Keithley 7752, 7753, 7751 User Manual

www.keithley.com
Model 7751/7752/7753 Switching Modules
User’s Guide
PA-823 Rev. B / Sept 2003
The Models 7751/7752/7753 can be used with Keithley’s Model 2790.
A GR
EA
TER MEASURE OF CONFIDENCE
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some in­struments and accessories would normally be used with non-haz­ardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recog­nize shock hazards and are familiar with the safety precautions re­quired to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection pro­vided by the product may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use
and maintenance of equipment, for ensuring that the equipment is operated within its specications and operating limits, and for en­suring that operators are adequately trained.
Operators use the product for its intended function. They must be
trained in electrical safety procedures and proper use of the instru­ment. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product
to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are de­scribed in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform
safe installations and repairs of products. Only properly trained ser­vice personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Installation Category I and Installation Category II, as de­scribed in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O sig­nals are Installation Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-volt­ages. Installation Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O con­nections are for connection to Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test xtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect
that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are pre­vented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human con­tact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of
the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When con­necting sources to switching cards, install protective devices to lim­it fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connect­ed to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power dis­connect device must be provided, in close proximity to the equip­ment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jump­ers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the com­mon side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specications and operating instructions or the safety of the equip­ment may be impaired.
Do not exceed the maximum signal levels of the instruments and ac­cessories, as dened in the specications and operating informa­tion, and as shown on the instrument or test xture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against re hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test xture, keep the lid closed while power is ap­plied to the device under test. Safe operation requires the use of a lid interlock.
5/02
If or is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should re­fer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or mea­sure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading in a manual explains dangers that might result in personal injury or death. Always read the associated infor­mation very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and
all test cables.
To maintain protection from electric shock and re, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instru­ments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that se­lected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments ofce for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to in­structions. If the board becomes contaminated and operation is af­fected, the board should be returned to the factory for proper cleaning/servicing.
Keithley Instruments, Inc.
28775 Aurora Road Cleveland, Ohio 44139 (440) 248-0400 www.keithley.com
Introduction
The Models 7751, 7752, and 7753 are used with the Model 2790 SourceMeter Switch System. Safety fea­tures are integrated into the design of the modules to help safeguard the user and the module. With proper use of the module and the Model 2790 mainframe, risk of ignition is minimal.
The 7751, 7752, and 7753 modules have a built-in programmable constant I-source to perform low resis­tance measurements. The 7751 and 7753 modules also have a built-in programmable V-source and an I/V converter. These features allow the 7751 and 7753 modules to be used with the Digital Multimeter (DMM) of the Model 2790 to measure high resistance.
NOTE Details on using the 7751, 7752, and 7753 modules are provided in Sections 1, 2, and 5
of the Model 2790 User’s Manual.
Models 7751/7752/7753 Switching Modules
User’s Guide
WARNING It is the responsibility of the user to ensure that external protection is provided, either
by an inherently safe electrical barrier and/or a safety barrier around the DUT or air­bag, to prevent injury in case of detonation. The user must take additional precautions when working with hazardous voltages. A shock hazard exists when >42V is present in the system.
WARNING DO NOT apply external sources to the 7751, 7752, or 7753 module. Only the sources
provided in the modules should be used for the test system.
Switching matrix
The switching matrix provides four banks of input for the DUT. Each bank provides inputs for 4-wire con­nections. Each bank also has a built-in short available for HIPOT testing. The switching scheme of the matrix provides the versatility needed to use the I-source, V-source (7751 and 7753), and the DMM of the Model 2790.
NOTE The schematic diagram of the 7751/7752/7753 module is provided in Section 2 of the
User’s Manual.
I-source
The programmable I-source of the 7751, 7752, and 7753 modules can be set from 0 to 50mA. The I-source is used with the DMM of the 2790 to measure low resistance (i.e., bridgewires and shunt bars). The I-source has a dry circuit clamp that can be used to measure oxide build-up on the contacts of a shunt bar. The dry circuit clamp limits the test voltage at 20mV to prevent oxide piercing during the measurement.
PA-823 Rev. B / 9-03
V-source and I/V converter (7751/7753)
The programmable V-source of the 7751 and 7753 modules can be set from 50 to 500V. The maximum out­put current for the 7751 V-source is 50µA for rated accuracy, while the 7753 V-source maximum output cur­rent is 500µA for rated accuracy. When the V-source and I/V converter are used with the 2790 DMM, high resistance measurements can be performed to test the insulation resistance of an inflator over the following ranges:
Model 7751: 1M to 1G
Model 7753: 0.1M to 1G
The V-source can also be used as an independent voltage source. It is typically used with the 7702 module to test multiple DUTs.
NOTE Details on using the V-source with the 7702 module are provided in Section 3 of the
Model 2790 User’s Manual.
Module wiring and installation
The 7751, 7752, and 7753 modules use screw terminals for DUT connections. The screw terminals are accessed by removing the top cover of the module. The quick-disconnect terminal blocks can be pulled off the PC-board to make the connections.
WARNING When connecting the DUT and before installing (or removing) the module in the main-
frame, make sure the Model 2790 is turned off, the line cord is disconnected, and there are no external sources connected to the module or the mainframe.
CAUTION Be sure to install only 7751/7753 module covers with the integral heat sink. Also be sure
that the thermal conductive pad, located on the voltage source (Figure 1), has not been dislodged or damaged. These thermal components are necessary to assure proper volt­age source operation. (Note that earlier 7751 modules do not have the thermal pad or heat sink.)
General wiring requirements
Cable wires should be mechanically durable, preferably 20AWG or larger with a high voltage insulation, such as rubber or silicon, that are rated for at least 1000V.
NOTE Details on making connections to the 7751/7752/7753 module and installing the module
in the mainframe are provided in Section 4 of the Model 2790 User’s Manual.
2
Figure 1 Model 7751/7753 thermal conductive pad location
A) Board location
SOURCE SENSE SRC
GUARD
HI HI
Thermal
Conductive
Pad
I / V
HIGNDGND HIINLOLO
GUARDISOURCEVSOURCE/ INTERLOCK
B) Side view
Thermal Pad
500V Supply (PS1)
Printed Circuit Board
LO LO
OUT I / V
BANK 1
B
SRC
SEN
SEN
A
LO
HI
HI
LO
BANK 2
SRC
B
SRC
SEN
SEN
A
HI
LO
HI
LO
LO
HI
HI
LO
SRC
B
SRC
SEN
SEN
A
BANK 4
LO
HI
HI
LO
LO
SRC
B
SRC
SEN
SEN
A
BANK 3
3
2790 SourceMeter Switch System Specifications
7751/7752/7753 SOURCE/SWITCH MODULE SPECIFICATIONS
2790 Resistance Mode Specifications with Cards
2,3
(Module function accuracy specifications are for 2 years, 23°C, ±5°C)
20
50
100
10
Resistance
Range
10M
100M
1G
1M
10M
100M
Resistance
Range
1M
10M
100M
1G
0.1M
1M
10M
100M
Typ. Open
Ckt.
Voltage
5.5V
5.5V
5.5V
20mV
Short Ckt.
Short Ckt.
Accuracy
%Rdg + Ohms
0.09% + 40m (0.002%+3mΩ)/°C
0.11% + 40m (0.003%+3mΩ)/°C
0.16% + 40m (0.004%+3mΩ)/°C
1.10% + 40m (0.026%+3mΩ)/°C
Max.
Current
<1mA 0.8% 0.03%/°C
<1mA 1.1% 0.05%/°C
<1mA 4.0% 0.12%/°C
<1mA 1.1% 0.04%/°C
<1mA 1.1% 0.06%/°C
<1mA 1.6% 0.13%/°C
Max.
Current
<1mA 0.8% 0.02%/°C
<1mA 0.9% 0.03%/°C
<1mA 1.3% 0.10%/°C
<1mA 6.7% 0.27%/°C
<1mA 1.1% 0.03%/°C
<1mA 1.1% 0.04%/°C
<1mA 1.3% 0.11%/°C
<1mA 4.5% 0.30%/°C
Temperature Coefficient
(0-18°C & 28-40°C)
Accuracy
%Rdg
Accuracy
%Rdg
Temperature
Coefficient
(0-18°C & 28-40°C)
Temperature
Coefficient
(0-18°C & 28-40°C)
Source
Maximum
Current
Resistance
50mA
20mA
10mA
(Dry Circuit Ohms 1mA max with 7751, 7752, or 7753 card)
1mA
(7751 Only)
Source
Voltage
500V
500V
500V
50V
50V
50V
(7753 Only)
Source
Voltage
500V
500V
500V
500V
50V
50V
50V
50V
Current Source Output
Output Level: Programmable 0 to 50mA (Ch. 27). Programming Resolution: 10µA. Output Voltage: 5.5V±10% compliance. Accuracy: 0.06% + 10µA (2 yr Specification). Settling Time: 1mS to 0.1% of final value (typ). Temperature Coefficient: (0-18°C & 28-40°C)
(0.001%+0.25µA)/°C
Dry Circuit Clamp (Ch. 24): 20mV±10%, Isource 1mA
Voltage Source Output (7751/7753 Only)
Output Level: Programmable 50V to 500V (CH. 28). Programming Resolution: 100mV. Output Current: (7751) - 50µA maximum for rated accuracy, <1mA typical
into short circuit. (7753) - 500µA maximum for rated accuracy, <1mA typical into short circuit.
Accuracy: 0.5% + 0.13V (2yr Specification). Settling Time: Rise Time: 50V to 500V step, 0.1% of final value,
Temperature Coefficient: (0-18°C & 28-40°C)
(0.001%+0.005V)/°C
Safety Limit: Current li mited ma ximum c urrent o f 1mA. Cable Discharge (Ch. 20): 100k Shunt Max Capacitance: 1nF
250 ms max. Fall Time: 500V to 50V step, 0.1% of final value, 1000 ms max.
Current Measure Input (7751/7753 Only)
Range: 7751 (0-50µA), 7753 (0-500µA). Accuracy: (7751) 0.5% (of reading) + 6nA (2yr Specification).
Temperature Coefficient: (0-18°C & 28-40°C) (0.02%+0.5nA)/°C Voltage Burden: <1mV.
(7753) 0.5% (of reading) + 60nA (2yr Specification).
Switching Capabilities (Bank 1 – Bank 4)
4 Channels: 1 Form A switch.
8 Channels: 4 4-pole or 8 2-pole signals into DMM or I/V converter.
Contact Check: 4-Wire contact check through internal DMM.
Relay Type: Latching electromechanical.
Actuation Time: <3ms.
Contact Life (typ): >10
Contact Resistance: <1 Ohm at end of contact life.
Contact Potential: <±2µV typical per contact pair, ±3µV max.
Connector Type: Plugable screw terminal, #22 AWG wire size.
Isolation Between any two Terminals
Isolation Between Terminals and Earth
Isolation Between Channel Groups
External Common Mode Voltage: 42V between any terminal and chassis.
6
operations at maximum source level.
8
>10
operations cold switching.
1
: >1GΩ, <100pF.
1
: >1GΩ, <200pF.
1
: >500GΩ, <100pF.
(Connect no external sources.)
7751, 7752, or 7753 Module Notes
1
Isolation for channels 1-12, only one channel closed at a time, or all channels
open.
2
See User's manual for ohm specifications at sources other than those specified.
3
All specifications valid for 1 NPLC ADC aperture setting.
System Throughput
(Connect, source, measure, calculate)
0.01 NPLC, Filter Off, over GPIB Bus
High Ohms (Source V): 13 Rdgs/Sec Low Ohms (Source I): 9 Rdgs/Sec
1 NPLC, Filter On, Over GPIB Bus
High Ohms (Source V): 11 Rdgs/Sec Low Ohms (Source I): 7 Rdgs/Sec
System Throughput Notes
1. Reset upon fixed Vsource level, no settling time.
1
1
HW 11/04/03
Rev. D
2790 SourceMeter Switch System Specifications
7751/7752/7753 SOURCE/SWITCH MODULE SPECIFICATIONS
Ch. 25
Ch. 13
Vsource
Ch. 28
(7751/7753 Only)
50-500V
Bank 1 J101
Bank 2 J102
Bank 3 J103
Bank 4 J104
+ 
Source
HI
LO
HI
LO
HI
LO
HI
LO
Sense
Source
Sense
Source
Sense
Source
Sense
Source
Sense
Source
Sense
Source
Sense
Source
Sense
Ch. 1
Ch. 2
A1
Ch. 3
B1
Ch. 4
Ch. 5
A2
Ch. 6
B2
Ch. 7
Ch. 8
A3
Ch. 9
B3
Ch. 10
Ch. 11
A4
Ch. 12
B4
Ch. 14
Ch. 15
Ch. 16 Ch. 18
Ch. 17
Ch. 23
Ch. 19
Source Enable
10K
10K
200K (7751)
20K (7753)
+
+5V
HI
Input
LO
HI
Sense
LO
Interlock
+5V
Interlock
Guard
Guard
Source HI
Source LO
Sense HI
Sense LO
I/V IN
I/V OUT
I/V LO
Vsrc HI
Vsrc LO
Isrc HI
Isrc LO
Guard
Guard
I/V Current/Voltage Amplifier (7751/7753 Only)
To Model 2790 Backplane
J106
J105
J108
J107
Isource Ch. 27
0-50mA
Ch. 24
20mV
Dry
Ch. 22
Ckt
Ch.22 Open: Isource Select Ch.22 Close: Vsource Select
Ch. 21
100KW
1W
Ch. 20
Cable Discharge (7751/7753 Only)
HW 11/04/03
Rev. D
2790 SourceMeter Switch System Specifications
2790 MAINFRAME FUNCTION SPECIFICATION
MAINFRAME FUNCTION ACCURACY SPECIFICATIONS ARE FOR 1 YEAR, 23°C±5°C DC MEASUREMENT SPECIFICATIONS
CONDITIONS: MED (1 PLC)2 or 10 PLC ACCURACY: ±(ppm of reading + ppm of range) or MED (1 PLC) with Digital Filter of 10 (ppm = parts per million) (e.g., 10ppm = 0.001%)
Function Range Resolution
Voltage
Resistance
100.0000 mV 0.1 µV
1.000000 V 1.0 µV
10.00000 V 10 µV
100.0000 V 100 µV
1000.000 V
6, 8
100.0000 100 µ
5
1 mV
1.000000 k 1 m
10.00000 kΩ 10 m
100.0000 k 100 m
1.000000 MΩ 1.0
10.00000 M7 10
100.0000 M7 100
Continuity (2W) 1.000 k 100 m
Current
Channel (Ratio)
20.00000 mA 10 nA <0.2 V 60 + 30 300 + 80 500 + 80 (50 + 5)/°C
100.0000 mA 100 nA <0.05 V 100 + 300 300 + 800 500 + 800 (50 + 50)/°C
1.000000 A 1.0 µA <0.3 V9 200 + 30 500 + 80 800 + 80 (50 + 5)/°C
3.000000 A 10 µA <1.0 V
10
Channel (Average)10
Test Current (±5%)
or Burden Voltage
1 mA 6.6 V 20 + 20 80 + 20 100 + 20 (8 + 1)/°C
1 mA 6.6 V 20 + 6 80 + 6 100 + 6 (8 + 1)/°C
100 µA 6.6 V 20 + 6 80 + 6 100 + 6 (8 + 1)/°C
10 µA 12.8 V 20 + 6 80 + 10 100 + 10 (8 + 1)/°C
10 µA 12.8 V 20 + 6 80 + 10 100 + 10 (8 + 1)/°C
Note 7 7.0 V 150 + 6 200 + 10 400 + 10 (70 + 1)/°C
Note 7 7.0 V 800 + 30 3000 + 30 3000 + 30 (385 + 1)/°C
1 mA 6.6 V 40 + 100 100 + 100 100 + 100 (8 + 1)/°C
Ratio Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
Average Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
1
Input
Resistance or
Open Ckt.
3
Voltage
>10 G
>10 G
>10 G
10 MΩ ±1%
10 ΜΩ ±1%
9
1000 + 15 1200 + 40 1200 + 40 (50 + 5)/°C
4
24 Hour 23°C ±1°
90 Day
23°C ±5°
1 Year
23°C ±5°
Temperature Coefficient
0°-18°C & 28°-40°C
15 + 30 25 + 70 30 + 70 (1 + 5)/°C
15 + 6 25 + 7 30 + 7 (1 + 1)/°C
10 + 4 20 + 5 30 + 5 (1 + 1)/°C
15 + 6 45 + 9 55 + 9 (5 + 1)/°C
20 + 6 35 + 9 50 + 9 (5 + 1)/°C
Temperature
(Displayed in °C, °F, or K. Exclusive of probe errors.)
Range Resolution
4-Wire RTD: (100 platinum [PT100], D100, F100, PT385, PT3916, or user type. Offset compensation On.)
-200°C to 630°C 0.01°C 0.06°C 0.003°C/°C
Thermistor: (2.2 k, 5k, and 10k)18
-80°C to 150°C 0.01°C 0.08°C 0.002°C/°C
Accuracy
1 Year (23°C ± 5°C)
Temperature Coefficient
0-18°C & 28°-40°C
DC Speed vs. Noise Rejection
Rate Filter Readings/s11 Digits
10 50 0.01 (0.08) 6.5 < 1.2 µV 110 dB
1 Off 15 (12) 6.5 < 4 µV 90 dB12 120 dB
0.1 Off 500 (400) 5.5 < 22 µV 80 dB
0.01 Off 2000 (1800) 4.5 < 150 µV 80 dB
RMS Noise 10V Range
NMRR CMRR13
12
120 dB
DC Operating Characteristics15
60Hz (50Hz) Operation
Function Digits Readings/s PLCs
DCV, DCI, Ohms (<10M), Thermistor
4W Ohms (<10M)
RTD
Channel (Ratio), Channel (AVG)
11, 15
6.5
5 (4) 10
6.515 30 (24) 1
11, 15
6.5
50 (40) 1
11, 15
5.5
100 (80) 0.1
15, 16
5.5
250 (200) 0.1
5.516 480 (400) 0.1
16
2000 (1800) 0.01
4.5
15
1.4 (1.1) 10
6.5
6.515 15 (12) 1
16
33 (25) 0.1
5.5
15
0.9 (0.7) 10
6.5
6.515 8 (6.4) 1
15, 16
5.5
18 (14.4) 0.1
15
6.5
2.5 (2) 10
6.515 15 (12) 1
16
25 (20) 0.1
5.5
HW 11/04/03
Rev. D
2790 SourceMeter Switch System Specifications
MAINFRAME FUNCTION SPECIFICATION
DC System Speeds
RANGE CHANGES15: 50/s (42/s).
FUNCTION CHANGES
AUTORANGE TIME
14, 17
15
: 50/s (42/s).
15
: < 30ms.
ASCII READINGS TO RS-232 (19.2k BAUD): 55/s.
MAX. INTERNAL TRIGGER RATE: 2000/s.
MAX. EXTERNAL TRIGGER RATE: 375/s.
DC MEASUREMENT CHARACTERISTICS
DC Volts
A-D LINEARITY: 2.0 ppm of reading + 1.0 ppm of range.
INPUT IMPEDANCE:
100mV–10V Ranges: Selectable >10G// with <400pF or
10M ±1%.
100V, 1000V Ranges: 10M ±1%.
INPUT BIAS CURRENT: <100pA at 23°C.
COMMON MODE CURRENT: <500nAp-p at 50Hz or 60Hz.
INPUT PROTECTION: Front, 1000V, Rear, 300V, 7702 card only.
Resistance
MAX 4W LEAD RESISTANCE: 10% of range per lead for 100 and
1k ranges; 1k per lead for all other ranges.
OFFSET COMPENSATION: Selectable on 4WΩ, 100, 1k, and
10k ranges.
CONTINUITY THRESHOLD: Adjustable 1 to 1000 Ω.
INPUT PROTECTION: Front, 1000V Source Inputs, 350V Sense
DC Current
Shunt Resistors: 100mA–3A, 0.1. 20mA, 5Ω.
Input Protection: 3A, 250V fuse.
Inputs, Rear, 300V, 7702 card only.
DC Notes
1
20% overrange except on 1000V and 3A.
2
Add the following to “ppm of range” uncertainty; 100mV 15ppm, 1V
and 100V 2ppm, 100 30ppm, <1M 2ppm, 10mA and 1A 10ppm, 100mA 40ppm.
3
±2% (measured with 10M input resistance DMM, >10G DMM on
10M and 100M ranges).
4
Relative to calibration accuracy.
5
For signal levels >500V, add 0.02ppm/V uncertainty for portion
exceeding 500V.
6
Specifications are for 4-wire Ω, 100Ω with offset compensation on.
With offset compensation on, OPEN CKT. VOLTAGE is 12.8V. For 2­wire add 1 additional uncertainty.
7
Must have 10% matching of lead resistance in Input HI and LO. Test current +0.7µA||10M
8
Add the following to “ppm of reading” uncertainty when using plug in
modules:
7702, 7751, 7752
9
Add 1V when used with plug-in modules.
10
For RATIO, DCV only. For AVERAGE, DCV only. Available with
plug-in modules only.
11
Auto zero off.
12
For LSYNC On, line frequency ±0.1 %. For LSYNC Off, use 60dB for
1PLC.
13
For 1k unbalance in LO lead.
14
Speeds are for 60Hz (50Hz) operation using factory defaults operating
10 k100 kΩ 1 M 10 M 100 M
220 ppm 2200 ppm
conditions (*RST). Autorange off, Display off, Limits off, Trigger delay = 0.
15
Speeds include measurements and binary data transfer out the GPIB.
16
Sample count = 1024, auto zero off.
17
Auto zero off, NPLC = 0.01.
18
For lead resistance >0, add the following uncertainty/ for
measurement temperatures of:
2.2 k
5.0 k
10 k
(44004) 0.22°C 1.11°C
(44007) 0.10°C 0.46°C
(44006) 0.04°C 0.19°C
70°–100°C 100°–150°C
HW 11/04/03
Rev. D
2790 SourceMeter Switch System Specifications
MAINFRAME FUNCTION SPECIFICATION
AC MEASUREMENT SPECIFICATIONS
Function Range Resolution
Voltage2
1.000000 V 1.0 µV
10.00000 V 10 µV 1 Year 0.35 + 0.03 0.06 + 0.03 0.12 + 0.05 0.6 + 0.08 4.0 + 0.5
100.0000 V 100 µV
750.000 V 1.0 µV
3 Hz-10 Hz 10 Hz-3 kHz 3 kHz – 5 kHz
Current2
3.0000014 A 10 µA 0.35 + 0.06 0.15 + 0.06 0.18 + 0.06
Frequency4 and Period
100.0000 mV 0.1 µV 90 Days 0.35 + 0.03 0.05 + 0.03 0.11 + 0.05 0.6 + 0.08 4.0 + 0.5
1.000000 A 1.0 µA 90 Day/1 Year 0.30 + 0.04 0.10 + 0.04 0.14 + 0.04
100 mV 0.333 ppm 90 Day/1 Year 80 ppm + 0.333 ppm (SLOW, 1s gate)
to 3.33 ppm 80 ppm + 3.33 ppm (MED, 100ms gate)
750 V 33.3 ppm 80 ppm + 33.3 ppm (FAST, 10ms gate)
(Temp. Coeff.)/°C
(Temp. Coeff.)/°C
Additional Uncertainty ±(% of reading)
Low Frequency Uncertainty MED FAST
20Hz - 30Hz 0.3
30Hz - 50Hz 0
50Hz - 100Hz 0 1.0
100Hz - 200Hz 0 0.18
200Hz - 300Hz 0 0.10
> 300Hz 0 0
CREST FACTOR5: 1–2 2–3 3–4
Additional Uncertainty: 0.05 0.15 0.30
All Ranges
Calibration Cycle
AC MEASUREMENT CHARACTERISTICS
AC Volts
MEASUREMENT METHOD: AC-coupled, True RMS.
INPUT IMPEDANCE: 1M ±2% // by <100pF.
INPUT PROTECTION: 1000Vp or 400VDC
module.
AC Current
MEASUREMENT METHOD: AC-coupled, True RMS.
SHUNT RESISTANCE: 0.1.
BURDEN VOLTAGE: 1A <0.3Vrms, 3A <1Vrms. Add 1Vrms when
used with 7702 modules.
INPUT PROTECTION: 3A, 250V fuse.
, 300Vrms with 7702
1
Accuracy: ±(% of reading + % of range), 23°C ±5 °C
3 Hz-
13
10 Hz
3
0.035 + 0.003 0.005 + 0.003 0.006 + 0.005 0.01 + 0.006 0.03 + 0.01
3
0.035 + 0.006 0.015 + 0.006 0.015 + 0.006
(3 Hz-500 kHz) (333 ms-2 µs)
10 Hz-
20 kHz
All Ranges
20 kHz-
50 kHz
50 kHz-
100 kHz
100 kHz-
Frequency and Period
MEASUREMENT METHOD: Reciprocal Counting technique.
GATE TIME: SLOW 1s, MED 100ms, and FAST 10ms.
AC General
AC CMRR6: 70dB.
MAXIMUM CREST FACTOR: 5 at full-scale.
VOLT HERTZ PRODUCT: <= 8 × 10
7
.
AC OPERATING CHARACTERISTICS7
60Hz (50Hz) Operation
Function Digits Readings/s Rate Bandwidth
ACV, ACI
Frequency, Period
6.58 2s/Reading SLOW 3 Hz-300kHz
6.58 1.4 (1.1) MED 30 Hz-300kHz
9
6.5
4.8 (4) MED 30 Hz-300kHz
6.59 35 (28) FAST 300 Hz-300kHz
6.5 1 (1) SLOW 3 Hz-300kHz
5.5 9 (9) MED 30 Hz-300kHz
4.5 35 (35) FAST 300 Hz-300kHz
4.510 65 (65) FAST 300 Hz-300kHz
300 kHz
HW 11/04/03
Rev. D
2790 SourceMeter Switch System Specifications
MAINFRAME FUNCTION SPECIFICATION
AC System Speeds
RANGE CHANGES12: 4/s (3/s).
FUNCTION CHANGES
7, 11
12
: 4/s (3/s).
AUTORANGE TIME: < 3s.
ASCII READINGS TO RS-232 (19.2k baud): 50/s.
MAX. INTERNAL TRIGGER RATE: 300/s.
MAX. EXTERNAL TRIGGER RATE: 250/s.
AC Notes
1 20% overrange except on 750V and 3A.
2
Specifications are for SLOW mode and sine wave inputs >5% of range.
SLOW and MED are multi-sample A/D conversions. FAST is DETector:BANDwidth 300 with nPLC = 1.0.
3
Applies to 0°–18°C and 28°–40°C.
4
Specifications are for square wave inputs only. Input signal must be
>10% of ACV range. If input is <20mV on the 100mV range then the frequency must be >10Hz.
5
Applies to non-sine waves >5Hz and <500Hz. (Guaranteed by design for
Crest Factors >4.3)
6
For 1k unbalance in LO lead.
7
Speeds are for 60Hz (50Hz) operation using factory defaults operating
conditions (*RST). Autorange off, Display off, Limits off, Trigger delay=0. Includes measurement and binary data transfer out GPIB.
8
0.01% of step settling error. Trigger delay = 400ms.
9
Trigger delay = 0.
10
Sample count = 1024.
11
DETector:BANDwidth 300 with nPLC = 0.01.
12
Maximum useful limit with trigger delay = 175ms.
13
Typical uncertainties. Typical represents two sigma or 95% of manufactured units measure < 0.35% of reading and three sigma or
99.7% < 1.06% of reading.
14
For signal levels >2.2A, add additional 0.4% to “of reading” uncertainty.
Internal Scanner Speeds:
Into and Out of Memory to GPIB1
7702 Scanning DCV: 60/s
Internal Scanner Speed Notes:
1 Speeds are 60Hz or 50Hz operation using factory default conditions
(*RST). NPLC = 0.01. Auto Zero off, Auto Range off, and Display off. Sample count =1024. Includes measurement and binary data transfer out GPIB.
GENERAL SPECIFICATIONS
POWER SUPPLY: 100V / 120V / 220V / 240V. LINE FREQUENCY: 50Hz to 60Hz and 400Hz,
automatically sensed at power-up.
POWER CONSUMPTION: 28VA. OPERATING ENVIRONMENT: Specified for 0°C to
50°C. Specified to 80% R.H. at 35°C. Altitude up to 2000 meters.
STORAGE ENVIRONMENT: -40°C to 70°C. BATTERY: Lithium battery-backed memory, 3 years @
23°C.
C: Conforms to European Union Directive 89/336/EEC
EM
EN61326-1.
SAFETY: Conforms to European Union Directive
73/23/EEC EN61010-1, CAT I.
VIBRATION: MIL-PRF-28800F Class 3, Random. WARM-UP: 2 hours to rated accuracy. DIMENSIONS:
Rack Mounting: 89mm high × 213mm wide × 370mm
deep (3.5 in. × 8.375 in. × 14.563 in.).
Bench Configuration (with handle and feet): 104mm
high × 238mm wide × 370mm deep (4.125 in. ×
9.375 in. × 14.563 in.).
SHIPPING WEIGHT: 6.5kg (14 lbs). DIGITAL I/O: 2 inputs, 1 for triggering and 1 for
hardware interlock. 5 outputs, 4 for Reading Limits and 1 for Master Limit. Outputs are TTL compatible or can sink 250mA, diode clamped to 33V.
EARTH ISOLATION: 500Vpeak, >10G and <150pF
any terminal to chassis.
TRIGGERING AND MEMORY:
Window Filter Sensitivity: 0.01%, 0.1%, 1%, 10%, or
Full-scale of range (none).
Reading Hold Sensitivity: 0.01%, 0.1%, 1%, or 10% of
reading.
Trigger Delay: 0 to 99 hrs (1ms step size). External Trigger Delay: <2ms. External Trigger Jitter: <1ms. Memory Size: 55,000 readings.
MATH FUNCTIONS: Rel, Min/Max/Average/Std
Dev/Peak-to-Peak (of stored reading), Limit Test, %, mX + b and m(1/X) +b with user defined units displayed.
REMOTE INTERFACE:
GPIB (IEEE-488.2) and RS-232C. SCPI (Standard Commands for Programmable
Instruments)
ACCESSORIES SUPPLIED: User Manual and
Reference Manual, Screw Driver.
MODULES SUPPORTED: Models 7751, 7752, and
7702.
HW 11/04/03
Rev. D
All Keithley trademarks and trade names are the property of Keithley Instruments, Inc.
All other trademarks and trade names are the property of their respective companies.
A GR
Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY (534-8453) • www.keithley.com
EA
Specifications are subject to change without notice.
E OF C
TER M
EASU
eithley Instruments, Inc.
K
R
FIDENCE
ON
12/04