KaVo INTRAsurg 300, INTRAsurg 300 plus Instructions For Use Manual

Page 1
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
Always be on the safe side.
Page 2
Distributed by: KaVo Dental GmbH Bismarckring 39 D-88400 Biberach Tel. +49 7351 56-0 Fax +49 7351 56-1488
Manufacturer: Kaltenbach & Voigt GmbH Bismarckring 39 D-88400 Biberach www.kavo.com
Page 3
Contents
1 User instructions ........................................................................................................................................ 3
1.1 User guide ......................................................................................................................................... 3
1.1.1 Symbols ................................................................................................................................ 3
1.2 Target group ...................................................................................................................................... 3
1.3 Service .............................................................................................................................................. 3
1.4 Warranty terms and conditions .......................................................................................................... 3
1.5 Transportation and storage ............................................................................................................... 4
1.5.1 Currently valid packaging regulations .................................................................................. 4
1.5.2 Damage in transit ................................................................................................................. 4
1.5.3 Information on the packaging: Storage and transportation .................................................. 5
2 Safety ........................................................................................................................................................ 7
2.1 Description of safety instructions ....................................................................................................... 7
2.1.1 Warning symbol .................................................................................................................... 7
2.1.2 Structure ............................................................................................................................... 7
2.1.3 Description of danger levels ................................................................................................. 7
2.2 Purpose – Proper use ........................................................................................................................ 7
2.2.1 Information on electromagnetic compatibility ....................................................................... 9
2.2.2 Risks from electromagnetic fields (implanted life-sustaining system) ................................ 10
2.2.3 Disposal of electronic and electrical devices ...................................................................... 10
2.3 Safety instructions ........................................................................................................................... 11
3 Product description .................................................................................................................................. 13
3.1 INTRAsurg 300/300 plus ................................................................................................................. 13
3.2 Controls ........................................................................................................................................... 14
3.3 Multifunctional foot control ............................................................................................................... 15
3.4 Rating plate for 300/300 plus .......................................................................................................... 16
3.5 Technical data for the 300/300 plus ................................................................................................ 16
4 First use ................................................................................................................................................... 18
4.1 Electrical connection ....................................................................................................................... 18
4.2 Unpacking ....................................................................................................................................... 19
4.3 Adjust the mains input voltage ........................................................................................................ 19
4.4 Mount the bottle holder and use the correct coolant container ...................................................... 21
4.5 Connect the surgery motor and coolant reservoir ........................................................................... 22
5 Operation ................................................................................................................................................. 25
5.1 Operation in general ........................................................................................................................ 25
5.1.1 Start up ............................................................................................................................... 25
5.2 Surgical motors ............................................................................................................................... 25
5.2.1 INTRA S 550 surgical motor ............................................................................................... 25
5.2.2 INTRA LUX SL 550 surgical motor .................................................................................... 26
5.2.3 Remove the handpiece or angle piece ............................................................................... 27
5.2.4 Start the motor .................................................................................................................... 27
5.3 Free use .......................................................................................................................................... 28
5.3.1 Parameters menu ............................................................................................................... 28
5.3.2 Set the torque limitation ...................................................................................................... 30
5.3.3 Preset the speed range ...................................................................................................... 30
5.3.4 Set the coolant flow ............................................................................................................ 31
5.3.5 Selecting the handpiece and contra-angle handpiece ....................................................... 32
5.3.6 Switching the direction of motor rotation ........................................................................... 33
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
Contents
1/66
Page 4
5.3.7 Rinsing function .................................................................................................................. 33
5.3.8 Surgical motor in free use .................................................................................................. 34
5.3.9 Request maximum torque .................................................................................................. 34
5.4 Program ........................................................................................................................................... 35
5.4.1 Start the program ............................................................................................................... 35
5.4.2 Switch forward step ............................................................................................................ 36
5.4.3 Go backward a step ........................................................................................................... 36
5.4.4 Set parameters ................................................................................................................... 36
5.4.5 Limit the number of steps ................................................................................................... 37
5.4.6 Setting up delivery status ................................................................................................... 38
5.4.7 Surgical motor in the program ............................................................................................ 38
5.4.8 Request maximum torque .................................................................................................. 39
5.5 Functions exclusively with the 300 plus program version ............................................................... 40
5.5.1 Recognition function ........................................................................................................... 40
5.5.2 Light functions .................................................................................................................... 42
5.5.3 Request maximum torque (free use) .................................................................................. 43
5.5.4 Request maximum torque (program) ................................................................................. 43
5.5.5 Recommended programming when setting a number of implants one after the other ....... 43
5.5.6 INTRAsurg calibration ........................................................................................................ 44
6 Preparation methods DIN EN ISO 17664 ................................................................................................ 46
6.1 Cleaning .......................................................................................................................................... 47
6.1.1 Machine cleaning ............................................................................................................... 47
6.1.2 Manual cleaning ................................................................................................................. 47
6.1.3 Rinse the hoses .................................................................................................................. 47
6.2 Disinfection ...................................................................................................................................... 49
6.2.1 Automated disinfection ....................................................................................................... 49
6.2.2 Manual disinfection ............................................................................................................. 49
6.2.3 Drying ................................................................................................................................. 50
6.2.4 Service, inspection and testing after preparation ............................................................... 50
6.3 Sterilisation in a steriliser in compliance with DIN EN 13060 .......................................................... 50
6.3.1 Packaging ........................................................................................................................... 51
6.3.2 Sterilisation ......................................................................................................................... 52
6.3.3 Storage ............................................................................................................................... 52
7 Troubleshooting ....................................................................................................................................... 53
7.1 Exchange the pump hose ................................................................................................................ 56
7.2 Exchange coolant hose ................................................................................................................... 57
7.2.1 Exchange the coolant hose from the handpiece or angle piece. ........................................ 58
7.2.2 Exchange the coolant hose in the motor hose ................................................................... 58
7.2.3 Disposable coolant hose .................................................................................................... 59
7.3 Device settings table ....................................................................................................................... 61
8 Accessories ............................................................................................................................................. 62
9 Details on electromagnetic compatibility ................................................................................................. 63
9.1 Guidelines and manufacturer's declaration - electromagnetic transmission ................................... 63
9.2 Guidelines and manufacturer's declaration - electromagnetic resistance to jamming ..................... 63
9.3 Guidelines and manufacturer's declaration - electromagnetic resistance to jamming ..................... 64
9.4 Recommended safe distance between portable and mobile HF telecommunications equipment
and the INTRAsurg 300 ................................................................................................................... 65
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
Contents
2/66
Page 5
1 User instructions
1.1 User guide
1.1.1 Symbols
Refer to chapter Safety/Warning
Important information for users and technicians
Thermodisinfectable
Sterilisable up to 135°C
CE mark (Communauté Européenne). A product with this mark meets the requirements of the relevant EC directives, i.e. the applicable standards in Europe. ESD warning plate
1.2 Target group
This document is for dentists and office personnel.
1.3 Service
Service hotline: +49 (0) 7351 56-2500 Service.Einrichtungen@kavo.com Please indicate the product serial number in all requests. Additional information can be obtained at: www.kavo.comAdditional information can be obtained at: www.kavo.com
1.4 Warranty terms and conditions
KaVo provides the end customer with a warranty that the product cited in the hand‐ over certificate will function properly and guarantees zero defects in the material or processing for a period of 12 months from data of purchase, subject to the following conditions: Upon justified complaints of flaws or a short delivery, KaVo will make good its war‐ ranty by replacing the product free of cost or repairing it according to the customer's wishes. Other claims of any nature whatsoever, in particular with respect to com‐ pensation, are excluded. In the event of default and gross negligence or intent, this shall only apply in the absence of mandatory legal regulations to the contrary. KaVo cannot be held liable for defects and their consequences due to natural wear, improper cleaning or maintenance, non-compliance with operating, maintenance or connection instructions, calcination or corrosion, contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with factory specifications.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
1 User instructions | 1.1 User guide
3/66
Page 6
The warranty does not usually cover bulbs, glassware, rubber parts and the colour‐ fastness of plastics. Defects or their consequences that can be attributed to interventions on or changes made to the product by the customer or a third party are excluded from the warranty. Claims from this warranty can only be asserted when the transfer form (copy) be‐ longing to the product has been sent to KaVo, and the original can be presented by the operator or user.
1.5 Transportation and storage
1.5.1 Currently valid packaging regulations
Note
Only valid for the Federal Republic of Germany.
Properly dispose of and recycle the sales packaging, in accordance with the relevant packaging regulations, through waste management businesses or recycling com‐ panies that run a comprehensive return system. KaVo has licensed its sales pack‐ aging in accordance with this directive. Please conform with the regional, public waste-disposal system regulations.
1.5.2 Damage in transit
In Germany
If the packaging is visibly damaged on delivery, please proceed as follows:
1. The recipient of the package must record the loss or damage on the delivery receipt. The recipient and the representative of the shipping company must sign this delivery receipt.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use the product.
4. Report the damage to the shipping company.
5. Report the damage to KaVo.
6. You must contact KaVo before returning a damaged product.
7. Send the signed delivery receipt to KaVo.
If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows:
1. Report the damage to the shipping company immediately and no later than 7 days after delivery.
2. Report the damage to KaVo.
3. Leave the product and packaging in the condition in which you received it.
4. Do not use a damaged product.
Note
Failure on the part of the recipient to comply with any of the above obligations will mean that the damage will be considered to have arisen following delivery (in accordance with ADSp. Art. 28).
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
1 User instructions | 1.5 Transportation and storage
4/66
Page 7
Outside Germany
Note
KaVo is not liable for damage arising from transportation. Immediately inspect the delivery after receipt!
If the packaging is visibly damaged on delivery, please proceed as follows:
1. The recipient of the package must record the loss or damage on the delivery receipt. The recipient and the representative of the shipping company must sign this delivery receipt. Without this evidence, the recipient will not be able to assert a claim for damages against the shipping company.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use the product.
If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows:
1. Report the damage immediately or at least 7 days after the delivery to the de‐ livery company.
2. Leave the product and packaging in the condition in which you received it.
3. Do not use a damaged product.
Note
Failure on the part of the recipient to comply with any of the above obligations will mean that the damage will be considered to have arisen following delivery (in accordance with CMR law, Chapter 5, Art. 30).
1.5.3 Information on the packaging: Storage and transportation
Note
Please keep the packaging in case you need to return the product for servicing or repair.
The symbols printed on the outside are for transportation and storage, and have the following meaning:
Transport upright with the arrows pointing upwards
Fragile - protect against impact.
Protect from moisture.
Permissible stacking load
Temperature range
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
1 User instructions | 1.5 Transportation and storage
5/66
Page 8
Humidity
Air pressure
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
1 User instructions | 1.5 Transportation and storage
6/66
Page 9
2 Safety
2.1 Description of safety instructions
2.1.1 Warning symbol
Warning symbol
2.1.2 Structure
DANGER
The introduction describes the type and source of the danger. This section illustrates the potential consequences of non-observance.
▶ The optional step covers necessary measures for avoiding hazards.
2.1.3 Description of danger levels
Safety instructions with three hazard levels are used in this document for avoiding personal and property damage.
CAUTION
CAUTION indicates a hazardous situation that can cause damage to property, or mild or moderate physical harm.
WARNING
WARNING indicates a hazardous situation that can cause death or serious injury.
DANGER
DANGER indicates the maximum hazard level. indicates a hazardous situation that can di‐ rectly cause death or serious injury.
2.2 Purpose – Proper use
Note
Only the INTRAsurg 300plus with the IPX8 foot switch is permitted for use in op‐ erating rooms.
The ambient temperature must lie between +10℃ and +35℃. The relative humidity must be between 30% and 75%.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
2 Safety | 2.1 Description of safety instructions
7/66
Page 10
This KaVo product is intended only for use in the field of dentistry, for surgery to expose and dissect oral tissue structures (such as the periodontal gap, gingiva, bone, the jaw, extractions and implantations). The product may not be used for a purpose for which it was not intended. "Proper use" includes following all the instructions for use and ensuring that all in‐ spections and service tasks are performed.
The overarching guidelines and/or national laws, national regulations and the rules of technology applicable to medical devices for start-up and use of the KaVo product for the intended purpose are to be applied and complied with.
The user must ensure that that the unit works properly and is in a satisfactory con‐ dition before each use.
During use, national legal regulations must be observed, in particular:
▪ the applicable health and safety regulations ▪ the applicable accident prevention regulations
Users have a duty to:
▪ Only use equipment that is operating correctly ▪ to protect himself, the patient and third parties from danger. ▪ to avoid contamination from the product
To guarantee constant readiness for use and maintenance of value of the KaVo product, the recommended annual servicing must be done. Yearly safety inspections are required.
KaVo recommends an annual service check. In this service check, the safety checks are performed according to
IEC 62353:2007
as well as calibration. The safety check involves the following tests: Visually inspect the medical device and accessories; check the ratings of fuses that are accessible from outside; pro‐ tective conductor check and leakage current measurements as per IEC 62353:2007; medical device function test with reference to accompanying docu‐ mentation.
Authorized to repair and service the KaVo product:
▪ Technicians with appropriate product training from KaVo branches ▪ Technicians of authorized dealers specially trained by KaVo
In Germany, operators, equipment managers and users are obliged to operate their equipment in accordance with the MPG regulations. The services encompass all the test tasks required in accordance with § 6 of the medical devices operator ordinance (Medizinprodukte-Betreiberverordnung, MPBetreibV).
After servicing, manipulation or repair of the device, the device must be tested ac‐ cording to IEC 62353:2007 (state of the art) before it is used again.
Check the parameters of the device settings after servicing. Enter the service pa‐ rameters in the table on the last page.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
2 Safety | 2.2 Purpose – Proper use
8/66
Page 11
Note
The product must be cleaned and serviced according to instructions if it is not to be used for a long period.
Note
Any waste which is generated must be recycled or disposed of in a manner which is safe both for people and for the environment. This must be done in strict com‐ pliance with all applicable national regulations. Questions on proper disposal of the KaVo product can be answered by the KaVo branch.
Note
A recycling pass can be downloaded from www.kavo.com.
2.2.1 Information on electromagnetic compatibility
Conformance level: The immunity test levels required in IEC 60601 are met.
CAUTION
Damage from incorrect accessories The use of other accessories, transformers and lines than those indicated (with the exception of transformers and lines that KaVo sells as replacement parts for in‐ ternal components) can increase transmission or reduce the electromagnetic im‐ munity of the product.
▶ Only use accessories recommended by KaVo.
CAUTION
Interactions between devices Effect on proper use
▶ Do not use the device directly next to or stacked with other devices ▶ If it is necessary to use it next to or stacked with other devices, check for proper
use in this set up.
Note
Based on EN 60601-1-2 concerning the electromagnetic compatibility of electro‐ medical devices, we need to point out that:
• medical electrical devices are subject to special measures regarding electromag‐ netic compatibility and must be operated in accordance with KaVo assembly in‐ structions.
• portable and mobile high-frequency communications devices can influence med‐ ical electronics.
Note
KaVo cannot guarantee that accessories, lines and transformers not delivered by KaVo will correspond with EMC requirements of EN 60601-1-2.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
2 Safety | 2.2 Purpose – Proper use
9/66
Page 12
2.2.2 Risks from electromagnetic fields (implanted life-sustaining system)
DANGER
Risks from electromagnetic fields. The functions of implanted systems (such as pacemakers) can be influenced by electromagnetic fields.
▶ Patients and users must inquire about implanted systems before using the
motors and check the use. ▶ Evaluate the risks and benefits. ▶ Do not bring the motors close to the systems. ▶ Take suitable emergency precautions and immediately react to any changes
in health. ▶ Symptoms such as elevated heart beat, irregular pulse and dizziness are signs
of problems with a pacemaker.
2.2.3 Disposal of electronic and electrical devices
Note
According to EC directive 2002/96 concerning used electrical and electronic devi‐ ces, this product is subject to the cited directive and must be disposed of accord‐ ingly within Europe. Before disassembling / disposing of the product, it must be completely processed (disinfected, sterilized) according to the section "Preparation methods" Additional information can be obtained from KaVo (www.kavo.com) or your dental supplier.
For final disposal, contact:
Germany
To return an electrical device, proceed as follows:
1. At the homepage www.enretec.de of enretec GmbH, you can download a form for a disposal request under the menu item eom, or you can use it as an online
request.
2. Fill out the request with the corresponding information, and send it as an online request or by fax (+49(0)3304 3919 590) to enretec GmbH. The following avenues are also available for questions and for initiating a dis‐ posal request: Telephone: +49 (0) 3304 3919 500 E-mail: pickup@eomRECYCLING.com and Post: enretec GmbH, eomRECYCLING Department Kanalstraße 17 16727 Velten
3. Yourmovabledevice will be picked up in your practice, and yourpermanently in‐ stalledunit will be picked up at the curb at your address on the agreed deadline. The owner or user of the device will bear the costs for disassembly, transporta‐ tion and packaging.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
2 Safety | 2.2 Purpose – Proper use
10/66
Page 13
International (EU)
For country-specific information on disposal, contact your dental supplier.
2.3 Safety instructions
CAUTION
Electrical sparks in the product Explosion.
▶ Do not use product in areas subject an explosion hazard.
CAUTION
Damaged network cable/missing protective conductor. Electrical shock.
▶ Check the network cable before use. The socket outlet must have a protective
contact and meet the respective national guidelines.
▶ Always completely inset the power cable into the device's socket to connect to
the mains.
CAUTION
Damage due to liquids Faults on electric components.
▶ Protect product openings from penetration of liquids. ▶ Remove liquids from the inside of the device.
CAUTION
Unintended penetration of liquids. Electrical shock.
▶ Do not immerse the product in a tub-like container.
▶ Check and make sure that the coolant containers and lines do not leak before
each use. If liquid is found on the device, do not touch it and disconnect the power cable from the mains. Completely dry the surface of the unit before re‐ inserting the power cable.
CAUTION
Damage to the instrument hoses from stickers. Instrument hoses can explode.
▶ Do not affix stickers or adhesive tape.
CAUTION
Rotating parts while the pump is operating Injury
▶ Do not stick anything in the pump. Turn off the device when the pump is open.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
2 Safety | 2.3 Safety instructions
11/66
Page 14
CAUTION
Risks from electromagnetic fields. The functions of implanted systems (such as pacemakers) can be influenced by electromagnetic fields.
▶ Ask patients before treatment!
CAUTION
Electrostatic charge. Destruction of electronic components.
▶ Before contacting the hose coupling, touch the bottle holder with your hand to
release static electricity.
KaVo recommends that only original KaVo parts® be used for operating and repairs since their safety, operation and specific suitability have been tested in extensive tests.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
2 Safety | 2.3 Safety instructions
12/66
Page 15
3 Product description
3.1 INTRAsurg 300/300 plus
① Controls ⑤ Surgical hose coupling ② Bottle holder ⑥ Motor holder ③ Unlocking the pump mechanics ⑦ Surgery motor ④ Hose pump ⑧ Multifunctional foot control
The KaVo INTRAsurg 300 plus has an additional lighting function and recognition function for the INTRA CL light, hand and contra-angle handpiece.
Note
Additional information on surgery and implantology can be obtained at: www.ka‐ vo.com
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
3 Product description | 3.1 INTRAsurg 300/300 plus
13/66
Page 16
3.2 Controls
A Program button B Parameter button C Minus button D Plus button E Enter button
① Step (1... 6) ② Transfer ratio of the handpiece and
contra-angle handpiece ③ Speed (rpm) ④ Direction of rotation (< - / - >) ⑤ Coolant pump setting (1... 6) ⑥ Maximum torque (Ncm)
PRG
Open the program. Use the PRG button to open the steps; repress the button to switch to free use.
Para­me ter
Open setting mode. And Use the parameter button to select changeable parame‐ ters. Press the parameter button until the asterisk (*) flashes at the desired position. Changeable parameters:
- max. torque
- max. speed
- coolant pump setting
- Transfer ratio of the handpiece and contra-angle handpiece
- Direction of motor rotation
Reduce * value.
Increase * value.
Enter button: Save set of values/leave setting mode INTRAsurg 300 Plus: Request maximum torque.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
3 Product description | 3.2 Controls
14/66
Page 17
Note
The motor can only be started after you leave the setting mode by pressing the ENTER button. This saves all the changes. Only after this is done is the motor released to start.
3.3 Multifunctional foot control
(W) Shift pedal (Y) Hinged pedal (X) Right button (Z) Left button
Shift pedal W: Move forward for the default counter-clockwise rotation of the surgical motor Only with the INTRAsurg 300 Plus: The light can be switched on/off by moving to the left, and the spot light can be turned on by moving to the right.
Right button X: Procedural step button for selections in the program and for advancing the individual steps. Press briefly = proceed forward; press for a long time = proceed backwards. Double-click = open selection. An acoustic signal confirms proceeding forward/ backward when the motor starts.
Hinged pedal Y: Footswitch to start the motor and control its speed, and to start withdrawing the coolant without the motor running and then and then regulate the amount conveyed. The process is started by pressing the pedal down. The intensity is decreased by pressing the pedal to the left, and increased by pressing it to the right.
Note
Always check the input values before each use.
Left button Z: Coolant button to turn the coolant on/off and select rinsing without the motor.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
3 Product description | 3.3 Multifunctional foot control
15/66
Page 18
3.4 Rating plate for 300/300 plus
Installation side and securing the mains input
CE mark
VDE mark
CSA mark
Classification, type BF
Follow instructions for use
Mode: Continuous operation with intermittent load
Supply voltage
SN: YYYY = Year manufactured
XXXXXXX = Serial number REF: Material number Type: Device type
For disposal information, see use in accordance with intended purpose
GOST R certification
3.5 Technical data for the 300/300 plus
Width 291 mm
Depth 306 mm
Height 126 mm
Weight of the device 6.85 kg
Weight of foot control 2.35 kg
Weight of motor approx. 170 g
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
3 Product description | 3.4 Rating plate for 300/300 plus
16/66
Page 19
Input voltage 100 V∼/120 V∼/230 V∼
Input frequency 50/60 Hz
Power consumption Max. 260 VA
True power Approximately 7 Watt
Speed 300 – 40000 rpm
Pump delivery rate 25 – 150 ml/min
delivery pressure 1.5 – 2.2 bar
Multifunctional foot control Protection class IPX1
Only KaVo INTRAsurg 300 Plus: Multifunctional foot control, identifica‐ tion: blue FBE stirrup, rating plate. Per‐ missible in surgery rooms.
Protection class IPX8
Standard-length hose for the multifunc‐ tional foot control
1.90 m
Special-length hose for the multifunc‐ tional foot control Only available with the INTRAsurg 300plus.
3.5 m
Standard motor hose length Mat. no. 1.001.2651
2.00 m
Special length motor hose Mat. no. 1.004.6825
3.0 m
Mode continuous operation duty type
30 sec. operation/ 9 min.sec pause
The torque precision of the KaVo angle piece 27:1 (TRA C09 + C3) ranges from 20 to 30 Ncm at 20 to 50 rpm. Greater deviations are possible with oth‐ er angle pieces.
± 3 Ncm
Transportation and storage conditions
Ambient temperature -20℃ to +70℃
Relative humidity 5% to 95%
Air pressure 700 hPa to 1060 hPa
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
3 Product description | 3.5 Technical data for the 300/300 plus
17/66
Page 20
4 First use
4.1 Electrical connection
CAUTION
Damaged network cable/missing protective conductor. Electrical shock.
▶ Check the network cable before use. The socket outlet must have a protective
contact and meet the respective national guidelines.
▶ Always completely inset the power cable into the device's socket to connect to
the mains.
Both lines L and N are individually fused with T 3.15 H. The network frequency can be 50 or 60 Hz.
Requirement Make sure that the supply voltage is correct.
▶ First plug the inlet connector for non-heating apparatuses into the device, and
then plug the other end of the line into a socket.
See also: 4.3 Adjust the mains input voltage, Page 19
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.1 Electrical connection
18/66
Page 21
4.2 Unpacking
Note
Do not kink the motor hose since it may damage it.
Note
Save the box with all the packaging materials so that the device can be safely shipped (for example for a yearly service check). Always send both hose boxes so that the device can be safely packaged to prevent damage.
Prevent damage by properly removing from the package.
▶ Open box. ▶ Remove the hose box. Remove the multifunctional foot control and additional
equipment.
Note
When removing, note that the multifunctional foot control is tightly connected to the device.
▶ Pull out the INTRAsurg 300/300 plus vertically upward and place on a flat sur‐
face.
4.3 Adjust the mains input voltage
CAUTION
Voltage-conducting parts Electrical shock.
▶ Unplug the power cable from the power supply.
A country-specific mains input voltage (a rating play) is set at the plant. Check the setting and change it if necessary to the existing mains input voltage. The voltage selector ① can be set to 100V, 120V or 230V. The existing setting can be read at the cutout ④.
Remove the drawer ③. Remove the fuse holder ② from the fuse drawer ③, and turn it so that the required voltage value appears in the cutout ④.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.2 Unpacking
19/66
Page 22
Shove the fuse holder② back in the drawer ③. Insert the drawer ③ in the device. A code protects the drawer from being incorrectly inserted.
1
2
3
4
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.3 Adjust the mains input voltage
20/66
Page 23
4.4 Mount the bottle holder and use the correct coolant container
▶ Plug-in the bottle holder ①Mat. no. 0.761.1872 and rotate it until the bottle holder
faces away from the device and always secure with screw ②Mat. no. 1.005.8527.
CAUTION
Tipping danger from excessively heavy coolant reservoir Malfunctions
▶ Do not use coolant reservoirs that can hold more than 1 litre! ▶ Ensure that the device is standing securely.
In the case of glass bottles, insert an aeration needle. Otherwise, the supplied amount decreases with prolonged removal of coolant.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.4 Mount the bottle holder and use the correct coolant container
21/66
Page 24
WARNING
Electric shock due to coolant ingress Injuries
▶ Rotate the bottle holder away from the device to ensure that no leaking coolant
drips on the device or the electrical connection.
4.5 Connect the surgery motor and coolant reservoir
Note
The delivered parts are not sterile. Sterilise them before the first patient treatment. All parts conducting liquids must be sterilised.
▶ Plug in the surgery motor ⑥ at these surgery hose coupling and tighten with a
union nut ⑤. Follow the separate instructions for use for the motor.
▶ Place the surgery motor on the motor holder ①.
Note
The kink-protection part on the coolant hose inlet may not be twisted. Twisting the kink protection can cause defects in the house and lead to a restriction of coolant flow and causes the coolant to drip afterwards.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.5 Connect the surgery motor and coolant reservoir
22/66
Page 25
ESD symbol
CAUTION
Electrostatic charge. Destruction of electronic components.
▶ Before contacting the hose coupling, touch the bottle holder with your hand to
release static electricity.
▶ Insert the hose coupling ② into the device connection until it audibly locks into
place and the marking ring on the plug is covered.
▶ Shove the coolant hose ④ onto the plug-in nipple ③ by twisting slightly.
▶ Open the pump mechanics by lifting the lock ⑧.
Note
Insert the pump hose into the pump so that it is not clamped or pinched by the lock. Run all hoses relaxed without tension. Shove the hoses firmly on the plug-in nipple to ensure a tight connection.
4
9
8
▶ Close the lock ⑧ in the direction of the arrow.
▶ Insert the insertion needle ⑨ into the coolant reservoir, and hang the coolant
reservoir on the bottle holder. Check the seal and seat of the insertion tip. Pre‐ vent fluid from exiting the device.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.5 Connect the surgery motor and coolant reservoir
23/66
Page 26
CAUTION
Tipping danger from excessively heavy coolant reservoir Malfunctions
▶ Do not use coolant reservoirs that can hold more than 1 litre! ▶ Ensure that the device is standing securely.
Insert a sterile ventilation needle into glass bottles. The flow will otherwise be re‐ duced if coolant is withdrawn for a long period. Check the level of the coolant res‐ ervoir several times after waiting a while. The system must be leak-free. No liquid may escape from the coolant hose or the side of the connections.
Note
The saline solution corrodes all the metal parts. Immediately remove residue.
▶ Fasten the self-adhesive cable clip ⑦ Mat. no. 1.001.3348 to the INTRAsurg
300/300 plus (for the location, see the illustration).
▶ Insert the coolant hose ④ into the cable clip and close.
Note
Always monitor the level of the coolant reservoir. The device itself does not monitor the level. Use transparent coolant reservoirs.
If the product is used on potentially infectious patients, use disposable products for supplying coolant and thorough infection protection. KaVo recommends using a sterile disposable protective hose over the motor and hose (approx. 1.2 m) and a transparent disposable protective film over the device. The control buttons and dis‐ play must always be identifiable for operation.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
4 First use | 4.5 Connect the surgery motor and coolant reservoir
24/66
Page 27
5 Operation
5.1 Operation in general
5.1.1 Start up
CAUTION
Product damage from incorrectly adjusting the tension. Destruction of electrical components.
▶ Adjust the voltage input for the product to the existing mains voltage.
▶ Switch on the device.
An acoustic signal confirms that the device is turned on. The device runs a self test. The firmware version and note "Please check parameters" is briefly shown on the display.
▶ Note the displays.
If the device is not monitored, we recommend turning it off for safety reasons and to save energy.
Note
The last saved values are displayed.
5.2 Surgical motors
CAUTION
Incorrectly set parameters Damage from incorrect input values
▶ Check all input values before use.
Note
Following instructions for use, service instructions and installation instructions in the motor, handpiece and contra-angle handpiece packaging.
5.2.1 INTRA S 550 surgical motor
▶ Plug straight or contra-angle handpiece onto the motor coupling. ▶ Check secure seat before each treatment.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.1 Operation in general
25/66
Page 28
▶ Attach coolant hose.
1
5.2.2 INTRA LUX SL 550 surgical motor
Note
The INTRAsurg 300 plus has a recognition function for INTRA CL light, hand and contra-angle handpieces. Handpiece and contra-angle handpiece-specific param‐ eters are automatically transferred. Check all input values before use.
Mount the handpiece or angle piece
1
CAUTION
Heat transferred from high-pressure. Burns from touching the high-pressure lamp.
▶ Only operate the motor with the retention ring ① mounted.
▶ Place the INTRA CL light, hand and contra-angle handpiece on the surgery
motor and turn it until the catch audibly locks in place.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.2 Surgical motors
26/66
Page 29
CAUTION
Damage from changing the handpiece and angle piece during operation. Wear to the catch on the handpiece and angle piece and the motor Unbalanced motor axis.
▶ Change the handpiece and angle piece only when the motor not running.
CAUTION
Damage from operating with open chucks The handpiece or angle piece locks and rotates on its own axis.
▶ Only operate the handpiece or angle piece when the chuck is locked.
5.2.3 Remove the handpiece or angle piece
CAUTION
Damage from changing the handpiece and angle piece during operation. Wear to the catch on the handpiece and angle piece and the motor Unbalanced motor axis.
▶ Change the handpiece and angle piece only when the motor not running.
Note
Prevent to coolant hose from dripping on the motor. Coolant may not penetrate the motor.
▶ Remove the coolant hose from the handpiece or angle piece.
▶ Twist the handpiece or contra-angle handpiece slightly while pulling it off.
See also: 5.1.1 Start up, Page 25
5.2.4 Start the motor
▶ Start the motor. Press the slide pedal and change the speed by moving it the
side. Left stop: minimum speed; right stop: maximum preselected speed. When the slide pedal is no longer actuated, the selected speed is set on the multifunc‐ tional foot control. Repress the slide pedal to retrieve the last selected speed.
See also:
5.3.3 Preset the speed range, Page 30
5.4.4 Set parameters, Page 36
Note
When the hinged pedal is pressed in an unchanged position (to the side), the last selected position with an assigned motor speed is retrieved. The motor immediately runs at the assigned speed. After the program step changes, check the speed setting, and move the hinged pedal to the required position before pressing.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.2 Surgical motors
27/66
Page 30
CAUTION
Excessive speed. Injuries.
▶ In applications where the speed is critical, move the hinged pedal to the left
position (minimum speed) before the motor starts and move up to the desired speed.
▶ Set the speed to a safe maximum for the application.
5.3 Free use
5.3.1 Parameters menu
The following values can be changed in the parameter menu:
▪ Transfer ratio of the handpiece and contra-angle handpiece ▪ max. speed ▪ Direction of motor rotation ▪ Coolant pump setting ▪ max. torque
Para­me ter
▶ Press the parameter button.
A star ① flashes next to the alterable parameter.
Para­me ter
▶ Select the desired parameter by pressing the parameter button several times.
An asterisk * flashes next to the alterable parameter.
▶ Set the parameters to the desired value with the plus or minus key.
▶ Press the Enter key.
The set values are saved.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
28/66
Page 31
Note
Always save the selected setting with enter button when leaving the menu. Other‐ wise, the motor will not start (safety circuit).
Examples of saving values of handpieces and contra-angle handpie‐ ces
Note
Default values for the respective transfer ratios can be selected with the Plus or Minus buttons.
Indication Transfer ra‐
tio
Combined unit Reducing shank
Head
Speed range rpm
Maximum torque Ncm
Grinding, cutting
1:2 3555
1:2
-- 600-80.000 2,7
Grinding, Drilling, cutting
1:1 3,610 N
1:1
-- 300-40.000 5,5
Grinding, Drilling, Thread, Implant
12:1 3624 N
4:1
C 3 3:1
25-3.300 40
Grinding, Drilling, Thread, Implant
16:1 3624 N
4:1
67 RIC 4:1
18-2.500 45
Grinding, Drilling, Thread, Implant
27:1 C 09
9:1
C 3 3:1
11-1.500 55
Grinding, Drilling, Thread, Implant
36:1 C 09
9:1
67 RIC 4:1
9-1.100 55
Note
The listed indications are only examples. To prevent unnecessary risk, observe the guideline speeds given by the manufacturer of the rotating instruments. KaVo rec‐ ommends setting a speed range of 11-25/min or 11-35/min for screwing tasks.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
29/66
Page 32
5.3.2 Set the torque limitation
Para­me ter
▶ Press the parameter button until the asterisk flashes at the torque display ③.
Note
The torque values ③ are only for KaVo handpieces and angle pieces that operate properly.
The maximum value can be changed in the parameter menu by using the Plus and Minus buttons.
Note
Once 90% of these set torque has been reached, a signal sounds. When the set torque is exceeded during treatment, the motor stops. The foot pedal has to be pressed again to start the motor.
Checking the handpieces and contra-angle handpieces: Check the set transfer ratio, and let the motor and handpiece or contra-angle hand‐ piece run for 20 seconds at maximum speed. The torque ③ is an indicator for the condition of the instrument. Display of 0 = status is good Display < 0.5 = improve maintenance Display > 0.5 perform service
5.3.3 Preset the speed range
Note
Only the maximum value ① can be preset. Use the multifunctional foot control to select the speed within the predetermined range by simultaneously pressing the slide pedal and moving it to the side.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
30/66
Page 33
Para­me ter
▶ Press the parameter button until the asterisk flashes at the maximum speed
①.
The maximum value can be changed in the parameter menu by using the Plus and Minus buttons.
CAUTION
Excessive speed. Injuries.
▶ In applications where the speed is critical, move the hinged pedal to the left
position (minimum speed) before the motor starts and move up to the desired speed.
▶ Set the speed to a safe maximum for the application.
5.3.4 Set the coolant flow
Para­me ter
▶ Press the parameter button until the asterisk flashes at the coolant flow ⑤.
The flow ⑤ can be changed in the parameter menu by using the Plus and Minus buttons. 1: Pump at the minimum flow for approximately 25 mL/min 6: Pump at the maximum flow for approximately 150 mL/min The flow is determined by the size of the outlet. When the outlet is small, back­pressure arises in the hose that can cause the hose connections to separate. Drip‐ ping coolant indicates that this may occur.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
31/66
Page 34
W
Y
Z
Note
The pump can only be turned on and off with the left button on the multi-functional for control. Display when the pump is turned off ⑤: 0
To save the pump at OFF (0), press the button on the multifunctional foot control and save it with the enter button. The previously set flow is retained. If you can turn it on again by pressing the left button on the multi-functional foot control and save it with the enter button.
5.3.5 Selecting the handpiece and contra-angle handpiece
▶ Press the parameter button until the asterisk flashes at the transfer ratio ③.
Use the Plus and Minus buttons to scroll through the specified table of transfer ratios ③ of the instruments in the parameter menu. Always check the selection in the display.
Note
The maximum permissible speed and torque of the selected handpieces and con‐ tra-angle handpieces are displayed.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
32/66
Page 35
5.3.6 Switching the direction of motor rotation
Note
The direction of motor rotation can only be switched with the multi-functional for control. Counterclockwise rotation <- ④ remains active until the next switch. If counterclockwise rotation is not saved by pressing the ENTER button, the motor rotates clockwise -> when the device is turned on. Wait for the acoustic anti-clock‐ wise signal.
W
Y
Z
▶ Move the shift pedal in the direction of the arrow.
With counterclockwise rotation <- the rotational direction display ④ flashes, and two acoustic signals sound. When the motor is started, three acoustic signals sound. In the PRG menu, counterclockwise rotation can be saved by pressing the enter but‐ ton.
Note
No display of the torque value ③ in counterclockwise rotation.
5.3.7 Rinsing function
This function can be selected from each step, but it cannot be saved.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
33/66
Page 36
W
Y
Z
▶ Press the left button for 3 seconds to select the rinsing function.
▶ Start the drawing of the coolant by pressing the slide pedal.
The flow can be changed in six steps by moving the slide pedal to the side.
▶ The rinsing function is shown in the display.
W
Y
Z
▶ Press the left button briefly to stop the rinsing function.
The device returns to the previously selected function.
5.3.8 Surgical motor in free use
Note
This function is only supported by the INTRAsurg 300 plus in conjunction with the
INTRA CL light, hand and angle pieces.
The recognition function of the INTRA CL light, handpiece and contra-angle hand‐ pieces enable parameters for the free use of handpiece and contra-angle handpie‐ ces to be recognised and adopted.
5.3.9 Request maximum torque
Note
This function is only supported by the INTRAsurg 300 plus.
▶ Hold down the enter button after the motor stops.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.3 Free use
34/66
Page 37
The maximum torque ① of the last motor operation is displayed in Ncm.
Note
The torques are overwritten when the motor restarts under a minimum load of 5 Ncm.
5.4 Program
5.4.1 Start the program
The product has a basic program with six programmable steps. The current step is shown in the display ①.
▶ Press the right button briefly.
or
PRG
▶ Press the program button once.
An acoustic signal confirms the program selection.
PRG
▶ To switch to free use from the program, press the program button again briefly.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
35/66
Page 38
5.4.2 Switch forward step
▶ Press the right button briefly.
or
▶ Press the plus button.
An acoustic signal confirms the entry.
5.4.3 Go backward a step
▶ Press the right button for a while.
or
▶ Press the minus button.
An acoustic signal confirms the entry.
5.4.4 Set parameters
▶ Select the step ① to be changed.
Para­me ter
▶ Switch to the parameter menu by pressing the parameter button.
See also: 5.3 Free use, Page 28
The values shown for a step are default values that you can use to start work im‐ mediately. They can all be changed in the parameter menu and can hence be adap‐ ted to the your individual approach.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
36/66
Page 39
Changed values can be saved in the parameter menu and are then available for the next use.
The following values can be changed in the parameter menu:
▪ Transfer ratio of the handpiece and contra-angle handpiece ▪ max. speed ▪ Direction of motor rotation ▪ Coolant pump setting ▪ max. torque
See also: 5.3 Free use, Page 28
CAUTION
Incorrect use. danger of injury.
▶ Always check the values before use.
5.4.5 Limit the number of steps
Six steps are saved upon first start up and after setting up the delivery status.
Example: Limiting the steps
The number of steps should be limited to four.
▶ Select step 4 ①.
PRG
▶ Press the program button for at least three seconds.
An acoustic signal confirms the entry.
Note
Now only the first four steps can be selected. When you continue to step 4, the message END appears.
If you continue, you go to free use.
Remove restriction
▶ Select the last step ① (4 in this example).
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
37/66
Page 40
PRG
▶ Press the program button for at least three seconds.
An acoustic signal confirms the entry.
5.4.6 Setting up delivery status
Note
All entries and changes made are deleted with this function.
Note
If the device was initialised, everything saved previously is deleted. Therefore, for simplicity, everything that was saved should be noted down for easy recovery.
PRG
▶ Turn device on while pressing the programme button.
After initialization, the device is ready for use in its condition at the time of de‐ livery.
Condition at the time of delivery: Programme
Step Instrument Speed (rpm) Torque
(Ncm)
Pump Direction of
motor rota‐
tion 1 27:1 11-1500 55 4 -> 2 27:1 11-50 40 4 -> 3 27:1 11-50 40 4 -> 4 27:1 11-50 40 4 -> 5 27:1 11-50 55 0 -< 6 27:1 11-50 10 0 ->
Condition at the time of delivery: Free application
Step Instrument Speed (rpm) Torque
(Ncm)
Pump Direction of
motor rota‐
tion 27:1 11-1500 55 4 ->
5.4.7 Surgical motor in the program
Requirement The mounted INTRA CL light, hand and contra-angle handpiece must be selected on the display.
If the mounted handpiece or contra-angle handpiece is not selected, the device asks for the corresponding input.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
38/66
Page 41
if the motor is turned on without the corresponding entry, three acoustic signals sound, and the read-in instrument flashes on the device display with question marks, for example: ?27:1? .
Note
The motor will not start (safety circuit). To start the motor, either the selected handpiece or contra-angle handpiece must be mounted, or the mounted handpiece or contra-angle handpiece must be saved using the ENTER button.
Save the light, hand and contra-angle handpieces (without the moun‐ ted and peace and contra-angle handpiece)
Para­me ter
▶ Press the parameter button until the asterisk flashes at the transfer ratio.
▶ Set the correct parameters with the Plus or Minus button.
▶ Save the set values with the enter button.
The data are now saved in this step.
5.4.8 Request maximum torque
The maximum torque is saved when you advance a step. When the motor is started in the next step, the highest torque is always saved.
▶ Press the enter button after closing the program.
In each step ①, the max. torque ② is shown in the display.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.4 Program
39/66
Page 42
▶ To leave the display, press the enter button.
All saved torques are deleted.
Note
If you leave the program buy pressing the plus button or the right button on the multifunctional foot control, all torques are deleted. If you accidently go too far and skip the torque display, the values can be reactivated and displayed by pressing the right button on the multifunctional foot control.
5.5 Functions exclusively with the 300 plus program version
Additional functions of the INTRAsurg 300 plus with an INTRA LUX SL550 motor:
▪ Light function. ▪ IPX8 multifunction foot control (blue stirrup, rating plate). ▪ Recognition of the INTRA CL light, hand and angle piece. The transfer ratio and
associated values are shown on the display.
▪ The user can perform additional calibration with the INTRAsurg Calibration
1.002.3569. (27:1 angle piece is required)
▪ Display of the maximum torques. The maximum torque is saved in each step.
The max. torque can be read on the display during surgery or after the program ends (only with clockwise rotation).
▪ Read-out table for the torque values at the end of applications.
5.5.1 Recognition function
Note
To prevent misinformation, the contact's provided for recognition always need to be regularly cleaned on the motor and instrument pusher. The contacts always must be shiny. Wipe off the contacts with a cloth soaked in disinfectant and rub dry.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
40/66
Page 43
▶ Place the INTRA CL light, hand and angle handpiece on the INTRA LUX SL 550
motor and turn it until the catch ① audibly locks in place.
The hand and angle piece is recognised, and the corresponding saved parameters are shown on the device display ②.
CAUTION
Incorrect input values. Danger of injury.
▶ The input values must be checked before each use.
These parameters are immediately activated in free use.
The recognised handpiece or angle piece must be selected in the program. If it is not selected, the device asks for the corresponding input. Three acoustic warning signals sound, and the read-in instrument flashes on the device display with a ques‐ tion mark, for example: ?27:1?
Note
For safety reasons, the motor does not start (safety circuit) since the input value on the display always has priority.
▶ Check the parameters. Values displayed by the device can be saved by pressing
the ENTER button.
The data of the mounted INTRA CL light, hand and angle piece are only saved in this step. The motor starts with these released parameters.
Note
This must be done for the saved parameters to be reliably maintained and contin‐ uously followed.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
41/66
Page 44
5.5.2 Light functions
Place the INTRA CL light, hand and contra-angle handpiece on the INTRA LUX SL 550 motor and turn it until the catch audibly locks in place. The INTRA CL light, hand and contra-angle handpiece is recognised and the corresponding parameters ap‐ pear on the display (this function is only possible with the INTRA CL light, hand and contra-angle handpiece).
Light on/off without motor and pump running (spot)
▶ Move the shift pedal to the right.
The light only shines when actuated (spotlight function).
Default light when the motor starts
▶ Move the shift pedal to the left.
The light is turned on or off. When the light is turned on, the following appears in the display: LUX (this function is only possible with INTRA CL light, handpieces and contra-angle handpieces).
Note
If there is no light function, check if the light is turned on and if the INTRA CL light, hand and contra-angle handpiece is properly mounted.
Afterglow time
The light shines for 3 sec. after the motor stops running.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
42/66
Page 45
Note
The motor can only operate with retention ring ①Mat. no. 1.001.5029.
1
Note
Only the light function is activated and a 1:1 hand and angle piece is displayed with hand and angle pieces with a metal snap-in button. The torque is not displayed.
5.5.3 Request maximum torque (free use)
▶ Hold down the enter button after the motor stops.
The maximum torque ① of the last motor operation is displayed in Ncm.
Note
The torques are overwritten when the motor restarts under a minimum load of 5 Ncm.
5.5.4 Request maximum torque (program)
See also: 5.4.8 Request maximum torque, Page 39
5.5.5 Recommended programming when setting a number of im‐ plants one after the other
Mode of operation: Perform all tasks for drilling an implant in Free use mode. Screw in the respective implant in PRG steps (max. 6). At the end of the task, the torque screw-in value can be requested and manually documented for evaluating primary stability (in the form of the following table, for example).
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
43/66
Page 46
Step Instrument Speed (rpm) Torque
(Ncm)
Pump Motor r/l
1 27:1 11 – 50 40 0 -> 2 27:1 11 – 50 40 0 -> 3 27:1 11 – 50 40 0 -> 4 27:1 11 – 50 40 0 -> 5 27:1 11 – 50 40 0 -> 6 27:1 11 – 50 40 0 ->
5.5.6 INTRAsurg calibration
The influence of the efficiency of the instruments is taken into account when testing the torque display. Store the INTRAsurg calibration device at 15°C to 30°C to provide precise calibra‐ tion. The calibration value refers to the temperature of the INTRAsurg calibration of 23°C. If this temperature is not maintained, the value needs to be corrected ac‐ cording to table. For example: Calibration at 20°C (read 15.2) results in a corrected input value (+0.3) of 15.5.
Tempera‐ ture
15℃ 16℃ 17℃ 18℃ 19℃ 20℃ 21℃ 22℃ 23℃ 24℃ 25℃ 26℃ 27℃ 28℃ 29℃ 30℃
correc‐ tion
+0.8 +0.7 +0.6 +0.5 +0.4 +0.3 +0.2 +0.1 0 -0.1 -0.2 -0.3 -0.4 -0.5 -0.6 -0.7
A sterile instrument can also be used for calibration. To prevent contamination, use a sterile catch pin.
Note
The INTRAsurg Calibration system can be disinfected.
▶ Pre-position cleaned and set up angle piece 27:1 and INTRAsurg Calibration
Mat. no. 1.002.3569.
▶ Press the parameter button for 3 seconds.
The following message appears on the display:
▶ Read the calibration affixed to the bottom of the INTRAsurg Calibration (such as
15.2), correct the value according to the temperature, and enter with the Plus and Minus keys.
▶ Confirm with the enter button.
▶ Insert the catch pin into the INTRAsurg Calibration. ▶ Place the INTRAsurg Calibration on a fixed base and hold.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
44/66
Page 47
▶ Start the motor. The motor always rotates to the right.
Note
Do not apply any pressure. The motor may not experience any additional load.
The process can be monitored on the display.
After calibration, the following appears on the display:
If calibration is unsuccessful, the following appears on the display:
Note
The output values are retained in case of interruption or error.
The calibration automatically ends. The INTRAsurg 300 plus is ready with the new setting.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
5 Operation | 5.5 Functions exclusively with the 300 plus program version
45/66
Page 48
6 Preparation methods DIN EN ISO 17664
Note
Before preparations, follow the instructions for use of the surgery motor, handpie‐ ces and contra-angle handpieces, and the scaler SONICflex and special tools (bone). The instructions for cleaning and sterilisation have been validated by the manu‐ facturer. Any deviation from the provided instructions by the person preparing the device should be checked to see if it is effective, and potential negative conse‐ quences should be evaluated. Sufficient safety that prevents microorganisms that have entered the instruments from becoming a possible cause of infection is only provided by carefully process‐ ing the outside and inner surfaces directly after use. The handpieces and angle pieces must be packaged and sterilised and sterile when used. If potentially infectious liquids and particles can contact the products, it is recom‐ mendable to cover and protect these areas with sterile disposable products. Frequent preparation with proper use does not significantly affect the unit and mo‐ tor. The product life normally ends due to wear and damage from use. Observe all manufacturer instructions. Treat KaVo gently when cleaning them. Remove deposits such as saline coolant residue, blood and saliva with a disposable cloth immediately after use. Only use fully functional products for treating patients. Observe all current national hygiene regulations. See also www.rki.de (infection protection)
Note
A German language CD on hygienic measures in medical practices is obtainable from: MEDIEN & Forum, Am Brunnen 8, 29229 CELLE – Tel. D 05141/370188/89. E-mail: hammer@medien-information-bildung.de. Additional information can be obtained at: www.rki.de
Site of use
Remove moisture-sensitive parts close to the disinfection site. The treatment sur‐ faces of the area should be hygienically smooth surfaces. Protect difficult to clean surfaces and objects from contamination by covering them.
Preparation for decontamination
▪ Wear suitable protective clothing. ▪ Wear eye protection. No eating, drinking or smoking. ▪ Disinfect hands properly. ▪ Turn off the device and unplug it. ▪ Prepare and follow a hygiene plan. ▪ Only use disinfectants suitable for this group of devices. ▪ Prepare the parts suitable for sterilisation: Care, rinsing and packaging. ▪ Use qualified trained personnel for preparing dental equipment and instruments.
Train the personnel according to these instructions for use Plan and provide a schedule.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 5.5 Functions exclusively with the 300 plus program version
46/66
Page 49
6.1 Cleaning
Always clean directly after using the medical device to prevent contaminants from hardening. Saline solution contains sodium chloride which corrodes metal. Quickly remove all residue to prevent surface damage.
6.1.1 Machine cleaning
Not applicable for INTRAsurg device parts.
CAUTION
Damage due to liquids Faults in electric components.
▶ Do not machine-clean the INTRAsurg device.
6.1.2 Manual cleaning
▶ Wipe off the supply unit (all visible surfaces), motor hose, motor holder, bottle
container, foot control surfaces and the connecting lines with a damp disposable cloth. Do not wipe the cloth on edges and press so that liquid will not penetrate. Remove all residue and dry the device.
Note
After uses that run the risk of infection, rinse the coolant lines of the handpieces and contra-angle handpieces with a disinfectant (approx. 20 ml) with a disposable syringe. Avoid damage. Rinse the coolant lines with demineralised water, then drain and dry them.
6.1.3 Rinse the hoses
Preparation for product decontamination: Directly after each use, rinse all the hoses to prevent malfunctions from the crys‐ tallisation of the saline solution. All of the hoses must be completely emptied. Then sterilise them and observe the hygienic requirements for storage and maintaining sterility.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning
47/66
Page 50
▶ Remove the pump hose needle from the coolant reservoir. Immerse the pump
hose needle in a clean container with demineralised water (50-60 ml). Place the spray hose from the instrument in an empty container or sink.
Note
Rinse until all the hoses are drained. Do not immerse the water outlet into the liquid. Observe the necessary measures for infection protection.
W
Y
Z
▶ Press left button 3 sec. The rinsing function is selected.
See also: 5.3.7 Rinsing function, Page 33
▶ The rinsing function is shown in the display.
▶ Hold down slide pedal and slide it to the right until the stop. The hose pump runs
at the max. rate.
▶ Ensure that the water jet is in the right direction and that it is the right amount.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning
48/66
Page 51
W
Y
Z
▶ Press the left button briefly.
You exit the rinsing menu.
6.2 Disinfection
Note
The motor and motor hose cannot be thermodisinfected. After treating each patient, disinfect the surfaces close to the patient that are contaminated by contact or aer‐ osols. Wipe disinfection should be used in all cases.
6.2.1 Automated disinfection
Not applicable for INTRAsurg device parts.
6.2.2 Manual disinfection
▶ Pull the insertion needle ③ out of the coolant reservoir ②. ▶ Open the lock ① and remove the hose ④.
▶ Remove all hoses marked gray including tips and surgical motor on the unit. ▶ Wipe-disinfect surfaces with a soft, disposable cloth and permissible disinfec‐
tants. Make sure that the device is completely wet. Let the disinfectant act for the prescribed time. Dry the surface.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.2 Disinfection
49/66
Page 52
Permissible disinfectants (the uses correspond to the existing manufacturer's in‐ structions and national guidelines. observe the safety data sheets.) KaVo recom‐ mends the following products based on material compatibility. The microbiological efficacy must be ensured by the disinfectant manufacturer.
▪ Incidin Liquid (Ecolab) ▪ FD 322 Dürr ▪ Schülke & Mayr Microcide ▪ Isopropanol 70%
6.2.3 Drying
▶ After disinfection and sterilisation: Let the surfaces and plug-in parts for the ho‐
ses and cables (contacts) completely dry in the room before they are used again.
6.2.4 Service, inspection and testing after preparation
▶ Test of cleanliness and intactness, care, repair: check the hose connections for
ease of plugging and secure attachment. Check the function of device settings and motor operation. Check hose pump for sufficient coolant. Check secure at‐ tachment of hose connections. Check control commands on foot control. Check recognition function of INTRAsurg 300 plus. Comply with hygiene requirements (sterility) during the check. If there are any sites of breakage or clearly recog‐ nised changes on the surface, the parts must be checked by the Service.
6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
CAUTION
Product damage due to improper sterilisation. Damage to the sterile product.
▶ No hot air sterilisation, no chemical cold sterilisation, do not sterilise with eth‐
ylene oxide!
CAUTION
Product damage due to improper sterilisation. Damage to the sterile product.
▶ An initial vacuum can shorten the product life.
CAUTION
Moisture Non-sterility
▶ Ensure dryness. Autoclaves with a subsequent vacuum ensure dryness. In
addition, drying can be accelerated with a 10 minute drying phase with the autoclave door open.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
50/66
Page 53
CAUTION
Product damage Contact corrosion
▶ Remove the sterilised item from the autoclave immediately after sterilising and
drying.
CAUTION
Crystallising saline solution Malfunctions
▶ Rinse all hoses and drain them completely before they are sterilised.
Note
The user is responsible for observing the regulations and conditions for sterility. After each patient, dispose of the coolant reservoir, rinse the hoses, and completely drain and sterilise them.
Note
When treating patients who may have an acute, critical infectious disease, be sure to observe the hygienic measures cited in applicable publications and reports. If possible, use suitable disposable products to avoid the transmission of critical pathogens. These protect the user, the patient and all participants in the surgery. A microbiologically appropriate coolant liquid must be used. Only use prescribed isotonic saline solution E NaCl 0.9 (identified as an infusion solution; follow the instructions on the package insert) for cooling and spraying wounds. All materials that are contaminated from the dental and medical fields must suitably processed and sufficiently identified after cleaning and sterilisation (RKI recommendation, www.rki.de).
See also: 6.1.3 Rinse the hoses, Page 47
The KaVo medical devices released for sterilisation have a maximum temperature resistance of 138 ℃.
6.3.1 Packaging
The sterilisation bag must be large enough for the items to be sterilised so that the bag is not stretched. The quality and use of the packaging of the items to be sterilised must satisfy the applicable standards and be appropriate for sterilising.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
51/66
Page 54
▶ Wind all hoses previously cleaned and disinfected that are marked grey including
tips and the surgical motor on the motor holder ⑥. Seal the motor holder, handpieces and contra-angle handpieces in sterile bags or place them on a surgical tray.
6.3.2 Sterilisation
▶ Sterilise in saturated steam at 135±1℃, gravity, for at least 4 minutes.
Note
Let the sterilised materials cool and try to the ambient temperature before using them again.
6.3.3 Storage
Observe all necessary measures for hygiene when storing sterile goods. Store pro‐ tected from dust, release with identification on the packaging. Monitor storage length.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
6 Preparation methods DIN EN ISO 17664 | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
52/66
Page 55
7 Troubleshooting
Note
If your problem cannot be rectified after reading the following troubleshooting doc‐ umentation, please contact a KaVo-trained Service Technician.
In case of a problem, a fault number is displayed with: "warning xx" or "remark xx" Fault numbers 1 to 67 are possible.
Note
Confirm all other error messages with the ENTER button. If the error message does not disappear or the error is reported again, please notify customer service. Please always write down what you want or the service you want performed when you send in the device. The settings must be rechecked after the device is returned.
Malfunction Cause Remedy Nothing works. Fuse blown. ▶ Check and if necessary replace
the fuse.
The unit must be switched off. ▶ Turn on the mains switch on the
back of the unit.
Both ends of the power line are not plugged in.
▶ Plug in the mains power line:
The wrong handpiece or angle piece is mounted or set on the de‐ vice.
▶ Adjust the right handpiece or an‐
gle piece on the device.
Unknown. ▶ Turn the machine off and on.
No coolant in the instrument. Coolant flow not preselected. ▶ Set the coolant flow.
See also: 5.3.4 Set the coolant flow, Page 31
The pump hose is sticky or crusty. ▶ Exchange the pump hose.
See also: 7.1 Exchange the pump
hose, Page 56 The coolant in the instrument is in‐ sufficient.
The spray nozzles are crusty or soiled.
▶ Clean the spray nozzle with the
nozzle needle.
The pump hose is sticky or crusty. ▶ Exchange the pump hose.
See also: 7.1 Exchange the pump
hose, Page 56
Glass bottle with saline solution is not ventilated.
▶ Ventilate the glass bottle.
Coolant is dripping Clogged coolant line or restricted
flow
▶ Check if the coolant hose and
connecting nipple are blocked.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
53/66
Page 56
Malfunction Cause Remedy
The coolant pump is set too high. ▶ Adjust the coolant flow.
See also:
5.3.4 Set the coolant flow, Page 31
7.1 Exchange the pump hose, Page
56
The pump hose is too long. ▶ To avoid dripping, keep the
pump hose length after the pump as short as possible.
Leaky hose pump housing. The pump hose is worn. ▶ Exchange the pump hose.
See also: 7.1 Exchange the pump
hose, Page 56 Leaky surgical hose. Worn or defective coolant hose. ▶ Exchange coolant hose.
See also: 7.2 Exchange coolant
hose, Page 57
▶ Shorten the hose by approx. 6
mm.
The motor shuts off during treat‐ ment without an error message.
The wrong handpiece or angle piece is mounted or set on the de‐ vice.
▶ Adjust the right handpiece or an‐
gle piece on the device.
The motor makes a grinding noise. The motor is not correctly plugged
on or screwed on.
▶ Firmly insert the motor hose into
the housing.
▶ Firmly screw on the motor hose
to the motor.
▶ Turn the machine off and on.
▶ Check if all the connections and
couplings are firmly seated.
The motor runs out of true. The motor is not correctly plugged
on or screwed on.
▶ Firmly insert the motor hose into
the housing.
▶ Firmly screw on the motor hose
to the motor.
▶ Turn the machine off and on.
▶ Check if all the connections and
couplings are firmly seated.
No light on the handpiece and angle piece.
The light is not turned on. ▶ Turn on the light.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
54/66
Page 57
Malfunction Cause Remedy
The hand and angle piece is im‐ properly attached.
▶ Mount the handpiece and angle
piece until the catch audibly locks.
The lamp is faulty. ▶ Replace the lamp
Mat. no. 1.002.2928.
See also: Instructions for use for the
motor.
The catch is dirty. ▶ Clean the catch (gold contacts).
See also: Instructions for use for the
motor and instrument.
No suitable light, hand and angle piece
▶ Use a suitable light, hand and
angle piece.
Display: >>remark 65
-motor is overloaded
Motor overload warning ▶ Stop treatment as soon as pos‐
sible and let the motor cool off for a few minutes.
Display: >>warning 50
-motor is overloaded
Motor is overloaded, the maximum permissible motor output was ex‐ ceeded.
▶ The motor is blocked for 10 mi‐
nutes; let it cool.
Display: >>remark 66
-motor not plugged
The motor hose is not correctly in‐ serted.
▶ Properly insert the motor hose.
The motor is not correctly connec‐ ted to the motor hose.
▶ Properly plug in the motor.
Display: >>remark 67
-motor not plugged.
When the motor starts, the discon‐ nection of the motor from the hose is detected.
▶ Plug the hose onto the motor
and screw it tight. This display disappears after eight seconds. The motor can be restarted after this time.
Display: >>warning 36
-missing foot control
Defective footswitch or defective connection to the footswitch, possi‐ bly by unintentional penetration of moisture.
▶ Dry the footswitch.
Turn the machine off and on
Display: error 3
Internal system error ▶ Exchange the pump hose. Con‐
tact service.
Display: error 4
Internal system error ▶ Exchange the pump hose. Con‐
tact service.
Display: error 29
The motor hose is damaged. ▶ Check the motor hose and ex‐
change it if necessary
Display: warning 49
The power consumption of the sur‐ gical motor was too high.
▶ Press the ENTER button and
continue. Exchange the motor hose. Exchange surgical motor.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060
55/66
Page 58
7.1 Exchange the pump hose
Note
The parts delivered for exchange are not sterile. Rise and sterilize them before the first treatment.
See also: 6.3 Sterilisation in a steriliser in compliance with DIN EN 13060, Page 50
Note
If the pump hose becomes brown after extended use, it must be exchanged.
▶ Make sure that all lines containing saline solution are completely empty. ▶ To drain the lines, remove the insertion needle ③ from the saline solution res‐
ervoir ② and activate the pump by using the multifunctional foot control until the hoses are free of saline solution. Remove the container and store it so that it does not drain.
▶ Turn off the INTRAsurg 300/300 plus.
CAUTION
Running, open hose pump. Injury hazard
▶ Turn off the device before opening the hose pump.
▶ Lift the lock and remove the pump hose.
Insert the new pump hose Mat. no. 1.001.3347.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.1 Exchange the pump hose
56/66
Page 59
Note
Insert the pump hose into the pump so that it is not clamped or pinched by the lock. Run all hoses relaxed without tension. Shove the hoses firmly on the plug-in nipple to ensure a tight connection.
Note
To avoid dripping, keep the pump hose length after the pump as short as possible.
▶ Carefully turn the lock ① downwards and turn it until it locks into place.
The pump hose length after the pump determines the dynamic pressure in the cool‐ ant hose after the motor or pump stops (drop formation). Pump hose ① short: less accumulated volume in the coolant hose (less potential dripping) Pump hose ② long: large accumulated volume in the coolant hose (increased drip‐ ping possible) Observe hygiene to prevent infection: The liquid in the coolant hose may not draw back to far (avoid contamination).
7.2 Exchange coolant hose
Note
The parts delivered for exchange are not sterile. Rise and sterilise them before the first treatment.
The coolant hose is run to the handpiece/contra-angle handpiece and consists of 2 pieces: Coolant hose ①(approx. 2.3 m long and obtainable as a section under Mat. no. 0.065.5279)
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.2 Exchange coolant hose
57/66
Page 60
Coolant hose ④(obtainable approx. 150 mm long under Mat. no. 0.593.0252 and as a section under Mat. no. 0.065.5188)
7.2.1 Exchange the coolant hose from the handpiece or angle piece.
▶ Pull off the coolant hose ④ Mat. no. 0.593.0252 (150 mm), or
Mat. no. 1.001.3436 (70 mm), or Mat. no. 0.065.5188 (cut section) off the cou‐ pling ③ Mat. no. 1.001.6462 and the handpiece or angle piece and exchange.
7.2.2 Exchange the coolant hose in the motor hose
Note
Run the new hose straight in the direction of tension so that it can be pulled in as easily as possible. Talcum powder can be used as a lubricant.
▶ Remove the coolant hose ① from the plug-in nipple ② and coupling ③.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.2 Exchange coolant hose
58/66
Page 61
▶ Connect the new and existing coolant hose ① to the coupling ③.
▶ Carefully remove the old hose from the motor end and simultaneously shove on
the new one until the coupling ③ piece is visible.
▶ Pull the coupling ③ out of the hose opening e.g. using a tweezers so that the
coupling piece is not clamped there.
7.2.3 Disposable coolant hose
Sterile-packaged coolant hoses can be ordered under Mat. no. 1.001.9902 (pack‐ age contents: 10 units). This should be used for applications with an increases infection risk or when the motor hose cannot be prepared according to the existing instructions. The disposable coolant hose consists of an insertion tip, pump hose, clip and hose holder. The length of the hose can be cut at the exit end with a sterile knife or sterile scissors. Observe the instructions regarding use on the package.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.2 Exchange coolant hose
59/66
Page 62
▶ Connect the coolant hose directly to the outside of the motor hose without loops
or kinks, and affix with the accompanying holders (evenly distributed).
▶ Before using the pump, completely open the inserted hose clip ①.
The coolant flow is also affected by the outlet in the handpiece or angle piece. A focused water jet should arise.
▶ Set the pump flow on the device.
Note
To avoid dripping, keep the pump hose length after the pump as short as possible.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.2 Exchange coolant hose
60/66
Page 63
7.3 Device settings table
Enter the set parameters in the table before each service check. Always check the parameters using the table after the service check.
Date: _______________________ Name: __________________
Step Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Free
Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Date: _______________________ Name: __________________
Step Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Free
Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Date: _______________________ Name: __________________
Step Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Free
Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Date: _______________________ Name: __________________
Step Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Free
Instrument Speed (rpm) Torque (Ncm) Pump Motor r/l
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
7 Troubleshooting | 7.3 Device settings table
61/66
Page 64
8 Accessories
The INTRAsurg 300/300 plus contains the follow permissible accessories:
▪ INTRA S 550 surgical motor Mat. no. 1.000.8072 ① (autoclavable collectorless
motor with a speed range of 300 to 40000 rpm)
▪ Alternative: INTRA LUX SL 550 surgical motor Mat. no. 1.001.3421 ① (auto‐
clavable collectorless motor with a speed range of 300 to 40000 rpm)
▪ Bottle holderMat. no. 0.761.1872 ②
Do not use any container with a volume in excess of 1 litre (danger of tipping
over). ▪ Pump hoseMat. no. 1.001.3347 ③ ▪ Disposable coolant hose sterileMat. no. 1.001.9902 ▪ Motor hose 2 m (standard equipment)Mat. no. 1.001.2651 ④
Motor hose 3 m (please inquire)Mat. no. 1.004.6825 ④
The connection hose with coolant guidance can be autoclaved. ▪ Motor-Steri supportMat. no. 0.726.2922 ⑤ ▪ IPX 8 foot control element with 3.5 m connection cable (please in‐
quire)Mat. no. 1.004.8276 ⑦ ▪ Steri-Set INTRAsurg 300 (please inquire)Mat. no. 1.001.4953 includes the fol‐
lowing parts: 1x motor hose 2 m Mat. no. 1.001.2651; 1x surgical motor INTRA
S 550; 1x motor-Steri support Mat. no. 0.726.2922; 1x hose 150mm
Mat. no. 0.593.0252; 1x pump hose Mat. no. 1.001.3347.
For the INTRAsurg 300 Plus:
▪ INTRAsurg Calibration Mat. no. 1.002.3569 ⑥ (disinfectable). ▪ Steri-Set INTRAsurg 300plus (only upon request) Mat. no. 1.001.4968 consist‐
ing of the following parts: 1 motor hose 2 m Mat. no. 1.001.2651; 1 surgical motor
INTRA LUX SL 550; 1 motor sterile holder Mat. no. 0.726.2922; 1 hose 150 mm
Mat. no. 0.593.0252; 1 pump hose Mat. no. 1.001.3347.
KaVo surgical, handpiece and angle piece line corresponding to the order. For news on surgery and implantology, see www.kavo.com.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
8 Accessories
62/66
Page 65
9 Details on electromagnetic compatibility
9.1 Guidelines and manufacturer's declaration - electromagnetic transmission
The IS 300 is for use in an environment like the one cited below. The IS 300 customer or user should ensure that use takes place in such an environment.
Measurements of noise transmis‐ sions
Conformance Electromagnetic environment -
hints HF-transmissions according to CISPR 11
Group 1 The IS 300 uses HF energy only for
its internal operation. Its HF trans‐
mission is therefore very low, and it
is improbable that neighbouring
electronic devices will be disturbed. HF-transmissions according to CISPR 11
Class B The IS 300 is for use in all facilities
including residential ones, and fa‐
cilities that are directly connected to
a public power supply that also sup‐
plies residential buildings. Transmissions of harmonics ac‐ cording to IEC 61000-3-2
Class A The IS 300 is for use in all facilities
including residential ones, and fa‐
cilities that are directly connected to
a public power supply that also sup‐
plies residential buildings. Transmissions of voltage fluctua‐ tions or flicker according to IEC 61000-3-3
fulfilled The IS 300 is for use in all facilities
including residential ones, and fa‐
cilities that are directly connected to
a public power supply that also sup‐
plies residential buildings.
Note
The device or system may not be used or stacked directly next to other devices. If it has to be used close to or stacked next to other devices, the device or system must be monitored to ensure that it is used properly in the existing arrangement.
Note
The immunity test levels required in IEC 60601 are met.
9.2 Guidelines and manufacturer's declaration - electromagnetic re‐ sistance to jamming
The IS 300 is for use in an environment like the one cited below. The IS 300 customer or user should ensure that use takes place in such an environment.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
9 Details on electromagnetic compatibility | 9.1 Guidelines and manufacturer's declaration - electromagnetic transmission
63/66
Page 66
Immunity tests IEC 60601 test level Conformance level Electromagnetic environ‐
ment - guidelines Electrostatic discharge (ESD) according to IEC 61000-4-2
± 6 kV contact discharge ± 8 kV atmospheric dis‐ charge
± 6 kV contact discharge ± 8 kV atmospheric dis‐ charge
Floors should be made of
wood or concrete or have
ceramic tiles. When the
floor is made of synthetic
material, the relative hu‐
midity must be at least
30%. Fast transient electrical disturbances/ Bursts ac‐ cording to IEC 61000-4-4
± 2 kV for power lines ± 1 kV for input and output lines
± 2 kV for power lines ± 1 kV for input and output lines
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment. Surges according to IEC 61000-4-5
± 1 kV Push-pull voltage (symmetrical) ± 2 kV common mode volt‐ age (unsymmetrical)
± 1 kV Push-pull voltage (symmetrical) ± 2 kV common mode volt‐ age (unsymmetrical)
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
Voltage interruptions, short-term interruptions and fluctuations of the supply voltage according to IEC 61000-4-11
< 5% U
T
for 1/2 period (>95% interruption) 40 % U
T
for 5 periods (60% interruption) 70 % U
T
for 25 periods (30 % interruption) < 5% U
T
for 5 s (>95% interruption)
< 5% U
T
for 1/2 period (>95% interruption) 40 % U
T
for 5 periods (60% interruption) 70 % U
T
for 25 periods (30 % interruption) < 5% U
T
for 5 s (>95% interruption)
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
When the user of the IS
300 needs continued
operation even when the
power supply is interrup‐
ted, it is recommended to
supply the IS 300 from an
uninterrupted power sup‐
ply or a battery.
Magnetic field with a sup‐ ply frequency (50/60 Hz) according to IEC 61000-4-8
3 A/m 3 A/m Magnetic fields at the
mains frequency should
correspond to typical val‐
ues in a business and
hospital environment.
NOTE: V T is the alternating mains voltage before the test level is used.
9.3 Guidelines and manufacturer's declaration - electromagnetic re‐ sistance to jamming
The IS 300 is for use in an environment like the one cited below. The IS 300 customer or user should ensure that use takes place in such an environment.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
9 Details on electromagnetic compatibility | 9.3 Guidelines and manufacturer's declaration - electromagnetic resistance to jamming
64/66
Page 67
Immunity tests IEC 60601 test level Conformance level Electromagnetic environment -
guidelines Conducted HF distur‐ bances according to IEC 61000-4-6 Radiated HF disturban‐ ces according to IEC 61000-4-3
3 V
eff
150 kHz to 80 MHz 10 V/m 80 MHz to 2.5 GHz
3 V
eff
3 V/m
Portable and mobile radio devices
should not be used closer to the IS
300 including the wires, than the re‐
commenced safe distance calcula‐
ted using the equation for the trans‐
mission frequency.
Recommended safe distance:
d = [3.5/3]P = 1.17
P
d= [3.5/3]P =1.17P for 80 MHz to
800 MHz
d= [7.0/3]P =2.33P for 800 MHz
to 2.5 GHz
with P as the maximum rated power
of the transmitter in Watts (W) ac‐
cording to the transmitter manufac‐
turer, and d as the recommended
safe distance in meters (m).
The field strength of stationary ra‐
dio transmitters should be less than
the conformance level at all fre‐
quencies in an on-site checka.
b
Disturbances are possible close to
devices that have the following
symbol.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not be applicable in every case. The spread of electromagnetic waves is absorbed and reflected by buildings, objects and people.
a
The field strength of stationary transmitters such as base stations of mobile tele‐ phones and land radio devices, amateur radio stations, AM and FM, radio and tel‐ evision broadcasters cannot be theoretically predetermined. To determine the elec‐ tromagnetic environment of stationary transmitters, a study of the location should be considered. When the measured field strength at the location on which the IS 300 is used, exceeds the conformity level, the IS 300 should be watched to ensure that it is functioning as per the correct usage. Should unusual performance features be observed, additional measures may be required, such as e.g. a different align‐ ment or another location for the IS 300.
b
Within the frequency range of 150 kHz to 80 MHz, the field strength should be less than 3V
eff
V/m.
9.4 Recommended safe distance between portable and mobile HF telecommunications equipment and the INTRAsurg 300
The IS 300 is for use in an environment like the one cited below. The customer or user of the IS 300 can help to avoid electromagnetic faults by keeping to the mini‐ mum distance apart between portable and mobile HF-telecommunication devices (transmitters) and the IS 300 - dependent on the output lines for the communication device - as given below.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
9 Details on electromagnetic compatibility | 9.4 Recommended safe distance between portable and mobile HF telecommunications equipment and the INTRAsurg 300
65/66
Page 68
Rated power of the trans‐ mitter in W
150 kHz to 80 MHz
d=1.17
P
80 MHz to 800 MHz d=1.17
P
800 MHz to 2.5 GHz d=2.33
P
0,01 0,12 0,12 0,23 0,1 0,37 0,37 0,74 1 1,17 1,17 2,33 10 3,70 3,70 7,37 100 11,70 11,70 23,30
For transmitters whose maximum rated power is not in the above table, the recom‐ mended safe distance d in meters (m) can be calculated using the equation for the respective gap, where P is the maximum rated power of the transmitter in Watts (W) according to the manufacturer's information.
NOTE 2: These guidelines may not be applicable in every case. The spread of electromagnetic waves is absorbed and reflected by buildings, objects and people.
Comment 1: To calculate the recommended safe distance from transmitters with a frequency range of 80 MHz to 2.5 GHz, an additional factor of 10/3 was used to reduce the probability that a mobile unintentionally brought into the patient area would cause malfunction.
Instructions for use INTRAsurg 300 / INTRAsurg 300 plus
9 Details on electromagnetic compatibility | 9.4 Recommended safe distance between portable and mobile HF telecommunications equipment and the INTRAsurg 300
66/66
Page 69
Page 70
1.001.3529 · Fk · 2010506_09 · en
Loading...