Keithley 7174A Service manual

Page 1
I
nstruction Manua
l
Model 7174A
Contains Operating and Servicing Information
7174A-901-01 Rev. A / 9-98
Page 2
W ARRANTY
Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty , write or call your local Keithley representative, or contact Keithle y headquarters in Cleveland, Ohio. Y ou will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the origi­nal warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDI­RECT , SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
Keithley Instruments, Inc. • 28775 Aurora Road • Cleveland, OH 44139 • 440-248-0400 • Fax: 440-248-6168 • http://www.keithley.com
CHINA: Keithley Instruments China • Yuan Chen Xin Building, Room 705 • 12 Yumin Road, Dewai, Madian • Beijing 100029 • 8610-62022886 • Fax: 8610-62022892 FRANCE: Keithley Instruments SARL • BP 60 • 3 Allée des Garays • 91122 Palaiseau Cédex • 33-1-60-11-51-55 • Fax: 33-1-60-11-77-26 GERMANY: Keithley Instruments GmbH • Landsberger Strasse 65 • D-82110 Germering, Munich • 49-89-8493070 • Fax: 49-89-84930759 GREAT BRITAIN: Keithley Instruments, Ltd. • The Minster • 58 Portman Road • Reading, Berkshire, England RG3 1EA • 44-1189-596469 • Fax: 44-1189-575666 ITALY: Keithley Instruments SRL • Viale S. Gimignano 38 • 20146 Milano • 39-2-48303008 • Fax: 39-2-48302274 NETHERLANDS: Keithley Instruments BV • Avelingen West 49 • 4202 MS Gorinchem • 31-(0)183-635333 • Fax: 31-(0)183-630821 SWITZERLAND: Keithley Instruments SA • Kriesbachstrasse 4 • 8600 Dübendorf • 41-1-8219444 • Fax: 41-1-8203081 TAIWAN: Keithley Instruments Taiwan • 1FL., 85 Po Ai Street • Hsinchu, Taiwan • 886-3-572-9077 • Fax: 886-3-572-9031
Page 3
Model 7174A 8 × 12 Low Current Matrix Card
Instruction Manual
©1998, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
First Printing September 1998
Document Number: 7174A-901-01 Rev. A
Page 4
Manual Print History
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revi­sions, contain important change information that the user should incorporate immediately into the manual. Addenda are num­bered sequentially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 7174A-901-01)............................................................................ September 1998
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand and product names are trademarks or registered trademarks of their respective holders.
Page 5

Safety Precautions

The following safety precautions should be observed before using this product and any associated instrumentation. Although some in­struments and accessories would normally be used with non-haz­ardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recog­nize shock hazards and are familiar with the safety precautions re­quired to avoid possible injury. Read the operating information carefully before using the product.
The types of product users are:
Responsible body is the individual or group responsible for the use
and maintenance of equipment, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be
trained in electrical safety procedures and proper use of the instru­ment. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product
to keep it operating, for example, setting the line voltage or replac­ing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator may per­form them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform
safe installations and repairs of products. Only properly trained ser­vice personnel may perform installation and service procedures.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present.
that hazardous voltage is present in any unknown circuit before measuring.
A good safety practice is to expect
Users of this product must be protected from electric shock at all times. The responsible body must ensure that users are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product users in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts,
exposed.
As described in the International Electrotechnical Commission (IEC) Standard IEC 664, digital multimeter measuring circuits (e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are Installation Category II. All other instruments’ signal terminals are Installation Category I and must not be connected to mains.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC mains. When con­necting sources to switching cards, install protective devices to lim­it fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connect­ed to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jump­ers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
no conductive part of the circuit may be
Page 6
Do not exceed the maximum signal levels of the instruments and ac­cessories, as defined in the specifications and operating informa­tion, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is ap­plied to the device under test. Safe operation requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should re­fer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or mea­sure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING heading in a manual explains dangers that might
result in personal injury or death. Alw ays read the associated infor ­mation very carefully before performing the indicated procedure.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instru­ments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that se­lected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
To clean the instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument.
The
CAUTION heading in a manual explains hazards that could
damage the instrument. Such damage may invalidate the warranty.
Page 7
7174A 8×12 Low Current Matrix Card Specifications
MATRIX CONFIGURATION: Single 8 rows×12 columns. Expanding
the columns can be done internally by connecting the rows of multiple 7174A cards together with coax jumpers.
CROSSPOINT CONFIGURATION: 2-pole Form A (Signal Guard). CONNECTOR TYPE: 3-lug triax (Signal, Guard, Chassis). MAXIMUM SIGNAL LEVEL:
Pin to Pin or Pin to Chassis: 200V. 2A carry current.
CONTACT LIFE: Cold Switching: 10
8
closures.
OFFSET CURRENT: 100fA max., 10fA typical (with 0V applied to
inputs and outputs).
ISOLATION: Path (Signal to Signal): >2×10
14
, 1pF.
Common (Signal to Chassis): >1014Ω, <10pF.
SETTLING TIME: <2.5s to 400fA (all pathways) after 10V applied
(typical).
CROSSTALK (1MHz,50Ω Load): <–70dB. INSERTION LOSS (1MHz, 50ΩLoad): <–0.2dB typical. 3dB BANDWIDTH:
(50Load, 50Source): 30MHz typical. (1MLoad, 50Source): 40MHz typical.
RELAY DRIVE CURRENT (per crosspoint): 17mA. RELAY SETTLING TIME: <1ms.
ENVIRONMENT:
Offset Current and Path Isolation Specifications: 23°C, <60%
R.H.
Operating: 0° to 50°C, up to 35°C at 70% R.H. Storage: –25° to +65°C.
MAXIMUM LEAKAGE:
Pin to Ground: 0.01pA/V. Pin to Pin: 0.005pA/V.
INSULATION RESISTANCE: 6.7×10
13
minimum.
CAPACITANCE: (Guard Driven): Path to Ground: <10pF. Path to
Path: 1pF typical.
ACCESSORY SUPPLIED: Instruction manual and eight MCX expan-
sion cables.
ACCESSORIES AVAILABLE:
7078-TRX-TBC 3-Lug Triax to BNC Adapter 7078-TRX-T 3-Lug Triax Tee Adapter 7078-TRX-3 3-Lug Triax Cable, 0.9m (3 ft.) 7078-TRX-10 3-Lug Triax Cable, 3m (10 ft.) 7078-TBC 3-Lug Female Triax Bulkhead Connector with Cap 7078-CSHP Cable Set to Connect 7174 to HP 4145, 4155, 4156
Specifications are subject to change without notice.
Columns
HGCHGCHGCHGCHGCHGCHGCHGCHGCHGCHGCHGC
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
222
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
222
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
User
connections
and expansion
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
H G C
H G C
H G C
H G C
H G C
H G C
H G C
H G C
Rows
Page 8

Table of Contents

1 General Information
1.1 Introduction ....................................................................................................................................................... 1-1
1.2 Features ............................................................................................................................................................. 1-1
1.3 Warranty information ........................................................................................................................................ 1-1
1.4 Manual addenda ................................................................................................................................................ 1-2
1.5 Safety symbols and terms ................................................................................................................................. 1-2
1.6 Specifications .................................................................................................................................................... 1-2
1.7 Unpacking and inspection ................................................................................................................................. 1-2
1.7.1 Inspection for damage ............................................................................................................................... 1-2
1.7.2 Shipment contents ..................................................................................................................................... 1-2
1.7.3 Instruction manual ..................................................................................................................................... 1-2
1.8 Packing for shipment ........................................................................................................................................ 1-2
1.9 Optional accessories .......................................................................................................................................... 1-2
2 Operation
2.1 Introduction ....................................................................................................................................................... 2-1
2.2 Handling precautions ........................................................................................................................................ 2-1
2.3 Environmental considerations ........................................................................................................................... 2-1
2.4 Card installation and removal ........................................................................................................................... 2-2
2.5 Connections ....................................................................................................................................................... 2-2
2.5.1 Card connectors .......................................................................................................................................... 2-2
2.5.2 Recommended cables and adapters ............................................................................................................ 2-3
2.5.3 Triax banana plug adapter .......................................................................................................................... 2-4
2.5.4 General instrument connections ................................................................................................................. 2-5
2.5.5 Keithley instrument connections .............................................................................................................. 2-11
2.5.6 Typical test fixture connections ............................................................................................................... 2-17
2.6 Matrix configuration ....................................................................................................................................... 2-18
2.6.1 Switching matrix ...................................................................................................................................... 2-18
2.6.2 Path isolators ............................................................................................................................................ 2-18
2.6.3 Internal matrix expansion ......................................................................................................................... 2-21
2.7 Measurement considerations ........................................................................................................................... 2-22
2.7.1 Magnetic fields ........................................................................................................................................ 2-22
2.7.2 Electromagnetic Interference (EMI) ....................................................................................................... 2-22
2.7.3 Ground loops ........................................................................................................................................... 2-22
2.7.4 Keeping connectors clean ....................................................................................................................... 2-23
2.7.5 Noise currents caused by cable flexing ................................................................................................... 2-23
2.7.6 Shielding ................................................................................................................................................. 2-23
2.7.7 Guarding .................................................................................................................................................. 2-24
2.7.8 Matrix expansion effects on card specifications ..................................................................................... 2-24
2.8 Coaxial jumper access ..................................................................................................................................... 2-25
i
Page 9
3 Applications
3.1 Introduction ........................................................................................................................................................ 3-1
3.2 CV measurements ............................................................................................................................................... 3-1
3.2.1 Stand alone system configuration ............................................................................................................... 3-1
3.2.2 Computerized system configuration ........................................................................................................... 3-1
3.2.3 Optimizing CV measurement accuracy ...................................................................................................... 3-3
3.2.4 Basic CV test procedure ............................................................................................................................. 3-3
3.2.5 Typical CV curves ...................................................................................................................................... 3-3
3.3 Semiconductor test matrix ................................................................................................................................. 3-5
3.3.1 System configuration .................................................................................................................................. 3-5
3.3.2 Testing common-source characteristic of FETs ......................................................................................... 3-6
3.4 Resistivity measurements ................................................................................................................................... 3-7
3.4.1 Test configuration ....................................................................................................................................... 3-7
3.4.2 Test procedure ............................................................................................................................................ 3-7
3.4.3 Resistivity calculations ............................................................................................................................... 3-9
3.5 Semiconductor IV characterization .................................................................................................................... 3-9
3.5.1 Test configuration ....................................................................................................................................... 3-9
3.5.2 Cable connections ..................................................................................................................................... 3-10
4 Service Information
4.1 Introduction ........................................................................................................................................................ 4-1
4.2 Handling and cleaning precautions .................................................................................................................... 4-1
4.3 Principles of operation ........................................................................................................................................ 4-2
4.3.1 Block diagram ............................................................................................................................................ 4-2
4.3.2 ID data circuits ........................................................................................................................................... 4-2
4.3.3 Relay control .............................................................................................................................................. 4-3
4.3.4 Power-on sequence ..................................................................................................................................... 4-3
4.3.5 Isolator relays ............................................................................................................................................. 4-4
4.4 Troubleshooting .................................................................................................................................................. 4-4
4.4.1 Recommended equipment .......................................................................................................................... 4-4
4.4.2 Gaining circuit access ................................................................................................................................. 4-4
4.4.3 Troubleshooting procedure ......................................................................................................................... 4-4
4.5 Special handling of static-sensitive devices ....................................................................................................... 4-5
4.6 Performance verification .................................................................................................................................... 4-5
4.6.1 Environment conditions .............................................................................................................................. 4-5
4.6.2 Recommended test equipment .................................................................................................................... 4-5
4.6.3 Offset current verification .......................................................................................................................... 4-6
4.6.4 Path isolation verification ........................................................................................................................... 4-7
4.6.5 Path resistance verification ......................................................................................................................... 4-9
4.7 Reed pack replacement...................................................................................................................................... 4-11
5 Replaceable Parts
5.1 Introduction ........................................................................................................................................................ 5-1
5.2 Parts list .............................................................................................................................................................. 5-1
5.3 Ordering information .......................................................................................................................................... 5-1
5.4 Factory service .................................................................................................................................................... 5-1
5.5 Component layout and schematic diagram ......................................................................................................... 5-1
ii
Page 10

List of Illustrations

2 Operation
Figure 2-1 Model 7174A installation .......................................................................................................................... 2-2
Figure 2-2 Card connectors ......................................................................................................................................... 2-3
Figure 2-3 Triax connector configuration ................................................................................................................... 2-3
Figure 2-4 Triax cable preparation ............................................................................................................................. 2-4
Figure 2-5 General instrument connections ................................................................................................................ 2-6
Figure 2-6 Model 617 electrometer connections ...................................................................................................... 2-11
Figure 2-7 Model 196 DMM connections ................................................................................................................ 2-12
Figure 2-8 Model 230 voltage source connections ................................................................................................... 2-13
Figure 2-9 Model 590 CV analyzer connections ...................................................................................................... 2-14
Figure 2-10 Model 220 current source connections ................................................................................................... 2-15
Figure 2-11 Model 236/237/238 source measure unit connections ............................................................................ 2-16
Figure 2-12 Typical test fixture connections .............................................................................................................. 2-17
Figure 2-13 Equivalent circuit of test fixture connections .......................................................................................... 2-18
Figure 2-14 Matrix configuration ............................................................................................................................... 2-19
Figure 2-15 Connecting three cards for an 8 × 36 matrix ........................................................................................... 2-21
Figure 2-16 Jumper connector locations ..................................................................................................................... 2-21
Figure 2-17 Two cards in daisy chain configuration .................................................................................................. 2-22
Figure 2-18 Power line ground loops ......................................................................................................................... 2-23
Figure 2-19 Eliminating ground loops ........................................................................................................................ 2-23
Figure 2-20 Shielded and guarded .............................................................................................................................. 2-24
Figure 2-21 Guarded circuit ........................................................................................................................................ 2-24
Figure 2-22 Coaxial jumper access ............................................................................................................................. 2-25
3 Applications
Figure 3-1 Stand alone CV system configuration ........................................................................................................ 3-2
Figure 3-2 Computerized CV system configuration .................................................................................................... 3-2
Figure 3-3 Typical quasistatic CV curve generated by Model 595 ............................................................................. 3-4
Figure 3-4 Typical high-frequency CV curve generated by Model 590 ...................................................................... 3-4
Figure 3-5 Semiconductor test matrix ......................................................................................................................... 3-5
Figure 3-6 System configuration for measuring common-emitter characteristics ....................................................... 3-6
Figure 3-7 Typical common-source FET IV characteristics ........................................................................................ 3-6
Figure 3-8 Resistivity test configuration ..................................................................................................................... 3-7
Figure 3-9 Resistivity measurement conventions ........................................................................................................ 3-8
Figure 3-10 Multi-unit test system using Models 236 and 237 source measure units ................................................. 3-10
iii
Page 11
4 Service Information
Figure 4-1 Model 7174A block diagram ...................................................................................................................... 4-2
Figure 4-2 ID data timing ............................................................................................................................................. 4-3
Figure 4-3 Offset verification test connections ............................................................................................................ 4-6
Figure 4-4 Connections for path isolation verification ................................................................................................ 4-7
Figure 4-5 Triaxial cable preparation ........................................................................................................................... 4-8
Figure 4-6 Connections for path resistance verification .............................................................................................. 4-9
Figure 4-7 Shorting measurement paths using triax tee adapter ................................................................................ 4-10
Figure 4-8 Cross point relays ..................................................................................................................................... 4-12
Figure 4-9 Isolator relays ........................................................................................................................................... 4-13
iv
Page 12

List of Tables

2 Operation
Table 2-1 Recommended cables and adapters ............................................................................................................ 2-3
Table 2-2 Parts for special triaxial cable .................................................................................................................... 2-4
Table 2-3 Column numbering by slot and unit ......................................................................................................... 2-20
3 Applications
Table 3-1 CV test crosspoint summary ...................................................................................................................... 3-3
Table 3-2 Crosspoint summary for resistivity measurements .................................................................................... 3-9
4 Service Information
Table 4-1 Recommended troubleshooting equipment ................................................................................................ 4-4
Table 4-2 Troubleshooting procedure ........................................................................................................................ 4-4
Table 4-3 Recommended verification equipment ...................................................................................................... 4-5
5 Replaceable Parts
Table 5-1 Model 7174A electrical parts list ................................................................................................................5-3
Table 5-2 Model 7174A mechanical parts list ........................................................................................................... 5-4
v
Page 13
1

General Information

1.1 Introduction

This section contains general information about the Model 7174A Low Current Matrix Card. The Model 7174A Low Current Matrix Card is designed for semiconductor research, development, and production applications that require high quality and performance switching I-V (current versus volt­age) and C-V (capacitance versus voltage) signals. The model 7174A is ideal for use with Keithley Model 236 Source Measure Unit for semiconductor testing and other low current switching applications. Model 237 and Model 238 Source Measure Units can also be used below the maxi­mum signal level (200V, 2A carry) of the Model 7174A Low Current Matrix Card (for full specifications, refer to para­graph 1.6). The Model 7174A also can be used with Models 590 and 595 C-V instruments.
Section 1 is arranged in the following manner:

1.2 Features

1.3 Warranty information

1.4 Manual addenda
1.5 Safety symbols and terms
1.6 Specifications
1.7 Unpacking and inspection
1.8 Packing for shipment
1.9 Optional accessories
1.2 Features
Key features of the Model 7174A Low Current Matrix Card include:
• Eight row by twelve column (8 × 12) switching matrix configuration, with signal and guard switched at each crosspoint
• Paths have offset currents of less than 100fA with typi­cal offset currents of 50fA
• Maximum Leakage Currents: Pin to Ground -- 0.01 pA/V
Pin to Pin -- 0.005 pA/V
• 3-lug Triaxial Connectors (Signal, Guard, Chassis) for all row and columns allow guarding of each signal path­way, minimizing effects of stray capacitance, leakage current, and leakage resistance
• Model 7174A cards can be connected together inter­nally using the supplied SMB to SMB cables (jumpers) to expand the number of columns in the matrix.
1.3 W arranty information
Warranty information is located on the inside front cover of this manual. Should your Model 7174A require warranty service, contact your Keithley representative or authorized repair facility in your area for further information.
1-1
Page 14
General Information

1.4 Manual addenda

Any improvements or changes concerning the matrix card or manual will be explained in an addendum included with the unit. Be sure to note these changes and incorporate them into the manual before using or servicing the unit.

1.5 Safety symbols and terms

The following symbols and terms may be found on an instru­ment or used in this manual.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the in­struction manual.
The symbol on an instrument shows that high voltage may be present on the terminal(s). Use standard safety pre­cautions to avoid personal contact with these voltages.
The WARNINGS heading used in this manual explains dan­gers that might result in personal injury or death. Always read the associated information very carefully before per­forming the indicated procedure.
1.7.2 Shipment contents
The following items are included with every Model 7174A order:
• Model 7174A Low Current Matrix Card.
• Model 7174A Instruction Manual.
• Coaxial jumper cables Model CA-121A (8) for matrix expansion.
• Additional accessories as ordered.
1.7.3 Instruction manual
The Model 7174A Instruction Manual is three-hole drilled so that it can be added to the system three-ring binder. After removing the plastic wrapping, place the manual in the binder after the mainframe instruction manual. Note that a manual identification tab is included and should precede the matrix card instruction manual.
If an additional manual is required, order the manual pack­age (Keithley part number 7174A-901-00). The manual package includes an instruction manual and any pertinent addenda.
The CAUTION heading used in this manual explains haz­ards that could damage the matrix card. Such damage may invalidate the warranty.
1.6 Specifications
Model 7174A specifications may be found at the front of this manual. These specifications are exclusiv e of the matrix card file specifications, which are located in the Model 707A Switching Matrix manual.

1.7 Unpacking and inspection

1.7.1 Inspection for damage
If you ordered the Model 7174A separately from a system, carefully unpack it from its shipping carton and inspect the card for any obvious signs of physical damage. Report any such damage to the shipping agent immediately. Save the original packing carton for possible future reshipment.

1.8 Packing for shipment

Should it become necessary to return the Model 7174A for repair, carefully pack the card in its original packing carton or the equivalent, and include the following information:
• Advise as to the warranty status of the matrix card.
• Write ATTENTION REPAIR DEPARTMENT on the shipping label.
• Fill out and include the service form located at the back of this manual.

1.9 Optional accessories

Cables
Model 7078-TRX-3 — A 0.9 m (3 ft.) triaxial cable termi-
nated at both ends with 3-slot male triax connectors. This type of cable is also available in 10 ft. (Model 7078-TRX-10) and 20 ft. (Model 7078-TRX-20) lengths.
Model CA-93-1 — BNC to right angle SMB (coaxial) cable.
1-2
Page 15
General Information
Model 7078-CSHP — Is a cable set containing:
Eight 10 ft. (3m) cables — Cables to connect the Model 7174A to a HP-4145 Semiconductor Parameter Analyzer.
Four BNC to triax adapters — Used with eight cables listed above
Four 3 slot triax cables— Cable set to connect the Model 7174A to Source Measurement Units
Four BNC to BNC coax cables — Adapt the measurement and source modules in the HP-4145 to the connectors of the Model 7174A.
Adapters
Model 7078-TRX-TBC — A 3-lug female triax bulkhead
connector (with cap). Use this connector for custom applica­tions and interface connections such as test fixtures.
Model 7078-TRX-T — 3-slot male to dual 3-lug female
triax tee adapter.
Model 237-BAN-3 — 3-slot male triax to male banana plug.
Model 237-ALG-2 — 3-slot male triax to alligator clips.
Model 7078-TRX-BNC — 3-slot male triax to BNC
adapter, connections to center and inner shell. For non­guarded applications, use Model 7078-TRX-GND.
Model 6171 — 3-slot male triax to 2-lug female triaxial
adapter.
Tools
Model 9172-314 — A tool designed to remove and install
internal coaxial jumpers between adjacent Model 7174A Low Current Matrix Cards.
1-3
Page 16
2

Operation

2.1 Introduction

WARNING
The installation and operation proce­dures in this section are intended for use only by qualified service personnel. Do not perform these procedures unless qualified to do so. Failure to recognize and observe normal safety precautions could result in personal injury or death.
This section contains information on matrix card connec­tions, installation and matrix programming, and is arranged as follows:
2.2 Handling precautions — Discusses precautions that
should be taken when handling the card to avoid con­tamination that could degrade performance.
2.3 Environmental considerations — Outlines environ-
mental aspects of using the Model 7174A.
2.4 Card installation and removal — Details installation
in and removal from the Model 707A Switching Matrix.
2.7 Measurement considerations — Reviews a number
of considerations when making low-level current and capacitance measurements.
2.8 Coaxial jumper access — Provides information on
jumper removal.

2.2 Handling precautions

To maintain high impedance isolation, care should be taken when handling the matrix card to avoid contamination from such foreign materials as body oils. Such contamination can substantially lower leakage resistance, degrading perfor­mance. The areas of the card that are most sensitive to con­tamination are those associated with Teflon® insulators. To avoid any possible contamination, always grasp the card by the handle or the card edges. Do not touch board surfaces, components, or card edge connectors.
Dirt and other particle build-up over a period of time are other possible sources of contamination. To avoid this prob­lem, operate the mainframe and matrix card only in a clean environment. If contamination is suspected, clean the card as discussed in Section 4.
2.5 Connections — Discusses card connectors, cables
and adapters, and typical connections to other instrumentation.
2.6 Matrix configuration — Discusses the switching
matrix, as well as matrix expansion by connecting two or more cards together.

2.3 Environmental considerations

For rated performance, the card should be operated within the temperature and humidity limits given in the specifica­tions at the front of this manual.
2-1
Page 17
Operation

2.4 Card installation and removal

Before making connections, the Model 7174A should be installed within the Model 707A Switching Matrix, as sum­marized below. Figure 2-1 shows the installation procedure.
WARNING
T urn off the system power before install­ing or removing matrix cards.
NOTE
The coaxial jumpers used to expand the matrix with two or more Model 7174A cards can not be installed before card insertion; an access door on top of the mainframe allows access to the connectors after the card is installed.
2. With one hand grasping the handle, and the other hold­ing the bottom of the card, line up the card with the tracks in the desired slot. Make certain that the compo­nent side of the card is facing the fan on the mainframe.
3. Slide the card into the mainframe until it is properly seated in the edge connectors at the back of the slot. Once the card is properly seated, secure it to the main­frame by finger tightening the spring-loaded screws.
WARNING
The mounting screws must be secured to ensure proper chassis ground connec­tions between the card and the main­frame. Failure to properly secure this ground connection may result in per­sonal injury or death due to electric shock.
4. To remove a card, first turn off the system power. Dis­connect all external and internal jumper cables (internal cables can be reached through the access door). Loosen the mounting screws, then pull the card out of the main­frame by the handle. When the back edge of the card clears the mainframe, support it by grasping the bottom edge near the back or back edge.
Model 707A
Switching Matrix
Model 7174A Low
Current Matrix Card
Figure 2-1
Model 7174A installation
1. Before installing the card, make sure the access door on top of the Model 707A Switching Matrix is fully closed and secured. The access door contains tracks for the card slots and must be in place to properly install the card.
CAUTION
Do not touch the card surfaces or any components to avoid contamination that could degrade card performance.

2.5 Connections

Card connectors, recommended cables and adapters, and typical connections to test instruments are discussed in the following paragraphs.
2.5.1 Card connectors
The card connectors are shown in Figure 2-2. Each pin is equipped with a 3-lug triax connector. As shown in Figure 2-3, the center conductor is signal, the inner shield is guard, and the outer shield is chassis ground.
CAUTION
Do not exceed 200V between any two pins or between any pin and chassis.
The Model 7174A has 12 columns (labeled 1 through 12) and 8 rows (labeled A through H).
2-2
Page 18
Chassis Ground
200V
Peak
Guard
200V Peak
Signal
200V
Peak
Caution:
Do not exceed maximum voltage levels shown.
Figure 2-3
Triax connector configuration
Operation
Mounting
Screw
KEITHLEY
7174A
8x12LOW
CURRENT MATRIX
SIGNAL
200VPK
200VPK
Carrying Handle
B
C
D
A
200VPK
ROWS
COLUMNS
GUARD
1
2
3
4
5
6
Caution:
Remove internal row jumpers before removing card.
Figure 2-2
Card connectors
Mounting
Screw
E
F
10
G
11
H
12
!
WARNING:
TIGHTEN MOUNTING SCREWS
TO ENSURE PROPER
CHASSIS GROUND
MADE INU.S.A.
7
2.5.2 Recommended cables and adapters
8
Table 2-1 summarizes the cables recommended for use with
9
the Model 7174A. Equivalent user-supplied items may be substituted as long as they are of sufficient quality (low of fset current, high leakage resistance). Using substandard cables and connectors may degrade the integrity of the measure­ments made. See paragraph 2.7 for a discussion of measure­ment considerations.
Table 2-1
Recommended cables and adapters
Model Description
7078-TRX-x
3-slot male triax connectors on both ends (x=3, 10 or 20 ft.)
237-BAN-3
3-slot male triax to male banana
plug 237-ALG-2 7078-TRX-BNC
3-slot male triax to alligator clips
3-slot male triax to BNC adapter,
connections to center and inner
shell 7078-TRX-GND
3-slot male triax to BNC adapter,
connections to center and outer
shell 7078-TRX-T
3-slot male to dual 3-lug female
triax tee adapter 6171
3-slot male triax to 2-lug female
triax adapter CA-93-1
BNC to right angle SMB cable
2-3
Page 19
Operation
2.5.3 T riax banana plug adapter
For instruments that use banana jacks, you need a triax cable terminated with a 3-slot male triax and a single banana plug. Use the Model 237-BAN-3 or prepare a special cable as out­lined below (Special triax to banana plug cable prepara-
tion) using the parts listed in Table 2-2.
Table 2-2
Parts for special triaxial cable
Keithley part or model number Description
7078-TRX-3 triax cable*
BG-10-2
*One connector must be cut off
Figure 2-4 shows the conductors and insulation layers of a triaxial cable. These layers must be carefully stripped back, cleaned thoroughly and insulated with high insulation resis­tance material such as Teflon  to maintain the integrity of the cable and measurement system. With the Model 237­BAN-3, the center conductor of the triax is connected to the banana plug. The inner and outer shields have no connection. With the special cable shown in Figure 2-4, the inner shield is shorted to the center conductor. Which cable to use depends on your application. The length of unshielded con­ductor that is connected to the banana plug should be mini­mized to maintain signal integrity. The topic of signal integrity is also discussed in paragraph 2.7 Measurement considerations.
Triax cable terminated with 3-slot male triax connectors Red banana plug
4. Strip the insulator back 1/2 inch, then twist the strands of the conductor together as shown in Figure 2-4(C).
5. Unscrew the cover from a banana plug, then slide the cover over the conductor.
6. Insert the stripped center conductor through the hole in the body of the banana plug, then wrap the wire around the plug body as shown in Figure 2-4(D).
7. Screw on the plastic cover as shown in Figure 2-4(E). Make certain the wire is secure by gently pulling on the plug.
Cut
1"
(A) Cut off insulation with knife.
Cut off outer shield.
Insulation Over Inner Shields
3/4"
(B) Strip insulation off inner shield.
(C) Twist inner shield then strip inner conductor. Twister inner shield and center conductor together,
slip on plastic cover.
Cut
Note that you can use either an unterminated triax cable, or cut a dual-connector cable (7078-TRX-10) in half to con­struct two special cables.
Special triax to banana plug cable preparation
The following steps outline a procedure for installing a banana plug on the end of a triaxial cable (with inner shield shorted to center conductor).
1. Using a knife, cut and strip back the jacket about 1-1/2 inches.
2. Remove the outer insulation, then cut away the outer shield as far as the insulation is stripped as shown in Fig­ure 2-4(A).
3. Carefully strip away the insulation over the inner shield one inch, then cut the inner shield and guard wire off even with the stripped insulation as shown in Figure 2-4(B).
2-4
(D) Insert wires into hole and wrap around body.
(E) Screw on plastic cover.
Figure 2-4
Triax cable preparation
Page 20
Operation
2.5.4 General instrument connections
The following paragraphs discuss connecting the Model 7174A to various general classes of instrumentation such as DMMs, electrometers, sources, and source/measure units. Because these configurations are generic in nature, some modification of the connecting schemes may be necessary for your particular instrumentation. Also, special cables or adapters may be necessary. In all cases, 3-lug triax cables must be used to make the connections.
WARNING
Do not use coaxial cables and adapters because hazardous voltage from guard sources may be present on the cable shields.
Figure 2-5 shows the general instrument connections for the discussions below. Note that DUT guarding or shielding are not included here; see Figures 2-22 and 2-23 for shielding and guarding information. As shown, all figure assume instruments are connected to rows, and the DUT is con­nected to columns.
DMM connections
General DMM connections are shown in Figure 2-5 (A), (B), and (C). Floating connections are shown in (A) with LO and HI routed to two separate jacks on the Model 7174A. The common LO connections in (Figure 2-5B) should be used only for non-critical applications because the performance of the GUARD pathway is not specified.
WARNING
Hazardous voltage from other guard sources may be present on LO or the DUT if other crosspoints are closed.
Electrometer connections
T ypical electrometer connections are shown in Figure 2-5(D) through (G). The unguarded volts connections in (D) show the HI signal path routed through one jack, and the LO path goes through the other connector. Both GUARD pathways are connected to electrometer LO. For guarded voltage (E), Model 7174A GUARD is connected to electrometer GUARD.
The connections for electrometer fast amps and resistance measurements are shown in Figure 2-5(F) and (G). These configurations are essentially the same as those discussed above. For the case of fast amps, both GUARD paths are connected to electrometer LO, while in the case of guarded resistance, one GUARD path is connected to electrometer GUARD, and the other GUARD path is connected to elec­trometer LO.
Source connections
Voltage and current source connections are shown in Figure 2-5(H) through (J). The HI and LO paths of the voltage source (H) are routed through two jacks, with both card GUARD pathways connected to voltage source LO. For the unguarded current source connections (I), card GUARD is again connected to source LO, with source HI and LO routed through two pathways. In the case of the guarded current source in (J), card GUARD of the HI signal path is connected to source GUARD, and the other GUARD path is connected to source LO.
Source/measure unit connections
Figure 2-5(K) shows typical connections for a source/mea­sure unit (SMU). In this instance, a remote-sensing type of a SMU is shown, requiring a total of four signal pathways to the DUT. For critical measurements, both source and sense HI pathways would be guarded as shown, with two of the four card GUARD pathways connected to SMU GUARD terminals. As with other instrument connections, the LO card GUARD pathways are connected to SMU LO terminals.
4-wire DMM connections are shown in Figure 2-5(C). In this case, a total of four jacks are required; HI, LO, SENSE HI, and SENSE LO.
2-5
Page 21
Operation
Rows Columns
S
ignal
HI
LO
DMM
A.) DMM Floating
Rows Columns
HI
LO
DMM
Warning:Hazardous voltage from guard
sources may be present on LO.
Guard
Signal
Guard
Signa
Guard
7174A
l
7174A
DUT
DUT
Note: Use this configuration only for
non-critical measurements.
B.) DMM Common LO
Figure 2-5
General instrument connections
2-6
Page 22
Operation
Sense HI Sense LO
DMM
HI LO
Rows Columns
Signal
Guard
Signal
Guard
DUT
Signal Guard
Signal
Guard
C.) DMM 4-Wire
LO
Electrometer
D.) Electrometer, Unguarded Volts
Figure 2-5
General instrument connections (cont.)
HI
7174A
Rows Columns
Signal
Guard
DUT
Signal
Guard
7174A
2-7
Page 23
Operation
Rows Columns
Signal
HI
Guard
LO
Electrometer
E.) Electrometer, Guarded Volts
HI
LO
Electrometer
Guard
DUT
Signal
Guard
7174A
Rows Columns
Signal Guard
DUT
Signal Guard
F.) Electrometer,Fast Current
HI LO
Guard
Electrometer
G.) Electrometer, Resistance (Guarded)
Figure 2-5
General instrument connections (cont.)
7174A
Signal
Guard
DUT
Signal
Guard
7174A
2-8
Page 24
ignal
S
Operation
HI LO
Voltage Source
H.) Voltage Source
HI LO
Electrometer
Guard
DUT
Signal Guard
7174A
Signal
Guard
DUT
Signal
Guard
I.) Current Source, Unguarded
Guard
Current Source
J.) Current Source, Guarded
Figure 2-5
General instrument connections (cont.)
HI LO
7174A
Rows Columns
Signal Guard
DUT
Signal Guard
7174A
2-9
Page 25
Operation
Force IorV
Sense
VorI
Source/Measure
Guard
Guard
Unit
HI
LO
HI
LO
Rows Columns
Signal
ua
G
Signal Guard
S
ignal
uard
G
Signal Guard
d
r
DUT
K.) Source/Measure Unit
Note: DUT shielding/guarding not shown. See Figures 2-20 and 2-21.
Figure 2-5
General instrument connections (cont.)
7174A
2-10
Page 26
Operation
2.5.5 Keithley instrument connections
The following paragraphs outline connecting typical Kei­thley instruments to the Model 7174A 8 × 12 Low Current Matrix Card. Other similar instruments can be connected using the same cabling as long as their input/output configu­rations are the same. Instrument connections covered include:
• Model 617 Electrometer/Source
• Model 196 DMM
• Model 230 Programmable Voltage Source
• Model 220 Programmable Current Source
• Model 590 CV Analyzer
• Model 236/237/238 Source Measure Unit
Model 617 electrometer connections
Connections for the Model 617 Electrometer are shown in Figure 2-6. The electrometer INPUT and COM can be con-
6172 2-slot to 3-lug Triax Adapter
Guard off
COM
INPUT
Model 617
nected to any row. Figure 2-6 shows connections to rows A and B.
1. Connect one end of a Model 7078-TRX- (3, 10, or 20) 3-lug triaxial cable to row A of the Model 7174A.
2. Connect the other end of the triax cable to the Model 617 INPUT connector using a Model 6172 adapter.
3. Connect the triax end of a triax/banana cable to row B of the Model 7174A.
4. Connect the banana plug end of the triax/banana cable to the COM terminal of the Model 617. The shorting link between COM and chassis ground should be re­moved for this application.
5. Place the GUARD switch in the OFF position.
6. T o connect the v oltage source to the Model 7174A, con­nect the V-SOURCE HI and LO connector of the Model 617 to the desired row connectors on the matrix card. Figure 2-6 shows connections to rows C and D.
KEITHLEY
7174A
8x12LOW
CURRENT MATRIX
SIGNAL
200VP K
COLUMNS
200VPK
GUARD
200VP K
7078 - TRX Triax
237-BAN-3 o r Special Triax
237-BAN-3
ROWS
1
A
2
B
3
C
4
D
5
L
6
7
Voltage Source Connections
Figure 2-6
Model 617 electrometer connections
237-BAN-3 o r Special Triax
Note: See paragraph 2.5.3 and Figure
2-4 f or special triax to banana cable.
8
E
9
F
10
G
H
!
WARNING:
TIGHTENMOUNTINGSCREWS
TOENSUREPROPER
CHAS SIS GRO UND
11
12
7174A Matrix Card
2-11
Page 27
Operation
Model 196 DMM connections
Connect the Model 196 or other similar DMM to the matrix card using the general configuration shown in Figure 2-7. The V OLTS OHMS HI and LO terminals should be connect­ed to the desired rows using triax/banana cables. For 4-wire ohms measurements, the OHMS SENSE HI and LO termi-
237-BAN-3
237-BAN-3 or Special Triax
HI
nals should be connected to two additional rows using the same type of cables.
NOTE
For low-level voltage measurements, con­nect the inner shield of the HI cable to VOLT OHMS LO to minimize noise.
KEITHLEY
7174A
8x12LOW
CURRENT MATRIX
SIGNAL
COLUMNS
200VP K
200VPK
GUARD
200VPK
ROWS
A
B
C
D
1
2
Note
3
4
5
L
6
: See paragraph 2.5.3
and Figur
special triax to banana
2-4 for
e
cable.
196 DMM
Connect inner shield to LO for low-levelmeasurements. (Modify the cable of Figure 2-4.)
Figure 2-7
Model 196 DMM connections
LO
7
8
E
F
G
H
TIGHTE N MOUNTING SCR EWS
WARNING:
TOENSUREPROPER
CHASSIS GROUND
9
10
11
12
!
7174A Matrix Card
2-12
Page 28
Operation
Model 230 voltage source connections
Connect the Model 230 OUTPUT and COMMON terminal to the desired rows using triax/banana plug cables, as shown
Common
230 C urrent Source
Note: See paragraph 2.5.3 and Figure
2-4 for special triax to banana cable.
Output
in Figure 2-8. For remote sensing applications, the SENSE OUTPUT and SENSE COMMON connectors can be routed through two additional rows using similar cables.
KEITHLEY
7174A
8x12LOW
CURRENT MATRIX
SIGNAL
COLUMNS
200VP K
200VP K
GUARD
200VPK
237-BAN-3 or Special Triax
237-BAN-3
ROWS
A
B
C
D
E
F
G
H
WARNING:
TIGHTENMOUNTING SCRE WS
TOE NSURE P R OPER
CHASSIS GROUND
1
2
3
4
5
6
7
8
9
10
11
12
!
Figure 2-8
Model 230 voltage source connections
7174A Matrix Card
2-13
Page 29
Operation
Model 590 CV analyzer connections
The Model 590 CV analyzer can be connected to any row or any column as shown in Figure 2-9. The BNC cables that are
. .
590 CV Analyzer
supplied with the Model 590 can be used; however, Model 7078-TRX-BNC adapters must be used at the Model 7174A end.
7078-TRX-BNC
Triax-to-BNC
Adapters
7051 BNC Cables
KEITHLEY
8x12LOW
CURRENT MATRIX
SIGNAL
200VPK
200VPK
GUARD
200VP K
ROWS
A
B
C
D
E
F
G
H
WARNING :
TIGHTENMOUNTINGSCREWS
TOENSUREPROPER
CHASSISGROUND
7174A
COLUMNS
!
1
2
3
4
5
6
7
8
9
10
11
12
Figure 2-9
Model 590 CV analyzer connections
7174A Matrix Card
2-14
Page 30
Operation
Model 220 current source connections
The Model 220 current source can be connected to the matrix card using the Model 6167 Guarded Adapter, as shown in Figure 2-10. This conguration guards the output signal to minimize the effects of distributed capacitance and leakage current.
NOTE
The Model 6167 adapter must be modified by internally disconnecting the inner shield connection of the input jack from the GUARDED/UNGUARDED selection switch. Otherwise, instrument LO will be
Model 6167
Guarded Adapter
connected to chassis ground through the adapter.
1. Connect the Model 6167 adapter to the Model 220 OUTPUT jack.
2. Connect a Model 7078-TRX-(3, 10 or 20) triax cable between the guarded adapter and the desired row of the Model 7174A.
3. Connect the Model 220 GUARD output to GUARD INPUT terminal of the adapter.
4. Connect the triax end of a triax/banana cable to the desired row on the Model 7174A.
5. Connect the banana plug end of the triax/banana cable to the OUTPUT COMMON jack of the Model 220.
KEITHLEY
7174A
8x12LOW
CURRENT MATRIX
SIGNAL
COLUMNS
200VP K
200VP K
GUARD
200VPK
7078-TRX T riax 237-BAN-3 or Special Triax
ROWS
1
A
2
B
3
C
4
D
5
6
Connect GUARD OUT
to GUARD
Note: See paragraph 2.5.3 and Figure
2-4 for special triax to banana cable.
Figure 2-10
Model 220 current source connections
220 C urrent Source
7
8
E
F
G
H
TIGHTENMOUNTING SCRE WS
WARNING:
TOE NSURE P R OPER
CHASSIS GROUND
9
10
11
12
!
7174A Matrix Card
2-15
Page 31
Operation
Model 236/237/238 source measure unit connections
Source measure units are connected to the matrix card using Model 7078-TRX cables. A Model 237-B AN-3 triax/banana cable can also be used to connect the output low binding post on the source measure unit to the matrix. Figure 2-11 shows connections for remote and local sensing applications.
LINERATING
TRIGGER
MAXOUTPUT
KEITHLEY
GUARD
!
110V
MADEIN U.S.A.
LO
HI
HI
OUTPUT
OUTPUT
SENSE
GUARD
A. Remote Sensing
OUT IN
OUTPUT
SENSE
LO
LO
LINEVOLTAGE
200V MAX
SELECTED
LINEF USE SLOWBLOW
INTERLOCK
50-60Hz
1
ACONLY
100VAMAX
2
IEEE-488INTERFACE
ADDRESSENTERED VIA FRONTPANEL PROGRAM
CAL
ENABLE
7078-TRX Triax
7078-TRX Triax 7078-TRX Triax
CAUTION
Models 237 and 238 source measure units can only be used within the speci­fied maximum signal levels of the Model 7174A (200V, 2A).
KEITHLEY
7174A
8x12LOW
CURRENTMATRIX
SIGNAL
200VPK
200VPK
COLUMNS
GUARD
200VPK
ROWS
1
A
2
B
3
C
4
D
5
6
7
8
E
9
F
10
G
11
H
12
WARNING:
!
TIGHTENMOUNTING SCREWS
TOENSUREPROPER
CHASSISGROUND
7174A Matrix Card
TRIGGER
OUT IN
MAXOUTPUT
KEITHLEY
GUARD
!
110V
MADEIN U.S.A.
OUTPUT
OUTPUT
SENSE
LO
LO
LO
HI
HI
200V MAX
OUTPUT
SENSE
GUARD
LINEVOLTAGE
SELECTED
LINEF USE SLOWBLOW
INTERLOCK
ENABLE
B. Local Sensing
Figure 2-11
Model 236/237/238 source measure unit connections
7078-TRX Triax 237-BAN-3 or Special Triax
LINERATING
50-60Hz
1
ACONLY
100VAMAX
2
IEEE-488INTERFACE
ADDRESSENTERED VIA FRONTPANELPROGRAM
CAL
Caution:
The models 237 and 238 Source Measure Units can only be used within the specified maximum signal levels of the Model 7174A (200V, 2A carry).
KEITHLEY
7174A
8x12LOW
CURRENTMATRIX
SIGNAL
COLUMNS
200VPK
200VPK
GUARD
200VPK
ROWS
1
A
2
B
3
C
4
D
5
6
7
8
E
9
F
10
G
11
H
12
!
WARNING:
TIGHTENMOUNTINGSCREWS
TOENSURE PROPER
CHASSISGROUND
7174A Matrix Card
2-16
Page 32
Operation
2.5.6 T ypical test fixture connections
Typically, one or more test xtures will be connected to desired columns of the Model 7174A. Typically, the test fix- tures will be equipped with card-edge connectors with wires soldered to them. In some cases, the test xture will be equipped with triax connectors; for those types, Keithley Model 7078-TRX-(3, 10, or 20) cables can be used, as shown in Figure 2-12.
WARNING
Do not use BNC cables and adapters in cases where hazardous voltages from guard sources could be present on the BNC cable shields.
KEI THLE Y
7174A
8x12LOW
CURRENTMATRIX
SIGNAL
COLUMNS
200VP K
200VP K
GUARD
200VP
K
ROWS
A
B
C
D
1
2
3
4
5
Internally, the test xture should be wired as shown in the equivalent circuit of Figure 2-13. SIGNAL should be con­nected to the probe or other device contact points, while GUARD should be carried through as close to the device as possible. If coaxial probes are to be used, connect GUARD to the probe shield if the probe shield is insulated from the xture shield.
Usually, the chassis ground terminal of the triax connector will automatically make contact with the xture shield by virtue of the mounting method. However, ground integrity should be checked to ensure continued protection against hazardous guard voltages.
3 - Lug Female
Test Fixture
Triax Connectors (or run cables through strain reliefs and connect internally)
6
7
8
E
F
G
H
TIGHT EN MOUNTING S CR EWS
WARNING:
TOENSUREPROPER
CHAS SIS GR OUND
9
10
11
12
!
7174A Matrix Card
Figure 2-12
Typical test fixture connections
®
Note: Teflon
-insulated connectors
recommended for specified performance.
Warning:Donot use BNC connectors
to avoid possible shock hazard.
2-17
Page 33
Operation
Triax Cable
Signal
Probe
From
7174A
Card
Guard
Chassis Ground
Figure 2-13
Equivalent circuit of test fixture connections
2.6 Matrix configuration
This section describes the matrix conguration of the Model 7174A. It also explains ways to expand the matrix by install­ing and connecting additional matrix cards in the Model 707A Switching Matrix.
2.6.1 Switching matrix
As shown in Figure 2-14, the Model 7174A is organized as an 8 × 12 (8 rows by 12 columns) matrix. The rows on the card are labeled A through H while the columns are num­bered 1 through 12. The actual column number to use when programming depends on the slot and unit number (Table 2-3). For example, card column number 2 on a card in slot 5 of unit 1 is accessed as matrix column 50.
Wafer
Test Fixture Chassis
Each intersecting point in the matrix is called a crosspoint that can be individually closed or opened by programming the Model 707A mainframe. By closing the appropriate cros­spoint, required pathways and pins may be connected. All crosspoints are congured for 2-pole switching, as shown in Figure 2-14. SIGNAL and GUARD are switched separately to any of the 12 columns on the card.
2.6.2 Path isolators
The path isolator relay switches shown in Figure 2-14 serve to isolate a given path from the rest of the matrix when no crosspoint (relay) is closed in that pathway. This topology minimizes leakage current and capacitance to pathways which are active. These isolators close automatically when any crosspoint on the path is closed and open automatically when all crosspoints on the path are opened.
2-18
Page 34
Columns
HGC HG C HG C HGC HGC HGC HGC HG C HGC HGC HGC
Operation
HGC
User
connections
and expansion
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
Path Isolator Relay Switch
Signal Guard
Row
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
Signal
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
Guard
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
Note: For schematic,
refer to Section 5.
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
H G C
H G C
H G C
H G C
H G C
H G C
H G C
H G C
Rows
Column
Figure 2-14
Matrix configuration
2-19
Page 35
Operation
Table 2-3
Column numbering by slot and unit
Unit Slot Columns (1 through 12)
1
1 2 3 4 5 6
2
1 2 3 4 5 6
1
3
2 3 4 5 6
1
4
2 3 4 5 6
1 13 25 37 49 61
73 85 97
109 121 133
145 157 169 181 193 205
217 229 241 253 265 277
2 14 26 38 50 62
74 86 98
110 122 134
146 158 170 182 194 206
218 230 242 254 266 278
3 15 27 39 51 63
75 87 99
111 123 135
147 159 171 183 195 207
219 231 243 255 267 279
4 16 28 40 52 64
76 88
100 112 124 136
148 160 172 184 196 208
220 232 244 256 268 280
5 17 29 41 53 65
77 89
101 113 125 137
149 161 173 185 197 209
221 233 245 257 269 281
6 18 30 42 54 66
78 90
102 114 126 138
150 162 174 186 198 210
222 234
2446
258 270 282
7 19 31 43 55 67
79 91
103 115 127 139
151 163 175 187 199 211
223 235 247 259 271 283
8 20 32 44 56 68
80 92
104 116 128 140
152 164 176 188 200 212
224 236 248 260 272 284
9 21 33 45 57 69
81 93
105 117 129 141
153 165 177 189 201 213
225 237 249 261 273 285
10 22 34 46 58 70
82
94 106 118 130 142
154 166 178 190 202 214
226 238 250 262 274 286
11 23 35 47 59 71
83
95 107 119 131 143
155 167 179 191 203 215
227 239 251 263 275 287
12 24 36 48 60 72
84
96 108 120 132 144
156 168 180 192 204 216
228 240 252 264 276 288
2-20
289
1
5
301
2
313
3
325
4
337
5
349
6
290 302 314 326 338 350
291 303 315 327 339 351
292 304 316 328 340 352
293 305 317 329 341 353
294 306 318 330 342 354
295 307 319 331 343 355
296 308 320 332 344 356
297 309 321 333 345 357
298 310 322 334 346 358
299 311 323 335 347 359
300 312 324 336 348 360
Page 36
Operation
2.6.3 Internal matrix expansion
Two to six Model 7174A cards can be connected together within the mainframe to yield an 8 × N matrix, where N depends on the number of cards. Figure 2-15 shows an inter­nally expanded matrix with three cards, resulting in an 8 × 36 (eight row by 36 column) matrix. As summarized in Table 2-3, the actual column number used when programming the unit is determined by the slot.
WARNING
The shells of the row jumpers are at guard potential. To avoid a possible shock hazard, always disconnect all cables from the row and column jacks before removing or installing jumpers.
Internal
9
10
Paths Jumpers
12
11
131415
Paths
1 2 3 4 5 6 7 8
123
Slot 1 Slot 2 Slot 3
8
6
4
5
7
Because of critical signal paths, rows A-H are not jumpered through the backplane. Instead, install the supplied coaxial jumpers between appropriate connectors on Model 7174A cards (for more critical signal paths, rows can be isolated from other cards by not installing these cables). Each card has two coaxial connectors for each row, allowing daisy chaining of card rows. These connectors can be reached by lifting the access door on the top of the mainframe; do not remove cards to install the jumpers. This expansion is sho wn schematically in Figure 2-15 for cards located in slots 1, 2 and 3.
Figure 2-16 shows the location of the connectors used for this expansion. These connectors can be reached by lifting the access door on the top of the Model 707A Switching Matrix. The jumpers must be installed and removed before installing or removing a Model 7174A from the 707A Switching Matrix. Figure 2-17 shows how two cards can be daisy chained together using the coaxial jumpers.
Internal
Paths Jumpers
18 19
20
212223
24 25 26 27 28
29 303132 333435
16 17
36
Figure 2-15
Connecting three cards for an 8 × 36 matrix
Figure 2-16
Jumper connector locations
ABCDEFGH
Measure
(For each path)
Warning:
Guard potential is on coaxial jumper shields.
2-21
Page 37
Operation
Figure 2-17
Two cards in daisy chain configuration

2.7 Measurement considerations

Most measurements made with the Model 7174A concern low-level signals. Such measurements are subject to various types of noise that can seriously affect low-level measure­ment accuracy. The following paragraphs discuss possible noise sources that might affect these measurements.
keep the switching and measuring circuits a good distance away from these potential noise sources.
2.7.2 Electromagnetic Interference (EMI)
The electromagnetic interference characteristics of the Model 7174A comply with the electromagnetic compatibil­ity (EMC) requirements of the European Union (EU) direc­tives as denoted by the CE mark. However, it is still possible for sensitive signals to be affected by external sources. In these instances, special precautions may be required in the test setup.
Sources of EMI include:
Radio and TV broadcast transmitters.
Communications transmitters, including cellular
phones and handheld radios.
Devices incorporating microprocessors and high-speed digital circuits.
Impulse sources as in the case of arcing in high-voltage environments.
The Model 7174A, signal source, measuring instrument, and signal leads should be kept as far away as possible from any EMI sources. Additional shielding of the card, signal leads, sources, and measuring instruments will often reduce EMI to an acceptable level. In extreme cases, a specially constructed screen room may be required to sufciently attenuate the troublesome signal.
2.7.1 Magnetic elds
When a conductor cuts through magnetic lines of force, a very small current is generated. This phenomenon will fre­quently cause unwanted signals to occur in the test leads of a switching matrix system. If the conductor has sufficient length, even weak magnetic elds like those of the earth can create sufcient signals to affect low-level measurements.
Two ways to reduce these effects are: (1) reduce the lengths of the test leads, and (2) minimize the exposed circuit area. In extreme cases, magnetic shielding may be required. Spe­cial metals with high permeability at low ux densities (such as mu metal) are effective at reducing these effects.
Even when the conductor is stationary, magnetically­induced signals may still be a problem. Fields can be produced by various signals such as the AC power line voltage. Large inductors such as power transformers can also generate substantial magnetic elds, so care must be taken to
2-22
Many instruments incorporate internal ltering that may help to reduce EMI effects in some situations. In other cases, additional external ltering may be required. Keep in mind, however, that ltering may have detrimental effects, such as increased settling time, on the measurement.
2.7.3 Ground loops
When two or more instruments are connected together, care must be taken to avoid unwanted signals caused by ground loops. Ground Loops usually occur when sensitive instru­mentation is connected to other instrumentation with more than one signal return path such as power line ground. As shown in Figure 2-18, the resulting ground loop causes unwanted signals to ow through the instrument LO signal leads and then back through power line ground. This circu­lating current develops a small but undesirable voltage between the LO terminals of the two instruments. This volt­age will be added to the source voltage, affecting the accu­racy of the measurement.
Page 38
Operation
SIGNALLEADS
INSTRUMENT 1 INSTRUMENT 2 INSTRUMENT 3
GROUND
LOOP
CURRENT
POWER LINE GROUND
Figure 2-18
Power line ground loops
Figure 2-19 shows how to connect several instruments to­gether to eliminate this type of ground loop problem. Here, there is only one connection to power line ground.
INSTRUMENT 1 INSTRUMENT 2 INSTRUMENT 3
these problems, never touch the connector insulating mate­rial. In addition, the matrix card should be used only in clean, dry environments to avoid contamination.
If the connector insulators should become contaminated, either by inadvertent touching or from air borne deposits, they can be cleaned with a cotton swab dipped in clean meth­anol or an HCFC. After thorough cleaning, they should be allowed to dry for several hours in a low-humidity environ­ment before use, or they can be dried more quickly using dry nitrogen.
2.7.5 Noise currents caused by cable exing
Noise currents can be generated by bending or exing coax­ial, triaxial, or quadraxial cables. Such currents, known as triboelectric currents, are generated by charges created between a conductor and insulator caused by friction.
Low-noise cable can be used to minimize these effects. Such cables have a special graphite coating under the shield to pro­vide lubrication and to provide a conduction path to equalize charges.
POWER LINE GROUND
Figure 2-19
Eliminating ground loops
Ground loops are not normally a problem with instruments having isolated LO terminals. However, all instruments in the test setup may not be designed in this manner. When in doubt, consult the manual for each instrument in the test set­up.
2.7.4 Keeping connectors clean
As is the case with any high-resistance device, the integrity of connectors can be damaged if they are not handled prop­erly. If the connector insulation becomes contaminated, the insulation resistance will be substantially reduced, affecting high-impedance measurement paths.
Oils and salts from the skin can contaminate connector insu­lators, reducing their resistance. Also, contaminants present in the air can be deposited on the insulator surface. To avoid
Even low-noise cable generates some noise current when exed or subjected to vibration. To minimize these effects, keep the cables as short as possible, and do not subject them to temperature variations that could cause expansion or con­traction. Tie down offending cables securely to avoid move­ment, and isolate or remove vibration sources such as motors or pumps.
2.7.6 Shielding
Proper shielding of all unguarded signal paths and devices under test is important to minimize noise pickup in virtually any switching matrix system. Otherwise, interference from such noise sources as line frequency and RF elds can seri­ously corrupt a measurement.
In order for shielding to be effective, the shield must com­pletely surround the source, measure and guard signals. It must also have a low impedance path to chassis ground. This is shown pictorially in Figure 2-20. The shield also functions as a safety shield since it is connected to chassis ground.
WARNING
Hazardous voltage may be present if LO on any instrument is floated above ground potential.
2-23
Page 39
Operation
To
7174A
Source
Measure
Guard
Chassis
Signal Source and Measure (Probe or Contact Point)
Figure 2-20
Shielded and guarded
2.7.7 Guarding
Guarding is important in high-impedance circuits where leakage resistance and capacitance could have degrading effects on the measurement. Guarding consists of using a shield surrounding a conductor that is carrying the high impedance signal. This shield is driven by a low-impedance amplier to maintain the shield at signal potential.
Guarding minimizes leakage resistance effects by driving the cable shield with a unity gain amplier, as shown in Figure 2-21. Since the amplier has a high input impedance, it min­imizes loading on the high-impedance signal lead. Also, the low output impedance ensures that the shield remains at sig­nal potential, so that virtually no leakage current flows through the leakage resistance, RL. Leakage between inner and outer shields may be considerable, but that leakage is of little consequence because that current is supplied by the buffer amplier rather than the signal itself.
DUT
Source or Measure
Signal
R
R
s
E
s
DUT
Figure 2-21
Guarded circuit
Inner Shield (Guard or LO)
Outer S hield
(Chassis Ground)
Inner Shield
X1
Buffer
L
E
M
Guard
Measuring Instrument
In a similar manner, guarding also reduces the effecti ve cable capacitance, resulting in much faster measurements on high­impedance circuits. Because any distributed capacitance is charged through the low impedance of the buffer amplifier rather than by the source, settling times are shortened consid­erably by guarding.
In order to use guarding effectively with the Model 7174A, the guard path of the matrix card should be connected to the guard output of the sourcing or measuring instrument. Figure 2-20 shows typical connections. Guard should be properly carried through the inner shield to the device under test to be completely effective. The shielded, guarded test fixture arrangement shown in Figure 2-20 is also recommended for safety purposes (guard voltage may be hazardous with some instruments).
2-24
2.7.8 Matrix expansion effects on card specications
Specications such as those given for path isolation and off­set current are with a single Model 7174A card installed in the mainframe. Expanding the matrix by internally connect­ing two or more Model 7174A cards together will degrade system performance specications (other types of cards do not affect the specications because they use different path­ways in the mainframe backplane). The extent depends on how many cards are used, as well as the amount of cabling used to connect them together.
With internal pathway expansion, isolation among paths is increased, and offset current is decreased, although the isola­tor relays on the card do help to minimize these effects.
Page 40
2.8 Coaxial jumper access
To gain access to jumpers, lift jumper across cover (Figure 2-22).
Operation
Figure 2-22
Coaxial jumper access
2-25
Page 41
3

Applications

3.1 Introduction

This section covers typical applications for the Model 7174A 8 × 12 Low Current Matrix Card and is organized as follows:
3.2 CV measurements — Outlines the test configuration
and procedure for making quasistatic and high­frequency CV measurements.
3.3 Semiconductor test matrix — Details a semiconduc-
tor test matrix that can be used to perform a variety of different tests on semiconductors such as FETs.
3.4 Resistivity measurements— Covers methods to mea-
sure the resistivity of semiconductor samples using the van der Pauw method.
3.5 Semiconductor IV characterization — Covers the
basic scheme and connections used to generate an IV curve of a bipolar or MOS transistor.

3.2 CV measurements

The Model 7174A can be used in conjunction with the Keithley Model 590 CV Analyzer, and the Keithley Model 595 Quasistatic CV Meter to perform quasistatic and high­frequency CV (capacitance vs. voltage) test on semiconductors. The resulting CV curves can be used to calculate important semiconductor parameters such as doping profile, band bending, and mobile ion concentration.
3.2.1 Stand alone system configuration
The stand alone system shown in Figure 3-1 can be used to make CV measurements without the aid of a computer. Sys­tem components perform the following functions.
Model 590 CV Analyzer: Measures CV data at 100kHz and
1MHz and sends the resulting data to the plotter for graphing.
Model 595 Quasistatic CV Meter: Measures quasistatic CV
data and sends the data to the plotter for graphing in real time.
Model 707A Switching Matrix: Controls the semiconductor
matrix card to close and open the desired crosspoints at the proper time.
Model 7174A 8 × 12 Low Current Matrix Card: Switches
the signa pathways to the six devices under test.
HP-GL Plotter: Plots CV and other curves directly from the
Models 590 and 595.
3.2.2 Computerized system configuration
Figure 3-2 shows a computerized version of the CV matrix test system. The addition of a computer allows greater sys­tem versatility and easier instrument control. Also, analysis functions such as doping profile and ion concentration can be added to the software to expand CV analysis capabilities.
3-1
Page 42
Applications
Devices Under
123456
A
B
C D
E
F G H
7174A Matrix Card
707A Switching Matrix
Figure 3-1
Stand alone CV system configuration
Test
1112345678910 12
Triax
Cables
4801 Low-Noise Cables
S G
S G
7078-TRX-BNC Triax BNC Adapters
S G
S G
7051 BNC
Cables
Model 595
Quasistatic CV Meter
Meter Input
V-Source Output
IEEE-488
Bus
Output
Input
Model 590
CV Analyzer
Note:
Connect plotter to only one instrument at a time.
HP-GL Plotter
Devices Under
Test
123456
A
B
C
D
E
F G
H
7174A Matrix Card
707A Switching Matrix
Triax Cables
4801 Low-Noise Cables
S G
S G
7078-TRX-BNC
Triax BNC
Adapters
S G
S G
Model 595
Quasistatic CV Meter
Meter Input
V-Source Output
IEEE-488
Bus
Output
1112345678910 12
Input
7051 BNC
Cables
Model 590
CV Analyzer
IEEE-488 Bus
Note: Remove jumpers to other 7174A cards (if installed)
to optimize Model 595 measurement accuracy.
Computer
HP-GL Plotter
Figure 3-2
Computerized CV system configuration
3-2
Page 43
Applications
3.2.3 Optimizing CV measurement accuracy
For accurate CV measurements, each Model 590 CV mea­surement pathway must be corrected using the procedure outlined in the Model 590 Instruction Manual. The pathways to each DUT must be cable corrected separately.
Also, for best quasistatic CV results, the corrected capaci­tance feature of the Model 595 should be used. Corrected capacitance compensates for any leakage currents present in the cables, switching matrix, or test fixture. However, care must be taken when using corrected capacitance to ensure that the device remains in equilibrium throughout the test sweep to avoid distorting the CV curves.
In order to minimize the effects of the switching network on quasistatic CV measurements, cables to the Model 595 and DUT should be kept as short as possible.
3.2.4 Basic CV test procedure
The fundamental CV test procedure is outlined below. Keep in mind that this procedure does not address many consider­ations and aspects of CV testing, which is fairly complex. The procedure given is for the stand alone system in Figure 3-1. Detailed instrument operating information may be found in the pertinent instruction manuals.
7. Set up the Model 590 for the expected CV sweep.
8. Close the crosspoints necessary to connect the Model 590 to the device under test. For example, to test device #1, close G1 and H2.
9. Run a high-frequency test sweep on the device to store the CV data in the Model 590 buffer.
10. Disconnect the plotter from the Model 595 and connect it to the Model 590.
11. Generate a plot from the data in the Model 590 buffer.
12. Repeat steps 2 through 11 for the remaining devices, as required.
Table 3-1
CV test crosspoint summary
Closed crosspoints
Wafer #
1 2 3 4 5 6
Quasistatic (595) High frequency (590)
A1, B2 A3, B4 A5, B6 A7, B8
A9, B10
A11, B12
G1, H2 G3, H4 G5, H6 G7, H8
G9, H10
G11, H12
1. Connect the HP-GL plotter to the IEEE-488 bus connec­tor of the Model 595 only.
2. Set up the Model 595 for the expected CV sweep.
3. Close the crosspoints necessary to connect the Model 595 to the device under test, as summarized in Table 3-1. For example, to test device #1, close A1 and B2.
4. Place the probes down on the wafer test dots.
5. Run a quasistatic sweep on the selected device and gen­erate a CV curve.
6. Open the crosspoints that are presently closed.
3.2.5 T ypical CV curves
Figure 3-3 and Figure 3-4 shows typical CV curves as generated by the Model 595 and 590 respectively. The quasistatic curve shows a fair amount of symmetry, while the high-frequency curve is highly asymmetrical. The asymmetrical nature of the high-frequency curve results from the inability of minority carries to follow the high­frequency test signal.
3-3
Page 44
Applications
+0.6E-10
+0.4E-10
-005.00
+005.00
KEITHLEY 595
Figure 3-3
Typical quasistatic CV curve generated by Model 595
1.50
1.42
1.35
9
0
-
^
0
1
1.27
X
)
d
a
r
a
1.20
F
( e
c
n
a
t
1.12
i
c
a
p
a C
1.05
0.97
0.90
0.82
0.75
- 4.99 - 3.99 - 2.99 - 1.99 - 0.99 0.00 1.00 1.99 2.99 3.9 9 4.99
Keithley 590: 00: 00: 10: 500 100KHz X1 Filter ------------- Para llel
Bias (Volt) X 10^+00
Figure 3-4
Typical high-frequency CV curve generated by Model 590
3-4
Page 45
1112345678910 12
A
C
B
D
E
F G H
G
SGS G
SGS G
SGS GSG SGSG
G
G
G
G
G
G
S S S
S
S S
HI
G
LO
LO
LO
617
I Measure
HI
HI
V Source
590
CV
HI HI HI LOLOLO
220
Current
Source
196
DMM
V Measure
230
V Source
Device Under Test
Source
Gate
Drain
7174A Matrix Card
707A Switching Matrix
Figure 3-5
Semiconductor test matrix
Applications

3.3 Semiconductor test matrix

Two important advantages of a matrix switching system are the ability to connect a variety of instruments to the device or devices under test, as well as the ability to connect any instrument terminal to any device test node. The following paragraphs discuss a typical semiconductor matrix test sys­tem and how to use that system to perform a typical test: common-source characteristic testing of a typical JFET.
3.3.1 System conguration
Figure 3-5 shows the configuration for a typical multi­purpose semiconductor test matrix. Instruments in the system perform the following functions.
Model 617 Electrometer/Source: Measures current, and
also could be used to measure voltages up to ±200VDC. The DC voltage source can supply a maximum of ±100V at cur­rents up to 2mA.
Model 230 Voltage Source: Sources DC voltages up to
±101V at a maximum current of 100mA.
Model 590 CV Analyzer: Adds CV sweep measurement
capability to the system.
Model 220 Current Source: Used to source currents up to a
maximum of 101mA with a maximum compliance voltage of 105V.
Model 196 DMM: Measure DC voltages in the range of
100nV to 300V. The Model 196 could also be used to mea­sure resistance in certain applications.
Device Under T est: A three-terminal picture for testing such
devices as bipolar transistors and FETs. Additional connec­tions could easily be added to test more complex devices, as required.
3-5
Page 46
Applications
617
Electrometer/Source
A12
HI
I
D
B9
Meter
(Current)
LO
D
F8 F10
HI
LO
Model 230
Voltage Source
= Closed Crosspoints on 7174A Card (Figure 3-5).
E9
E11
G
V
GS
S
Figure 3-6
System configuration for measuring common-emitter characteristics
3.3.2 T esting common-source characteristic of FET s
The system shown in Figure 3-5 could be used to test a vari­ety of characteristics including I and V
DS[OFF]
. To demonstrate a practical use for the system, we will show how it can be used to generate common source characteristic curves of a particular JFET.
In order to generate these curves, the instrument must be connected to the JFET under test, as shown in Figure 3-6. The advantage of using the matrix is, of course, that it is a simple matter of closing specific crosspoints. The cross­points that must be closed are also indicated on the diagram.
GSS
, I
D[OFF]
, I
G[ON]
, I
DSS
,
V
DS
I
(µA)
D9
HI
Voltage
Source
LO
C11
100
90
V=0V
80 70 60
D
50 40 30 20 10
0
0 23456789
1
V(Volts)
DS
GS
V=-0.5V
GS
V=-1V
GS
10
To run the test, V increments of 0.25V. At each V voltage (V
DS
drain current, I
is set to specific values, for example in
GS
value, the drain-source
GS
) is stepped across the desired range, and the
, is measured at each value of V
D
. Once all
DS
data are compiled, it is a simple matter to generate the common-source IV curves, an example of which is shown in Figure 3-7. If the system is connected to a computer, the test and graphing could all be done automatically.
3-6
Figure 3-7
Typical common-source FET IV characteristics
Page 47
Applications

3.4 Resistivity measurements

The Model 7174A 8 × 12 Low Current Matrix Card can be used in conjunction with a Model 220 Current Source and a Model 196 DMM to perform resistivity measurements on semiconductors. Such measurements can yield such impor­tant information as doping concentration.
3.4.1 T est configuration
Figure 3-8 shows the basic test configuration to make resis­tivity measurements on van der Pauw samples. The Model 220 sources current through the samples, while the Model 196 measures the voltage developed across the samples. The matrix card, of course, switches the signal paths as neces­sary. In order to minimize sample loading, which will reduce accuracy, the Model 196 DMM should be used only on the 300mV or 3V ranges. Also, this configuration is not recom-
1
3
due to the
2
4
mended for resistance measurements above 1M Ω accuracy-degrading effects of DMM loading.
1
Sample 1
4
2
3
Sample 2 Sample 3
4
3.4.2 T est procedure
In order to make van der Pauw resistivity measurements, four terminals of a sample of arbitrary shape are measured. A current (from the Model 220) is applied to two terminals, while the voltage is measured (by the Model 196) across the two opposite terminals, as shown in Figure 3-9. A total of eight such measurements on each sample are required, with each possible terminal and current convention. The resulting voltages are designated V1 through V8.
In order to source current into and measure the voltage across the sample, specific crosspoints must be closed. Table 3-2 summarizes the crosspoints to close for each voltage measurement on all three samples from the test configuration shown in Figure 3-8.
1
2
3
SGSG
A
B
C
D
E
F
G
H
Figure 3-8
Resistivity test configuration
SG
SGSG SGSGSG
SG SGSGSG
7174A Matrix Card
707A Switching Matrix
1112345678910 12
S
G
S
G
S
G
S
G
HI G
LO
220 Current Source (Sources Current through Sample)
HI
LO
196 DMM (Measures Voltage Across Sample)
3-7
Page 48
Applications
I
12
43
V
(A)
V
3
(C) (D)
1
I
12
43
(B)
I
12
43
V
2
1
V
4
43
2
I
V
5
12
43
I
(E) (F)
1
43
I
(G) (H)
2
V
7
Figure 3-9
Resistivity measurement conventions
V
6
12
43
I
I
12
V
43
8
3-8
Page 49
Table 3-2
Crosspoint summary for resistivity measurements
Crosspoint closed
Voltage
Applications
Current between
Voltage
betweenSample #1 Sample #2 Sample #2
V V V V V V V V
A1
B4
E3
F2
A5
B9
1
A4
B1
E3
F2
2
A4
B3
3
A3
4
A3
5
A2
6
A2
7
A1
8
B4 B2 B3 B1 B2
E2 E2 E1 E1 E4 E4
F1 F1 F4 F4 F3 F3
A8 A8 A7 A7 A6 A6 A5
B5 B7 B8 B6 B7 B5 B6
E7 E7 E6 E6 E5 E5 E8 E8
3.4.3 Resistivity calculations
Once the eight voltage measurements are known, the resis­tivity can be calculated. Two values of resistivity, ρ are initially computed as follows:
1.1331 fAtSV2+V4–V1–V
------------------------------------------------------------------------ -=
ρ
A
1.1331 fBtSV6+V8–V5–V
------------------------------------------------------------------------ -=
ρ
B
()
I
()
I
3
7
and ρ
A
F6 F6 F5 F5 F8 F8 F7 F7
A9 A12 A12 A11 A11 A10 A10
A9
B12
B9 B11 B12 B10 B11
B9 B10
E11 E11 E10 E10
E9
E9 E12 E12
F10 F10
F9
F9 F12 F12 F11 F11
1-2 2-1 2-3 3-2 3-4 4-3 4-1 1-4
3-4 3-4 4-1 4-1 1-2 1-2 2-3 2-3

3.5 Semiconductor IV characterization

A source measure unit such as the Model 236, 237, or 238 is used to test and characterize many types of devices. One of
B
these is semiconductor devices. The following paragraphs explain the basic scheme and connections used to generate an IV curve of a bipolar or MOS transistor. Figure 3-10 shows FET devices connected in a test fixture.
3.5.1 T est configuration
Rows A and B are used to switch the Model 237 Source Mea­sure Unit; rows C and D are used for the Model 236.
Where: ρ
and ρ
A
t
is the sample thickness in cm
S
V
1
are the resistivities in Ω -cm
B
through V
are the voltages measured by the
8
Model 196 I is the current through the sample in amperes f
and f
are geometrical factors based on sample
B
=f
=1 for perfect symmetry).
A
B
are known, the average resistivity, ρ
B
Once ρ
A
symmetry (f
and ρ
A
can be determined as follows:
ρ
AVG
ρAρB+
------------------ -=
CAUTION
To prevent card damage, do not exceed the 200 volt maximum rating of the Model 7174A when switching the Model 237, which is capable of sourcing up to 1100 volts.
AVG
connected in a 4-wire sensing configuration. This connection
At the test fixture, the drain and source leads of the FETs are
,
scheme allows the Model 237 to use remote sensing to accurately apply Vds to the FETs. The Model 236 uses local sensing and is used to supply the bias to the gates of the
2
FETs. Since the gates are low current, remote sensing is not necessary.
If more DUT pins are needed, the system is easily expanded by adding more Model 7174A matrix cards. Each additional card will add 12 columns to the system.
3-9
Page 50
Applications
3.5.2 Cable connections
Source Measure Unit and test fixture connections to the matrix card are accomplished using Model 7078-TRX. These are three slot triax cables. On each Source Measure Unit, the banana jack (5-way binding post) is used to access OUTPUT LO. This connection is made using a Model 237-
Note:
Source Measure Units
Model 237 Model 236
DUT #1
Output HI/Guard
Sense HI/Guard
Sense LO/Output LO
Output LO
Output HI/Guard
Output LO
7078-TRX
Cables
123456789101112
Rows
A
X
B
X
C D E F G H
X
X
X
X
X
X
123456789101112
BAN-3 or the special cable constructed using the informa­tion in Figure 2-4. This allows OUTPUT LO to be applied to a signal pathway and independently switched. The guard pathways of the matrix cards are used exclusively to extend the driven guards of the Source Measure Units to the DUT to eliminate the effects of leakage current.
Do not connect guards to the DUT or short them together. If guarding the inner panel, only connect one guard line to it.
Test Fixture
DUT #2
DUT #3
Inner Panel
Warning:
Columns
Test fixture chassis must be connected to a safety earth ground.
7174A
Matrix Card
=
Model 237-BAN-3 or Triax to banana plug Modification (see Figure 2-4)
X
=
Crosspoints closed to apply VDS to DUT #1
=
Crosspoints closed to bias gate of DUT #1
Figure 3-10
Multi unit test system using Models 236 and 237 source measure units
3-10
7174A
Matrix Card
Signal Guard
Signal
Crosspoint Switching for all Rows
Guard
Page 51
4

Service Information

4.1 Introduction

WARNING
The service procedures in this section are intended for use only by qualified service personnel. Do not perform these procedures unless qualified to do so. Failure to recognize and observe normal safety precautions could result in per­sonal injury or death.
This section contains information necessary to service the Model 7174A Low Current Matrix Card and is arranged as follows:
4.2 Handling and cleaning precautions — Discusses
handling precautions and methods to clean the card should it become contaminated.
4.3 Principles of operation — Briefly discusses circuit
operation.
4.4 Troubleshooting — Presents some troubleshooting
tips for the Model 7174A.
4.5 Special handling of static-sensitive devices —
Reviews precautions necessary when handling static­sensitive devices.
4.6 Performance verification — Describes conditions
and provides references to determine if the card is op­erating properly.
4.7 Reed pack replacement — Provides a procedure for
replacing faulty reed packs.

4.2 Handling and cleaning precautions

Because of the high-impedance circuits on the Model 7174A, care should be taken when handling or servicing the card to prevent possible contamination. The following pre­cautions should be taken when servicing the card.
1. Handle the card only by the edges and handle (do not touch the edge connectors). Do not touch any board sur­faces or components not associated with the repair.
2. Do not store or operate the card in an environment where dust could settle on the circuit board. Use dry ni­trogen gas to clean dust off the board if necessary.
3. When making repairs on the circuit board, use aqua core solder and OA-based (organic acti v ated) flux. Use warm deionized water along with clean cotton swabs or a clean, soft brush to remove the flux. Take care not to spread the flux to other areas of the circuit board. Once the flux has been removed, blow dry the board with dry nitrogen gas.
NOTE
Removal of skin oils and other non­organic contaminants can be done with methanol or an HCFC.
4. After cleaning, the card should be placed in a 50°C low­humidity environment for several hours before use.
4-1
Page 52
Service Information

4.3 Principles of operation

The following paragraphs discuss the basic operating princi­ples for the Model 7174A. A schematic diagram of the matrix card may be found in drawings 9174-106 (mother board) and 9174-126 (air matrix relay board) located at the end of Section 5.
4.3.1 Block diagram
Figure 4-1 shows a simplified block diagram of the Model 7174A. Key elements include the buf fer (U410), ID data cir ­cuits (U406, U408, and U410), relay drivers (U101 through U113), relays (K101-K204), and the power-on safe guard (U409). The major elements are discussed below.
4.3.2 ID data circuits
At power up, the card identification data information from each card is read by the mainframe. This ID data includes such information as card ID, hardware settling time for the card, and a relay configuration table, which tells the main­frame which relays to close for a specific crosspoint. This configuration table is necessary because some cards (such as the Model 7174A) require the closing of more than one relay to close a specific crosspoint.
ID Data is contained within an on-card ROM, U406. In order to read this information, the sequence below is performed upon power up. Figure 4-2 shows the general timing of this sequence.
1. The CARDSEL line is brought low, enabling the ROM outputs. This line remains low throughout the ID data transmission sequence.
2. The CLRADDR line is pulsed clearing the address counter to zero. At this point, a ROM address of zero is selected. This pulse occurs only once.
3. The NEXTADDR line is set low. NEXTADDR going low increments the counter and enables parallel loading of the parallel-to-serial converter. NEXTADDR is kept low long enough for the counter to increment and the ROM outputs to stabilize. This sequence functions because the load input of the parallel-to-serial converter is level sensitive rather than edge sensitive. The first ROM address is location 1, not 0.
4. The CLK line clocks the parallel-to-serial converter to shift all eight data bits from the converter to the main­frame via the IDDATA line.
The process in steps 3 and 4 repeats until all the necessary ROM locations hav e been read. A total of 498 bytes of infor­mation are read by the mainframe during the card ID sequence.
Address Counter
U407
CLRADDR
To
Mainframe
Figure 4-1
Model 7174A block diagram
A0-A11
NEXTADDR
Buffer
U410
ROM
U406
CARDSEL
IDDATA
RELAYDATA
STROBE
Power-On
Safeguard
U409
D0-D7
CLK
NEXTADDR
Drivers
U101-U113
Output Enable
Converter
U408
Relay
Parallel to Serial
Relays
K101-K204
Columns
1-12
Rows A-H
4-2
Page 53
CARDSEL
CLRADDR
NEXTADDR
Service Information
CLK
IDDATA
HI-Z
Note: ID data sequence occurs on power-up only.
CLRADDR pulse occurs only once.
D7 D6 D5 D4 D3 D2 D1 D0
Figure 4-2
ID data timing
4.3.3 Relay control
The relays are controlled by serial data transmitted via the RELAY DATA line. A total of 16 bytes for each card are shifted in serial fashion into latches located in the 16 relay drivers, (U101 through U113). The serial data is fed in through the DATA lines under control of the CLK signal. As data overflows one register, it is fed out the Q’s line of that register to the next IC down the chain.
Once all 16 bytes have been shifted into each card in the mainframe, the STROBE line is set high to latch the relay information into the Q outputs of the relay drivers, and the appropriate relays are energized (assuming the driver outputs are enabled, as discussed below). Logic convention is such that the corresponding relay driver output must be low to energize the associated relay, while the output is high when the relay is de-energized. For example, if the Q1 output of U113 is low, relay K197 will be energized.
HI-Z
4.3.4 Power-on sequence
A power-on safeguard circuit, made up of U409 and associ­ated components, ensures that relays do not randomly ener­gize upon power-up. The tw o ANG gates, U409, make up an R-S flip-flop. Initially, the Q output of the flip-flop (pin 3 of U409) is set high upon power up. Since the OEN terminals of the relay drivers (U101 through U113) are held high, their outputs are disabled, and all relays remain de-energized regardless of the relay data information preset at that time.
The first STROBE pulse that comes along (in order to load relay data) clears the R-S flip-flop, setting the OEN lines of the relay drivers low to enable their outputs. This action allows the relays to be controlled by the transmitted relay data information.
A hold-off period of approximately 470msec is included in the safeguard circuit to guard against premature enabling of the relays. The time constant of the hold-off period is deter­mined by the relative values of R419 and C419.
4-3
Page 54
Service Information
4.3.5 Isolator relays
Row isolator relays are necessary in addition to the cross­point relays in order to ensure the integrity of signal path­ways. Path isolator relays include K198-K204. The necessary isolator relay is closed in addition to the selected crosspoint to complete the entire pathway. For example, if crosspoint C10 (Row C, Column 10) is closed, relays K199 and K183 would be energized.

4.4 T roubleshooting

4.4.1 Recommended equipment
T able 4-1 summarizes the recommended equipment for gen­eral troubleshooting.
Table 4-1
Recommended troubleshooting equipment
Manufacturer
Description
5 ½ digit DMM Oscilloscope Extender card
and model Application
Keithley 199 TEK 2243 Keithley 7070
Measure DC voltages View logic waveforms Allow circuit access
4.4.2 Gaining circuit access
In order to gain access to the test points and other circuitry on the Model 7174A, the card must be plugged into the Model 7070 Extender Card, which, in turn, must be plugged into the desired slot of the mainframe. The Model 7070 must be configured as an extender card by placing the configura­tion jumper in the EXTEND position. See the documentation supplied with the Model 7070 for complete details on using the card.
NOTE
Do not use the Model 7070 for performing verification tests because its presence will affect the results.
4.4.3 T roubleshooting procedure
Table 4-2 summarizes the troubleshooting procedure for the Model 7174A Low Current Matrix Card. Some of the trou­bleshooting steps refer to the ID data timing diagram shown in Figure 4-2. Refer to paragraph 4.3 for an overview of oper ­ating principles.
Table 4-2
Troubleshooting procedure
Step Test point/component Required condition Comments
1
DGND
2
4-4
10
11
3 4 5 6 7 8 9
+6V +5V NXTADR CLRADR IDDATA STRB RLDAT CLK OE
U101-U113 pins 10-18
+6VDC +5VDC NEXT ADDR pulses CLR ADDR pulse ID data pulses STROBE pulse Relay data (128 bits) CLK pulses High on power up until first STROBE sets low. Low with relay energized, high with relay de-energized.
All voltages referenced to DGND (digital common) Relay voltage Logic voltage Power up only (Fig. 4-2) Power up only (Fig. 4-2) Power up only (Fig. 4-2) End of relay data sequence Present when updating relays Present during relay data or ID data (Fig. 4-2) Power on safe guard
Relay driver outputs
Page 55
Service Information

4.5 Special handling of static-sensitive devices

CMOS and other high-impedance devices are subject to pos­sible static discharge damage because of the high-impedance levels in volved. When handling such devices, use the precau­tions listed below.
NOTE
In order to prevent damage, assume that all parts are static sensitive.
1. Such devices should be transported and handled only in containers specially designed to prevent or dissipate static build-up. Typically , these devices will be received in anti-static containers made of plastic or foam. Keep these parts in their original containers until ready for in­stallation or use.
2. Remove the devices from their protective containers only at a properly-grounded work station. Also ground yourself with an appropriate wrist strap while working with these devices.
3. Handle the devices only by the body; do not touch the pins or terminals.
4. Any printed circuit board into which the device is to be inserted must first be grounded to the bench or table.
5. Use only anti-static type de-soldering tools and grounded-tip soldering irons.
4.6 Performance verification
The following paragraphs discuss performance verification procedures for the Model 7174A, including relay testing, contact resistance, contact potential, path isolation, and leak­age current.
4.6.1 Environment conditions
All verification measurements except for path isolation and offset current should be made at an ambient temperature of 21°C-25°C and a relative humidity of less than 0% R.H. Path isolation and offset current verification must be performed at an ambient temperature of 23°C and a relative humidity of less than 60% R.H. If the card has been subjected to temper­atures or humidities outside of this range for even a short time, allow it to stabilize within these ranges for at least one hour before performing any tests.
4.6.2 Recommended test equipment
Table 4-3 summarizes the equipment necessary to make the performance verification tests, along with the application for each item.
Table 4-3
Recommended verification equipment
Qty. Description Application
1
Model 617 Electrometer
1
Model 196 6½ Digit DMM
1
Model 707A Switching Matrix
4
Model 7078-TRX-10 triax cables*
2
Model 7078-TRX-3 triax cables
1
Model 6172 2-slot male to 3-lug female triaxial adapter
3
Model 7078-TRX-T triax tee adapter
5
Banana plugs (part #BG-10-2*)
1
Model 263 Calibrator/Source
1
BNC to Right-angle SMB Cable (part #CA-93-1)
1
BNC to Dual Banana Adapter (Pomona part #1269)
*These items are used to construct special cables; see text and Figure 4-9.
Offset current; path isolation Path resistance; electrometer All tests Offset current; path resistance Path isolation; offset current; electrometer Offset current; electrometer Path resistance Path isolation and resistance Electrometer Electrometer Electrometer
4-5
Page 56
Service Information
4.6.3 Offset current verification
Recommended equipment
• Model 707A Switching Matrix
• Model 617 Electrometer
• Model 7078-TRX-3 Triax Cable
• Model 6172 2-slot male to 3-lug female triaxial adapter
Test connections
Figure 4-3 shows the test connections for offset current verification. The Model 7174A row being tested is to be connected to the Model 617 Electrometer input through the triaxial cable and the triaxial adapter. Note that the electrometer ground strap is to be removed, and the electrometer should be operated in the unguarded mode.
Procedure
NOTE
The following procedure should be per­formed at an ambient temperature of 23°C and at a relative humidity of less than 60%.
1. Turn on the Model 617 power and allow it to warm up for two hours before beginning the verification procedure
6172 2-Slot to
3-Lug Triax Adapter
Guard off
2. With the power off, install the Model 7174A in the desired slot of the Model 707A Switching Matrix. Remove all other cards from the instrument, and install the slot covers.
3. After the prescribed warm up period, select the amps function and the 2pA range on the Model 617. Zero cor­rect the instrument, and then select autoranging.
4. Connect the Model 617 to row A of the Model 7174A, as shown in Figure 4-4.
5. Close crosspoint A1 by using the Model 707A front panel controls.
6. Disable zero check on the Model 617, and allow the reading to settle.
7. Verify that the offset current reading is <500fA.
8. Enable zero check on the Model 617, and open cross­point A1.
9. Repeat steps 5 through 8 for crosspoints A2 through A12. Only one crosspoint at a time should be closed.
10. Disconnect the triax cable from row A, and connect it instead to row B.
11. Repeat steps 5 through 8 for crosspoints B1 through B12. Only one crosspoint at a time should be closed.
12. Connect the triax cable to each succeeding row and repeat steps 5 through 8 for each of the row’s crosspoints.
Connect Cable
to Row Under
Test
7078-TRX-3 Triax Cable
Ground Link
Removed
200VPK
200VPK
A
B
C
D
KEITHLEY
8x12LOW
CURRENT MATRIX
SIGNAL
200VPK
GUARD
ROWS
7174A
COLUMNS
1
2
3
4
5
617
7174A
A
Equivalent Circuit
Figure 4-3
Offset verification test connections
4-6
617 Electrometer
E
F
G
H
WARNING:
TIGHTENMOUNTING SCREWS
TOENSURE PROPER
CHASSISGROUN
Matrix Card7174A
6
7
8
9
10
11
12
!
D
Page 57
Service Information
4.6.4 Path isolation verication
The procedure for verifying path isolation is discussed be­low. Should the card fail any of the tests, clean it using the procedures outlined in paragraph 4.2.
Recommended equipment
• Model 707A Switching Matrix
• Model 617 Electrometer
• Model 7078-TRX-3 Triax Cable
• Unterminated 3-slot triaxial cable (cut connector off 7078-TRX-3)
• Banana plug (Keithley part #BG-10-2)
• #16-18AWG insulated stranded wire (6 in. length)
Test connections
Figure 4-4 shows the test connections for the path isolation tests. One row being tested is to be connected to the Model 617 Electrometer input through a Model 6172 2-slot female
6172
Adapter
Guard off
to 3-lug male triaxial adapter. The other row is to be con­nected to the voltage source HI terminal using a specially prepared 3-slot triax-to-banana plug cable, the construction of which is shown in Figure 4-5. Note that both the inner shield and the center conductor are to be connected to the banana plug as shown.
COM and LO terminal of the electrometer voltage source must be connected together as shown. Also, the ground link between COM and chassis must be removed, and the Model 617 guard must be turned off for current measurements.
Procedure
WARNING
Hazardous voltage from the electrome­ter voltage source will be used in the fol­lowing steps. Take care not to contact live circuits, which could cause personal injury or death.
KEITHLEY
7174A
8x12LOW
CURRENT MATRIX
SIGNAL
200VPK
COLUMNS
200VPK
GUARD
200VPK
7078-TRX-3 Triax Cable
Ground Link
Removed
ROWS
1
A
2
B
3
C
4
D
5
617
7174A
A
F
Equivalent Circuit
Figure 4-4
Connections for path isolation verification
Row A
Row B
617 Electrometer
User-Prepared Triax Cable
(See Figure 4-5)
Warning: Hazardous voltage from
the electrometer source may be present on terminals.
6
7
8
E
9
F
10
G
11
H
!
WARNING:
TIGHTENM OUNTINGSCREWS
TOENSURE PROPER
CHASSISGROUND
12
7174A Matrix Card
4-7
Page 58
Service Information
Cut
(A) Cut off insulation with knife.
Cut off outer shield.
Insulation Over Inner Shields
3/4"
(B) Strip insulation off inner shield.
(C) Twist inner shield then strip inner conductor.
Twister inner shield and center conductor together, slip on plastic cover.
(D) Insert wires into hole and wrap around body.
(E) Screw on plastic cover.
Figure 4-5
Triaxial cable preparation
NOTE
The following procedure must be per­formed at an ambient temperature of 23°C and at a relative humidity of less than 60%.
Cut
1. Turn on the Model 617 and allow it to warm up for two hours for rated accuracy.
2. With the mainframe power turned off, plug the Model
1"
7174A into slot 1 of the mainframe. Remove all other cards from the mainframe, and install the slot covers.
3. After the prescribed warm up period, select the Model 617 amps function, and enable zero check. Select the 2pA range, and zero correct the instrument.
4. Connect the Model 617 to rows A and B of the matrix card, as shown in Figure 4-4.
5. Program the Model 617 voltage source for a value of +100V, but do not yet turn on the voltage source output.
6. Close crosspoints A1 and B2 by using the switching ma­trix front panel controls.
7. With the Model 617 in amps, enable suppress after the reading has settled.
8. Turn on the Model 617 voltage source output, and en­able the V/I ohms function on the electrometer.
9. After the reading has settled, verify that the resistance is >67T Ω (6.7 × 10
13
).
10. Turn off the voltage source, and enable zero check. Disable suppress, and select the amps function on the electrometer.
11. Open crosspoints A1 and B2, and close crosspoints A3 and B4.
12. Repeat steps 7 through 11 for A3 and B4.
13. Repeat steps 7 through 12 for crosspoint pairs A5 and B6, A7 and B8, A9 and B10, and A11 and B12.
14. Disconnect the electrometer from rows A and B, and connect it instead to rows C and D.
15. Repeat steps 7 through 13 for rows C and D. The path isolation for these rows should be 67T Ω (6.7 × 10
3
).
16. Repeat steps 7 through 14 for row pairs E and F, and G and H. For each row pair, step through the crosspoint pairs 1 and 2, 3 and 4, 5 and 6, 7 and 8, 9 and 10, and 11 and 12. The complete procedure outlined in steps 7 through 11 should be repeated for each crosspoint pair. Each resistance measurement for rows E through H should be 67T Ω (6.7 × 10
13
).
4-8
Page 59
Service Information
4.6.5 Path resistance verication
The following paragraphs discuss the equipment, connec­tions, and procedure to check path resistance. Should a par­ticular pathway fail the resistance test, the relay (or relays) for that particular crosspoint is probably defective. See the schematic diagram at the end of Section 5 to determine which relay is defective.
NOTE
The following procedure verifies the resis­tance of the HI signal path. To verify the resistance of the GUARD path, modify the cable of Figure 4-5 to connect the inner shield to the banana plug and have no con­nection to the triax center conductor.
Recommended equipment
• Model 196 DMM
• 7078-TRX-T triax tee adapters (3)
• 237-BAN-3 triax to banana cables (4)
Connections
Figure 4-6 shows the connections for the path resistance tests. The Model 196 is to be connected to the row and col­umn jacks using Model 237-BAN-3 triax/banana cables. These cables differ from the one in Figure 4-5 in that the inner shield and center conductor are not connected together.
7078-TRX-T
Triax Tee Adapters
237-BAN-3
Triax/Banana
196 DMM
Triax
Figure 4-6
Connections for path resistance verification
7174A
KEITHLEY
8x12LOW
CURRENT MATRIX
SIGNAL
COLUMNS
200VPK
200VPK
GUARD
200VPK
ROWS
B
C
D
E
F
G
H
!
WARNING:
TIGHTENM OUNTINGSCREWS
TOENSURE PROPER
CHASSISGROUND
7174A Matrix Card
4-9
Page 60
Service Information
Procedure
1. Turn on the Model 196 DMM and allow it to warm up for at least one hour before beginning the test.
2. With the po wer off, install the Model 7174A card in slot 1 of the mainframe.
3. Connect the four triaxial cables to the Model 196 and the two triax tee adapters (Figure 4-6), but do not yet connect the adapters to the Model 7174A.
4. Temporarily connect the two triax tee connectors to­gether using a third triax tee adapter, as shown in Figure 4-7.
5. Select the ohms function, 300 Ω range, and 6 ½ digit res­olution on the Model 196.
6. After the reading settles, enable zero on the Model 196 DMM. Leave zero enabled for the remainder of the tests.
7. Disconnect the two triax tee adapters from the shorting adapter, and connect the two adapters with the cable to
To
196
the row A and column 1 connectors on the Model 7174A (see Figure 4-6).
8. Close crosspoint A1, and allow the reading to settle.
9. Verify that the resistance reading is <1.5 Ω
.
10. Open the crosspoint, and disconnect the triax adapter from column 1. Connect the adapter to column 2.
11. Repeat steps 8 through 10 for columns 2 through 12. In each case, the column adapter must be connected to the column under test, and the crosspoint must be closed.
12. Disconnect the row adapter from row A, and connect it instead to row B.
13. Repeat steps 8 through 10 for row B. The crosspoints of interest here are B1 through B12. Also, the row adapter must be connected to the row being tested.
14. Repeat steps 8 through 13 for rows C through H. In each case, the crosspoint to close is the one corresponding to the row and column connections at that time. In all cases, the measured resistance should be <1.5 Ω
.
To
196
Figure 4-7
Shorting measurement paths using triax tee adapter
4-10
Triax Tee
Adapters
Page 61
Service Information

4.7 Reed pack replacement

If after performing troubleshooting or verification, a reed switch is thought to be faulty , it should be replaced. Note that reeds can only be replaced as a pack of three. The part num­ber for a reed pack is RL-179. A reed pack maintenance kit, part number 9174-MK is also suggested.
Cross point relay
Use the following procedure and Figure 4-8 to replace cross point relay reed packs.
NOTE
Repairs should only be done from the component side of the board.
1. Remove the card and place on an anti-static surface.
2. Put on the gloves that came in the maintenance kit.
3. Carefully pry off the cover of the guard enclosure of the crossbar that includes the suspect reed. Take care not to break the wire that connects the cover to the front panel.
4. Carefully pry off the cross point interconnect board. Take care not to break the wires that connect the inter­connect board to the front panel.
5. Using clean needle-noise pliers, carefully remove the reed pack in question.
6. Inspect the relay coil. If there is any debris or contamination, clean with methanol and blow out with dry nitrogen.
NOTE
When inserting reed pack, make sure long leads are inserted into relay coil (bobbin).
7. Using the reed pack insertion tool from the maintenance kit, install the new reed pack into bobbin making sure the long leads align with the holes at the base of the relay coil (bobbin).
8. Reinstall any reed packs that may have been removed when the cross point interconnect board was removed.
9. Reinstall the cross point interconnect board. Be sure that it is properly seated.
10. Reinstall the guard enclosure cover.
11. Allow the card to stabilize for one hour.
12. Retest the card to verify operation and specifications.
Isolator relay
Use the following procedure and Figure 4-9 to replace isola­tor relay reed packs.
NOTE
Repairs should only be done from the component side of the board.
1. Remove the card and place on an anti-static surface.
2. Put on the gloves that came in the maintenance kit.
3. Carefully pry off the cover of the guard enclosure. Take care not to break wires connecting the relay to the path­way interconnect board.
4. Using clean needle nose pliers, carefully disconnect the three wires that attach to the reed pack, taking note of the locations for each.
5. Remove the reed pack in question.
6. Inspect the reed pack socket. If there is any debris or contamination, clean with methanol and blow out with dry nitrogen.
NOTE
When inserting reed pack, make sure long leads are inserted into relay coil (bobbin).
7. Using the reed pack insertion tool from the maintenance kit, install the new reed pack into bobbin making sure the long leads align with the holes at the base of the relay coil (bobbin).
8. Reconnect the wires to the top of the reed pack.
9. Reinstall the guard enclosure cover.
10. Allow the card to stabilize for one hour.
11. Retest the card to verify operations and specifications.
4-11
Page 62
Service Information
Isolator
Relays (8)
N
ote
When inserting reed
:
pack, make sure long leads are inserted into bobbin.
Figure 4-8
Cross point relays
Clear
Triax
Connector
Yellow
TP1
Cross point
Relays (12 sets)
Bobbin
Cross point
Interconnect
Board
Cover
Crosspoint
Relay
Reed Pack
4-12
Page 63
Pathway Interconnect Board
Service Information
Isolator
Relays (8)
Clear
Red
Bobbin
Isolator
Relay
Reed Pack
Note:
When inserting reed pack, make sure long leads are inserted into bobbin.
Black
Red
Black Yellow
Yellow
Clear
Figure 4-9
Isolator relays
Cross point
Relays (12 sets)
4-13
Page 64
5

Replaceable Parts

5.1 Introduction

This section contains a list of replaceable electrical and mechanical parts for the Model 7174A, as well as a compo­nent layout drawing and schematic diagram of the matrix card.

5.2 Parts list

Electrical parts are listed in order of circuit designation in Table 5-1. Table 5-2 summarizes mechanical parts.

5.3 Ordering information

To place an order, or to obtain information about replace­ment parts, contact your Keithley representative or the fac­tory (see the front of this manual for addresses). When ordering parts, be sure to include the following information:
1. Matrix card model number 7174A.
2. Card serial number
3. Part description
4. Circuit designation, if applicable
5. Keithley part number

5.4 Factory service

If the matrix card is to be returned to Keithley Instruments for repair, perform the following:
1. Call the Repair Department at 1-800-552-1115 for a Return Authorization (RMA) number.
2. Complete the service form located at the back of this manual, and include it with the unit.
3. Carefully pack the card in the original packing carton or the equivalent.
4. Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping label. Note that it is not necessary to return the matrix mainframe with the card.

5.5 Component layout and schematic diagram

Mother board
• Component layout — 9174-100
• Schematic — 9174-106
Relay board
• Component layout — 9174-120
• Schematic — 9174-126 (sheets 1 and 4 only)
Pathway interconnect board
• Component layout — 7174-160
• Schematic — 7174-166
5-1
Page 65
Table 5-1
Model 7174A electrical parts list
Circuit Desig. Description Keithley Part No.
Replaceable Parts
C101-113 C415-418, 421 C413 C414, 425 C419 C420, 423 C422, 424
CR403 CR402
J1001 J1001-1003
K101-204 K101-204
R101-128 R417, 418 R419 R420 R421 R422 R423
CAP, .1µF, 20%, 50V, CERAMIC CAP, .1µF, 20%, 50V, CERAMIC CAP, .01µF, 10%, 1000V, CERAMIC CAP, 10µF, -20+100%, 25V, ALUM ELEC CAP, 47µF, 10%, 16V, ALUM ELEC CAP, .01µF, 20%, 50V, CERAMIC CAP, 270pF, 20%, 100V, CERAMIC/FERRITE
DIODE, SILICON, IN4148 (DO-35) DIODE, SCHOTTKY, IN5711
CONNECTOR CONNECTOR, MALE
REED PACK 3 FORM, DUAL GUARDED RELAY, AIR MATRIX BOBBIN
RES, 22M, 5%, 1/4W, COMPOSITION OR FILM RES, 10K, 5%, 1/4W, COMPOSITION OR FILM RES, 47K, 5%, 1/4W, COMPOSITION OR FILM RES, 200, 5%, 1/4W, COMPOSITION OR FILM RES, 680, 5%, 1/4W, COMPOSITION OR FILM RES, 120K, 5%, 1/4W, COMPOSITION OR FILM RES, 11K, 5%, 1/4W, COMPOSITION OR FILM
C-365-.1 C-365-.1 C-64-.01 C-314-10 C-321-47 C-237-.01 C-386-270P
RF-28 RF-69
CS-598 CS-389-2
RL-179 RL-177
R-76-22M R-76-10K R-76-47K R-76-200 R-76-680 R-76-120K R-76-11K
U101-113 U406 U407 U408 U409 U410
W401
*Order present firmware revision level.
IC, 8-BIT SERIAL-IN LATCH DRIVER, 5841A PROGRAM IC, 12 STAGE BINARY COUNTER, 74HCT4040 IC, 8-BIT PARALLEL TO SERIAL, 74HCT165 IC, QUAD 2 INPUT NAND, 74HCT00 IC, OCTAL BUFFER/LINE DRIVER, 74HC244
STIFFENER, BOARD
IC-536 7174-800A02* IC-545 IC-548 IC-399 IC-489
J-16
5-3
Page 66
Replaceable Parts
Table 5-2
Model 7174A mechanical parts list
Description Keithley Part No.
#4-40X1/4 PHILLIPS PAN HD SEMS SCREW (9174-162 BOARD TO BRACKETS) #6-32X3/8LG. PHIL FLAT HD SCR (HANDLE MTG) #6-32X5/16PHIL PAN HED SEMS SCR (R. PANEL TO MB; BRKT TO BD) #6-32 PEM NUT 8 POS INTERCONNECT 13 POS INTERCONNECT BRACKET BRACKET CABLE CABLE ASSEMBLY, 16 CONDUCTOR CAP, PROTECTIVE CONN, TEST POINT (DONE,IDDATA,NXTADR,OR,P,/F,RDAT,STRB) CONN, TEST POINT (+5V, +VTH, -VTH,AGND,CLK,CLRADR,DGND) CONNECTOR (ON SC-120-4) FASTENER FEMALE, BULKHEAD MOUNT RECEPTACLE GROUND CLIP GUARD TUBE, BASE GUARD TUBE, LID GUARD TUBE, BASE GUARD TUBE, LID GUARD TUBE, BASE GUARD TUBE, LID HANDLE LOCKNUT (REL BD TO FASTENERS, BRKT TO BD) LUG (ON SC-83, SC-120-4) LUG (ON SC-120-4) REAR PANEL ASSEMBLY SOCKET (SOLDERED ON SC-83 AND SC-120-4) SOCKET, I.C. 28 PIN (FOR U406) SPACER, LID STRIP TERMINAL (FOR SC-111) SUPPORT TEFLON TWISTED PAIR SHLD TERMINAL, BIFURCATED (TEFLON) TEST PIN TRIAX CONNECTOR (FOR SC-111 WIRE, SINGLE CONDUCTOR
4-40X1/4PPHSEM 6-32X3/8PFH 6-32X5/16PPHSEM FA-135 9174-315B 9174-314B 9174-307A 9174-308A CA-121A CA-27-13C CAP-30-1 CS-553 CS-553 CS-236 FA-257-1 CS-824 9174-313A 9174-301A 9174-302A 9174-303A 9174-304A 9174-305A 9174-306A HH-33-1 FA-226-1 LU-108-2 LU-108-2 7174A-301A SO-144 SO-69 9174-316A CS-812 9174-309A SC-83 TE-119 TE-108 CS-827 SC-120-4
5-4
Page 67
M1
Page 68
Page 69
Page 70
Page 71
Page 72
Page 73
Page 74

Index

A
Applications 3-1
B
Basic CV test procedure 3-3 Block diagram 4-2
C
Cable connections 3-10 Card connectors 2-2 Card installation and removal 2-2 Coaxial jumper access 2-25 Component layout and schematic diagram
5-1 Computerized system configuration 3-1 Connections 2-2 CV measurements 3-1
E
Electromagnetic Interference (EMI) 2-22 Environment conditions 4-5 Environmental considerations 2-1
F
Factory service 5-1 Features 1-1
G
Gaining circuit access 4-4 General information 1-1 General instrument connections 2-5 Ground loops 2-22 Guarding 2-24
H
Handling and cleaning precautions 4-1 Handling precautions 2-1
I
ID data circuits 4-2 Inspection for damage 1-2 Instruction manual 1-2 Internal matrix expansion 2-21 Introduction 1-1, 2-1, 3-1, 4-1, 5-1 Isolator relays 4-4
K
Keeping connectors clean 2-23 Keithley instrument connections 2-11
M
Magnetic fields 2-22 Manual addenda 1-2 Matrix configuration 2-18 Matrix expansion effects on card specifica-
tions 2-24
Measurement considerations 2-22
N
Noise currents caused by cable flexing
2-23
O
Offset current verification 4-6 Operation 2-1 Optimizing CV measurement accuracy
3-3 Optional accessories 1-2 Ordering information 5-1
P
Packing for shipment 1-2 Parts list 5-1 Path isolation verification 4-7 Path isolators 2-18 Path resistance verification 4-9 Performance verification 4-5 Power-on sequence 4-3 Principles of operation 4-2
R
Recommended cables and adapters 2-3 Recommended equipment 4-4 Recommended test equipment 4-5 Reed pack replacement 4-11 Relay control 4-3 Replaceable parts 5-1 Resistivity calculations 3-9 Resistivity measurements 3-7
S
Safety symbols and terms 1-2 Semiconductor IV characterization 3-9 Semiconductor test matrix 3-5 Service information 4-1 Shielding 2-23 Shipment contents 1-2 Special handling of static-sensitive devices
4-5 Specifications 1-2 Stand alone system configuration 3-1 Switching matrix 2-18 System configuration 3-5
T
Test configuration 3-7, 3-9 Test procedure 3-7 Testing common-source characteristic of
FETs 3-6 Triax banana plug adapter 2-4
i-1
Page 75
Troubleshooting 4-4 Troubleshooting procedure 4-4 Typical CV curves 3-3 Typical test fixture connections 2-17
U
Unpacking and inspection 1-2
W
Warranty information 1-1
i-2
Page 76
Service Form
Model No. Serial No. Date Name and Telephone No. Company
List all control settings, describe problem and check boxes that apply to problem.
Intermittent
❏ ❏
IEEE failure Front panel operational
Display or output (check one)
Drifts
Unstable
❏ ❏
Overload
Calibration only Data required
(attach any additional sheets as necessary) Show a block diagram of your measurement system including all instruments connected (whether power is turned on or not).
Also, describe signal source.
Analog output follows display
❏ ❏
Obvious problem on power-up All ranges or functions are bad
Unable to zero
Will not read applied input
Certificate of calibration required
Particular range or function bad; specify
❏ ❏
Batteries and fuses are OK Checked all cables
Where is the measurement being performed? (factory, controlled laboratory, out-of-doors, etc.)
What power line voltage is used? Ambient temperature? °F
Relative humidity? Other?
Any additional information. (If special modifications have been made by the user, please describe.)
Be sure to include your name and phone number on this service form
.
Page 77
Keithley Instruments, Inc.
28775 Aurora Road Cleveland, Ohio 44139
Printed in the U.S.A.
Loading...