Any unauthorized reproduction, photocopy, or use the information herein, in whole or in part,
without the prior written approval of Keithley Instruments, Inc. is strictly prohibited.
All Keithley Instruments product names are trademarks or registered trademarks of Keithley
Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective
holders.
Document number: 2200S-901-01 Rev. A / November 2011
11/07
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using
the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained
service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I and
Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most
measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or
to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient
over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are
for connection to Category I sources unless otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any
unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If
the circuit is capable of operating at or above 1000V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective
devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth)
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the
voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions, or the safety of
the equipment may be impaired.
11/07
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating
information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions located in the
user documentation in all cases where the symbol is marked on the instrument.
The symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal
contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be
properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national
safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be
purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be
purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the
applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled
according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the
factory for proper cleaning/servicing.
Thank you for using a Keithley Instruments product. The Series 2200 Programmable DC Power
Supplies provide a wide range of voltage outputs to address the testing and characterization of
components, circuits, modules, and complete devices whether you are in a research laboratory, in
design and development, or in production test. The Series 2200 consists of five models with output
voltages from 20 V to 72 V that can deliver 86W, 96W, 100W, and 150W of power. In addition, these
power supplies can act as constant current sources as well as constant voltage sources. The Series
2200 power supplies offer an excellent combination of performance, versatility, and ease of use that
allow you to obtain quality test data as quickly as possible. They perform as effectively in automated
test systems as they do in manual instrument configurations.
Extended warranty
Additional years of warranty coverage are available on many products. These valuable contracts
protect you from unbudgeted service expenses and provide additional years of protection at a fraction
of the price of a repair. Extended warranties are available on new and existing products. Contact your
local Keithley Instruments representative for details.
Contact information
If you have any questions after reviewing this information, please use the following sources:
1.
2. Keithley web forum
3. Call Keithley Instruments corporate headquarters (toll-free inside the U.S. and Canada only) at
Section 1: Introduction Series 2200 Programmable DC Power Supplies Reference Manual
1-2 2200S-901-01 Rev. A/November 2011
CD-ROM contents
Model
Description
2200-20-5
Programmable DC power supply. 20 V, 5 A, 1 channel, USB, GPIB
2200-30-5
Programmable DC power supply. 30 V, 5 A, 1 channel, USB, GPIB
2200-32-3
Programmable DC power supply. 32 V, 3 A, 1 channel, USB, GPIB
2200-60-2
Programmable DC power supply. 60 V, 2.5 A, 1 channel, USB, GPIB
2200-72-1
Programmable DC power supply. 72 V, 1.2 A, 1 channel, USB, GPIB
Accessory
Keithley part number
Rear panel mating connector
CS-1638-12
Documentation and Driver CD
2200S-950-01
Certification of calibration
A CD-ROM is shipped with each Series 2200 order.
The Series 2200 Product Information CD-ROM contains:
Reference Manual: Includes advanced operation topics and maintenance information. Programmers
looking for a command reference, and users looking for an in-depth description of the way the
instrument works (including troubleshooting and optimization), should refer to the Reference Manual.
Drivers and release notes: IVI Instrument Driver, driver for National Instruments LabVIEW™, and
related release notes.
Additional product information: The product specifications are also included on the CD-ROM.
Key features
The Series 2200 offers:
Single-output, DC power
Linear regulation
0.03 percent basic voltage accuracy
0.05 percent basic current accuracy
Less than 1 mVrms ripple and noise
40 user-defined setup memories
USB 2.0 device port on rear panel for PC connectivity and remote programming
IEEE-488.2 General Purpose Interface Bus (GPIB) interface for PC connectivity and remote
programming
Standard accessories
Series 2200 Programmable DC Power Supplies Reference Manual Section 1: Introduction
1-year factory warranty extended to 3 years from date of shipment
2200-3Y-EW-STD
3 calibrations within 3 years of purchase
C/2200-3Y-STD
3 (ANSI-Z540-1 compliant) calibrations within 3 years of purchase
C/2200-3Y-DATA
Services available
Specifications
This topic provides information about:
Electrical ratings for the power connection
Environmental performance
Nominal Voltage Ratings
Float Voltage Rating
For additional specifications, refer to Series 2200 Programmable DC Power Supplies Specifications
(document number SPEC-2200S) on the Product Information CD-ROM that came with your Series
2200.
Section 1: Introduction Series 2200 Programmable DC Power Supplies Reference Manual
1-4 2200S-901-01 Rev. A/November 2011
Electrical ratings for the power connection
Model
Line selector
switch setting
Frequency
Fuse rating
Max power
2200-20-5
110 V / 220 V
50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
350 VA
2200-30-5
110 V / 220 V
50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
500 VA
2200-32-3
110 V / 220 V
50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
350 VA
2200-60-2
110 V / 220 V
50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
500 VA
2200-72-1
110 V / 220 V
50 / 60 Hz
For 110 V: 5 A TH, 250 V
For 220 V: 2.5 A TH, 250 V
350 VA
Parameter
2200-20-5
2200-30-5
2200-32-3
2200-60-2
2200-72-1
Temperature
Temperature Operating: +0 °C to +40 °C
Humidity
Operating:
5 percent to 95 percent relative humidity ( percent RH) at up to 40 °C, non-
condensing
Pollution degree 2, For indoor use only.
Altitude
Operating: 100 percent capability up to 2,000 meters.
Environmental performance
Nominal voltage ratings
There are two ranges for all models, selectable by the line-voltage selector switch. Check the voltage
select switch on the bottom of the power supply before turning it on:
110 V setting, 110 / 115 / 120 VAC
220 V setting, 220 / 230 / 240 VAC
Fluctuations must not exceed 10 percent of the nominal rated voltage.
Float voltage rating
Float voltage rating: up to 100 V (DC + peak AC) between earth ground and any output terminal.
Clear and sanitize procedures ................................................ 2-34
Operating requirements
Section 2
General operation
This section provides information about operating requirements for the Series 2200 instruments.
Observe all safety precautions listed in Safety Precautions (on page 1-1) and in this topic before
using the Series 2200 instruments and any associated instruments.
Read and follow all installation, operation, and maintenance information carefully before using the
Series 2200 instruments.
Refer to the following sources for complete product specifications:
Specifications (on page 1-3)
Series 2200 Programmable DC Power Supplies Specifications (document number SPEC-2200S)
on the Product Information CD-ROM that came with your Series 2200.
1. Place the bottom of the instrument on a bench or similar surface. Do not stand the instrument on
it's rear panel. Refer to the following figure for an illustration of how to position the instrument.
Do not stand the instrument on its rear panel. This position is unstable and the instrument
might tip over and be damaged or fall on someone and cause injury. To avoid damage or
bodily injury, place the bottom of the instrument flat on a stable surface.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual
2-2 2200S-901-01 Rev. A/November 2011
Figure 1: Place bottom of instrument on flat stable surface
2. Before operating, ensure that the ambient temperature is between +0 °C and +40 °C (+32 °F to
+104 °F).
To ensure proper cooling, keep the front, sides, and rear of the instrument clear of
obstructions
Use properly rated load wires. All load wires must be heavy enough not to overheat when
carrying the maximum short-circuit output current of the power supply. If there is more than
one load, then any pair of load wires must be capable of safely carrying the full-rated shortcircuit output current of the power supply.
To reduce risk of fire and shock, ensure the mains supply voltage fluctuations do not exceed
10% of the operating voltage range.
Do not loosen any screw on this product, except those in the rear connector, which are intended to
retain attached external wires to the connector. There are no user serviceable components inside.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-3
Installing the system
This section contains information on how to install your Model 2200 power supply.
Unpack the instrument and check that you received all the items listed as standard accessories.
Check that you also received any other accessories that you ordered with your instrument.
To verify that the power supply is ready to use, follow the procedures to:
Power the instrument on and off (on page 2-3)
Check the output (on page 2-3)
Power the instrument on and off
To power on the instrument follow these steps:
1. Make all the connections.
2. Properly set the 110 V / 220 V selector switch located on the bottom of the instrument.
Figure 2: Selector switch
3. Connect the power cord that was provided with the instrument to the power connector on the rear
panel. Connect the power cord plug to a properly grounded electrical outlet.
4. Press the power switch on the front panel.
To power off the instrument, press the power switch on the front panel.
To satisfy safety requirements, always use load wires heavy enough not to overheat when
carrying the maximum short-circuit output current of the power supply. If there is more than
one load, then every pair of load wires must be capable of safely carrying the full-rated current
of the power supply.
Check the output
The following procedures check that the power supply develops its rated outputs and properly
responds to operation from the front panel.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual
2-4 2200S-901-01 Rev. A/November 2011
Voltage output check
To verify basic voltage functions without a load, follow these steps.
1. Remove all leads from the output connectors.
2. Turn on the power supply.
3. Press Shift and Menu (1). >Default Set shows on the display.
4. Press Enter to bring up the default settings options. No and Yes appear on the display.
5. Press the right arrow key to select Yes. Press Enter to enable the default settings.
6. Press the front-panel On/Off key to enable the output. The OFF message on the display turns off
and the CV message on the display turns on. The upper line of the display shows the actual
output voltage and current. The lower line shows the settings.
7. Check that the front-panel voltmeter properly responds to pressing keys on the numeric keypad.
Press V-set, use the numeric keypad to set the voltage value to 0 and press Enter. Check if the
displayed voltage value is approximately 0 V and check if the displayed current value is
approximately 0 A. You can confirm the 0 V setting with a voltmeter.
8. Press V-set and use the numeric keys and Enter to set the voltage value to the maximum
allowable for your power supply, as indicated on the instrument’s front panel.
9. Check if the displayed voltage value is approximately the same value as the voltage setting.
Current output check
To verify the basic current functions with a short across the power supply output, follow these steps:
1. Remove all leads from the output connectors.
2. Turn on the power supply.
3. Press Shift and Menu (1). >Default Set shows on the display.
4. Press Enter to bring up the default settings options. No and Yes appear on the display.
5. Press the right arrow key to select Yes. Press Enter to enable the default settings.
6. Ensure that the output is disabled and the display OFF message is turned on. If needed, press the
On/Off key to ensure that the output is disabled and the OFF message is turned on.
7. Use an insulated test lead to connect a short across the (+) and (-) output terminals.
Use a wire size sufficient to handle the maximum current. You should use at least a 22 gauge
wire.
To satisfy safety requirements, always use load wires heavy enough not to overheat when
carrying the maximum short-circuit output current of the power supply. If there is more than
one load, then every pair of load wires must be capable of safely carrying the full-rated current
of the power supply.
8. Press the On/Off key to enable the output. The CC message should be lit.
9. Press I-set and use the numeric keypad and Enter key to set the current value to 0 A. Check if
the displayed current value is approximately 0 A.
10. Press I-set and use the numeric keys and Enter to set the current value to the maximum
allowable for your current supply. Check if the displayed current value is approximately the same
value as the maximum allowable.
11. Turn off the power supply and remove the short wire from the (+) and (-) output terminals.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-5
Operating basics
Front panel at a glance
The front panel of the Keithley Instruments Series 2200 contains the following items:
Power switch
Display
Arrow keys
The navigation wheel
Output connectors
Keys
Number keys
You can use the keys, display, and the navigation wheel to access, view, and edit the
configuration menu items and set values.
For additional information about the front panel, see the Front panel operation (on page 2-15) section.
The front panel of the Series 2200 is shown in the following figure.
Figure 3: Front panel
(1) Display
The display shows information.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual
2-6 2200S-901-01 Rev. A/November 2011
(2) Information
During operation, the display provides information about the output (top row) and settings (bottom
row) for volts (left column) and current (right column). During configuration, the display shows the
configuration menus (bottom row).
(3) Arrow keys
(4) Navigation wheel
Up, down, right, and left arrow keys (▲ and ▼) and Enter key.
Rotate the navigation wheel to increase or decrease digits or to select menu items.
(5) Output connectors
Three banana jack output connectors provide high, low, and chassis ground connections.
(6) Save and recall keys
See Key descriptions (on page 2-7) for descriptions of these keys.
(7) Numeric keypad
The numeric keypad provides number keys (0 to 9 and Esc) for direct numeric entry.
(8) V-Set, I-Set, Shift, and Output On/Off keys
See Key descriptions (on page 2-7) for descriptions of these keys.
(9) Power switch
Press the power switch to turn the instrument on (|). Press it again to turn the instrument off (0).
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-7
Key descriptions
Keys
Description
V-set
Sets the voltage limit. See Setting the voltage limit (on page 2-17) for more
information.
I-set
Sets the current limit. See Setting the current limit (on page 2-17) for more information.
Save
Saves the present settings to a specified setup memory location (1 to 40). See Save
and recall setups (on page 2-18) for more information.
Recall
Recalls a saved setting from a specified setup memory location (1 to 40). See Save
and recall setups (on page 2-18) for more information.
Shift
Used to access secondary functions of the numeric keys. For example, when the Shift
key is used with the 1 key, the combination displays the configuration menu. See the
following table for more information on Shift + numeric keypad descriptions.
Output
On/Off
Turns on or off the output channel of the power supply. The regulation mode, constant
voltage (CV) or constant current (CC), is shown on the display when the output is
turned on.
Esc
Backs out of the selected function. If the instrument’s focus is in a menu, pressing Esc
will back up one menu level.
Shift + numeric keypad descriptions
Key
Numeric
keypad
Description
Shift
1
Brings up the configuration menus. Turn the navigation wheel or
press the up or down keys to scroll through the list of menus. Press
Enter to select which menu to use.
Shift
2
Brings up the list selection. Turn the navigation wheel or press the
up or down arrow keys to scroll through the various lists you can run.
Shift
3
Generates a manual trigger event.
Shift
7
Returns control from remote operation to local front panel control. This
operation generates a return to local message.
Key descriptions
The keys provide front panel control and configuration. See Front panel at a glance (on page 2-5) for
an illustration of each key location. The following tables describe how to use each key.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual
2-8 2200S-901-01 Rev. A/November 2011
Operation
Key press
Display the instrument firmware version.
Shift + 8
Display the internal temperature of the instrument.
Shift + . (decimal point) + 1
Perform a factory reset. All setups and lists are erased.
Shift + . (decimal point) + 2
Unlock the keypad.
If the instrument's keypad and front panel controls
are locked, you will need a password to unlock
them. If you forget the password, enter the
appropriate default code for the model instrument
you are using:
Model
Default code
2200-20-5
4205
2200-30-5
4305
2200-32-3
4323
2200-60-2
4602
2200-70-1
4721
Cycling the instrument power also unlocks the
keypad.
Keypad special operations
See the following table for the Series 2200 keypad special operations.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-9
Primary
menu
Secondary menu
First option
Additional
option(s)
Allows you to ...
>Default Set
No
Yes
Return the instrument to the factory
default configuration. See Initializing
to the default setup (on page 2-17) for
more information.
>Protect
>Reset Protect
No
Yes
Reset Protect menu selections to the
factory default configuration.
>Max Volt Set
Off
Turn maximum output voltage setting
off.
On
Max =
Set or edit maximum output voltage
for the instrument. See Setting the
maximum voltage (on page 2-18) for
more information.
>OVP Set
Off
Turn overvoltage protection (OVP)
off.
On
OVP =
Turn overvoltage protection on and
set or edit the voltage threshold. See
Setting the overvoltage protection (on
page 2-19) for more information.
>Out Time Set
Off
Turn out time mode off.
On
Time =
Turn out time mode on and set or edit
the time period. See Timer example
(on page 2-30) for an example of
setting a timer with Out Time Set.
>Key Lock
Password =
unlock
Indicates that no password is set to
lock all front panel controls except
power switch and output on/off key.
Enter up to 4 digits to set a password.
Lck
Indicates that a password has been
set.
>Edit
List
>Recall 1-7
Empty
Recall or edit lists saved in up to 7
locations. Empty indicates no saved
list. See Define a list of voltage and
current steps (on page 2-24) for more
information.
Continuous
Repeat =
Set instrument to continuous mode. If
set, the instrument executes the
entire sequence of steps as soon as it
receives the next trigger signal.
Repeat allows you to set or edit
number of times to repeat the list.
List Steps allows you to indicate the
number of steps in the list.
See List mode example (on page 2-
31) for a list example.
Step
Repeat =
Set instrument to step mode. If set,
Configuration menus description
Use the configuration menus to configure (setup) the instrument. Sometimes the configuration menus
are called the settings menus. See the following table for information on the primary menu selections,
secondary menu selections, options, and a brief description of each selection. The following table
indicates default settings in bold.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual
2-10 2200S-901-01 Rev. A/November 2011
Primary
menu
Secondary menu
First option
Additional
option(s)
Allows you to ...
the instrument will advance to each
step only after it receives a trigger
signal.
Repeat allows you to set or edit
number of steps.
List Steps allows you to indicate the
number of steps in the list.
>User
>Reset User
No
Yes
Reset User menu selections to the
factory default configuration.
>Output Recall
Off
Turn output recall mode off.
On Set power output state to that which
was in use before power was last
turned off. See Recall the power
supply ON/OFF output state at power
on (on page 2-20) for more
information.
>Save Last
On
Save last operating parameters in
use before turning instrument off.
Restores saved parameters when
instrument turned on. See Recall the
power supply operating parameters at
power on (on page 2-21) for more
information.
Off
Turns save last mode off.
>Key Beep
Off
Turns off key press notification
sound.
On
Turns on key press notification
sound. See Setting the key sound (on
page 2-21) for more information.
>Knob Lock
Off
Unlocks navigation wheel .
On
Locks navigation wheel . See Lock
the navigation wheel (on page 2-22)
for more information.
>System
>Reset System
No
Yes
Reset System menu selections to the
factory default configuration.
>Port Mode
Trigger
Set port mode input to trigger. You
must also set the trigger source (Trig
Source).
RI/Def
Set port mode input and output to
RI/Def. You must also set the remote
inhibit mode (RI mode).
Digital
Set port mode input and output to
digital. This setting is for
programmable interface commands
described in Commands listed in
alphabetical order (on page 4-2).
>Trig Source
Manual
Set rear panel port mode to manual
trigger.
External
Set rear panel port mode to external
trigger.
Bus
Set rear panel port mode to bus
trigger.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-11
Primary
menu
Secondary menu
First option
Additional
option(s)
Allows you to ...
Immediate
Set rear panel port mode to
immediate trigger.
>RI Mode
Off
Turn remote inhibit (RI) mode off.
Latching
Set remote inhibit mode to latching.
Latching mode turns the output
channel off when the instrument
detects a high-to-low transition (5 V to
0 V).
Live
Set remote inhibit mode to live. Live
mode the output channel off when the
RI input is low (0 V) and on when the
RI input is high (5 V)
>DFI Source
Off
Turn the discrete fault input (DFI)
source mode off.
Ques
Set the DFI source to question
(Ques).
Oper
Set the DFI source to operation
(Oper).
ESB
Set the DFI source to error status
byte (ESB).
RQS
Set the DFI source to request service
(RQS.
>Communication
USBTMC
Set the communications mode to
USBTMC.
GPIB
Address=22
Set the communications mode to
GPIB. Set or edit the GPIB primary
address. See To change GPIB
address settings (on page 3-2) for
more information.
RS232
9600,8,N,1
Set the communication mode to
RS232. Set or edit the RS232
address. USBTCM and GPIB are the
recommended communication
modes.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual
2-12 2200S-901-01 Rev. A/November 2011
Upper display messages
Operating conditions
Messages
Description
OFF
Power supply output is off.
CV
Instrument is operating in constant voltage (CV) mode.
CC
Instrument is operation in constant current (CC) mode.
RI
Instrument output has been turned off by the remote inhibit (RI) input. See
[SOURce:]DIGital:FUNCtion (on page 4-12) for more information.
Error conditions
Messages
Description
OVP
Instrument has sensed a voltage level above the threshold set for overvoltage control
and turned off the output. See Setting the overvoltage protection (on page 2-19) for
more information.
RSV
Instrument has sensed a reverse remote sense polarity condition and turned the output
off. Polarities of source and sense connections must be the same. See Using remote
sense (on page 2-23) for more information.
OTP
Instrument has sensed an excessive internal temperature and turned off the output.
Message
Description
OVP
If overvoltage protection (OVP) is set, the instrument displays the OVP message to
Indicates that OVP is set. Flashes when OVP has tripped. See Setting the overvoltage
protection (on page 2-19) for more information.
Lck
If the front-panel controls are locked, the instrument displays the Lck message. See
Configuration menus description (on page 2-9) for information on the >Protect >Key
Lock configuration. See Keypad special operations (on page 2-8) for information on
unlocking the front panel controls.
The instrument shows messages on the upper row of the display to indicate output operating and
error conditions. See Configuring the instrument for your application (on page 2-16) for general
information about these messages. See the following tables for the messages and a description of
what the messages indicate.
If the instrument displays the OTP message, the instrument should be shut down immediately .Do not
operate the instrument when temperatures are above the internal temperature trip level to avoid
damaging internal components.
Lower display messages
The instrument shows the following messages on the lower row of the display.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-13
Rear panel at a glance
The rear panel of the Series 2200 contains the following items:
Cooling vents
Factory test port
USB device port
GPIB IEEE-488 port
110 V/220 V power connector
12-pin connector; includes remote sense
Shorting clip. Use this for proper regulation when not using remote sense. This connector comes
installed. It shorts the + drive to the + sense and the - drive to the - sense. To use remote sense,
remove the clip.
Figure 4: Rear panel
The rear panel of the Series 2200 is shown in the following figure.
(1) Cooling vents
Exhaust vent for the internal cooling fan. Keep the vent free of obstructions to prevent overheating.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual
2-14 2200S-901-01 Rev. A/November 2011
(2) Factory test port
Do not use the factory test port.
Unauthorized use of the factory test port could damage this product.
(3) USB device port
The USB 2.0 high-speed device port supports remote control and data transfer to a PC.
(4) General purpose interface bus (GPIB) connector
The GPIB supports remote control and data transfer to a PC.
(5) 110 V/220 V power connector
Using the supplied line cord, connect to a grounded AC power outlet.
(6) 12-pin connector
The 12-pin connector includes remote sense. The illustration of the 12-pin connector (see Rear panel
at a glance (on page 2-13).) shows the location of the following output and control connectors:
Two "Output Drive + " connectors are equivalent to the front-panel output + connector
Two "Output Drive - " connectors are equivalent to the front-panel output - connector
Two connectors: "Output Sense + " and "Output Sense - " are used for remote sensing
Two No Connect connectors on the rear panel terminal strip are unused
Four connectors "Control In + ", Control In - ", "Control Out + ", and Control In - " on the rear panel
terminal strip are configured by the port mode controls. They support the trigger, RI/DFI, and digital I/O
functions. See >System in Configuration menus description (on page 2-9) for more information about
the port mode controls.
(7) Shorting clip
Use the shorting clip for proper regulation when not using remote sense. This connector comes
installed. It shorts the + output drive to the + sense and the - output drive to the - sense for local
sensing at the output terminals. To use remote sense, remove the clip.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-15
110 V/220 V selector switch
The 110 V/220 V selector switch is located on the bottom of the instrument as shown in the following
figure.
Figure 5: Selector switch
Front panel operation
Within a few seconds after turning on, the power supply displays the actual voltage and current output
value on the display’s top line and the voltage and current settings on the bottom line.
You can enable or disable the output of the power supply from the front panel by pressing the Output On/Off key. When the output is off, the OFF message displays to the right of the current and voltage
values.
The display shows the present operating status of the power supply with display messages. When the
power supply operates in constant voltage mode, the CV message displays. When the power supply
operates in the constant current mode, the CC message displays.
If the front panel was locked with a password, enter the correct password after you press a key (Vset, I-set, Save, Recall, or Shift), then you can change the settings.
To cancel a function operation (V-set, I-set, Save, Recall, or Shift), press Esc.
Loading...
+ 142 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.