Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 3 years from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation
prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written
consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable
batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE
REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR
ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT
OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS
BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED
DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND
INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE
TO PROPERTY.
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Level letter increases alphabetically as the manual undergoes
subsequent updates. Addenda, which are released between Revisions, contain important change
information that the user should incorporate immediately into the manual. Addenda are
numbered sequentially. When a new Revision is created, all Addenda associated with the
previous Revision of the manual are incorporated into the new Revision of the manual. Each new
Revision includes a revised copy of this print history page.
Revision A (Document Number 2750-900-01).............................................................. March 2001
Revision B (Document Number 2750-900-01) ........................................................November 2001
Revision C (Document Number 2750-900-01) ........................................................November 2002
Revision D (Document Number 2750-900-01).............................................................August 2003
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
S
afety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before
using the product. Refer to the manual for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product may be impaired.
The types of product users are:
Responsible body
ment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel
voltage or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state
if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel
trained service personnel may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category
II, as described in the International Electrotechnical Commission (IEC) Standard IEC 60664. Most measurement, control, and
data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with
high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often
associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to
Category I sources unless otherwise marked or described in the Manual.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures.
The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS,
42.4V peak, or 60VDC are present.
circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock.
If the circuit is capable of operating at or above 1000 volts,
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited
sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective
devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be provided, in close proximity to the equipment and within easy reach of the operator.
is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equip-
use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
perform routine procedures on the product to keep it operating properly, for example, setting the line
are trained to work on live circuits, and perform safe installations and repairs of products. Only properly
A good safety practice is to expect that hazardous voltage is present in any unknown
no conductive part of the circuit may be exposed.
5/03
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground.
Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being
measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating
information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of
normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The symbol indicates a connection terminal to the equipment frame.
The
WARNING
information very carefully before performing the indicated procedure.
The
CAUTION
warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer,
test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses, with applicable national safety
approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from
other suppliers as long as they are equivalent to the original component. (Note that selected parts should be purchased only
through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability
of a replacement component, call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for
proper cleaning/servicing.
heading in a manual explains dangers that might result in personal injury or death. Always read the associated
heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the
Table of Contents
1Getting Started
General information ................................................................... 1-2
Warranty information .......................................................... 1-2
Contact information ............................................................ 1-2
Safety symbols and terms ................................................... 1-2
acquaint yourself with fundamental instrument operations. Use
yourself with front panel controls, use
QS3
QS1 •
QS2 •
— Of the following section topics, three can be used immediately to quickly
QS1
to familiarize
QS2
to power-up the instrument, and, finally, use
to perform exercises to operate the instrument.
•
General information
information, contact information, safety symbols and terms, inspection, and
available options and accessories.
•
Model 2750 features
•
Plug-in switching modules
Model 77XX series switching modules.
Front and rear panel familiarization
of the instrument.
•
Rack mounting
in a standard 19-inch rack.
Power-up
power line frequency, and the power-up sequence.
•
Display
•
Defaults and user setups
covers the three setup configurations available to the user.
— Covers line power connection, line voltage setting, fuse replacement,
— Provides information about the display of the Model 2750.
— Covers general information that includes warranty
— Summarizes the features of Model 2750.
— Summarizes the capabilities of the Keithley
— Summarizes the controls and connectors
— Covers the options available for rack mounting the Model 2750
— Lists the *RST and factory default settings, and
QS3 •
•
Remote programming information
presented in this manual.
Quick start exercises
(front panel and remote programming) to acquaint a user with operation basics.
— Provides abbreviated operating information and exercises
— Explains how SCPI commands are
1-2Getting StartedModel 2750 Multimeter/Switch System User’s Manual
General information
Warranty information
Warranty information is located at the front of this manual. Should your Model 2750
require warranty service, contact the Keithley representative or authorized repair facility in
your area for further information. When returning the instrument for repair, be sure to fill
out and include the service form at the back of this manual to provide the repair facility
with the necessary information.
NOTE
The service form requires the serial number of the Model 2750. The serial
number is located on the rear panel. It can also be read from the front panel by
selecting the SNUM item of the SETUP menu. Press SHIFT and then SETUP to
access the menu. For remote operation, the serial number can be read using the
*IDN? command (see Section 12 for details).
Contact information
Worldwide phone numbers are listed at the front of this manual. If you have any questions,
please contact your local Keithley representative or call a Keithley Application Engineer at
1-800-348-3735 (U.S. and Canada only).
Safety symbols and terms
The following symbols and terms may be found on the instrument or used in this manual:
!
Thesymbol on an instrument indicates that the user should refer to the operating
instructions located in the manual.
The symbol on the instrument shows that high voltage may be present on the
terminal(s). Use standard safety precautions to avoid personal contact with these voltages.
The
WARNING
personal injury or death. Always read the associated information very carefully before
performing the indicated procedure.
heading used in this manual explains dangers that might result in
The
CAUTION
instrument. Such damage may invalidate the warranty.
heading used in this manual explains hazards that could damage the
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-3
Inspection
Model 2750 was carefully inspected electrically and mechanically before shipment. After
unpacking all items from the shipping carton, check for any obvious signs of physical
damage that may have occurred during transit. (There may be a protective film over the
display lens, which can be removed). Report any damage to the shipping agent
immediately. Save the original packing carton for possible future shipment. The following
items are included with every Model 2750 order:
•Model 2750 with line cord.
•Safety test leads (Model 1751).
•Accessories as ordered.
•Hardware for rack mounting.
•Certificate of calibration.
•Model 2750 User’s Manual (P/N 2750-900-00).
•Manual Addenda (pertains to any improvements or changes concerning the
instrument or manual).
•Software CD containing the following:
•TestPoint Runtime – Provides basic data logging capabilities. This can be
modified with the TestPoint application development package (optional
software).
•KE2700 IVI Instrument Driver – Provided for programmers, designed for use
with application development environments.
Optional “Software” available from Keithley is summarized on page 1-6.
If an additional manual is required, order the appropriate manual package. The manual
packages include a manual and any pertinent addenda.
Options and accessories
Plug-in switching modules
NOTETable 1-1 provides a side-by-side comparison of the following Keithley
switching modules. All multiplexer modules can be configured as two
independent multiplexers.
NOTEThe Model 77XX Series Switching Modules Instruction Manual provides
operating and service information for the switching modules. This manual is
supplied with each switching module.
Model 7700 — This differential multiplexer provides 20 channels of 2-pole input, or 10
channels of 4-pole input. The internal cold junction allows direct-connection of
thermocouples. It also has two 2-pole channels used exclusively for current input.
1-4Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Model 7701 — This differential multiplexer provides 32 channels of 2-pole input, or
16 channels of 4-pole input.
Model 7702 — This differential multiplexer provides 40 channels of 2-pole input, or
20 channels of 4-pole input. It also has two 2-pole channels used exclusively for current
input.
Model 7703 — This differential multiplexer provides 32 channels of 2-pole input, or
16 channels of 4-pole input.
Model 7705 — This control module provides 40 independent 1-pole switching (SPST)
channels that are isolated from the internal DMM.
Model 7706 — This all-in-one module provides 20/10 channels of 2/4-pole input,
16 digital outputs, two analog outputs, one 32-bit counter with gating and totalizer.
Model 7707 — This module provides 10 channels of 2-pole input, or 5 channels of 4-pole
input. Also provides 32 digital inputs/outputs.
Model 7708 — This differential multiplexer provides 40 channels of 2-pole input, or 20
channels of 4-pole input. The internal cold junction allows direct-connection of
thermocouples for temperature measurements.
Model 7709 — This module is configured as a 6 × 8 matrix (six rows, eight columns). The
matrix consists of 48 crosspoint channels and two backplane isolation channels. For
system channel operation, row 1 is connected to DMM Input. For 4-wire measurements,
row 2 is connected to DMM Sense.
Model 7710 — This differential multiplexer provides 20 channels of 2-pole input or
10 channels of 4-pole input. The internal cold junction allows direct-connection of
thermocouples for temperature measurements. This module provides high-speed
switching and uses long-life relays.
Model 7711 — The Model 7711 is a 50Ω, 2GHz, single-pole dual 1 × 4 RF Multiplexer
module (eight channels, no measurement capability). This 1 × 4 multiplexer is a cascading
tree design — one of the channels of each is always connected to a common out. It can be
used to connect one instrument to multiple devices or multiple instruments to a single
device.
Model 7712 — The Model 7712 is a 50Ω, 3.5GHz, single-pole dual 1 × 4 RF Multiplexer
module (eight channels, no measurement capability). This 1 × 4 multiplexer is a cascading
tree design — one of the channels of each is always connected to a common out. It can be
used to connect one instrument to multiple devices or multiple instruments to a single
device.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-5
Cables and connector kits for switching modules
Model 7788 DB-50 connector kit — Contains two male DB-50 solder cup connectors
with strain relief connector shells. These connectors mate to the female connectors of the
Models 7703 and 7705 switching modules.
Model 7789 50/25-pin solder cup connector kit — Contains one male DB-50 and one
male DB-25 solder cup connectors. These connectors mate to the female connectors on the
Models 7701 and 7709 switching modules.
Model 7790 ribbon cable adapter kit — Contains one female DB-50, one male DB-50
and one male DB-25 IDC ribbon cable connectors. These connectors are used with the
Models 7701, 7707 and 7709 switching modules.
Model 7051-X — BNC cable (male to male). 7051-2 is 2 ft long, 7051-5 is 5 ft long, and
7051-10 is 10 ft long. These cable are used with the Model 7711 switching module.
Model 7712-SMA-1 — SMA cable (male to male), 1.0m (3.3 ft) long. This cable is used
with the Models 7711 and 7712 switching modules.
Model 7712-SMA-N — Female SMA to male N-type adapter. This adapter is used with
the Models 7711 and 7712 switching modules.
S46-SMA-X — SMA cable (male to male). S46-SMA-1 is one foot long and
S46-SMA-0.5 is one-half foot long. This cable is used with the Models 7711 and 7712
switching modules.
Cables and adapters (GPIB and trigger link)
Models 7007-1 and 7007-2 shielded GPIB cables — Connect Model 2750 to the GPIB
bus using shielded cables and connectors to reduce electromagnetic interference (EMI).
Model 7007-1 is one meter long; Model 7007-2 is two meters long.
Models 8501-1 and 8501-2 trigger link cables — Connect Model 2750 to other
instruments with Trigger Link connectors (e.g., Model 7002 Switch System).
Model 8501-1 is one meter long; Model 8501-2 is two meters long.
Model 8502 trigger link adapter — Lets you connect any of the six trigger link lines of
Model 2750 to instruments that use the standard BNC trigger connectors.
Model 8503 DIN to BNC trigger cable — Lets you connect trigger link lines one
(Voltmeter Complete) and two (External Trigger) of Model 2750 to instruments that use
BNC trigger connectors. Model 8503 is one meter long.
1-6Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Software
The following optional software is available from Keithley:
ExceLINX-1A – This is an economical, easy-to-use, add-in utility for Microsoft Excel ®
and Keithley Integra Series Multimeter/Switch systems. No programming is required —
configure your measurements quickly using pop-up menus and eliminate time-consuming
and error prone programming. Acquire data into a spreadsheet on the fly during a scan or
transfer data into a spreadsheet after a scan is completed. A few mouse clicks are all it
takes to configure channels, set parameters, triggers, and scan lists. There is no need to
launch a separate data logging or data-crunching application; live data streams
automatically into an Excel workbook, ready for analysis or charting using all of Excel's
powerful built-in tools.
TestPoint application development package — This powerful and economical
programming environment uses object-oriented technology through a drag-and-drop
interface to build a basic system quickly and without in-depth programming. Optional
toolkits (database and statistical process control) are available to expand ExceLINX-1A
capability.
Rack mount kit
Model 4288-7 rack mount kit — Mounts a Model 2750 in a standard 19-inch rack.
Includes rear brackets to provide additional support for a mainframe that has two or more
switching modules installed.
NOTEThe Model 2750 includes hardware that allows it to be mounted to the front rails
of a standard 19-inch rack. With two or more switching modules installed, rear
support brackets may be required. The Model 4288-7 rack mount kit includes
rear support brackets.
Model 2750 features
Model 2750 is a 6½-digit high-performance multimeter/data acquisition system. It can
measure voltage (DC and AC), current (DC and AC), resistance (2- and 4-wire),
temperature (thermocouple, thermistor, and 4-wire RTD), frequency and period, and test
continuity.
The Model 2750 has five slots that will accommodate Keithley Model 7700 series
switching modules (Table 1-1). Each channel of a switching module that is closed or
scanned is measured by the Model 2750. For scanning, each channel can have its own
unique setup (i.e., function, range, digits, etc.).
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-7
More information on the measurement capabilities of the Model 2750 is provided in
“DMM measurement capabilities,” page 3-2. A connection guide for the Model 7700 is
provided in Appendix B. Specifications for the Model 2750 and 7700 switching module
are provided in Appendix A.
Additional features of Model 2750 include:
•Setup storage — Five instrument setups (three user, *RST defaults and factory
defaults) can be saved and recalled.
•Offset compensated ohms — A two-measurement process for 4-wire ohms to
cancel the effects of thermal EMFs. Available for the 1Ω, 10Ω, 100Ω, 1kΩ, and
10kΩ, ranges.
•Dry circuit ohms — Low voltage mode (20mV open-circuit clamping) for 4-wire
ohms measurements. Available for the 1Ω, 10Ω, 100Ω, and 1kΩ ranges.
•Math — mX+b, percent, and reciprocal (1/X) calculations provide mathematical
manipulation of readings.
•Relative — Null offsets or establish baseline values.
•Ratio and channel average — Ratio and average calculations for two switching
module channels.
•Buffer — Store up to 110,000 readings in the internal buffer.
•Limits — Two sets of high and low reading limits to test devices.
•Digital I/O port — Five digital limit test output lines to control external circuitry.
The digital trigger link and hardware interlock input can also be accessed at this
port.
•Monitor — The Model 2750 can monitor a selected channel. A scan can be
triggered to start when the Monitor detects a reached reading limit.
•Remote interface — Model 2750 can be controlled using the IEEE-488 interface
(GPIB) or the RS-232 interface.
Plug-in switching modules
Up to five Keithley Model 77XX series switching modules can be installed in the
Model 2750. A side-by-side comparison of the switching modules is provided in
Table 1-1.
Basic close/open operation for switching module channels is provided in Section 2,
while scanning is covered in Section 7. Connection information for the Model 7700
switching module is provided in Appendix B. For all other switching modules,
connection information is provided in the packing list that was shipped with each
switching module.
1-8Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Table 1-1
Model 77xx series switching modules
Model 7700Model 7701Model 7702Model 7703
2-pole Operation20 channels32 channels40 channels32 channels
4-pole Operation10 channel pairs16 channel pairs20 channel pairs16 channel pairs
1-pole OperationN/AN/AN/AN/A
Measure Volts300V maximum150V maximum300V maximum300V maximum
Measure AmpsCh 21 & 22, 3A Max NoCh 41 & 42, 3A Max No
Measure Ohms2/4-wire2/4-wire2/4-wire2/4-wire
Thermocouple
1. Latching relays hold their open/close state after the mainframe is turned off. When turned on, all relays open after a few
seconds.
2. All multiplexers can be configured as two independent multiplexers
3. The Models 7711 and 7712 have no measurement capabilities.
Pseudocards
Using remote programming, you can assign a pseudocard to an empty switching module
slot. With a pseudocard “installed,” the Model 2750 will operate as if the switching
module is installed in the Model 2750. This feature allows you to configure your system
without having the actual switching module installed in the unit. There is a pseudocard for
every Keithley Model 77xx series switching module. For details, see “Pseudocards,”
page 2-5.
1-10Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Identifying installed switching modules
On power-up, the model numbers of installed switching modules are displayed briefly.
While in the normal display state, slot indicators on the right side of the display indicate
which slots have a switching module or pseudocard installed.
NOTEIf a Model 7700, 7701, 7702, 7703, 7705, 7708, 7709, 7710, 7711, or 7712
switching module is removed while the Model 2750 is on, the slot indicator for
that slot will remain on and the instrument will operate as if the module is
installed. That is, the Model 2750 will operate as if the pseudocard is installed.
If a Model 7706 or 7707 is removed while power is on, error +523 “Card hardware error” will occur, and the module will be removed from the system.
In general, it is not recommended to install or remove switching modules with
the power on.
The CARD menu and remote query commands can be used to identify modules installed
in the mainframe. For details, see “Switching module installation and connections,”
page 2-3.
Front and rear panel familiarization
Front panel summary
The front panel of Model 2750 is shown in Figure 1-1.
Figure 1-1
Model 2750 front panel
5
SLOT 1
TEMP
RANGE
AUTO
RANGE
SLOT 2
SLOT 3
SLOT 4
SLOT 5
MODEL 2750 MULTIMETER/SWITCH SYSTEM
INTEGRA SERIES
3
4
CONT
CH AVG
DCIACV
LIMITS ON/OFFDELAY
RECALL
CONFIG HALT
STEP SCAN
ACIΩ2Ω4
TYPE
TEST
DIGITS RATE
OCOMP
MONITOR
LSYNC
RELFILTER
PERIOD SENSOR
FREQ
CH-OFF CARD
RS-232
GPIB
EXIT ENTER
RATIO
MATH
OUTPUT
SHIFT
DCV
DRYCKT
LOCAL
EX TRIG
TRIG
1
POWER
SAVE SETUP
OPEN
STORE
CLOSE
2
350V
PEAK
SENSE
Ω 4 WIRE
INPUT
F
FRONT/REAR
CAT I
INPUT
HI
1000V
!
PEAK
LO
500V
PEAK
R
AMPS
3A, 250V
7
6
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-11
NOTEMost keys provide a dual function or operation. The nomenclature on a key
indicates its unshifted function/operation, which is selected by pressing the key.
Nomenclature (in blue) above a key indicates its shifted function. A shifted
function is selected by pressing the SHIFT key and then the function/operation
key.
1 Special keys and power switch:
SHIFTUse to select a shifted function or operation.
LOCALCancels GPIB remote mode.
POWERPower switch. In position turns 2750 on (I), out position turns it off (O).
2 Function and operation keys:
Top Row
Unshifted
DCVSelects DC voltage measurement function.
ACVSelects AC voltage measurement function.
DCISelects DC current measurement function.
ACISelects AC current measurement function.
FREQSelects frequency measurement function.
TEMPSelects temperature measurement function.
Shifted
MATHConfigures and controls mX+b, percent, or reciprocal (1/X) calculation.
OUTPUTConfigures and controls digital and audio (beeper) output for limits.
RATIOEnables/disables channel ratio.
CH-AVGEnables/disables channel average.
CONTConfigures and controls continuity test.
OCOMPEnables/disables offset compensated ohms with Ω4 function selected.
PERIODSelects period measurement function.
SENSORConfigures temperature measurements.
Middle Row
Unshifted
EXTRIGSelects external triggering (front panel, bus, trigger link) as the trigger source.
TRIGTriggers a measurement when in external triggering (EX TRIG).
STORESets the number of readings to store and enables the buffer.
RECALLDisplays stored readings and buffer statistics. Use the, , , and keys to navigate
through buffer.
FILTEREnables/disables filter for selected function.
RELEnables/disables relative for selected function.
and Dual function — Manually scans switching channels. When in a menu, these keys
control cursor position for making selections or change values.
1-12Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Shifted
DELAYSets user delay between trigger and measurement.
DRYCKTSelects dry circuit ohms (Ω4 must first be selected).
LIMITSSets upper and lower limits for readings.
ON/OFFEnables/disables limits.
TYPEConfigures and enables filter for selected function.
MONITORSelects and enable/disables monitor channel.
CH-OFFDisables channel for a scan (must be in scan channel setup mode).
CARDIdentifies switching modules installed in mainframe. Set up switching modules that
require configuration. View closed channels and channel settings for switching
modules that require configuration.
Bottom Row
Unshifted
OPENOpens closed channel.
CLOSECloses specified channel.
STEPSteps through channels; sends a trigger after each channel.
SCANScans through channels; sends a trigger after last channel.
DIGITSSets display resolution for all functions.
RATESets measurement speed (fast, medium, or slow) for all functions.
EXITCancels selection, moves back to measurement display.
ENTERAccepts selection, moves to next choice or back to measurement display.
Shifted
SAVESaves up to three instrument setups for future recall, and selects power-on setup.
SETUPRestores a default setup (factory or *RST) or a saved setup. Enables/disables buffer
auto clear, auto scan, and auto channel configuration. Sets timestamp, date, and
time. Displays serial number of Model 2750.
CONFIGSelects and configures a simple scan or an advanced scan.
HALTDisables step/scan.
TESTSelects the calibration menu, display test or the key-press test.
LSYNCEnables/disables line cycle synchronization. When enabled, noise induced by the
power line is reduced at the expense of speed.
GPIBEnables/disables GPIB and selects address.
RS-232Enables/disables RS-232 interface; selects baud rate, flow control, and terminator.
3 Range keys:
and Dual function — Selects the next higher/lower measurement range for the selected
function. When in a menu, these keys make selections or change values.
AUTOEnables/disables autorange for the selected function.
4 Display annunciators:
* (asterisk)Readings being stored in buffer.
↔ (more)Indicates additional selections are available.
))) (speaker) Beeper on for continuity or limits testing.
Digital input/output or analog output active (set to non-default value).
4W4-wire resistance or 4-wire RTD temperature reading displayed.
DCKTDry circuit resistance reading displayed.
~ACAC function selected (ACV, dB, or ACI).
AUTOAuto range enabled.
BUFFERRecalling readings stored in buffer.
CHANSetup or a reading for a switching channel displayed.
DELTAChannel average enabled.
ERRQuestionable reading, or invalid cal step.
FASTFast reading rate selected.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-13
FILTFilter enabled for selected function.
HIGHReading has reached or exceeded the enabled high limit.
LSTNInstrument addressed to listen over GPIB.
LOWReading has reached or exceeded the enabled low limit.
MATHmX+b, percent, or reciprocal (1/X) calculation enabled.
MEDMedium reading rate selected.
MON Monitor channel displayed.
OCOMP4-wire offset compensated ohms enabled.
RATIOChannel ratio enabled.
REAR Front panel input terminals disconnected.
RELRelative enabled for selected function.
REMInstrument in GPIB remote mode.
SCANScanning operation being performed.
SHIFTAccessing a shifted key.
SLOWSlow reading rate selected.
SRQService request over GPIB.
STATDisplaying buffer statistics.
STEPStepping operation being performed.
TALKInstrument addressed to talk over GPIB bus.
TIMERTimer controlled triggering in use.
TRIGExternal triggering selected (trigger link, TRIG key, or GPIB).
5 Slot indicators:
Lit lamp indicates that the slot has a switching module or pseudocard installed in it. When the VIEW
option of the CARD menu is used, only the lamp that corresponds to the selected slot is turned on.
6 INPUTS switch:
Use to select front panel inputs (out; F) position, or switching module inputs (in; R) position.
NOTEFor remote programming, the following command queries the INPUTS switch
INPUT HI and LOUsed for DCV, ACV, Ω2, CONT, FREQ, PERIOD, and thermocouple/thermistor
TEMP measurements.
SENSE HI and LOUse with INPUT HI and LO for Ω4 and RTD TEMP measurements.
AMPSUse with INPUT LO for DCI and ACI measurements.
Amps fuse holderHolds current fuse for front panel amps input.
1-14Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Rear panel summary
The rear panel of Model 2750 is shown in Figure 1-2. As shown, slot covers are installed
on unused slots.
WARNINGSlot covers must be installed on unused slots to prevent personal con-
tact with high voltage circuits.
Figure 1-2
Model 2750 rear panel
4321
DIGITAL I/O TRIG. LINK
!
RS-232
IEEE-488
SLT
1
MADE IN
U.S.A.
SLT
2
CAT I
KEITHLEY
SLOT COVER
KEITHLEY
SLOT COVER
KEITHLEY
SLOT COVER
SLT
3
SLT
4
!
SLT
5
6
1 DIGITAL I/O
Male DB-9 connector for digital input (trigger link in) and digital outputs.
2 TRIG LINK
Eight-pin micro-DIN connector for sending and receiving trigger pulses among connected instruments.
Use a trigger link cable or adapter, such as Models 8501-1, 8501-2, 8502, and 8503.
3 RS-232
Female DB-9 connector for RS-232 operation. Use a straight-through (not null modem) DB-9 shielded
cable.
4 IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as Models 7007-1 and 7007-2.
5 Power module
Contains the AC line receptacle, power line fuse, and line voltage setting. The instrument can be
configured for line voltages of 100V/120V/220V/240VAC at line frequencies of 50 or 60Hz.
5
WARNINGSlot covers must be installed on unused slots to prevent personal
contact with high voltage circuits.
6 Slots 1 through 5
Five slots to accommodate Keithley Model 77XX series switching modules. The Model 2750 is shipped
from the factory with slot covers installed. Please note additional slot covers can be requested from
Keithley Instruments.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-15
Rack mounting
The Model 2750 can be mounted in a standard 19-inch rack. For a mainframe that has one
or no switching modules installed, the Model 2750 can be secured to the front rails of the
rack.
The hardware necessary to secure the mainframe to the front rails of the rack is supplied
with the Model 2750. The supplied hardware kit includes four dress screws and four
retaining clips. Each retaining clip has a captive nut.
Perform the following steps to mount the Model 2750 to front rails of the rack:
WARNINGMake sure the Model 2750 is turned off, the line cord is disconnected,
and it is not connected to any external circuitry.
1.Select a location in the rack. The mainframe takes up 3½-inches of vertical space.
2.Using Figure 1-3 as a guide, install two retaining clips on the left front rail. Slide
each retaining clip over a mounting hole such that the captive nut is positioned on
the inside of the rack cabinet. In a similar manner, install two retaining clips on the
right front rail.
3.Remove the four foot assemblies from the bottom of the Model 2750. The retaining
screw for an assembly is located under the rubber foot. Simply pull off the rubber
feet to gain access to the screws. Retain these foot assemblies for future use.
4.Position the Model 2750 in the rack and loosely attach the front panel to the rack
rails using the four supplied dress screws.
5.Tighten the four dress screws.
CAUTIONFor a Model 2750 that has two or more installed switching modules,
rear brackets are required to support the additional weight. The
Model 4288-7 is a rack mount kit for the Model 2750 that uses rear
support brackets.
Figure 1-3
Rack preparation
Left Front
Rack Rail
1.75”
Retaining Clips
with Nuts
1-16Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Power-up
Line power connection
Follow the procedure below to connect the Model 2750 to line power and turn on the
instrument.
1.Check to see that the line voltage indicated in the window of the fuse holder
assembly (Figure 1-4) is correct for the operating voltage in your area. If not, refer
to the next procedure, “Setting line voltage and replacing fuse,” page 1-17.
CAUTIONOperating the instrument on an incorrect line voltage may cause
damage to the instrument, possibly voiding the warranty.
2.Before plugging in the power cord, make sure that the front panel power switch is
in the off (O) position.
3.Connect the female end of the supplied power cord to the AC receptacle on the rear
panel. Connect the other end of the power cord to a grounded AC outlet.
WARNINGThe power cord supplied with the Model 2750 contains a separate
ground wire for use with grounded outlets. When proper connections
are made, instrument chassis is connected to power line ground
through the ground wire in the power cord. Failure to use a grounded
outlet may result in personal injury or death due to electric shock.
4.Turn on the instrument by pressing the front panel power switch to the on (I)
position.
Line frequency
The Model 2750 will operate at line frequencies from 45Hz to 66Hz, and 360Hz to 440Hz.
There are no user-settings for line frequency. It is automatically sensed at power-up. The
following command can be used to read the line frequency:
SYSTem:LFRequency?' Query power line frequency.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-17
Figure 1-4
Power module
Model 2750
DIGITAL I/O TRIG. LINK
!
RS-232
IEEE-488
SLT
1
MADE IN
U.S.A.
SLT
2
CAT I
KEITHLEY
SLOT COVER
KEITHLEY
SLOT COVER
KEITHLEY
SLOT COVER
Fuse
Line Voltage
Selector
220
240
120
100
Spring
Window
Fuse Holder Assembly
SLT
3
SLT
4
!
SLT
5
Setting line voltage and replacing fuse
A rear panel fuse located next to the AC receptacle protects the power line input of the
instrument. If the line voltage setting needs to be changed or the line fuse needs to be
replaced, perform the following steps.
WARNINGMake sure the instrument is disconnected from the AC line and other
equipment before changing the line voltage setting or replacing the line
fuse.
1.Place the tip of a flat-blade screwdriver into the power module by the fuse holder
assembly (Figure 1-4). Gently push in and up. Release pressure on the assembly
and its internal spring will push it out of the power module.
1-18Getting StartedModel 2750 Multimeter/Switch System User’s Manual
2.Remove the fuse and replace it with the type listed in Table 1-2.
CAUTIONFor continued protection against fire or instrument damage, only
replace fuse with the type and rating listed. If the instrument
repeatedly blows fuses, locate and correct the cause of the trouble
before replacing the fuse.
3.If configuring the instrument for a different line voltage, remove the line voltage
selector from the assembly and rotate it to the proper position. When the selector is
installed into the fuse holder assembly, the correct line voltage appears sideways in
the window.
4.Install the fuse holder assembly into the power module by pushing it in until it
locks in place.
Table 1-2
Fuse ratings
Line voltageFuse ratingKeithley P/N
100/120V
220/240V
Power-up sequence
On power-up, the Model 2750 performs self-tests on its EPROM and RAM and
momentarily lights all segments and annunciators. If a failure is detected, the instrument
momentarily displays an error message and the ERR annunciator turns on. (Error
messages are listed in Appendix C).
NOTEIf a problem develops while the instrument is under warranty, return it to
Keithley Instruments, Inc., for repair.
If the instrument passes the self-tests, the firmware revision levels are displayed. An
example of this display is:
REV: A01 A01
where: First A01 is the main board ROM revision.
Second A01 is the display board ROM revision.
Installed switching modules are then displayed. For example, if there is a Model 7700
switching module installed in all five slots, the following messages will be displayed:
1: 77002: 7700
3: 77004: 7700
5: 7700
0.630A, 250V, slow-blow 5×20mm
0.315A, 250V, slow-blow 5×20mm
FU-106-.630
FU-106-.315
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-19
If a slot is empty, the message “NONE” will be displayed instead.
If the saved power-on setup is not the factory defaults setup (SYSTem:POSetup PRESet),
a message to identify the setup will be briefly displayed (“Defaults and user setups,”
page 1-21).
After the power-up sequence, the instrument begins its normal display of readings.
NOTEThe serial number of the Model 2750 can be displayed by selecting the SNUM
item of the SETUP menu. Press SHIFT and then SETUP to access the menu. For
remote operation, the serial number can be read using the *IDN? command (see
Section 12 for details).
Keyclick
With keyclick enabled, an audible “click” will sound when a front panel key is pressed.
Perform the following steps to disable or enable keyclick:
1.Press SHIFT and then LOCAL to display the present state of KEYCLICK (ON or
OFF).
2.Press
or to display the desired keyclick state and press ENTER.
Display
Status and error messages
Remote programming
The following command controls keyclick:
SYSTem:KCLick <b>' Enable or disable keyclick.
where: <b> = ON or OFF
NOTEKeyclick ON is the FACTORY, *RST, and SYSTem:PRESet default.
Readings are displayed in engineering units (i.e., 100.23mV), while annunciators indicate
various states of operation. See “Front panel summary,” page 1-10, for a complete listing
of display annunciators.
NOTEThe display test allows you to test display digit segments, annunciators and the
red slot indicator LEDs. The key test checks the functionality of front panel keys.
These tests are accessed by pressing SHIFT and then TEST. Refer to the
Model 2750 Service Manual for details.
Status and error messages are displayed momentarily. During operation and programming,
you will encounter a number of front panel messages. Typical messages are either of status
or error variety, as listed in Appendix C.
1-20Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Remote programming — display
Using remote programming, the Model 2750 can display a custom ASCII message (up to
12 characters). Also, the front panel display and controls can be disabled.
Display commands
The commands are listed in Table 1-3. Details on these commands follow the table.
NOTEOptional command words and queries are not included in Table 1-3. Table 15-2
provides an unabridged list of all display commands.
Table 1-3
Display commands
CommandDescriptionDefault*
DISPlay:TEXT:DATA <a>Define message (<a> = ASCII characters, up to 12).(none)
DISPlay:TEXT:STATe <b>Enable or disable message mode (<b> = ON or OFF).OFF
DISPlay:ENABle <b>Enable or disable the front panel display (<b> = ON or OFF). ON
*SYSTem:PRESet and *RST have no effect on DISPlay commands. The listed defaults are power-on defaults.
DISPlay:TEXT:DATA <a>Define text message
This command defines the text message for display. A message can be as long as 12
characters. A space counts as a character. Excess message characters results in an error.
The characters must be enclosed in either single quotes (‘ ’) or double quotes (“ ”).
DISPlay:TEXT:STATe ON | OFFControl (on/off) message for display
This command enables and disables the text message mode. When enabled, a defined
message is displayed. When disabled, the message is removed from the display.
A user defined text message remains displayed only as long as the instrument is in remote.
Taking the instrument out of remote (by pressing the LOCAL key or sending GTL)
cancels the message and disables the text message mode.
DISPlay:ENABle ON | OFFControl display circuitry
This command is used to enable and disable the front panel display circuitry. When
disabled, the instrument operates at a higher speed. While disabled, the display is blanked.
All front panel controls (except LOCAL) are disabled. Normal display operation can be
resumed by using the ENABle command to enable the display or by putting the
Model 2750 into local mode (press LOCAL).
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-21
Programming example
The following command sequence displays the text message “TESTING”:
DISP:TEXT:DATA 'TESTING'' Define text message.
DISP:TEXT:STAT ON' Enable text message mode.
Defaults and user setups
Model 2750 can be restored to one of two default setup configurations (FACTory or
*RST), or three user-saved (SAV0, SAV1, or SAV2). As shipped from the factory,
Model 2750 powers up to the factory (FACT) default settings.
NOTEClosed channels can be saved in a user setup (SAV0, SAV1, or SAV2). When the
setup is restored, those channels (and only those channels) will be closed. FACT
and *RST defaults opens all channels.
The factory default setup provides continuous triggering, while the *RST default setup
places the Model 2750 in the one-shot trigger mode. With one-shot triggering, a measurement is performed whenever the TRIG key is pressed or an initiate command is sent over
the remote interface.
The factory and *RST default settings are listed in Table 1-4. Setting differences (Set Diff)
between the two default setups are indicated by checkmarks (✓).
For remote programming, the SYSTem:PRESet and *RST commands are used to reset the
instrument. The *RST command returns the instrument to the *RST defaults and, for the
most part, the SYSTem:PRESet command returns the instrument to the factory default
conditions. The exceptions are explained as follows:
•Auto scan and auto channel configuration — FACTory defaults disable auto scan
and auto channel configuration, while SYSTem:PRESet has no effect. The *RST
defaults (front panel and remote operation) have no effect.
•Memory buffer auto clear — FACTory defaults enable buffer auto clear, while
SYSTem:PRESet has no effect. The *RST defaults (front panel and remote operation) have no effect.
The instrument will power up to whichever default setup is saved as the power-on setup.
NOTEAt the factory, the factory default setup is saved as the SAV0, SAV1, or SAV2
setups.
1-22Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Saving and restoring setups
Saving a user setup
1.Configure Model 2750 for the desired measurement application.
2.Press SHIFT and then SAVE to access the save setup menu.
3.Press
4.Use the
Saving a power-on setup
1.Configure Model 2750 for the desired measurement application.
2.Press SHIFT and then SAVE to access the save setup menu.
3.Press the
4.Press
5.Use the
to place the cursor on the present setup (SAV0, SAV1, or SAV2).
or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
key to display the present power-on (PWR-ON) setup: FACT, *RST,
SAV0, SAV1, or SAV2.
to place the cursor on the present power-on setup.
or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Restoring a setup
1.Press SHIFT and then SETUP to access the restore setup menu.
2.Press
3.Use the
NOTEIf the settings for a user setup or power-on setup do not match the switching
to place the cursor on the present RESTORE setup (FACT, *RST, SAV0,
SAV1, or SAV2).
or key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
module types presently installed in the Model 2750, error +520 (Saved setup
scancard mismatch) occurs when the setup is recalled. The scan list will reset to
the factory defaults and all channels will open. However, the saved setup is still
retained in memory and can be restored when the matching switching module is
later installed.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-23
Table 1-4
Default settings
SettingFactory*RSTSet Diff
Auto channel configuration No (off)No effect✓
AutozeroOnOn
BufferNo effectNo effect
Auto clearYes (on) No effect✓
Channel AverageOffOff
Closed channels NoneNone
Closure count intervalNo effectNo effect
Continuity
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-27
Programming example
*SAV 2' Save present setup in memory location 2.
SYST:POS SAV2' Specify SAV2 setup as the power-on setup.
*RST' Return 2750 to RST defaults.
*RCL 2 ' Return 2750 to setup stored in memory location 2.
Remote programming information
Remote programming information is integrated with front panel operation throughout this
manual. Programming commands are listed in tables, and additional information that
pertains exclusively to remote operation is provided after each table. The tables may
reference you to other sections of this manual.
NOTEExcept for Section 11 through Section 15, most programming tables in this
manual are abridged. That is, they exclude most optional command words and
query commands. Optional command words and query commands are
summarized as
follows.
Optional command words — In order to be in conformance with the IEEE-488.2
standard, Model 2750 accepts optional command words. Any command word that is
enclosed in brackets ([]) is optional and does not have to be included in the program
message.
Query commands — Most command words have a query form. A query command is
identified by the question mark (?) that follows the command word. A query command
requests (queries) the programmed status of that command. When a query command is
sent and Model 2750 is addressed to talk, the response message is sent to the computer.
NOTEFor complete details, see “Programming syntax,” page 10-10.
1-28Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Quick start exercises
This section topic summarizes the following basic instrument operations and provides
simple exercises to perform them:
•Basic DMM measurements — front panel inputs
•Closing and opening channels — system channel operation
•Simple scanning
•Trigger and return readings — remote programming
WARNINGFor the exercises, it is not necessary to connect an input signal or DUT
to the instrument (front panel inputs or switching module inputs).
However, if you decide to use an input signal, it is recommended that
you keep it at a nonhazardous level (<42V) while learning to use the
instrument.
NOTEWhen using the front panel input terminals, the INPUT switch must be in the
“F” (out) position. The INPUT switch is located on the right side of the front
panel near the input terminals. When using a switching module, the switch must
be in the “R” (in) position.
Basic DMM measurements — front panel inputs
NOTESee Section 3 for details on basic DMM operation.
The Model 2750 is shipped from the factory to power-up to factory defaults. The
instrument powers up to a setup that continuously measures DC volts. Some of the default
settings for the DCV function include auto range enabled, 6∫-digit resolution, filter
enabled, and slow reading rate. These settings provide a good starting point and in many
cases, do not need to be changed.
“Starting-point” default settings are also provided for the other measurement functions.
Therefore, to perform basic measurements, simply select the desired function, and
“tweak” the setup (range, rate, filter, digits, etc.) as required.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-29
For remote programming, the instrument is typically used in a non-continuous
measurement mode. In this mode, the user (via remote command programming) specifies
the number of measurements to perform. *RST defaults place the instrument in a noncontinuous measurement mode. Most of the other settings for factory and *RST defaults
are the same.
For remote programming, the following command is used to select function.
NOTEItems in brackets ([]) are optional and do not need to be included. Upper case
characters are required. Lower case characters are optional and need not be
included.
Each function can have its own unique setup configuration (i.e., range, digits, speed, etc.).
For example, the following command words select range and digits:
RANGe[:UPPer] <n>' Specify expected reading.
RANGe:AUTO <b>' Enable (ON) or disable (OFF) auto range.
DIGits' Set display resolution; 3.5, 4.5, 5.5 or 6.5 (digits).
The following examples demonstrate how to include the function name in the command
string for configuration commands.
VOLT:RANG 10' Select 10V range for DCV.
RES:RANG:AUTO ON' Enable auto range for W2.
CURR:DIG 4.5' Set DCI for 4½- digit resolution.
NOTESee Section 4 for details on setting range, digits, rate, bandwidth, and filter.
1-30Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Exercise 1 — Basic DMM measurements
The exercise in Table 1-6 measures ACV on the 10V range and stores 15 readings in the
buffer.
Table 1-6
Exercise 1 — Measure AC volts - store readings in buffer
Front panel operationCommand sequence
1For front panel operation, proceed to step 2.
For remote programming, clear the buffer
1
:
TRAC:CLE
2Restore defaults2:
Press SHIFT > press SETUP > select RESTORE: FACT.
*RST
3Select ACV function:
Press ACV.
FUNC 'VOLT:AC'
4Select 10V range:
Press RANGE to display “RANGE: 10V”.
VOLT:AC:RANG 10
6Store 15 readings in buffer3:
Press STORE > set for 000015 RDGS > press ENTER.
SAMP:COUN 15
READ?
7Recall buffer readings4:
Press RECALL > use edit keys to display readings. Press EXIT to exit
CALC1:DATA?
recall mode.
1. To avoid problems with remote programming, it is good practice to routinely clear the buffer (TRAC:CLE) at the beginning
of a program that performs multiple measurements (
the buffer.
2. FACTory defaults place the instrument in a continuous measurement mode. *RST places the instrument in a noncontinuous measurement mode.
3. READ? triggers and returns 15 readings. These 15 readings are automatically stored in the buffer. See Exercise 4 and 5 for
more information on the READ? command.
4. Statistics for buffer readings are also stored in the buffer. For remote programming,
ings that were stored. It does not return buffer statistics.
(see Section 6 for details).
SAMP:COUN >1). Restoring *RST or FACTory defaults does not clear
CALC1:DATA? only returns the read-
CALC2 commands are used to calculate and return buffer statistics
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-31
Closing and opening channels — system channel operation
NOTESee Section 2 for details on closing and opening switching module channels.
NOTEThe following discussion assumes a multiplexing switching module
(i.e., Model 7700) installed in slot 1 of the mainframe. Switching module
installation is covered in “Switching module installation and connections,”
page 2-3.
An alternative to installing a switching module is to assign slot 1 as a
pseudocard using remote programming. The instrument will operate as if a
switching module is installed in slot 1. To “install” a 7700 pseudocard in slot 1,
send the following command:
SYST:PCAR1 C7700
System channel operation is used to connect input channels to the DMM of the
Model 2750:
•For a 2-wire function (i.e., DCV), closing a system channel connects the input to
DMM Input of the Model 2750.
Figure 1-5 shows system channel 1 closed. For the Ω2 function, the resistance
(DUT) would be connected to DMM Input as shown Figure 1-5.
•For a 4-wire function (i.e., Ω4), a channel pair is connected to the DMM when a
system channel is closed. The system channel is connected to DMM Input and the
paired channel is connected to DMM Sense.
Figure 1-6 shows system channel 6 closed. For a 4-wire function, the paired
channel also closes. For the Model 7700, channels 1 through 10 are paired to
channels 11 through 20. When channel 6 is closed, channel 16 also closes.
Figure 1-6 shows how the DUT is connected to the DMM for the 4-wire function.
NOTEFigure 1-5 and Figure 1-6 show simplified schematics of the switching module.
They show a single switch closed to connect an input channel to the DMM. In
reality, multiple switching to is used to make proper connections to the DMM.
However, for system channel operation, the user need not be concerned about
which switches in the module close.
Figure 1-5
Connection to DMM for 2-wire function (system channel 101 closed)
Switching ModuleSwitching Module
HI
Ch 1
LO
DMM
HI
Input
LO
DUT
HI
Ch 1
LO
DMM
HI
Input
LO
1-32Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Figure 1-6
Connection to DMM for 4-wire function (system channel 106 closed)
Switching Module
HI
Ch 6
LO
HI
Ch 16
LO
7700
HI
Input
LO
HI
Sense
LO
HI
Ch 6
LO
DUT
HI
Ch 16
LO
7700
Switching Module
HI
Input
LO
DMM DMM
HI
Sense
LO
NOTESwitching module channels can also be controlled using multiple channel
operation. This allows individual control of all module channels (switches).
Multiple channel operation should only be used by experienced service
personnel who recognize the dangers associated with multiple channel closures.
See Section 2 for details.
Close/open operation
The following points on operation pertain to system channel operation only:
•Only one input channel (or channel pair) is closed at one time. When you close an
input channel, the previously closed input channel(s) will open.
•When a system channel is closed, the channel number will be displayed on the
Model 2750. The slot number for the module is also displayed. For example, “103”
indicates that system input channel 3 for a module in slot 1 is closed.
•The paired channel for a 4-wire function is not displayed. Only the system channel
number is displayed. For example, in Figure 1-6, channel number 106 will be
displayed with the Model 7707 installed in slot 1 of the mainframe.
•Switching modules that have current measurement capability have separate
channels reserved exclusively for the DCI and ACI functions. For example, the
Model 7700 has channels 21 and 22 reserved for amps measurements. With the
DCI or ACI function selected, only channels 21 and 22 can be closed. These
channels cannot be accessed on any other function.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-33
Figure 1-7 shows the front panel keys used to close and open system channels.
Figure 1-7
Front panel keys to close and open system channels
Close next
measurement
channel
CLOSEOPEN
Press CLOSE key
CLOSEOPEN
Press OPEN key
CLOSE:SINGLE
Close previous
measurement
channel
A. Sequencing through channnels B. Specifiying channel to closeC. Opening all channels
CLOSE CH: XXX
Display SINGLE option
and press ENTER
Specify channel
number (XXX) and
press ENTER
OPEN: ALL
Display ALL
option and press
OPEN again
For remote programming, the following three commands are used for basic system
operation to open and close input channels:
ROUTe:CLOSe <clist>' Close specified system channel1.
ROUTe:CLOSe?' Query closed system channel2.
ROUT:OPEN ALL' Open all channels.
1.Only one channel can be specified in the <clist>. For example, to close input
channel 3 for a module in slot 1, the following command would be sent:
ROUTe:CLOSe (@103)
2.Only the closed system channel is returned by ROUTe:CLOSe?. The paired channel
for a 4-wire function is not returned. For example, assume channel 2 in slot 1 is
closed. The following response message will be returned:
(@102)
1-34Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Exercise 2 — Closing and opening channels (system channel operation)
The exercise in Table 1-7 demonstrates a sequence to close and open channels of a
Model 7700 installed in slot 1 of the mainframe.
Table 1-7
Exercise 2 — Close and open channels (system channel operation)
Front panel operationCommand sequence
1Open all channels*:
Press OPEN > display OPEN:ALL > Press OPEN.
ROUT:OPEN ALL
2Select Ω2 function:
Press Ω2.
FUNC 'RES'
3Close system channel 101:
Press the
key. Channel 1 connects to DMM Input (see Figure 1-5).
ROUT:CLOS (@101)
4Close system channel 102:
Press the
key. Channel 2 connects to DMM Input.
ROUT:CLOS (@102)
5Close system channel 106:
Press CLOSE > select CLOSE:SINGLE > key in channel 106 > press
ROUT:CLOS (@106)
ENTER. Channel 6 connects to DMM Input.
6Select Ω4 function:
Press Ω4. 4W annunciator turns on, and channels 6 and 16 connects to
FUNC 'FRES'
DMM Input and Sense (see Figure 1-6).
7Open all channels*:
Press OPEN > display OPEN:ALL > Press OPEN ROUT:OPEN ALL
*It is a good, safe practice to start and end a switching sequence by opening all channels.
Simple scanning
NOTESee Section 7 for details on scanning.
With at least one multiplexer switching module (i.e., Model 7700) installed in the
mainframe, the instrument can scan channels that are valid for the selected function.
For front panel operation, Figure 1-8 shows the three basic steps to configure and run a
simple scan. The differences between the STEP function and the SCAN function involve
the reading count and the timer.
Reading count (RDG CT) — For both STEP and SCAN, the reading count specifies the
number of readings to store in the buffer. For STEP, the reading count determines the
number of channels to scan.
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-35
For SCAN, the reading count also determines the number of scans to perform and is best
explained by an example. Assume there are 10 channels in the scan list (i.e., 101 through
110). If you set the reading count to 10 or less, one scan of the 10 channels will be
performed. If you set the reading count to any value from 11 to 20, two scans will be
performed. A reading count from 21 to 30 gives you three scans, and so on.
Timer interval (TIMER) — For the STEP function, the timer specifies the time delay
between scanned channels. For the SCAN function, the interval specifies the time delay
between scans. The timer starts when the scan is started. For SCAN, the next scan will not
start until the timer interval expires.
NOTEThe Model 2750 can also be configured to run an advanced scan. For an
advanced scan, each channel can have its own unique setup (i.e., function,
range, etc.). Advanced scanning is covered in Section 7.
Figure 1-8
Simple scan operation
Step 1. Configure simple scan:Step 2. Run simple scan:
Press SHIFT
SHIFT
Step 3. Disable scan mode:
Press SHIFT
SHIFT
CONFIG
SCANSTEP
INT: SIMPLE
MIN CHAN: XXX
MAX CHAN: YYY
TIMER? NO/YES
YESNO
xxH:xxM:xx.xxxS
RDG CT:xxxxxx
HALT
SCANSTEP
Press CONFIG (STEP)
Display SIMPLE option
and press ENTER
Specify minimum channel
(XXX) and press ENTER
Specify maximum channel
(YYY) and press ENTER
Display NO or YES and
press ENTER
Set timer interval in
hr:min:sec format and
press ENTER
Specify reading count and press
ENTER
Press HALT (SCAN)
SCANSTEP
Press STEP or SCAN to start
scan
Timer interval specifies time
between scans.
Reading count:
Specifies number of scans
to be performed.
Specifies number of readings
to store in buffer.
Timer interval specifies time between
scanned channels.
Reading count:
Specifies number of channels to be
scanned.
Specifies number of readings to store
in buffer.
1-36Getting StartedModel 2750 Multimeter/Switch System User’s Manual
For remote programming, the following commands are used for simple scanning:
ROUTe:SCAN <clist>' Define scan list*.
TRIGger:COUNt <NRf>' Specify number of scans (1 to 11000 or
' INFinity).
SAMPle:COUNt <NRf>' Specify number of channels to scan (1 to 11000).
ROUTe:SCAN:LSELect <name> ' Enable (INT) or disable (NONE) scan.
*Any valid switching module channel can be included in the scan list. Make sure to list
them from the lowest numbered channel to the highest. For example, to scan channels 1
through 8 of a Model 7700 installed in slot 1, send the following command to define the
scan list:
ROUTe:SCAN (@101:108)
Exercise 3 — Simple scanning
The scanning example in Table 1-8 assumes a Model 7700 installed in slot 1 of the
mainframe. The scan will use default settings (DCV) to scan eight channels and store the
readings in the buffer.
Table 1-8
Exercise 3 — Simple scanning
Front panel operation Command sequence
1For front panel operation, proceed to step 2.
For remote programming, clear the buffer:
TRAC:CLE
2Restore defaults1:
Press SHIFT > press SETUP > select RESTORE: FACT.
*RST
3Configure scan:
Press SHIFT > press CONFIG > select INT: SIMPLE > set MIN
CHAN101 > set MAX CHAN: 108 > select TIMER? NO > set RDG
ROUT:SCAN (@101:108)
SAMP:COUN 8
CT:000008.
4Enable and start scan2:
Press STEP.
ROUT:SCAN:LSEL INT
INIT
5Halt (disable) scanner:
Press SHIFT > press HALT.
ROUT:SCAN:LSEL NONE
6Recall the eight stored readings:
Press RECALL > use edit keys to display readings. Press EXIT to
CALC1:DATA?
exit recall mode.
7Open all channels:
Press OPEN > display OPEN:ALL > Press OPEN
1. Factory and *RST defaults opens all channels, select the DCV function and sets TRIG:COUN to 1. The trigger count specifies the number of scans to be performed.
2. ROUT:SCAN:LSEL INT enables the scan, and INIT triggers the start of the scan.
ROUT:OPEN ALL
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-37
Trigger and return readings — remote programming
There are several commands used to trigger and return readings. The proper commands
and sequence to use depend on trigger state (continuous or non-continuous) and what you
are trying to accomplish.
Presented here are three fundamental command sequences that can be used to “trigger and
return readings.” These three command sequences (exercises) will accommodate most
basic measurement scenarios. Simply use the command sequence (exercise) that satisfies
your needs:
•Exercise 4 — Trigger and return a single reading
•Exercise 5 — Trigger and return multiple readings
•Exercise 6 — Return a single reading (continuous triggering)
Details on the commands to trigger and return readings are provided in other sections of
this manual. For details, refer to the following sections:
Section 3 — See “Trigger and retrieve readings” in Table 3-7.
Section 7 — For scanning, see “Trigger commands” in Table 7-1.
Section 8 — Explains the triggering process.
Section 13 — Covers Signal Oriented Measurement Commands (i.e., FETCh?, READ?).
Section 15 — See Table 15-9 (Trigger command summary).
Appendix D — Shows how trigger and read commands control data flow within the
instrument.
NOTEEach exercise indicates the commands used to configure triggering. Once
triggering is configured, the commands to trigger and/or return readings can be
repeated as often as desired (unless noted otherwise).
Exercise 4 — Trigger and return a single reading
Exercise 5 — Trigger and return multiple readings
Trigger controlled measurements — The instrument is typically used in a noncontinuous trigger mode. In this mode, commands are used to trigger one or more
readings. After the specified number of readings are completed, the measurement process
stops.
Exercise 4 in Figure 1-9 provides a command sequence to trigger and return one reading.
Exercise 5 in Figure 1-10 provides a command sequence to trigger and return multiple
readings.
1-38Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Exercise 6 — Return a single reading (continuous triggering)
Readings can be returned while the instrument is in the continuous measurement (trigger)
mode. Each time a read command is sent, the latest reading is returned. Exercise 6 in
Figure 1-11 provides a command sequence to return a single reading while in the
continuous trigger state.
Figure 1-9
Exercise 4 — Trigger and return a single reading
INIT:CONT OFF
TRIG:COUN 1
Trigger Configuration
SAMP:COUN 1
INITTrigger Reading
Trigger and Return
Reading
FETCh?
OR
CALC:DATA?
Return Result of MATH
Calculation
1. If a MATH function (mX+B, percent or 1/X) is enabled, the result of the calculation will
be returned (MATH functions are covered in Section 5).
2. If there is no MATH function enabled, FETCh?and CALC:DATA? will return the basic
reading.
3. FETCh?, CALC:DATA?, and DATA? do not trigger readings. They simply return the last
reading. If you again send one of these commands before triggering a new reading, the
old reading will be returned.
4. DATA:FRESh? can only be used once to return the same reading. Sending it again
1, 2, 3
without first triggering a new reading will cause error -230 (data corrupt or stale).
DATA ?
OR
DATA:FRESh?
Return Basic Reading
3,4
Place 2750 in
non-continuous
trigger state
Set 2750 to perform
one measurement
READ?
1
Trigger and Return
Reading
Model 2750 Multimeter/Switch System User’s ManualGetting Started1-39
Figure 1-10
Exercise 5 — Trigger and return multiple readings
1
Trigger Configuration
TRAC:CLE
INIT:CONT OFF
TRIG:COUN 1
SAMP:COUN x
Clear Buffer
Place 2750 in non-continuous
trigger state
Set 2750 to perform “x”
number of measurements
(x = 2 to 110000)
INIT
FETCh?
OR
READ?
Trigger and Return
Readings
2, 3
Trigger and Return
Readings
TRAC:DATA?
Return Stored Readings
4
1. In order to trigger and return multiple readings, the buffer must first be cleared of
readings that were stored by the TRACe command or front panel operation (see
Section 6 for details on buffer operation).
2. INIT triggers the measurements, and FETCh? returns the readings. Again sending
FETCh? without first sending INIT will return old readings.
3. READ? performs an INIT to trigger the measurements, and then FETCh? to return
the reading(s).
4. Triggered readings are automatically stored in the buffer. Statistics for buffer readings
are also stored in the buffer. CALC2 commands are used to calculate and return
buffer statistics (see Section 6 for details).
1-40Getting StartedModel 2750 Multimeter/Switch System User’s Manual
Figure 1-11
Exercise 6 — Return a single reading (continuous triggering)
Place 2750 in
continuous
trigger state.
DATA ?
OR
DATA:FRESh?
2, 3
Trigger Configuration
Return Readings
SAMP:COUN 1
INIT:CONT ON
FETCh?
OR
CALC:DATA?
Return result of MATH
Calculation
1, 2
Return Basic Reading
1. If a MATH function (mX+B, percent or 1/X) is enabled, the result of the calculation will
be returned. If there is no MATH function enabled, FETCh? and CALC:DATA? will return
the basic reading.
MATH functions are covered in Section 5.
2. None of these read commands trigger measurements. They simply return the lastest
reading. If FETCH?, CALC:DATA?, or DATA? is again sent before a new reading is
triggered, the ld reading will be returned.
3. DATA:FRESh? can only be used once to return the same reading. Sending it again before
a new reading is triggered will cause error -230 (data corrupt or stale).
2
Closing and Opening
Switching Module Channels
•Close/open overview — Summarizes the two operating modes to control switching modules: system channel operation and multiple channel operation.
•Switching module installation and connections — Explains how to install a
switching module (or pseudocard) into the Model 2750 mainframe. Also explains
where to find connection information which should only be performed by qualified
service personnel.
•Channel assignments — Explains the format for specifying the mainframe
channel assignment which is made up of the slot number and switching module
channel number.
•System channel operation — Provides detailed information for using system
channel operation.
•Multiple channel operation — Provides detailed information for using multiple
channel operation. Due to safety considerations, this operating mode should only
be used by experienced test engineers.
•Identifying installed modules and viewing closed channels — Explains how to
use the CARD menu to identify installed switching modules and view closed
channels. Explains how to remotely identify installed modules (*OPT?) and
summarizes other query commands that can be used to acquire information about
the installed modules.
•Relay closure count — Explains how to determine the number of times switching
module relays have been closed.
•Model 7700 switching module — Covers operating characteristics that are unique
to the Model 7700. Also includes a simplified schematic diagram of the switching
module.
2-2Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Close/open overview
NOTEThis section covers basic close/open operations for switching module channels.
It also covers the operating characteristics that are unique to the Model 7700
switching module. Unique operating characteristics for all switching modules
are provided in the Model 77XX Series Switching Modules Instruction Manual.
There are two modes of close/open operation:
•System channel operation — This is the mode of operation that should be used
exclusively by most (if not all) users. When you close an input channel (or channelpair), other channels on the switching module close automatically to internally
connect it the DMM of the Model 2750.
•Multiple channel operation — This mode of operation provides additional
flexibility by providing individual control of each switching module channel.
However, careless operation could create a safety hazard and/or damage the
switching module and other equipment. Multiple channel operation should only be
used by experienced test engineers.
CAUTIONTo prevent damage to a switching module, do not exceed the maximum
signal level input for that module. Most switching modules are rated
for 303V. The following command queries maximum module voltage:
SYSTem:CARDx:VMAX?‘ Request maximum allowable voltage for
‘ CARDx (where x is the slot number for
‘ the module).
For system channel operation, the instrument will display the
“OVERFLOW” message when the maximum allowable voltage for
the module is being exceeded.
However, for multiple channel operation, the “OVERFLOW” message
will not occur until the maximum voltage of the mainframe (not
module) is exceeded. Therefore, the “OVERFLOW” message would
occur only if 1010V is exceeded.
WARNINGCareless multiple channel operation could create an electric shock
hazard that could result in severe injury or death. Improper operation
can also cause damage to the switching modules and external circuitry.
Multiple channel operation should be restricted to experienced test
engineers who recognize the dangers associated with multiple channel
closures.
NOTEThe Model 2750 can scan switching module channels. Each channel in the scan
can have its own unique setup configuration. Scanning is covered in Section 7.
NOTEWhen a setup is saved as a user setup (SAV0, SAV1, or SAV2), closed channels
are also saved. When the setup is restored, those channels (and only those
channels) will be closed (see “Defaults and user setups,” page 1-21).
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-3
Switching module installation and connections
In order to exercise close/open operations explained in this section, a switching module
(or pseudocard) must be installed in the mainframe. A switching module can be installed
by the user, however external connections to the switching module are only to be
performed by qualified service personnel.
NOTEFor inexperienced users, it is recommended that DUT and external circuitry not
be connected to switching modules. This will allow you to exercise close/open
operations without the dangers associated with live test circuits.
WARNINGTo prevent electric shock that could result in injury or death, NEVER
handle a switching module that has power applied to it:
•Before installing (or removing) a switching module, make sure the
Model 2750 is turned off and disconnected from line power.
•If the switching module is already connected to DUT, make sure
power is removed from all external circuitry.
Module installation
WARNINGSlot covers must be installed on unused slots to prevent personal
contact with high voltage circuits.
Perform the following steps to install a switching module into the Model 2750 mainframe:
1.Turn the Model 2750 off and disconnect the power line cord and any other cable
connected to the rear panel.
2.Position the Model 2750 so you are facing the rear panel.
3.Remove the slot cover plate from the desired mainframe slot. Retain the plate and
screws for future use.
4.With the top cover of the switching module facing up, slide the module into an
empty slot. For the last ¼ inch or so, press in firmly to mate the module connector
to the mainframe connector.
5.On each side of the module, there is a mounting screw. Tighten these two screws to
secure the module to the mainframe. Do not overtighten.
6.Reconnect the power line cable and any other cables to the rear panel.
7.When you turn on the Model 2750, the model number of the switching module will
be briefly displayed.
2-4Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Connections
WARNINGConnection information for switching modules is intended for qualified
service personnel. Do not attempt to connect DUT or external circuitry
to a switching module unless qualified to do so.
To prevent electric shock that could result in serious injury or death,
adhere to following safety precautions:
•Before making or breaking connections to the switching module,
make sure the Model 2750 is turned off and power is removed
from all external circuitry.
•Do not connect signals that will exceed the maximum
specifications of switching module. Specifications for the Model
7700 are provided in Appendix A.
If both the front panel terminals and the switching module terminals
are connected at the same time, the test lead insulation must be rated
to the highest voltage that is connected. For example, if 1000V is
connected to the front panel input, the test lead insulation for the
switching module must also be rated for 1000V.
Dangerous arcs of an explosive nature in a high energy circuit can
cause severe personal injury or death. If the multimeter is connected to
a high energy circuit when set to a current range, low resistance range,
or any other low impedance range, the circuit is virtually shorted.
Dangerous arcing can result even when the multimeter is set to a
voltage range if the minimum voltage spacing is reduced in the
external connections. For details to safely make high energy
measurements, see “High energy circuit safety precautions,” page 3-3.
As described in the International Electrotechnical Commission (IEC)
Standard IEC 664, the Model 2750 is Installation Category I and must
not be connected to mains.
For the Model 7700, detailed connection and wiring information is provided in
Appendix B of this manual (Model 7700 Connection Guide).
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-5
Pseudocards
Using remote programming, you can assign a pseudocard to an empty switching module
slot. With a pseudocard installed, the Model 2750 will operate as if the switching module
is installed in the Model 2750. This feature allows you exercise open/close/scan
operations, or configure your system without having the actual switching module installed
in the unit. There is a pseudocard for every Keithley Model 77XX series switching
module.
A pseudocard cannot be installed from the front panel. However, once it is installed, you
can take the Model 2750 out of remote and use the front panel. Pressing the LOCAL key
takes the Model 2750 out of remote.
When a pseudocard is installed, the appropriate front panel slot indicator lamp turns on.
When the instrument is turned off, the pseudocard will be lost (uninstalled). Use the
following commands to install pseudocards:
SYSTem:PCARd1 <name>' Install pseudocard in slot 1.
SYSTem:PCARd2 <name>' Install pseudocard in slot 2.
SYSTem:PCARd3 <name>' Install pseudocard in slot 3.
SYSTem:PCARd4 <name>' Install pseudocard in slot 4.
SYSTem:PCARd5 <name>' Install pseudocard in slot 5.
<name> = C7700, C7701, C7702, C7703, C7705, C7706, C7707, C7708, C7709,
C7710, C7711, or C7712
Programming example — The following command sets up the Model 2750 to operate as
if a Model 7700 switching module is installed in slot 2, which must be empty. You cannot
assign a pseudocard to a slot that already has a switching module installed in it.
SYSTem:PCAR2 C7700' "Install" pseudocard 7700 for slot 2.
Channel assignments
The Model 2750 has five slots for switching modules. To control the appropriate switching
module, the slot number must be included with the switching module channel number
when you specify a channel. The channel assignment is formatted as follows:
NOTEFor remote operation, the 3-digit channel assignment is included in the channel
list parameter for the commands. Format examples for the channel list
parameter are provided in Table 2-1 and Table 2-2.
2-6Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
System channel operation
The system channel is a closed measurement channel that is internally connected to the
internal DMM Input of the Model 2750. The system channel number is displayed on the
Model 2750. For a 4-wire function (i.e., Ω4), the paired channel for the system channel is
internally connected to DMM Sense. The paired channel is not displayed on the
Model 2750. When triggered, the DMM performs a measurement and displays it on the
Model 2750.
The system channel is selected by closing a measurement channel using the system
channel close keys. These include the
option). See “Controlling the system channel,” page 2-9, for details.
Other important points about system channel operation include the following:
•There can only be one system channel. This is the channel that is presently
displayed (and closed) on the Model 2750. When a channel is not displayed, there
is no system channel.
•When a measurement channel is closed, the input backplane isolation channel also
closes to connect the system channel to DMM input. For a 4-wire function, the
paired channel and the sense backplane isolation channel also close to make the
sense connections to the DMM.
•When a different measurement channel is closed, the previous system channel
opens. The newly closed (and displayed) measurement channel becomes the
system channel.
•The system channel close keys can only close measurement channels that will
automatically connect to the DMM. Non-measurement channels cannot be closed
by the system channel close keys.
and keys, or the CLOSE key (SINGLE menu
NOTEUse the VIEW option of the CARD menu to display all closed channels in the
mainframe (see “CARD menu,” page 2-29).
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-7
2-wire functions
Figure 2-1 shows an example of how the system channel is connected to the DMM Input
of the Model 2750. Assume a Model 7700 switching module is installed in slot 1 of the
mainframe. When channel 101 is closed using the system channel close keys, both the
Channel 1 relay and the backplane isolation relay (Channel 25) closes to connect the
channel to the DMM. The complete simplified schematic of the Model 7700 is provided in
Figure 2-12.
Figure 2-1
2-wire system channel connections to Model 2750 DMM
Model 2750
Model 7700 Switching Module
Channel 1
Relay
HI
Channel 1
LO
System Channel Operation:
Close Channel 101
Slot 1
DMM
Channel 25
HI
Input
LO
Backplane
Isolation
Relay
2-8Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
4-wire functions (paired channels)
A 4-wire function, such as Ω4, requires that another measurement channel be paired to the
system channel. For example, if the switching module has 20 measurement channels,
channels 1 through 10 can be used as the system channel, while channels 11 through 20
are used as the paired channel. For a switching module that has 20 measurement channels,
channel 1 is paired to channel 11, channel 2 is paired to channel 12, channel 3 is paired to
channel 13, and so on.
Figure 2-2 shows an example of system channel connections for a 4-wire function.
Assume a Model 7700 switching module is installed in slot 1 of the mainframe, and a
4-wire function, such as Ω4, is selected. When channel 101 is closed using the system
channel close keys, the Channel 1 relay and the input backplane isolation relay
(Channel 25) closes to connect the channel to DMM Input. Also, the Channel 11 relay and
the sense backplane isolation relay (Channel 24) closes to connect the paired channel to
DMM Sense. Also note in Figure 2-2 that the Channel 23 relay closes to isolate channel 1
from channel 11.
The complete simplified schematic of Model 7700 is provided in Figure 2-12.
Figure 2-2
4-wire system channel connections to Model 2750 DMM
Model 7700 Switching Module
Channel 1
Relay
Channel 1
System Channel Operation:
Close Channel 101
Channel 11
HI
LO
Channel
(Closed Position
Shown
Channel 11
Relay
HI
LO
Model 2750
23
)
Channel 25
Backplane
Isolation
Relay
2-Pole/4-Pole
Relay
Channel 24
Backplane
Isolation
Relay
Slot 1
DMM
HI
Input
LO
HI
Sense
LO
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-9
Controlling the system channel
When a measurement channel is closed, a previous system channel (and, for a 4-wire
function, its paired channel) is first opened. The closed measurement channel becomes the
system channel. When a 4-wire function is selected, the paired channel for the system
channel also closes.
and keys
These front panel keys (Figure 2-3) can be used to select the next or previous
measurement channel as the system channel. If there are no measurement channels
available, one of the following messages will be briefly displayed when one of these keys
is pressed:
NO SCAN CARD — This message indicates that there are no switching modules (or
pseudocards) installed. All five slots are empty.
NO MEAS CARD — This message indicates that none of the installed switching
modules (or pseudocards) have measurement channels. For example, the Model 7705
switching module does not have any measurement channels. Those channels cannot be
internally connected to the DMM.
NOTEThe
and keys can also be used to open all channels in the mainframe.
Simply increment or decrement the channel number until there is no channel
displayed.
Figure 2-3
System channel operation — closing next or previous measurement channel
Close previous
measurement
channel
Close next
measurement
channel
2-10Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
CLOSE key (SINGLE menu option)
The SINGLE menu option for the CLOSE key can be used to select a measurement
channel as the system channel (Figure 2-4). Perform the following steps to select the
system channel:
1.Press the CLOSE key. The message “CLOSE:SINGLE” message will be
displayed.
NOTEIf the “CLOSE:MULTI” message is instead displayed when CLOSE is pressed,
it indicates that there are no measurement modules installed in the mainframe.
See “Multiple channel operation,” page 2-16, to close the channels of a non-
measurement module (i.e., Model 7705).
2.Press ENTER to display the prompt to close a channel (CLOSE CH: XXX).
3.Using
4.Press ENTER. The channel closes and the CHAN annunciator turns on.
An invalid channel cannot be closed and will cause one of the following error messages to
be briefly displayed:
INVALID CHAN — This message indicates that the channel is not a valid measurement
channel. The following actions will cause this error:
, , , and , key in the three-digit channel you want to select.
•Trying to close a non-measurement channel, such as a backplane isolation channel,
a channel that sets the pole mode, or a channel that cannot be internally connected
to the DMM.
•Trying to close an amps channel while on a non-amps function. The DCI or ACI
function must be selected in order to close an amps channel.
•Trying to close a paired-channel while on a 4-wire function. For the Model 7700,
channels 1 through 10 are paired to channels 11 through 20 for a 4-wire function.
If, for example, you try to close channel 12 while on the Ω4 function, the
INVALID CHAN error will occur.
•Trying to close a switching module channel that does not exist.
Figure 2-4
System channel operation — specifying measurement channel to close
CLOSEOPEN
CLOSE:SINGLE
CLOSE CH: XXX
Press CLOSE key
Display SINGLE option
and press ENTER
Specify channel number (XXX)
and press ENTER
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-11
TOO SMALL or TOO LARGE — These messages also indicate an invalid channel. TOO
SMALL indicates that the specified channel and any other lower numbered channel is
invalid. TOO LARGE indicates that the specified channel and any other higher numbered
channel is invalid.
OPEN key (ALL menu option)
The ALL menu option of the OPEN key opens all channels for all switching modules
installed in the Model 2750 (Figure 2-5). For example, if a Model 7700 switching module
is installed in slot 1, OPEN: ALL will open all measurement channels (101 to 120, 121,
and 122), the backplane isolation channels (124 and 125) and the 2-pole/4-pole channel
(123). Figure 2-2 shows the backplane isolation channels and the 2-pole/4-pole channel
for the Model 7700.
Perform the following steps to open all channels:
1.Press the OPEN key to display “OPEN: ALL.”
2.Press OPEN a second time (or press ENTER) to open all channels.
NOTEOpening the system channel disables Ratio or Channel Average. Ratio and
Channel Average operation are covered in Section 5.
Figure 2-5
System channel operation — opening all channels in mainframe
CLOSEOPEN
OPEN: ALL
Press OPEN key
Display ALL option
and press OPEN again
2-12Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Remote programming — system channel control commands
The commands to close and open the system channel are listed in Table 2-1. When a
system channel reading is returned, the system channel number will be included in the
data string if the CHANnel data element is selected. The FORMat:ELEMents command is
used to specify the data elements to be included in the data string (see FORMat commands
in Section 14).
Table 2-1
System channel control commands
CommandsDescriptionRef
ROUTe:CLOSe <clist>Specify one measurement channel to close. a
ROUTe:CLOSe:STATe? <clist> Query closed channels in specified list
(1 = closed).
ROUTe:CLOSe?Returns a <clist> of closed measurement
channels.
ROUTe:OPEN:ALLOpen all channels, and disable ratio and channel
average.
Channel list parameter:
<clist> = (@SCH)
where: S = Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
b
c
d
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-13
Reference:
a.ROUTe:CLOSe <clist>
This command functions the same as the front panel CLOSE key (SINGLE menu
option) to select the system channel. Only one measurement channel can be
specified in the <clist>.
Trying to close an invalid channel (such as a non-measurement channel) with this
command will result in error -222 (Parameter data out of range).
b.ROUTe:CLOSe:STATe? <clist>
This query returns a “0” (open) or “1” (closed) for every measurement channel
specified the <clist>. For example, assume <clist> = (@101, 104, 107, 102). The
response message “0, 0, 1, 0” indicates that channel 107 is closed.
The state of non-measurement channels cannot be checked with this command.
c.ROUTe:CLOSe?
This query command returns a <clist> of closed measurement channels, including
paired channels for 4-wire functions.
This query command will not return non-measurement channels, such as
backplane isolation channels and the pole-mode channel.
d.ROUTe:OPEN:ALL
This command functions the same as the front panel OPEN key (ALL menu
option). It simply opens all channels (including non-measurement channels)
installed in the mainframe.
Remote programming example (system channel operation)
The following example assumes a Model 7700 installed in slot 1, and the Ω4 function of
the Model 2750 is selected. This command sequence connects channel 101 and its paired
channel (111) to DMM Input and Sense as shown in Figure 2-2.
ROUT:OPEN:ALL' Open all channels.
ROUT:CLOS (@101)' Close channels 101, 111, 123,
' 124 and 125.
2-14Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Non-amp and non-measure switching modules
There are Keithley switching modules that do not support current measurements and there
are modules that do not support any measurements at all.
Non-amps module — With an amps function selected (DCI or ACI), system channel
operation cannot be used to close channels on that module.
Non-measure module — For front panel operation, system channel operation cannot be
used to close channels. For remote programming, system channel operation can be used,
but only the one specified channel will close. All other channels on the module will open.
Non-amps switching modules
NOTEPresently, non-amps Keithley modules include the Models 7701, 7703, 7706,
7707, 7708, and 7709. You can check the Keithley website (www.keithley.com)
for new modules.
A non-amp module does not support amps measurements. System channel operation
cannot be used to close channels while an amps function (DCI or ICI) is selected.
If an amps function (DCI or ACI) is selected and you attempt to close a system channel,
the message “NO AMPS CHAN” will be displayed briefly. For remote programming,
error -222 (Parameter data out of range) is generated. Example:
SYST:PRES‘ Restores system preset defaults.
SENS:FUNC ‘CURR:DC’‘ Selects DCI function.
ROUT:CLOS (@101) ‘ Attempts to close system channel 101 – Generates
‘ error -222.
If a system channel is already closed and you attempt to select the DCI or ACI function,
the message “INVALID FUNC” will be displayed briefly. For remote programming,
error -221 (Settings conflict) is generated. Example:
SYST:PRES‘ Restores system preset defaults.
ROUT:CLOS (@101) ‘ Close system channel 101.
SENS:FUNC ‘CURR:DC’‘ Attempts to select DCI function – Generates
‘ error -221.
Making amps measurements — In order to perform amps measurements, you must use
the front panel inputs of the 2750 mainframe. You can still use the non-amps module for
other aspects of the test, but you must use multiple channel operation to close channels.
Example:
NOTEIn order to use the front panel inputs, make sure the INPUT switch is in the out
(F) position.
SYST:PRES‘ Restores system preset defaults.
ROUT:MULT:CLOS (@101) ‘ Closes channel 101.
ENS:FUNC ‘CURR:DC’‘ Selects DCI function – Legal operation.
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-15
Non-measure switching modules
NOTEPresently, non-measure Keithley modules include the Models 7705, 7711, and
7712. You can check the Keithley website (www.keithley.com) for new modules.
Keep the following in mind when using an non-measure module:
•For a non-measure card, no channels are connected to the internal DMM (the
channels cannot be connected to the backplane).
•Multiple channel operation should be used to close channels on a non-measure
module. For remote operation, the ROUT:MULT commands are used to close
channels.
•Front panel system (single) channel operation cannot be used to close channels on
a non-measure module. For front panel operation, system channel operation will
cause message “NO MEAS CARD” to be displayed.
•A non-measure module may have open/close operations that are specific only to
that module. Refer to the appropriate module manual (packing list) for details on
operation.
•In order to perform measurements, you must use the front panel inputs of the 2750
mainframe. You can still use the non-measure module to control other operations.
2-16Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Multiple channel operation
The capability to individually control channels provides you with added flexibility in how
you use a switching module. For example, assume you want to route a signal into channel
1 and out channel 20 of a Model 7700 switching module. You would do this by closing
channels 1, 20, and 23. If you open channels 24 and 25, you will isolate the input signal
from the DMM of Model 2750.
Multiple channel operation allows any channel (or channels) in the test system to be
closed or opened. It allows more than one measurement channel to be closed at the same
time. It also allows individual control of non-measurement channels, such as backplane
isolation channels. Multiple channel operation should only be performed by experienced
test system engineers.
WARNINGCareless multiple channel operation could create an electric shock
hazard that could result in severe injury or death. Improper operation
can also cause damage to the switching modules and external circuitry.
Multiple channel operation should be restricted to experienced test
engineers who recognize the dangers associated with multiple channel
closures.
NOTEMultiple channel operation cannot be used to perform thermocouple
temperature measurements using the internal or external reference junction. The
simulated reference junction will instead be used and the integrity of the
temperature reading will be questionable (“ERR” annunciator turns on). See
“Temperature measurements,” page 3-35, for details.
Some other key points for multiple channel operation include the following:
•Closing a channel using multiple channel operation has no affect on other closed
channels. Whatever channels were previously closed, remain closed.
•A channel closed using multiple channel operation is not displayed on the
Model 2750. Also, the CHAN annunciator does not turn on when a channel is
closed.
•Opening a channel using multiple channel operation has no affect on other closed
channels. Only the specified channel opens.
NOTEUse the VIEW option of the CARD menu to display closed channels (see “CARD
menu,” page 2-29).
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-17
Controlling multiple channels
WARNINGWhen using multiple channel operation, you must be very careful
when switching hazardous voltages. If you inadvertently close the
wrong channel(s), you could create a shock hazard and/or cause
damage to the equipment.
Most switching modules use latching relays. That is, closed channels
remain closed when the Model 2750 is turned off. Never handle a
switching module that is connected to an external source that is turned
on. Turn off all power sources before (1) making or breaking
connections to the module, and (2) installing (or removing) the module
into (or out of) the Model 2750.
Avoiding corrupt measurements
Aside from the safety issues, improper use of multiple channel operation can result in
corrupt measurements. For example, assume two Model 7700s installed in slots 1 and 2,
and a 2-wire function selected. If you use multiple channel operation to close channels 201
and 225, you will connect the input at channel 201 to the DMM for measurement.
If you then use system channel operation to close channel 101, channel 125 will also close
to connect the input at channel 101 to the DMM. You now have two input channels (101
and 201) connected to DMM Input at the same time, inviting all sorts of problems.
The above problem can be avoided by opening channels 201 and/or 225 before closing
channel 101 (and 125) as demonstrated by the following sequence:
1.Multiple channel operation — Close channels 201 and 225 for connection to
DMM.
2.Multiple channel operation — Open channels 201 and/or 225 to disconnect from
DMM.
3.System channel operation — Close system channel 101 to connect to DMM.
2-18Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
CLOSE key (MULTI menu option)
The MULTI menu option for the CLOSE key can be used to close any individual channel
in the mainframe (Figure 2-6). Perform the following steps to close a channel:
NOTEChannels closed by the MULTI option of the CLOSE key are not displayed. Use
the VIEW option of the CARD menu to display closed channels (see “CARD
menu,” page 2-29).
1.Press the CLOSE key and then use the
or key to display the “CLOSE:MULTI”
message.
2.Press ENTER to display the prompt to close a channel (CLOSE MLT:XXX).
3.Using
, , , and , key in the three-digit channel you want to select.
4.Press ENTER to close the channel.
An invalid channel cannot be closed. The error messages associated with system channel
operation also apply to multiple channel operation.
Figure 2-6
Multiple channel operation — specifying a channel to close
CLOSEOPEN
CLOSE:MULTI
CLOSE MLT:XXX
Press CLOSE key
Display MULTI option
and press ENTER
Specify channel number
(XXX) and press ENTER
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-19
OPEN key
The OPEN key has two options to open channels: ALL and MULTI. The ALL option
simply opens all channels in the mainframe. The MULTI option opens only the specified
channel. All other closed channels remain closed. Figure 2-7 summarizes OPEN key
operation.
OPEN: ALL — Perform the following steps to open all channels in the mainframe:
1.Press the OPEN key to display “OPEN: ALL.”
2.Press OPEN again (or press ENTER) to open all channels.
OPEN: MULTI — Perform the following steps to open only the specified channel:
1.Press the OPEN key. The “OPEN: ALL” message will be displayed.
2.Press the
3.Press ENTER to display the prompt to open a channel (OPEN MLT:XXX).
4.Using
5.Press ENTER to open the channel.
NOTEIf the channel you open using OPEN: MULTI is the system channel (channel
or key to display the “OPEN: MULTI” message.
, , , and , key in the three-digit channel you want to select.
number displayed on the Model 2750), the channel will open, but the system
channel number will still be displayed (see “Multiple channel operation
anomalies,” page 2-22).
Figure 2-7
Multiple channel operation — opening one or all channels
Press OPEN key
Display MULTI option
and press ENTER
Specify channel number
(XXX) and press ENTER
Display ALL option
and press OPEN again
OPEN: ALL
CLOSEOPEN
OPEN: MULTI
OPEN MULTI:XXX
2-20Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Remote programming — Multiple channel control commands
The commands to close and open the system channel are listed in Table 2-2.
Table 2-2
Multiple channel control commands
CommandsDescriptionRef
ROUTe:MULTiple:CLOSe <clist>Specify one or more channels to
close.
ROUTe:MULTiple:OPEN <clist>Open channels specified in list. Unlisted
channels not affected.
ROUTe:OPEN:ALLOpen all channels.c
ROUTe:MULTiple:CLOSe?Returns a <clist> of all closed channels.d
ROUTe:MULTiple:CLOSe:STATe?
<clist>
Query closed channels in specified list
(1 = closed).
Channel list parameter:
<clist> = (@SCH)
where: S = Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Reference:
a.ROUTe:MULTiple:CLOSe <clist>
This command functions like the front panel CLOSE key (MULTI menu option) to
close channels. When you send this command to close the channels specified in the
<clist>, only those listed channels will close. Channels not specified are not
affected, and channel pairing is disabled.
a
b
e
NOTES Channels closed by ROUT:MULT:CLOS are not displayed.
The ROUT:MULT:CLOS command cannot be used to perform
thermocouple temperature measurements using the internal or
external reference junction. The simulated reference junction will
instead be used and the integrity of the temperature reading will be
questionable (“ERR” annunciator on). See “Temperature
measurements,” page 3-35, for details.
NOTEFor RS-232 operation (and in some cases, GPIB operation), *OPC
or *OPC? should be used with :ROUT:MULT:CLOS if the <clist> is
large. Details on *OPC and *OPC? are provided in Section 12.
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-21
b.ROUTe:MULTiple:OPEN <clist>
With this command, you can open one or more switching module channels. When
you send this command to open the channels specified in the <clist>, only those
listed channels will open. Channels not specified are not affected.
NOTE For RS-232 operation (and in some cases, GPIB operation), *OPC
or *OPC? should be used with :ROUT:MULT:OPEN if the <clist>
is large. Details on *OPC and *OPC? are provided in Section 12.
c.ROUTe:OPEN:ALL
This command functions the same as the front panel OPEN key (ALL menu
option). It simply opens all channels (including non-measurement channels) in the
mainframe.
d.ROUTe:MULTiple:CLOSe?
This query command returns a <clist> of all closed channels, including
non-measurement channels and paired channels for 4-wire functions.
e.ROUTe:MULTiple:CLOSe:STATe? <clist>
This query returns a “0” (open) or “1” (closed) for every channel specified in the
<clist>. It is valid for both measurement and non-measurement channels.
For example, assume channel 125 is closed, and you use this command to query
channels 101, 104, and 125 (<clist> = (@101, 104, 125)). The response message
returns “0, 0, 1” to indicate that channels 101 and 104 are open, and channel 125 is
closed.
Remote programming example (multiple channel operation)
The following example assumes a Model 7700 installed in slot 1. This command sequence
connects channel 101 to channel 111 (through channel 123). Note that these two closed
channels will be internally isolated from the DMM since the backplane isolation channels
(124 and 125) will be open.
NOTEThe following example can be run from the KE2700 Instrument Driver using the
example named “CloseChannels” in Table H-1 of Appendix H.
ROUT:OPEN:ALL' Open all channels.
ROUT:MULT:CLOS (@101,111,123)' Close channels 101, 111, and 123.
When finished with multiple channel operation it is a good, safe practice to open all channels (ROUT:OPEN:ALL).
2-22Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Multiple channel operation anomalies
•Anomaly #1 — When you use multiple channel operation to open the system
channel, the channel will open but the system channel number will still be
displayed on the Model 2750. For details, see “Anomaly #1 example — wrong
channel displayed.”
•Anomaly #2 — For a 4-wire function, you can use multiple channel operation to
open the paired channel. If you then use system channel operation to again select
the already closed system channel, it will not re-close the paired channel. For
details, see “Anomaly #2 example — opening the paired channel.”
NOTEThe following anomaly examples assume a Model 7700 installed in slot 1.
Anomaly #1 example — wrong channel displayed
The following example closes channel 102 and connects it to the DMM Input. However,
the Model 2750 will not display the measurement channel that is closed. It will display
channel 101 instead of channel 102.
1.Use the ALL option for the OPEN key to open all channels in the mainframe.
Remote programming:
2.Press the key to close (and display) channel 101. This closes channel 101 (which
is the system channel) and channel 125 to connect it to the DMM Input
(Figure 2-1).
Remote programming: ROUT:CLOS (@101)
3.Use the MULTI option for the CLOSE key to close channel 102. The system
channel is not affected. Channels 101, 102, and 125 are now closed.
Remote programming: ROUT:MULT:CLOS (@102)
4.Use the MULTI option for the OPEN key, open channel 101. Even though channel
101 is still being displayed on the Model 2750, it is channel 102 that is actually
connected to the DMM Input (channels 102 and 125 closed).
Remote programming: ROUT:MULT:OPEN (@101)
ROUT:OPEN:ALL
To correctly display the channel that is closed (channel 102) repeat step 1 above to open
all channels, and then use the key or the ROUT:CLOS (@102) command to close (and
display) channel 102. This closes channel 102 (which is the system channel) and channel
125 to connect it to the DMM Input.
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-23
Anomaly #2 example — opening the paired channel
Assume 4-wire connections to a 1kΩ resistor using channels 1 and 11 of the Model 7700
switching module. Also assume the Ω4 function is selected. The following procedure
demonstrates how careless multiple channel operation can cause an overflow reading even
though everything else from the front panel “looks right.”
1.Use the ALL option for the OPEN key (OPEN: ALL) to open all channels in the
mainframe.
Remote programming: ROUT:OPEN:ALL
2.Press the key to close (and display) channel 101. The following channels close
(see Figure 2-2):
•Channel 101 (system channel).
•Channel 125 (connects channel 101 to DMM Input).
•Channel 111 (paired channel for 4-wire measurements).
•Channel 124 (connects channel 111 to DMM Sense).
•Channel 123 (isolates channel 101 from channel 111).
The Model 2750 will display the 1kΩ reading for system channel 101.
Remote programming:
3.Using the MULTI option for the OPEN key, open channel 111. This opens the
connection to DMM Sense and causes an OVRFLW reading. Keep in mind that
channel 101 is still closed and displayed as the system channel.
Remote programming:ROUT:MULT:OPEN (@111)
4.In an attempt to clear the overflow reading problem, use the SINGLE option of the
CLOSE key to again close channel 101. You might think that this will again close
channel 111 to reconnect it to DMM Sense. However, that is not the case. Since
channel 101 is still the system channel, selecting it again in this manner is a “no
action.” Channel 111 does not close.
Remote programming:ROUT:CLOS (@101)
ROUT:CLOS (@101)
A simple way to resolve the above problem is to repeat step 1 to open all channels, and
then repeat step 2 to close channel 101. All the listed channels in step 2 will close to make
the 4-wire connection to the 1kΩ resistor.
2-24Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Dual independent multiplexers
Using multiple channel operation, any multiplexer switching module can be configured as
two independent multiplexers. For example, the Model 7700 is normally used as a single
1 × 20 multiplexer, but it can also be configured as two 1 × 10 multiplexers.
NOTEThermocouple temperature measurements using the internal or external
reference junction cannot be performed when using multiple channel operation
to connect an input channel to the DMM. The simulated reference junction will
instead be used resulting in invalid readings (“ERR” annunciator turns on). See
“Temperature measurements,” page 3-35, for details.
A multiplexer switching module is configured as two multiplexers by using multiple
channel operation to close the 2-pole/4-pole relay. The Model 7700 is configured as two
independent multiplexers by closing channel 23. As shown in Figure 2-8, the closed
position of channel 23 isolates Multiplexer A (channels 1 through 10) from Multiplexer B
(channels 11 through 20).
For the dual multiplexer configuration, only Multiplexer A channels can be internally
connected to the DMM of the Model 2750. For the Model 7700, closing channel 25 allows
channels 1 through 10 to be measured by the DMM.
When using the dual multiplexer configuration, the sense backplane isolation relay must
be kept open to isolate Multiplexer B channels from the sense terminals of the DMM. For
the Model 7700, channel 24 must be kept open (Figure 2-8).
Figure 2-8
Dual multiplexer configuration (Model 7700)
HI
Ch 1
LO
Multiplexer A
(1x10)
Multiplexer B
(1x10)
Channels
HI
LO
HI
LO
Channels
HI
LO
2–9
Ch 10
(Closed)
Ch 11
12–19
Ch 20
Ch 25
Ch 23
Ch 24
For the dual multiplexer configuration,
Ch 23 must be closed, and Ch 24 must
remain open.
HI
Input
LO
To
Model 2750
DMM
HI
Sense
LO
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-25
Dual multiplexer application
This application demonstrates how to use the Model 7700 as a dual multiplexer to bias and
measure 10 DUT. An external source powers the DUT, while the DMM of the Model 2750
measures the output of the DUT. To prevent overloading of the external source, each DUT
is powered (and measured) separately.
Figure 2-9 shows the connections for this application. The external source is connected to
the Sense terminals of the switching module, and DUT is connected to channels 1 through
10. Channels 11 through 20 are used to connect external power to each DUT.
For this application, the 2-pole/4-pole relay and backplane isolation relays of the switching module are to be controlled as follows:
•Closing channel 23 connects the External Source to DUT via channels 11 through
20. Closing channel 23 also isolates measure channels (1 through 10) from the
source channels (11 through 20). This channel must remain closed while testing
DUT.
•Opening channel 24 isolates the external source from the backplane of the
Model 2750. This channel must remain open while testing DUT.
•Closing channel 25 connects an input channel (1 through 10) to the DMM.
In Figure 2-9, channels 1 and 11 are closed to test DUT 1. A more detailed view of the test
for DUT 1 is shown in Figure 2-10. The test for the other DUTs is similar except that
different source and measure channels are closed. Closed channels for each DUT test are
listed as follows:
Tested
deviceClosed channels
DUT 11, 11, 23 and 25 DUT 66, 16, 23 and 25
DUT 22, 12, 23 and 25DUT 77, 17, 23 and 25
DUT 33, 13, 23 and 25DUT 88, 18, 23 and 25
DUT 44, 14, 23 and 25DUT 99, 19, 23 and 25
DUT 55, 15, 23 and 25DUT 1010, 20, 23 and 25
NOTEDo not use this application to measure the temperature of the DUT using a
thermocouple with the INTernal or EXTernal reference junction selected. The
SIMulated reference junction will instead be used resulting in invalid readings.
The “ERR” annunciator will turn on to indicate that the integrity of the
temperature reading is questionable.
Tested
deviceClosed channels
2-26Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Figure 2-9
Dual multiplexer application connections
Model 2750
Model 7700 Switching Module
H1
External
Source
DUT
1
DUT
2
Sense
LO
H1
LO
H1
LO
Ch 1
Ch 2
DUT
10
H1
LO
H1
LO
H1
LO
H1
LO
Ch 10
Ch 11
Ch 12
Ch 20
Ch 25
Ch 23
(Closed)
Ch 24
HI
Input
LO
DMM
HI
Sense
LO
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-27
Figure 2-10
Testing DUT 1
Model 2750
Model 7700 Switching Module
HI
External
Source
DUT
1
Sense
LO
HI
Ch 1
LO
Ch 25
Slot 1
HI
Input
LO
DMM
HI
Sense
LO
DMM
HI
Ch 11
LO
Mutliple channel operation:
Open all channels
Close channel 123
Close channel 125
Close channel 101
Close channel 111
Ch 23
(Closed)
External
Source
Equivalent Circuit
Ch 24
DUT
1
Test procedure:
NOTES The following test procedure assumes a Model 7700 switching module installed
in slot 1 of the mainframe.
The procedure assumes that the instrument is operating in the continuous
measurement (triggering) mode (see “Defaults and user setups,” page 1-21).
Do not use the following procedure to perform thermocouple temperature
measurements with the INTernal or EXTernal reference junction selected. The
SIMulated reference junction will instead be used resulting in invalid readings.
The “ERR” annunciator will turn on to indicate that the integrity of the
temperature reading is questionable.
2-28Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
1.Open all channels. For most switching modules, channels remain closed after the
Model 2750 is turned off. Therefore, it is good, safe practice to open all channels at
the start and end of the test.
Front panel operation: Press OPEN > Display ALL > Press OPEN
Remote programming:
ROUT:OPEN:ALL
2.Close channels 23 and 25.
Front panel operation: Press CLOSE > Select MULTI > Key in 123 >
Press ENTER
Press CLOSE > Select MULTI > Key in 125 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@123,125)
3.Close channels 1 and 11 to connect DUT #1 to the DMM and bias supply.
Front panel operation: Press CLOSE > Select MULTI > Key in 101 >
Press ENTER
Press CLOSE > Select MULTI > Key in 111 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@101,111)
4.Measure DUT #1.
Front panel operation: Take reading from display
Remote programming:
DATA?
5.Open channels 1 and 11.
Front panel operation: Press OPEN > Select MULTI > Key in 101 >
Press ENTER
Press OPEN > Select MULTI > Key in 111 >
Press ENTER
Remote programming:
ROUT:MULT:OPEN (@101,111)
6.Modify steps 3, 4, and 5 to test DUT #2. That is, close channels 2 and 12, measure
DUT #2, and then open channels 2 and 12.
7.Test the remaining eight DUT in a similar manner. That is, close the appropriate
channels for the DUT, make the measurement, and then open the channels.
8.After the last DUT is tested, repeat step 1 to open all channels.
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-29
Identifying installed modules and viewing
closed channels
On power-up, the model numbers of installed switching modules are displayed briefly.
While in the normal display state, red slot indicators on the right side of the display
indicate which slots have a switching module or pseudocard installed.
NOTEIf a Model 7700, 7701, 7702, 7703, 7705, 7708, or 7709 switching module is
removed while the Model 2750 is on, the slot indicator for that slot will remain
on, and the instrument will operate as if the module is installed. That is, the
Model 2750 will operate as if the pseudocard is installed.
NOTEIf a Model 7706 or 7707 is removed while power is on, error +523 (“Card
hardware error”) will occur, and the module will be removed from the system.
NOTEIn general, it is not recommended to install or remove switching modules with
the power on.
CARD menu
The CARD menu identifies the switching modules installed in the mainframe, and is used
for the following operations:
•Configure digital inputs and outputs, and analog outputs for switching modules that
have one or more of those capabilities (i.e., Models 7706 and 7707).
•View the analog input channels that are presently closed. Also, read digital input
and output ports, and analog output values for switching modules that have one or
more of those capabilities.
Menu navigation keys — Once in the menu structure, the manual range keys (
and the cursor keys (
parameter values. With the desired item, option or setting displayed, press the ENTER key
to select it. You can cancel a pending selection (and exit the menu structure) by pressing
the EXIT key.
Press the SHIFT key and then the CARD key to display the CARD menu. The Card menu
tree is shown in Figure 2-11. The items and options of the menu are explained as follows:
NOTEIdentifying installed modules — If you simply want to identify installed modules
or pseudocards, select CONFIG or VIEW and use the
slot. While in the menu structure, the slot indicator for the selected slot will be
on. The other indicators will be off. When finished, press EXIT.
and ) are used to display menu items and options, and set
or key to check each
and )
2-30Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
CARD: CONFIG — This menu item is used to configure switching modules. The
channels of the Model 7700 switching module and other similar type modules do not need
to be configured.
SLOTX: 77xx — Use to configure the switching module in Slot X (where X = 1, 2, 3, 4,
or 5). If configuration is not necessary, the instrument will exit from the menu when
ENTER is pressed.
NOTEFor switching modules that require configuration, refer to the packing list that
was shipped with each module.
CARD: VIEW — This menu item is used to view all analog input channels that are
presently closed. These include both measurement and non-measurement channels.
The channels are built into a string that scrolls the display. Four dots identify the end of the
string. Model 7700 example (Slot 1) — Assume the Ω4 function is selected and system
channel 101 is closed. The following string will scroll across the display:
101, 111, 123, 124, 125 . . . .
Channels 101 and 111 are the paired channels for the 4-wire measurement. Channel 123 is
the 4-pole relay setting, and channels 124 and 125 connect input and sense to the DMM of
the Model 2750 (Figure 2-2).
NOTESome switching modules have analog outputs, digital inputs, and/or digital out-
puts. The values for these channels are also displayed from the VIEW menu item.
For details on a particular switching module, refer to the packing list that was
shipped with each module.
SLOTX: 77xx — Use to scroll the closed channels and channel settings (if applicable) for
the switching module in Slot X (where X = 1, 2, 3, 4, or 5).
Scrollingspeed — The scrolling speed of the channel string is adjustable, or can be
paused. The
key slows down scrolling speed and the key speeds it up. The ENTER
key pauses scrolling. Press ENTER a second time to resume scrolling.
Exiting VIEW —
key will also exit VIEW, but it will also perform the operation associated with the key. For
example, pressing Ω2 will exit VIEW, and select the Ω2 function.
To exit from VIEW, press the EXIT key. Pressing an instrument setting
NOTEWhen a command is received while the display is scrolling, the instrument exits
from the CARD menu and the command is executed.
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-31
Figure 2-11
CARD menu tree
SHIFT
CARD
CONFIG
SLOT1: 77XXSLOT5: 77XX
77XX = Model number of installed
switching module.
SLOT1: 77XX
Scrolls
Channels
Switching module queries (remote operation)
For remote operation there are commands to identify installed switching modules and
channels that are closed. There are also commands to acquire general information about
the installed modules.
*OPT?
For remote operation, the *OPT? command can be used to determine which switching
modules (or pseudocards) are installed in the Model 2750. For example, assume a
Model 7700 is installed in slot 1, a Model 7706 is installed in slot 2 and the other slots are
empty. After sending *OPT? and addressing the Model 2750 to talk, the following
response message will be sent to the computer:
7700, 7706, NONE, NONE, NONE
VIEW
SLOT5: 77XX
Scrolls
Channels
2-32Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
These query commands are used to determine closed switching module channels.
ROUT:CLOS? is used to return a list of closed measurement channels including the paired
channel for 4-wire measurements. It will not return non-measurement channels. For
details, see Table 2-1 and related reference information.
ROUT:MULT:CLOS? is used to return all closed channels (measurement and nonmeasurement). For details, see Table 2-2 and related reference information.
ROUT:MULT:CLOS:STAT? is used to return the state (open or closed) of each specified
channel. A “0” is returned for an open channel, and a “1” is returned for a closed channel.
For details, see Table 2-2 and related reference information.
SYSTem:CARD commands
There is a series of SYSTem:CARD commands that can be used to acquire the following
information about a switching module installed in the Model 2750:
•Return the serial number and firmware revision.
•Determine the maximum allowable voltage.
•Determine if the module supports multiplexer or isolated channels.
•Determine if the module has built-in temperature sensors for internal cold junction,
thermocouple temperature measurements.
•Determine which channels are used for volts/2-wire measurements and which are
used for amps.
•Determine which channels are used for analog or digital output.
•Determine the totalizer channel (Model 7706 only).
The SYSTem:CARD commands are covered in Table 15-7.
Relay closure count
The Model 2750 keeps an internal count of the number of times each module relay has
been closed. The total number of relay closures are stored in EEPROM on the card. This
count will help you determine if and when any relays require replacement (see module
contact life specifications).
Relay closures are counted only when a relay cycles from open to closed state. If you send
multiple close commands to the same channel without sending an open command, only
the first closure will be counted.
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-33
Relay closure count can only be read via remote operation. The commands are
summarized in Table 2-3. Details follow the table.
Table 2-3
Relay closure count commands
CommandsDescriptionDefault
ROUTe:CLOSe:COUNt? <clist>Query close count for specified
channels.
ROUTe:CLOSe:COUNt:INTerval <NRf> Set count update interval in minutes
where: S = Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Note: Relay count interval set to 15 minutes at the factory. SYSTem:PREset and *RST have no effect on the
set interval.
NOTEThe relay closure count can be reset to zero. For details, see Model 2750 Service
Manual, “Plug-in module relay closure count.”
Reading relay closure count
To determine the closure count of specific channels, send this query via remote:
ROUTe:CLOSe:COUNt? <clist>
Here, <clist> is the summary of channels. For example, to determine the closure count of
channels 1 and 4 of a module in slot 1, the following query would be sent:
ROUT:CLOS:COUN? (@101,104)
The following query would determine the closure count of slot 1 module channels 1
through 10:
ROUT:CLOS:COUN? (@101:110)
2-34Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Setting count update interval
Relay closure counts are updated in temporary RAM every time a channel is closed
regardless of how it was closed: by an SCPI command, front panel control, or during a
scan. These counts are permanently written to the EEPROM on the card only at a user-set
time interval (which is initially set to 15 minutes at the factory), or whenever the counts
are queried. Valid intervals (set in integer number of minutes) are between 10 and 1440
minutes (24 hrs).
The lower the interval, the less chance there is of losing relay counts due to power
failures. However, writing to the EEPROM more often may reduce scanning throughput.
The higher the interval, the less scanning throughput is reduced, but more relay counts
may be lost in the event of a power failure.
NOTEIf the Model 2750 is turned off before the updated count is written to EEPROM,
the relay counts will be lost. It is good practice to add the ROUT:CLOS:COUN?
<clist> command at the end of a program to manually update the count.
To set the count update interval, send this command:
ROUTe:CLOSe:COUNt:INTerval <NRf>
where; <NRf> = 10 to 1440 (minutes)
For example, to set the interval to 30 minutes, send this command:
ROUT:CLOS:COUN:INT 30
Model 7700 switching module
NOTEConnection and wiring procedures for the Model 7700 are to be performed by
qualified service personnel. This information is provided in Appendix B
(Model 7700 Connection Guide).
Switching module capabilities
Channels 1 through 20 — The Model 7700 can multiplex one of 20 2-pole signals, or one
of 10 4-pole signals into the input of the Model 2750.
Channels 21 and 22 — The Model 7700 can multiplex one of two 2-pole current signals
into the input of the Model 2750.
Model 2750 Multimeter/Switch System User’s ManualClose/Open Switching Module Channels2-35
CAUTIONTo prevent damage to the Model 7700 switching module, do not exceed
these maximum signal levels:
Channels 1-20:300VDC or 300V RMS (425V peak) for AC
waveforms, 1A switched, 60W, 125VA
Channels 21, 22:60VDC or 30V RMS, 3A switched, 60W, 125VA
NOTESystem channel operation – Of the 22 measurement channels, only one channel
(or channel pair) can be closed at the same time. When you close a channel (or
channel pair), all other measurement channels will open. The user has no
control of channels 23, 24, and 25. The open/close state of these channels are
determined by the selected function.
The Model 7700 has six temperature transducers to monitor the cold junction temperature
at the screw terminals. For temperature measurements, this internal reference junction
allows thermocouples to be connected directly to the screw terminals of the module.
When the Model 2750 is on the DCV, ACV, Ω2, CONT, Ω4, FREQ, PERIOD, or TEMP
function, channels 1 through 20 are available. When on a current function (DCI or ACI),
channels 21 and 22 are the only available channels.
The Model 7700 can accommodate 4-wire measurements by using channel pairs. Primary
channels 1 through 10 become paired to channels 11 through 20. For example, with the Ω4
function selected, channel 1 becomes paired to channel 11. For example, when you close
channel 1, channel 11 will also close.
The 2-wire functions include DCV, ACV, DCI, ACI, Ω2, CONT, FREQ, PERIOD, and
TEMP (thermocouple and thermistor). The 4-wire functions/operations include Ω4,
TEMP (4-wire RTD), RATIO, and CH AVG (ratio and channel average are covered in
Section 5).
With a 4-wire function/operation selected, channels are paired as follows:
• CH1 and CH11• CH6 and CH16
• CH2 and CH12• CH7 and CH17
• CH3 and CH13• CH8 and CH18
• CH4 and CH14• CH9 and CH19
• CH5 and CH15• CH10 and CH20
2-36Close/Open Switching Module ChannelsModel 2750 Multimeter/Switch System User’s Manual
Schematic diagram
The simplified schematic diagram of the Model 7700 is shown in Figure 2-12. Channels 1
through 20 are used for all measurements except amps. Channels 21 and 22 are used for
amps only.
There are two backplane relays (channels 24 and 25) to connect the input channel(s) to the
backplane of the Model 2750. With a 2-wire function (except amps) selected, channel 25
will close, and with a 4-wire function selected, both channels 24 and 25 will close.
There is a 2-pole/4-pole relay (channel 23) between channels 1-10 and channels 11-20.
When a 2-wire function (i.e., DCV) is selected, channel 23 opens (2-pole position) to
allow any of the 20 channels to be connected to the input backplane.
When a 4-wire function is selected, channel 23 closes (4-pole position) to isolate channels
1 through 10 from channels 11 through 20. With a system channel (1 through 10) closed,
its paired channel (11 through 20) will also close to connect the sense channel to the sense
backplane.
For the two current channels (21 and 22), signal HI and LO are routed directly to the backplane when the channel is closed.
As shown in Figure 2-12, there are also screw terminals labeled “Input,” “Sense,” and
“Amps.” The Input and Sense terminals are connected to the inputs of channels 24 and 25
(isolation relays). If channels 1 through 20 are not intended to be connected to the internal
DMM, channels 24 and 25 can be controlled independently using multiple channel
operation. The Amps terminals are connected directly to the DMM.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.