KaVo Primus 1058 Life TM, KaVo Primus 1058 Life S, KaVo Primus 1058 Life C Instructions For Use Manual

Page 1
Instructions for use
Primus 1058 Life
Page 2
Manufacturer:
Kaltenbach & Voigt GmbH Bismarckring 39 D-88400 Biberach www.kavo.com
Distributed by:
KaVo Dental GmbH Bismarckring 39 D-88400 Biberach Phone +49 (0) 7351 56-0 Fax +49 (0) 7351 56-1488
Page 3
Table of contents
1 User instructions ................................................................................................................ 6
1.1 User guide .................................................................................................................... 6
1.1.1 Abbreviations ..................................................................................................... 6
1.1.2 Symbols ............................................................................................................ 6
1.1.3 Target group...................................................................................................... 6
1.2 Service ......................................................................................................................... 7
1.3 Terms and conditions of warranty.................................................................................... 7
1.4 Transportation and storage............................................................................................. 7
1.4.1 Currently valid packaging regulations.................................................................... 7
1.4.2 Damage in transit............................................................................................... 8
1.4.3 Information on the packaging: Storage and transportation...................................... 9
2 Safety..................................................................................................................................10
2.1 Description of safety instructions ..................................................................................... 10
2.1.1 Warning symbol ................................................................................................. 10
2.1.2 Structure ........................................................................................................... 10
2.1.3 Description of hazard levels ................................................................................. 10
2.2 Purpose – Proper use ..................................................................................................... 10
2.2.1 General ............................................................................................................. 10
2.2.2 Product-specific .................................................................................................. 12
2.3 Safety instructions ......................................................................................................... 13
2.3.1 General information ............................................................................................ 13
2.3.2 Product-specific .................................................................................................. 15
3 Product description ............................................................................................................18
3.1 Treatment unit versions.................................................................................................. 18
3.1.1 KaVo Primus 1058 Life TM................................................................................... 18
3.1.2 KaVo Primus 1058 Life S ..................................................................................... 18
3.1.3 KaVo Primus 1058 Life C ..................................................................................... 19
3.1.4 KaVo Primus 1058 Life C with right-side set-up kit ................................................. 19
3.2 Patient chair Standard and COMPACTchair........................................................................ 20
3.3 Device body with patient unit.......................................................................................... 21
3.4 Dentist unit versions ...................................................................................................... 22
3.4.1 TM/C table ......................................................................................................... 22
3.4.2 S table............................................................................................................... 23
3.5 Assistant element – Versions ......................................................................................... 24
3.5.1 Standard assistant unit........................................................................................ 24
3.5.2 Assistant element right, left (optional, only in conjunction with patient chair
Standard) .......................................................................................................... 25
3.6 Three function handpiece (3F handpiece) ......................................................................... 26
3.7 Multifunctional handpiece (MF handpiece)......................................................................... 26
3.8 X-ray viewer 1440 ......................................................................................................... 27
3.9 Controls........................................................................................................................ 27
3.9.1 Dentist elements ............................................................................................... 27
3.9.2 Assistant unit ..................................................................................................... 28
3.9.3 Groups of keys ................................................................................................... 28
3.9.4 Foot control........................................................................................................ 30
3.10Rating plate and identification plate ................................................................................. 31
Instructions for use Primus 1058 Life
Table of contents
3 / 126
Page 4
3.10.1 Rating plate 1440 ............................................................................................... 36
3.11Technical data............................................................................................................... 36
3.12KaVo Service table 1568 (optional accessory)................................................................... 43
4 Operation............................................................................................................................ 44
4.1 Switching the device on and off....................................................................................... 44
4.2 Adjusting the dental chair............................................................................................... 45
4.2.1 Adjusting the arm rest (optional).......................................................................... 45
4.2.2 Adjust head rest ................................................................................................. 46
4.2.3 Positioning the dental chair manually .................................................................... 48
4.2.4 Automatic positioning of dental chair .................................................................... 49
4.2.5 Safety shut-off ................................................................................................... 53
4.3 Moving the patient chair................................................................................................. 56
4.4 Move the dentist's unit ................................................................................................... 57
4.4.1 Move the cart..................................................................................................... 58
4.5 Move the patient unit..................................................................................................... 59
4.5.1 Swing the patient unit by hand............................................................................. 59
4.6 Moving the assistant element.......................................................................................... 59
4.6.1 Adjusting the height of the Standard assistant element........................................... 59
4.6.2 Moving the assistant element right, left (optional).................................................. 60
4.7 Using functions through the menu................................................................................... 62
4.7.1 Using the user menu........................................................................................... 62
4.7.2 Standby menu.................................................................................................... 66
4.7.3 Operating the MEMOspeed menu (optional)........................................................... 69
4.7.4 Using the CONEXIOcom (optional)........................................................................ 75
4.8 Using function through the dentist or assistant unit ........................................................... 76
4.8.1 Using the hygiene functions................................................................................. 76
4.8.2 Using lamp and X-ray image viewer...................................................................... 78
4.8.3 Using the bell (optional) ...................................................................................... 78
4.8.4 Using the timer................................................................................................... 78
4.8.5 Saving handpiece settings (without MEMOspeed)................................................... 79
4.9 Operating the foot switch................................................................................................ 82
4.9.1 General functions................................................................................................ 82
4.9.2 Positioning the patient chair with the foot control ................................................... 82
4.9.3 Preselect dentist ................................................................................................. 82
4.9.4 Start and regulate instruments............................................................................. 82
4.9.5 Setting the cooling condition................................................................................ 83
4.9.6 Activate blown air............................................................................................... 83
4.9.7 Preselect counterclockwise motor rotation ............................................................. 83
4.9.8 Adjusting the instrument light.............................................................................. 83
4.9.9 Using CONEXIOcom (fee-based additional option).................................................. 84
4.10Service table 1568 (optional accessory) ........................................................................... 84
4.10.1 Moving the service table...................................................................................... 85
5 Preparation methods DIN EN ISO 17664...........................................................................87
6 Accessories and kits ........................................................................................................... 88
6.1 Device.......................................................................................................................... 88
6.2 Dental chair .................................................................................................................. 88
6.3 Assistant unit ................................................................................................................ 89
Instructions for use Primus 1058 Life
Table of contents
4 / 126
Page 5
6.4 Dentist unit................................................................................................................... 89
7 Safety check - Test instructions ......................................................................................... 90
7.1 Introduction.................................................................................................................. 90
7.1.1 General instructions ............................................................................................ 90
7.1.2 Notes for medical electrical systems...................................................................... 91
7.1.3 Components of the safety check........................................................................... 92
7.1.4 Testing intervals ................................................................................................. 92
7.1.5 Notes on the test method in accordance with IEC 62353......................................... 92
7.1.6 Notes on repeat testing ....................................................................................... 93
7.2 Instructions for the safety check...................................................................................... 93
7.2.1 Preparatory measures to be undertaken on the device............................................ 93
7.2.2 Visual inspection (inspection by examination)........................................................ 94
7.2.3 Measurements.................................................................................................... 96
7.2.4 Functional test....................................................................................................106
7.2.5 Assessment and documentation ..........................................................................109
8 Appendix - Additional measuring sites...............................................................................111
8.1 Additional scanning sites SL X in the protective conductor measurement .............................111
8.2 Additional measuring sites AP X for EUL/EPL measurement ................................................112
8.3 Additional connection sites ACP X (additional earth connections).........................................113
9 Troubleshooting .................................................................................................................114
10Data on electromagnetic compatibility according to EN 60601-1-2 ..................................119
10.1Electromagnetic Transmissions........................................................................................119
10.2Resistance to electromagnetic interference .......................................................................119
10.3Recommended safe distance between portable and mobile HF telecommunications
equipment and the treatment unit...................................................................................120
10.4Immunity to electromagnetic interference ........................................................................121
Instructions for use Primus 1058 Life
Table of contents
5 / 126
Page 6
Instructions for use Primus 1058 Life
1 User instructions | 1.1 User guide
6 / 126
1 User instructions
1.1 User guide
Requirement
Read these instructions prior to first use to avoid misuse and prevent damage.
1.1.1 Abbreviations
Ab­brevi­ation
Explanation
IFU Instructions for use CI Care instructions AI Assembly instructions TI Technician's instructions SC Safety checks IEC International Electrotechnical Commission RI Repair instructions RK Retrofitting kit AS Assembly kit CK Conversion kit EP Enclosed parts EMC Electromagnetic compatibility PI Processing instructions
1.1.2 Symbols
See the Safety/Warning Symbols section
Important information for users and technicians
CE mark according to Medical Devices Directive EC 93/42
Action required
eLabeling ID
1.1.3 Target group
This document is for dentists and dental office staff.
Page 7
Instructions for use Primus 1058 Life
1 User instructions | 1.2 Service
1.2 Service
KaVo Customer Service: +49 (0) 7351 56-1000 service.einrichtungen@kavokerr.com or service.treatmentunits@kavokerr.com Please refer to the serial number of the product in all inquiries! For further information, please visit: www.kavo.com
1.3 Terms and conditions of warranty
KaVo provides the final customer with a warranty that the product cited in the handover certificate will function properly and guarantees zero defects in the material or processing for a period of 12 months from data of purchase, subject to the following conditions: Upon justified complaints of flaws or a short delivery, KaVo will make good its warranty by replacing the product free of cost or repairing it according to the customer's wishes. Other claims of any nature whatsoever, in particular with respect to compensation, are excluded. In the event of default and gross negli­gence or intent, this shall only apply in the absence of mandatory legal regula­tions to the contrary. KaVo cannot be held liable for defects and their consequences due to natural wear, improper cleaning or servicing, non-compliance with operating, servicing or connection instructions, calcification or corrosion, contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with factory specifications. The warranty does not usually cover bulbs, glassware, rubber parts and the col­ourfastness of plastics. Defects or their consequences that can be attributed to interventions on or changes made to the product by the customer or a third party are excluded from the warranty. Defects or their consequences that can be attributed to interventions on or changes made to the product by the customer or a third party are excluded from the warranty.
1.4 Transportation and storage
1.4.1 Currently valid packaging regulations
Note
Only valid for the Federal Republic of Germany.
Dispose of and recycle the sales packaging appropriately in accordance with current packaging regulations, employing waste management or recycling com­panies. Comply with the comprehensive return system. KaVo has had its sales packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
7 / 126
Page 8
Instructions for use Primus 1058 Life
1 User instructions | 1.4 Transportation and storage
8 / 126
1.4.2 Damage in transit
In Germany
If the packaging is visibly damaged on delivery, please proceed as follows:
1. The recipient of the package must record the loss or damage on the delivery receipt. The recipient and the representative of the shipping company must sign this delivery receipt.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use the product.
4. Report the damage to the shipping company.
5. Report the damage to KaVo.
6. Consult with KaVo first, before returning a damaged product.
7. Send the signed delivery receipt to KaVo.
If the product is damaged but there was no discernable damage to the pack­aging on delivery, proceed as follows:
1. Report the damage to the shipping company immediately and no later than 7 days after delivery.
2. Report the damage to KaVo.
3. Leave the product and packaging in the condition in which you received it.
4. Do not use a damaged product.
Note
Failure on the part of the recipient to comply with any of the above-men­tioned obligations will mean that the damage will be considered to have arisen following delivery (in accordance with the General German Freight For­warders' Terms and Conditions, Art. 28).
Outside Germany
Note
KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival.
If the packaging is visibly damaged on delivery, please proceed as follows:
1. The recipient of the package must record the loss or damage on the delivery receipt. The recipient and the representative of the shipping company must sign this delivery receipt. Without this evidence, the recipient will not be able to assert a claim for damages against the shipping company.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use the product.
If the product is damaged but there was no discernable damage to the pack­aging on delivery, proceed as follows:
1. Report any damage to the shipping company immediately and no later than 7 days after delivery.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use a damaged product.
Note
If the recipient fails to comply with any of the above-mentioned obligations, the damage will be considered to have arisen following delivery (in accordance with CMR law, Chapter 5, Art. 30).
Page 9
Instructions for use Primus 1058 Life
1 User instructions | 1.4 Transportation and storage
1.4.3 Information on the packaging: Storage and
transportation
Note
Please keep the packaging in case you need to return the product for servi­cing or repair.
The symbols printed on the outside are for transportation and storage, and have the following meaning:
Transport upright with the arrows pointing upwards!
Fragile - protect against impact!
Protect from moisture!
Permissible stacking load
Temperature range
Humidity
Air pressure
9 / 126
Page 10
Instructions for use Primus 1058 Life
2 Safety | 2.1 Description of safety instructions
10 / 126
2 Safety
2.1 Description of safety instructions
2.1.1 Warning symbol
Warning symbol
2.1.2 Structure
DANGER
The introduction describes the type and source of the hazard.
This section describes potential consequences of non-compliance.
The optional step includes necessary measures for hazard prevention.
2.1.3 Description of hazard levels
The warning and safety notes in this document must be observed to prevent personal injury and material damage. The warning notes are designated as shown below:
DANGER
In cases which – if not prevented – directly lead to death or severe in­jury.
WARNING
In cases which – if not prevented – could lead to death or severe in­jury.
CAUTION
In cases which – if not prevented – could lead to minor or moderate injury.
NOTICE
In cases which – if not prevented – could lead to material damage.
2.2 Purpose – Proper use
2.2.1 General
The functional safety and proper condition of the device must be checked be­fore each use of the device.
The KaVo Primus 1058 Life equipment system is a dental treatment centre in accordance with ISO 7494 equipped with a patient chair in accordance with ISO
6875. This KaVo product is designed for use in dentistry only and may only be
used by trained medical personnel. Any other type of use is not permitted.
Page 11
Instructions for use Primus 1058 Life
2 Safety | 2.2 Purpose – Proper use
"Proper use" includes compliance with all information in the Instructions for Use and ensuring that all inspections and service tasks are performed.
The KaVo service table 1568 is an accessory of KaVo treatment centres that do not feature a dentist element mounted to the unit. The KaVo Service table 1568 serves as a support surface for the dentist during the treatment. The service table 1568 can be used as a support for trays and for attachment of the X-ray viewer 1440. The mobile support system allows the KaVo Service table 1568 to be moved to the proper position.
The overarching guidelines and/or national laws, national regulations and the rules of technology applicable to medical devices for start-up and use of the KaVo product for the intended purpose must be applied and followed.
KaVo accepts liability for the safety, reliability, and performance of components supplied by KaVo, provided:
▪ installation, instructions, expansions, adjustments, changes or repairs were
carried out by technicians trained by KaVo or third parties authorised by KaVo, or by the personnel of authorised distributors.
▪ the unit was operated in accordance with the instructions for use, care and
installation.
▪ the IT components supplied by the operator meet the technical require-
ments in the present instruction for use for hardware and software, and are installed and set up according to the descriptions of these components.
▪ in the case of repairs, the requirements of IEC 62353 "Recurrent tests and
tests before start-up of electrical items of medical electrical equipment and systems - general regulations" are met in full.
It is a responsibility of the user:
▪ to only use equipment that is operating correctly, ▪ to protect him or herself, the patient and third parties from hazards, and ▪ to prevent contamination from the product
The applicable national legal regulations must be observed during the use of the device, in particular the following:
▪ Applicable regulations governing the connection and startup of medical
devices.
▪ Current occupational safety regulations. ▪ Current accident prevention regulations.
Regular servicing and safety checks are essential for the permanent assurance of the operating and functional safety of the KaVo product and for the preven­tion of damage and hazards. Testing and maintenance intervals: Maintenance must be performed once a year, the safety checks at intervals of 2 years. Shorter intervals for the safety checks may be specified by the tester if necessary.
The following persons are authorised to conduct repairs and servicing and the safety check on the KaVo product:
▪ Technicians of KaVo branch offices after appropriate product training. ▪ Specifically KaVo-trained technicians of KaVo franchised dealers.
In Germany, operators, equipment managers and users are obliged to operate their equipment in accordance with the MPG regulations.
11 / 126
Page 12
Instructions for use Primus 1058 Life
2 Safety | 2.2 Purpose – Proper use
12 / 126
The services encompass all the test tasks required in accordance with § 6 of the medical devices operator ordinance (Medizinprodukte-Betreiberverordnung, MP­BetreibV).
Note
The product must be cleaned and serviced according to instructions if it is not to be used for an extended period of time.
Note
The MULTIflex couplings, the current K/KL motors, and the ultrasonic scaler hoses of KaVo are equipped as standard with a protective device to prevent treatment water from being drawn back into the treatment centre via the handpieces. If products from other manufacturers are used at the standard­ised interfaces, it must be ensured that they are equipped with an appropri­ate protective device! If this is not the case, they may not be used!
Information about electromagnetic compatibility
Note
Based on IEC 60601-1-2 (DIN EN 60601-1-2) concerning the electromag­netic compatibility of electrical medical devices, we must draw your attention to the following points:
• Medical electrical devices are subject to special precautions concerning the electromagnetic compatibility and must be installed and operated in accord­ance with the KaVo assembly instructions.
• High-frequency communications devices may interfere with electrical med­ical devices.
See also:
2 10 Data on electromagnetic compatibility according to EN 60601-1-2, Page
119
Note
KaVo cannot guarantee the compliance of accessories, cables, and other components not supplied by KaVo with the EMC requirements of IEC 60601-1-2 (DIN EN 60601-1-2).
Disposal
Note
Any waste which is generated must be recycled or disposed of in strict com­pliance with all applicable national regulations in a manner which is safe both for people and the environment.
If you have any questions regarding proper disposal of the KaVo product, please contact the KaVo branch.
Disposal of electronic and electrical devices
2.2.2 Product-specific
Designated use and target group
KaVoPrimus 1058 Life is designed for dental treatment of children and adults.
Page 13
Instructions for use Primus 1058 Life
2 Safety | 2.3 Safety instructions
The KaVo Primus 1058 Life equipment system is a dental treatment centre in accordance with ISO 7494 equipped with a patient chair in accordance with ISO
6875. KaVo three-way and multifunctional handpieces are dental handpieces in
accordance with EN 1639. They support the dental application in the mouth of the patient by supplying air, water or spray. In addition, the multifunctional handpiece supplies light and heated media. The KaVo X-ray viewers 1440 are designed for radiograph viewing in dentistry and comply with the requirements of DIN 6856-3. These KaVo products are designed for use in dentistry only and must be used by trained medical personnel only.
Connecting devices
KaVo-approved accessories for patient communication. These accessories must be used exclusively.
Accessories Use Name Material number
Monitors Monitor 22“ KaVo Screen HD 1.011.0302
Monitor 19“ KaVo Screen One 1.011.0300
Cameras Intraoral camera ERGOcam One 130
ERGOcam One 160
1.011.2130
1.011.2129
Cables between unit, accessories and PC
USB extension cord - 5 metres
USB extension cord 5m with 1:1 hub
1.004.6953
USB extension cord - 10 metres
USB extension cord 2x5m with 1:1 hub
1.011.3745
Display port cable
- 5 metres
LTG Display port 5m Standard
1.011.3583
Display port cable
- 10 metres
LTG Display port 10m Standard
1.011.0298
Note
The USB interfaces of the system may only be connected to IT devices ap­proved by KaVo.
Note
When connecting IT equipment to the the medial electrical system, observe EN 60601-1.
2.3 Safety instructions
2.3.1 General information
Note
The safety and reliability of the system can only be ensured when the de­scribed procedure is followed.
DANGER
Explosion hazard.
Risk of fatal injury.
Do not use KaVo product in areas subject an explosion hazard.
13 / 126
Page 14
Instructions for use Primus 1058 Life
2 Safety | 2.3 Safety instructions
14 / 126
WARNING
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
WARNING
Application of un-authorised accessories or un-authorised modifica­tions of the product.
Accessories that have not been approved and/or inadmissible modifications of the product could lead to hazards and/or personal injury or property damage.
Only use accessories that have been approved for combination with the
product by the manufacturer or are equipped with standardised interfaces (e. g. MULTIflex couplings, INTRAmatic).
Do not make any modifications to the device unless these have been ap-
proved by the manufacturer of the product.
WARNING
Injury or damage from damaged functional parts.
If functional parts are damaged, it can cause additional damage or personal in­jury.
Check the device, electrical cables and any accessories regularly for pos-
sible damage to the insulation and replace them according to need.
If functional parts are damaged: discontinue your work and
repair the damage or notify a service technician!
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
Before disposal, reprocess and sterilise the product and accessories appro-
priately.
CAUTION
Health hazard and property damage due to non-compliance with servi­cing schedule.
Infection hazard to users and patients. Product damage.
Comply with servicing schedule.
NOTICE
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
Perform regular proper care and servicing!
CAUTION
Risks from electromagnetic fields.
Electromagnetic fields might interfere with the functions of implanted systems (such as pacemakers).
Ask patients if they have a cardiac pacemaker or other system implanted
before you start the treatment!
Page 15
Instructions for use Primus 1058 Life
2 Safety | 2.3 Safety instructions
CAUTION
Malfunctions due to electromagnetic fields.
The product meets the applicable requirements regarding electromagnetic fields. Given the complex interactions between equipment and cell phones, the product may be influenced by a cell phone that is in use.
Do not use cell phones in medical offices, hospitals or laboratories!Put electronic devices such as e.g. computer storage media, hearing aids
etc. down during operation!
NOTICE
Damage from liquids.
Residual liquids of any type can cause stains on or damage to cushions and parts of the housing.
Remove any residual liquids without delay.
Note
The operator may only carry out repair work if the device is switched off and no patient is being treated.
2.3.2 Product-specific
WARNING
Injury or infection hazard from laid down instruments.
Given the arrangement of the instruments, injury or infections in the hand and underarm may occur when reaching for the tray holder or operating device. In­creased risk of infection from ill patients.
Be aware of the arrangement of the instruments when you reach for the
tray holder or operating device.
WARNING
Health damage due to back-suction on the instruments.
Infection hazard. It is feasible to use, on standardised interfaces, products from other manufac-
turers, which are not equipped with a protective device preventing back-suc­tion of treatment water via the instruments into the dental treatment centre.
If products from other manufacturers are used at the standardised inter-
faces, you must make sure that the products are equipped with an appro­priate protective device!
Do not use products with no protective device.
CAUTION
Sitting down on a dental chair that is in horizontal orientation is asso­ciated with a risk of injury.
Do not sit on the head or foot end of the patient chair when it is in a hori-
zontal position.
CAUTION
Risk of injury if you support yourself on the swivel arm.
Exposing the swivel arm to an overload may damage the swivel arm and cause injury to the patient or user.
Never strain the swivel arm, spring arm, assistant element and dentist ele-
ment by using it as a support!
15 / 126
Page 16
Instructions for use Primus 1058 Life
2 Safety | 2.3 Safety instructions
16 / 126
CAUTION
Risk of injury by suspended instruments (S table).
Patients may get injured by sharp instrument tips.
When you move the dentist's unit, make sure that nobody is injured.Alert patients and care providers to the risk of injury.
CAUTION
Risk of injury during cleaning of the treatment unit.
Lack of instructions to the cleaning staff and lack of preparation of the treat­ment unit can lead to the cleaning personnel sustaining injuries.
Only trained professionals and instructed cleaning personnel may be
present in the treatment rooms.
Position the chair for cleaning and turn the device off.
CAUTION
Electrical power.
Electrical shock.
Set up the external PC outside of the patient environment keeping a min-
imal distance of 1.5 m.
Connect the PC and equipment connected to the PC in accordance with IEC
60601-1 / 60950.
CAUTION
Electrical power
Electrical shock from incorrectly connecting a non-medical system to the freely usable USB interfaces of the device (if any).
Connect any IT device to the medical system in accordance with IEC
60601-1.
Use USB devices with no additional power supply (USB-powered) only.Applied parts connected to the USB interface of the dentist element must
comply with the requisite insulation.
USB-powered devices failing to meet the requisite insulation for applied
parts must be placed appropriately such that direct contact of the USB device and the patient is excluded.
It is not permissible to touch USB-powered devices failing to meet the re-
quisite insulation for applied parts and the patient at the same time.
CAUTION
Health damage due to germ formation.
Infection hazard.
Before starting, rinse all the water drain lines with no handpieces attached.Before start-up and after downtimes (weekends, holidays, vacations, etc.),
rinse or purge the air and water lines with air.
Option: perform intensive germ reduction (if assembly kit is available).Actuate the tumbler filler several times.
Page 17
Instructions for use Primus 1058 Life
2 Safety | 2.3 Safety instructions
CAUTION
Third-party device connection assembly kit (optional): Risk of reverse contamination by stagnant water.
Infections. If a water consumer is connected to the third-party device connection as-
sembly kit, always perform the following actions on the device:
Before starting, rinse all the water drain lines with no handpieces attached
(if applicable).
Before start-up and after downtimes (weekends, holidays, vacations, etc.),
rinse or purge the air and water lines with air.
Mind the H2O2 resistance of the water consumer since the water is doped
with OXYGENAL 6 (concentration up to 0.02 %).
CAUTION
Long stay in the patient chair.
Decubitus formation.
Take precautions against the formation of decubitus in long treatments.
CAUTION
Risk of injury when the dental chair or headrest is moved.
Hair of the patient or practice personnel may get caught when the headrest of the dental chair is moved.
Mind the hair of the patient or practice personnel when moving the dental
chair or the headrest.
CAUTION
Risk of injury when moving the patient or patient chair.
The patient or treatment personnel can be pinched or crushed.
Position all moving parts, such as dentist element, assistant element, oper-
ating light, screens, etc., outside the collision range when you move the pa­tient or patient chair.
NOTICE
Damage to the instrument hoses from stickers.
Instrument hoses can burst.
Do not affix stickers or adhesive tape.
17 / 126
Page 18
Instructions for use Primus 1058 Life
3 Product description | 3.1 Treatment unit versions
18 / 126
3 Product description
3.1 Treatment unit versions
3.1.1 KaVo Primus 1058 Life TM
3.1.2 KaVo Primus 1058 Life S
Page 19
Instructions for use Primus 1058 Life
3 Product description | 3.1 Treatment unit versions
3.1.3 KaVo Primus 1058 Life C
3.1.4 KaVo Primus 1058 Life C with right-side set-up kit
19 / 126
Page 20
Instructions for use Primus 1058 Life
3 Product description | 3.2 Patient chair Standard and COMPACTchair
20 / 126
3.2 Patient chair Standard and COMPACTchair
Headrest ② BackrestChair base ④ SeatArm rest (optional)
Page 21
Instructions for use Primus 1058 Life
3 Product description | 3.3 Device body with patient unit
3.3 Device body with patient unit
Patient element Unit body
The central control is housed in the unit body.
Pressurised water bottle
(supplementary equipment)
Spittoon bowl
Tumbler filler ⑥ Supply element
Customer connection of power, wa­ter, compressed air, wastewater, and suction air
Foot control
21 / 126
Page 22
Instructions for use Primus 1058 Life
3 Product description | 3.4 Dentist unit versions
22 / 126
3.4 Dentist unit versions
3.4.1 TM/C table
Handle ② Turbine (multiflex coupling)INTRA LUX motor KL 703 or
INTRA LUX motor KL 701
Ultrasonic scaler
Three-function handpiece or multi-
functional handpiece
ERGOcam One
Tray support ⑧ Control element
Page 23
Instructions for use Primus 1058 Life
3 Product description | 3.4 Dentist unit versions
3.4.2 S table
Note
The holder assignment and arrangement of the instruments can be changed as needed and does not have to follow the picture.
Control element Triple function handpiece or multi-
functional handpiece
Turbine (multiflex coupling) INTRAlux motor KL 703 LED or IN-
TRA LUX motor KL 701
Ultrasonic scaler
23 / 126
Page 24
Instructions for use Primus 1058 Life
3 Product description | 3.5 Assistant element – Versions
24 / 126
3.5 Assistant element – Versions
3.5.1 Standard assistant unit
Three-function or multifunctional
handpiece
Spray mist suction
Control element ④ Saliva ejectorSatelec Mini LED
(polymerisation handpiece)
Tray holder for dental assistant
Page 25
Instructions for use Primus 1058 Life
3 Product description | 3.5 Assistant element – Versions
3.5.2 Assistant element right, left (optional, only in
conjunction with patient chair Standard)
Triple function handpiece ② Spray mist suctionControl element ④ Saliva ejectorSatelec Mini LED
(polymerisation handpiece)
25 / 126
Page 26
Instructions for use Primus 1058 Life
3 Product description | 3.6 Three function handpiece (3F handpiece)
26 / 126
3.6 Three function handpiece (3F handpiece)
MF handpiece hose ② Gripping sleeveMedia buttons (air/water) ④ Labelled blue: Three-function hand-
piece (3F handpiece)
Cannula
3.7 Multifunctional handpiece (MF handpiece)
MF handpiece hose ② Gripping sleeveMedia buttons (air/water) ④ Labelled gold: Multifunctional hand-
piece (MF handpiece)
Cannula
Page 27
Instructions for use Primus 1058 Life
3 Product description | 3.8 X-ray viewer 1440
3.8 X-ray viewer 1440
X-ray viewer 1440
Note
The KaVo X-ray viewer 1440 is a type 1 radiological viewing device in accord­ance as defined in DIN 6856-3.
3.9 Controls
3.9.1 Dentist elements
Dentist element TM/C table
A Group of keys for the dental chair B Group of keys for menu selection
(MEMOspeed optional) C Group of keys for hygiene D Group of keys for the timer E Group of keys for the handpieces F Group of keys for illumination
27 / 126
Page 28
Instructions for use Primus 1058 Life
3 Product description | 3.9 Controls
28 / 126
Dentist element S-table
A Group of keys for menu selection
(MEMOspeed optional)
B Group of keys for the timer
C Group of keys for the handpieces D Group of keys for the dental chair E Group of keys for illumination F Group of keys for hygiene
3.9.2 Assistant unit
A Group of keys for hygiene B Group of keys for illumination C Group of keys for the timer D Group of keys for the dental chair
3.9.3 Groups of keys
Group of keys for the dental chair
The keys of the assistant unit each have two functions and show two symbols.
Page 29
Instructions for use Primus 1058 Life
3 Product description | 3.9 Controls
Assistant element key Dentist element key Name
"Chair up" key
"AP 0" key (automatic position 0)
"Chair down" key
"SP" key (rinsing position)
"LP" key (last position)
"AP" key (activate automatic posi­tion)
"Backrest down" key
"AP 1" key (automatic position 1)
"Backrest up" key
"AP 2" key (automatic position 2)
"Collapsed position" key
Group of keys for illumination
Key Name Control element
"Operating light" key Assistant element
"X-ray image viewer" key Dentist element
Group of keys for hygiene
Key Name Control element
"Tumbler filler" key Dentist element and
assistant element
29 / 126
Page 30
Instructions for use Primus 1058 Life
3 Product description | 3.9 Controls
30 / 126
Key Name Control element
"Bowl rinsing" key Dentist element and
assistant element
"Intensive germ reduc­tion" key
Assistant element (op­tional)
"HYDROclean" key Assistant element
Group of keys for the handpieces/timer
Key Name Control element
"Preselected spray" but­ton
Dentist element
"Direction of motor rota­tion" button
Dentist element
"Timer" key Dentist element and
assistant element
Group of keys for the menu
Standby menu with MEMOspeed Standby menu without MEMOspeed
Menu function selection keys Display
3.9.4 Foot control
The footswitches of the foot control have two functions. The functions of the footswitches depend on if an instrument is mounted or removed.
Page 31
Instructions for use Primus 1058 Life
3 Product description | 3.10 Rating plate and identification plate
Item Name Function with hand-
piece mounted
Function with hand­piece removed
"Spray pre-selection/
AP" footswitch
Moves the dental chair into automatic posi­tion.
Sets the spray pre-se­lection.
U-shaped switch Turns the safety
shutoff On.
Switches the foot­switches to the "Chair motion" function.
"Blown air/AP" foot-
switch
Moves the dental chair into automatic posi­tion.
Sets the preset blown air (chip blower).
Cross-switch: "Coun-
terclockwise motor ro­tation"
Changes the position of the dental chair.
Selects the direction of motor rotation (for IN­TRA LUX motor KL 701/703 or COM­FORTdrive 200XD).
"Handpieces" foot-
pedal
Generates a video/ freeze frame if CONEX­IOcom is installed.
Starts the motor and controls the speed/ in­tensity of the handpieces.
3.10 Rating plate and identification plate
Rating plate
Rating plates inside and outside
31 / 126
Page 32
Instructions for use Primus 1058 Life
3 Product description | 3.10 Rating plate and identification plate
32 / 126
Internal attachment site for rating plate
Outside attachment site for rating plate
SN Serial number
Read and take note of the content of accompanying documents
Please note the instructions for use
Follow the instructions for use!
Type B applied part
Type BF applied part
Page 33
Instructions for use Primus 1058 Life
3 Product description | 3.10 Rating plate and identification plate
Operating mode: Operating time of the patient chair: 25 seconds Pause time of the patient chair: 400 seconds (The permissible operating times correspond to common dental pro­cedure.)
Fuse ratings: The "?????" depend on the mains voltage and are either T10 H or T6.3H. 100 V~ ,110 V~, 120 V~, 130 V~ = T10H 220 V~ , 230 V~, 240 V~ = T6.3H
For disposal information, see also: Purpose - Intended use
CE mark according to Medical Devices Directive EC 93/42
VDE mark
DVGW certification DVGW CERT registration number AS-0630BT0111
HIBC Code
eLabeling ID
Identification plates
Rating plate and dentist element ID
Site of attachment of rating plate and type BF applied parts ID on dentist element
33 / 126
Page 34
Instructions for use Primus 1058 Life
3 Product description | 3.10 Rating plate and identification plate
34 / 126
Nameplate dentist element (e.g. table TM) / marking of the applied parts of type BF
Type Device type SN Year of manufacture - serial number REF Material number
Labelling and marking of the three-function handpiece and multifunctional handpiece
Company logo of the manufacturer
SN Serial number
CE mark according to 93/42/EEC medical devices
Sterilisable up to 135 oC
Disposal instructions according to Directive WEEE 2002/96/EG Annex N
Follow instructions for use
Page 35
Instructions for use Primus 1058 Life
3 Product description | 3.10 Rating plate and identification plate
Rating plate Service table 1568
Site for affixing the rating plate
4
5
6
7
Device type ② Material numberYear and month manufactured -
Serial number
HIBC Code
Read and take note of the content
of accompanying documents
CE mark
GOST ID
35 / 126
Page 36
Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
36 / 126
3.10.1 Rating plate 1440
1 3
2 4
5
6
8
9
7
Rating plate X-ray viewer 1440
Device type SN: Year and month manufactured
- Serial number
Material number ④ Supply voltage, frequencyPower ⑥ Disposal instructionsCE mark ⑧ Serial numberYear and month manufactured
3.11 Technical data
Drilling template and setup plan
Installation plan (Mat. no.
3.002.4533)
2 sheets each right-handed and 2 sheets left-handed
Installation plan with COMPACTchair (Mat. no. 1.003.6767)
2 sheets each right-handed and 2 sheets left-handed
Page 37
Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
Electrical system
Electrical lead 3 x 2.5 mm
2
Free end above the floor 1 000 mm
Input voltages 100/110/120/130/220/230/240 V AC
Frequency 50/60 Hz
Input voltage set by the manufac­turer
See rating plate
Power consumption at 100 to 240 V 100 to 600 VA – with appropriate
device configuration, deviations in this range are possible!
Customer-provided fuse protection Automat C 16 or screw-plug fuse
10 A
Protective conductor above floor See DIN VDE 0100-710, 1000 mm
Heat emission 360 to 3,240 kJ/h
Heat emission Ø 900 kJ/h
Mark of approval CE / DVGW / VDE
Foot control IPX1: Protection against water drips
Triple-function handpiece and multifunctional handpiece
Flush the water and air passages for 20 to 30 seconds before working at the beginning of the day.
Water pressure 1.5 ± 0.3 bar; Flow pressure; 4 x
manometer
Max. static pressure water 2.5 ± 0.3 bar
Water flow 80 ± 10 ml/min
Air pressure 3.3 ± 0.1 bar; Flow pressure; 4 x
manometer
Max. static pressure air 4 + 0.5 bar
Air flow at least 16 Nl/min
Operating time (multifunctional handpiece only)
1 minute
Interval (multifunctional handpiece only)
3 minutes
37 / 126
Page 38
Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
38 / 126
Electrical multifunctional handpiece
Safety extra-low voltage according to DIN EN 60601-1:
24 V AC ± 10% (non-grounded voltage)
Frequency 50/60 Hz
Type of use BF
Heat output for water approx. 90 W
Heat output for air approx. 20 W
Lamp voltage max. 3.2 V ± 0.15 V
High-pressure lamp power max. 2.5 W
Water supply
Note
If the water is very hard (above 12 °dH), a water softening device must be fitted in the ion-exchange process.
Insufficient water hardness (below 8.4 °dH) can promote the formation of al­gae.
Note
The "water inlet block" assembly kit does not include a separation between the treatment water and water supplied by the local mains. The operator must observe and comply with relevant national directives concerning the prevention of backflow. Upon failure to comply with these rules, the manu­facturer accepts no liability for the quality of the treatment water and the mi­crobial re-contamination of the public drinking water network.
Note
In conjunction with the "DVGW water block with integrated water germ re­duction" a water germ reduction unit is installed in dental units from KaVo. The germ reduction liquid OXYGENAL 6 is continually added to the water at a concentration which is harmless for persons, but hygienically effective to maintain the quality of the treatment water. The handling is described in the care instructions of the treatment centres. Supplementary measures such as the rinsing of water conducting lines and intensive germ reduction must be carried out according to the instructions of the manufacturer.
WARNING
There is a risk of infection if you fail to comply with national regula­tions.
Contamination of the treatment water and/or drinking water supply with germs.
Note and comply with national regulations concerning the quality of water
for human use (drinking water) - if applicable.
Note and comply with national regulations concerning the prevention of
backflow (flow from the treatment centre back to the public water supply) ­if applicable.
Page 39
Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
WARNING
Risk of infection if the "Water block, compact" is used without addi­tional safeguards.
Contamination of the treatment water and/or drinking water supply with germs.
With regard to the "Water block, compact" assembly kit, please note that
no germ reduction facility is installed in the unit, and take appropriate safe­guards. KaVo recommends to use the "Water block DVGW with integrated water germ reduction system" in combination with KaVo OXYGENAL 6 (Mat. no. 0.489.3451).
If the Water bottle kit is used with the enclosed dosing attachment (Mat.
no. 1.002.0287), add the proper amount of KaVo OXYGENAL 6 (Mat. no.
0.489.3451) with each filling. For the correct amount, please refer to the
instructions of the dosing attachment for water germ reduction.
According to DIN EN 1717, each unit that is not listed by DVGW must be provided with an upstream type AA, AB or AD safety device. (The DVGW water bottle kit is certified; see the following list.) When establishing a water connection, make sure that there are no sections of brackish water with standing water (also in the house plumbing).
For further information, please refer to www.dvgw.de
Free drainage according to DIN EN 1717 - DVGW certified
Water block DVGW, water bottle DVGW, register no.: AS-0630BT0111
Water quality Tap water, cold water connection
Water hardness 1.5 to 2.14 mmol/l 8.4 to 12 °dH
pH 7.2 to 7.8
Customer-site water filtering 80 microns
Water connection Shut-off valve with brass cone com-
pression screw connection 3/8" to Ø 10 mm provided
Above-floor water connection min. 50 mm, max. 105 mm with
valve opened
Water inlet pressure 2.0 to 6.0 bar
Water inflow 4 l/min
Diameter of the drain connection 40mm
Above-floor drain connection 20mm
Outflow quantity max. 4 l/min
Slope of water drain pipe downstream from device: at least 10
mm per metre
39 / 126
Page 40
Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
40 / 126
Air supply
WARNING
Non-compliance with national guidelines concerning the quality of the dental air.
Infection hazard.
Note and comply with the national guidelines concerning the quality of the
dental air - if any.
Blow through the air line prior to commissioning.
Air inlet pressure 5.2 to 7 bar
Minimum air flow rate min. 80 Nl/min
Pressure dew point < -30 ºC (compressor with
dry air system)
Oil content < 0.1 mg/m3 (oil-free compressor)
Contamination < 100 particles/m3 at particle sizes
of 1 to 5 µm
Customer-site air filtration 50 microns
Air connection Shut-off valve with brass cone com-
pression screw connection 3/8" to Ø 10 mm provided
Air connection above floor level min. 50 mm, max. 105 mm with
valve opened
Suction
Pressure drop at the connecting point
Semi-dry and wet suction Dry suction
Note
If the negative static pressure is > 180 mbar, the unit must be equipped with the negative pressure regulating valve assembly kit.
Page 41
Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
Cannula connecting pieces according to DIN EN ISO 7494 - 2
Diameter of cannula of saliva ejector and surgical suction
7mm
Diameter of cannula of spray mist suction
15mm
Suction air quantity at spray mist cannula
minimum 250Nl/min (suction system with high flow rate), recommended 300Nl/min
Diameter of the suction connection 40mm
Suction connection above floor 20mm
Operating environment
WARNING
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
Floor quality The quality of the flooring must meet
the load bearing ability for buildings DIN 1055 page 3 and have a pres­sure resistance in accordance with DIN 18560 T 1.
Ambient temperature +10 to +40 oC
Relative humidity 30 to 75%
Air pressure 700 hPa - 1,060 hPa
Max. elevation for operation up to 3,000 m
41 / 126
Page 42
Instructions for use Primus 1058 Life
3 Product description | 3.11 Technical data
42 / 126
Maximum loads
Max. patient weight load on patient chair Standard
185 kg
Max. patient weight load COMPACTchair
135 kg
Tray holder of the dentist element ­loadable up to
2 kg
Assistant element standard tray ­loadable up to
1 kg
Dentist element - loadable up to 2 kg
Service table 1568 without locking system
2 kg
Service table 1568 with locking sys­tem
5 kg
Transportation and storage conditions
Ambient temperature -20 to +55 oC
Relative Humidity 5% to 95% non-condensing
Air pressure 700 hPa to 1,060 hPa
Weight
Treatment centre (TM) with patient chair Standard
223 kg gross, 182 kg net
Includes steel mounting plate and patient communication
333 kg gross, 287 kg net
Treatment centre (TM) with COM­PACTchair
206 kg gross, 158 kg net
With steel mounting plate and pa­tient communication
316 kg gross, 263 kg net
Package KaVo Service table 1568
Length approx. 1040 mm
Width approx. 810 mm
Height approx. 240 mm
Gross weight approx. 25 kg
Net weight approx. 19 kg
For more information about the packages, please refer to Assembly Instructions
Page 43
Instructions for use Primus 1058 Life
3 Product description | 3.12 KaVo Service table 1568 (optional accessory)
X-ray viewer 1440
Input voltage 24 V AC
Frequency 50/60 Hertz
Power consumption max. 35 VA
ON-time 100 %
Lights 2x Osram Lumilux de Luxe G5 Day-
light L8W/954
Light field dimensions 300 mm x 150 mm in accordance
with DIN 6856-3
Case dimensions 340x216x48 in accordance with DIN
6856-3
Operating light KaVoLUX 540 LED
See also:
2 Instructions for Use KaVoLUX 540 LED
3.12 KaVo Service table 1568 (optional accessory)
Swivel arm ② Spring armRotary knob (brake) ④ Rotary knob (lockable)Service table ⑥ Non-slip matCup holder ⑧ HandleRating plate ⑩ X-ray viewer 1440
43 / 126
Page 44
Instructions for use Primus 1058 Life
4 Operation | 4.1 Switching the device on and off
44 / 126
4 Operation
4.1 Switching the device on and off
Note
Always switch the machine off before leaving the office.
Switch on the device using the main switch.
ð The display of the dentist element shows the preselected basic menu. ð The green LED "Device turned on" lights up on the assistant element ②.
Basic menu without MEMOspeed / basic menu with MEMOspeed / assistant element
Note
Activate the KaVoLUX 540 LED operating light using the "Operating light" key on the assistant element. Only then the operating light can be operated by means of the sensor and the control panel of the operating light.
Page 45
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
4.2 Adjusting the dental chair
4.2.1 Adjusting the arm rest (optional)
Armrest for the standard dental chair
To make it easier for the patient to sit in the chair, the armrest can be swung up.
CAUTION
The patient's hands are in a bad position when the chair is rising
Danger of crushing fingers between the backrest and arm rest.
Make sure that the patient is sitting in the right position (especially chil-
dren).
Arm rest for patient chair COMPACTchair
To make it easier for the patient to sit in the chair, the arm rest of the patient chair can be swiveled forward.
Swivel the arm rest forwardThen swivel the arm rest back into place.
45 / 126
Page 46
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
46 / 126
4.2.2 Adjust head rest
Adjust double-jointed knob headrest with knob
CAUTION
Adjusting the headrest.
Injury to the neck muscles.
Make sure that the patient is aware of the headrest being adjusted.Patients need to raise their head slightly during adjustment.Use both hands to adjust the headrest.
Push in or pull out the headrest depending on the patient's size.
To swing the headrest, turn the locking dial to the left, move the headrest
into position, and turn the dial to the right to lock it.
To remove the headrest cushion, remove the screw ②, pull the cushion
up slightly, and remove it to the front.
Page 47
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
Setting the push button of 2-joint headrest (optional)
CAUTION
Adjusting the headrest.
Injury to the neck muscles.
Make sure that the patient is aware of the headrest being adjusted.Patients need to raise their head slightly during adjustment.Use both hands to adjust the headrest.
The bar length and angle of the headrest can be adjusted.
Press the lock button and push in or pull out the headrest depending on the
patient's height.
Note
The service technician can adjust the braking force.
Press the lock button and swing the headrest into the desired position.
When swinging the headrest back into position, make sure that there is nothing between the area A and head cushion.
Turning the head cushion
The head rest cushion is a rotating cushion. It can be turned to offer better neck support, for example when treating children.
Evenly pull the cushion up and rotate it 180o. ▶ Then mount and push the head cushion back on.
47 / 126
Page 48
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
48 / 126
4.2.3 Positioning the dental chair manually
CAUTION
Danger of injury from overload or dynamic load.
Patient chair may be damaged by overloading it.
Do not subject the patient chair to a load exceeding its limit (Standard pa-
tient chair 185 kg/ COMPACTchair patient chair 135 kg).
Do not subject the patient chair to dynamic loads.
CAUTION
Motorised movement of the chair
The patient or treatment personnel can be clamped or crushed.
Monitor the patient and treatment personnel when changing the patient's
position.
CAUTION
Risk of injury when moving the patient or patient chair.
The patient or treatment personnel can be pinched or crushed.
Position all moving parts, such as dentist element, assistant element, oper-
ating light, screens, etc., outside the collision range when you move the pa­tient or patient chair.
Positioning the chair and backrest manually using the dentist or assistant unit
Use the following buttons to adjust the chair height and position of the back­rest:
Key Dentist element
Key Assistant element
Feature
The chair moves up.
The chair moves down.
The backrest moves up­ward.
The backrest moves downward.
Press the related key.
ð The chair or backrest moves in the desired direction.
Page 49
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
Positioning the chair and backrest manually using the foot control
The cross switch of the foot control assumes the function of the button wheel on the dentist element during manual positioning of the patient chair.
Requirement
All instruments are in their holder.
▶ Chair up: Move the cross switch on the foot control in direction ①. ▶ Chair down : Move the cross switch on the foot control in direction ③. ▶ Backrest up: Move the cross switch on the foot control in direction ②. ▶ Backrest down: Move the cross switch on the foot control in direction ④.
4.2.4 Automatic positioning of dental chair
CAUTION
Danger of injury from overload or dynamic load.
Patient chair may be damaged by overloading it.
Do not subject the patient chair to a load exceeding its limit (Standard pa-
tient chair 185 kg/ COMPACTchair patient chair 135 kg).
Do not subject the patient chair to dynamic loads.
CAUTION
Danger of crushing during automatic chair movement.
The patient or treatment personnel can be clamped.
Monitor the patient and treatment personnel when changing the chair posi-
tion.
CAUTION
Risk of injury when moving the patient or patient chair.
The patient or treatment personnel can be pinched or crushed.
Position all moving parts, such as dentist element, assistant element, oper-
ating light, screens, etc., outside the collision range when you move the pa­tient or patient chair.
Gradually adjust the chair position
Save chair positions
The chair positions can be saved and retrieved at any time by the press of a button. Win the position is retrieved, the chair automatically moves to the saved position (the so-called "automatic position," or "AP" for short).
49 / 126
Page 50
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
50 / 126
The four chair positions can be saved on the control panels. Two of these four positions can be saved with the foot control. It is for example recommendable to save the sitting down and getting up posi­tion using the "AP 0" key and the rinsing position with the "SP" key.
Recalling automatic positions with the dentist unit
The following keys can be used to recall saved chair positions.
Key Operation
Move to the rinsing position.
The last position before actuating the SP is assumed.
Move to automatic position 0.
Move to automatic position 1.
Move to automatic position 2.
Move to the collapsed position.
Briefly press the desired button.
ð Chair automatically moves to the stored position. ð Upon arrival at the stored position, the display diode on the button is turned
on.
Saving automatic positions with the dentist unit
Recommended assignment of buttons: "SP" button: rinsing position "AP 0" button: entry and exit position "AP 1" button: treatment position, e.g. for lower jaw treatment "AP 2" button: treatment position, e.g. for upper jaw treatment "Collape position" button: collapse position
Move the chair to the desired position.To save the chair position, press "AP 0", "AP 1", "AP 2", "SP" or "Collapsed
position" button until you hear a signal.
ð The display diode of the pressed button is turned on. The chair position is
saved.
Last position
After the "LP" button is pressed, the chair moves into its position before the "SP" button was pressed.
Page 51
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
Note
The memory is erased when you turn off the device. After turning on the device again (for example in the morning or after lunch), the chair does not execute a specific movement when you press the "LP" button.
Recalling automatic positions with the assistant unit
Briefly press the "AP" key.
ð The LEDs of the "AP 0", " AP 1", " AP 2", "SP", and "LP" keys flash for ap-
proximately four seconds.
During these four seconds, briefly press the "AP 0", " AP 1", " AP 2", "SP" or
"LP" key.
ð The chair moves into the selected automatic position.
Saving automatic positions with the assistant unit
Note
The automatic position "Last position" is saved on the "LP" button. Press the "LP" button for the chair to automatically move to the last position before the rinsing position. The "LP" button cannot be assigned to another automatic position.
Move the chair to the desired position.
Briefly press the "AP" key.
ð The LEDs of the "AP 0", " AP 1", " AP 2", "SP", and "LP" keys flash for ap-
proximately four seconds.
During these four seconds, press the "AP 0", "AP 1", "AP 2", "SP" or "LP"
button, until a signal sound is transmitted.
ð LED of the pressed button lights up. The chair position is saved.
Recalling automatic positions with the foot control
Note
If an instrument is removed, the chair functions of the foot control are blocked. The blocking can be removed by briefly pressing the stirrup switch. The functions are then available.
51 / 126
Page 52
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
52 / 126
Spray preselection/AP footswitch ② Blown air/AP footswitchFoot pedal
The chair positions can be recalled with two foot switches; the standard setting is as follows:
▪ "Spray selection" foot switch: automatic position "LP" (last position) ▪ "Blown air" foot switch: automatic position "SP" (rinsing position)
Move the chair when the instrument is mounted
Press the "SP" foot-operated button.
or
Press the "LP" foot-operated button.
ð The chair moves into the selected automatic position.
Move the chair when the instrument is removed
Note
If an instrument is removed, the chair functions of the foot control are blocked. The blocking can be removed by briefly pressing the stirrup switch. The functions are then available.
Press the stirrup switch and then the "Preselected spray" or "Blown air" foot
switch.
ð The chair moves into the selected automatic position.
Page 53
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
Saving an automatic position with the foot control
Spray preselection/AP footswitch ② Blown air/AP footswitchFoot pedal
The chair positions can be saved on two footswitches; the standard setting is as follows:
▪ "Spray default" footswitch: "LP" automatic position (last position) ▪ "Blown air" footswitch: "SP" automatic position (rinsing position)
Hold down the foot pedal and foot-operated button "SP", and simultan-
eously press any button for an automatic position ("AP 0", "AP 1", "AP 2" or "SP") on the dentist or assistant unit until you hear a beep.
ð The automatic position is saved to the foot-operated button.
or
Hold down the foot pedal and foot-operated button "LP", and simultan-
eously press any button for an automatic position ("AP 0", "AP 1", "AP 2" or "SP") on the dentist or assistant unit until you hear a beep.
ð The automatic position is saved to the foot-operated button.
4.2.5 Safety shut-off
To prevent collisions arising from the movement of the patient chair, safety shutoff switches are installed to protect the patient and practice personnel from injury and the treatment centre from damage.
NOTICE
Damage to the assistant element and dental chair.
Despite some safety shut-downs being present, certain positions of the assistant element may collide with the dental chair.
Keep the assistant element out of the motion range of the patient chair.Always monitor the chair movement.
53 / 126
Page 54
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
54 / 126
CAUTION
Pinching from the treatment chair.
The safety shutoff of the treatment chair is activated by lifting the respective component. Depending on the patient's body weight and the leverage, more force can be exerted on the object to be triggered than is necessary to trigger the switching function.
The treatment personnel must move outside of the chair's swivelling range
whenever the chair moves.
The safety cut-offs can be found at the following places on the treatment unit.
Safety shutoff for the standard patient chair
Patient unit pivoted over the patient
chair
Assistant element
Backrest ④ Bracket on the foot controlKick plate ⑥ Seat
Item No.
Safety switch-off ac­tuated
LED on assistant element
LED on dentist ele­ment
Patient unit pivoted
over the patient chair
Assistant element
Backrest
Bracket on the foot
control
Base plate
Bench
Page 55
Instructions for use Primus 1058 Life
4 Operation | 4.2 Adjusting the dental chair
Safety shutoff for the COMPACTchair patient chair
Patient element pivoted over dental
chair
Assistant element
Backrest ④ Bracket on the foot controlBench support / seat cushion ⑥ Foldable part of the seat
Item No.
Safety switch-off ac­tuated
LED on assistant element
LED on dentist ele­ment
Patient unit pivoted
over the patient chair
Assistant element
Backrest
Bracket on the foot
control
Bench support / seat
cushion
Foldable part of the
bench
The safety shutoff occurs if a movement angle has been exceeded or if a part of the treatment centre collides with an object.
If a person or object actuates a safety shutoff, the chair immediately stops moving. The fact that the safety shutoff has been activated is displayed by the corres­ponding display flashing on the dentist or assistant element.
55 / 126
Page 56
Instructions for use Primus 1058 Life
4 Operation | 4.3 Moving the patient chair
56 / 126
Note
The chair's position cannot be changed with the key wheels when a safety shutoff is activated.
Exception: The patient unit safety switch only stops the upward and down­ward movement of the patient chair. The backrest can be moved up and down.
4.3 Moving the patient chair
Patient chair Standard
Patient chair COMPACTchair
Page 57
Instructions for use Primus 1058 Life
4 Operation | 4.4 Move the dentist's unit
4.4 Move the dentist's unit
NOTICE
Damage from overloading the dentist element.
Exceeding the maximum weight of more than 2 kg by adding handpieces, ac­cessories, etc., can cause damage.
Do not overload the dentist element!
CAUTION
Risk of injury when the dentist or assistant element is moved.
The patient or office staff may be injured or bruised.
Monitor the patient and office staff when moving the dentist or assistant
element.
The swinging range of the dentist unit is limited by stops.
Note
Do not pull the dentist unit by the instrument hose.
To adjust the dentist unit height, release the brake, adjust the height, and
reset the brake.
Dentist element TM
Dentist element S
57 / 126
Page 58
Instructions for use Primus 1058 Life
4 Operation | 4.4 Move the dentist's unit
58 / 126
4.4.1 Move the cart
CAUTION
Moving and overloading the cart.
Danger of tipping and damaging the cart.
Only use the card on a continuously smooth floor.Do not overextend the supply hose for the cart.Make sure that there are no obstructions on the floor.Do not sit on the dentist element or step on the castor.
Note
The area in which the cart can be move is restricted by the length of the lines and hoses that connect the cart to the base of the device. Only move the cart within this range.
To change the position of the cart, hold the cart by the bow-type handle
and move it to the desired position. Make sure that there are no obstruc­tions on the floor.
The top part of the dentist's unit can be positioned in 9 levels.
Note
Do not lift the dentist's unit using the handle. The handle is only for horizont­ally positioning the dentist unit.
739 - 946
Lift the top part of the dentist's unit until it locks into place.To release the lock, move the top part all the way up and then move it
down.
Page 59
Instructions for use Primus 1058 Life
4 Operation | 4.5 Move the patient unit
4.5 Move the patient unit
4.5.1 Swing the patient unit by hand
The swinging range is about 250o.
CAUTION
The left armrest can collide with the manually adjusted patient's unit when the chair moves.
Injury hazard.
Each time before the chair is adjusted (automatic and manual), swing the
manually adjusted patient's unit into resting position.
Note
When the patient unit is swung over the patient chair, the safety shutoff is activated.
4.6 Moving the assistant element
4.6.1 Adjusting the height of the Standard assistant element
The assistant unit can be vertically positioned in four levels.
To set a higher level, pull the assistant unit upward gently until it audibly
locks in place.
59 / 126
Page 60
Instructions for use Primus 1058 Life
4 Operation | 4.6 Moving the assistant element
60 / 126
To set a lower level, pull the assistant unit all the way up until the lock re-
leases, and then lower the assistant unit.
Mounting the tray holder
Mount the tray holder on the assistant element.
Tray support ② Air nozzle holder
The support for the tray holder ①is an optional accessory.
4.6.2 Moving the assistant element right, left (optional)
CAUTION
Pinching from the treatment chair.
The treatment staff can get pinched or crushed.
The treatment personnel must move outside of the chair's swinging range
whenever the chair moves.
NOTICE
Material damage caused by overloading.
Do not rest your foot near the pivot point and/or transverse arm of the as-
sistant element.
Page 61
Instructions for use Primus 1058 Life
4 Operation | 4.6 Moving the assistant element
Pivoting range of assistant element r, l (optional)
Move the backrest up before swinging the assistant element.Move the assistant element to the desired position in its swinging range.
Adjusting the height of the assistant element right, left (optional)
Note
Handpieces may drop out of the holders while the assistant element is being moved, especially during adjustment of the height. In order to prevent hand­pieces from being damaged, make sure that no handpiece drops down while you move the assistant element.
Undo the clamping screw and push the assistant element into the desired
position.
Re-tighten the clamping screw.
61 / 126
Page 62
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
62 / 126
4.7 Using functions through the menu
4.7.1 Using the user menu
The following options can be opened in the user menu:
Op­tion
Feature Description
1 Firmware Display current firmware version. 2 Time of day Set time of day. 3 Date Set date. 4 Time display mode Setting the time of day display
mode:
▪ Time of day only ▪ Time of day without seconds
5 Language Set menu language:
▪ Deutsch ▪ English ▪ Italiano
▪ Français 6 LCD Set contrast of LCD display. 7 Licenses Display of activated licenses
The functions in the menu are used through the selections keys (S1 to S4) on the display
User menu with MEMOspeed / without MEMOspeed
Press the "Next" key (S4) to start-up the user menu.
ð The user menu displays options and parameters that can be set and
changed by the user.
Press the "Next" key (S4) to switch to the next option.
Page 63
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
Option 1: Displaying the firmware version
The firmware version is displayed
Option 2: Setting the time of day
Press the "SET" (S2) key to change the values of minutes and hours.
ð The value to be changed flashes.
Press the "Save" (S4) key to save the selection made.
Press the "reduce value" or "increase value" key to set the marked time of
day.
Press the "SS:MM" (S1) key to switch between hours and minutes.
Press the "Save" (S4) key to save the values and switch to the SET display.
63 / 126
Page 64
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
64 / 126
Option 3: Set the date
Press the "SET" (S2) key to change the values of day, month, and year.
ð The value to be changed flashes.
Press the "Save" (S1) key to save the selection made.
Press the "Decrease value" or "Increase value" key to set the marked value.
Press the "SS:MM" (S1) key to switch between day, month, and year.
Press the "Save" (S4) key to save the values and switch to the SET display.
Option 4: Setting the display mode for time of day
Press the "Decrease value" or "Increase value" keys to set the time of day
display mode.
The following views can be selected:
- Time of day only
- Time of day (without seconds)
Page 65
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
Option 5: Setting the language
Press the "Decrease value" or "Increase value" key to select the menu lan-
guage. The following languages can be selected: Deutsch, English, Italiano, Francais.
Press the "Save" (S1) key to save the values.
Option 6: Setting the display contrast
Press the "reduce value" or "increase value" key to set the contrast of the
LCD display.
Press the "Save" (S1) key to save the values.
Option 7: Displaying the licenses
Displays the activated licenses:
▪ PLED: PiezoLED ▪ FCR: Foot control ▪ BELL: Bell ▪ MS: MEMOspeed
65 / 126
Page 66
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
66 / 126
4.7.2 Standby menu
Standby menu as default setting
The unit starts in the standby menu.
Select function
The display shows display fields with symbols for the operating functions. Below each display field, there is a key for selecting the displayed operating function.
Standby menu with MEMOspeed Standby menu without MEMOspeed
Menu function selection keys Display
Select dentist
The first symbol in the stand-by menu displays the current user.
Press the "S1" key to select dentist 1 or dentist 2.
Permitting level switching (available with MEMOspeed only)
Level switching is deactivated in the basic state. The level switching symbol displays the current dentist.
Page 67
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
Note
The device acts like on level E when level switching is deactivated.
To switch between levels, keep the "Direction of motor rotation" and the
"Bowl flushing" keys and the foot pedal depressed, until a beep can be heard.
After activating level switching, the level switching symbol shows the level (E, 1 2 or 3 - level E is selected in the example shown). The pre-selected dentist is only displayed very small in the level switching symbol.
Note
The device automatically saves the activation of level switching for the cur­rent dentist.
Note
Level switching is deactivated using the same key combination as activation.
To select a level, briefly press the selection button for "Preselect level".
Selection of dentist with level switching activated
Press the "Pre-select level" key long to select dentist 1 or dentist 2.
Status display in the Standby menu
If a status message is available, the standby menu shows an exclamation mark on selection key "S2" .
1
1
Press the "S2" selection key ① to display status messages.
67 / 126
Page 68
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
68 / 126
Press the selection keys "+" ② and "-" ① to switch between multiple status
messages.
Press the "OK" selection key ③ to exit from the display of status messages.
Error messages in the status display
See also:
2 9 Troubleshooting, Page 114
Page 69
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
4.7.3 Operating the MEMOspeed menu (optional)
The MEMOspeed menu is used to display and set handpiece-specific values.
The display depends on which instrument was withdrawn.
To save the handpiece-specific values, there are 3 memory levels (1, 2, 3) each available for two dentists (dentist 1 and dentist 2).
Save instrument-specific settings
The following settings can be individually:saved for the instruments:
Handpiece Setting
Turbine Speed range (available with level switching only)
Preselected spray
Motor INTRA LUX KL 701/703, COMFORTdrive
Speed range (available with level switching only) Direction of motor rotation Preselected spray
Ultrasonic scaler Spray on/off*
Intensity (with level switching only)
Multifunctional handpiece Heating on/off
* only with a corresponding setting in service mode
Changing turbine settings in the menu
Note
Following instructions for use, service instructions and installation instructions in the instrument packaging.
Note
The speed can be set at level E only with the foot pedal. The speed cannot be saved in level E.
Remove the turbine from the holder.Briefly press the "Preselect level" (S1) key to select the level.Press the "Preselect level" (S1) key for 4 seconds to change settings.
ð The display switches to the turbine settings menu.
Settings menu for minimal/maximal speed
Press the "min/max" (S4) keyed to toggle between the settings menus for
minimal and maximal speed.
Press the "Decrease value" key to decrease the speed.
or
69 / 126
Page 70
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
70 / 126
Press the "Increase value" key to increase the speed.
ð The intensity is shown on the display.
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð Saving is acknowledged by a beep. ð This closes the "Settings" menu.
Setting the cooling level
Requirement
Settings menu for turbine is selected.
Press the "Preselected spray" button.
or
T
M
Press the "Preselected spray" footswitch. (with level switching only)
ð The set values are saved each for the set memory level and the set dentist
level.
Key Feature
No LED is on: No cooling
One LED is on: Spray air cooling status
Both LEDs are on: Spray cooling status
Changing motor settings in the menu
Note
Following instructions for use, service instructions and installation instructions in the instrument packaging.
Note
The motor mode is equivalent to 2 minutes operating time and 5 minutes pause.
This represents the possible maximum load of the motor (full load at max­imum speed).
In practice, pulse loads lasting seconds or pause times lasting seconds or minutes are realistic given that the maximum possible motor current is not normally reached. This equates to the dentist's normal way of working.
Remove the motor from the holder.Briefly press the "Preselect level" (S1) key to select the level.Press the "Preselect level" (S1) key for 4 seconds to change settings.
ð The display switches to the motor settings menu.
Page 71
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
Setting the speed
Note
The speed can be set at level E only with the foot pedal. The speed cannot be saved in level E.
Motor KL 701/KL 703 COMFORTdrive 200XD
Minimum 100 rpm 30,000 rpm Maximum 40,000 rpm 200,000 rpm
Settings menu for minimal/maximal speed
Press the "min/max" (S4) keyed to toggle between the settings menus for
minimal and maximal speed.
Press the key for "Decrease value" to decrease the speed.
or
Press the "Increase value" key to increase the speed.
ð The speed is shown in the display.
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð Saving is acknowledged by a beep. ð This closes the "Settings" menu.
Setting the cooling level
Requirement
Settings menu for motor is selected.
Press the "Preselected spray" button.
or
T
M
Press the "Preselected spray" footswitch.
ð The set values are saved each for the set memory level and the set dentist
level.
Key Feature
No LED is on: No cooling
71 / 126
Page 72
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
72 / 126
Key Feature
One LED is on: Spray air cooling status
Both LEDs are on: Spray cooling status
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð Saving is acknowledged by a beep. ð This closes the "Settings" menu.
Setting the direction of motor rotation
Note
The direction of motor rotation can only be changed when the motor is at rest.
Requirement
Settings menu for motor is selected.
Press the "Direction of motor rotation" key.
or
T
M
Press the "Direction of motor rotation" foot pedal.
ð The direction of motor rotation is reversed each time the cross-switch and/
or the "Direction of motor rotation" key is actuated: counterclockwise rota­tion - clockwise rotation.
ð The LED is on when CCW motor rotation is set.
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð Saving is acknowledged by a beep. ð This closes the "Settings" menu.
Changing the ultrasonic Scaler PiezoLED in the menu
Note
Please comply with the enclosed "PiezoLED" Instructions for Use.
Take the PiezoLED off the holder.Briefly press the "Preselect level" (S1) key to select the level.Press the "Preselect level" (S1) key for 4 seconds to change settings.
ð The display switches to the settings menu of the PiezoLED.
Setting the intensity
The intensity is set in steps of 0.25; the minimum intensity is 1.0, and the max­imum is 10.0.
Page 73
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
Settings menu for minimal/maximal intensity
Press the "min/max" (S4) keyed to toggle between the settings menus for
minimal and maximal intensity in the set operating mode.
Press the "Decrease value" key to decrease the intensity.
or
Press the "Increase value" key to increase the intensity.
ð The intensity is shown on the display.
Define operating mode (PiezoLED only)
Note
The selection of the mode depends on the treatment method and the tip used. For information about the selection of an operating mode, please refer to the "Operating modes P1 / P2 / P3 and E" section of the "PiezoLED In­structions for Use".
Take the PiezoLED off the holder.Press the "Mode" key to select the operating mode.
Modes P1 / P2 / P3 / E are available for selection.
Setting the cooling status
Requirement
Settings menu for PiezoLED is selected.
Press the "Preselected spray" button.
or
T
M
Press the "Preselected spray" footswitch.
73 / 126
Page 74
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
74 / 126
Key Feature
No LED is on: No cooling
Both LEDs are on: Spray cooling status
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð Saving is acknowledged by a beep. ð This closes the "Settings" menu.
Changing the settings of the multifunctional handpiece in the menu
Briefly press "Preselect level" key to select the level.Take the multifunctional handpiece off the holder.Press "Preselect level" button for 4 seconds in order to change settings.
ð The display changes to the multifunctional handpiece settings menu.
Adjusting the air/water heating
Settings menu for multifunctional handpiece
Adjust the heating using the "Air/water heating" button.
Symbol Function
Air/water heating "on"
Air/water heating "off"
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð Saving is acknowledged by a beep. ð This closes the "Settings" menu.
Page 75
Instructions for use Primus 1058 Life
4 Operation | 4.7 Using functions through the menu
4.7.4 Using the CONEXIOcom (optional)
Note
To start the CONEXIOcom menu, no handpiece may be removed.
Note
For all CONEXIOcom functions, the dental unit must be connected to an in­stallation of the KaVo "CONEXIO" software.
The function of the CONEXIOcom menu is to control the display of previously recorded and saved images and videos. In order to use the function, the unit must have access to the data of the KaVo Software "CONEXIO" software. For details on the configuration, please refer to the "CONEXIO" installation instruc­tions.
Opening the CONEXIOcom menu
To display existing images, take the camera off the holder. Select the proper patient on the corresponding PC for this purpose. It is also feasible to automat­ically transfer the patient from your invoicing programme to CONEXIO. For de­tails on the configuration, please refer to the "CONEXIO" installation instruc­tions.
If no patient is selected, images from the clipboard are displayed. If the clip­board is empty, no image is displayed. The clipboard is deleted automatically when the patient is logged off the corresponding PC. The CONEXIOcom menu is opened automatically for recording of images or videos as soon as a device (DIAGNOcam U, ERGOcam One) is taken out.
To close CONEXIOcom: Replace the active device.
Display CONEXIOcom
No Icon Setting
1 - Info line
This line displays the active patient name (if selected in CONEXIO) under which the acquired data are stored. If no patient is selected, images and videos are stored temporarily in the clipboard under "unas­signed patient".
2 - If a device is active, the device type is shown. The
following is implemented at this time: DIAGNOcam U ERGOcam One
75 / 126
Page 76
Instructions for use Primus 1058 Life
4 Operation | 4.8 Using function through the dentist or assistant unit
76 / 126
No Icon Setting
3 Next image/video
To be able to communicate efficiently with the pa­tient, individual images can be selected and dis­played directly. This uses a rolling system that ad­vances from left to right and from top to bottom.
4 Screen display:
This button changes the display on the monitor. The following settings can be made: 1/2/4/6 – number of images displayed. The live image is always shown as the last image in split view.
Note
CONEXIO can show only those patients who have been entered in the CON­EXIO database.
Patient files can be transferred to the treatment centre by one of 2 ways:
▪ Transfer patient via VDDS media or Slida interface to CONEXIO in the re-
spective treatment room. The patient is now automatically activated on the treatment centre. This might be an option of your patient management sys­tem that is subject to a charge; please contact your IT specialist.
▪ Opening the patient file of the patient on the PC workstation in the treat-
ment room. The patient is now automatically activated on the treatment centre.
4.8 Using function through the dentist or assistant unit
4.8.1 Using the hygiene functions
The following buttons are available for the hygiene functions:
Key Name Feature Control ele-
ment
"Tumbler filler" key
The tumbler is being filled. Filling time can be set
Dentist element and assistant element/
"Bowl rinsing" key
The bowl is be­ing rinsed. Rins­ing time can be set. Leaving the rins­ing position (SP), the bowl is rinsed for the full rinsing time (function can be activated by ser­vice technician).
Dentist element and assistant element/
Page 77
Instructions for use Primus 1058 Life
4 Operation | 4.8 Using function through the dentist or assistant unit
Key Name Feature Control ele-
ment
"Intensive germ reduction" key
Intensive germ reduction/rinsing function
See also:
2 Servicing in-
structions
Assistant element/
"HYDROclean" key
HYDROclean function
See also:
2 Servicing in-
structions
Assistant element/
The following applies to the hygiene functions, "Tumbler filling" and "Bowl rins­ing":
Press key to activate the function.Press key again to discontinue the function.
Note
The preparation methods can be found in the care instructions.
The following settings can be changed:
▪ Tumbler filling time ▪ Bowl rinsing time
Using the tumbler filling
Press the "Tumbler" button briefly to start filling the tumbler.
ð Tumbler filling is started and then stopped after the saved period of time. ð Default = 5 s. ð An on/off operation is not supported.
Press the "Tumbler filling" key for more than 4 seconds to start the pro-
gramming mode. Set the period of time in 200 ms increments. Minimum: 0.4 s.
ð If the key stays depressed, the elapsed time is counted in 200 ms steps and
an acoustical signal is issued each second.
ð Once the key is released, the current value is saved.
Using the bowl flush
Press the "Bowl flush" key briefly to start the bowl flush.
ð The bowl flush is started and then stopped after the saved period of time. ð Default = 7 s. An on/off operation is not supported.
Press the "Bowl flush" key for more than 4 seconds to start the program-
ming mode. Set the period of time in 200 ms increments. Minimum: 0.4 s.
ð If the key stays depressed, the elapsed time is counted in 200 ms steps and
an acoustical signal is issued each second.
77 / 126
Page 78
Instructions for use Primus 1058 Life
4 Operation | 4.8 Using function through the dentist or assistant unit
78 / 126
ð Once the key is released, the current value is saved.
4.8.2 Using lamp and X-ray image viewer
The following buttons are available for the light functions:
Key Feature Control element
Turning the operating light On / Off.
Assistant element
Turning the X-ray image viewer (accessory equip­ment) On / Off.
Dentist element
Note
Activate the KaVoLUX 540 LED operating light using the "Operating light" key on the assistant element. Only then the operating light can be operated by means of the sensor and the control panel of the operating light.
4.8.3 Using the bell (optional)
Press the "S3" key ("Bell" function key) to activate the bell relay.
ð The bell relay is activated for as long as the key is being pressed.
"Bell" function key without MEMOspeed /with MEMOspeed
4.8.4 Using the timer
Press the "Timer" key briefly to start or stop the timer.
ð LED flashes while the timer counts down.
The elapsed time is shown in the display on the top right. A beep is issued after the timer time is elapsed.
Timer elapsing without MEMOspeed / with MEMOspeed
Page 79
Instructions for use Primus 1058 Life
4 Operation | 4.8 Using function through the dentist or assistant unit
Setting the timer
Requirement
Standby menu is selected.
To set a timer time (e.g. Timer 1), press the "Timer" key until you hear a
beep.
ð The display switches to the settings menu for the timer time.
Key Settings
S1 Saves the parameters.
Quits the programming mode. S2 Decreases the value. S3 Increases the value. S4 Switches counter/timer function. (Direction of counting)
4.8.5 Saving handpiece settings (without MEMOspeed)
The following settings can be individually:saved for the instruments:
Handpiece Setting
Turbine Speed
Preselected spray
Motor INTRA LUX KL 701/703, COMFORTdrive
Speed Direction of motor rotation Preselected spray
Ultrasonic scaler Spray on/off*
Intensity
Multi-functional piece Heating on/off
* only with a corresponding setting in service mode
Setting the turbine
Setting the speed
Remove the turbine from the holder.
T
M
To reduce or increase the speed, move the foot pedal to the left or right.
79 / 126
Page 80
Instructions for use Primus 1058 Life
4 Operation | 4.8 Using function through the dentist or assistant unit
80 / 126
Note
The speed is not shown in the display cannot be saved. The minimum and maximum speed depends upon the type of turbine that is
used.
Setting the cooling status
Remove the turbine from the holder.
Press the "Preselected spray" button.
or
T
M
Press the "Preselected spray" footswitch.
Key Feature
No LED is on: No cooling
One LED is on: Spray air cooling status
Both LEDs are on: Spray cooling status
Save the cooling status
Press the "Save" (S1) key until you hear a beep.
Setting the motor
Note
The speed is not shown in the display cannot be saved. The minimum and maximum speed depends on the motor that is used and
the attached straight or contra-angle handpiece.
The speed, spray preselection are set and the values are saved in the same manner as with the turbine.
See also:
2 4.8.5.1 Setting the turbine, Page 79
Setting the direction of motor rotation
Note
The direction of motor rotation can only be changed when the motor is at rest.
Take motor off the holder.
Press the "Direction of motor rotation" key.
or
Page 81
Instructions for use Primus 1058 Life
4 Operation | 4.8 Using function through the dentist or assistant unit
T
M
Press the "Direction of motor rotation" foot pedal.
ð The direction of motor rotation is reversed each time the cross-switch and/
or the "Direction of motor rotation" key is actuated: counterclockwise rota­tion - clockwise rotation.
ð The LED is on when CCW motor rotation is set.
Save the direction of motor rotation
Press the "Save" (S1) key until you hear a beep.
Setting the ultrasonic scalers, PiezoLED and PIEZOsoft
Set the intensity the same way you set the speed of the turbine.
See also:
2 4.8.5.1 Setting the turbine, Page 79
Selecting the operating mode (PiezoLED only)
Note
The selection of the mode depends on the treatment method and the tip used. For information about the selection of an operating mode, please refer to the "Operating modes P1 / P2 / P3 and E" section of the "PiezoLED In­structions for Use".
Take the PiezoLED off the holder.Press the "Mode" key to select the operating mode.
Modes P1 / P2 / P3 / E are available for selection.
Setting the multifunctional handpiece
Take the multifunctional handpiece off the holder.
Adjust the heating using the "Air/water heating" button.
Symbol Function
Air/water heating "on"
Air/water heating "off"
Save the heating status
Press the "Save" (S1) key until you hear a beep.
81 / 126
Page 82
Instructions for use Primus 1058 Life
4 Operation | 4.9 Operating the foot switch
82 / 126
4.9 Operating the foot switch
4.9.1 General functions
The footswitches of the foot control have two functions. The function of the con­trol depends on whether an instrument is in its holder or whether it has been removed.
See also:
2 Foot control
4.9.2 Positioning the patient chair with the foot control
See also:
2 Automatic positioning of patient chair 2 Position the dental chair using the button cross or 4-way switch
4.9.3 Preselect dentist
Requirement
All instruments are in their holder.
Hold down the foot pedal and press the stirrup switch.
ð Each time the stirrup switch is pressed, the selection advances to the next
dentist (dentist 1 to 2).
4.9.4 Start and regulate instruments
Remove the handpiece (such as turbine, motor) from the holder.
ð The handpiece is active.
Press the foot pedal.
ð The removed handpiece runs at the set speed or intensity.
Changing speed or intensity with the foot pedal.
ð The left stop corresponds to the minimum speed/intensity. ð The right stop corresponds to the maximum speed/intensity.
Page 83
Instructions for use Primus 1058 Life
4 Operation | 4.9 Operating the foot switch
4.9.5 Setting the cooling condition
Remove the handpiece (e.g turbine, motor) from the holder.
ð The handpiece is active.
Press the "Preselected spray" footswitch.
ð The cooling status is switched each time the foot switch is pressed: spray air
- spray.
ð The cooling status is displayed on the dentist element.
4.9.6 Activate blown air
Remove the handpiece (such as turbine, motor) from the holder.
ð The handpiece is active.
Press the "Blown air" foot-operated button.
ð As long as the foot-operated button is pressed, blown air exits from the re-
moved handpiece (does not apply to PiezoLED).
4.9.7 Preselect counterclockwise motor rotation
Take motor off the holder.
ð The handpiece is active.
Slide the cross switch upward.
ð The direction of motor rotation is reversed each time the cross-switch is ac-
tuated: counterclockwise rotation - clockwise rotation.
ð The direction of motor rotation is displayed on the dentist element.
4.9.8 Adjusting the instrument light
Slide the cross switch to the right. (spotlight function)
ð Cold light "On".
83 / 126
Page 84
Instructions for use Primus 1058 Life
4 Operation | 4.10 Service table 1568 (optional accessory)
84 / 126
4.9.9 Using CONEXIOcom (fee-based additional option)
No Setting
U-shaped switch
Discard image/video
Press briefly - deletes the selected image/video
Press long - all images/videos in the clipboard are deleted Previous image/video
Select previous image/video Next image/video
Select next image/video Screen display
The number of displayed images (Split View) is reduced:
The live image is always shown as the last image in split view. Screen display
The number of displayed images (Split View) is increased:
The live image is always shown as the last image in split view Capture Mode
Toggles between the recording modes, video recording and image re-
cording Screen display
Toggles between full screen and normal view Save image/video
Press briefly - freezes the live image
Press long - saves the live image directly.
If no patient is selected, the images are stored directly under "unas-
signed patient".
Note
If no patient is selected, the images stay in the "Swap Tray" and are not saved permanently. As soon as a patient is selected, these temporary data in the "Swap Drying" are deleted. When an active patient is logged off (or a new patient is logged on) in CONEXIO, a query is shown asking whether the images shall be deleted or saved. Data deleted at this point cannot be re­stored subsequently.
4.10 Service table 1568 (optional accessory)
CAUTION
Exceeding the load limits.
Damage to the service table.
Comply with maximal load limits.
Page 85
Instructions for use Primus 1058 Life
4 Operation | 4.10 Service table 1568 (optional accessory)
CAUTION
Over-travel beyond swivel range.
Property damage.
Comply with rotary knob swivel range of 180°.
The service table 1568 can be locked in 4 snap-in positions using the rotary button . This locks only the downward movement to allow for higher loads. In the absence of the locking mechanism, the maximum load of the service table is 2 kg. The presence of the locking mechanism increases the maximum load to 5 kg.
1
2
Rotate the rotary knob ① in counterclockwise direction to lock the service
table in place.
Rotate the rotary knob ① in clockwise direction to unlock the service table.Use the rotary button ② to adjust the break for vertical motion.
4.10.1 Moving the service table
CAUTION
Collision with persons or pieces of equipment.
Collisions could be caused by the required degrees of freedom and the large swivel range.
Always move or swivel the service table with great care.
5
1
2
°
2
4
°
3
6
°
4
0
°
Swing ranges
85 / 126
Page 86
Instructions for use Primus 1058 Life
4 Operation | 4.10 Service table 1568 (optional accessory)
86 / 126
Dimensions and swing ranges (in mm)
Page 87
Instructions for use Primus 1058 Life
5 Preparation methods DIN EN ISO 17664
5 Preparation methods DIN EN ISO 17664
Note
The preparation methods can be found in the care instructions.
87 / 126
Page 88
Instructions for use Primus 1058 Life
6 Accessories and kits | 6.1 Device
88 / 126
6 Accessories and kits
6.1 Device
Name Description
Water block DVGW with integrated water germ re­duction system
With a DVGW permit and electronic monitoring of the filling level of the disinfection container.
Water block compact Without DVGW permit.
With water filter and shutoff valve.
Water bottle DVGW with water block, compact With DVGW permit.
Features a tumbler and handpiece water supply that is independent of external water supply, in­cludes Oxygenal dosing attachment for manual dosing of the germ reduction liquid into the water
bottle. Steel mounting plate For installation on the left or right. Connector for third-party equipment To connect or supply third-party devices such as
an airflow through the quick couplings. Amalgam separator DÜRR CAS 1 Approved amalgam separation systems with a
separation > 95 %. Separation DÜRR CS1 Separation using a solids collector. Solids collector kit Wastewater solids collector for wet suctioning. External suction Wastewater and wet suction air are drawn from a
central location. Water jet pump For saliva ejector. Operating light EDI / KaVoLUX 540 LED T / MAIA
LED
Operating light.
Tray support For the small handpiece tray. Warm water heater Heats the tumbler water. Low-pressure regulator Regulator for suction air when the suction vacuum
is too high. Selective support kit Turns on the saliva ejector and/or spray mist suc-
tion. Intensive germ reduction Only in combination with DVGW water block kit. Monitor support arm The monitor support arm is either affixed to the
lamp mount pole or a Centro 1540. Monitor KaVo Screen One and KaVo Screen HD Service table 1568 It can be mounted on a device stand (cart ver-
sion).
Service table accessories:
▪ X-ray viewer 1440 ▪ Instrument tray ▪ Cup holder
Radiograph viewer Röbi 1440 For installation on the light mounting pole.
6.2 Dental chair
Name Description
Armrest The patient chair can be fitted with one or two arm
rests.
Page 89
Instructions for use Primus 1058 Life
6 Accessories and kits | 6.3 Assistant unit
Name Description
2-hinge head rest with rotary knob Controlled by rotary knob 2-hinge headrest with push button Controlled by push button Backrest Progress Correct working posture and optimum access for
the dentist, especially during pediatric procedures
6.3 Assistant unit
Name Description
Satelec Mini LED LED curing light. Triple function handpiece Multifunctional handpiece featuring air, water, no
heating, and no cold light. Multifunctional handpiece Multifunctional handpiece featuring air, water,
heating, and cold light. Saliva ejector, water-operated With water jet pump. Second saliva ejector The second saliva ejector kit is mounted on the
sieve housing that is included in the basic configur-
ation.
6.4 Dentist unit
Mat. no. Description
Multiflex LUX hose For connection of turbine and SONICflex and all
handpieces fitting on the multiflex coupling. Motor hose, COMFORTbase 404L
COMFORTbase 404S
For connection of INTRA LUX motor KL 701,
motor KL 703 LED, COMFORTdrive 200XD. Assembly kit INTRA LUX motor KL 703 LED Brushless motor with light. Assembly kit INTRA LUX motor KL 701 Brushless motor with light. KaVo COMFORTdrive 200 XD Dental handpiece for the high speed range up to
200,000 rpm. It can be attached only to the KaVo
COMFORTbase coupling. Triple function handpiece Multifunctional handpiece featuring air, water, no
heating, and no cold light.
Also available as an "upright" version. Multifunctional handpiece Multifunctional handpiece featuring air, water,
heating, and cold light.
Also available as an "upright" version. PiezoLED Handpiece for the removal of dental calculus with
the tip sets, Scaler / Paro / Endo / Prep. PIEZOsoft Handpiece for the removal of dental calculus with
the Scaler tip sets. Spray heater for instruments without handpieces Heater for spray water heating. Tray holder for a standard tray / US tray / 2x-
standard tray
Standard tray, US tray, and/or 2x-standard trays
(install on left or right side of dentist element). ERGOcam One Intraoral camera for documentation and patient
communication.
89 / 126
Page 90
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.1 Introduction
90 / 126
7 Safety check - Test instructions
7.1 Introduction
7.1.1 General instructions
Note
The safety check may only be carried out by one or more electricians (as defined in IEC 61140) who have been appropriately trained for the device to be inspected.
Note
The contents and specified tests described in this document are based on the international standard, IEC 62353. This standard applies to the testing and inspections of medical electrical devices or medical electrical systems comply­ing with IEC 60601-1 (DIN EN 60601-1).
Note
In order to evaluate the safety of medical devices, systems or components of medical devices or systems, the safety check must be carried out at the fol­lowing times:
Prior to startupduring servicingduring inspections and maintenancefollowing repairson the occasion of recurrent tests
Note
With regard to devices that have not been manufactured in accordance with IEC 60601-1 (DIN EN 60601-1), these requirements can be applied taking the mandatory safety standards for the production of these devices into con­sideration.
Note
If several medical electrical devices (ME device) or electrical devices from several manufacturers combined into a system are connected to the KaVo dental unit, the manufacturer data contained in the instructions for use for all products subject to the safety checks must also be noted.
Note
Accessories of ME devices that might impact the safety of the device to be tested or the measured results must be included in the safety checks.
Note
All tests on accessories included in the safety checks must be documented.
Note
Furthermore, the manufacturer data contained in the instructions for use must be adhered to in all products to be tested and inspected.
Page 91
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.1 Introduction
Note
KaVo offers a medical device book for keeping an inventory and recording es­sential master data on the medical device. The medical device book is only available in German (Mat. no. 0.789.0480).
Note
The following tests and measurements must be documented, for example in the medical device book. We recommend using the templates at the end of the document.
Note
The tests must be performed in the order specified by the manufacturer!
7.1.2 Notes for medical electrical systems
Note
An ME System is the combination of individual devices (as defined by manu­facturers) that must meet the following conditions:
At least one of these devices must be a medical electrical device.The devices must be functionally connected or at least they should be
connected by the application of a multiple socket outlet.
Note
With ME systems, the person responsible for putting the system together must employ the necessary measuring parameters and measuring proced­ures defined in IEC 60601-1 (DIN EN 60601-1).
Note
Each individual device in an ME system, which has a separate connection to the power supply mains, or which can be connected to or separated from the power supply mains without the aid of a tool, must be checked individually. Moreover, the ME system must be checked as one unit to avoid the situation, in which the „aging“ of individual devices lead to unacceptable values in sum.
Note
An ME system that is connected to the power supply mains by means of a multiple socket outlet must be treated as one device during checks and test­ing.
Note
If the ME system or part of the system is connected to the power supply mains by means of an isolating transformer, the transformer must be in­cluded in the measurements.
Note
In ME systems, in which more than one ME device are interconnected via data lines or otherwise, e. g. via electrically conductive attachments or coolant tubes, the earth wire resistance of every single device must be checked.
91 / 126
Page 92
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.1 Introduction
92 / 126
Note
If it should be impossible to check single ME devices that are functionally connected to an ME system individually for technical reasons, the ME system must be checked as a whole.
7.1.3 Components of the safety check
Visual inspection (inspection by examination)
Optical appraisal of the safe and usable condition of the medical device and its accessories.
Measurements
▪ Measurement of the protective earth resistance in accordance with IEC
62353
▪ Measurement of the equipment leakage current - Alternative measuring
method in accordance with IEC 62353
▪ Measurement of the equivalent patient leakage current - Alternative meas-
uring method in accordance with IEC 62353
Note
A measurement of the insulation resistance in accordance with IEC 62353 need not be carried out. This check is covered by the measurement of the leakage current provided a safety tester specified in IEC 62353 Annex C is used!
Functional test
Medical device function test as well as testing of all safety shutdowns with refer­ence to accompanying documentation/instructions for use.
7.1.4 Testing intervals
▪ Testing interval every 2 years according to device type IIa (without HF sur-
gery)
7.1.5 Notes on the test method in accordance with IEC 62353
▪ Protection class 1 ▪ Type BF ▪ The device is permanently installed / threshold: PE conductor test < 0.3 Ω ▪ Measurement of the equipment leakage current - alternative measure-
ment / threshold: < 10 mA*
▪ Measurement of the equivalent patient leakage current - Alternative meas-
urement / threshold: < 5 mA
*The limit of the equipment leakage current corresponds to the value defined in IEC 60601 (DIN EN 60601), taking comment 2 from table 2 into consideration.
Page 93
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
7.1.6 Notes on repeat testing
Note
The value determined in these tests must be documented and evaluated to­gether with the measuring processes. The measured values may not over­shoot the specified values.
Note
Comparisons with previous measurements must be carried out if the meas­ured values undershoot the threshold values by more than 10 %. The test in­tervals should be reduced if a deterioration in values is determined!
7.2 Instructions for the safety check
7.2.1 Preparatory measures to be undertaken on the device
WARNING
Electrical power.
Death or injury from electric shock.
Before servicing, pull the mains plug out of the socket or completely discon-
nect the device from the power to de-energise it!
After conversion, check the electrical safety according to DIN EN 62353
[IEC 62353].
Turn off the main switch before any maintenance work.Unscrew the fastening screw on the bracket of the main switch.
Take off the cover ② proceeding in upward direction.Release the rear cover ① below and remove it.Unscrew the fastening screws (see: arrows) of the cladding and take off the
covers.
93 / 126
Page 94
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
94 / 126
7.2.2 Visual inspection (inspection by examination)
Check the following items in advance:
▪ Has the equipment of the ME device or the ME system been changed since
the last inspection?
▪ Has the change been documented and approved (test protocol of safety
check STK)?
▪ Are there any indications of insufficient safety?
Check the ratings of fuses that are accessible from outside
Note
A check of whether or not the main fuse on the mains socket corresponds to the rated values is no longer included in DIN EN 62353:2015 [IEC 62353:2015].
Visual inspection and appraisal of the medical device and accessories
The following list is for exemplary purposes and makes no claim of being com­plete.
Check the following items:
▪ Stability of the device ▪ No damage to the cladding or casing (cracks, breakage) ▪ Functioning of the carrier systems on dentist and assistant side, treatment
lamp, and display (brakes, height adjustment, etc.) ▪ Condition of the handpiece and suction hoses ▪ Condition of all installed application parts ▪ Condition of the control panels ▪ Condition of the threads for the fitting of tips to the ultrasound scaler hand-
piece ▪ Condition of the operating light ▪ Absence of leaks on the body of the device ▪ Condition of the power connection provided by the treatment centre
Page 95
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
▪ Condition of air and water connections ▪ Any damage on the sight window and the casing of the camera ERGOcam ▪ Expiry date of the water bottle inserted in the BS water bottle not exceeded
Check of legibility and completeness of the safety­related labels
Check if all safety-related markings (plates and labels) are present and
legible.
Check if the rating plate and serial number plates are present and legible.
Attachment locations: nameplate, markings BF and note "Comply with the instructions for use"
Control of the availability of the necessary documents
Check if the required instructions for use and care instructions are available
in the surgery.
95 / 126
Page 96
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
96 / 126
Note
Any irregularities determined in the visual inspection must be recorded in the test protocol. It is essential to determine whether defects and deficiencies could have an adverse impact on the safe operation of the unit. If the de­termined irregularities present a safety hazard and cannot be rectified dir­ectly, the unit must be closed down until safe operation is restored.
7.2.3 Measurements
WARNING
Danger to persons due to a lack of care exercised during the safety checks and testing.
Prior to connecting the treatment centre to the safety tester, disconnect it
from the mains supply network.
Carry out all safety checks and tests in a manner that will ensure that there
will be no danger to the testing personnel, patients or other persons.
Note
The safety tester must comply with the requirements defined in DIN EN 62353 [IEC 62353], Annex C.
Note
If no other specifications have been made, all values relating to voltage and current are effective values of alternating voltage, direct voltage or pulsating voltage res. alternating current, direct current or pulsating current.
Note
Connection cables such as data cables and cables for the functional earth could simulate protective conductor connections. These types of supplement­ary but unintentional protective earth connections could lead to erroneous measurements.
Note
Cables and wires, e.g. power supply cords, measuring circuits and data lines, must be arranged appropriately such that their influence on measurements is minimised.
Note
The following measuring aids can be ordered: KaVo measuring cable (Mat. no. 0.411.8811)
Page 97
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
Using the measuring cable ① the unit is disconnected from the mains supply and connection of the treatment centre to the safety tester is enabled. Hence, the customer-provided mains supply L & N on the power input board need not be disconnected. The adapter cable ② is included in the delivery of the KaVo measuring cable and is required for older treatment centres that are not equipped with an X2 connector.
Connecting the safety tester with KaVo measuring cables to the treatment centre
Remove plug X2 from the power input board and plug it into the matching
connector X2 of the KaVo- measuring cable (Mat. no. 0.411.8811).
Plug the second plug X2 of the KaVo measuring cable into the network card
(X2).
Insert the protective contact plug of the KaVo measuring cable into the
safety tester.
97 / 126
Page 98
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
98 / 126
Connecting the safety tester without the KaVo measuring cable to the treatment centre
Switch L + N of the on-site power supply cord to be voltage-free.Disconnect L + N on terminals X1.1 and X1.2.Connect the safety tester directly to terminals X1.1 (L) and X1.2 (N) and
protective earth conductor terminal (PE).
Note
The main switch of the ME device / ME system must be turned on during the measurement.
Connect the application parts [AP] to the safety tester:
Connect ① to ④ to the safety tester.Connect the safety tester to additional measuring points AP X.
Page 99
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
Note
Additional measuring points AP X must be taken into consideration in the presence of accessories: e.g. accessories such as PIEZO ultrasonic scaler, etc.
See also:
2 8 Annex - Additional measuring points, Page 111
Connect accessible conductive parts [ACP] with PE
ACP = accessible conductive parts
ACP
ACP
Note
Additional measuring points ACP X must be taken into consideration in the presence of accessories.
See also:
2 8 Annex - Additional measuring points, Page 111
ACPs on the treatment centre
No ACPs need to be connected to the protective conductor (PE) during the measurement on the treatment centre Primus 1058 Life, as all relevant parts are connected to the PE at the factory and are included in the test.
ACPs on treatment lamps
No ACPs need to be connected to the operating lights during the measurement with protective conductor (PE) because all relevant parts have already been connected to the protective conductor (PE) at the factory and are included in the test.
Measure protective conductor resistance
Limit < 0,3 Ω (maximum value!)
Note
The integrity of the power supply cable, in particular the protective earth wire of the power cable must be ensured. As this is a fixed installation, the evalu­ation can be conducted by means of a visual inspection. If damage is determ­ined, the further procedure to be taken is specified in the general instruc­tions.
99 / 126
Page 100
Instructions for use Primus 1058 Life
7 Safety check - Test instructions | 7.2 Instructions for the safety check
100 / 126
Note
In this measurement the resistance of the protective earth connection of the supply network can be taken into consideration.
Note
If applicable: all removable supply connection lines, which are kept handy for possible use, should be taken into consideration and the respective PE meas­ured.
Protective earth measurement
The protective conductor resistance must be measured at the following parts of the device:
▪ Treatment centre ▪ Treatment lamp ▪ Accessories
Note
Additional measuring points SL X need to be taken into consideration in the presence of accessories: e.g. additional devices, such as third-party con­nector, USB port of the intraoral camera, etc.
See also:
2 8 Annex - Additional measuring points, Page 111
Loading...