11.1Technical data ..........................................................................................49
1.1Warning signs
Danger
This symbol is used when there may be danger to personnel if the
instructions are ignored or not followed correctly!
Caution
This symbol is used when there may be damage to equipment or data if the
instructions are ignored or not followed correctly!
Caution
This symbol is used where special care is required when handling components
liable to damage through electrostatic discharge.
1.2Note signs
Note
This symbol is used to draw your special attention to a remark.
1 Typographical conventions
1
abc
✱Action instruction
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes
consist of two parts:
Marking in the text and the footnote text.
The markers in the text are arranged as sequential superscript numbers.
This symbol indicates that an action to be performed is described.
The individual steps are marked by this asterisk.
Example:
✱ Loosen Phillips-head screws.
5
2 Description
2.1General
❏ Depending on its design, the instrument measures relative or absolute
(1)
pressure in
liquid and gaseous media.
❏ The pressure is displayed digitally.
❏ Depending on the design, the following outputs are available:
- 1 PNP switching output
- 2 PNP switching outputs
- 1 PNP switching output + 1 analog output 4 ... 20 mA
- 1 PNP switching output + 1 analog output 0 ... 20 mA
- 1 PNP switching output + 1 analog output 0 ... 10 V
❏ The instrument is also available in a design for use at elevated medium
temperatures.
❏ The instrument can be adjusted directly on site or can be configured via PC
with a setup program.
The protection type specified for the device (See section 11.1 "Technical
data", page 49) can only be achieved with the control opening (1) closed.
1
1
1
1
The output is freely configurable.
6
3.1Nameplate
on the
pressure switch
(1)
3 Identifying the instrument version
Made in Germany
(2)
www.jumo.net
DELOS SI
405052/000
TN 12345678
(3)
(5)
(1)
Ty pe
(2)
Part number
(3)
Measuring range
(4)
Power supply voltage
3.2Block diagram
Pressure
(4)
(6)
0...250 mbar
DC 12...30 V
4...20 mA/OC: 250 mA
F.-Nr.: 001234567801008350123
(5)
(6)
(7)
}
(7)
Output signal
Manufacturing number
Date of manufacture
(year and calendar week)
Interface
Setup program
Rotary encoder
To operate: turn and press
Power supply
12 30 V DC?
1
optional
Display
2 x 4 characters, 2 colors
Switching output 1
Open collector
Type 405052
Analog output
0 (4) 20 mA oder 0 10 V??
Switching output 2
Open collector
(1)
(1)
7
3 Identifying the instrument version
3.3Type description
(1) Basic type
405052 JUMO DELOS SI
Electronic Pressure Switch with Display
(2) Basic type extension
000None
004For elevated medium temperatures up to 200°C
999Special design
(3) Nominal (rated) measuring range
Measuring range - pressure
4520 ... 400 mbar relative pressure
4540 ... 1 bar relative pressure
4570 ... 4 bar relative pressure
4590 ... 10 bar relative pressure
4610 ... 25 bar relative pressure
4630 ... 60 bar relative pressure
Measuring range - vacuum
447-400 ... 400 mbar relative pressure
449-1 ...1 bar relative pressure
481-1 ... 3 bar relative pressure
483-1 ... 9 bar relative pressure
485-1 ... 24 bar relative pressure
1
Measuring range absolute pressure
4860 ... 400 mbar absolute pressure
4880 ... 1 bar absolute pressure
4910 ... 4 bar absolute pressure
4930 ... 10 bar absolute pressure
4950 ... 25 bar absolute pressure
5060 ... 60 bar absolute pressure
(4) Output
4701 x PNP switching output
4712 x PNP switching output
4751 x PNP switching output + analog output 4 ... 20 mA,
three wires
2
4761 x PNP switching output + analog output 0 ... 20 mA,
three wires
2
4771 x PNP switching output + analog output 0 ... 10 V,
three wires
2
1
Front-flush only for process connection.
2
Factory setting - the output is freely configurable.
8
3 Identifying the instrument version
(5) Process connection
Not front-flush
504G 1/2
5111/4-18 NPT
521G 1/4 to DIN 3852 T11
523G 1/2 to DIN 3852 T11
998Suitable for connecting to pressure measuring instrument
Front-flush
571G 3/4
575G 3/4 seal positioned in front
576G1 seal positioned in front
603Tapered adapter with groove nut, to DIN11 851, DN20
604Tapered adapter with groove nut, to DIN11 851, DN25
606Tapered adapter with groove nut, to DIN11 851, DN40
607Tapered adapter with groove nut, to DIN11 851, DN50
612Clamp to DIN 32 676, DN20
613Clamp to DIN 32 676, DN25
616Clamp to DIN 32 676, DN50
623Small flange connection
4
4
4
4
652Tank connection with groove union nut4, DN25
997JUMO PEKA (EHEDG-certified)
3
998Suitable for connecting to pressure measuring instrument
4
4
4
4
(6) Process connection material
20Stainless steel 316L
(7) Electrical connection
36Round plug M 12 x 1
(8) Measuring system filling medium
01FDA-compliant oil
99Special filling medium
(9) Extra codes
000None
591Throttle in pressure channel
624Free of oil and grease
691Cast version
(1)(2)(3)(4)(5)(6)(7)(8)(9)
Order code/------/
Sample order405052/000-459-471-504-20-36-1/000
3
For matching process connection adapter see data sheet 40.9711
4
For measuring ranges up to 25 bar only
9
3 Identifying the instrument version
3.4Accessories
Sales No.Designation
40/004045854-pin cable connector (straight) M 12 x 1 with 2-m PVC cable
40/004093344-pin angle box M 12 x1 with 2-m PVC cable
40/00522384Setup program
40/00507861Connecting cable (only required for programming with the
setup program)
70/00456352PC interface line including USB/TTL converter and
adapter (USB connecting cable)
1
1
Please order a connecting cable as well.
10
4.1Installation instructions
The electrical connection must only be performed by
qualified personnel!
❏ The load circuits must be fused for the maximum load currents in each case to
prevent the instrument from being destroyed.
❏ Electromagnetic compatibility meets the requirements of EN 61326,
❏ No other consumers can be connected to the power supply of the instrument.
❏ The instrument is not suitable for installation in areas with an explosion hazard.
❏ Apart from faulty installation, incorrect settings on the instrument may also
affect the proper functioning of the subsequent process or lead to damage.
You should therefore always provide safety equipment that is independent of
the instrument and it should only be possible for qualified personnel to make
settings.
4 Electrical connection
4.2Color assignment of M12 x 1 round plug
The following color assignment applies only to A-coded standard cables!
2
3
5
1
4
1Brown
2White
3Blue
4Black
5Gray
11
4 Electrical connection
4.3Terminal assignment for output 470
One PNP switching output
2
3
5
1
4
K1
Power supply voltage
1 L+12 ... 30 V DC
3 L-GND
Output
4 K1Highside Open Collector
2nc
5Interface
L+
L-
maximum 0.25 A
4.4Terminal assignment for output 471
Two PNP switching outputs
2
3
5
1
4
K1K2
Power supply voltage
1 L+12 ... 30 V DC
3 L-GND
Output
4 K1Highside Open Collector
2 K2
5Interface
L+
L-
maximum 0.25 A
12
4 Electrical connection
4.5Terminal assignment for outputs 475, 476, and 477
One PNP switching output + one analog output
2
3
5
1
4
K1
Power supply voltage
1 L+12 ... 30 V DC
3 L-GND
Output
4 K1Highside Open Collector
2 Analog0(4)...20 mA / 0...10 V
5Interface
L+
L-
maximum 0.25 A
13
5 Mounting
5.1General information
The compatibility of the instrument with the measuring medium must be
tested,
See section 11.1 "Technical data", page 49.
Mounting
location
Installation
position
- Find a location that ensures easy accessibility for later operation.
- The fastening must be secure and must ensure low vibration for the
instrument.
- Avoid direct sunlight!
- Permissible ambient temperature at the installation location See section
11.1 "Technical data", page 49.
The instrument can be mounted in any position.
The "vertical" installation position is recommended, see illustration.
5.1.1Rotating the display
The display image can be rotated 180° in the software, See section "Display
position", page 25. This may make it easier to read when the instrument is
installed overhead, for example.
14
5.1.2Rotating the housing
The instrument housing can be rotated a maximum of 320° with the
combination tool (1).
5 Mounting
(1)
320° max.
15
5 Mounting
5.2Dimensions of electronic pressure switches
35
30
75
ø42
M12x1
40
Type 405052/000-...
Type 405052/004-...
The overall height is 40 mm greater for instruments with basic type extension
004 (for increased medium temperature up to 200°C). See drawing
16
5 Mounting
5.3Dimensions of process connections, not front-flush
504511
E
31
20
G1/2
G1/2
19
1.6
ø7
15
11.07
E
26
15
1/4-18 NPT
1/4-18 NPT
10.4
521523
E
E
G1/4
ø19
G1/4
A = Profile seal DN G3/4
B = Profile seal DN G1/2
E = SW 27
F = SW 32
12
1.5
12
1.6
A
23.5
G1/2
ø27
G1/2
B
14
1.6
14
17
5 Mounting
5.4Dimensions of process connections, front-flush
571575
1.6
G3/4
G3/4
576
16
16
F
A
32
ø24
G3/4
D
1.6
13
16
20.5
G
G3/4
ø24.5
1.6
F
A
34
20.5
9
ø30
G1
20.5
D
1.6
17.5
14
G1
30.5
A = profile seal DN G3/4
B = profile seal DN G1
C = O-ring 26.7 x 1.78
B
35
20.5
7
C
1.6
D = Drill out after tapping
F = SW 32
G = SW 41
18
L1
603 to 607
D2
D1
D3
D4
5 Mounting
Process
connection
DNø D1ø D2ø D3ø D4L1
6032036.530RD 44 x 1/65413
604254435RD 52 x 1/663
606405648RD 65 x 1/678
6075068.561RD 78 x 1/69216
612 to 616
3
15
D
1
D
2
15
Process
connection
612
DN
DIN 32676
20
15
25
613
32
40
616502“
DN
(inches)
1“
1.5“
Nominal Size
ISO 2852
12
12.7
17.2
21.3
25
33.7
38
40
51
ø D1ø D2
27.534
43.550.5
56.564
19
5 Mounting
623652
ø19
ø26
ø40
997
45
35
15
3
21
31
RD52 x 1/6
55
63
E
11
E = SW 27
28
Process connection 997 is EHEDG-certified
For detailed information about this process connection system, see
data sheet 40.9711.
20
6.1Controls
6 Operation
(1)
(2)
(1) Protective screw
(2) Hexagon socket
✱ Unscrew the protective screw (1).
✱ "Turn / push" the control element (2) with the enclosed combination tool
(or a 0.5x3 screwdriver).
6.2LC display
6.2.1Measurement mode (normal display)
Example:
The display is lit yellow.
21
6 Operation
6.2.2Settingmode
Example:
The display is lit red.
Operation
ContinuePress the combination tool less than 1 second ( < 1s)
Yes (accept)Press the combination tool less than 1 second ( < 1s)
No (Cancel)Press the combination tool more than 3 seconds ( > 3s)
TimeoutNo activity for more than 60 seconds ( >60 s)
6.3Levels
To return to measuring mode:
- "No (Cancel)" or
- Wait for timeout = no activity performed for 60 seconds.
Process value display
(Normal display)
< 1 s
Code = 0
Enter
code
Cancel
or timeout
< 1 s
> 3 s
22
< 1 s
Select
parameter
edit
parameter
Cancel
or
save
Cancel
or timeout
> 3 s
< 1 s
Cancel
or timeout
> 3 s
> 3 s
= "Press"
6.4Parameter
6.4.1Input
6 Operation
ParameterDisplaySetting range
Pressure unitbar
kPa
MPa
psi
mbar
Note:
The units kPa and mbar cannot be configured for
all measuring ranges.
Offset
(zero-point correction)
Damping
(filter time constant)
-20.00 ... 0.00 ... +20.00% of the measuring range
Note:
Automatic offset correction See section 7.5
"Setting the zero point (offset) (Off.P)", page 31.
0.00 ... 99.99 s
6.4.2Analog output
ParameterDisplaySetting range
Signal type
(for analog output)
4 ... 20 mA
0 ... 20 mA
0 ... 10 V
1
1
Scaling start
(for analog output)
Scaling end
(for analog output)
Signal for error
(for analog output)
1
The default setting is marked in bold.
0.00 ... 75.00% of nominal measuring range
25.00 ... 100% of nominal measuring range
3.4 mA or 22 mA for output signal 4...20 mA
0 mA or 22 mA for output signal 0...20 mA
0 V or 10.7 V for output signal 0...10 V
Note:
Depending on the configured output signal.
23
6 Operation
6.4.3Binary output 1
ParameterDisplaySetting range
Switching function
(for switching output only)
Switching point
(for switching output only)
Reset point
(for switching output only)
Hysteresis
(for switching output and
configured switching point
or reset point only)
Switching delay
(for switching output only)
Hysteresis, make contact
Hysteresis, break contact
Window, make contact
Window, break contact
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
Note:
Used only with window switching functions.
0.00 ... 99.99 s
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
1
6.4.4Binary output 2
ParameterDisplaySetting range
Switching function
(for second switching
output only)
Switching point
(for second switching
output only)
Reset point
(for switching output only)
Hysteresis
(for second switching
output and configured
switching point or reset
point only)
Hysteresis, make contact
Hysteresis, break contact
Window, make contact
Window, break contact
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
Note:
Used only with window switching functions.
1
1
The default setting is marked in bold.
24
6 Operation
ParameterDisplaySetting range
Switching delay
(for second switching
output only)
0.00 ... 99.99 s
See section 7.10 "Setting the switching function
(B.Fct)", page 37.
1
6.4.5Display and operation
ParameterDisplaySetting range
Display positionNormal (for normal operation)
Rotated (for overhead operation)
See section 7.15 "Setting the display alignment
(D.Dir)", page 41.
Unit of actual value display
(for analog output only)
Version DSoftware version of the operating device
Version SSoftware version of the signal stage
Pressure unit (see parameter "Uni.P")
Percentage of the scaled range
See section 7.16 "Setting the display unit (D.Uni)",
page 42.
See section 7.17 "Displaying the version of the
operating device software (SW.Di)", page 43.
1
Code
(can only be edited via
setup program)
See section 7.18 "Displaying the version of the
signal stage software (SW.Si)", page 43.
0000 ... 0072 .... 9999
See section 7.2 "Unlocking the instrument (code
entry)", page 27.
25
7 Commissioning
7.1Getting started
✱ Mounting the instrument, See section 5 "Mounting", page 14.
✱ Installing the instrument, See section 4 "Electrical connection", page 11.
✱ Unlocking the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ Selecting the unit of the measured value, See section 7.4 "Selecting the unit
of the measured value (Uni.P)", page 28.
✱ Adjusting the output signal, See section 7.7 "Setting the output signal
(S.TyP)", page 32.
✱ Adjusting the scaling of the output signal (restricting the measuring range),
See section 7.8 "Setting scaling", page 33.
This is a suggestion for configuring the instrument reliably in little
time.
By checking the setting options of this list before starting the
configuration, you can avoid timeouts during the configuration.
✱ Setting the switching function, See section 7.10 "Setting the switching
function (B.Fct)", page 37.
✱ Setting the switching point, See section 7.11 "Setting the switching point
(B.SP)", page 39.
26
7.2Unlocking the instrument (code entry)
The instrument is protected by a code to prevent unauthorized operation.
The code is set to 0072 in the factory. It can only be changed with the setup
program. If the code is set to 0000 with the setup program, the instrument is
unprotected.
Unlocking
(1)
7 Commissioning
(2)
✱ Unscrew the protective screw (1)
✱ Continue briefly pressing the combination tool (2) until the third "0" from the
left is flashing.
The color of the display also changes to "red."
✱ Turn the combination tool until "7" is displayed.
✱ Continue briefly pressing the combination tool until the fourth "0" from the
left is flashing.
✱ Turn the combination tool until "2" is displayed.
✱ Press the combination tool briefly - the instrument switches to the
27
7 Commissioning
parameter level.
After an incorrect code is entered:
7.3Cancel operation
✱ Press and hold the combination tool (2) longer than 3 seconds or
✱ wait for timeout (no activity for longer than 60 seconds).
7.4Selecting the unit of the measured value (Uni.P)
✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "Uni.P".
✱ "Press"
✱ The measured pressure is shown in millibar.
✱ "Press"
28
Flashing
Continuous
The measured pressure is shown in bar.
7 Commissioning
✱ "Rotate"
The measured pressure is shown in Kilopascal.
✱ "Rotate"
The measured pressure is shown in Megapascal (MPa).
✱ "Rotate"
The measured pressure is shown in psi.
To confirm setting: "Press" until the display is no longer flashing.
29
7 Commissioning
7.4.1Display and setting options of the instrument
Measuring rangeUnit
mbar-400.0400.0
bar-0.4000.400
-0.4 ... 0.4 bar
-1 ... 3 bar
0...60bar
-1 ... 9 bar
kPa-40.0040.00
MPa-0.0400.040
psi-5.8025.802
mbar-10003000
bar-1.0003.000
kPa-100.0300.0
MPa-0.1000.300
psi-14.5043.51
mbar00009999
bar00.0060.00
kPa00006000
MPa0.0006.000
psi000.0870.2
mbar-10009000
bar-1.0009.000
kPa-100.0900.0
MPa-0.1000.900
psi-14.5130.5
Display
StartEnd
-1 ... 24 bar
Gray cells are units that cannot be configured! Display overflow values do not appear in operation or in setup!
mbar-10009999
bar-1.0024.00
kPa-1002400
MPa-0.1002.400
psi-14.5348.1
30
7.5Setting the zero point (offset) (Off.P)
7.5.1Automatic offset adjustment
This setting is used to accept the current measured value as the new zero
point.
Automatic offset adjustment is only possible for instruments with a
relative pressure measuring range!
✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "Off.P".
✱ "Press" twice in quick succession
The current measured value is accepted as the zero point.
7.5.2Edited offset setting
7 Commissioning
This setting is used to increase or reduce the measured pressure selectively by
an adjustable value.
✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "Off.P".
✱ "Press"
"-" Flashing
Continuous
"-" means: the offset is negative - the measured pressure is
reduced.
Enter the value "digit by digit."
31
7 Commissioning
7.6Setting the filter time constant (damping) (DamP)
The filter time constant (damping) can be used to smooth the measured value.
Small filter time constant: the display is refreshed quickly.
Large filter time constant: Display refresh is slower.
The value is entered in seconds with two places after the decimal.
✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "Dam.P".
✱ "Press"
7.7Setting the output signal (S.TyP)
✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "S.TyP".
✱ "Press"
Example
4.20A = output signal 4 ... 20 mA
0.20A = output signal 0 ... 20 mA
0.10U = output signal 0 ... 10 V
32
7.8Setting scaling
7 Commissioning
Customerspecific
measuring
range
The customer measuring range (2) is defined by:
- Range start (4)
-Range end (5)
-Span (MSP)
ExampleActual
The instrument has a nominal measuring range (1) from 0 to 4 bar
Tar ge t
The customer would like to measure the pressure in the range from 1 to 2 bar
(25% of the nominal measuring range).
Range start (5) is 1 bar
Range end (6) is 2 bar
Span (MSP) is 1 bar
0
1
2
54
3
4 bar
1
2
ScalingThe scaling of the instrument's output signal describes how the measured
pressure is converted into an output signal.
Simple
example
Actual
The instrument has a nominal measuring range (1) from 0 to 4 bar and
the instrument has an output signal from 4 to 20 mA (3).
Tar ge t
The customer would like:
the "Customer measuring range" (2) from 0 to 4 bar (100% of nominal
measuring range (1)
should correspond to the output signal (3) from 4 to 20 mA (100%).
The scaling is 1: 1 (100% to 100%).
0
4
1
2
100%
100%
3
4 bar
1
2
3
20 mA
Customerspecific scaling
It is often useful to scale part of the nominal measuring range to the
output signal.
33
7 Commissioning
ExampleActual
The instrument has a nominal measuring range (1) from 0 to 4 bar and
the instrument has an output signal from 4 to 20 mA (3).
Tar ge t
The customer would like:
the "Customer measuring range" (2) from 1 to 2 bar (25% of nominal
measuring range (1)
should correspond to the output signal from 4 to 20 mA (100%).
The scaling is 1: 4 (25% to 100%).
Inversion of the
output
signal
Example
20 to 4 mA
0
4
1
2
100%
3
4 bar
1
2
3
20 mA
The instrument provides the option of inverting the output signal (3).
The output signal
- 0 to 20 mA becomes output signal 20 to 0 mA
- 4 to 20 mA becomes output signal 20 to 4 mA
- 0 to 10 V becomes output signal 10 to 0 V
0
1
2
3
4 bar
1
Abbreviations
34
100%
20
4mA
(1) Nominal measuring range (NMB)(4) Range start (MA)
(2) Customer measuring range (MB)(5) Range end (ME)
(3) Output signalSpan (MSP)
2
3
7 Commissioning
7.8.1Setting the starting value of scaling (Sc.Lo)
The output signal can only be scaled for instruments with
analog output!
Setting range:0 to 75% of the nominal measuring range
Factory setting:Initial value of measuring range
ExampleThe instrument has a nominal measuring range -400 ... +400 mbar
The output signal of the instrument is 0 ...20 mA
Objective:The range from 0 to 200 mbar (customer's measuring
range) will be represented on the output side by
0...20mA.
Setting:The initial value of scaling (Sc.Lo) = 0.000
The final value of scaling (Sc.Hi) = 200.0
Result:At a pressure of less than 0 mbar the instrument reports
an error (value below lower measuring range limit) and
makes the corresponding error signal (0 mA) available at
the output.
At a pressure of 0 mbar the instrument makes 0 mA
available at the output.
At a pressure of +200 mbar the instrument makes 20 mA
available at the output.
At a pressure greater than +200 mbar the instrument
reports an error (measuring range exceeded) and makes
the corresponding error signal (22 mA) available at the
output.
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "Sc.Lo".
✱ "Press"
"-" Flashing
Continuous
Enter the value "digit by digit."
35
7 Commissioning
7.8.2Setting the final value of scaling (Sc.Hi)
The output signal can only be scaled for instruments with
analog output!
Explanation See section 7.8.1 "Setting the starting value of scaling
(Sc.Lo)", page 35.
Setting range:25 to 100% of the nominal measuring range
Factory setting:Final value of measuring range
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "Sc.Hi".
✱ "Press"
"-" Flashing
Continuous
Enter the value "digit by digit."
7.9Setting the error signal (S.Err)
Only for instruments with analog output, an error signal is
generated for overrange or underrange!
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "S.Err".
✱ "Press"
36
7 Commissioning
Example
3.40nA =For underrange
error signal = 0 mA for measuring range 0 ... 20 mA
error signal = 3.4 mA for measuring range 4 ... 20 mA
error signal = 0 V for measuring range 0 ... 10 V
22nA =For overrange
error signal = 22 mA for measuring range 0 ... 20 mA
error signal = 22 mA for measuring range 4 ... 20 mA
error signal = 10.7 V for measuring range 0 ... 10 V
7.10Setting the switching function (B.Fct)
GeneralThe response of the instrument's switching output can be selected:
- Hysteresis make contact
- Hysteresis break contact
- Window function make contact
- Window function break contact
7.10.1 Hysteresis
Relay behavior
(1)SPSwitching point(4)noSPST (make)
(2)RSPReset point(5)ncSPST (break)
(3)Hysteresis
p
1
2
1
0
1
0
1
0
1
0
1
0
1
0
3
t
4
5
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
37
7 Commissioning
✱ "Rotate" until the bottom line shows "B.Fct".
✱ "Press"
1
7.10.2 Window
Relay response
0
B.RSp
B.Sp
0 = Hysteresis of make contact (switching difference) (factory setting)
1
0
B.RSp
B.Sp
1 = Hysteresis of break contact (switching difference) = min. contact
p
2
1
p
p
38
1
2
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
4
5
(1)SPSwitching point(4)noSPST (make)
(2)RSPReset point(5)ncSPST (break)
t
2 = Window function make contact
7 Commissioning
B.Hys
1
B.Hys
3 = Window function break contact
7.11Setting the switching point (B.SP)
See section 7.10 "Setting the switching function (B.Fct)", page 37.
Setting range:0 to 100% of the nominal measuring range
Factory setting:50% of the nominal measuring range
0
B.RSp
B.Hys
1
0
B.RSp
B.Hys
B.Sp
B.Sp
p
p
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "B.SP".
✱ "Press"
"-" Flashing
Continuous
Enter the value "digit by digit."
39
7 Commissioning
7.12Setting the reset point (B.RSP)
See section 7.10 "Setting the switching function (B.Fct)", page 37.
Setting range:0 to 100% of the nominal measuring range
Factory setting:40% of the nominal measuring range
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "B.RSP".
✱ "Press"
"-" Flashing
Continuous
Enter the value "digit by digit."
7.13Setting the switching difference (hysteresis) (B.HYS)
See section 7.10 "Setting the switching function (B.Fct)", page 37.
Setting range:0 to 100% of the nominal measuring range
Factory setting:40% of the nominal measuring range
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "B.HYS".
✱ "Press"
"-" Flashing
40
Continuous
Enter the value "digit by digit."
7 Commissioning
7.14Setting the switching delay (B.DLY)
Setting range:0.00 to 99.99 s
Factory setting:0.00 s
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "B.DLY".
✱ "Press"
"0" flashes
Continuous
Enter the value "digit by digit."
7.15Setting the display alignment (D.Dir)
Setting range:std = standard = instrument upright
turn = turned = instrument overhead
Factory setting:std
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "D.Dir".
✱ "Press"
or
41
7 Commissioning
7.16Setting the display unit (D.Uni)
Setting range:Uni.P = pressure unit set as for "Uni.P", See section 7.4
"Selecting the unit of the measured value (Uni.P)", page
28.
Pro2 = percentage of scaled measuring range = "Sc.Hi"
minus "Sc.Lo", See section 7.8.1 "Setting the starting
value of scaling (Sc.Lo)", page 35 and See section 7.8.2
"Setting the final value of scaling (Sc.Hi)", page 36.
Factory setting:std
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "D.Uni".
✱ "Press"
Uni.P = the measured value is displayed in the unit that was selected, See
section 7.4 "Selecting the unit of the measured value (Uni.P)", page
28.
or
Pro.2 = the measured value is displayed as a percentage of the scaled
measuring range, See section 7.8.1 "Setting the starting value of
scaling (Sc.Lo)", page 35 and See section 7.8.2 "Setting the final
value of scaling (Sc.Hi)", page 36.
ExampleThe measuring range of the instrument was set to from -50 to +350 mbar and
the scaled measuring range is 300 mbar.
If the instrument measures a pressure of +150 mbar, 50% is displayed.
42
7 Commissioning
7.17Displaying the version of the operating device software
(SW.Di)
Setting range:Read only
Factory setting:-
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "SW.Di".
✱ "Press"
"Alternating"
7.18Displaying the version of the signal stage software
(SW.Si)
Setting range:Read only
Factory setting:-
Setting✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
✱ "Rotate" until the bottom line shows "SW.Sir".
✱ "Press"
"Alternating"
43
8 Calibration
8.1Setting the zero point (offset)
8.1.1Automatic offset adjustment
Automatic offset adjustment is only possible for instruments with a
relative pressure measuring range!
On the instrument
See section 7.5.1 "Automatic offset adjustment", page 31.
By setup program
not possible.
8.1.2Edited offset setting
On the instrument
See section 7.5.2 "Edited offset setting", page 31.
By setup program
Connect the instrument with the PC and start the setup program,
See section 9 "Setup program", page 45 and following.
Input / offset.
44
9.1Function
9 Setup program
Configurable
parameters
Connection
The optionally available PC setup software (TN 20/00522384) can be used to
operate the instrument conveniently from a PC.
Depending on the device design, the following settings are possible, for
example:
- Measuring range and limits of measuring range.
- Response of outputs when the measuring range is exceeded.
- Functions of switching outputs K1 and K2.
- Setting special functions (for example tables for special linearizations).
Data can be transferred from or to the transmitter if it is connected
to the power supply; See section 4 "Electrical connection", page
11 and following.
(2)
(3)
(c)
(b)
(a)
(1)
DC 24 V
(1) Power supply DC 24 V(a) 4-pin cable socket (straight)
M12x1 with 2-m PVC cable
TN 40/00404585 or 4-pin angle
box M12x1 with 2-m PVC cable
TN 40/00409334
(2) Pressure switch type 405052(b) Connecting cable
TN 70/00507861
(4)
(3) USB/TTL converter
Part of TN 70/00456352
(4) Notebook / PC
(c) PC interface line (gray)
Part of (3)
45
9 Setup program
(2)
(1)
DC 24 V
(a)
(b)
(3)
(c)
(4)
✱ Install the setup program software on the notebook / PC.
During the installation, the driver for the USB/TTL converter is also
installed on the PC.
✱ Screw the connecting cable (b) onto the plug of the pressure switch (2).
✱ Connect the USB/TTL converter (3) to the connecting cable (b) and PC
interface line (gray) (c).
✱ Connect the PC interface line (gray) (c) to the notebook (4).
✱ Connect the cable (a) to the power supply (1) and connecting
Configurable in the nominal measuring range (> reset point)
Configurable in the nominal measuring range (< switching point)
Configurable in the nominal measuring range
99.99 sec
99.99 sec
R
≥ (UB-6.5 V) / 0.022 A
i
R
≥ (UB-6.5 V) / 0.022 A
i
R ≥ 10 kΩ
Mechanical properties
Process connection
Material
Surface
Process seal
- Process connection
521, 523, 571, 575,
576, 652
Stainless steel 316L
Ra ≤ 0.8 µm
All flange connections are welded and therefore have no seals
FPM standard
- Process connection
997 (JUMO PEKA)
Measurement
diaphragm
Material
Surface
Enclosure
Material
Surface
Threaded sleeve M12x1
Enclosure seal
Display
Screw plug
Material
Surface
Enclosure seal
Weight0.2 kg with process connection 504 (G1/2)
FPM, VMQ (silicon), EPDM also available, see Data Sheet 40.9711
Stainless steel 316L
Ra ≤ 0.8 µm
Stainless steel 316L
Ra ≤ 0.8 µm
Stainless steel 316L
VMQ (silicon)
PA
Aluminum 3.2315
Eloxal coating
VMQ (silicon)
51
11 Instrument description
Ambient conditions
Permissible
temperatures
Measuring material
-for basic type
extension 004
Environment
-25 ... +100°C (+135°C max. 1 hour per day)
-25 ... +200°C
-25 ... +75°C
-Ambient
temperature -50°C
Storage
Permissible
relative humidity
-in operation
-in storage
Permissible
mechanical loading
-Vibration
strength
- Shock resistance
Electromechanical
compatibility
- Interference emission
- Interference immunity
ProtectionIP 67 to DIN 60529
Restricted function: Stationary use only, danger of broken cable,
display does not function
-40 ... +85°C
100% incl. condensation on instrument outer sleeve
90% without condensation
20 g, 10 ... 2000 Hz per IEC 60068-2-6
Performance characteristic A per EN 61326
(Only with 4-pin connecting cable and grounded enclosure)
Class A per EN 61326
50 g for 11 ms / 100 g for 1 ms per IEC 60068-2-27
52
Accuracy
Relative pressure
Nominal (rated)
measuring range
Linearity
Accuracy at
+20°C
Accuracy at
-20...+75°C
1
2
3
11 Instrument description
0.414102560
0.150.150.10.10.10.1% of end val. of
0.350.30.250.250.250.25% of end val. of
0.70.60.50.50.50.5% of end val. of
bar
nom. mess. range
nom. mess. range
nom. mess. range
Nominal (rated)
-0.4...0.4
-1...1-1...3-1...9-1...24bar
measuring range
Linearity
Accuracy at
+20°C
Accuracy at
-20...+75°C
1
0.150.150.10.10.1% of end val. of
0.350.30.250.250.25% of end val. of
2
0.70.60.50.50.5% of end val. of
3
Absolute pressure
Nominal (rated)
0.41 4 102560bar
measuring range
Linearity
Accuracy at
+20°C
Accuracy at
-20...+75°C
1
0.150.150.10.10.10.1% of end val. of
0.350.30.250.250.250.25% of end val. of
2
0.70.60.50.50.50.5% of end val. of
3
Long-term stability 0.2% with reference conditions to EN 61298-1
nom. mess. range
nom. mess. range
nom. mess. range
nom. mess. range
nom. mess. range
nom. mess. range
1
Linearity based on limit point setting
2
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of
measuring range.
3
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of
measuring range, thermal effect on initial value of measuring range and span.
53
JUMO GmbH & Co. KGJUMO Instrument Co. Ltd.JUMO Process Control, Inc.