increase current sourcing capability
CAUTION
Carefully consider and configure the appropriate output-off state, source, and compliance levels
before connecting the Model 2651A High Power System SourceMeter® Instrument to a device that
can deliver energy (for example, other voltage sources, batteries, capacitors, solar cells, or other
Series 2600A System SourceMeter® instruments). Program recommended instrument settings before
making connections to the device. Failure to consider the output-off state, source, and compliance
levels may result in damage to the instrument or to the device under test (DUT).
To combine two Model 2651A High Power System SourceMeter® Instruments to output a pulse that is
greater than that of a single instrument, the current sources are connected in parallel. This delivers
the sum of the currents to the DUT, which allows a sweep of the DUT up to 100 A. The following
figure illustrates this principle of combining current sources.
Figure 23: Sum of parallel current sources
®
I1+I
2
I
1
I
2
DUT
Because each Model 2651A can produce up to 50 A, placing the instruments in parallel can deliver up
to 100 A. In this example, we will sweep the DUT using each Model 2651A to source exactly half of
the desired total current level for each point in the sweep.
7-2 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 7: Increasing SMU current sourcing ability
Equipment needed for this example
To run this test, you will need the following equipment:
• Two Model 2651A High Power System SourceMeter
• One Series 2600A System SourceMeter
SourceMeter
2611A/2612A, or 2635A/2636A.
•Two TSP-Link
®
Instrument refers to any of the following instrument models: Models 2601A/2602A,
®
cables (Keithley Instruments part number CA-180-3A)
• One GPIB cable or Ethernet cable to connect one Model 2651A to a computer
• Additional cable and connector assemblies as required to make connections to the device under
test (DUT) (for example, Model 2651A-KIT and Model 2600-KIT); see Device connections
page 7-4) for a schematic of required connections
The Model 2651A High Power System SourceMeter® Instruments must be at the same firmware
revision.
Set up communication
The communication setup is illustrated in the following diagram. GPIB is used as an example, but this
application can be run using any of the supported communication interfaces for the instruments. A
TSP-Link
Model 2651A SMU #2 (node #2) and the Series 2600A SMU #3 (node #3) are sent over the TSP-Link
interface.
®
connection enables communication between the two instruments. Commands for the
®
Instruments
®
Instrument; for this application, Series 2600A System
(on
NOTE
Figure 24: GPIB communication example for Rds(on) sweep
GPIB
Model 2651A SMU #1
(TSP-Link Node #1)
Controller
Model 2651A SMU #2
TSP-Link
* Model 260xA, Model 261xA,
2651A-900-01 Rev. A / March 2011 7-3
or Model 263xA
(TSP-Link Node #2)
Series 26xxA* SMU #3
(TSP-Link Node #3)
Section 7: Increasing SMU current sourcing ability Model 2651A High Power System SourceMeter® Instrument User's Manual
To set the TSP-Link node number using the front-panel interface:
1. Press the MENU key.
2. Select TSPLink.
3. Select NODE.
4. Use the navigation wheel
5. Press the ENTER key to save the TSP-Link node number.
On the Model 2651A SMU #1 (TSP-Link node #1), perform a TSP-Link reset to update it with the
linked instruments:
1. Press the MENU key.
2. Select TSPLink.
3. Select RESET.
If error 1205 is generated during the TSP-link reset, ensure that the Series 2600A has a unique
TSP-Link node number.
You can also perform a TSP-Link reset from the remote command interface by sending
tsplink.reset() to the Model 2651A. This method is used in the example program.
Device connections
to adjust the node number.
NOTE
Quick Tip
To run the test, make the connections as illustrated in the following figure. To ensure good contact,
use care when making connections.
CAUTION
Carefully consider and configure the appropriate output-off state, source, and compliance levels
before connecting the Model 2651A High Power System SourceMeter® Instrument to a device that
can deliver energy (for example, other voltage sources, batteries, capacitors, solar cells, or other
Series 2600A System SourceMeter® instruments). Program recommended instrument settings before
making connections to the device. Failure to consider the output-off state, source, and compliance
levels may result in damage to the instrument or to the device under test (DUT).
7-4 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 7: Increasing SMU current sourcing ability
NOTE
For best results, all connections should be left floating (no connections should be tied to ground).
Figure 25: Connections for Rds(on) sweep
Model 2651A
(master node #1)
Model 2651A
(subordinate node #2)
SENSE/GUARD
S
GGGG
LO
Series 26xxA*
(subordinate node #3)
LO G HI G G G
S
LO
S
HI
Gate resistor
(if required)
S
HI
* Model 260xA, Model 261xA,
or Model 263xA
OUTPUT
LO HI
40V, 50A MAX.
LO
S HIS HI
D
G
S
S LO
LO
HIHI
N-Channel
power
MOSFET
S LO
LO
S LO
LO
SENSE/GUARD
S
GGGG
LO
OUTPUT
S
HI
LO HI
40V, 50A MAX.
Equivalent
HI
S HI
SMU
(subordinate
node #3)
S LO
LO
Gate resistor
(if required)
D
HI
G
S
S HI
(master,
node #1)
S LO
LO
SMU
HI
S HI
SMU
(subordinate,
node #2)
S LO
LO
NOTE
During high-current pulsing through the channel of the device, the gate may begin to oscillate. This
oscillation creates an unstable voltage on the gate, which may cause unstable measurements through
the device channel. You can install a resistor to dampen this oscillation and stabilize the gate. Install
the resistor between the gate of the device and the force and sense HI of the Model 2651A. If the
gate still remains unstable, enable High-C mode on the Model 2651A source-measure unit (SMU).
2651A-900-01 Rev. A / March 2011 7-5
Section 7: Increasing SMU current sourcing ability Model 2651A High Power System SourceMeter® Instrument User's Manual
Configuring the trigger model
When the trigger model is correctly configured, you can achieve pulsed measurements with precise
timing and current levels up to 100 A for the combined Model 2651A instruments. Refer to the
Example program code
In this trigger model, the master Model 2651A node[1] is configured to control the overall timing of
the sweep. Subordinate Model 2651A node[2] is configured to wait for signals from the master
Model 2651A node[1] that prompt it to pulse.
In this example, the Series 2600A is controlled by script, so its trigger model is not used.
(on page 7-9) for an implementation of this trigger model.
The following figure illustrates the overall trigger model configuration of this test, which synchronizes
the two Model 2651A High Power System SourceMeter
®
Instruments to within 1 µs to provide a single
100 A pulse.
7-6 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 7: Increasing SMU current sourcing ability
Figure 26: Trigger model configuration for Rds(on) sweep
Model 2651A (master)
node[1].
trigger.timer[1].
stimulus
passthrough = true
trigger.timer[2].
stimulus
passthrough = false
tsplink.trigger[1].
mode = tsplink.TRIG_FALLING
50e-3 s
500e-6 s
Model 2651A (subordinate)
node[2].
tsplink.trigger[1].
mode = tsplink.TRIG_FALLING
EVENT_ID
count = 99
EVENT_ID
count = 1
EVENT_ID
stimulus
EVENT_ID
stimulus
smua.trigger.
Idle
Arm layer
arm.stimulus
Trigger layer
source.stimulus
measure.stimulus
endpulse.stimulus
smua.trigger.
Idle
Arm layer
arm.stimulus
Trigger layer
IDLE_EVENT_ID
SWEEPING_EVENT_ID
arm.count = 1
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
count = 100
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
IDLE_EVENT_ID
SWEEPING_EVENT_ID
arm.count = 1
SWEEP_COMPLETE_EVENT_ID
ARMED_EVENT_ID
count = 100
source.stimulus
trigger.timer[1].
stimulus
passthrough = false
3e-5 s
EVENT_ID
count = 1
measure.stimulus
endpulse.stimulus
SOURCE_COMPLETE_EVENT_ID
MEASURE_COMPLETE_EVENT_ID
PULSE_COMPLETE_EVENT_ID
2651A-900-01 Rev. A / March 2011 7-7
Section 7: Increasing SMU current sourcing ability Model 2651A High Power System SourceMeter® Instrument User's Manual
Master Model 2651A node[1] trigger model operation
In the master Model 2651A (node[1]) trigger model, timer 1 controls the period of the pulse, while
timer 2 controls the pulse width. TSP-Link
®
trigger 1 prompts Model 2651A (subordinate, node[2]) to
output its pulse.
When the trigger model of the master Model 2651A (node[1]) is initialized, the following occurs:
1. The source-measure unit (SMU) trigger model leaves the idle state, flows through the Arm Layer,
enters the Trigger Layer, outputs the ARMED event trigger, and then reaches the Source Event,
where it waits for an event trigger.
2. The ARMED event trigger is received by timer 1, which begins its countdown and passes the
trigger through to be received by TSP-Link trigger 1, timer 2, and the SMU Source Event
Detector.
3. TSP-Link trigger 1 receives the event trigger from timer 1 and sends a trigger through the
TSP-Link interface to Model 2651A (subordinate, node[2]) to instruct it to output the pulse.
4. The SMU Source Event Detector receives the event trigger from timer 1, begins to output the
pulse, waits the programmed source delay (if any), outputs the SOURCE_COMPLETE event to
timer 2, and then allows the SMU trigger model to continue.
5. Timer 2 receives the SOURCE_COMPLETE event trigger from timer 1 and begins to count down.
6. The SMU trigger model continues to the Measure Event, where it waits a programmed measure
delay (if any), takes a measurement, and continues until it encounters the End Pulse Event,
where it waits for an event trigger.
7. The timer 2 countdown expires and timer 2 outputs an event trigger to the SMU End Pulse Event.
8. The SMU End Pulse Event Detector receives the event trigger from timer 2, outputs the falling
edge of the pulse, and allows the SMU trigger model to continue.
9. The SMU trigger model compares the current trigger layer loop iteration with the trigger count.
•If the current iteration is less than the trigger count, the trigger layer repeats and the SMU trigger model
reaches Source Event, where it waits for another trigger from timer 1. Because timer 1 had its count set
to one less than the trigger count, timer 1 will continue to output a trigger for each iteration of the Trigger
Layer loop. The trigger model then repeats from step 3.
•If the current iteration is equal to the trigger count, the SMU trigger model exits the Trigger Layer,
passes through the Arm Layer, and returns to an idle state.
Model 2651A node[2] trigger model operation
In the Model 2651A instrument (subordinate, node[2]) trigger model, timer 1 controls the pulse width
and is programmed with the same delay as timer 2 in the Model 2651A (master, node[1]).
®
The pulse period is controlled by TSP-Link
instrument (master, node[1]) timer 1, so the same timer in the system controls the pulse period.
trigger 1, which gets its triggers from the Model 2651A
7-8 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 7: Increasing SMU current sourcing ability
When the trigger model of Model 2651A (subordinate, node[2]) is initialized, the following occurs:
1. The source-measure unit (SMU) trigger model leaves the idle state, flows through the arm layer,
enters the trigger layer, and reaches the Source Event, where it waits for an event trigger.
2. TSP-Link trigger 1 receives a trigger from the TSP-Link interface and outputs an event trigger to
the SMU Source Event.
3. The SMU Source Event Detector receives the event trigger from TSP-Link trigger 1, begins to
output the pulse, waits the programmed source delay (if any), and outputs the
SOURCE_COMPLETE event to timer 1 before allowing the SMU trigger model to continue.
4. Timer 1 receives the SOURCE_COMPLETE event trigger from TSP-Link trigger 1 and begins its
countdown.
5. The SMU trigger model continues to the Measure Event, where it waits for a programmed
measure delay (if any), takes a measurement, and then continues until it stops at the End Pulse
Event, where it waits for an event trigger.
6. The timer 1 countdown expires and timer 1 outputs an event trigger to the SMU End Pulse Event.
7. The SMU's End Pulse Event Detector receives the event trigger from timer 1, outputs the falling
edge of the pulse, and then allows the SMU trigger model to continue.
8. The SMU trigger model compares the current trigger layer loop iteration with the trigger count.
•If the current iteration is less than the trigger count, the trigger layer repeats and the SMU trigger model
reaches Source Event, where it waits for another trigger from TSP-Link trigger 1. The trigger model then
repeats from step 2.
•If the current iteration is equal to the trigger count, the SMU trigger model exits the trigger layer, passes
through the arm layer, and returns to an idle state.
Series 2600A trigger model operation
In this example, the Series 2600A is controlled by script, so its trigger model is not used.
Example program code
The example code is designed to be run from Test Script Builder or TSB Embedded. It can also be
run from other programming environments, such as Microsoft® Visual Studio® or National Instruments
LabVIEWTM. However, modification of the example code may be required to do so.
NOTE
2651A-900-01 Rev. A / March 2011 7-9
Section 7: Increasing SMU current sourcing ability Model 2651A High Power System SourceMeter® Instrument User's Manual
The following script contains all the code necessary to perform an Rds(on) sweep up to 100 A using
two Model 2651A High Power System SourceMeter
SourceMeter
•Initializes the TSP-Link
®
Instrument. This script:
®
connection
®
Instruments and a Series 2600A System
• Configures the source-measure units (SMUs)
• Configures the trigger models of the two Model 2651A instruments
• Prepares the reading buffers
• Executes the test
• Stores the collected data in a format that can be copied from the instrument console and pasted
into a Microsoft
The script is written using Test Script Processor (TSP®) functions rather than as a single block of
inline code. TSP functions are similar to functions in other programming languages, such as
Microsoft
®
Visual C® or Visual Basic®, and must be called before the code contained in them is
executed. Because of this, running the script alone will not execute the test. To execute the test, run
the script to load the functions into test script memory and then call the functions. Refer to the
documentation for Test Script Builder or TSB Embedded for directions on how to run scripts and enter
commands using the instrument console.
The script contains comments that describe what is being performed by the lines of code and
document the functions in the script. Lines starting with node[2]. are commands that are being sent
to Model 2651A on node 2 through the TSP-Link interface. Lines starting with node[3]. are
commands that are being sent to the Series 2600A through the TSP-Link interface. All other
commands are executed on the master Model 2651A on node [1].
--[[
Title: Combining SMUs for 100A Example
Description: This script is designed to perform an Rds(on)sweep on a power
MOSFET device. It combines two 2651A SMUs in parallel to perform a current
sweep up to 100A. Data collected from the sweep is then returned in a
Microsoft Excel compatible format for plotting and analysis.
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 7: Increasing SMU current sourcing ability
--[[
Name: DualSmuRdson(gateLevel, dstart, dstop, dsteps, pulseWidth,
pulsePeriod, pulseLimit)
Description: This function uses two 2651A SMUs to perform a pulsed Rds(on)
sweep with currents up to 100A.
Parameters:
gateLevel: The gate level to be used during the sweep
dstart: The starting current level of the drain sweep
dstop: The ending current level of the drain sweep
dsteps: The number of steps in the drain sweep
pulseWidth: The width of the drain pulse in seconds
pulsePeriod: The time from the start of one drain pulse to
the next in seconds
pulseLimit: The voltage limit of the drain pulse in volts
Note: Actual pulse limit will be 10% lower than setting
to protect SMUs in a compliance condition
Example Usage:
DualSmuRdson(10, 1, 100, 100, 500e-6, 50e-3, 10)
]]
function DualSmuRdson(gateLevel, dstart, dstop, dsteps, pulseWidth, pulsePeriod,
pulseLimit)
tsplink.reset(3) -- Verify that at least three nodes are present
reset()
-- Timer 1 controls the pulse period by triggering the pulse to begin
trigger.timer[1].count = dsteps - 1
trigger.timer[1].delay = pulsePeriod
trigger.timer[1].passthrough = true
trigger.timer[1].stimulus = smua.trigger.ARMED_EVENT_ID
trigger.timer[1].clear()
-- Each unit will source half the current, so divide the start
-- and stop values by 2
smua.trigger.source.lineari(dstart / 2, dstop / 2, dsteps)
smua.trigger.source.limitv = pulseLimit - (pulseLimit * 0.1)
-- Again, keep the limit SMU #1 lower than the limit of SMU #2
-- to prevent parallel V-sources
smua.trigger.measure.iv(smua.nvbuffer1, smua.nvbuffer2)
smua.trigger.measure.action = smua.ENABLE
-- Return to the bias level at the end of the pulse/sweep
smua.trigger.endpulse.action = smua.SOURCE_IDLE
smua.trigger.endsweep.action = smua.SOURCE_IDLE
-- SMU will be a 0A current source with 10V limit when output is turned off
node[2].smua.source.rangei =
math.max(math.abs(dstart / 2), math.abs(dstop / 2))
node[2].smua.source.leveli = 0 -- Sets the DC bias level
node[2].smua.source.limitv = 10 -- Sets the DC bias limit
-- Return the output to the bias level at the end of the pulse/sweep
node[2].smua.trigger.endpulse.action = node[2].smua.SOURCE_IDLE
node[2].smua.trigger.endsweep.action = node[2].smua.SOURCE_IDLE
node[2].smua.trigger.count = dsteps
node[2].smua.trigger.arm.stimulus = 0
node[2].smua.trigger.source.stimulus = node[2].tsplink.trigger[1].EVENT_ID
node[2].smua.trigger.measure.stimulus = 0
node[2].smua.trigger.endpulse.stimulus = node[2].trigger.timer[1].EVENT_ID
node[2].smua.trigger.source.action = node[2].smua.ENABLE
2651A-900-01 Rev. A / March 2011 7-13
Section 7: Increasing SMU current sourcing ability Model 2651A High Power System SourceMeter® Instrument User's Manual
-- If you find your gate oscillating even with a dampening resistor
-- in place, try enabling high-C mode to help stabilize the gate.
-- Prepare the reading buffers
node[3].smua.nvbuffer1.clear()
node[3].smua.nvbuffer1.collectsourcevalues = 1
if node[3].smua.nvbuffer1.fillmode ~= nil then
node[3].smua.nvbuffer1.fillmode = node[3].smua.FILL_ONCE
end
node[3].smua.nvbuffer2.clear()
node[3].smua.nvbuffer2.collectsourcevalues = 1
if node[3].smua.nvbuffer2.fillmode ~= nil then
node[3].smua.nvbuffer2.fillmode = node[3].smua.FILL_ONCE
end
Model 2651A High Power System SourceMeter® Instrument User's Manual Section 7: Increasing SMU current sourcing ability
--[[
Function: PrintDualSmuRdsonData()
Description:
This function processes the data stored in the SMU reading buffers by
function DualSmuRdson() and prints back the individual SMU data and the
combined SMU data and Rds(on) readings in a format that is copy and paste
compatible with Microsoft Excel.
]]
function PrintDualSmuRdsonData()
-- Combined Voltage = Average(SMU1 Voltage reading, SMU2 Voltage reading)
combinedVoltage = (smua.nvbuffer2[i] + node[2].smua.nvbuffer2[i]) / 2
-- Combined Current = SMU1 Current reading + SMU2 Current reading
combinedCurrent = smua.nvbuffer1[i] + node[2].smua.nvbuffer1[i]
-- Rds(on) = Combined Voltage / Combined Current
rdson = combinedVoltage / combinedCurrent
-- Print a row of data
print(string.format("%g\t%g\t%g\t%g\t%g\t%g\t%g\t%g\t%g",
smua.nvbuffer1.timestamps[i],
sourceLevel,
smua.nvbuffer2[i],
smua.nvbuffer1[i],
node[2].smua.nvbuffer2[i],
node[2].smua.nvbuffer1[i],
combinedVoltage,
combinedCurrent,
rdson))
end
end
2651A-900-01 Rev. A / March 2011 7-15
Section 7: Increasing SMU current sourcing ability Model 2651A High Power System SourceMeter® Instrument User's Manual
Example program usage
The functions in this script allow the sweep parameters of the test to be adjusted without having to
rewrite and rerun the script. A test can be executed by calling the function
DualSmuRdson(gateLevel, dstart, dstop, dsteps, pulseWidth, pulsePeriod,
pulseLimit) with the appropriate values passed to it in its parameters.
DualSmuRdson() parameters
Parameter Units Description
gateLevel
dstart
dstop
dsteps
pulseWidth
pulsePeriod
pulseLimit
An example call to this function is as follows:
DualSmuRdson(10, 1, 100, 100, 500e-6, 50e-3, 10)
Volts The voltage level to hold the gate
Amperes The level of the first step in the drain sweep
Amperes The level of the last step in the drain sweep
Not
applicable
Seconds The width of the pulse in the drain sweep
Seconds The time between the start of consecutive pulses in the drain sweep
Volts The voltage limit of the pulses in the drain sweep
The number of steps in the drain sweep
This call will set the gate source-measure unit (SMU) output to 10 V and then sweep the drain of the
device from 1 A to 100 A in 100 points. The points of this sweep will be gathered using pulsed
measurements with a pulse width of 500 µs and a pulse period of 50 ms for a 1 percent duty cycle.
These pulses will be limited to a maximum of 10 V. At the end of the sweep, all SMU outputs will be
turned off. The resulting data from this test will be returned in a Microsoft
®
Excel® compatible format
(you can cut and paste the output from the console) that can be used for graphing and analysis.
7-16 2651A-900-01 Rev. A / March 2011
In this section:
About this section..................................................................... 8-1
Need different line frequency or voltage................................... 8-1
Can I use external trigger lines?............................................... 8-1
The internal web page of the instrument is not accessible ....... 8-1
Save the present state of the instrument.................................. 8-2
LabVIEW driver for the Model 2651A ....................................... 8-2
About this section
The intent of this section is to troubleshoot the most common problems encountered with the Model
2651A. For additional troubleshooting information, please see the "Troubleshooting" section in the
Model 2651A Reference Manual.
Section 8
Troubleshooting FAQs
Need different line frequency or voltage
I need to use a Model 2651A at a different line frequency or voltage. What do I need to do?
The Model 2651A supports a line frequency of either 50 Hz or 60 Hz. The factory configures the
Model 2651A to automatically detect and operate at line frequencies (either 50 Hz or 60 Hz) each
time the instrument power is turned on. In noisy environments, it may be necessary to manually
configure the instrument to match the actual line frequency. See the Model 2651A Reference Manual,
"Line frequency configuration."
Can I use external trigger lines?
Yes, these signals appear on the digital I/O connector. For information about the digital I/O
capabilities, refer to "Controlling digital I/O lines" in the Model 2651A Reference Manual.
The internal web page of the instrument is not accessible
I cannot access the internal web page of the instrument. How do I fix this?
Ensure that the instrument has been assigned a valid and unique IP address.
Section 8: Troubleshooting FAQs Model 2651A High Power System SourceMeter® Instrument User's Manual
•To view the IP address of the Model 2651A, press MENU. Select LAN > STATUS >
IP-ADDRESS. If a valid IP address is displayed, the instrument configuration is correct. If 0.0.0.0
is displayed, the IP address is not valid.
•If you are using the automatic configuration method over a corporate network, contact your
network administrator for assistance.
•If you are using the automatic configuration method and you have a one-to-one connection
between the instrument and the computer, verify that the computer is set to receive an IP address
automatically.
•If you are using the manual configuration method over a corporate network, verify that you have
entered the IP address, gateway, and subnet mask provided by your network administrator.
•If you are using the manual configuration method and a one-to-one connection with the computer,
verify that the subnet mask agrees with the computer's settings.
•If the above suggestions do not work, see the Model 2651A Reference Manual, "Communication
interfaces."
Save the present state of the instrument
How do I save the present state of the instrument?
You can save the present state of the instrument using the front-panel menus or from a command
interface. See the Model 2651A Reference Manual, "User setups."
LabVIEW driver for the Model 2651A
Where do I find the LabVIEW driver for the Model 2651A?
8-2 2651A-900-01 Rev. A / March 2011
In this section:
Additional Model 2651A information......................................... 9-1
Additional Model 2651A information
This manual has prepared you to start using your new Model 2651A for your real-world applications.
For more detailed information about the Model 2651A, refer to the Keithley Instruments Model 2651A
Reference Manual, part number 2651A-901-01.
For additional information on the Model 2651A, refer to:
•The Product Information CD-ROM (ships with the product): Contains software tools, drivers, and
product documentation, including documentation for switch cards that are compatible with the
Model 2651A
Section 9
Next steps
•
•The Knowledge Center, which contains the following handbooks:
•The Low Level Measurements Handbook: Precision DC Current, Voltage, and Resistance
Measurements
•Switching Handbook: A Guide to Signal Switching in Automated Test Systems
•Application notes
•Updated drivers
•Information on related products, including:
•The Model 4200-SCS Semiconductor Characterization System
®
•The Series 2600A System SourceMeter
Instruments
•Your local Field Applications Engineer can help you with product selection, configuration, and
usage. Check the website for contact information.
Index
A
Action keys
RECall • 2-2
STORE • 2-2
C
Communication
Set up • 5-1, 6-2, 7-3
Compliance
Setting front panel compliance limit • 4-2
Connections
Device • 4-1, 5-2, 6-4, 7-4
Current
Increase sourcing capability • 7-1, 7-2
Curves
Example test script • 6-10, 6-16
I-V curves • 6-1, 6-4, 6-5, 6-6, 6-7, 6-9
IV_Curves() parameters • 6-16
External trigger lines • 8-1
IP address • 3-1, 3-2, 8-1
LabVIEW driver • 8-2
Line frequency • 8-1
Save present state • 8-2
Voltage • 8-1
Web page • 3-1, 3-2, 8-1
TSB Embedded • 3-3
Script, create with • 3-3
TSP Express • 3-8
Index-2 2651A-900-01 Rev. A / March 2011
Model 2651A High Power System SourceMeter® Instrument User's Manual Index
TSP-Link
Nodes • 6-2
V
Voltage
Source, set level • 4-2
W
Warranty • 1-1
Web interface • 3-1
Accessing • 3-1
Browser requirements • 3-1
Home page • 3-2
IP address configuration • 3-2
Reading buffers • 3-7
TSB Embedded • 3-3
TSP Express • 3-8
Using • 3-1
2651A-900-01 Rev. A / March 2011 Index-3
12/06
Specifications are subject to change without notice.
All Keithley trademarks and trade names are the property of Keithley Instruments, Inc.
All other trademarks and trade names are the property of their respective companies.
A GREATER MEASURE OF CONFIDENCE