Johnson Controls T-4002, T-4054, T-4003, T-4100, T-4110 Adjusting Instructions

...
Pneumatic Control Manual 717.1
Archived Document
Room Devices Section
T-4000 Series Room Thermostat
Technical Bulletin T-4000
Issue Date 0984
The T-4000 Series Pneumatic Room Thermostats are factory calibrated to the values specified in Table 1. If these settings are suitable to the application, no adjustments need to be made. If different settings are required, turn the set point dial to the desired value.
If the room temperature is out of control, determine the cause of the problem. Possible causes not involving the thermostat are listed below.
1. Check for proper operation of the controlled device.
*
This bulletin provides adequate instructions for making basic adjustments to T-4000 series room thermostats. For more detailed information, contact the local branch office.
2. Check to see if the controlled medium (water, for example) is hot or cold enough to maintain the desired temperature.
3. Check for unusual or extreme loads in the space which may overcome the efforts of the thermostat to maintain control.
If it is determined that the thermostat is out of adjustment, follow the appropriate instructions to recalibrate the instrument.
Sensitivity Adjustment (See Fig. 1)
All T-4000 series thermostats have an integral output pressure test port. The newer high volume output thermostats have a plug-in test port fitting which accepts a hypodermic needle test probe and gage (X-200-140 and G­2010-11). On older models, the test port cap screw must be removed to accept a screw-in test gage (X-200-19). For these units, a test port fitting (T-4000-
615) can be ordered. All low volume output thermostats have hypodermic needle test points. For units that require a screw-in gage, remember to replace the test port cap screw after adjustment.
The sensitivity of the T-4000 series instruments is factory set at 2.5 PSI/F° (32 kPa/C°). It is adjustable from 1 to 3 PSI/F (13 to 38 kPa/C°) for low volume output thermostats and from . 75 to 4 PSI/F° (10 to 51 kPa/C°) for high volume output thermostats. (Sensitivity of the T­4003 is fixed at 2.5 PSI/F°).
1. Furnish the appropriate supply pressure to the instrument.
2. Insert the appropriate gage.
3. Note the output pressure.
Note:
The output pressure for T-4054 thermostats must be within one of the two control ranges (not within the deadband range).
4. For high volume output thermostats, move the sensitivity slider toward the control port to increase or away from the control port to decrease the sensitivity. For low volume output thermostats, move the sensitivity slider down to
°
© 1984 Johnson Controls, Inc.
Code No. LIT-7171089X
1
increase or up to decrease
Archived Document
the sensitivity.
5. Check the output pressure. If it is different from the value found in step 3, turn the sensitivity slider adjusting screw until the output pressure equals that value. For low volume output (plastic body) instruments, use the output adjusting screw to adjust the output pressure.
6. Check the sensitivity. Change the control setting on the dial by at least 5F° and note the new output pressure. The change in output pressure divided by the change in dial setting is the actual sensitivity.
7. Repeat steps 4 through 6 until the desired sensitivity is obtained.
T-4003 Range of Remote Readjustment
The range of remote readjustment is adjustable from .15 to 2.5 F°/PSI (.012 to .2C°/kPa) and is factory set at
0.5 F°/PSI (.04 C°/kPa). To
determine if the range of remote readjustment needs to be adjusted, proceed as follows:
1. Insert the test gage into the test port fitting and note the output pressure.
2. Set the master pressure to 4 psig (28 kPa).
3. Turn the set point dial to 80F and turn the calibration screw to obtain a 3 psig (21 kPa) output pressure.
4. Set the master pressure to 14 psig (98 kPa).
5. Turn the set point dial until the output pressure returns to 3 psig.
6. The change in dial setting divided by the 10 psig (70 kPa) change in master pressure (14 psig - 4 psig = 10 psig) is the actual range of readjustment. If this is not the desired range, move the range of readjustment slider (sensitivity slider in Fig. 1a) toward the control port to decrease or away from the control port to increase the range of readjustment. Readjust the slider screw to the pressure noted in step 1 above. Repeat steps 2 through 6 until the desired range is obtained.
Set Point Adjustment
Before calibrating the instrument, check to see if the output pressure is already at an appropriate value according to the ambient temperature. For
example, if the actual room temperature is below the set point and the output signal is calling for more heat, the thermostat is doing its job and does not require adjustment.
T-4002 and T-4100 (See Fig. 1)
1. Insert the test gage.
2. Furnish supply air to the instrument (restricted supply for T-4100).
3. Note the ambient temperature in the space.
4. Turn the set point dial to that temperature.
5. Turn the calibration screw until the output pressure is at the mid spring range of the controlled device.
6. Turn the set point dial to the desired set point.
T-4003 Set Point Adjustment
Insert the gage and set the master pressure to one half of the supply pressure. Follow steps 3 through 6 under T-4002 Set Point Adjustment.
2 T-4000 Technical Bulletin
Fig. 2: T-4054 Adjustments
Archived Document
Fig. 3: Back View of T-4054
T-4054 Deadband and Set Point Adjustment (See Fig. 2)
The T-4054 can easily be adjusted to suit the application by changing the settings on the heating and cooling dials. If the thermostat has been tampered with or is out of adjustment, proceed with the following instructions to recalibrate the instrument.
Note: On deadband applications which have overlapping spring ranges, it is suggested that a C-9200 or other sequencing device be used to bias the cooling valve spring range to compensate for spring range shift.
Deadband Setting
If it is necessary to check or reset the deadband setting, proceed as follows:
Turn the lower set point dial to 85F on direct acting models or 55F on reverse acting models. Turn the upper set point dial to 55F on direct acting models or 85F on reverse acting models. Insert a hypodermic needle test probe into the test connection. The reading on the test gage is the actual deadband pressure.
If it is necessary to adjust this pressure to meet application requirements, proceed as follows:
1. Remove the T-4054 from the mounting bracket and pull it away from the wall so the deadband adjusting screw on the back of the instrument is exposed (see Fig. 3).
2. Turn the screw clockwise to increase or counter­clockwise to decrease the deadband pressure until the gage reads the desired pressure. Approximately one half turn will produce a 1 PSI (7 kPa) change in deadband pressure.
3. Return the set point dials to the original settings.
Set Point Calibration (See Fig. 2)
1. Set the lower set point dial to 85F for direct or 55F for reverse acting models.
2. Note the ambient temperature at the element.
3. Set the upper dial to this temperature (direct acting model: heating, reverse acting model: cooling).
4. Turn the upper calibration screw until the control pressure is in the middle of the spring range of the controlled device (heating).
5. Set the lower dial to the temperature noted in step 2 (direct acting model: cooling, reverse model: heating).
6. Turn the lower calibration screw until the control pressure is in the middle of the spring range of the controlled device (cooling).
7. Turn the upper dial to the desired set point.
T-4054 Operational Graph
T-4000 Technical Bulletin 3
Loading...
+ 5 hidden pages