This sign is used when there may be
are disregarded or not followed accurately!
Caution
This sign is used when there may be
instructions are disregarded or not followed accurately!
1.2Note signs
Note
This sign is used when your
Reference
➩
This sign refers to
abc
1
Footnote
Footnotes
Markings in the text and the footnote text.
special attention
further information
refer to certain points
danger to personnel
damage to equipment or data
is drawn to a remark.
in other chapters or sections.
in the text. Footnotes consist of two parts:
if the instructions
if the
✱
The markings in the text are arranged as continuous superscript numbers.
Action
This sign indicates that an
The individual steps are marked by this asterisk.
Example:
✱
Remove crosspoi nt s crew.
✱
Press the key.
action to be perform e d
is described.
4
2.1Preface
2 General
Please read these Operating Instructions before commissioning the instrument. Keep the manual in a place which is accessible to all users at all times.
Please assist us to improve these operating instructions where necessary.
Your suggestions are welcome.
PhoneGermany (0661) 6003-714
abroad(+49) 661 6003-0
FaxGermany (06 61) 6003-605
abroad(+49) 661 603-607
All necessary settings are described in this manual. If any difficulties should still arise during commissioning, you are asked not to
carry out any unauthorized manipulations on the unit. You could
endanger your rights under the instrument warranty!
Please contact the nearest subsidiary or the main factory in such a
case.
When returning modules, assemblies or components, the regulations of EN 100 015 “Protection of electrostatically sensitive components” must be observed. Use only the appropriate ESD packaging for transport.
Please note that we cannot accept any liability for damage caused
by ESD.
ESD=electrostatic discharge
5
2 General
2.2Application
GeneralThe type JUMO dTRANS p02 is used to measure the pressure of corrosive
and non-corrosive gases, vapors and liquids. The pressure transmitter uses
the piezo-resistive effect to make the measurement. The output signal is a proportional DC current which is linearly proportional to the input pressure.
Version with
Ex protection
The JUMO dTRANS p02 pressure transmitter in intrinsically safe version
marked II 1/2 G EEx ia IIC T6, is suitable for mounting within the hazardous
area Zone 1 for connection to Zone 0.
The EC Type-Examination Certificate for the JUMO dTRANS p02 is:
PTB 98 ATEX 2194
If this pressure transmitter is used in areas with potentially explosive atmospheres, the appropriate regulations for installing and operating equipment in
such areas must be observed.
The diaphragms of the JUMO dTRANS p02 pressure transmitter
must not be damaged. Do not insert any objects into the bores of
the pressure connections!
6
3 Delivery package
Operating
Instructions
B 40.4385
JUMO
calibration
certificate
Setup
program
These operating instructions are a comprehensive guide for the installation,
electrical connection, commissioning, operation and parameterization of the
JUMO dTRANS p02 pressure transmitter.
A JUMO calibration certificate and a setup print-out comes with each pressure
transmitter. The se documents contain information on the parameter settings
and/or the measured characteristic data for the transmitter concerned.
If this calibration certificate is lost, or should you need an additional one, it can
be ordered from JUMO stating the serial No. of the pressure transmitter (see
nameplate),
➩
see addresses of suppliers on the back of the operating instructions
All the parameters of the transmitter can be conveniently checked and
adjusted by using the setup program. Furthermore, additional functions are
available such as:
- recording of measurements
- graphical display of temperature and pressure
- comprehensive diagnostic messages
- indication of the complete order code and instrument configuration
(for repeat orders).
The setup program is available on request, at extra cost,
Sales No. 40/00365072.
®
HART
Power supplyA supply unit with a supply isolator is available on request, at extra cost,
modemThe HART
HART
Sales No. 40/00345666.
Sales No. 40/00389710.
®
modem serves to connect the JUMO dTRANS p02 to a PC. The
®
modem is available on request, at extra cost,
7
4 Instrument identification
4.1Nameplate
Nameplate on a pressure transmitter without approval for use within the
hazardous (Ex) area.
Nameplate on a pressure transmitter with approval for use wi thin the hazardous (Ex) area.
8
4.2Type designation
(1) Basic type
404385pressure transmitter dTRANS p02 with sensing element
(2) Basic type extensions
0none
1Protection EEx ia IIC (PTB 98 ATEX 2194)
4for increased temperature of medium up to 200°C
9special version
(3) Nominal input range
414-100 to +100 mbar gauge pressure
453-0.6 to +0.6mbar gauge pressure
457-1 to +4 bar gauge pressure
461-1to +25 bar gauge pressure
464-1to +100 bar gauge pressure
468-1to +600 bar gauge pressure
4870 to +0.6 bar absolute pressure
4910 to +4 bar absolute pressure
4950 to +25 bar absolute pressure
(4) Output
4054 — 20mA with HART
(5) Process connection
504
512
564
571
604taper connection with slotted ring nut DN25 to DIN 11 851
606taper connection with slotted ring nut DN40
613clamp connection DN25 to DIN 32 676
616clamp connection DN50 to DIN 32 676
(6) Material of process connection
20stainless steel
82Hastelloy C276, Mat. Ref. 2.4819
99special material
(7) Electrical connection
06screw terminals
33Harting connector Han 7D /Han 8D (8U)
(8) Filling medium of measuring system
0none
1silicone oil
2halogenized filling oil
1
not for nominal range -1 to 100 bar and -1 to 600 bar gauge pressure
2
not with Protection EEX ia II C
4 Instrument identification
in piezo-resistive or thin-film technology
1
(only with process connections 571, 604, 606, 613 and 616)
®
protocol
1
/2" pipe to EN 837
1
/2" - 14 NPT to DIN 837
1
/2" - 14 NPT female
3
/4" pipe flush to DIN 837
1
1
1,2
1,2
1,2
1
2
1,2
(1)(2)(3)(4)(5)(6)(7)(8)
Order code404385 /405
Order example404385 / 0 - 457 - 405 - 504 - 20 - 06 -1
9
4 Instrument identification
4.3Dimensions/details
(2)
131
(1)
(3)
(6)(7)
(4)
Ø76
76
170
69
G 1/2
(5)
(8)
(9)
(10)
G 1/2
NTS 527
(1)Cover for electrical connection(6)Electrical connection,
M 20x1.5 cable gland
(2)PE connection(7)Blanking plug (electrical connection or
-30 to +120°C
(with halogenized filling oil: -40 to +85°C)
-20 to +85°C, to DIN 16086
(at temperatures below -20°C, the LCD display may not be readable).
60 bar:stainless steel, Mat. Ref. 1.4571, 1.4542
Overload
limit
Bursting
pressure
12
4 x full scale
2 x full scale with range
10 x full scale, 2000 bar max.
≥
100 bar
5 Instrument description
5.1.2For pressure transmitters with intrinsic safety
Explosion
protection
EC TypeExamination
Certificate
Permissible
ambient
temperature
EEx ia IIC T4—T6, to EN 50 014 and EN 50 020 (CENELEC)
PTB 98 ATEX 2194
Temperature class
T4T5T6
+85°C+75°C+60°C
13
6 Mounting
6.1General
Open cover flapThe cover flap can be opened with a coin.
The keys for operating the JUMO dTRANS p02 are underneath the cover flap.
For details on the setting options see:
➩
"The level concept", page 32ff.
✱
List of
functions
Affix one of the enclosed lists of functions to the cover flap.
6.2Prior to mounting
De-pressurize the system befor e mounting the
JUMO dTRANS p02 pressure transmitter!
The installation site should be readily accessible, as close as possible to the measurement point, and largely free from vibration. The
permissible ambient temperature limits must be observed (note
possible radiant heat).
The JUMO dTRANS p02 pressure transmitter may be mounted
above or below the pressure tapping.
The measurement arrangements below serve as an example and
can be altered according to the requirements of the measurement
task.
14
6 Mounting
Operating
position
The normal operating position of the JUMO dTRANS p02 pressure transmitter
is vertically upright.
Depending on the conditions of the measurement site, the pressure transmitter can also be mounted in a different orientation. In accordance with the
required operating position, the LC D can be rotated in 90° steps, see
➩
"Rotating the instrument", page 42.
6.3Mounting by the pressure connection
6.4Mounting by the housing
There are M 20x1. 5 threads on either side of the housing. One of the threads
is used to screw in the cable gland (optionally right or left). The other thread
can be used to mount the pressure transmitter.
p
15
6 Mounting
6.5Pressure connection
SealsSeals to EN 837 or DIN 16 258 may be used.
Tightening
torque
Check for
tightness
200 Nm max.
The correct tightening torque depends on the size, materi al and shape of the
seal that is used, as well as on the pressure connection of the pressure transmitter.
After establishing the pressure connection, it must be checked for tightness.
Note the sequence when opening or closing the valves!
➩
"Measuring the absolute pressure", page 26ff
or
➩
"Measuring the gauge pressure", page 29ff.
6.6Use inside the Ex area
.
Seals
With intrinsically safe protection EEx ia IIC, when used in Zone 0
with process connection 504, a flat metal gasket ring, e.g. to
EN 837 or DIN 16 258, must be used
16
7.1Electrical connection
Earth the instrument!
HARTTEST
+
-
1
2
+
-
✱
Unscrew the cover from the housing
➩
"Unscrewing the bezel ring or housing cover", page 41.
7 Installation
General notes- cable diameter 6 to 12 mm
- conductor cr oss- se c tion up t o 1.5mm
- route signal lines separately from cables with voltages above 60 V
- use shielded cables with twisted cores
- avoid the neighborhood of large electrical systems
- the full specification according to HART
with shielded cables
2
‚ version 5.1, will only be achieved
7.1.1Connection with screw terminals/Harting connector Han 7D/Han 8D (8U)
Connection
diagram
HARHARTTEST
-
+
8
2+
1
2
+
-
1-
Screw terminalsHarting connector
17
7 Installation
Ω
ConnectionTerminals
Screw terminalsAssignment
Supply
11.5 — 36 V DC
11.5 — 30 V DC for
intrinsically safe version
Output
4 — 20 mA 2-wire
proportional current 4 to 20 mA in supply
Test connection for current output
internal resistance of ammeter
≤
Ω
10
Test connection for HART
®
burden must be present!
Potential equalization
(with intrinsically safe circuit)
1 L+
2 L-
1 L+
2 L-
1 L+
2 L-
1 L+
2 L-
TEST +
TEST -
TEST +
®
HART
Shield8
Connection
HARTTEST
+
-
1
2
+
-
1
Burden
1
Burden ≤ (UB - 11.5 V) / 0.022 A
Burden with HART
-
Supply
voltage
11.5—36 VDC
+
®
min. 250 Ω, max. 1100
18
✱
Unscrew the housing cover
➩
"Unscrewing the bezel ring or housing cover", page 41.
✱
Insert the cable through the cable gland.
✱
Make the connection according to the connection diagram.
✱
Close the housing cover.
7.2Testing the output signal
HARTTEST
+
-
1
2
+
-
7 Installation
Output signal
-
+
Burden
1
1
Burden ≤ (UB - 11,5 V) / 0.022 A
Burden with HART
®
min. 250 Ω, max. 1100 Ω
test
4—20 mA
-
Supply
voltage
11.5—36 VDC
+
Test connection via faston connectors 2.8 x 0.8 mm or test terminals.
7.3Connection of a HART® modem
1L+
2L-
I
®
HART
modem
RS-232-C
up to 15 transmitters
(multidrop mode)
1
Burden ≤ (UB - 11.5 V) / 0.022 A
Burden with HART
®
min. 250 Ω, max. 1100
P
Pressure transmitter
JUMO dTRANS p0 2
Ω
Burden
PC or note-
book
+
-
1
Setup
program
Supply voltage
11.5 —36 V DC
19
7 Installation
7.4Connection of a HART® communicator
P
up to 15 transmitters
(multidrop mode)
1
Burden ≤ (UB - 11.5 V) / 0.022 A
Burden with HART
®
min. 250 Ω, max. 1100
Pressure transmitter
JUMO dTRANS p0 2
Ω
7.5Testing the HART
1L+
2L-
I
®
HART
communicator
®
communication
HART
Burden
®
communicator
or
HART® modem
+
-
1
Supply voltage
11.5 —36 V DC
-
Supply
voltage
HARTTEST
+
-
1
2
+
-
1
Burden
1
Burden ≤ (UB - 11.5 V) / 0.022 A
Burden with HART
®
min. 250 Ω, max. 1100 Ω
11.5 — 36 VDC
+
Test connection via faston connectors 2.8 x 0.8 mm or test terminals.
20
7 Installation
7.6Electrical connection in the hazardous (Ex) area
GeneralThe appropriate regulations must be observed when making the electrical
connection; inside the area with an explosion hazard this is, in particular:
-the regulations concerning electrical equipment in areas with an explosion hazard
(Elex V)
- the regulation for the installation of electrical equipment in areas with an
explosion hazard EN 60 079-14:1997 (VDE 0165)
- EC Type Examination Certificate
The HART® modem must not be used in the hazardous area!
The supply must be intrinsically safe and must not exceed the
following maximum values:
: 30 V DCIi: 100 mAPi: 750 mW
U
i
The connection of the HART® communicator or the HART
modem is optional.
For fault-free communication, the signal circuit must be loaded
with a minimum burden of 250
the burden is in most cases already integrated.
7.6.1Connection diagram “EX”
Site
+
-
1
2
+
-
®
Ω
. When supply isolators are used,
Control room
intrinsically
safe circuit
(2) +(3)
R
HARTTEST
(1)
not intrinsically safe
connections
+
-
+
Supply voltage
-
+
-
Output
Other
units
(2)
R
PC
or
notebook
®
HART
communicator
intrinsically safe
HART® communicator
intrinsically safe
RS-232-C
HART
modem
®
(1) Supply unit with isolating transformer for connecting a transmitter with
explosion protection
®
(2) Total resistance for HART
Ω
1100
max. The current limiting resistor which is integrated in the supply
communicator or HART® modem: 250Ω min.
unit must be included in this calculation.
21
7 Installation
PE connectionFor operating the pressure transmitter in the Ex area, it must be earthed via the
external ground terminal. In addition, the ground termi nal inside the housing
can be used to connect the cable shielding.
22
8.1Characteristic
8 Commissioning
Output signal
and display
Example:
Nominal range 0 — 600 mbar.
Range setting 50 — 450 mbar.
Output current
[mA]
Fault current
(21.5 or 3.85)
x (x < -600)
Range end
(meas. error)
Underrange
(meas. error)
Err
Saturation
(low)
21.50
20.00
4.00
3.85
0
Range start
(nominal)
Below
meas. start
(saturation)
17.5
50
Meas. start
(zero)
Meas. end
(zero+span)
Normal range
124.5
mbar
450
Saturation
(high)
600
Range end
(nominal)
Meas. end
exceeded
(saturation)
517.5
Fault current
(21.5 or 3.85)
x (x > 600)
Range end
(meas. error)
Overrange
(meas. error)
Err
Pressure
[mbar]
Range limitsDepending on the range setting and the applied pressure, the output signal
can lie within different ranges:
- normal range (the example shows the linear correlation between pressure
and output signal)
- saturation range
- measurement error range
The different ranges are shown in the display as different symbols,
➩
see above
23
8 Commissioning
8.2Self-test
ActiveAfter the supply has been switched on, the pressure transmitter carries out a
self-test.
OKIf the result of the self-test is positive, the display that was most recently
selected at the display level will be shown after approx. 20 seconds (on new
pressure transmitters, the display “pressure with unit”).
➩
"The level concept", page 32ff.
The 4 — 20 mA signal that is available on the output side is proportional to
pressure.
24
ErrorOn over/underrange, an error signal appears,
➩
see below
and
➩
"Trouble-shooting", page 43.
8 Commissioning
25
8 Commissioning
8.3Measuring the absolute pressure
Gases
Pressure
application
Transmitter above
the pressure tapping
(standard arrangement)
(1) Transmitter
(2) Shut-off fitting
2A shut-off valve to process
2B shut-off valve for test connection
(3) Pressure line
(4) Shut-off valve
(5) Shut-off valve
(6) Condensation vessel
(7) Purge valve
Initial state: all valves closed
Operate the shut-off fittings in the following order:
✱
Open shut-off valve (4) at the pressure tapping.
✱
Open shut-off valve (2A).
✱
Check measurement start.
✱
Close shut-off valve (2A).
✱
Open shut-off valve (2B).
✱
Apply a pressure that corresponds to the measurement start
to the transmitter, via the test connection of the shut-off fitting (2).
✱
Check output current at the measurement start and correct it if required,
➩
"Output current, measurement start and end", page 38.
✱
Close shut-off valve (2B).
✱
Open shut-off valve (2A).
Transmitter below
the pressure tapping
(exception)
26
Va pors
8 Commissioning
(1) Transmitter
(2) Shut-off fitting
2A shut-off valve to process
2B shut-off valve for test connection
(3) Pressure line
(4) Shut-off valve
(5) Blow-down valve
(8) Pressure equaliza tion reservoir
Pressure
application
Initial state: all valves closed.
Operate the shut-off fittings in the following order:
✱
Open shut-off valve (4) at the pressure tapping.
✱
Wait until steam has condensed in the pressure line.
✱
Open shut-off valve (2A).
✱
Check measurement start.
✱
Close shut-off valve (2A).
✱
Open shut-off valve (2B).
✱
Open drain / purge valve on the transmitter (1)
and drain liquid.
✱
Close drain / purge valve on the transmitter (1).
✱
Apply a pressure that corresponds ot the measurement start to the
transmitter (1), via the test connection of the shut-off fitting (2).
✱
Check output current at the measurement start and correct it, if required
➩
"Output current, measurement start and end", page 38.
✱
Close shut-off valve (2B).
✱
Open shut-off valve (2A).
27
8 Commissioning
Liquids
(1) Transmitter
(2) Shut-off fitting
2A shut-off valve to process
2B shut-off valve for test connection
Pressure
application
(3) Pressure line
(4) Shut-off valve
(5) Blow-down valve
Initial state: all valves closed
Operate the shut-off fittings in the following order:
✱
Open shut-off valve (4) at the pressure tapping.
✱
Open shut-off valve (2A).
✱
Check measurement start.
✱
Close shut-off valve (2A).
✱
Open shut-off valve (2B).
✱
Open drain/ purge valve on the transmitter (1) and drain liquid.
✱
Close drain/ purge valve on the transmitter (1).
✱
Apply a pressure that corresponds to the measurement start to the
transmitter (1), via the test connection of the shut-off fitting (2).
✱
Check the output current at the measurement start and correct it, if required
➩
"Output current, measurement start and end", page 38.
28
✱
Close shut-off valve (2B).
✱
Open shut-off valve (2A).
8.4Measuring the gauge pressure
Gases
8 Commissioning
Pressure
application
Transmitter above
the pressure tapping
(standard arrangement)
(1) Transmitter
(2) Shut-off fitting
2A shut-off valve to process
2B shut-off valve for test connection
or vent screw
Initial state: all valves closed
Operate the shut-off fittings in the following order:
✱
Open shut-off valve (2B).
✱
Apply a pressure that corresponds to the measurement start to the
transmitter, via the test connection of the shut-off fitting (2).
✱
Check measurement start and correct it, if required
➩
"Output current, measurement start and end", page 38.
✱
Close shut-off valve (2B).
✱
Open shut-off valve (4) on the pressure tapping.
Transmitter below
the pressure tapping
(exception)
Initial state: all valves closed
Operate the shut-off fittings in the following order:
✱
Open shut-off valve (2B).
✱
Apply a pressure that corresponds to the measurement start to the
transmitter, via the test connection of the shut-off fitting (2).
✱
Check measurement start and correct it, if required
➩
"Output current, measurement start and end", page 38.
✱
Close shut-off valve (2B).
✱
Open shut-off valve (4) on the pressur e tapping.
✱
Open shut-off valve (2A).
30
Liquids
8 Commissioning
(1)Transmitter
(2)Shut-off fitting
2A shut-off valve to process
2B shut-off valve for test connection or vent screw
Pressure
application
(3)Pressure line
(4)Shut-off valve
(6) Blow-down valve
Initial state: all valves closed
Operate shut-off fitting in the following order:
✱
Open shut-off valve (2B).
✱
Apply a pressure that corresponds to the measurement start to the
transmitter, via the test connection of the shut-off fitting (2).
✱
Check measurement start and correct it, if required
➩
"Output current, measurement start and end", page 38.
✱
Close shut-off valve (2B).
✱
Open shut-off valve (4) on the pressure tapping.
✱
Open shut-off valve (2A).
31
9 Adjustment options
9.1The level concept
Two levelsIn order to make the operation of the pressure transmitter simple and transpar-
ant, the functions of the JUMO dTRANS p02 have been divided into two levels.
Both levels can be accessed from the keypad of the transmitter.
The keypad can be found underneath the cover fl ap of the pressure transmitter.
➩
"Open cover flap", page 14.
As an alternative to operation from the keys, all the measured values and parameters can be easi ly displayed and / or set up us ing
the setup program. Th e setup program is available on request, at
extra cost
➩
"Setup program", page 7.
Display levelIn normal operation, the instrument is in the display level.
Parameter
level
The display options are: pressure, measurement, output current, sensor temperature or peak-reading pointer. Step thr ough with the or keys.
After applying the supply voltage, and after the end of the self-test,
the instrument is in the display level. The display that was most
recently selected will be shown (for new instruments ex-works, this
is “Pressure with unit”).
At this level, the para meters that can be checked or altered include the following: measurement start, measurement end, zero, dimensional unit, damping
and output current in a fault situation.
32
9.2Sequence for keypad operation
Layout
after complete run or press "P" key foror pause for > 120 sec> 3 sec
Pressure
with
dim. unit
or
Pressure values
and
sensor temp.
or
Max. pressure
or
Min. pressure
or
Temperature
[°C / °F]
or
Current
[mA]
or
Measurement
[%]
(freely scalable)
or
9 Adjustment options
Parameter levelDisplay level
(Pmin)
Reset
min. pressure
P
(Pmax)
Reset
max. pressure
P
(P0)
Density
correction
P
(P1)
Dimensional
unit
P
(P2)
Meas. start with
given pressure
P
(P3)
Meas. end with
given pressure
P
(P4)
Damping
P
(P5)
Meas. start
(blind setting)
P
(P6)
Meas. span
(blind setting)
> 1 s
> 1 s
> 1 s
> 1 s
> 1 s
> 1 s
> 1 s
> 1 s
(P7)
Zero
adjustment
P
(P8)
Current generator
P
(P9)
Current on fault
P
(P10)
Key inhibit
P
(P11)
Characteristic
(P12)
Start of
square root
(P13)
Version
(P14)
Display unit
temperature
> 1 s
P
> 1 s
P
> 1 s
P
> 1 s
> 1 s
> 1 s
> 1 s
P
> 1 s
P
> 3 s
P
> 1 s
Every altered value is accepted after modification ( signalled by a
brief display switch-off).
The function “Adjust zero point” is inactive on transmitters for
absolute pressure.
33
9 Adjustment options
9.3Display level
After switching on the pressure transmitter, the last display that
was selected will be shown at the display level.
Sequence
Action
1
FunctionDisplay
(Example)
Display of pressure with
dimensional unit
Display of pressure value
and sensor temperature
in the units selected
Display of the stored
maximum pressure
Display of the stored
minimum pressure
Display of the sensor
temperature in °C or °F
Display of the output
current in mA
Display of
measurement in %
2
or
scaled with freely
selectable dim. unit
1
The sequence can also be stepped through in the opposite direction
by using the key.
2 Factory setting: measurement in %. On request, freely scalable through setup program
(0—100% as -9999 to +9999). Unit ca n be s et (maximum 5 digits, restricted A SCI I
character set).
34
9.4Parameter level
9 Adjustment options
ActionFunctionDisplay
(Example)
Reset the stored
minimum
> 1
pressure
second
Reset the stored
maximum
> 1
pressure
second
Display and set
density
> 1
correction
second
Display and set
dimensional unit
> 1
of pressure
second
Possible
setting
reset
reset
0.100—5.000
mbar
bar
psi
mmHg
mmH
ftH
0
2
0
2
inHg
0
inH
2
mH
0
2
MPa
Torr
kPa
2
kgcm
with keys
+
> 2
seconds
+
> 2
seconds
or
oder
or
oder
➩
"Detailed explanations", page
38ff.
➩
"Peak-reading
pointer", page 38.
➩
"Peak-reading
pointer", page 38.
➩
"Densitiy correction", page 40.
➩
"Dimensional
unit", page 38.
> 1
second
> 1
second
> 1
second
Display or set
output current at
measurement
start (with
pressure input)
Display or set
output current at
measurement
end (with
pressure input)
Display or set
damping
(time constant)
3.85—
21.50mA
4mA
3.85—
21.50mA
20mA
0.0—100.0sec
or
oder
+
> 2
seconds
or
oder
+
> 2
seconds
oder
or
➩
"Output current,
measurement
start and end",
page 38.
➩
"Output current,
measurement
start and end",
page 38.
➩
"Damping",
page 38.
35
9 Adjustment options
ActionFunctionDisplay
(Example)
Display or set
measurement
> 1
second
start
(blind setting)
Display or set
measurement
> 1
second
span
(blind setting)
Zero
adjustment
> 1
second
(not for absolutepressure
transmitters)
Set and activate
current generator
> 1
second
Possible
setting
-110% to +110%1
or
-10 to +110%
of nominal
range
-210% to
+210%
of nominal
range
0.000
Select:
3.85 mA
4.00 mA
12.00 mA
20.00 mA
with keys
2
> 2
seconds
or
oder
oder
or
or
oder
➩
"Detailed explanations", page
38ff.
➩
"Measurement
start, blind setting", page 38.
➩
"Measurement
span, blind setting", page 38.
➩
+
"Zero adjustment", page 39.
➩
"Current generator", page 39.
21.50 mA
and:
activate
+
> 2
seconds
➩
in fault situation",
page 39.
> 1
Display and set
output current
in fault situation
21.5 mA
3.85 mA
or
oder
second
1
only with transmitters for relative pressure ➩ "Instrument identification", page 8ff.
2
only with transmitters for absolute pressure ➩ "Instrument identification", page 8ff.
"Output current
36
9 Adjustment options
ActionFunctionDisplay
(Example)
Display and
set key inhibit
> 1
second
Possible
setting
0:
no inhibit
LA:
all inhibited
LALL:
all inhibited
(adjustable via
setup
program only)
LS:
all inhibited
except
measurement
start and end
LO:
all inhibited
except
measurement
start
with keys
oder
or
➩
"Detailed explanations", page
38ff.
➩
"Keys inhibited", page 39.
> 1
second
> 1
second
> 1
second
> 1
second
Display and set
characteristic
Display and
set starting point
of square-root
characteristic in
percent of range
Display
instrument
version
Display and set
unit of
temperature
Lin:
linear
SLin:
square root,
linear up to
starting point
SoFF:
square root,
switched off
up to starting
point
5.0%—15.0%
°C
°F
oder
or
oder
or
or
oder
➩
"Characteristic", page 39.
➩
"Characteristic", page 39
➩
"Instrument version", page 40.
➩
"Unit of temperature displa y " ,
page 40.
37
9 Adjustment options
9.5Detailed explanations
Peak-reading
pointer
Dimensional
unit
Output current,
measurement
start and end
DampingTo dampen an unsteady display caused by rapidly changing pressures, the
Minimum and maximum pressures are saved. Both values can be individually
display e d and deleted.
If the damping is set to 0 .0sec, minimum and maximum pressures
will not be saved.
➩
"Detailed explanations", page 38.
If the damping is 0.0sec, or after a reset of the display,
“- - - - -” will be shown.
The pressure can be indicated directly in bar or in another pressure unit.
The output current (in mA) that is available at the start (or end) of the measurement can be set within the range +3.85 to +21.5mA. At the same time, the
pressure applied is saved as measurement start or end.
time constant for the display, T
At a damping of 0.0sec, the pressure transmitter will not save the
minimum and maximum pressures, i.e. the peak-reading pointer
function is switched off.
, can be adjusted from 0.0 to 100.0sec.
63
Measurement
start,
blind setting
Measurement
span,
blind setting
The measurement start can be set in % of the nominal pressure range. The
nominal measurement range is given on the nameplate.
➩
"Nameplate", page 8 .
Example:
The pressure transmitter has a nominal range of 0 to 25bar, but the output signal needs to be proportional in the range 5 to 20bar.
The necessary setting for the measurement start is derived as follows:
Nominal range:0—25bar
Meas. start required:5bar, corresponding to 20% of the nominal range
Setting:20%
The measurement span can be set in % of the nominal measurement range.
Example 1:
The pressure transmitter has a nominal measurement range from 0 to 25bar,
but the output signal needs to be proportional in the range 5 to 20bar.
The necessary setting for the measurement span is derived as follows:
Nominal range:0— 25bar
Range required:5—20bar => required span: 15bar
Span required:15bar, corresponding to 60% of the nominal range
Setting:60%
38
9 Adjustment options
Example 2:
Nominal range:0— 600mbar
Range required:-300 to +300mbar
Measurement start:-50%
Span:100%
Zero
adjustment
Current
generator
Output current
in fault situation
When a zero adjustment is made, the transmitter display is set to 0.000. The
preset measurement start remains unaffected.
Setting the measurement start:
➩
"Output current, measurement start and end", page 38.
This function is inhibited for absolute-pressure transmitters.
➩
"Nameplate", page 8.
If the “current generator” function is selected, an adjustable constant current
is present at the transmitter output.
There is no time-out for this function; the current can remain indefinitely.
This function can only be ended by pressing the key. It is not
possible in multidrop mode.
➩
"Connection of a HART® communicator", page 20. In this case,
a constant current of 4mA is present at the output.
In order to indicate measurement errors in the event of over /underrange, or
other errors, (e.g. for a connected PLC), the pressure transmitter gives out
either 3.85mA or 21.5mA.
P
➩
"Output signal and display", page 23.
Keys inhibitedTo prevent an unauthorized modification of parameters, some or all of the
parameters can be inhibited.
To remove an inhibit (LA, L0, LS), the key must be pressed for
longer than 5 seconds in the display “Output current in fault sit-uation” (P9).
The inhibit “LALL” can only be set or deleted in the setup program.
CharacteristicThe characteristic of the output current can be selected as:
❏
Lin:linear (proportional to pressure).
❏
SLin:linear up to the starting point, then square root
(proportional to flow-through),
❏
SoFF: square-root (proportional to flow), but switched off up to
the starting point.
The starting point of the square-root characteristic can be adjusted
between 5.0 and 15.0% of the output current.
P
39
9 Adjustment options
Characteristics
Output current
[%]
100
80
60
40
20
0
020406080
20
(b)
(a)
100
Pressure [%]
15
10
5
0
012 34
20
(b)
Starting point
linear
(a)
Instrument
version
Densitiy
correction
(a) = Lin= linear
(b) = SLin = square root, linear up to
15
10
(c)
Starting point
starting point
(c) = SoFF = square root, switched off
up to starting point
5
0
012 34
Display format: xx.yy
xx = hardware ve rsion
yy = software version
Density correction is important for the hydrostatic level measurement for
media with densities other than 1 (e.g. pure water).
Possible range setting: 0.100—5.000 (kg/dm
3
).
Example
Range: 0—1 bar
Scaling:0—100% => 0—5000 l
Medium DensityMeasured pressureOutput currentDisplay LCD
Water1.00.5 bar12 mA2500 liters
Gasoline0.80.5 bar14 mA3125 liters
Unit of
temperature
display
40
The display unit of the measured sensor temperature can be selected as:
❏
°C
❏
°F
10 Opening the instrument
10.1Unscrewing the bezel ring or housing cover
The bezel ring and the rear housing cover can be unscrewed.
(1)
(1) Housing cover
(2) Bezel ring
(2)
✱
Open with a screwdriver
✱
Close by hand without any tools
41
11 Rotating the instrument
11.1Rotating
Rotating the
LCD display
Depending on the operating position required, the LCD display can be rotated
in steps of 90°.
✱
Unscrew the bezel ring.
➩
"Unscrewing the bezel ring or housing cover", page 41.
✱
Use a small-bladed screwdriver to carefully push out the catch of the LCD
fixing and lever out the display board.
✱
Set the LCD display to the required position (90° steps) in the notches
provided.
Rotating the
housing
✱
Screw on the bezel ring finger-tight,
rotate housing
The housing can be rotated through ± 170° .
✱
Loosen the grub screw (1) with a 2 mm hexagon socket wrench
(about half a turn is enough).
✱
Rotate the housing into the required po sition.
(1)
42
✱
Screw up grub screw tightly.
12 Maintenance
12.1Trouble-shooting
Error/faultPossible causeRemedy
Display:noneno supply voltageswitch on supply
instrument faultysend instrument to supplier for
repair
Display:overrange,
excessive pressure
Display:underrange,
pressure to o low
Display:pressure can no longer be
measured; overrange, excessive
pressure
instrument faultysend instrument back to the
Display:pressure can no longer be
measured; overrange, pressure
too low.
instrument faultysend instrument to supplier for
bring pressure back to within
range, or adjust range
bring pressure back to within
range, or adjust range
bring pressure back to within
range, or use instrument with a
wider nominal range
supplier for repair
bring pressure back to within
range, or use instrument with a
wider nominal range
repair
Display:an error in the electronics was
detected during the last self-test
Display:temperature probe faultysend instrument to supplier for
Display:span too smallincrease span
instrument faultysend instrument to supplier for
send instrument to supp l ier for
repair
repair
(at least 5%)
➩
"Measurement start, blind set-
ting", page 38.
repair
43
12 Maintenance
Display:span too smallincrease span
(at least 5%)
➩
"Measurement start, blind set-
ting", page 38.
instrument faultysend instrument to supplier for
repair
key not respondinginstrument faultysend instrument to supplier for
P
repair
key
or
key not responding
keys inhibitedreset key inhibit
➩
"Keys inhibited", page 39.
instrument faultysend instrument to supplier
for repair
1
You can find the supplier addresses on the back page of the operating
instructions.
1
44
12 Maintenance
45
13 Appendix
13.1EC Type-Examination Certificate
46
Translation (for information only):
Physikalisch-Technische BundesanstaltPTB
(Federal Physical-Technical Laboratory)
Braunschweig and Be rlin
13 Appendix
(1)
(2)Equipment and Protective Systems Intended for Use in
(4)Equipment:Pressure transmitter Type dTRANS p02
(5)Manufacturer:M. K. JUCHHEIM GmbH & Co.
(6)Address:Moltkestraße 13/31, D-36039 Fulda, Germany
(7)This equipment and any acceptable variation thereto are specified in the
schedule to this Type-Examination Certificate.
(8)The Physikalisch-Technische Bundesanstalt, notified body No. 0102 in accordance
with Article 9 of the Council Directive 94/9/EC of 23 March 1994, certifies that this
equipment has been found to comply with the Essential Health and Safety Requirements
relating to the design and construction of equipment and protective systems intended
for use in potentially explosive atmospheres, given in Annex II to the Directive.
The examination and test results are reco rded in the confidential report PTB Ex 98-28372.
(9)Compliance with the Essential Health and Safety Requirements has been assured
by compliance with
EN 50 014:1997EN 50 020:1994EN 50 284:1998
EC Type-Examination Certificate
PTB 98 ATEX 2194
(10)If the sign “X” is placed after the certificate number, it indicates that the equipment
is subject to special conditions for safe use specified in the schedule to this certificate.
(11)Thi s EC Type-Examination Certificate relates only to the design and construction of
the specified equipment in accordance with Directive 94/9/EC. Further requirements
of this Directive apply to the manufacture and supply of this equipment.
(12)The marking of the equipment shall include the following:
ExII 1/2GEEx ia IIC T6
Zertifizierungsstelle ExplosionsschutzBrunswick, 11 December 1998
(Certifying Body for Explosion Protection)
signed
Dr.-Ing. U. Johannsmeyer(stamp)
RegierungsdirektorPage 1/3
EC Type-Examination Certificates are not valid without signature and stamp. The certificates may
only be circulated without alteration. Extracts or alterations are subject to the approval by the
Physikalisch-Technische Bundesanstalt.
The pressure transmitter Type dTRANS p02 is used to convert a physical measurement
variable into a standard electrical signal. The transmitter is intended for use in potentially
explosive atmospheres.
The electronics housing is installed in hazardous zones which require category 2
equipment. The elements for the process connection are installed in the separating wall
which separates the zones requiring category 2 equipment from those with category 1.
The maximum permissible ambient temperature is related to the temperature class as
follows:
Electrical data
Supply circuitintrinsically safe EEx ia IIC
(terminals X401+, X402-)only for connection to a certified intrinsically safe circuit
Schedule
Temperature classAmbient temperature
T485°C
T575°C
T660°C
maximum values:U
Sensor circuit and
display circuitintegral to device, intrinsically safe protection EEX ia IIC
Examination and test report:
EC Type-Examination Certificates are not valid without signature and stamp. The certificates may
only be circulated without alteration. Extracts or alterations are subject to the approval by the
Physikalisch-Technische Bundesanstalt.
Braunschweig and Berlin
Schedule to EC Type-Examination Certificate PTB 98 ATEX 2194
(17)Special conditions
not applicable
(18)Essential health and safety requirements
are assured by compliance with the standards cited above
Zertifizierungsstelle Explosionsschutz
(Certifying Body for Explosion Protection)
signedBrunswick, 11 December 1998
Dr.-Ing. U. Johannsmeyer(stamp)
Regierungsdirektor
Page 3/3
EC Type-Examination Certificates are not valid without signature and stamp. The certificates may
only be circulated without alteration. Extracts or alterations ar e subject to the approval by the
Physikalisch-Technische Bundesanstalt.
to EC Type-Examination Certificate PTB 98 ATEX 2194
Equipment:Pressure transmitter Type dTRANS p02
Marking:Ex II 1/2G EEx ia IIC T6
Manufacturer:M. K. Juchheim GmbH & Co.
Address:Moltkestraße 13 - 31
D-36039 Fulda, Germany
13 Appendix
Description of additions and alterations
The pressure transmitter Type dTRANS p02 will in future be manufactured and operated in
accordance with the test documentation listed in the test report. The alterations concern the
internal and external construction, for the introduction of other equipment variations, as well as to
the “Electrical data” and the type designation.
For the equipment version with differential pressure sensor this is:
Pressure transmitter Type dTRANS p02 DELTA
Electrical data
Supply circuitintrinsically safe protection EEx ia IIC
(terminals P601+, P602-,
or plug connection)only for connection to a certified intrinsically safe circuit
maximum values:U
Display circuitintegral to device, in intrinsically safe protection EEx ia IIC
Sensor circuitintrinsically safe protection EEx ia IIC
integral to device, or only for connection to the corresponding sensor through the appropriate attached cable
= 30 V
i
I
100 mA
i =
750 mW
P
i =
C
=11 nF
i
L
= 170 µH
i
All other data and specifications remain unaffected.Page 1/2
EC Type-Examination Certificates are not valid without signature and stamp. The certificates may
only be circulated without alteration. Extracts or alterations are subject to the approval by the
Physikalisch-Technische Bundesanstalt.
Braunschweig and Be rlin
Supplement 1 to EC Type-Examination Certificate PTB 98 ATEX 2194
13 Appendix
Test and examination report:
Zertifizierungsstelle ExplosionsschutzBrunswick, 5 April 2001
(Certifying Body for Explosion P rotection)
signed
Dr.-Ing. U. Gerlach
PTB Ex 01-20368
(stamp)
Page 2/2
EC Type-Examination Certificates are not valid without signature and stamp. The certificates may
only be circulated without alteration. Extracts or alterations are subject to the approval by the
Physikalisch-Technische Bundesanstalt.
Measurement span (blind setting)P6 %
Zero adjustmentP70.000
Individual settings can be recorded in this table.
setting
or mbar
1
0
1
100
Customer’s
setting
not adjustable
not adjustable
Current generatorP8 mA3.85
Output current in fault situationP9 mA21.50
Key / function lockingP100
CharacteristicP11Lin
Starting point (EP = Entry Point)P12 %9.400
Version displayP1301.12
Display unit for temperatureP14°C
1
Different factory settings are possible in accordance with customers’ specification!
not adjustable
58
M. K. JUCHHEIM GmbH & Co
Street address:
Moltkestraße 13 - 31
36039 Fulda, Germany
Delivery address:
Macken ro d tstr aß e 1 4
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone:+49 661 60 03-0
Fax:+49 661 60 03-6 07
E-mail:mail@jum o.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Ban k, Riverw ay
Harlow, Essex CM20 2TT, UK