Short instructions for use 1 Controls4
1 Controls
CAUTION
These brief instructions for use only contain the essential operating
functions.
It does not substitute for the Instructions for Use and Servicing that are part of the
scope of delivery.
▶ The Instructions for Use and Servicing include all information and safety notes
that must be observed to ensure intended use and safe operation of the product!
▶ The user must ensure that the unit works properly and is in satisfactory condition
before each use.
1.1 Dentist element TM/C table
Dentist element TM/C table
A
Group of keys for the dental chair
C
Group of keys for hygiene
E
Group of keys for the handpieces
B
Group of keys for menu selection
(MEMOspeed optional)
D
Group of keys for the timer
F
Group of keys for illumination
Page 5
Short instructions for use 1 Controls5
1.2 Dentist element S-table
Dentist element S-table
A
Group of keys for menu selection
(MEMOspeed optional)
C
Group of keys for the handpieces
E
Group of keys for illumination
B
Group of keys for the timer
D
Group of keys for the dental chair
F
Group of keys for hygiene
1.3 Assistant element
A
Group of keys for hygiene
C
Group of keys for the timer
Note
Activate the KaVoLUX 540 LED operating light using the "Operating light" key on
the assistant element. Only then the operating light can be operated by means of
the sensor and the control panel of the operating light.
B
Group of keys for illumination
D
Group of keys for the dental chair
Page 6
Short instructions for use 1 Controls6
1.4 Groups of keys
Group of keys for the dental chair
The keys of the assistant unit each have two functions and show two symbols.
Assistant element keyDentist element keyName
"Chair up" key
"AP 0" key
(automatic position 0)
"Chair down" key
"SP" key
(rinsing position)
"LP" key
(last position)
"AP" key
(activate automatic position)
"Backrest down" key
"AP 1" key
(automatic position 1)
"Backrest up" key
"AP 2" key
(automatic position 2)
"Collapsed position" key
Group of keys for illumination
KeyNameControl element
"Operating light" keyAssistant element
"X-ray image viewer" keyDentist element
Group of keys for hygiene
KeyNameControl element
"Tumbler filler" keyDentist element and
assistant element
Page 7
Short instructions for use 1 Controls7
KeyNameControl element
"Bowl rinsing" keyDentist element and
"Intensive germ reduction" keyAssistant element (optional)
"HYDROclean" keyAssistant element
assistant element
Group of keys for the handpieces/timer
KeyNameControl element
"Preselected spray" buttonDentist element
"Direction of motor rotation" but‐
ton
"Timer" keyDentist element and
Dentist element
assistant element
Group of keys for the menu
Standby menu with MEMOspeedStandby menu without MEMOspeed
① Menu function selection keys
② Display
Page 8
Short instructions for use 1 Controls8
1.5 Foot control
2
3
1
4
5
ItemNameFunction with handpiece
mounted
①"Spray pre-selection/AP"
footswitch
Moves the dental chair into
automatic position.
Function with handpiece re‐
moved
Sets the spray pre-selection.
②U-shaped switchTurns the safety shutoff On.Switches the footswitches to
the "Chair motion" function.
③"Blown air/AP" footswitchMoves the dental chair into
automatic position.
④Cross-switch: "Counterclock‐
wise motor rotation"
Changes the position of the
dental chair.
Sets the preset blown air
(chip blower).
Selects the direction of motor
rotation (for INTRA LUX mo‐
tor KL 701/703 or COM‐
FORTdrive 200XD).
⑤"Handpieces" foot-pedalGenerates a video/freeze
frame if CONEXIOcom is in‐
stalled.
Starts the motor and controls
the speed/ intensity of the
handpieces.
"Chair position" cross switch
①
③
Chair up
Chair down
②
Backrest up
④
Backrest down
Page 9
Short instructions for use 2 Moving the patient chair9
2 Moving the patient chair
2.1 Positioning the dental chair manually
Key
Dentist element
Key
Assistant element
Feature
The chair moves up.
The chair moves down.
The backrest moves upward.
The backrest moves downward.
2.2 Positioning the patient chair with the foot control
Recalling automatic position through the dentist element
▶ Briefly press the desired button.
ð Chair automatically moves to the saved chair position.
ð When the saved chair position is reached, the LED on the button lights up.
Saving the automatic position through the dentist element
▶ Move the chair to the desired position.
▶ To save the chair position, press the "AP 0", "AP 1", "AP 2", "SP" or "Collapsed
position" button until you hear a signal.
ð LED of the pressed button lights up. The chair position is saved.
Page 10
Short instructions for use 2 Moving the patient chair10
Recalling an automatic position through the assistant element
▶ Briefly press the "AP" key.
ð The LEDs of the "AP 0", " AP 1", " AP 2", "SP", and "LP" keys flash for approxi‐
mately four seconds.
▶ During these four seconds, briefly press the "AP 0", " AP 1", " AP 2", "SP" or "LP"
key.
ð The chair moves into the selected automatic position.
Saving an automatic position through the assistant element
▶ Move the chair to the desired position.
▶ Briefly press the "AP" key.
ð The LEDs of the "AP 0", " AP 1", " AP 2", "SP", and "LP" keys flash for approxi‐
mately four seconds.
▶ During these four seconds, press the "AP 0", "AP 1", "AP 2" or "SP" button, until a
beep is issued.
ð LED of the pressed button lights up. The chair position is saved.
Recalling an automatic position through the foot control
▶ Press the "SP" foot-operated button.
or
▶ Press the "LP" foot-operated button.
ð The chair moves into the selected automatic position.
Saving an automatic position with the foot control
▶ Hold down the foot pedal and foot-operated button "SP", and simultaneously press
any button for an automatic position ("AP 0", "AP 1", "AP 2" or "SP") on the dentist
or assistant unit until you hear a beep.
ð The automatic position is saved to the foot-operated button.
or
▶ Hold down the foot pedal and foot-operated button "LP", and simultaneously press
any button for an automatic position ("AP 0", "AP 1", "AP 2" or "SP") on the dentist
or assistant unit until you hear a beep.
ð The automatic position is saved to the foot-operated button.
Page 11
Short instructions for use 3 User menu11
3 User menu
The following options can be opened in the user menu:
OptionFeatureDescription
1FirmwareDisplay current firmware version.
2Time of daySet time of day.
3DateSet date.
4Time display mode
5Language
6LCDSet contrast of LCD display.
7LicensesDisplay of activated licenses
Setting the time of day display mode:
• Time of day only
▪ Time of day without seconds
Set menu language:
• Deutsch
• English
• Italiano
▪ Français
3.1 Starting and exiting from the user menu
▶ Press the "Next" key (S4) to switch to the user menu.
ð Displays option 1 "Firmware".
Page 12
Short instructions for use 3 User menu12
▶ Press the "Save" key (S1) to exit from the user menu.
ð The changed settings are saved.
3.2 Selecting and setting an option
▶ Press the "Decrease value" or the "Increase value" keys to select the desired pa‐
rameter or value.
See also:
2 Instructions for use Primus 1058 Life
Page 13
Short instructions for use 4 Switching between levels and selecting a dentist13
4 Switching between levels and selecting a dentist
Standby menu
The unit starts in the standby menu.
Available with MEMOspeed licence only:
The device automatically switches to its standby menu when you close the handpiece menu and the
patient communication menu.
Standby menu with MEMOspeedStandby menu without MEMOspeed
① Menu function selection keys
② Display
Switching between levels
Available with MEMOspeed licence only:
Level switching is deactivated in the basic state.
The level switching symbol displays the current dentist.
Note
The device acts like on level E when level switching is deactivated.
▶ To switch between levels, keep the "Direction of motor rotation" and the "Bowl
flushing" keys and the foot pedal depressed, until a beep can be heard.
After activating level switching, the level switching symbol shows the level (E, 1 2 or 3 - level E is se‐
lected in the example shown). The pre-selected dentist is only displayed very small in the level switch‐
ing symbol.
Page 14
Short instructions for use 4 Switching between levels and selecting a dentist14
Note
The device automatically saves the activation of level switching for the current den‐
tist.
Note
Level switching is deactivated using the same key combination as activation.
▶ Briefly press "Preselect level" key to select the level.
Selection of dentist
▶ Level switching blocked: press "Dentist" key to select dentist 1 or dentist 2.
or
▶ Level switching permitted: press "Preselect level" key long to select dentist 1 or
dentist 2.
Page 15
Short instructions for use 5 Setting the timer15
5 Setting the timer
5.1 Select the timer time
▶ Press the "Timer" key briefly to start or stop the timer.
ð LED flashes while the timer counts down.
The elapsed time is shown in the display on the top right.
A beep is issued after the timer time is elapsed.
5.2 Set the timer time
A maximum 59:59 minutes can be set using the timer.
Requirement
Standby menu is selected.
▶ To set a timer time (e.g. Timer 1), press the "Timer" key until you hear a beep.
ð The display switches to the settings menu for the timer time.
▶ Press the "Decrease value" key to reduce the time.
or
▶ Press the "Increase value" key to increase the time.
ð The set time is shown on the display.
SymbolFunction
The timer counts down (for example: 0:30 to 0)
The timer counts up (for example: 0 to 0:30)
▶ Press the "Save" button to save the value.
ð A beep confirms that the value has been saved.
or
▶ Press the "Back" button to return to the default (without saving).
Page 16
Short instructions for use 6 Setting the handpieces16
6 Setting the handpieces
6.1 Adjusting functions
Switch to the settings menu of the active element
Level switching must be activated for the selection of a level.
See also:
2 4 Switching between levels and selecting a dentist, Page 13
▶ Remove the handpiece from the holder.
▶ Briefly press "Preselect level" key to select the level.
▶ Press "Preselect level" button for 4 seconds in order to change settings.
ð The display switches to the Settings menu of the selected handpiece. (Available
only with MEMOspeed licence with activated level switching in level 1-3)
Saving the values
▶ Press the "Save" key to save the values. You can save after setting each value, or
after setting all values.
ð Saving is acknowledged by a beep.
ð This closes the "Settings" menu.
Set the speed/power intensity (in levels 1 to 3 only)
▶ Press the "Decrease value" key to decrease the speed.
or
▶ Press the "Increase value" key to increase the speed.
ð The speed is shown in the display.
Set the power intensity (in levels 1 to 3 only)
▶ Press the "Decrease value" key to decrease the speed/power intensity.
or
▶ Press the "Increase value" key to increase the speed/power intensity.
ð The speed/power intensity is shown on the display.
Setting the minimum and maximum speed
▶ Press the "min/max" (S4) keyed to toggle between the settings menus for minimal
and maximal speed.
Page 17
Short instructions for use 6 Setting the handpieces17
Setting the cooling level
Requirement
Settings menu for turbine is selected.
▶ Press the "Preselected spray" button.
KeyFeature
No LED is on: No cooling
One LED is on: Spray air cooling status
Both LEDs are on: Spray cooling status
Setting the direction of motor rotation
Note
The direction of motor rotation can only be changed when the motor is at rest.
Requirement
Settings menu for motor is selected.
▶ Press the "Direction of motor rotation" key.
IconFeature
Clockwise rotation
CCW rotation
The LED is on when CCW motor rotation is set.
Setting the water/air heating (only with the multifunctional handpiece)
▶ Adjust the heating with the key for "Air/water heating".
SymbolFunction
Air/water heating "on"
Air/water heating "off"
Page 18
Short instructions for use 6 Setting the handpieces18
Define operating mode (PiezoLED only)
Note
The selection of the mode depends on the treatment method and the tip used. For
information about the selection of an operating mode, please refer to the "Operat‐
ing modes P1 / P2 / P3 and E" section of the "PiezoLED Instructions for Use".
▶ Press the "Mode" (S4) key for 4 seconds to select the operating mode.
ð Modes P1 / P2 / P3 / E are available for selection.
SymbolFeature
Mode P1
Mode E
6.2 Turbine
The following settings can be changed:
• Speed
▪ Cooling status
Settings menu for minimal/maximal speed
6.3 INTRA LUX Motor KL 703 LED and COMFORTdrive
The following settings can be changed:
• Direction of motor rotation
• Speed
▪ Cooling status
Settings menu for minimal/maximal speed
6.4 PiezoLED
Note
Please comply with the enclosed "PiezoLED" Instructions for Use.
Page 19
Short instructions for use 6 Setting the handpieces19
The following settings can be changed:
• Output intensity
• Operating mode (P1 / P2 / P3 / E)
▪ Cooling status (no cooling / spray water cooling)
Settings menu for minimal/maximal intensity
6.5 Multifunctional handpiece
The following settings can be changed:
▪ Air/water heating
Settings menu for multifunctional handpiece
Page 20
Short instructions for use 7 Using handpieces20
7 Using handpieces
▶ Take the handpiece (e.g., turbine, motor, PiezoLED, etc.) from the holder.
ð The handpiece is active.
▶ Press the foot pedal.
ð The removed handpiece runs at the set speed or intensity.
▶ Changing speed or intensity with the foot pedal.
ð The left stop corresponds to the minimum speed/intensity.
ð The right stop corresponds to the maximum speed/intensity.
Preselect counterclockwise motor rotation
▶ Take motor off the holder.
ð The handpiece is active.
▶ Slide the cross switch upward.
ð The direction of motor rotation is reversed each time the cross-switch is actuated:
counterclockwise rotation - clockwise rotation.
ð The direction of motor rotation is displayed on the dentist element.
Setting the instrument light
▶ Slide the cross switch to the right. (spotlight function)
ð Cold light "On".
Page 21
Short instructions for use 8 Operation of the operating light KaVoLUX 540 LED (op‐
tional)21
8 Operation of the operating light KaVoLUX 540 LED (optional)
The KaVoLUX 540 LED operating light can be used in the following modes:
• Normal light: preset at 5,500 Kelvin and 30,000 Lux corresponding to daylight quality
• COMPOsave-Modus: enables longer processing periods for composites by filtering the blue com‐
ponents of the light
• Dimmed light: approx. 4,000 Kelvin; equivalent to the light of a halogen operating lamp
▪ Laser mode: Light mode, which has no negative influence on the KaVo KEY Laser III, the KEY
Laser 3+ or the KaVo DIAGNOdent.
Turning the operating light On and Off
Note
Activate the KaVoLUX 540 LED operating light using the "Operating light" key on
the assistant element. Only then the operating light can be operated by means of
the sensor and the control panel of the operating light.
Sensor / control panel KaVoLUX 540 LED T
①
"Operating light" key
③
"Mode/ colour temperature" button
▶ Press the "Treatment light" button.
ð
or
▶ Hold your hand just in front of the sensor.
ð Treatment light is switched on, Display LED shining.
▶ Press the "operating light" button once again.
or
▶ Hold your hand just in front of the sensor.
ð Treatment light is switched off, Display LED not shining.
②
"Operating light dimming" key
Setting the brightness
▶ Press and hold down the "Treatment light" key.
ð Signal buzzes, brightness changes from dark to bright in stages.
▶ When the desired brightness is reached, release the "Treatment light" key.
Page 22
Short instructions for use 8 Operation of the operating light KaVoLUX 540 LED (op‐
tional)22
Switch on COMPOsave mode
▶ Press the "Dim treatment light" key.
or
▶ Hold your hand in front of the sensor for 2 sec.
ð The COMPOsave mode switches on.
ð The COMPOsave mode can be recognised by the yellow light.
▶ Press the "Dim treatment light" key.
or
▶ Hold your hand in front of the sensor for 2 sec.
ð The treatment light switches back to normal light mode.
Setting the brightness of the dimmer (COMPOsave mode or normal light)
▶ Press and hold down the "operating light dimming" button.
ð Signal buzzes, brightness changes from dark to bright in stages.
▶ Once the desired brightness is reached, release the "operating light dimming" button.
Switching between COMPOsave mode and dimmed light and normal light
▶ Press the "Mode/colour temperature" button
ð The treatment light switches from COMPOsave mode to dimmed normal light.
▶ Press the "Mode/colour temperature" button once again.
ð The treatment light switches back to COMPOsave mode.
ð The COMPOsave mode can be recognised by the yellow light.
Setting the colour temperature
▶ Press and hold down the "Mode/colour temperature" button
ð Colour temperature changes in five steps.
▶ Release the button when the desired colour temperature is achieved.
Turning the laser mode On and Off
▶ Press the „operating light“ button and the "operating light dimming" button on the
operating light simultaneously.
ð The laser mode switches on.
ð Laser mode is activated: the operating light shines for 1 second in green and then
changes to a white light.
or
▶ Hold your hand in front of the sensor for 3 sec.
ð The laser mode switches on.
ð Laser mode is activated: the operating light switches initially in COMPOsave mode and then
shines for 1 second in green and then changes to a white light.
Page 23
Short instructions for use 8 Operation of the operating light KaVoLUX 540 LED (op‐
tional)23
Operation of the 3D joint
▶ Turn the switching ring to the left until it snaps into place.
ð Treatment light can be rotated 45° to the left or 45° to the right.
▶ Turn the switching ring to the right and it will spring back to its original position.
ð If the treatment light is turned to the centre position (zero position), it will automatically lock into
place in the centre position.
Page 24
Short instructions for use 9 Using the CONEXIOcom (optional)24
9 Using the CONEXIOcom (optional)
Note
For all CONEXIOcom functions, the dental unit must be connected to an installa‐
tion of the KaVo "CONEXIO" software.
The function of the CONEXIOcom menu is to control the display of previously recorded and saved im‐
ages and videos. In order to use the function, the unit must have access to the data of the KaVo Soft‐
ware "CONEXIO" software. For details on the configuration, please refer to the "CONEXIO" installation
instructions.
Display CONEXIOcom
NoIconSetting
1-Info line
2-If a device is active, the device type is shown. The following is im‐
3Next image/video
4Save image/video
5Discard image/video
6Screen display:
This line displays the active patient name (if selected in CONEXIO)
under which the data obtained are stored.
If no patient is selected, images and videos are stored in the clip‐
board under "unassigned patient".
plemented at this time:
DIAGNOcam U
ERGOcam One
To be able to communicate efficiently with the patient, individual im‐
ages can be selected and displayed directly. This uses a rolling sys‐
tem that advances from left to right and from top to bottom.
Press briefly - saves the selected image/video.
Press long - all images/videos are saved in the swap tray.
If no patient is selected, the images stay in the clipboard and cannot
be saved permanently. As soon as a patient is selected, these tem‐
porary data in the clipboard are deleted.
When an active patient is logged off (or a new patient is logged on)
in CONEXIO, a query is shown asking whether the images in the
shall be deleted or saved. Data deleted at this point cannot be re‐
stored subsequently.
Press briefly - deletes the selected image/video
Press long - all images/videos in the clipboard are deleted
This button changes the display on the monitor. The following set‐
tings can be made:
1/2/4/6 – number of images displayed.
The live image is always shown as the last image in split view.
Page 25
Short instructions for use 9 Using the CONEXIOcom (optional)25
NoIconSetting
▶ To activate a function, press the key below the info line briefly or long.
Starting and exiting the patient communication menu
▶ Make sure that the treatment unit has access to the KaVo CONEXIO software.
ð As soon as an imaging device is activated, the patient communication menu opens automati‐
cally.
▶ Select the right patient on the corresponding PC.
ð The display switches to the patient communication menu.
ð If no patient was selected, images from the clipboard are displayed.
ð When the patient is logged off the corresponding PC, the clipboard is deleted.
9.1 Using the patient communication menu with the foot control
The patient communication menu can also be controlled with the foot control.
Item
Name
①U-shaped switch
②Cross key "up"Split View:
③Cross key "down"
④Cross key "left"
⑤Cross key "right"Switches screen display between full-screen
⑥"SP" foot-operated buttonSelect next image/video
Feature
Delete image/video:
• Press briefly: Deletes the selected im‐
age/video
▪ Press long: All images/videos in the
clipboard are deleted
▪ Show several images
Split View:
▪ Show fewer images
Switching between the modes:
• Picture
▪ Video
and normal view
Page 26
Short instructions for use 9 Using the CONEXIOcom (optional)26
If no patient is selected, the images stay in the "Swap Tray" and are not saved per‐
manently. As soon as a patient is selected, these temporary data in the "Swap Dry‐
ing" are deleted. When an active patient is locked off for a new patient is logged on
in the "CONEXIO" software, a query is shown asking whether the images in the
"Swap Tray" shall be deleted or saved. If the data is deleted, it cannot be restored
subsequently.
Feature
Save image/video:
• Press briefly: Saves the image/video
▪ Press long: All images/videos are saved
in the clipboard.
Page 27
Short instructions for use 10 Hygiene27
10 Hygiene
10.1 Tumbler filling and bowl flushing
The following settings can be changed:
• Tumbler filling time
▪ Bowl rinsing time
▶ Press the "Tumbler filler" button briefly to start filling the tumbler.
or
▶ Press the "Bowl flush" key briefly to start the bowl flush.
▶ Press the "Set hygiene function" button for more than 4 seconds to start the programming mode.
Set the period of time in 200 ms increments. Minimum: 0.4 s.
An acoustic signal is heard every second.
▶ Release the button to save the selected value.
10.2 Rinsing program
▶ Briefly press the "Intensive germ reduction" button.
Preparation
▶ Position the germ reduction attachment on the porcelain spittoon bowl.
▶ Place the dental instruments on the germ reduction attachment.
The cannula of the triple-function and multifunctional handpieces must be removed for the rinsing pro‐
gram and the intensive germ reduction.
Page 28
Short instructions for use 10 Hygiene28
S-table
▶ Lock the swing-arm into the front catching position to ensure that the handpiece is removed.
▶ In the case of swing-arms without a catching function, insert the swing-arm lock (Mat. no.
1.008.1511) to ensure that the handpiece is removed.
▶ Hold the 3-way or multifunctional handpiece at the gripping sleeve and take off the cannula with a
slight twisting motion.
▶ Insert the triple-function or multifunctional handpieces without the cannula into the clamping de‐
vice of the germ reduction attachment. Make sure that the "W" button (water) of the triple-function
or multifunctional handpiece is activated, and that no air is exiting.
▶ Place the instrument hoses of the dentist and assistant elements in the corresponding places of
the germ reduction attachment.
▶ Support tubing with the enclosed holder.
Note
All water regulation mechanisms for the motor, instruments, and multiflex couplings
must be open.
Page 29
Short instructions for use 10 Hygiene29
Set rinsing time
▶ Press the "decrease value" or "increase value" keys to select rinsing program 01
"Programm vorbereiten: 01" or rinsing program 02 "Programm vorbereiten: 02".
ð The standard value for rinsing program 01 is 00:20 minutes per instrument.
ð The standard value for rinsing program 02 is 2:00 minutes per instrument.
M
T
▶ Press the foot pedal to display the time for rinsing the instrument hoses and the
three-function or multifunctional handpiece.
With the foot pedal of the foot control being pressed down, the rinsing time for the re‐
spective rinsing programme can be set in the range from 00:01 to 4:00 minutes using
the "Decrease value" and "Increase value" buttons.
The rinse program proceeds in four steps. For rinse program 01, the cycle starts in step 2, and for
rinse program 02, the cycle starts in step 4.
Start rinsing programme
▶ Briefly press the "Intensive germ reduction" button.
Discontinue the rinsing programme
The rinsing programme can be stopped at any time by pressing the "intensive germ
reduction" key.
If the device is turned off and then on again, the rinsing program also stops.
10.3 Intensive germ reduction
An additional intensive germ reduction must be carried out during extended down‐
times (> 4 weeks).
The need for intensive disinfection is automatically displayed in the top LED of the
"Intensive germ reduction" button at first start-up and after downtimes (weekend, holi‐
days, vacation, etc.).
If an audible signal is issued every 10 seconds and the bottom LED of the "Intensive
germ reduction" button flashes, the KaVo OXYGENAL 6 must be replenished. If a
beep is issued during the filling process (10 times at a one-second interval), the con‐
tainer is full.
Note
If the amalgam separator malfunctions during intensive germ reduction, the entire
process is stopped. The bottom LED of the "HYDROclean" button flashes.
▶ Press the "Intensive germ reduction" button for more than 4 seconds until you hear
an acoustic signal and the upper LED of the "intensive germ reduction" button
flashes. The display switches to the intensive germ reduction menu.
Page 30
Short instructions for use 10 Hygiene30
▶ Perform the preparations for intensive germ reduction.
See also:
2 10.2 Rinsing programme, Page 27
▶ Press the "Intensive germ reduction" button.
ð Intensive germ reduction starts and runs though several steps. The process can
be followed on the display.
Intensive germ reduction is completed when "01 intensive germ reduction completed"
appears on the display.
▶ Put all handpieces in holders and take off the germ reduction attachment.
Interrupt intensive germ reduction during the dwell time
So that no one will have to remain with the device throughout the entire intensive germ reduction proc‐
ess, intensive germ reduction can be interrupted in program step 4 (e.g. on the last workday before
the weekend). Program step 4 is the dwell time, in which the microorganisms are killed by an in‐
creased OXYGENAL 6 concentration.
Requirement
intensive germ reduction is in program step 4.
You hear a beep for 4 seconds.
▶ Turn the device off.
When the device is switched on after 30 minutes of dwell time, the remaining program steps run imme‐
diately and automatically.
If the device is turned on before the 30 minute period is elapsed, the dwell times finishes first before
the remaining steps of the intensive germ reduction program run.
Skip exposure phase
This function is required when the treatment unit needs to be used (for example in an emergency, at
hospitals, etc.) even though the germ reduction procedure has already started.
Intensive germ reduction is in step 4:
▶ Press and hold down the "Hydroclean" and "Intensive germ reduction" keys.
▶ Simultaneously press the foot pedal on the foot control.
M
T
ð The intensive germ reduction switches to step 3.7.
ð The automatic rinsing process starts up.
10.4 Reprocessing the suction system including suction hoses
Note
Clean the suction hoses after each treatment and disinfect them with DEKASEP‐
TOL Gel daily.
Page 31
Short instructions for use 10 Hygiene31
Application of DEKASEPTOL Gel
▶ Aspirate one tumbler of cold water with each suction hose.
▶ With the suction handpiece attached or by hand, press down the Dekaseptol dosing facility once
each for the saliva ejector hose and the spray mist ejector.
▶ Aspirate DEKASEPTOL Gel (Mat. no. 10007205) from the dosing facility.
▶ Aspirate another tumbler of cold water with each suction hose.
▶ Then aspirate another dose of Dekaseptol and allow it to act.
▶ Place the suction hoses in the holder.
10.5 HYDROclean function
The following processes are triggered by the HYDROclean function:
• The suction lines (from the filter housing to the separator system) are rinsed automatically for 150
seconds with pure water by activating the selective valves and the suction machine.
▪ The bowl rinse is turned on and off several times.
Requirement
Suction hoses remain hanging in the holder.
▶ Press the "HYDROclean" button.
ð The HYDROclean function is in the preparatory stage.
▶ Press the "HYDROclean" button.
ð This starts-up the HYDROclean function.
The HYDROclean function is completed when the top LED goes dark.
The process can be discontinued at any time by pressing the "HYDROclean" button.
Page 32
Short instructions for use 10 Hygiene32
10.6 Replenishing Oxygenal
When the Oxygenal container is empty, a beep is issued every 10 seconds.
▶ Remove the tumbler filler ① and tumbler holder ②.
▶ Top off the Oxygenal (max. 1 litre).
Note
Do not fill through the centre opening (tumbler drain).
▶ Quit replenishing immediately if you hear a signal sound being issued every second (10 sounds).
ð Reservoir is full.
▶ Attach the tumbler holder and the tumbler filler.
Page 33
Short instructions for use 11 Warning messages33
11 Warning messages
11.1 Safety shut-off
Safety shutoff for the standard patient chair
Item No. Safety switch-off actuatedLED on assistant elementLED on dentist element
①Patient unit swung over the
②Assistant element
③Backrest
patient chair
④Bracket on the foot control
⑤Kick plate
⑥Seat
▶ To deactivate an activated safety shutoff, remove the triggers from the to the range of movement
of the stool.
Page 34
Short instructions for use 11 Warning messages34
11.2 Additional warning messages
MalfunctionCauseRemedy
A beep is issued every
ten seconds and the "In‐
tensive germ reduction"
LED (green) flashes.
Ten beeps are issued.The Oxygenal container
A melody sounds.The amalgam separator
LED on "HYDROclean"
button (red) flashes.
The Oxygenal container
is empty.
is too full.
CAS1 is almost full
(95%).
The CAS1 amalgam
separator is defective.
Malfunction in the amal‐
gam separator.
Emergency shut off of
the bowl valve (only
when external suction is
installed)
▶ Refill the Oxygenal container.
▶ Stop filling the Oxygenal container.
▶ Exchange the amalgam container.
▶ Also refer to: Instructions for use for the CAS
1 or
▶ Call a Service technician.
▶ Call a service technician.
▶ Note the amalgam separator warning notice.
Also refer to: Operating instructions of the
amalgam separator
See also:
2 Operating instructions of the amalgam sepa‐
rator
▶ Call a technician.
Page 35
Short instructions for use 12 Service and error messages in the stand-by menu35
1
12 Service and error messages in the stand-by menu
If a status message is available, the standby menu shows an exclamation mark on selection key "S2"
①.
▶ Press the "S2" selection key ① to display status messages.
▶ Press the selection keys "+" ② and "-" ① to switch between multiple status messages.
▶ Press the "OK" selection key ③ to exit from the display of status messages.
Error messages in the status display
MalfunctionCauseRemedy
Display shows: ID 33No CAN node and/or in‐
ternal communication is
faulty.
Display shows: ID 64Water is shut off.
Water works leaks
strongly.
Water works malfunc‐
tion
Display shows: ID 65Safety switch of bowl
suction has been
reached.
Display shows: ID 66Amalgam separator
malfunction
Display shows: ID 67The oxygenal container
is empty.
Display shows: ID 68Call for service
▶ Call a service technician.
▶ Turn water on.
▶ Call a technician.
▶ Turn external suction on.
▶ Check and clean, if required, the bowl valve.
▶ Remedy the malfunction.
See also:
2 Instructions for Use of amalgam separator
▶ Refill the Oxygenal container.
See also:
2 Servicing instructions
▶ Have a service performed.
▶ Call a technician.
Page 36
Short instructions for use 12 Service and error messages in the stand-by menu36
MalfunctionCauseRemedy
Display shows: ID 69Intensive germ reduc‐
tion must be done.
Display shows: ID 70Dekaseptol empty.
Display shows: ID 72Dekaseptol bottle.
Display shows: ID XXThis error is not descri‐
bed in this chapter.
Display shows: CAN fail Internal communication
error.
Display shows: System
State
Unit does not work.
▶ Carry out an intensive germ reduction.
See also:
2 Servicing instructions
▶ Replenish Dekaseptol.
See also:
2 Servicing instructions
▶ Insert the DEKASEPTOL bottle.
See also:
2 Care instructions
▶ Call a technician.
▶ Turn the unit off and on again, consult a serv‐
ice technician according to need.
▶ Call a service technician.