JUMO dTRANS AS 02, 202553 Operating Manual

Page 1
JUMO dTRANS AS 02
Transmitter/controller
for standard signals and temperature
Type 202553
Operating Manual
20255300T90Z002K000
V1.00/EN/00541524
Page 2
WARNING:
EXIT
PGM
PGM
PGM
Note:
Please read these Operating Instructions before placing the instrument in operation. Keep the manual in a place which is accessible to all users at all times.
Resetting the brightness of the LC display:
If the brightness setting has been adjusted so that the display text is no longer legible, the basic setting can be restored as follows:
Switch off the voltage supply.Switch on the voltage supply and immediately press and hold the and
keys simultaneously.
Operator language selection:
Press the key for longer than 3 seconds.Select the appropriate language with the and keys.Briefly press the key.
Reset to factory settings:
To get to the Administrator level, proceed as follows:
Press the key for longer than 2 seconds.Use the or keys to select "ADMINISTR. LEVEL".Use the and keys to enter the password 8192.
Confirm the key.
WARNING:
Customer-specific settings will be lost!
Page 3
Contents
1 Typographical conventions ...................................................... 5
1.1 Warning signs ..............................................................................................5
1.2 Reference signs ...........................................................................................5
2 Description ................................................................................ 6
3 Instrument identification .......................................................... 8
3.1 Nameplate ....................................................................................................8
3.2 Order details .................................................................................................8
3.3 Accessories (included in delivery) ..............................................................10
3.4 Accessories (optional) ................................................................................10
4 Mounting .................................................................................. 11
4.1 General information ....................................................................................11
4.2 Dimensions ................................................................................................11
5 Installation ............................................................................... 12
5.1 Installation instructions ..............................................................................12
5.2 Electrical isolation ......................................................................................13
5.3 Connection .................................................................................................14
6 Operation ................................................................................. 18
6.1 Controls ......................................................................................................18
6.2 Display .......................................................................................................19
6.3 Principle of operation .................................................................................20
6.4 Measuring mode ........................................................................................23
6.5 Input/output information ............................................................................24
6.6 User level ...................................................................................................29
6.7 Administrator level .....................................................................................30
6.8 MANUAL mode/Simulation mode ..............................................................33
6.9 HOLD mode ...............................................................................................36
7 Commissioning ....................................................................... 38
7.1 Getting started ...........................................................................................38
7.2 Setting examples .......................................................................................39
8 Calibration ............................................................................... 45
8.1 Notes ..........................................................................................................45
8.2 General information ....................................................................................45
8.3 Linear operating mode ...............................................................................48
8.4 pH operating mode ....................................................................................52
Page 4
Contents
8.5 Conductivity operating mode .....................................................................56
8.6 Concentration operating mode ..................................................................62
8.7 Chlorine measurement operating mode, pH-compensated 64
9 Calibration logbook ................................................................ 66
9.1 General information ....................................................................................66
10 Controller ................................................................................. 67
10.1 General information ....................................................................................67
10.2 Controller functions ....................................................................................67
10.3 Software controllers and outputs ...............................................................68
10.4 Configuration of higher order controllers ...................................................70
10.5 Parameter sets ...........................................................................................70
10.6 Sample configurations ...............................................................................71
11 Setup program ........................................................................ 74
11.1 Configurable parameters ...........................................................................74
11.2 Documenting the instrument configuration ................................................75
11.3 Special features for "Datalogger" ..............................................................76
12 Eliminating faults and malfunctions ...................................... 78
13 Technical data ......................................................................... 80
14 Retrofitting optional boards ................................................... 83
15 Appendix .................................................................................. 86
15.1 Glossary .....................................................................................................86
15.2 Parameters of the User level ......................................................................99
16 Index ....................................................................................... 111
Page 5

1.1 Warning signs

Danger
This symbol is used when there may be danger to personnel if the instructions are ignored or not followed correctly!
Caution
This symbol is used when there may be damage to equipment or data if the instructions are ignored or not followed correctly!
Read documentation! This symbol – placed on the device – indicates that the associated device
documentation has to be observed. This is necessary to recognize the kind
of the potential hazards as well as to take the measures to avoid them.

1 Typographical conventions

1.2 Reference signs

Note
This symbol is used to draw your special attention to a remark.
1
abc
Instruction
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes consist of two parts:
A marker in the text and the footnote text. The markers in the text are arranged as consecutive superscript numbers.
This symbol indicates the description of an action to be performed. The individual steps are marked by this asterisk. Example: Briefly press the key.
5
Page 6

2 Description

Inputs/outputs In addition to the main input (conductivity, TDS, resistance) and the secondary
input (temperature compensation), the basic instrument alone has two binary inputs, two relays, one power supply for external sensors and a setup interface.
Input signals can be shown as numbers or as a bar graph on the graphic display. Parameters are displayed in plain text for easily comprehensible and reliable operation.
Optional Three further slots can be fitted with extensive additional configurable inputs
and outputs and interfaces.
Application The instrument is suitable, for example, for displaying, measuring and
controlling:
- Conductivity, TDS and resistance.
- Free chlorine, chlorine dioxide, ozone, hydrogen peroxide and peracetic acid, in combination with sensors as per data sheet 202630.
- (Hydrostatic) liquid levels with 2-wire transmitters (level probes) as per data sheet 402090 or data sheet 404390.
- Flow rate in conjunction with transmitters as per data sheet 406010.
- Two temperature measuring points.
- Most sensors and transmitters that output standard signals (0 to 10 V or 0(4) to 20 mA).
Because temperature measurement is integrated, temperature compensation takes place quickly and precisely, which is particularly important for many analytical measurements.
6
Page 7
2 Description
Special features - Display: mS/cm, µS/cm, MOhm × cm, mg/l, pH, mV, etc. Special settings
are also possible with the setup program
- Configurable display text (operator level)
- A choice of display visualizations: large numbers, bar graph or tendency (trend) display
- Four limit controllers
- Integrated calibration routines: with 1, 2 and 3 points
- Math and logic module (optional)
- Calibration logbook
- Three optional slots
- Selectable languages: English, German, French, etc.
- Setup program provides: convenient programming, system documentation
- RS422/485 interface (optional)
- PROFIBUS-DP interface (optional)
7
Page 8

3 Instrument identification

Typ: 202553/01-8-01-4-8-0-23/000
TN: 00566868
JUMO GmbH & Co. KG dTRANS AS 02
36039 Fulda Germany
AC .. 0V 48..63Hz/DC 20 3
max 14VA
F-Nr.: 0169924101016010001

3.1 Nameplate

on the transmitter
The date of manufacture is encoded in the "F No." (serial number): 1601 means year of manufacture 2016 / calendar week 01

3.2 Order details

(1) Basic type
202553 JUMO dTRANS AS 02 - Transmitter/controller
(2) Basic type extension
01 In the panel enclosure 05 In the surface-mounted enclosure
(3) Version
8 Standard with factory setting 9 Programming to customer specification
(4) Operating language
01 German 02 English 03 French 04 Dutch 05 Russian 06 Italian 07 Hungarian 08 Czech 09 Swedish 10 Polish 13 Portuguese 14 Spanish 16 Rumanian
a
8
Page 9
3 Instrument identification
(5) Optional slot 1
0Not used 1 Analog input (universal) 2 Relay (1× changeover) 3 Relay (2× normally open) 4 Analog output 52 PhotoMOS
®
relays 6 Solid state relay 1 A 8 Voltage supply output DC 12 V (e.g. for inductive proximity switch)
(6) Optional slot 2
0Not used 1 Analog input (universal) 2 Relay (1x changeover) 4 Analog output 52 PhotoMOS
®
relays 6 Solid state relay 1 A 8 Voltage supply output DC 12 V (e.g. for inductive proximity switch)
(7) Optional slot 3
0Not used 1 Analog input (universal) 2 Relay (1x changeover) 3 Relay (2x normally open) 4 Analog output 52 PhotoMOS
®
relays 6 Solid state relay 1 A 8 Voltage supply output DC 12 V (e.g. for inductive proximity switch)
10 RS485 interface 11 Datalogger with interface RS485 12 PROFIBUS-DP interface
(8) Power supply
23 AC 110 to 230 V, -15/+10 %, 48 to 63 Hz 25 AC/DC 20 to 30 V, 48 to 63 Hz
(9) Extra codes
d
000 None
a
All languages are available on the instrument and can be changed by the customer at any time. Factory default setting to a language (other than "German") is available for a charge.
b
PhotoMOS® is a registered trademark of Panasonic Corporation.
c
The only way to read files is with the PC setup software!
d
List extra codes in sequence, separated by commas.
b
c
(1) (2) (3) (4) (5) (6) (7) (8) (9) Order code /- -----/,... Order example 202553 / 01 - 8 - 01 - 2 - 2 - 4 - 23 / 000
9
Page 10
3 Instrument identification

3.3 Accessories (included in delivery)

4× fastening elements, complete 3× CON plug-in link 3× jumper wire 1× seal for panel 1× fastening elements, complete
- 1× DIN rail fastening left
- 1× DIN rail fastening right
- 3× wall mount
- 3× fastening screw
a
For basic type extension 01 only (in the panel enclosure)
b
For basic type extension 05 only (in the surface-mounted enclosure)
a
b
a
a
b

3.4 Accessories (optional)

Type Part no.
Holder for C rail 00375749 Dummy cover 96 mm × 48 mm 00069680 Pipe mounting set 00398162 Weather protection roof complete for basic type extension 05 00401174 PC setup software 00560380 PC interface cable including USB/TTL converter and two adapters
(USB connecting cable)
00456352
Optional board Code Part no.
Analog input (universal) 1 00442785 Relay (1× changeover) 2 00442786 Relay (2× NO) 3 00442787 Analog output 4 00442788 2 PhotoMOS Solid state relay 1 A 6 00442790 Voltage supply output DC ±5 V (e.g. for ISFET) 7 00566681 Voltage supply output DC 12 V (e.g. for inductive proximity switch) 8 00566682 Interface - RS422/485 10 00442782 Data logger with RS485 interface 11 00566678 PROFIBUS-DP interface 12 00566679
®
relays 5 00566677
10
Page 11

4.1 General information

4 Mounting

Mounting location
Installation position
Find a location that ensures easy accessibility for the later calibration. The fastening must be secure and must ensure low vibration for the
instrument. Avoid direct sunlight! Permissible ambient temperature at the installation location: -10 to +55 °C
with max. 95 % rel. humidity, no condensation. The instrument can be mounted in any position.

4.2 Dimensions

Close mounting
Minimum spacing of panel cutouts Horizontal Vertical Without setup connector: 30 mm 11 mm With setup connector (see arrow): 65 mm 11 mm
11
Page 12

5 Installation

5.1 Installation instructions

The electrical connection must only be performed by qualified personnel!
The choice of cable, the installation and the electrical connection must
conform to the requirements of VDE 0100 “Regulations on the Installation of Power Circuits with Nominal Voltages below 1000 V” and the relevant local regulations.
At maximum load, the cable must be heat resistant up to at least 80 °C.The instrument shall be operated by mains protected with a branch circuitry
overcurrent protection device not more than 20 Amps. For servicing/repairing a Disconnecting Device shall be provided to disconnect all conductors.
The load circuits must be fused for the maximum load currents in each case
to prevent the relay contacts from becoming welded in the event of a short circuit.
Electromagnetic compatibility meets the requirements of EN 61326.Lay the input, output, and supply lines so they are physically separated
from each other and are not parallel.
Use twisted and shielded probe cables. If possible, do not lay these cables
close to components or cables through which current is flowing. Ground the shielding at one end.
The probe cables must have an uninterrupted run (do not route them via
terminal blocks or similar arrangements).
No other consumers can be connected to the power terminals of the
instrument.
The instrument is not suitable for installation in areas with an explosion
hazard.
Apart from faulty installation, incorrect settings on the instrument may also
affect the proper functioning of the subsequent process or lead to damage. You should therefore always provide safety equipment that is independent of the instrument and it should only be possible for qualified personnel to make settings.
Mounting information for conductor cross-sections and ferrules
Ferrule Conductor cross-section Minimum length of ferrule or
Mi ni mu m Ma xi mu m
Without ferrule 0.34 mm Without collar 0.25 mm With collar up to 1.5 mm
2
0.25 mm
Twin, with collar 0.25 mm
2
2
2
2
2.5 mm
2.5 mm
1.5 mm
1.5 mm
2
2
2
2
stripping
10 mm (stripping) 10 mm 10 mm 12 mm
12
Page 13

5.2 Electrical isolation

3700 V AC
Setup interface
Power supply
Power supply
for a 2-wire transmitter
3700 V AC
Semi Semiconductor relay Triac
Binary inputs
30 V AC 50 V DC
Continuous outputs
30 V AC 50 V DC
Input 2 (option)
3700 V AC
Relay outputs
Main input for standard signals
30 V AC 50 V DC
PhotoMOS®relay
30 V AC 50 V DC
Interface RS422/485 PROFIBUS-DP Datalogger
Power supply for ISFET sensor
Extra-low voltage
30 V AC 50 V DC
5 Installation
13
Page 14
5 Installation
1
2
3
4
1
2
3
4
5
6
7
8
6
7
8
9
10
9
10
11
12
4
L1(L+)
5
6
8
9
11
12
13
15
16
17
N(L-)
11
12
13
14
15
(a)
(b)(c)
(1)
(3)
(2)
J
J
E
S
+
-

5.3 Connection

5.3.1 Terminal assignment

(1) Row 1 (a) Option 1 (b) Option 2 (c) Option 3 (2) Row 2 Main input board
(standard signal and temperature)
(3) Row 3 PSU board
(power supply/2× relays)

5.3.2 Optional board (row 1, slot a, b or c)

Function Symbol Terminal
Analog input Temperature sensor
in a two-wire circuit
Pt100 or Pt1000
Temperature sensor in a three-wire circuit
Pt100 or Pt1000
Resistance transmitter 2
for slot (a)
2 4
3 2 4
3 4
Te rm in al
for slot (b)
6 8
7 6 8
6 7 8
Ter mi na l
for slot (c)
10 12
11 10 12
10 11 12
Electrical current 3
14
4
7 8
11 12
Page 15
5 Installation
+
-
+
-
+
-
RxD-
RxD+
TxD-
TxD+
R
xD/TxD-
RxD/TxD+
RxD/TxD-P(B)
VP(+5V)
DGND
RxD/TxD-N(A)
RxD/TxD-
RxD/TxD+
S
P
O
S
P
S
Function Symbol Terminal
for slot (a)
Voltage
0(2) to 10 V
Voltage
0to1V
1 2
2 3
Te rm in al
for slot (b)
5 6
6 7
Ter mi na l
for slot (c)
9
10 10
11
Continuous output Current or voltage 2
3
6 7
10 11
Modbus interface
RS422 9
10 11 12
RS485 11
12
PROFIBUS-DP interface
9 10 11 12
Datalogger interface
RS485 10
11
Relay (1× changeover)
K3 1
K4 5 2 3
6 7
K5 9
Relay (2× NO, common pin)
K3 1
K5 9
2
K6 3
K8 11
Triac (1 A)
PhotoMOS
®
relay (0.2 A)
K3 2
3
K3 1
2
K4 6
K4 5
K5 10
7
K5 9
6
10 11
10
11
10
K6 3
K7 7 4
8
K8 11
12
15
Page 16
5 Installation
+
-
+
-
+
-
J
J
J
E
S
A
Function Symbol Terminal
for slot (a)
Te rm in al
for slot (b)
Ter mi na l
for slot (c)
Power supply voltage for proximity switch
DC +12 V GND
1 2
5 6
9
10

5.3.3 Main board (row 2)

Function Symbol Terminal Standard signal input for electrical current
0(4) to 20 mA
Standard signal input for voltage
0(2) to 10 V or 10 to 0(2) V
Temperature sensor in a two-wire circuit
Pt100 or Pt1000
Temperature sensor in a three-wire circuit
Pt100 or Pt1000
Temperature sensor in a four-wire circuit
Pt100 or Pt1000
Resistance transmitter 4
3 4
1 2
2 3 4
2 3 4
1 2 3 4
3 2
Binary inputs
Binary input 1 6+
10
Binary input 2 7+
10
16
Page 17
5 Installation
S
P
O
S
P
O

5.3.4 PSU board (row 3)

Function Symbol Terminal Power supply for JUMO dTRANS 02
Power supply: AC 110 to 240 V
Power supply: AC/DC 20 to 30 V
n.c. 4
Supply voltage for external 2-wire transmitter
DC 24 V (-15/+20 %) 8 L+
Relay 1
Switching output K1 (floating)
1 L1 (L+)
2 N (L-)
5 6
9 L
11 12 13
Relay 2
Switching output K2 (floating)
15 16 17
17
Page 18

6 Operation

(8)
(5)
(2)
(4)
(3)
(1)
(6)
(7)
PGM
EXIT
Operation via the instrument keypad is described below. Instrument operation via the optional set-up program, See section 11 "Setup
program", page 74.

6.1 Controls

(1) Measurement unit
(2) Temperature
(3) Operating mode
(4) Measured value
(5) key Increase numerical value/Forward selection
(6) key Decrease numerical value/Forward selection
(7) key Change level/Forward selection/Confirm selection
(8) key Cancel entry/Exit level
18
Page 19

6.2 Display

(1) (2) (3) (5)(4) (6)
(8)
(9)
(7)
EXIT

6.2.1 Measuring mode (normal display)

Example
6 Operation
(1) Binary output (relay) K1 is active (2) Binary output (relay) K2 is active (3) Binary input is active (4) Keypad is locked (5) Instrument status
ALARM (flashing): Broken sensor or overrange, etc. AL R1: Controller monitoring alarm from controller channel 1 AL R2: Controller monitoring alarm from controller channel 2 CALIB: Calibration mode active CALIB (flashing): Calibration timer elapsed
(6) Output mode
MAN.: Manual mode and/or simulation mode active HOLD: Hold mode active
(7) Top display
Measured value and unit of the variable set by parameter "Top display"
(8) Bottom display
Measured value and unit of the variable set by parameter "Bottom display"
(9) Operating mode
MEASURING: Standard measuring mode is active
To return to Measuring mode (MEASURING): Press the key or wait for a "timeout".
19
Page 20
6 Operation

6.3 Principle of operation

6.3.1 Operation in levels

Measurement mode
Normal display 23 Min/max values of the main input 25 Min/max values of the optional inputs 26 Output display 26 Current values of the main input 26 Current values of the optional inputs 27 Current values of the math channels 27 States of the binary inputs and outputs 27 Manual mode overview 28 Hardware information 28 Instrument information 29 User data 75 Calibration (depending on the basic setting) 45 Manual mode/simulation 33 Hold mode 36
Main menu
User level 29
Analog input 99 Input temperature 100 Optional inputs 100
Analog input 1, 2, 3
Binary inputs 102
Binary input 1, 2
Controllers 102
Controller 1
Controller 2
Controller special functions 104
Limit value control 105
Limit value 1, 2, 3
Binary outputs 102
Binary output 1, 2, 3, ... 8
Analog outputs 107
Analog output 1, 2, 3 Interface 108 Wash timer 108 Datalogger 108
See page
Parameter set 1, 2 Configuration
Parameter set 1, 2 Configuration
20
Page 21
6 Operation
Display 109
Administrator level (password) 30
Parameter level 30
Parameters as above for "User level" Release level 30
Parameters as above for "User level" Basic setting 30 Calibration level 33
Main input (depending on the basic setting)
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point 3-point
Optional input 1, 2, 3
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point 3-point
Calibration release 33
Main input (depending on the basic setting)
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point 3-point
Optional input 1, 2, 3
Temperature coefficient, linear Temperature coefficient, curve Relative cell constant Zero point Limit point 2-point 3-point
Delete min/max values 33
Main input
Optional input 1, 2, 3
21
Page 22
6 Operation
Calibration level (depending on the basic setting) 45
Calibration logbook 66
Instrument information 29
Delete logbook 33
Main input
Optional input 1, 2, 3 Delete daily batch 33 Delete total batch 33
Main input
Temperature coefficient, linear
Temperature coefficient, curve
Relative cell constant
Zero point
Limit point
2-point
3-point Optional input 1, 2, 3 100
Temperature coefficient, linear
Temperature coefficient, curve
Relative cell constant
Zero point
Limit point
2-point
3-point
Main input Optional input 1, 2, 3
22
Page 23

6.4 Measuring mode

EXIT
(3)
(1)
(2)
Different display types can be configured, See "Display of measured values STANDARD" page 86.
To return to Measuring mode: press the key or wait for a "timeout".
Measurements with "out of range" are ignored. The min./max. value memory can be reset:
Administrator level/Delete min/max. When the basic setting is changed, the min and max values are deleted.

6.4.1 Normal display

Visualization
The following are displayed in Measuring mode:
- Analog input signal
- Unit (for example ppm)
- Temperature of the sample medium
6 Operation
(1) MEASURING -> Measuring mode (2) 23.7 °C -> Temperature of the sample medium (3) 1.58 ppm -> measured value
23
Page 24
6 Operation

6.5 Input/output information

24
Page 25

6.5.1 User data

Measuring mode (normal display)
EXIT
Manual mode
(controller)
or
T
ime-out
(adjustable)
> 3 s
PGM
EXIT
+
EXIT
Hold mode
or
T
ime-out
(adjustable)
EXIT
+
Hardware information
+
> 3 s
EXIT
Calibration
or
Time-out
(adjustable)
EXIT
+
Only if released
Keys release
1
1 1
PGM
PGM
6 Operation
Up to 8 parameters that are frequently changed by the user can be combined in the user level under "User data" (via setup program only).
Activating the display
The instrument is in Measuring mode (normal display)
Briefly press the key.Select the required "quick setting" with the and keys.
Editing
Briefly press the key.Edit the setting with the and keys.

6.5.2 Min/max values of the main input

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
Minimum and maximum values of the main value "1:" (mS/cm, µS/cm,
25
Page 26
6 Operation
MOhm × cm, mV, %, ppm) and the temperature "T:" are displayed.
The extreme values of the main measurement variable and the temperature are
not mutually assigned (e.g. not 813 µS/cm at 24.3 °C).

6.5.3 Min/max values of the optional inputs

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
Minimum and maximum values of the optional inputs (1, 2 and 3) are displayed

6.5.4 Output level

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
The current output levels of the controller outputs.

6.5.5 Current values of the main entries

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
26
The current values of the main output are displayed.
Page 27

6.5.6 Curgent values of the optional entries

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
The current values of the optional inputs (1, 2 and 3) are displayed

6.5.7 Current values of the math channels

6 Operation
Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary).
The current values of the main output are displayed.

6.5.8 States of the binary inputs and outputs

Activating the display
The instrument is in Measuring mode (normal display) Briefly press the or key (several times if necessary.
The states of binary inputs E1 and E2 and of relays K1 through K8 are displayed. In the example shown here, relay K1 is active.
27
Page 28
6 Operation
EXIT
PGM

6.5.9 Manual mode overview

Analog outputs (optional boards)
In this example, analog outputs 2 and 3 are working normally.
Switching outputs (PSU board and optional boards)
In this example relay output 2 is in Manual mode.
The instrument is in "normal display" mode Briefly press the or key (several times if necessary).
Manual mode can only be displayed if at least one output is in Manual mode. For example Administrator level/Parameter level/Binary outputs/ Binary output 1/Manual mode "Active" or "Simulation".
To return to Measuring mode: press the key or wait for a "timeout".

6.5.10 Hardware info

These displays are required for phone support.
The instrument is in Measuring mode (normal display) Press and hold the and keys.
Alternating display
28
Page 29

6.5.11 Device info

PGM
PGM
PGM
These displays provide an overview of fitted hardware options and the settings of inputs (helpful for troubleshooting, etc.).
Press the key for longer than 3 seconds. Briefly press the or key (several times if necessary). Select Device info
6 Operation
Press the keys.
Briefly press the or key (several times if necessary).

6.6 User level

All the parameters that the Administrator (See section 6.7 "Administrator level", page 30) has released can be edited at this level. All the other parameters (marked by a key ) are read only.
Press the key for longer than 2 seconds.Select "USER LEVEL".
For further information about the inputs, press the or keys.
29
Page 30
6 Operation
PGM
PGM
All possible parameters are accessed below. Depending on the configuration of a specific instrument, some of these parameters may not appear.

6.6.1 Parameters of the User level

See section 15.2 "Parameters of the User level", page 99

6.7 Administrator level

- All the parameters can be edited at this level.
- At this level, it is also possible to define which parameters can be edited by a "normal" user (operator) and which calibrations can be performed.
To get to the Administrator level, proceed as follows:
Press the key for longer than 2 seconds.Use the or keys to select "ADMINISTR. LEVEL".Use the and keys to enter the password 300 (factory setting).Confirm the key.

6.7.1 Parameter level

The settings that can be made here are the same as those at the User level, See "User level" page 29. As the operator (user) has administrator rights here, the parameters that are locked in the User level can now also be modified.

6.7.2 Release level

All parameters can be released (modification possible) or locked (no modification possible) for editing at the User level.

6.7.3 Basic setting

The JUMO dTRANS 02 AS has a basic setting wizard, to make it easier for the user to configure the extensive setting options of the instrument and to avoid configuration conflicts. The basic settings are reached via ADMINISTR.-LEVEL/PASSWORD/BASIC SETTING.
All the important settings are systematically polled here. At the end, once a request for conformation has been acknowledged, the instrument is initialized with the new settings. Dependent parameters are checked and adjusted.
30
Page 31
Basic setting wizard
6 Operation
31
Page 32
6 Operation
32
Page 33

6.7.4 Calibration level

Depending on which operating mode has been configured (in the Basic setting menu), one or more of the following calibration options will be available:
- Cell constant
- Temperature coefficient

6.7.5 Calibration release

Which calibration procedure may be performed directly and which may not can be configured here, See section 8.2.2 "Ways to start the calibration", page
46.

6.7.6 Delete min/max values

If required, the values can be deleted once a request for confirmation has been acknowledged.
See "Min/max values of the main input" page 25 or See "Min/max values of the optional inputs" page 26.
6 Operation

6.7.7 Delete logbook

The last five calibration processes for each input are archived in the calibration logbook. If a "Datalogger" optional board is fitted, the date and time are also archived. If necessary the logbook can be deleted after a confirmation prompt.

6.7.8 Delete daily batch

If required, the counter can be deleted once a request for confirmation has been acknowledged.

6.7.9 Delete total batch

If required, the counter can be deleted once a request for confirmation has been acknowledged.

6.8 MANUAL mode/Simulation mode

These functions can be used to set the switching outputs and analog outputs of the instrument manually to a defined state. This facilitates dry startup, troubleshooting and customer service, etc.
33
Page 34
6 Operation
EXIT
Simulation mode accesses the analog outputs and binary outputs directly. When simulation mode has been selected, MANUAL mode is not possible!
In MANUAL mode the settings for "higher order controllers" are taken into consideration.

6.8.1 MANUAL mode only via "higher order" controller functions

Select Manual mode
In the factory setting of the instrument the MANUAL mode parameter is locked and can only be activated by the administrator! This parameter must first be released for other users, See "Release level" page 30.
Set ADMINISTR. LEVEL/PARAMETER LEVEL/CONTROLLER/
CTRL.SPEC. FUNCT./MANUAL MODE "Locked, Coding or Switching.
Locked = No Manual mode, control is via device. Coding = The outputs are active as long as the or key is pressed. Switching = the outputs are active if the or key is pressed. If the
corresponding key is pressed again, the output becomes inactive again.
Activate Manual mode
The instrument is in Display mode Press the and keys for less than 2 seconds.
The word MANUAL appears in the status line of the display.
34
Page 35
Deactivation
EXIT
EXIT
EXIT
EXIT
6 Operation
If the keys (alone) are pressed for longer than 3 seconds, the instrument switches to language selection.
If the and keys are pressed for longer than 3 seconds, the instrument goes into HOLD mode.
Then the outputs of the instrument respond according to the default settings. To exit HOLD mode, press the and keys for longer than 3 seconds.
Control is not longer via the instrument. The output level of the controllers is 0%.
Controller 1 is activated by the key. In this case the output level of controller 1 is 100 %.
Controller 2 is activated by the key. In this case the output level of controller 2 is 100 %.
Press the key. Control is once again through the outputs of the instrument.
The word MANUAL appears in the status line of the display.

6.8.2 Simulation of binary outputs

Activate simulation
In the factory setting of the instrument the MANUAL mode parameter is set to "No simulation" and can only be activated by the administrator! This parameter must first be released for other users, See "Release level" page 30.
If a higher order switching function has been assigned to an output, Simulation mode is not possible for that output.
Set ADMINISTR. LEVEL/PARAMETER LEVEL/BINARY OUTPUTS/
BINARY OUTPUT1 (... 8) "Manual mode no simulation, Inactive or Active".
No simulation = No Manual mode, control is via device. Inactive = Relay K1 or K2 is de-energized;the word MANUAL
appears in the status line of the display
Active = Relay K1 or K2 is energized;the word MANUAL appears
in the status line of the display
Deactivate Manual mode
No simulation = No Manual mode, control is via device. When the instrument is in display mode, the word MANUAL disappears from
the status line of the display.
35
Page 36
6 Operation
EXIT
EXIT
EXIT

6.8.3 Simulation of analog outputs via MANUAL mode

Release and activation
Select activation of simulation of the actual value output:
ADMINISTR. LEVEL/PARAMETER LEVEL/ANALOG OUTPUTS/
/SIMULATION/ON.
/SIMULATION/OFF.
Deactivation
ANALOG OUTPUT 1 (2, 3)
With "On" the output takes on the value of the "Simulation value" parameter. When the instrument is in display mode, the word MANUAL appears in the
status line of the display.
ADMINISTR. LEVEL/PARAMETER LEVEL/ANALOG OUTPUTS/ ANALOG OUTPUT 1 (2, 3)
The corresponding output of the instrument works again. When the instrument is in display mode, the word MANUAL disappears from
the status line of the display.

6.9 HOLD mode

In HOLD status the outputs take on the states programmed in the relevant parameter (controller channel, switching output or analog output).
This function can be used to "freeze" switching outputs and the analog outputs of the instrument. This means the current status of the output will be retained even when the measured value changes. Control is not via the instrument.
If MANUAL mode is activated while HOLD mode is activated, MANUAL mode takes precedence and MANUAL then appears in the status line of the display! MANUAL mode can be terminated by pressing the key. If HOLD mode is still activated (by the binary input or by keyboard), the instrument then returns to HOLD mode!
HOLD mode can be activated by pressing the key or by the binary input.
Activation by pressing key
Press and hold the and keys longer than 3 seconds.
Then the outputs of the instrument respond according to the default settings. The word HOLD appears in the status line of the display.
If the and keys are pressed for less than 3 seconds, the instrument goes into Manual mode.
Then the outputs of the instrument respond according to the default settings.
36
Page 37
Pressing a key to deactivate HOLD mode
EXIT
EXIT
Press the and keys for longer than 3 seconds.
If the and keys are pressed for less than 3 seconds, the instrument goes into Manual mode.
Then the outputs of the instrument respond according to the default settings.
Control is through the outputs of the instrument again. The word MANUAL disappears from the status line of the display.
6 Operation
37
Page 38

7 Commissioning

7.1 Getting started

Some suggestions follow for configuring the instrument reliably in little time.
Mount the instrument, See section 4 "Mounting", page 11.Install the instrument, See section 5 "Installation", page 12 ff.Call up Administrator level (ADMINISTR. LEVEL).Enter password 0300 (factory setting).Call up PARAMETER LEVEL/DISPLAY/OPERAT. TIMEOUT.Set OPERAT. TIMEOUT to 0 minutes (no timeout).Leave the Display level with "EXIT"Leave the Parameter level with "EXIT"Select BASIC SETTING and work through all the menu items, See section
6.7.3 "Basic setting", page 30.
Answer "YES" to the "Reinitialize device" query Configure the required additional parameters.Calibrate the instrument to the sensor and sample medium.
38
Page 39
7 Commissioning
(1)
(2)
(5)
(5)
(4)
(3)

7.2 Setting examples

7.2.1 Indicator/controller for pH (temperature compensated)

Measurement of drinking water.
Layout
Electrical connection
Ta sk
Data sheet (1) Transmitter/controller type 202553 202553 (2) 2-wire transmitter for pH on main board 202701 (3) pH combination electrode 201020 (4) Compensation thermometer on optional board 201085 (5) Connecting cables 202950
See section 5 "Installation", page 12.
Measuring range: 2 to 12.00 pH Output signal: 4 to 20 mA Temperature measurement Pt100 Limit monitoring: Limit value function Limit value1: 7.5 pH
39
Page 40
7 Commissioning
Basic setting
Start the basic settingSee section 6.7.3 "Basic setting", page 30 Diagrammatic overview, See section "Basic setting wizard", page 31.
Operating mode pH measurement Signal type 4 to 20 mA Scaling start -600 mV Scaling end +600 mV Temperature compensation source Optional input 1 Supply frequency 50 Hz
Reinitialize device Yes
Temperature input
Administrator level/Password/Parameter level/Optional inputs/Analog input 3
Analog output
Controller settings
Operating mode Temperature Signal type Pt100 Connection type 2-wire Filter time constant 2.0 sec.
Administrator level/Password/Parameter level/Analog outputs/Analog output 1
Signal source Main variable Signal type 4 to 20 mA Start of scaling 2.00 pH End of scaling 12.00 pH
See section 10.6.1 "Simple limit monitoring (min. function)", page 71.
40
Page 41
7 Commissioning
(1)
(2)
(3)
(4)
(4)

7.2.2 Indicator/controller for free chlorine, chlorine dioxide, hydrogen peroxide, peracetic acid or ozone (with flow monitoring)

Swimming pool water monitoring
Layout
Electrical connection
Ta sk
Data sheet (1) Transmitter/controller type 202553 202553 (2) Conductivity sensor for free chlorine on main board 202630 (3) Flow monitor on main board 202630 (3) Connecting cables 202950
See section 5 "Installation", page 12.
Measuring range: 0 to 2.00 ppm Output signal: 4 to 20 mA Flow monitoring: Flow monitor Limit monitoring: Limit value function Limit value 1: 0.60 ppm
41
Page 42
7 Commissioning
Basic setting
Start the basic settingSee section 6.7.3 "Basic setting", page 30 Diagrammatic overview, See section "Basic setting wizard", page 31.
Operating mode Linear Display format XX.xx Unit ppm Signal type 4 to 20 mA Scaling start 0.00 ppm Scaling end 2.00 ppm Supply frequency 50 Hz
Reinitialize device Yes
Analog output
Binary inputs
Controller settings
Administrator level/Password/Parameter level/Analog outputs/Analog output 1
Signal source Main variable Signal type 4 to 20 mA Start of scaling 0.00 ppm End of scaling 2.00 ppm
Administrator level/Password/Parameter level/Binary inputs/Binary input 1
Function HOLD mode inverse
See section 10.6.2 "Simple limit monitoring (max. function)", page 71.
42
Page 43
7 Commissioning
(1)
(2)
(3)

7.2.3 Indicator/controller for level or liquid level measurement

Tank monitoring.
Layout
(1) Transmitter/controller type 202553 202553 (2) Level measurement probe 404390 (3) Connecting cable (part of the level measurement probe)
Electrical connection
See section 5 "Installation", page 12.
Ta sk
Measuring range: 0 to 100 % (0 to 1 bar) Output signal: 4 to 20 mA Limit monitoring : Limit value function Limit value 1: 10 % Limit value 2: 80 %
Data sheet
43
Page 44
7 Commissioning
Basic setting
Start the basic settingSee section 6.7.3 "Basic setting", page 30 Diagrammatic overview, See section "Basic setting wizard", page 31.
Operating mode Linear Display format XXXX Unit % Signal type 4 to 20 mA Scaling start 0 % Scaling end 100 % Supply frequency 50 Hz
Reinitialize device Yes
Analog output
Controller settings
Administrator level/Password/Parameter level/Analog outputs/Analog output 1
Signal source Main variable Signal type 4 to 20 mA Start of scaling 0 % End of scaling 100 %
See section 10.6.3 "Controller with limit value function", page 72.
44
Page 45

8.1 Notes

During calibration, relays and analog output signals adopt their configured states!
The sensors connected to the instrument should be cleaned and the instrument itself calibrated, at regular intervals (subject to the sample medium).
Every successfully completed calibration is documented in the calibration logbook, See section 9 "Calibration logbook", page 66.

8.2 General information

The electrical properties of all sensors vary slightly from instance to instance and also change during operation (due to deposits or wear, etc.). This changes the output signal of the sensor.

8 Calibration

8.2.1 Requirements

- The instrument must be supplied with voltage, See section 5 "Installation",
page 12 ff.
- A sensor must be connected to the transmitter.
For a configuration example See section 7.2.1 "Indicator/controller for pH (temperature compensated)", page 39.
A sensor (via a transmitter) can be
- connected directly to the main input or
- to the optional board "Analog input (universal)"
- The appropriate operating mode (Linear, Temperature, Conductivity,
Concentration, etc.) must be configured in the basic setting.
- The instrument is in Measuring mode.
45
Page 46
8 Calibration
PGM
PGM
PGM

8.2.2 Ways to start the calibration

Select the input to which the sensor (transmitter) is connected.
If Calibration level is not released
- Press the key for longer than 3 seconds/ADMINISTR. LEVEL/
PASSWORD/CALIBR. LEVEL/MAIN INPUT or ANALOG INPUT.
If Calibration level is released
- Press the and keys simultaneously/MAIN INPUT or
ANALOG INPUT.
If Calibration level is released
- Press the key for longer than 3 seconds/CALIBR. LEVEL/
MAIN INPUT or ANALOG INPUT.

8.2.3 Operating modes

The operating mode selection depends on which sensor (transmitter) is connected.
Linear
For example sensor for free chlorine, redo, pressure, liquid level or humidity
Temperature (for main board only)
For example Pt100
pH measurement
For example pH sensor
Conductivity
For example sensor for conductivity
Concentration (for main board only)
For example sensor for conductivity
Customer specs.
For sensors with non-linear characteristics. Up to 20 interpolation points can be defined in an instrument table. This allows for an excellent approximation of a non-linear characteristic.
Stroke retransmission (for main board only)
For example valve settings
Chlorine, pH and temperature-compensated
Combination of chlorine sensor and pH sensor and temperature sensor. The measured value for chlorine often depends to a great extent on the pH value of the solution.
46
Page 47
8 Calibration
The chlorine measurement is compensated depending on the pH value in this operating mode. The pH measurement is temperature-compensated
Off (for main board only)
If only the optional boards will be used.

8.2.4 Calibration options

Different calibration options are available depending on the operating mode.
Operating mode Calibration options Page
1-point 2-point Limit point Rel.
cell const.
Linear X X X - - 48 Temperature No calibration required pH XX---52 Conductivity - - - X X 56 Concentration - - - X 62 Customer specs. Due to the table with interpolation points, no calibration is required Stroke
retransmission Chlorine,
pH-compensated
--X--64
No calibration required
Te mp .
coeffic.
-With one-point (offset) calibration, the zero point of the sensor is
calibrated.
-With two-point calibration, the zero point and slope of the sensor are
calibrated. This is the recommended calibration for most sensors.
-With one-point final value calibration, the slope of the sensor is
calibrated. This is the recommended calibration for chlorine sensors, for example.
- Calibration of relative cell constant
With conductivity sensors only.
- Calibration of the temperature coefficient
With conductivity sensors only.
47
Page 48
8 Calibration
PGM
PGM
PGM
PGM
EXIT

8.3 Linear operating mode

8.3.1 1-point calibration

This example is based on a liquid level measurement (as a %). The input signal is provided by a pressure transmitter.
- The transmitter is in "Measuring mode".
Now bring the system to a defined state (e.g. when measuring liquid level,
empty the container).
Start the calibration, See "Ways to start the calibration" page 46.Select the zero point calibration with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the required value (usually 0 %) with the and
keys; then press to continue.
The zero point determined by the instrument is displayed. Use the key to accept the value or
the key to reject it.
48
Page 49
The instrument returns to Measuring mode.
Anzeige =
Eingangswert
Steilheit
+Nullpunkt
PGM
PGM
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

8.3.2 Two-point calibration

The values determined during calibration (zero point and slope) work out as follows:
This example is based on a liquid level measurement. The input signal is provided by a pressure transmitter.
8 Calibration
- The transmitter is in "Measuring mode".
Now bring the system to a defined state (e.g. when measuring liquid level,
empty the container).
Start the calibration, See "Ways to start the calibration" page 46.Select the 2-point calibration with the key.
Wait until the display value has stabilized; then press
to continue.
Set the displayed value to the required value (usually 0) with the and
49
Page 50
8 Calibration
PGM
PGM
PGM
PGM
EXIT
Now bring the system to a second defined state (e.g. when measuring
Set the displayed value to "Maximum" (usually 100 %) with the and
keys; then press to continue.
liquid level, container full). Wait until the display value has stabilized; then press to continue
keys; then press to continue.
The zero point and slope determined by the instrument are displayed.
Use the key to accept the calibrated values or
The instrument returns to Measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
reject them with the key.
50
Page 51

8.3.3 Calibration limit point

PGM
PGM
PGM
This example is based on a measurement of free chlorine. The input signal is provided by a corresponding transmitter.
- The transmitter is in "Measuring mode".
-
The process must now be brought to the state that is as relevant as
possible to the final value (e.g. when measuring chlorine, the required concentration).
Start the calibration, See "Ways to start the calibration" page 46.Select the limit point calibration with the key.
8 Calibration
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the measured reference value with the or keys; then press to continue.
The slope determined by the instrument is displayed.
51
Page 52
8 Calibration
PGM
EXIT
Use the key to accept the value or the key to reject it.
The instrument returns to Measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

8.4 pH operating mode

8.4.1 Zero point (1-point) calibration

This example is based on a glass combination electrode with a connected two-wire transmitter.
- The transmitter is in "Measuring mode".
Perform calibration as follows.
Zero point (1-point) calibration
Make preparations, See section 8.2 "General information", page 45 .Start calibration, See section 8.2.2 "Ways to start the calibration", page 46.Select zero point calibration.
52
Page 53
8 Calibration
PGM
PGM
PGM
PGM
PGM
EXIT
Immerse the combination electrode in a buffer solution with a known pH
value.
Start the zero point calibration with the key.
Now the source of temperature acquisition can be selected (manually, or using the temperature input of the PSU board, or the temperature input via the optional board). This source will be active for the duration of the calibration.
An example follows: manual temperature entry.
To enter the temperature manually, use the and keys to set the
calibration solution temperature and confirm your entry with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the buffer solution value with the or keys;
then press to continue.
Use the key to accept the zero point or the key to reject it.
53
Page 54
8 Calibration
PGM
The instrument returns to Measuring mode.

8.4.2 2-point calibration

This example is based on a glass combination electrode with a connected two-wire transmitter.
- The transmitter is in "Measuring mode".
2-point calibration
Perform calibration as follows:
The buffer solutions (reference solutions) used for calibration must differ by at least 2 pH! During the calibration, the temperature of the two buffer solutions must be identical and remain constant!
Make preparations, See section 8.2 "General information", page 45 .Start calibration, See section 8.2.2 "Ways to start the calibration", page 46.Select 2-point calibration.
Immerse the combination electrode in the first buffer solution with the
known pH value.
Start the two-point calibration with the key.
54
Page 55
8 Calibration
PGM
PGM
PGM
PGM
Now the source of temperature acquisition can be selected (manually, or using the temperature input of the PSU board, or the temperature input via the optional board). This source will be active for the duration of the calibration.
An example follows: manual temperature entry.
To enter the temperature manually, use the and keys to set the
calibration solution temperature and confirm your entry with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the value of the first buffer solution with the
and keys; then press to continue.
Rinse and dry the pH combination electrode.Immerse the pH combination electrode in the second buffer solution.Wait until the display value has stabilized; then press to continue.
Set the displayed value to the second buffer solution value with the or
55
Page 56
8 Calibration
PGM
PGM
EXIT
PGM
The zero point and slope determined by the instrument are displayed. Use the key to accept the calibrated values or
The instrument returns to Measuring mode.
keys; then press to continue.
reject them with the key.

8.5 Conductivity operating mode

8.5.1 Calibration of the relative cell constant

This example is based on a conductivity sensor with a connected two-wire transmitter.
- The transmitter is in "Measuring mode".
Immerse the conductivity sensor in a reference solution with a known
conductivity.
Start the calibration, See "Ways to start the calibration" page 46.Select REL. CELL CONST.Press the key.
56
Page 57
8 Calibration
PGM
PGM
PGM
EXIT
When the measured value is stable, press the key
The measured conductivity value flashes on the display.
Use the or keys to set the value to the actual conductivity.Press the key;
Use the key to accept the temperature coefficient or
The current measurement value and the temperature are displayed.
Calibration is complete
the relative cell constant determined by the instrument is displayed (as a %).
the key to reject it.
After rinsing, the sensor can again be used to take measurements.
57
Page 58
8 Calibration
(1)
(2)

8.5.2 Calibration of the temperature coefficient

Linear temperature coefficient
This example is based on a conductivity sensor with a connected two-wire transmitter.
- The transmitter is in "Measuring mode".
Immerse the conductivity sensor in the sample medium. Start the calibration, See "Ways to start the calibration" page 46.
Select "TEMP.COEFF. LIN.".
The current sensor temperature flashes in the display (1).
The working temperature must be at least 5 °C above or below the reference temperature (25.0 °C).
Enter the required working temperature and confirm your entry.
The LC display now shows the selected working temperature (flashing) (2).
58
Page 59
8 Calibration
PGM
PGM
PGM
EXIT
Press the key.
The conductivity (399 µS/cm) at the current temperature (24.3 °C) now appears on the right of the LC display. The temperatures T1 (25 °C) and T2 (70.0 °C) that have yet to be triggered are shown on the left.
Press the key.Heat the sample medium until the working temperature is reached.
During calibration, the rate of temperature change in the measurement solution must not exceed 10 °C/min.
Calibration is also possible in the cooling process (with a falling temperature). It starts above the working temperature and ends below the working temperature.
As soon as the temperature of the sample medium exceeds T1 (25 °C), this is hidden on the display. The uncompensated conductivity at the current temperature is displayed on the right.
If the temperature of the medium exceeded T2 (73.0 °C), the instrument determines the temperature coefficient.
The LC display now shows the determined temperature coefficient as %/K.
Use the key to accept the temperature coefficient or
the key to reject it.
59
Page 60
8 Calibration
PGM
(1)
(2)
The transmitter is in "Measuring mode" and displays the compensated conductivity of the solution.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
With non-linear temperature coefficient (TEMP. COEF. CURVE)
This example is based on a conductivity sensor with a connected two-wire transmitter.
The non-linear temperature coefficient can only be calibrated with a rising temperature! The start temperature must be below the configured reference temperature (usually 25 °C)!
The "TEMP.COEF. CURVE" menu item is only displayed if a temperature sensor is connected and "TEMP.COEFF. CURVE" is configured as the type of temperature compensation.
- The transmitter is in "Measuring mode".
Immerse the conductivity sensor in the sample medium. Start the calibration, See "Ways to start the calibration" page 46.
Select "TEMP.COEFF. CURVE " and press the key.
Enter the required start temperature (1) for the temp. coef. curve.
Enter the required end temperature (2) for the temp. coef. curve.
60
Page 61
8 Calibration
(3)
(4)
(5)
PGM
EXIT
PGM
EXIT
Heat the sample medium continuously
(3) the current uncompensated conductivity (4) the current temperature of the sample medium (5) the first target temperature
During calibration, the rate of temperature change in the measurement solution must not exceed 10 °C/min.
During the calibration process, the instrument displays values for the following five temperature interpolation points.
The end temperature has been reached
Use the key to accept the temperature coefficients or the key to reject the calibration result.
The LC display now shows the determined temperature coefficients as %/K. Use the key to accept the temperature coefficients or
the key to reject the values.
The transmitter is in "Measuring mode" and displays the compensated conductivity of the solution.
61
Page 62
8 Calibration
PGM
PGM
Calibration is complete
After rinsing, the sensor can again be used to take measurements.

8.6 Concentration operating mode

8.6.1 Calibration of the relative cell constant

This example is based on a conductivity sensor with a connected two-wire transmitter.
The conductivity of a caustic solution is converted into a concentration value [%] by the instrument.
- The transmitter is in "Measuring mode".
Immerse the conductivity sensor in a sample medium with a known
conductivity.
Start the calibration, See "Ways to start the calibration" page 46.Press the key.
The measured conductivity value is displayed.
Wait until the measurement value has stabilized.Press the key.
Use the and keys to set the value to the actual conductivity.
62
Page 63
8 Calibration
PGM
PGM
EXIT
Press the key;
the relative cell constant determined by the instrument is displayed (as a %).
Use the key to accept the relative cell constant or
the key to reject the values.
The transmitter is in "Measuring mode" and displays the compensated conductivity of the solution.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
63
Page 64
8 Calibration
PGM
PGM
PGM

8.7 Chlorine measurement operating mode, pH-compensated

8.7.1 Final value calibration

The pH signal and temperature signal are supplied via the main input, the chlorine signal (standard signal) via the optional input.
- The transmitter is in "Measuring mode".
-
Calibrate pH sensor
Perform calibration, See "pH operating mode" page 52.
Calibrate chlorine sensor
The process must now be brought to the state that is as relevant as
possible to the final value (e.g. when measuring chlorine, the required concentration).
Start the calibration, See "Ways to start the calibration" page 46.Select the limit point calibration with the key.
Wait until the display value has stabilized; then press to continue.
Set the displayed value to the measured reference value with the or keys; then press to continue.
64
Page 65
8 Calibration
PGM
EXIT
The slope determined by the instrument is displayed. Use the key to accept the value or the key to reject it.
The instrument returns to Measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
65
Page 66

9 Calibration logbook

PGM
PGM

9.1 General information

The characteristic data for the last 5 successful calibration processed are documented in the calibration logbook.
Calling up
The instrument is in Measuring mode. Press the key for longer than 3 seconds.
Select input
Briefly press the key.
Most recent successful calibration
The "time stamp" in the following screen printouts (top left, for example 11­06-06 12:02) only appears if optional slot 3 is fitted with the "Datalogger with interface RS485"!
Briefly press the key.
Next most recent successful calibration
Briefly press the key.
66
Page 67

10.1 General information

L
H
d
a
(1)
(2)
(3)
(4)
Apart from faulty installation, incorrect settings on the instrument may also affect the proper functioning of the subsequent process or lead to damage. You should therefore always provide safety equipment that is independent of the instrument and it should only be possible for qualified personnel to make settings.

10.2 Controller functions

"Software" control functions are assigned to "Hardware" outputs for this instrument.

10 Controller

1 Software controller for "simple" switching functions (e.g. alarm control) 2 Software controller for "higher order" switching functions (e.g. PID
controller) 3 "Switching" hardware output (e.g. relay) 3 "Continuous" hardware output (analog output)

10.2.1 Simple switching functions

Up to four switching functions can be set (limit value 1, 2, 3, 4) ADMINISTR.-LEVEL/PARAMETER LEVEL/LIMIT VALUE CONTR./ LIMIT VALUE x.

10.2.2 Higher order switching functions (PID)

Higher order switching functions are configured at the parameter level via the parameters of "Controller 1 or 2".
ADMINISTR.-LEVEL/PARAMETER LEVEL/CONTROLLER/CONTROLLER 1(2)/ CONFIGURATION/CONTROLLER TYPE/e.g. PULSE LENGTHS
67
Page 68
10 Controller
L
H
d
a
d
a
d
a
d
a
(1)
(2)

10.2.3 Typical operator level parameters

Binary outputs
Explanation
Signal source No signal No switching function desired Limit control
1 to 4 Alarm function (AF1)
Alarm function (AF2)
Alarm function (AF7)
Alarm function (AF8)
"Simple" switching functions
Controller 1(2) "Higher order" switching functions Limit value
Pulse width Pulse frequency Steady Modulating

10.3 Software controllers and outputs

Simple controller functions
1 Main board 2 Optional board L Simple controller H Higher order controller d Digital output a Analog output
68
Page 69
If "Simple controller functions" have been configured, only the digital outputs
(1)
(2)
L
H
d
a
d
a
d
a
d
a
can be controlled! The operator must configure which of the digital outputs will be controlled -
the main board or optional board 1, 2 or 3.
Higher order controller functions
10 Controller
1Main board 2 Optional board L Simple controller H Higher order controller d Digital output a Analog output
If "higher order controller functions" have been configured, both the digital outputs and the analog outputs can be controlled.
The operator must configure which of the outputs will be controlled - the main board or optional board 1, 2 or 3.
Additional explanations, See section 15.1 "Glossary", page 86.
69
Page 70
10 Controller

10.4 Configuration of higher order controllers

10.4.1 Structure

10.5 Parameter sets

Different process steps may require different controller settings. The instrument offers the option of creating two parameter sets and then switching between them by means of a binary input.
Defining a parameter set
ADMINISTR.-LEVEL/PARAMETER LEVEL/CONTROLLER 1(2)/ PARAMETER SET 1(2) See "controller" page 102.
Configuring parameter set switchover
ADMINISTR.-LEVEL/PARAMETER LEVEL/BINARY INPUTS/ BINARY INPUT 1(2)/PARAMET. SWITCHOVER
See "Binary inputs" page 102.
70
Page 71

10.6 Sample configurations

10.6.1 Simple limit monitoring (min. function)

Configuration
Limit monitoring Limit value 1
Signal source: Main value Switching function: Alarm function (AF8) Switching point: 7.50 pH Hysteresis: 0.30 pH
Configuration of binary output, e.g. relay)
Binary outputs Binary output 1
Signal source: Limit monitoring 1 At calibration: Standard operation
10 Controller
Error: Inactive HOLD mode: Frozen Turn-on delay: 0 seconds Turn-off delay: 0 seconds Wiper time: 0 seconds Manual mode: No simulation

10.6.2 Simple limit monitoring (max. function)

Configuration
Limit monitoring Limit value 1
Signal source: Main value Switching function: Alarm function (AF7) Switching point: 0.60 ppm Hysteresis: 0.05 ppm
Configuration of binary output, e.g. relay)
Binary outputs Binary output 1
Signal source: Limit monitoring 1 At calibration: Standard operation Error: Inactive
71
Page 72
10 Controller
HOLD mode: Frozen Turn-on delay: 0 seconds Turn-off delay: 0 seconds Wiper time: 0 seconds Manual mode: No simulation

10.6.3 Controller with limit value function

Configuration of software controllers
Controller 1 Configuration
Controller type: Pulse value Controller actual value: Main variable Stroke retransmission: No signal Additive disturbance: No signal Multiplicative disturbance: No signal Min./max. contact: Min. contact Inactive/active contact: Active contact HOLD mode 0 % HOLD output: 0 % Error: 0 % Alarm control: Off
Parameter set 1
Min. setpoint: As required Max. setpoint: As required Setpoint: 10% Hysteresis: As required On-delay: As required Delayed release: As required Alarm delay: As required 2
Controller 2 Configuration
Controller type: Pulse value Controller actual value: Main variable Stroke retransmission: No signal Additive disturbance: No signal
72
Multiplicative disturbance: No signal Min./max. contact: Max. contact Inactive/active contact: Active contact
Page 73
HOLD mode 0 % HOLD output: 0 % Error: 0 % Alarm control: Off
Parameter set 1
Min. setpoint: As required Max. setpoint: As required Setpoint: 80% Hysteresis: As required On-delay: As required Delayed release: As required Alarm delay: As required
Configuration of binary output, e.g. relay)
Binary outputs Binary output 1
10 Controller
Signal source: Controller 1 output 1
Binary output 2
Signal source: Controller 2 output 1
73
Page 74

11 Setup program

(1)
(2)
(3)

11.1 Configurable parameters

Both the setup program (Part no. 00560380) and the PC interface cable with USB/TTL converter (Part no. 00456352) are available as options and provide a convenient way to adapt the transmitter to meet requirements:
- Setting the measuring range.
- Setting the behavior of outputs when the measuring range is exceeded.
- Setting the functions of switching outputs K1 to K8.
- Setting the functions of the binary inputs.
- Setting a customized characteristic
-etc.
Data can only be transferred from or to the transmitter if it is supplied with voltage, See section 5 "Installation", page 12ff.
Connection
(1) JUMO dTRANS 02 AS (2) PC interface cable with USB/TTL converter, Part no.: 00456352 (3) PC or notebook
74
Page 75
11 Setup program
(2)
(1)

11.2 Documenting the instrument configuration

Start the setup programEstablish the connection to the instrument (1).
Read the instrument configuration (2).
75
Page 76
11 Setup program
(2)
(1)
(3)
(4)

11.3 Special features for "Datalogger"

Start the setup programEstablish the connection to the instrument (1).Read the instrument configuration (2).
Read data from datalogger (for example table view)
- Mark datalogger icon (3)
- Read values from the instrument (4)
76
Page 77
11 Setup program
Export data (for processing in an external program).
77
Page 78

12 Eliminating faults and malfunctions

Problem Possible cause Action
No measurement display or current output
Measurement display 0000 or current output 4 mA
Incorrect or fluctuating measurement display
There is no supply voltage Check the power supply
Sensor not immersed in medium; level in container too low
Flow-through fitting is blocked Clean the flow-through fitting Sensor faulty Replace the sensor Sensor faulty Replace the sensor Sensor positioning incorrect Choose another installation
Air bubbles Optimize assembly Measurement overrange
Measurement underrange
Main input: Measurement range "out of range"
Top up the container
location
Choose a suitable measuring range
Compensation range has been left
Measurement overrange
Measurement underrange
Temperature input: Measurement range "out of range"
Compensation range has been left
Temperature input: Measurement range "out of range"
Configuration change OK
Choose a suitable measuring range
Choose a suitable measuring range
78
Page 79
12 Eliminating faults and malfunctions
Configuration change OK
Inhibit via binary contact Check configuration and unlock if
necessary
Do not release If appropriate release in the
release level
Te st
Inhibit via binary contact Check configuration and unlock if
necessary
Cancel in basic setting OK
Instrument had no power supply for a very long time
Check hardware
Check fitting, adjust if necessary
Establish power supply Set the datalogger time
79
Page 80

13 Technical data

Inputs (main board)

Measuring range/control range Accuracy Effect of temperature

Standard signal 0(4) to 20 mA or 0 to 10 V 0.25 % of range 0.2 %/10 K

Secondary input

Temperature Pt100/1000 -50 to +250 °C Temperature NTC/PTC 0.1 to 30 kΩ
Entry via table with 20 value pairs
Resistance transmitter Minimum: 100 Ω
Maximum: 3 kΩ
a
Selectable in °F.
Resistance thermometer inputs (optional board)
a
0.25 % of range 0.2 %/10 K1.5 % of range 0.2 %/10 K
±5 Ω 0.1 %/10 K

Designation Connection type Measuring range Measuring accuracy Effect of ambient

3-wire/4-wire 2-wire
Pt100 DIN EN 60751 (factory-set)
Pt1000 DIN EN 60751 (factory-set)
Sensor lead resistance Maximum 30 Ω per line with 3- and 4-wire circuit Measurement current Approx. 250 µA Lead compensation Not required for 3- and 4-wire circuit. With a 2-wire circuit, lead resistance can be compensated
2-wire/3-wire/ 4-wire
2-wire/3-wire/ 4-wire
in the software by correcting the process value.
-200 to +850 °C 0.05 % 0.4 % 50 ppm/K
-200 to +850 °C 0.1 % 0.2 % 50 ppm/K
temperature
Standard signals inputs (optional board)

Designation Measuring range Measuring accuracy Effect of ambient

Voltage 0(2) to 10 V
0to1V Input resistance
Electrical current 0(4) to 20 mA,
voltage drop 1.5 V
Resistance transmitter Minimum: 100 Ω
Maximum: 4 kΩ
> 100 kΩ
E
0.05 % 100 ppm/K
0.05 % 100 ppm/K
±4 Ω 100 ppm/K
temperature
Temperature compensation

Type of compensation Range

Linear 0 to 8 %/K -10 to +160 °C ASTM D1125 - 95 (ultra-pure water) 0 to 100 °C Natural waters (ISO 7888) 0 to 36 °C

Reference temperature

Adjustable from 15 to 30 °C; preset to 25 °C (default)

a
Note the sensor operating temperature range!
a
Measuring circuit monitoring

Inputs Underrange/overrange Short circuit Broken lead

Temperature Yes Yes Yes Voltage 2 to 10 V
2to10V
Current 4 to 20 mA
0to20mA
Resistance transmitter No No Yes
Ye s Ye s
Ye s Ye s
Ye s No
Ye s No
Ye s No
Ye s No
80
Page 81
Binary input
13 Technical data

Activation Floating contact is open: function is not active

Function Key lock, manual mode, HOLD, HOLD inverse, alarm suppression, freeze measured value,

Floating contact is closed: function is active
level lock, reset partial quantity, reset total quantity, parameter set switchover
Controller

Controller type Limit comparators, limit controllers, pulse length controllers, pulse frequency

controllers, modulating controllers, continuous controllers

Controller structure P/PI/PD/PID

Outputs
Relay (changeover)
Contact rating Contact service life
Voltage supply for 2-wire transmitter
Voltage supply for inductive proximity switch
Relay (changeover)
Contact rating Contact service life
Relay SPST (normally open)
Contact rating Contact service life
Semiconductor relay
Contact rating Protective circuit
PhotoMOS® relays Optional board U ≤ AC/DC 50 V
Voltage
Output signals Load resistance
Accuracy Electrical current
Output signals Load resistance Accuracy
PSU board
5 A at AC 240 V resistive load 350,000 operations at nominal load/750,000 operations at 1 A
PSU board Electrically isolated, non-controlled
DC 17 V at 20 mA, open-circuit voltage approx. DC 25 V
Optional board DC 12 V; 10 mA
Optional board
8 A at AC 240 V resistive load 100,000 operations at nominal load/350,000 operations at 3 A
Optional board
Optional board
Optional board
Optional board
3 A at AC 240 V resistive load 350,000 operations at nominal load/900,000 operations at 1 A
1 A at 240 V Varistor
I 200 mA
0to10V or 2to10V R
500 Ω
load
0.5 %
0to20mA or 4to20mA R
500 Ω
load
0.5%
Display
Type LC graphic display, blue with background lighting, 122 × 32 pixels
Electrical data
Voltage supply (switch-mode PSU) AC 110 to 240 V +10/-15 %; 48 to 63 Hz or
AC/DC 20 to 30 V; 48 to 63 Hz
Electrical safety To DIN EN 61010, Part 1
overvoltage category II, pollution degree 2 Power consumption Max. 14 VA (20 A fuse max.) Data backup EEPROM Electrical connection On the back via screw terminals,
Electromagnetic Compatibility (EMC)
Interference emission Interference immunity
conductor cross-section up to max. 2.5 mm
DIN EN 61326-1
Class A
To industrial requirements
2
81
Page 82
13 Technical data
Enclosure
Enclosure type Plastic enclosure for panel mounting to DIN IEC 61554 (indoor use) Depth behind panel 90 mm Ambient temperature
Storage temperature Climatic rating Rel. humidity 90 % annual mean, no condensation Site altitude Up to 2000 m above sea level Operating position Horizontal Enclosure protection
In the panel enclosure In the surface-mounted enclosure
Weight (fully fitted) About 380 g
Interface

Modbus

Interface type RS422/RS485 Protocol Modbus, Modbus Integer Baud rate 9600, 19200, 38400 Device address 0 to 255 Max. number of nodes 32

PROFIBUS-DP

Device address 0 to 255
-5 to +55 °C
-30 to +70 °C
To DIN EN 60529
Front IP65, rear IP20
IP65
Approvals/marks of conformity
Mark of conformity
c UL us Underwriters Laboratories E 201387 UL 61010-1
Testing laboratory Certificates/certification
numbers
Test basis valid for
CAN/CSA-C22.2 No. 61010-1
Type 202553/01...
82
Page 83

14 Retrofitting optional boards

Caution:
The instrument must be de-energized on the input and output sides! Optional boards must only be retrofitted by qualified specialists.
ESD:
Optional boards can be damaged be electrostatic discharge. You must therefore prevent electrostatic charges from accumulating during installation and removal. Optional boards should be retrofitted at a grounded workstation.
14.1 Identifying an optional board
The packaging of the optional board is identified by a sales number.
Optional board Code Part no. Board view
Analog input (universal) 1 00442785
Relay (1× changeover) 2 00442786
Relay (2× NO) This board must only be
inserted in optional slot 1 or 3!
Analog output 4 00442788
2 PhotoMOS
®
-Relais 5 00566677
3 00442787
83
Page 84
14 Retrofitting optional boards
Optional board Code Part no. Board view
Solid state relay 1 A 6 00442790
Supply voltage output DC ±5 V (e.g. for ISFET)
Supply voltage output DC 12 V (e.g. for inductive proximity switch)
Interface - RS422/485 This board must only be
inserted in optional slot 3!
Datalogger with interface RS422/485 and real-time clock
This board must only be inserted in optional slot 3!
7 00566681
8 00566682
10 00442782
11 00566678
PROFIBUS-DP interface This board must only be
inserted in optional slot 3!
Note:
The optional boards detected by the instrument are displayed in "Device information" (See section 6.5.11 "Device info", page 29).
12 00566679
84
Page 85
14 Retrofitting optional boards
1
2
3
14.2 Removing a plug-in module
(1) Squeeze the front panel together by the left and right sides and remove
the plug-in module.
14.3 Inserting a plug-in module
Caution:
No "3" relays (2× SPST/normally open) may be inserted in slot 2!
n
(1) Slot 1 for optional board (2) Slot 2 for optional board (3) Slot 3 for optional board
(1) Push the optional board into the slot until it locks in place.
(2) Push the device plug-in into the enclosure until it locks in place.
85
Page 86

15 Appendix

(3)
(1)
(2)

15.1 Glossary

Display of measured values STANDARD
The measurement value, measurement variable and temperature of the measuring material are shown in standard display.
Operating mode
(1)
Display bottom (temperature input)
(2)
Display top (analog input measurement value)
(3)
Display of measured values TENDENCY
The operator can quickly see the direction in which the measurement is changing.
86
Rising
Steady
Greatly Moderately Slightly Slightly Moderately Greatly
The measurement tendency (trend) is calculated over the last 10 measurement values. So with a sampling interval of 500 ms, the last 5 seconds are considered.
Falling
Page 87
Display of measured values BARGRAPH
PGM
PGM
PGM
EXIT
PGM
- Values of the main inputs, input options or math channels (signal source) can be represented as a variable bar (a bar graph).
Scaling the bar
Activate "BARGRAPH" as the display of measured values.Select "SCALE START" with .Confirm the selection with .Use and to enter the lower limit of the range to be displayed.Confirm the selection with .
15 Appendix
Select "SCALE END" with . Use or to enter the upper limit of the range to be displayed.Confirm the selection with .
To return to measuring mode: Press the key repeatedly or wait for a "timeout".
Display of measured values TREND CHART
Values of the main inputs, input options or math channels (signal source) can be represented as a graph. The current values appear to the right on the screen.
Scaling the display
Activate "TREND CHART" as the display of measured values.Select "SCALE START" with .Confirm the selection with .Use and to enter the lower limit of the range to be displayed.
87
Page 88
15 Appendix
PGM
PGM
EXIT
Confirm the selection with .Select "SCALE END" with . Use or to enter the upper limit of the range to be displayed.Confirm the selection with .
To return to measuring mode: Press the key repeatedly or wait for a "timeout".
Display of measured values LARGE DISPLAY
Values of the main inputs, input options or math channels (signal source) can be displayed in large format.
Display of measured values 3 MEAS. VALUES
Three values of the main inputs, input options or math channels (signal source) can be displayed simultaneously.
The position of the value to be displayed can be set to "Top", "Center" or "Bottom".
Relative cell constant
Mechanical or chemical effects can change the electrical properties of a conductivity sensor. This will result in a measurement error. This deviation (and thus the measurement error as well) can be compensated for by adjusting the relative cell constant in the transmitter. The relative cell constant defines the deviation of the actual cell constant of the conductivity sensor from its nominal value.
88
Page 89
Zero point (1-point) calibration
pH
mV
123 4 5 6 89 10 11 12 13 140
7
B
A
pH
mV
123 4 5 6 89 10 11 12 13 140
7
B
A
- With one-point offset calibration, the zero point of the pH combination electrode is calculated, See section "Zero point (1-point) calibration", page 52. Recommended only for special applications, such as ultra-pure water.
15 Appendix
2-point calibration
- With two-point calibration, the zero point and slope of the combination electrode are calibrated. This is the recommended calibration for most sensors.
Temperature compensation (conductivity or resistance)
The conductivity of a measurement solution is temperature-dependent (the conductivity of a solution rises as the temperature increases). The dependency of conductivity and temperature describes the temperature coefficient of the measurement solution. As conductivity is not always measured for the reference temperature, automatic temperature compensation is integrated in
89
Page 90
15 Appendix
g
Ref.
T =T
Start 1
Temperature
uncompensated
measurement value
g
2
T
2
T
Reference
T =T
End 6
T
5
T
3
T
4
1
2
3
4
5
this instrument. The transmitter uses the temperature coefficient to calculate the conductivity that would exist for a reference temperature from the current conductivity and the current temperature. This is then displayed. This process is called temperature compensation. Modern transmitters offer different ways to perform this temperature compensation.
- Linear compensation (constant temperature coefficient).
- Natural water (EN27888 or ISO 7888).
- ASTM1125-95.
This type of compensation can be applied to many kinds of normal water, with acceptable accuracy. The temperature coefficient used is then approx.
2.2 %/°C
In this case, so-called non-linear temperature compensation is used. According to the standard cited above, the relevant type of compensation can be applied to natural groundwater, spring water and surface water. The definition range for the water temperature is as follows: 0°C ≤ T < 36 °C Conductivity of the water is compensated in the range from 0 °C to 36 °C.
This type of temperature compensation is used in measurements of ultra­pure water. The highly non-linear nature of the temperature dependency for neutral, acidic and alkaline impurities is taken into consideration in accordance with the standard. The definition range for the water temperature is as follows: 0 °C < T < 100 °C. Conductivity of the water is compensated in the range from 0 °C to 100 °C.
Determining the temp. coef. curve
90
Page 91
Calculating a temperature coefficient
α = temperature coefficient (TC) γ = uncompensated measurement value
Temp. coef. curve
15 Appendix
Temperature compensation with the temp. coef. curve
The relevant temperature coefficient is determined from the temp. coef. curve by means of the current temperature of the medium.
Intermediate values, such as (α
) and (α4 at T4) are linearly approximated.
at T
3
As with linear temperature compensation, the compensated measurement value is calculated with the ascertained TC.
If the measured temperature is lower than the temperature at the beginning, the first TC is used for compensation. If the measured temperature is higher than the temperature at the end, the last TC is used for compensation.
at Tx) between the two ascertained values (α3
x
91
Page 92
15 Appendix
Temperature compensation (pH or ammonia)
The pH value of a measurement solution depends on the temperature. Since the pH value is not always measured at the reference temperature, the instrument is able to perform a temperature compensation.
The sensor signal for the ammonia measurement is temperature-dependent. The instrument can perform temperature compensation.
The redox potential of a measurement solution is not temperature-dependent! Temperature compensation is not required.
Concentration measurement
The instrument can calculate the concentration of different measuring materials from the current measurement values of uncompensated conductivity and temperature.
Choice of concentration calculations:
NaOH (caustic soda)
- Range 1: 0 to 12 % by weight
- Range 2: 25 to 50 % by weight
HNO
(nitric acid)
3
- Range 1: 0 to 25 % by weight
- Range 2: 36 to 82 % by weight
H
- Range 1: 0 to 28 % by weight
- Range 2: 36 to 85 % by weight
- Range 3: 92 to 99 % by weight
HCl (hydrochloric acid)
- Range 1: 0 to 18 % by weight
- Range 2: 22 to 44 % by weight
(sulfuric acid)
2SO4
92
Page 93
Limit value (alarm) function of the binary outputs
HySt
x
0
w
AF
1
HySt
0
w
AF
x
1
HySt
x
w
0
1
HySt
w
x
0
1
Process value X
Setpoint W
Proportional band X
P
100%
50%
0%
Output level y
Switching period
10%
90%
90%
10%
t
On
50%
50%
t
Off
10%
90%
X
P
X - W
0
1
AF1
AF2
15 Appendix
AF7
AF8
Pulse length controller (output active with x > w and P control structure)
If actual value x exceeds setpoint W, the P controller will control in proportion
93
Page 94
15 Appendix
100%
50%
0%
No pulses
50% of pulse frequency
Maximum pulse frequency
Setpoint W
Proportional band X
P
X
P
X - W
0
1
Output level y
Process value X
to the control deviation. When the proportional range is exceeded, the controller operates with an output level of 100 % (100 % clock ratio).
Pulse frequency controller (output active with x > w and P control structure)
If actual value x exceeds setpoint W, the P controller will control in proportion to the control deviation. When the proportional range is exceeded, the controller operates with an output level of 100 % (maximum switching frequency).
Special controller functions: Separate controllers
This function is normally deactivated (factory setting or select "No"). In the deactivated state, the software prevents the two controller outputs from
being able to work "against each other". So, for example, it is not possible to dose acid and lye at the same time.
If the controllers are separate ("Yes" selection), each controller can be freely configured.
Switch-off of the I-component
This function is normally deactivated (factory setting or select "No"). In the deactivated state, the controller works in accordance with general
controller theory. When I-component switch-off is activated ("Yes" selection), the part of the
output level that can be traced back to the I-component is set to zero when the setpoint is reached.
This can be useful with mutual neutralization (acid and lye dosing both possible) in one treatment tank.
Calibration timer
The calibration timer indicates (on request) a required routine calibration. The calibration timer is activated by entering the number of days that must expire before there is a scheduled re-calibration (specified by the system or the operator).
Wash timer
94
The wash timer can be used to implement automated sensor cleaning. To do this, the function is assigned to a switching output.
Page 95
The cycle time (cleaning interval) can be adjusted in the range from 0.0 to
240.0 hours.
A cycle time of "0.0" means the wash timer is deactivated. The wash time (cleaning duration) is adjustable from 1 to 1800 seconds. During the wash time the controller goes into the HOLD state, which is
maintained for 10 seconds after completion of the wash time. A sensor calibration within the cycle time restarts the wash timer.
USP contact (for ultra-pure water)
The USP contact makes it possible to monitor the quality of ultra-pure water according to the requirements of USP <645>. USP <645> contains a table that assigns a limit value for conductivity depending on the temperature. If the conductivity stays below this limit value, the ultra-pure water meets the requirements of USP <645>.
If the conductivity of the water is greater than what is specified in the USP table for a given temperature, the USP contact switches the instrument.
Limit values are defined in levels. For example, a value of 5 °C is used at 8 °C.
Note:
During monitoring, temperature compensation must be turned off (temperature coefficient = 0)! To do this, select Administrator Level/Basic Setting/Temperature Compensation/None.
15 Appendix
Excerpt from USP <645>
Te mp er at ur e
°C
0 0.6 55 2.1
5 0.8 60 2.2 10 0.9 65 2.4 15 1.0 70 2.5 20 1.1 75 2.7 25 1.3 80 2.7 30 1.4 85 2.7 35 1.5 90 2.7 40 1.7 95 2.9 45 1.8 100 3.1 50 1.9
USP warning alarm
µS/cm (uncompensated)
If the conductivity is exceeded at the relevant temperature, the configured contact switches.
The USP warning alarm switches before the water quality reaches the set limit value. This parameter (0 to 100) is used to set the distance as a percentage (relative to the active limit value) to be maintained from the USP limit.
Max. conductivity
Te mp er at ur e
°C
Max. conductivity
µS/cm (uncompensated)
95
Page 96
15 Appendix
Ultra-pure water per Ph. Eur.
The limit comparators of the instrument switch, depending on the corresponding configuration, according to the limit valued of the European Pharmacopeia (Ph. Eur.) for purified water.
Ph. Eur. warning alarm
The Ph. Eur. warning alarm switches before the water quality reaches the set limit value. This parameter (0 to 100) is used to set the distance as a percentage (relative to the active limit value) to be maintained from the USP limit.
Temperature
°C
00.6 10 0.9 15 1.0 20 1.1 25 1.3 30 1.4 35 1.5 40 1.7 45 1.8 50 1.9
Max. conductivity
µS/cm
TDS
Display/control with the unit ppm. The specific TDS factor can also be entered in this mode. TDS (Total Dissolved Solids, also commonly referred to in Germany as
filtrate dry residue (Filtrattrockenrückstand). This value is important in areas such as groundwater analysis and power plants. The value is also used in evaluating drinking water quality (for example in the USA, Arab and Asian countries). Various organizations have published limit values on this topic.
-WHO (World Health Organization) <1000mg/l
-USEPA (United States Environmental Protection Agency) <500mg/l Standardized determination is performed gravimetrically, i.e.:
-Filter sample
- Evaporate filtrate
-Weigh residue A conductivity measurement is used for the online measurement. A single time
is sufficient to determine the conversion factor. It corresponds to the ratio of the conductivity value of the water to the value of the gravimetrically determined filtrate dry residue (TDS). The factor moves within the range from
0.55 to 1.0. A typical value for drinking water is about 0.67. With modern instruments, this factor can be entered individually to achieve the
most accurate measurement possible.
96
Page 97
Customer specs. table
In this mode, the input value can be displayed based on a table (max. 20 value pairs). This function is used to display and linearize non-linear input variables. Values can only be entered in the table using the optional setup program.
Cust. specs. characteristic
In this mode, the instrument can model a monotonically increasing input variable to any output value.
15 Appendix
The optional setup program is used to enter the requisite value table.
97
Page 98
15 Appendix
Min./max. value memory
This storage records the minimum and maximum input quantities that have occurred. This information can be used, for example, to assess whether the design of the connected sensor is suitable for the values that actually occur.
The max./min. value memory can be reset, See section 6.7.6 "Delete min/max values", page 33:
Datalogger
Recording duration = about 10 hours with a storage interval of 1 second Recording duration = about 150 days with a storage interval of 300 seconds
Parameter set switchover
In some processes (different process steps) is is advantageous to have two complete parameter sets available.
Define the parameter sets See section 10.5 "Parameter sets", page 70. The predefined parameter sets are activated by a binary input.
98
Page 99

15.2 Parameters of the User level

When there are numerous instrument parameters to configure, it is advisable to make a note in the table below of all the parameters to be changed and to work through these parameters in the given order.
The following list shows the maximum number of parameters that can be modified.
Some of these parameters will not be visible (and therefore not editable) for your particular instrument, depending on the configuration.
15 Appendix
Parameter Selection/value range
Factory setting
Conductivity input
Cell constant 0.01/0.1/0.5/1.0/3.0/10.0 Relative cell constant
and Relative cell constant MB 2
Offset and offset MB 2
Temperature compensation and temperature compensation MB 2
Temperature compensation source
Temperature coefficient and temperature coefficient MB 2
Reference temperature 15.0 to 25.0 to 35.0 °C Pollution recognition Off
Broken sensor detection Off
Filter time constant 0.0 to 2.0 to 25.0 seconds Calibration interval 0 to 99 days (0 = timer not active)
20.0 to 100.0 to 500.0
-20.00 to 0.00 to 20.00 % of the display range
None
Linear
Natural waters ASTM 1125 neutral ASTM 1125 acidic ASTM 1125 alkaline
Temperature input
Option input 1 Option input 2 Option input 3 Manual temperature input
0.00 to 2.20 to 8.00 %/K
On
On
New setting
99
Page 100
15 Appendix
Parameter Selection/value range
Factory setting
Differential measurement Off
Main input - (minus) Option input 1 Main input - (minus) Option input 2 Main input - (minus) Option input 3 Option input 1 - (minus) Main input Option input 2 - (minus) Main input Option input 3 - (minus) Main input
Supply frequency 50 Hz
60 Hz
Temperature input
Temperature sensor No sensor
Pt100
Pt1000 Cust. specs. 0to20mA 4to20mA 0to10V 2to10V Resistance transmitter
Unit °C/°F
% Without unit Cust. specs.
Scaling start -100.0 to 0.0 to 499.9 °C Scaling end -99.9 to 100.0 to 500.0 °C Filter time constant 0.0 to- 2.0 to 25.0 seconds Manual temperature -99.9 to 25.0 to +99.9 °C Offset -99.9 to 0.0 to +99.9 °C
Optional inputs Analog inputs 1 to 3
Operating mode Off
Linear Te mp er at ur e pH measurement Conductivity Concentration Cust. specs. Stroke feedback Chlorine, pH-compensated
Signal type 0to20mA
4to20mA 0to10V 2to10V 0to1V Pt100 Pt1000 Cust. specs.
New setting
100
Loading...