Keithley 169 Service manual

Instruction Manual
Model 169
Digital Multimeter
01979, Keithley Instruments, Inc.
Cleveland, Ohio, U.S.A
Document Number 30161
SECTION
1.
GENERAL INFORMATION.
1.1.
INTROOUCTION
1.3.
FEATURES .......................... l-1
1-5.
WARRANTY INFDKMATION ....................
l-7.
CHANGE NOTICES
1-9.
SAFETY SYMBOLS
l-11. SPECIFICATIONS
2.
OPERATION. 2-l. INTRODUCTION
2-3. 2-5. 2-6. 2-7. 2-8.
2-10. Front Panel Switches 2-12. Input Connections. 2-14. Tilt Bail/Handle 2-16. INITIAL FUNCTIONAL CHECK 2-18. OPERATING INSTRUCTIONS 2-20. DC VOLTAGE MEASUREMENT 2-22. AC VOLTAGE MEASUKEMENT 2-24. AC OR DC CUKRENT MEASUREMENT 2-26. RESISTANCE (n) MEASUREMENT
3.
PERFORWINCE VERIFICATION
3-1.
3.3.
3-5. ENVIRONMENTAL CONDITIONS .................. 3-l
3-7.
3-9. Initial Conditions 3-11. DC Volts Checkout. 3-12. AC Volts Checkout.
3-13. Resistance Checkout. 3-14. DC Current Checkout. 3-15. ANALYSIS
4.
ACCESSORIES.
4-1. 4-3. 4-5. 4-7.
4-9.
4-11.
4-13. MODEL 1684 CARRYING CASE 4-15. MODEL 1635 CLAMP-ON AC CURRENT PROBE 4-17. MODEL
4-19.
4-21. MODELS
............................ 2-l
UNPACKING AND INSPECTION .................. 2-l
PREPARATION FOR USE. .................... 2-l
Installation of Batteries, if required ........... 2-1
CONTROLS, INDICATORS AND CONNECTORS. ............
Display ........................... 2-1
GENERAL ........................... 3-l
RECOMMENDED TEST EQUIPMENT ................. 3-1
PERFORMANCE VERIFICATION PROCEDURE ............. 3-1
...........................
GENERAL ........................... 4-l
MODEL
1600
MODEL
1651
MODEL 1681 CLIP-ON TEST LEAD SET .............. 4-l
MODEL 1682 RF PROBE. ....................
MODEL 1683 UNIVERSAL TEST LEAD KIT
1691
MODEL
1699
1010
.......................
........................
.......................
.......................
.......................
........................
.................... 2-2
.....................
......................
..................
...................
...................
...................
................
.................
.....................
.....................
..................... 3-2
.....................
.................... 3-3
....................
..........................
HIGH VOLTAGE PROBE. ............... 4-l
50-AMPERE SHUNT ................. 4-1
............. 4-2
..................
............ 4-2
GENERAL PURPOSE TEST LEAD SET
SPARE PARTS KIT
and 1017 RACK MOUNTING KITS.
................. 4-2
.......... 4-2
.......... 4-2
PAGE
l-l 1-l
1-2 1-2 l-2 1-2
2-l
2-l
2-2 2-4 2-4 2-4 2-5 2-5 2-5 2-6
3-l
3-2
3-2
3-3 3-3
4-l
4-1
4-2
5.
THEORY OF OPERATION. 5-l.
GENERAL.
5-3.
OVERALL OPERATION.
5-6.
SIGNAL CONDITIONING.
5-7.
AC/DC Voltage Measurements
5-10.
AC/DC Current Measurements
5-12.
Resistance Measurements.
5-14.
A/D CONVERTER.
ii
I
.......................
................
.............
...........
..........
.......
.......
........
........
........ 5-l
........ 5-l
........
........ 5-I
........ 5-2
........ 5-2
6-l 5-l
5-I
i i
TABLE OF CONTENTS (Cont'd)
SECTION
5-16. DISPLAY CIRCUITS 5-18. REFERENCE SOURCE 5-20. LOW BATTERY DETECTOR 5-22. INPUT PROTECTION 5-24. Voltage Protection 5-27. Current Protection 5-29. Ohms Protection.
6. MAINTENANCE.
GENERAL ...........................
6-1.
CALIBRATION. ........................
6-3.
Recormended Calibration Equipment.
6-5.
Environmental Conditions
6-7.
Case Cover Removal
6-9. 6-11. Calibration Adjustment 6-14. TROURLESHOOTING. 6-16. Special tiandling of Static Sensitive Devices 6-18. Power supply 6-20. A/D Converter and Display Driver 6-22. AC Converter 6-24. Voltage Divider, Reference Source and Ohms Circuit b-26. Current Circuitry. 6-31. LIQUID CRYSTAL DISPLAY REPLACEMENT 6-33. BATTERY REPLACEMENT.
6.35. CURRENT FUSE REPLACEMENT
7. REPLACEABLE PARTS.
GENERAL ...........................
7-I.
ORDERING INFORMATION
7-3.
FACTORY SERVICE.
J-5.
SCHEMATIC. .........................
7-J.
COMPONENTS LAYOUT.
7-9.
J-11.
MODEL
...........................
1699
......................
......................
....................
......................
.....................
.....................
......................
.....................
...................
......................
........................
........................
.....................
....................
........................
....................
......................
.....................
SPARE PARTS KIT
.............
..................
..............
.............
..................
.................
........
.....
PAGE
5-2 5-3 5-3 5-3 5-3 5-3
5-3 6-1 6-l 6-l h-l h-l 6-1 6-I 6-2 6-2 6-3 6-3 h-3 6-3 6-3 6-4 6-9 6-9
7-D
7-o 7-D
7-O
7-O 7-D 7-O
FIGURE
l-l
2-l 2-2 2-3
4-l 4-2 4-3 4-4 5-I 5-2 5-3 5-4
6-1
6-2 6-3 6-4 6-5
7-l
ILLUSTRATIONS
TITLE JhGJ
Model lG9.
Installation of Batteries. Model 169 Front Panel Display. Operating Controls High Voltage Probe and RF Probe. Current Probe and Current Shunt. Convenience Cable and Connectors Carrying Case and Hack Mounting Kits Model 169 DMM Overall Signal Flow Block IDiagram. Simplified AC/DC Voltage Measurement Simplified AC/DC Current Measurement Simplified Resistance Measurement. Calibration Adjustment Location. LCD Assembly
Integrator Waveform. AC/DC Converter Waveform Current Fuse Removal/Replacement Covers and Panels.
.......................
...............
.............
...................
............
............
............
..........
..........
..........
...........
............
......................
..................
................
............
...................
.... 5-5
I-1 2-o 2-O 2-3
4.2
4.3
4.3
4.4
5-6
5-J
5-8 6-2
6-4
6-6
6-J
6-9
7-4
MOOEL 169
GENERAL INFORMATION
SECTION 1. GENERAL INFORMATION
l-l.
INTRODUCTION.
The Model 169 is a rugged, low cost, battery-
l-2. operated digital multimeter that provides full measurement capability for many general purpose applications. IOOuVldigit to 1OOOV on five dc and five ac voltage ranges, lOOnA/digit to 2000mA (2 amperes) on five dc
and five ac current ranges, and lOIlmn/digit to 20 megohms on six resistance ranges. Ranges and ~ccurxy are listed in the Table of Specifications on page 1-3.
FEATURES.
l-3.
The Model 169 incorporates features that help
l-4. prevent instrument damage through customer abuse and provide maximum instrument reliability, operator safety and simplicity of operation. features are:
A ZOOO-count (3-l/2 digit) liquid crystal dis-
e
play (LCD) with large 0.6" numerals, function and range annunicators, and a low battery indi-
It provides dccurate readings from
Some of these
cator (BAT). minus sign (-) are also displayed, positive polarity is implied.
l
The Model 169 is built rugged. The full bench­sire cdse is molded from impact-resistant dnd flame-retardant plastic.
tection prevents damage on all Voltage ranges dre protected up to 1400V peak, resistance ranges to 300VRMS and current ranges dre protected by dn externally accessible 2A
(25OV) fuse.
0 Pushbutton switches dre color coded to the
front panel for quick and edsy selection of
function and range. The decimal point is duto­matically positioned by the rdngc pushbutton. Annunciators on the display selected function and range. and function combinations are indicated by contradicting function and/or range d""u"cid­tars appearing dt the sdme time.
Appropriate decimal point and
Effective input pro-
functions.
Indicate
Improper range
the
FIGURE l-l. Model
169. l-l
GENERAL INFORMATION
The latest LSI technology and stable precision
l
components have been used in the Model 169 to
provide long term accuracy and minimize main-
tenance.
required, once a year. If alkaline batteries are used, battery life can be the same as the calibration cycle. an MTRF of up to 20,000 hours can be expected.
Only one calibration adjustment is
Barring destructive misuse,
MODEL
l-l.
CHANGE NOTICES.
l-8. which occur after printing of the Instruction Manual will be explained on a Change Notice sheet attached to the inside back cover.
l-9.
Improvements or changes to the instrument
SAFETY SYMBOLS.
169
Automatic zeroinq, automatic polarity and only
l
two input terminals simplify operation by eliminating potentiometer zeroing and lead
changing.
Optional accessories can be ordered to extend
l
the measurement capability of your Model
Some of these are:
High Voltage Probe allows your DMM to measure from 1000V to 40kV dc.
High Frequency (RF) Probe allows your DMM to measure from 0.25V to 30V rms ac over a fre­quency range of IOOkHZ to IOOMHZ.
Clamp-On AC Current Probe allows your DMM to measure from zero to 200A rms ac.
50.Ampere Current Shunt allows your IXIM to neasure from zero to 50A. ac and dc.
NOTE
Refer to Section 4 for more detailed
information on accessories.
l-5.
WARRANTY INFORMATION.
169.
l-10.
follows:
l-11.
1-12.
given in Table I-I.
Safety symbols used in this manual we as
IMPORTANT
The A symbol can be found in various places in this manual. Carefully read the associated CAUTION statements with regard to proper use and handling of the instru­ment. Damage to the instrument may occur if these precautions are ignored.
This places in this manual. This svmbal indi­cates those areas on the instrument which are potential shock hazards. Carefully read the associated WARNING statements with regard to proper use and handling of the instrument. Serious personal injury
result if
w ignored.
SPECIFICATIONS.
Detailed specifications for the Model 169 are
symbol can be found in various
these precautions are
l-6. cover of this Instruction Manual.
need to exercise the warranty, contact the Keithley Representative in your area to determine the proper action to be taken. facilities in the United Kingdom and West Germany, as well as in the United States.
front cover of this manual for addresses.
1-2
The warranty is given on the inside front
If there is a
Keithley maintains service
Check the inside
MODEL 169
GENERAL INFORMATION
TABLE 1-1.
Specifications
Mode
1169
MAXIM”M ALLOWABLE INPUT: 10”“” ims. ,400” prak.
,“*“.Hr. cO”ti”“o”S except 2”“m” range: 35”” m”,i”uuLkr. 15 sec. rndXirnUnl above 35”“.
1-3
BATTERY
ADJUSTMENT RlO,
l-
FIGURE 2-1.
NEGATIVE DISPLAYED,
POSITIVE
IWLIED WHEN LW
MINUS l-1 CFF
AC DIPLAYEO,
DC IWLIE"
FIGURE 2-2.
BATTERY
I
3-l/2 DIGITS
WITH DECIMAL
-__
Installation of Batteries.
MILLI (10-3)
MICRO I 10-6)
I
POINT
Model 169 Front Panel Display.
2-O
I
MODEL 169
SECTION 2. OPERATION
OPERATION
2-1. INTRODUCTION,
2-2. This section provides information needed for incoming inspection, preparation for use and opera­tion of the Model 169 and its accessories.
2-3. UNPACKING AN0 INSPECTION.
2-4. The Model 169 was carefully inspected, both mechanically and electrically before shipment. Upon receiving the Model 169, unpack all items from the
shipping container and check for any obvious damage which (may have occured during transit. damages to the shipping agent. Retain and use the original packaging materials if reshipment is re-
quired. Model 169 orders:
a. Model 169 DMM
b. six carbon zinc "C' cells (may be shipped
installed in the DMM in some units).
c.
d.
2-5.
2-6.
The following items are shipped with all
A copy of this Manual.
Separate optional accessories, as ordered.
PREPARATION FOK USE.
Installation of Batteries, if required.
neport any
NOTE
Carbon-zinc batteries are supplied. These batteries will operate your DMM for ap­proximately 1000 hours. If it is desired to change batteries only once a year when calibration is required, alkaline cells
may be substituted. These cells will pro-
vide approximately 2000 hours of opera-
tion.
indicator will automatically come on with
approximately 5% of battery life remaining
(20 hours for carbon-zinc and 40 hours for
alkaline cells).
t. Install the batteries in the holder ds shown
in Figure 2-1.
d. Press the battery clip onto the terminals of the battery holder; then position battery holder on the two spacers as illustrated.
e. Reinstall the top cover.
f. After the batteries are installed. you should become familiar with the Model I69 by reading the section titled Controls, Indicators and Connectors (Paragraph Z-7); then perform the Initial Functional
Check (Paragraph Z-16).
In either case,
the low battery
WARNING
3
To prevent a shock hazard, all test leads
should be removed from the INPUT terminals
before removing the instrument top cover.
a. Turn the DMM bottom side-up and loosen the four screws in the bottom cover until the threads are not engaged.
and they will fall out when the instrument is re­turned to its normal position.
b. Hold the top and bottom ‘over together to prevent their separation and turn the DMM cover to normal position.
NOTE: The scrws are not retained
Lift off the top cover.
2-7.
CONTROLS, INDICATORS AND CONNECTORS.
2-B.
Display.
2-9. The operating status of the Model 169 is con­tinuously provided by the 3-l/2 digit (2000 count) liquid crystal display located on the front panel.
In addition to the digitized input signal, with appropriate decimal point and polarity, the display provides indications of the selected function and range, battery condition and overrange. All indica­tors on the display are shown in Figure 2-2 and described as follows.
2-1
OPERATION
MODEL 169
a. Numerals.
maximum of 1999 counts.
pressed for most significant (l/Z) digit.
b. Decimal Point. The decimal point is indicated in three positions and implied when behind the least significant digit. The decimal point is properly positioned when a range pushbutton is pressed.
C. Overrange. A "1" followed by three blanked digits is displayed as an overrange indication for all ranges and functions, except the 1000 volt ranges which read beyond their maximum allowable input voltages. The minus sign and a decimal point
may also be displayed, if appropriate.
d. Polarity. For dc measurements the minus sign
(-) is displayed, positive is implied. The minus sign may flash on any function at zero since it is not suppressed. It is also normal for it to flash in ohms while the instrument is recovering from overrange.
e. Low Battery Indication. BAT is displayed when approximately 5% of battery life remains. indicates that the batteries should be replaced, however the Model 169 will continue to operate pro­perly for 40 hours if alkaline cells are installed, or 20 hours if carbon-zinc cells are installed.
f. Function and Range Annunciators.
cators provide a continuous display of the selected
funCtiOn and range, and allow direct reading of the
display. For example: current on the highest current range would read AC
IOOOmA.
Large 0.6" numerals display up to a
Leading zero is only sup-
This
These indi-
a displayed half-scale ac
Z-10. Front Panel Switches.
Z-11.
eleven pushbutton switches which are color coordi­nated to the front panel. These switches are shown in Figure 2-3 and described as follows.
that, when in, applies power to the DMM.
buttons provide selection of the five functions of the Model 169. that selects whether the function is ac or dc, in for ac or out for dc. annuncmtor is on.
is implied. The AC/DC pushbutton is not effective when ohms (0) is selected. buttons are push-on switches that select volts,
amperes and ohms, respectively. these pushbuttons all the way in selects the new function, turns on the proper annunciator and can­cels the previously selected function. Note that it
is possible, by forcing, to push two pushbuttons in at the same time. This will be evidenced by con­flicting range/function annunciators appearing, and the unwanted function can be canceled by pressing the desired function pushbutton again.
are push-on switches which select the new range and cancel the previous one. The decimal point is posi­tioned and the appropriate annunciator is turned on
when the range is selected. The 20M pushbutton is only used for ohms function. If it is inadvertently
selected with volts or current, contradicting annun-
ciators will appear.
Control of the Model 169 is provided by
a. ON/OFF Pushbutton. Push on - push off switch
b. Function Pushbuttons. The four function push-
AC/DC is a push on - push off switch
When ac is selected, the AC
When the annunciator is off, dc
The Y, A and " push-
Pressing one of
c.
Range Pushbuttons. The six range pushbuttons
2-2
NOTE
Extended exposure of the liquid crystal display to direct sunlight, or high tem­perature and high humidity conditions can cause the display to temporarily blacken.
Also, the display may become sluggish if
exposed to extremely cold temperatures. Even though recovery occurs at normal Operating temperatures. these conditions
should be avoided for maximum display life.
2-12.
Z-13. tions to the Model high impedance input and is normally connected to the point in the test circuit most removed from sig­nal or earth ground.
common or low inout.
Input connections.
Two banana jacks provide the input connec-
169.
The HI (red) jack is the
The LO (black) jack is the
I
MODEL 169
TABLE 2-1
Initial Function Check
OPERATION
STEP FUNCTION RANGE INPUT READING
1
2 ACV 200"V Shorted 00.0 ACmV 3 DC"
4
5
6
SELECTION DMM
200mV Shorted
A and n* 2k
A and n*
A and n* 2DOn and
200n
and Open 1.00.0 ,,An
2mA
open
open .I00 mAkn
Open 1.0.0 uAfl Display reads 20mA current shunt on 200n range.
DISPLAY
000
00.0 "V
2olTd (100)
7
8 n 9 " 20DOk Open
2DM 2DOOk Shorted
Shorted
*Press both pushbuttons simultaneously.
0.00 M" 000 k"
1
kn Overrange indication, leading 1 with all other
REMARKS
Power on, all other switches out. (-1 mdy flash.
Display reads 2mA current shunt on 2kn range. Checks current fuse.
Display reads 2mA current shunt on 200n range.
(approx. loon)
digits blanked.
SET PWER Q4
@SELECT F"NCTI(N
FIGURE 2-3. operating COntPOlS.
@SELECT
RANGE
@CONNECT
SOJRCE
IEZ302,A
z-3
OPERATION
MODEL 169
Z-18. OPERATING INSTRUCTIONS.
WARNING
t
To avoid electrical shock and/or instru­ment damage, do not connect the LO input to any source of more than 1400 volts
(peak) above earth ground.
2-14. Tilt Bail/Handle.
2-15. The 16 position tilt bail/handle allows you to position the instrument for the best viewing angle of the display.
2-16.
Z-17.
have become familiar with your OMM, the initial checkout procedure given in Table 2-1 should be per­formed. This procedure provides d quick and easy functional check of the DMM. Only test leads are
required. ciators to appear, except for low battery (BAT). All decimal point locations are checked. All digit segments are checked, except for the horizontal cen­ter segment on the last three digits. minus sign (-) is not suppressed on any function, it should periodically flash on one or snore of the
funcrions at zero with the input shorted. By
selecting two functions and/or ranges in the test, the current fuse is checked and d rough indication of the condition of the respective current shunt and
range resistor is obtained.
not intended to check the DMM accuracy, and thus, the indicated readings in the table are nominal with
*5 digits allowed. If it is desired to check the DMM accuracy specifications, perform the Performance Verification, Section 3. Check is begun with power on and all other push­buttons wt.
necessary to press unselected function and range
pushbuttons slightly to cancel previous selections. After completing the functional check, operate the Model 169 in accordance with the Operating Instruc­tions beginning with Paragraph 2-18.
INITIAL FUNCTIONAL CHECK.
Now that the batteries are installed and you
The check is designed to cause all annun-
Since the
However, this test is
The Initial Functional
To obtain this condition it may be
2-19.
Model there are individual instructions provided in this section which describe how to nakt specific function meas"rements. DMM as follows.
button. useable immediately to rated accuracy. ambient temperature is within 18-28°C range, but the DMM has been subjected to temperature extremes, up to I hour may be required to obtain rated accuracy.
pushbuttons.
range pushbutton.
the medsupement. should be used ds required.
The basic operating instructions for the
169
DMM are outlined below. In addition,
Refer to Figure 2-3 and operate the
CAUTION
A
Do not exceed the maximum input limits given in Table 2-2.
a.
Turn on power by depressing the ON/OFF push-
If the instrument is within I&28“C, it is
If
the
b. Select the function with the AC/DC, V, A or n
c. Select the range by depressing the appropriate
d. Connect the source to the INPUT jacks and make
Accessories described in Section 4
TABLE 2-2
Sumnary of Maximum Inputs
-
MAXIMUM
INPUT
1OOOV DC, 1400V Peak AC
350V rms continuous, 15
sec. maximum above 350V
IOOOV rms, 1400 peak,
lO%.Hz
ZA, 250V DC or rms (fuse
protected)
2-4
I
MODEL 169
OPERATION
Z-20. Z-21.
digit to 1000 volts.
is 1999 with appropriate decimal point. Polarity is automatically sensed and the minus sign (-) will be displayed, if applicable. a "1" and three blanked digits, except on the 1OOOV range.
beyond the maximum allowable input voltage. Maximum allowable input: 1OOOV dc, 1400 peak ac. Use the Model 169 to measure dc voltage as follows:
select dc volts with the V and AC/DC pushbuttons.
The "V" annunciator should appear.
able.
pushbutton.
2QOmV range is selected.
INPUT HI and LO jacks; observe the voltage reading
on the display.
2-22.
2-23. The Model 169 reads ac voltages from lOOuV/
digit to 1000 volts.
responding, calibrated to read the root mean square value of a sine wave within a frequency of 45Hz to 5kHz. decimal point. lowed by three blanked digits, except on the IOOOV range. beyond the maximum allowable input voltage. Maximum allowable input voltage: 106V.Hz maximum. ac voltage as follows.
OC VOLTAGE MEASUREMENT.
The Model
On the 1OOOV range, the display can read
00 not exceed the maximum allowable input voltage. Instrument damage may occur.
Turn on power with the ON/OFF pushbutton;
a.
Select the range from the five ranges avail-
b.
The decimal point is positioned by the range
c. Connect the signal to be measured between the
AC VOLTAGE MEASUREMENT.
The maximum reading is 1999 with appropriate
On the 1OOOV range, the instrument can read
169 reads dc voltage from 1OOpV/
The maximum displayed reading
Overrange is indicated by
CAUTION
A
The 'm' annunciator will appear if
The instrument is average
Overrange is indicated by a "1" fol-
1ooov Pm,
Use the Model
1400V peak,
169 to measure
A
CAUTION
00 not exceed the maximum allowable input
voltage.
Turn on power With the ON/OFF pushbutton;
a.
select dc volts with the V and AC/DC pushbuttons.
Select the range from the five ranges avail-
b. able. pushbuttons. 200mV range selected.
INPUT Hi and LO jacks; observe the voltaqe readin on the display.
2-24.
2-25.
lOOnA/digit to 2000 milliamperes (ZA). For ac mea­surements the instrument is average rsponding, cali­brated to read the root mean square value of d sine wave of d frequency within 4%~ to mum reading is 1999. with appropriate decimal point. Overrange is indicated by d blanked digits. vided by ZA, 250V dc or rms, quick-blow fuse. the Model 169 to measure, ac or dc current as follows.
The decimal point is positioned by the range
C. Connect the signal to be measured between the
AC OR DC CURRENT HEASUREIIENT.
The Model
00 not install .3 larger capacity current
fuse than the one originally supplied (ZA,
25OV).
Turn on power with the ON/OFF pushbutton.
a.
Select ac or dc with the AC/DC pushbutton; depress A pushbutton. with the "AC" annunicator if dc was selected.
instrument damage may occur.
The "rn' annunciator will appear if
169
reads ac and dc currents from
1kHz.
“1”
followed by three
Input overload protection is pro-
CAUTION
A
Instrument damage may occur'.
The "A" annunciator should appear along
The maxi-
Use
2-5
I
OPERATION
Select the range from the five ranges avail-
b.
able.
pushbutton. 200uA range was selected; otherwise, the W annun­ciator will appear.
voltage drop across the current shunt in the instru­ment (called voltage burden) can affect the circuit
under test if it has a low source voltage.
voltage burden can be reduced by selecting the
highest current range that will resolve the desired meas"rement.
HI and LO jacks; observe the current reading on the
display.
The decimal point is positioned by the range
The '11" annunciator will appear if
It should be noted that the
C. Connect the signal to be measured to the INPUT
2-26.
2-27. digit to 20 megohms. with appropriate decimal point. cated by a Maximum Allowable Input: 3OOV dc or rms. Use the Model 169 to measure resistance as follows.
RESISTANCE (n) MEASURENENl
The Model 169 reads resistance from 1OOnW
The maximum reading is 1999
Overrange is indi-
"1" followed by three blanked digits.
CAUTION
A
Turn the test circuit off and discharge
all capacitors before attempting in­circuit measurements. present in the circuit if either the minus sign is displayed, along with a steady displayed number other than zero, OP the
reading changes more than two digits when
the leads are reversed.
A voltage may be
This
MODEL 169
c.
Read the resistance an the display. ZOOn range is selected, disconnect the test leads at the test circuit, short them together and subtract the reading obtained from the reading obtained above.
NOTE
As previously mentioned far in-circuit meas"rements, the presence of a voltage will cause a change in the reading when the leads are reversed. The presence of a forward biased semiconductor junction will also cause a significant error, and there-
fore, it is a good practice to reverse the leads and make comparative measurements of in-circuit resistances. should be noted that the bottom two re­sistance ranges will not turn on most semiconductors, except in an overrange condition.
d. Diode Tests. Any of the top four ohms ranges
will turn on semiconductor junctions, with an
on-scale reading. but the 20Kn range is recommended for out-of-circuit tests of semiconductors because its test current is higher. To test a junction, first forward bias it and then reverse bias it while observing the display reading for both conditions. The HI jack is positive. Compare the results as follows:
INDICATION
In addition, it
If the
JUNCTION CONDITION
Turn on power and depress the 0 pushbutton.
a.
The "n" annunciator should appear.
b. Connect the circuit to be measured to INPUT HI and LO and select the desired range from the six ranges available. by the range pushbutton and the applicable range annunciator should appear.
2-6
The decimal point is positioned
1. Overrange. both Polarities
2. Low reading, both polarities
3. Overrange when reverse biased and an on-range reading when forward biased
OPEN
SHORTEO
PROBABLY GOOD
MODEL 169
PERFORMANCE VERIFICATION
SECTION 3.
3-l.
GENERAL.
3-2. Performance verification inay be performed upon
receipt of the instrument to ensure that no damage
or misadjustment has occured during transit.
fication may also be performed whenever there is question of the instrument's accuracy, and following calibration, if desired.
NOTE 3-5.
For instruments that are still under war­ranty (less than 12 months since date of shipment), contact your Keithley repre­sentative or the factory immediately if the instrument falls outside of specifica­tions as determined by the Performance Verification Procedure in paragraph 3-7.
PERFORMANCE VERIFICATION
Veri-
RECDHMENOEO TEST EQUIPMENT.
3-3.
3-4. Recommended test equipment for perfornance
verification (and calibration) is listed in Table
3-l. Alternate test equipment may be used.
ever, if the accuracy of the alternate test equip-
ment is not at least 10 times better than the
instrument specifications, additional allowance must
bc made in the readings obtained.
ENVIRONMENTAL CONDITIONS.
3-6. All measurements should be made at an ambient temperature within the range of la0 to 28OC (65" to BZ'F), and a relative humidity of less than 80%.
3-7.
PERFORWNCE VERIFICATION PROCEDURE.
3-B. basic accuracy of the Model resistance and current measurements.
Use the following procedures to verify the
169 for vo:tage,
HOW-
ITEM
-
A
B
C
0
E
TABLE 3-l
Reformended Test Equipment for Performance Verification.
DESCRIPTION
DC Calibrator
SPECIFICATIONS
O.lV, lV, lOV, IOOV
Fluke
1ooov , to .005%
AC Calibrator
AC Calibrator/Amplifier
Decade Resistor
O.l”, 1”. lo”, 1oov
*0.0223
1000" 20.04%
1.9k", 19kn,
190kn,
H-P
H-P
ES1
1.9Hfl, lOMn,
?0.02%
Current Calibrator
lOOpA, lmA, lOmA.
IOOmA, 1A. to.034
Valhalla
IMFR .
MODEL
341A
745A
745R/746A
0862
2500
3-l
PERFORMANCE VERIFICATION
I
MODEL 169
NOTE
Performance verification should be per­formed by qualified personnel using accurate and reliable test equipment.
3-9. Initial Conditions.
3-10. Before beginning the verification procedure the instrument must !meet the following conditions:
a. If the instrument has been subjected to extremes of temperature, allow internal temperature to stabilize for one (1) hour minimum at the environmental conditions specified in paragraph 3-5.
b. Turn the instrument on and check for low battery indication. If low battery indicator (BAT)
is on, remove and replace the batteries with fresh
O”L?S.
t
Some procedures require the use of high
voltage. Take care to prevent contact with live circuits which could cause electrical shock resulting in injury or death.
3-11.
DC Volts Checkout.
WARNING
TABLE 3-2
DC voltage Perfornance Check
Applied
Range
ZOOmV
20 v
200 v
1000 v
3-12.
a. Select ac voltage readings with the AC/DC and
V pushbuttons.
b. Connect the AC Calibrator (Item B, Table 3-1)
to the OMM.
c. Set the DMM to the 200 mV range and apply
mV ac to the OMM. The reading must be within the
limits specified in Table 3-3.
d. Select the 2, 20 and 200 volt ranges and apply the required voltages as specified in Table 3-3. Verify that the readings are within specifications.
Voltage
100.00 mv
2v
1.0000 v
10.000 v
100.00 v
1000.0 v
AC Volts Checkout.
Set the calibrator frequency to
Allowable Reading at
180 to 2rc
99.6 to 100.4
.996 to 1.004
9.96 to 10.04
99.6 to 100.4 996 to 1004
1kHz.
100
a. Select dc voltage readings with the AC/DC and
V pushbuttons.
b. Connect the DC Calibrator (Item A, Table 3.1)
to the instrument.
c.
Select the 200mV range, and apply positive 1OOmVdc to the OMM. The reading must be within the limits specified in Table 3-2.
d.
Select each remaining range and apply the
required voltage as specified in Table 3-2, verify
that the reading is within specifications.
e. Repeat all checks with negative voltage.
3-2
e.
To check the Calibrator Amplifier (Item C, Table 3-l) to the output of 1000 volts ac rms and verify that the OMM
reading is within the specified limits.
AC Voltage Performance Check
Applied
Range
200mV
20 v
200 v
1000 v
2v
Voltage
100.00 mv
1.0000 v
10.000 v
100.00 v
1000.0 v
1000
volt range, connect the AC
TABLE 3-3
Allowable Reading at
18’ to 28’C
98.7 to 101.3
.987 to 1.013 V
9.87 to 10.13 V
98.7 to 101.5
987 to 1013
mV
V
V
I
MODEL 169
PERFORMANCE VERIFICATION
3-13. Resistance Checkout.
a. Select resistance readings by pressing the n
pushbutton.
b. Select the 2000 range.
C. Connect the decade resistor (Item 0, Table
3-1) to the OMM.
d. Set the decade resistor to zero and measure the resistance of the test leads. reading from the display in all of the following steps.
Set the decade resistance to
e. that the reading is within the limits specified in
Table 3-4.
f. Continue the resistance check by selecting the
next range and measuring the next resistance as specified in Table 3-4. within specifications.
Resistance Perfonance Check
Ran e
Resistance
Verify that the reading is
Test each item in the table.
TABLE 3-4
Allowable Reading at
Subtract this
190.0.
18’ to 28’C
Verify
3-14. DC Current Checkout.
a. Select dc current readings with the AC/DC and
A pushbuttons.
b. Connect the dc current source (Item E. Table
3-1) to the
C.
lOO.OOpA to the OHM. The reading must be within the
limits in Table 3-5.
d. Select each range and apply the required cuv­rent as specified in Table 3-5. reading is within specifications.
3-15. ANALYSIS.
3-16. limits given in Tables 3-2 through 3-5. either troubleshooting and repair or described in Section 6, may be required. recalibration will only be the indicated solution when the instrument is out of speciflcbtlon on ~lofe than one function (except for the higher ranges of DC volts which are tht most critical). instrument is still under warranty, refer to the Note in Paragraph 3-2.
DMM.
Select the 200uA range and apply a current of
Verify that the
If the instiwwnt is out of the specified
calibration, a5
NO~Il~lly,
If the
kn
kn
kn
190.0 ”
1.9OOkn
19.0Okn 190
.Okn
200 n
r--t--
2
20
200
I
189.1
to
190.9 n
1.695 to 1.905 KZI
18.95 to 19.05
189.5 to 190.5
Kn
Kn
1895 to 1905 Kn
9.19 to 10.21
Mn
DC Current Performance Check
Applied
Range current
PA
nA
mA
mA
mA
100.00
1 .OOOO
10.000
100.00
1000.0
200
2
20
200
1000
TABLE 3-5
Allowable Reading at
18' to 28°C
"A
mA
mA
mA
mA
99.1 to 100.9
.991 to 1.009 mA
9.91 to 10.09 mA
99.1 to 100.9 mA
991 to 1009 mA
uA
3-3
MODEL 169
ACCESSORIES
SECTION 4.
4-l. GENERAL.
4-2.
This section describes the various accessories and options available for use with the Model 169 OMM .
4-3.
MODEL 1600 HIGH VOLTAGE PROBE.
4-4. The Model 1600 High Voltage Probe (shown in Figure 4-l) extends the measurable dc voltage range up to 40 kilovolts. It has a 1OOO:l division ratio, so that a reading of 1 volt on the DMM corresponds to 1 kilovolt (1000 volts). To "se the probe, select DCV and the required range, connect the high voltage probe banana plug to the instrument, connect
the alligator clip to soui-ce low, and touch the
probe tip to source high.
WARNING
t
Be sure the alligator clip is connected to source low before touching probe tip to souw? high. A shock hazard or damage to
instrument may result.
Specifications
Voltage Range:
input Resistance: Division Ratio: 1OOO:l. Ratio Accuracy:
t1.5% at 25kV, decreasing to Q.O% at 20k" and 30k", e3.04 at 1OkV and 40kV, and
t4.0% at
Ratio Stability: Heating Effects: Self-heating due to application
of high voltage for period in excess of 1 minute will cause a maximum of 0.2% additional error at 40kV (error is less at lower voltage).
IkV.
0 to 40,000
1000 megohms.
?O.Ol%
per Y; *0.1x per year.
volts DC.
ACCESSORIES
4-5.
MODEL 1651 50-AMPERE SHUNT
4-6.
Figure 4-2) permits current measurements of up to 50 amperes ac or dc.
0.001 ohm ?l%, so that a 50-ampere current vill car­respond to a reading of 50 millivolts (0.0500 volt). Set the "NM to ACV or DCV and select the required rangc. TO use the shunt,
nished with the shunt from the shunt screw terminals to the OMM input terminals. Use separate leads (not furnished) to connect the source to the hex head bolts. Be sure to use leads with a capacity of 50 amperes, or as needed.
4-7.
4-8.
leads with banana plugs at one end and spring-action clip-on probes at the other end. TO use, insert the banana plugs into the "NM and attach the prober to the source.
4-9.
4-10.
permits measurement of dc voltages dt frequencies of 20 kilohertz to 100 megahertz. Connect the probe to the input terminals and select 3CV and the appro­priate range.
Specifications
The Model 1651 50-Ampere Shunt (shown in
The shunt has a resistance of
connect the leads fur-
MODEL
1681
CLIP-ON TEST LEAD SET
This set (shown in Figure 4-3) contains two
MODEL 1682 RF PROBE
The Model 1682 RF Probe (shown in Figure 4-1)
Voltage Range: Transfer Accuracy: i0.5dS. 1OflkHr to 10OWr peak
responding calibrated in rms of d sinewave.
Input Impedance: 4 megohm shunted by 3pF. Maximum Allowable input: 30V rms AC, ZOOV DC. Accessories Supplied: straight-tip, hook tip,
ground clip, hi adapter, banan; plug adapter.
0.25 to 30 volts nns.
4-1
ACCESSORIES
I
MODEL 169
4-11.
4-12. test leads and 12 screw-in adapter tips. The tips consist of two alligator tips with boots, two banana
plugs, two needle tips with chucks, two spade lugs
and four heavy duty tip plugs which permit con-
nection of the DMM to virtually any source within its measurement range.
4-13. MODEL 1684 CARRYING CASE.
4-14. The Model 1684 Carrying Case (shown in Figure 4-4) is a hard vinyl case with a fitted foam insert to help protect the Model 169 from damage. There is also room in the case for the Instruction manual and
small accessories
4-15.
4-16.
(shown in Figure 4-2) permits measurement of ac
current by clamping around a single conductor,
eliminating the need to interrupt the current path. Plug the ac current probe into the OMM and select
MODEL 1683 UNIVERSAL TEST LEA0 KIT.
This kit (shown in Figure 4-3) contains two
MODEL 1685 CLAMP-ON AC CURRENT PROBE.
The Model 1685 Clamp-On AC Current Probe
ACV and the appropriate range. The DMM will display
0.1
volt rms per ampere.
4-17. MODEL
4-18. The Model
(shown in Figure 4-3) consists of two .9lm (36 inches) test leads with probe tips terminated in banana plugs.
4-19.
4-20.
parts that will mdintain up to one year. The parts we listed in Table 7-l of Section 7, Replaceable Parts.
4-21. MODELS
4-22.
permit mounting one or two Model
19-inch rack for convenient viewing.
1691
GENERAL PURPOSE TEST LEAD SET.
1691
General Purpose Test Lead Set
MODEL
1699
SPARE PARTS KIT.
The Model 1699 contains a complement of spare
10
Model
169
DMMs for
1010
and 1017 RACK HOUNTING KITS.
The rack mounting kits (shown in Figure 4-4)
169 DMMs
in
a
FIGURE 4-l. High Voltage Probe and RF Probe.
MODEL 169
ACCESSORIES
MODEL 1651
50.AMPEKE SiiUIIT
CLAMP-ON AC CURRENT PROBE
MODEL 1685
FIGURE 4-2.
Current Probe and Current Shunt.
MODEL
CLIP-Ok TCST LEAD SET
MODEL 1691
GE,,ER,,L PURPOSE TEST
1681
LEAD SET
MODEL 1683
UNIVERSAL TEST LEAD KIT
FIGURE 4-3.
Convenience Cable and Connectors.
4-3
ACCESSORIES MODEL 169
I
\- /’
MODEL 1684
CARRYING CASE
FIGURE 4-4. Carrying Case and Rack Mounting Kits.
“; :
;:I
:o i
SINGLE RACK I~lOIINTIllT; KIT
MODEL 1010
DUAL RACK ElOUNTIN(; KIT
I
4-4
I
MOOEL 169
TtlEORY OF OPERATION
I
I
SECTION 5.
5-1.
GENERAL.
5-2. This section contains circuit descriptions for
the Model 169 OMM. An overall block diagram of sig-
nal flow is provided in Figur‘e 5-I. schematic diagram, drawing 301800, is contained in Section 7 of this m~nudl.
5-3.
OVERALL OPERATION.
5-4.
As shown in Figure 5-1, the heart of the Model 169 is d single chip A/D converter with built-in liquid crystal display drivers. reading is the ratio of two floating input voltages to this converter. one voltage (Reference) is IOOmV. The other voltage (VINHI) is the unknown input cuwent 01‘ voltage
to the Model
the input signal conditioning circuitry to a dc Voltage between zero and ?200m". reading is then determined by the formula:
(YINHI +
also automatically determines the polarity of the input signal and detects whether an overrange condition exists displayed, plus is implied, and an ove~~ange is displayed by a I with the last three digits blanked. The annunciators ape displayed by the combination Of
range dnd function switching, and the decimal point is positioned by the range switching.
For the ohms function, d reference resistor is
5-5. placed in series with the unknown resistor and a
voltage is applied. through both resistors, the ratio of their voltages is the sdme as the ratio Of their resistances.
Thus, the voltage w1‘oss each resistor is measured
by the A/O converter and their ratio is displayed per the formula: IO00 (VR, i VRREFERENCE).
169
VREFERENCE).
On all functions except ohms,
which is converted and/or scaled by
The
(>1999
counts).
Since the Same current flows
The overall
The displayed
The displayed
1000
Conwrter
A/O
A minus sign is
THEORY OF OPERATION
5-6. SIGNAL CONDITIONING.
5-l. AC/DC Voltage Measure,wnts.
5.8. For voltage measurements, ds shown in Figure 5-2, the input is divided by 10,000 by the IO megohm resistive divider. resulting output from the divider is d 0 to 200mV
voltage. direct to the A/D co"vel‘teP along with the IOOmY reference. gain or attenuator stages are used. the dc dccurdcy
is determined primarily by the precision resistor
accuracies (actually their ratios) and the reference adJ"stment.
5-9. For an x input, the output of the resistive divider is first applied to the AC Converter. AC conver‘ter is a half wave rectifier with IOOUV
resolution and sufficient gain to provide d positive
dc output voltage equivalent to the nns value of d sinusoidal input. lOOmVac, the output will be IOOmVdc. The important gain determining elements of the converter dw
precision resistors so that d gain adjustment is not required.
5-10.
5-11.
5-3, the appropriate shunt resistor is placed across the OMM input in accordance with the selected range. As with voltage medsurcments, the voltage drop across the current shunt is designed to be 2OOmV for a full scale input current on any range. Therefore, after the input current is converted to d voltage, the meaSUPement process for current is identical to that for voltage measurements.
Current meaSurCments is primarily determined by the
ilccUrXies of the precision shunt resistors and the reference adjustment. ac accuracy along with any error contributed by the
AC Converter.
For a dc input, this voltage is applied
It should be noted that, since no active
Thus, for sinusoidal input of
AC/DC Current lkasuremnt.
FOT Current medsurements, its Shown in Figure
These two items also affect
1, 10, 100, 1000, or
Accuracy for dc
The
The
5-1
THEORY OF OPERATION
MOOEL
169
5-12.
5-13. the voltage drop across the unknown resistance to the voltage drop of an internal reference resistor as shown in Figure 5-4. For this ratio measurement technique, d voltage is connected to the input divider used for voltage measurements and the range switching selects a l/2 of full scale value refer­ence resistance to be placed in series with the unknown resistor.
pared by the A/D converter and displayed on the LCD.
A 2 kilohm thermistor (RTlOl) provides overvoltage
input protection for the low resistance ranges and
it is in series with the reference resistor and the unknown resistor on all but the 20 megohm range. For this reason, the +V (t2.8V nominal) top of the
reference is selected on the 200 ohm range. On all other ranges tl.ZV is used. Since a ratio technique
is used, determine the measurements, and no adjustments are necessary.
5-14. A/D CONVERTER.
5-15.
verter and display drivers ape contained on a single integrated circuit (U102). This chip also contains an oscillator and the regulators that are used to establish the analog and digital grounds. components are used to establish the timing and reference levels required far A/O operation. employs the dual slope principle of A/D conversion. This method provides d ~asurement cycle consisting of three consecutive time periods, autozero. signal integrate and reference integrate (read). for these periods, and thus the measurement period,
is a function of the oscillator frequency. oscillator components are selected to provide 40 kHz which results in a period for best rejection of 50Hz and 60Hz. To begin each measurement cycle. the autozero period is set to a minimum of period, the reference voltage input to the A/D is
Resistance
Resistance measurements are made by comparing
actual values of the reference resistors
As shown in Schematic
Measurements.
The voltage drops are then com-
primary accuracy for
301800,
100
millisecond signal integrate
100
milliseconds.
resistance
the A/O con-
External
During this
u102
Timing
The
stored on capacitor referenced to analog cornnon. is then stored on capacitor Cl02 for use during the remainder of the measurfment cycle, if any amplifier offset exists. The second phase of the measurement cycle is the 100mS signal integration period as previously mentioned. differential voltage between IN HI and IN LO is integrated. of the integrated signal is also determined. final phase is reference integrate, or read. During this period, reference capacitor will be connected with the correct polarity to return the integrator output to the autozero level. output to return to this level is proportional to the unknown input signal. from zero to 200 milliseconds. corresponding to displayed counts of zero to 2000 full scale as determined by input signals less than full cmwrtw rewrts to wtozero far the remainder of the reference integrate period. measurement cycle is 400ms. which results in a display update rate of 2-l/2 per second.
5-16.
5-17. associated circuitry is shown in Schematic As previously mentioned, the seven segment decoders and drivers are provided by the A/D converter chip u102. on the output lines to the LCD. driven by a square wave having the same amplitude and frequency as the BACKPLANE line (nominally 0 to
t5V at 50Hz). ments are driven out of phase with BACKPLANE the segments are ON. segments are OFF. points are turned ON and OFF similarly, with the phasing of their lines controlled by the range and function switches through logic gates. A swnary of annunciator switching is given in Table 5-1.
At the end of this period, the polarity
DISPLAY CIRCUITS.
The liquid crystal display (LCD) and its
The digitized measurement data is presented
Cl11
and the A/D converter is
A correction voltage
During this period, the
The
internal circuitry ensures that the
The time required for the
The read period can be
1000 (VI, + VREF).
scale,
Thus,
a complete
FW
the A/D
301800.
These lines are
When the lines to the display seg-
Conversely. when in-phase the
The annunciators and decimal
5-2
MODEL 169
THEORY OF OPERATION
5-18. 5-19.
lated and calibrated) to the A/D converter for
voltage and current measurements. approximately 2.8V to the 2008~ range and approxi­mately 1.2V to all other n ranges during resistance meas"rements. The heart of the reference source is a band-gap reference (VRlOl) that regulates the
voltage to a value between
approximately SOOvA. applied to the reference source depends on the
internal regulator in the A/O converter which establishes analog ground nominally 2.8 volts below vt. Thus, the value of RlOZG is selected to provide
a minimum current of 400uA to "8101 to ensure a good temperature coefficient. tion adjustment, adjustment span and are stable precision resistors to ensure the 1 year calibration cycle.
5-20.
5-21.
tor switch (QlO2) which compares the negative supply to digital cornnon (ground). 9102 is set for a nomi­nal supply voltage of 6.9 volts. (Actual battery voltage will be about 7.2 volts due to the drop of CRlO5.) In the Model 169, the primary limit to bat­tery usefulness is the integrator swing on the high ohms ranges with a full scale input. For this rea-
son, the low battery indicator (RAT) should turn on with a battery voltage of 6.5 volts, or above, for the proper functioning of the A/O Converter and the
low battery indicator.
5-22.
5-23. Effective input protection circuitry has been designed into the Model 169 which provides protec­tion against accidental input overloads on all functions and ranges. in the following paragraphs for each function.
REFERENCE SOURCE.
The reference source supplies lOOmV (regu-
It also provides
1.20 and 1.25 volts at
Recall that the voltage
RlOl
provides the calibra-
while R1020 and R102E limit the
LOW BATTERY DETECTOR.
The low battery detector is a simple transis-
INPUT PROTECTION.
This circuitry is described
5-24. Voltage Protectlo".
5-25. during dc voltage measurements by limiting the cur­rent to *ImA.
5-26. tected by using R107 for current limiting and CR103 and CR104 to clamp the voltage to digital cornnon and vt. through ClD7,
5-27. current Protection.
5-28. opens for input current overloads with up to 250 Volts.
Clamp the input voltage to about 1 volt until F101
"PWS.
5-29. Ohms Protection.
5-30. basic components used to provide protection during
resistance meas"reme"ts. leakage zener to clamp any unintentially applied
voltage to ground. RTlOl is a positive temperature Coefficient thermistor which bears the applied volt­age by increasing its resistance. This timits cur-
rent to
of the A/O converter as in OC volts.
Voltage across 9101 can
tive tects REF HI by limiting the current to +lOOuA. Since tected by RlOZG on the 200n range and by R102F plus the divider resistance on the other ranges. negative input overloads the circuit operates the same, except to ANALOG COMMON.
R107 and K103 protect the A/D converter
For ac voltage medsuremeotn, UlOI is pro-
Overload currents are returned to anal09 cornnon
Cl13
and the AID converter.
On the higher current ranges,
On the lower current ranges, CR106 and CR107
Transistor
4101
and thermfstor RTlOl are the
9101
protect
supply.
YRlOl has a current
QlOl.
R102C protects REF LO dnd RlOKE pro-
qlO1
RlO3 is used to protect IN Hi
be 10 volts above the posi-
limit of IOmA. it is pro-
is then forward biased and clamps
FlOl
(2A. 25OVl
is used as d low
Since the
For
5-3
THEORY OF OPERATION
MOUEL 169
TABLE 5-1
Annunciator Truth Table
\NN!JNCIATOR
BAT
AC
DPl
DPZ
DP3
V
a
k
SWITCH
Q102
DC/AC
R
2K
20K or 20M
200
or ZOOK
V
a
200
and 20M
R
CONDITIOh
OFF
AC OFF
ON
ON
ON
ON
ON
OFF
ON
20M
A
A and 200
A
1. 200
V and 200
2.
I
5-4
ON
ON
ON
ON
OFF
ON
MODEL
169
THEORY OF OPERATION
INPUT
HI>
NOTES: 1.
n
TO CKTS
Output voltage from current shunts
and voltage attenuator are zen, to
200mVAC and zero to t200mVdc.
,-, +2.8V
AND ANNUNCIATOR
200n
DIGIT SECML,,T
DATA
DISPLAY
t
2. Output of AC Converter is zero to +2OOmVdc.
FIGURE 5-1.
Model 169 DMM Overall Signal Flow Block Diagram.
THEORY OF OPERATION
MODEL 169
DECADE
RESISTIVE
DIVIDER
LO
9M
900K
90K
+
FIGURE 5-2.
0v to t200mV
Simplified AC/LX Voltage Measurement.
I
5-6
HI
INPUI 0
MODEL 169
CURRENT
SHUllTS
zooun
900
2oA
90
2omA
LO>---
200mR
2oooaiA
0
RANGE
S~IITCHING
FIGURE 5-3.
9
.9
.l
Simplified AC/DC Current Meaurement.
THEORY OF OPERATION
REFERENCE
SOURCE
MODEL 169
I
100.1
90K
900K
9M
9K
200
2K
20K
200K
2000K
20M
RANGE
SWITCHING
0
.
RTlOl 2K
,C c
I
5-8
RX
HI
-1n A
REFERENCE
RESISTORS
FIGURE 5-4. Simplified Resistance Measurement.
I
MODEL 169 MAINTENANCE
SECTION 6. MAINTENANCE
6-l. GENERAL.
6-2. This section contains information necessary to maintain
the Model 169 DMM. AdjustmenVcalibration, troubleshooting,
battery replacement, fuse replacement and LCD replacement procedures are provided.
6-3. CALIBRATION.
6-4. Calibration should be performed yearly (every 12 months) or whenever Performance Verification (see Section 3) indicates
that the Model 169 is out of specifications. If the Calibration Pro-
cedure cannot be periormed properly, r&r to Troubleshooting information in this section or contact your Keithley representa-
tive or the factory.
NOTE Calibration should be performed by qualified Peru sonnel using accurate and reliable equipment.
6-5. Recommended Calibration Equipment.
6-6. The Fluke Model 341A DC Calibrator is recommended for
calibrating the Model 169. Alternate calibration equipment may be used. However, the accuracy of the alternate 190 mVDC
source must be .025% or better.
a. Turn off power and disconnect ail leads from the 169 DMM
Input.
b. Loosen the four screwy in the bottom of lhe DMM until lhe
threads are disengaged. NOTE: These screws are not rep tained and they will fall out when the DMM 1s returned to the upright position.
c. Hold the top and bottom cover together to prevent ihew
separation and turn the DMM over lo !ls normal p&Ivan. Remove the top cover.
d. Turn power on by depressing the ON/OFF pushbutlon.
and verify that the low battery indicator (BAT) does not apt pear. (NOTE: It is recommended that fresh batteries be ins stalled at this time to provide the longest possible use of the Model 169 before additional maintenance is required. Refer to Installation of Batteries, Paragraph 2-6).
6-11. Calibration Adjustment.
6-12. Only two adjustments are needed to calibrate the Model
169. Proceed as follows:
a. Place the Model 169 on the 200mVDC range.
6-7. Environmental Conditions.
6-8. Calibration should be performed under environmental con-
ditions having an ambient temperature of 23” i2”C and a rela­tive humidity lessthan 70%. If the Model 169 has been exposed to extremes of temperature or humidity, allow at least one hour for the instrument to stabilize.
6-9. Case Cover Removal.
6-10. To gain access to the calibration pot and circuit compo­nents, remove the case cover as follows:
WARNING Disconnect all te?,t circuits from the INPUT connec­tions of the Model 169 before removing the case CO”H.
b. Apply-190.0mVDCandadjusf RlOl (see Figure&l)far a
reading of -190.0.
c. Apply +19O,OmVDC and note reading.
d. Adjust I?1 01 for a reading halfway between the noted read-
ing and 190.0 (e.g., if noted reading is 190.2, adjust Al01
for a reading of 190.1).
e. Place the Model 169 on the ZVAC range.
f. Apply 1 .OVAC at 5kHz and adjust C1 17 for a reading be-
tween 1.025 and 1.035VAC. (The trimmer capacitor is lo­cated at the center of the PC board, between the ribbon ca­bles.)
6-13. This completes calibration of the Model 169. Reinstall the DMM top cover and verify performance as outlined in Section 3.
MAINTENANCE
MODEL 169
SPACERS (21
CAL I
BRA; I ON ADJUSTMENT Rlcl
6-14.
6-15.
FIGURE 6-l.
TROUBLESHOOTING.
The following troubleshooting instructions
Calibration Adjustment Location.
are intended for qualified personnel having a basic understanding of analog and digital
electronic
principles and components used in a precision electronic test instrument.
Instructions have been
written to assist in isolating the defective circuit or Subcircuit. component has been left to the technician. the case cover ds explained in Paragraph 6-9 to gain
access to
locations.
Isolation of the specific defective
circuit
components
and test
Remove
point
6-16. Special Handling of Static Sensitive Devices.
6.17.
CMOS devices a-e designed to function at very high impedance levels for low power consumption. For this reason, a normal static charge build up on
your person or clothing can be sufficient to destroy
these devices.
The following steps list the static
sensitive devices in your Model 169 and provide instruction on how to avoid damaging them when they
must be removed/replaced.
6-Z
I
MODEL 169
MAINTENANCE
a. Static sensitive devices:
Kfithley
Part Number
LSI-22
IC-102 IC-226 IC-228
OZ-62
b.
The above integrated circuits should be handled and transported only in protective con­tainers. tubes or static protective foam. Keep the devices in their original containers until ready for use.
c. tainers only at d properly grounded work bench or table, and only after grounding yourself by using a wrist strap.
d. Handle the devices only by the body.
touch the pins.
e. Any printed circuit board into which a device
is to be inserted must a'150 be grounded to the bench
or table.
f. Use only anti-static type solder suckers.
g. Use only grounded tip soldering irons.
h. After soldering the device into the board, or
properly inserting it into the mating receptacle,
the device is adequately protected and normal
handling can be resumed.
Typically they will be rcceivcd in metal
Remove the devices from their protective con-
Reference
Designation
u102
u103
IJ104, u105
“106
VRlOl
DO not
6-21.
Display Driver circuit should be verified before troubleshooting the signal conditioning circuits. Check these circuits per Tables 6-2 and 6-3. If it
is determined that the Liquid Crystal Display is defective, 6-31.
6-22.
6-23.
Converter circuitry.
Proper Operation of the A/D Converter and
replace per instructions in Paragraph
AC
Converter.
Problems with dc voltage may
Check this circuit per Table
involve
the AC
6-4.
6-24.
Circuit.
6-25.
may involve these Signal conditioning circuits. Check out the Reference Source and Ohms Circuit per
Tables 6-S and 6-6.
checked by measuring the resistance of each resistor individually with the instrument off and the push­buttons in the out position. Voltage Divider resistors are as follows:
Voltage Olvlder. Reference Source and Ohms
Problems with voltage or resistance ranges
The Voltage Divider can be
The tolerances of the
a. R108, 100.1n, 0.1%
b.
A109, 900n, 0.1%
c. RllO,
d.
e. H112, 900Kn,
f. R113, 9Mn, 0.08%
Rlll,
9Kn, 0.1%
90Kn, 0.1%
0.1%
6-18.
6-19. In general,
6-1,
and power supply are providing the correct voltages to the electronic components. that the battery needs replacing, refer to Paragraph 6-33, Battery Replacement.
6-20. A/O Converter and Oisplay Driver.
Power supply.
start troubleshooting with Table
Power Supply Checks, to verify that the battery
If it is determined
6-26. current Circuitry. 6-27.
check the current circuitry ds directed in the fol­lowing pangraphs. the current shunts are cornnon to both ac and dc cvr­rent,
current ranges if there is d fault in the current
shunt circuitry.
If problems occur with current readings.
It should be noted that, since
problems should exist on the same ac dad dc
6-3
MAINTENANCE
MODEL 169
6-28. simultaneously depressing the A and n pushbuttons while on the 2K range. good fuse is nominally .lOOmAKn. A blown fuse Will be indicated by an overranged display. current fuse is blown, replace per instructions in Paragraph 6-35.
6-29.
suring the resistance of each resistor Individually
with the instrument off and all the pushbuttons in
the out position. as follows:
The cuwent fuse can be checked wt by
The display reading for a
If the
The Current Shunts can be checked by "ea-
The tolerances of the shunts are
a. R114A, 0.9Q, 0.5%
b. R1146, 0.10, 0.5%
c. R115, 9n, 0.5%
d. R116, 90n, 0.1%
e. R117, 9OOfl. 0.1%
6-30.
and CR108) as follows:
resistor to the input of the Model 169.
or less.
6-31. LIQUID CRYSTAL DISPLAY REPLACEMENT.
b-32. Figure 6-4, to replace the display of the Model 169.
graph 6-9.
to the front panel.
careful of the LCII display (Keithley Part NO. DD-18)
as it will be loose and may fall out of the front panel.
Check the current protection diodes (CR107
a. Select ZOO,,AOC range.
Apply 5VAC through a IKn current limiting
b.
Clamping across the diodes must occur at tlV
c.
Perform the following procedure, referring to
Remove the case cover as explained in Para-
a.
b. Remove the three 318" pal nuts holding PC-509
Separate PC-509 from the front panel being
c.
6-4
I
FIGURE 6-2.
CONNECTOR
(CS-376-1)
LCD Assembly.
I
MODEL 169
TABLE 6-l
Power Supply Checks
STEP / ITEM/COMPONENT REQUIRED CONDITION' REMARKS
1
Battery Voltage 7.5V min.
2 CR105 0.5V drop max.
MAINTENANCE
3 UlO2, pin 40 and 15 7.OV min. 4 Cl07 2.4 to 3.2V output Of u102
0.8 to 2.6V output of u102
!
'Above checks with fresh batteries and power switch ON.
TABLE 6-2
A/O Converter
STEP ITEM/COMPONENT
1 2 Display Short input, 00.0 tld 3 Refewnce (UlO2, pin 5)
REWIRED CONDITIONS' REMARKS
power ON, ZOO","DC ,uan9e
Nominally 98mV using
voltmeter with 1OMn Input Impedance
Supply to u102
(Analog Gnd)
(Oigital Gnd)
I
rero Check
IHigh ImpeddnCP reference
4 Ratio Test Short "102 pin 5 to
5 Display
6 UlO6, Pin 3
7 External Voltage source
8 u102 Pin 14 to Input Lo
*Measurements made with respect to input lo
**Non-symmetrical in most cases
pin 10
100.0
*2d
40kHz f6kHz**
Approx. 5Vp-p
Apply +190.0mVAC Calibration
See waveform (Figure ,ntegrator
6-3) Check
connects refer­ence to Input
If steps 3 and 5 are Correct AID Converter
is functioning properly
Clock check
poi nr
6-5
MAINTENANCE
MODEL 169
190mVDC Input
FIGURE 6-3.
STEP
1 2 "102 pin 40 to pin 4 4.0 to 6.0 Volts 3 "102 pin 3 to Input Lo 4OkHz +6kHz Clack Check
4 U102 pin 20 to Input Lo
5 UlO2 - any display Approx. 50H.7 Square
6 U104, U105, UlO6,
ITEMICONPONENT
driver line wave, approx. svp-p backplane for
decimal point and wave, approx. svp-p
annunciator driver lines
Integrator Waveform.
TABLE 6-3
Display Driver
REQUIRED CONDITIONS REMARKS
Power ON
Digital Supply
Apprax. S'ip-p
Approx. 50Hz Square
wave, approx. svp-p
In phase with
seqment Off, out of phase with backplane
for segment on
Approx. 50Hz square
6-6
I
MODEL
169
MAINTENANCE
TABLE 6-4
AC Converter
STEP
c-
1
2
UlOl
External "oltagc Source Apply 1OOmVAC at 100Hz Cl08
Ulol pin 6 to Input Lo
6
UlOl pin 6 to Input Lo
7
ITEM/COMPONENT REQUIRED CONDITIONS
Power 01,. 20h"AC Kangc
pin
7 to
pin 4
7V min.
98mVOC Nominal using
1OMfl input impedance voltmeter
See Waveform (Figure
6-4)
Short Input
*SmVDC offset max.
Supply Voltage Calibrated Input IHigh Impedance
+440mV-
1OOnVAC Input
FIGURE 6-4.
AC/DC Converter Waveform.
6-l
MAINTENANCE
TABLE 6-5
Reference source
MODEL
169
STEP ITEM/COMPONENT
1 2 U102 pin 40 to Input Lo
VRlOl
3
4
STEP
(<lOOpf probe)
R102
pin 8 to Input Lo
ITEM/COMPONENT
REQUIRE0 CONOITIONS
power ON
2.4 to 3.2 "DC
1.2 to 1.25v Nominally 1OOmVDC using
voltmeter with ,lOMn input impedance
TABLE 6-6
Ohms Circuit
REQUIRE0 CONDITIONS* REMARKS
REMARKS
Output Of u102
(Analog Supply)
Reference sourc
e
6-8
1
2 RTlOl l.OKn to 3.OKn (can
3 Reference Resistors R108 thru R113 can
4 Power ON, 200n
5
(1101
'Do not remove RTlOl unless absolutely necessary because the leads fall
off device easily while soldering.
Power OFF, DCV func-
tion, all range buttons out
now be checked in circuit)*
be checked in cir- R109 affects 2Kn range wit for XC"t-KY
range, Input open
0.1%
‘2 Volts across
errmitter ­collector junction
Resistance Check
R108 affects 2005, range RI10 affects ZOK" range
Rlll
affects 200Kn range
It112
affects 2000Kn ran<
R113 affects 20Mn range
Voltage drop check
MODEL 169
d. Make sure the rubber gasket is in place before placing the new display in the slot of the front panel.
e. The zebra strip should be installed after the
new display is placed in the front panel window. Care must be taken to assure the zebra strip is kept clean. The zebra strip should be aligned such that the black stripes contact the conductors on the display and the conductors on PC-509.
f. Reassemble PC-509 and the front panel using the three 318" pal nuts. are snug.
Turn the unit "ON" and check the display to
'3. see that it is working properly. If the zebra strip
is not aligned properly or if it is dirty there will
be some segments on the display that will not light.
Tighten them until they
K-33. BATTERY REPLACEMENT.
6-34. The low battery indicator will automatically come on with approximately 5% of battery life remaining (20 hours for carbon-zinc and 40 hours for alkaline cells). Replace per directions in Para-
graph 2-6.
6-35. CURRENT FUSE REPLACEMENT (Figure 6-5).
t
To prevent a shock hazard, disconnect all
circuits from the Model 169 INPUT termi-
nals before removing current fuse.
6-36. To remove the current fuse, turn the DMM over and lay it carefully on its top. Insert d small
screwdriver blade into the slot on the fuse holder, press gently and turn l/4-turn counter-clockwise.
Lift the holder and fuse out of the receptacle. The fuse can now be removed for checking or replacement. Replace with one of the following types:
WARNING
'FIGURE 6-5.
Installing a higher rated fuse than the one specified could result in damage to the instrument.
Current Fuse Removal/Replacement.
A
CAUTION
a. U.S.A. Use - 3AG, 25OV, 2A, Normal Blow
b. Europe Use - 5 x 20mm, 25OV, 2A, Normal Blow
6-9
I
MODEL 169
SECTION 7. REPLACEABLE PARTS
REPLACEABLEPARTS
7-i. GENERAL.
7-2. This section contains information for ordering replacement parts. Panel and covers are shown separately on Figure 7-l. The Replaceable Parts List is arranged in alphabetical order of the Circuit Designations of the components.
7-3. ORDERING INFORMATION
7-4. To place an order or to obtain information concerning re­placement parts contact your Keithley representative or the fac­tory. See the inside front cover for addresses. When ordering,
include the following information:
a. Instrument Model Number.
b. Instrument Serial Number.
c. Part Description. d. Ci’cuit Designation (if applicable). e. Keithley Part Number.
7-5. FACTORY SERVICE.
7-6 If the instrument is to be returned to the factory for service, please complete the Service Form which follows this section, and return it with the instrument.
7-7. SCHEMATIC.
7-8. The Model 169 3.112 Digit DMM: Schematic No. 30180D.
7-9. COMPONENT LAYOUT.
7-10. Model 169 PC-506 Component Layout No. 301760.
7-11. MODEL 1699 SPARE PARTS KIT.
7-12. A spare parts kit is available that contains a complement
of spare parts that can maintain up to ten Model 169’s for ap-
proximately one year. A list of the spare parts is given in Table 7-1.
TABLE 7-1
Model 1699 Parts List
iiy
Keithley P/N
-
2
IJZ-62.
2
FH-12
5
FU-13
5
FU-48
1
IC-1028
1
IC-226’
2
IC-227
1
IC-228*
1
LSI-22’
1
R-277.9M
3
RF-28
2
RF-34
1
RF-39
1
RT-5
3
TG-47
-
‘Anti-Static Protection and Handling required. Refer to Section 6 of this Manual for Special Handling Procedures for Static-Sensitive
Devices.
Schematic Designation VRlOl
Fuse Holder
FiOl (3AG) FlOl (5mm,x 20mm) u103
u104, u105
UlOl U106 u102
RI13 CR101 thru CR104
CR106, CR107 CR105
RTlOl QIOI, Cl102
7-o
I
REPLACEABLEPARTS
MODEL 169
Table 7-2
REPLACEABLE PARTS LIST
Circuit m?sig.
BTlOl thru BT106
Cl01 Cl02 Cl03 Cl04 Cl05 Cl 06 Cl07 Cl06 Cl09 Cl10 Cl11 Ci12 Cl13 Cl14 Cl15 Cl16 Cl17 Cl16
CR101 CR102 CR103 CR104 CR105 CR106 Rectifier, Silicon, 3A. 50V CR107
Description
Battery, 1.5V, “c” cell, Carbon Zinc
Cap., .luF, 16OV, Mstalized Polypropylene Cap., .22uF, lOOV, Metalized Polyester Cap., .047uF, IOOV, Metalized Polyester Cap., .I uF, lOOV, Metalized Polyester Cap., 4.7uF, 25V, Aluminum Electrolytic Cap., .luF. lOOV, Metalized Polyester Cap., 4.7uF, 25V, Aluminum Electrolytic Cap., 1 .OuF, 16V, Aluminum Electrolytic Cap., .02uF, IOOOV, Ceramic Disc Cap., 3.3pF. 15OOV, Ceramic Cap., .047uF, IOOV, Metalized Polyester Cap., lZOpF, 5OOV, Silver Mica Cap., 4.7uF, 25V, Aluminum Electrolytic Cap., .22uF, lOOV, Metalized Polyester Cap., 39OOpF, lOOOV, Ceramic Disc Cap., 4.7uF, 25V, Aluminum Electrolytic Cap., Adjustable (1.5pF, nominal) Cap., O.luF. 5OV, Ceramic Film
Rectifier, Silicon, 75mA. 75V Rectifier, Silicon, 75mA, 75V Rectifier, Silicon, 75mA, 75V Rectifier. Silicon, 75mA, 75V Rectifier, Germanium 15mA. 25V
Rectifier, Silicon, 3A, 50V
Keithley Part No.
BA-6
~ c-305-.047
c-305-.1 c-314-4.7 c-305.. 1
1 c-314-4.7
C-325-l C-296-.02 c-30%3.3P c-305-.047 C-320-1 IOP c-314-4.7 C-305-.22 C-64.3900P c-314-4.7 c-1 64 C-237-.1
RF-26 RF-28 RF-26 RF-26 RF-39 32453 32453
DSlOl
Liquid Crystal Display
Zebra Connector Strip
Fuse, 2A, 25OV, 3AG (U.S. Models) Fuse. 2A, 25OV, 5 x 20mm (European Models)
Banana Jack, Red
Banana Jack, Black Battery Connector
Transistor. NPN
DD-16 CS-376-l
FU-13 FU-46
BJ-1 l-2 BJ-11-O
BH-26
TG-47 TG-47
7-1
MODEL 169
REPLACEABLEPARTS
Table 7-2
REPLACEABLE PARTS LIST (Cont.)
Zircuit Desig.
RlOi RI02 RI 03 RI04 RI05 HI 06 RI07 RIO8 RI OS RI10 RI11 Fill2 RI13 RI14
R115 RI16 RI17
RTlOl SlOl
UlOl u102 u103 u104 uio5 U106
Description Potentiometer, 500 ohm
Thick Film Resistor Network
Resistor, 1 M ohm, 10%. 1 W. Comp Resistor, lOkohm, .I%, .lW. MtF Resistor,4,71kohm, .1%. .lW. MtF Thick Film Resistor Network
Resistor, tOOk ohm, 10%. 2W. Comp.
Resistor, 100.1 ohm. .I% .lW, MtF
Resistor. 900 ohm, .I%, .1 W, MtF
Resistor. Sk ohm, .I%. .iW, MtF
Resistor,90kohm, .I%. .lW, MtF
Resistor, 900k ohm, .i%, .lW, MtF
Resistor, 9M ohm, .08X 2W. CMtF
Resistor, Dual
Resistor, 9 ohm, 5% l/ZW. MtF
Resistor, 90 ohm, .l%, 1/8W, MtF
Resistor, SOOohm, .I%, .lW, MtF
Thermistor, 2k ohm, 40% 8mA, 5OOV, PTC
Switch, 12 Station, Pushbutton
Low-Power JFET-Input OP~AMP
3-i/2 Digit Single Chip A/D Converter
CMOS Quad Z-Input NAND Gate
CMOS Quad Exclusive-OR Gate
CMOS Quad Exclusive-OR Gate
CMOS Quad Exclusive-NOR Gate
Keithley Part No.
RP-97.500 TF-93 R-2-l M R~283-10k R-283~4.71 k TF-92
R-3.IOOk R~283-100 R-283-900 R..283~Sk R-283.90k R-283~900k R~277-9M R-278 R-61 -9 R~l79-90 R-283-900
RT-5 303148 ,sl
IC-277
?3.shwA-
IC-102’ IC~226’ IC-226” IC-228’
I
7-2
VRlOl VW 02
‘Anti-static protection and handling required
Diode, Low voltage ref, 1.2V. IOmA Diode, Zener, S.lV nominal. .25W
DZ-62 DZ-14
I
REPLACEABLE PARTS
MODEL 169
FIGURE 7-1. Covers and Panels.
7-3
“I I “r I
OY
\
I -
SERVICE FORM
Model No.
Serial No. Name and Telephone No. Company
Lisl all control settings, describe problem and check boxes that apply 10 problem.
q
intermittent
q
IEEE failure
q
Front panel operational
Display or output (circle one)
q
Drifts
OLInstable
q
0verlond
lJCalihration only
q
Dara required
(autach any addilional sheets as necessary.)
q
Analog output follows display
q
Obvious problem on power-up
q
All mnges or functions are bed q Checked all cables
q
Unahle to xl-o
q
WiU not read applied input
UC of C required
q
l’anicular wngrt or function had: specify
q
Batteries and fuses are OK
Date
Show a block diagram of your measurement system including all instruments connected (wlxrher power is tumcd on or nor). Also, describe signal source.
Where is the measurement being performed? (factory, controlled laboratory, out-of-doors. etc.)
What power line voltage is used? Ambient TenlperaturG
Relative humidity?
Any additional information. (If special modifications have been made by [he user. pleasc describe.)
Other‘?
B INSTRUMENTS
Model 169 Digital Multimeter Addendum
INTRODUCTION
This addendum to the Model 169 Digital Multimeter Instn~tion Manual is being provided in order to supply you with the latest information in the least possible time. Pleaseincorporate this information into the
immediately.
Page 7-O; Replace Table 7-1 with the following:
Table 7-1. Model 1699 Parts List
manual
!&entity
1 5 5
1
1 2
1
1 i
2
1
1
1
‘Anti-Static Protection and Handling required. Refer lo the Model 169 Instruction Manual for Special Procedures for Static-Sensitive Devices
Keithley PIN
DZ-62* FU-13 FU-46 K-102
10.226 IC-227 IC-226 LSI-122’ R277-9M RF-26 RF-34 RF-39 UT-5
TG-47
Schematlc Designation VRiOl
FlOl (3AG) Fl 01 (5mm x 20mm) u103 u104, u105 UlOl U106 u102 R113 CR101 thru CR104 CR106, CR107 CR105 RT-101 Q102
MODEL 169
- RECOMMENDED SPARE PARTS
PART NUMBER DESCRIPTION
DZ-62
IC-102 IC-226 IC-227 IC-228
LSI-22
R-277-W RF-28 RF-34 RF-39 RT-5
TG-47
DD-18C~ c179-4.7 BA-28 30169
30170
Diode
Integrated Circuit Integrated Circuit Integrated Circuit Integrated Circuit Integrated Circuit Resistor Rectifier Rectifier Rectifier Thermistor Transistor Readout
Cap Battery
Front Panel
Overlay
QTY
-
2 1
1
2
1
1
1
3 2
I 1 3 1 2
1 1
1
KH/ntb:09/10/82
m INSTRUMENTS
Model 169 Digital Multimeter Addendum
INTRODUCTION Thisaddendum to the Model 169 Digital Multimeter Inshuction Manual is being provided in
you with the latest information in the least possible time. Pleaseincorporate this information into immediately.
Page 7-O; Replace Table 7-1 with the following:
Table 7-1. Model 1699 Paris List
Quantity
i 5
1 1
2
1 1
1 3 2
1 1
r
1
Keithley P/N Schematic Designation DZ-62’
FU-13 FU-48 IC-102’ IC-226 IO-227 K-228* LSI-122 R277-9M RF-28
RF-34 RF-39 RT-5 TG-47
VRlOl FlOl (3AG) FlOl (5mm x 20mm) u103 u104, u105 UlOl U106 u102
R113 CR101 thru CR104 CR106, CR107 CR105
RT-101 Q102
order
tosupply
the manual
‘Anti-Static Protection and Handling required. Refer lo
Instruction Manual for Special Procedures for Static-Sensitive Devices
the Model 169
m INSTRUMENTS
Model 169 Digital Multimeter Addendum
INTRODUCTION Thisaddendum to the Model 169 DigitalMultimeter Inshuction Manual is being provided in order to supply
you with the latest information in the least possible time. Please incorporate this information into the immediately.
manual
Page 7-O; Replace Table 7-l with
I
Zwantity
1
5
5
1
1
2
1 1
1 3 2
1
1
1
‘Anti-Static Protection and Handling required. Refer to the Model 169 instruction Manual for Swcial Procedures for Static-Sensitive Devices,
the
following:
Table 7-1. Model 1699 Parts List
Keithley PIN DZ-62’
FU-13
FU-46 K-102’ K-226 IC-227
IC-228’ LSI-122’ R277-9M
RF-26
RF-34
RF-39 RT-5 TG-47
Schematic Designation VRlOl
FlOl (3AG)
Fl 01 (5mm x 20mm)
u103 u104, u105 UlOl U106 u102 R113 CR101 thru CR104 CR106, CR107 CR105 RT-101
Q102
Loading...