JUMO 902424 User Manual

Data sheet 902424 Page 1/6
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road East Syracuse, NY 13057, USA Phone: 315-437-5866
1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com
Screw-in RTD Temperature Probe for Heat and Cold Meters with Terminal Head for Direct Mounting (Type DL)
• For temperatures between 0 and 180 °C
• Approval for heat meters as replaceable temperature probes; MID and domestic approvals
• Fulfills requirements of DIN EN 1434, AGFW 202, and FW211
• Paired and calibrated according to the German calibration regulations
• Production certified according to module D of MID (CE and metrology identification marking)
• Approval for cold meter
Heat and cold meter RTD temperature probes are used for temperature measurement in closed pipeline systems. Direct mounting without an additional thermowell is recommended to achieve optimum thermal coupling to the measurement medium in the heating system. The temperature probes are approved according to the EU directive 2004/22/EC (MID) as well as according to the German calibration regulations (domestic approval). The temperature probes are available in paired and calibrated versions. The officially recognized testing agency for heat KF2 is available for this. Moreover, the production is certified according to module D of MID, and the temperature probes are also available with the CE and metrology identification marking.
Technical data
Terminal head Form J or B, aluminum diecast,
M16 × 1.5 or M24 × 1.5, IP65,
ambient temperature -20 to +100 °C
Process connection Thread G 1/2, stainless steel
Protection tube Stainless steel, Ø 6 mm
Measuring insert Platinum temperature sensor according to DIN
Temperature 0 to 180 °C
Temperature difference 3 to 180 K
Minimum immersion depth 25 mm
Insertion length 85 to 280 mm
Admissible rated pressure PN 25 bar
Maximum flow velocity 2 m/s (water) Thermoelectric voltage 5µV
Response behavior t
Environmental conditions Climate -25 to +70 °C
2014-03-18/00386325
or Ø 8 mm stepped down to Ø 6 mm
EN 60751
Nominal value: Pt100, Pt500, Pt1000; class B
Connection: two-wire or four-wire circuit
5s
0.5
Protection type IP65
Electromagnetic E1
Mechanical M3
Environment class C
Heat meter
Cold meter
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road East Syracuse, NY 13057, USA Phone: 315-437-5866
1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com
Dimensions
~33
D
G
~52
EL
10
~33
G
~52
35
D
EL
10
L = 60 (100)
75
10
Ø29
50
D
G
EL 22
27
L = 60 (100)
75
10
Ø29
50
G
EL 22
D
35
27
Data sheet 902424 Page 2/6
Basic type 90.238-F40 (DL) Basic type 902424/10 (DL) Basic type 902427/10 (DL)
Basic type 90.238-F40 (DL) Basic type 902424/11 (DL) Basic type 902427/11 (DL)
Basic type 902454/10 (DL) Basic type 902454/11 (DL)
2014-03-18/00386325
Loading...
+ 4 hidden pages