JUMO DELOS SI, 405052 Operating Manual

JUMO DELOS SI
Electronic Pressure Switch with Display
B 40.5052.0
Operating Manual
2009-06-15/00526948
Contents
1 Typographical conventions ...................................................... 5
1.1 Warning signs ..............................................................................................5
2 Description ................................................................................ 6
2.1 General .........................................................................................................6
3 Identifying the instrument version ........................................... 7
3.1 Nameplate ....................................................................................................7
3.2 Block diagram ..............................................................................................7
3.3 Type description ..........................................................................................8
3.4 Accessories ................................................................................................10
4 Electrical connection .............................................................. 11
4.1 Installation instructions ..............................................................................11
4.2 Color assignment of M12 x 1 round plug ..................................................11
4.3 Terminal assignment for output 470 ..........................................................12
4.4 Terminal assignment for output 471 ..........................................................12
4.5 Terminal assignment for outputs 475, 476, and 477 .................................13
5 Mounting .................................................................................. 14
5.1 General information ....................................................................................14
5.2 Dimensions of electronic pressure switches .............................................16
5.3 Dimensions of process connections, not front-flush .................................17
5.4 Dimensions of process connections, front-flush .......................................18
6 Operation ................................................................................. 21
6.1 Controls ......................................................................................................21
6.2 LC display ..................................................................................................21
6.3 Levels .........................................................................................................22
6.4 Parameter ...................................................................................................23
Contents
7 Commissioning ....................................................................... 26
7.1 Getting started ...........................................................................................26
7.2 Unlocking the instrument (code entry) .......................................................27
7.3 Cancel operation ........................................................................................28
7.4 Selecting the unit of the measured value (Uni.P) .......................................28
7.5 Setting the zero point (offset) (Off.P) ..........................................................31
7.6 Setting the filter time constant (damping) (DamP) .....................................32
7.7 Setting the output signal (S.TyP) ................................................................32
7.8 Setting scaling ............................................................................................33
7.9 Setting the error signal (S.Err) ....................................................................36
7.10 Setting the switching function (B.Fct) ........................................................37
7.11 Setting the switching point (B.SP) .............................................................39
7.12 Setting the reset point (B.RSP) ..................................................................40
7.13 Setting the switching difference (hysteresis) (B.HYS) ................................40
7.14 Setting the switching delay (B.DLY) ...........................................................41
7.15 Setting the display alignment (D.Dir) ..........................................................41
7.16 Setting the display unit (D.Uni) ...................................................................42
7.17 Displaying the version of the operating device software (SW.Di) ..............43
7.18 Displaying the version of the signal stage software (SW.Si) ......................43
8 Calibration ............................................................................... 44
8.1 Setting the zero point (offset) .....................................................................44
9 Setup program ........................................................................ 45
9.1 Function .....................................................................................................45
9.2 Start the setup program .............................................................................46
10 Eliminating errors and faults .................................................. 48
10.1 Possible errors ...........................................................................................48
11 Instrument description ........................................................... 49
11.1 Technical data ..........................................................................................49

1.1 Warning signs

Danger
This symbol is used when there may be danger to personnel if the instructions are ignored or not followed correctly!
Caution
This symbol is used when there may be damage to equipment or data if the instructions are ignored or not followed correctly!
Caution
This symbol is used where special care is required when handling components liable to damage through electrostatic discharge.

1.2 Note signs

Note This symbol is used to draw your special attention to a remark.

1 Typographical conventions

1
abc
Action instruction
Footnote
Footnotes are remarks that refer to specific points in the text. Footnotes consist of two parts:
Marking in the text and the footnote text. The markers in the text are arranged as sequential superscript numbers.
This symbol indicates that an action to be performed is described. The individual steps are marked by this asterisk. Example: Loosen Phillips-head screws.
5

2 Description

2.1 General

Depending on its design, the instrument measures relative or absolute
(1)
pressure in liquid and gaseous media.
The pressure is displayed digitally.Depending on the design, the following outputs are available:
- 1 PNP switching output
- 2 PNP switching outputs
- 1 PNP switching output + 1 analog output 4 ... 20 mA
- 1 PNP switching output + 1 analog output 0 ... 20 mA
- 1 PNP switching output + 1 analog output 0 ... 10 V
The instrument is also available in a design for use at elevated medium
temperatures.
The instrument can be adjusted directly on site or can be configured via PC
with a setup program.
The protection type specified for the device (See section 11.1 "Technical data", page 49) can only be achieved with the control opening (1) closed.
1 1
1
1
The output is freely configurable.
6

3.1 Nameplate

on the pressure switch
(1)

3 Identifying the instrument version

Made in Germany
(2)
www.jumo.net
DELOS SI 405052/000 TN 12345678
(3)
(5)
(1)
Ty pe
(2)
Part number
(3)
Measuring range
(4)
Power supply voltage

3.2 Block diagram

Pressure
(4)
(6)
0...250 mbar
DC 12...30 V
4...20 mA/OC: 250 mA
F.-Nr.: 001234567801008350123
(5) (6) (7)
}
(7)
Output signal Manufacturing number Date of manufacture
(year and calendar week)
Interface
Setup program
Rotary encoder
To operate: turn and press
Power supply
12 30 V DC?
1
optional
Display
2 x 4 characters, 2 colors
Switching output 1
Open collector
Type 405052
Analog output
0 (4) 20 mA oder 0 10 V??
Switching output 2
Open collector
(1)
(1)
7
3 Identifying the instrument version

3.3 Type description

(1) Basic type
405052 JUMO DELOS SI
Electronic Pressure Switch with Display
(2) Basic type extension
000 None 004 For elevated medium temperatures up to 200°C 999 Special design
(3) Nominal (rated) measuring range
Measuring range - pressure
452 0 ... 400 mbar relative pressure 454 0 ... 1 bar relative pressure 457 0 ... 4 bar relative pressure 459 0 ... 10 bar relative pressure 461 0 ... 25 bar relative pressure 463 0 ... 60 bar relative pressure
Measuring range - vacuum
447 -400 ... 400 mbar relative pressure 449 -1 ...1 bar relative pressure 481 -1 ... 3 bar relative pressure 483 -1 ... 9 bar relative pressure 485 -1 ... 24 bar relative pressure
1
Measuring range absolute pressure
486 0 ... 400 mbar absolute pressure 488 0 ... 1 bar absolute pressure 491 0 ... 4 bar absolute pressure 493 0 ... 10 bar absolute pressure 495 0 ... 25 bar absolute pressure 506 0 ... 60 bar absolute pressure
(4) Output
470 1 x PNP switching output 471 2 x PNP switching output 475 1 x PNP switching output + analog output 4 ... 20 mA,
three wires
2
476 1 x PNP switching output + analog output 0 ... 20 mA,
three wires
2
477 1 x PNP switching output + analog output 0 ... 10 V,
three wires
2
1
Front-flush only for process connection.
2
Factory setting - the output is freely configurable.
8
3 Identifying the instrument version
(5) Process connection
Not front-flush
504 G 1/2 511 1/4-18 NPT 521 G 1/4 to DIN 3852 T11 523 G 1/2 to DIN 3852 T11 998 Suitable for connecting to pressure measuring instrument
Front-flush
571 G 3/4 575 G 3/4 seal positioned in front 576 G1 seal positioned in front 603 Tapered adapter with groove nut, to DIN11 851, DN20 604 Tapered adapter with groove nut, to DIN11 851, DN25 606 Tapered adapter with groove nut, to DIN11 851, DN40 607 Tapered adapter with groove nut, to DIN11 851, DN50 612 Clamp to DIN 32 676, DN20 613 Clamp to DIN 32 676, DN25 616 Clamp to DIN 32 676, DN50 623 Small flange connection
4 4 4
4
652 Tank connection with groove union nut4, DN25 997 JUMO PEKA (EHEDG-certified)
3
998 Suitable for connecting to pressure measuring instrument
4 4 4 4
(6) Process connection material
20 Stainless steel 316L
(7) Electrical connection
36 Round plug M 12 x 1
(8) Measuring system filling medium
01 FDA-compliant oil 99 Special filling medium
(9) Extra codes
000 None 591 Throttle in pressure channel 624 Free of oil and grease 691 Cast version
(1) (2) (3) (4) (5) (6) (7) (8) (9)
Order code /------/
Sample order 405052 / 000 - 459 - 471 - 504 - 20 - 36 - 1 / 000
3
For matching process connection adapter see data sheet 40.9711
4
For measuring ranges up to 25 bar only
9
3 Identifying the instrument version

3.4 Accessories

Sales No. Designation 40/00404585 4-pin cable connector (straight) M 12 x 1 with 2-m PVC cable 40/00409334 4-pin angle box M 12 x1 with 2-m PVC cable 40/00522384 Setup program 40/00507861 Connecting cable (only required for programming with the
setup program)
70/00456352 PC interface line including USB/TTL converter and
adapter (USB connecting cable)
1
1
Please order a connecting cable as well.
10

4.1 Installation instructions

The electrical connection must only be performed by qualified personnel!
The load circuits must be fused for the maximum load currents in each case to
prevent the instrument from being destroyed.
Electromagnetic compatibility meets the requirements of EN 61326,No other consumers can be connected to the power supply of the instrument.The instrument is not suitable for installation in areas with an explosion hazard.Apart from faulty installation, incorrect settings on the instrument may also
affect the proper functioning of the subsequent process or lead to damage. You should therefore always provide safety equipment that is independent of the instrument and it should only be possible for qualified personnel to make settings.

4 Electrical connection

4.2 Color assignment of M12 x 1 round plug

The following color assignment applies only to A-coded standard cables!
2
3
5
1
4
1Brown 2White 3Blue 4Black 5Gray
11
4 Electrical connection

4.3 Terminal assignment for output 470

One PNP switching output
2
3
5
1
4
K1
Power supply voltage
1 L+ 12 ... 30 V DC 3 L- GND
Output
4 K1 Highside Open Collector
2nc 5 Interface
L+
L-
maximum 0.25 A

4.4 Terminal assignment for output 471

Two PNP switching outputs
2
3
5
1
4
K1 K2
Power supply voltage
1 L+ 12 ... 30 V DC 3 L- GND
Output
4 K1 Highside Open Collector 2 K2 5 Interface
L+
L-
maximum 0.25 A
12
4 Electrical connection

4.5 Terminal assignment for outputs 475, 476, and 477

One PNP switching output + one analog output
2
3
5
1
4
K1
Power supply voltage
1 L+ 12 ... 30 V DC 3 L- GND
Output
4 K1 Highside Open Collector
2 Analog 0(4)...20 mA / 0...10 V 5 Interface
L+
L-
maximum 0.25 A
13

5 Mounting

5.1 General information

The compatibility of the instrument with the measuring medium must be tested, See section 11.1 "Technical data", page 49.
Mounting location
Installation position
- Find a location that ensures easy accessibility for later operation.
- The fastening must be secure and must ensure low vibration for the instrument.
- Avoid direct sunlight!
- Permissible ambient temperature at the installation location See section
11.1 "Technical data", page 49.
The instrument can be mounted in any position. The "vertical" installation position is recommended, see illustration.

5.1.1 Rotating the display

The display image can be rotated 180° in the software, See section "Display position", page 25. This may make it easier to read when the instrument is installed overhead, for example.
14

5.1.2 Rotating the housing

The instrument housing can be rotated a maximum of 320° with the combination tool (1).
5 Mounting
(1)
320° max.
15
5 Mounting

5.2 Dimensions of electronic pressure switches

35
30
75
ø42
M12x1
40
Type 405052/000-...
Type 405052/004-...
The overall height is 40 mm greater for instruments with basic type extension 004 (for increased medium temperature up to 200°C). See drawing
16
5 Mounting

5.3 Dimensions of process connections, not front-flush

504 511
E
31
20
G1/2
G1/2
19
1.6
ø7
15
11.07
E
26
15
1/4-18 NPT
1/4-18 NPT
10.4
521 523
E
E
G1/4
ø19
G1/4
A = Profile seal DN G3/4 B = Profile seal DN G1/2 E = SW 27 F = SW 32
12
1.5
12
1.6
A
23.5
G1/2
ø27
G1/2
B
14
1.6
14
17
5 Mounting

5.4 Dimensions of process connections, front-flush

571 575
1.6
G3/4
G3/4
576
16
16
F
A
32
ø24
G3/4
D
1.6
13
16
20.5
G
G3/4
ø24.5
1.6
F
A
34
20.5
9
ø30
G1
20.5
D
1.6
17.5
14
G1
30.5
A = profile seal DN G3/4 B = profile seal DN G1 C = O-ring 26.7 x 1.78
B
35
20.5
7
C
1.6
D = Drill out after tapping F = SW 32 G = SW 41
18
L1
603 to 607
D2
D1 D3 D4
5 Mounting
Process
connection
DN ø D1 ø D2 ø D3 ø D4 L1
603 20 36.5 30 RD 44 x 1/6 54 13 604 25 44 35 RD 52 x 1/6 63 606 40 56 48 RD 65 x 1/6 78 607 50 68.5 61 RD 78 x 1/6 92 16
612 to 616
3
15
D
1
D
2
15
Process
connection
612
DN
DIN 32676
20 15
25
613
32 40
616 50 2“
DN
(inches)
1“
1.5“
Nominal Size
ISO 2852
12
12.7
17.2
21.3 25
33.7 38
40 51
ø D1 ø D2
27.5 34
43.5 50.5
56.5 64
19
5 Mounting
623 652
ø19
ø26
ø40
997
45
35
15
3
21
31
RD52 x 1/6
55
63
E
11
E = SW 27
28
Process connection 997 is EHEDG-certified For detailed information about this process connection system, see data sheet 40.9711.
20

6.1 Controls

6 Operation

(1)
(2)
(1) Protective screw
(2) Hexagon socket
Unscrew the protective screw (1)."Turn / push" the control element (2) with the enclosed combination tool
(or a 0.5x3 screwdriver).

6.2 LC display

6.2.1 Measurement mode (normal display)

Example: The display is lit yellow.
21
6 Operation

6.2.2 Settingmode

Example: The display is lit red.
Operation
Continue Press the combination tool less than 1 second ( < 1s) Yes (accept) Press the combination tool less than 1 second ( < 1s) No (Cancel) Press the combination tool more than 3 seconds ( > 3s) Timeout No activity for more than 60 seconds ( >60 s)

6.3 Levels

To return to measuring mode:
- "No (Cancel)" or
- Wait for timeout = no activity performed for 60 seconds.
Process value display
(Normal display)
< 1 s
Code = 0
Enter
code
Cancel or timeout
< 1 s
> 3 s
22
< 1 s
Select
parameter
edit
parameter
Cancel or save
Cancel or timeout
> 3 s
< 1 s
Cancel or timeout
> 3 s
> 3 s
= "Press"

6.4 Parameter

6.4.1 Input

6 Operation
Parameter Display Setting range
Pressure unit bar
kPa MPa psi mbar
Note:
The units kPa and mbar cannot be configured for all measuring ranges.
Offset (zero-point correction)
Damping (filter time constant)
-20.00 ... 0.00 ... +20.00% of the measuring range
Note:
Automatic offset correction See section 7.5 "Setting the zero point (offset) (Off.P)", page 31.
0.00 ... 99.99 s

6.4.2 Analog output

Parameter Display Setting range
Signal type (for analog output)
4 ... 20 mA
0 ... 20 mA 0 ... 10 V
1
1
Scaling start (for analog output)
Scaling end (for analog output)
Signal for error (for analog output)
1
The default setting is marked in bold.
0.00 ... 75.00% of nominal measuring range
25.00 ... 100% of nominal measuring range
3.4 mA or 22 mA for output signal 4...20 mA 0 mA or 22 mA for output signal 0...20 mA 0 V or 10.7 V for output signal 0...10 V
Note:
Depending on the configured output signal.
23
6 Operation

6.4.3 Binary output 1

Parameter Display Setting range
Switching function (for switching output only)
Switching point (for switching output only)
Reset point (for switching output only)
Hysteresis (for switching output and configured switching point or reset point only)
Switching delay (for switching output only)
Hysteresis, make contact
Hysteresis, break contact Window, make contact Window, break contact
See section 7.10 "Setting the switching function (B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range See section 7.10 "Setting the switching function
(B.Fct)", page 37.
Note:
Used only with window switching functions.
0.00 ... 99.99 s See section 7.10 "Setting the switching function
(B.Fct)", page 37.
1

6.4.4 Binary output 2

Parameter Display Setting range
Switching function (for second switching output only)
Switching point (for second switching output only)
Reset point (for switching output only)
Hysteresis (for second switching output and configured switching point or reset point only)
Hysteresis, make contact
Hysteresis, break contact Window, make contact Window, break contact
See section 7.10 "Setting the switching function (B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range See section 7.10 "Setting the switching function
(B.Fct)", page 37.
0.00 ... 100.00% of nominal measuring range See section 7.10 "Setting the switching function
(B.Fct)", page 37.
Note:
Used only with window switching functions.
1
1
The default setting is marked in bold.
24
6 Operation
Parameter Display Setting range
Switching delay (for second switching output only)
0.00 ... 99.99 s See section 7.10 "Setting the switching function
(B.Fct)", page 37.
1

6.4.5 Display and operation

Parameter Display Setting range
Display position Normal (for normal operation)
Rotated (for overhead operation) See section 7.15 "Setting the display alignment
(D.Dir)", page 41.
Unit of actual value display (for analog output only)
Version D Software version of the operating device
Version S Software version of the signal stage
Pressure unit (see parameter "Uni.P") Percentage of the scaled range
See section 7.16 "Setting the display unit (D.Uni)", page 42.
See section 7.17 "Displaying the version of the operating device software (SW.Di)", page 43.
1
Code (can only be edited via
setup program)
See section 7.18 "Displaying the version of the signal stage software (SW.Si)", page 43.
0000 ... 0072 .... 9999
See section 7.2 "Unlocking the instrument (code entry)", page 27.
25

7 Commissioning

7.1 Getting started

Mounting the instrument, See section 5 "Mounting", page 14.Installing the instrument, See section 4 "Electrical connection", page 11.Unlocking the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
Selecting the unit of the measured value, See section 7.4 "Selecting the unit
of the measured value (Uni.P)", page 28.
Adjusting the output signal, See section 7.7 "Setting the output signal
(S.TyP)", page 32.
Adjusting the scaling of the output signal (restricting the measuring range),
See section 7.8 "Setting scaling", page 33.
This is a suggestion for configuring the instrument reliably in little time.
By checking the setting options of this list before starting the configuration, you can avoid timeouts during the configuration.
Setting the switching function, See section 7.10 "Setting the switching
function (B.Fct)", page 37.
Setting the switching point, See section 7.11 "Setting the switching point
(B.SP)", page 39.
26

7.2 Unlocking the instrument (code entry)

The instrument is protected by a code to prevent unauthorized operation. The code is set to 0072 in the factory. It can only be changed with the setup
program. If the code is set to 0000 with the setup program, the instrument is unprotected.
Unlocking
(1)
7 Commissioning
(2)
Unscrew the protective screw (1)Continue briefly pressing the combination tool (2) until the third "0" from the
left is flashing. The color of the display also changes to "red."
Turn the combination tool until "7" is displayed.Continue briefly pressing the combination tool until the fourth "0" from the
left is flashing.
Turn the combination tool until "2" is displayed.
Press the combination tool briefly - the instrument switches to the
27
7 Commissioning
parameter level.
After an incorrect code is entered:

7.3 Cancel operation

Press and hold the combination tool (2) longer than 3 seconds or wait for timeout (no activity for longer than 60 seconds).

7.4 Selecting the unit of the measured value (Uni.P)

Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "Uni.P"."Press"
The measured pressure is shown in millibar."Press"
28
Flashing
Continuous
The measured pressure is shown in bar.
7 Commissioning
"Rotate"
The measured pressure is shown in Kilopascal. "Rotate"
The measured pressure is shown in Megapascal (MPa). "Rotate"
The measured pressure is shown in psi. To confirm setting: "Press" until the display is no longer flashing.
29
7 Commissioning

7.4.1 Display and setting options of the instrument

Measuring range Unit
mbar -400.0 400.0
bar -0.400 0.400
-0.4 ... 0.4 bar
-1 ... 3 bar
0...60bar
-1 ... 9 bar
kPa -40.00 40.00
MPa -0.040 0.040
psi -5.802 5.802
mbar -1000 3000
bar -1.000 3.000
kPa -100.0 300.0
MPa -0.100 0.300
psi -14.50 43.51
mbar 0000 9999
bar 00.00 60.00
kPa 0000 6000
MPa 0.000 6.000
psi 000.0 870.2
mbar -1000 9000
bar -1.000 9.000
kPa -100.0 900.0
MPa -0.100 0.900
psi -14.5 130.5
Display
Start End
-1 ... 24 bar
Gray cells are units that cannot be configured! Display overflow ­values do not appear in operation or in setup!
mbar -1000 9999
bar -1.00 24.00
kPa -100 2400
MPa -0.100 2.400
psi -14.5 348.1
30

7.5 Setting the zero point (offset) (Off.P)

7.5.1 Automatic offset adjustment

This setting is used to accept the current measured value as the new zero point.
Automatic offset adjustment is only possible for instruments with a relative pressure measuring range!
Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "Off.P"."Press" twice in quick succession
The current measured value is accepted as the zero point.

7.5.2 Edited offset setting

7 Commissioning
This setting is used to increase or reduce the measured pressure selectively by an adjustable value.
Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "Off.P"."Press"
"-" Flashing
Continuous
"-" means: the offset is negative - the measured pressure is reduced.
Enter the value "digit by digit."
31
7 Commissioning

7.6 Setting the filter time constant (damping) (DamP)

The filter time constant (damping) can be used to smooth the measured value. Small filter time constant: the display is refreshed quickly. Large filter time constant: Display refresh is slower.
The value is entered in seconds with two places after the decimal. Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "Dam.P"."Press"

7.7 Setting the output signal (S.TyP)

Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "S.TyP"."Press"
Example
4.20A = output signal 4 ... 20 mA
0.20A = output signal 0 ... 20 mA
0.10U = output signal 0 ... 10 V
32

7.8 Setting scaling

7 Commissioning
Customer­specific measuring range
The customer measuring range (2) is defined by:
- Range start (4)
-Range end (5)
-Span (MSP)
Example Actual
The instrument has a nominal measuring range (1) from 0 to 4 bar
Tar ge t
The customer would like to measure the pressure in the range from 1 to 2 bar (25% of the nominal measuring range).
Range start (5) is 1 bar Range end (6) is 2 bar Span (MSP) is 1 bar
0
1
2
54
3
4 bar
1
2
Scaling The scaling of the instrument's output signal describes how the measured
pressure is converted into an output signal.
Simple example
Actual
The instrument has a nominal measuring range (1) from 0 to 4 bar and the instrument has an output signal from 4 to 20 mA (3).
Tar ge t
The customer would like: the "Customer measuring range" (2) from 0 to 4 bar (100% of nominal measuring range (1) should correspond to the output signal (3) from 4 to 20 mA (100%).
The scaling is 1: 1 (100% to 100%).
0
4
1
2
100%
100%
3
4 bar
1
2
3
20 mA
Customer­specific scaling
It is often useful to scale part of the nominal measuring range to the output signal.
33
7 Commissioning
Example Actual
The instrument has a nominal measuring range (1) from 0 to 4 bar and the instrument has an output signal from 4 to 20 mA (3).
Tar ge t
The customer would like: the "Customer measuring range" (2) from 1 to 2 bar (25% of nominal measuring range (1) should correspond to the output signal from 4 to 20 mA (100%).
The scaling is 1: 4 (25% to 100%).
Inversion of the output signal
Example 20 to 4 mA
0
4
1
2
100%
3
4 bar
1
2
3
20 mA
The instrument provides the option of inverting the output signal (3). The output signal
- 0 to 20 mA becomes output signal 20 to 0 mA
- 4 to 20 mA becomes output signal 20 to 4 mA
- 0 to 10 V becomes output signal 10 to 0 V
0
1
2
3
4 bar
1
Abbreviations
34
100%
20
4mA
(1) Nominal measuring range (NMB) (4) Range start (MA)
(2) Customer measuring range (MB) (5) Range end (ME)
(3) Output signal Span (MSP)
2
3
7 Commissioning

7.8.1 Setting the starting value of scaling (Sc.Lo)

The output signal can only be scaled for instruments with analog output!
Setting range: 0 to 75% of the nominal measuring range Factory setting: Initial value of measuring range
Example The instrument has a nominal measuring range -400 ... +400 mbar
The output signal of the instrument is 0 ...20 mA Objective: The range from 0 to 200 mbar (customer's measuring
range) will be represented on the output side by
0...20mA.
Setting: The initial value of scaling (Sc.Lo) = 0.000
The final value of scaling (Sc.Hi) = 200.0
Result: At a pressure of less than 0 mbar the instrument reports
an error (value below lower measuring range limit) and makes the corresponding error signal (0 mA) available at the output. At a pressure of 0 mbar the instrument makes 0 mA available at the output. At a pressure of +200 mbar the instrument makes 20 mA available at the output. At a pressure greater than +200 mbar the instrument reports an error (measuring range exceeded) and makes the corresponding error signal (22 mA) available at the output.
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "Sc.Lo"."Press"
"-" Flashing
Continuous
Enter the value "digit by digit."
35
7 Commissioning

7.8.2 Setting the final value of scaling (Sc.Hi)

The output signal can only be scaled for instruments with analog output!
Explanation See section 7.8.1 "Setting the starting value of scaling (Sc.Lo)", page 35.
Setting range: 25 to 100% of the nominal measuring range Factory setting: Final value of measuring range
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "Sc.Hi"."Press"
"-" Flashing
Continuous
Enter the value "digit by digit."

7.9 Setting the error signal (S.Err)

Only for instruments with analog output, an error signal is generated for overrange or underrange!
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "S.Err"."Press"
36
7 Commissioning
Example
3.40nA = For underrange
error signal = 0 mA for measuring range 0 ... 20 mA error signal = 3.4 mA for measuring range 4 ... 20 mA error signal = 0 V for measuring range 0 ... 10 V
22nA = For overrange
error signal = 22 mA for measuring range 0 ... 20 mA error signal = 22 mA for measuring range 4 ... 20 mA error signal = 10.7 V for measuring range 0 ... 10 V

7.10 Setting the switching function (B.Fct)

General The response of the instrument's switching output can be selected:
- Hysteresis make contact
- Hysteresis break contact
- Window function make contact
- Window function break contact

7.10.1 Hysteresis

Relay behavior
(1) SP Switching point (4) no SPST (make)
(2) RSP Reset point (5) nc SPST (break)
(3) Hysteresis
p
1
2
1
0
1
0
1
0
1
0
1
0
1
0
3
t
4
5
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
37
7 Commissioning
"Rotate" until the bottom line shows "B.Fct"."Press"
1

7.10.2 Window

Relay response
0
B.RSp
B.Sp
0 = Hysteresis of make contact (switching difference) (factory setting)
1
0
B.RSp
B.Sp
1 = Hysteresis of break contact (switching difference) = min. contact
p
2
1
p
p
38
1
2
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
4
5
(1) SP Switching point (4) no SPST (make)
(2) RSP Reset point (5) nc SPST (break)
t
2 = Window function make contact
7 Commissioning
B.Hys
1
B.Hys
3 = Window function break contact

7.11 Setting the switching point (B.SP)

See section 7.10 "Setting the switching function (B.Fct)", page 37. Setting range: 0 to 100% of the nominal measuring range Factory setting: 50% of the nominal measuring range
0
B.RSp
B.Hys
1
0
B.RSp
B.Hys
B.Sp
B.Sp
p
p
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "B.SP"."Press"
"-" Flashing
Continuous
Enter the value "digit by digit."
39
7 Commissioning

7.12 Setting the reset point (B.RSP)

See section 7.10 "Setting the switching function (B.Fct)", page 37. Setting range: 0 to 100% of the nominal measuring range Factory setting: 40% of the nominal measuring range
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "B.RSP"."Press"
"-" Flashing
Continuous
Enter the value "digit by digit."

7.13 Setting the switching difference (hysteresis) (B.HYS)

See section 7.10 "Setting the switching function (B.Fct)", page 37. Setting range: 0 to 100% of the nominal measuring range Factory setting: 40% of the nominal measuring range
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "B.HYS"."Press"
"-" Flashing
40
Continuous
Enter the value "digit by digit."
7 Commissioning

7.14 Setting the switching delay (B.DLY)

Setting range: 0.00 to 99.99 s Factory setting: 0.00 s
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "B.DLY"."Press"
"0" flashes
Continuous
Enter the value "digit by digit."

7.15 Setting the display alignment (D.Dir)

Setting range: std = standard = instrument upright
turn = turned = instrument overhead
Factory setting: std
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "D.Dir"."Press"
or
41
7 Commissioning

7.16 Setting the display unit (D.Uni)

Setting range: Uni.P = pressure unit set as for "Uni.P", See section 7.4
"Selecting the unit of the measured value (Uni.P)", page
28. Pro2 = percentage of scaled measuring range = "Sc.Hi" minus "Sc.Lo", See section 7.8.1 "Setting the starting value of scaling (Sc.Lo)", page 35 and See section 7.8.2 "Setting the final value of scaling (Sc.Hi)", page 36.
Factory setting: std
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "D.Uni"."Press"
Uni.P = the measured value is displayed in the unit that was selected, See
section 7.4 "Selecting the unit of the measured value (Uni.P)", page
28.
or
Pro.2 = the measured value is displayed as a percentage of the scaled
measuring range, See section 7.8.1 "Setting the starting value of scaling (Sc.Lo)", page 35 and See section 7.8.2 "Setting the final value of scaling (Sc.Hi)", page 36.
Example The measuring range of the instrument was set to from -50 to +350 mbar and
the scaled measuring range is 300 mbar. If the instrument measures a pressure of +150 mbar, 50% is displayed.
42
7 Commissioning

7.17 Displaying the version of the operating device software (SW.Di)

Setting range: Read only Factory setting: -
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "SW.Di"."Press"
"Alternating"

7.18 Displaying the version of the signal stage software (SW.Si)

Setting range: Read only Factory setting: -
Setting ✱ Unlock the instrument, See section 7.2 "Unlocking the instrument (code
entry)", page 27.
"Rotate" until the bottom line shows "SW.Sir"."Press"
"Alternating"
43

8 Calibration

8.1 Setting the zero point (offset)

8.1.1 Automatic offset adjustment

Automatic offset adjustment is only possible for instruments with a relative pressure measuring range!
On the instrument
See section 7.5.1 "Automatic offset adjustment", page 31.
By setup program
not possible.

8.1.2 Edited offset setting

On the instrument
See section 7.5.2 "Edited offset setting", page 31.
By setup program
Connect the instrument with the PC and start the setup program, See section 9 "Setup program", page 45 and following.
Input / offset.
44

9.1 Function

9 Setup program

Configurable parameters
Connection
The optionally available PC setup software (TN 20/00522384) can be used to operate the instrument conveniently from a PC.
Depending on the device design, the following settings are possible, for example:
- Measuring range and limits of measuring range.
- Response of outputs when the measuring range is exceeded.
- Functions of switching outputs K1 and K2.
- Setting special functions (for example tables for special linearizations).
Data can be transferred from or to the transmitter if it is connected to the power supply; See section 4 "Electrical connection", page 11 and following.
(2)
(3)
(c)
(b)
(a)
(1)
DC 24 V
(1) Power supply DC 24 V (a) 4-pin cable socket (straight)
M12x1 with 2-m PVC cable TN 40/00404585 or 4-pin angle box M12x1 with 2-m PVC cable TN 40/00409334
(2) Pressure switch type 405052 (b) Connecting cable
TN 70/00507861
(4)
(3) USB/TTL converter
Part of TN 70/00456352
(4) Notebook / PC
(c) PC interface line (gray)
Part of (3)
45
9 Setup program
(2)
(1)
DC 24 V
(a)
(b)
(3)
(c)
(4)
Install the setup program software on the notebook / PC.
During the installation, the driver for the USB/TTL converter is also installed on the PC.
Screw the connecting cable (b) onto the plug of the pressure switch (2).Connect the USB/TTL converter (3) to the connecting cable (b) and PC
interface line (gray) (c).
Connect the PC interface line (gray) (c) to the notebook (4).Connect the cable (a) to the power supply (1) and connecting
cable (b).

9.2 Start the setup program

Start / Programs / JUMO instruments / Setup program JUMO DELOS K, SI,
46
HP
9 Setup program
(1)
47

10 Eliminating errors and faults

10.1 Possible errors

t
Display Possible cause Measure
- Overrange or underrange
- Broken sensor.
Device error:
- 1 = Internal communication error
- 2 = Error analog output
- 3 = Short circuit Switching output 1
- 4 = Short circuit Switching output 2
- 5 = VCC 8 V outside of working range
- 6 to 8 = Internal communica­tion error
- 9 = Invalid configuration
Display overflow: Upper display: "- - - -" Lower display: Parameter name
Value is less than -9999 or greater than 9999.
Configure other measuring range, Chapter 7.4 "Selecting the unit of the measured value (Uni.P)".
1), 6), 7), 8): Call Customer Service; see the back of the Operating Manual.
2): Check the ambient temperature. Check output for broken line. Output burden is too high (for current output) or too low (for voltage output).
3), 4): Check the corresponding switching output.
5): Check the power supply.
9): Check the configuration.
Check the corresponding switching output.
Check the power supply.
48

11.1 Technical data

General

Reference conditions DIN 16086 and EN 60770 Sensor system
Construction
Silicon sensor with stainless steel separating diaphragm

11 Instrument description

Pressure transfer medium
Permissible load change
Location
Mounting location Calibration location Location-dependent
zero point offset
- Basic type extension 000 (standard design)
- Basic type extension 004 (for elevated medium temperature)
Display
Ty pe Alignment
Size Color
Synthetic oil, FDA-compliant more than 10 million
Any Device standing vertically, process connection on bottom Zero point correction possible locally or via setup (20% of nominal
measuring range)
1mbar
10 mbar
Backlit LCD Display can be rotated 180° (via setup)
Housing can be turned 320° (mechanically) Display field 16x26 mm / font size 7 mm / 2 x 4 places Normal operation:amber
Error: red
Setup mode: red Switching state indicator Measuring unit
Operation
Local via setup
K1, K2
mbar, bar, kPa, MPa, psi, %
With accompanying combination tool or screwdriver 0.5x3
Pin 12 of the M12 round plug
49
11 Instrument description

Input

All measuring ranges can be overloaded to -1 bar (vacuum-proof)
Relative pressure
Measuring range 0.4 1 4 10 25 60
Bursting pressure2 5 2050125300 Measuring range -0.4...0.4 -1...1 -1...3 -1...9 -1...24
Bursting pressure2 5 2050125
Absolute pressure
Measuring range 0.4 1 4 10 25 60
Bursting pressure2 5 2050125300

Outputs

Analog output
barOverload capacity 1.6 4 16 40 100 240
barOverload capacity 1.6 4 16 40 100
barOverload capacity 1.6 4 16 40 100 240
Freely configurable
Jump response T
Switching output
Number
Switching type Switching function
Switching capacity
- Voltage drop from U
- Contact rating
- Switching cycles Response time Short-circuit proof
Current load check
- Time interval
- Periodic protective
circuit in case of overcurrent
Setting range
Analog output Turn down 1:4
90
4 ... 20 mA + 1 x PNP switching output standard
0 ... 20 mA + 1 x PNP switching output
0 ... 10 V + 1 x PNP switching output
100 ms
1 x PNP switching output
2 x PNP switching output
Break contact / make contact
Window / hysteresis
PNP 2V
B
On 250 mA / Off 1mA
> 10 million
20 ms
Ye s
2s; T
f=0.5 Hz
Display: Err3 switching output K1, Err4 switching output K2
ON
40 ms
50
Switching output
11 Instrument description
- Switching point
- Reset point
- Hysteresis
-Damping
- Delay
Burden
4to20mA 0to20mA 0to10V
Configurable in the nominal measuring range (> reset point)
Configurable in the nominal measuring range (< switching point)
Configurable in the nominal measuring range
99.99 sec
99.99 sec
R
(UB-6.5 V) / 0.022 A
i
R
(UB-6.5 V) / 0.022 A
i
R 10 kΩ

Mechanical properties

Process connection
Material Surface Process seal
- Process connection
521, 523, 571, 575, 576, 652
Stainless steel 316L
Ra 0.8 µm
All flange connections are welded and therefore have no seals
FPM standard
- Process connection
997 (JUMO PEKA)
Measurement diaphragm
Material Surface
Enclosure
Material Surface Threaded sleeve M12x1 Enclosure seal Display
Screw plug
Material Surface Enclosure seal
Weight 0.2 kg with process connection 504 (G1/2)
FPM, VMQ (silicon), EPDM also available, see Data Sheet 40.9711
Stainless steel 316L
Ra 0.8 µm
Stainless steel 316L
Ra 0.8 µm
Stainless steel 316L
VMQ (silicon)
PA
Aluminum 3.2315
Eloxal coating
VMQ (silicon)
51
11 Instrument description

Ambient conditions

Permissible temperatures
Measuring material
-for basic type
extension 004
Environment
-25 ... +100°C (+135°C max. 1 hour per day)
-25 ... +200°C
-25 ... +75°C
-Ambient
temperature -50°C
Storage
Permissible relative humidity
-in operation
-in storage
Permissible mechanical loading
-Vibration
strength
- Shock resistance
Electromechanical compatibility
- Interference emission
- Interference immunity
Protection IP 67 to DIN 60529
Restricted function: Stationary use only, danger of broken cable,
display does not function
-40 ... +85°C
100% incl. condensation on instrument outer sleeve
90% without condensation
20 g, 10 ... 2000 Hz per IEC 60068-2-6
Performance characteristic A per EN 61326
(Only with 4-pin connecting cable and grounded enclosure)
Class A per EN 61326
50 g for 11 ms / 100 g for 1 ms per IEC 60068-2-27
52

Accuracy

Relative pressure
Nominal (rated) measuring range
Linearity
Accuracy at +20°C
Accuracy at
-20...+75°C
1
2
3
11 Instrument description
0.4 1 4 10 25 60
0.15 0.15 0.1 0.1 0.1 0.1 % of end val. of
0.35 0.3 0.25 0.25 0.25 0.25 % of end val. of
0.7 0.6 0.5 0.5 0.5 0.5 % of end val. of
bar
nom. mess. range
nom. mess. range
nom. mess. range
Nominal (rated)
-0.4...0.4
-1...1 -1...3 -1...9 -1...24 bar
measuring range
Linearity
Accuracy at +20°C
Accuracy at
-20...+75°C
1
0.15 0.15 0.1 0.1 0.1 % of end val. of
0.35 0.3 0.25 0.25 0.25 % of end val. of
2
0.7 0.6 0.5 0.5 0.5 % of end val. of
3
Absolute pressure
Nominal (rated)
0.41 4 102560bar
measuring range
Linearity
Accuracy at +20°C
Accuracy at
-20...+75°C
1
0.15 0.15 0.1 0.1 0.1 0.1 % of end val. of
0.35 0.3 0.25 0.25 0.25 0.25 % of end val. of
2
0.7 0.6 0.5 0.5 0.5 0.5 % of end val. of
3
Long-term stability 0.2% with reference conditions to EN 61298-1
nom. mess. range
nom. mess. range
nom. mess. range
nom. mess. range
nom. mess. range
nom. mess. range
1
Linearity based on limit point setting
2
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of measuring range.
3
Includes: linearity, hysteresis, repeatability, deviation from initial and final values of measuring range, thermal effect on initial value of measuring range and span.
53
JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc.
Street address: Moritz-Juchheim-Straße 1 36039 Fulda, Germany Delivery address: Mackenrodtstraße 14 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO House Temple Bank, Riverway Harlow - Essex CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 63 52 62 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-JUMO
1-800-554-JUMO Fax: 315-697-5867 E-mail: info@jumo.us Internet: www.jumo.us
Loading...