JUMO tecLine ClO2, tecLine O3 Operating Manual

JUMO tecLine ClO2 / JUMO tecLine O3
Sensors for chlorine dioxide and ozone
Type 202634/45, type 202634/46
type 202634/50, type 202634/51
B 202634.0
Operating manual
2012-07-13/00585740
Screws coated with threadlocking paint must not be adjusted! The threadlocking paint must not be damaged. If it is, claims under the warranty will not be honored!
Contents
1 Typographical conventions ...................................................... 5
1.1 Warning signs ..............................................................................................5
1.2 Note signs ....................................................................................................5
2 General ....................................................................................... 6
2.1 Preface .........................................................................................................6
2.2 Scope of application ....................................................................................6
2.3 Output signal ................................................................................................6
3 Device identification ................................................................. 7
3.1 Nameplate ....................................................................................................7
3.2 Order details .................................................................................................7
4 Description ................................................................................ 8
4.1 Scope of delivery .........................................................................................8
4.2 Sensor for chlorine dioxide - type 202634/45 ..............................................8
4.3 Sensor for chlorine dioxide - type 202634/46
(resistant to chemicals and surfactants) ......................................................8
4.4 Sensor for ozone - type 202634/50 .............................................................8
4.5 Sensor for ozone - type 202634/51
(resistant to chemicals and surfactants) ......................................................9
4.6 Details ..........................................................................................................9
4.7 Technical data ............................................................................................11
5 Mounting .................................................................................. 13
5.1 Notes ..........................................................................................................13
5.2 Flow-through fitting (part no. 00392611) ...................................................14
5.3 Flow monitoring .........................................................................................16
6 Electrical connection .............................................................. 17
6.1 Sensor for chlorine dioxide or ozone .........................................................17
6.2 Flow monitor ..............................................................................................18
6.3 Example of a measurement section ...........................................................19
7 Startup ..................................................................................... 21
7.1 Screwing the membrane cap off and on ....................................................21
7.2 Topping up reference electrolyte ...............................................................21
Contents
8 Maintenance ............................................................................ 27
8.1 Cleaning the electrode finger tip / replacing the membrane cap and
electrolyte ...................................................................................................27
8.2 Remove scale that has accumulated on the membrane cap ....................28
8.3 Storage / transport .....................................................................................28
8.4 Consumables / accessories .......................................................................29
9 Calibration ............................................................................... 30
9.1 General .......................................................................................................30
9.2 Calibration with a indicator/control instrument ..........................................30
10 Eliminating errors and faults .................................................. 32

1.1 Warning signs

Caution
This sign is used when there may be danger to personnel if the instructions are ignored or not followed correctly!
Attention
This sign is used when there may be damage to equipment or data if the instructions are ignored or not followed correctly!

1.2 Note signs

Note This sign is used to draw your attention to something of special importance.

1 Typographical conventions

1
abc
Action instruction
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes consist of two parts:
Identification marking in the text and the footnote text. The identification markings in the text are arranged as sequential superscript
numbers.
This sign indicates that an action to be performed is described. The individual steps are marked by this asterisk. Example: Press key.
5

2 General

2.1 Preface

Please read this operating manual before placing the device in operation. Keep the operating manual in a place which is accessible to all users at all times.
Danger:
Sudden failure of the sensor may result in dangerous overdosing of chlorine dioxide or ozone. Suitable preventive measures must be in place as needed to prevent this from happening. Check the plant for odor of chlorine dioxide or ozone and check water for any unusual discoloration.

2.2 Scope of application

These membrane-covered, amperometric sensors are used to measure the concentration of chlorine dioxide and ozone (for example in drinking and swimming water, industrial, process and cooling water). The sensor for chlorine dioxide can be used to measure chlorine dioxide from chlorite / chlorine and chlorite / hydrochloric acid plants. The sensor for ozone can be used for example to measure electrolytically generated ozone. The sensors are not suitable for detecting the absence of chlorine dioxide and ozone. The integrated electronics of the sensors provides a temperature­compensated current signal of 4 to 20mA. A downstream device (indicator, controller, recorder, PLC, etc.) is used for calibration.

2.3 Output signal

Since the measurement signal of amperometric sensors is temperature­dependent, an integrated NTC resistor is provided for automatic temperature compensation. The recommended temperature ranges are as follow
- For sensors of types 202634/45 and 202634/50): +5 to +45 °C.
- For sensors of types 202634/46 and 202634/51): +5 to +50 °C. The integrated electronics of the sensors provides a current signal of 4 to
20 mA. A downstream device (indicator, controller, recorder, PLC, etc.) is used for calibration. The sensors can be connected directly to a suitable indicator and controller. The JUMO AQUIS 500 AS indicator/controller (type sheet 202568) and the JUMO dTRANS AS 02 indicator/controller (type sheet 202553) are especially suitable for combining with these sensors. They provide the voltage required for the power supply of the sensor and make for an easy way to calibrate the measuring system.
6

3.1 Nameplate

Sensor for chlorine dioxide Type: 202634/45-20 Measuring range: 0.00 to 2.00 mg/l F No. 00000000 00 0 1226 0005 Serial No.: 01 01 0002
JUMO tecLine ClO2
JUMO GmbH & Co. KG
Fulda, Germany www.jumo.net
Note:
The date of manufacture is encoded in the "F No." (manufacturing number): 1226 signifies: year of manufacture 2012 / calendar week 26.

3.2 Order details

(1) Basic type
202634 Sensor

3 Device identification

(2) Basic type extension
45 for chlorine dioxide 46 for chlorine dioxide (resistant to chemicals and surfactants) 50 for ozone 51 for ozone (resistant to chemicals and surfactants)
(3) Measuring range
o o 10 0.00 to 0.50 mg/l (ppm) oooo 20 0.00 to 2.00 mg/l (ppm)
25 0.00 to 5.00 mg/l (ppm) 35 0.00 to 10.00 mg/l (ppm)
o o 37 0.00 to 20.00 mg/l (ppm)
(1) (2) (3)
Order code /- Order example 202634 / 45 - 20
7

4 Description

4.1 Scope of delivery

Two-wire sensor with membrane cap, electrolyte and special abrasive paper for cleaning the electrode finger.

4.2 Sensor for chlorine dioxide - type 202634/45

This sensor features a membrane-covered, amperometric 2-electrode measuring system.
The membrane of this sensor is made of hydrophobic, microporous PTFE material. These sensors should only be used in water with drinking or swimming pool quality. The water should not under any circumstances contain surfactants (many cleaning agents and disinfectants contain surfactants). Surfactants will eliminate the hydrophobic properties of the membrane. Then proper functioning can no longer be guaranteed.
The membrane cap that seals off the electrode space is filled with an aqueous electrolyte containing alkali halides.
The working electrode (cathode) is made of gold (Au). The anode, which performs the function of a combined reference electrode and counter electrode, is made of silver (Ag) fitted with a coating of silver halide (AgHal).
The sensor for chlorine dioxide can be used to measure chlorine dioxide from chlorite/chlorine and chlorite/hydrochloric acid plants.

4.3 Sensor for chlorine dioxide - type 202634/46 (resistant to chemicals and surfactants)

The sensor measures chlorine dioxide from chlorite/chlorine plants and chlorite/hydrochloric acid plants.

4.4 Sensor for ozone - type 202634/50

This sensor features a membrane-covered, amperometric 2-electrode measuring system.
The membrane of these sensors is made of hydrophobic, microporous PTFE material. This sensor should only be used in water with drinking or swimming pool quality. The water should not under any circumstances contain surfactants (many cleaning agents and disinfectants contain surfactants). Surfactants will eliminate the hydrophobic properties of the membrane. Then proper functioning can no longer be guaranteed.
The membrane cap that seals off the electrode space is filled with an aqueous electrolyte containing alkali chlorides.
The working electrode (cathode) is made of gold (Au). The anode, which performs the function of a combined reference electrode and counter electrode, is made of silver (Ag) fitted with a coating of silver halide (AgHal).
The sensor for ozone can be used to measure electrolytically generated ozone.
8
4.5 Sensor for ozone - type 202634/51
(1)
(2)
(3)
(4)
(5)
(6) (7)
(8) (9)
(10)
(11)
(12)
(resistant to chemicals and surfactants)
This sensor measures ozone, for example electrolytically generated ozone
). This sturdy sensor can be used in all water qualities.
(O
3

4.6 Details

4.6.1 Types 202634/45 and 202634/50

4 Description
(1) Pg screw connection (2) Cover (3) O-ring (4) 2-pin terminal for measuring cable connection (5) Electrode shaft with integrated electronics (6) O-ring (7) Electrode finger (reference electrode) (8) Measurement electrode (9) Transparent cover (10) Membrane cap (11) PTFE membrane (12) Valve opening
9
4 Description
(1)
(2)
(3)
(4)
(5)
(7)
(8) (9)
(10)
(11)
(13)
(14)
(12)
(6)
Attention:
- This measurement is only possible in a suitable flow-through fitting (part no. 00392611).
- The process medium must be similar in quality to drinking or swimming pool water.
- The sample medium must not contain any surfactants (surface-active substances, for example those in cleaning agents, detergents and disinfectants).
- The process medium must not contain any hydrophobic substances (for example oils and greases).

4.6.2 Types 202634/46 and 202634/51

10
(1) Pg screw connection (2) Cover (3) O-ring (4) 2-pin terminal for measuring cable connection (5) Electrode shaft with integrated electronics (6) Counter electrode (stainless steel) (7) O-ring (8) Electrode finger (reference electrode) (9) Measurement electrode (10) Transparent cover (11) Membrane cap (12) Membrane holder (stainless steel) (13) Membrane (14) Valve opening

4.7 Technical data

UB – 7.5 V
0.02A

4.7.1 Types 202634/45 and 202634/50

4 Description
Analyte Chlorine dioxide (ClO2)Ozone (O Membrane type Hydrophobic PTFE membrane
Measuring cable connection 2-pin terminal, polyamide PG7 screw connection; conductor cross section 2x 0.25 mm Voltage supply U Electromagnetic
compatibility
Output signal 4to20mA Burden
Settling time 1h 2h Incident flow velocity approx. 15cm / s
Measuring ranges
(other ranges on request)
Resolution 0.001mg / l, for measurement range 0 to 0.5mg/l;
Response time t Operating temperatures /
temperature compensation Zero point adjustment Not required pH value operating range 1to11pH 2to11pH pH dependency
(loss of slope) Disruptive substances /
cross sensitivities Pressure resistance p
Material Shaft, cover, cap: PVC Shaft, cover, cap: PVC Dimensions Diameter: 25 mm, length: 220 mm Weight approx 125 g Maintenance
90
Check the measurement signal:regularly, at least once a week
0to2.0/0to5/0to10mg/l (ppm)
Type 202634/45 Type 202634/50
12 to 30V DC (electrical isolation recommended)
B
Interference immunity: To industrial requirements
If the sensor is installed in a JUMO flow-through fitting 00392611,
this is equivalent to a flow rate of about 30l / h.
approx. 15 s approx. 15 s
+5 to +45 °C +5 to +45 °C
Chlorine is disruptive
Ozone is disruptive
No pressure fluctuations are admissible when operating under pressure.
0.01mg / l, for measurement range 0 to 2.0mg / l
We recommend unpressurized operation (atmospheric pressure).
According to EN 61326-1
Interference emission: Class B
No pH dependency
max. 2 bar
abs
p
max. 1 bar
rel
0to0.5/0to2.0/0to5/ 0to10
mg / l (ppm)
Chlorine is disruptive
Chlorine dioxide is disruptive
)
3
2
, cable diameter approx. 4 mm
Replace the membrane cap:once a year (subject to water quality) Change the electrolyte: every 3 to 6 months
Storage
Sensor: frost-free, dry and without electrolyte, can be stored for an unlimited time at
+5 to +45 °C Membrane cap: Used membrane caps cannot be stored! Electrolyte: In the original bottle and protected against sunlight at +5 to +25 °C
11
4 Description
UB – 7.5 V
0.02A

4.7.2 Types 202634/46 and 202634/51

Analyte Chlorine dioxide (ClO2)Ozone (O Membrane type Membrane resistant to chemicals and surfactants
Measuring cable connection 2-pin terminal, polyamide PG7 screw connection; conductor cross section 2x 0.25 mm Voltage supply U Electromagnetic
compatibility
Output signal 4to20mA Burden
Settling time 1h 1h Incident flow velocity approx. 15cm / s
Measuring ranges
(other ranges on request)
Resolution 0.001mg / l, for measurement range 0 to 0.5mg/l;
Response time t Operating temperatures /
temperature compensation Zero point adjustment Not required pH value operating range 1to11pH 2to11pH pH dependency
(loss of slope) Disruptive substances /
cross sensitivities Pressure resistance p
Material Shaft, cover, cap: PVC
Dimensions Diameter: 25 mm, length: 220 mm Weight approx 125 g Maintenance
90
Check the measurement signal: regularly, at least once a week
Type 202634/46 Type 202634/51
12 to 30V DC (electrical isolation recommended)
B
Interference immunity: To industrial requirements
If the sensor is installed in a JUMO flow-through fitting 00392611,
this is equivalent to a flow rate of about 30l / h.
approx. 1.5 min approx. 50 s
Chlorine is not disruptive
Ozone is disruptive
No pressure fluctuations are admissible when operating under pressure.
0.01mg / l, for measurement range 0 to 2.0mg / l
We recommend unpressurized operation (atmospheric pressure).
According to EN 61326-1
Interference emission: Class B
0 to 2.0mg / l (ppm)
+5 to +50 °C
No pH dependency
max. 2 bar
abs
max. 1 bar
p
rel
Membrane disk holder:
Stainless steel
Chlorine is disruptive
Chlorine dioxide is disruptive
)
3
2
, cable diameter approx. 4 mm
Replace the membrane cap: once a year (subject to water quality) Change the electrolyte: every 3 to 6 months
Storage
Sensor: Frost-free, dry and without electrolyte, can be stored for an unlimited time at
+5 to +45 °C Membrane cap: Used membrane caps cannot be stored! Electrolyte: In the original bottle and protected against sunlight at +5 to +25 °C
12

5.1 Notes

5 Mounting

Attention:
- The measurement is only possible with a suitable flow-through fitting (for example a flow-through fitting for sensors according to type sheets 202630, 202631, 202634, 202636, part no. 00392611, see type sheet 202634).
- The sensor should be operated unpressurized if possible with free discharge of the process medium. If that is not possible, the sensor can also be operated at a constant pressure of up to 1 bar (p Pressure fluctuations must be prevented!
- Air bubbles upstream from the membrane can distort the measurement signal.
- Both the transmitter and the sensor connected to it must remain continuously in operation even if the system is only operated at intervals. The supply voltage of the connected sensor must not be turned off!
- The sensor must not be left dry.
) or 2 bar (p
rel
abs
).
13
5 Mounting
(4)
(3)
(5)
(1)
(2)

5.2 Flow-through fitting (part no. 00392611)

5.2.1 Mounting the fitting

The flow-through fitting can be fastened to a wall with an optional support bracket (part no. 00455706) or fastened to an installation table.
14
(1) Sensor (2) Support bracket (optional) (3) Connection G1/4 for hose ø 8mm x 6mm (4) Fitting (5) Removable measuring vessel (inspection glass)

5.2.2 Sensor installation

(1)
(2)
(5)
(6)
(4)
(8)
(10)
(9)
(7)
(3)
Overview
5 Mounting
Installation
(1) Sensor (6) Inspection glass (2) Union nut (7) Inlet G1/4A or DN10 (3) Flow-through fitting (8) Stepped collar 1" (4) Outlet G1/4A or DN10 (9) Compression ring (5) O-ring (10) O-ring
1
Part of the fitting.
Caution: When assembling and installing the sensor (1) make certain the O-
1
1
1
rings and threads are clean and fully operational! The transparent inspection glass (6) can be unscrewed from the fitting housing
for maintenance purposes.
1. First push the O-ring (10) onto the sensor (1), then the compression ring (9) and over it the 1" stepped collar (8) (starting from the Pg screw connection). The stepped collar (8) must engage in the groove.
15
5 Mounting
(3)
(1)
(2)
2. After the sensor has been prepared in this manner, insert it into the flow­through fitting (3) and fasten it in place with the union nut (2).

5.3 Flow monitoring

5.3.1 Minimum inflow

For proper operation, the incident flow of the process medium on the sensor must be at least 15 cm/s. This corresponds to a minimum flow rate of 30 l/h in the flow-through fitting.
Below this minimum incident flow velocity, the sensor will indicate values that are too low. This could result in dangerous overdosing in a connected control system.
Above this minimum incident flow velocity the measurement signal is only slightly affected by the incident flow velocity.
The monitoring system consists of a flow fitting (2) part no.: 00396470 and the flow monitor (1) (part no.: 00396471), See section 5.3.2 "Mounting", page 16.
The instrument can be used to monitor the minimum incident flow velocity of 15 cm/s.

5.3.2 Mounting

16
The O-ring (3) must be resting in the groove of the flow monitor (1).
1. Screw the flow monitor (1) into the fitting (2).

6 Electrical connection

(1)
(2)
(3)
(4)

6.1 Sensor for chlorine dioxide or ozone

6.1.1 General

- Cable diameter approx. 4 mm
- Wire cross section 2 x 0.25mm
- Lay signal lines isolated from cables with voltages of > 60 V
- Use a shielded cable with twisted wires
- Keep clear of large electrical installations

6.1.2 Terminal assignment

Connection Screw terminals
Voltage supply DC 12 to 30 V
2
1 L+
2 L-
Output 4 to 20 mA two wires
Impressed current 4 to 20 mA in voltage supply

6.1.3 Connection

1 L +
2 L-
1. Push the cover (2) over the connecting cable.
2. Connect the wires to the terminals (4) according to the terminal assignment.
3. Screw on the cover (2) by hand until the O-ring (3) seals.
4. Tighten the Pg screw connection (1).
Attention:
When disconnecting, undo the Pg screw connection (1) first!
17
6 Electrical connection
4
1
3 -
NC
C
br
ws
NO
gn signal
1
4
3
2
Stecker 4polig

6.2 Flow monitor

6.2.1 Function

The contact (3+4) is opened starting at a flow velocity of 15 cm/s.
18
6 Electrical connection
4
1
3
+
-
1
4
3
2
(A)
(B)
(3)
(4)
(1)

6.3 Example of a measurement section

6.3.1 Sensor

The electronics integrated into the shaft of the sensor (2) returns an uncalibrated 4 to 20 mA signal.

6.3.2 Indicator/controller

The signal can be further processed with the JUMO AQUIS 500 AS (1) or the JUMO dTRANS AS 02. The device provides the required supply voltage and makes for an easy way to calibrate the measuring system. The sensor can also be connected to other indicator, controller, recorder or PLC systems, however, provided the power supply and calibration allow it.

6.3.3 Connection

Danger:
Before making the connection, always read the JUMO AQUIS 500 AS (B 202568) operating manual!
(1) JUMO AQUIS 500 AS, type 202568/20-888-888-888-310-310-23/000,
see type sheet and operating manual 202568 (3) Sensor for chlorine dioxide or ozone, see data sheet 202634 (4) Optional flow monitor part no.: 00396471,
see type sheet 202634 (A) Row 1 (B) Row 2
19
6 Electrical connection
(1)
(2)
(3)

6.3.4 Flow monitor

If the flow monitor (3) indicates too small a flow from the sensor (2), the binary input of the JUMO AQUIS 500 AS (1) switches, the instrument goes to "Hold" and an alarm is generated.
(1) JUMO AQUIS 500 AS, type 202568/20-888-888-888-310-310-23/000,
see type sheet and operating manual 202568 (2) Sensor for chlorine dioxide or ozone, see data sheet 202634 (3) Flow monitor part no.: 00396471, see type sheet 202634
20

7.1 Screwing the membrane cap off and on

(1)
(3)
(2)
(4)
(5) (6) (7)
(8)
Caution: The electrolyte may spray out of the valve opening (5) when the membrane cap is screwed on! Wear safety goggles!
Wash off sprayed liquid (aqueous solution of an alkali halide) under flowing water.
Attention:Before screwing off the membrane cap, move the transparent cover(s) (6) to the side so that the valve opening (5) is exposed! Otherwise the membrane will be destroyed. Do not touch the electrode finger (3/4) or get it dirty!
The steps must be performed exactly as they are described below! The membrane of the sensor is very sensitive. Make certain when screwing the
membrane cap on and off that the membrane is not destroyed by vacuum or excess pressure.
The service life of the electrolyte is about 3 to 6 months. Types 202634/45 and /50 have one transparent cover (6). Types 202634/46 and /51 have two transparent cover (6).

7 Startup

7.2 Topping up reference electrolyte

7.2.1 For types 202634/45 and /50

21
7 Startup
1. Use a small screwdriver or similar tool to raise the transparent cover (6) of the valve opening (5) close to the valve opening and push it somewhat to the side. The valve opening (5) is now open.
2. Screw the membrane cap (7) off the shaft (1) of the sensor.
3. Push the transparent cover (6) back in place until the cover is resting in the slot again and the valve opening closes.
4. Discard the used electrolyte, clean the membrane cap with clean water and then dry it.
5. Fill the membrane cap (7) up to the brim with enclosed electrolyte with no bubbles.
6. Holding the shaft (1) of the sensor vertically, tightly screw it (manually) onto the filled membrane cap (7), making certain as you do so that no excess electrolyte can escape from the valve opening (5).
22
7 Startup
Attention:
Always wait before screwing the membrane cap onto the sensor until just before it is installed in a fitting! The sensor is very sensitive when it is completely assembled!
When it is in a condition ready for measuring (membrane cap screwed on completely), the sensor must not be pushed open with the membrane (8). This will damage it and make it unusable!
Screw on the membrane cap (7) completely (as far as it will go)! The sealing O­ring (2) offers the first resistance against screwing in the membrane cap. Continue screwing in the membrane cap (7) until it is touching the shaft (1) of the sensor!
23
7 Startup
(1)
(3)
(2)
(4)
(8)
(7)
(6)
(5)

7.2.2 For types 202634/46 and /51

The membrane cap comes in a can filled with water. This ensures that a layer of air
cannot form on the inside of membrane.
1. Open the can with the membrane cap, drain the water and remove the membrane cap. There must only be one transparent cover (6) in the groove, closing the valve opening (see arrow). The second transparent cover must be visible as shown in the picture.
2. Fill the membrane cap (7) up to the brim with enclosed electrolyte.
If there are still air bubbles trapped in the membrane area (only visible against backlight), please apply the tapping procedure described below. This can be used to remove any air bubbles that may be trapped inside the membrane, which would interfere with the measuring function:
24
2a.Tap with the electrode shaft on the filled membrane cap (7) until no more air
bubbles form and rise. This displaces a layer of air that may have formed on the inside of the membrane. Then
add electrolyte. if necessary
7 Startup
3. Remove the black protective tube from the electrode finger (3). The enclosed special abrasive paper is only used to clean the tip (4) of the dry electrode finger (=measurement electrode). To do this hold the soft support with the special abrasive paper and move the electrode tip (4) of the sensor over the abrasive paper with the sensor at a slight angle. Then turn the sensor slightly on its axis and move across the abrasive paper again. Repeat this process several times.
4. Hold the electrode shaft vertically and place it on the filled membrane cap. If necessary turn it counter-clockwise first until the thread engages, then slowly screw the electrode shaft into the membrane cap. Excess electrolyte will escape from the valve opening (5) during this process not keep the valve opening (arrow) closed! Make certain the membrane cap electrode shaft
Attention:
The electrolyte may spray out of the valve opening (5). Wash off spilled electrolyte or electrolyte that has gotten on the skin or in the eyes with water. The electrolyte may contain diluted acids. Observe the warnings
! Flush off the electrolyte adhering to the outside with water.
on the electrolyte bottle!
is screwed (manually) tightly against the
. Do
25
7 Startup
Attention:
Always wait before screwing the membrane cap onto the sensor until just before it is installed in a fitting! The sensor is very sensitive when it is completely assembled!
When it is in a condition ready for measuring (membrane cap screwed on completely), the sensor must not be pushed open with the membrane (8). This will damage it and make it unusable!
Screw on the membrane cap (7) completely (as far as it will go)! The sealing O­ring (2) offers the first resistance against screwing in the membrane cap. Continue screwing in the membrane cap (7) until it is touching the shaft (1) of the sensor!

7.2.3 Settling time

The sensor will not provide a constant value until after the settling time of 2 hours has passed. It cannot be calibrated until then. The calibration procedure should be repeated on the day after initial commissioning.
26

8 Maintenance

(1)
(3)
(2)
(4)
(8)
(7)
(6)
(5)

8.1 Cleaning the electrode finger tip / replacing the membrane cap and electrolyte

Attention:
Do not rub the brown coating on the combined counter electrode and reference electrode (3) with special abrasive paper!
Do not touch the electrode finger (3/4) or get it dirty! The steps must be performed exactly as they are described below!
1. Unscrew the membrane cap from the shaft of the sensor, See section 7.1 "Screwing the membrane cap off and on", page 21.
2. Clean the electrode finger (3) and membrane cap (7) with distilled water and dry See section 7.2 "Topping up reference electrolyte", page 21.
3. Using the special abrasive paper, clean only the tip of the dried electrode finger (4). To do this place the special abrasive paper on a cleaning tissue with the matte side facing up, hold one corner firmly in place and while holding the sensor vertically, move the electrode tip two or three times over the special abrasive paper.
4. If necessary use a new membrane cap.
5. Fill the membrane cap (7) up to the brim with enclosed electrolyte with no bubbles.
27
8 Maintenance
6. Screw the membrane cap onto the sensor shaft, See section 7.1 "Screwing the membrane cap off and on", page 21.
Note:
The service life of the electrolyte is 3 to 6 months. The service life of the membrane depends very much on the water quality.
Avoid heavy contamination of the membrane! If the sensor still indicates values that are too low after the electrode finger tip
has been cleaned, a new membrane cap must be used.

8.2 Remove scale that has accumulated on the membrane cap

1. Unscrew the membrane cap from the shaft of the sensor, See section 7.1 "Screwing the membrane cap off and on", page 21.
2. Flush away the electrolyte.
3. Place the membrane cap (7) in a 1% solution of hydrochloric acid for a few hours.
4. Before startup rinse off with clean water.
5. Screw the membrane cap filled with electrolyte onto the shaft of the sensor.

8.3 Storage / transport

Attention:
Used membrane caps that have been in operation for longer than 1 day cannot be stored and reused!
Preparation
1. Unscrew the membrane cap from the shaft of the sensor, See section 7.1 "Screwing the membrane cap off and on", page 21.
2. Flush away the electrolyte.
3. Rinse off the membrane cap and electrode finger (3) with clean water and dry so no dust is left.
4. Screw the dry membrane cap (7) loosely onto the electrode shaft (1). The membrane (8) must not be resting against the electrode tip (4)!
Placing in operation again
1. Clean the electrode tip (4) with the enclosed special abrasive paper See "Cleaning the electrode finger tip / replacing the membrane cap and electrolyte", page 27.
2. Use a new membrane cap (7)!
28
8 Maintenance

8.4 Consumables / accessories

Designation Part no.
Flow-through fitting for sensors according to type sheets 202630, 202631, 202634, 202636
Support bracket for flow-through fitting 00455706 Flow monitor 00396471 Fitting for flow monitor 00396470 Special electrolyte for 202634/45 00392332 Special electrolyte for 202634/46 00441316 Special electrolyte for 202634/50 00392333 Special electrolyte for 202634/51 00441311 Spare parts set for 202634/45 and 202634/50 (1x membrane cap, fine abrasive
paper) Spare parts set for 202634/46 (1x membrane cap, fine abrasive paper) 00409344 Spare parts set for 202634/51 (1x membrane cap, fine abrasive paper) 00441309 Matching indicator/controller: JUMO AQUIS 500 AS,
Type: 202568/20-888-888-888-310-310-23/000 (for other versions see type and price sheet 202568)
Matching transmitter / controller: JUMO dTRANS AS 02, Type: 202553/01-8-01-4-0-00-23/000 (for other versions see type and price sheet 202553)
00392611
00392331
00528718
00550842
29

9 Calibration

9.1 General

Depending on requirements, the sensor should be checked and calibrated on a regular basis at set time intervals. Recommendation: weekly or more frequently depending on accuracy requirements.
Note:
The process for verification and calibration is as follows:
1. Take a sample of the medium in the immediate vicinity of the sensor.
2. Determine the concentration of the substance with a reference method.
3. Use the reference value that was determined to adjust the slope of the indicator.

9.2 Calibration with a indicator/control instrument

Reference method
Note:
Suitable reference methods for calibrating can be found for example in standards
- Chlorine dioxide: DIN 38408-5
Initial condition
Procedure
- Ozone: DIN 38408-3
The DPD method of photometric determination is frequently used (DPD = N,N­diethyl-1,4-phenylenediamine). Suitable testing systems are commercially available. Providers include VWR International (previously Merck) (Spectroquant), Macherey-Nagel (Nanocolor) etc.
- The display format and measurement range have been set; see the operating manual of the indicator/control instrument in use.
- The sensor is installed in a suitable flow-through fitting, See section 5 "Mounting", page 13.
- The settling time of the sensor (2 hours) has elapsed and the measured value is stable.
1. Take a water sample at the flow-through fitting outlet (or in the immediate vicinity of it).
2. Determine the analyte concentration (chlorine dioxide or ozone) of the sample with a suitable reference method.
3. Adjust the indicator to the reference value; see the operating manual of the transmitter/controller you are using.
30
Check the saved slope
Many transmitters/controllers (for example JUMO AQUIS 500 AS) have a "calibration log". Relevant data from each calibration is saved in the log.
Attention:
If the value for the slope falls below 40%, the membrane cap and electrolyte must be replaced and the electrode tip must be cleaned, See section 8.1 "Cleaning the electrode finger tip / replacing the membrane cap and electrolyte", page 27.
Setting the slope manually
See the operating manual for the transmitter/controller you are using.
Zero-point adjustment
A zero-point adjustment is not required for the sensors described in this operating manual. If there is no analyte present in the process medium, approximately zero is indicated. The zero point is independent of changes in the flow rate, conductivity, temperature and pH value.
9 Calibration
31

10 Eliminating errors and faults

Error/fault Possible cause Remedy Preventive
measures
(1) Sensor output signal too high or too low.
Incorrect calibration. Repeat calibration
using the DPD method See "Initial condition", page 30.
If necessary calibrate the sensor more frequently.
(2) Sensor output signal is too low.
Sensor cannot be adjusted to the DPD value.
(3) Sensor output signal is too low.
Sensor cannot be adjusted to the DPD value.
or else the sensor output signal decreases or remains constant while the DPD value increases
or the signal fluctuates
(4) Sensor output signal is too high.
Sensor cannot be adjusted to the DPD value.
Settling time too short. Wait at least 2 hours Deposit on the tip of
the electrode finger (measuring electrode)
Incident flow on sensor is too low.
Membrane destroyed: Electrolyte escaping ­process medium entering.
Effect of surfactants in detergents and cleaning agents on the membrane. The hydrophobic properties of the membrane are lost and passage of water or ions is possible.
Coating on the membrane cap.
Gas bubbles on the outside of the membrane.
No electrolyte in the membrane cap
In addition to the analyte, the process medium also contains other oxidizing agents, for example ClO
2
, O
Clean tip of electrode finger.
Increase incident flow. Monitor minimum
Replace membrane cap.
Replace membrane cap. Replace the water as well before using the chlorine sensor again (to eliminate all traces of surfactants).
Replace membrane cap.
Increase the flow briefly.
Fill the membrane cap with electrolyte, See "Topping up reference electrolyte", page 21
Avoid adding these substances. Change the water.
3
If necessary shorten maintenance intervals.
inflow. Avoid damaging the
membrane. Do not knock the sensor when the membrane cap is screwed on. Avoid incident flow containing coarse pieces or glass splinters.
Avoid allowing surfactants to affect the membrane. After cleaning the system, thoroughly flush out all detergents or other disinfectants.
Check installation and change if necessary.
Remove cleaning agents and disinfectants completely after use.
32
10 Eliminating errors and faults
Error/fault Possible cause Remedy Preventive
measures
(5) The DPD and sensor values match and the trend of the redox measurement is correct, but the setpoint value is not maintained.
(6) Sensor value and DPD value do not match, sensor values fluctuate: too much / not enough.
Incorrect control parameters.
The amount of disinfectant dosed per unit of time is too high. Concentration is exceeded before the process medium reaches the sensor.
System circulation is too slow.
System circulation is too slow.
Optimize the control parameters.
Reduce the amount added per unit of time. Reduce the concentration of disinfectant in the solution added.
Improve mixing. Make structural
changes to ensure more thorough mixing.
Improve mixing. Make structural
changes to ensure more thorough mixing.
(7) Sensor is unusually slow in responding
(8) Output signal of the sensor is 0
Membrane partially blocked by contamination such as lime or oil. This interferes with the disinfectant reaching the sensor.
Sensor connected to the transmitter/ controller with reverse polarity.
Measuring lead broken.
Sensor faulty. Send the sensor in to
Transmitter/controller faulty
Replace membrane cap.
Connect the sensor correctly, See section 6 "Electrical connection", page 17
Replace measuring lead.
the manufacturer to have it checked/ repaired.
Send the transmitter/ controller in to the manufacturer to have it checked/repaired.
Take steps to improve the water quality.
33
JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc.
Street address: Moritz-Juchheim-Straße 1 36039 Fulda, Germany Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO House Temple Bank, Riverway Harlow - Essex CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 63 52 62 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-

1-
800-554-JUMO Fax: 315-697-5867 E-mail: infoUS@jumo.NET Internet: www.jumousACOM
Loading...