JUMO flowTRANS MAG I02 Operating Manual

JUMO flowTRANS MAG I02
Electromagnetic flowmeter
for liquids
Operating Manual
40601100T90Z001K000
V3.00/EN/00690082
Contents
Contents
1.1 Symbols and signal words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Definition of terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3 Basic safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.1 Warranty conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
4.2 Information available online . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5 Design and function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.1.1 Design and function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.1.2 Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.2 Areas of application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.3 Nameplate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.1 Measuring range and accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.2 Measurement media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.3 Mechanical features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.3.1 Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.3.2 Pressure temperature diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.4 Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.5 Connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.6 Environmental influences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.7 Compliance with standards and directives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.2 Installation into the pipeline. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.2.1 Recommendation for installing into the pipeline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.2.2 Installation in a pipeline with G 2" union nut . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
7.3 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.3.1 Make sure that the potential is equal in the installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.3.2 Use of a cable holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Contents
7.3.3 Terminal assignment and use of a selection switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.3.4 Connecting the digital input DI1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.3.5 Connection for the current output AO1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7.3.6 Connecting the transistor output DO1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
7.3.7 Connecting the relay outputs DO2 and DO3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8 Startup and parameterization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.2 Operating levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.3 Display, navigation keys and status LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
8.4 Using the navigation keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.5 Process level details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.6 Parameter menu details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.6.1 Choosing the display language. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8.6.2 Choosing the flow rate unit, the number of decimals, and the counter display unit . . . . . . . . . 38
8.6.3 Entering the K-factor for the fitting (406090) used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
8.6.4 Determining the fitting's K-factor with a calibration procedure (Teach-In). . . . . . . . . . . . . . . . . 41
8.6.5 Configuring outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
8.6.6 Configuring current output AO1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8.6.7 Configuring the transistor output DO1 as a pulse output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
8.6.8 Configuring the transistor output DO1 to switch a load depending on two threshold values . . 47
8.6.9 Configuring the transistor output DO1 to switch a load when the direction of flow changes . . 50
8.6.10 Configuring the transistor output DO1 to switch a load when the device generates a warning. . 51
8.6.11 Configuring the relay outputs DO2 and DO3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
8.6.12 Configuring the digital input DI1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.6.13 Adjusting the filter to the measured flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
8.6.14 Resetting both counters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
8.6.15 Adjusting the mains frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.6.16 Adjusting the CUT-OFF flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.6.17 Setting display backlight and how long it stays on, or deactivating the backlight . . . . . . . . . . . 64
8.7 Test menu details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
8.7.1 Adjusting the current output AO1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
8.7.2 Calibrate the flow rate zero point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
8.7.3 Checking proper response of the outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
8.7.4 Changing the sensor's KW value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
8.7.5 Monitoring the flow rate in the pipeline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
8.8 Information menu details. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
9 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9.2 Returning devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.3 Cleaning the device and sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
9.4 Replace the seal (device with G 2" union nut) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
9.5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Contents
9.5.1 Solving a problem with the device status LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
9.5.2 Solving a problem without a warning or error message with the device status LED. . . . . . . . . 76
9.5.3 Solving a problem without a warning or error message with a green device status LED . . . . . 78
9.5.4 Solving a problem with a warning or error message with the red device status LED . . . . . . . . 79
9.5.5 Solving a problem with a warning or error message with the orange device status LED . . . . . 80
10 Spare parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
11 Packaging and transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
12 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
13 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
14 China RoHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
1 General information and notes for the reader
?

1 General information and notes for the reader

Before mounting and starting up the unit, this manual must be read carefully.
The manual is an integral part of the product and must be stored for subsequent use. In the interest of clarity, the manual does not contain all the detailed information about all product versions and cannot take into consideration every conceivable case involving the installation, operation or maintenance.
If you would like further information or if problems occur that are not covered by the manual, the required information can be obtained from the manufacturer.
The contents of this manual are not part of or a change to a previous or existing agreement, assurance or legal relationship.
This product is built based on state-of-the-art technology and safe to use. It has been tested and was shipped from the factory in perfect working order. To maintain this condition for its service life, the infor­mation contained in these instructions must be observed and followed.
Modifications and repairs to the product may only be performed if expressly permitted by this manual.
Optimum protection of the personnel and the environment and the safe and smooth operation of this product are only ensured by observing all the safety information as well as all safety and warning sym­bols contained in this manual.
Notes and symbols attached directly on the product must be observed. They may not be removed and must be fully legible at all times.
The symbol on the nameplate refers to:
6
1 General information and notes for the reader
?

1.1 Symbols and signal words

DANGER – Serious damage to health / risk to life!
This symbol in connection with the signal word "danger" indicates an imminent threat of danger. Failure to observe the safety information results in death or serious injuries.
WARNING – Injuries to persons!
This symbol in connection with the signal word "warning" indicates a potentially dangerous situation. Failure to observe the safety information may result in death or serious injuries.
CAUTION – Minor injuries!
This symbol in connection with the signal word "caution" indicates a potentially dangerous situation. Fail­ure to observe the safety information may result in minor or moderate injuries. It may also be used for warnings against property damage.
CAUTION – Property damage!
This symbol indicates a potentially harmful situation. Failure to observe the safety information may dam­age or destroy the device and/or other system parts.
NOTE!
This symbol indicates user tips, particularly useful or important information about the device; or its addi­tional uses. This is not a signal word for a dangerous or harmful situation.
NOTE!
This symbol is used in tables and indicates that further information is provided after the table.

1.2 Definition of terms

In this manual, the term "device" always refers to the electromagnetic flowmeter (406011).
In this manual, the terms "medium", "media", "measurement media", "fluid" and "fluids" always refer to liquid or liquids.
7
2 Intended use

2 Intended use

NOTE!
Failure to use this device correctly may pose a danger to people, surrounding machinery, and the envi­ronment.
This device is intended solely for the purpose of measuring the flow rate of liquids.
During application, the admissible data and operating conditions specified in the legal documents and operating manual must be observed.
Protect the device from electromagnetic interference, UV radiation, and the weather (when using outside).
Only ever use the device when it is free from faults.
Proper storage, transport, installation and handling are all essential pre-requisites for safe and cor­rect device operation.
Only ever use the device for its intended purpose.
8
3 Basic safety informati on

3 Basic safety information

This safety information does not account for:
Random events or incidents that may occur during installation, operation, and maintenance of the device
Local safety requirements; the operator is responsible for compliance with any such regulations, in­cluding with regard to installation and maintenance specialists.
Risk of injury resulting from high pressure levels within the plant!
Before working on the system, switch off the pressure and bleed/empty the lines.
Risk of electric shocks!
Restrict the maximum operating voltage to DC 35 V if the unit is used in a humid environment or out­doors.
Before working on the system or device, switch off the voltage and secure it so that it cannot switch on again.
Always observe the relevant accident prevention regulations and safety requirements for electrical de­vices.
Do not touch electronic components when they are live!
Risk of injury due to high fluid temperatures!
Always use safety gloves to touch the device.
Stop circulation of medium and empty the pipelines before disconnecting the process connections.
Risk of injury resulting from the type of medium!
When using dangerous media, always observe the specifications on the safety sheet and the relevant accident prevention regulations.
General dangerous situations
Observe the following to protect against injuries:
Do not use the device in potentially explosive areas.
Do not use the device in an environment that is not compatible with the materials used to make the device.
Do not use any media that are not compatible with the materials used to make the device.
Do not place the device under mechanical strain.
Do not modify the device.
Do not operate the system unsupervised.
All installation and maintenance work must always be performed by authorized and qualified person­nel using suitable tools.
If the power supply is interrupted, the restart process must be defined or controlled.
Always observe general technical rules when planning application of the device and when operating it.
9
3 Basic safety information
NOTE!
The device may be damaged by the medium.
Systematically check the chemical compatibility of the materials used to make the device and the media with which the device comes into contact (for example: alcohols, strong or concentrated acids, alde­hydes, bases, ester, aliphatic compounds, ketone, aromatic or halogenated hydrocarbons, oxidizing agents, and chlorinated substances).
NOTE!
Components/modules at risk of static electricity
The device contains elements that are sensitive to electrostatic discharge. Any such element shall be at risk of damage if it comes into contact with electrostatically charged persons or objects. In the worst case scenario, the component could be destroyed immediately or fail after startup.
Observe the requirements under EN 61340-5-1 to minimize or eliminate potential damage in the event of sudden electrostatic damage.
10
4 General information

4.1 Warranty conditions

Improper use, failure to observe this manual, the use of underqualified personnel, or unauthorized mod­ifications releases the manufacturer from liability for any resulting damage. In these cases, the manufac­turer's warranty no longer applies.

4.2 Information available online

NOTE!
All the documentation, declarations of conformity and certificates are also available in the download area at www.jumo.de.

4 General information

11
5 Design and func tion

5 Design and function

5.1 General information

5.1.1 Design and function

The device consists of a transducer (sensor) and a transmitter with a display. The transducer operates according to Faraday's law of induction.
The flowmeter is equipped with an analog output (current output AO1, 4 to 20 mA), a digital output (tran­sistor output DO1, basic setting: pulse output), and two counters.
It also has two digital outputs (relay outputs DO2 and DO3) and one digital input (DI1).
The device operates as a three-wire system and requires a voltage supply between DC 18 to 36 V.
The electrical connection takes the form of two cable fittings (M20 × 1.5) connected to the electronic board's terminal strips inside the transmitter.

5.1.2 Block diagram

5.2 Areas of application

This flowmeter is intended solely for the purpose of measuring the flow rate of liquids (media or mea­sured media), see chapter 2 "Intended use", page 8.
The adjustable transistor output and the two relay outputs enable the device to switch a solenoid valve or activate an alarm. The current output (4 to 20 mA) allows a control circuit to be created.
The digital input enables a function to be triggered remotely.
12

5.3 Nameplate

5 Design and function
(1) Manufacturer and device designation (2) Product group number (3) Sensor version (4) Part no. (5) Conformity label (6) Current/relay output data (7) Barcode (8) Fabrication number (9) Voltage supply / digital input data (10) Flow direction
13
6 Technical data

6 Technical data

6.1 Measuring range and accuracy

Flow velocity 0.2 to 10 m/s (0,66 to 32,8 ft/s) Measuring deviation
with standard K-factor ≤ ±3.5 % of the measured value
after Teach-In ≤ ±0.5 % of measured value Linearity ≤ ±0.5 % of measuring range limit value Repeatability ≤ ±0.25 % of measured value
a
under reference conditions, i. e. measurement medium = water, ambient and water temperature = 20 °C, observing the necessary inlet and outlet sections and associated fitting (406090)
b
10 m/s (32,8 ft/s)

6.2 Measurement media

Medium type Neutral, contaminated, and aggressive liquids Minimum conductivity of the medi-um20 µS/cm
a
a
b
a
Viscosity < 1000 mPas Medium temperature
a
in conjunction with fitting made of
PVC 0 to 50 °C (32 to 122 °F)
PE 0to70°C (32to158°F)
PP 0to80°C (32to176°F)
Stainless steel -15 to +110 °C (5 to 230 °F) Medium pressure
a
in conjunction with fitting made of
PVC PN 10 (145 psi)
PE PN 10 (145 psi)
PP PN 10 (145 psi)
Stainless steel PN 16 (232 psi)
a
see also "pressure temperature diagram", page 15
14
6 Technical data

6.3 Mechanical features

6.3.1 Materials

Housing/seal PPA, black/NBR Union nut PPA Safety cap/seal PSU/silicone Front film Polyester M20 × 1.5 cable fittings/
seal Screws Stainless steel Components in contact with the
medium sensor Stainless steel 316 L (1.4404), FKM, or EPDM, PEEK Fitting See data sheet 406090

6.3.2 Pressure temperature diagram

Relationship between the medium pressure and temperature for a device with a stainless steel sensor and a 406090 fitting made from stainless steel, PVC, PP, or PE:
PA/neoprene
15
6 Technical data

6.4 Electrical data

Voltage supply DC 18 to 36 V, filtered and regulated Tolerance of the applied voltage ±0,5 % Current consumption Max. 300 mA at DC 18 V
Digital input DI1
Switching voltage DC 18 to 36 V Safety Protected against polarity reversal and voltage peaks, galvanically isolated Input impedance 15 kΩ Minimum pulse length 200 ms Switching thresholds PLC level: logical „0“ < 7 V, logical „1“ > 10 V
Analog output AO1
Type Current output Signal range 4 to 20 mA, sink or source (depending on the connection), 22 mA for reporting
faults
Load resistance
at DC 18 V 450 Ω
at DC 24 V 700 Ω
at DC 30 V 1000 Ω
at DC 36 V 1300 Ω
Digital output DO1
Type Transistor output (open-collector output), NPN/PNP (depending on connec-
tion)
Function (adjustable) Pulse output (default setting), hysteresis/window mode, reverse flow direction
message, generate a warning message Pulse frequency 0 to 250 Hz Switching voltage DC 5 to 36 V Switching current Max. 100 mA Duty cycle at f > 2 Hz 0,5 Minimum pulse length at f < 2 Hz 250 ms Safety Protected against excess voltage, polarity reversal, and short circuiting, gal-
vanically isolated
Digital outputs DO2 and DO3
Type Relay output Function (adjustable) Hysteresis (default setting), hysteresis/window mode, reverse flow direction
message, generate a warning message Switching output N/O contact Switching capacity 3 A at AC 250 V resistive load, 3 A at DC 40 V resistive load Operating life 100000 cycles Galvanic isolation Mixed switching of mains voltage AC 230 V and SELV or PELV voltage is
not admissible due to the basic insulation between the relays!
16

6.5 Connecting cable

Connection type via two cable fittings M20 × 1.5 Cable data Cable type Shielded Cross-section 0.5 to 1.5 mm Cable diameter When using one cable per cable fit-
ting When using two cables per cable
fitting
6to12mm
4 mm, with supplied multiway seal
2

6.6 Environmental influences

Wire diameter DN 15 to DN 400 (1/2 to 16″) Fitting Type 406090 Inlet and outlet sections see "Inlet and outlet sections ", page 21 Operating temperature range -10 to +60 °C (14 to 140 °F) Storage temperature range -20 to +60 °C (-4 to +140 °F) Relative humidity < 85 %, non-condensing
Height above sea level: max. 2000 m (6562 ft) Protection type according to
EN 60529
EMC EN 61000-6-3, EN 61000-6-2 Safety EN 61010-1
IP65,
With connected device, screwed cable glands and screwed flap cover
6 Technical data

6.7 Compliance with standards and directives

The standards applied to verify conformity with EU directives are listed in the EU type examination cer­tificate and/or the EU declaration of conformity (if applicable).
Pressure:
According to Article 4 (1) of the Pressure Equipment Directive 2014/68/EU, the device may only be ap­plied under the following conditions (depending on the maximum pressure and the DN of the piping used for the medium):
Type of liquid (medium) Requirements
Group 1 fluids, Article 4 (1) c.i. Prohibited Group 2 fluids, Article 4 (1) c.i. ≤ DN 32
Or > DN 32 and PN × DN ≤ 1000 Group 1 fluids,
Article 4 (1) c.ii.
Group 2 fluids, Article 4 (1) c.ii.
≤DN25
Or PN × DN ≤ 2000
DN ≤ DN 200
Or ≤ PN 10
Or PN × DN ≤ 5000
17
7 Mounting

7 Mounting

7.1 Safety information

Risk of injury resulting from high pressure levels within the system!
Before working on the system, switch off the pressure and bleed/empty the lines.
Risk of electric shocks!
Restrict the maximum operating voltage to DC 35 V if the unit is used in a humid environment or out­doors.
Before working on the system or device, switch off the voltage and secure it so that it cannot switch on again.
Always observe the relevant accident prevention regulations and safety requirements for electrical de­vices.
Risk of injury due to high fluid temperatures!
Always use safety gloves to touch the device.
Stop medium circulation and empty the pipelines before disconnecting the process connections.
Do not touch electronic components when they are live!
Risk of injury resulting from the type of medium!
When using dangerous media, always observe the specifications on the safety sheet and the relevant accident prevention regulations.
Risk of injury due to improper installation!
Mounting and electrical installation processes must always be carried out by authorized and qualified personnel using suitable tools.
Refer to the operating manual for the fitting used (406090).
Install a circuit breaker or disconnect switch into the electronics circuit in the building in which the device is going to be used.
Install the circuit breaker or disconnect switch in an easily accessible location.
Mark the circuit breaker or disconnect switch as a control device for the device's power supply.
Always use suitable safety equipment (fuses and/or circuit breakers in the correct size).
Observe standard NF C 15-100/IEC 60364.
Risk of injury resulting from the system switching on unintentionally and restarting without su­pervision!
Secure the system so that it cannot be operated unintentionally.
After conducting any work on the device, always make sure that a controlled restart is carried out.
18
Risk of injury resulting from a failure to comply with the relationship between the pressure and temperature of the measurement medium!
Always observe the relationship between the temperature and pressure of the medium depending on the material of the fitting (406090).
See chapter 6 "Technical data", page 14 and the operating manual for the fitting used (406090).
Observe the Pressure Equipment Directive 2014/68/EU.
7 Mounting
Risk of injury due to improper startup!
Failing to operate the device correctly can lead to injury and damage to the device and its surroundings.
Prior to startup, make sure that the operating staff are familiar with and understand the contents of the operating manual.
Any safety information and notes on proper use are particularly important.
The device/system must always be started up by staff with a sufficient level of training.
NOTE!
Protect the device from electromagnetic interference, UV radiation, and the weather (when using out­side).
19
7 Mounting

7.2 Installation into the pipeline

Risk of injury resulting from high pressure levels within the system!
Before working on the system, switch off the pressure and bleed/empty the lines.
Risk of injury resulting from the type of medium!
When using dangerous media, always observe the specifications on the safety sheet and the relevant accident prevention regulations.
The flowmeter is inserted into a fitting (406090) mounted on the pipeline.

7.2.1 Recommendation for installing into the pipeline

Flow velocity requirements
Select the fitting 406090 according to the flow velocity requirements in the pipeline.
Example A in the diagram:
Requirement: If the flow rate is 10 m3/h, the flow veloc­ity should ideally be between 2 and 3 m/s.
Solution:
Select a pipe with DN 40.
20
Installation
50xDN
5xDN
20xDN
5xDN
40xDN
5xDN
18xDN
5xDN
25xDN
5xDN
15xDN
5xDN
7 Mounting
The device can be installed in either a horizontal or ver­tical pipeline.
Make sure that:
The section of pipe around the sensor is always filled.
The direction of flow (arrow) points upwards when installing vertically.
The formation of bubbles is avoided in the pipeline containing the device.
The device is always installed upstream from dis­charge points for liquids with increased conductivity (e.g.: acids, bases, salt solutions).
Inlet and outlet sections
To calm the flow in the pipeline, the specified inlet and outlet sections are required at the very least.
For increased accuracy, you can make these sections of contact longer.
Direction of flow: from left to right.
DN = Nominal pipe width
Correct Incorrect
Control valve 1× 90-° manifold
or T-piece
45° angle installation position
We recommend installing the flow rate transmitter at a 45° to the pipe's central horizontal axis. This helps to avoid the majority of deposits on the mea­suring electrodes and measuring errors caused by air bubbles.
2× 90-° manifolds
Pipe extension
three-dimensional
2× 90-° manifolds Pipe reduction
21
7 Mounting

7.2.2 Installation in a pipeline with G 2" union nut

NOTE!
Please refer to the recommended installation type!
See chapter 7.2 "Installation into the pipeline", page 20 and operating manual "Fittings for flow rate sensors" (406090)
NOTE!
To ensure more accurate measurements and good stability of the flow rate zero point, bring the sensor into contact with the measurement medium at least 24 hours prior to calibration.
(1) Flow direction (2) Measuring electrodes (3) Circlip (4) Union nut (5) Groove (6) Fitting 406090 (7) Seal (8) Device
1. Install the fitting 406090 into the pipeline, see chapter 7.2.1 "Recommendation for installing into the pipeline", page 20.
2. Check that the seal (7) is on the device (8).
3. Place the union nut (4) on the fitting (6).
4. Place the circlip (3) into the groove (5).
5. Position the device so that the arrow on the side of the housing is pointing in the direction of flow (1): the counters are incremented.
6. Place the device (8) into the fitting (6).
7. Manually tighten the union nut (4) on the device (8).
22

7.3 Electrical connection

Risk of electric shocks!
Restrict the maximum operating voltage to DC 35 V if the unit is used in a humid environment or out­doors.
Before working on the system or device, switch off the voltage and secure it so that it cannot switch on again.
Always observe the relevant accident prevention regulations and safety requirements for electrical de­vices.
NOTE!
The device may leak if you do not use at least one of the cable fittings.
1. Insert the provided plugs into the cable fittings not in use.
2. Unscrew any cable fittings not in use.
3. Remove the transparent pane.
4. Insert the plug.
5. Tighten the cable fitting's union nut.
7 Mounting
NOTE!
Use a high-quality power supply (filtered and controlled).
Protect the potential equalization of the installation, see chapter 7.3.1 "Make sure that the potential is equal in the installation", page 24.
Use a shielded cable with an operating temperature limit above 80 °C.
Avoid laying the cable near the high-voltage or high-frequency cables; if you cannot avoid doing so, maintain a minimum gap of 30 cm.
Protect the power supply with a fuse (300 mA) and circuit breaker.
Protect the relay with a fuse (max. 3 A) and an automatic circuit break (suited to the process).
Do not connect a dangerous voltage and a protective low voltage supply to the relay at the same time.
NOTE!
If you are using two cables in one cable fitting, insert the multi-way seal into the cable fitting first.
Connect the device as follows:
1. Unscrew the screw on the flap and open the flap.
2. Loosen the four screws on the housing lid.
3. Set the lid to one side.
4. Screw on the cable fittings.
5. Guide the cable through the union nut through the cable fitting.
6. Check whether the grounding cable (from the housing) is connected as shown in the figure in chapter 7.3.3 "Terminal assignment and use of a selection switch", page 26.
7. Lay the cable according to chapter 7.3.1 "Make sure that the potential is equal in the installation", page 24 to chapter 7.3.7 "Connecting the relay outputs DO2 and DO3", page 30.
23
7 Mounting

7.3.1 Make sure that the potential is equal in the installation

Measures for making sure the potential is equal in the installation (voltage supply  device  measure- ment medium):
Connect the various grounding points in the installation to make sure any difference in potential be­tween two grounding points is eliminated.
Make sure both ends of the supply cable's shielding are grounded correctly.
Connect the negative terminal on the voltage supply to the ground to suppress the effects of common mode currents. If you cannot connect the two elements directly, you can switch a capacitor with 100 nF/50 V between the voltage supply's negative terminal and the ground.
If the device is installed on plastic pipes, please pay particular attention to the following: direct grounding is not possible in this case. For correct grounding, connect all metallic objects close to the device (valves or pumps) to the same grounding point. If there are no metallic objects near the de­vice, insert grounding rings into the device in both the flow direction and against the flow direction; connect these rings with the same ground. The grounding rings must be in contact with the measure­ment medium.
Wiring diagrams for potential equalization in plastic pipelines
24
(1) Voltage supply (2) If direct grounding is not possible, connect a capacitor with 100 nF/50 V between the voltage sup-
ply's negative terminal and the ground.
(3) Pumps, motors, valves
Alternative: Grounding rings inserted in the pipeline (not included in the scope of delivery) (4) Plastic pipeline (5) Shielding on the supply cable
Wiring diagrams for potential equalization in metal pipelines
(1) Voltage supply (2) If direct grounding is not possible, connect a capacitor with 100 nF/50 V between the voltage sup-
ply's negative terminal and the ground. (3) Metal pipeline (4) Shielding on the supply cable
7 Mounting

7.3.2 Use of a cable holder

NOTE!
Before connecting the device, insert the supplied cable holder onto the electronics board.
(1) Cable holder (included in the scope of delivery)
25
7 Mounting

7.3.3 Terminal assignment and use of a selection switch

HINWEIS!
Galvanic isolation of relay outputs DO2 and DO3
Mixed switching of mains voltage AC 230 V and SELV or PELV voltage is not admissible due to the basic insulation between the relays!
(1) Terminal strip I out Current output AO1 (4 to 20 mA)
L+ V+ (positive power supply) L- 0 V (ground for the power supply) PE Functional ground, wired at the plant ( see "Terminal strip 1: Connecting
the grounding cable ", page 27) P+ Positive transistor output DO1 P- Negative transistor output DO1
(2) Terminal strip PE Shielding on the supply cable and output cables of AO1 and DO1 (3) Terminal strip PE Functional ground for the cable for the DI1 digital input
- Negative signal for the digital input DI1 + Positive signal for the digital input DI1
(4) Terminal strip Connection for the relay output DO2 (5) Terminal strip Connection for the relay output DO3 (6) Terminal strip 4-pin ribbon cable for connecting the sensor (A) Switch see "Use of the sink selection switch (right) or source selection switch
(left) ", page 27
(B) Switch see "Use of the selection switch to lock or unlock the CONFIRM key ", page 27
26
Use of the sink selection switch (right) or source selection switch (left)
OFF ON
B
OFF ON
A
7 Mounting
CURRENT
SOURCESINK
A Selection switch A is used to configure the 4 to 20 mA output as a source or sink.
Use of the selection switch to lock or unlock the CONFIRM key
B Selection switch B is used to lock or unlock the key in order to avoid unauthorized configu-
ration of the device.
Terminal strip 1: Connecting the grounding cable
CURRENT
SOURCESINK
(1) Grounding cable (from the housing) (2) Grounding cable (from the sensor)
27
7 Mounting
SOURCESINK
L+ L-PEP+P-Iout
Supply
18...36 Vdc
PULSE
DO1
CURRENT
PE
DI1
PE
DO2
DO3
AO1
+
-
+
-
a
SOURCESINK
L+ L-PEP+P-Iout
Supply
18...36 Vdc
PULSE
DO1
CURRENT
PE
DI1
PE
DO2
DO3
AO1
+
-
+
-
a

7.3.4 Connecting the digital input DI1

Possible connection options for the digital input DI1
28
a If direct grounding is not possible, connect a capacitor with 100 nF/50 V between the voltage sup-
ply's negative terminal and the ground.

7.3.5 Connection for the current output AO1

SOURCESINK
L+ L-PEP+P-Iout
Supply
18...36 Vdc
PULSE
DO1
CURRENT
PE
AO1
A
+
-
300 mA
I
+
-
a
SOURCESINK
L+ L-PEP+P-Iout
Supply
18...36 Vdc
PULSE
DO1
CURRENT
PE
AO1
A
+
-
300 mA
I
+
-
a
NOTE!
For safety reasons, secure the cable with a conductive cable holder.
The current output AO1 (4 to 20 mA) can be connected as a source or a sink.
NOTE!
Adjust switch A to "SOURCE".
Connecting the current output AO1 (4 to 20 mA) as a source
7 Mounting
a If direct grounding is not possible, connect a capacitor with 100 nF/50 V between the voltage sup-
ply's negative terminal and the ground.
NOTE!
Adjust switch A to "SINK".
Connecting the current output AO1 (4 to 20 mA) as a sink
a If direct grounding is not possible, connect a capacitor with 100 nF/50 V between the voltage sup-
ply's negative terminal and the ground.
29
7 Mounting
L+ L-PEP+P-Iout
Supply
18...36 Vdc
PULSE
DO1
PE
AO1
300 mA
a
+
-
+
-
+
-
L+ L-PEP+P-Iout
Supply
18...36 Vdc
PULSE
DO1
PE
AO1
300 mA
+
-
a
+
-
+
-
DO2
DO3
AC 230 V
m
3A
3A
AC 230 V

7.3.6 Connecting the transistor output DO1

NPN connection for the transistor output DO1
a If direct grounding is not possible, connect a capacitor with 100 nF/50 V between the voltage sup-
ply's negative terminal and the ground.
PNP connection for the transistor output DO1
a If direct grounding is not possible, connect a capacitor with 100 nF/50 V between the voltage sup-
ply's negative terminal and the ground.

7.3.7 Connecting the relay outputs DO2 and DO3

Possible connection options for the relay outputs DO2 and DO3
30
8 Startup and paramet erizatio n

8.1 Safety information

Risk of injury due to improper operation!
Failing to operate the device correctly can lead to injury, and damage to the device and its surroundings.
The device/system must always be operated by staff with a sufficient level of training.
Make sure that the operating staff are familiar with and understand the contents of the operating manual.
Any safety information and notes on proper use are particularly important.
NOTE!
Before putting the sensor into operation for the first time, immerse it into the measuring material for 24 hours.

8 Startup and parameterization

31
8 Startup and parameterization

8.2 Operating levels

The device's operating levels consist of a process level and configuration level.
Level Functions
Process level Extracting the flow rate measured by the device, the current value
emitted at the 4 to 20 mA current output, and the values for the main counter and daily counter
Extracting the daily counter Switching to the configuration level
Configuration level with parameterization, testing, and information menu
Function Basic setting
LANGUAGE English UNIT of the flow rate l/min. UNIT for the counter Liters K-FACTOR 1000 OUTPUT AO1 4 mA = 0.000
OUTPUT DO1 Pulse
OUTPUT DO2 Hysteresis
OUTPUT DO3 Hysteresis
INPUT DI1 Inactive FILTER 5, slow FREQUENC. 50 Hz CUT-OFF 0.000 BACKLIT Filter 9, activation period: 30 s K-SENSOR KW = 1.000 FLOW-W. W- = 0.000
Adjusting the device's parameters Testing certain device parameters Calibrating the device Extracting any warnings or error messages generated by the de-
vice if the device status LED is orange or red
20 mA = 0.000
PU = 0.00 liters
2- = 0.000 2+ = 0.000 Not inverted Delay = 0
3- = 0.000 3+ = 0.000 Not inverted Delay = 0
W+ = 0.000
32
8 Startup and parameterization
++
+
Process level Configuration level
12.6 L/S
>5s >2s
16.45 mA
87654 L
231 L.
> 2 s K-FACTOR FLOW NEG.FLOW
To reset the daily counter (identified by a dot after the volume units)
> 5 s Parameters
menu
LANGUAGE CAL AO1 MEAS.OVF
UNIT CALIB 0 CAL.FAIL
OUTPUT K-SENSOR
INPUT FLOW-W.
FILTER END END
TOTAL
FREQUENC.
CUT-OFF
BACKLIT
Test menu Information
menu
b
a
. . .
or
END SAVE N/Y Process level
a
Only accessible when the device status is orange or red ( see chapter 8.3 "Display, navigation keys and status LEDs", page 34).
b
When the CONFIRM key is unlocked
33
8 Startup and parameterization

8.3 Display, navigation keys and status LEDs

(1) Digital display with eight digits (four numeric, four alphanumeric) (2) CONFIRM key: To select the displayed function, to confirm settings (3) Status LED for the relay DO3 (LED on = Contact closed) (4) Status LED for the relay DO2 (LED on = Contact closed) (5) To read messages, to scroll down through the functions, to select the digit on the left (6) Device status LED see table below (7) To scroll up through the functions, to increase the selected digit
Device status LED Device status
Green The device is working without any faults. Orange A warning message has been generated.
Press the key in the process level for two seconds to read the message ( see chapter 9.5.5 "Solving a problem with a warning or error message with the orange device status LED", page 80).
Otherwise, the relay output DO2 or DO3 or the transistor output DO1 will switch if it is configured for "WARNING" mode ( see chapter 8.6.10 "Configuring the transistor output DO1 to switch a load when the device generates a warning", page 51 or "Config­uring the relay outputs DO2 and DO3 to switch a load when the device generates a warning ", page 54).
Red An error message has been generated and the current output
emits 22 mA. Press the key in the process level for two seconds to read the
message ( see chapter 9.5.4 "Solving a problem with a warning or error message with the red device status LED", page 79).
Flashing, regardless of the color
The digital input DI1 is active or The checking function for proper output response is active ( see
chapter 8.7.3 "Checking proper response of the outputs", page 70) or
Calibration of the zero flow rate is active ( see chapter 8.7.2 "Calibrate the flow rate zero point", page 67) or
The daily counter is locked at zero.
34

8.4 Using the navigation keys

You can... Button functions
Move between the functions in a level or menu
Display the parameters menu + at the same time for 5 seconds in the process level Display the test menu + + at the same time for 5 seconds in the process level Display the information menu for 2 seconds in the process level if the device status LED is
Reset the daily counter + at the same time for 2 seconds if the daily counter is dis-
Select the displayed function Confirm the displayed value Change a numerical value to increase the selected digit
Next function:
Previous function:
orange or red
played in the process level
8 Startup and parameterization
to select the previous digit + to move the comma
35
8 Startup and parameterization
+

8.5 Process level details

This level is active when the device is switched on.
12.6 L/S or 0 L/S. A dot after the flow unit indicates that the CUT-OFF
Flow rate value, selecting and adjusting the unit, see chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
16.45 mA Value of the current output, proportional to the measured flow rate
87654 L Value of the main counter, volume of fluid counted by the device since the
last reset.
function is active. The displayed flow rate is set to 0 because the mea­sured flow rate is below the threshold value defined in the CUT-OFF parameter of the parameters menu, see chapter 8.6.16 "Adjusting the CUT­OFF flow rate", page 63.
231 L. Value of the daily counter (identified by a dot after the volume units), volume
of fluid counted by the device since the last reset.
Resetting the daily counter
>2s
36

8.6 Parameter menu details

1. In order to access the parameters menu, press the + keys at the same time for more than 5 seconds.
This menu allows you to adjust the following device parameters:
LANGUAGE Choosing the display language
UNIT Choosing the flow rate unit, the number of decimals, and the unit the count-
ers are displayed in
K-FACTOR Entering the K-factor for the fitting (406090) used or defining it using Teach-
In calibration
OUTPUT Parameterizing the current output AO1 (4 to 20 mA) and configuring the
transistor output DO1 and both relay outputs DO2 and DO3
INPUT Configuring the digital input DI1
8 Startup and parameterization
FILTER Choosing the filter level of the measured flow rate, which affects the dis-
played flow rate and the current output AO1.
TOTAL Resetting both counters
FRE-
QUENC.
CUT-OFF Entering the measured flow rate value below which the device sets the
BACKLIT Setting the brightness of the display and how long it stays ON, or deacti-
END SAVE N/Y 12.6 L/S
Adjusting the mains frequency
measured flow rate to 0
vating the backlight
Process level
Saving the changes made within the parameters menu or not. If the changes are saved, the device operates with the new settings
37
8 Startup and parameterization

8.6.1 Choosing the display language.

The menu is in English when switched on.
LANGUAGE ENGLISH
DEUTSCH
FRANCAIS 1. Confirm the displayed language.
The selected language is immediately ac-
ITALIANO
ESPANOL
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.

8.6.2 Choosing the flow rate unit, the number of decimals, and the counter display unit

tive.
UNIT
NOTE!
Only the counters are automatically converted when the flow rate unit is adjusted.
Convert any other flow rate settings manually if necessary.
NOTE!
The maximum flow value that can be displayed depends on the decimal places selected:
9999 if decimal places = 0 or AUTO
999.9 if decimal places = 1
99.99 if decimal places = 2
9.999 if decimal places = 3
The UNIT function enables you to select the following:
Flow rate unit
Fixed point (selection 0, 1, 2 or 3) to display the flow rate value in the process level or a sliding point (selection AUTO): In this case, the device selects the position of the decimal point depending on the unit selected and the flow rate value measured
Counters' volume unit if the flow rate unit selected above is liters or m
3
38
8 Startup and parameterization
+
UNIT FLOW LIT/SEC 1. Choose the flow rate unit
2. Confirm
LIT/MIN
LIT/H
M3/MIN
M3/H
US GAL/S AUTO 1. Choose the
number of decimal places
US GAL/M DEC PT 3
US GAL/H DEC PT 2
IMP GA/S DEC PT 1
IMP GA/M DEC PT 0
IMP GA/H
If the selected unit is in liters or m
TOTAL LITERS 1. Choose the counter
M3
RETURN
3
unit
2. Confirm
2. Confirm
If
the se-
lected
unit
is gallons
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
39
8 Startup and parameterization

8.6.3 Entering the K-factor for the fitting (406090) used

The device calculates the medium's flow rate in the pipeline with the K-factor for the fitting (406090) used.
You can enter the K-factor for the fitting used here. The device can also determine the K-factor in a cal­ibration procedure (Teach-In), s see chapter 8.6.4 "Determining the fitting's K-factor with a calibration procedure (Teach-In)", page 41.
NOTE!
The device starts using the new K-factor as soon as the function SAVE YES is confirmed when leaving the parameters menu.
K-FACTOR K = 10.000 K = 2.8500
The display shows the K-factor for the fitting (406090) used.
1. Confirm the parameter.
OUTPUT RETURN VALID N/Y
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
1. Adjust the K-factor for the fitting (406090) used (value between
0.0001 and 9999.9). Changing a numerical value, see chapter 8.4 "Using the naviga­tion keys", page 35.
2. Confirm the displayed value.
Displayed only if the K-factor has been changed.
1. Confirm the entered K-factor or not.
40
8 Startup and parameterization

8.6.4 Determining the fitting's K-factor with a calibration procedure (Teach-In)

NOTE!
Prior to Teach-In:
Calibrate the flow rate zero point ( see chapter 8.7.2 "Calibrate the flow rate zero point", page 67).
Check that the KW value for the sensor has not been changed ( see chapter 8.7.4 "Changing the sen­sor's KW value", page 70).
The device calculates the medium's flow rate in the pipeline with the K-factor for the fitting used.
The function TEACH V. or TEACH F. is a calibration process (Teach-In) for determining the fitting's K­factor. The K-factor can also be entered, see chapter 8.6.3 "Entering the K-factor for the fitting (406090) used", page 40.
The calibration procedure is applied either in relation to a known volume (TEACH V.) or in relation to the current flow rate (TEACH F.), as measured with a reference instrument.
41
8 Startup and parameterization
Determining the fitting's K-factor with a Teach-In process in relation to a volume (TEACH V.)
NOTE!
The device starts using the new K-factor as soon as the function SAVE YES is confirmed when leaving the parameters menu.
1. Prepare a container with a known volume.
2. Stop medium circulation.
3. Confirm TEACH V.: FILL. END is displayed.
K-FACTOR K = 2.8500 The display shows the K-factor for the fitting (406090)
used, whether entered or determined using Teach-In.
TEACH V. FILL. END
1. Charge the pipe to fill the tank.
2. When the tank is full, confirm FILL. END.
0000.0 L
TEACH F. 1. Enter the volume (value between 0.1 and
9999.9) of medium that flowed past the device and confirm.
K = 2.9000
The device calculates the K-factor of the fitting and displays it.
1. Confirm the displayed value.
OUTPUT RETURN VALID N/Y
1. Confirm the K-factor determined through Teach­In or not.
42
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
8 Startup and parameterization
Determining the fitting's K-factor with a Teach-In process in relation to a flow rate (TEACH F.)
NOTE!
The device starts using the new K-factor as soon as the function SAVE YES is confirmed when leaving the parameters menu.
1. Charge the pipe.
2. Wait for the flow rate to stabilize.
3. Confirm TEACH F.: MEASURE \ is displayed.
K-FACTOR K = 10.000 The display shows the K-factor for the fitting (406090)
used, whether entered or determined using Teach-In.
TEACH V.
TEACH F. MEASURE \
The device calculates the mean flow rate in the pipe over a period of approximately 50 s.
0.000 L/S
1. Enter the value of the flow rate in the pipe (value between 0.001 and 9999) and confirm.
The device calculates the fitting's K­factor and displays it.
K = 2.8500
1. Confirm the displayed value.
OUTPUT RETURN VALID N/Y
1. Confirm the K-factor determined through Teach­In or not.
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
43
8 Startup and parameterization

8.6.5 Configuring outputs

OUTPUT AO1 see chapter 8.6.6 "Configuring current output AO1", page 45.
DO1 PULSE see chapter 8.6.7 "Configuring the transistor
output DO1 as a pulse output", page 46
HYSTERES. see chapter 8.6.8 "Configuring the transistor
output DO1 to switch a load depending on two
WINDOW
DIRECTIO. see chapter 8.6.9 "Configuring the transistor
WARNING see chapter 8.6.10 "Configuring the transis-
threshold values", page 47
output DO1 to switch a load when the direction of flow changes", page 50
tor output DO1 to switch a load when the device generates a warning", page 51
DO2 HYSTERES. see "Configuring the relay outputs DO2 and DO3
WINDOW
DIRECTIO. see "Configuring the relay outputs DO2 and
WARNING see "Configuring the relay outputs DO2 and
RETURN
DO3 for switching a load depending on two threshold values ", page 52
DO3 to switch a load when the direction of flow changes ", page 53
DO3 to switch a load when the device generates a warning ", page 54
44

8.6.6 Configuring current output AO1

NOTE!
The current output AO1 emits a 22 mA-signal when the device emits an error, even if the output is de­activated.
The current output AO1 (4 to 20 mA) emits a signal, the value of which represents the flow rate mea­sured by the device.
Example:
Current output signal (4 to 20 mA) in relation to the measuring range:
1. To invert the output signal, enter a flow rate value lower than the 4 mA value for the 20 mA value.
2. To deactivate the output, set the 4 mA and 20 mA limits to zero. In this case, the current output emits a constant signal of 4 mA.
8 Startup and parameterization
NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
OUTPUT AO1 4 = 28.00
1. Enter a flow rate that corresponds to a 4mA current.
2. Confirm
DO1 20 = 6,000
1. Enter a flow rate that corresponds to a 20 mA current.
2. Confirm
45
8 Startup and parameterization
+

8.6.7 Configuring the transistor output DO1 as a pulse output

If the transistor output DO1 is configured as a pulse output, a pulse is generated at this output each time a selected medium volume is reached.
NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
Is only displayed if the unit chosen in the UNIT parameter is in liters or m
OUTPUT DO1 PULSE LITERS
M3
DO2 PU = 01.000
1. Enter the volume of medium (value be­tween 0.000 and 9999.9) at which a pulse is transmitted at the transistor output. If the unit chosen is in gallons, enter a pulse in gallons ( see 8.6.2).
2. Deactivating the function: Set PU = 0. The output is then always open or no pulse is emitted.
3. Confirm
3
1. Choose the volume unit of a pulse.
.
46
8 Startup and parameterization
8.6.8 Configuring the transistor output DO1 to switch a load depending on two thresh­old values
NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
OUTPUT DO1 HYSTERES. 1- = 0.000
WINDOW 1. Enter a flow rate value
rate value corresponds to the low­er threshold.
2. Confirm
1+ = 0.000
1. Enter a flow rate value rate value corresponds to the low­er threshold.
2. Confirm
INV YES
INV NO
1. Choose the operating mode (in­verted or not inverted) for the tran­sistor output.
a
The flow
a
. The flow
DO2 DEL. 1 = 00
In versions WITH relays 1. Enter the time delay before the
switching point (value between 0 and 99 s).
2. Confirm
RETURN
In versions WITHOUT relays
a
In hysteresis mode: Value for 2- ≤ 2+ or 3- ≤ 3+. In window mode: Value for 2- < 2+ or 3- < 3+.
47
8 Startup and parameterization
ON
OFF
X- X+
ON
OFF
X- X+
ON
OFF
X- X+
ON
OFF
X- X+
Hysteresis mode
The output switches when a threshold is reached:
When the flow rate increases, the output status changes when the high threshold X+ is reached.
When the flow rate decreases, the output status changes when the low threshold X- is reached.
Not inverted Inverted Contact Contact
Flow Flow
X- = Low switching threshold of the transistor or the relay (1-, 2- or 3-) X+ = High switching threshold of the transistor or the relay (1+, 2+ or 3+)
Window mode
The status changes as soon as any threshold (X- or X+) is reached.
Not inverted Inverted Contact Contact
Flow Flow
X- = Low switching threshold of the transistor or the relay (1-, 2- or 3-) X+ = High switching threshold of the transistor or the relay (1+, 2+ or 3+)
48
8 Startup and parameterization
Example of operating the transistor output DO1 with switching thresholds
Flow
High switching threshold
Low switching threshold
Hysteresis mode
DEL. = 0 s Inverted
DEL. = 2 s Inverted
Window mode
DEL. = 0 s Inverted
DEL. = 2 s Inverted
Not inverted
Not inverted
Not inverted
Not inverted
49
8 Startup and parameterization

8.6.9 Configuring the transistor output DO1 to switch a load when the direction of flow changes

NOTE!
If the measured flow rate is in the range of the CUT-OFF flow rate ( see chapter 8.6.16 "Adjusting the CUT-OFF flow rate", page 63), the flow rate is regarded as zero and positive. The following diagram shows how the DO output responds when it is configured to report a change in flow direction and when the CUT-OFF function is used.
Flow
+CUTOFF
0
-CUTOFF
DO output active
DO output inactive
As long as the measured flow rate is lower than -CUTOFF, the device status LED is orange and the message NEG. FLOW is generated.
(1) Measured flow rate (2) Displayed flow rate
OUTPUT DO1 DIRECTIO. INV YES
INV NO
1. Choose the operating mode for the relay output (inverted or not in­verted).
DO2 DEL. 2 = 00
1. Enter the time delay before the switching point (value between 0 and 99 s).
50
2. Confirm
8 Startup and parameterization
8.6.10 Configuring the transistor output DO1 to switch a load when the device gener­ates a warning
When the device generates a warning message, the device status LED is orange.
The generation of a warning message can also be indicated by the switching of the transistor output DO1.
NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
OUTPUT DO1 WARNING INV YES
INV NO
1. Choose the operating mode for the transistor (inverted or not in­verted).
DO2 DEL. 1 = 00
In versions WITH relays 1. Enter the time delay before the
switching point (value between 0 and 99 s).
2. Confirm
RETURN
In versions WITHOUT relays
51
8 Startup and parameterization

8.6.11 Configuring the relay outputs DO2 and DO3

NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
Configuring the relay outputs DO2 and DO3 for switching a load depending on two threshold values
OUTPUT DO2 HYSTERES. 2- = 0.000
WINDOW 1. Enter a flow rate value
rate value corresponds to the low­er threshold.
2. Confirm
2+ = 0.000
1. Enter a flow rate value rate value corresponds to the up­per threshold.
2. Confirm
INV YES
INV NO
1. Choose the operating mode for the relay output (inverted or not in­verted).
a
The flow
a
. The flow
52
DO3 DEL. 2 = 00
1. Enter the time delay before the switching point (value between 0 and 99 s).
2. Confirm
a
In hysteresis mode: Value for 2- ≤ 2+ or 3- ≤ 3+. In window mode: Value for 2- < 2+ or 3- < 3+.
NOTE!
See also chapter 8.6.8 "Configuring the transistor output DO1 to switch a load depending on two threshold values", page 47
8 Startup and parameterization
Configuring the relay outputs DO2 and DO3 to switch a load when the direction of flow changes
NOTE!
If the measured flow rate is in the range of the CUT-OFF flow rate ( see chapter 8.6.16 "Adjusting the CUT-OFF flow rate", page 63), the flow rate is regarded as zero and positive. The following diagram shows how the DO output responds when it is configured to report a change in flow direction and when the CUT-OFF function is used.
Flow
+CUTOFF
0
-CUTOFF
DO output active
DO output inactive
As long as the measured flow rate is lower than -CUTOFF, the device status LED is orange and the message NEG. FLOW is generated.
(1) Measured flow rate (2) Displayed flow rate
OUTPUT DO2 DIRECTIO. INV YES
INV NO
1. Choose the operating mode for the relay output (inverted or not in­verted).
DO3 DEL. 2 = 00
1. Enter the time delay before the switching point (value between 0 and 99 s).
2. Confirm
53
8 Startup and parameterization
Configuring the relay outputs DO2 and DO3 to switch a load when the device generates a warning
When the device generates a warning message, the device status LED is orange.
The generation of a warning message can also be indicated by the switching of the transistor out­put.
OUTPUT DO2 WARNING INV YES
INV NO
1. Choose the operating mode for the relay output (inverted or not in­verted).
DO3 DEL. 2 = 00
1. Enter the time delay before the switching point (value between 0 and 99 s).
2. Confirm
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
54

8.6.12 Configuring the digital input DI1

The digital input DI1 allows one of the four following functions to be triggered remotely:
INPUT DI1 DISABLE The digital input is not used with
8 Startup and parameterization
this option.
CALIB.0 Configure the digital input to re-
motely trigger the calibration of the flow zero point. The calibration can also be performed in the test menu, see "Configuring the dig­ital input DI1 to trigger flow rate zero point calibration ", page 56.
HOLD Configuring the digital input to
freeze the flow rate measure­ments,  see "Configuring the dig- ital input DI1 to trigger calibration for the device's HOLD mode ", page 57.
RETURN
SET FLOW Configuring the digital input to con-
trol the status of the outputs and the display depending on a flow rate replacement value, see "Configuring the digital input DI1 to switch from a measured value to a replacement value ", page 58.
HOLD TOT. Configuring the digital input to
freeze the counters, see "Con­figuring the digital input DI1 to freeze the counters ", page 59.
RES.TOT Configuring the digital input to re-
set the daily counter, see "Con­figuring the digital input DI1 to reset the daily counter ", page 60.
55
8 Startup and parameterization
Configuring the digital input DI1 to trigger flow rate zero point calibration
NOTE!
Calibrating the flow rate zero point, see chapter 8.7.2 "Calibrate the flow rate zero point", page 67
INPUT DI1 CALIB 0 INV YES
1. Choose INV NO to trigger the cal-
2. Choose INV YES to trigger the
3. Confirm
INV NO
ibration of the zero flow point on the leading edge.
calibration of the zero flow point on the trailing edge.
RETURN
56
8 Startup and parameterization
Configuring the digital input DI1 to trigger calibration for the device's HOLD mode
NOTE!
Triggering of HOLD mode is ignored if the checking function for proper output response is active ( see chapter 8.7.3 "Checking proper response of the outputs", page 70).
The HOLD mode is used to carry out maintenance work without interrupting the process.
In practice, when the device is in HOLD mode:
The device status LED flashes.
The current transmitted at the 4 to 20 mA output is frozen to the value of the last measured flow rate.
The displayed flow rate is frozen to the value of the last measured flow rate.
Each relay or transistor output is frozen to the status it had when Hold mode was activated.
The counters do not increment any more (frozen).
The device remains in Hold mode until the digital input switches again.
INPUT DI1 HOLD INV YES
INV NO
1. Choose INV NO to trigger the in­put on the leading edge.
2. Choose INV YES to trigger the in­put in the trailing edge.
3. Confirm
RETURN
57
8 Startup and parameterization
Configuring the digital input DI1 to switch from a measured value to a replacement value
NOTE!
The replacement value is ignored if the checking function for proper output response is active ( see chapter 8.7.3 "Checking proper response of the outputs", page 70).
NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
This function enables a replacement value specified by the user to be emitted instead of the mea­sured value.
When the digital input is activated:
The device status LED flashes.
The signal transmitted at the 4 to 20 mA output is set to the replacement value.
The displayed flow rate is set to the replacement value.
Each transistor or relay output adopts the status depending on the replacement value.
The counters do not increment any more.
The replacement flow rate is active until the digital input switches again.
INPUT DI1 SET FLOW SF = 0.000
1. Enter a flow rate value.
2. Confirm
INV YES
INV NO
1. Choose INV NO to trigger the in­put on the leading edge.
2. Choose INV YES to trigger the in­put in the trailing edge.
3. Confirm
RETURN
58
8 Startup and parameterization
Configuring the digital input DI1 to freeze the counters
Hold Tot. mode enables maintenance work to be performed without interrupting the process.
When the device is in Hold Tot. mode:
The counters do not increment any more (frozen).
The device status LED flashes.
The display appears along with the 4 to 20 mA current output, the status of each transistor out­put and the relay, and the measured value of the flow rate.
The device is in Hold Tot. mode until the digital input switches again.
INPUT DI1 HOLD TOT. INV YES
INV NO
1. Choose INV NO to trigger the in­put on the leading edge.
2. Choose INV YES to trigger the in­put in the trailing edge.
RETURN
3. Confirm
59
8 Startup and parameterization
Configuring the digital input DI1 to reset the daily counter
NOTE!
The daily counter remains at zero until the digital input switches again.
INPUT DI1 RES. TOT. INV YES
RETURN
INV NO
1. Choose INV NO to trigger the in­put on the leading edge.
2. Choose INV YES to trigger the in­put in the trailing edge.
3. Confirm
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
60
8 Startup and parameterization

8.6.13 Adjusting the filter to the measured flow rate

This function enables you to dampen the following fluctuations in the transmission of measured values:
On the display
At the current output AO1
NOTE!
If "fast" filtering has been selected and the flow rate changes by ±30 % (e.g., when switching the medium circulation on or off), the filter is deactivated: The new flow rate is emitted in non-filtered form.
NOTE!
If the filter selected is too high, the device does not recognize jumps in the flow rate. This can lead to large differences between the flow rate in the pipeline and the displayed flow rate or the value emitted at the current output.
FILTER FILTER 0 RES. TOT. INV YES
FILTER 9 INV NO
1. Choose the filter.
1. Choose the slow or fast filter.
2. Confirm
TOTAL
There are ten filters available. The following table lists the response times (10 to 90 %) for each filter:
Filter Response time Filter Response time
01s 58s 12s 615s 23s 728s 34s 870s 45s 9145s
2. Confirm
61
8 Startup and parameterization
Available filters
Filter 0: measured flow rate
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
Filter 3 Filter 6 Filter 9

8.6.14 Resetting both counters

This function enables both counters to be reset.
NOTE!
Both counters are only reset if SAVE YES is confirmed when leaving the parameters menu.
Slower measurand output
Faster measurand output
NOTE!
The daily counter can be reset in the process level ( see chapter 8.4 "Using the navigation keys", page 35) or via the digital input ( see chapter 8.6.12 "Configuring the digital input DI1", page 55).
TOTAL RESET N
RESET Y
1. Reset both counters or not.
2. Confirm
FREQUENC.
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
62

8.6.15 Adjusting the mains frequency

This function enables the mains frequency to be adjusted so that the device can filter interfering signals in the supply voltage.
NOTE!
Always adjust this parameter, even if the device is supplied with direct current (DC).
FREQUENC. F = 50 HZ
CUT-OFF
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
8 Startup and parameterization
F = 60 HZ
1. Choose the mains frequency.
2. Confirm

8.6.16 Adjusting the CUT-OFF flow rate

This function is used to adjust the flow rate value below which the device emits the flow rate as zero.
The display shows a flow rate of zero; the flow rate unit is marked with a dot after it in the display
Outputs and counters respond to the zero flow rate
NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
CUT-OFF CO = 0.300
BACKLIT
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
1. Enter the measured flow rate value below which the device emits the flow rate as zero.
2. To disable the function, set CO = 0.
3. Confirm
63
8 Startup and parameterization

8.6.17 Setting display backlight and how long it stays on, or deactivating the backlight

This function allows
The brightness of the backlight to be activated and adjusted and an activation period to be selected after pressing a button;
The backlight be deactivated.
BACKLIT BKLG = 1 1. Set the backlight brightness,
(from 1 to 9).
BKLG = 9
DELAY = 30
1. Set the duration the display is backlit after a key press (00 s: the display is constantly backlit).
2. Confirm
END
BKLG OFF 1. Deactivate the backlight.
2. Confirm
END
1. If you do not wish to change any further parameters, go to the END function in the parameters menu and press the key in order to save the settings and return to the process level.
64

8.7 Test menu details

To access the test menu, press the + + keys at the same time for more than 5 seconds.
This menu enables:
CAL. AO1 The current output (4 to 20 mA) to be adjusted.
CALIB.0 The flow zero point of the device to be calibrated.
FLOW The outputs to be checked for proper operation.
K-SENSOR The KW value of the sensor to be changed in order to adjust the device's
FLOW-W. The flow rate range to be adjusted outside which a warning message is
8 Startup and parameterization
accuracy.
generated.
END SAVE N/Y 12.6 L/S
Saving the changes made within the test menu or not. If the changes are saved, the de­vice operates with the new settings.
Process level
65
8 Startup and parameterization

8.7.1 Adjusting the current output AO1

This function allows for the adjustment of the signal values emitted at the analog output.
CAL. AO1 OFFSET
OF = 4.05
The transmitter generates a current of 4 mA.
1. Measure the current emitted at the 4 to 20 mA output using a multi­meter.
2. Enter the value displayed by the multimeter. The permitted offset range is 3 to 5 mA.
3. Confirm
SPAN
SP = 19.95
The transmitter generates a current of 20 mA.
1. Measure the current emitted at the 4 to 20 mA output using a multi­meter.
2. Enter the value displayed by the multimeter. The permitted span range is 18.50 to 21.50 mA.
3. Confirm
VALID N/Y 1. Confirm the entered value or not.
CAL. AO1
1. If you do not wish to change any further parameters, go to the END function in the test menu and press the key in order to save the settings and return to the process level.
66

8.7.2 Calibrate the flow rate zero point

NOTE!
Adjust these parameters in the following cases:
Before using the Teach-In for the K-factor
After performing maintenance work
If the measured flow rate is not zero, even though medium circulation has been stopped
NOTE!
Make sure there are no air bubbles in the pipeline.
NOTE!
Before calibrating the flow rate zero point
and before starting up the sensor for the first time, immerse it into the measuring material for 24 hours
and after performing maintenance work, immerse the sensor into the measuring medium for one hour
NOTE!
During calibration
The device status LED flashes
The outputs are frozen at the last measured flow rate
The device cannot be adjusted
8 Startup and parameterization
67
8 Startup and parameterization
Calibrating the flow rate zero point via the digital input
NOTE!
see also chapter 8.6.12 "Configuring the digital input DI1", page55
1. Charge the pipe.
2. Stop medium circulation.
3. Make sure the device status LED is green.
4. Make sure the device is in the process level and not in the configuration level.
5. Activate the digital input configured to trigger the flow zero point calibration.
MEASURE / The device calibrates the flow rate zero point for approxi-
mately 30 seconds.
Process level 0.000 L/S
If the device status LED remains green, the calibration of the flow rate zero point has succeeded.
If the device status LED is orange, the calibration of the flow rate zero point failed: the message CAL.FAIL is displayed in the warning message list. Restart the calibration.
68
8 Startup and parameterization
Calibrating the flow rate zero point using the CALIB 0 function in the test menu
1. Charge the pipe.
2. Stop medium circulation.
CALIB.0 CALIB.N
CALIB.Y
1. Choose CALIB.Y.
MEASURE / The device calibrates the flow rate zero point.
FLOW CAL.FAIL
The calibration of the flow rate zero point has succeed­ed.
The device has detected a non-nil flow rate in the pipe.
The device status LED turns orange.
1. If you do not wish to change any further parameters, go to the END function in the test menu and press the key in order to save the settings and return to the process level.
69
8 Startup and parameterization

8.7.3 Checking proper response of the outputs

This function can be used to test whether the outputs are responding according to the settings.
NOTE!
The counters are incremented according to the measured flow rate value and not according to the sim­ulated value.
The device status LED flashes as long as the checking function for proper output response is active.
FLOW SIM = 12.50
1. Enter a flow rate value in the selected unit.
2. Confirm by pressing the key.
3. Check that the outputs are responding as expected.
4. To test another value, press .
5. To exit simulation, press or .
1. If you do not wish to change any further parameters, go to the END function in the test menu and press the key in order to save the settings and return to the process level.

8.7.4 Changing the sensor's KW value

This parameter allows the device's precision to be adjusted.
NOTE!
Changing the sensor's KW value can affect the determination of the K-factor for the fitting (406090) with a calibration procedure, see chapter 8.6.4 "Determining the fitting's K-factor with a calibration proce­dure (Teach-In)", page 41.
K-SENSOR KW = 1.000
1. Enter the sensor's KW value (value between 0.850 and 1.150).
2. Confirm
FLOW-W.
Setting example: to increase the measured flow rate value by 1.5 %, enter KW = 1.015.
70
1. If you do not wish to change any further parameters, go to the END function in the test menu and press the key in order to save the settings and return to the process level.
8 Startup and parameterization

8.7.5 Monitoring the flow rate in the pipeline

A problem in the process or with the sensor can be detected if the flow rate measurement is too low or too high.
This function enables the medium flow rate to be monitored and a report to be triggered if the flow rate is too low or too high.
NOTE!
Deactivating flow rate measurement monitoring: Set W- = W+ = 0.
NOTE!
The units of the displayed values and the values to be entered manually correspond to the values se­lected in chapter 8.6.2 "Choosing the flow rate unit, the number of decimals, and the counter display unit", page 38.
FLOW-W. W- = 0.000
1. Enter a flow rate value, below which a WARN LO warning message is generated by the device.
2. Confirm
W+ = 0.800
1. Enter a flow rate value (W+ > W-) above which a WARN HI warning message is generated by the device.
2. Confirm
END
To trigger a report if the flow rate is too low or too high, set a flow rate range outside of which the device generates a WARN LO or WARN HI warning and switches on the orange device status LED.
If a WARN LO or WARN HI warning is generated
1. check the process.
2. If the process is not the cause, check the condition of the sensor and clean it if necessary.
3. If the flow rate measurement still is not correct, contact the manufacturer.
NOTE!
The transistor or relay output can be configured to switch a load when the device generates a warning ( see chapter 8.6.5 "Configuring outputs", page 44 and chapter 9.5 "Troubleshooting", page 76).
1. If you do not wish to change any further parameters, go to the END function in the test menu and press the key in order to save the settings and return to the process level.
71
8 Startup and parameterization

8.8 Information menu details

NOTE!
This menu is only accessible when the device status LED is orange or red.
Meaning of a message see chapter 9.5.4 "Solving a problem with a warning or error message with the red device status LED", page 79 and chapter 9.5.5 "Solving a problem with a warning or error message with the orange device status LED", page 80.
To access the information menu, press the key for longer than 2 seconds in the process level.
This menu can be used to read off warnings and error messages generated by the device.
MESSAGE 1
MESSAGE 2
MESSAGE Z
.
.
.
END
a
When the CONFIRM key is unlocked
a
or
Process level
72
9 Maintenance

9.1 Safety information

Risk of injury resulting from high pressure levels within the system!
Before working on the system, switch off the pressure and bleed/empty the lines.
Risk of electric shocks!
Before working on the system or device, switch off the voltage and secure it so that it cannot switch on again.
Always observe the relevant accident prevention regulations and safety requirements for electrical de­vices.
Risk of injury due to high fluid temperatures!
Always use safety gloves to touch the device.
Stop medium circulation and empty the pipelines before disconnecting the process connections.

9 Maintenance

Risk of injury resulting from the type of medium!
When using dangerous media, always observe the specifications on the safety sheet and the relevant accident prevention regulations.
Risk of injury due to improper maintenance!
Ma int enance work must alw ays be carrie d ou t by authoriz ed a nd q ualified personnel using suitable tools.
After conducting any work on the device, always make sure that the restart process is supervised.
NOTE!
Startup after maintenance
After performing maintenance work, immerse the device in the measuring medium for one hour.
73
9 Maintenance

9.2 Returning devices

NOTE!
All documents important for the return as well as the return address of the manufacturer are available at http://reparaturdienst.jumo.info.
NOTE!
According to the EU Directive for hazardous materials, the owner of hazardous waste is responsible for its disposal or must observe the following regulations when shipping:
All devices delivered to the manufacturer must be free of any hazardous materials (acids, alkalis, sol­vents, etc.).
The following applies for returning devices for repairs or recalibration:
Use the original packaging or suitably secure shipping containers.
Enclose the completed accompanying repair form with the device.
The following also applies to devices that have come into contact with hazardous materials:
Enclose the completed accompanying repair form and the declaration of decontamination with the device.

9.3 Cleaning the device and sensor

NOTE!
Before cleaning the device/sensor, always observe the basic safety notes!  See chapter 3 "Basic safe- ty information", page 9
NOTE!
Always use a cleaning agent that is compatible with the materials used to make the device and sensor. See also chapter 6.3.1 "Materials", page 15
NOTE!
After cleaning the sensor
Rinse the sensor,
Check the seals and replace if necessary.
Calibrate the flow rate zero point before startup ( see chapter 8.7.2 "Calibrate the flow rate zero point", page 67).
Clean the device/sensor with a cloth slightly dampened with water or an agent that is compatible with the materials used to make the device.
To avoid any measurement errors caused by contamination of the measuring electrodes, clean the com­ponents that come into contact with the medium (define the cleaning frequency according to the pro­cess).
74

9.4 Replace the seal (device with G 2" union nut)

NOTE!
Do not scratch the seal's groove.
(1) Device (1) Groove (2) Union nut (2) Seal (3) Fitting (type 406090)
1. Unscrew the device's union nut.
2. Pull the device out of the fitting.
9 Maintenance
3. Remove the seal out of the groove.
4. Clean the seal's groove.
5. Place the new O-ring seal into the groove ( see chapter 10 "Spare parts and accessories", page 82).
6. Place the device into the fitting
7. Tighten the union nut onto the device by hand.
75
9 Maintenance

9.5 Troubleshooting

9.5.1 Solving a problem with the device status LED

Device sta­tus LED
Off 0 mA Low signal status PWRFAIL The voltage supply
Off 0 mA Not switched - The device is not
Analog out­put AO1
Digital outputs DO1/DO2/DO3
Displayed message
Meaning Measure
Make sure that the volt-
is too low. The device is not
working.
supplied with pow­er.
age supply is between DC 18 and 36 V.
If the problem arises again, contact the man­ufacturer
Check connection
Check the installation's fuse and replace if nec­essary
Check that the sys­tem's switch-off device is deactivated
Check that the voltage supply is working cor­rectly

9.5.2 Solving a problem without a warning or error message with the device status LED

Device sta­tus LED
Regardless of the color
Regardless of the color
Regardless of the color
Analog out­put AO1
4 to 20 mA Depending on
0 mA Depending on the
4 mA Depending on the
Digital outputs DO1/DO2/DO3
switching thresh­olds or switched
switching thresh­olds or switched
switching thresh­olds or switched
a
a
a
Problem Measure
Unable to access the parameters menus and test menu
The current output emits a 0 mA cur­rent.
The current output emits a signal be­tween 0 and 4 mA.
The current output emits a 4 mA sig­nal, regardless of the flow rate dis­played.
Check the position of the selection switch to lock or unlock the CONFIRM key, see "Use of the selection switch to lock or unlock the CONFIRM key ", page 27
Check the current output's connection
Check the installation's fuse and re­place if necessary
Check the position of the sink/source selection switch, see chapter 7.3.5 "Connection for the cur­rent output AO1", page 29
If the problem arises again, contact the manufacturer.
Switch the device's power supply off and on.
If the problem arises again, contact the manufacturer.
Check the settings for the current out­put, see chapter 8.6.6 "Configuring current output AO1", page 45
76
9 Maintenance
Device sta­tus LED
Regardless of the color
Analog out­put AO1
Digital outputs DO1/DO2/DO3
20 mA Depending on the
switching thresh­olds or switched
a
Problem Measure
The current output emits a 20 mA sig­nal, regardless of
Check the settings for the current out­put, see chapter 8.6.6 "Configuring current output AO1", page 45
the flow rate dis­played.
Regardless of the color
Regardless of the color
22 to 30 mA Depending on the
switching thresh­olds or switched
a
30 mA Depending on the
switching thresh­olds or switched
a
The current output emits a signal be­tween 22and30mA.
The current output emits a signal >30mA.
Switch the device's power supply off and on.
If the problem arises again, contact the manufacturer.
Check the current output's connec­tion.
If the problem arises again, contact the manufacturer.
Regardless of the color
4 to 20 mA Depending on the
switching thresh­olds or switched
a
The signal value emitted by the cur­rent output does not match the flow rate measured.
Check the settings for the current out­put, see chapter 8.6.6 "Configuring current output AO1", page 45
Check the OFFSET and SPAN set­tings for the current output, see chapter 8.7.1 "Adjusting the current output AO1", page 66
Regardless of the color
4 to 20 mA The relay outputs DO2 and DO3 do not
switch, regardless of the flow rate dis­played.
Check the settings for the relay out­puts DO2 and DO3, see chapter 8.6.11 "Configuring the relay outputs DO2 and DO3", page 52
Check the flow rate unit
Check the outputs' response, see chapter 8.7.3 "Checking proper re­sponse of the outputs", page 70
a
If the output has been configured to switch upon generation of a warning, see chapter 8.6.5 "Configuring out­puts", page 44
77
9 Maintenance

9.5.3 Solving a problem without a warning or error message with a green device status LED

Device sta­tus LED
Green 4 to 20 mA Depending on
Green 4 to 20 mA Depending on the
Green 4 to 20 mA Depending on the
Green 4 to 20 mA Depending on the
Green 4 to 20 mA Depending on the
a
If the output has been configured to switch upon generation of a warning, see chapter 8.6.5 "Configuring out­puts", page 44
Analog out­put AO1
Digital outputs DO1/DO2/DO3
switching thresh­olds or switched
switching thresh­olds or switched
switching thresh­olds or switched
switching thresh­olds or switched
switching thresh­olds or switched
Meaning Measure
The device is not measuring the flow
a
rate correctly.
The displayed flow rate is not zero
a
even though the flow rate in the pipe­line is zero.
The displayed flow rate is always zero.
a
The displayed flow rate is not stable.
a
The flow rate dis­play is changing at
a
a very slow place.
Check that the K-factor matches the
Check that the KW value has not been
Perform a Teach-In process to deter-
Check that the flow rate in the pipeline
Check that there are no air bubbles in
Check the filter level-
Calibrate the flow rate zero point.
Check that the flow rate in the pipeline
Check that the K factor or KW value
Check that the measuring electrodes
Select a smaller flow rate unit or in-
Check whether the medium is flowing
Select a higher filter
Check whether the medium is flowing
Select a smaller filter
fitting used
switched
mine the K-factor for the fitting used
is zero-
the pipeline-
is not zero.
are not too low.
are mounted so that they are perpen­dicular to the direction of flow.
creased the number of decimal places displayed.
in the pipeline
in the pipeline
78
9 Maintenance

9.5.4 Solving a problem with a warning or error message with the red device status LED

Device sta­tus LED
Red 22 mA Depending on the
Red 22 mA Depending on the
Red 22 mA Depending on the
Red 22 mA Depending on the
Red 22 mA Depending on the
Red 22 mA Depending on the
Red 22 mA Depending on the
Analog out­put AO1
Digital outputs DO1/DO2/DO3
thresholds
thresholds
thresholds
thresholds
thresholds
thresholds
thresholds
Displayed message
"ERROR3" The user settings
ERROR4 The counter values
ERROR5 Both
ERROR6 Counter values
ERROR7 Both
MEAS.OVF Flow rate in the
BAD MEAS. Faulty measuring
Meaning Measure
and factory calibra­tion have been lost.
The device is mea­suring the wrong values.
have been lost. The values saved
from the previous exterior voltage mains supply are used.
ERROR3 and ER­ROR4
have been lost completely
Both counters are reset.
ERROR3 and ER­ROR6
pipeline is > 12 m/s.
signal The flow rate is not
measured correctly.
Restart the device
If the problem arises again, contact the man­ufacturer
If the message LIN.LOST is generated at the same time, con­tact the manufacturer.
Restart the device
If the problem arises again, contact the man­ufacturer.
Contact the manufac­turer
Restart the device
If the problem arises again, contact the man­ufacturer
Contact the manufac­turer
Check the flow rate in the pipeline
Reduce the flow rate if necessary
If the problem arises again, contact the man­ufacturer
Make sure that the me­dium is in the pipeline
Make sure there are no air bubbles in the medi­um
Check the potential equalization of the in­stallation
If the problem arises again, contact the man­ufacturer
79
9 Maintenance

9.5.5 Solving a problem with a warning or error message with the orange device status LED

Device sta­tus LED
Orange 4to20mA Switched
Orange 4to20mA Switched
Orange 4to20mA Switched
Orange 4to20mA Switched
Analog out­put AO1
Digital outputs DO1/DO2/DO3
Displayed message
a
a
a
a
PULS.OVF The value set for
NEG. FLOW
WARN. LOW
WARN.HIG The measured flow
Meaning Measure
Select a higher value the pulse output is incorrect (the fre­quency generated is > 250 Hz).
The measured flow rate is negative (even though the value on the display is positive).
The measured flow rate is below the se­lected minimum flow rate.
This message is only generated when flow rate monitoring is active ( see chapter 8.7.5 "Moni toring the flow rate in the pipeline", page 71).
rate is above the selected maximum flow rate.
This message is only generated when flow rate monitoring is active ( see chapter 8.7.5 "Moni toring the flow rate in the pipeline", page 71).
for the parameter PU, see see chapter 8.6.7 "Configur ing the transistor output DO1 as a pulse output", page 46
Make sure that revers-
ing the direction of flow does not have a nega­tive impact on the pro­cess
If it does, mount the de-
vice on the pipeline so that the arrow on the side of the housing is pointing in the direction of flow
Check the flow rate in
the pipeline and any potential consequenc­es
If necessary, clean the
sensor and calibrate the flow rate zero point
Check the flow rate in
the pipeline and any potential consequenc­es
If necessary, clean the
sensor and calibrate the flow rate zero point
80
9 Maintenance
Device sta­tus LED
Orange 4to20mA Switched
Orange 4to20mA Switched
Orange 4to20mA Switched
a
If the output has been configured to switch upon generation of a warning, see chapter 8.6.5 "Configuring out­puts", page 44
Analog out­put AO1
Digital outputs DO1/DO2/DO3
a
a
a
Displayed message
DISP.OVF The measured flow
LIN.LOST The factory calibra-
CAL.FAIL Calibration of the
Meaning Measure
rate display in the process level is full and does not corre­spond to the actual flow rate.
Apart from the dis­play, the device is working in line with the actual value for the flow rate.
tion has been lost. The device is mea-
suring the wrong values.
flow rate zero point failed.
Change the unit or dec-
imal places in the UNIT function in the parame­ters menu so that the display can show high­er values
Contact manufacturer
Refer to the conditions
for calibration, see chapter 8.7.2 "Calibrat e the flow rate zero point", page 67
81
10 Spare parts and accessories

10 Spare parts and accessories

Risk of injury or material damage due to unsuitable parts!
Using the wrong accessories and unsuitable spare parts can cause injuries and damage to the device and its environment.
Always use genuine accessories and genuine spare parts supplied by the manufacturer.
Designation Part no.
PC lid, with flap, window, screws and adhesive film 693539 PPA lid, with flap, window, screws and adhesive film 693548 Set with 693568
2 cable fittings M20 × 1.5 2 flat seals made from CR for cable fittings or screw plugs 2 screw plugs M20 × 1.5 2 multiway seals 2 × 6 mm
Set with 693590
2 reductions M20 × 1.5/NPT 1/2" (with mounted seal) 2 flat seals made from CR for screw plugs 2 screw plugs M20 × 1.5
Set with 693607
1 plug for cable fitting M20 × 1.5 1 multiway seal 2 × 6 mm, for cable fitting 1 green seal made from FKM 1 installation instruction
Set with 693610
1 green seal made from FKM 1 black seal made from EPDM
Set with 693612
1 plug for cable fitting M20 × 1.5
1 multiway seal 2 × 6 mm, for cable fitting Circlip 693620 PC union nut for PC housing 693625 PPA union nut for PPA housing 693627
82
11 Packaging and tran sport

11 Packaging and transport

NOTE!
Transport damage!
Failing to protect the device properly can cause damage during transport.
Transport the device in an impact-proof packaging solution that is protected from moisture and contam­ination.
Do not subject the device to temperatures outside of the admissible storage temperature range, see chapter 12 "Storage", page 84.
Seal any electrical interfaces with protective caps to protect them against damage.
83
12
Storage

12 Storage

NOTE!
Incorrect storage can cause damage to the device!
Store the device in a dry and dust-free environment!
Device storage temperature range: -20 to +60 °C.
84
13 Disposal

13 Disposal

NOTE!
Environmental damage caused by improper disposal
The country-specific laws and regulations for handling and disposing of waste must be observed!
NOTE!
At the end of its service life, the device and any batteries present do not belong in the trash!
85
14 China RoHS
żżżżżż
żżżżżż
żżżżżż
ż
ℶ❐兓Ⓔ
3URGXFWJURXS
ℶ❐₼㦘⹂䓸德䤓⚜䱿♙⚺摞
&KLQD((3+D]DUGRXV6XEVWDQFHV,QIRUPDWLRQ
捷ↅ⚜䱿
&RPSRQHQW1DPH
3E
+J
&G
⏼ↆ杻
&U9,
⮩䅃勣啾 ᧤3%%
⮩䅃ℛ啾搩
3%'(
ཆ༣
+RXVLQJ
*HKlXVH
䗷〻䘎᧕
3URFHVVFRQQHFWLRQ 3UR]HVVDQVFKOXVV
㷪⇽
1XWV
0XWWHU
㷪ṃ
6FUHZ
6FKUDXEH
żżżżż
㦻嫷㫋∬㗽^:dϭϭϯϲϰ䤓屓⸩冥Ⓟᇭ
7KLVWDEOHLVSUHSDUHGLQDFFRUGDQFHZLWKWKHSURYLVLRQV6-7 ż᧶嫷䯉年㦘⹂䓸德⦷年捷ↅ㓏㦘⧖德㧟㠨₼䤓⚺摞⧖⦷*%7屓⸩䤓棟摞尐㻑ⅴₚᇭ ,QGLFDWHWKHKD]DUGRXVVXEVWDQFHVLQDOOKRPRJHQHRXVPDWHULDOV¶IRUWKHSDUWLVEHORZWKHOLPLWRIWKH *%7
î᧶嫷䯉年㦘⹂䓸德咂⺠⦷年捷ↅ䤓㩟₏⧖德㧟㠨₼䤓⚺摞怔⒉*%7屓⸩䤓棟摞尐㻑ᇭ ,QGLFDWHWKHKD]DUGRXVVXEVWDQFHVLQDWOHDVWRQHKRPRJHQHRXVPDWHULDOV¶RIWKHSDUWLVH[FHHGHGWKH OLPLWRIWKH*%7

14 China RoHS

86
JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc.
Street address: Moritz-Juchheim-Straße 1 36039 Fulda, Germany
Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany
Postal address: 36035 Fulda, Germany
Phone: Fax: Email: Internet:
+49 661 6003-0 +49 661 6003-607 mail@jumo.net www.jumo.net
JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK
Phone: Fax: Email: Internet:
+44 1279 63 55 33 +44 1279 62 50 29 sales@jumo.co.uk www.jumo.co.uk
6733 Myers Road East Syracuse, NY 13057, USA
Phone: Fax: Email: Internet:
+1 315 437 5866 +1 315 437 5860 info.us@jumo.net www.jumousa.com
Loading...