JUMO 902435 User Manual

JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-5866
1-800-554-JUMO Fax: 315-697-5867 E-mail: info@jumo.us Internet: www.jumo.us
Push-in RTD temperature probe for heat meter with connection cable for immersion sleeves (type PS/PL)
For temperatures from 0 to 180 °CApproval for heat meter MID and domestic approval
as exchangeable temperature probes
Meets the requirements as per DIN EN 1434, AGFW FW202 and FW211Approval for cold meter and cold / heat meter combinationsPaired and calibrated as per the German calibration regulationsProduction certified as per module D of the MID (CE and metrology sign)Adapted accessoried for installation (data and price sheet 902440)
Heat meter RTD temperature probes are used for temperature measurement in closed-loop pipe systems. Thanks to the installation in thermowells with tolerance on fit for optimum thermal coupling, the system need no longer be discharged at the time of the regular exchange once the certification period has elapsed.
The temperature probes are approved in accordance with the EU directive 2004/22/EC (MID) as well as according to the German calibration regulations (domestic approval). The tempera­ture probes are available paired and calibrated. The state-approved test body for heat KF2 is responsible for this purpose. Moreover, the production is certified according to module D of MID, and the temperature probes are also available with the CE and metrology sign.
Data sheet 902435 Page 1/8
Note:
In Germany, temperature probes, type PS, with MID approval may no longer be installed in new installations smaller than DN 25 pipe systems (4th. amendment of the Calibration Regulations).
Technical Data
Process connection Push-in RTD temperature probe for thermowells (immersion sleeves) Connection Wire end ferrules Connection cable PVC, PUR, TPE, Silicone; unshielded or shielded
Thermowell Type PS: stainless steel Ø 5.0 mm, Ø 5.2 mm or Ø 6.0 mm
Measuring insert Platinum temperature probe as per DIN EN 60751
Temperature Type PS: 0 to 150 °C
Temperature difference Type PS: 3 to 150 K
Minimum immersion depth Type PS: > 15 mm
Fitting length Type PS: 45 to 85 mm
Response behavior Type PS: t
Ambient conditions Climatic 0 to 70 °C
Accessories For the thermowell, refer to the data and price sheet 902440
2010-09-08/00530105
For the max. connection cable lengths, refer to the table on the following page
Type PL: stainless steel Ø 6.0 mm Thermowell with tolerance to fit for thermowells
Rated value: Pt100, Pt500, Pt1000; class B Connection: 2-wire/4-wire circuit
Type PL: 0 to 180 °C
Type PL: 3 to 180 K
Type PL: 30 mm
Type PL: 85 to 450 mm
= 15 s (installed in thermowell)
Type PL: t
Protection class IP65 Electromagnetic E1 Mechanical M3
0.5
= 12 s (installed in thermowell)
0.5
Data sheet 902435 Page 2/8
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-5866
1-800-554-JUMO Fax: 315-697-5867 E-mail: info@jumo.us Internet: www.jumo.us
Cable cross-section Max. connection cable length AL
for Pt100
0.22 mm
0.34 mm
0.50 mm
2
2
2
2500 mm 12500 mm 25000 mm 3500 mm 17500 mm 35000 mm 5000 mm 25000 mm 50000 mm
Max. connection cable length AL for Pt500
Max. connection cable length AL for Pt1000
Note:
Also take into account a possible connection cable length limitation due to the design approval of the arithmetic unit/counter. For 4-wire circuit, the max. connection cable lengths are limited due to the requirements of the arithmetic unit/counter.
Approval overview
Version Basic type Approval
Type Number
Figure: PS 90.276-F40 (PS)
902435/05 (PS)
902435/10 (PS) Domestic (Germany) 22.30/99.10 (105 °C) 902435/11 (PS) 902435/50 (PS) Domestic (Germany) 22.30/02.03 (150 °C) 902435/51 (PS) 902435/10 (PS)
902435/11 (PS) 902435/50 (PS)
902435/51 (PS) 902438/50 (PS) Module B as per MID
Figure: PL 90.276-F80 (PL) Domestic (Germany) 22.30/91.05 (180 °C)
902435/30 (PL) Domestic (Germany)
902438/30 (PL) Module B as per MID 902465/30 (PL) Domestic (Germany; Cold) 22.77/09.01 (150 °C)
Domestic (Germany) 22.30/89.07 (Pt100) (150 °C)
22.30/89.08 (Pt500) (150 °C)
22.30/89.09 (Pt1000) (150 °C)
Domestic (Germany)
22.70/99.10 (105 °C)
as per DIN EN 1434 Domestic (Germany)
22.70/02.03 (150 °C)
as per DIN EN 1434
1
A0445-2112-2007 (150 °C)
22.70/00.01 (180 °C)
as per DIN EN 1434
1
DE-06-MI004-PTB011 (180 °C)
1
MID = Measuring Instruments Directive, European measuring instrument directive 2004/22/EC
2010-09-08/00530105
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-5866
1-800-554-JUMO Fax: 315-697-5867 E-mail: info@jumo.us Internet: www.jumo.us
Dimensions
D
EL AL
D
15 ALEL
Data sheet 902435 Page 3/8
Basic type 90.276-F40, 902435/05 (PS) Basic type 902435/10 (PS) Basic type 902435/50 (PS) Basic type 902438/50 (PS)
Basic type 90.276-F80 (PL) Basic type 902435/30 (PL) Basic type 902438/30 (PL) Basic type 902465/30 (PL)
2010-09-08/00530105