JVC KY-F75U User Manual

For Model KY-F75U
Supplement for medical Use
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the Accompany­ing documents.
Note:Please refer to the “In case of use as a medical equipment”(page E17) in the Instructions.
portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
WARNING:The use of ACCESSORIES, transducers and cables other than those specified,with the exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement parts for inter­nal components ,may result in increase EMISSIONS or decreased IMMUNITY of the EQUIPMENT or SYSTEM.
Information of Cables and Accessories
Make sure to use specified Accessories as described below, in order to complied with EMC directive
93/42/EEC.
Accessory name
Power Cable
Clamp filter Earth Cable
1394 Cable
Note:Please refer to a part of “SAFETY PRECAUTIONS” and “Accessories and attachments”(page E3 and E5) in the Instructions besides above information.
WARNING:The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary,the EQUIPMENT or SYSTEM should be observed to verify normal operation in the configuration in which it will be used.
WARNING: The use of ACCESSORY, transducer and cable with EQUIPMENT and SYSTEM other than those specified may result in increase EMISSIONS or decreased IMMUNITY of the EQUIPMENT or SYSTEM.
Exclusive Cable,SCV2839-201
ZCAT2032-0930
Single wire,SCV3063-001
1394 N1 2.9M
Type
length
2 meters
-
2 meters
2.9 meters
Manufacturer,etc
For DC IN terminal
TDK
,for IEEE1394 cable,1 pcs and for Power cable, 1 pcs.
Nitto Densen Industry Co.,Ltd.AWG20
Oki Electric Cable Co.,Ltd.
Guidance or Manufacturer’s declaration-electromagnetic emissions
The Model KY-F75U is intended for use in the electromagnetic environment specified below. The customer or the user of the Model KY-F75U should assure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic environment-guidance
RF emissions 
CISPR11/EN55011 
RF emissions 
CISPR11/EN55011
Harmonic emissions 
IEC/EN 61000-3-2 
Voltage Fluctuations/
Flicker emissions 
IEC/EN 61000-3-3 
Group 1 
Class B 
Class A 
Complies 
The Model KY-F75U uses RF energy only for its internal func- tion. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The Model KY-F75U is suitable for use in all establishments,including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Guidance or Manufacturers declaration-electromagnetic immunity
The Model KY-F75U is intended for use in the electromagnetic environment specified below.
The customer or the user of the Model KY-F75U should assure that it is used in such an environment.
Immunity Test 
Electrostatic dis-
charges(ESD) 
IEC/EN 61000-4-2
Electrical fast
Transients/burst 
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines 
IEC/EN 61000-4-11
IEC/EN 60601
Test Level
+/- 6 kV:Contact
+/- 8 kV:Air 
+/- 2 kV:AC mains
+/- 1 kV:Signal lines 
+/- 2 kV :Common mode
+/- 1 kV :Differential mode 
<5 % 230 V (>95 % dip in 230 V) : for 0.5 cycle
40 % 230 V (60 % dip in 230 V) : for 5 cycles
70 % 230 V (30 % dip in 230 V) : for 25 cycles
Compliance Level
+/- 6 kV:Contact
+/- 8 kV:Air 
+/- 2 kV:AC mains
+/- 1 kV:Signal lines 
+/- 2 kV :Common mode
+/- 1 kV :Differential mode 
<5 % 230 V (>95 % dip in 230 V) : for 0.5 cycle
40 % 230 V (60 % dip in 230 V) : for 5 cycles
70 % 230 V (30 % dip in 230 V) : for 25 cycles
Electromagnetic
environment-guidance
Floor should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital enviroment. If the user of the Model KY-F75U requires continued op­eration during power mains inter­ruptions, it is recommended that the Model KY-F75U be powered from an uninterruptible power sup­ply or a battery.
Power frequency
(50 Hz/60 Hz)
Magnetic Field
IEC/EN 61000-4-8
<5 %230 V (>95 % dip in 230 V) : for 5 s 
3 A/m
<5 %230V (>95 % dip in 230 V) : for 5 s 
3 A/m
If image distortion occurs, it may be necessary to position the Model KY-F75U further from sources of power frequency magnetic fields or to install magnetic shielding. The power frequency magnetic field should be measured in the in­tended installation location to as­sure that it is sufficiently low.