Julabo SL-8K Data Sheet

Product data sheet 1/3
SL-8K Calibration Bath
The calibration baths are perfectly suitable for applications in calibration laboratories and conform to the requirements specified by DIN ISO
9001. The instruments provide highest temperature stability and temperature uniformity. An optional precision reference sensor with integrated measuring electronics and RS232 connection is available.
Your advantages
VFD COMFORT DISPLAY LCD DIALOG DISPLAY backlit for convenient interactive operation Keypad for setpoints, warning/safety values and menu functions ICC (Intelligent Cascade Control), self-optimizing temperature control TCF Temperature Control Features to optimize the control behavior ATC3 3-Point-Calibration Pt100 External sensor connection for measurement and control SMART PUMP, electronically adjustable pump stages Early warning system for low liquid level (DBGM 203 06 059.8) Adjustable high temperature cut-out, visible via display RS232/RS485 interface for online communication Integrated programmer for 6 x 60 program steps Highest temperature stability to ±0.005 °C (±0.02 °C at +200 °C) Temperature uniformity better than ±0.01 °C Optional Pt100 sensor for reference temperature measurements Temperature chamber with overflow at maximum liquid level
SL-8K
Pump capacity (Water)
Technical Data
Order No. 9352508 Model series HighTech Category Calibration Baths Working temperature range (°C) +50 ... +300 Temperature stability (°C) ±0.005 Setting / display resolution 0.01 °C Integrated programmer 6x60 steps Temperature Display VFD, LCD Heating capacity (kW) 3 Pump capacity flow rate (l/min) 22-26 Pump capacity flow pressure (bar) 0.4-0.7 Pump capacity flow suction (bar) 0.2-0.4 Bath opening / bath depth (W x L / D cm) Ø = 12 / 17 cm Filling volume liters 8
Product data sheet 2/3
SL-8K
External Pt100 sensor connection integrated Digital interface RS232
Optional: Profibus Ambient temperature 5...40 °C Dimensions W x L x H (cm) 22 x 46 x 47 Weight (kg) 16 Classification according to DIN12876-1 Classification III (FL) Included with each unit Including bath cover with 7 openings . Temperature
stability at +200°C: ±0.02°C Order information Optional: Electronic module with analog connections for
external programming, temperature recorder, standby,
alarm (Order no. 8 900 100). Bath cover integrated Available voltage versions 208 - 230 V / 50 - 60 Hz
Characteristics
Display
A perfect view
Ample, easy to read VFD Comfort display for simultaneous display of 3 values, warning functions, high temperature cut-off, pump stages (resolution 0.01 °C)
Additional plain text information
Comfortable LCD dialog display for interactive operation with plain text display
Pump stage and liquid level
Backlit indicator for selected pump stages and filling volume on Presto® PLUS, Magnum 91 & Forte HT
Operation
Comfortable and detailed
Comfortable keypad with additional menu functions for pump stages, calibration, control parameters, programmer, warnings, etc.
Temperature Control
For perfect results
Intelligent Cascade Control’, automatic & self optimizing adjustment of PID control parameters, temperature stability ±0.005 °C internal, <±0.05 °C external
Full control
Temperature Control Features’ for individual optimization, access to all important control parameters, additional settings for band limit, limits, Co-Speedfactor etc.
Highest measuring accuracy
Absolute Temperature Calibration’ for manual compensation of a temperature difference, 3-point calibration
Refrigeration Technology
Consistent cooling capacity
Easily removable venting grid for quick and easy cleaning
100 % Cooling capacity
Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Condensation and ice protection
A heated cover plate prevents condensation or ice build-up in the bath
Technical Features
Clever pump system
Reliable and consistent pump capacity, electronically adjustable pump stages
Control from the external application
External Pt100 sensor connection for precise measurement and control directly in the external application
Connection compliant to standard
RS232/RS485 dual-interface for serial data transmission according to EIA-485 industry standard (2-wire bus technology), upgradable with Profibus DP
Optimal program control
For the execution of time and temperature dependant profiles, 6 temperature profiles with 60 steps max., with real time clock
Automatic control of operating time
Electronic countdown-timer function for timer-programmed unit shut-down, standby mode after programmed time expires
Warning & Safety Functions
Early warning system for low liquid level
Maximum safety for applications, optical and audible alarm, allows user to refill bath fluid before the unit shuts down