Keithley 7711, 7712 User Manual

Model 7711/7712
User’s Guide
PA-818 Rev. B / 8-03
A GREATER MEASURE OF CONFIDENCE
WARRANTY
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PRO­VIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIM­ITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
Keithley Instruments, Inc.
Sales Offices: BELGIUM:
CHINA:
Yuan Chen Xin Building, Room 705 • 12 Yumin Road, Dewai, Madian • Beijing 100029 • 8610-82251886 • Fax: 8610-82251892
FINLAND: FRANCE: GERMANY: GREAT BRITAIN: INDIA: ITALY: JAPAN: KOREA: NETHERLANDS: SWEDEN: TAIWAN:
3, allée des Garays • 91127 Palaiseau Cédex • 01-64 53 20 20 • Fax: 01-60 11 77 26
1/5, Eagles Street • Langford Town • Bangalore 560 025 • 080 212 80-27 • Fax: 080 212 80 05
Viale San Gimignano, 38 • 20146 Milano • 02-48 39 16 01 • Fax: 02-48 30 22 74
New Pier Takeshiba North Tower 13F • 11-1, Kaigan 1-chome • Minato-ku, Tokyo 105-0022 • 81-3-5733-7555 • Fax: 81-3-5733-7556
2FL., URI Building • 2-14 Yangjae-Dong • Seocho-Gu, Seoul 137-888 • 82-2-574-7778 • Fax: 82-2-574-7838
c/o Regus Business Centre • Frosundaviks Allé 15, 4tr • 16970 Solna • 08-50 90 46 00 • Fax: 08-655 26 10
13F-3, NO. 6, Lane 99, Pu-Ding Road, Hsinchu, Taiwan, ROC. • 886-3-572-9077 • Fax: 886-3-572-9031
28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168
1-888-KEITHLEY (534-8453) • www.keithley.com
Bergensesteenweg 709 • B-1600 Sint-Pieters-Leeuw • 02-363 00 40 • Fax: 02-363 00 64
Halsuantie 2 • 00420 Helsinki, Finland • 09-53 06 65 60 • Fax: 09-53 06 65 65
Landsberger Strasse 65 • 82110 Germering • 089-84 93 07-40 • Fax: 089-84 93 07-34
Unit 2 Commerce Park, Brunel Road • Theale, Berkshire RG7 4AB • 0118 -929 75 00 • Fax: 0118- 929 75 19
Postbus 559 • 4200 AN Gorinchem • 0183-63 53 33 • Fax: 0183-63 08 21
5/03
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some in­struments and accessories would normally be used with non-haz­ardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recog­nize shock hazards and are familiar with the safety precautions re­quired to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection pro­vided by the product may be impaired.
The types of product users are:
Responsible body
and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for en­suring that operators are adequately trained.
Operators
trained in electrical safety procedures and proper use of the instru­ment. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel
to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are de­scribed in the manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel
safe installations and repairs of products. Only properly trained ser­vice personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and data I/O sig­nals are Measurement Category I and must not be directly connect­ed to mains voltage or to voltage sources with high transient over­voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless other­wise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present.
that hazardous voltage is present in any unknown circuit before measuring.
is the individual or group responsible for the use
use the product for its intended function. They must be
perform routine procedures on the product
are trained to work on live circuits, and perform
A good safety practice is to expect
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are pre­vented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human con­tact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts,
the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When con­necting sources to switching cards, install protective devices to lim­it fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connect­ed to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power dis­connect device must be provided, in close proximity to the equip­ment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jump­ers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the com­mon side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equip­ment may be impaired.
Do not exceed the maximum signal levels of the instruments and ac­cessories, as defined in the specifications and operating informa­tion, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is ap­plied to the device under test. Safe operation requires the use of a lid interlock.
no conductive part of
5/03
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should re­fer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or mea­sure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The symbol indicates a connection terminal to the equipment frame.
The
WARNING
result in personal injury or death. Always read the associated infor­mation very carefully before performing the indicated procedure.
The
CAUTION
damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
heading in a manual explains dangers that might
heading in a manual explains hazards that could
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instru­ments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that se­lected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to in­structions. If the board becomes contaminated and operation is af­fected, the board should be returned to the factory for proper cleaning/servicing.
Keithley Instruments, Inc.
28775 Aurora Road Cleveland, Ohio 44139 (440) 248-0400 Fax: (440) 248-6168 www.keithley.com
The information in this User’s Guide is organized as follows:
Introduction on page 2
• Available accessories on page 2
Card configuration — schematic on page 3
Expansion on page 4
• Multiplexer examples on page 4
•4 × 4 Blocking matrix example on page 7
•4 × 4 Non-blocking matrix example on page 8
Wiring procedure on page 10
Operation on page 11
• Channel assignments on page 11
Close/Open channels on page 11
• Multiple channel operation on page 11
S-parameters on page 12
• Unlocking calibration on page 12
• CALibration subsystem on page 13
SYSTem subsystem on page 14
Measurement considerations on page 15
• Termination of unused connectors on page 15
• Power handling on page 15
• Additional measurement uncertainty on page 16
Application example on page 18
Service on page 22
• Performance verification on page 22
• Replaceable parts on page 24
Model 7711/7712
User’s Guide
PA-818 Rev. B / 8-03
Introduction
If you have any questions after reviewing this information, please contact your local Keithley representative or call one of our Applications Engineers at 1-800-KEITHLEY (U.S. and Canada only).
The Model 7711/7712 is a single-pole dual 1 × 4 RF Multiplexer module (8 channels, no measurement capability). The 1 × 4 multiplex is a cascading tree design — one of the channels of each bank is always connected to a common out. The 1 × 4 multiplex is also referred to as a single-pole 4-throw (SP4T) switch. Refer to the simplified schematic in Figure 1 or the specifications. The multiplex switching can be used to connect one instrument to multiple devices (1:N) or multiple instruments to a single device (N:1). This card expands the Integra Series’ general purpose AC switching capability by providing an RF switching capability above 1MHz. Each card has the following features:
• 10 SMA connectors with 50 impedance
• Frequencies (refer to specifications) Model 7711 — up to 2 GHz Model 7712 — up to 3.5 GHz
• Two channels can be closed at one time—one channel to OUT A (channels 1–4) and one to OUT B (channels 5–8)
• Designed specifically for use with the Integra Series including Keithley’s Models 2700, 2701 and 2750 Multimeter / Data Acquisition Systems
NOTE All references to the Model 27xx apply to the Models 2700, 2701, and 2750.
With this in mind, the capability of the Model 27xx is expanded for applications such as the following:
• General AC and digital signal routing to and from test equipment. This includes oscilloscopes, function generators, pulse generators, counter/timers, and signal analysis tools including network and jitter analyzers. Also, this includes diverse applications that require RF and digital routing plus DC measurements for research, development, burn-in and production testing, AC ripple and noise test for switching power supplies, and DC-DC converters.
• Research, development, and burn-in of temperature compensated oscillators.
• Research, development, and production test of communications or networking products including Ethernet, DSL, DS3, T1/E1, etc.
WARNING Before operating the Model 27xx with an accessory card, verify that the card is properly
installed and the mounting screws are tightly fastened. If the mounting screws are not properly connected, an electrical shock hazard may be present.
Available accessories
Model 7711 only
7711-BNC-SMA Male SMA to Female BNC, five 0.15m (0.5ft) cables
(Insertion loss < 1dB@1GHz, VSWR < 1.5@ 1GHz) 7051-2 BNC cable, male to male, 0.6m (2ft.) 7051-5 BNC cable, male to male, 1.5m (5ft.) 7051-10 BNC cable, male to male, 3.0m (10ft.)
2
Model 7711/7712
7712-SMA-1 SMA cable, male to male, flexible, low loss, 1m (3.3ft)
(Insertion loss < 0.65dB @ 2GHz, VSWR < 1.11 @2GHz) 7712-SMA-N Female SMA to Male N-Type Adapter S46-SMA-1 SMA cable, male to male, 0.3m (1ft) S46-SMA-0.5 SMA cable, male to male, 0.15m (0.5ft)
Card configuration—schematic
Figure 1 shows a simplified schematic diagram of the Model 7711/7712 module. As shown, the Model 7711/7712 has channels
that are grouped into dual 1 × 4 multiplexers. The multiplexer’s design allows 1 channel of each set of 4 to be closed at a time (1 of 4 tree structure). Also, channels 1 and 5 are normally closed to OUT A and OUT B, respectively.
NOTE The Model 7711/7712 relays are set to close channels 1 and 5 a few seconds after either a power
cycle or a *RST command is issued (see Figure 1 for a schematic).
Figure 1
Simplified schematic for Model 7711/7712
Ch1
Ch2
OUT A
Ch3
Ch4
Ch5
Ch6
OUT B
Ch7
Ch8
NOTE To close channels 1 and 5 on a Model 7711/7712 in slot 1, send: :ROUT:MULT:CLOS
(@101,105)
The channels are controlled over the bus or from the front panel. The grounds for these channels are non-isolated. Detailed connection information is contained later in this guide. By using the ROUT:MULT commands (refer to Section 2 of the Model 27xx User’s Manual), they can be manually configured.
3
Expansion
One Model 7711/7712 configuration is a dual (2 bank) 1 × 4 multiplexer. A single Model 7711/7712 module can also be configured as a 4 × 4 blocking matrix. The 7711/7712 can be combined with additional modules to further expand its multiplexing and matrix switching capabilities (specifications may degrade when the Model 7711/7712 is used in multiple module configurations). Refer to Table 1 for possible expansion information. Examples of multiplexer and blocking matrix expansion follow Table 1.
Table 1
Matrix and multiplex expansion
Multiplexer Blocking Matrix
Number of 7711/7712 Cards 1 ××
1X X
2 XXX XX
3 XXXX X* XXX
*Using four cards in a Model 2750, a 4 × 4 non-blocking matrix can be created (see the 4 × 4 non-blocking matrix example on page 9).
NOTE The commands in this section are formatted as follows: Command (@Model 27xx Slot #,
Channel #)
××
41 ××
××
81 ××
××
12 1 ××
××
16 4 ××
××
44 ××
××
84 ××
××
12 4 ××
××
16
Multiplexer examples
NOTE To minimize noise within the system, terminate unused channels with 50
not available, terminating unused connectors with a cable may also reduce noise within the system.
Figure 2 shows how to create a 1 × 8 SP8T MUX (single-pole eight-throw multiplex) using two Model 7711 cards. Channels 1
and 2 are connected to OUT A and OUT B of the second card in slot two. Terminate unused channels (in the example, channels 3 and 4 on the Model 7712 in slot one) with a 50 load.
Figure 3 shows how to create a 1 × 12 MUX using two Model 7711 cards.
loads. If 50
loads are
NOTE Channels 3 and 4 (of 7711 in slot 1) may also be connected to other banks that will create a 1
multiplexer (or a 1 sample commands needed to operate the 1
NOTE The command “ROUTe:MULTiple:CLOSe (@<channel list>)” closes one channel per bank (two
banks per Model 7711/7712 module), and opens all other channels within the same bank.
4
×
16 multiplexer). See Figure 4 for a 1
×
8 multiplexer.
×
×
12 example. The following is a list of
12
Figure 2
1 x 8 MUX expansion
5
OUT A
7711/7712
Slot 1
>>
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
NOTE: The simplified schematic is shown in the
power up state (Channels 1 and 5 are closed for cards in slots 1 and 2).
Grayed channels are not used in this 1 × 8 SP8T MUX.
S46-SMA-0.5 Cable (Qty 2)
7711/7712 Slot 2
Ch1
Ch2
Ch3
OUT A
>>
Ch4
SLOT 1
SLOT 2
Ch1
Ch2
Ch3
Ch4
Ch5
Ch6
Ch7
Ch8
Ch5
OUT B
>>
Ch6
Ch7
Ch8
To close to this crosspoint (CH7) send:
ROUTe:MULTiple:CLOSe(@102,207)
Figure 3
××
1
××
12 MUX expansion
6
OUT A
>>
7711/7712 Slot 1
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
NOTE: The simplified schematic is shown in the
power up state (Channels 1 and 5 are closed for cards in slots 1 and 2).
Grayed channels are not used in this 1 × 12 SP12T MUX.
Ch1
Ch2
Ch3
Ch4
S46-SMA-0.5 Cable (Qty 3)
OUT A
OUT B
7711/7712 Slot 2
>>
>>
SLOT 1
SLOT 2
To close to this crosspoint (CH3) send:
ROUTe:MULTiple:CLOSe(@101,203)
Ch1
Ch2
Ch3
Ch4
To close to this crosspoint (CH7) send:
ROUTe:MULTiple:CLOSe(@102,207)
Ch5
Ch6
Ch7
Ch8
Terminate this unused channel in bank 1 with a 50 load or an unconnected S46-SMA-0.5 cable.
NOTE: Channels 5-8 on the 7711/7712
in slot 1 correspond to crosspoints Channels 9-12 of the 1 × 12 SP12T MUX.
OUT B
>>
Ch5 (Ch9)
Ch6 (Ch10)
Ch7 (Ch11)
Ch8 (Ch12)
To close to this crosspoint (CH11) send:
ROUTe:MULTiple:CLOSe(@103,107)
4
××
××
4 Blocking matrix example
Figure 4 shows how to connect a single Model 7711 in a 4 × 4 blocking matrix. This configuration provides a convenient way
to connect four DUTs to four different test instruments. In this configuration, one channel of bank 1 (channels 1 – 4) may be routed to any channel of bank 2 (channels 5 – 6). Use a short coaxial cable to connect OUT A to OUT B (S46-SMA-0.5). con­tains example commands.
Figure 4
××
4
4 blocking matrix expansion
××
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
S46-SMA-0.5 Cable (Qty 1)
7711/7712 Slot 1
Ch1
Ch2
Ch3
Ch4
OUT A
OUT B
Ch5
Ch6
Ch7
Ch8
To route the signal between Ch4 and Ch8 send:
ROUTe:MULTiple:CLOSe(@104,108)
NOTE: The simplified schematic is shown in the
power up state (Channels 1 and 5 closed).
Table 2
××
4
××
4 Blocking matrix commands
Command Signal path
ROUTe:MULTiple:CLOSe (@101,105) Routes signal through channels 1 and 5
ROUTe:MULTiple:CLOSe (@103,106) Routes signal through channels 3 and 6
7
4 x 4 Non-blocking matrix example
Using a Model 2750 and four Model 7711/7712 cards, a 4 × 4 non-blocking matrix can be created. Figure 5 shows how to connect the 7711/7712 modules. This configuration provides a convenient way to simultaneously connect up to 4 DUTs to 4 different test instruments. Use a short coaxial cable to connect the OUTs of each module as shown. contains a couple of sample commands to control the matrix.
To use , select the IN connection (IN 1 – IN 4), and then in the same square select the OUT connection (OUT1 - OUT4). Send the
ROUTe:MULTiple:CLOSe command with the channels in both “Ch” columns.
Table 3
××
4 Sample non-blocking matrix commands
4
××
Command Signal path
ROUTe:MULTiple:CLOSe (@101,405) Routes signal from IN1(OUT A of slot 1) through
OUT4 (OUT B of slot 4).
ROUTe:MULTiple:CLOSe (@101,405,108,302) Routes signal from IN1 (OUT A of slot 1) through
OUT4 (OUT B of slot 4), and also from IN2 (OUT B of slot 1) through OUT1 (OUT A of slot 4).
Table 4
××
4
××
4 Non-blocking matrix channels
IN Ch Ch OUT IN Ch Ch OUT
IN 1 101 405 OUT4 IN 3 201 407 OUT4
102 401 OUT3 202 403 OUT3
103 305 OUT2 203 307 OUT2
104 301 OUT1 204 303 OUT1
IN Ch Ch OUT IN Ch Ch OUT
IN 2 105 406 OUT4 IN 4 205 408 OUT4
106 402 OUT3 206 404 OUT3
107 306 OUT2 207 308 OUT2
108 302 OUT1 208 304 OUT1
8
Figure 5
4 x 4 Non-blocking matrix expansion
7711
9
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
IN1
OUT A OUT B
SLOT 1 SLOT 2
1
2 3 4 5
SLOT 1 SLOT 2
IN2
7 8Channel 1 2 3 4 5
6
IN3
OUT A OUT B
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
IN4
6
7 8
SLOT 3
7711
NOTE: The simplified schematic
is shown in the power up state (Channels 1 and 5 closed for cards in all slots).
Channel
1
2
3
SLOT 3
4
5
6
7
8
1
2
3
SLOT 4
4
5
6
7
8
SLOT 4
OUT A OUT B OUT A OUT B
OUT1
OUT2
OUT3
OUT4
Wiring procedure
WARNING Do not exceed the maximum specifications for the Model 7711/7712 module. Refer to
Appendix A for specifications.
The default closed channels are 1 and 5 — the lowest number in each bank. At power-up, the lowest numbered channel in each bank will always be closed. If using a saved setup that calls for another channel to be closed, the card will be switched to that channel but only after the firmware has finished "waking up." Therefore, the lowest channel will ALWAYS be closed for at least a brief time at power-up. The user needs to verify that nothing is connected to the lowest channel in each bank that could be damaged.
The connectors on the card are standard SMA type connectors. When making or breaking connections, use a torque wrench to produce mating torque of 0.9 N•m (8in•lbs).
NOTE Use a box-spanner torque wrench similar to Suhner Type 74 Z-1-0-21, Identification number
543130. This wrench is for SMA connectors with a 1.00 N•m torque with 8.0mm (0.315 in.) across the flats.
Figure 6
Model 7711/7712 channel connections
7711
CH7 CH1CH2OUT ACH3CH4CH5CH8 CH6OUT B
WARNING Do not touch live RF (radio frequency) conductors, even at low voltages! RF behaves
differently than DC or low frequency AC. Low voltage RF can cause severe burns. RF signals spread out over body areas generating substantial currents at the points of contact with RF conductors.
Make all connections using correct wire size. If BNC connections are required, use the SMA to BNC adapter cables (see
Available accessories on page 2).
Specifications
Full Model 7711/7712 specifications are included at the end of this User Guide.
10
Operation
NOTE The Model 7711/7712 modules cannot be included in scanlists.
Detailed information to close and open switching module channels is provided in Section 2 of the Model 27xx User’s Manual. The following summarizes basic operation and provides operating information specific to the Model 7711/7712.
Channel assignments
NOTE Measurement functions cannot be assigned to Model 7711/7712 modules. For example, if a
Model 7711/7712 is installed in slot 1 and the SENSE:FUNC ‘VOLT:AC’, (@101) command is sent, a -221, “Settings Conflict” error would be reported.
The Model 2700/2701 has two slots for switching modules, and the Model 2750 has five slots. To control the appropriate switching module, the slot number must be included with the switching module channel number when you specify a channel. The channel assignment is formatted as follows:
SCH where: S is the slot number
CH is the channel number
Examples: 101 = Slot 1, Channel 1 210 = Slot 2, Channel 10 506 = Slot 5, Channel 6 (Model 2750)
NOTE For remote operation, the 3-digit channel assignment is included in the channel list parameter for
the commands.
Close/Open channels
Multiple channel operation
WARNING Multiple channel operation should only be performed by experienced test engineers who
recognize the dangers associated with multiple channel closures.
For the Model 7711/7712, multiple channel operation provides control of one or two Model 7711/7712 modules in a Model 2700/2701 (or up to 5 in a Model 2750). OUT A automatically switches as appropriate for channels 1 – 4, and OUT B automatically switches as appropriate for channels 5 – 8 for each module. When you close two channels in a single module (one for OUT A and one for OUT B), the appropriate channels will close channels outside of the banks (or on other modules) are not affected by channel closures.
The following paragraphs summarize multiple channel operation for the front panel and from the remote interface (bus).
Front panel
• For the Models 2700/2701 and 2750, the MULTI option of the CLOSE key can be used to close a channel.
• For the Models 2700/2701 and 2750, the ALL option of the OPEN key closes channels 1 and 5 to OUT A and OUT B respectively.
11
Remote interface
ROUT:MULT:CLOS <clist> ROUT:MULT:CLOS? ROUT:MULT:CLOS:STAT? <clist> ROUT:OPEN:ALL
The following command is not available:
ROUT:MULT:OPEN <clist>
NOTE For example, if a Model 7711/7712 is installed in slot 1 and the ROUT:MULT:OPEN (@101) com-
— Closes channels 1 and 5 to OUT A and OUT B respectively
mand is sent, a -221, “Settings Conflict” error would be reported.
— Closes specified channels (unlisted channels not affected unless in the same bank).
— Returns list of all closed channels.
— Query closed channels in list (1 = closed).
— Invalid command for the Model 7711/7712 module.
DMM measurements
None of the channels of the 7711/7712 module are connected to the internal DMM (the channels cannot be connected to the backplane). If your test system requires DMM measurements, you can use a measure card in one of the other mainframe slots or you can use the front panel inputs of the Model 27xx.
Remember to use multiple channel operation to control the 7711/7712 module and system channel operation to control a measure card for DMM measurements. If you attempt to close a 7711/7712 module channel using front panel system channel operation, the message “NO MEAS CARD” will be displayed briefly.
If you use front panel inputs for measurements, make sure that the front panel INPUT switch is in the out (F) position.
S-parameters
S-parameters allow a text string of up to 238 ASCII characters to be stored on a module. This can be used for system calibration of cable and module losses. The calibration text string is stored on the module, independent of the Model 27xx mainframe.
Unlocking calibration
Use the following sequence of commands to unlock calibration (factory default code):
CAL:PROT:CODE “KI0027xx” CAL:PROT:CARD1:INIT
CAL:PROT:CARD1:SPAR “......”
CAL:PROT:CARD1:SAVE
where: xx = 00 (for 2700)
xx
= 01 (for 2701) = 50 (for 2750)
xx
NOTE
Your mainframe calibration unlock code will remain the factory default code unless it is changed.
12
CALibration subsystem
Commands to perform output operations are listed in Table 5. Details on these commands follow the table.
NOTE
The following commands for the CALibration subsystem are specific to the Model 7711/7712 module. It is only available if a Model 7711/7712 is installed.
Table 5
CALibration subsystem commands
Command Description Default
2700/2701 Commands
CALibration
:PROTected
:CARD1 Path to CARD1 commands.
:INITiate Initiate command—required before sending s-parameters.
:SPARameters <string> Enter up to 238 characters for storage of S-parameter information about
the card in slot 1.
:SPARameters? Query the ASCII information for CARD1.
:CARD2 Path to CARD2 commands.
:SPARameters? Query the ASCII information for CARD2.
2750 Commands (only)
CALibration
:CARD1 Path to CARD commands.
:INITiate Initiate command—required before sending s-parameters.
:SPARameters <string> Enter up to 238 characters for storage of S-parameter information about
the card in slot 1.
:CARD Path to s-parameter queries.
:SPARameters? <NRf> Query the ASCII information for card in slot <NRf>.
Empty
Empty
:SPARameters <string>
If the card in this slot is not a valid RF card (Model 7711, etc.), then a -221, “Settings Conflict” error is returned. Any valid ASCII information can be stored in the string space; it is not error checked in any way.
NOTE
S-parameter data can only be
written
to a card in slot 1. The S-parameters are treated like card-cal constants, meaning that you must send a CAL:PROT:CARD1:INIT command before sending the SPARameters, and they are only saved on the card after a CAL:PROT:CARD1:SAVE command. The S-parameter can be
:SPARameters? <NRf> Query S-parameter
read
from any slot.
S-parameter data can only be written to a card in slot 1. This is consistent with other Integra series type card calibrations that are allowed in slot 1 only. Also, the Model 2750 can be queried using CAL:PROT:CARD1:SPAR
NOTE
Errors +518 and +519 are provided to flag corrupted S-parameter information or flag S-parameter information that was not properly initialized.
13
SYSTem subsystem
Commands to perform system operations are listed in Table 6. Details on these commands follow the table.
NOTE The following commands for the Model 7711/7712 SYSTem subsystem are included here for
reference.
Table 6
SYSTem subsystem commands
Command Description Default
2700/2701 Commands
:SYSTem
:CARD1 Path to queries for slot 1 card.
:CARD2 Path to queries for slot 2 card.
:SNUMber? Request serial number of card in Slot x.
:SWRevision? Request firmware revision of card in Slot x.
:VMAX? Query the highest allowed voltage on this card.
:MUX? Return TRUE if this card support ANY mux channels.
:SNOpen? Query whether the card is of the "single, no-open" type.
:BANKs? This query returns the number of banks on the card. If the card is not of the
Single, No-Open type, then a -221 "Settings Conflict" error results with this query.
2750 Commands (only)
:CARD Path to queries (slot is designated by <NRf> after query command).
:SNUMber? <NRf> Request serial number of card in Slot x.
:SWRevision? <NRf> Request firmware revision of card in Slot x.
:VMAX? <NRf> Query the highest allowed voltage on this card.
:MUX? <NRf> Return TRUE if this card support ANY mux channels.
:SNOpen? <NRf> Query whether the card is of the "single, no-open" type.
:BANKs? <NRf> This query returns the number of banks on the card. If the card is not of the
Single, No-Open type, then a -221 "Settings Conflict" error results with this query.
:CARD1 Model 2700/2701 commands :CARD2
If no card is in slot 1, then all commands in Table 6 generate a -241 “hardware missing” error. If the card does not support the feature or channels being queried, then 0 will be returned.
:MUX?
Returns either a 1 or 0. (For a Model 7711/7712 card, a 0 is returned.)
:SNOpen?
"single, no-open" type — The card is organized into banks of channels, and one channel must always be closed within that
bank. The Model 7711 is representative of this class with two banks of 4 channels each. Returns either a 1 or 0. (For a 7711/7712 card, a 1 is returned.)
14
:CARD Model 2750 commands only
If no card is in slot <NRf>, then all of these commands will generate a -241 “hardware missing” error. If the card does not support the feature or channels being queried, then 0 will be returned.
Unlike the Model 2700/2702, the Model 2750 SYST:CARD queries have the slot passed as a <NRf> following the query (rather than incorporated into the command name). For compatibility with existing Model 2700/2702 software, the Model 2750 also accepts SYST:CARD1 or SYST:CARD2 (not recommended for queries when developing new software).
Example: To query the maximum voltage allowed on the card in slot 3, send:
:SYST:CARD:VMAX? 3
Measurement considerations
Termination of unused connectors
To minimize noise within the system, terminate unused channels with 50 loads. If 50 loads are not available, terminating unused connectors with a cable may also reduce noise within the system.
Power handling
The graph in Figure 7 shows the maximum amount of power per channel that can be routed using the Model 7711 or Model 7712 while maintaining proper digital multimeter accuracy. Power levels of up to 10 watts at frequencies > 200MHz can be used, but this may cause measurement errors. For example, routing 10 watts of power at 1GHz may cause the Model 2700/2702 to have an additional 6µV offset uncertainty when measuring DC voltages. For the Model 2750, the precision low level circuitry may cause uncertainty of 30µV. Moving the Model 7711/7712 to a higher number slot will reduce this uncertainty.
Figure 7
Carry power (per channel) vs. frequency
12
10
8
6
4
Power (watts)
2
0
10 100 1000
200MHz
Additional DMM Uncertainty
500MHz
3500
MHz
15
Additional measurement uncertainty
Table 7 lists additional uncertainty to the published specifications for the Model 7708. Other modules, such as the Model 7700
(which measures thermocouple temperature), are also affected by this heat but are still within published specification. The amount of uncertainty is dependent on:
• Model 7711/7712 slot location
• The amount of time the channel is programmed as an output
• Which other model 77XX module(s) are being used
The listed uncertainty in Table 7 is listed for channels 4 and 8 closed. The uncertainty is linear—for each channel programmed as input or off, there will be approximately 25 percent reduction in the listed on uncertainty. Table 8 lists the percentage of additional uncertainty for specific channel pairs closed.
Table 7
Additional uncertainty for Model 7711/7712
Thermocouple Range 7708
J -200 to 0˚C
0 to +768˚C
K -200 to 0˚C
0.4
0 to +1372˚C
N -200 to 0˚C
0 to +1300˚C
T -200 to 0˚C
0.75
0.15
0.4
0 to +400˚C
E -200 to 0˚C
0.15
0 to +1000˚C
R 0 to +400˚C
1.5
0 to +1768˚C
S 0 to +400˚C
1.5
0 to +1768˚C
B 350 to 1100˚C
0 to +1820˚C
Notes:
1. The above listed uncertainties are guaranteed by design for thermocouple types J, K, N, T, E, R, S, B.
2. When the Model 7711/7712 and Model 7708 are used together, the Model 7711/7712 must be installed in the lowest numbered slot. For example: Model 2700/2701: 7711/7712 in slot 1, 7708 in slot 2
Model 2750: 7711/7112 in slot 2, 7708 in slot 3
2.4
0.6
16
Table 8
Percent of additional uncertainty
Percent of additional
Channels closed
uncertainty (from Table 7)
4 and 8 2 and 8 3 and 8 4 and 6 4 and 7 1 and 8 2 and 6 2 and 7 3 and 6 3 and 7 4 and 5 1 and 6 1 and 7 2 and 5 2 and 6 1 and 5
100%
75% 75% 75% 75% 50% 50% 50% 50% 50% 50% 25% 25% 25% 25%
0%
17
Application example
The following application example is for power supply testing. For basic overview and switching, refer to Figure 8. For the tests performed, refer to the provided code (Figure 9). Basic knowledge for Visual Basic functions, such as the CStr() function, as well as how forms work in Visual Basic, is required to fully understand the example.
Figure 8
Simplified connection schematic (one power supply shown)
Oscilloscope
OUT A
Ch1
7711-SMA-BNC
7711
Cable
Thermocouple
Power Supply
Input Output
HI
LO
7712-SMA-1 Cable
Ch2
Ch4
Ch1
Ch3
HI
LO
HI
LO
DCV
CONT
10% Load
7705 Ch21
90%
Load
7705 Ch31
High Line
132V
~ ~
7705 Ch1
7705 Ch11
Low Line
108V
ACV
TEMP
HI
LO
HI
LO
7700
Figure 9
Visual basic application example—power supply testing
'' --------------------------------------------------------------------------------------------------------------------­'' 7711 Application Example '' --------------------------------------------------------------------------------------------------------------------­'' This program tests a set of power supplies using a 2750 loaded with a 7700, '' 7711, and a 7705. '' '' The following occurs: '' 1) The 7700 takes DCV, ACV, 2W OHMS (CONT), and TEMP measurements. '' 2) Limits are used in each of the measurements by the 7700. '' A) DCV limits are set to 10% of the power supply's output. '' B) ACV limits are 108 and 132V. '' C) 2W OHMS limits are -0.01 and 2 ohms. '' D) TEMP limits are 70 and 125 C. '' '' 3) RF switching to an oscilloscope is obtained using a 7711 where the AC ripple '' is measured. '' '' 4) A 7705 is used to switch loads and to control the line input voltage '' to the DUTs (power supplies). '' '' Please note: '' The maximum number of power supplies that can be used with this example is 4. '' ---------------------------------------------------------------------------------------------------------------------
18
Option Explicit '' --------------------------------------------------------------------------------------------------------------------­'' Constants '' --------------------------------------------------------------------------------------------------------------------­Const MAXLEN = 6400 ' Constant used with ENTER function. Const LENGTH = 6400 ' Constant used with ENTER function. Const ENTER_TIMEOUT = 3000 ' Length of time the ENTER function will ' wait for a query. Const ADDRESS = 16 ' DMM setup address.
'' Slot assignments. Const SLOT_7700 = 1 Const SLOT_7711 = 2 Const SLOT_7705 = 3
'' Number of configurations, power supply units, and functions. Const NUMBER_OF_DUTS = 4 Const NUMBER_OF_FUNCTIONS = 4 ' DCV, ACV, CONT, and TEMP. Const NUMBER_OF_SUPPLY_CONFIGURATIONS = 4 ' Includes HIGH/ LOW line and 10%/90% load. Const N_READINGS = NUMBER_OF_DUTS * NUMBER_OF_FUNCTIONS * NUMBER_OF_SUPPLY_CONFIGURATIONS
'' Constants to define limits for the different measurements. Const POWER_SUPPLY_VOLTAGE = 5 ' Voltage output of power supplies. Const DCV_MAX = POWER_SUPPLY_VOLTAGE * 1.1 ' Allow power supply to vary by 10%. Const DCV_MIN = POWER_SUPPLY_VOLTAGE * 0.9 Const ACV_MAX = 132 ' ACV input voltage to power supply limits. Const ACV_MIN = 108 Const CONT_MAX = 2 ' Limit failure will occur at 2 ohms or more. Const CONT_MIN = -0.01 Const TEMP_MAX = 125 ' Max. and min. limits for temperature. Const TEMP_MIN = 70
'' Constants to define power supply configurations. Const HIGH_LINE = 1 Const LOW_LINE = 2 Const TEN_PERCENT_LOAD = 1 Const NINETY_PERCENT_LOAD = 2
' Global variables. Dim Status As Integer ' Variable used by the send() function. Dim Readings(N_READINGS) As String ' Array to store readings. Dim Reading_Counter As Integer ' Index to Readings().
'' --------------------------------------------------------------------------------------------------------------------­'' 7700 Channel Configurations '' --------------------------------------------------------------------------------------------------------------------­'' Configure_Limits(,,) ­'' Sets the limits for a given channel given by the constants defined above. '' '' Configure_7700_DCV(,) ­'' Sets a channel to measure DCV. '' '' Configure_7700_ACV() ­'' Configures a channel to measure ACV. '' '' Cofigure_CONT() ­'' Configures the channel to measure 2W ohms. '' This measurement is used instead of CONT, so we can select the ohm's limits. '' '' Configure_7700_TEMP() ­'' Sets a channel to measure temperature using *RST default. '' '' Configure_Channel() ­'' Uses the above functions to do the following on the 7700: '' 1. Sets channels 1,5,9, and 13 to DCV. '' 2. Sets channels 2,6,10, and 14 to ACV. '' 3. Sets channels 3,7,11, and 15 to CONT. '' 4. Sets channels 4,8,12, and 16 to TEMP. '' 5. Places the box into Auto-Configuration Mode. '' '' Note: Auto-configuration mode brings in the individual channels' '' settings when that channel is closed. '' --------------------------------------------------------------------------------------------------------------------­Private Sub Configure_7700_Limits(Channel_Num As Integer, Ulimit As Integer, Llimit As Integer) Dim Channel_Str As String Dim Upper_Limit As String Dim Lower_Limit As String
Channel_Str = ",(@" & CStr(Channel_Num) & ")" ' Converts a number to an ASCII channel parameter. Upper_Limit = CStr(Ulimit) Lower_Limit = CStr(Llimit)
Call send(ADDRESS, "CALC3:LIM1:STAT ON" & Channel_Str, Status) ' Enable Limit1 for DCV. Call send(ADDRESS, "CALC3:LIM1:UPP:DATA " & Upper_Limit & Channel_Str, Status) ' Set upper limit. Call send(ADDRESS, "CALC3:LIM1:LOW:DATA " & Lower_Limit & Channel_Str, Status) ' Set lower limit. End Sub
Private Sub Configure_7700_DCV(Channel_Num As Integer, Power_Supply_DCV As Integer) Dim Channel_Str As String
Channel_Str = ",(@" & CStr(Channel_Num) & ")" Call send(ADDRESS, "SENS:FUNC 'VOLT:DC'" & Channel_Str, Status) ' Set channel to measure DCV. Call send(ADDRESS, "SENS:VOLT:DC:RANG 100" & Channel_Str, Status) ' Set voltage range(Power Supply is at 5V).
Call Configure_7700_Limits(Channel_Num, DCV_MAX, DCV_MIN) End Sub
Private Sub Configure_7700_ACV(Channel_Num As Integer) Dim Channel_Str As String
Channel_Str = ",(@" & CStr(Channel_Num) & ")" Call send(ADDRESS, "SENS:FUNC 'VOLT:AC'" & Channel_Str, Status) ' Set channel to measure ACV. Call send(ADDRESS, "SENS:VOLT:AC:RANG 150" & Channel_Str, Status) ' Set voltage range. Call Configure_7700_Limits(Channel_Num, ACV_MAX, ACV_MIN) End Sub
19
Private Sub Configure_7700_CONT(Channel_Num As Integer) Dim Channel_Str As String
Channel_Str = ",(@" & CStr(Channel_Num) & ")" Call send(ADDRESS, "SENS:FUNC 'RES'" & Channel_Str, Status) ' Set channel to measure 2W ohms. Call send(ADDRESS, "SENS:RES:RANG 10" & Channel_Str, Status) ' Set resistance range. Call send(ADDRESS, "SENS:RES:NPLC 0.1" & Channel_Str, Status) ' Set meaurement speed to fast.
Call Configure_7700_Limits(Channel_Num, CONT_MAX, CONT_MIN) End Sub
Private Sub Configure_7700_TEMP(Channel_Num As Integer) Dim Channel_Str As String
Channel_Str = ",(@" & CStr(Channel_Num) & ")" Call send(ADDRESS, "SENS:FUNC 'TEMP'" & Channel_Str, Status) ' Set channel to TEMP measurement. Call send(ADDRESS, "SENS:TEMP:TRAN TC" & Channel_Str, Status) ' Select thermocouple measurement. Call send(ADDRESS, "SENS:TEMP:TC:TYPE T" & Channel_Str, Status) ' Use T thermocouples. Call send(ADDRESS, "SENS:TEMP:TC:RJUN:RSEL INT" & Channel_Str, Status) ' Use internal reference junction.
Call Configure_7700_Limits(Channel_Num, TEMP_MAX, TEMP_MIN) End Sub
Private Sub Configure_7700_Channels() Dim I As Integer Dim POWER_SUPPLY_VOLTAGE As Integer
For I = 0 To NUMBER_OF_DUTS - 1 Call Configure_7700_DCV(1 + (I * NUMBER_OF_DUTS) + (SLOT_7700 * 100), POWER_SUPPLY_VOLTAGE) Call Configure_7700_ACV(2 + (I * NUMBER_OF_DUTS) + (SLOT_7700 * 100)) Call Configure_7700_CONT(3 + (I * NUMBER_OF_DUTS) + (SLOT_7700 * 100)) Call Configure_7700_TEMP(4 + (I * NUMBER_OF_DUTS) + (SLOT_7700 * 100)) Next
Call send(ADDRESS, "ROUT:CLOS:ACON ON", Status) ' Place instrument in auto-configure mode. End Sub '' --------------------------------------------------------------------------------------------------------------------­'' 7705 Configurations '' --------------------------------------------------------------------------------------------------------------------­'' Configure_Power_Supplies(,) ­'' Uses the 7705 switch card to change the input voltage to the power supplies. '' They may run at either high line or low line. A load of 10% or 90% is also '' selected in this function. '' '' Assume the following for the 7705: '' 1) Channels 1 to 4 connect each of the possible DUTS to HIGH LINE. '' 2) Channels 11 to 14 connect each of the possible DUTS to LOW LINE. '' 3) Channels 21 to 24 connect each of the possible DUTS to 10% load. '' 4) Channels 31 to 34 connect each of the possible DUTS to 90% load. '' --------------------------------------------------------------------------------------------------------------------­Private Sub Configure_Power_Supplies(Line As Integer, Load As Integer) Dim Min_Channel, Max_Channel As String
' Insert code here to turn off power supply outputs.
Min_Channel = CStr((SLOT_7705 * 100) + 1) Max_Channel = CStr((SLOT_7705 * 100) + 40) Call send(ADDRESS, "ROUT:MULT:OPEN (@" & Min_Channel & ":" & Max_Channel & ")", Status) ' Open all 7705 channels.
Select Case (Line) Case HIGH_LINE Min_Channel = CStr((SLOT_7705 * 100) + 1) Max_Channel = CStr((SLOT_7705 * 100) + NUMBER_OF_DUTS) Call send(ADDRESS, "ROUT:MULT:CLOS (@" & Min_Channel & ":" & Max_Channel & ")", Status)
Case LOW_LINE Min_Channel = CStr((SLOT_7705 * 100) + 1 + 10) Max_Channel = CStr((SLOT_7705 * 100) + NUMBER_OF_DUTS + 10) Call send(ADDRESS, "ROUT:MULT:CLOS (@" & Min_Channel & ":" & Max_Channel & ")", Status) End Select
Select Case (Load) Case TEN_PERCENT_LOAD Min_Channel = CStr((SLOT_7705 * 100) + 1 + 20) Max_Channel = CStr((SLOT_7705 * 100) + NUMBER_OF_DUTS + 20) Call send(ADDRESS, "ROUT:MULT:CLOS (@" & Min_Channel & ":" & Max_Channel & ")", Status)
Case NINETY_PERCENT_LOAD Min_Channel = CStr((SLOT_7705 * 100) + 1 + 30) Max_Channel = CStr((SLOT_7705 * 100) + NUMBER_OF_DUTS + 30) Call send(ADDRESS, "ROUT:MULT:CLOS (@" & Min_Channel & ":" & Max_Channel & ")", Status) End Select
' Insert code here to configure and/ or turn on power supply outputs. End Sub
'' --------------------------------------------------------------------------------------------------------------------­'' Measurement Functions '' --------------------------------------------------------------------------------------------------------------------­'' Take_Measurements() ­'' Closes the appropriate channels and takes a measurement using READ? command. '' ---------------------------------------------------------------------------------------------------------------------
20
Private Sub Take_Measurements() Dim I, J As Integer Dim Channel_Num As Integer Dim Channel_Str As String Dim Reading As String
For I = 1 To NUMBER_OF_DUTS ' Loop to take measurements for all DUTs. Channel_Num = (SLOT_7711 * 100) + I ' Determine channel number. Channel_Str = "(@" & CStr(Channel_Num) & ")" ' Convert channel to ASCII string parameter. Call send(ADDRESS, "ROUT:MULT:CLOS " & Channel_Str, Status) ' Route channel on 7711 to Out A.
For J = 1 To NUMBER_OF_FUNCTIONS ' Loop for each measurement function. Channel_Num = (SLOT_7700 * 100) + J + ((I - 1) * NUMBER_OF_DUTS) Channel_Str = "(@" & CStr(Channel_Num) & ")" Call send(ADDRESS, "ROUT:CLOS " & Channel_Str, Status) ' Close measurement channel on the 7700. Call send(ADDRESS, "READ?", Status) ' Take measurement. Call enter(Reading, MAXLEN, LENGTH, ADDRESS, Status)
Readings(Reading_Counter) = Reading ' Store reading. Reading_Counter = Reading_Counter + 1 Next
' Insert code here to take oscilloscope measurements. Next End Sub
'' --------------------------------------------------------------------------------------------------------------------­'' Main Function '' --------------------------------------------------------------------------------------------------------------------­Private Sub Start_Click() Reading_Counter = 0 ' Reset index counter. Call settimeout(ENTER_TIMEOUT) ' Set timeout valuefor enter function.
Call initialize(21, 0) ' Initialize GPIB.
Call send(ADDRESS, "*RST", Status) ' Restore *RST defaults. Call send(ADDRESS, "INIT:CONT OFF", Status) ' Ensure INIT continuous trigger is off. Call send(ADDRESS, "FORM:ELEM READ,UNIT", Status) ' Display only reading and units.
Call Configure_7700_Channels ' Setup measurement channels on 7700. Call Configure_Power_Supplies(HIGH_LINE, TEN_PERCENT_LOAD) ' DUT input voltage is high line and 10% load. Call Take_Measurements Call Configure_Power_Supplies(HIGH_LINE, NINETY_PERCENT_LOAD) ' DUT input voltage is high line and 90% load. Call Take_Measurements Call Configure_Power_Supplies(LOW_LINE, TEN_PERCENT_LOAD) ' DUT input voltage is low line and 10% load. Call Take_Measurements Call Configure_Power_Supplies(LOW_LINE, NINETY_PERCENT_LOAD) ' DUT input voltage is low line and 90% load. Call Take_Measurements End Sub
21
Service
WARNING All service information is intended only for qualified service personnel. Do not attempt
to service the Model 7711/7712 unless you are qualified to do so.
Performance verification
The following tests are used to verify the RF characteristics of the Model 7711/7712 card. S-parameters are obtained by measuring VSWR and insertion loss among the common output and channel. The following equipment is recommended for the tests in this section:
1. Agilent HP 8753ES Network Analyzer (6GHz) with options 6 and 10.
2. 50 3.5MM test port cables (1m cables).
3. Agilent HP 85033D, 3.5mm Calibration Kit.
4. Phase matched 3.5mm female-female thru-adapter.
Instrument setup
1. Turn on the network analyzer and the mainframe with the Model 7711/7712 installed. Allow the analyzer to warm-up for at least 1 hour.
2. Calibrate the network analyzer from 100MHz to 3.5GHz using the appropriate calibration kit. The calibration routine should include a load, short, open, and pass-through procedure. Use SMA style connectors in the calibration process.
VSWR
1. Connect cables as shown in Figure 10.
2. Close channel 1.
3. Select S11 or S22 measurement on the analyzer.
4. Display VSWR graph on network analyzer.
5. Verify specifications. For the Model 7711, verify at 100MHz, 500MHz, 1GHz, 1.5GHz, and 2GHz. For the Model 7712, verify at 500MHz, 1GHz, 2.5GHz, and 3.5GHz. Refer to the VSWR section of the Model 7711 or 7712 specification values at the end of this User’s Guide for acceptable limits.
6. Repeat procedure for remaining channels. Change the network analyzer connection from OUT A to OUT B when verifying channels 5–8.
22
Figure 10
VSWR and insertion loss verification
Network analyzer
Port 1 Port 2
Model 7711/7712
Ch1
Ch2
OUT A
Ch3
Ch4
Ch5
Ch6
OUT B
Ch7
Ch8
Insertion loss
1. Connect cables as shown in Figure 10.
2. Close channel 1.
3. Select S12 or S21 measurement on the analyzer.
4. Display Log Magnitude plot on the network analyzer.
5. Verify specifications. For the Model 7711, verify at 100MHz, 500MHz, 1GHz, 1.5GHz, and 2GHz. For the Model 7712, verify at 500MHz, 1GHz, 2.5GHz, and 3.5GHz. Refer to the Insertion Loss section of the Model 7711 or 7712 specifications values for acceptable limits.
6. Repeat procedure for remaining channels. Change the network analyzer connection to Port 2 to OUT B when verifying channels 5–8.
23
Cross-talk
1. Connect cables as shown in Figure 11 (Channel 1 to port 1 and channel 2 to port 2, 50 load on OUT A).
2. Close channel 1 (all other channels will open).
3. Display Log Magnitude plot on the network analyzer.
4. Select S12 measurement.
5. Verify specifications. For the Model 7711, verify at 100MHz, 500MHz, 1GHz, 1.5GHz, and 2GHz. For the Model 7712, verify at 500MHz, 1GHz, 2.5GHz, and 3.5GHz. Refer to the Ch-Ch cross-talk section of the Model 7711 or 7712 specifications values for acceptable limits.
6. Repeat procedure for remaining channels in both banks. There are six possible combinations for each bank of 4 channels (combinations in bank 1: 1-2, 1-3, 1-4, 2-3, 3-4). When verifying cross-talk in the second bank, connect the 50 load on OUT B.
Figure 11
Cross-talk verification
Network analyzer
Port 1 Port 2
Model 7711/7712
Ch1
Ch2
OUT A
50 load
Ch3
Ch4
Ch5
Ch6
OUT B
Ch7
Ch8
Replaceable parts
This section contains replacement parts information and the component layout drawing for the Model 7711/7712.
Parts list
Replaceable parts for the Model 7711/7712 are listed in Table 9.
24
Ordering information
To place an order, or to obtain information concerning replacement parts, contact your Keithley representative or the factory (see back cover for addresses). When ordering parts, be sure to include the following information:
• Card model number (Model 7711/7712).
• Card serial number.
• Part description.
• Component designation (if applicable).
• Keithley part number.
Factory service
If the instrument is to be returned to Keithley Instruments for repair, perform the
following:
• Call the Repair Department at 1-888-KEITHLEY for a Return Material Authorization (RMA) number.
• Carefully pack the instrument in the original packing carton.
• Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping label.
Component layout
The component layout for the Model 7711 is provided in Figure 12. Figure 13 is the Model 7712 component layout.
25
Figure 12
Model 7711 component layout
26
Figure 13
Model 7712 component layout
27
Table 9
Model 7711/7712 parts list
Circuit designation Description Keithley part no.
+5VD+5VR,GNDD SURFACE MOUNT PCB TEST POINT CS-1026
C1,C3,C6,C19 CAP, 47P, 5%, 100V, CERAMIC (0805) C-465-47P
C2,C4,C5,C18,C20-C24 CAP, .1UF, 20%, 50V CERAMIC(1206) C-418-.1
C7-C17 CAP, 470U, 20%, 25V, ALUM ELEC C-622-470
CR1,CR2 DIODE MBR5130LT3 RF-115
J100 CONN, RT ANGLE DUAL ROW RECEPT CS-1065-1
J1-J8,J10A,J10B* CONN, RT ANGLE JACK SMA 50 OHM CS-1136
K1-K6: Model 7711 HIGH FREQ RELAY RL-238
Model 7712) RF RELAY RL-268
L1,L2 FERRITE CHIP 600 OHM BLM32A07(1206) CH-62
Q1-Q6 TRANS, NPN SILICON TG-389
R1,R2,R11 RES, 1K, 1%, 100MW, THICK FILM (0805) R-418-1K
R10 RES, 10K, 1%, 100MW, THICK FILM (0805) R-418-10K
R12-R17 RES, 2.55K, 1%, 1W THICK FILM R-418-2.55K
R3-R8 RES, 43.2, 1%, 100MW THICK FILM (0805) R-418-43.2
R9 RES, 357K, 1%, 100MW, THICK FILM(0805) R-418-357K
U1,U2 IC, QUAD 2 IN AND, 74HCT08(SOIC) IC-837
U10 IC, 8 STAGE SHIFT/STORE,MC14094BD(SOIC) IC-772
U3 IC, 2.5V CASCADABLE SERIAL EEPROM LSI-212
U5 IC, +5V VOLTAGE REGULATOR, LM2940CT IC-576
U7 IC, DARLINGTON ARRAY, ULN2003L (SOIC) IC-969
U8 IC, RETRIG., MULTIVIB, 74HC123AM (SOIC) IC-788
U9 IC, HEX SCHMITT INVERT TRIGGER IC-1397
Hardware Model 7711/7712
Model 7711 Specific hardware
Model 7712 Specific hardware
MODIFIED TOP COVER V-7711-304A
MODIFIED BOTTOM COVER V-7711-303A
STANDOFF MALE-FEMALE ST-166-25
BAR CODE S/N LABEL MC-612A
SERIAL NUMBER LABEL MC-285
4-40 PEM NUT FA-131
MODIFIED TOP COVER 7711-304A
MODIFIED BOTTOM COVER 7711-303A
TOP CARD COVER 7703-302C
4-40X3/16 PHIL. PAN HD SEMS 4-40X3/16PPHSEM
#4-40 X 1/4 LG. PHIL.FLAT HD.SCREW 4-40X1/4PFH
#4-40 X 1/4 LG. PHIL.FLAT HD.SCREW 4-40X1/4PFH
BOTTOM CARD COVER 7703-301B
SCANNER BOARD ASSEMBLY 7711-100B
CONNECTOR BRACKET 7711-301-1A
SCANNER BOARD ASSEMBLY 7712-100A
CONNECTOR BRACKET 7711-301-2A
28
HW 2/05/02
Rev. A
of 2
7711 50Ω 2GHz Multiplexer Specifications
GENERAL
Relay Type: High Frequency Electromechanical Contact Configuration: Dual 1x4 Multiplexer, single pole four
throw, Channels 1 and 5 are normally closed.
Notes: One channel in each multiplex bank is always closed to
the corresponding OUT connector.
Close Channel: ROUTe:CLOSe allows a single channel in a
multiplex bank to be closed. ROUTe:MULTiple:CLOSe allows two channels (One in each bank) to be closed at one time.
Open Channel: ROUTe:OPEN:ALL closes CH1 and CH5 to
OUT A and OUT B respectively.
Actuation Time: <10mS Firmware: Specified for Model 2700 rev. B04 and
Model 2750 rev. A03 or higher.
Connector Type: Ten External Rear Panel SMA connectors. Mating Torque: 0.9 Nm (8 in-lb).
INPUTS (CHANNELS 1-8)
Maximum Signal Level: Any channel to any channel or chassis (1-8)
Maximum Power: 20W per module, 10W per channel.* Safety: Conforms to European Union Directive 73/23/EEC
EN61010-1, CAT I.
EMC: Conforms with European Union Directive 89/336/EEC;
EN61326-1.
Isolation:
Multiplexer to Multiplexer: >1GΩ Center to Shield: >1GΩ, < 25pF Channel to Channel: >100MΩ
Contact Life: 1x10 Contact Potential: <6µV Contact Resistance: <0.5Ω (Initial), <1Ω (End Of Life) Rise Time: <300ps (Guaranteed by design) Signal Delay: <3ns
MODEL ACCESSORIES AVAILABLE
7711-SMA-BNC Male SMA to Female BNC
Cables (Five, 0.15m (0.5ft) cables)
7051-2 BNC cable, male to male, 0.6m (2ft.) 7051-5 BNC cable, male to male, 1.5m (5ft.) 7051-10 BNC cable, male to male, 3.0m (10ft.) 7712-SMA-1 SMA cable, male to male, 1.0m (3.3 ft.) 7712-SMA-N Female SMA to Male N-Type Adapter S46-SMA-1 SMA cable, male to male, 0.3m (1 ft.) S46-SMA-0.5 SMA cable, male to male, 0.15m (0.5 ft.)
* Refer to 7711/7712 User’s Guide (PA-818) for measurement
considerations.
30Vrms (42V peak for AC waveforms) or 60VDC, 0.5A.
6
No Load, 1x105 Rated Load (resistive load)
Frequency (MHz)
Frequency (MHz)
Frequency (MHz)
Page 1
HW 2/05/02
Rev. A
of 2
100MHz
7711 50Ω 2GHz Multiplexer Specifications
AC Performance: (Includes end of life)
For Z
= Z
= 50
source
<<
0.4dB 0.6dB 1.0dB 1.2dB 2.0dB
1.1 1.2 1.2 1.3 1.7
-85dB -65dB -55dB -45dB -35dB
Specifications subject to change without notice.
load
Insertion
Loss
MAX
VSWR
MAX
Ch-Ch
Cross-talk1
MAX
1
Specification assumes 50 termination.
Environmental
Operating Environment: Specified for 0°C to 50°C
Specified for 80% RH at 35°C
Storage Environment: −25°C to 65°C Weight: <0.50Kg (1.1 lb.)
500MHz 1GHz 1.5GHz 2GHz
Page 2
HW 2/05/02
Rev. A
of 2
7712 50Ω 3.53.5GHz Multiplexer
GENERAL
Relay Type: High Frequency Electromechanical Contact Configuration: Dual 1x4 Multiplexer, single pole four
throw, Channels 1 and 5 are normally closed
Notes: One channel in each multiplex bank is always closed to the
corresponding OUT connector.
Close Channel: ROUTe:CLOSe allows a single channel in a
multiplex bank to be closed. ROUTe:MULTiple:CLOSe allows two channels (One in each bank) to be closed at one time.
Open Channel: ROUTe:OPEN:ALL closes CH1 and CH5 to OUT A
and OUT B respectively.
Actuation Time: <10mS Firmware: Specified for Model 2700 rev. B04 and
Model 2750 rev. A03 or higher.
Connector Type: Ten External Rear Panel SMA connectors Mating Torque: 0.9 Nm (8 in-lb).
INPUTS (CHANNELS 1-8)
Maximum Signal Level: Any channel to any channel or chassis (1-8)
Maximum Power: 20W per module. 10W per channel.* Safety: Conforms to European Union Directive 73/23/EEC
EN61010-1, CAT I.
EMC: Conforms with European Union Directive 89/336/EEC;
EN61326-1.
Isolation:
Multiplexer to Multiplexer: >1GΩ Center to Shield: >1GΩ, <20pF Channel to Channel: >100MΩ
Contact Life: 5x10 Contact Potential: <12µV Contact Resistance: <0.5Ω (Initial), <1Ω (End Of Life) Rise Time: <200ps (Guaranteed by design) Signal Delay: <1.5ns
MODEL ACCESSORIES AVAILABLE
7712-SMA-1 SMA cable, male to male, 1m (3.3 ft) 7712-SMA-N Female SMA to Male N-Type Adapter S46-SMA-1 SMA cable, male to male, 0.3m (1ft) S46-SMA-0.5 SMA cable, male to male, 0.15m (0.5 ft.)
* Refer to 7711/7712 User’s Guide (PA-818) for measurement
considerations.
30Vrms (42V peak for AC waveforms) or 42VDC, 0.5A.
6
No Load, 1x105 Rated Load (Resistive load)
Frequency (MHz)
Frequency (MHz)
Frequency (MHz)
Page 1
HW 2/05/02
Rev. A
of 2
AC Performance: (Includes end of life)
For Z
= Z
source
MAX
VSWR
MAX
Ch-Ch
Crosstalk
MAX
= 50
1
<<500MHz
0.5dB 0.65dB 1.1dB 1.3dB
1.15 1.2 1.45 1.45
-75dB -70dB -50dB -45dB
1GHz 2.4GHz 3.5GHz
load
Insertion Loss
1
Specification assumes 50 termination.
ENVIRONMENTAL
Operating Environment: Specified for 0°C to 50°C
Specified for 80% RH at 35°C
Storage Environment: −25°C to 65°C Weight: <.50Kg (1.1 lb.)
Specifications subject to change without notice.
Page 2
Specifications are subject to change without notice.
All Keithley trademarks and trade names are the property of Keithley Instruments, Inc. All other trademarks and trade names are the property of their respective companies.
Keithley Instruments, Inc. 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168
1-888-KEITHLEY (534-8453) • www.keithley.com
Sales Offices: BELGIUM: Bergensesteenweg 709 • B-1600 Sint-Pieters-Leeuw • 02-363 00 40 • Fax: 02-363 00 64
CHINA: Yuan Chen Xin Building, Room 705 • 12 Yumin Road, Dewai, Madian • Beijing 100029 • 8610-82251886 • Fax: 8610-82251892 FINLAND: Halsuantie 2 • 00420 Helsinki, Finland • 09-53 06 65 60 • Fax: 09-53 06 65 65 FRANCE: 3, allée des Garays • 91127 Palaiseau Cédex • 01-64 53 20 20 • Fax: 01-60 11 77 26 GERMANY: Landsberger Strasse 65 • 82110 Germering • 089-84 93 07-40 • Fax: 089-84 93 07-34 GREAT BRITAIN: Unit 2 Commerce Park, Brunel Road • Theale, Berkshire RG7 4AB • 0118 -929 75 00 • Fax: 0118- 929 75 19 INDIA: 1/5, Eagles Street • Langford Town • Bangalore 560 025 • 080 212 80-27 • Fax: 080 212 80 05 ITALY: Viale San Gimignano, 38 • 20146 Milano • 02-48 39 16 01 • Fax: 02-48 30 22 74 JAPAN: New Pier Takeshiba North Tower 13F • 11-1, Kaigan 1-chome • Minato-ku, Tokyo 105-0022 • 81-3-5733-7555 • Fax: 81-3-5733-7556 KOREA: 2FL., URI Building • 2-14 Yangjae-Dong • Seocho-Gu, Seoul 137-888 • 82-2-574-7778 • Fax: 82-2-574-7838 NETHERLANDS: Postbus 559 • 4200 AN Gorinchem • 0183-63 53 33 • Fax: 0183-63 08 21 SWEDEN: c/o Regus Business Centre • Frosundaviks Allé 15, 4tr • 16970 Solna • 08-50 90 46 00 • Fax: 08-655 26 10 TAIWAN: 13F-3, NO. 6, Lane 99, Pu-Ding Road, Hsinchu, Taiwan, ROC. • 886-3-572-9077 • Fax: 886-3-572-9031