1 User instructions .............................................................................................................................................. 4
1.1 User guide................................................................................................................................................ 4
3 Assembly | 3.1 Assembly of the GENTLEpower LUX Angle Piece 25 LP
3 Assembly
3.1 Assembly of the GENTLEpower LUX Angle Piece 25 LP
3.1.1 Assembly instructions
Material numbers for spare parts are listed in the separate spare parts sheet.
See also:
2 Spare parts list for the GENTLEpower LUX Angle Piece 25 LP
Requirement
All parts must be cleaned thoroughly.
Glued joints must be clean; there must be no traces of grease or adhesive residue.
Check air and water hoses for free passage.
Note
Once assembled, the contra-angle handpiece has to be sprayed thoroughly with
KAVOspray.
3.1.2 Assembling the bearing cartridge
ToolName
Mounting device A (Mat. no. 1.002.3804)
Mounting device B (Mat. no. 1.002.3802)
▶ Apply a thin layer of grease to the O-ring ③, insert into the groove on mounting
device A, pull back the slide on mounting device A and hold in place.
▶ Insert mounting device A into the drilled hole on the bearing sleeve ④ and slide
the O-ring ③ into the first groove on the bearing sleeve.
▶ Apply a thin layer of grease to the O-ring ②, insert into the groove on mounting
device B, pull back the slide on mounting device B and hold in place.
▶ Insert mounting device B into the drilled hole on the bearing sleeve ④ and slide
the O-ring ② into the second groove on the bearing sleeve.
▶ Push the intermediate drive ① into the bearing sleeve ④ as far as the stop.
3 Assembly | 3.1 Assembly of the GENTLEpower LUX Angle Piece 25 LP
▶ Place the contra-angle handpiece in a hot-air furnace and allow the adhesive 30
minutes
to set at 120 oC.
▶ Remove the head casing ⑤ from the sleeve ⑧ again.
▶ Use Loctite 306 adhesive to fix the inside of the spray sleeve ⑧ to position ⑥ and
the fibre-optic bundle ④ making sure that the fibre-optic bundle is situated such as
to be flush at the front. Remove any residual adhesive from the end of the fibreoptic bundle.
▶ Place the angle piece in a hot-air furnace and allow the adhesive to set for 30 mi‐
nutes at
120 oC.
3.1.6 Assembling the head
ToolDescription
Wrench (Mat. no. 1.001.3288)
Universal bit holder (Mat. no. 1.002.4577
Torque handle (Mat. no. 1.002.4578) with
100 Ncm
Circlip pliers (Mat. no. 0.411.2421)
▶ Fit one O-ring ② in the groove on the spray insert ③ and one in the attachment of
the spray insert ③.
▶ Insert the spring washer ④ and drive axle ⑤ into the spray insert ③.
▶ Insert the spray insert ③ into the head casing ①.
▶ Screw the cover ⑥ into the head casing ① and tighten with a torque of 100 Ncm
with the wrench, the universal bit holder and the torque handle.
Carry out a general visual inspection
of external condition.
There should be no visual evidence
of damage.
There should be no spots on the fi‐
bre-optic bundle.
Check instrumentation connection▪ Motor
Connect the angle piece to the mo‐
tor coupling and pull off. The con‐
nection should be made, lock and
disconnect with ease.
Check chuck▪ Test drill
Insert the test drill into the chuck
and pull out.
There are no obstructions to inser‐
tion and removal.
▪ Retention force gauge
Use a retention force gauge to
measure the axial retention force of
the chuck.
Axial retention force: at least 17 N
Check pushbutton▪ Pushbutton tester
Insert the pushbutton tester into the
chuck system and push it as far as
the stop.
The pushbutton must remain de‐
pressed.
Check running characteristics▪ Motor
Run the angle piece under test con‐
ditions CW and CCW with radial and
axial load.
There should be no noticeable vi‐
brations or running noise.
Under axial load, the adapter does
not graze the pushbutton and the
pushbutton does not heat up.
Run the angle piece with test drill
off-load at 200,000 rpm.
There should be no noticeable vi‐
brations or running noise.
Check power consumption▪ Test drill
Measure the power consumption
under test conditions with head and
test drill.
Power consumption: max. 150 mA
Check seals▪ Leakage air hose
Pressurize the media chamber with
compressed air
(2 bar) and pressure-test the spray
drilled hole for 15 seconds.
The compressed air must remain
constant.
Run a spray test▪ Spray test drill
Carry out a visual inspection of the
spray mist.
The spray jets should hit the ball of
the spray test drill.
▪ Test device
Measure air and water quantity un‐
der test conditions.
Air quantity: at least 2.5 Nl/min.
Water quantity: at least 50 ml/min.
18 / 20