Keithley 7752, 7753, 7751 User Manual

www.keithley.com
Model 7751/7752/7753 Switching Modules
User’s Guide
PA-823 Rev. B / Sept 2003
The Models 7751/7752/7753 can be used with Keithley’s Model 2790.
A GR
EA
TER MEASURE OF CONFIDENCE
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some in­struments and accessories would normally be used with non-haz­ardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recog­nize shock hazards and are familiar with the safety precautions re­quired to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection pro­vided by the product may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use
and maintenance of equipment, for ensuring that the equipment is operated within its specications and operating limits, and for en­suring that operators are adequately trained.
Operators use the product for its intended function. They must be
trained in electrical safety procedures and proper use of the instru­ment. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product
to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are de­scribed in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform
safe installations and repairs of products. Only properly trained ser­vice personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Installation Category I and Installation Category II, as de­scribed in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O sig­nals are Installation Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-volt­ages. Installation Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O con­nections are for connection to Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test xtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect
that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are pre­vented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human con­tact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of
the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When con­necting sources to switching cards, install protective devices to lim­it fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connect­ed to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power dis­connect device must be provided, in close proximity to the equip­ment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jump­ers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the com­mon side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specications and operating instructions or the safety of the equip­ment may be impaired.
Do not exceed the maximum signal levels of the instruments and ac­cessories, as dened in the specications and operating informa­tion, and as shown on the instrument or test xture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against re hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test xture, keep the lid closed while power is ap­plied to the device under test. Safe operation requires the use of a lid interlock.
5/02
If or is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should re­fer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or mea­sure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading in a manual explains dangers that might result in personal injury or death. Always read the associated infor­mation very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and
all test cables.
To maintain protection from electric shock and re, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instru­ments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that se­lected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments ofce for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to in­structions. If the board becomes contaminated and operation is af­fected, the board should be returned to the factory for proper cleaning/servicing.
Keithley Instruments, Inc.
28775 Aurora Road Cleveland, Ohio 44139 (440) 248-0400 www.keithley.com
Introduction
The Models 7751, 7752, and 7753 are used with the Model 2790 SourceMeter Switch System. Safety fea­tures are integrated into the design of the modules to help safeguard the user and the module. With proper use of the module and the Model 2790 mainframe, risk of ignition is minimal.
The 7751, 7752, and 7753 modules have a built-in programmable constant I-source to perform low resis­tance measurements. The 7751 and 7753 modules also have a built-in programmable V-source and an I/V converter. These features allow the 7751 and 7753 modules to be used with the Digital Multimeter (DMM) of the Model 2790 to measure high resistance.
NOTE Details on using the 7751, 7752, and 7753 modules are provided in Sections 1, 2, and 5
of the Model 2790 User’s Manual.
Models 7751/7752/7753 Switching Modules
User’s Guide
WARNING It is the responsibility of the user to ensure that external protection is provided, either
by an inherently safe electrical barrier and/or a safety barrier around the DUT or air­bag, to prevent injury in case of detonation. The user must take additional precautions when working with hazardous voltages. A shock hazard exists when >42V is present in the system.
WARNING DO NOT apply external sources to the 7751, 7752, or 7753 module. Only the sources
provided in the modules should be used for the test system.
Switching matrix
The switching matrix provides four banks of input for the DUT. Each bank provides inputs for 4-wire con­nections. Each bank also has a built-in short available for HIPOT testing. The switching scheme of the matrix provides the versatility needed to use the I-source, V-source (7751 and 7753), and the DMM of the Model 2790.
NOTE The schematic diagram of the 7751/7752/7753 module is provided in Section 2 of the
User’s Manual.
I-source
The programmable I-source of the 7751, 7752, and 7753 modules can be set from 0 to 50mA. The I-source is used with the DMM of the 2790 to measure low resistance (i.e., bridgewires and shunt bars). The I-source has a dry circuit clamp that can be used to measure oxide build-up on the contacts of a shunt bar. The dry circuit clamp limits the test voltage at 20mV to prevent oxide piercing during the measurement.
PA-823 Rev. B / 9-03