JUMO Logoline 500 junior Operating Manual

Pen recorder with text printing
B 95.3535
Operating Manual
4.99/00351782
Levels and key functions
Key functions
- Chart fast forward when recording is stopped
-Abort parameter input (Exit)
- Level change backwards
S
- Start/Stop recording
-Decimal place selection during value input
F
- not used
l
- Move cursor to the left (select digit)
h
- Parameter selection
- Channel selection
-Value selection from value table
- Increment current digit
E
- Initiate parameter input
-Enter input
- Acknowledge error message
h+E
- Level change forwards
Contents
1 About this Manual
1.1 Preface ...................................................................................................................... 3
1.2 Arrangement of the documentation ......................................................................... 4
1.3 Typographical conventions ...................................................................................... 5
1.3.1 Warnings ................................................................................................................... 5
1.3.2 Notes ........................................................................................................................ 5
1.3.3 Presentation .............................................................................................................. 6
2 Identifying the instrument version
2.1 Instrument description .............................................................................................. 7
2.2 Type designation ...................................................................................................... 8
3 Installation
3.1 Location and climatic conditions ........................................................................... 11
3.2 Fitting in position .................................................................................................... 12
4 Electrical connection
4.1 Notes on installation ............................................................................................... 13
4.2 Connection diagram ............................................................................................... 14
5 Starting up
5.1 Display and controls ............................................................................................... 15
5.2 Opening and closing the door ............................................................................... 15
5.3 Marking the channel label ...................................................................................... 15
5.4 Fitting the fibre pens ............................................................................................... 16
6 Preparation
6.1 Operating modes and status .................................................................................. 17
6.2 Operating principle ................................................................................................. 18
7 Text printing
7.1 Printing priorities ..................................................................................................... 23
7.2 Time ........................................................................................................................ 24
7.3 Change of chart speed ........................................................................................... 25
7.4 Recording start and end ......................................................................................... 26
7.5 Print test .................................................................................................................. 27
7.6 Service print ............................................................................................................ 28
Contents
8 Programming
8.1 Basic status ............................................................................................................ 29
8.2 Operating level ....................................................................................................... 30
8.3 Chart speed ............................................................................................................ 31
8.4 Print test .................................................................................................................. 32
8.5 Service print ............................................................................................................ 33
8.6 Level inhibit and code request ............................................................................... 34
8.7 Parameter level ....................................................................................................... 35
8.8 Language ................................................................................................................ 36
8.9 Date and time ......................................................................................................... 37
8.10 Summer time .......................................................................................................... 38
8.11 Display of time ........................................................................................................ 40
8.12 Configuration level .................................................................................................. 41
8.13 Signal inputs ........................................................................................................... 42
9 Consumables
9.1 Summary of consumables ..................................................................................... 45
9.2 Removing and replacing the chart cassette .......................................................... 46
9.2.1 Changing the roll chart ........................................................................................... 47
9.2.2 Changing the fanfold chart ..................................................................................... 48
10 Accessories
10.1 Converting the chart cassette ................................................................................ 49
11 Fault finding
11.1 What to do if... ........................................................................................................ 51
12 Appendix
12.1 Error messages ...................................................................................................... 53
12.2 Hardware fault ........................................................................................................ 54
12.3 Status messages .................................................................................................... 55
12.4 Overview of the parameterss ................................................................................. 56
Index
3
1 About this Manual
1.1 Preface
Please read this Operating Manual before commissioning the instru­ment. Keep the manual in a place which is at all times accessible to all users.
Please assist us to improve this manual where necessary.
Your suggestions will be most wel­come.
Phone Germany (0661) 6003-725 abroad (int. +49) 661 6003-0 Fax Germany (0661) 6003-681 abroad (int. +49) 661 6003-607
All necessary settings and, where appropriate, alterations inside the instrument are described in this manual. If any difficulties should still arise during start-up, you are asked not to carry out any prohibited ma­nipulations. You could endanger your rights under the instrument warranty!
Please contact the nearest JUMO office or the main factory.
E
When returning chassis, modules or components, the rules of EN 100 015 “Protection of electrostati­cally sensitive components” have to be observed. Use only the appropri­ate ESD packaging material for transport.
Please note that we can not be held liable for any damage caused by ESD (electrostatic discharges).
1 About this Manual
4
1.2 Arrangement of the documentation
This Operating Manual is supplied with the instrument. It is addressed to the OEM (orig­inal equipment manufacturer) and users with appropriate technical know-how.
In addition to installation and electrical con­nection, it contains all the information on in­strument start-up, operation and parameter setting.
1 About this Manual
5
1.3 Typographical conventions
1.3.1 Warnings The signs for Danger and Warning are
used in this manual under the following con­ditions:
V
Danger This symbol is used when there may be
damage to personnel if the instruction is disregarded or not followed accurately.
Warning This symbol is used when there may be
damage to equipment or data if the instruction is disregarded or not followed accurately.
E
Warning This symbol is used where precautions
have to be observed during the handling of components which may be damaged by electrostatic discharges.
1.3.2 Notes
Note This symbol is used if your special
attention is drawn to a remark.
Reference This symbol refers to additional information
in other handbooks or sections.
abc
1
Footnote Footnotes are notes which refer to certain
points in the text. Footnotes consist of 2 parts:
Markings in the text and the footnote text. The markings in the text are arranged as
continuous raised numbers. The footnote text (in smaller typeface) is
placed at the bottom of the page and starts with a number and a full stop.
1 About this Manual
6
* Action This sign indicates that an action to be per-
formed is described. The individual steps are indicated by this
asterisk, e.g.:
* press key h *enter with E
1.3.3 Presentation
h+E
Keys Keys are shown as boxes. Both symbols
and texts are possible. Where a key has multiple functions, the text shown is the one corresponding to the actual function dis­cussed.
7
2 Identifying the instrument version
2.1 Instrument description
The pen recorder provides up to three chan­nels for recording measurements, which are isolated from each other by optocouplers. Channel 1 can be used to write text in addi­tion to the signal trace. All channels are ze­roed using Hall sensors. The measurements can be indicated by pointers against scales.
A 4-digit 7-segment display is available for programming. Operation is by six keys on the instrument front. The configuration data are stored permanently in EEPROM.
Input signals include current and voltage signals (standard signals). The watchdog monitors the recorder function and triggers a restart in the event of a fault. On a power failure, the real-time clock is supplied by a lithium battery.
An overview of all the parameters of the pen recorder is given in the Appendix (Section 12.4).
2 Identifying the instrument version
8
2.2 Type designation
The label is affixed to the housing. The in­strument version can be identified from the type designation.
The extra Codes are listed in sequence and separated by a comma.
The voltage of the supply must correspond to the voltage indicated on the label.
LL. v-44uj/... Pen recorder
with scales
1 1 channel
with text writing
2 2 channels
(channel 1 with text writing)
3 3 channels
(channel 1 with text writing)
vamplifier
-44 bezel 144mm x 144mm u convertible chart cassette
Extra Codes
input circuit
recording
sk special scales,
e. g. in m3/h, bar etc. (to calibration curve if non-linear)
fp cassette for fanfold chart 16m long r32 cassette for roll chart 32m long
2 Identifying the instrument version
9
housing
Ordering example
(1) (2) (3) (4) (5) (6) LL 3 v-44uj/as, fp, tm
Standard accesssories
as chart run-out slot in housing door ab housing for wall mounting.
The panel-mounting housing is fitted with a carrier and can be swung out through 90°.
tm housing with carrying handle,
rubber feet and terminal cover, also 3 m mains cable with plug
TG-35 portable recording station
(1), (3) pen recorder (2) 3 channels (3) bezel 144mm x 144mm (4) run-out slot in housing door (5) fanfold chart cassette (6) housing with carrying handle
1 Operating Manual B 95.3535 2 mounting brackets 1 fibre pen, disposable, for each
channel
2 chart rolls 16m long
or
1 chart roll 32m long
(with Code r32) or
1 fanfold chart pack 16m long
(with Code fp)
2 Identifying the instrument version
10
11
3 Installation
3.1 Location and climatic conditions
The instrument location should as far as possible be free from shock and vibration. Stray electromagnetic fields, e.g. from mo­tors, transformers etc., should be avoided.
The ambient temperature at the location may be between -10 and +50°C at a relative humidity not exceeding 75%, without con­densation.
Corrosive air and fumes may interfere with the function and life of the recorder.
Section 4.1
3 Installation
12
3.2 Fitting in position
Side view
Front view
Panel cut-out
* Insert the pen recorder from the front into
the panel cut-out.
* From the back of the panel, hook the
mounting brackets (1) into the cut-outs in the sides of the housing. The flat bracket faces must lie against the housing.
* Place the brackets against the rear of the
panel and tighten them evenly.
13
4 Electrical connection
4.1 Notes on installation
The choice of cable, the installation and
the electrical connection of the instru­ment must conform to the requirements of VDE 0100 “Regulation on the Installa­tion of Power Circuits with nominal volt­ages below 1000V” or the appropriate local regulations.
Earth the instrument at the PE terminal
to the earth conductor. This line should have at least the same cross-section as the supply lines. Earth lines should be run in a star layout to a common earth point which is connected to the earth conductor of the supply. Do not loop the earth connections, i.e. do not run them from one instrument to another.
Work inside the instrument must only be
carried out to the extent described and, like the electrical connection, only by properly qualified personnel.
Do not connect any additional loads to
the supply terminals of the instrument.
If contact with live parts is possible when
working on the instrument, it has to be isolated on both poles from the supply.
The instrument is not suitable for instal-
lation in hazardous areas.
Electromagnetic compatibility (EMC)
conforms to the standards and regula­tions listed under Technical Data.
Inductive loads in the neighbourhood of
the instrument, such as contactors or solenoid valves, should be fitted with RC modules for interference suppression.
Run input, output and supply lines sepa-
rately and not parallel to each other.
Sensor and interface lines should be
arranged as twisted and screened cables. Do not run them close to current-carrying components or cables. Earth the screen at one end at the instrument on terminal PE.
4 Electrical connection
14
4.2 Connection diagram
Rear view with screw-clamp connectors
V
The electrical connection must only be carried out by properly qualified person­nel.
Terminals
Supply as on label N neutral
L1 line PE protective earth
N(L-) L1 (L+) PE
Analogue inputs Input 1 Input 2 Input 3
Field Field Field
Voltage input up to 170mV
1. 2. 3.
Voltage input above 170mV
Current input
Inputs 8 and 9 are not available for LOGOLINE 500 junior.
15
5 Starting up
5.1 Display and controls
(1) 4-digit 7-segment LED display (2) fibre pen, channel 3, green (3) fibre pen, channel 2, red (4) fibre pen, channel 1, blue (5) chart (6) keys for operation and programming (7) door (8) housing to DIN 43700 for flush-panel
mounting, galvanised steel
(9) channel marker label
5.2 Opening and closing the door
For opening and closing the door, press as shown in the illustration until a click can be heard.
5.3 Marking the channel label
Please mark the channel label (1) with the measurement channel designation and the corresponding range.
(1)
5 Starting up
16
5.4 Fitting the fibre pens
* Open the recorder door * Stop recording
(press S key)
* Swing scale upwards * Slide the fibre pen into the holder
up to the stop
* Swing scale down again
17
6 Preparation
6.1 Operating modes and status
Operating mode/status Notes Basic status Basic status of the pen recorder with signal acquisition
and processing. The display shows
- time (can be switched off via the parameter
parameter leveldisplay time)
- and system error messages, where appropriate.
In the stop status the display flashes “
StOP
”.
If there is a system error message, it is shown flashing at regular intervals instead of the time.
Stop
Stop by key
Recording is stopped and continued using key S. In the stop status, the display flashes “
StOP
”.
Chart speeds
Normal chart speed The chart is advanced with the speed programmed
under chart speed. Section 8.3
6 Preparation
18
6.2 Operating principle
The individual parameters and functions are divided into four levels for clearer operation of the recorder.
Basic status
Basic status of the pen recorder with signal acquisition, recording and processing.
The display shows:
- the time (can be switched off via the para-
meter parameter leveldisplay time)
- and system error messages,
where appropriate.
If there is a system error message, it is shown flashing at regular intervals instead of the current display.
Operating level
At this level, signal acquisition and process­ing remain activated.
The following parameters can be activated at the operating level:
-chart speed
- print test
- service print
6 Preparation
19
Parameter level
The parameter level and the subsequent configuration level are protected by a code­number to prevent unauthorised access.
The codenumber is: 9200.
If a wrong codenumber is input, the individ­ual parameters can be viewed, but can not be programmed.
Following input of the correct codenumber, signal acquisition and recording are inter­rupted at this level.
The following parameters are altered at the parameter level:
-language
- date and time
-summer time
- display time
Configuration level
At the configuration level, the following pa­rameters are altered:
- signal inputs
An overview of all parameters of the record­er is given in the Appendix.
Section 8.6
Section 12.4
6 Preparation
20
The levels, parameters and sub-parameters are arranged in a tree structure. Starting from the basic status, the program branch­es out into the individual levels, and from there into the corresponding parameters and, if there are any, into the sub-parame­ters.
To alter a specific parameter, run through the relevant levels up to this parameter.
The editing process is started with E. If a parameter has several sub-parameters,
these are reached with E.
Coding the parameters
The parameters and sub-parameters are in­dicated in code form on the 7-segment dis­play. Coding involves up to four digits. The first digit is a letter, the others are numbers.
Digit Notes
1
A
= operating level
b
= parameter level
C
= configuration level
2 indicates the parameter
3 indicates the sub-parameter
4 channel number
6 Preparation
21
Entering parameters
If there are no further sub-parameters,
E transfers all the data of the parame-
ter to the memory.
Aborting programming
Programming can be aborted within a pa­rameter by pressing the key e.
Error messages during programming
If there are error messages during program­ming due to incorrect inputs, these have to
be acknowledged with E before pro­gramming can be repeated.
If the last sub-parameter has not yet been entered, the parameter data which have already been edited are rejected. The old contents of the sub-parameters are retained.
Key functions
- Chart fast forward when recording is stopped
- Abort parameter input (Exit)
- Level change backwards
S
- Start/Stop recording
- Decimal place selection during value input
F
- not used
l
- Move cursor to the left (select digit)
h
- Parameter selection
- Channel selection
- Value selection from value table
- Increment current digit
E
- Initiate parameter input
- Enter input
-Acknowledge error message
h+E
- Level change forwards
6 Preparation
22
Selection
Selection consists of a list of several options.
Two keys are used to select an option: * select option with the h key * enter selection with E
Value input
Three keys are used to input values: * select the digit to be altered with the
l keyy
* increment the selected digit with the
h keyy
* shift the decimal point with the S key * enter value input with E
An incorrect input produces an error message on the display.
Section 12.1
The error message must be ac­knowledged with E.
The value can then be entered again.
23
7 Text printing
The pen recorder can print text in addition to the trace using the fibre pen of channel 1. Text printing is used for comments on the trace and for event recording. The charac­ters are written in dots on a 7 x 9 matrix.
7.1 Printing priorities
There are various types of text with different priorities. These priorities determine abort criteria where printing of several texts is re­quested simultaneously.
The printing of text with low priority is abort­ed when text with high priority is to be print­ed.
List of priorities:
All printing requests which are still present are rejected and fresh ones ignored in the following events:
- Pen recorder moves to stop status
- Print test is started
- Service print is started
- Pen recorder is switched off
Priority Text
higher
lower
- Stop by key
S
, no chart, configuration by keys, end-of-recording text
- Service print
-Print test
- start-of-recording text
- chart speed change
-time
7 Text printing
24
7.2 Time
Printing is performed cyclically at intervals of approx. 4 cm. Since time is not printed at “irregular” times, the interval mentioned above is only an approximation.
With every fourth time printing, the current chart speed, the instrument name or the date are printed alternately following the time.
At a change of date, the date is generally in­cluded in the printing at 00:00 hours.
In order to obtain an exact time reference on the chart, the time is preceded by a time ref­erence mark (“_”).
If a “?” is printed after the time, the time has to be checked and, if nec­essary, freshly programmed. (Section 8.9).
Time printing only occurs when the chart speed is in the range of 5 — 360mm/h.
7 Text printing
25
7.3 Change of chart speed
Every change of chart speed is reported by printing a line, the current time and the new chart speed.
Setting the printing priorities ( Section 7.1) affects not only the printing of the chart speed change marking, but also the change itself.
There is no chart speed change as long as a text with a higher priority is being printed.
The line for marking the chart speed change is always printed, the text (time and new speed) only when the chart speed is in the range of 5 — 360mm/h.
7 Text printing
26
7.4 Recording start and end
Recording start and end are reported by a start and end text.
Printing the end text is omitted when the chart sensor recognises “end of chart”.
Printing the end text is aborted by pressing the e key.
7 Text printing
27
7.5 Print test
The print test can only be initiated on the re­corder using the keys (operating level). It is used to check the function of the writing system and the fibre pens.
Section 8.4
Started text printing is aborted and not continued after the print test.
All text printing requests arriving during the print test are rejected.
7 Text printing
28
7.6 Service print
The service print can be initiated on the re­corder using the keys (operating level).
It prints the software version, the number of supply interruptions and the total operating time of the recorder in hours.
Section 8.5
Started text printing is aborted and not continued after the service print.
All text printing requests arriving during the service print are rejected.
29
8 Programming
8.1 Basic status
The recorder is in the basic status after the supply has been applied, and it has been in­itialised. The measurement signals are ac­quired, processed and recorded.
In the basic status the following are indicated:
- the clock time and
- status/error messages, as appropriate. Status and error messages are flashing reg-
ularly instead of the current display. Using the S key, the recording is stopped
and re-started. In the stop status, the dis­play flashes “
StOP
”.
Pressing the e key in the stop status ac­tivates the chart fast forward.
It is possible to change into the operating level by simultaneously pressing the
h+Ekeys.
8 Programming
30
8.2 Operating level
The operating level serves for performing simple operating actions.
It is reached from the basic status by press­ing the h+Ekeys simultaneously, from the parameter level by pressing the
e key.
The operating level is identified in the dis­play by the letter “A”.
The operating level includes the following parameters:
-chart speed
- print test
- service print
A summary of all parameters of the recorder is given in the Appendix (Section 12.4).
While the operating level is activat­ed, signal acquisition, processing and recording are continuing.
Display Edit Selection/input with keys Enter Continue with
A
h
forwards
8 Programming
31
8.3 Chart speed
The chart speed for recording the measure­ment is set here.
The chart speed is selected from a table. The chart speeds are: 0, 5, 10, 20, 60, 120, 240, 300, 360, 600,
720, 1800, 3600 und 7200mm/h.
Display Edit Selection/input with keys Enter Continue with
A1
E
240 Select chart speed: 0, 5, 10, 20, 60, 120, 240,
300, 360, 600, 720, 1800, 3600, 7200mm/h
hEh
forwards
8 Programming
32
8.4 Print test
Print test is used to check the function of the writing system and the fibre pens.
PRINTTEST = On continues until it is terminated by PRINTTEST =
OFF
.
Started text printing is aborted and not continued after the print test.
All text printing requests arriving during the print test are ignored or rejected.
Display Edit Selection/input with keys Enter Continue with
A2
E
OFF, On Select status of print test
hEh
forwards
8 Programming
33
8.5 Service print
The service print informs about the software version, the number of supply interruptions and the total operating time of the recorder. .
Started text printing is aborted and not continued after the service print.
All text printing requests arriving during the service print are ignored or rejected.
Display Edit Selection/input with keys Enter Continue with
A3
E
no, YES Service print no,
yes
hEh
forwards
8 Programming
34
8.6 Level inhibit and code request
The parameter level is protected by a 4-digit codenumber.
The codenumber is: 9200. If the codenumber is entered, the following
are interrupted:
- signal acquisition and
-recording.
The parameters can now be programmed.
If an incorrect codenumber has been entered, the recorder remains in normal operation. Signal acquisi­tion, processing and recording re­main activated.
The parameters can be viewed, but not programmed.
8 Programming
35
8.7 Parameter level
At the parameter level, it is possible to con­figure general parameters.
From the operating level, the parameter lev­el is reached by simultaneously pressing
keys h+E, from the configuration lev- el by pressing the e key at the parame-
ter level. The parameter level comprises the following
parameters:
-language
- date and time
-summer time
- display time
If an incorrect codenumber is en­tered, the recorder remains in nor­mal operation. Signal acquisition, processing and recording are acti­vated.
The parameters can be viewed, but not programmed.
Following entry of a correct code­number, the normal operation is in­terrupted.
There is no signal acquisition, processing or recording.
Display Edit Selection/input with keys Enter Continue with
COdE
E
0000 Input codenumber
h, lE
2
2
b
h
forwards
8 Programming
36
8.8 Language
The following languages can be selected:
- English
-German
-French
The language setting affects the texts which are printed.
Display Edit Selection/input with keys Enter Continue with
b1
E
Select language: GEr (GERMAN)
EnGL (ENGLISH) FrEn (FRENCH)
hEh
forwards
8 Programming
37
8.9 Date and time
Programming the system clock of the re­corder. The following are programmed: day, month, year, weekday, hour and minute.
The current clock time is printed at certain events and is shown on a 7-segment dis­play in the basic status.
Date and time are checked for plau­sibility after input. An error message is output, when appropriate.
The error message is acknowl­edged with E and the data can then be re-entered.
Display Edit Selection/input with keys Enter Continue with
b2
E
2
2
b21
E
09.02. Input day and month:
DD.MM.
h, lE
3
3
b22
E
98 Input year: YY
h, lE
Check date
4
4
b23
E
1 Select weekday: 1Monday
2Tuesday 3Wednesday 4Thursday 5Friday 6Saturday 7Sunday
hE
5
5
b24
E
00.00 Input clock time:
hh.mm
h, lE
Check time
h
forwards
8 Programming
38
8.10 Summer time
Input of a time period during which the sys­tem clock of the recorder is changed to summer time.
Example: Summer time start: 29.03.98, 2:00 hrs
Summer time end: 25.10.98, 3:00 hrs On 29.03.98 at 2:00 hrs the system clock is
put forward automatically by an hour to 3:00 hrs.
On 25.10.98 at 3:00 the clock is put back automatically by an hour to 2:00 hrs.
The time interval for the summer time always applies only to the peri­od indicated.
It has to be re-programmed for each year.
8 Programming
39
Display Edit Selection/input with keys Enter Continue with
b3
E
2
2
b31
E
Off, On Input status
hE
On ⇒ 3 Off:
h forwards
3
b32
(On)
E
29.03. Input day and month of
start date: DD.MM.
h, lE
4
4
b33
E
98 Input start year:
YY
h, lE
Check date
5
5
b34
E
02.00 Input time for start:
hh.mm
h, lE
Check time
6
6
b35
E
25.10. Input date and month
of end date: DD.MM.
h, lE
7
7
b36
E
98 Input year of end date:
YY
h, lE
Check date
8
8
b37
E
03.00 Input time for end:
hh.mm
h, lE
Check time
h forwards
8 Programming
40
8.11 Display of time
The parameter “Display time” has been added to the parameter level with effect from the instrument version 116.01.02.2.
The time is indicated on the display when the recorder is in the basic status. The time display can be suppressed (status = OFF) by using this parameter.
The instrument version can be determined by the parameter
Operating level Service print.
Display Edit Selection/input with keys Enter Continue with
b5
E
Off, ON Select status
hEh
forwards
8 Programming
41
8.12 Configuration level
The signal inputs are configured at the con­figuration level.
It can be reached from the parameter level by simultaneously pressing the h+ E keys.
If an incorrect codenumber has been entered, the recorder remains in normal operation. Signal acquisi­tion, processing and recording re­main activated.
The parameters can be viewed, but not programmed.
If a correct codenumber is entered, the normal operation is interrupted.
There is no signal acquisition, processing or recording.
Display Edit Selection/input with keys Enter Continue with
C
h
forwards
8 Programming
42
8.13 Signal inputs
The input signal is selected, and the ranges as well as the filter constants for each signal input are determined here.
Display Edit Selection/input with keys Enter Continue with
C1
E
C101 Select input channel: C101 channel 1 C102 channel 2 C103 channel 3
hE
2n
with n = 1 — 3
2n
C111 C112 C113
E
CA00 Select input signal: CA00 voltage Cb00 current
hE
Voltage
31n
Current ⇒ 41n
Voltage:
31n
CA11 CA12 CA13
E
1 Select unit: 0 (mV)
1 (V)
hE
32n
32n
CA21 CA22 CA23
E
0.000 Enter range start
h, l
,
S
E
check range start (–10.0V
range start < 10.0V or
0.0V
range start < 22.0V)
33n
8 Programming
43
33n
CA31 CA32 CA33
E
1.000 Input range end
h, l, S
E
Check (–10.0V < range end 10.0V or
0.0V < range end 22.0V) span
5mV
5n
(FILTER)
Current: 41n
Cb11 Cb12 Cb13
E
0.000 Input range start
h, l, S
E
Check range start (–20.5mA range start < 20.5mA or
0.0mA range start < 45.0mA or –4.0mA
range start < 21.0mA)
42n
42n
Cb21 Cb22 Cb23
E
020.0 Input range end
h, l, S
E
Check range end (–20.5mA < range end 20.5mA or
0.0mA < range end 45.0mA or –4.0mA
< range end 21.0mA)
span
0.5mA
5n
(FILTER)
5n
C121 C122 C123
E
0.800 Input filter constant
(value range:
0.0 — 10.0 sec)
h, l, S
E
Check value range
0.0 — 10.0sec
h forwards
Display Edit Selection/input with keys Enter Continue with
8 Programming
44
45
9 Consumables
9.1 Summary of consumables
Fibre pens, disposable
blue, Part No.: 00309750 red, Part No.: 00309751 green, Part No.: 00309753
Roll chart
overall width: 120mm no name, % graduation, linear
overall length: 16m Packing unit: 5 rolls Part No.: 00331497
no name, % graduation, linear overall length: 32m Packing unit: 5 rolls Part No.: 00331499
no name, special graduation (printing as order specification) overall length: 16m/32m
Fanfold chart
overall width: 120mm overall length: 16m
no name, % graduation, linear Packing unit: 5 packs Part No.: 00331490
no name, special graduation (printing as order specification)
9 Consumables
46
9.2 Removing and replacing the chart cassette
9 Consumables
47
9.2.1 Changing the roll chart
1
5
26
3
7
4
8
9 Consumables
48
9.2.2 Changing the fanfold chart
1
4
2
5
3
5 cm10 cm
15 cm
****** ***** ***** ****
1 m
END
E
END FIN
49
10 Accessories
10.1 Converting the chart cassette
13
24
10 Accessories
50
51
11 Fault finding
11.1 What to do if...
Fibre pens on 0% - The value is outside the measuring range
(Out Of Range)
- The transducer has been connected incorrectly
- The signal inputs are configured incorrectly
- Sensor break
No recording - Key S has been pressed
Section 6.1
- The chart cassette has not been fitted correctly Section 9.2 Caution: Insert the chart cassette by applying light up­wards pressure
- The take-up roll has not properly engaged with the chart cassette Section 9.2.1
- The end of the chart has been reached Section 9.2.1, Section 9.2.2
- The chart speed has been programmed as 0 mm/h Section 8.3
Fibre pen does not write - The fibre pen has not been fitted correctly
Section 5.4
- The fibre pen has run out of ink
No chart feed - The chart cassette has not been fitted correctly
Section 9.2
- The chart has not been fitted correctly
- The chart speed has been programmed as 0mm/h
- The pin roller does not engage with the chart perforations
- The tear-off edge in not correctly engaged
- The pen recorder is in the STOP status
Measurements are not recorded
- Check that the connecting terminals are tightened properly
- Check the supply
- Check the input configuration (range)
- Check the transducers and their cables, measure them where appropriate
11 Fault finding
52
53
12 Appendix
12.1 Error messages
All error messages flash on the 7-segment display at regular intervals. All other instrument functions remain unaf­fected as far as possible.
Display Cause / Remedy
Status message
Er13
The chart cassette has been removed or the end of the chart has been reached; a fresh chart must be fitted.
Section 9.2 ff
Error or fault on module
Er10
The battery for the real time clock and for RAM backup is discharged.
Please contact the nearest office or the factory.
Errors on parameter input
Err
- Date is invalid An invalid date has been input. The input must be repeated.
- Time is invalid An invalid time has been input. The input must be repeated.
- Error in value input The input is outside the range of values. The input must be repeated.
54
12 Appendix
12.2 Hardware fault
If one of the following faults occurs, the re­cording is aborted and the error message flashes on the display.
The recorder does not react to any event and can not be operated.
Please contact the nearest office or the main factory.
Display Cause / Remedy
Er12
The EEPROM in the recorder is faulty, the configura­tion date can no longer be stored.
Er17
The A/D converter of the recorder is faulty.
Er18
One of the Hall sensors in the recorder is faulty.
55
12 Appendix
12.3 Status messages
The following status messages are indicat­ed on the 7-segment display:
Display Description
InIt
The recorder is being initialised. Please wait.
StOP
The recorder is in the Stop status, because the S key has been pressed.
bUSY
The configuration data are being written to the EEPROM. During this time the recorder does not respond to any inputs.
56
12 Appendix
12.4 Overview of the parameters
s
Parameter Description Chart speed Chart speed in mm/h Language
Language (for printing)
Date and time
System clock of the recorder
Summer time
Start and end of summer time
Display time Switch on/off display of time in the basic status Signal inputs Selection of the analogue input signals
57
Index
A
Analogue inputs 14
B
Basic status 18, 29
C
Change of chart speed 25 Channel label 15 Chart cassette
changing fanfold chart 48 changing roll chart 47 converting 49
removing and replacing 46 Chart fast forward 29 Chart speed 31 Clock time 24, 37 Codenumber 19, 34 Configuration level 41 Connection diagram 14 Consumables 45 Current 14, 42
D
Decimal place 22 Display and controls 15 Display time 40 Documentation 4 Door 15
E
Earth line, protective 13 Electrical connection 14 Electrostatic discharge (ESD) 3 Entering parameters 21 Error messages 21, 53 Extra Codes 8
F
Fanfold chart 45 Faults, remedy of 51 Fibre pen 45 Fibre pen, fitting of 16 Filter constants 42 Fitting in position 12 Front view 12
H
Hardware fault 54 Hazardous area (EX) 13
I
Input signals 42 Installation notes 13 Instrument description 7 Interface lines 13
K
Key functions 21
L
Language 36 LED display 15 Level inhibit and code request 34 Levels 18
basic status 18 configuration level 19 operating level 18 parameter level 19
Location of recorder 11
O
Operating level 18, 30 Operating mode
basic status 17, 18
58
Index
chart speed 17
stop 17 Operation 18 Overview of parameters 56
P
Panel cut-out 12 Parameter coding 20 Parameter level 19, 35 Parameter overview 56 Parameters, entering 21 Print test 27, 32 Printing priority 23 Programming
abort 21
chart speed 31
date and time 37
language 36
level inhibit and code request 34
print test 32
selection 22
service print 33
signal inputs 42
summer time 38
value input 22
R
Range 42 Recording start 26 Recording start and end 26 Returning the recorder 3 Roll chart 45
S
Selection 22 Sensor 13 Sensor break 51 Service print 28, 33
Side view 12 Signal inputs 42 Starting up 3 Status messages 55 Stop 17 Sub-parameters 20 Summer time 38 System clock 37, 38
T
Text printing 23
abort criteria 23 chart speed 25 clock time 24 print test 27 printing priority 23
service print 28 Time reference mark 24 Tree structure 20 Type designation 8 Typographical conventions 5
V
Value input 22 Voltage 14, 42
W
Warranty 3
M.K. JUCHHEIM GmbH & Co.
36035 Fulda Germany Phone ++49 6 61-60 03-0 Fax ++49661-6003-607 Telex 49 701 juf d email JUMO_de@e-mail.com
USA
JUMO PROCESS CONTROL INC. 735 Fox Chase, Coatesville, PA 19320 Phone 610-380-8002,
800-554 JUMO
Fax 610-380-8009
United Kingdom
JUMO Instrument Co. Ltd. Tem ple Ban k, R iv er way GB-Harlow, Essex CM20 2TT Phone (0 12 79) 63 55 33 Fax (0 12 79) 63 52 62
Loading...