KaVo PHYSIO Evo, PHYSIO One, PHYSIO Evo F Instructions For Use Manual

Instructions for use

PHYSIO Evo/Evo F/One

Always be on the safe side.
Distributed by: KaVo Dental GmbH Bismarckring 39 D-88400 Biberach Phone +49 (0) 7351 56-0 Fax +49 (0) 7351 56-1488
Manufacturer: Kaltenbach & Voigt GmbH Bismarckring 39 D-88400 Biberach www.kavo.com
Instructions for use PHYSIO Evo/Evo F/One

Table of contents

Table of contents
1 User instructions............................................................................................................................................... 5
1.1 User guide................................................................................................................................................ 5
1.2 Abbreviations............................................................................................................................................ 5
1.3 Symbols.................................................................................................................................................... 5
1.4 Target group ............................................................................................................................................ 5
1.5 Service...................................................................................................................................................... 5
1.6 Terms and conditions of warranty............................................................................................................. 6
1.7 Transportation and storage....................................................................................................................... 6
1.7.1 Currently valid packaging regulations......................................................................................... 6
1.7.2 Damage in transit........................................................................................................................ 6
1.7.3 Information on the packaging: Storage and transportation......................................................... 7
2 Safety............................................................................................................................................................... 9
2.1 Description of safety instructions.............................................................................................................. 9
2.1.1 Warning symbol........................................................................................................................... 9
2.1.2 Structure...................................................................................................................................... 9
2.1.3 Description of hazardous steps................................................................................................... 9
2.2 Safety instructions.................................................................................................................................... 9
3 Description of the Product.............................................................................................................................. 11
3.1 Purpose – Intended use ......................................................................................................................... 11
3.2 Operator stool PHYSIO Evo/Evo F/One................................................................................................. 12
3.3 Rating plates and instruction labels........................................................................................................ 13
3.4 Technical specifications PHYSIO Evo.................................................................................................... 14
3.5 Technical specifications PHYSIO Evo F................................................................................................. 14
3.6 Technical specifications PHYSIO One................................................................................................... 15
4 Installation...................................................................................................................................................... 16
4.1 Packaging............................................................................................................................................... 16
4.1.1 Transportation and storage....................................................................................................... 16
4.1.2 Packed items............................................................................................................................. 16
4.2 Assembly of operator stool PHYSIO Evo/Evo F/One............................................................................. 17
4.2.1 Mounting the castors................................................................................................................. 17
4.2.2 Mounting the upholstery............................................................................................................ 18
4.2.3 Mounting the backrest upholstery............................................................................................. 19
4.2.4 Adjust the gas pressure spring.................................................................................................. 20
4.3 Assembly of operator stool PHYSIO Evo F............................................................................................ 21
4.3.1 Mounting the top part of the chair.............................................................................................. 21
4.3.2 Mounting the upholstery............................................................................................................ 22
4.3.3 Mounting the backrest cushion.................................................................................................. 22
4.3.4 Adjust the gas pressure spring foot control............................................................................... 23
5 Dismantling..................................................................................................................................................... 24
5.1 Dismantling the operator stool PHYSIO Evo/Evo F/One........................................................................ 24
5.2 Removing the intervention protection cover of the gas pressure spring................................................. 24
5.3 Dismantling the gas pressure spring...................................................................................................... 25
5.3.1 Dismantling the gas pressure spring Evo / One........................................................................ 25
5.3.2 Dismantling the gas pressure spring Evo F............................................................................... 27
6 Operation........................................................................................................................................................ 30
3 / 38
Instructions for use PHYSIO Evo/Evo F/One
Table of contents
6.1 PHYSIO Evo........................................................................................................................................... 30
6.1.1 Adjust the seat height................................................................................................................ 30
6.1.2 Adjusting the seat...................................................................................................................... 30
6.1.3 Adjust the backrest angle.......................................................................................................... 31
6.2 PHYSIO Evo F........................................................................................................................................ 32
6.2.1 Adjust the seat height................................................................................................................ 32
6.2.2 Adjusting the seat...................................................................................................................... 32
6.2.3 Adjust the backrest angle.......................................................................................................... 33
6.3 PHYSIO One.......................................................................................................................................... 34
6.3.1 Adjust the seat height................................................................................................................ 34
6.3.2 Adjusting the seat and backrest................................................................................................ 34
7 Reprocessing methods according to DIN EN ISO 17664............................................................................... 35
7.1 General care instructions........................................................................................................................ 35
7.2 Cleaning and disinfection........................................................................................................................ 35
8 Troubleshooting.............................................................................................................................................. 36
4 / 38
Instructions for use PHYSIO Evo/Evo F/One
1 User instructions | 1.1 User guide

1 User instructions

1.1 User guide

Requirement
Read these instructions before the initial startup to prevent misuse and damage.

1.2 Abbreviations

Abbre‐ viation
IfU Instructions for use CI Care instructions AI Assembly instructions TI Technician's instructions SC Safety checks IEC International Electrotechnical Commission RI Repair instructions RK Retrofitting kit AS Assembly set EP Enclosed parts EMC Electromagnetic compatibility PI Processing instructions
Explanation

1.3 Symbols

See the Safety/Warning Symbols section
Important information for users and technicians
Action required

1.4 Target group

This document is for dentists, the dental office personnel, and the service technician.

1.5 Service

KaVo Customer Service: +49 (0) 7351 56-1000 Service.Einrichtungen@kavo.com Please refer to the serial number of the product in all inquiries! For further information, please visit: www.kavo.com
5 / 38
Instructions for use PHYSIO Evo/Evo F/One
1 User instructions | 1.6 Terms and conditions of warranty

1.6 Terms and conditions of warranty

KaVo provides the final customer with a warranty that the product cited in the hand‐ over certificate will function properly and guarantees zero defects in the material or processing for a period of 12 months from the date of purchase, subject to the follow‐ ing conditions: Upon justified complaints of flaws or a short delivery, KaVo will make good its warran‐ ty by replacing the product free of cost or repairing it according to the customer's wish‐ es. Other claims of any nature whatsoever, in particular with respect to compensation, are excluded. In the event of default and gross negligence or intent, this shall only ap‐ ply in the absence of mandatory legal regulations to the contrary. KaVo cannot be held liable for defects and their consequences due to natural wear, improper cleaning or servicing, non-compliance with operating, servicing or connection instructions, calcification or corrosion, contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with factory specifications. The warranty does not usually cover bulbs, glassware, rubber parts and the colour‐ fastness of plastics. Defects or their consequences that can be attributed to interventions on or changes made to the product by the customer or a third party are excluded from the warranty. Claims from this warranty can only be asserted when the transfer form (copy) belong‐ ing to the product has been sent to KaVo, and the original can be presented by the operator or user.

1.7 Transportation and storage

1.7.1 Currently valid packaging regulations

Note
Only valid for the Federal Republic of Germany.
Dispose of and recycle the sales packaging appropriately in accordance with current packaging regulations, employing waste management or recycling companies. Com‐ ply with the comprehensive return system. KaVo has had its sales packaging licensed for this purpose. Please comply with the regional public waste-disposal system.

1.7.2 Damage in transit

In Germany
If the packaging is visibly damaged on delivery, please proceed as follows:
1. The recipient of the package must record the loss or damage on the delivery re‐ ceipt. The recipient and the representative of the shipping company must sign this delivery receipt.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use the product.
4. Report the damage to the shipping company.
5. Report the damage to KaVo.
6. Consult with KaVo first, before returning a damaged product.
7. Send the signed delivery receipt to KaVo.
6 / 38
Instructions for use PHYSIO Evo/Evo F/One
1 User instructions | 1.7 Transportation and storage
If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows:
1. Report the damage to the shipping company immediately and no later than 7 days after delivery.
2. Report the damage to KaVo.
3. Leave the product and packaging in the condition in which you received it.
4. Do not use a damaged product.
Note
If the recipient fails to comply with any of the above-mentioned obligations, the damage will be considered to have arisen after delivery (in accordance with the General German Freight Forwarders´ Terms and Conditions, Art. 28) .
Outside Germany
Note
KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival.
If the packaging is visibly damaged on delivery, please proceed as follows:
1. The recipient of the package must record the loss or damage on the delivery re‐ ceipt. The recipient and the representative of the shipping company must sign this delivery receipt.
Without this evidence, the recipient will not be able to assert a claim for damages
against the shipping company.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use the product.
If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows:
1. Report any damage to the shipping company either immediately or no later than 7 days after delivery.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use a damaged product.
Note
If the recipient fails to comply with any of the above-mentioned obligations, the damage will be considered to have arisen after delivery (in accordance with CMR law, Chapter 5, Art. 30).

1.7.3 Information on the packaging: Storage and transportation

Note
Please keep the packaging in case you need to return the product for servicing or repair.
The symbols printed on the outside are for transportation and storage, and have the following meaning:
Transport upright with the arrows pointing upwards!
Fragile - protect against impact!
7 / 38
Instructions for use PHYSIO Evo/Evo F/One
1 User instructions | 1.7 Transportation and storage
Protect from moisture!
Permissible stacking load
Temperature range
Humidity
Air pressure
8 / 38
Instructions for use PHYSIO Evo/Evo F/One
2 Safety | 2.1 Description of safety instructions

2 Safety

2.1 Description of safety instructions

2.1.1 Warning symbol

Warning symbol

2.1.2 Structure

DANGER
The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance.
▶ The optional step includes necessary measures for hazard prevention.

2.1.3 Description of hazardous steps

Safety instructions distinguishing between three hazard levels are used in this docu‐ ment to prevent personal and property damage.
CAUTION
CAUTION indicates a hazardous situation that can cause damage to property or mild to moder‐
ate injuries.
WARNING
WARNING indicates a hazardous situation that can lead to serious or fatal injury.
DANGER
DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal
injury.

2.2 Safety instructions

CAUTION
Damage from unsuitable accessories. The use of accessories, other components, and cables other than those specified
(with the exception of components and cables sold by KaVo as spare parts for inter‐ nal components) can increase emissions or reduce the electromagnetic immunity of the product.
▶ Use accessories recommended by KaVo only!
9 / 38
Instructions for use PHYSIO Evo/Evo F/One
2 Safety | 2.2 Safety instructions
Damage due to working on the pressure vessel. Forceful opening of the pressure vessel or extensive heat exposure can lead to injury
and damage.
▶ Only instructed specialists may work on or exchange parts in the vicinity of the
seat height adjusting element.
Screws are pre-tensioned. Inappropriate loosening of screws on the support plate beneath the seat upholstery
can lead to injury and damage.
▶ Only instructed specialists may dismantle screws from the support plate.
CAUTION
CAUTION
10 / 38
Instructions for use PHYSIO Evo/Evo F/One
3 Description of the Product | 3.1 Purpose – Intended use

3 Description of the Product

3.1 Purpose – Intended use

The PHYSIO Evo/Evo F/One is an operator stool for dentists in a dental practice in accordance with DIN EN ISO 7493. All versions of the PHYSIO Evo/Evo F/One operator stool are tested and approved in accordance with the currently applicable guidelines. All versions of the PHYSIO Evo/Evo F/One operator stool are compliant with DIN EN 1335 - 2 and 3, DIN 4550 and DIN EN ISO 7493. Intended use also comprises the observation of all indications in the instructions for used as well as adherence to prescribed inspections and maintenance. Application of and compliance with the general guidelines and/or national laws, nation‐ al regulations, and the rules of technology applicable to this product for start-up and use of the KaVo product for the intended purpose are to be applied and complied with. The user must ensure that the unit works properly and is in satisfactory condition be‐ fore each use.
The applicable national legal regulations must be observed during the use of the de‐ vice, in particular the following:
▪ Current occupational safety regulations
▪ Current accident prevention regulations
Users have a duty to:
▪ Only use equipment that is operating correctly
▪ to protect himself, the patient and third parties from danger.
▪ to avoid contamination from the product
11 / 38
Instructions for use PHYSIO Evo/Evo F/One
3 Description of the Product | 3.2 Operator stool PHYSIO Evo/Evo F/One

3.2 Operator stool PHYSIO Evo/Evo F/One

KaVo PHYSIO Evo/Evo F/One The operator stool PHYSIO Evo/Evo F/One is equipped with soft castors Mat. no.
1.010.3896 for hard flooring upon delivery. The running surface and housing of the
castors are bi-chrome. Hard castors for soft flooring, e.g. fitted carpet are available as a kit Mat. no.
1.010.3827. The running surface and housing of the castors are monochrome.
Note
Use the castors that are appropriate for the flooring. Only use soft castors for hard flooring. Only use hard castors for soft flooring.
12 / 38