This symbol is used when there may be danger to personnel if the
instructions are ignored or not followed correctly!
Caution
This symbol is used when there may be damage to equipment or data if the
instructions are ignored or not followed correctly!
Caution
This symbol is used where special care is required when handling components
liable to damage through electrostatic discharge.
1.2Note signs
Note
This symbol is used to draw your special attention to a remark.
1
abc
✱Action instruction
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes
consist of two parts:
Marking in the text and the footnote text.
The markers in the text are arranged as continuous superscript numbers.
This symbol indicates that an action to be performed is described.
The individual steps are marked by this asterisk.
Example:
✱ Loosen Phillips-head screws.
6
2 Notes
To protect the device from static electrical
discharge, the user must be electrostatically
discharged before touching the instrument!
All necessary settings are described in these
operating instructions. If any difficulties should
nevertheless arise during startup, please do not
tamper with the instrument in any way. By doing so,
you could endanger your rights under the instrument
warranty! Please contact the nearest subsidiary or the
head office in such a case.
Please read these operating instructions before
placing the instrument in service. Keep the manual in
a place which is accessible to all users at all times.
Please assist us in improving these operating
instructions where necessary.
For technical questions
Service hotline:
Phone:0661 6003-300 or
0661 6003-653
Fax:0661 6003-881300 or
0661 6003-881653
E-mail:Service@jumo.net
7
3 Identifying the instrument version
The nameplate with the order code glued onto the side of the
instrument. The connected supply voltage must match the voltage
specified on the nameplate.
3.1 Type description
(1)Basic type
202723JUMO ecoTRANS pH 03,
Microprocessor/transmitter / switching device for
pH value / redox voltage and temperature
(2)Output I (pH value / redox voltage)
888Analog actual value output, freely programmable
(3)Output II (temperature)
000None
888Analog actual value output, freely programmable
(4)Output III (switching)
000None
1011 x relay, switching contact
(5)Extra codes
000None
024With PC setup software
(1)(2)(3)(4)(5)
Order code/--/
Order example202723/888-888-101/000
8
4 Mounting
≤ 75
-10
+60
°C
5 Electrical connection
The choice of cable, the installation, and the electrical connection
must conform to the requirements of VDE 0100 “Regulations on the
Installation of High-Voltage Systems with Nominal Voltages below
1000 V” or the appropriate local regulations.
❏ To protect the device from static electrical discharge, the user
must be electrostatically discharged before touching the
instrument!
❏ The electrical connection must only be carried out by qualified
personnel.
❏ Electromagnetic compatibility conforms to the standards and
regulations cited in the technical data.
❏ Operates only on SELV or PELV circuits.
❏ The instrument is not suitable for installation in areas with an
explosion hazard.
Apart from faulty installation, incorrect settings on the instrument
may also affect the proper functioning of the subsequent process or
lead to damage. Safety devices should always be provided that are
independent of the instrument (such as overpressure valves or
9
temperature monitors/limiters) and only capable of adjustment by
specialist personnel. Please observe the relevant safety regulations
for such matters.
❏ The load circuit must be fused for the maximum relay current, in
order to prevent the output relay contacts from becoming
welded in the event of a short circuit.
❏ The supply voltage must be fed into the instrument through a
125-mA fuse, medium time lag or comparable protection by
means of a separate branch.
❏ Do not connect any additional loads to the screw terminals for
the supply of the instrument.
❏ Any electrical connection deviating from the connection diagram
may cause the instrument to be destroyed.
❏ Lay the input, output, and supply lines so they are physically
separated from each other and are not parallel.
❏ Lay probe cables only as continuous lines, twisted and shielded
(not via series terminals or similar arrangements).
❏ Fluctuations in supply voltages are only permitted within
specified tolerances (see datasheet 20.2723).
❏ The instrument must not be mounted or dismounted unless it is
in dead state and/or the lines are not connected.
5.1 Connecting the sensor
To connect the pH redox electrode, special coaxial lines are
recommended as specified by JUMO datasheet 202990. Standard
commercial antenna or computer lines with coaxial construction are
generally not suitable. They may result in faulty readings or could
destroy the pH or redox electrode.
The semiconducting layer (see Fig. (2)) of the special coaxial line must
be at a sufficient distance to prevent a short circuit with the inner
conductor on the transmitter terminal. The coaxial line may be sensitive
to touch or motion due to the high internal resistance of a pH electrode.
This must be taken into consideration when laying the cable.
10
5.2 Customizing the special coaxial line
~40 mm
(1)
(2)
(1)
(2)
Example:
✱ Strip insulation from the line as shown in the drawing. Remove
the black semiconducting layer (2) from the insulation of the inner
conductor as indicated.
✱ Twist the shield (1) and insulate with a shrink tubing.
✱ Insulate the black semiconducting layer with shrink tubing (2) to
prevent short circuits.
✱ Place ferrules on the line ends.
11
P
K1
EXIT
4321
5876
13L-L+
16
9121110
Terminal assignment
12
Measurement
1613
13
16
12
ϑ
9
10
inputs
Terminal assignmentSymbol
pH electrode
or
Redox
combination
electrode
pH glass electrode
or
Metal electrode
(with isolated
reference
electrode)
Reference
electrode
(with isolated
electrodes)
1613Reference system
(braiding)
Glass electrode/metal
electrode (inner conductor)
13Glass electrode/metal
electrode (inner conductor)
16Reference system
Liquid potential
(connect only with
symmetrical
connection)
Resistance
thermometer
in 2-wire circuit
12
9
10
13
Binary input11
12
11
56
+-
78
+-
413
P
S
Ö
L-L+
-+
12
OutputsTerminal assignmentSymbol
I
Analog process
value output pH/
redox
(galvanically
isolated)
II
Analog process
value output temperature
(galvanically
isolated)
III
Relay
5
6
7
8
1
3
+
-
+
-
Pin
Break (SPST-NC)
Power
supply
Power
supply
(with reverse
polarity protection)
DC 20 to 30 V
4
Make (SPST-NO)
Terminal assignmentSymbol
LL+
14
6 Commissioning
6.1 Brief description of the instrument
The instrument measures and, depending on the configuration,
regulates the pH or redox voltage in aqueous solutions. The range
of applications includes general water and wastewater
management, measuring drinking and wastewater, process water,
surface and ocean water, swimming pool and fountain water, and
professional aquariums, etc.
The transmitter has two analog inputs. The first analog input (main
input for pH value or redox voltage) is provided for connecting
electrodes with an isolated reference electrode. The instrument is
also prepared for connecting antimony electrodes. A resistance
thermometer (Pt 100 or Pt 1000) can be connected to the second
analog input.
Up to two analog outputs and one relay switching contact are
available. The analog outputs are galvanically isolated and assigned
to the inputs. Either the main value (pH or redox voltage) or the
temperature can be assigned to the relay contact.
The keys and integrated LD display are used to operate and
configure the instrument. This can also be done very conveniently
with the setup connection (Notebook / PC) using the optional setup
program. The setup program can also be used to print configuration
data. This makes system documentation easier.
Instruments are delivered with a calibration verification tool that is
used to document instrument and calibration data.
With the correct wiring of device and sensor, and
correct configuration of the device, a measurement without calibration should be possible.
If this is not the case, influencing factors such as
short circuit, open circuit, EMC and flow conditions should also be considered.
Electrode or
electrode with isolated
reference electrode
6.3 Applying the supply voltage
If the instrument has been correctly connected, all LCD segments are
briefly lit immediately after the supply voltage is applied.
Note:
After the instrument is started, the output signal is 0 V or 0 mA. The
relay is in idle state (inactive). After about 2 seconds, the JUMO
ecoTRANS pH 03 begins working according to the configuration.
16
7 Adjusting / changing instrument functions
About 3
P
P
Changes can be made in the setup program or with the JUMO
ecoTRANS pH 03 keys.
7.1 Process value display
The process value display can be either
- in static mode or in
- alternating mode
Static display (default setting)
Compensated pH valueredox voltage
Alternating display (at a fixed rate of about 3 seconds)
Compensated pH valueCurrent temperature
with current unit
Change from static display to alternating display
✱ Press the key (for less than 2 sec)
Change from alternating display to static display
✱ Press the key (for less than 2 sec)
17
7.2 Operation
The instrument can be operated on different levels.
Access to all levels (except the user level) is protected by different
2
codes
On the User level (USER) all parameters can be viewed and/or
modified depending on user rights
On the Calibration level (CALIB) the electrode zero point (intercept
point) and/or the electrode slope can be calibrated.
On the Enable level (RIGHT) user rights can be defined.
On the Administrator level (ADMIN) all parameters can be set
(configured).
The different codes and settings on the Enable level make it
possible to assign different user rights to users.
1
The factory setting for all parameters is "READ," i.e. all
.
1
(see Enable level).
parameters can only be read on the User level, not changed.
2
Codes for the Administrator and Enable levels can only be
changed with the setup program
(see Section 13 "Operation via setup interface", page 64).
7.3 Key function
After 60 seconds with no user activity (no keystrokes),
the display reverts to process value display.
This timeout function is not active during calibration!
- The UP and DOWN keys are used to select a submenu in the
main menu and to scroll forward and backward.
- The P key switches to the corresponding submenu.
- To change (edit) a parameter, press the P key.
- When the parameter is released for editing, the value flashes. If
the parameter is locked, "LOCK" appears in the display.
- To change the parameter, it must be unlocked on the Enable level
(changed from "rEAd" to "Edit").
18
- To increase or reduce the value, press the UP or DOWN key.
- The value can then be accepted with the P key.
- The EXIT key is used to cancel the entry and switch to the next
higher level.
19
7.4 Selecting levels
Process value display
(Operator level)
2
(Enable level)
4
CALIB
(Calibration level)
5
(Administrator level)
3
Selection level
> 2 seconds
EXIT
1
P
P
EXIT
P
EXIT
P
EXIT
P
EXIT
USER
RIGHT
ADMIN
1
Timeout (automatically jumps back after 60 sec. with no activity).
2
See Section 7.5 "User level (USER)", page 21.
3
See Section 7.6 "Administrator level (ADMIN)", page 22.
4
See Section 7.7 "Enable level (RIGHT)", page 23.
5
See Section 7.8 "Calibration level (CALIB)", page 24.
20
7.5 User level (USER)
USER
PH
SENS.T
FCT.IN
FCT.1
SIM.1
ERR.1
CAL.1
T.PUL1
T.OFF2
T.ON2
HYS.1
ALAR.1
OUT.CO
SIM.CO
ERR.CO
CAL.CO
SCH.CO
SCL.CO
OUT.TE
SIM.TE
ERR.TE
CAL.TE
SCH.TE
SCL.TE
CA.LVL
CA.INT
SLOPE
NULL
DF.PH
SYM.PH
UNIT.P
CAL
TE.OUT
PH.OUT
BIN.1
BIN.IN
TEMP
OFF.TE
MAN.TE
UNIT.T
DF.TE
DST1
SENS.P
P
EXIT
21
7.6 Administrator level (ADMIN)
ADMIN
CODE
20
USER
PH
SENS.T
FCT.IN
FCT.1
SIM.1
ERR.1
CAL.1
T.PUL1
T.OFF2
T.ON2
HYS.1
ALAR.1
OUT.CO
SIM.CO
ERR.CO
CAL.CO
SCH.CO
SCL.CO
OUT.TE
SIM.TE
ERR.TE
CAL.TE
SCH.TE
SCL.TE
CA.LVL
CA.INT
SLOPE
NULL
DF.PH
SYM.PH
UNIT.P
CAL
TE.OUT
PH.OUT
BIN.1
BIN.IN
TEMP
OFF.TE
MAN.TE
UNIT.T
DF.TE
DST1
SENS.P
P
EXIT
22
7.7 Enable level (RIGHT)
RIGHT
CODE
300
USER
PH
SENS.T
FCT.IN
FCT.1
SIM.1
ERR.1
CAL.1
T.PUL1
T.OFF2
T.ON2
HYS.1
ALAR.1
OUT.CO
SIM.CO
ERR.CO
CAL.CO
SCH.CO
SCL.CO
OUT.TE
SIM.TE
ERR.TE
CAL.TE
SCH.TE
SCL.TE
CA.LVL
CA.INT
SLOPE
NULL
DF.PH
SYM.PH
UNIT.P
CAL
TE.OUT
PH.OUT
BIN.1
BIN.IN
TEMP
OFF.TE
MAN.TE
UNIT.T
DF.TE
DST1
SENS.P
P
EXIT
Parameters on the User level (USER)
ValueVisibleVariable
EDITXX
READX-
23
7.8 Calibration level (CALIB)
CALIB
1
CODE
NULL
110
CAL
SLOPE
CAL
See Section 9.5.1 "One-point calibration
(zero point)", page 44
See Section 9.5.2 "Two-point calibration
(zero point and slope)", page 45
1
See Section 9 "Calibration", page 40.
The timeout function is not active during calibration!
24
8 Configurable parameters
Parameters can be set with the setup program or on
the instrument.
Since some parameters are independent of each
other, it may be necessary when one parameter is
changed to adjust others as well.
Example:
When the measurement variable is changed (from pH
to redox voltage or vice versa) the display format,
setpoints, and other parameters are adjusted.
While parameters are being adjusted internally, the
display appears as follows:
25
1
1 =pH antimony
2 =redox
1 =%
1 =On
.
bold
8.1 Inputs
8.1.1 Measurement input - main value (submenu "PH")
ParameterDisplaySetting range
Measurement variable SENS.P0 =pH (default)
UNIT.P0 =mV
Unit of measure
(only for "redox"
measurement
variable)
26
SYM.PH0 = Off (default mode is asymmetrical)
Activation of
symmetrical pH value
recording.
Caution:
Before activation,
terminal 12 must be
connected with the
liquid potential!
1
The default setting is marked in
1
-2.00...0.00...2.00 pH(pH antimony)
-999...0...999 mV(redox)
10.0...100.0...110.0%(pH antimony)
-999.9...100.0...999.9% (redox)
.
bold
DF.PH0...1...99 seconds
Filter constant
main value
ParameterDisplaySetting range
(2nd order filter)
Electrode zero pointZERO5.00...7.00...9.00 pH(pH default)
8.1.2 Measurement input - temperature (submenu "TEMP")
ParameterDisplaySetting range
Probe typeSENS.T0 = Manual temperature entry
Filter constant
temperature
(2nd order filter)
Temperature unitUNIT.T0 = °C
28
MAN.TE-10...25...150.0°C
Manual
temperature entry
The default setting is marked in
1
1
or
-36...0...36°F
Note:
The default setting depends on which
temperature unit "UNIT.T" is selected.
1
1 = Freeze switching output
2 = Freeze analog outputs
3 = Freeze all outputs
4 = Freeze process value recording
.
bold
OFF.TE-20.00...0...20.00°C
Process value
correction
ParameterDisplaySetting range
Temperature (offset)
FCT.IN0 = No function
Activation by
potential-free
8.1.3 Binary input
ParameterDisplaySetting range
contact
29
The default setting is marked in
1
2
2
2
2
2
2
1
(active inside a window)
1 = Window contact main value
(active outside a window)
2 = Window contact main value
(normally open (NO), similar to LK7)
3 = Max contact main value
(normally closed (NC), similar to LK8)
4 = Min contact main value
2
(active inside a window)
5 = Window contact temperature
2
(active outside a window)
6 = Window contact temperature
(normally open (NO), similar to LK7)
7 = Max contact temperature
(normally closed (NC), similar to LK8)
8 = Min contact temperature
9 = Error output
10 = Calibration timer elapsed
see Section 11 "Relay output", page 54
2
.
8.2 Relays
8.2.1 Binary output 1 (submenu "BIN.1")
ParameterDisplaySetting range
FCT.10 = No function
Switching function
Note:
Changes to this
parameter affect:
-ALAR.1
30
- HYS.1.
bold
The default setting is marked in
1
1
pH:-2.00 ... 0 ...16.00 pH
Redox:-1500 ... 0 ... 1500 mV
+14.0 ... 0 ...+302°C
Temperature:-10.0 ... 0 ...+150°C
0 ... 1 ... 1500 mV
0 ... 1 ... 50%
0 ... 1 ... 80°C
0 ... 1 ... 144°F
0 ... 1 ... 1500 mV
0 ... 1 ... 50%
0 ... 1 ... 80°C
0 ... 1 ... 144°F
.
2
ParameterDisplaySetting range
Switching pointALAR.1
HysteresisHYS.1.0.00 ... 1.00 ... 9.00 pH
31
Half window widthDST.10.00 ... 1.00 ... 9.00 pH
bold
The default setting is marked in
Energizing delayT.ON 10...2..0.999 seconds
1
The setting range depends on the switching function FCT.1 selected
2
1
Note:
This parameter is only active if the pulse time is
"T.PUL1" 0.
page 54
1 = active
2 = remains in current state
1 = active
2 = frozen (relay remains unchanged)
1 = active
2 = no manual mode
.
bold
ParameterDisplaySetting range
De-energizing delayT.OFF10...1..0.999 seconds
CAL.10 = inactive
Pulse timeT.PU L10...999 seconds, see Section 11 "Relay output",
Behavior of
binary output 1
in calibration mode
ERR.10 = inactive
Response in case of
error
Manual modeSIM.10 = inactive
32
The default setting is marked in
1
1
1 = 4...20 mA
2 = 20...0 mA
3 = 20...4 mA
4 = 0...10 V
5 = 2...10 V
6 = 10...0 V
7 = 10...2 V
Note:
If there is a change in the type of unit signal
"OUT.PH", manual mode of the output is
deactivated.
.
bold
OUT.PH0 = 0...20 mA
Type of
8.3 Analog outputs
8.3.1 Main value (submenu "PH.OUT")
ParameterDisplaySetting range
unit signal
33
The default setting is marked in
1
1
-15.00 ... 1200 mV(Redox)
0 ... 90%(Redox voltage)
Note:
The interval between the initial value of scaling
"SCL.PH" and the final value of scaling "SCH.PH"
must be at least 10% of the measurement range.
-1200 ... 1500 mV(redox voltage)
10 ... 100%(redox voltage)
Note:
The interval between the initial value of scaling
"SCL.PH" and the final value of scaling "SCH.PH"
must be at least 10% of the measurement range.
1 = Current status is maintained
.
bold
SCL.PH-2.00 ... 14.20 pH(pH)
Initial value of
ParameterDisplaySetting range
scaling
SCH.PH-0.20 ... 16.00 pH(pH)
Final value of
scaling
34
CAL.PH0 = Concurrent
Behavior in
calibration mode
The default setting is marked in
1
1
1 = HIGH (e.g. 10 V)
2 = LOW NAMUR (1.4 V / 3.4 mA)
3 = HIGH NAMUR (10.7 V / 22 mA)
0...22 mA or 0...10.7 V
.
bold
ERR.PH0 = LOW (e.g. 0 V)
Response in case of
ParameterDisplaySetting range
error
SIM.pHOFF = no manual mode
Manual mode of the
analog output -
conductivity
35
The default setting is marked in
1
1
1 = 4...20 mA
2 = 20...0 mA
3 = 20...4 mA
4 = 0...10 V
5 = 2...10 V
6 = 10...0 V
7 = 10...2 V
Note:
If there is a change in the type of unit signal
"OUT.TE", manual mode of the output is
deactivated.
.
bold
OUT.TE0 = 0...20 mA
Type of
8.3.2 Temperature (submenu "TE.OUT")
ParameterDisplaySetting range
unit signal
36
The default setting is marked in
1
1
or
50.0 ... 273.0°F
Note:
The setting range and default setting depend on the
temperature unit "UNIT.T" that is selected.
The interval between the initial value of scaling
"SCL.TE" and the final value of scaling "SCH.TE"
must be at least 10% of the measurement range.
or
42.8 ... 302.0°F
Note:
The setting range and default setting depend on the
temperature unit "UNIT.T" that is selected.
The interval between the initial value of scaling
"SCL.TE" and the final value of scaling "SCH.TE"
must be at least 10% of the measurement range.
.
bold
SCL.TE-10.0 ... +134.0°C
Initial value of
ParameterDisplaySetting range
scaling
37
SCH.TE6.0 to 150.0°C
Final value of
scaling
The default setting is marked in
1
1
1 = Current status is maintained
CAL.TE0 = Concurrent
1 = HIGH (e.g. 10 V)
2 = LOW NAMUR (1.4 V / 3.4 mA)
3 = HIGH NAMUR (10.7 V / 22 mA)
ERR.TE0 = LOW (e.g. 0 V)
0...22 mA or 0...10.7 V
.
bold
SIM.TEOFF = no manual mode
Behavior in
ParameterDisplaySetting range
calibration mode
Response in case of
The default setting is marked in
error
Manual mode of the
analog output -
temperature
1
38
1
P
1 = Rapid access enabled
CAL.LVL0 = Rapid access disabled
Rapid access to
calibration with
8.3.3 Calibration settings
ParameterDisplaySetting range
and keys
Calibration intervalCA.INT0 ... 999 days
39
9 Calibration
9.1 General information
The zero point (intercept) and slope of pH sensors stray from copy
to copy. During the service life of the sensor, up to the time it must
be replaced by a new sensor, the zero point and slope change. To
allow for precise measurements, the transmitter must be adjusted
(calibrated) to the current sensor parameters. The calibration is
performed using buffer solutions.
Only the zero point needs to be calibrated for redox sensors.
9.1.1 When is calibration required?
The electrochemical sensors should be cleaned and the transmitter
should be calibrated at regular intervals (depending on the
measuring medium)!
With the correct wiring of device and sensor, and
correct configuration of the device, a measurement without calibration should be possible.
If this is not the case, influencing factors such as
short circuit, open circuit, EMC and flow conditions should also be considered.
9.2 Activating and starting calibration mode
During the calibration process, the display flashes.
The analog outputs respond as they were calibrated
on the USER LEVEL / ANALOG OUTPUT x /
FOR CALIBRATION.
The response of the relay depends on the
configuration of the switching output!
40
Calibration can be cancelled at any time with the EXIT
P
P
P
P
P
key. Old calibration data will not be lost
.
Calibration is performed using the instrument keys.
The zero point and slope of a sensor can also be
entered manually. This should only be done in
exceptional cases. Some sensors come with a test
report that specifies the zero point and slope. These
values simply provided documentation that the
sensor was in proper condition upon delivery. Since
these values change during storage, they are not
suitable for manual entry. We always advise
performing calibration with buffer solutions.
The timeout function is not active during calibration!
9.2.1 Starting calibration from the calibration level "CALIB"
Use the menu to switch to the calibration level (see
Section 7.4 "Selecting levels", page 20 and Section
7.8 "Calibration level (CALIB)", page 24).
✱
✱ Press and hold (for less than 1 sec)
Press and hold
(for more than 2 sec)
/
then select CALIB.
/ then enter
code 110.
✱ Confirm with ,
continue with Section 9.4 "One-point or two-point calibration",
page 43.
9.2.2 Starting calibration with the hot key (rapid access)
Activation of the calibration mode by rapid access
must be previously enabled:
Press and hold for 2 seconds / Set ADMIN / CAL /
>
CA.LVL to 1.
✱ Press the and keys,
continue with Section 9.4 "One-point or two-point calibration",
page 43.
41
9.3 Canceling calibration and error messages
Calibration can be cancelled at any time with the EXIT
key. Old calibration data will not be lost
During the calibration process, the transmitter calculates the
electrode parameters zero point and if applicable slope. If the
calculated values fall outside the permissible parameter limits, an
error message is generated.
For the permitted value ranges of parameters NULL and SLOPE see
Section 8.1.1 "Measurement input - main value (submenu "PH")",
page 26.
9.3.1 Zero point error
.
Parameter limits for NULL have been violated.
9.3.2 Slope error
Parameter limits for SLOPE have been violated.
9.3.3 General error during calibration
42
The parameter limits for NULL and SLOPE have been violated and/
P
or the two calibration points Ref.1 and Ref.2 are too close to each
other.
For two-point calibration of a pH sensor, the minimum spacing is
2 pH.
For two-point calibration of a redox sensor, the minimum spacing is
2 mV.
The greater the spacing, the more accurate the calibration.
9.3.4 Acknowledging errors
✱ Press the keys or EXIT.
The error message is deleted.
The faulty parameters NULL and/or SLOPE are not saved.
The instrument continues working with old calibration data.
9.3.5 Additional measures
✱ Check the quality and condition (age) of the buffer solutions.
✱ Check whether the sensor is dirty or needs to be replaced.
✱ Check whether the sensor plug is moist or the wiring is faulty.
✱ Observe the minimum spacings for buffer or reference values.
9.4 One-point or two-point calibration
For one-point calibration, the transmitter is only adjusted to the zero
point of the sensor.
In two-point calibration the transmitter is adjusted to both the zero
point of the sensor and its slope. We expressly recommend twopoint calibration!
9.4.1 Selecting one-point calibration
Select CAL I-PT.
For more details see Section 9.5.1 "One-point calibration (zero
point)", page 44.
43
9.4.2 Selecting two-point calibration
P
P
P
Press the key and select CAL 2-PT.
For more details see Section 9.5.2 "Two-point calibration (zero
point and slope)", page 45.
9.5 Calibrating a pH measurement chain
9.5.1 One-point calibration (zero point)
✱ Confirm the selection with .
Display or editing option of the buffer temperature.
✱ Confirm the selection with .
The first reference value is then measured.
✱ Wait until the measurement value has stabilized.
Confirm the measurement value with .
44
✱ Enter the actual buffer value with or and confirm with .
Flashing
P
P
EXIT
The calculated zero point of the measurement chain appears.
✱ Accept the value with or press
to cancel the calibration.
The instrument goes into measurement mode.
If an error message appears,
see Section 9.3
"Canceling calibration and error messages", page 42.
9.5.2 Two-point calibration (zero point and slope)
Calibrate zero point and slope.
The buffer solutions (reference solutions) used for
calibration must differ by at least 2 pH!
During the calibration, the temperature of the two
buffer solutions must be identical and remain
constant!
45
✱ Confirm the selection with .
P
P
P
Flashing
P
Display and editing option of the buffer temperature.
✱ Confirm the selection with .
The first reference value is then measured.
✱ Immerse the measurement chain in the first buffer (for example
pH 7.00)
✱ Wait until the measurement value has stabilized.
Confirm the measurement value with .
✱ Enter the actual buffer value with or and
confirm with .
46
✱ Remove the measurement chain from the first buffer and rinse
P
Flashing
P
P
EXIT
with distilled water.
✱ Immerse the measurement chain in the second buffer (for
example pH 4.00)
✱ Wait until the measurement value has stabilized.
Confirm the measurement value with .
✱ Enter the actual buffer value with or and
confirm with .
The calculated zero point (top line) and the calculated slope (bottom
line) appear.
✱ Accept the values with or press
to cancel the calibration.
The instrument goes into measurement mode.
If an error message appears,
see Section 9.3
"Canceling calibration and error messages", page 42.
47
9.6 pH antimony measurement chain
P
Antimony measurement chains are calibrated similarly to "normal"
pH measurement chains, see Section 9.5 "Calibrating a pH
measurement chain", page 44.
9.7 Redox measurement chain
9.7.1 General information
The instrument offers two calibrating options for adjusting it to the
redox measurement chain.
- One-point calibration
If "mV" was configured as UNIT.
- Two-point calibration
If "%" was configured as UNIT.
The display flashes during calibration.
The analog outputs respond the same way as on
USER LEVEL / ANALOG OUTPUT x / FOR
CALIBRATION.
The response of the relay depends on the
configuration of the switching output!
✱ Confirm the selection with .
✱ Rinse off the measurement chain with distilled water or clean if
necessary (see operating instructions for electrode).
✱ Immerse the measurement chain in the test solution (e.g. 468 mV).
48
The reference value is then measured.
P
Flashing
P
P
EXIT
✱ Wait until the measurement value has stabilized.
✱ Confirm the measurement value with .
✱ Enter the actual buffer value with or and
confirm with .
The calculated zero point of the measurement chain appears.
Accept the value with or press
to cancel the calibration.
The instrument goes into measurement mode.
If an error message appears,
see Section 9.3
"Canceling calibration and error messages", page 42.
49
9.7.3 Two-point calibration
P
P
Flashing
In this type of calibration the display range can be scaled freely from
0 to 100%.
Example:
A span of -10 mV ... +1000 mV can be scaled to 0 ... 100% werden.
The zero point can fall within the range of
-999 ... +999 mV.
Two different reference fluids must be used for the
calibration. The redox voltage of a measurement
solution is not temperature-dependent!
)
Confirm the selection with .
The first reference value is then measured.
✱ Immerse the measurement chain in the first solution (for example
59 mV)
✱ Wait until the measurement value has stabilized.
Confirm the measurement value with .
50
✱ Enter the desired value with or (for example 20)
P
P
Flashing
P
P
EXIT
and confirm with .
✱ Remove the measurement chain from the first solution and rinse
with distilled water.
✱ Immerse the measurement chain in the second solution (for
example 295 mV).
✱ Wait until the measurement value has stabilized.
Confirm the measurement value with .
✱ Enter the desired value with or (for example 80)
and confirm with .
✱ The calculated zero point (top line in mV) and the calculated
slope (bottom line) appear.
Accept the values with or press to cancel the calibration.
The instrument goes into measurement mode.
If an error message appears,
see Section 9.3
"Canceling calibration and error messages", page 42.
51
10 Analog output
Analog outputs are configured on the user level
(USER) or on the administrator level (ADMIN) in
PH.OUT (pH or redox output) and TE.OUT
(temperature output); see Section 7.5 "User level
(USER)", page 21.
10.1 Behavior of the output signal during calibration
Two options are possible, "Concurrent" and "Unchanged"
(constant).
10.2 Behavior of the output signal in case of error
If one of the following errors occurs, the output signal assumes the
defined status (see Section 10.3 "Output signal in case of error",
page 53):
Analog output of pH value / redox voltage with non-active
temperature compensation
- Underrange pH value / redox voltage
- Overrange pH value / redox voltage
Analog output of pH value / redox voltage with active
temperature compensation
- Underrange pH value / redox voltage
- Overrange pH value / redox voltage
- Underrange temperature
- Overrange temperature
Analog output - temperature
- Underrange temperature
- Overrange temperature
52
10.3 Output signal in case of error
Depending on the configuration, the output signal may assume the
"LOW" or "HIGH" state in case of error.
Output signal
nominal
0...20 mA22.0 mA0 mA
4...20 mA22.0 mA3.4 mA
0...10 V10.7 V0 V
2...10 V10.7 V1.4 V
Output signal
HIGH
Output signal
LOW
10.4 Output signal when leaving the scaling range
When leaving the scaling range, the output returns a proportional
signal up to a defined limit (in compliance with NAMUR NE43). The
limits are listed in the table below:
Value below scaling
range
0.0 mA0...20 mA20.5 mA
In the scaling
range
Scaling range was
exceeded
3.8 mA4...20 mA20.5 mA
0.0 V0...10 V10.2 V
20.5 mA20...0 mA0.0 mA
20.5 mA20...4 mA3.8 mA
10.2 V10...0 V0.0 V
1.8 V2...10 V10.2 V
10.2 V10...2 V1.8 V
10.5 Manual mode of the analog output
For test purposes or startup, a constant analog signal can be
generated by the instrument
(see also Section 11.3 "Manual mode of the relay output", page
57).
After a power supply failure, manual mode is
deactivated.
53
11 Relay output
11.1 Relay behavior
Depending on the setting, the JUMO ecoTRANS pH 03 monitors a
limit value.
11.2 Binary output 1 (submenu "BIN.1")
Setting options
0=No function (default setting)
1=Window contact main value (active inside a window)
2=Window contact main value (active outside a window)
3=Max contact main value (normally open (NO), similar to LK7)
4=Min contact main value (normally closed (NC), similar to LK8)
5=Window contact temperature (active inside a window)
6=Window contact temperature (active outside a window)
7=Max contact temperature (normally open (NO), similar to
LK7)
8=Min contact temperature (normally closed (NC), similar to
LK8)
9=Error output (every device error causes the relay to switch)
10=Calibration timer elapsed
54
Contact function for settings 1 and 5
Hysteresis
HYS.1.
pH
On
Off
Distance
DST.1
Switching point
ALAR.1
Hysteresis
HYS.1.
pH
On
Off
Distance
DST.1
Switching point
ALAR.1
Contact function for settings 2 and 6
55
Contact function for settings 3 and 7
Hysteresis
HYS.1.
pH
On
Off
Switching point
ALAR.1
Hysteresis
HYS.1.
Switching point
ALAR.1
pH
On
Off
Contact function for settings 4 and 8
56
11.3 Manual mode of the relay output
About 3
132
For test purposes or when starting up systems, a constant signal
can be generated by the transmitter.
Manual mode can be set with parameter: USER / BIN.1 / SIM.1 to
0= inactive
1= active => LED "K1" is lit
2= No manual mode.
Indication of manual mode
If one of the outputs is in manual mode, this is indicated in the
change to process value display by an "S" or "-".
SThe relevant output is in manual mode.
-The relevant output is not in manual mode.
1st place Analog output - pH/redox voltage
2nd place Analog output - temperature
3rd place Binary output 1
In the example above, the analog temperature output is in manual
mode. No other outputs are in manual mode.
When the user exits manual mode, the output signal immediately
assumes the value proportional to the pH value/redox voltage or the
temperature process value.
57
After "Supply voltage On" manual mode is always deactivated.
11.4 Behavior of the relay during calibration
The behavior of the relay with the parameter:
USER / BIN.1 / CAL.1 set to
0 =Relay inactive
1 =Relay active
2 =Relay unchanged
(during calibration the relay status remains set to the status
that was valid before the calibration process)
11.5 Pulse function of the relay output
The limit comparator is reset after an adjustable "pulse time." The
parameter for this is: USER / BIN.1 / T.PUL1 .
It can be set from 0 = 0 seconds (no pulse function)
to 999 = 999 seconds.
LED "K1" is lit red as long as the switching condition is met.
No OFF delay is possible in pulse mode.
58
Triggering condition is longer than pulse duration
Time
On
Off
Pulse duration
T.PUL1
Pulse
contact
Time
On
Off
Triggering
condition
Time
On
Off
Pulse
contact
Time
On
Off
Triggering
condition
Pulse duration
T.PUL1
Triggering condition is shorter than pulse duration
59
11.6 Behavior of the relay in case of error
The behavior of the relay can be adjusted with the following
parameters:
USER / BIN.1 and ERR.1 set to
0 =Relay inactive
1 =Relay active
2 =Relay unchanged
("frozen": the relay status remains set to the status that was
valid before the error)
FunctionpH value
redox voltage
Under-
range
Limit comparator
pH/redox voltage
xx
without temperature
compensation
Limit comparator
pH/redox voltage
xxxx
with temperature
compensation
Limit comparator
temperature
Calibration timer elapsedxxxx
Over-
range
Temperature
Under-
range
xx
Overrange
Error outputxxxx
11.7 Error detection
The relay output is active for the following errors:
Limit comparator for pH value/redox voltage with non-active
temperature compensation
- Underrange pH value/redox voltage
- Overrange pH value/redox voltage
60
Limit comparator of pH value/redox voltage with active
temperature compensation
- Underrange pH value/redox voltage
- Overrange pH value/redox voltage
- Underrange temperature
- Overrange temperature
Limit comparator - temperature
- Underrange temperature
- Overrange temperature
Calibration timer
-Timeout
61
12 Display and LED messages
12.1 Operating states of JUMO ecoTrans pH 03
Two LED indicate operating states
Instrument statusRed LED (top)Yellow LED (bottom)
Normal modeOff On if LC1 is active
ErrorFlashingOn if LC1 is active
InitializationOffOff
12.2 Underrange
Value is below lower measurement range limit (ORP means redox).
12.3 Overrange
Measurement range exceeded (ORP means redox).
12.4 Broken sensor
Broken sensor, wrong or no temperature sensor connected;
see Section 8.1.2 "Measurement input - temperature
(submenu "TEMP")", page 28.
62
If no automatic temperature compensation or temperature
measurement is required parameter SENS.T must be set
accordingly; see Section 8.1.2 "Measurement input temperature (submenu "TEMP")", page 28.
12.5 Short-circuit
12.6 Initialization of dependent parameters
After a parameter has been changed, other dependent parameters
are automatically changed.
Please check all dependent parameters!
12.7 Calibration timer elapsed
Depending on specifications (for example of the equipment
manufacturer), calibration of the measurement chain should be
performed.
After a correct calibration, the calibration timer is reset and
automatically restarted.
63
13 Operation via setup interface
P
K1
EXIT
(1)
(2)
(3)
(4)
(1)JUMO ecoTRANS pH 03
(2)PC interface line (optional accessory)
(3)JUMO PC setup software, multilingual D / GB / F (optional
accessory)
(4)PC or Notebook with USB interface
®,
operating system: Windows 2000
®
NT
4.0 and up, Windows Vista
Windows XP® or Windows
®
64
13.1 Operation with PC setup program
(1)
(2)
(3)
(1)Navigation tree
The navigation tree provides fast access (double click) to
individual setting options.
(2)Diagnostics window
As soon as a connection is established with an instrument,
current data is displayed here.
(3)Working area
Clicking the arrow () shows possible settings.
Double clicking the text opens the appropriate editing window.
65
14 Technical data
Inputs
Analog input 1 (pH / redox)
- Electrodes
- Glass or metal electrodes with isolated reference electrode
- Antimony electrode
Measurement ranges for pH / redox
-2 ... 16 pH or
-1500 ... +1500 mV
Accuracy for pH / redox
± 1% of the measurement range
Analog input 2 (temperature)
- Resistance thermometer Pt100 or Pt1000
The temperature sensor can be connected in a 2-wire circuit.
Measurement display can be switched between °C / °F
Temperature offset for analog input 2
A process value can be corrected by an offset in the range from 20 ... +20°C.
Temperature measurement range
-10 ... +150°C or 14 ... 302°F
Deviation from characteristic temperature curve
for Pt 100 / Pt 1000: ≤ 1.5 K
Outputs
Two analog outputs:
Freely configurable:
0(2) … 10VR
10 … (2)0VR
load
load
≥ 2kΩ or ≥ 2kΩ or else
66
0(4) … 20mAR
20 … (4)0mAR
≤ 400 Ω or
load
≤ 400 Ω
load
galvanically isolated from inputs:
ΔU ≤ 30V AC or ΔU ≤ 50 V DC
Scope of scaling at least 10% of the scope of the measurement
range
deviation from the characteristic curve of the output signal
≤ 0.075% of the measurement range
Relay output:
Switching contact
switching capacity: 8 A, 250 V AC or 8 A, 24 V DC
with resistive load
contact life: > 100,000 operations at normal load