Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Model 2010 Multimeter
User’s Manual
WARRANTY
Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 3 years from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation
prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written
consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable
batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE
REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR
ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT
OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS
BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED
DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND
INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE
TO PROPERTY.
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Level letter increases alphabetically as the manual undergoes
subsequent updates. Addenda, which are released between Revisions, contain important change
information that the user should incorporate immediately into the manual. Addenda are
numbered sequentially. When a new Revision is created, all Addenda associated with the
previous Revision of the manual are incorporated into the new Revision of the manual. Each new
Revision includes a revised copy of this print history page.
Revision A (Document Number 2010-900-01)............................................................ January 1996
Revision B (Document Number 2010-900-01) .......................................................... February 1996
Addendum B (Document Number 2010-900-02).................................................... September 1996
Revision C (Document Number 2010-900-01) ................................................................. June 1998
Revision D (Document Number 2010-900-01)................................................................ April 1999
Revision E (Document Number 2010-900-01) .............................................................August 2003
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
S
afety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before
using the product. Refer to the manual for complete product specifications
If the product is used in a manner not specified, the protection pr vided by the product may be impaired.
The types of product users are:
Responsible body
ment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained
Operators
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel
voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state
if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel
trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category
II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and
data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with
high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often
associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to
Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixture .
The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS,
42.4V peak, or 60VDC are present.
circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock.
If the circuit is capable of operating at or above 1000 volts,
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective
devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equip-
use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
perform routine procedures on the product to keep it operating properly, for example, setting the line
are trained to work on live circuits, and perform safe installations and repairs of products. Only properly
A good safety practice is to expect that hazardous voltage is present in any unknown
no conductive part of the circuit may be exposed.
5/03
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground.
Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being
measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating
information, and as shown on the instrument or test fixture panels, or switching card
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of
normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The symbol indicates a connection terminal to the equipment frame.
The
WARNING
information very carefully before performing the indicated procedure.
The
CAUTION
warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer,
test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety
approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from
other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only
through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability
of a replacement component, call a Keithley Instruments office for information
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for
proper cleaning/servicing.
heading in a manual explains dangers that might result in personal injury or death. Always read the associated
heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the
This section contains general information about the Model 2010 Multimeter. The information
is organized as follows:
•Feature overview
•Warranty information
•Manual addenda
•Safety symbols and terms
•Specification
•Inspection
•Options and accessories
If you have any questions after reviewing this information, please contact your local
Keithley representative or call one of our Applications Engineers at 1-800-348-3735
(U.S. and Canada only). Worldwide phone numbers are listed at the front of this manual.
Feature overview
The Model 2010 is a 7½-digit high-performance digital multimeter. It has 0.0018% 90-day
basic DC voltage accuracy and 0.0032% 90-day basic resistance accuracy. At 6
multimeter delivers 50 triggered readings/sec over the IEEE-488 bus. At 4
up to 2000 readings/sec into its internal buffer. The Model 2010 has broad measurement ranges:
½
digits, the
½
digits, it can read
•DC voltage from 10nV to 1000V.
•AC (RMS) voltage from 0.1
•DC current from 10nA to 3A.
•AC (RMS) current from 1
•Two and four-wire resistance from 1µ
•Frequency from 3Hz to 500kHz.
•Thermocouple temperature from -200°C to +1372°C.
•RTD temperature from -200°C to +630°C.
Some additional capabilities of the Model 2010 are:
•Full range of functions — In addition to those listed above, the Model 2010 functions
include period, dB, dBm, continuity, diode testing, mX+b, and percent.
•Optional scanning — For internal scanning, options include the Model 2000-SCAN, a
10-channel, general-purpose card, and the Model 2001-TCSCAN, a 9-channel,
thermocouple card with a built-in cold junction. For external scanning, the Model 2010
is compatible with Keithley's Model 7001 and 7002 switch matrices and cards.
•Programming languages and remote interfaces — The Model 2010 offers two
programming language choices (SCPI and Keithley Models 196/199), and two remote
interface ports (IEEE-488/GPIB and RS-232C).
•Reading and setup storage — Up to 1024 readings and two setups (user and factory
defaults) can be stored and recalled.
•Closed-cover calibration — The instrument can be calibrated either from the front panel
or remote interface.
µ
V to 750V, 1000V peak.
µ
A to 3A.
Ω
to 120MΩ.
Warranty information
Warranty information is located at the front of this instruction manual. Should your
Model 2010 require warranty service, contact the Keithley representative or authorized
repair facility in your area for further information. When returning the instrument for repair,
be sure to fill out and include the service form at the back of this manual to pr vide the
repair facility with the necessary information.
Manual addenda
Any improvements or changes concerning the instrument or manual will be explained in
an addendum included with the manual. Be sure to note these changes and incorporate them
into the manual.
Safety symbols and terms
General Information1-3
The following symbols and terms may be found on the instrument or used in this manual.
Thesymbol on the instrument indicates that the user should refer to the operating
instructions located in the manual.
Thesymbol on the instrument shows that high voltage may be present on the terminal(s).
Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING
injury or death. Always read the associated information very carefully before performing the
indicated procedure.
The
instrument. Such damage may invalidate the warranty.
Specifications
Full Model 2010 specif cations are included in Appendix A.
!
CAUTION
heading used in this manual explains dangers that might result in personal
heading used in this manual explains hazards that could damage the
1-4General Information
Inspection
The Model 2010 was carefully inspected electrically and mechanically before shipment.
After unpacking all items from the shipping carton, check for any obvious signs of physical
damage that may have occurred during transit. (Note: There may be a protective film ver the
display lens, which can be removed.) Report any damage to the shipping agent immediately.
Save the original packing carton for possible future shipment. The following items are included
with every Model 2010 order:
•Model 2010 Multimeter with line cord.
•Safety test leads (Model 1751).
•Accessories as ordered.
•Certificate of calibration.
•Model 2010 User's Manual (P/N 2010-900-00).
•Model 2010 Service Manual (P/N 2010-902-00).
•Model 2010 Support Software Disk including TestPoint run-time applications, TestPoint
instrument libraries for GPIB and RS-232, and QuickBASIC examples.
If an additional manual is required, order the appropriate manual package. The manual
packages include a manual and any pertinent addenda.
Options and accessories
The following options and accessories are available from Keithley for use with the
Model 2010.
Scanner cards
Model 2000-SCAN —
Model 2010. Channels can be conf gured for two-pole or four-pole operation. Included are two
pairs of leads for connection to Model 2010 rear panel inputs (Keithley P/N CA-109).
Model 2001-TCSCAN —
Model 2010. The card has nine analog input channels that can be used for high-accuracy, highspeed scanning. A built-in temperature reference allows multi-channel, cold-junction
compensated temperature measurements using thermocouples.
General purpose probes
Model 1754 Universal Test Lead Kit
lugs, two banana plugs, two hooks, and two alligator clips.
Model 8605 High Performance Modular Test Leads
(1000V) test probes and leads. The test leads are terminated with a banana plug with a
retractable sheath on each end.
A ten-channel scanner card that installs in the option slot of the
A thermocouple scanner card that installs in the option slot of the
— Consists of one set of test leads (0.9m), two spade
— Consists of two high voltage
General Information1-5
Model 8606 High Performance Probe Tip Kit
clips, and two spring hook test probes. (The spade lugs and alligator clips are rated at 30V RMS,
42.4V peak; the test probes are rated at 1000V.) These components are for use with high
performance test leads terminated with banana plugs, such as the Model 8605.
The following test leads and probes are rated at 30V RMS, 42.4V peak:
Models 5805 and 5805-12 Kelvin Probes —
probes with banana plug termination. Designed for instruments that measure four-terminal
resistance. The Model 5805 is 0.9m long; the Model 5805-12 is 3.6m long.
Model 5806 Kelvin Clip Lead Set —
plug termination. Designed for instruments that measure four-terminal resistance. A set of eight
replacement rubber bands is available (Keithley P/N GA-22).
Model 8604 SMD Probe Set —
surface mount device “grabber” clip on one end and a banana plug with a retractable sheath on
the other end.
Low thermal probes
Model 8610 Low Thermal Shorting Plug —
1-inch square circuit board, interconnected to provide a short circuit among all plugs.
Cables and adapters
Models 7007-1 and 7007-2 Shielded GPIB Cables —
bus using shielded cables and connectors to reduce electromagnetic interference (EMI). The
Model 7007-1 is 1m long; the Model 7007-2 is 2m long.
— Consists of two spade lugs, two alligator
Consists of two spring-loaded Kelvin test
Includes two Kelvin clip test leads (0.9m) with banana
Consists of two test leads (0.9m), each terminated with a
Consists of four banana plugs mounted to a
Connect the Model 2010 to the GPIB
Models 8501-1 and 8501-2 Trigger Link Cables —
instruments with Trigger Link connectors (e.g., Model 7001 Switch System). The Model 8501-1
is 1m long; the Model 8501-2 is 2m long.
Model 8502 Trigger Link Adapter —
the Model 2010 to instruments that use the standard BNC trigger connectors.
Model 8503 DIN to BNC Trigger Cable —
(Voltmeter Complete) and two (External Trigger) of the Model 2010 to instruments that use
BNC trigger connectors. The Model 8503 is 1m long.
Lets you connect any of the six Trigger Link lines of
Connect the Model 2010 to other
Lets you connect Trigger Link lines one
1-6General Information
Rack mount kits
Model 4288-1 Single Fixed Rack Mount Kit —
19-inch rack.
Model 4288-2 Side-by-Side Rack Mount Kit —
486, 487, 2000, 2001, 2002, 2010, 6517, 7001) side-by-side in a standard 19-inch rack.
Model 4288-3 Side-by-Side Rack Mount Kit —
side-by-side in a standard 19-inch rack.
Model 4288-4 Side-by-Side Rack Mount Kit —
instrument (Models 195A, 196, 220, 224, 230, 263, 595, 614, 617, 705, 740, 775, etc.)
side-by-side in a standard 19-inch rack.
Carrying case
Model 1050 Padded Carrying Case —
and shoulder strap.
Mounts a single Model 2010 in a standard
Mounts two instruments (Models 182, 428,
Mounts a Model 2010 and a Model 199
Mounts a Model 2010 and a 5.25-inch
A carrying case for a Model 2010. Includes handles
2
Basic Mea-
surements
2
Basic
Measurements
2-2Basic Measurements
Introduction
This section summarizes front panel operation of the Model 2010. It is organized as follows:
•
Front panel summary —
connections.
•
Rear panel summary —
•
Power-up —
the warm-up time, and default conditions.
•
Display —
instrument.
•
Measuring voltage —
level voltage considerations.
•
Ratio
with the SENSE terminals.
•
Measuring current —
fuse replacement.
•
Measuring resistance —
shielding considerations, dry circuit measurement, and offset compensation.
•
Measuring frequency and period —
connections.
•
Measuring temperature —
temperature measurements.
•
Math —
readings.
•
Measuring continuity —
•
Testing diodes —
Includes an illustration and summarizes keys, display, and
Includes an illustration and summarizes connections.
Describes connecting the instrument to line power, the power-up sequence,
Discusses the display format and messages that may appear while using the
Covers DC and AC voltage measurement connections and low
— Details ratio function connections for DC voltages and voltage measurement
Covers DC and AC current measurement connections and current
Details two and four-wire measurement connections,
Covers frequency and period measurement
Describes the use of thermocouples and four-wire RTDs for
Covers the mX+b, percent, dBm, and dB math functions performed on single
Explains setting up and measuring continuity of a circuit.
Describes testing general-purpose and zener diodes.
Front panel summary
Fi
1
M
p
The front panel of the Model 2010 is shown in Figure 2-1. This figure includes important
abbreviated information that should be reviewed before operating the instrument.
Basic Measurements2-3
gure 2-
odel 2010 front
anel
5
1
3
SHIFT
LOCAL
POWER
1Function keys
STEPCH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
CH1REM
SCAN
TALK
LSTN
SRQ
SHIFT
TIMER
HOLD TRIG FAST MED SLOWAUTO ERR
%
MX+B
DCV
ACV
HOLD
EX TRIG
TRIG
SAVE SETUP
OPEN CLOSE
dBm
DCI
LIMITS ON/OFFDELAY
STORE
RECALL
CONFIG HALT
STEP SCAN
REL FILT
dB
ACI
CONT
Ω2Ω4
TYPE
RATIO
RELFILTER
RS232
GPIB
DIGITS RATE
2
(shifted and unshifted)
BUFFER
STAT
PERIOD SENSOR
FREQ
DRYCKT O COMP
CAL TEST
EXIT ENTER
8
MATH
REAR
4W
2010 MULTIMETER
TEMP
RANGE
AUTO
RANGE
SENSE
INPUT
Ω 4 WIRE
HI
350V
PEAK
INPUTS
F
FRONT/REAR
4
7
1000V
!
PEAK
LO
500V
PEAK
R
3A 250V
AMPS
6
Select measurement function (DC and AC voltage, DC and AC current, two-wire and
four-wire resistance, frequency, period, temperature with thermocouples or four-wire RTDs),
math function (mX+b, %, dBm, dB), or special function (continuity, diode test).
2Operation keys
EXTRIGSelects external triggers (front panel, bus, trigger link) as the trigger source.
TRIG Triggers a measurement from the front panel.
STORE Enables reading storage.
RECALLDisplays stored readings and buffer statistics (maximum, minimum, average,
standard deviation). Use ▲ and ▼ to scroll through buffer; use and to
toggle between reading number and reading.
FILTERDisplays digital filter status for present function and toggles filter on/ f.
RELEnables/disables relative reading on present function.
and Moves through selections within functions and operations. If scanner card
installed, manually scans channels.
OPENOpens all channels on internal scanner card; stops scanning.
CLOSECloses selected internal channel.
STEPSteps through channels; sends a trigger after each channel.
SCANScans through channels; sends a trigger after last channel.
DIGITSChanges number of digits of resolution.
RATEChanges reading rate: fast, medium, slow.
EXITCancels selection, moves back to measurement display.
ENTERAccepts selection, moves to next choice or back to measurement display.
SHIFTUsed to access shifted keys.
LOCALCancels GPIB remote mode.
2-4Basic Measurements
3Shifted operation keys
mX+BManipulates normal display readings(X) using the equation Y=mX+b.
dBmConverts a value to the decibels above or below a 1mW reference.
dBCompresses a large range of DC or AC voltage measurements into a much
CONTMeasures circuit continuity on the 1kΩ range.
SENSORChooses temperature sensor (thermocouple or four-wire RTD).
PERCENTSelects the percentage calculations and lets you specify a reference value.
PERIODMakes period measurements from 2µs to 333ms on voltage ranges of 100mV,
LOCALBrings into remote mode for front panel control.
DELAYSets user delay between trigger and measurement.
HOLDHolds reading when the selected number of samples is within the selected
LIMITSSets upper and lower limit values for readings.
ON/OFFEnables/disables limits; selects beeper operation for limit testing.
TYPESelects the number of readings to be taken and the filter type, moving average
RATIOPerforms ratio function between sense inputs (denominator) and measure
DRY CKTEnables/disables dry circuit testing.
O COMPEnables/disables offset compensation
SAVESaves present configuration for p wer-on user default.
SETUPRestores factory or user default configuration
CONFIGSelects minimum/maximum channels, timer, and reading count for step/scan.
HALTTurns off step/scan.
GPIBEnables/disables GPIB interface; selects address and language.
RS232Enables/disables RS-232 interface; selects baud rate, fl w control, terminator.
TESTSelects display or key test.
CALAccesses calibration.
4Range keys
▲Moves to higher range, increments digit, and moves to next selection.
▼Moves to lower range, decrements digit, and moves to previous selection.
AUTOEnables/disables autorange.
5Annunciators
*(asterisk)Reading being stored.
(diode)Instrument is in diode testing function.
)))
(speaker)Beeper on for continuity or limits testing.
(more)Indicates additional selections are available.
4WFour-wire resistance reading displayed.
AUTOAutoranging enabled.
BUFFERRecalling stored readings.
CH 1-10Displayed internal channel is closed.
ERRQuestionable reading; invalid cal step.
FASTFast reading rate.
FILTDigital filter enabled
smaller scope.
Measures the forward voltage drop of general-purpose diodes, the zener
voltage of zener diodes, and the test current range from the front panel.
1V, 10V, 100V, and 750V
tolerance.
or repeating.
inputs (numerator) for DC volts only.
Basic Measurements2-5
HOLDInstrument is in hold mode.
LSTNInstrument addressed to listen over GPIB.
MATHMath function (mX+b, %, dB, dBm) enabled.
MEDMedium reading rate.
REARReading acquired from rear inputs.
RELRelative reading displayed.
REMInstrument is in GPIB remote mode.
SCANInstrument is in scan mode.
SHIFTAccessing shifted keys.
SLOWSlow reading rate.
SRQService request over GPIB.
STATDisplaying buffer statistics.
STEPInstrument is in step mode.
TALKInstrument addressed to talk over GPIB.
TIMERTimed scans in use.
TRIGIndicates external trigger (front panel, bus, trigger link) selected.
6Input connections
INPUT HI and LOUsed for making DC volts, AC volts, two-wire resistance
AMPSUsed in conjunction with INPUT LO to make DC current and AC
SENSE Ω4 WIREUsed with INPUT HI and LO to make four-wire resistance measureHI and LOments and RATIO measurements in conjunction with INPUT HI and
measurements.
current measurements. Also holds current input fuse (3A, 250V, fast
blow, 5×20mm).
LO.
7INPUTS
Selects input connections on front or rear panel.
8Handle
Pull out and rotate to desired position.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Fi
2
M
p
2-6Basic Measurements
Rear panel summary
The rear panel of the Model 2010 is shown in Figure 2-2. This figure includes important
abbreviated information that should be reviewed before operating the instrument.
gure 2-
odel 2010 rear
anel
345
2
HI
350V
PEAK
SENSE
Ω 4W
1
!
1000V
PEAK
!
LO
500V
PEAK
INPUT
OPTION SLOT
TRIGGER
LINK
!
3 5
1
VMC
4 6
2
EXT TRIG
FUSE LINE
250mAT
(SB)
125mAT
(SB)
6
8
7
5
2
1
#2
EXTERNAL TRIGGER INPUT
Trigger Reading
TTL HI
>2µsec
TTL LO
VOLTMETER COMPLETE OUTPUT
MADE IN
!
100 VAC
120 VAC
220 VAC
240 VAC
U.S.A.
RS232
120
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
LINE RATING
50, 60
400HZ
22 VA MAX
6
#7, #8
34
DIGITAL COMMON
#1
Reading
Complete
>10µsec
TTL HI
TTL LO
Basic Measurements2-7
1Option slot
An optional scanner card (Model 2000-SCAN or 2001-TCSCAN) installs in this slot.
2Input connections
INPUT HI and LOUsed for making DC volts, AC volts, two-wire resistance
SENSE Ω4 WIREUsed with INPUT HI and LO to make four-wire resistance measureHI and LOments, for connecting scanner card, and RATIO measurements in
measurements, and for connecting scanner card.
conjunction with INPUT HI and LO.
3TRIGGER LINK
One eight-pin micro-DIN connector for sending and receiving trigger pulses among other
instruments. Use a trigger link cable or adapter, such as Models 8501-1, 8501-2, 8502, 8503.
4RS-232
Connector for RS-232 operation. Use a straight-through (not null modem) DB-9 shielded
cable.
5IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as Models 7007-1 and
7007-2.
6Power module
Contains the AC line receptacle, power line fuse, and line voltage setting. The Model 2010
can be configured for line voltages of 100V/120V/220V/240VAC at line frequencies of 45Hz
to 66Hz or 360Hz to 440Hz.
7,8Digital Common
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Fi
3
P
2-8Basic Measurements
Power-up
Line power connection
Follow the procedure below to connect the Model 2010 to line power and turn on the
instrument.
1.Check to be sure the line voltage selected on the rear panel (see Figure 2-3) is correct for
the operating voltage in your area. If not, refer to the next procedure, “Setting line
voltage and replacing fuse.”
CAUTIONOperating the instrument on an incorrect line voltage may cause damage to
2.Before plugging in the power cord, make sure the front panel power switch is in the off
(0) position.
3.Connect the female end of the supplied power cord to the AC receptacle on the rear
panel. Connect the other end of the power cord to a grounded AC outlet.
WARNINGThe power cord supplied with the Model 2010 contains a separate ground
the instrument, possibly voiding the warranty.
wire for use with grounded outlets. When proper connections are made,
instrument chassis is connected to power line ground through the ground
wire in the power cord. Failure to use a grounded outlet may result in
personal injury or death due to electric shock.
gure 2-
ower module
4.Turn on the instrument by pressing the front panel power switch to the on (1) position.
Model 2010
HI
1000V
350V
PEAK
!
PEAK
LO
500V
3 5
1
PEAK
INPUT
SENSE
Ω 4W
!
4 6
2
TRIGGER
MADE IN
U.S.A.
LINK
RS232
!
VMC
EXT TRIG
!
FUSE LINE
250mAT
100 VAC
(SB)
120 VAC
220 VAC
125mAT
240 VAC
(SB)
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
120
LINE RATING
50, 60
400HZ
17 VA MAX
Line Voltage Selector
Fuse
220
240
120
100
Spring
Window
Fuse Holder Assembly
Setting line voltage and replacing fuse
A rear panel fuse located next to the AC receptacle protects the power line input of the
instrument. If the line voltage setting needs to be changed or the line fuse needs to be replaced,
perform the following steps.
WARNINGMake sure the instrument is disconnected from the AC line and other
equipment before changing the line voltage setting or replacing the line fuse.
1.Place the tip of a f at-blade screwdriver into the power module by the fuse holder
assembly (see Figure 2-3). Gently push in and move to the left. Release pressure on the
assembly, and its internal spring will push it out of the power module.
2.Remove the fuse, and replace it with the type listed in Table 2-1.
CAUTIONFor continued protection against fire or instrument damage, only replace
fuse with the type and rating listed. If the instrument repeatedly blows fuses,
locate and correct the cause of the trouble before replacing the fuse. See the
Model 2010 Service Manual for troubleshooting information.
3.If configuringthe instrument for a different line voltage, remove the line voltage selector
from the assembly and rotate it to the proper position. When the selector is installed into
the fuse holder assembly, the correct line voltage appears inverted in the window.
4.Install the fuse holder assembly into the power module by pushing it in until it locks in
place.
Basic Measurements2-9
Table 2-1
Fuse ratings
Line voltageFuse ratingKeithley P/N
100/120V
220/240V
0.25A slow-blow 5×20mm
0.125A slow-blow 5×20mm
FU-96-4
FU-91
2-10Basic Measurements
Power-up sequence
On power-up, the Model 2010 performs self-tests on its EPROM and RAM and momentarily
lights all segments and annunciators. If a failure is detected, the instrument momentarily
displays an error message and the ERR annunciator turns on. (Error messages are listed in
Appendix B.)
NOTEIf a problem develops while the instrument is under warranty, return it to Keithley
If the instrument passes the self-tests, the f rmware revision levels are displayed. An example
of this display is:
REV: A01 A02
where: A01 is the main board ROM revision.
After the power-up sequence, the instrument begins its normal display of readings.
Instruments, Inc., for repair.
A02 is the display board ROM revision.
High energy circuit safety precautions
To optimize safety when measuring voltage in high energy distribution circuits, read and use
the directions in the following warning.
WARNINGDangerous arcs of an explosive nature in a high energy circuit can cause
severe personal injury or death. If the multimeter is connected to a high
energy circuit when set to a current range, low resistance range, or any other
low impedance range, the circuit is virtually shorted. Dangerous arcing can
result even when the multimeter is set to a voltage range if the minimum
voltage spacing is reduced in the external connections.
When making measurements in high energy circuits, use test leads that meet the following
requirements:
•Test leads should be fully insulated.
•Only use test leads that can be connected to the circuit (e.g., alligator clips, spade lugs,
etc.) for hands-off measurements.
•Do not use test leads that decrease voltage spacing. These diminish arc protection and
create a hazardous condition.
Use the following sequence when testing power circuits:
Basic Measurements2-11
1.De-energize the circuit using the regular installed connect-disconnect device, such as a
circuit breaker, main switch, etc.
2.Attach the test leads to the circuit under test. Use appropriate safety rated test leads for
this application.
3.Set the multimeter to the proper function and range.
4.Energize the circuit using the installed connect-disconnect device and make
measurements without disconnecting the multimeter.
5.De-energize the circuit using the installed connect-disconnect device.
6.Disconnect the test leads from the circuit under test.
WARNINGThe maximum common-mode voltage (voltage between INPUT LO and the
chassis ground) is 500V peak. Exceeding this value may cause a breakdown
in insulation, creating a shock hazard.
2-12Basic Measurements
Power-on defaults
Power-on defaults are the settings the instrument assumes when it is turned on. The
Model 2010 offers two choices for the settings: factory and user. The power-on default will be
the last conf guration you saved. The SAVE and SETUP keys select the two choices of power-on
defaults.
To save present conf guration as user settings:
1.Configure the instrument as desired for USER de ault.
2.Press SHIFT then SAVE.
3.Use the ▲ and ▼ keys to select YES or NO.
4.Press ENTER.
To restore factory or user settings:
1.Press SHIFT then SETUP.
2.Use the ▲ and ▼ keys to select FACTory or USER.
3.Press ENTER.
Since the basic measurement procedures in this manual assume the factory defaults, reset the
instrument to the factory settings when following step-by-step procedures. Table 2-2 lists the
factory default settings.
Table 2-2
Factory defaults
SettingFactory default
Basic Measurements2-13
Autozero
Buffer
Continuity
Beeper
Digits
Rate
Threshold
Current (AC and DC)
Digits (AC)
Digits (DC)
Filter
Count
Mode
Range
Relative
Value
Rate (AC)
Rate (DC)
Diode test
Digits
Range
Rate
Frequency and Period
Digits
Range
Relative
Value
Rate
Function
GPIB
Address
Language
Key click
Limits
Beeper
High limit 1
Low limit 1
High limit 2
Low limit 2
mX+b
Scale factor
Offset
On
No effect
On
4½
Fast (0.1 PLC)
10Ω
5½
7½
On
10
Moving average
Auto
Off
0.0
Medium*
Medium (1 PLC)
6½
1mA
Medium (1 PLC)
6½
10V
Off
0.0
Slow (1 sec)
DCV
No effect
(16 at factory)
(SCPI at factory)
On
Off
Never
+1
-1
+2
-1
Off
1.0
0.0
2-14Basic Measurements
Table 2-2 (cont.)
Factory defaults
SettingFactory default
Percent
Reference
Resistance (two-wire and four-wire)
Digits
Filter
Count
Mode
Range
Relative
Value
Rate
Dry circuit
Offset compensation
RS-232
Baud
Flow
Tx term
Scanning
Channels
Mode
Temperature
Digits
Filter
Count
Mode
Junction
Temperature
Relative
Value
Rate
Sensor
Thermocouple
Four-wire RTD
Units
Triggers
Continuous
Delay
Source
Off
1.0
7½
On
10
Moving average
Auto
Off
0.0
Medium (1 PLC)
Off
Off
Off
No effect
No effect
No effect
Off
1-10
Internal
5½
On
10
Moving average
Simulated
23°C
Off
0.0
Medium (1 PLC)
Thermocouple
J
PT100
°C
On
Auto
Immediate
Table 2-2 (cont.)
Factory defaults
SettingFactory default
Voltage (AC and DC)
dB reference
dBm reference
Digits (AC)
Digits (DC)
Filter
Count
Mode
Range
Relative
Value
Rate (AC)
Rate (DC)
Ratio (DC)
Sensein
*DETector:BANDwidth 30
No effect
75Ω
5½
7½
On
10
Moving average
Auto
Off
0.0
Medium*
Medium (1 PLC)
Off
Off
Basic Measurements2-15
GPIB primary address
The GPIB primary address of the instrument must be the same as the primary address you
specify in the controller’s programming language. The default primary address of the instrument
is 16, but you can set the address to any value from 0 to 30 by using the following instructions.
1.Press SHIFT then GPIB.
2.Use the ▲ and ▼ keys to select ADDRess, or press ENTER. Once you have pressed
ENTER, the unit automatically displays the address selection.
3.Use the and keys to toggle from ADDRess to the numeric entry. Notice the
values are blinking.
4.Use the ▲ and ▼ keys to change the numeric entries to the desired address.
5.Press ENTER.
See Section Four — Remote Operation for more GPIB information.
Warm-up time
The Model 2010 is ready for use as soon as the power-up sequence has completed. However,
to achieve rated accuracy, allow the instrument to warm up for two hours. If the instrument has
been subjected to extreme temperatures, allow additional time for internal temperatures to
stabilize.
2-16Basic Measurements
Display
The display of the Model 2010 is primarily used to display readings, along with the units and
type of measurement. Annunciators are located on the top, bottom, right, and left of the reading
or message display. The annunciators indicate various states of operation. See Figure 2-1 for a
complete listing of annunciators.
Status and error messages
Status and error messages are displayed momentarily. During Model 2010 operation and
programming, you will encounter a number of front panel messages. Typical messages are either
of status or error variety, as listed in Appendix B.
Measuring voltage
The Model 2010 can make DCV measurements from 10nV to 1000V and ACV
measurements from 0.1µV to 750V RMS, 1000V peak.
Connections
Assuming factory default conditions, the basic procedure is:
1.Connect test leads to the INPUT HI and LO terminals. Either the front or rear inputs can
2.Select the measurement function by pressing DCV or ACV.
3.Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
4.Connect test leads to the source as shown in Figure 2-4.
WARNINGMaximum common mode voltage (voltage between LO and chassis ground) is
CAUTIONDo not apply more than 1000V peak to the input or instrument damage may
5.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
6.Take readings from the display.
be used; place the INPUTS button in the appropriate position.
autoranging. If you want manual ranging, use the RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected voltage.
500V peak. Exceeding this value may cause a shock hazard.
occur. The voltage limit is subject to the 8
until an on-scale reading is displayed (or press AUTO for autoranging). Use the lowest
possible range for the best resolution.
Caution: Maximum Input = 750V RMS, 1000V peak, 8 x 107 V•Hz
Maximum Common Mode = 500V peak.
0
STATREL FILT4WBUFFER
MATH
REAR
2001 MULTIMETER
AC Voltage
Source
Crest factor
AC voltage and current accuracies are affected by the crest factor of the waveform, the ratio
of the peak value to the RMS value. 500Hz is the maximum fundamental frequency at which the
corresponding crest factor must be taken into account for accuracy calculations.
Low level considerations
For sensitive measurements, external considerations beyond the Model 2010 affect the
accuracy. Effects not noticeable when working with higher voltages are signif cant in microvolt
signals. The Model 2010 reads only the signal received at its input; therefore, it is important that
this signal be properly transmitted from the source. The following paragraphs indicate factors
that affect accuracy, including stray signal pick-up and thermal offsets.
2-18Basic Measurements
Shielding
AC voltages that are extremely large compared with the DC signal to be measured may
produce an erroneous output. Therefore, to minimize AC interference, the circuit should be
shielded with the shield connected to the Model 2010 INPUT LO (particularly for low level
sources). Improper shielding can cause the Model 2010 to behave in one or more of the
following ways:
•Unexpected offset voltages.
•Inconsistent readings between ranges.
•Sudden shifts in reading.
To minimize pick-up, keep the voltage source and the Model 2010 away from strong AC
magnetic sources. The voltage induced due to magnetic f ux is proportional to the area of the
loop formed by the input leads. Therefore, minimize the loop area of the input leads and connect
each signal at only one point.
Thermal EMFs
Thermal EMFs (thermoelectric potentials) are generated by thermal differences between the
junctions of dissimilar metals. These can be large compared to the signal that the Model 2010
can measure. Thermal EMFs can cause the following conditions:
•Instability or zero offset is much higher than expected.
•The reading is sensitive to (and responds to) temperature changes. This effect can be
demonstrated by touching the circuit, by placing a heat source near the circuit, or by a
regular pattern of instability (corresponding to changes in sunlight or the activation of
heating and air conditioning systems).
To minimize the drift caused by thermal EMFs, use copper leads to connect the circuit to the
Model 2010. A banana plug generates a few microvolts. A clean copper conductor such as #10
bus wire is ideal for this application. The leads to the input may be shielded or unshielded, as
necessary.
Widely varying temperatures within the circuit can also create thermal EMFs. Therefore,
maintain constant temperatures to minimize these thermal EMFs. A shielded enclosure around
the circuit under test also helps by minimizing air currents.
The REL control can be used to null out constant offset voltages.
NOTEAdditional thermals may be generated by the optional scanner cards.
Basic Measurements2-19
AC voltage offset
The Model 2010, at 5½ digits resolution, will typically display 100 counts of offset on AC
volts with the input shorted. This offset is caused by the offset of the TRMS converter. This offset
will not affect reading accuracy and should not be zeroed out using the REL feature. The following equation expresses how this offset (V
) is added to the signal input (VIN):
OFFSET
Displayed readingVIN()2V
()
+=
OFFSET
2
Example: Range = 1VAC
Offset = 100 counts (1.0mV)
Input = 100mV RMS
Displayed reading100mV()21.0mV()
Displayed reading0.01V()1106–V×()+=
Displayed reading0.100005 =
+=
2
The offset is seen as the last digit, which is not displayed. Therefore, the offset is negligible.
If the REL feature were used to zero the display, the 100 counts of offset would be subtracted
from V
, resulting in an error of 100 counts in the displayed reading.
IN
See Section 3 for information on the conf guration options for DC and AC voltage
measurements.
2-20Basic Measurements
Ratio
The Model 2010 can perform a quotient calculation between the sense input (denominator)
and the measure input (numerator). This calculation can only be performed for DC voltages.
This function can be useful when comparing several voltages to a single voltage in a piece of
equipment. The sense input is used as the reference input. With this function, the sense terminals
can be used to measure DC volts in 100mV, 1V, and 10V ranges.
Connections
Assuming factory default conditions, the basic procedure is:
1.Connect test leads to the INPUT HI and LO terminals. Either the front or rear inputs can
be used; place the INPUTS button in the appropriate position.
2.Connect test leads to the SENSE HI and LO terminals. Use the same inputs (front or
rear) as in the previous step.
3.Press DCV.
4.Connect SENSE LO and LO together. SENSE LO and LO cannot have a voltage
difference of greater than 5% of either lowest range selected.
5.Press AUTO to toggle autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected voltages.
6.Press SHIFT then RATIO. Use the ▲, ▼, , and keys to toggle RATIO to ON and
SENSEIN to OFF. The display will read RA for RATIO.
NOTERATIO takes priority if both RATIO and SENSE IN are toggled to ON, and the display
will read RS at the far right. If only SENSE IN is turned ON, the Model 2010 reads
only the voltage present at the SENSE terminals.
7.Connect test leads from the INPUT HI and LO terminals to the source to be measured.
8.Connect test leads from the SENSE HI and LO terminals to the reference source.
CAUTIONDo not apply more than 1000V peak to the INPUT terminals or more than
350V peak to the SENSE terminals, or instrument damage may occur.
9.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until an on-scale reading is displayed (or press AUTO for autoranging). Use the lowest
possible range for the best resolution.
10.Take readings from the display.
NOTETo use autorange with the RATIO function, the following applies. When both RATIO
and SENSE IN are turned on (the display will show RS), the AUTO key applies to the
sense inputs. When RATIO is on and SENSE IN is off (display will show RA), the
AUTO key applies to the input terminals. To have autorange apply to both functions,
go into each function first and select AUTO before RATIO ON is enabled.
Measuring voltage with the SENSE terminals
The SENSE terminals can be used to measure DC voltage in the 100mV, 1V, and 10V ranges.
Assuming factory default conditions, make the connections as follows:
1.Connect test leads to the SENSE HI and LO terminals. Either the front or rear inputs can
be used; place the INPUTS button in the appropriate position.
2.Connect SENSE LO to LO. SENSE LO and LO cannot have a voltage difference of
greater than 5% of the lowest range selected.
3.Press DCV.
4.Press SHIFT then RATIO. Use the ▲, ▼, , and keys to toggle RATIO to OFF and
SENSEIN to ON. Note that the display will read VS for voltage on SENSE terminals.
5.Press AUTO to toggle autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected voltages.
NOTEOnly 100mV, 1V, and 10V ranges are available in either AUTO or manual ranging.
6.Connect test leads to the source.
CAUTIONDo not apply more than 350V peak to the SENSE terminals, or instrument
damage may occur.
7.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until an on-scale reading is displayed (or press AUTO for autoranging). Use the lowest
possible range for the best resolution.
8.Take readings from the display.
Basic Measurements2-21
Using ratio with the relative function
The relative (rel) function is normally used to null offsets or to subtract a baseline reading
from present and future readings. (See “Relative” in Section 3 for complete details.) When
relative is used with the ratio mode, the instrument calculates the ratio reading as follows:
Measure Input
Ratio = – Measure Input Rel Value
Sense Input
For example, assume the following:
Measure Input: 5V
Sense Input:2V
Rel Value: 1V
The ratio value is:
Ratio = (5/2) – 1 = 1.5
Fi
5
D
2-22Basic Measurements
Measuring current
The Model 2010 can make DCI measurements from 10nA to 3A and ACI measurements from
1µA to 3A RMS.
NOTESee the previous discussion about crest factor in “Measuring voltage” of this section.
Connections
Assuming factory default conditions, the basic procedure is:
1.Connect test leads to the AMPS and INPUT LO terminals. The front inputs must be used;
place the INPUTS button in the FRONT position.
2.Select the measurement function by pressing DCI or ACI.
3.Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected current.
4.Connect test leads to the source as shown in Figure 2-5.
CAUTIONDo not apply more than 3A, 250V to the input or the AMPS fuse will
open-circuit.
gure 2-
C and AC current
measurements
5.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until an on-scale reading is displayed (or press AUTO for autoranging). Use the lowest
possible range for the best resolution.
WARNINGMake sure the instrument is disconnected from the power line and other
equipment before replacing the AMPS fuse.
1.Turn off the power and disconnect the power line and test leads.
2.From the front panel, gently push in the AMPS jack with your thumb and rotate the fuse
carrier one-quarter turn counter-clockwise. Release pressure on the jack and its internal
spring will push the jack out of the socket.
3.Remove the fuse and replace it with the same type (3A, 250V, fast blow, 5 × 20mm). The
Keithley part number is FU-99-1.
CAUTIONDo not use a fuse with a higher current rating than specified or instrument
damage may occur. If the instrument repeatedly blows fuses, locate and
correct the cause of the trouble before replacing the fuse. See the Model 2010
Service Manual for troubleshooting information.
4.Install the new fuse by reversing the procedure above.
See Section 3 for information on the conf guration options for DC and AC current
measurements.
Basic Measurements2-23
2-24Basic Measurements
Measuring resistance
The Model 2010 can make two-wire and four-wire resistance measurements from 1µΩ to
120MΩ.
Connections
Assuming factory default conditions, the basic procedure is:
1.Connect test leads to the Model 2010 as follows:
A. For Ω2-wire, connect the test leads to INPUT HI and LO.
B. For Ω4-wire, connect the test leads to INPUT HI and LO, and SENSE Ω4 WIRE
HI and LO. Recommended Kelvin test probes include the Keithley Models 5805
and 5806. Either the front or rear inputs can be used; place the INPUTS button in
the appropriate position.
2.Select the measurement function by pressing Ω2 or Ω4.
3.Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE ▲ and ▼ keys to select a
measurement range consistent with the expected resistance.
4.Connect test leads to the resistance as shown in Figure 2-6.
CAUTIONDo not apply more than 1000V peak between INPUT HI and LO or 350V
peak between SENSE HI and SENSE LO, or instrument damage may occur.
5.Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until an on-scale reading is displayed. Use the lowest possible range for the best
resolution.
Note: Source current flows from the INPUT
HI to INPUT LO terminals.
Model 2010
CH1REM
SCAN
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
TALK
LSTN
SRQ
SHIFT
HOLD TRIG FAST MED SLOW AUTO ERR
TIMER
0
STATREL FILT4WBUFFER
MATH
REAR
2001 MULTIMETER
Shielded
Cable
Note: Source current flows from the INPUT HI
to INPUT LO terminals.
Optional Shield
Resistance
Under Test
Optional Shield
Resistance
Under Test
Shielding
To achieve a stable reading, shield resistances greater than 100kΩ. Place the resistance in a
shielded enclosure and connect the shield to the INPUT LO terminal of the instrument
electrically.
See Section 3 for information on the conf guration options for two-wire and four-wire
resistance measurements.
Low resistance measurements
The Model 2010 can be used for low resistance measurements normally handled by a microohmmeter. The following paragraphs discuss the Model 2010’s dry circuit testing and offset
compensation modes.
2-26Basic Measurements
Dry circuit testing
Many low resistance measurements are made on contact devices such as switches and relay
contacts. The purpose of these tests is to determine whether oxidation has increased the
resistance of the contacts. If the voltage across the contacts during the test is too high, the
oxidation will be punctured and render the test meaningless.
Dry circuit testing limits the voltage across the DUT to 20mV or less.
NOTEThis function is only available in four-wire ohms.
Offset compensation
Offset compensation is used to compensate for voltage potential, such as thermal offsets,
across the device under test. In offset compensation, a full-scale source current is applied to the
resistance being measured during part of the measurement cycle. Figure 2-7 shows the measurement cycle. During the f rst half of the measurement cycle, the reduced source current is applied
and the voltage being measured is any thermal EMFs present in the circuit plus the voltage
across R
S
with the reduced source current:
V
M2
V
+=
EMFISRRS
During the second half of the measurement cycle, the full-scale source current is on, and the
total voltage measured includes the voltage drop across the resistor and any thermal EMFs. This
is defined as follows:
V
V
M1
+=
EMFISFSRS
Offset compensation is available up to 100MΩ. However, compensation is only being used
on the 10KΩ and lower ranges. An ’o’ will flash on the display if offset compensation is turned
on and measuring 100KΩ and higher ranges, indicating offset compensation has no effect.
Since the thermal EMF voltage is measured during the f rst and second half of the cycle, it
can be subtracted from the voltage measurement made during the f rst half of the cycle. The
result is the offset-compensated voltage measurement:
–V=
V
M1VM2
V
+()V
V
M
EMFISFSRS
RSI
–()=
SFSISR
M
+()–=
EMFISRRS
Therefore,
V
R
S
M
-------------------------=
I
–
SFSISR
Fi
7
Basic Measurements2-27
gure 2-
Offset-compensated
One measurement cycle
ohms measurement
Source
Current
Thermal offset
measurement
Voltage measurement with
source current off
VM1
Voltage measurement with
source current on
VEMF
ISR
RS
VM1
VFMF
ISFS
RS
2-28Basic Measurements
Measuring frequency and period
The Model 2010 can make frequency measurements from 3Hz to 500kHz on voltage ranges
of 100mV, 1V, 10V, 100V, and 750V. Period measurements can be taken from 2µs to 333ms on
the same voltage ranges as the frequency.
The instrument uses the volts input terminals to measure frequency. The AC voltage range can
be changed with the RANGE ▲ and ▼ keys. The signal voltage must be greater than 10% of
the full-scale range.
CAUTIONThe voltage limit is subject to the 8
Trigger level
Frequency and Period use a zero-crossing trigger, which means a count is taken when the
frequency crosses the zero level. The Model 2010 uses a reciprocal counting technique to
measure frequency and period. This method generates constant measurement resolution for any
input frequency. The multimeter’s AC voltage measurement section performs input signal
conditioning.
Gate time
The gate time is the amount of time the Model 2010 uses to sample frequency or period
readings. All settings of the RATE key (FAST, MEDium, SLOW) yield a gate time of one
second.
The Model 2010 completes a reading when it receives its first zero-crossing after the ate
time expires. In other words, the reading is completed 1/2 cycle after the gate time has expired.
For example, with a 1 second gate time to sample a 3Hz frequency, you may wait up to 3 seconds
before the Model 2010 returns a reading.
×
107V•Hz product.
Connections
Fi
8
F
Assuming factory default conditions, the basic procedure is:
1.Connect test leads to the INPUT HI and LO terminals of the Model 2010. Either the front
2.Select the FREQ or PERIOD function.
3.Connect test leads to the source as shown in Figure 2-8.
CAUTIONDo not exceed 1000V peak between INPUT HI and INPUT LO or
4.Take a reading from the display.
See Section 3 for information on the conf guration options for frequency and period
measurements.
Basic Measurements2-29
or rear inputs can be used; place the INPUTS button in the appropriate position.
The Model 2010 can measure temperature with a four-wire RTD sensor or a thermocouple.
The temperature measurement ranges available depend on the type of RTD or thermocouple
chosen.
RTDs can be connected to the banana jacks on the front or rear panel.
Thermocouples can be connected to the Model 2001-TCSCAN card, which plugs into the
option slot of the Model 2010, or to an external thermocouple card, such as a Model 7057A,
7402, or 7014 installed in a Model 7001 or 7002 Switch System.
If the Model 2001-TCSCAN card is not used then an estimate of the panel temperature must
be made (usually 2°C above room temperature). Connect the thermocouple card directly to the
front panel Input HI and LO as shown in Figure 2-9. To input the panel temperature estimate,
choose the TCOUPLE conf guration option and select JUNC. Pick the SIM option and input the
estimate.
Connections
Fi
9
gure 2-
Thermocouple and RTD
temperature measurements
Basic Measurements2-31
2001-TCSCAN
+
-
Note: This thermocouple card
must be inserted into a
Keithley Model 2010.
Note: Front or rear inputs
can be used.
Model 2010
CH1REM
SCAN
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
TALK
0
LSTN
SRQ
SHIFT
STATRELFILT4WBUFFER
HOLDTRIG FAST MED SLOW AUTO ERR
TIMER
MATH
REAR
2001 MULTIMETER
Input
HI
Input
LO
CH 2
OUT A HI
OUT A LO
A. Thermocouple Connections
Model 2010
SCAN
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8CH9 CH1
CH1REM
TALK
LSTN
SRQ
SHIFT
HOLD TRIG FAST MED SLOW AUTO ERR
TIMER
0
STATREL FILT4WBUFFER
MATH
REAR
2001 MULTIMETER
B. 4-Wire RTD Connections
Input HI
Input LO
Sense Ω4-wire HI
Platinum
RTD
Sense Ω4-wire LO
2-32Basic Measurements
Configuration
The following provides the various configuration options for temperature measurements. To
select and conf gure either a thermocouple or four-wire RTD measurement:
Press SHIFT then SENSOR. Four choices are available using the ▲ and ▼ keys:
•UNITS — C, K, F (Centigrade, Kelvin, Fahrenheit). This parameter selects the displayed
units for temperature measurements.
•SENSOR — TCOUPLE, 4W-RTD (sensor type). This parameter selects the type of
sensor being used.
•TYPE — J, N, T, K (thermocouple type) or PT100, USER, PT3916, PT385, F100, D100
(4W-RTD type). Note that with USER selected, you must set the Alpha, Beta, Delta, and
RZero values from over the GPIB bus or the RS-232 Interface (see “USER RTD Type”
for details).
•JUNC — SIM, CH1 (simulated or referenced at Channel 1). Typically, a thermocouple
card uses a single reference junction. The Model 2010 can simulate a reference junction
temperature or use the reference junction on a thermocouple switching card. Typical
reference junction temperatures are 0°C and 23°C. In order to keep the reference
calculations updated and accurate, Channel 1 needs to be read periodically.
To assign a value to a parameter, use the ▲ and ▼ keys to scroll to the desired parameter.
Select the key, and use the ▲ and ▼ keys to scroll through and choose the preferred value.
Select the ENTER key to save your changes.
Math
USER RTD Type
Alpha, Beta, Delta, and RZero values for the USER RTD type cannot be set from the front
panel. These values can only be set remotely from the GPIB bus or the RS-232 Interface. After
selecting USER, use the following commands to set the RTD factors:
[:SENSe[1]]:TEMPerature:FRTD:ALPHa <NRf>Specify alpha value (0 to 0.01)
[:SENSe[1]]:TEMPerature:FRTD:BETA <NRf>Specify beta value (0 to 1.00)
[:SENSe[1]]:TEMPerature:FRTD:DELTa <NRf>Specify delta value (0 to 5.00)
[:SENSe[1]]:TEMPerature:FRTD:RZERo <NRf>Specify resistance at 0° C (0 to 10000)
NOTEFor details on these commands, see “FRTD commands” in the SENSe Subsystem
(Section 5).
Model 2010 math operations are divided into four categories:
•mX+b and percent
•dBm and dB calculations
•Statistics of buffered readings
•Limit testing
mX + b
Basic Measurements2-33
The first two categories are discussed in the following paragraphs. Buffered reading statistics
and reading limit testing are described in Section 3.
The procedure to select and conf gure a math operation is:
1.Press SHIFT then the appropriate math key.
2.Configure the parameters for the math operation. Press ENTER when nished. (Press
SHIFT then the related math function to end the calculation.)
NOTES Once enabled for a function, the mX+b and percentage calculations are in effect
across function changes.
The Model 2010 uses IEEE-754 floating point format for math calculations.
This math operation lets you manipulate normal display readings (X) mathematically
according to the following calculation:
Y= mX + b
where: X is the normal display reading.
m and b are user-entered constants for scale factor and offset.
Y is the displayed result.
Configuration
To configure the mX+b calculation, perform the foll wing steps:
1.Press SHIFT then MX+B to display the present scale factor:
M: +1.000000 ^
2.Enter a value and units pref x. Use the and keys to choose a numerical place and
use the ▲ and ▼ keys to increment or decrement the digits.
3.Press ENTER to conf rm the M value and display the B value:
B: +00.00000 m
4.Enter a value and units pref x.
5.Press ENTER to conf rm the B value and display the UNITS designation:
MX
6.Scroll through the letters to change and press ENTER when f nished.
The Model 2010 will display the result of the calculation.
2-34Basic Measurements
Percent
Percent selects the percentage calculation and lets you specify a reference value. The
displayed reading will be expressed as a percent deviation from the reference value. The
percentage calculation is performed as follows:
where: Input is the normal display reading.
Configuration
To configure the percent calculation, perform the foll wing steps:
1.Press SHIFT then % to display the present value:
REF:+1.000000^
2.Enter a reference sign, value, and units pref x. Use the and keys to choose a
numerical place and use the ▲ and ▼ keys to increment or decrement the digits.
3.Press ENTER when done.
Input - Reference
Percent
-----------------------------------------Input
Reference is the user entered constant.
Percent is the displayed result.
100%×=
The Model 2010 will display the result of the calculation. The result is positive when the input
exceeds the reference and negative when the input is less than the reference. Engineering units
are used to show values in the range 1 nano to 1000G. Exponential notation is used above that
range.
dBm calculation
dBm is defined as decibels above or below a 1mW reference. With a user-programmable
reference impedance, the Model 2010 reads 0dBm when the voltage needed to dissipate 1mW
through the reference impedance is applied. The relationship between dBm, a reference
impedance, and the voltage is def ned by the following equation:
Where: V
Z
NOTEDo not confuse reference impedance with input impedance. The input impedance of
the instrument is not modified by the dBm parameter.
If a relative value is in effect when dBm is selected, the value is converted to dBm, and then
REL is applied to dBm. If REL is applied after dBm has been selected, dBm math has REL
applied to it.
2
/Z
V
IN
dBm = 10 log
is the DC or AC input signal.
IN
is the specif ed reference impedance.
REF
--------------------------------1mW
REF
Configuration
To set the reference impedance, perform the following steps:
1.After selecting dBm, the present reference impedance is displayed (1-9999Ω):
REF: 0075
2.To change the reference impedance, use the and keys to select the numeric
position. Then use the ▲ and ▼ keys to select the desired value. Be sure to press ENTER
after changing the reference impedance.
NOTES dBm is valid for positive and negative values of DC volts.
dB calculation
Expressing DC or AC voltage in dB makes it possible to compress a large range of
measurements into a much smaller scope. The relationship between dB and voltage is def ned
by the following equation:
Basic Measurements2-35
The mX+b and percent math operations are applied after the dBm or dB math. For
example, if mX+b is selected with m=10 and b=0, the display will read 10.000 MX
for a 1VDC signal. If dBm is selected with Z
V
dB= 20 log
IN
-----------------V
REF
= 50Ω, the display will read 130MX.
REF
where: V
is the DC or AC input signal.
IN
is the specif ed voltage reference level.
V
REF
The instrument will read 0dB when the reference voltage level is applied to the input.
If a relative value is in effect when dB is selected, the value is converted to dB then REL is
applied to dB. If REL is applied after dB has been selected, dB has REL applied to it.
Configuration
To set the reference voltage, perform the following steps:
1.After selecting dB, the voltage applied between HI and LO is acquired and presented as
the reference voltage. This level can then be adjusted.
REF: +1.000000^
2.To change the reference level, use the and keys to select the numeric position.
Then use the ▲ and ▼ keys to select the desired value. Move the cursor to the rightmost
position (^) and use the ▲ and ▼ keys to move the decimal point. Be sure to press
ENTER after changing the reference voltage.
NOTEThe largest negative value of dB is -160dB. This will accommodate a ratio of
V
= 10µV and V
IN
= 1000V.
REF
Fi
10
C
2-36Basic Measurements
Measuring continuity
The Model 2010 uses the 1kΩ range to measure circuit continuity. After selecting continuity,
the unit prompts you for a threshold resistance level (1Ω-1000Ω). The Model 2010 alerts you
with a beep when a reading is below the set level.
To measure the continuity of a circuit, press SHIFT then CONT, set the threshold resistance
level, and connect the circuit.
NOTEContinuity has a non-selectable reading rate of FAST (0.1 PLC).
Connections
Connect the circuit you want to test to the INPUT HI and INPUT LO terminals of the
Model 2010. The test current f ows from the INPUT HI as shown in Figure 2-10.
Note: Source current flows from the INPUT
HI to INPUT LO terminals.
0
STATREL FILT4WBUFFER
MATH
REAR
2001 MULTIMETER
Resistance
Under Test
You can define a threshold resistance from 1Ω to 1000Ω. The factory setting is 10Ω. Follow
1.Press SHIFT then CONT.
2.Use the and keys to choose a numerical place and use the ▲ and ▼ keys to
increment or decrement the digits. Enter a value from 1 to 1000.
3.Press ENTER to conf rm your setting.
Testing diodes
Fi
11
D
With a Model 2010, you can measure the forward voltage drop of general-purpose diodes and
the zener voltage of zener diodes. To test diodes, press SHIFT then , set the test current
range, connect the diode, and take a reading from the display.
NOTEDiode test has a non-selectable reading rate of MEDium (1 PLC).
Connections
Connect the diode leads to the INPUT HI and INPUT LO terminals on the Model 2010. The
test current f ows from the INPUT HI terminal as shown in Figure 2-11.
Note: Source current flows from the
INPUT HI to INPUT LO terminals.
Zener
Diode
You can set the test current range from the front panel. The choices are 1mA, 100µA, and
10µA. The factory test current setting is 1mA. To set the test current, perform the following
steps:
1.Press SHIFT and then.
2.Use the ▲ and ▼ keys to scroll through the three test current selections.
The diode test measures voltages up to 10V for the 1mA test current, 5V for 100µA, and 10V
for the 10µA range. If a reading is more than 10V, the Model 2010 displays the “OVERFLOW”
status message.
2-38Basic Measurements2-38Basic Measurements
3
Measurement
Options
3
Measurement
Options
3-2Measurement Options
Introduction
This section describes the front panel features of the Model 2010. For those measurement
options accessible only by a remote interface, refer to Sections 4 and 5. This section is organized
as follows:
•Measurement configuration — Describes ranging, f ltering, relative readings, digits of
resolution, and measurement rate.
•Trigger operations — Uses a trigger model to explain trigger modes and sources.
•Buffer operations — Discusses the reading storage buffer and buffer statistics.
•Limit operations — Defines how to set reading limits.
•Scan operations — Explains the internal and external scanning capabilities.
•System operations — Gives details on setup saving and restoring, selecting a remote
interface, and accessing test and calibration.
Measurement configuration
The following paragraphs discuss conf guring the multimeter for making measurements. See
the end of Appendix A for information about optimizing readings for speed or accuracy.
Range
The selected measurement range affects both the ultimate digits and accuracy of the
measurements as well as the maximum signal that can be measured. The range setting (f xed or
auto) for each measurement function is saved when changing functions.
Maximum readings
The full scale readings for every range on each function are 20% overrange except for the
1000VDC, 750VAC, 3ADC, 3AAC, 1MΩ two-wire and four-wire, and diode test ranges.
Input values more than the maximum readings cause the "OVERFLOW" message to be
displayed.
Manual ranging
To select a range, press the RANGE ▲ or ▼ key. The instrument changes one range per
keypress. The selected range is displayed for one second.
If the instrument displays the "OVERFLOW" message on a particular range, select a higher
range until an on-range reading is displayed. Use the lowest range possible without causing an
overfl w to ensure best accuracy and resolution.
Note that the temperature and continuity functions have just one range.
Filter
Measurement Options3-3
Autoranging
To enable autoranging, press the AUTO key. The AUTO annunciator turns on when
autoranging is selected. While autoranging is selected, the instrument automatically chooses the
best range to measure the applied signal. Autoranging should not be used when optimum speed
is required.
Note that up-ranging occurs at 120% of range, while down-ranging occurs at 10% of nominal
range.
To cancel autoranging, press AUTO or the RANGE ▲ or ▼ key. Pressing AUTO to cancel
autoranging leaves the instrument on the present range.
The AUTO key has no effect on the temperature, continuity, and diode test functions.
Filter lets you set the f lter response to stabilize noisy measurements. The Model 2010 uses a
digital filte . The displayed, stored, or transmitted reading is simply an average of a number of
reading conversions (from 1 to 100).
To select a filter
1.Use the FILTER button to enable the f lter. The FILT annunciator will come on when
FILTER is enabled.
2.Press SHIFT then TYPE.
3.Use the , , ▲, and ▼ keys to select the number of readings and then press ENTER.
4.Use the , , ▲, and ▼ keys to select the type of f lter desired (moving average or
repeating) and then press ENTER.
The FILT annunciator turns on. When a filter is enabled, the selected f lter configuration for
that measurement function is in effect.
Pressing FILTER once disables the f lter.
NOTEThe filter can be set for any measurement function except frequency, period,
continuity, and diode test.
Fi
1
M
3-4Measurement Options
Filter types
The moving average filter(Figure 3-1) uses a first-in, first-outstack. When the stack becomes
full, the measurement conversions are averaged, yielding a reading. For each subsequent
conversion placed into the stack, the oldest conversion is discarded, and the stack is re-averaged,
yielding a new reading.
For the repeating f lter (Figure 3-1), the stack is f lled and the conversions are averaged to
yield a reading. The stack is then cleared and the process starts over. Choose this f lter for
scanning so readings from other channels are not averaged with the present channel.
Response time
The filter parameters have speed and accuracy tradeoffs for the time needed to display, store,
or output a f ltered reading. These affect the number of reading conversions for speed versus
accuracy and response to input signal changes.
gure 3-
oving average and
repeating filters
Conversion #10
#9
Conversion #1
Conversion #10
#9
Conversion #1
#8
#7
#6
#5
#4
#3
#2
A. Type - Moving Average, Readings = 10
#8
#7
#6
#5
#4
#3
#2
B. Type - Repeating, Readings = 10
Reading
#1
Reading
#1
Conversion #11
#10
Conversion #2
Conversion #20
#19
Conversion #11
#9
#8
#7
#6
#5
#4
#3
#18
#17
#16
#15
#14
#13
#12
Conversion #12
#11
#8
Reading
#2
Conversion #3
Reading
#2
#10
#9
#7
#6
#5
#4
Conversion #30
#29
#26
Conversion #21
#28
#27
#25
#24
#23
#22
Reading
#3
Reading
#3
Relative
Measurement Options3-5
The rel (relative) function can be used to null offsets or subtract a baseline reading from
present and future readings. When rel is enabled, the instrument uses the present reading as a
relative value. Subsequent readings will be the difference between the actual input value and the
rel value.
You can definea rel value for each function. Once a rel value is established for a measurement
function, the value is the same for all ranges. For example, if 50V is set as a rel value on the
100V range, the rel is also 50V on the 1000V, 10V, 1V, and 100mV ranges.
NOTEWhen a rel value is larger than the range selected, the display is formatted to
maximum resolution and range information is lost.
Thus, when you perform a zero correction for DCV, Ω2, and Ω4 measurements by enabling
REL, the displayed offset becomes the reference value. Subtracting the offset from the actual
input zeroes the display, as follows:
Actual Input – Reference = Displayed Reading
A rel value can be as large as the highest range.
Selecting a range that cannot accommodate the rel value does not cause an overfl w
condition, but it also does not increase the maximum allowable input for that range. For
example, on the 10V range, the Model 2010 still overfl ws for a 12V input.
Digits
To set a rel (relative) value, press REL key when the display shows the value you want as the
relative value. The REL annunciator turns on. Pressing REL a second time disables rel.
You can input a REL value manually using the mX+b function. Set M for 1 and B for any
value. See Section 2 for more information on the mX+b function.
The display resolution of a Model 2010 reading depends on the DIGITS setting. It has no
effect on the remote reading format. The number of displayed digits does not affect accuracy or
speed. Those parameters are controlled by the RATE setting.
Perform the following steps to set digits for a measurement function:
1.Press the desired function.
2.Press the DIGITS key until the desired number of digits is displayed (3½ to 7½).
NOTEFrequency and period can be displayed with four to seven digits. ACV, AC Amps, and
dryckt ohms are limited to 6½ digits resolution.
3-6Measurement Options
Rate
The rate operation sets the integration time of the A/D converter, the period of time the input
signal is measured (also known as aperture). The integration time affects the usable digits, the
amount of reading noise, as well as the ultimate reading rate of the instrument. The integration
time is specif ed in parameters based on a number of power line cycles (NPLC), where 1 PLC
for 60Hz is 16.67msec and 1 PLC for 50Hz and 400Hz is 20msec.
In general, the fastest integration time (FAST (0.1 PLC) from the front panel, 0.01 PLC from
the bus) results in increased reading noise and fewer usable digits, while the slowest integration
time (SLOW (5PLC) from the front panel, 10PLC from the bus) provides the best commonmode and normal-mode rejection. In-between settings are a compromise between speed and
noise.
The RATE parameters are explained as follows:
•FAST sets integration time to 0.1 PLC. Use FAST if speed is of primary importance (at
the expense of increased reading noise and fewer usable digits).
•MEDium sets integration time to 1 PLC. Use MEDium when a compromise between
noise performance and speed is acceptable.
•SLOW sets integration time to 5PLC. SLOW provides better noise performance at the
expense of speed.
NOTEThe integration time can be set for any measurement function except frequency
(SLOW), period (SLOW), continuity (FAST), and diode test (MEDium). For frequency
and period, this value is gate time or aperture.
For the AC functions, MEDium and SLOW have no effect on the number of power line
cycles.
Measurement Options3-7
Bandwidth
The rate setting for AC voltage and current measurements determines the bandwidth setting:
•Slow — 3Hz to 300kHz.
•Medium — 30Hz to 300kHz.
•Fast — 300Hz to 300kHz.
Bandwidth is used to specify the lowest frequency of interest. When the Slow bandwidth
(3Hz to 300kHz) is chosen, the signal goes through an analog RMS converter. The output of the
RMS converter goes to a fast (1kHz) sampling A/D and the RMS value is calculated from 1200
digitized samples (1.2s).
When the Medium bandwidth (30Hz to 300kHz) is chosen, the same circuit is used. However,
only 120 samples (120ms) are needed for an accurate calculation because the analog RMS
converter has turned most of the signal to DC.
In the Fast bandwidth (300Hz to 300kHz), the output of the analog RMS converter (nearly
pure DC at these frequencies) is simply measured at 1 PLC (16.6ms), 60Hz line frequency.
Table 3-1 lists the rate settings for the various measurement functions. The FAST, MED, and
SLOW annunciators are only lit when conditions in the table are met. In other case, the
annunciators are turned off.
Table 3-1
Rate settings for the measurement functions
Function
DCV, DCI
ACV, ACI
Ω2W, Ω4W
FREQ, PERIOD
dB, dBm (ACV)
dB, dBm (DCV)
Continuity
Diode test
Notes:
NPLC = number of power line cycles.
BW = lower limit of bandwidth (in Hz).
APER = aperture in seconds.
N/A = not available.
X = setting ignored.
The following paragraphs discuss front panel triggering, the programmable trigger delay, the
reading hold feature, and external triggering.
Trigger model
The fl wchart in Figure 3-2 summarizes triggering as viewed from the front panel. It is called
a trigger model because it is modeled after the SCPI commands used to control triggering. Note
that for stepping and scanning, the trigger model has additional control blocks. These are
described later in this section.
gure 3-
ront panel triggering
without stepping/scanning
Idle
The instrument is considered to be in the idle state whenever it is not performing any
measurements or scanning functions. From the front panel, the unit is considered idle at the end
of a step or scan operation when the reading for the last channel remains displayed. To restore
triggers, use the SHIFT-HALT keys.
Once the Model 2010 is taken out of idle, operation proceeds through the f owchart.
Control source and event detection
The control source holds up operation until the programmed event occurs and is detected. The
control sources are described as follows:
Control
Source
Immediate
External
Idle
Event
Detection
Delay
Device
Action
Output
Trigger
•Immediate — With this control source, event detection is immediately satisf ed allowing
operation to continue.
•External — Event detection is satisf ed for any of the following three conditions:
• An input trigger via the Trigger Link line EXT TRIG is received.
• A bus trigger (GET or *TRG) is received.
• The front panel TRIG key is pressed. (The Model 2010 must be taken out of remote
before it will respond to the TRIG key. Use the LOCAL key or send LOCAL 716 over
the bus.)
Measurement Options3-9
Delay
A programmable delay is available after event detection. It can be set manually or an auto
delay can be used. With auto delay, the Model 2010 selects a delay based on the function and
range. The AUTO settings are listed in Table 3-2.
Table 3-2
Auto delay settings
FunctionRange and delay
DCV
ACV
FREQ
DCI
ACI
Ω2W, Ω4W
dryckt
w/ & w/o ocomp
Continuity
Diode testing
The delay function is accessed by pressing the SHIFT-DELAY keys. The present delay
setting (AUTO or MANual) is displayed. Use the ▲ and ▼ keys to select the type of delay. If
MANual is chosen, also enter the duration of the delay. The maximum is 99H:99M:99.999S.
Press ENTER to accept the delay or EXIT for no change.
100mV
1ms
100mV
400ms
100mV
1ms
10mA
2ms
10Ω
3ms
10Ω
3ms
1V
1ms
1V
400ms
1V
1ms
100mA
2ms
100Ω
3ms
100Ω
13ms
1kΩ
3ms
1mA
1ms
10V
1ms
10V
400ms
10V
1ms
1A
2ms
1A
400ms
1kΩ
3ms
100µA
1ms
100V
5ms
100V
400ms
100V
1ms
3A
2ms
3A
400ms
10kΩ
13ms
10µA
1ms
1000V
5ms
750V
400ms
750V
1ms
100kΩ
25ms
1MΩ
100ms
10MΩ
150ms
100MΩ
250ms
Changing the delay to MANual on one function changes the delays on all functions to
MANual.
3-10Measurement Options
Device actions
The primary device action is a measurement. However, the device action block could include
the following additional actions:
•Filtering — If the repeating f lter is enabled, the instrument samples the specif ed
number of reading conversions to yield single f ltered reading. Only one reading
conversion is performed if the f lter is disabled, or after the specif ed number of reading
conversions for a moving average filter is reached. The output of f lter feeds hold.
•Hold — With hold enabled, the f rst processed reading becomes the “seed” reading and
operation loops back within the device action block. After the next reading is processed,
it is checked to see if it is within the selected window (0.01%, 0.1%, 1%, 10%) of the
“seed” reading. If the reading is within the window, operation again loops back within
the device action block. This looping continues until the specif ed number (2 to 100)
consecutive readings are within the window. If one of the readings is not within the
window, the instrument acquires a new “seed” reading and the hold process continues.
•Channel closure — When stepping or scanning, the last device action is to open the
previous channel (if closed) and close the next channel. Using the hold feature provides
an auto settling time for the scanner relays. Each open/close transition will restart the
hold process and a reading for each channel will not occur until the relay settles.
Output trigger
After the device action, an output trigger occurs and is available at the rear panel Trigger Link
connector. This trigger can be used to trigger another instrument to perform an operation (e.g.,
select the next channel for an external scan).
Counters
The trigger model for stepping and scanning contains additional blocks for counting samples
(the number of channels to scan) and counting triggers. These counters are explained later in this
section.
Reading hold (autosettle)
When a hold reading is acquired, an audible beep is sounded (if enabled) and the reading is
considered a “true measurement”. The reading is held on the display until an “out of window”
reading occurs to restart the hold process.
When operating remotely or scanning, the hold process seeks a new “seed” once it has been
satisfied and the reading has been released. When operating from the front panel, the hold
process does not seek a new “seed” until the held condition is removed.
Hold example
Fi
3
R
1.Enable HOLD, select a window percentage, and enter a count.
2.Apply test probes to a signal. Once the signal becomes stable enough to satisfy the hold
condition, the reading is released, and the beeper sounds (if enabled).
3.Remove the hold condition by lifting the probes. Hold will then seek a new “seed”.
External triggering
The EXT TRIG key selects triggering from two external sources: trigger link and the TRIG
key. When EXT TRIG is pressed, the TRIG annunciator lights and dashes are displayed to
indicate the instrument is waiting for an external trigger. From the front panel, press the TRIG
key to trigger a single reading. Pressing the EXT TRIG key again toggles back to continuous
triggers.
The Model 2010 uses two lines of the Trigger Link rear panel connector as External Trigger
(EXT TRIG) input and Voltmeter Complete (VMC) output. The EXT TRIG line allows the
Model 2010 to be triggered by other instruments. The VMC line allows the Model 2010 to
trigger other instruments.
At the factory, line 1 is conf gured as VMC and line 2 as EXT TRIG. (Changing this
configuration is described in the Model 2010 Service Manual.) A connector pinout is shown in
Figure 3-3.
Measurement Options3-11
gure 3-
ear panel pinout
Rear Panel Pinout
5
Pin 2
External
Trigger
Input
6
8
7
34
2
1
Pin 1
Voltmeter
Complete
Output
Pin NumberDescription
1
2
3
4
5
6
7
8
* Either pin 3 or 5 may be configured as an output instead of pin 1.
Either pin 4 or 6 may be configured as an input instead of pin 2.
See the Model 2010 Service Manual for details.
Voltmeter Complete Output
External Trigger Input
No Connection *
No Connection *
No Connection *
No Connection *
Signal Ground
Signal Ground
Fi
4
p
Fi
5
p
3-12Measurement Options
External trigger
The EXT TRIG input requires a falling-edge, TTL-compatible pulse with the specif cations
shown in Figure 3-4. In general, external triggers can be used to control measure operations. For
the Model 2010 to respond to external triggers, the trigger model must be conf gured for it.
gure 3-
Trigger link input
ulse specifications
(EXT TRIG)
gure 3-
Trigger link output
ulse specifications
(VMC)
Triggers on
Leading Edge
TTL High
(2V-5V)
TTL Low
(≤0.8V)
2µs
Minimum
Voltmeter complete
The VMC output provides a TTL-compatible output pulse that can be used to trigger other
instruments. The specifications for this trigger pulse are shown in Figure 3-5. Typically, you
would want the Model 2010 to output a trigger after the settling time of each measurement.
Meter
Complete
TTL High
(3.4V Typical)
TTL Low
(0.25V Typical)
10µs
Minimum
External triggering example
In a typical test system, you may want to close a channel and then measure the DUT
connected to the channel with a multimeter. Such a test system is shown in Figure 3-6, which
uses a Model 2010 to measure ten DUTs switched by a Model 7011 multiplexer card in a
Model 7001/7002 Switch System.
Fi
gure 3-
6
D
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Fi
7
UT test system
DUT
#1
Measurement Options3-13
1
OUTPUT
gure 3-
Trigger link
connections
DUT
#2
2
2000 MULTIMETER
2010 Multimeter
DUT
#10
10
Card 1
7011 MUX Card
The Trigger Link connections for this test system are shown in Figure 3-7. Trigger Link of
the Model 2010 is connected to Trigger Link (either IN or OUT) of the Model 7001/7002. Note
that with the default trigger settings on the Model 7001/7002, line #1 is an input and line #2 is
an output. This complements the trigger lines on the Model 2010.
7001 or 7002 Switch System
MADE IN USA
OUT
Trigger
Link
IN
350V
PEAK
!
Trigger
Link Cable
(8501)
SENSE
Ω 4W
2010 Multimeter
HI
1000V
PEAK
!
LO
500V
PEAK
INPUT
MADE IN
U.S.A.
IEEE-488
(CHANGE IEEE ADDRESS
RS232
120
Trigger
Link
FROM FRONT PANEL)
LINE RATING
50, 60
400HZ
17 VA MAX
TRIGGER
LINK
!
3 5
1
VMC
4 6
2
EXT TRIG
!
FUSE LINE
250mAT
100 VAC
(SB)
120 VAC
220 VAC
125mAT
240 VAC
(SB)
For this example, the Model 2010 and 7001/7002 are conf gured as follows:
Model 2010:
Factory defaults restored (accessed from SHIFT-SETUP)
External scanning, channels 1 - 10, no timer, 10 readings (accessed from SHIFT-CONFIG)
External triggers (accessed from EXT TRIG)
Model 7001 or 7002:
Factory defaults restored
Scan list = 1!1-1!10,
Number of scans = 1
Channel spacing = TrigLink
Fi
8
3-14Measurement Options
To run the test and store readings in the Model 2010 with the unit set for external triggers,
press STEP or SCAN. The Model 2010 waits (with the asterisk annunciator lit) for an external
trigger from the Model 7001/7002.
Press STEP on the Model 7001/7002 to take it out of idle and start the scan. The scanner's
output pulse triggers the Model 2010 to take a reading, store it, and send a trigger pulse. The
following explanation on operation is referenced to the operation model shown in Figure 3-8.
gure 3-
Operation model for
triggering example
7001or 7002
Press STEP to start scan
Idle
Bypass
C
D
No
B
Wait for
Trigger Link
Trigger
Scan
Channel
Output
Trigger
Scanned
10
Channels
?
Yes
Trigger
Trigger
2010
Idle
A
Wait for
Trigger Link
Trigger
Make
Measurement
Output
Trigger
Made
10
Measurements
?
Yes
E
F
No
A
Pressing EXT TRIG then STEP or SCAN on the multimeter places it at point A in the
fl wchart, where it is waiting for an external trigger.
B
Pressing STEP takes the Model 7001/7002 out of the idle state and places operation at
point B in the f owchart.
C
For the first pass through the model, the scanner does not wait at point B for a trigger.
Instead, it closes the f rst channel.
D
After the relay settles, the Model 7001/7002 outputs a Channel Ready pulse. Since the
instrument is programmed to scan ten channels, operation loops back up to point B,
where it waits for an input trigger.
Fi
9
D
Measurement Options3-15
EF
and The Model 2010 operation is at point A waiting for a trigger. The output
Channel Ready pulse from the Model 7001/7002 triggers the multimeter to measure
DUT #1 (point E). After the measurement is complete, the Model 2010 outputs a
completion pulse (point F) and then loops back to point A, where it waits for another
input trigger.
The trigger applied to the Model 7001/7002 from the Model 2010 closes the next channel in
the scan. This triggers the multimeter to measure the next DUT. The process continues until all
ten channels are scanned and measured.
External triggering with BNC connections
An adapter cable is available to connect the micro-DIN Trigger Link of the Model 2010 to
instruments with BNC trigger connections. The Model 8503 DIN to BNC Trigger Cable has a
micro-DIN connector at one end and two BNC connectors at the other end. The BNC cables are
labeled VMC (trigger line 1) and EXT TRIG (trigger line 2).
Figure 3-9 shows how a Keithley Model 706 Scanner can be connected to the Trigger Link
of the Model 2010 using the adapter cable. With this adapter, a Model 706 could be substituted
for the Model 7001/7002 in the previous example. With the Model 706 set for External
Triggering, the test would start when the single scan mode is selected and initiated.
If the Model 2010 trigger line conf guration has been changed from the factory setting, the
Model 8502 Trigger Link Adapter must be used to interface with instruments having BNC
trigger connections. It has two micro-DIN connectors and six BNC connectors, one for each
trigger line.
gure 3-
Model 8503 DIN to BNC Trigger Cable
IN to BNC trigger
cable
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
HI
1000V
350V
PEAK
!
PEAK
LO
500V
1 3 5
INPUT
SENSE
PEAK
W 4W
!
2
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
2010 Multimeter
MADE IN
KEITHLEY
U.S.A.
IEEE-488
(CHANGE IEEE ADDRESS
TRIGGER
LINK
!
VMC
4 6
EXT TRIG
!
FUSE LINE
250mAT
100 VAC
(SB)
120 VAC
125mAT
220 VAC
(SB)
240 VAC
FROM FRONT PANEL)
RS232
LINE RATING
50, 60
400HZ
17 VA MAX
Channel
Ready
External
Trigger
706 Scanner
3-16Measurement Options
Buffer operations
The Model 2010 has a buffer to store from two to 1024 readings and units. It also stores the
channel number for scanned readings and overfl w readings. In addition, recalled data includes
statistical information, such as minimum, maximum, average, and standard deviation.
NOTEStatistics are not calculated when an overflow reading has been stored in the buffer.
The buffer fills with the requested number of readings and stops. Readings are placed in the
buffer after any math operations are performed. Buffered data is overwritten each time the
storage operation is selected. The data is volatile; it is not saved through a power cycle.
The following paragraphs discuss storing and recalling buffered data.
Storing readings
Use the following procedure to store readings:
1.Set up the instrument for the desired conf guration.
2.Press the STORE key.
3.Use the , , ▲, and ▼ keys to select the number of readings desired.
4.Press ENTER. The asterisk (*) annunciator turns on to indicate a data storage operation.
It will turn off when the storage is f nished.
Recalling readings
Use the following steps to view stored readings and buffer statistics:
1.Press RECALL. The BUFFER annunciator indicates that stored readings are being
displayed. The arrow annunciator indicates that more data can be viewed with the ,
, ▲, and ▼ keys.
2.As shown in Figure 3-10, use the cursor keys to navigate through the reading numbers,
reading values, and statistics. For any of the buffer statistics (maximum, minimum,
average, standard deviation), the STAT annunciator is on.
3.Use the EXIT key to return to the normal display.
Fi
gure 3-
10
B
uffer locations
Buffer statistics
Measurement Options3-17
RDGNO.10Reading Value
RDGNO.9Reading Value
RDGNO.8Reading Value
RDGNO.7Reading Value
RDGNO.6Reading Value
RDGNO.5Reading Value
RDGNO.4Reading Value
RDGNO.3Reading Value
RANGE
RANGE
RDGNO.2Reading Value
RDGNO.1Reading Value
STDDEVStandard Deviation Value
AverageAverage Value
MinAtXXMinimum Value
MaxAtXXMaximum Value
The MAX AT and MIN AT values are the maximum and minimum values in the buffer. The
AVERAGE value is the mean of the buffered readings. The equation used to calculate the mean
is:
n
X
∑
i
i1=
-----------------=
y
n
where: Xi is a stored reading.
n is the number of stored readings.
The STD DEV value is the standard deviation of the buffered readings. The equation used to
calculate the standard deviation is:
NOTEThe Model 2010 uses IEEE-754 floating point format for math calculations.
3-18Measurement Options
Limit operations
Limit operations set and control the values that determine the HI/IN/LO status of subsequent
measurements. Limits can be applied to all measurement functions except continuity. The limit
test is performed after mX+b and percent math operations. Unit pref xes are applied before the
limit test, for example:
•Low limit = -1.0, High limit = 1.0
A 150mV reading equals 0.15V (IN).
•Low limit = -1.0, High limit = 1.0
A 0.6kΩ reading equals 600Ω (HI).
You can configure the multimeter to beep when readings are inside or outside of the limit
range.
Setting limit values
Use the following steps to enter high and low limit values:
1.Press the SHIFT-LIMITS keys to view the present HI1 limit value:
HI1:+1.000000
This value represents the absolute value of that function.
2.Use the or keys to move to the number f eld. Use the , , ▲, and ▼ keys to
enter the desired value. Move the cursor to the rightmost position (^) and use the ▲ and ▼ keys to move the decimal point.
3.Press ENTER to view the present LO1 limit value:
LO1:-1.000000
This value represents the absolute value of that function.
4.Enter the desired value for this low limit.
5.Press ENTER to view the present HI2 limits value:
HI2: +2.000000^
This value represents the absolute value of that function.
6.Enter the desired value for this high limit.
7.Press ENTER to view the present LO2 limit value:
LO2: -2.000000^
This value represents the absolute value of that function.
8.Enter the desired value for the low limit. Pressing ENTER returns to the normal display.
^
^
Enabling limits
Fi
11
Use the following procedure to turn on the limits operation:
1.Press the SHIFT-ON/OFF keys to view the present beeper status:
BEEP: NEVER
2.Use the ▲ and ▼ keys to change the beeper status (NEVER, OUTSIDE, INSIDE). Press
ENTER when f nished.
When the multimeter returns to the normal display, the HI/IN/LO status is displayed along
with the reading. To disable limit testing, press SHIFT-ON/OFF again. An example of using
limits to sort resistors is shown in Figure 3-11.
Measurement Options3-19
gure 3-
Using limits test to
sort 100Ω, 10%
resistors
LOINHI
90Ω
LO Limit
110Ω
HI Limit
The CALC3:LIMit2 subsystem has all the same commands available as the CALC3:LIMit[1]
subsystem. From the front panel, the same menu is used to control the beeping state and
conditions (inside or outside) for both limits. Since there is only one beeper, there are two
distinct tones used for the two limits, but limit set 1 will take precedence.
Example: Power up with default limits (HLIM1 = +1, LLIM1 = -1,HLIM2 = +2, LLIM2 = -2).
Set the beeper to beep inside. Then, apply 0.9 volts. The beep will be higher in pitch. When the
voltage is increased past 1V, the input is no longer inside limit set 1 but is still inside limit set 2.
At that point, the tone of the beep will change, indicating that you are still inside limit set 2.
NOTE:Limit 1 takes priority over Limit 2 when beeper is set to outside. No change in tone
will be detected.
3-20Measurement Options
Scan operations
The Model 2010 can be used with an internal scanner card (Model 2000 SCAN or
2001-TCSCAN) or with external scanner cards installed in switching mainframes such as the
Models 707, 7001, and 7002. The following paragraphs discuss various aspects of using
scanning with the Model 2010.
Scanning overview
A scanner lets you switch among a number of input signals to the Model 2010 for
measurement. The channel control and scanning capabilities depend on whether an internal or
external card is being used, as well as on the capabilities of the scanner card. Refer to the
documentation supplied with the scanner card for specif c connection information.
Using an internal scanner card
The optional Model 2000-SCAN scanner card lets you step through or scan up to ten
two-pole channels or f ve four-pole channels.
The optional Model 2001-TCSCAN Thermocouple/General Purpose Scanner Card lets you
multiplex one of nine two-pole or one of four four-pole analog signals into the Model 2010, and/
or any combination of two- or four-pole analog signals.
Using external scanner cards
When using external channels, the switching mainframe controls the opening and closing of
individual channels. To synchronize Model 2010 measurements with external channel closures,
connect the Trigger Link lines of the multimeter and switching mainframe. Refer to “Trigger
operations” earlier in this section for details and an example on using external triggering.
Front panel scanner controls
In addition to the trigger keys discussed previously, front panel keys that affect scanner card
operation include:
• and — Lets you manually step through consecutive internal card channels.
•OPEN and CLOSE — Let you selectively open and close internal card channels.
•SHIFT-CONFIG — Selects internal or external scanning, scan list, time between scans,
and reading count.
•STEP — Starts a stepping operation of consecutive channels, where output triggers are
sent after every channel closure.
•SCAN — Starts a scanning operation of consecutive channels, where an output trigger
is sent at the end of the scan list.
•SHIFT-HALT — Stops stepping or scanning and restores the trigger model to a nonscanning mode.
Using the and keys
The and keys can be used to manually scan through channels on the internal scanner
card. With a scanner card installed in the option slot, press the key to manually increment
channels or the key to manually decrement channels. The annunciator of the closed channel
is lit. Hold down either key to manually scan through channels continuously. Press OPEN to
open all channels.
Using OPEN and CLOSE keys
The OPEN and CLOSE keys control channels on the internal scanner card only. The keys
allow you to directly:
•Close a specif c channel (or channel pair for four-wire resistance).
•Immediately open any internal closed channel (or channel pair for four-wire resistance).
With a scanner card installed in the option slot of the Model 2010, the following prompt is
displayed when the CLOSE key is pressed:
CLOSE CHAN:01
Use the , , ▲, and ▼ keys to display the desired channel (1 to 10) and press ENTER.
The annunciator of the closed channel will be displayed on the front panel along with normal
readings. Selecting a different channel from the one that is presently closed will cause the closed
channel to open and allow a settling time before closing the selected channel.
Channel relays will be closed according to the presently selected function. If a four-wire
function is selected, both the selected channel relay and the matching relay pair will be closed.
Fixed four-pole relay pairs are:
•1 and 6 (not available for Model 2001-TCSCAN)
•2 and 7
•3 and 8
•4 and 9
•5 and 10
Pressing the OPEN key will immediately open any closed scanner card channel or channel
pair for four-wire resistance.
Measurement Options3-21
Fi
12
F
3-22Measurement Options
Stepping and scanning trigger model additions
The trigger model presented in “Trigger operations” earlier in this section has some
additional capabilities when stepping or scanning. These are outlined below:
•Timer — With this control source, event detection is immediately satisf ed on the initial
pass. Each subsequent detection is satisf ed when the programmed timer interval (up to
99H:99M:99.99S) elapses.
•Reading counter — For both stepping and scanning, the reading count can be entered
from SHIFT-CONFIG. (This is referred to as the trigger counter over the bus.) The
reading counter can bypass the idle state. Operation will wait until the programmed
control source event occurs.
•Channel counter — For scanning, the scan list length (maximum channel less minimum
channel) is used to bypass the control source allowing a specif ed number of device
actions to occur. (This counter is referred to as the sample counter over the bus.)
These additional blocks are shown in the trigger models of Figure 3-12 and Figure 3-13. Uses
of the timer control source, reading counter, and channel counter are shown in the scanning
examples later in this section.
gure 3-
ront panel triggering
with stepping
Control
Source
Immediate
External
Timer
Idle
Event
Detection
Delay
Device
Action
Yes
No
More
Readings
?
Output
Trigger
Reading
Count
(Trigger Counter)
Fi
13
F
Measurement Options3-23
gure 3-
ront panel triggering
with scanning
Using SHIFT-CONFIG to configure stepping and scanning
Control
Source
Immediate
External
Timer
Idle
Yes
Event
Detection
Yes
Delay
Device
Action
No
More
Readings
?
Output
Trigger
No
More
Channels
?
Reading
Count
(Trigger Counter)
Scan List
Length
(Sample Counter)
Using the SHIFT-CONFIG key combination, you can select internal or external scanning, the
minimum and maximum channels in the scan list, the time between scans, and the reading count.
To configure stepping or scanning, perform the foll wing:
1.Select the desired measurement function.
2.Press the SHIFT-CONFIG keys to access the step/scan conf guration.
3.Select the type of scan (INTernal or EXTernal) by using the ▲ and ▼ keys and pressing
ENTER.
4.Select the f rst channel in the scan list (MINimum CHANnel) by using the , , ▲,
and ▼ keys and pressing ENTER.
5.Select the last channel in the scan list (MAXimum CHANnel) and press ENTER to
confirm
6.The next selection is for timed scans. This is the Timer control source in the trigger
model. It sets a user-specified interval for starting scans. If you choose timed scans, the
Model 2010 prompts for a time interval:
00H:00M:00.000S
Use the , , ▲, and ▼ keys to select a time interval and press ENTER to conf rm.
Fi
14
I
3-24Measurement Options
7.Next, you are prompted for a reading count (RDG CNT). This can be less than, equal to,
or greater than the scan list length (up to 1024). It is the number of readings that will be
stored in the buffer. The effects of these choices are further described in the scanning
examples.
8.Press ENTER when f nished to return to the normal display. Note that scanned readings
are always stored in the buffer, up to the setting for RDG CNT.
Scanning examples
The following examples demonstrate the use of reading count, timed scans, delay, and
external scanning.
Counters
One of the conf guration options for stepping and scanning is the reading count. The example
in Figure 3-14 shows how different settings of RDG CNT affect these operations.
gure 3-
nternal scanning
example with reading
count option
STEP
10 channel closures
10 output triggers
SCAN
10 channel closures
1 output triggers
RECALL
10 Readings
SHIFT-CONFIG
TYPE: INT
MIN CHAN: 1
MAX CHAN: 10
TIMER? OFF
00100002
RDG CNT:
0020
STEP
20 channel closures
20 output triggers
SCAN
10 channel closures (x2)
2 output triggers
RECALL
20 Readings
Note: "Factory setup" on the
Model 2010 is assumed.
STEP
2 channel closures
2 output triggers
SCAN
10 channel closures
1 output triggers
RECALL
2 Readings
Measurement Options3-25
•With a reading count (0010) equal to the scan list length (10), a step operation
consecutively closes ten channels and sends an output trigger after each channel. A scan
operation also consecutively closes ten channels but sends an output trigger only at the
end of the scan.
•With a reading count (0020) greater than the scan list length (10), stepping yields 20
channel closures and 20 output triggers. Scanning also goes through the scan list twice
but sends an output trigger only at the end of each scan.
•With a reading count (0002) less than the scan list length (10), stepping yields two
channel closures and output triggers. Scanning goes through the entire scan list and
sends an output trigger but only two readings are stored.
NOTEIf the reading count divided by the scan list length is not an integer, it is rounded up.
For example, if the reading count is 15 and the scan list length is 10, there will be two
output triggers for scanning.
The differences between stepping and scanning counters for bus commands are summarized
in Table 3-3.
Table 3-3
Bus commands parameters for stepping and scanning counters
Operation:SAMPle:COUNt:TRIGger:COUNt
STEP1reading count
SCANscan list length(reading count) / (scan list length)
3-26Measurement Options
Timing
Another conf guration option for stepping and scanning is the timing of channel closures. The
example in Figure 3-15 shows how different settings of TIMER and DELAY affect these
operations. These are the timer control source and the delay block shown in the trigger models
in Figure 3-12 and Figure 3-13.
•With the timer ON and set to f ve seconds and delay set to AUTO, channels are stepped
through at fi e second intervals with an output trigger after each closure. A scan
operation yields ten channels scanned immediately with an output trigger at the end of
the scan.
•With the timer OFF and the delay set to MANual for f ve seconds, stepping and scanning
through the channels is timed the same. The difference is in the number of output
triggers, with stepping sending a trigger after each channel closure and scanning sending
a trigger at the end of the scan.
When using both the timer and delay parameters, the timer is not started until after the delay.
For example, if the timer is two minutes and the delay is ten seconds, the timer is not started until
ten seconds after pressing SCAN. Each successive scan will occur at 2:10.0, 4:10.0, etc.
If the total delay time per scan is greater than or equal to the timer setting, the timer condition
is already satisf ed and is ignored.
Fi
15
I
Measurement Options3-27
gure 3-
nternal scanning
example with timer and
delay options
SHIFT-CONFIG
TYPE:INT
MIN CHAN: 1
MAX CHAN: 10
TIMER?
TIMER? ON
00H:00M:05.000S
RDG CNT: 0010
SCAN
10 channel closures
1 output trigger
STEP
10 channel closures
at 5-second intervals
10 output triggers
RECALL
10 readings
OFF
ON
Note: "Factory setup" on the
Model 2010 is assumed.
RDG CNT: 0010
DELAY: MAN
00H:00M:05.000S
STEP
10 channel closures
at 5-second intervals
10 output triggers
SCAN
10 channel closures
at 5-second intervals
1 output trigger
RECALLL
10 readings
3-28Measurement Options
External scanning
The example in Figure 3-16 shows the front panel operations to conf gure an external scan.
The trigger and signal connections were shown previously in “Trigger operations”. Both
instrument setups assume factory defaults. Set the Model 2010 for the desired measurement
function.
1
On the Model 7001 Switch System, enter a scan list of channels 1 to 10 on card 1.
2
Also on the Model 7001, conf gure the instrument for Trigger Link triggers and one
scan of ten channels.
3
On the Model 2010 Multimeter, configure an xternal scan of the f rst ten channels.
4
Set the Model 2010 for external triggers by pressing EXT TRIG. The display will be
dashes.
5
Press STEP or SCAN on the Model 2010. The asterisk and STEP or SCAN
annunciator will light.
6
Press STEP on the Model 7001 to start channel closures.
7
After the scan, you can recall ten readings from the Model 2010 buffer.
NOTEWhen using an external thermocouple scanner card and channel 1 as a reference, the
Model 2010 only recognizes channel 1 when a step or scan is performed. If using a
Model 7001 or 7002 to close channel 1 manually, the Model 2010 will not interpret
that channel as the reference junction without a step or scan operation.
SHIFT-CONFIG
TYPE:EXT
MIN CHAN: 001
MAX CHAN: 010
TIMER? OFF
RDG CNT: 0010
ENTER
4
EX TRIG
5
STEP or SCAN
7
RECALL (10 readings)
, , ,
EXIT
3-30Measurement Options
System operations
The Model 2010 has other front panel operations. Saving and restoring setup information is
described in Section 2. Selecting the remote interface and language is covered in Section 4.
Self-test
The TEST selections are used as diagnostic tools to isolate problems within the Model 2010.
Information on using these test procedures is included in the Model 2010 Service Manual.
Calibration
The CAL selections are used to view the calibration date and next due date, to perform
calibration, and to view the number of times calibration has been performed. Some of the items
are password-protected to prevent unintended changing of calibration constants.
To view the calibration dates, press SHIFT-CAL. Press ENTER at the DATES prompt. The
first date is the last time calibration was performed. The NDUE date is the calibration due date.
Running calibration is password-protected. Refer to the Model 2010 Service Manual for
details.
To view the calibration count, press ENTER at the COUNT prompt.
4
Remote Oper-
ation
4
Remote
Operation
4-2Remote Operation
Introduction
This section includes the following information:
•Selecting an interface
•Selecting a language
•RS-232 operation
•GPIB bus operation and reference
•Status structure
•Trigger model (GPIB operation)
•Programming syntax
•Common commands
Selecting an interface
The Model 2010 multimeter supports two built-in remote interfaces:
•GPIB bus
•RS-232 interface
You can use only one interface at a time. The factory interface selection is the GPIB bus. You
can select the interface only from the front panel. The interface selection is stored in non-volatile
memory; it does not change when power has been off or after a remote interface reset.
RS-232
Before you select a remote interface, consider the programming language you want to use.
NOTEChanging the interface (GPIB to RS-232 or RS-232 to GPIB) will clear the data
buffer.
You can connect a controller to the RS-232 interface. Some considerations for selecting the
RS-232 interface are:
•You must define the baud rate, enable or disable software handshake XON/XOF.
•You can only use the SCPI programming language with the RS-232 interface.
To select RS-232 as the remote interface, perform the following:
1.Access the RS-232 conf guration by pressing SHIFT then RS232.
You see: RS232: OFF
2.Move to the on/off selection by pressing the key.
You see the OFF selection blinking.
3.Turn on the RS-232 interface by toggling the selection to ON using the ▼ or ▲ key and
press ENTER.
You can exit the configuration menu by pressing EXIT.
GPIB bus
The GPIB bus is the IEEE-488 interface. You must select a unique address for the Model 2010
multimeter. The address is displayed when the multimeter is turned on. At the factory, the
address is set to 16.
Since GPIB is the interface selection def ned by the factory, only follow these steps to select
the GPIB interface if you have been previously using the RS-232 remote programming interface:
1.Select the GPIB option by pressing SHIFT then GPIB.
You see: GPIB: OFF.
2.Move to the on/off selection by pressing the key.
You see the OFF selection blinking.
3.Turn on the GPIB interface by toggling the selection to ON using the ▼ or ▲ key and
press ENTER.
Turning off the RS-232 interface automatically selects GPIB as the remote programming
interface.
Selecting a language
Choose one of the following languages to program the Model 2010 multimeter:
Remote Operation4-3
•SCPI (Signal Oriented Measurement Commands)
•Keithley Models 196/199 Digital Multimeter
NOTEThe 196/199 language is intended for use only over the IEEE-488 bus. Using front
panel controls with this language may cause erratic operation. In this case, results
cannot be guaranteed.
The factory sets the language selection as SCPI.
You only can select a programming language from the front panel. The language selection is
stored in non-volatile memory, which means it does not change when power has been off or after
a remote interface reset.
Table 4-1 shows the languages supported by the two available interfaces:
Table 4-1
Language support
LanguageGPIBRS-232
SCPI
Keithley Models 196/199
The language you select determines the remote operations allowed.
Ye s
Ye s
Ye s
No
4-4Remote Operation
To select a programming language, follow these steps:
1.Access the GPIB conf guration options by pressing SHIFT then GPIB.
2.Select the language conf guration option by pressing the ENTER key twice.
3.Move to the language selection f eld by pressing the key.
4.Select the programming language you want by pressing the ▼ or ▲ key until you see the
5.Confirm your selection by pressing ENTER. The multimeter returns to the measurement
SCPI
Standard Commands for Programmable Instruments (SCPI) is fully supported by the GPIB
and RS-232 interfaces. Always calibrate the Model 2010 Multimeter using the SCPI language.
You see GPIB:ON with GPIB blinking.
You see: LANG:<name>.
appropriate language.
The menu scrolls through SCPI and 199/6 (Keithley Models 196/199).
mode.
Keithley Models 196/199 Digital Multimeter
The Model 2010 Multimeter implements virtually all commands available in the Keithley
Models 196/199 Digital Multimeter, except for the self-test and calibration commands. The
commands are listed in Appendix D.
See the Models 196/199 Digital Multimeter User’s Manuals for more information about
remote programming.
RS-232 operation
Sending and receiving data
The RS-232 interface transfers data using eight data bits, one stop bit, and no parity. Make
sure the controller you connect to the multimeter also uses these settings.
You can break data transmissions by sending a ^C or ^X character string to the multimeter.
This clears any pending operation and discards any pending output.
Selecting baud rate
The baud rate is the rate at which the Model 2010 Multimeter and the programming terminal
communicate. Choose one of the following available rates:
•19.2k
•9600
•4800
•2400
•1200
•600
•300
Remote Operation4-5
The factory selected baud rate is 9600.
Make sure that the programming terminal that you are connecting to the Model 2010
Multimeter can support the baud rate you selected. Both the multimeter and the other device
must be conf gured for the same baud rate. To select a baud rate, follow these steps:
1.Access the RS-232 conf guration by pressing SHIFT then RS232.
You see: RS232: ON (assuming you have already selected the RS-232 interface).
2.Go to the baud rate f eld by pressing the ▼ key.
You see BAUD:<rate>.
3.Access the baud rate list by pressing the key. You see the rate selection blinking.
4.Scroll through the available rates by pressing the ▼ and ▲ key until you findthe rate you
want.
5.Confirm your selection by pressing ENTER. The multimeter prompts you to def ne
signal handshaking. Continue for information about handshaking. You can return to
measurement mode by pressing EXIT.
4-6Remote Operation
Selecting signal handshaking (flow control)
Signal handshaking between the controller and the instrument allows the two devices to
communicate to each other regarding being ready or not ready to receive data. The Model 2010
does not support hardware handshaking (f ow control).
Software fl w control is in the form of X__ON and X__OFF characters and is enabled when
XonXoFF is selected from the RS232 FLOW menu. When the input queue of the Model 2010
becomes more than 3/4 full, the instrument issues an X_OFF command. The control program
should respond to this and stop sending characters until the Model 2010 issues the X_ON, which
it will do once its input buffer has dropped below half-full. The Model 2010 recognizes X_ON
and X_OFF sent from the controller. An X_OFF will cause the Model 2010 to stop outputting
characters until it sees an X_ON. Incoming commands are processed after the <CR> character
is received from the controller.
If NONE is the selected f ow control, then there will be no signal handshaking between the
controller and the Model 2010. Data will be lost if transmitted before the receiving device is
ready.
Perform the following steps to set f ow control:
1.Access the RS-232 conf guration by pressing SHIFT and then RS232. You see: RS 232:
2.Go to the fl w control field by using the ▲ or ▼ key. You see FLOW: <control>.
3.Access the f ow control options by pressing the key. You see the fl w control
4.Use the ▲ or ▼ key to display the desired f ow control (NONE or XonXoFF) and press
ON (assuming you have already selected the RS-232 interface).
selection blinking.
ENTER. You will then be prompted to set the terminator. Continue for information about
the terminator. You can return to the measurement mode by pressing EXIT.
Setting terminator
The Model 2010 can be conf gured to terminate each program message that it transmits to the
controller with any combination of <CR> and <LF>. Perform the following steps to set the
terminator:
1.Access the RS-232 conf guration by pressing SHIFT and then RS232.
You see: RS 232: ON (assuming you have already selected the RS-232 interface).
2.Go to the terminator f eld by using the ▲ or ▼ key.
You see TX TERM: <terminator>.
3.Access the terminator options by pressing the key.
You see the terminator selection blinking.
4.Use the ▲ or ▼ key to display the desired terminator (LF, CR, CRLF, or LFCR) and
press ENTER. The instrument will return to the measurement mode.
RS-232 connections
Fi
1
R
The RS-232 serial port can be connected to the serial port of a controller (i.e., personal
computer) using a straight through RS-232 cable terminated with DB-9 connectors. Do not use a null modem cable. The serial port uses the transmit (TXD), receive (RXD), and signal ground
(GND) lines of the RS-232 standard. It does not use the hardware handshaking lines CTS and
RTS. Figure 4-1 shows the rear panel connector for the RS-232 interface, and Table 4-2 shows
the pinout for the connector.
If your computer uses a DB-25 connector for the RS-232 interface, you will need a cable or
adapter with a DB-25 connector on one end and a DB-9 connector on the other, wired straight
through (not null modem).
Remote Operation4-7
gure 4-
S-232 interface
connector
54321
9876
RS232
Rear Panel Connector
Table 4-2
RS-232 connector pinout
Pin numberDescription
1
2
3
4
5
6
7
8
9
1
CTS and RTS signals are not used.
no connection
TXD, transmit data
RXD, receive data
no connection
GND, signal ground
no connection
CTS, clear to send
RTS, ready to send
no connection
1
1
Error messages
See Appendix B for RS-232 error messages.
Fi
2
I
4-8Remote Operation
GPIB bus operation and reference
Introduction
The following paragraphs contain information about connecting to and using the GPIB
(IEEE-488) bus.
GPIB bus standards
The GPIB bus is the IEEE-488 instrumentation data bus with hardware and programming
standards originally adopted by the IEEE (Institute of Electrical and Electronic Engineers) in
1975. The Model 2010 multimeter conforms to these standards:
•IEEE-488-1987.1
•IEEE-488-1987.2
This standard def nes a syntax for sending data to and from instruments, how an instrument
interprets this data, what registers should exist to record the state of the instrument, and a group
of common commands.
•SCPI 1991 (Standard Commands for Programmable Instruments)
GPIB bus connections
gure 4-
EEE-488 connector
This standard def nes a command language protocol. It goes one step farther than
IEEE-488-1987.2 and def nes a standard set of commands to control every programmable aspect
of an instrument.
To connect the Model 2010 Multimeter to the GPIB bus, use a cable equipped with standard
IEEE-488 connectors as shown inFigure 4-2.
To allow many parallel connections to one instrument, stack the connector. Two screws are
located on each connector to ensure that connections remain secure. Current standards call for
metric threads, which are identif ed with dark-colored screws. Earlier versions had different
screws, which were silver-colored. Do not use these types of connectors on the Model 2010
Multimeter, because it is designed for metric threads.
Figure 4-3 shows a typical connecting scheme for a multi-unit test system.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.