JUMO dTRANS CR 02 Operating Manual

JUMO dTRANS CR 02
Transmitter/controller for conductivity, TDS,
resistance, temperature and standard signals
Type 202552
Operating Manual
20255200T90Z002K000
V2.00/EN/00541516
WARNING:
EXIT
PGM
PGM
PGM
Note:
Please read these Operating Instructions before placing the instrument in operation. Keep the manual in a place which is accessible to all users at all times.
Resetting the brightness of the LC display:
If the brightness setting has been adjusted so that the display text is no longer legible, the basic setting can be restored as follows:
Switch off the voltage supply.Switch on the voltage supply and immediately press and hold the and
keys simultaneously.
Operator language selection:
Press the key for longer than 3 seconds.Select the appropriate language with the and keys.Briefly press the key.
Reset to factory settings:
To get to the Administrator level, proceed as follows:
Press the key for longer than 2 seconds.Use the or keys to select "ADMINISTR. LEVEL".Use the and keys to enter the password 8192.
Confirm the key.
WARNING:
Customer-specific settings will be lost!
Contents
1 Typographical conventions ...................................................... 7
1.1 Warning signs ..............................................................................................7
1.2 Reference signs ...........................................................................................7
2 Description ................................................................................ 8
3 Identifying the device version ................................................ 10
3.1 Nameplate ..................................................................................................10
3.2 Order details ...............................................................................................10
3.3 Accessories (included in delivery) ..............................................................12
3.4 Accessories (optional) ................................................................................12
4 Mounting .................................................................................. 13
4.1 General information ....................................................................................13
4.2 Dimensions ................................................................................................13
5 Installation ............................................................................... 14
5.1 Installation instructions ..............................................................................14
5.2 Electrical isolation ......................................................................................15
5.3 Connection .................................................................................................16
5.3.1 Terminal assignment ..................................................................................16
5.3.2 Optional board (row 1, slot a, b or c) .........................................................16
5.3.3 Main board (row 2) .....................................................................................18
5.3.4 PSU board (row 3) ......................................................................................19
6 Operation ................................................................................. 20
6.1 Controls ......................................................................................................20
6.2 Display .......................................................................................................21
6.2.1 Measuring mode (normal display) ..............................................................21
6.3 Principle of operation .................................................................................22
6.3.1 Operation in levels .....................................................................................22
6.4 Measuring mode ........................................................................................25
6.4.1 Normal display ...........................................................................................25
6.5 Input/output information ............................................................................26
6.5.1 User data ....................................................................................................27
6.5.2 Min/max values of the main input ..............................................................27
6.5.3 Min/max values of the optional inputs .......................................................28
6.5.4 Output level ................................................................................................28
6.5.5 Current values of the main entries .............................................................28
6.5.6 Curgent values of the optional entries .......................................................29
6.5.7 Current values of the math channels .........................................................29
6.5.8 States of the binary inputs and outputs .....................................................29
Contents
6.5.9 Manual mode overview ..............................................................................30
6.5.10 Hardware info .............................................................................................30
6.5.11 Device info .................................................................................................31
6.6 User level ...................................................................................................31
6.6.1 Parameters of the User level ......................................................................32
6.7 Administrator level .....................................................................................32
6.7.1 Parameter level ..........................................................................................32
6.7.2 Release level ..............................................................................................32
6.7.3 Basic setting ..............................................................................................32
6.7.4 Calibration level ..........................................................................................35
6.7.5 Calibration release .....................................................................................35
6.7.6 Delete min/max values ...............................................................................35
6.7.7 Delete logbook ...........................................................................................35
6.7.8 Delete daily batch ......................................................................................35
6.7.9 Delete total batch .......................................................................................35
6.8 MANUAL mode/Simulation mode ..............................................................36
6.8.1 MANUAL mode only via "higher order" controller functions .....................36
6.8.2 Simulation of binary outputs ......................................................................37
6.8.3 Simulation of analog outputs via MANUAL mode .....................................38
6.9 HOLD mode ...............................................................................................38
7 Commissioning ....................................................................... 40
7.1 Getting started ...........................................................................................40
7.2 Setting examples .......................................................................................41
7.2.1 Conductivity measurement, temperature compensated ...........................41
7.2.2 Measurement of ultra-pure water with 2-electrode measuring sensor ......43
7.2.3 Measurement of ultra-pure water with 2-electrode measuring sensor ......45
8 Calibrating a conductivity sensor .......................................... 47
8.1 Notes ..........................................................................................................47
8.2 General information ....................................................................................47
8.2.1 Measurements in highly-purified water ......................................................47
8.2.2 Requirements .............................................................................................48
8.2.3 Ways to start the calibration ......................................................................48
8.2.4 Calibration options .....................................................................................48
8.3 Calibration of the temperature coefficient of the sample medium .............49
8.4 Calibrating the relative cell constant ..........................................................51
8.4.1 Entering the cell constant manually ...........................................................52
8.4.2 Cell constants ............................................................................................52
9 Calibrating a sensor with a standard signal ......................... 53
9.1 General information ....................................................................................53
Contents
9.1.1 Operating modes .......................................................................................53
9.1.2 Calibration options .....................................................................................54
9.1.3 Ways to start the calibration ......................................................................54
9.2 Linear operating mode ...............................................................................55
9.2.1 1-point calibration ......................................................................................55
9.2.2 2-point calibration ......................................................................................56
9.2.3 Calibration limit point .................................................................................58
9.3 pH operating mode ....................................................................................59
9.3.1 Zero-point (1-point) calibration ..................................................................59
9.3.2 2-point calibration ......................................................................................61
9.4 Conductivity operating mode .....................................................................63
9.4.1 Calibration of the relative cell constant ......................................................63
9.4.2 Calibration of the temperature coefficient .................................................65
9.5 Concentration operating mode ..................................................................69
9.5.1 Calibration of the relative cell constant ......................................................69
9.6 Chlorine measurement operating mode,
pH-compensated .......................................................................................71
9.6.1 Final value calibration ................................................................................71
10 Calibration logbook ................................................................ 73
10.1 General information ....................................................................................73
11 Controller ................................................................................. 74
11.1 General information ....................................................................................74
11.2 Controller functions ....................................................................................74
11.2.1 Simple switching functions ........................................................................74
11.2.2 Higher order switching functions (PID) ......................................................74
11.2.3 Typical operator level parameters ..............................................................75
11.3 Software controllers and outputs ...............................................................75
11.4 Configuration of higher order controllers ...................................................77
11.4.1 Structure ....................................................................................................77
11.5 Parameter sets ...........................................................................................77
11.6 Sample configurations ...............................................................................78
11.6.1 Simple limit monitoring ..............................................................................78
11.6.2 Limit monitoring to USP .............................................................................78
11.6.3 Controller with limit value function .............................................................79
12 Setup program ........................................................................ 80
12.1 Configurable parameters ...........................................................................80
12.2 Documenting the instrument configuration ................................................81
12.3 Special features for "Datalogger" ..............................................................82
Contents
13 Eliminating faults and malfunctions ...................................... 84
14 Technical data ......................................................................... 86
15 Retrofitting optional boards ................................................... 90
16 Appendix .................................................................................. 93
16.1 Glossary .....................................................................................................93
16.2 Parameters of the User level ....................................................................105
17 Index ....................................................................................... 116

1.1 Warning signs

Danger
This symbol is used when there may be danger to personnel if the instructions are ignored or not followed correctly!
Caution
This sign indicates that components could be destroyed by electrostatic discharge(ESD=Electro Static Discharge), if the respective cautionary measu­res are not taken.Only use the ESD packages intended for this purpose to return device inserts, assemblygroups or assembly components.
Caution
This symbol is used when there may be damage to equipment or data if the instructions are ignored or not followed correctly!

1 Typographical conventions

Read documentation! This symbol – placed on the device – indicates that the associated device
documentation has to be observed. This is necessary to recognize the kind
of the potential hazards as well as to take the measures to avoid them.

1.2 Reference signs

Note This symbol is used to draw your special attention to a remark.
Instruction
This symbol indicates the description of an action to be performed. The individual steps are marked by this asterisk. Example: Briefly press the key.
7

2 Description

Power supply
Optional board 2
Optional board 1
Optional board 3
2 Binary inputs
1 Main input
(conductivity, TDS and resistance)
1 Analog input
(compensation)
Setup interface
2 Relay (changeover)
Power supply
for a 2-wire transmitter
Inputs/outputs In addition to the main input (conductivity, TDS, resistance) and the secondary
input (temperature compensation), the basic instrument alone has two binary inputs, two relays, one voltage supply for external sensors and a setup interface.
Input signals can be shown as numbers or as a bar graph on the graphic display. Parameters are displayed in plain text for easily comprehensible and reliable operation.
Optional Three further slots can be fitted with extensive additional configurable inputs
and outputs and interfaces.
Application The instrument is suitable, for example, for displaying, measuring and
controlling:
- Conductivity, TDS and resistance.
- Free chlorine, chlorine dioxide, ozone, hydrogen peroxide and peracetic acid, in combination with sensors as per data sheet 202630.
- (Hydrostatic) liquid levels with 2-wire transmitters (level probes) as per data sheet 402090 or data sheet 404390.
- Flow rate in conjunction with transmitters as per data sheet 406010 or
406020.
- Two temperature measuring points.
- Most sensors and transmitters that output standard signals (0 to 10 V or 0(4) to 20 mA).
Because temperature measurement is integrated, temperature compensation takes place quickly and precisely, which is particularly important for many analytical measurements.
8
2 Description
Special features - Display: mS/cm, µS/cm, MOhm × cm, mg/l, pH, mV, etc.
Special settings are also possible with the setup program
- Configurable display text (operator level)
- A choice of display visualizations: large numbers, bar graph or tendency (trend) display
- Four limit controllers
- Integrated calibration routines: with 1, 2 and 3 points
- Math and logic module (optional)
- Calibration logbook
- Three optional slots
- Selectable languages: English, German, French, etc.
- Setup program provides: convenient programming, system documentation
- RS422/485 interface (optional)
- PROFIBUS-DP interface (optional)
9

3 Identifying the device version

Typ: 202552/01-8-02-0-0-0-25/000
TN: 00617042
JUMO GmbH & Co. KG dTRANS CR 02
36039 Fulda Germany
AC .. 0V 48..63Hz/DC 20 3
max 14VA
F-Nr.: 0168122901016010001

3.1 Nameplate

on the transmitter
The date of manufacture is encoded in the "F No." (serial number): 1601 means year of manufacture 2016, calendar week 01

3.2 Order details

(1) Basic type
202552 JUMO dTRANS CR 02 - Transmitter/controller
(2) Basic type extension
01 In the panel enclosure 05 In the surface-mounted enclosure
(3) Version
8 Standard with factory setting 9 Programming to customer specification
(4) Operating language
01 German 02 English 03 French 04 Dutch 05 Russian 06 Italian 07 Hungarian 08 Czech 09 Swedish 10 Polish 13 Portuguese 14 Spanish 16 Rumanian
a
10
3 Identifying the device version
(5) Optional slot 1
0Not used 1 Analog input (universal) 2 Relay (1× changeover) 3 Relay (2× normally open) 4 Analog output 52 PhotoMOS
®
relays 6 Solid state relay 1 A 8 Voltage supply output DC 12 V (e.g. for inductive proximity switch)
(6) Optional slot 2
0Not used 1 Analog input (universal) 2 Relay (1× changeover) 4 Analog output 52 PhotoMOS
®
relays 6 Solid state relay 1 A 8 Voltage supply output DC 12 V (e.g. for inductive proximity switch)
(7) Optional slot 3
0Not used 1 Analog input (universal) 2 Relay (1× changeover) 3 Relay (2× normally open) 4 Analog output 52 PhotoMOS
®
relays 6 Solid state relay 1 A 8 Voltage supply output DC 12 V (e.g. for inductive proximity switch)
10 RS485 interface 11 Datalogger with interface RS485 12 PROFIBUS-DP interface
(8) Voltage supply
23 AC 110 to 230 V, +10/-15 %, 48 to 63 Hz 25 AC/DC 20 to 30 V, 48 to 63 Hz
(9) Extra codes
0None
a
All languages are available on the instrument and can be changed by the customer at any time. Factory default setting to a language (other than "German") is available for a charge.
b
PhotoMOS® is a registered trademark of Panasonic Corporation.
c
The only way to read files is with the PC setup software!
b
c
(1) (2) (3) (4) (5) (6) (7) (8) (9) Order code /- -----/ Order example 202552 / 01 - 8 - 01 - 2 - 2 - 4 - 23 / 000
11
3 Identifying the device version

3.3 Accessories (included in delivery)

4× fastening elements, complete 3× CON plug-in link 3× jumper wire 1× seal for panel 1× fastening elements, complete
- 1× DIN rail fastening left
- 1× DIN rail fastening right
- 3× wall mount
- 3× fastening screw
a
For basic type extension 01 only (in the panel enclosure)
b
For basic type extension 05 only (in the surface-mounted enclosure)
a
b
a
a
b

3.4 Accessories (optional)

Type Part no.
Holder for C rail 00375749 Dummy cover 96 mm × 48 mm 00069680 Pipe mounting set 00398162 Weather protection roof complete for basic type extension 05 00401174 PC setup software 00560380 PC interface cable including USB/TTL converter and two adapters
(USB connecting cable)
00456352
Optional board Code Part no.
Analog input (universal) 1 00442785 Relay (1× changeover) 2 00442786 Relay (2× NO) 3 00442787 Analog output 4 00442788 2 PhotoMOS Solid state relay 1 A 6 00442790 Voltage supply output DC ±5 V (e.g. for ISFET) 7 00566681 Voltage supply output DC 12 V (e.g. for inductive proximity switch) 8 00566682 Interface - RS422/485 10 00442782 Datalogger with RS485 interface 11 00566678 PROFIBUS-DP interface 12 00566679
®
relays 5 00566677
12

4.1 General information

4 Mounting

Mounting location
Installation position
Find a location that ensures easy accessibility for the later calibration. The fastening must be secure and must ensure low vibration for the
instrument. Avoid direct sunlight! Permissible ambient temperature at the installation location: -10 to +55 °C
with max. 95 % rel. humidity, no condensation. The instrument can be mounted in any position.

4.2 Dimensions

Close mounting
Minimum spacing of panel cutouts Horizontal Vertical Without setup connector: 30 mm 11 mm With setup connector (see arrow): 65 mm 11 mm
13

5 Installation

5.1 Installation instructions

The electrical connection must only be performed by qualified personnel!
The choice of cable, the installation and the electrical connection must
conform to the requirements of VDE 0100 “Regulations on the Installation of Power Circuits with Nominal Voltages below 1000 V” and the relevant local regulations.
At maximum load, the cable must be heat resistant up to at least 80 °C.The device is intended to be installed in electrical cabinets. It shall be
operated by mains protected with a branch circuitry overcurrent protection device not more than 20 Amps. For servicing/repairing a Disconnecting Device shall be provided to disconnect all conductors.
The load circuits must be fused for the maximum load currents in each case
to prevent the relay contacts from becoming welded in the event of a short circuit.
Electromagnetic compatibility meets the requirements of EN 61326.Lay the input, output, and supply lines so they are physically separated
from each other and are not parallel.
Use twisted and shielded probe cables. If possible, do not lay these cables
close to components or cables through which current is flowing. Ground the shielding at one end.
The probe cables must have an uninterrupted run (do not route them via
terminal blocks or similar arrangements).
No other consumers can be connected to the power terminals of the
instrument.
The instrument is not suitable for installation in areas with an explosion
hazard.
Apart from faulty installation, incorrect settings on the instrument may also
affect the proper functioning of the subsequent process or lead to damage. You should therefore always provide safety equipment that is independent of the instrument and it should only be possible for qualified personnel to make settings.
Mounting information for conductor cross-sections and ferrules
Ferrule Conductor cross-section Minimum length of ferrule or
Mi ni mu m Ma xi mu m
Without ferrule 0.34 mm Without collar 0.25 mm With collar up to 1.5 mm
2
0.25 mm
Twin, with collar 0.25 mm
2
2
2
2
2.5 mm
2.5 mm
1.5 mm
1.5 mm
2
2
2
2
stripping
10 mm (stripping) 10 mm 10 mm 12 mm
14

5.2 Electrical isolation

3700 V AC
Setup interface
Power supply
Power supply for a 2-wire transmitter
3700 V AC
Solid state relay Triac
Binary inputs
30 V AC 50 V DC
Continuous outputs
30 V AC 50 V DC
Input 2 (option)
3700 V AC
Relay outputs
Secondary input (Pt100/Pt1000)
Main input cond. conductivity
30 V AC 50 V DC
Extra-low voltage
30 V AC 50 V DC
Interface RS422/485 PROFIBUS-DP Datalogger
Power supply for ISFET sensor
PhotoMOS relay
®
30 V AC 50 V DC
5 Installation
15
5 Installation
1
2
3
4
1
2
3
4
5
6
7
8
6
7
8
9
10
9
10
11
12
4
L1(L+)
5
6
8
9
11
12
13
15
16
17
N(L-)
11
12
13
14
15
(a)
(b)(c)
(1)
(3)
(2)
J
J
E
S
+
-

5.3 Connection

5.3.1 Terminal assignment

(1) Row 1 (a) Option 1 (b) Option 2 (c) Option 3 (2) Row 2 Main input board
(conductivity/resistance/temperature/standard signal)
(3) Row 3 PSU board
(voltage supply/2× relays)

5.3.2 Optional board (row 1, slot a, b or c)

Function Symbol Terminal
Analog input Temperature sensor
in a two-wire circuit
Pt100 or Pt1000
Temperature sensor in a three-wire circuit
Pt100 or Pt1000
Resistance transmitter 2
for slot (a)
2 4
2 3 4
3 4
Te rm in al
for slot (b)
6 8
6 7 8
6 7 8
Ter mi na l
for slot (c)
10 12
10 11 12
10 11 12
Electrical current 3
16
4
7 8
11 12
5 Installation
+
-
+
-
+
-
RxD-
RxD+
TxD-
TxD+
R
xD/TxD-
RxD/TxD+
RxD/TxD-P(B)
VP(+5V)
DGND
RxD/TxD-N(A)
RxD/TxD-
RxD/TxD+
S
P
O
S
P
S
Function Symbol Terminal
for slot (a)
Voltage
0(2) to 10 V
Voltage
0to1V
1 2
2 3
Te rm in al
for slot (b)
5 6
6 7
Ter mi na l
for slot (c)
9
10 10
11
Continuous output Current or voltage 2
3
6 7
10 11
Modbus interface
RS422 9
10 11 12
RS485 11
12
PROFIBUS-DP interface
9 10 11 12
Datalogger interface
RS485 10
11
Relay (1× changeover)
K3 1
K4 5 2 3
6 7
K5 9
Relay (2× NO, common pin)
K3 1
K5 9
2
K6 3
K8 11
Triac (1 A)
PhotoMOS
®
relay (0.2 A)
K3 2
3
K3 1
2
K4 6
K4 5
K5 10
7
K5 9
6
10 11
10
11
10
K6 3
K7 7 4
8
K8 11
12
17
5 Installation
+
-
+
-
J
J
E
S
A

5.3.3 Main board (row 2)

Function Symbol Terminal Standard signal input for electrical current
0(4) to 20 mA
Standard signal input for voltage
0(2) to 10 V or 10 to 0(2) V
Temperature sensor in a two-wire circuit
Pt100 or Pt1000
Temperature sensor in a three-wire circuit
Pt100 or Pt1000
Resistance transmitter 4
Conductivity sensor
Conductivity sensor (2-electrode system) Terminals 6+7 and 8+9 can be bridged on the instrument;
2-wire cable routing up to the head of the conductivity sensor. For concentric cells, terminal 6 must be connected with the outer
electrode. Conductivity sensor (2-electrode system)
Wiring for highest accuracy; 4-wire cable routing to the head of the conductivity sensor.
For concentric cells, terminal 6 must be connected with the outer electrode.
Conductivity sensor (4-electrode system) 6 - Outer electrode 1 7 - Inner electrode 1 8 - Inner electrode 2 9 - Outer electrode 2
Shield connection
Conductivity sensor 10 GND
3 4
1 4
2 3 4
2 3 4
3 2
6 7 8 9
6 7 8 9
6 7 8 9
Binary inputs
Binary input 1 12+
14
Binary input 2 13+
14
18
5 Installation
S
P
O
S
P
O

5.3.4 PSU board (row 3)

Function Symbol Terminal Voltage supply for JUMO dTRANS 02
Voltage supply: AC 110 to 240 V
Voltage supply: AC/DC 20 to 30 V
n.c. 4
Voltage supply for external 2-wire transmitter
DC 24 V (+20/-15 %) 8 L+
Relay 1
Switching output K1 (floating)
1 L1 (L+)
2 N (L-)
5 6
9 L-
11 12 13
Relay 2
Switching output K2 (floating)
15 16 17
19

6 Operation

(8)
(5)
(2)
(4)
(3)
(1)
(6)
(7)
PGM
EXIT
Operation via the instrument keypad is described below. Instrument operation via the optional set-up program, see chapter 12 "Setup
program", page 80.

6.1 Controls

(1) Measurement unit
(2) Temperature
(3) Operating mode
(4) Measured value
(5) key Increase numerical value/Forward selection
(6) key Decrease numerical value/Forward selection
(7) key Change level/Forward selection/Confirm selection
(8) key Cancel entry/Exit level
20

6.2 Display

(1) (2) (3) (5)(4) (6)
(8)
(9)
(7)
EXIT

6.2.1 Measuring mode (normal display)

Example
6 Operation
(1) Binary output (relay) K1 is active (2) Binary output (relay) K2 is active (3) Binary input is active (4) Keypad is locked (5) Instrument status
ALARM (flashing): Broken sensor or overrange, etc. AL R1: Controller monitoring alarm from controller channel 1 AL R2: Controller monitoring alarm from controller channel 2 CALIB: Calibration mode active CALIB (flashing): Calibration timer elapsed
(6) Output mode
MAN.: Manual mode and/or simulation mode active HOLD: Hold mode active
(7) Top display
Measured value and unit of the variable set by parameter "Top display"
(8) Bottom display
Measured value and unit of the variable set by parameter "Bottom display"
(9) Operating mode
MEASURING: Standard measuring mode is active
To return to Measuring mode (MEASURING): Press the key or wait for a "timeout".
21
6 Operation

6.3 Principle of operation

6.3.1 Operation in levels

Measurement mode
Normal display 25 Min/max values of the main input 27 Min/max values of the optional inputs 28 Output display 28 Current values of the main input 28 Current values of the optional inputs 29 Current values of the math channels 29 States of the binary inputs and outputs 29 Manual mode overview 30 Hardware information 30 Instrument information 31 User data 81 Calibration (depending on the basic setting) 47, 53 Manual mode/simulation 36 Hold mode 38
Main menu
User level 31
Conductivity input 105 Temperature input 106 Optional inputs 106
Analog input 1, 2, 3
Binary inputs 108
Binary input 1, 2
Controllers 108
Controller 1
Controller 2
Controller special functions 110
Limit value control 110
Limit value 1, 2, 3
Binary outputs 108
Binary output 1, 2, 3, ... 8
Analog outputs 112
Analog output 1, 2, 3 Interface 113 Wash timer 113 Datalogger 113
See page
Parameter set 1, 2 Configuration
Parameter set 1, 2 Configuration
22
6 Operation
Display 114
Administrator level (password) 32
Parameter level 32
Parameters as above for "User level" Release level 32
Parameters as above for "User level" Basic setting 32 Calibration level 35
Main input (depending on the basic setting)
Temperature coefficient, linear Relative cell constant
Optional input 1, 2, 3
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point
Calibration release 35
Main input (depending on the basic setting)
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point 3-point K factor
Optional input 1, 2, 3
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point 3-point
Delete min/max values 35
Main input
Optional input 1, 2, 3 Delete logbook 35
Main input
Optional input 1, 2, 3 Delete daily batch 35 Delete total batch 35
23
6 Operation
Calibration level 47
Calibration logbook 73
Instrument information 31
Main input
Temperature coefficient, linear Temperature coefficient, curve
Optional input 1, 2, 3 106
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point
Main input Optional input 1, 2, 3
24

6.4 Measuring mode

EXIT
(3)
(1)
(2)
Different display types can be configured, see "Display of measured values STANDARD", page 93.
To return to Measuring mode: press the key or wait for a "timeout".
Measurements with "out of range" are ignored. The min./max. value memory can be reset:
Administrator level/Delete min/max. When the basic setting is changed, the min and max values are deleted.

6.4.1 Normal display

Visualization
The following are displayed in Measuring mode:
- Analog input signal
- Unit (for example pH)
- Temperature of the sample medium
6 Operation
(1) MEASURING -> Measuring mode (2) 21.6 °C -> Temperature of the sample medium (3) 2032 µS/cm -> the measured value calculated from the standard signal
at the input
25
6 Operation
Administrator
level
User level
Measuring mode (normal display)
Calibration
level
Calibration
logbook
Instrument information
PGM
Min/Max Values Main input
> 3 s
+
PGM
< 2 s
EXIT
User data
or timeout
(adjustable)
EXIT
or timeout
(adjustable)
EXIT
or timeout
(adjustable)
EXIT
or timeout
(adjustable)
EXIT
or timeout
(adjustable)
EXIT
or timeout (adjustable)
EXIT
or timeout (adjustable)
EXIT
or timeout (adjustable)
EXIT
or timeout (adjustable)
Min/Max Values Option input
EXIT
or timeout
(adjustable)
Main variable
Temperature input
EXIT
or timeout
(adjustable)
Optional input 1 Optional input 2 Optional input 3
EXIT
or timeout
(adjustable)
Math 1
2Math
EXIT
or timeout
(adjustable)
output Controller 1 Controller 2
EXIT
or timeout
(adjustable)
Manual overview Binary outputs
Manual overview
Analog outputs
EXIT
or timeout
(adjustable)
Binary signal
overview
2
< 2 s
Can only be activated with setup program
2

6.5 Input/output information

26

6.5.1 User data

Measuring mode (normal display)
< 2 s
EXIT
Manual mode
(controller)
or timeout
(adjustable)
> 3 s
PGM
EXIT
+
EXIT
Hold mode
or timeout
(adjustable)
EXIT
+
Hardware information
+
> 3 s
EXIT
Calibration
or timeout
(adjustable)
EXIT
+
Only if released
Keys
release
1
1 1
PGM
PGM
6 Operation
Up to 8 parameters that are frequently changed by the user can be combined in the user level under "User data" (via setup program only).
Activating the display
The instrument is in Measuring mode (normal display)
Briefly press the key.Select the required "quick setting" with the and keys.
Editing
Briefly press the key.Edit the setting with the and keys.

6.5.2 Min/max values of the main input

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
27
6 Operation
Minimum and maximum values of the main value "1:" (mS/cm, µS/cm, MOhm x cm, mV, %, ppm) and the temperature "T:" are displayed.
The extreme values of the main measurement variable and the temperature are
not mutually assigned (e.g. not 813 µS/cm at 24.3 °C).

6.5.3 Min/max values of the optional inputs

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
Minimum and maximum values of the optional inputs (1, 2 and 3) are displayed

6.5.4 Output level

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
The current output levels of the controller outputs.

6.5.5 Current values of the main entries

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
28
The current values of the main output are displayed.

6.5.6 Curgent values of the optional entries

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
The current values of the optional inputs (1, 2 and 3) are displayed

6.5.7 Current values of the math channels

6 Operation
Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
The current values of the main output are displayed.

6.5.8 States of the binary inputs and outputs

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary.
The states of binary inputs E1 and E2 and of relays K1 through K8 are displayed. In the example shown here, relay K1 is active.
29
6 Operation
EXIT
PGM

6.5.9 Manual mode overview

Analog outputs (optional boards)
In this example, analog outputs 2 and 3 are working normally.
Switching outputs (PSU board and optional boards)
In this example relay output 2 is in Manual mode.
The instrument is in "normal display" mode Briefly press the or key (several times if necessary).
Manual mode can only be displayed if at least one output is in Manual mode. For example Administrator level/Parameter level/Binary outputs/Binary output 1/Manual mode "Active" or "Simulation".
To return to Measuring mode: press the key or wait for a "timeout".

6.5.10 Hardware info

These displays are required for phone support.
The instrument is in Measuring mode (normal display) Press and hold the and keys.
Alternating display
30

6.5.11 Device info

PGM
PGM
PGM
These displays provide an overview of fitted hardware options and the settings of inputs (helpful for troubleshooting, etc.).
Press the key for longer than 3 seconds. Briefly press the or key (several times if necessary). Select Device info
6 Operation
Press the keys.
Briefly press the or key (several times if necessary).

6.6 User level

All the parameters that the Administrator (see chapter 6.7 "Administrator level", page 32) has released can be edited at this level. All the other parameters (marked by a key ) are read only.
Press the key for longer than 2 seconds.
For further information about the inputs, press the or keys.
31
6 Operation
PGM
PGM
Select "USER LEVEL".
All possible parameters are accessed below. Depending on the configuration of a specific instrument, some of these parameters may not appear.

6.6.1 Parameters of the User level

See chapter 16.2 "Parameters of the User level", page 105.

6.7 Administrator level

- All the parameters can be edited at this level.
- At this level, it is also possible to define which parameters can be edited by a "normal" user (operator) and which calibrations can be performed.
To get to the Administrator level, proceed as follows:
Press the key for longer than 2 seconds.Use the or keys to select "ADMINISTR.-LEVEL".Use the and keys to enter the password 300 (factory setting).Confirm the key.

6.7.1 Parameter level

The settings that can be made here are the same as those at the User level, see "User level", page 31. As the operator (user) has administrator rights here, the parameters that are locked in the User level can now also be modified.

6.7.2 Release level

All parameters can be released (modification possible) or locked (no modification possible) for editing at the User level.

6.7.3 Basic setting

The JUMO dTRANS 02 CR has a basic setting wizard, to make it easier for the user to configure the extensive setting options of the instrument and to avoid configuration conflicts. The basic settings are reached via ADMINISTR.-LEVEL/PASSWORD/BASIC SETTING.
All the important settings are systematically polled here. At the end, once a request for conformation has been acknowledged, the instrument is initialized with the new settings. Dependent parameters are checked and adjusted.
32
Basic setting wizard
Cell type
2-wire 4-wire
Cell constant
Cell constant
0.01 / 0.1 / 0.5 /
1.0 / 3.0 / 10.0
0.01 / 0.1 / 0.5 /
/ 3.0 / 10.01.0
Operating mode
Pollution recogn.
Off / On
Sens. break recog.
Off / On
TDS
Temp. coefficient
0.0 - - 8.0%/K2.20
0.5 /1.0
Temperature compensation source
Temperature input,(manual temperature,
optional input 1, 2, 3)
Table
Input variable
µS/cm / mS/cm /
kOhm*cm / MOhm*cm
Temp. comp.
None Nat. waters ASTM D1125 neutral ASTM D1125 acidic ASTM D1125 alkal.
Linear
customer-specific factor
Input variable
µS/cm / mS/cm /
kOhm*cm / MOhm*cm
Linear
Temp. comp.
None Nat. waters ASTM D1125 neutral ASTM D1125 acidic ASTM D1125 alkal.
Conductivity
Temp. comp.
None Nat. waters ASTM D1125 neutral ASTM D1125 acidic ASTM D1125 alkal.
Linear
6 Operation
33
6 Operation
Unit MB 2
µS/cm / mS/cm
Supply frequency
- 50 Hz
- 60 Hz
Initialize all
dependent parameters
No change
to parameters
Temperature compensation source
Temperature input,(manual temperature, optional input 1, 2, 3)
Reinitialize device
Ye s
No
Unit MB1
µS/cm / mS/cm /
kOhm*cm / MOhm*cm
Display format MB1
XXXX / XXX.x /
XX.xx / X.xxx
2nd measuring range
Off
Auto.
Temp. coefficient
0.0 - - 8.0%/K2.20
Manual
Unit MB 2
µS/cm / mS/cm /
k cm / M cmWW
Display format
MB 2
XXXX / XXX.x /
XX.xx / X.xxx
TDS factor
0.01 - - 2.000.67
Unit
Cust. specs. / ppm
Display format
XXXX / XXX.x /
XX.xx / X.xxx
Temp. coefficient
0.0 - - 8.0%/K2.20
TDS factor
0.01 - - 99.990.67
Unit
Cust. specs. / ppm /
1
µS/cm / mS/cm /
kOhm*cm / MOhm*cm
Display format
XXXX / XXX.x /
XX.xx / X.xxx
Temp. coefficient
0.0 - - 8.0%/K2.20
Unit
Cust. specs. / ppm /
1
µS/cm / mS/cm /
kOhm*cm / MOhm*cm
Display format
XXXX / XXX.x /
XX.xx / X.xxx
Temp. coefficient
0.0 - - 8.0%/K2.20
Temp. coefficient
0.0 - - 8.0%/K2.20
Temp. comp.
None Nat. waters ASTM D1125 neutral ASTM D1125 acidic ASTM D1125 alkal.
Linear
34

6.7.4 Calibration level

Depending on which operating mode has been configured (in the Basic setting menu), one or more of the following calibration options will be available:
- Cell constant
- Temperature coefficient

6.7.5 Calibration release

Which calibration procedure may be performed directly and which may not can be configured here, see chapter 8.2.3 "Ways to start the calibration", page
48.

6.7.6 Delete min/max values

If required, the values can be deleted once a request for confirmation has been acknowledged,
see chapter 6.5.2 "Min/max values of the main input", page 27 or see chapter 6.5.3 "Min/max values of the optional inputs", page 28.
6 Operation

6.7.7 Delete logbook

The last five calibration processes for each input are archived in the calibration logbook. If a "Datalogger" optional board is fitted, the date and time are also archived. If necessary the logbook can be deleted after a confirmation prompt.

6.7.8 Delete daily batch

If required, the counter can be deleted once a request for confirmation has been acknowledged.

6.7.9 Delete total batch

If required, the counter can be deleted once a request for confirmation has been acknowledged.
35
6 Operation
Simulation modeMANUAL mode
"Higher order"
controller
Binary outputs
Analog outputs

6.8 MANUAL mode/Simulation mode

These functions can be used to set the switching outputs and analog outputs of the instrument manually to a defined state. This facilitates dry startup, troubleshooting and customer service, etc. .
Simulation mode accesses the analog outputs and binary outputs directly. When simulation mode has been selected, MANUAL mode is not possible!
In MANUAL mode the settings for "higher order controllers" are taken into consideration.

6.8.1 MANUAL mode only via "higher order" controller functions

Select Manual mode
In the factory setting of the instrument the MANUAL mode parameter is locked and can only be activated by the administrator! This parameter must first be released for other users, see "Release level", page 32.
Set ADMINISTR.-LEVEL/PARAMETER LEVEL/CONTROLLER/CTRL.SPEC.
FUNCT./MANUAL MODE "Locked, Coding or Switching.
Locked = No Manual mode, control is via device. Coding = The outputs are active as long as the or key is pressed. Switching = the outputs are active if the or key is pressed. If the
corresponding key is pressed again, the output becomes inactive again.
36
Activate Manual mode
EXIT
EXIT
EXIT
EXIT
EXIT
The instrument is in Display mode Press the and keys for less than 2 seconds.
If the keys (alone) are pressed for longer than 3 seconds, the instrument switches to language selection.
If the and keys are pressed for longer than 3 seconds, the instrument goes into HOLD mode.
Then the outputs of the instrument respond according to the default settings. To exit HOLD mode, press the and keys for longer than 3 seconds.
Control is not longer via the instrument. The output level of the controllers is 0%.
Controller 1 is activated by the key. In this case the output level of controller 1 is 100%.
Controller 2 is activated by the key. In this case the output level of controller 2 is 100%.
6 Operation
The word MANUAL appears in the status line of the display.
Deactivation
Press the key. Control is once again through the outputs of the instrument.
The word MANUAL appears in the status line of the display.

6.8.2 Simulation of binary outputs

Activate simulation
In the factory setting of the instrument the MANUAL mode parameter is set to "No simulation" and can only be activated by the administrator! This parameter must first be released for other users, see "Release level", page 32.
If a higher order switching function has been assigned to an output, Simulation mode is not possible for that output.
Set ADMINISTR. LEVEL/PARAMETER LEVEL/BINARY OUTPUTS/BINARY
OUTPUT1 ( ... 8) "Manual mode no simulation, Inactive or Active".
No simulation = No Manual mode, control is via device. Inactive = Relay K1 or K2 is de-energized;the word MANUAL
appears in the status line of the display
Active = Relay K1 or K2 is energized;the word MANUAL appears
in the status line of the display
37
6 Operation
EXIT
EXIT
Deactivate Manual mode
No simulation = No Manual mode, control is via device. When the instrument is in display mode, the word MANUAL disappears from
the status line of the display.

6.8.3 Simulation of analog outputs via MANUAL mode

Release and activation
Select activation of simulation of the actual value output:
ADMINISTR.-LEVEL/PARAMETER LEVEL/ANALOG OUTPUTS/ANALOG
Deactivation
OUTPUT 1 (2, 3)
With "On" the output takes on the value of the "Simulation value" parameter. When the instrument is in display mode, the word MANUAL appears in the
status line of the display.
ADMINISTR.-LEVEL/PARAMETER LEVEL/ANALOG OUTPUTS/ANALOG OUTPUT 1 (2, 3)
/SIMULATION/ON.
/SIMULATION/OFF.
The corresponding output of the instrument works again. When the instrument is in display mode, the word MANUAL disappears from
the status line of the display.

6.9 HOLD mode

In HOLD status the outputs take on the states programmed in the relevant parameter (controller channel, switching output or analog output).
This function can be used to "freeze" switching outputs and the analog outputs of the instrument. This means the current status of the output will be retained even when the measured value changes. Control is not via the instrument.
If MANUAL mode is activated while HOLD mode is activated, MANUAL mode takes precedence and MANUAL then appears in the status line of the display! MANUAL mode can be terminated by pressing the key. If HOLD mode is still activated (by the binary input or by keyboard), the instrument then returns to HOLD mode!
HOLD mode can be activated by pressing the key or by the binary input.
Activation by pressing key
Press and hold the and keys longer than 3 seconds.
Then the outputs of the instrument respond according to the default settings. The word HOLD appears in the status line of the display.
38
If the and keys are pressed for less than 3 seconds, the instrument
EXIT
EXIT
EXIT
goes into Manual mode. Then the outputs of the instrument respond according to the default settings.
Pressing a key to deactivate HOLD mode
Press the and keys for longer than 3 seconds.
If the and keys are pressed for less than 3 seconds, the instrument goes into Manual mode.
Then the outputs of the instrument respond according to the default settings.
Control is through the outputs of the instrument again. The word MANUAL disappears from the status line of the display.
6 Operation
39

7 Commissioning

7.1 Getting started

Some suggestions follow for configuring the instrument reliably in little time.
Mount the instrument, see chapter 4 "Mounting", page 13.Install the instrument, see chapter 5 "Installation", page 14 ff.Call up Administrator level (ADMINISTR. LEVEL).Enter password 0300 (factory setting).Call up PARAMETER LEVEL/DISPLAY/OPERAT. TIMEOUT.Set OPERAT. TIMEOUT to 0 minutes (no timeout).Leave the Display level with "EXIT"Leave the Parameter level with "EXIT"Select BASIC SETTING and work through all the menu items, see chapter
6.7.3 "Basic setting", page 32.
Answer "YES" to the "Reinitialize device" query Configure the required additional parameters.Calibrate the instrument to the conductivity sensor and sample medium,
see chapter 8 "Calibrating a conductivity sensor", page 47 or see chapter 9 "Calibrating a sensor with a standard signal", page 53.
40
7 Commissioning
(1)
(2)
(3)

7.2 Setting examples

7.2.1 Conductivity measurement, temperature compensated

Measurement of drinking water.
Layout
Electrical connection
See chapter 5 "Installation", page 14.
Ta sk
Measurement range: 0 to 1.00 mS/cm Cell constant K: 1.0 1/cm Output signal: 4 to 20 mA Temperature measurement Pt100 Limit monitoring: Limit function Limit value 1: 0.80 mS/cm
Data sheet (1) Transmitter/controller type 202552 202552 (2) Conductivity sensor on the main board 202925 (3) Conductivity cable 202990
41
7 Commissioning
Basic setting
Start the basic settings, see chapter 6.7.3 "Basic setting", page 32 Diagrammatic overview, see "Basic setting wizard", page 33.
Cell type 2-wire Cell constant 1.0 Broken sensor detection Off Operating mode Conductivity Temperature compensation Linear Temperature compensation source Temperature input Temperature coefficient 2.20 (factory setting) Unit mS/cm Display format XX.xx 2nd measuring range Off Supply frequency 50 Hz
Temperature input
Analog output
Controller settings
Reinitialize device Yes
Administrator level/Password/Parameter level/Temperature input
Temperature sensor Pt100
Administrator level/Password/Parameter level/Analog outputs/Analog output 1
Signal source Main variable Signal type 4 to 20 mA Start of scaling 0.00 mS/cm End of scaling 1.00 mS/cm
See chapter 11.6.3 "Controller with limit value function", page 79.
42
7 Commissioning
(1)
(2)
(3)

7.2.2 Measurement of ultra-pure water with 2-electrode measuring sensor

USP limit monitoring
Layout
Electrical connection
See chapter 5 "Installation", page 14.
Ta sk
Measurement range: 0 to 2.00 µS/cm Cell constant K: 0.01 1/cm Output signal: 4 to 20 mA Temperature measurement Pt100 Limit monitoring: Limit value function Limit value 1: USP
Data sheet (1) Transmitter/controller type 202552 202552 (2) Conductivity sensor on the main board 202924 (3) Conductivity cable 202990
43
7 Commissioning
Basic setting
Start the basic settings, see chapter 6.7.3 "Basic setting", page 32 Diagrammatic overview, see chapter "Basic setting wizard", page 33.
Cell type 2-wire Cell constant 0.01 Broken sensor detection Off Operating mode Conductivity Temperature compensation None Temperature compensation source Temperature input Unit µS/cm Display format X.xxx 2nd measuring range Off Supply frequency 50 Hz
Temperature input
Analog output
Controller settings
Reinitialize device Yes
Administrator level/Password/Parameter level/Temperature input
Temperature sensor Pt100
Administrator level/Password/Parameter level/Analog outputs/Analog output 1
Signal source Main variable Signal type 4 to 20 mA Start of scaling 0.00 µS/cm End of scaling 2.00 µS/cm
See chapter 11.6.2 "Limit monitoring to USP", page 78.
44
7 Commissioning
(1)
(2)
(3)

7.2.3 Measurement of ultra-pure water with 2-electrode measuring sensor

Display in MOhm × cm.
Layout
Electrical connection
See chapter 5 "Installation", page 14.
Ta sk
Measurement range: 0 to 20.00 MOhm × cm Cell constant K: 0.01 1/cm Output signal: 4 to 20 mA Temperature measurement Pt100 Limit monitoring: Limit value function Limit value 1: 10.00 MOhm × cm
Data sheet (1) Transmitter/controller type 202552 202552 (2) Conductivity sensor on the main board 202924 (3) Conductivity cable 202990
45
7 Commissioning
Basic setting
Start the basic settings, see chapter 6.7.3 "Basic setting", page 32 Diagrammatic overview, see chapter "Basic setting wizard", page 33.
Cell type 2-wire Cell constant 0.01 Broken sensor detection Off Operating mode Conductivity Temperature compensation None Temperature compensation source Temperature input Unit MOhm × cm Display format XX.xx 2nd measuring range Off Supply frequency 50 Hz
Temperature input
Analog output
Controller settings
Reinitialize device Yes
Administrator level/Password/Parameter level/Temperature input
Temperature sensor Pt100
Administrator level/Password/Parameter level/Analog outputs/Analog output 1
Signal source Main variable Signal type 4 to 20 mA Start of scaling 0.00 MOhm × cm End of scaling 20.00 MOhm × cm
See chapter 11.6.1 "Simple limit monitoring", page 78.
46

8 Calibrating a conductivity sensor

8.1 Notes

During calibration, relays and analog output signals adopt their configured states!
When is calibration required?
- The temperature coefficient of the sample medium must be determined once.
- The cell constant must be calibrated at regular intervals (depending on the sample medium and requirements).
Every successfully completed calibration is documented in the calibration logbook, see chapter 10 "Calibration logbook", page 73.

8.2 General information

The electrical properties of all sensors vary slightly from instance to instance and also change during operation (due to deposits or wear, etc.). This changes the output signal of the sensor.

8.2.1 Measurements in highly-purified water

Measurements in highly-purified water (measured values < approx. 10 µS/cm) make special demands on the metrology and the measurement environment.
The following points should therefore be considered and checked first before attempting a calibration:
- Basically sensors with ASTM certificate are recommended for measure­ments in highly-purified water. Their cell constants are measured by the manufacturer and can be found in the certificate.
- Ready-to-use calibration solutions in the range < 5 µS/cm are difficult or impossible to get. Effort and error rate are very high when handling these.
- Reliable comparative measurements are often problematic due to unknown or insufficient quality of the comparison device. In addition, the reference junction is often not close enough to the actual measuring point.
- If minor measurement errors exist despite of entering the exact cell con­stant, these can manually be adjusted in the range of several percent by changing the relative cell constant. Possible causes are installation conditi­ons and flow dependencies.
- Larger deviations (> approx. 10 %) mostly have other causes, such as contamination of the sensor by mishandling or EMC.
More information on highly-purified water measurement in form of a scientific paper can be found on the Internet at www.jumo.de.
For this purpose, enter the keyword "FAS 614" into the search box.
47
8 Calibrating a conductivity sensor
PGM
PGM
PGM

8.2.2 Requirements

- The instrument must be supplied with voltage, see chapter 5 "Installation", page 14 ff.
- A conductivity sensor must be connected to the transmitter.
For a configuration example see chapter 7.2.1 "Conductivity measurement, temperature compensated", page 41.
A conductivity sensor be
- connected directly to the main input or
- connected to the "Analog input (universal)" optional board via a transmitter.
- "Conductivity" must be configured as operating mode in the basic setting.
- The instrument is in Measuring mode.

8.2.3 Ways to start the calibration

Select the input to which the conductivity sensor is connected.
If Calibration level is not released
Press the key for longer than 3 seconds/ADMINISTR.-LEVEL/ PASSWORD/CALIBR.-LEVEL/MAIN INPUT or ANALOG INPUT.
If Calibration level is released
Press the and keys simultaneously/MAIN INPUT or ANALOG INPUT.
If Calibration level is released
Press the key for longer than 3 seconds/CALIBR.-LEVEL/MAIN INPUT or ANALOG INPUT.

8.2.4 Calibration options

The instrument provides two calibration options for adjusting the JUMO dTRANS 02 CR to the measuring point:
Calibration of the temperature coefficient
See chapter 8.4 "Calibrating the relative cell constant", page 51.
Calibration of the cell constant
See chapter 8.4 "Calibrating the relative cell constant", page 51.
48
8 Calibrating a conductivity sensor
(1)
PGM

8.3 Calibration of the temperature coefficient of the sample medium

Make preparations, see chapter 8.2 "General information", page 47.Start calibration, see chapter 8.2.3 "Ways to start the calibration", page 48.Select "TEMP.COEFF. LIN.".
Now the source of temperature acquisition can be selected (manually, or using the temperature input of the PSU board, or the temperature input via the optional board). This source will be active for the duration of the calibration.
An example follows: automatic temperature acquisition using the temperature sensor integrated into the conductivity sensor.
The current sensor temperature appears in the display (+ flashing) (1).
Enter the required working temperature and confirm your entry with the
key.
The working temperature must be at least 5 °C above or below the reference temperature (25.0 °C).
49
8 Calibrating a conductivity sensor
PGM
EXIT
The conductivity (399 µS/cm) at the current temperature (24.3 °C) now appears on the right of the LC display. The temperatures T1 (25 °C) and T2 (70.0 °C) that have yet to be triggered are shown on the left.
Heat the sample medium until the working temperature is reached.
During calibration, the rate of temperature change in the measurement solution must not exceed 10 °C/min.
Calibration is also possible in the cooling process (with a falling temperature). It starts above the working temperature and ends below the working temperature.
As soon as the temperature of the sample medium exceeds T1 (25 °C), this is hidden on the display. The uncompensated conductivity at the current temperature is displayed on the right.
If the temperature of the medium exceeded T2 (73.0 °C), the instrument determines the temperature coefficient.
The LC display now shows the determined temperature coefficient as %/K.
Use the key to accept the temperature coefficient or
the key to reject it.
The transmitter is in "measuring mode" and displays the compensated conductivity of the solution.
50
8 Calibrating a conductivity sensor
PGM
PGM
The currently measured conductivity can be coerced manually by pressing the
key. This may be useful if the reference or working temperature cannot be reached precisely. However, the calibration result incorporates a certain amount of inaccuracy!

8.4 Calibrating the relative cell constant

Make preparations, see chapter 8.2 "General information", page 47.Start calibration, see chapter 8.2.3 "Ways to start the calibration", page 48.Select the relative cell constant.
Immerse the conductivity sensor in a reference solution with a known
conductivity.
The measurement solution must maintain a constant temperature during calibration! The conductivity sensor must be kept at a distance of at least 20 mm from the container wall during the calibration and must not be moved!
The current measurement value and the temperature are displayed.
When the measurement value is steady, press the key;
the conductivity measurement flashes in the display.
Set the value to the actual conductivity.Press the PGM key. The relative cell constant determined by the instrument
is displayed (as a %).
51
8 Calibrating a conductivity sensor
PGM
EXIT
Use the key to accept the value or the key to reject it. The current measurement value and the temperature are displayed.

8.4.1 Entering the cell constant manually

If the exact cell constant is known (for example a conductivity sensor with the ASTM test report), the value can be entered directly.
ADMINISTR.-LEVEL/PARAMETER LEVEL/INPUT CONDUCT./ REL. CELL CONST.

8.4.2 Cell constants

Two-electrode systems
Cell constant
Four-electrode systems
Cell constant
[1/cm]
0.01
Setting range of the
relative cell constant
Resulting
usable range [1/cm]
0.002 to 0.05
0.1 0.02 to 0.5
1.0 0.2 to 5
20 - 500 %
3.0 0.6 to 15
10.0 2.0 to 50
Setting range of the
[1/cm]
0.5
1.0 0.2 to 1.5
relative cell constant
20 - 150 %
Resulting
usable range [1/cm]
0.1 to 0.75
52

9 Calibrating a sensor with a standard signal

9.1 General information

During calibration, relays and analog output signals adopt their configured states!
Sensors with a standard signal output can only be connected to an "Analog input (universal)" optional board!
The sensors connected to the instrument should be cleaned and the instru­ment itself calibrated, at regular intervals (subject to the sample medium).
Every successfully completed calibration is documented in the calibration logbook, see chapter 10 "Calibration logbook", page 73.

9.1.1 Operating modes

The operating mode selection depends on which sensor (transmitter) is connected.
Linear operating mode
For example sensor for free chlorine, redox, pressure, liquid level or humidity
pH operating mode
For example pH sensor
Conductivity operating mode
For example sensor for conductivity, concentration
Customer specs.
For sensors with non-linear characteristics. Up to xx interpolation points can be defined in an instrument table. This allows for an excellent approximation of a non-linear characteristic.
Chlorine, pH and temperature-compensated
Combination of chlorine sensor and pH sensor and temperature sensor. The measured value for chlorine often depends to a great extent on the pH value of the solution. The chlorine measurement is compensated depending on the pH value in this operating mode. The pH measurement is temperature-compensated
53
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM

9.1.2 Calibration options

Different calibration options are available depending on the operating mode.
Operating mode Calibration options Page
1-point 2-point Limit point Rel.
cell const.
Linear X X X - - 55
a
pH
XX---59 Conductivity - - - X X 63 Concentration - - - X 69 Customer specs. Due to the table with interpolation points, no calibration is required Chlorine,
--X--71
pH-compensated
a
When configuring the device: the parameter "zero point" for the operating mode "pH" of the re­spective optional board has to be set – one time – to value "7".
-With one-point (offset) calibration, the zero point of the sensor is
calibrated.
-With two-point calibration, the zero point and slope of the sensor are cali-
brated. This is the recommended calibration for most sensors.
-With one-point final value calibration, the slope of the sensor
is calibrated. This is the recommended calibration for chlorine sensors, for example.
Te mp .
coeffic.
- Calibration of relative cell constant
With conductivity sensors only.
- Calibration of the temperature coefficient
With conductivity sensors only.

9.1.3 Ways to start the calibration

Select the input to which the sensor is connected.
If Calibration level is not released
Press the key for longer than 3 seconds/ADMINISTR.-LEVEL/ PASSWORD/CALIBR.-LEVEL/OPTION INPUT.
If Calibration level is released
Press the and keys simultaneously/OPTION INPUT.
If Calibration level is released
Press the key for longer than 3 seconds/CALIBR.-LEVEL/ OPTION INPUT.
54
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM
PGM
EXIT

9.2 Linear operating mode

9.2.1 1-point calibration

This example is based on a liquid level measurement (as a %). The input signal is provided by a pressure transmitter.
The transmitter is in "Measuring mode".
Now bring the system to a defined state (e.g. when measuring liquid level,
empty the container).
Start the calibration, see "Ways to start the calibration", page 54.Select the zero point calibration with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the required value (usually 0%) with the and keys; then press to continue.
The zero point determined by the instrument is displayed. Use the key to accept the value or
the key to reject it.
55
9 Calibrating a sensor with a standard signal
Display =
Input value
Slope
+ Zero point
PGM
PGM
The instrument returns to Measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

9.2.2 2-point calibration

The values determined during calibration (zero point and slope) work out as follows:
This example is based on a liquid level measurement. The input signal is pro­vided by a pressure transmitter.
The transmitter is in "Measuring mode".
Now bring the system to a defined state (e.g. when measuring liquid level,
empty the container).
Start the calibration, see "Ways to start the calibration", page 54.Select the 2-point calibration with the key.
Wait until the display value has stabilized; then press
to continue.
56
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM
PGM
EXIT
Set the displayed value to the required value (usually 0) with the and
keys; then press to continue.
Now bring the system to a second defined state (e.g. when measuring
liquid level, container full). Wait until the display value has stabilized; then press to continue
Set the displayed value to "Maximum" (usually 100%) with the and
keys; then press to continue.
The zero point and slope determined by the instrument are displayed. Use the key to accept the calibrated values or
reject them with the key.
The instrument returns to Measuring mode.
57
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

9.2.3 Calibration limit point

This example is based on a measurement of free chlorine. The input signal is provided by a corresponding transmitter.
The transmitter is in "Measuring mode".
-
The process must now be brought to the state that is as relevant as pos-
sible to the final value (e.g. when measuring chlorine, the required concen­tration).
Start the calibration, see "Ways to start the calibration", page 54.Select the limit point calibration with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the measured reference value with the or keys; then press to continue.
The slope determined by the instrument is displayed.
58
9 Calibrating a sensor with a standard signal
PGM
EXIT
Use the key to accept the value or the key to reject it.
The instrument returns to Measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

9.3 pH operating mode

9.3.1 Zero-point (1-point) calibration

This example is based on a glass combination electrode with a connected two-wire transmitter.
The transmitter is in "Measuring mode".
Perform calibration as follows.
Zero point (1-point) calibration
Make preparations, see chapter 8.2 "General information", page 47 .Start calibration, see chapter 8.2.3 "Ways to start the calibration", page 48.
59
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM
PGM
Select zero point calibration.
Immerse the combination electrode in a buffer solution with a known pH
value.
Start the zero point calibration with the key.
Now the source of temperature acquisition can be selected (manually, or using the temperature input of the PSU board, or the temperature input via the optio­nal board). This source will be active for the duration of the calibration.
An example follows: manual temperature entry.
To enter the temperature manually, use the and keys to set the cali-
bration solution temperature and confirm your entry with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the buffer solution value with the or keys;
then press to continue.
60
9 Calibrating a sensor with a standard signal
PGM
EXIT
Use the key to accept the zero point or the key to reject it.
The instrument returns to Measuring mode.

9.3.2 2-point calibration

This example is based on a glass combination electrode with a connected two-wire transmitter.
2-point calibration
The transmitter is in "Measuring mode".
Perform calibration as follows:
The buffer solutions (reference solutions) used for calibration must differ by at least 2 pH! During the calibration, the temperature of the two buffer solutions must be identical and remain constant!
Make preparations, see chapter 8.2 "General information", page 47 .Start calibration, see chapter 8.2.3 "Ways to start the calibration", page 48.Select 2-point calibration.
61
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM
PGM
PGM
Immerse the combination electrode in the first buffer solution with the
known pH value.
Start the two-point calibration with the key.
Now the source of temperature acquisition can be selected (manually, or using the temperature input of the PSU board, or the temperature input via the optio­nal board). This source will be active for the duration of the calibration.
An example follows: manual temperature entry.
To enter the temperature manually, use the and keys to set the cali-
bration solution temperature and confirm your entry with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the value of the first buffer solution with the
and keys; then press to continue.
Rinse and dry the pH combination electrode.Immerse the pH combination electrode in the second buffer solution.Wait until the display value has stabilized; then press to continue.
62
9 Calibrating a sensor with a standard signal
PGM
PGM
EXIT
Set the displayed value to the second buffer solution value with the or
keys; then press to continue.
The zero point and slope determined by the instrument are displayed. Use the key to accept the calibrated values or
reject them with the key.
The instrument returns to Measuring mode.

9.4 Conductivity operating mode

9.4.1 Calibration of the relative cell constant

This example is based on a conductivity sensor with a connected two-wire transmitter.
The transmitter is in "Measuring mode".
Immerse the conductivity sensor in a reference solution with a known con-
63
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM
PGM
EXIT
ductivity.
Start the calibration, see "Ways to start the calibration", page 54.Select REL. CELL CONST.Press the key.
When the measured value is stable, press the key
The measured conductivity value flashes on the display.
Use the or keys to set the value to the actual conductivity.Press the key;
the relative cell constant determined by the instrument is displayed (as a %).
Use the key to accept the temperature coefficient or
the key to reject it.
The current measurement value and the temperature are displayed.
64
9 Calibrating a sensor with a standard signal
(1)
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

9.4.2 Calibration of the temperature coefficient

Linear temperature coefficient
This example is based on a conductivity sensor with a connected two-wire transmitter.
The transmitter is in "Measuring mode".
Immerse the conductivity sensor in the sample medium. Start the calibration, see "Ways to start the calibration", page 54.
Select "LINEAR TEMP. COEF.".
The current sensor temperature flashes in the display (1).
The working temperature must be at least 5 °C above or below the reference temperature (25.0 °C).
Enter the required working temperature and confirm your entry.
65
9 Calibrating a sensor with a standard signal
(2)
PGM
PGM
PGM
EXIT
The LC display now shows the selected working temperature (flashing) (2).
Press the key.
The conductivity (399 µS/cm) at the current temperature (24.3 °C) now appears on the right of the LC display. The temperatures T1 (25 °C) and T2 (70.0 °C) that have yet to be triggered are shown on the left.
Press the key.Heat the sample medium until the working temperature is reached.
During calibration, the rate of temperature change in the measurement solu­tion must not exceed 10 °C/min.
Calibration is also possible in the cooling process (with a falling temperature). It starts above the working temperature and ends below the working tempe­rature.
As soon as the temperature of the sample medium exceeds T1 (25 °C), this is hidden on the display. The uncompensated conductivity at the current tempe­rature is displayed on the right.
If the temperature of the medium exceeded T2 (73.0 °C), the instrument deter­mines the temperature coefficient.
The LC display now shows the determined temperature coefficient as %/K.
Use the key to accept the temperature coefficient or
the key to reject it.
66
9 Calibrating a sensor with a standard signal
PGM
(1)
The transmitter is in "Measuring mode" and displays the compensated con­ductivity of the solution.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
With non-linear temperature coefficient (TEMP. COEFF. CURVE)
This example is based on a conductivity sensor with a connected two-wire transmitter.
The non-linear temperature coefficient can only be calibrated with a rising temperature! The start temperature must be below the configured reference temperature (usually 25 °C)!
The "TEMP.COEFF. CURVE" menu item is only displayed if a temperature sensor is connected and "TEMP.COEFF. CURVE" is configured as the type of temperature compensation.
The transmitter is in "Measuring mode".
Immerse the conductivity sensor in the sample medium. Start the calibration, see "Ways to start the calibration", page 54.
Select "TEMP. COEFF. CURVE " and press the key.
Enter the required start temperature (1) for the temp. coef. curve.
67
9 Calibrating a sensor with a standard signal
(2)
(3)
(4)
(5)
PGM
EXIT
PGM
Enter the required end temperature (2) for the temp. coef. curve.
Heat the sample medium continuously
(3) the current uncompensated conductivity (4) the current temperature of the sample medium (5) the first target temperature
During calibration, the rate of temperature change in the measurement solu­tion must not exceed 10 °C/min.
During the calibration process, the instrument displays values for the following five temperature interpolation points.
The end temperature has been reached
Use the key to accept the temperature coefficients or the key to reject the calibration result.
The LC display now shows the determined temperature coefficients as %/K. Use the key to accept the temperature coefficients or
68
9 Calibrating a sensor with a standard signal
EXIT
PGM
the key to reject the values.
The transmitter is in "Measuring mode" and displays the compensated con­ductivity of the solution.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

9.5 Concentration operating mode

9.5.1 Calibration of the relative cell constant

This example is based on a conductivity sensor with a connected two-wire transmitter.
The conductivity of a caustic solution is converted into a concentration value [%] by the instrument.
The transmitter is in "Measuring mode".
Immerse the conductivity sensor in a sample medium with a known con-
ductivity.
Start the calibration, see "Ways to start the calibration", page 54.Press the key.
The measured conductivity value is displayed. Wait until the measurement value has stabilized.
69
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM
EXIT
Press the key.
Use the and keys to set the value to the actual conductivity.
Press the key; the relative cell constant determined by the instrument is
displayed (as a %).
Use the key to accept the relative cell constant or
The transmitter is in "Measuring mode" and displays the compensated con­ductivity of the solution.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
the key to reject the values.
70
9 Calibrating a sensor with a standard signal
PGM
PGM
PGM

9.6 Chlorine measurement operating mode, pH-compensated

9.6.1 Final value calibration

The pH signal and temperature signal are supplied via the main input, the chlorine signal (standard signal) via the optional input.
The transmitter is in "Measuring mode".
-
Calibrate pH sensor
Perform calibration, see "pH operating mode", page 59.
Calibrate chlorine sensor
The process must now be brought to the state that is as relevant as pos-
sible to the final value (e.g. when measuring chlorine, the required concen­tration).
Start the calibration, see "Ways to start the calibration", page 54.Select the limit point calibration with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the measured reference value with the or keys; then press to continue.
71
9 Calibrating a sensor with a standard signal
PGM
EXIT
The slope determined by the instrument is displayed. Use the key to accept the value or the key to reject it.
The instrument returns to Measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
72

10.1 General information

PGM
PGM
The characteristic data for the last 5 successful calibration processed are documented in the calibration logbook.
Calling up
The instrument is in Measuring mode. Press the key for longer than 3 seconds.
Select input
Briefly press the key.

10 Calibration logbook

Most recent successful calibration
The "time stamp" in the following screen printouts (top left, for example 11­06-06 12:02) only appears if optional slot 3 is fitted with the "Datalogger with interface RS485"!
Briefly press the key.
Next most recent successful calibration
Briefly press the key.
73

11 Controller

L
H
d
a
(1)
(2)
(3)
(4)

11.1 General information

Apart from faulty installation, incorrect settings on the instrument may also affect the proper functioning of the subsequent process or lead to damage. You should therefore always provide safety equipment that is independent of the instrument and it should only be possible for qualified personnel to make settings.

11.2 Controller functions

"Software" control functions are assigned to "Hardware" outputs for this instrument.
1 Software controller for "simple" switching functions (e.g. alarm control) 2 Software controller for "higher order" switching functions (e.g. PID
controller) 3 "Switching" hardware output (e.g. relay) 3 "Continuous" hardware output (analog output)

11.2.1 Simple switching functions

Up to four switching functions can be set (limit value 1, 2, 3, 4) ADMINISTR.-LEVEL/PARAMETER LEVEL/LIMIT VALUE CONTR./ LIMIT VALUE x.

11.2.2 Higher order switching functions (PID)

Higher order switching functions are configured at the parameter level via the parameters of "Controller 1 or 2".
ADMINISTR.-LEVEL/PARAMETER LEVEL/CONTROLLER/CONTROLLER 1(2)/ CONFIGURATION/CONTROLLER TYPE/e.g. PULSE LENGTHS
74

11.2.3 Typical operator level parameters

L
H
d
a
d
a
d
a
d
a
(1)
(2)
11 Controller
Binary outputs
Explanation
Signal source No signal No switching function desired Limit control
1 to 4 Alarm function (AF1)
Alarm function (AF2)
Alarm function (AF7)
Alarm function (AF8)
"Simple" switching functions
Controller 1(2) "Higher order" switching functions Limit value
Pulse width Pulse frequency Steady Modulating

11.3 Software controllers and outputs

Simple controller functions
1 Main board 2 Optional board L Simple controller H Higher order controller d Digital output a Analog output
75
11 Controller
(1)
(2)
L
H
d
a
d
a
d
a
d
a
If "Simple controller functions" have been configured, only the digital outputs can be controlled!
The operator must configure which of the digital outputs will be controlled ­the main board or optional board 1, 2 or 3
Higher order controller functions
1Main board 2 Optional board L Simple controller H Higher order controller d Digital output a Analog output
If "higher order controller functions" have been configured, both the digital outputs and the analog outputs can be controlled.
The operator must configure which of the outputs will be controlled - the main board or optional board 1, 2 or 3
Additional explanations, see chapter 16.1 "Glossary", page 93.
76

11.4 Configuration of higher order controllers

Controllers
Controller 1 Controller 2
Parameter set 2
Parameter set 1
Parameter set 2
Parameter set 1
Controller 1
Output 1
Controller 1
Output 2
Controller 2
Output 1
Controller 2
Output 2
Software
Configuration Configuration
Controller special functions
(if necessary)

11.4.1 Structure

11 Controller

11.5 Parameter sets

Different process steps may require different controller settings. The instrument offers the option of creating two parameter sets and then switching between them by means of a binary input.
Defining a parameter set
ADMINISTR.-LEVEL/PARAMETER LEVEL/CONTROLLER 1(2)/ PARAMETER SET 1(2) see "Controller", page 108.
Configuring parameter set switchover
ADMINISTR.-LEVEL/PARAMETER LEVEL/BINARY INPUTS/ BINARY INPUT 1(2)/PARAMET. SWITCHOVER
see "Binary inputs", page 108.
77
11 Controller

11.6 Sample configurations

11.6.1 Simple limit monitoring

Configuration
Limit monitoring Limit value 1
Signal source: Main value Switching function: Alarm function (AF8) Switching point: 10.00 MOhm × cm Hysteresis: 0.50 MOhm × cm
Configuration of binary output, e.g. relay)
Binary outputs Binary output 1
Signal source: Limit monitoring 1 At calibration: Standard operation Error: Inactive HOLD mode: Frozen Turn-on delay: 0 seconds Turn-off delay: 0 seconds Wiper time: 0 seconds Manual mode: No simulation

11.6.2 Limit monitoring to USP

Configuration
Limit monitoring Limit value 1
Signal source: Main value Switching function: USP Switching point: derived automatically from table, see "Excerpt
Hysteresis 0.50 µS/cm
from USP <645>", page 100
Configuration of binary output, e.g. relay)
Binary outputs Binary output 1
Signal source: Limit monitoring 1 At calibration: Standard operation
78
Error: Inactive HOLD mode: Frozen Turn-on delay: 0 seconds Turn-off delay: 0 seconds Wiper time: 0 seconds Manual mode: No simulation

11.6.3 Controller with limit value function

Configuration of software controllers
Controller 1 Configuration
Controller type: Pulse value Controller actual value Stroke retransmission Additive disturbance Multiplicative disturbance Min./max. contact: Max. contact Inactive/active contact: Active contact
1
:Main variable
1
: No signal
1
: No signal
1
: No signal
11 Controller
HOLD mode 0 % HOLD output: 0 % Error: 0 % Alarm control: Off
Parameter set 1
Min. setpoint: As required Max. setpoint: As required Setpoint: 0.80 mS/cm Hysteresis: As required On-delay: As required Delayed release: As required Alarm delay: As required
Configuration of binary output, e.g. relay)
Binary outputs Binary output 1
Signal source: Controller 1 output 1
1
This parameter only appears if "Separate controllers" has been configured in special controller functions.
79

12 Setup program

(1)
(2)
(3)

12.1 Configurable parameters

Both the setup program (00560380) and the PC interface cable with USB/TTL converter (00456352) are available as options and provide a convenient way to adapt the transmitter to meet requirements:
- Setting the measuring range.
- Setting the behavior of outputs when the measuring range is exceeded.
- Setting the functions of switching outputs K1 to K8.
- Setting the functions of the binary inputs.
- Setting a customized characteristic
-etc.
Data can only be transferred from or to the transmitter if it is supplied with voltage, see chapter 5 "Installation", page 14ff.
Connection
(1) JUMO dTRANS 02 CR (2) PC interface cable with USB/TTL converter,
Part no. 00456352
(3) PC or notebook
80
12 Setup program
(2)
(1)

12.2 Documenting the instrument configuration

Start the setup programEstablish the connection to the instrument (1).
Read the instrument configuration (2).
81
12 Setup program
(2)
(1)
(3)
(4)

12.3 Special features for "Datalogger"

Start the setup programEstablish the connection to the instrument (1).Read the instrument configuration (2).
Read data from datalogger (for example table view)
- Mark datalogger icon (3)
- Read values from the instrument (4)
82
12 Setup program
Export data (for processing in an external program).
83

13 Eliminating faults and malfunctions

Problem Possible cause Action
No measurement display or current output
Measurement display 0000 or current output 4 mA
Incorrect or fluctuating measurement display
There is no voltage supply Check the voltage supply
Sensor not immersed in medium; level in container too low
Flow-through fitting is blocked Clean the flow-through fitting Sensor faulty Replace the sensor Sensor faulty Replace the sensor Sensor positioning incorrect Choose another installation
Air bubbles Optimize assembly Measurement overrange
Measurement underrange
Main input: Measurement range "out of range"
Top up the container
location
Choose a suitable measuring range
Compensation range has been left
Measurement overrange
Measurement underrange
Temperature input: Measurement range "out of range"
Compensation range has been left
Temperature input: Measurement range "out of range"
Coating
Choose a suitable measuring range
Choose a suitable measuring range
Clean electrodes. Replace conductivity sensor.
84
13 Eliminating faults and malfunctions
Configuration change OK
Configuration change OK
Inhibit via binary contact Check configuration and unlock if
necessary
Do not release If appropriate release in the
release level
Te st
Inhibit via binary contact Check configuration and unlock if
necessary
Cancel in basic setting OK
Check hardware
Check fitting, adjust if necessary
Instrument had no voltage supply for a very long time
Establish voltage supply Set the datalogger time
85

14 Technical data

Inputs (main board)
Main input Measuring range/
µS/cm 0.000 to 9.999
mS/cm 0.000 to 9.999
kΩ × cm 0.000 to 9.999
MΩ × cm 0.000 to 9.999
Secondary input
Temperature Pt100/1000 -50 to +250 °C Temperature NTC/PTC 0.1 to 30 kΩ
Standard signal 0(4) to 20 mA or 0 to 10 V 0.25 % of range 0.2 %/10 K Resistance transmitter Minimum: 100 Ω
a
In the range between 1 to 10 S the accuracy is 1 % of the measuring range.
b
Selectable in °F
control range
00.00 to 99.99
000.0 to 999.9 0000 to 9999
00.00 to 99.99
000.0 to 999.9 0000 to 9999
00.00 to 99.99
000.0 to 999.9 0000 to 9999
00.00 to 99.99
000.0 to 999.9 0000 to 9999
Entry via table with 20 value pairs
Maximum: 3 kΩ
a
b
Accuracy Effect of temperature
0.6 % of range + 0.3 µS × cell constant (K)
0.6 % of range + 0.3 µS × cell constant (K)
0.6 % of range + 0.3 µS × cell constant (K)
0.6 % of range + 0.3 µS × cell constant (K)
0.25% of range 0.2 %/10 K1.5 % of range 0.2 %/10 K
±5 Ω 0.1%/10K
0.2%/10K
0.2%/10K
0.2%/10K
0.2%/10K
Resistance thermometer inputs (optional board)
Designation Connection type Measuring range Measuring accuracy Effect of ambient
3-wire/4-wire 2-wire
Pt100 DIN EN 60751 (factory-set)
Pt1000 DIN EN 60751 (factory-set)
Sensor lead resistance Maximum 30 Ω per line with three- and four-wire circuit Measurement current Approx. 250 µA Lead compensation Not required for three- and four-wire circuit. With a 2-wire circuit, lead resistance can be
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
compensated in the software by correcting the process value.
-200 to +850 °C 0.05 % 0.4 % 50 ppm/K
-200 to +850 °C 0.1 % 0.2 % 50 ppm/K
temperature
Standard signals inputs (optional board)
Designation Measuring range Measuring accuracy Ambient temperature
Voltage 0(2) to 10 V
Electrical current 0(4) to 20 mA,
Resistance transmitter Minimum: 100 Ω
0to1V Input resistance
voltage drop 1.5 V
Maximum: 4 kΩ
> 100 kΩ
E
0.05 % 100 ppm/K
0.05 % 100 ppm/K
±4 Ω 100 ppm/K
effect
86
14 Technical data
Temperature compensation
Ye s Ye s Ye s Ye s
a
Ye s No Ye s No
Ye s No Ye s No
Type of compensation Range
Linear 0 to 8 %/K -10 to +160 °C ASTM D1125 - 95 (ultra-pure water) 0 to 100 °C Natural waters (ISO 7888) 0 to 36 °C
Reference temperature
Adjustable from 15 to 30 °C; preset to 25 °C (default)
a
Note the sensor operating temperature range!
Measuring circuit monitoring
Inputs Underrange/overrange Short circuit Broken lead
Conductivity Yes Depends on measuring range Depends on measuring range Temperature Yes Yes Yes Voltage 2 to 10 V
0to10V
Current 4 to 20 mA
0to20mA
Resistance transmitter No No Yes
Two-electrode systems
Cell constant
[1/cm]
0.01
0.1 0.02 to 0.5
1.0 0.2 to 5
3.0 0.6 to 15
10.0 2.0 to 50
Setting range of the relative cell constant
20 to 500 %
Resulting usable range
[1/cm]
0.002 to 0.05
Four-electrode systems
Cell constant
[1/cm]
0.5
1.0 0.2 to 1.5
Setting range of the relative cell constant
20 to 150 %
Resulting usable range
[1/cm]
0.1 to 0.75
Binary input
Activation Floating contact is open: function is not active
Function Key lock, manual mode, HOLD, HOLD inverse, alarm suppression, freeze measured value,
Floating contact is closed: function is active
level lock, reset partial quantity, reset total quantity, parameter set switchover
Controller
Controller type Limit comparators, limit controllers, pulse length controllers, pulse frequency
Controller structure P/PI/PD/PID
controllers, modulating controllers, continuous controllers
87
14 Technical data
Outputs
Relay (changeover)
Contact rating Contact service life
Voltage supply for 2-wire transmitter
Voltage supply for inductive proximity switch
Relay (changeover)
Contact rating Contact service life
Relay SPST (normally open)
Contact rating Contact service life
Solid state relay
Contact rating Protective circuit
PhotoMOS® relay Optional board U ≤ AC/DC 50 V
Voltage
Output signals Load resistance Accuracy
Electrical current
Output signals Load resistance Accuracy
PSU board
PSU board Electrically isolated, non-controlled
Optional board DC 12 V; 10 mA
Optional board
Optional board
Optional board
Optional board
Optional board
5 A at AC 240 V resistive load 350,000 operations at nominal load/750,000 operations at 1 A
DC 17 V at 20 mA, open-circuit voltage approx. DC 25 V
8 A at AC 240 V resistive load 100,000 operations at nominal load/350,000 operations at 3 A
3 A at AC 240 V resistive load 350,000 operations at nominal load/900,000 operations at 1 A
1 A at 240 V Varistor
I 200 mA
0to10V or 2to10V R
500 Ω
load
0.5 %
0to20mA or 4to20mA R
500 Ω
load
0.5 %
Display
Type LC graphic display, blue with background lighting, 122 × 32 pixels
Electrical data
Voltage supply (switch-mode PSU) AC 110 to 240 V +10/-15 %; 48 to 63 Hz or
AC/DC 20 to 30 V; 48 to 63 Hz
Electrical safety To DIN EN 61010, Part 1
Overvoltage category II, pollution degree 2 Power draw Approx. 14 VA (20 A fuse max.) Data backup EEPROM Electrical connection On the back via screw terminals,
Electromagnetic Compatibility EMC)
Interference emission Interference immunity
conductor cross-section up to max. 2.5 mm
DIN EN 61326-1
Class A
To industrial requirements
2
Enclosure
Enclosure type Plastic enclosure for panel mounting to DIN IEC 61554 (indoor use) Depth behind panel 90 mm Ambient temperature
Storage temperature Climatic rating Rel. humidity 90 % annual mean, no condensation Site altitude Up to 2000 m above sea level Operating position Horizontal Enclosure protection
In the panel enclosure In the surface-mounted enclosure
Weight (fully fitted) About 380 g
-5 to +55 °C
-30 to +70 °C
To DIN EN 60529
Front IP65, rear IP20
IP65
88
Interface
Modbus
Interface type RS422/RS485 Protocol Modbus, Modbus Integer Baud rate 9600, 19200, 38400 Device address 0 to 255 Max. number of nodes 32
PROFIBUS-DP
Device address 0 to 255
Approvals/marks of conformity
14 Technical data
Mark of conformity
c UL us Underwriters Laboratories E 201387 UL 61010-1
Testing laboratory Certificates/certification
numbers
Test basis valid for
CAN/CSA-C22.2 No. 61010-1
Type 202552/01...
89

15 Retrofitting optional boards

Caution:
The instrument must be de-energized on the input and output sides! Optional boards must only be retrofitted by qualified specialists.
ESD:
Optional boards can be damaged be electrostatic discharge. You must therefore prevent electrostatic charges from accumulating during installation and removal. Optional boards should be retrofitted at a grounded workstation.
15.1 Identifying an optional board
The packaging of the optional board is identified by a sales number.
Optional board Code Part no. Board view
Analog input (universal) 1 00442785
Relay (1× changeover) 2 00442786
Relay (2× NO) This board must only be
inserted in optional slot 1 or 3!
Analog output 4 00442788
2 PhotoMOS
®
relays 5 00566677
3 00442787
90
15 Retrofitting optional boards
Optional board Code Part no. Board view
Solid state relay 1 A 6 00442790
Voltage supply output DC ±5 V (e.g. for ISFET)
Voltage supply output DC 12 V (e.g. for inductive proximity switch)
Interface - RS422/485 This board must only be
inserted in optional slot 3!
Datalogger with interface RS422/485 and real-time clock
This board must only be inserted in optional slot 3!
7 00566681
8 00566682
10 00442782
11 00566678
PROFIBUS-DP interface This circuit board must only
be inserted into option slot 3!
Note:
The optional boards detected by the instrument are displayed in "Device information" (see section 6.5.11 "Device info", page 31).
12 00566679
91
15 Retrofitting optional boards
1
2
3
15.2 Removing a plug-in module
(1) Squeeze the front panel together by the left and right sides and remove
the plug-in module.
15.3 Inserting a plug-in module
Caution:
No "3" relays (2× SPST/normally open) may be inserted in slot 2!
n
(1) Slot 1 for optional board (2) Slot 2 for optional board (3) Slot 3 for optional board
(1) Push the optional board into the slot until it locks in place.
(2) Push the device plug-in into the enclosure until it locks in place.
92

16.1 Glossary

(3)
(1)
(2)
Display of measured values STANDARD
The measurement value, measurement variable and temperature of the measuring material are shown in standard display.
Operating mode
(1)
Display bottom (temperature input)
(2)
Display top (analog input measurement value)
(3)
Display of measured values TENDENCY
The operator can quickly see the direction in which the measurement is changing.

16 Appendix

Rising
Greatly
The measurement tendency (trend) is calculated over the last 10 measurement values. So with a sampling interval of 500 ms, the last 5 seconds are considered.
Moderately
Slightly Slightly
Steady
Falling
Moderately
Greatly
93
16 Appendix
PGM
PGM
PGM
EXIT
PGM
Display of measured values BARGRAPH
Values of the main inputs, input options or math channels (signal source) can be represented as a variable bar (a bar graph).
Scaling the bar
Activate "BARGRAPH" as the display of measured values.Select "SCALE START" with .Confirm the selection with .Use and to enter the lower limit of the range to be displayed.Confirm the selection with .Select "SCALE END" with . Use or to enter the upper limit of the range to be displayed.Confirm the selection with .
To return to Measuring mode: Press the key repeatedly or wait for a "timeout".
Display of measured values TREND CHART
Values of the main inputs, input options or math channels (signal source) can be represented as a graph. The current values appear to the right on the screen.
Scaling the display
Activate "TREND CHART" as the display of measured values.Select "SCALE START" with .Confirm the selection with .Use and to enter the lower limit of the range to be displayed.
94
Confirm the selection with .
PGM
PGM
EXIT
Select "SCALE END" with . Use or to enter the upper limit of the range to be displayed.Confirm the selection with .
To return to Measuring mode: Press the key repeatedly or wait for a "timeout".
Display of measured values LARGE DISPLAY
Values of the main inputs, input options or math channels (signal source) can be displayed in large format.
16 Appendix
Display of measured values 3 MEAS. VALUES
Three values of the main inputs, input options or math channels (signal source) can be displayed simultaneously.
The position of the value to be displayed can be set to "Top", "Center" or "Bottom".
Relative cell constant
Mechanical or chemical effects can change the electrical properties of a conductivity sensor. This will result in a measurement error. This deviation (and thus the measurement error as well) can be compensated for by adjusting the relative cell constant in the transmitter. The relative cell constant defines the deviation of the actual cell constant of the conductivity sensor from its nominal value.
95
16 Appendix
pH
mV
123 456 89 10 11 12 13 140
7
B
A
pH
mV
123 456 89 10 11 12 13 140
7
B
A
Zero point (1-point) calibration
With one-point offset calibration, the zero point of the pH combination electrode is calculated, see chapter 8.4 "Calibrating the relative cell constant", page 51. Recommended only for special applications, such as ultra-pure water.
2-point calibration
With two-point calibration, the zero point and slope of the combination electrode are calibrated. This is the recommended calibration for most sensors.
Temperature compensation (conductivity or resistance)
The conductivity of a measurement solution is temperature-dependent (the conductivity of a solution rises as the temperature increases). The dependency of conductivity and temperature describes the temperature coefficient of the measurement solution. As conductivity is not always measured for the reference temperature, automatic temperature compensation is integrated in this instrument. The transmitter uses the temperature coefficient to calculate
96
16 Appendix
the conductivity that would exist for a reference temperature from the current conductivity and the current temperature. This is then displayed. This process is called temperature compensation. Modern transmitters offer different ways to perform this temperature compensation.
- Linear compensation (constant temperature coefficient). This type of compensation can be applied to many kinds of normal water, with acceptable accuracy. The temperature coefficient used is then approx.
2.2 %/°C
- Natural water (EN27888 or ISO 7888). In this case, so-called non-linear temperature compensation is used. According to the standard cited above, the relevant type of compensation can be applied to natural groundwater, spring water and surface water. The definition range for the water temperature is as follows: 0°C ≤ T < 36 °C Conductivity of the water is compensated in the range from 0 °C to 36 °C.
- ASTM1125-95. This type of temperature compensation is used in measurements of ultra­pure water. The highly non-linear nature of the temperature dependency for neutral, acidic and alkaline impurities is taken into consideration in accordance with the standard. The definition range for the water temperature is as follows: 0 °C < T < 100 °C. Conductivity of the water is compensated in the range from 0 °C to 100 °C.
Temperature compensation (pH or ammonia)
The pH value of a measurement solution depends on the temperature. Since the pH value is not always measured at the reference temperature, the instrument is able to perform a temperature compensation.
The sensor signal for the ammonia measurement is temperature-dependent. The instrument can perform temperature compensation.
The redox potential of a measurement solution is not temperature-dependent! Temperature compensation is not required.
97
16 Appendix
HySt
x
0
w
AF
1
HySt
0
w
AF
x
1
HySt
x
w
0
1
HySt
w
x
0
1
Process value X
Setpoint W
Proportional band X
P
100%
50%
0%
Output level y
Switching period
10%
90%
90%
10%
t
On
50%
50%
t
Off
10%
90%
X
P
X - W
0
1
Limit value (alarm) function of the binary outputs
AF1
AF2
AF7
AF8
Pulse length controller (output active with x > w and P control structure)
98
If actual value x exceeds setpoint W, the P controller will control in proportion
16 Appendix
100%
50%
0%
No pulses
50% of pulse frequency
Maximum pulse frequency
Setpoint W
Proportional band X
P
X
P
X - W
0
1
Output level y
Process value X
to the control deviation. When the proportional range is exceeded, the controller operates with an output level of 100 % (100 % clock ratio).
Pulse frequency controller (output active with x > w and P control structure)
If actual value x exceeds setpoint W, the P controller will control in proportion to the control deviation. When the proportional range is exceeded, the controller operates with an output level of 100 % (maximum switching frequency).
Special controller functions: Separate controllers
This function is normally deactivated (factory setting or select "No"). In the deactivated state, the software prevents the two controller outputs from
being able to work "against each other". So, for example, it is not possible to dose acid and lye at the same time.
If the controllers are separate ("Yes" selection), each controller can be freely configured.
Switch-off of the I-component
This function is normally deactivated (factory setting or select "No"). In the deactivated state, the controller works in accordance with general
controller theory. When I-component switch-off is activated ("Yes" selection), the part of the
output level that can be traced back to the I-component is set to zero when the setpoint is reached.
This can be useful with mutual neutralization (acid and lye dosing both possible) in one treatment tank.
Calibration timer
The calibration timer indicates (on request) a required routine calibration. The calibration timer is activated by entering the number of days that must expire before there is a scheduled re-calibration (specified by the system or the operator).
Wash timer
The wash timer can be used to implement automated sensor cleaning. To do this, the function is assigned to a switching output.
99
16 Appendix
The cycle time (cleaning interval) can be adjusted in the range from 0.0 to
240.0 hours.
A cycle time of "0.0" means the wash timer is deactivated. The wash time (cleaning duration) is adjustable from 1 to 1800 seconds. During the wash time the controller goes into the HOLD state, which is
maintained for 10 seconds after completion of the wash time. A sensor calibration within the cycle time restarts the wash timer.
USP contact (for ultra-pure water)
The USP contact makes it possible to monitor the quality of ultra-pure water according to the requirements of USP <645>. USP <645> contains a table that assigns a limit value for conductivity depending on the temperature. If the conductivity stays below this limit value, the ultra-pure water meets the requirements of USP <645>.
If the conductivity of the water is greater than what is specified in the USP table for a given temperature, the USP contact switches the instrument.
Limit values are defined in levels. For example, a value of 5 °C is used at 8 °C.
Note:
During monitoring, temperature compensation must be turned off (temperature coefficient = 0)! To do this, select Administrator Level/Basic Setting/ Temperature Compensation/None.
Excerpt from USP <645>
Te mp er at ur e
°C
0 0.6 55 2.1
5 0.8 60 2.2 10 0.9 65 2.4 15 1.0 70 2.5 20 1.1 75 2.7 25 1.3 80 2.7 30 1.4 85 2.7 35 1.5 90 2.7 40 1.7 95 2.9 45 1.8 100 3.1 50 1.9
USP warning alarm
µS/cm (uncompensated)
If the conductivity is exceeded at the relevant temperature, the configured contact switches.
The USP warning alarm switches before the water quality reaches the set limit value. This parameter (0 to 100) is used to set the distance as a percentage (relative to the active limit value) to be maintained from the USP limit.
Max. conductivity
Te mp er at ur e
°C
Max. conductivity
µS/cm (uncompensated)
100
Loading...